1. Operating State Default Communication Speed 9600 bps Code JIS8 Parity None STOP 1 bit Xon Xoff Disabled Local Echo Disabled ESC key function Standard Set H Tab Stop 1 CR operation CR Shift operation Simultaneous Auto Repeat No CS Yes DSR Yes Remote Control Messages Printing This message appears when images are being printed Error opening SCSI card or printer off or printer SCSI ID not 1 This message appears when there is a problem with the SCSI interface and or when the printer is off 1 Grab 2 Print 3 Burn 0 Exit This is the main screen of the remote control Press the 1 key to grab images from the video device Press the 2 key to print images press the 3 key to write images to a CD R disc and press the 0 key to conclude the patient examination Converting image This message appears when you capture images with the UPA P100MD It means that the data being captured is being converted to TIFF format Burning a CD This message appears when the UPA P100MD is writing images to a CD R disc An error occurred writing a CD This message appears when an error occurred when writing images to a CD R disc Enter Doctors name This message appears when you need to enter the doctor s name directly or by pressing a function key on the remote control unit 38 Enter Patients name This message appears when you need to enter the patient s name directly or by pressing a
2. 2 Keyboard The keyboard is composed of 26 alphanumerical keys 8 function keys a backspace key an escape key a space key and an enter key Escape key The ESC key allows you to cancel an operation such as typing the name of a doctor 4 Backspace key The BS key allows you to go back one space to correct mistakes ABC key Allows you to enter uppercase text The operation of this key can be changed Refer to the RM P110 Remote Control Unit User s Guide for details 6 abc key Allows you to enter lowercase text The operation of this key can be changed Refer to the RM P110 Remote Control Unit User s Guide for details Function keys You can assign strings of up to 16 characters to each of the 8 function keys This allows you to assign doctor names patient names or procedure names to the function keys Refer to the RM P110 Remote Control Unit User s Guide for details 11 Preliminary Connections Before using the UPA P100MD locally or through a network you need to make certain connections To make these connections use the following procedure RM P110 Remote Control Unit AC Power Inlet Locate the UPA P100MD s power cord On the rear panel of the device locate the AC inlet Connect the female end of the power cord
3. Do not write except with a felt tip marker or affix labels to discs Do not store discs in dusty dirty or humid locations in direct sunlight or near heat generating devices Avoid spilling liquid on discs To protect important data always keep discs in their protective case Digital Capture Unit User s Guide UPA P100MD Table of Contents Before Using the Digital Capture Unit 4 About the Hard Disk sees 4 About the rear connectors eese 4 About the CD R eee reete etus 4 Overview Digital Capture Unit Overview 7 lic el 8 Built in 8 Remote Control Unit eese 8 Linux Operating System sese 8 Networking Capabilities esses 8 Direct Printout Capabilities e 8 Presentation of the Digital Capture Unit Front Panel 2 2 ener 9 Rear Panel oiicscccccccseccecccccecentececescccssceseccesstescuene 10 Remote Control eese 11 Setup Preliminary Connections 12 Preliminary Settings 13 Connecting the Digital Capture Unit to a Network 14 Configuring the UPA P100MD 15 The Configuration Screen 16 Operation Grabbing Images eene 19 Vi
4. sends the hexadecimal character codes 1B and OD to the UPA P100MD The ESC key has the same effect as the ENT key Display ET H TAB STOP R OPERATION r aamu SET H TAB STOP Select the position to which the cursor moves when a horizontal tab is received Available choices are 1 5 and 10 By default this is 1 CR OPERATION Select how the cursor moves when a carriage return is received Available choices are Cr carriage return only and Cr Lf carriage return and line feed By default this is Cr Keyboard SHIF OPERATION Bimultaneous AUTO REPEAT No SHIFT OPERATION Select how the ABC and abc keys operate Available choices are Simultaneous and Changeover By default this is Simultaneous Changeover In this mode press the ABC or abc key to lock the input mode the uppercase or lowercase input mode remains enabled until the ABC or abc key is pressed again When the input mode is locked you can toggle directly to the opposite mode by pressing the corresponding shift key i e when locked in the uppercase input mode press abc to switch directly to the lowercase input mode Simultaneous In this mode you need to hold the ABC or abc key to input characters AUTO REPEAT Select whether characters input from the keyboard can be repeated by holding down the key Available choices are Yes and No When this
5. Be sure to select Multi session in the configuration screen Verify that the CD R disc in the tray is of a supported type Refer to General Specifications for details 30 Remote Control Unit User s Guide RM P110 Table of Contents Overview Remote Control Unit Overview 33 Fealtfes oci ricco oui cia VeRO UE EIL Sa ON 33 Palin Sizes 33 Large Display eite 33 8 Function Keys oerte 33 R5 232C Pott reos reris 33 Setup Connecting the Remote Control Unit 34 Operation Remote Control Setup Mode 35 Programming Function Keys 37 Specificati ons ee t ss 38 Display sd Eee eere er ta 38 Controller eie ees ae Nees 38 Keyboard 38 Default Operating States esse 38 Remote Control Messages esses 38 32 Overview Remote Control Unit Overview The RM P110 Remote Control Unit is used to control the operation of the UPA P100MD Digital Capture Unit It is connected to the Digital Capture Unit through an RS 232C port This section takes you through a quick overview of the unit Features Palm Size The RM P110 is compact and light making it very easy to use Large Display The remote control unit is equipped with an LCD that can display 4 rows of 20 characters each for easy readability 8 Function Keys The remote control comes with 8 fun
