1. SIEMENS POLYDOROS SX 65 80 A AX Troubleshooting Guide POLYDOROS SX SiemensAG 2001 The reproduction transmission or use of this document or its contents is not permitted without express written authority Offenders will be liable for damages All rights including rights created by patent grant or registration of a utility model or design are reserved Print No RX63 055 840 01 01 02 English Replaces RX63 055 072 01 02 02 Doc Gen Date 05 01 Table of Contents General 5 Product specific ReMarkS o oooooooccccooo eee ee 5 Required Documents o 5 Required Tools and Measurement Equipment occccocccccoccooo o 5 Safety Information sisaceksehe nt a See Gu Yh Been es 5 Safety PIECAUIONS rd Al atan 5 kV mAs and tube current tolerances ooooooornonronor nao 6 Abbreviations and symbols used 0 00 eee eee eee 7 Troubleshooting 8 Error000 0 p30 205 Seeded as Ret A a eed bss d 8 EMORDO Na ana cea Boia ath sn dais ddr ENS aran ae ad ta ae 9 EXODUS 10 ENCTORDOS A A lle Se AEE A R bE E aA 11 ErrOFOD ia eh oe eae eos Leh Bees eee as Sa E ANE a 12 Error DO i334 cetera poe A A eG ten ere bed aoe 13 EnrOmODit 221000031000 a rd SSO a A a nt ae od 14 EMOS ra a Owe Re pa dt 15 EMORODS ia at E eg a ES oe Pes 16 Emor oSA 22 ote EA EA E nto eee ine natty ate a tee es 17 Error O55 ins cho tiie Re ea Rs Rs Ce en ee PS a TM ee E AN 18 Error 05Oras Ses bee eed
2. AG Medical Solutions RX63 055 840 01 01 02 Page 103 of 104 05 01 TD SD 21 POLYDOROS SX 65 80 104 Changes to Previous Version Restructuring of document and new layouted POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 104 of 104 Siemens AG 05 01 SD 21 Medical Solutions
3. Medical Solutions Troubleshooting 65 Error 575 Meaning lontomat Checksum error density correction Possible causes and action e D190 Siemens AG RX63 055 840 01 01 02 Page 65 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 66 Troubleshooting Error 580 Meaning lontomat bad dose value Possible causes and action e Check system configuration e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 66 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 67 Error 581 Meaning lontomat bad lead time value Possible causes and action e Check system configuration e D190 Siemens AG RX63 055 840 01 01 02 Page 67 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 68 Troubleshooting Error 582 Meaning lontomat bad voltage correction Possible causes and action e Check system configuration e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 68 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 69 Error 590 Meaning lontomat mAs value not sent Possible causes and action e Check system configuration e D190 e D100 Siemens AG RX63 055 840 01 01 02 Page 69 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 70 Troubleshooting Error 591 Meaning lontomat bad tube AP selection Possible causes and action e Check system configuration POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 70 of 104 Siemens AG 05 01 TD SD 2
4. Setthe main system switch to OFF to switch off power to all components generator and connected equipment e To avoid unintentional release of high voltage or radiation switch OFF S1 SS on the D100 board Siemens AG RX63 055 840 01 01 02 Page 5 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 6 General e Install or remove assemblies only with the generator switched off and in accordance with ESD guidelines Z WARNING Tests and adjustments that must be made with radiation switched on are indicated by the radiation warning symbol VAN c gt Take appropriate radiation protection measures when carrying out such work kV mAs and tube current tolerances NOTICE All kV mAs and tube current values are stated in these instruc tions with 5 Tolerance POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 6 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions General 7 Abbreviations and symbols used ADC Analog digital converter AP Workstation selection FL Fluoroscopy FO Large focal spot filament current F1 Small focal spot filament current HS High voltage cable HT High voltage generator IGBT Transistor Insulated Gate Bipolar Transistors LS Charging contactor LSR Charging contactor relay MPS Serial interface multiprocessor interface M_TK Door contact NAK No acknowledgment N TU1 Workstation switchover OUTA Actuation of bridge
5. communikation between filament D220 and master D100 is made via serial link MPS D100 X1 28 D220 X1 28 If an interruption of the communication between D100 and D200 occurs Error 119 is out put by D100 Possible causes and action Serial interface interrupted gt check D100 D220 defetive replace Arcing tube H1 High tension transformer Check screening and shielding of cabling Check protective ground wire all protective ground wire connections intended by the manufacturer must be made Check firmware level Check voltage power supplies 5 V 0 2 V on D100 D160 and D220 Check schematics of generator for LED s Ignore error 119 if the red LED D220 is blinking and no firmware is loaded gt perform download POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 24 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 25 Error 120 Meaning No communication vie D220 to D100 via serial line transmit Description The communikation between filament D220 and master D100 is made via serial link MPS D100 X1 28 D220 X1 28 In communication errors between D100 and D220 Error 120 is output by D100 Possible causes and action e Serial interface interrupted gt check e D100 D220 defective replace e Arcing tube e Check screening and shielding of cabling e Check protective ground wire all protective ground wire connections intended by the manufacturer must be made e Check firmware
6. error once again If error does not occur then 20 mA wiring POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 88 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 89 Error 711 Meaning Bridge short circuit in the main inverter current in the inverter too high Description If more than 3 bridge short circuits occur during an exposure or in fluoroscopy within 3 6 s of if the current in the inverter is too high then Error 711 is output Possible causes and action 1 Checking D165 The intermediate circuit voltage is too low but higher than 400V 400V 630V UZ moni toring does not respond Error 604 with lt 400V Error 605 with gt 630V N Generator OFF Connect instrument to D110 X5 X6 Generator ON Uz must be approx 550 V Checking the charging current Generator OFF Main switch OFF Check freedom from voltage at the mains fuses Disconnect leads L1 L2 L3 at the PS contactor Fita current transformer with 10 ohms terminating resistor over each lead Connect the leads again Connect oscilloscope to current transformers over L1 L2 L3 Trigger Int Main switch and generator ON The mains currents must be the same in all phases pay attention to symmetry In the case of an error Look for the error in the phase in which the smallest amplitude is measured bad con nections burnt contacts Checking the tube assembly Generator OFF Connect
