1. Troubleshooting EMBEDDING CENTER No response when powered on Power line not connected Switch isn t turned on Fuse isn t connected well or may be burned out Connect power line turn on again Press the switch indicator light on Check if fuse is good or not change with another Light is not on Possible burnt fuse Check fuse or have technician check No heating Possible burnt fuse Heating element defective Sensor not detecting Contact technical support Check fuse or have technician check No flow of paraffin liquid Wax block is not melted enough Magnetic valve is blocked or switch is stuck with paraffin Not enough time for melting continue heating Check switch Use hair dryer to melt wax around switch if necessary Contact technical support No flow of paraffin liquid or leak Faulty magnetic valve Contact technical support Sound is good as magnetic valve starts up but no flow Paraffin low Other things in magnetic valve Clean the paraffin screen Not enough time for melting continue heating Contact technical support Light is twinkling when turned on Unstable power Check the power Starts to heat up when turned on but turns off under work mode SSR is broken Contact technical sup
2. User Manual TN1500 Embedding Center Read Prior to Operation Always keep this manual near the unit TANNER scientific P46818 TANNER scientific User Manual TN1500 Embedding Center 888 708 5233 Info TannerScientific com www TannerScientific com Contents 1 General Summary uo Page 2 2 Installation 3 3 Technical Parameters 4 4 Safety Precautions 5 5 Equipment Overview 6 6 Operation eee 10 6 1 Main Control Panel Setup Page 11 6 2 Temperature Setting 11 6 3 Timer Setting 12 6 4 Clock Setting 13 6 5 Running Manual Mode 13 6 6 Cooling Plate Operation 14 7 Cleaning amp Maintenance 15 8 Warranty 8 Service 15 9 Troubleshooting 16 Tanner Scientific TN 1500 Embedding Center 1 1 General Summary This instrument was a collaborative effort developed using input from over 100 Histologists in the United States Tanner Scientific TN 1500 Tissue Embedding Center and Cooling Plate are intended to embed and cool pathological tissue Features e Pre set On Off timer for every day of the week e Histo trimmers on both ends of the hot plate are convenient for both left and right hand users e Adjustable magnifier for easy viewing e Two modes of operation automatic and manual e Temperature control functions located on 5 parts e Multiple over heat protection devices e Keys with light indicator e Low voltage and adj
3. on the power and press AUTO MAN key 2 Press SET key choose 2 TIMER 3 Press P to choose Tuesday and press or to set 7 30 ON 16 30 OFF NOTE If on and off times are both 00 00 that means the instrument will not turn on that day NOTE If the instrument is turned on according to a pre set starting time it will automatically start to work at that time in automatic mode 12 Tanner Scientific TN1500 Embedding Center 6 4 Clock Setting 1 Select 3 CLOCK using the gt keys 2 Press SET key to enter into current time setting mode 3 4 Use or to increase or decrease time figures 5 Press the AUTO MAN key to return back to the home screen Use to choose minute hour or week SET CLOCK SELECT 1 3 1 MINUTE 05 2 HOUR 09 3 WEEK FRI 6 5 Running Manual Mode 1 To manually run the instrument press AUTO MAN key on the main control panel to enter into the manual interface 2 Press PB to enter into running mode Heating icons will appear meaning all areas are warming up 3 Press lt Q to return back to waiting mode AUTO MODE Running FRI 18 40 PARAFFIN 65 C RIGHT 65 c HOTPLATE 55 c fi LEFT ssec liN DISPENSER 679c i CRYO Ni MAN MODE Off Work FRI 18 40 PARAFFIN 65 C f RIGHT esec liN HOT PLATE 55 E LEFT sec liN DISPENSER 679c i CRYO mi Tanner Scientifice TN1500 Embedding Center 13 6 6 Cooling Pl
4. to 6 liters of paraffin e Temperature can be adjusted in a range from 0 to 859C e The reservoir has overheat protection in the eventofa temperature control failure 2 Main Control Panel e To light up screen press any key except CRYO e The screen will enter into protection and power save mode after 15 minutes of no contact 3 LED Light Control e Press key the LED light will become brighter e Press lt q key the LED light will become darker 4 Thermal Chambers e Used to keep the blocks warm e Has programmable temperature control e Removable transfer trays 5 Paraffin Dispenser e Has programmable temperature control e Flow control is adjustable 6 Small Cryo Plate e Located directly in front of the embedding area to ensure consistent low temperatures during the filling ofthe molds e Settemperature allows cryo plate to cool to maximum ability 7 Paraffin Waste Trays e Two removable trays are located under the heated work areas to collect the excess paraffin drained from the surface e Empty and clean both trays daily 8 Hot Plate e Features a paraffin trimmer on both the left and right side e Has programmable temperature control 9 Forceps Holder e Maintains independenttemperature Tanner Scientifice TN1500 Embedding Center 7 10 Rear Panels Image 5 3 Rear Panel of Embedding Center A B C D O TANNER O o lt a gt