6. Mode of operation Continuous Design and specifications are subject to change without notice Before Using the Digital Capture Unit About the Hard Disk Compared to a floppy disk a hard disk has a greater memory capacity and needs less time to write or read data However dust and vibrations affect a hard disk s performance Furthermore the same as for a floppy disk a hard disk should never be placed near magnetic fields Hard disks contain built in safety mechanisms to protect data against dust and vibrations However to avoid losing data you should take the following precautions Do not use the hard disk in areas prone to vibrations or on unstable surfaces Do not move the hard disk while it is on Do not turn the power off while the hard disk is performing an operation Do not use the hard disk in areas prone to sudden temperature fluctuations i e more than 10 C Note that if the hard disk is damaged data may not be recoverable Do not forget to back up the data on the hard disk Make sure to back up images saved on the hard disk of the UPA P100MD on a regular basis Sony denies any responsibility if data saved on the hard disk becomes unrecoverable Make sure that the power cord and other connectors are properly connected Improperly connected power cord or connectors constitute a risk of fire or electric shock About the rear connectors Confirm that the cable connectors you are usi
7. Power Requirements 100 240 V 50 60 Hz Power Consumption 60W Input Current 0 6A 0 25A Temperature Range operating 10 to 35 C Temperature Range stored 20 to 60 C Humidity Range operating 20 to 80 96 No condensation Dimensions W x D x H 300 x 330 x 78 mm Weight 5 5 kg 26 UPA P100MD Audible Signals The UPA P100MD emits various audible signals in different situations often accompanied by a remote control message Here is a list of events and the UPA P100MD audible signals that accompany them Event SCSI or printer error Audible signal Low pitch beep 200 Hz 0 5 s Unit becomes ready to capture images 3 consecutive high pitch beeps 1000 Hz 0 25 s each Start of image capture or pressing 1 on the remote control unit High pitch beep 1000 Hz 0 15 s UPA P100MD searching for a video signal Repetitive high pitch beep 1000Hz 0 15 s each Connection of a printer to the UPA P100MD and when each full page of images is captured Medium pitch beep 750 Hz 0 15 s Start of printing or saving images to a CD R disc When using the RM P110 Remote Control Unit this signal is output when you push 2 or 3 Medium pitch beep 750 Hz 0 15 s Problem when writing images to a CD R disc Low pitch beep 200 Hz 0 5 s Printer initialization error Low pitch beep 200 Hz 0 5 s Print or Burn is selected Medium pitch beep 750 Hz 1
8. Save Config To access the configuration screen once more you need to type the address specified in the IP Address box i e http network IP address in the Address box of you Internet browser click Configuration then click Refresh Do not click Back on your browser e Star Over Click this button to cancel all the changes that you ve made in the configuration screen The configuration screen settings return to those at the time of opening the configuration screen 18 Using the Digital Capture Unit consists of 4 major operations all of which can be done using the remote control These operations are grabbing images viewing images printing images and saving images to CD R disc For details about remote control operations refer to the RM P110 User s Guide Grabbing Images To capture images from a video device using the RM P110 Remote Control Unit use the following procedure If you are using a custom made device to control the UPA P100MD the procedure to grab images depends on your control device Before you begin using the device as described below please make sure that a video device is properly connected to the UPA P100MD and turned on The connection procedure depends on your video device f you are using a digital video printer UPA D50 or UPA D70XR also make sure that it is properly connected and turned on 1 Turn on the Digital Capture Unit After approximately 1 minute you hear 3
9. and hold the 1 key then turn on the UPA P100MD The following screen appears SETUP MODE find Communication Display Keyboard RS 232C signals 2 Using the or gt keys move the cursor to the setting that you want to modify Press the ENT key to select a mode Refer to the table on the next page for the meaning of the display items in this screen 3 Use the and gt keys to move to the desired item Use the f and l keys to change the setting Repeat for other settings When you finish making all the settings in one screen press the ENT key This takes you to the next screen of the selected Setup Mode item or back to the SETUP MODE screen Repeat steps 2 and 3 as necessary 4 When you finish making all the desired settings move the cursor to End in the SETUP MODE screen and press the ENT key The RM P110 returns to the screen displayed just before entering the Setup Mode The operating state settings made in the Setup Mode are stored in EEPROM electrically erasable ROM so that the settings are retained even when the remote control is off 35 SETUP MODE screen item End Contents Explanation Select to exit the Setup Mode Communication COMMUNICATION SPEED Select the communication speed for data transfer Available choices are 110 150 300 600 1200 2400 4800 9600 and 19200 bps By default this is 9600