7. oscilloscope to the following points CH1 MAjsz Ir ist D100 X64 MA_ACT CH2 KViszt D100 X61 kV_ACT Trigger D100 X64 SWR SS switch on D100 OFF Generator ON Check trigger SS switch on D100 ON Release exposure with 81 kV 20 mAs if no error increase kV stepwise and observe current Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 89 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 90 Troubleshooting If step peaks are observed and the generator climbs out with ERROR 711 AP ID 80 or 119 AP ID 80 The tube assembly is defective Replace tube assembly Checking D110 inverter e Generator OFF e Connect oscilloscope to following points e CH1 D100 X62 _LOAD 1V 2 50A e Trigger X64 SWR time base 10 us e Generator ON 9 e In SSW lt DIAGNOSTIK gt conduct lt Inverter Test gt and compare measurement results with the following wiring diagrams POLYDOROS SX 65 80 e The current load is to be measured with single pulse and fluoro attachment It must show the following values Pos Signal 500 mV 4 50A Neg Signal 500 mV 450A Ou Is Sier Ou Y Fig 1 e The following requirements are to be checked in both oscillograms Forward oscillation 120A 10 Backward oscillation 50A 10 e On deviations in the amplitude e Generator OFF e Connect oscilloscope to following points CH1 D100 X60 OUT_A CH2 D100 X60 OUT_B Trigger D100 X64 SWR e Switch on D100 OF
8. see wiring diagram X2206 16 The fluoro attachment feedback DLR_U is output via D111 X50 7 to D100 X50 7 Possible causes and action e D100 e D111 e Contactor K6 e Ribbon cable X50 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 20 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 21 Error 058 Meaning Polydoros Host SW is mismatching the type of generator configured in the XCS configura tions SW E g D100 PROM is LX lite and configured type is LX 80 Possibe causes and action e Configure the correct generator in the XCS service software Siemens AG RX63 055 840 01 01 02 Page 21 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 22 Troubleshooting Error 077 Meaning Timeout exposure release Description HS trigger of image system ASU is missing during pulsed fluoroscopy or indirect tech nique Possible causes and action e Check of HS trigger POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 22 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 23 Error 100 Meaning Invalid task state D100 Description The software on the D100 is in an impermissible task state Possible causes and action e PROM D100 J15 for D100 Part No 37 75 256 e D100 Siemens AG RX63 055 840 01 01 02 Page 23 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 24 Troubleshooting Error 119 Meaning Time out filament circuit Description The
9. 00 Troubleshooting Error 804 Meaning Invalid tube data Description Error when selecting 2 phase or 3 phase tube assembly or error in nominal speed 5 Hz 6 Hz POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 100 of 104 Siemens AG 05 01 SD 21 Medical Solutions Troubleshooting 101 Error 805 Meaning No current in the main phase no rotation no ROT Description The current in the main phase is measured during the run up test point D100 X60 _ANL 1V 2 5A The current must be greater than 1 5 A The error is displayed no ROT green LED V36 on D100 not on Possible causes and action Check activation ANO AN5 of the IGBT modules from the D100 to the D115 see wi ring diagram X2206 31 POLYDOROS OFF OnD100 switch S2 ZK OFF POLYDOROS ON Check activation signals ANO AN5 on the D100 X65 with oscilloscope The indivi dual voltage pulse of the activation signals ANO AN5 should be approximately 13 V If the activation signals are smaller than 10 V the D100 is to be replaced D115 defective Interruption in the stator cable Disconnect the cable at the tube assembly and measu re ohmically there must be no connection between the leads Il O Test the currents in the stator leads with a current transformer Measure intermediate circuit voltage at D115 X6 and X5 and on D110 X6 and X5 UZacr 1 4 X Umains Caution Direct voltage Check intermediate circuit voltage connection
10. 00 X1 28 D220 X1 28 In communication errors between D100 and D220 Error 122 is output by D100 Possible causes and action e Serial interface interrupted gt check e D100 D220 defective replace e Arcing tube e Check screening and shielding of cabling e Check protective ground wire all protective ground wire connections intended by the manufacturer must be made e Check firmware level e Check voltage power supplies 5 V 0 2 V on D100 D160 and D220 e Check schematics of generator for LED s e Ignore error 122 if the red LED D220 is blinking and no firmware is loaded gt perform download Siemens AG RX63 055 840 01 01 02 Page 27 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 28 Troubleshooting Error 123 Meaning Invalid answer from filament Description The communikation between filament D220 and master D100 is made via serial link MPS D100 X1 28 D220 X1 28 In communication errors between D100 and D220 Error 123 is output by D100 Possible causes and action e Serial interface interrupted gt check e D100 D220 defective replace e Arcing tube e Check screening and shielding of cabling e Check protective ground wire all protective ground wire connections intended by the manufacturer must be made e Check firmware level e Check voltage power supplies 5 V 0 2 V on D100 D160 and D220 e Check schematics of generator for LED s e Ignore error 123 if the red L
11. 01 TD SD 21 Medical Solutions Troubleshooting 59 Error 562 Meaning lontomat invalid measuring field Description No scan field selected Possible causes and action e Check system configuration e D190 Siemens AG RX63 055 840 01 01 02 Page 59 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 60 Troubleshooting Error 563 Meaning lontomat Invalid fluoro detector no PDA MPL or B signal Possible causes and action e Check system configuration POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 60 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 61 Error 570 Meaning lontomat Checksum error detector Possible causes and action e D190 Siemens AG RX63 055 840 01 01 02 Page 61 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 62 Troubleshooting Error 571 Meaning lontomat Checksum error HUD Possible causes and action e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 62 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 63 Error 572 Meaning lontomat Checksum error lead time value Possible causes and action e D190 Siemens AG RX63 055 840 01 01 02 Page 63 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 64 Troubleshooting Error 573 Meaning lontomat Checksum error voltage correction Possible causes and action e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 64 of 104 Siemens AG 05 01 TD SD 21