5. O O O O O O A Main Power Switch amp Socket C Grounding Wire Connector B Foot Switch amp Socket D Socket for Cooling Plate Image 5 4 Rear Panel of Cooling Plate A Power Switch B Fuse Pocket C Power Socket D Grounding Wire Connector E Back Cover Panel components may change depending on date of manufacture 8 Tanner Scientifice TN1500 Embedding Center Cooling Plate amp Cooling Plate Control Panel Cooling Plate e Power switch is located on the rear ofthe instrument Reference Image 5 4 e Has programmable temperature control e Prior to operation the cooling plate should be dry and free of paraffin to prevent frost buildup Image 5 5 Cooling Plate Control Panel TANNER scientific A O LCD Display Screen A Setting Indicator B Running Indicator C Set Key D Increase E Decrease Image 5 6 Tanner Scientific TN1500 Embedding Center 9 6 Operation A Please ensure that all preparatory steps have been completed and safety precautions have been reviewed prior to use Getting Started Plugging In e Before you connect the instrument to the main electrical source please check that the power switch is in the O OFF position e Only use the power supply line provid
6. ate Operation 1 Reference Image 5 6 on page 9 2 Connect the cooling plate to power source Turn on the power switch located on the rear of the machine Green light will illuminate 3 Press SET key the compressor will stop running 4 Press or to adjust the temperature 5 Press the SET key to enter running mode Turn on 20 minutes prior to use to allow for temperature to be achieved e There are two indicator lights beside the screen If the indicator light underneath is on this means the compressor is working If the indicator light on top is on this means the compressor has stopped working e If the actual temperature is lower than setting temperature the compressor will automatically stop running When the actual temperature is higher than setting temperature the compressor will restart e For convenience the preset temperature range is 5 to 10 C 14 Tanner Scientific TN1500 Embedding Center 7 Cleaning amp Maintenance Cleaning e Equipment should be cleaned daily e Before performing all maintenance or cleaning of the equipment turn off the power switch and unplug the instrument e Only household grade cleaners should be used for cleaning the instrument e Acetone and benzene will damage paint on the surface of the instrument e Wipe area using a soft cloth dampened in alcohol 70 75 isopropyl or ethyl solution e Do NOT wipe dry Allow area to air dry to maximize disinfection Cha
7. chine e Only use the power supply line supplied with the instrument to connect with the socket e Make sure the grounding line of the 3 line power is securely connected before starting the machine in compliance with the GB9706 1 standards Tanner Scientific TN1500 Embedding Center 3 3 Technical Parameters Power Supply 110V 60Hz Fuses 15 AMP x2 Power Consumption 1000VA Operating Temperature Range 15 Cooling System DC 12 24 V Compressor Refrigerant R134A Dimensions LxDxH 26 5 x 22 75 x 16 5 Weight 77lbs Paraffin Reservoir 6L Thermal Chambers Approx 1 5L Cold Plate Approx 80 Cassettes Working Environment e Operate in an ambient temperature range of 59 409C e Humidity should be less than 80 e Atmospheric Pressure 860 hpa 1060 hpa e All heated components include overheat protection Temperature Control Functional Temperature Heated Separate Warm Up Time Unit Range Separately Temperature Control Prior to Operation Paraffin Reservoir 0 C to 85 C 2 hrs Paraffin Dispenser 0 C to 85 C 15 20 min Thermal Chambers 0 C to 85 C 15 20 min Hot Plate 0 C to 85 C 15 20 min Cold Plate 0 C to 10 C 30 min Forceps Holders 0 C to 85 C 15 20 min 4 Tanner Scientific TN1500 Embedding Center 4 Safety Precautions e This instrument should only be operated by qualified personnel e Read the user manual in full before use