10. beeps and Enter Patients name appears on the remote control display Enter Patients name Note that if you have cleared the Remote Control Unit Support check box in the configuration screen the system assigns patient doctor and procedure names automatically f you press ENT without typing a patient s name the system assigns a name to the patient automatically f Video Printer Support is selected in the UPA P100MD configuration screen and if there is no printer connected or it is off the message Error opening SCSI card or printer off or printer SCSI ID not 1 appears on the display before the screen above appears 2 Using the remote control keyboard enter the patient s name The patient s name has a maximum length of 18 alphanumerical characters Press ENT Enter Doctors name appears Enter Doctors name 3 Using the remote control keyboard enter the name of the doctor performing the procedure The doctor s name has a maximum length of 18 alphanumerical characters Press ENT Enter Procedure name appears Enter Procedure name Note that if you press ENT without entering any information the system assigns the name Doctor to the user Also if you have selected Never ask in the configuration screen this information does not appear and the system goes directly to the next item Also note that if you specify a default doctor name and select Use default i
11. disk of the UPA P100MD The date that an image was captured appears above the image when you open it 3 Click the image The download process begins Follow the instructions on the screen 4 Repeat for other images as necessary 5 When you finish saving images to your computer select the address of the UPA P100MD in the Address box of your Internet browser The Sony UPA P100MD screen appears 6 Click Configuration to access the configuration screen 24 7 In the Expiration Time days box type a number of days lower than the current setting Be sure to type an expiration time that will only delete the images that you have saved in steps 2 to 4 If you do not images that have not been backed up but exceed the new expiration time are deleted by the UPA P100MD when you click Save Config in step 8 resulting in the loss of images Click Save Config The UPA P100MD restarts When the UPA P100MD restarts all the images on the UPA P100MD hard disk that exceed the new expiration time are deleted clearing space on the hard disk Access the configuration screen again return Expiration Time days to its original setting and click Save Config Saving Images to CD R discs 1 Follow the procedure outlined in Viewing Images to access the images of specific doctors and patients on the UPA P100MD Note the names of doctors and patients that you want to save to CD R disc When Enter Patients name
12. function key on the remote control unit Enter Procedure name This message appears when you need to enter the procedure name directly or by pressing a function key on the remote control unit No Video present This message appears when there is no video device connected to the UPA P100MD Printer offline This message appears when there is no printer connected to the UPA P100MD or when the connected printer is off CD not configured This message appears when you try to write images to a CD R disc that is not formatted Printer not configured This message appears when the Video Printer Support check box in the configuration screen is not selected and you push the 2 key print command on the remote control keyboard Rebooting system This message appears when you click the Save config button indicating that the unit is restarting Booting system This message appears while the UPA P100MD is starting when you turn it on Hard Disk 80 used See user manual This message appears when the UPA P100MD hard disk is 80 full See Managing UPA P100MD Disk Hard Space for details about what to do when this message appears Hard Disk FULL See user manual This message appears when the UPA P100MD hard disk is completely full and no further images can be captured See Managing UPA P100MD Hard Disk Space for details about what to do when this message appears Sony Corporation http Awww world sony com
13. like except when writing information on a CD R disc or the hard disk Networking Capabilities Although the Digital Capture Unit is connected to a single computer during setup it can also be configured for use through a local area network LAN This is highly useful when it is necessary for several users to share view images on the unit s hard disk However you can only control image capture and recording with the remote control unit or a custom made control unit not through the network Direct Printout Capabilities Connecting a digital color printer such as the UP D50 or UP D70XR to the Digital Capture Unit allows images to be printed directly from the Digital Capture Unit Presentation of the Digital Capture Unit Front Panel This section describes the unit s front panel SONY DIGITAL CAPTURE UNIT UPA P100MD 3x 9954 SE o 4 6 4 Power switch and indicator 4 CD R drive Busy indicator This button turns the unit on or off The green This indicator blinks when the CD R drive is busy indicator lights up when the unit is on writing information on a CD R disc 2 CD R drive IrDA infrared receptor The CD R drive is used to write the images Currently not supported captured on the unit s hard disk to CD R discs The images on the CD R disc created with the unit s 6 HDD LAN access indicator CD R drive can be accessed from any CD ROM This in