12. 1 Medical Solutions Troubleshooting 71 Error 600 Meaning 15V 10 faulty Possible causes and action e 5V 0 2V present e Shortcircuit on board D100 D220 D110 e LED V72 on D160 is not on replace SNT power pack e Check voltage on D160 X4 4 and X4 1 230V 10 if present replace SNT power pack Siemens AG RX63 055 840 01 01 02 Page 71 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 72 Troubleshooting Error 601 Meaning 15V faulty Possible causes and action e Short circuit on board D100 D220 e LED V73 on board D160 is not on replace SNT power pack e Check voltage on D160 X4 4 and X4 1 230V 10 if present replace SNT power pack POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 72 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 73 Error 602 Meaning Power contactor PS not OK Description e Oninitialization D100 switches the power contactor on and wait 0 5s for the acknowledgment If this noes not occur Error 602 is output e In operation Ifthe power contactor drops out during operation Error 602 is also output Possible causes and action e Drive Power relay defective D160 K5 i e no low signal OV PSR at D160 X5 7 D100 defective e Ifthe above named signal OK and the power contactor does not switch relay K5 on D160 defective replace D160 e Check cable from D100 X5 to D160 X5 e Actuation of the PS contactor A2 A1 24V 10 a
13. 1 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 42 Troubleshooting Error 433 Meaning Current in the filament inverter FO large focal spot too high Possible causes and action e Short circuit in the charging circuit e Wiring X41 filament transformer high voltage cable filament or D220 defective e Ejectleads X41 3 and X41 11 if error 433 occurs anew after reset D220 must be replaced if error 402 occurs D220 is OK e Check filament transformer Primary winding R lt 10hm transformation ratio 32 9 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 42 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 43 Error 434 Meaning Current in the filament inverter F1 small focal spot too high Possible causes and action e Short circuit in the charging circuit e Wiring X41 filament transformer high voltage cable filament or D220 defective e Ejectleads X41 9 and X41 12 if error 434 occurs anew after reset D220 must be replaced if error 402 occurs D220 is OK e Check filament transformer Primary winding R lt 10hm transformation ratio 32 9 Siemens AG RX63 055 840 01 01 02 Page 43 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 44 Troubleshooting Error 485 Meaning Filament download error when deleting Flash Prom Possible causes and action e Repeat filament download e D220 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 44 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Tro
14. 5 01 TD SD 21 96 Troubleshooting Error 716 Meaning Short circuit in main inverter one incident Description During an exposure or a fluoro there are in between 3 6 sec ONE inverter short circuit This will result in under exposed images or in a DR series in a blank image plus an under exposed one If the short circuit continues error 711 will be caused POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 96 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 97 Error 800 Meaning Temperature switch of tube has been switched Description Tube is too hot the temperature switch of tube has been switched Possible causes and action e Tellthe user to wait for next exposure until tube has cooled down a bit e With POLYDOROS SX65 80 Temperature switch of the tube is only connected for OPTI154 30 50R 100 and ME GALIX tube assemblies for all other tube assemblies there is a jumper on D160 X61 1 and 2 for AP1 or D160 X62 1 and 2 for AP2 or D160 X63 1 and 2 Siemens AG RX63 055 840 01 01 02 Page 97 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 98 Troubleshooting Error 803 Meaning Bridge short circuit in the rotating anode starter inverter Description A bridge short circuit has occurred in a branch of the inverter The R_KURZ signal is indicated by the LED s V35 on D100 and V114 on D115 Possible causes and action e Check stator resistances at the tube assembly connection partic
15. D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 52 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 53 Error 515 Meaning lontomat impermissible plug position Description After selecting a iontomatized mode of operation it is checked whether or not the plug position stored in EEPROM is permissible Possible causes and action e Check system configuration e D190 Siemens AG RX63 055 840 01 01 02 Page 53 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 54 Troubleshooting Error 550 Meaning Dose monitoring has responded not enough dose detected after 100 ms Description In an iontomated exposure the dose counter on D100 is read after 100 ms exposure time If one half of the necessary does is not reached by the max exposure time error 550 is signalled Possible causes and action e Faulty operation customer Tube not directed onto selected IONTOMAT chamber Collimator closed Wrong choice of the exposure kV e Errors Wrong lIONTOMAT workstation programmed check system configuration Nosignal from IONTOMAT chamber at measuring point D100 X63 DL_IN Check detector and cabling D100 defective replace POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 54 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 55 Error 551 Meaning lontomat Dose supervision min for tomo Description This error is similar to error 550 in normal mode However t
16. D220 6 HCheck high voltage cable Check the plug of the HV cable for tracking If tracking is present the HV cable must be replaced If there is no tracking on the HV cable clean the HV plug replace corona disks and gas ket and insert in the tube assembly with silicone oil No corona disks may be used on the transformer side Approx 10 mm oil must be present in the transformer receptacle The ohmic value of the shielding braid may be 1 Ohm per meter If there is suspicion the the HV cable is no longer voltage proof the HV cable must be replaced Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 91 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 92 Troubleshooting Error 712 Meaning Plus kV max error Description The positive side of the high voltage is monitored for an absolute value of 80 kV Display LED D100 V168 KV MAX Possible causes and action e Checking the KV actual value D220 X18 A7 and D100 X61 kV see wiring diagram X2206 15 at both test points 1V 10 kV e FLattachmentis out of adjustment Perform adjustment e Check ribbon cable between D220 X1 and D100 X1 e kV controller defective replace D100 e kV actual value sensing on D220 defective e Tube assembly or high voltage cable defective e High voltage generator defective e Ifthe tube assembly is replaced perform tube adjustment POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 92 of 104 Siemens AG 05 01 TD SD 21 Medical Solution
17. ED D220 is blinking and no firmware is loaded gt perform download POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 28 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 29 Error 125 Meaning Timeout lontomat Possible causes and action e Check cabling from D100 to lontomat or replace cable Siemens AG RX63 055 840 01 01 02 Page 29 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 30 Troubleshooting Error 401 Meaning Time out ADC of filament circuit Description A D converter in the filament circuit does not convert Possible causes and action e Measure Vcc 5 V 0 2 Volt on D220 X18 A1 e fvoltage is within tolerance replace board D220 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 30 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 31 Error 402 Meaning Minimum filament current Description The minimum filament current 1 2 nominal value is monitored in standby Possible causes and action e Intermediate circuit voltage missing in the filament circuit at X41 7 and X41 10 mearu re 230V 10 e Check fuse F21 D160 e Interruption in the filament path Check wiring X41 H1 Workstation selected AP relay in H1 actuated 1 H1 test point 50 OV 2 Ap 1 test point 51 24V 415 3 AP2 test point 52 24V 415 e Check filament transformer primary winding R lt 1Ohm transformation ratio 32 9 e Check filament and high voltage cable e Ifno er