8. ed with the instrument e Electrical socket must be a grounded socket e Connect power plug into the instrument before plugging into the electrical socket Starting Up the Machine e Turn on the main switch of the instrument Reference Image 5 3 on page 8 for power switch location e LCD display screen on the main control panel will show the following picture HOME SCREEN Paraffin AUTO MODE Tank PARAFFIN 65 C HOT PLATE 55 C Working DISPENSER 679C Table Paraffin Dispenser Running FRI 18 40 ly a E Right Thermal Chamber RIGHT 65 C gt e LEFT 55 C Ld gg Left Thermal CRYO gt Chamber Small Cryo Plate 10 Tanner Scientific TN1500 Embedding Center 6 1 Main Control Panel Setup Mode 1 Reference Image 5 2 on page 7 for the main control panel 2 Place instrument into manual mode by pressing the AUTO MAN key Light will illuminate under MAN 3 Press SET key to enter into set up mode 4 The main control screen will show the following picture SETUP SCREEN SETUP MODE SELECT 1 3 1 TEMPERATURE 2 TIMER 3 CLOCK e All settings temperature and time parameters will be automatically recorded on the computer chip When restarting the embedding center and pressing the AUTO MAN key it will run according to the pre set parameters 6 2 Temperature Setting 1 Select 1 TEMPERATURE using the gt keys 2 Press SET key to e
9. nging the Fuse e Only replace with the same model Refer to the Technical Parameters Failure to comply could invalidate the warranty 8 Warranty amp Service Tanner Scientific warrantees this product for a period of one year from the date of purchase provided that the customer complies with the rules ofthis manual Extended warranties are available Contact Tanner Scientific at 888 708 5233 for more information Tanner Scientific will not be responsible for the damage and other problems caused by abuse and misuse of the instrument Service Information If you require any service or parts during the warranty period please contact Tanner Scientific at 888 708 5233 or the sales agent who sold you the unit Prior to calling please have the instrument model and serial number ready Amendment Tanner Scientific reserves the right to change the technical parameters of any model for improving the function of our instruments Quality Guarantee e Tanner Scientific will ensure that every instrument sold has been strictly examined to ensure it meets our stringent quality and technical standards e Service terms are only provided for those who regularly use the instrument and operate the instrument according to its instructions Disposal of Unwanted Instruments Discarding the instrument and its parts should be done according to current laws and regulations Tanner Scientific TN1500 Embedding Center 15 9
10. nter into temperature setting status 3 Choose the working zone 4 Use or to increase or decrease the temperature figures 5 Once you have selected your desired temperature ranges press the AUTO MAN key to return back to the home screen Tanner Scientific TN1500 Embedding Center 11 Suggested Pre Setting Temperatures for these 6 Working Zones 1 Paraffin Tank 65 C 2 Right Thermal Chamber 65 C 3 Hot Plate 55 C 4 Left Thermal Chamber 55 C 5 Paraffin Dispenser 67 C 6 Small Cryo Plate NOTE Paraffin dispenser should always be 2 3 C higher than paraffin tank means section is heating means heating has stopped NOTE When the instrument reaches the desired set temperature the heating symbol will change from to E If the actual temperature is lower than the setting temperature the symbol will automatically change as temperature adjusts The unit is working only when these symbols are present and lit 6 3 Timer Setting 1 Select 2 TIMER using the d gt keys 2 Press SET key to enter into time setting mode 3 Use gt toselect time and day Press PB to choose time setting for Thursday Sunday or lt Q to choose time setting for Monday Wednesday 4 Press or to increase or decrease the time figure 5 Press AUTO MAN key to return back to the home screen Example To turn on at 7 30 and turn off at 16 30 on Tuesday follow these steps 1 Turn