14. mode is Yes pressing a key on the keyboard for one second or more results in the character being input repeatedly By default this is No RS 232C signals Control RS 232C Inpu t Signals CS DSR Hes Yes CS Select whether the Clear to Send signal is effective when received Available choices are Yes and No By default this is Yes DSR Select whether the Data Set Ready signal is effective when received Available choices are Yes and No By default this is Yes 36 Programming Function Keys The remote control comes with 8 function keys F1 to F8 that can be programmed with up to 16 alphanumerical characters The setting procedure of the function keys is similar to that of the other modes of the remote control except for a selection menu allowing you to choose the function key F1 to F8 that you want to program Follow the procedure below to program the function keys 1 Press and hold the 3 key and turn on the UPA P100MD The following screen appears Select Function fl F2 F3 F4 F5 F6 F7 F8 End 2 Using the or keys move the cursor to the function key where you want to preset a name 3 Press the ENT key to select the function key Three lines of the 165 programmable characters appear at the top of the 4 line display W 525 0123 456789 lt gt ABCDEFG HIJKLMNOPORS TUVWXYZ El 4 Use the arrow
15. the unit try these measures Relocate the unit with respect to the susceptible equipment Plug this unit and the susceptible equipment into different branch circuit Consult your dealer According to Standard EN60601 1 2 and CISPR11 Class B Group 1 Caution When you dispose of the unit or accessories you must obey the law in the relative area or country and the regulation in the relative hospital Warning on power connection for medical use Use a proper power cord for your local power supply United States Canada Continental Europe Plug Type HOSPITAL HOSPITAL LP 34A GRADE GRADE Female end E62405 LR53182 LS 60 Cord type E159216 LL112007 1 HO5VV F Min Type SJT Min Type SJT Min 18AWG Min 18AWG Minimum cord 10A 125V 10A 125V 10A 250V set rating Safety UL Listed CSA VDE approval Note Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked Hospital Only or Hospital Grade The following specifications are applied only to the UPA P100MD Storage and transport temperature 20 C to 60 C 4 F to 140 F Storage and transport humidity 20 96 to 80 96 no condensation allowed Protection against electric shock Class I Protection against harmful ingress of water Ordinary Degree of safety in the presence of flammable anesthetics or oxygen Not suitable for use in the presence of flammable anesthetics or oxygen
16. to the AC inlet A WN Connect the RM P110 Remote Control Unit connector to the remote control port on the rear panel When using a foot switch or other such external control device connect the foot switch to the parallel port on the rear panel of the UPA P100MD 5 On the rear panel of the UPA P100MD locate the network port Connect a LAN cross cable to this port 6 Connect the other end of the LAN cross cable to the network port of your computer You are now ready to make preliminary settings Before connecting the unit make sure that you are using the appropriate type of cross cable If you are not sure of what type of cable to use ask your system administrator f your computer is connected to a network to be able to make the preliminary settings of the UPA P100MD it is necessary to isolate your computer from the network so that the connection between the UPA P100MD and your computer is local The Remote control port outputs 5 V therefore do not connect it to the RM P110 Remote Control Unit when the UPA P100MD is on 12 Preliminary Settings Before using the UPA P100MD you need to make certain settings on your computer Once you make these settings you can access the device s configuration screen Use the following procedure to make the preliminary settings Note Please use Windows computers to configure and access the UPA P100MD Other operating environments are not supported by the UPA P
17. type a valid IP address In the Netmask box type a valid subnet mask address In the Gateway box type a valid gateway address f you are unsure or do not know what to type in any of the three boxes above consult your system administrator before making any changes Entering inappropriate values may make the system inaccessible resulting in loss of all data on the unit s hard disk If this occurs please contact a Sony service center near you f you forget the IP address of the UPA P100MD after changing it you will not be able to access the unit To prevent this please write down the IP address and store it in a safe place Click Save Config The Configuration saved message appears and the UPA P100MD restarts 10 Close your browser 1 1 Disconnect the unit from your computer and connect it to the network If you are unsure of how to connect the unit to your network ask your system administrator Configuring the UPA P100MD Before using the UPA P100MD you need to configure it through the configuration screen Follow the procedure below to configure the UPA P100MD 1 Turn on a computer connected to the network to which the UPA P100MD is connected Or turn on the computer directly connected to the UPA P100MD if connected locally 2 Start your Internet browser 3 Type the following address in the address bar of your Internet browser f using the UPA P100MD through a network http net
18. 100MD The procedures in this guide are based on Windows 98 dialog boxes and windows Screen appearance may differ slightly in other Windows versions 1 Start the computer to which the Digital Capture Unit is connected On the Start menu point to Settings and then click Control Panel In Control Panel double click Network The Network dialog box appears A c N On the Configuration tab click TCP IP Network 2 x Configuration identification Access Control The following network components are installed E Client for Microsoft Networks 3 corega FastEther Il PCI TX Ethernet Adapter Y NetBEUI Add Remove Properties Primary Network Logon Client for Microsoft Networks ivi File and Print Sharing f Description TCP IP is the protocol you use to connect to the Internet and wide area networks Note If there is more than one TCP IP entry in the Network dialog box make sure that you select the TCP IP that belongs to the network adapter 5 With TCP IP selected click Properties The TCP IP Properties dialog box appears Click the IP Address tab On the IP Address tab click Specify an IP address In the IP Address box type the following 192 168 3 201 Co N Oo In the Subnet Mask box type the following 255 255 255 0 TCP IP Properties 2 x Bindings Advanced NeBlos DNS Configuration Gateway WINS C