18. ENOL 4I it ii A E A E A aula eae tees 42 EXNOR DA lt 4 tat ee eis Poh ie Pa oe Sy ae eS 43 EXOR SO eo es cdo a A eee eee e e ee ee ee eae 44 Eror 48O iaa ida a od as 45 ENOS Tc is A Dd a Ad wae eee 46 Error488 Ii aaa tada da a AA A a i 47 Error505 oie cn SS Ok ee RS ee ee 48 Error 51 Obits See pute wa be ata kins ey bed ant h dog Pals Sade ad Read he Pee ees 49 ErmOro ly oh ami ested Dota oe ok peeves ae YO kee eed gas 50 ITO D1 2s 2 dadas a oe e a nd oD Sk Fo cl Dn SIA a 51 Eor ST 3 aai ada a bale it and PEA sia alo 52 ENOED A ai is A a A eared a ana 53 ErONSS OF a 8 Ste ef Sh ets RO ote Pe hee Fae ees Ot Re Boe 54 Errore Lite cote ei Eek a es Pate ano Sie od Geena nav eten bats SIM als eek a 55 Error 552er A eee hae Pale bw Ee bs was deed 56 Error 560 2 cicae io Gas fot fon oe ae Bia ae eed Foie ke ores TERh Ss 57 A eet ae A ee ee ee ee ee ene eae 58 Error 562 ua be ead Md hh ee A Dye ast dees 59 Error 563 a a eats As tied a dea st es ceased el Se 60 EON D108 aa Po a at ert Sk Sh cae Ps Ne epee AO REE 61 EMO 57 de ioc ce AE ee ee te Oh Se Fae Ve ns eet 62 Error 572er ae Seer cutee we th Tata gas A dag ks Bae aad ee a Pte at 63 Error 5 98 sat eea ae daw ee Pee eet Sa tenet ewe ga 64 EXOrRO O 0 dd ot os ad a a ee a o OD AA a 65 Eror 580 2 24 55 fort di dae onl haere bb Raed Seb hee ed PE ene ee Da Saba oad de 66 EMOS di Hee io Ae a Pere eee dal AA a ad PRES 67 EMOn O24 sc A o o ge Ome Pent ase 68 Error 590 ss cata a ei e
19. F e Generator ON e Release FL or exposure POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 90 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 91 Compare pulses with diagram X2206 16 replace D100 in the case of deviation If pulses are OK proceed as follows Replace inverter The FL attachment must be adjusted anew after replacement of the inverter 4 Checking D111 FL attachment 5 Check contacts of the K6 contactor Check screw connections Checking D220 filament board If ERROR messages 711 and 119 AD IP 80 occur sporadically and if no error is found when checking D100 D110 D111 D115 D165 and the tube assembly D220 can be defective Perform check as follows Ribbon cable between D100 and D220 must be run in front of the retaining bracket on H1 Generator OFF Connect oscilloscope to following points CH1 MAacr Ir act D100 X64 MA_ACT CH2 kV acr D100 X61 kV ACT Trigger D100 X64 SWR SS switch on D100 OFF Generator ON Check trigger SS switch on D100 ON Perform measurements with both focal spots Release exposure with 81 kV 32 mAs 100 ms and observe the tube current If the cur rent wave shape is linear increase tube current with the mAs button and retain the 100 ms If a too high current is measured in the measurement the tube must be adju sted again If abrupt changes of push factor occur during the setting and if the current can not be adjusted replace
20. Medical Solutions Troubleshooting 83 Error 629 Meaning No AP relay Acknowledgment Actuation of the oil switch wrong Description The N TU1 OK signal monitors the workstation relay switchover The workstation relay switchover is actuated by the TU1 N TU2 signal TU1 N TU2 1 tube1 selected N TU1 OK 0 Possible causes and action e Check signal path according to wiring diagram X2206 19 D220 X19 A5 reversed to X19 A6 D220 OK e g API selected X19 A6 24V 15 X19 A5 0V e D100 defective e D220 defective Siemens AG RX63 055 840 01 01 02 Page 83 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 84 Troubleshooting Error 633 Meaning Uist gt 2kV not OK Description Radiation is present There is a blockage Consequence LS PS oil switch or tube contactor are blocked Possible causes and action Actual value sensing D220 H1 defective Check ribbon cable D220 X1 D100 X1 D100 defective High voltage does not decay no load tube assembly defective oil switch side NOTICE Caution on pulling out the high voltage cables first discharge cables POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 84 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 85 Error 650 Meaning Invalid tube Description XCU sends impermissible tube assembly workstation to D100 Siemens AG RX63 055 840 01 01 02 Page 85 of 104 POLYDOROS SX 65 80 Me
21. POLYDOROS SERVICE SW gt ADJUSTMENTS gt GENERATOR PARAME TER If this is done with generator using older D100 SW this error might be caused especially in Tomography Set current back to 10 mA Siemens AG RX63 055 840 01 01 02 Page 79 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 80 Troubleshooting Error 611 Meaning Specified value of tube voltage outside tolerance limits 40 kV lt U lt 150 kV for radiography 40 kV lt U lt 110 kV for fluoroscopy Possible causes and action e SWerror POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 80 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 81 Error 612 Meaning Timeout for filament to get ready Description Error can occur if filament error is acknowledged and radiation is released again immedia tely Possible causes and action e D220 Siemens AG RX63 055 840 01 01 02 Page 81 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 82 Troubleshooting Error 628 Meaning DOOR error door contact open Description The DOOR M_TK signal monitors the door contact The error is output only if the door is opened with radiation switched on DOOR 1 door closed OK Possible causes and action e The message line shows Door open during standby e Check the signal path according to drawing X2206 1 1 e Check the oil pressure switch POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 82 of 104 Siemens AG 05 01 TD SD 21
22. ROS SX 65 80 RX63 055 840 01 01 02 Page 16 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 17 Error 054 Meaning Invalid fluoro data Description Probably the configuration has been modified without turn off on the generator so that an unknown fluoro curve has been selected known to XCU but not yet known to the genera tor This might also be due to the fact that MANUAL kV fluoroscopy is selected in systems supporting pulsed fluoroscopy Siemens AG RX63 055 840 01 01 02 Page 17 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 18 Troubleshooting Error 055 Meaning Invalid parameters for characteristic kV curve Description XCU sends characteristic kV curve with more than 10 points POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 18 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 19 Error 056 Meaning Invalid exposure mode Description XCU sends request for impermissible mode of operation Siemens AG RX63 055 840 01 01 02 Page 19 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 20 Troubleshooting Error 057 Meaning Fluoro attachment feedback signal DLR_U wrong Description With fluoroscopy the activation of the fluoroscopy relay DLR_ is displayed on the D100 X50 5 This signal comes to D111 X50 5 and activates the relay K1 Through the activated relay contact of K1 1 2 the contactor K6 is activated and the fluoro attachment is added on