11. of this instrument e Turn off and disconnect the instrument from its power supply before changing or cleaning any of its parts e Be aware there is a risk of burning when operating The paraffin reservoir mold warmer cassette bath work area and forceps holders are all hot e Make sure the power supply complies with the specified requirements before use Only connect to a grounded socket e The unit must not be exposed to direct sunlight or set up near an air condition vent e The back of the instrument must be placed at least 15cm from the wall Attention e Use heating elements only as needed e Turn to sleep mode or away when not in use Tanner Scientific TN1500 Embedding Center 5 5 Equipment Overview Tanner Scientific TN1500 Embedding Center Image 5 1 1 Paraffin Tank 7 Paraffin Waste Trays 2 Main Control Panel 8 Hot Plate 3 LED Light Control 9 Forceps Holder 4 Thermal Chamber 10 Rear Panels On Back of Instrument 5 Paraffin Dispenser 11 Cooling Plate 6 Small Cryo Plate 12 Cooling Plate Control Panel 6 Tanner Scientific TN1500 Embedding Center Image 5 2 TANNER A Left Arrow scientific B Right Arrow C Auto Manual LCD Display Screen a O D Small Cryo Plate On Off A B E Minus Key F Plus Key G Set Key 1 Paraffin Tank e Holds up
12. port COOLING PLATE Will not cool down Bad wire connection Controller or compressor may be broken Check the wires Contact technical support Will cool down but temperature cannot reach setting temperature Fan does not work or airflow is blocked Not enough refrigerating fluid Contact technical support 16 Tanner Scientific TN1500 Embedding Center TANNER scientific Information The information notes and figures that appear in this user manual represent the present state of knowledge as we comprehend it Tanner Scientific reserves the right to modify any specifications as well as manufacturing procedures without prior notice This makes it possible to continuously improve the technology and manufacturing technique used in our products For inaccurate statements drawings and technical illustrations in this user manual we exclude liability as far as permissible No liability is accepted for any monetary loss or consequential damage caused by or related to compliance with statements or other information in this user manual Tanner Scientific has no obligation to update the present manual periodically and on a continuing basis according to the latest technical developments nor to provide customers with extra copies or updates of this user manual This document is protected under copyright laws Any reproduction of illustrations or text
13. requires prompt documented consent by Tanner Scientific Statements drawings illustrations and other information regarding contents or technical details of the present user manual are not to be considered as warranted characteristics of our products These are determined only by the contract provisions agreed between us and our customers Refer to the name plate on the back of the unit for the serial number and manufacturing date of this instrument Tanner Scientific 1073 Tallevast Rd Sarasota Florida 34243 888 708 5233 Info TannerScientific com www TannerScientific com TANNER scientific 1073 Tallevast Rd Sarasota Florida 34243 888 708 5233 Info TannerScientific com www TannerScientific com
14. ustable LED illumination system w adjustable flow tube e Two module configurable options left to right or right to left Tanner Scientific TN1500 Embedding Center 2 Tanner Scientific TN1500 Embedding Center 2 Installation Transportation e Only transport the instrument vertically Any other position can cause damage Unpacking e Examine the package before opening If damaged do NOT open and immediately contact Tanner Scientific and the transportation carrier e Unpack the carton removing all components and user manual e Remove the unit from the carton using the two black securing straps e When lifting the instrument be sure to hold both sides by its base plate Placement e Place the instrument on an even and steady surface to avert vibration e Do NOT operate the instrument under a fan or directly below an A C duct e Do NOT operate the instrument around any electromagnetic interference e Do NOT place the instrument near any explosive or flammable materials e Do NOT place the instrument in an area where it would be exposed to direct sunlight e The back of the instrument MUST be placed at least 15cm away from the wall If the instrument is placed closer than 15cm from the wall it can cause substantial damage to the instrument and can invalidate the warranty Power Supply e An independent grounding power socket should be used for the ma
Download Pdf Manuals
Related Search
Related Contents
Enduit de réparation Progress Lighting P5501-31 Instructions / Assembly Moen YB9790 User's Manual "user manual" Pelco C747M User's Manual ECG Palmare Cardio B Cardio-B Palm ECG ECG de Grandstream Networks, Inc. V481 Hardware User`s Manual v2 SDメモリーカード SD-2GB CK-5 / CKE-5 CK-8 / CKE-8 Copyright © All rights reserved.
Failed to retrieve file