19. 30 Save Config Start Over E lel Done bn lo Internet e Hospital Unit Name Hospital Video NTSC C PAL Type the name of the hospital clinic or department associated with the Digital Capture Unit This area can be composed of a maximum of 7 alphanumerical characters When capturing images this name automatically becomes part of the file name saved on the hard disk therefore a short name is preferable When consulting your files it makes it easier to determine their origin Note You can only use alphanumerical characters upper and lowercase and hyphens with the UPA P100MD All other characters are rejected including spaces In the configuration screen an error message appears to tell you so when you click Save Config to inform you if there are illegal characters in use or a name exceeds the maximum permitted length This setting needs to be adjusted according to the type of incoming video signal PAL or NTSC This usually depends on the country in which the device is used Color Color Mono This setting needs to be adjusted according to the type of incoming video signal Color saves images as 24 bit color images whereas Mono saves images as 8 bit black and white images e Remote Control Unit Support M Be sure to select this box when you use the RM P110 Remote Control Unit and want to input your patient doctor and procedure names When you clear Remote Control Unit Support the unit assigns names automati
20. 5 s 27 Pin Assignments Control signals are available at the parallel port of the UPA P100MD making it possible to design a customized external control device Refer to the illustrations and chart below for the pin assignments Parallel Port Pin Assignments 1 0000000000060 0000000000 25 14 Female connector on the rear panel of the UPA P100MD Pin number Signal 1 STROBE Not supported by the unit at this time 2 DATAO Not supported by the unit at this time 3 DATA1 Not supported by the unit at this time 4 DATA Not supported by the unit at this time 5 DATAS Not supported by the unit at this time 6 DATAA Not supported by the unit at this time 7 DATAS Not supported by the unit at this time 8 DATA6 Not supported by the unit at this time 9 DATA7 Not supported by the unit at this time 10 ACK BURN 11 BUSY EXIT 12 PE PRINT 13 SELECT GRAB 14 AUTOFEED Not supported by the unit at this time 15 ERR Not supported by the unit at this time 16 INIT Not supported by the unit at this time 17 SLIN Not supported by the unit at this time 18 GND 19 GND 20 GND 21 GND 22 GND 23 GND 24 GND 25 GND piniO BURN pint1 EXIT pin12 PRINT pin13 GRAB GRAB PRINT EXIT BURN pin18 19 20 21 22 23 24 25 GND 28 Remote Control Port Pin Assign
21. America New_York J You need to select the local time zone here Select the time zone in your area from the list e Date Time yyyy mm dd 2001 fo for fia fee Set the date and time in these boxes e Set Date After setting the date and time be sure to select this box Otherwise the new settings are not applied when you click Save Config eLanguagefEnglish m Select the language that you want to use for the remote control display You can choose one of 12 languages from the list as the display language By default this language is English Also note that the configuration screen language is not affected by this setting e Delay Capture Be sure to select this check box when you want to capture images from an analog printer synchronized with the UPA P100MD The capture of images is delayed by the time specified in the Delay value in milliseconds box By default this is 250 e Delay value in milliseconds 250 In this box enter the delay capture time in milliseconds By default this is 250 Debounce value in milliseconds Bo Enter the minimum length of time during which the switch of an external control unit connected to the UPA P100MD parallel port must be depressed The shorter the time set here the faster the key response By default this is 30 e Save Config Click this button to save the changes that you make in the configuration screen The device restarts automatically when you click
22. CD R disc remains open for additional recording sessions In other words you can record several sessions consecutively each in a different folder for several different patients on the same CD R disc Be careful of the value you specify hore When you modify this setting and click Save Config the UPA P100MD deletes images older than the expiration time If the expiration time is set to a value lower than the age of images that you need to keep you will lose those images e Video Printer Support C 2X2 2X3 Be sure to select this box when using a UP D50 or UP D70XR digital video printer Leave it clear otherwise The Digital Capture Unit starts faster when this box is not selected This setting also determines the layout of the images on one page Dod when printing Be sure to back up the image files on the hard disk of The UP D50 and UP D70XR default SCSI settings Be te ee MUR pe the UPA P100MD before reaching the Expiration CUM Time See Managing UPA P100MD Hard Disk Terminator on Space for details eID 001 Refer to the appropriate printer user s guide for e Doctors Name Always ask C Never ask C Use default details Specify whether you want the doctor s name to be IP Address 521683200 requested every time you begin a new procedure Always ask never requested Never ask or that the system use the default doctor name Use default If in doubt as to what IP subnet mask or gateway