23. ad ew bk Bae e 95 EMO do a a te ae e e o lO e ets ys 96 Error 800 ae oid Pee a ee is E 97 Error 808 da aes dee Slane ets ees a A ewe SS a Se wes a 98 Error 80At a eis Soeur SA A aie ee ee Ses 100 EO S05 te cto See a a o as a a eb Rak A a A 101 Erron80G pass a edie bd 102 EMOS TD gone ma ts di o a aE eae Ed eat aran 103 3 Changes to Previous Version 104 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 4 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions General 5 Product specific Remarks Required Documents e Generator Wiring Diagram e Startup instructions Required Tools and Measurement Equipment e Standard installation tool kit e Service PC e PC connection cable 5m 99 00 440 Safety Information NOTICE When carrying out the work steps and tests the product specific safety information contained in the documents as well as the general safety information must be observed Safety Precautions e Use the power OFF switch on board D 160 to switch off the generator before working on it Z WARNING Line voltage is still present in transformer T1 and switch on circuit D 160 when the generator is switched off After the generator is switched off approx 600 V DC for the inverter is still present gt This is indicated by LEDs V35 and V36 on D110 and LED V89 on D220 goes on The line voltage decays in approx 1 5 minutes to 0 V the LED s go out when the voltage is approx 30 V e
24. ae OS ead See Sa a ate Pe ek ett 69 Error 591 ii e de Re aed aay ere a a dag Pia eb tae Doe ee be dee tee 70 Error600 2 Gicae io E dan Pele ate A a Fah 71 EXOROO Le bea ete add e a o e dal at A al ek OD Ne 72 10 8 5 02 QRO phen oa Saws HOS a dated Obed band be a eee bata Pa aed 73 Efror 603 2 6 203 00 ke rs a RR AR EE lag eet e 74 EXI A gale hacen e RO De 75 Error GOS ii E A A A ia 76 Emor GOO ti ie wae ead id wae we A A ee wa betes 77 Siemens AG RX63 055 840 01 01 02 Page 3 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 4 Table of Contents Error 608 302 5 a A is o dd 78 Error G Diva a A e A EA A a 79 EMOS diles an a E A ata POE BER Ra e 80 EGLO O12 eet See Ne Bi N r dee Pe Renee Mente ns ate uate OAS 81 ErrOr628 vies td hatin bag alg ks Pe aa by es eae tas a A ae es be te 82 Error 629 cart ek FOR AS Pan cA Sea es ee Soe 83 EO O a fe ee a ba Ne Baad a a eb den oe te eee 84 Error 650 he ce sich oases oak de hs See ek Phin es ned Se wee PRS Rhee eed dee 85 Error G00 sec 2 reast eee Pig O eee hae ae be wae Behe 86 EMmOR 6882 votos rea aa rs acetates welsh ale a wen a ae Gat ea 87 Error 690 nse yoink ke eee es Se ee Pe ee ee EE Ea 88 O ee eae wiles es eae Bo eke Bee hy Pinky Sa hd ee we Pe ee 89 Error FA DA O ee he tists Gees PON Eek te eee ro 92 EMO Voc tee ten bOI E A e 0 ea te ee a A IO E 93 Error tase Baloo td diated Be aie toh te aie ead deh uh bid 94 Erot ZIS a A Be ded acta Oa Maes keh Se bared
25. branch A OUTB Actuation of bridge branch B PS Power contactor PSU Powers contact acknowledgment PSR Power contactor relay RANGE Current range ON FL R short Jumper short circuit ROT Rotating anode rotation SNT Switching power supply STRSU Tube assembly switchover in starter UZ Intermediate circuit voltage WR Inverter ZB Exposure preparation Refer to the generator wiring diagram for further signal names Siemens AG RX63 055 840 01 01 02 Page 7 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 8 Troubleshooting Error 000 Meaning Invalid Init Data Description At least one init block is wrong This error will be repeated at each following init block even if the init data of this block are correct You may try to configure and select all fluoro curves new or configure the complete system from the scratch POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 8 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 9 Error 001 Meaning Start of Service Session Description e The Service PC has contacted the XCU e No error for information only e This error will be shown as error 41 of AP ID 240 XCU in the errorlog Siemens AG RX63 055 840 01 01 02 Page 9 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 10 Troubleshooting Error 002 Meaning End of Service Session Description e The Service PC has terminated the connection to the XCU
26. dical Solutions 05 01 TD SD 21 86 Troubleshooting Error 666 Meaning Switch S3 X44 on D100 is in Service position Possible causes and action Switch over switch S3 to Normal POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 86 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 87 Error 688 Meaning No dose signal from the VIDEOMED DI Possible causes and action Check connection between VIDEOMED DI and generator VIDEOMED DI defective Check image intensifier output Remove camera head from image intensifier Release FL in the Service mode The image must appear in the image intensifier output window with FL ON If not then the l l circuit 1 1 voltage supply is defective lfyes VIDEOMED DI is defective D100 defective Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 87 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 88 Troubleshooting Error 690 Meaning No high voltage trigger signal from ASU of FL TOP or FL Compact to generator Possible causes and action e Check connection between ASU FL Compact and SK 111 generator e ASU defective FL Compact defective e Wrong fluoro mode pulsed instead of continuous e D100 defective e FL Compact takes longer than 1 second to send a trigger pulse after a radiation re quest e 20 mA safety current loop is not closed or closed too late when pulsed fluoroscopy starts For a test actuate service switch S3 and try to provoke
27. ding relays are monitored via the RANGE O and RANGE 1 signals Possible causes and action Check function at SS OFF on D220 Test points D220 X19 A1 A2 wiring diagram X2206 18 and 22 With ZB ON measuring range 200 mA RANGE 0 X19 A1 0V gt RANGE 0 OK X19 A4 24V 15 RANGE 1 without significance With normal FL ON measuring range 1 mA RANGE 0 X19 A1 24V 15 RANGE 0 OK X19 A4 OV RANGE 1 X19 A2 24V 15 DANN RANGE 1 OK X19 A3 OV Ifthe RANGE 0 and RANGE 1 signals do not agree with the operating mode there is a fault in the cable X1 to D100 or on D100 If the signals RANGE 0 _OK and RANGE 1 _OK are wrong with correct actuation D220 must be replaced Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 77 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 78 Troubleshooting Error 608 Meaning AP contactor not present oil switch acknowledgment missing Possible causes and action e Check plug on D160 X44 3 4 24V 10 at 50 Hz 29V 10 at 60 Hz POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 78 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 79 Error 610 Meaning Invalid range of tube current 10 mA lt I lt 800 mA Description This error can be caused by an invalid setting of the tube current in Polydoros Service SW For LX generators and SX generators it is possible to set the minimum tube current to 1 mA