23. Screen Close your browser and restart the UPA P100MD Cannot print Verify connection with the printer Verify that the printer is on Verify that the printer in online The printer must be turned on before the UPA P100MD Cannot write on a CD R Verify that there is a CD R disc in the CD R drive Verify that your are using a blank CD R disc If you use a previously written CD R disc check that it was written in the UPA P100MD Multi session recording mode CD R discs written in the Single session recording mode or on other systems cannot be used Verify that the CD R disc in the tray is of a supported type Refer to General Specifications for details Cannot capture images The hard disk is full Refer to Managing Hard Disk Space for details The message No Video present appears on the RM P110 display Verify the connection with the video device Verify that the video device is on The message An error occurred writing a CD appears on the RM P110 display Verify that there is a CD R disc in the CD R drive Verify that your are using a blank CD R disc If you use a previously written CD R disc check that it was written in the UPA P100MD Multi session recording mode CD R discs written in the Single session recording mode or on other systems cannot be used If you use a previously written CD R disc in Single session recording mode this message appears
24. a address to specify ask your system administrator Type the default doctor name to be used by the Netmask 582582580 system here The default doctor name can be l composed of up to 18 alphanumerical characters Specify a valid IP address in this box Specify the subnet mask address of your network here e Procedure Name Always ask C Never ask Use default 3218838 Specify whether you want the procedure name to be requested every time you begin a new procedure Always ask never requested Never ask or that the system use the default procedure name Use default e Gateway If necessary specify a gateway address here 17 e Default Name pocedue Type the default procedure name to be used by the system here The default procedure name can be composed of up to 10 alphanumerical characters Note To prevent the names of the image files on the UPA P100MD from being truncated when you view them in the browser file list use a short default doctor name and procedure name Also note that when you specify a default doctor name and or default procedure name and select Use default the default names appear on the RM P110 Remote Control Unit display where you can modify them if necessary If you do not need to make any modifications simply press the ENT key to proceed See Grabbing Images for details about entering names using the remote control e Time Zone
25. appears enter the name of one patient noted in step 2 then press the ENT key When Enter Doctors name appears enter the name of the doctor related to the patient entered in step 3 then press the ENT key When Enter Procedure name appears enter any procedure name then press the ENT key It does not matter what procedure name is entered Insert a blank CD R disc in the UPA P100MD CD R drive Without capturing any additional images press 3 on the remote control keyboard All the images for the doctor s and patient s entered in steps 3 and 4 are written to the CD R disc 8 Repeat for other patients and or doctors as necessary 9 When you finish writing images to CD R discs select the address of the UPA P100MD in the Address box of your Internet browser The Sony UPA P100MD screen appears 10 Click Configuration to access the configuration screen 1 1 In the Expiration Time days box type a number of days lower than the current setting Be sure to type an expiration time that will only delete the images that you have written to CD R disc in steps 3 to 8 If you do not images that have not been backed up but exceed the new expiration time are deleted by the UPA P100MD when you click Save Config in step 12 resulting in the loss of images 12 Click Save Config The UPA P100MD restarts When the UPA P100MD restarts all the images on the UPA P100MD hard disk that exceed the new expiration tim
26. bps CODE Select the type of code used to transfer data Available choices are JIS7 JIS8 and ASCII By default this is JIS8 PARITY Select the type of parity used to transfer data Available choices are None Even and Odd When CODE is set to JIS7 or ASCII two more choices are available Spaced and Marked By default this is None COMMUNICATION SPEED E600 bps CODE PARITY STOP JIS8 None 1Bit ENT XON XOFF LOCAL ECHO Disable Disable ENT STOP Select the number of stop bits added to transferred data Available choices are 1Bit and 2Bit By default this is 1Bit ESC FUNCTION Standard XON XOFF Select whether to use handshaking to transfer data Available choices are Enable perform handshaking and Disable no handshaking By default this is Disable LOCAL ECHO Select whether data input from the keyboard is to be echoed to the display and stored in the remote control memory or displayed without storage Available choices are Enable and Disable By default this is Disable ESC FUNCTION Select the effect of the ESC key Available choices are Standard and Non standard By default this is Standard Standard In this mode pressing the ESC key sends a hexadecimal character code 1B to the UPA P100MD The display status is unaffected Non standard In this mode pressing the ESC key