28. he dose is not checked after 100 ms but at the end of the tomo exposure It can be caused as a follow up error if the object to be exposed requires more mAs than the generator is allowed to give E G 800 mAs is max mAs At 1 2 s tomo time max mA would be 666 mA However the genera tor may use 800 mA in order to have a reserve for the different angles and projections during the tomo turn If the object is that thick that the generator is using 800 mA all the 1 2 sec the 800 mAs will be reached after 1 0 s and the generator will stop X raying After 1 2 s it will state that the dose was too low Siemens AG RX63 055 840 01 01 02 Page 55 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 56 Troubleshooting Error 552 Meaning lontomat Dose supervision max for tomo Description Dose counter terminates exposure before the tomographic time ends Possible causes and action e Select higher exposure parameters POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 56 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 57 Error 560 Meaning lontomat kV value impermissible Possible causes and action e D190 Siemens AG RX63 055 840 01 01 02 Page 57 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 58 Troubleshooting Error 561 Meaning lontomat invalid max time gt 10s Possible causes and action e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 58 of 104 Siemens AG 05
29. level e Check voltage power supplies 5 V 0 2 V on D100 D160 and D220 e Check schematics of generator for LED s e Ignore error 120 if the red LED D220 is blinking and no firmware is loaded gt perform download Siemens AG RX63 055 840 01 01 02 Page 25 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 26 Troubleshooting Error 121 Meaning No communication vie D220 to D100 via serial line receive Description The communikation between filament D220 and master D100 is made via serial link MPS D100 X1 28 D220 X1 28 In communication errors between D100 and D220 Error 121 is output by D100 Possible causes and action e Serial interface interrupted gt check e D100 D220 defective replace e Arcing tube e Check screening and shielding of cabling e Check protective ground wire all protective ground wire connections intended by the manufacturer must be made e Check firmware level e Check voltage power supplies 5 V 0 2 V on D100 D160 and D220 e Check schematics of generator for LED s e Ignore error 121 if the red LED D220 is blinking and no firmware is loaded gt perform download POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 26 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 27 Error 122 Meaning Buffer overflow in receive queue to filament Description The communikation between filament D220 and master D100 is made via serial link MPS D1
30. lled 2cables installed The terminal connector has Part No 86 06 352 Cable not connected from XCU to generator Cable has a short circuit Hybrid circuit of ARCNET is defective on any XCS component SW of a component sends ARCNET Reconfiguration interrupts intermittently Watchdog of Polydoros is active D100 or power supply defective Possible causes and action Check cabling for connection and shorts Check Hybrids and terminators D100 D200 defect replace Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 13 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 14 Troubleshooting Error 051 Meaning D100 J15 PROM checksum error with D100 Part No 37 75 256 Description D100 checks the PROM s check sum during initialization Possible causes and action e PROM D100 J15 defekt with D100 Part No 37 75 256 e D100 defect POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 14 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 15 Error 052 Meaning Heating Error during download Description D220 sends telegram not acknowledge Possible causes and action e Repeat filament download e Replace D220 Siemens AG RX63 055 840 01 01 02 Page 15 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 16 Troubleshooting Error 053 Meaning Image intensifier invalid zoom level Description XCU sends zoom level gt 3 to D100 Possible causes and action e Check configuration POLYDO
31. logged off e No error for information only e This error will be shown as error 42 of AP ID 240 XCU in the errorlog POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 10 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 11 Error 003 Meaning Start of Remote Service Session Description e The Service PC has contacted the XCU via Modem e No error for information only e This error will be shown as error 46 of AP ID 240 XCU in the errorlog Siemens AG RX63 055 840 01 01 02 Page 11 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 12 Troubleshooting Error 004 Meaning End of Remote Service Session Description e The Service PC has terminated the remote connection to the XCU logged off e No error for information only e This error will be shown as error 47 of AP ID 240 XCU in the errorlog POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 12 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 13 Error 007 Meaning Polydoros is no longer in the XCS network Error 007 is most of the times accompanied by a 1 on the D100 in case of HW defects In case of SW Resets 007 will heal itself Description The communication between Polydoros and XCU is down Possible reason Checking the terminal connectors At the terminal D320 X3 S and US or D320 X4 S and US there must be either 2terminal connectors installed 1terminal connector and 1 cable insta
32. ltage at D220 V65 for 15V 10 e Check connection to D100 lead X1 10 X1 38 40 e Check voltage on D160 Checkpoints X30 15V 10 X29 15V 10 X28 0V Siemens AG RX63 055 840 01 01 02 Page 39 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 40 Troubleshooting Error 427 Meaning 15V digital 10 is less than 12V on D220 Possible causes and action e Check D220 V64 for 15V 10 e Check connection to D100 lead X1 22 X1 38 40 e Check voltage on D160 Checkpoints X30 15V 10 X29 15V 10 X28 0V POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 40 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 41 Error 428 Meaning 24V 15 is less than 20V on D220 Possible causes and action Check voltage on D220 X19 A8 24V 15 Check connection to D100 lead X1 21 X1 38 40 Check voltage on D160 D160 X13 14 24V 15 D160 X28 0V It might as well be the case that the power fail signal X1 on D320 is not detected and therefore each time you switch on the error is entered into the error log Input check Remove cable D160 X1 coming from D320 The Hex display on D320 should indicate F4 If not D320 is defective Output check Connect a voltmeter DC to D160 X121 2 power fail and X121 1 when switching off the generator voltage should suddenly change from 5 V to 0 V Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 4
33. m X2206 16 UZacr 1 4 X Umains Caution Direct voltage Actual value sensing D220 defective kV controller defective Tube arcing High voltage generator H1 defective High voltage cable OK Filament current too high perform tube adjust If the tube assembly is replaced perform tube adjust POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 94 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 95 Error 715 Meaning Minus kV min error Description The minus kV min monitoring becomes active if the kV actual value is more than 10 kV less than the kV nominal value Display LED D100 V165 kV MIN Possible causes and action Check minus KV actual value at D220 X18 A8 and D100 X61 Mef3punkt KV see wi ring diagram X2206 15 at both test points 1V 4 10kV Check ribbon cable between D220 X1 and D100 X1 Perform Diagnostic Inverter Test Check oscillating current on D100 X62 I_LAST 1V 2 50 A and compare with dia gram ERROR 711 Check connection screws in the intermediate circuit and inverter Measure intermediate circuit voltage at D110 X2 and X1 X2206 16 UZacr 1 4 X Umains Caution Direct voltage Actual value sensing D220 defective kV controller defective Tube arcing High voltage generator H1 defective High voltage cable OK Filament current too high perform tube adjust Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 95 of 104 POLYDOROS SX 65 80 0
34. ng tube current value in fluoroscopy Possible causes and action Error in the master gt perform system configuration and POLYDOROS adjustment POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 36 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 37 Error 422 Meaning Maximum filament current in tube current control Description The tube current control permits a filament current of 1 25 times the max filament current for 200 ms If the controller finds no adjustment with max filament current after this time a warning is sent to the notice memory Possible causes and action e Check the filament current with cathotest e The tube is not adjusted correctly e The tube emits verly badly e Perform tube adjust e Check filament transformer primary winding R lt 1Ohm transformation ratio 32 9 Siemens AG RX63 055 840 01 01 02 Page 37 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 38 Troubleshooting Error 425 Meaning 15V 10 is less than 12V on D220 Possible causes and action e Check voltage at D220 V66 15V 10 e Check connection to D100 lead X1 12 X1 38 40 e Check voltage on D160 Checkpoints X30 15V 10 X29 15V 10 X28 0V POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 38 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 39 Error 426 Meaning 15V analog 10 is less than 12V on D220 Possible causes and action e Check vo
35. oa E Be ea ea te dead wie a ea Sa wad ae ws 19 Error O57 523 saci a SON iia VS Pee A Sa ee tee ee eee See 20 Error O58 ics 4 ccs saddened Seb oe A he bdo ae TA RAE See 21 Error OZ ei aade ant tan Dd 0G 6 oid beets 8 hace e Say eR es bbe baa Pe es 22 Error TOOS ornai cat sr as ceases arroba ERG ae a aati 23 EMOTIO ena SMe e aa A a nee Is Scone ga i Meret a AS eter os Oe 24 Erroi 1208 ni le A A Re A ole Saeed A ae ee ae ERNA 25 Error eed ey we hoe eae eels Pewee Beha Bee ib eats Sau a Pe wees es 26 Error 122020 et OF a tak ON 2 a FON o e E te ae nee 1 co E ed 27 Error 128 ca ate 28 Error Dc 200 o0 Rio Gd haved beta enter OY Rae dwt eet RO ose Pa Bae od was 29 ErrorRdO Ma daras o A a a eine 30 EMOL4O2 trena a a ote Bot cee Se a an de a a ey oats a ae ate ee aay ed ko tee it as 31 ENCORE A bee ged De Oe ee eee Ws ANA 32 Error 404i coh au eani dee dba E chs ts Sy Pa eed de e s 33 Emors IO 2d a co RO soe ee nr Dd e ro a RC a A a re eee ee 34 ErrordlB xc cani ads 35 Error 42d 2 Ss Cok tan Paw ie See ote eRe Pies deaths bbw Bae wad 36 Er Or42 2n 2603 a ar a a rn wa Gees a a et ea ahs 37 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 2 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Table of Contents 3 Eror 425 ease aa adds a a ia 38 EmOr4a2Gi ui is we oe A A ek ew A A a a a 39 Error 7 set te tie na Bae eee Rees AOL ee amines ee es POSH Ae Seas 40 OMA 2 Be se Nar oe cecre apts a a aie arg ah Sener Starke arene he aes watt nee ae ale 41
36. on e Are the IONTOMAT workstations correctly programmed Siemens AG RX63 055 840 01 01 02 Page 49 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 50 Troubleshooting Error 511 Meaning lontomat Impermissible film screen system Description After selecting an iontomatized mode of operation the lontomat checks whether or not the film screen system H U D transferred to it is permissible If no several or a wrong film screen system were transferred Error 511 is displayed Possible causes and action e Check system configuration e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 50 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 51 Error 512 Meaning lontomat invalid tomo time selection Description After selecting the Plani lontomat mode of operation the lontomat checks whether or not the selected tomographic time is permissible If the transferred tomographic time was too long or too short Error 512 is displayed Possible causes and action e Check system configuration e D190 Siemens AG RX63 055 840 01 01 02 Page 51 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 52 Troubleshooting Error 513 Meaning lontomat invalid detector Description After selecting a iontomatized mode of operation or fluoroscopy the lontomat checks whe ther or not the detector stored in EEPROM is permissible Possible causes and action e Check system configuration e
37. ror is found replace board D220 Siemens AG RX63 055 840 01 01 02 Page 31 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 32 Troubleshooting Error 403 Meaning Maximum tube current in radiography Description The tube current is exceeded by 50 100 ms after high voltage ON Possible causes and action Perform tube adjust POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 32 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 33 Error 404 Meaning Maximum tube current in fluoroscopy Description The tube current is exceeded by 50 100 ms after high voltage ON Possible causes and action Perform tube adjust Siemens AG RX63 055 840 01 01 02 Page 33 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 34 Troubleshooting Error 410 Meaning Timeout processor Possible causes and action e Check power supply on D220 V66 15V 10 V65 15V 10 V64 15V 10 X19 A8 24V 15 e Replace D220 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 34 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 35 Error 413 Meaning Tube current nominal value to large Possible causes and action e Errorin the master gt perform system configuration and POLYDOROS adjustment e Replace D220 Siemens AG RX63 055 840 01 01 02 Page 35 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 36 Troubleshooting Error 421 Meaning Wro
38. rror 604 is output e In operation Ifthe intermediate circuit voltage drops during operation Error 604 is also output Possible causes and action e Phase is missing intermediate circuit is charged too slowly e Check line voltages e Check UZacz signal at D100 X20 17 and X20 19 1V 4 50V UZacT 1 4 X U mains e Check ribbon cable from D100 X20 to D115 X20 e Charging resistors R1 R2 defective Replace resistors Siemens AG RX63 055 840 01 01 02 Page 75 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 76 Troubleshooting Error 605 Meaning Intermediate circuit voltage UZ gt 630V UZact 1 4 x Umains Caution Direct voltage Description On initialization The charging contactor is switched in by D160 The status of the interme diate circuit voltage is queried after 5 s provided there has been a charging contactor acknowledgment At a too high intermediate circuit voltage gt 630 V Error 605 is out put Possible causes and action e Check line voltages e Measuring the line resistance e Check UZacz Signal at D100 X20 17 and X20 19 UZact 1 4 X U mains e Check ribbon cable from D100 X20 to D115 X20 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 76 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 77 Error 606 Meaning RANGE Error Description The signals RANGEO OK and RANGE1 OK monitor the switchover of the tube current measuring ranges on D220 The correspon