27. cable 10Base T or 100Base TX connect the UPA P100MD directly to the appropriate network interface on your computer 2 Turn on the unit and wait until it is ready for operations 3 On your computer start your Internet browser Before you proceed to step 4 make sure that your Internet browser does not use a proxy server 4 In the Address bar of your browser type http 192 168 3 200 The Sony UPA P100MD screen opens 5 Click Configuration The Sony Digital Capture Unit UPA P100MD Configuration screen opens SONY UPA P100MD configuration Microsoft Internet Explorer 8 xj Ele Edt View Favorites Tools Help EIS 8 2 9 B a Gm J3 B5 Back Forward Stop Refresh Home Search Favorites History Mail amp ddress http 192168 159 143 cai bin loadconfig cgi y Go Links Video NTSC C PAL Color Color Mono r Support IV C Single session Multi session ter Support C2X2 2X3 ress 192 168 159 143 Netmask 255 255 255 0 Gateway melaya A Always ask C Never ask C Use default m doctor Name Always ask C Never ask C Use default Default Name procedure Time Zone America New York z Date Time vyyy mm dd Language English m Delay Capture Delay value in millisec Debounce value in millisecond 30 Save Config Start Over 192 168 159 860 Bom for fm pee Set Date F nds 250 lel Done Bs lo Internet 14 6 7 8 9 In the IP Address box
28. cally 16 e CD Recorder Support M Single session C Multi session e Expiration Time days ha Be sure to select this box when you want to write Type the number of days after which the files and information to a CD R disc It is possible to select folders saved on the unit s hard disk are deleted The either Single session or Multi session maximum number of days you can set is 60 For When you select Single session you can only write examples if you type 10 days in this box all the files the images taken for one patient on a CD R disc and folders on the Digital Capture Unit hard disk are Patient deleted after 10 days Users that save large numbers l Grab 2 Print of images every day should be careful of the number Tue Jul e EU 30 of days that they set here It should be set so that there are never more than 5000 images on the unit s hard disk at any one time In other words 5000 images saved per day maximum safe number of days that you can set for Expiration Time If you allow too many images to accumulate on the unit s hard disk it will become full and further image capture with the Digital Capture Unit will be impossible By default this is 14 days When you finish an examination that is when you press 0 on the remote control keyboard the disc can no longer be used to write further information When you select Multi session images are written to CD R disc but when you finish an examination the
29. ction keys that can be used to save strings of up to 16 characters to each of the 8 function keys This allows you to assign doctor names patient names or procedure names to the function keys RS 232C Port The RM P110 handles data through an RS 232C port making it easy to design custom control switches 33 Connecting the Remote Control Unit To connect the RM P110 Remote Control Unit to the UPA P100MD Digital Capture Unit follow the procedure below 1 On the rear panel of the Digital Capture Unit locate the remote control port 2 Connect the remote control connector to the remote control port 3 Secure the connector by tightening the screws Do not overtighten the screws as this could damage the connector or the remote control port 34 With the RM P110 Remote Control Unit it is possible to perform 3 operations on the UPA P100MD Digital Capture Unit These operations are grabbing images printing images and saving images to CD R It is also possible to assign strings of up to 16 characters to each of the 8 function keys This allows you to assign doctor names patient names or procedure names to the function keys This section contains an overview of the functions of the major keys Operation of the unit with the Digital Capture Unit is explained in the UPA P100MD User s Guide Refer to it for more details Remote Control Setup Mode Use the following procedure to switch to the Setup Mode 1 Press
30. ctoredoctortPatient patient O0TImage tSession SNlon eUSE dg eG Patient patient 001 Procedure procedure Date Time 07 04 01 1325 94610389 nemoto akira zla J Jl 2 Done fg Inemet Z Sample images When you click a thumbnail it opens as a JPEG file When you click the JPEG it opens as a TIFF files and can be saved To view other files click another thumbnail lt lt or gt P Click Top to go back to the Choose a doctor screen Whenever you access a screen that contains image files click Refresh to update it and make sure you are looking at the latest contents of the screen To view the images of another patient click the Back button of your browser click Refresh then select the desired patient 21 Printing Images If a digital color printer is connected to the device s SCSI port you can print images from the unit During the examination of a patient you can print the captured images at any time Follow the procedure below to print images from the unit 1 Proceed with the capture of images as outlined in Grabbing Images 2 When you are ready to print the images captured during the current examination press 2 on the keyboard of the remote control When you do one of two things happens All the images captured from the beginning of the current examination up to the present are printed All the images captured during the current examinat
31. dicator blinks amber when the hard disk of drive the unit is being accessed It blinks green when the LAN is accessed CD R drive eject button This button is used to open or close the CD R drive When writing to a CD R disc this button should not be used Because the CD R tray opens automatically after writing on a disc do not place any obstruction in front of the CD R tray Rear Panel This section describes the unit s rear panel H3 na m mo 9 8 7 Remote control port D sub 9 pin When using the RM P110 Remote Control Unit sold separately connect it to this port This connector outputs 5 V therefore do not connect it to any device other than the RM P110 Remote Control Unit or other control device designed for use with the UPA P100MD If you connect a device other than the remote control to this connector there is a risk of damage to the device 2 Equipotential terminal When connected this terminal ensures that all the system components are at the same electrical potential AC power inlet Connect to a power outlet using the provided power cord 4 Video input RCA PIN jack Input analog video signals Connect to the optical device s video output S video input mini DIN 4 pin Input analog s video signals Connect to the optical device s s video output 6 SCSI port half pitch 50 pin This port connects t