39. s Troubleshooting 93 Error 713 Meaning Minus kV max error Description The negative side of the high voltage is monitored for an absolute value of 80 kV Display LED D100 V167 KV MAX Possible causes and action Checking the KV actual value D220 X18 A8 and D100 X61 kV see wiring diagram X2206 15 at both test points 1V 10 kV Check ribbon cable between D220 X1 and D100 X1 kV controller defective kV actual value sensing on D220 defective Tube assembly or high voltage cable defective High voltage cable defective D100 D220 defective replace If the tube assembly is replaced perform tube adjustment Siemens AG Medical Solutions RX63 055 840 01 01 02 Page 93 of 104 POLYDOROS SX 65 80 05 01 TD SD 21 94 Troubleshooting Error 714 Meaning Plus kV min error Description The plus kV min monitoring becomes active if the kV actual value is more than 10 kV less than the kV nominal value Display LED D100 V166 kV MIN Possible causes and action Check KV actual value at D220 X18 A7 or D100 X61 KV see wiring diagram X2206 15 at both test points 1V 4 10 kV Check ribbon cable between D220 X1 and D100 X1 Perform Diagnostic Inverter Test Check oscillating current on D100 X62 I_LAST 1V 50A and compare with dia gram ERROR 711 Check connection screws in the intermediate circuit and inverter Measure intermediate circuit voltage at D110 X2 and X1 see wiring diagra
40. t 50 Hz 29V 10 at 60 Hz e Check contactor acknowledgment PSU at D100 X5 25 24V 10 e Contactor sticks or jams Siemens AG RX63 055 840 01 01 02 Page 73 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 74 Troubleshooting Error 603 Meaning Charging contactor LS not OK Description On initialization D100 switches the charging contactor closed There must be an acknow ledgment after 5 s otherwise Error 603 is output Possible causes and action Actuation of charging relay D160 K4 defective i e no low signal OV LSR at D160 X5 5 D100 defective If the above named signal is OK and K4 does not switch relay K4 on D160defective replace D160 Check cable from D100 X5 to D160 X5 Check actuation of the LS contactor A2 A1 24V 10 at 50 Hz 29V 10 at 60 Hz Check contactor acknowledgment LSU at D100 X5 27 or D160 X3 9 24V 15 Contactor sticks or jams POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 74 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 75 Error 604 Meaning Intermediate circuit voltage UZ lt 400V UZact 1 4 x Umains Caution Direct voltage Description e Oniinitialization The charging contactor is switched in by D160 The status of the intermediate circuit voltage is queried after 5s provided there has been a charging contactor acknow ledgment At a too low intermediate circuit voltage lt 400V E
41. terminals D165 D110 D115 Check stator resistances at the tube assembly connection at 20 C Inthe 3 phase stator 0 2 0 2 6 ohms 0 Il 2 0 2 6 ohms Inthe 2 phase stator 0 13 16 ohms 0 Il 18 20 ohms Inthe Opti 154 0 1 ca 10 ohms 0 Il ca 10 ohms Siemens AG RX63 055 840 01 01 02 Page 101 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 102 Troubleshooting Error 806 Meaning Intermediate circuit is switched off when switching on the generator Description During power up inspect whether or not the intermediate circuit is switched on If this is not the case and a liquid bearing X ray tube is connected to the generator then this tube will not be accelerated The system is no longer ready for operation Possible causes and action e Switch off system switch on intermediate circuit and switch on system again POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 102 of 104 Siemens AG 05 01 SD 21 Medical Solutions Troubleshooting 103 Error 810 Meaning Stator contactor acknowledgment SRTRSU missing Description Stator contactor K3 wiring diagram X2206 30 not changed over Possible causes and action Contactor K3 defective Acknowledgment interrupted Check signal on D100 X20 23 STRSU Signal OK K3 ON 0V K3 OFF 24V 15 Replace D100 Check whether a third workstation is configured and order contactor attachment if ne cessary Siemens
42. ubleshooting 45 Error 486 Meaning Filament download error in check sum of the Flash PROM Possible causes and action e Repeat filament download e D220 Siemens AG RX63 055 840 01 01 02 Page 45 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 46 Troubleshooting Error 487 Meaning Filament download Error by bank switching Possible causes and action e Repeat filament download e D220 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 46 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 47 Error 488 Meaning Filament download Error during programming Flash Proming Possible causes and action e Repeat filament download e D220 Siemens AG RX63 055 840 01 01 02 Page 47 of 104 POLYDOROS SX 65 80 Medical Solutions 05 01 TD SD 21 48 Troubleshooting Error 505 Meaning Invalid iontomat configuration Description Storage of configuration data into D190 RAM is not possible Possible causes and action e D190 POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 48 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 49 Error 510 Meaning Inadmissible IONTOMAT workstation Description After selection of an iontomated operating mode or fluoroscopy a check is made whether a permissible IONTOMAT workstation has been transferred by the master D100 In the event of an error no or a wrong workstation is output Possible causes and action e Check system configurati
43. ulars at 20 Inthe 3 phase stator 0 2 0 2 6 ohms 0 Il 2 0 2 6 ohms Inthe 2 phase stator 0 13 16 ohms 0 Il 18 20 ohms Inthe Opti 154 0 ca 10 ohms 0 Il ca 10 ohms e Stator defective e Correct stator configured e Check activation ANO AN5 of the IGBT modules from the D100 to the D115 see wi ring diagram X2206 31 POLYDOROS OFF OnD100 switch S2 ZK OFF POLYDOROS ON Check activation signals ANO AN5 on the D100 X65 with oscilloscope The indivi dual voltage pulse of the activation signals ANO AN5 should be approximately 13 V If the activation signals are smaller than 10 V the D100 is to be replaced e D115 defective e Check phase shifter Connection 2 phase tube assembly connection see wiring dia gram X2206 30 e Test stator cable disconnect cable at tube assembly and measure with ohmmeter the re must be no connection between the leads Il O e Test the currents in the stator leads with a current transformer POLYDOROS SX 65 80 RX63 055 840 01 01 02 Page 98 of 104 Siemens AG 05 01 TD SD 21 Medical Solutions Troubleshooting 99 Measure intermediate circuit voltage at D115 X6 and X5 D110 X6 and X5 UZact 1 4 X Umains Caution Direct voltage Check intermediate circuit voltage connection terminals D165 D110 D115 Siemens AG Medical Solutions RX63 055 840 01 01 02 05 01 Page 99 of 104 TD SD 21 POLYDOROS SX 65 80 1