32. e Q N Y 3 206 230 11 1 Digital Capture Unit Remote Control Unit User s Guides UPA P100MD RM P110 2001 Sony Corporation WARNING To prevent fire or shock hazard do not expose the unit to rain or moisture To avoid electrical shock do not open the cabinet Refer servicing to qualified personnel only THIS APPARATUS MUST BE EARTHED For UPA P100MD Symbol on the products This symbol indicates the equipotential terminal which brings the various parts of a System to the same potential This symbol is intended to alert the user to the presence of important operating and maintenance servicing instructions in the literature accompanying the appliance When using a LAN cable For safety do not connect to the connector for peripheral device wiring that might have excessive voltage For the customers in the U S A for UPA P100MD This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference i
33. e are deleted clearing space on the hard disk 13 Access the configuration screen again return Expiration Time days to its original setting and click Save Config 25 Specifications and Troubleshooting The following section contains information about general specifications pin assignments and troubleshooting General Specifications Item Specifications Notes CPU Celeron 566 MHz or faster Main Memory 64 MB or more HDD 20 GB or more OS Linux Recording Media CD R discs CD R disc 12x speed Orange Book part Il 650 700 MB ISO 9660 format Joliet Rock Ridge extension Track Once Single session or Multi session recording mode File Format TIFF JPEG Data Size NTSC 640 x 480 pixels PAL 758 x 576 pixels PAL 712 x 480 CCIR601 CD Write Read Speed Write 12x Read 32x PC Card Slot Composite In 1 0 V p p 75 Q RCA PIN jack e Y C In Y 1 0 V p p 75 Q Mini DIN 4 pin C 1 0 V p p 75 Q SCSI SCSI 2 Half pitch 50 pin O Connection e PS2 For service use only Mini DIN 6 pin Remote RM P110 Remote Control Unit only D sub 9 pin Parallel D sub 25 pin IEEE 1394 Not supported Series A 4 pin x 2 USB Not supported x 2 Analog RGB For service use only D sub 15 pin 10Base T 100Base TX RJ 45 IrDA Not supported Composite Out Not supported RCA PIN jack Y C Out Not supported Mini DIN 4 pin Speaker Out Not supported PIN
34. emote Control Unit UP D50 or UP D70XR Digital Color Printers External Control Device not provided by Sony Network if applicable The illustration above presents the connections of the UPA P100MD For explanations about the UPA P100MD connectors and rear panel see Rear Panel on page 10 For more about the RM P110 Remote Control Unit see Remote Control on page 11 and consult the Remote Control Unit User s Guide on page 3 For explanations about how to connect the UPA P100MD to a network see Connecting the Digital Capture Unit to a Network on page 14 Features Built in CD R The built in CD R drive of the Digital Capture Unit allows you to save images to CD R discs Using the Digital Capture Unit s configuration screen you can choose to write images to CD R discs using either single session or multi session recording Remote Control Unit The remote control allows one touch control of the unit With its standard 80 character display its 8 function keys and uppercase and lowercase typing capabilities the remote control unit is versatile and easy to use Linux Operating System The flexibility of the Linux operating system does not require long shutdown and start up procedures This means that you can turn the unit on or off as you
35. ewing Images 20 Printing Images 22 Saving Images to CD R Discs 22 Accessing Images on a CD R Disc 23 Finishing Examinations 24 Managing UPA P100MD Hard Disk Space 24 Saving Images to Your Computer 24 Saving Images to CD R discs 25 Specifications and Troubleshooting General Specifications 26 UPA P100MD Audible Signals 27 Pin Assi nments ceeeeeeeee 28 Parallel Port Pin Assignments 28 Remote Control Port Pin Assignments 29 Troubleshooting 30 Troubleshooting eere 30 Microsoft and Windows are registered trademarks of Microsoft Corporation Other company names and product names mentioned in this guide are also trademarks and registered trademarks Overview Digital Capture Unit Overview The Digital Capture Unit is used to record analog images from video devices in JPEG and TIFF format As need arises it is also possible to save the recorded images on CD with the built in CD R drive or print the images with a digital color printer UP D50 UP D70XR Furthermore if the unit is connected to a network users can preview the images on the unit s hard disk using an Internet browser RM P110 R
36. ion since the last printout are printed This way you can print images in batches or all at once During printing the unit cannot be used You must wait until the message Printing disappears and the remote control display returns to the function screen 3 Continue capturing images for the same patient until you are finished 4 When you have finished the examination press 0 Enter Patients name appears on the remote control display and the unit is ready to begin a new operation When Video Printer Support is selected in the configuration screen the Digital Capture Unit emits a beep when capturing images When you select 2x2 the unit emits a beep after it captures 4 images When you select 2x3 the unit beeps after it captures 6 images Therefore you can use this beep as an indicator of when one page of images can be printed Saving Images to CD R Discs Itis possible to write images from the unit to a CD R disc ISO 9660 format Joliet Rock Ridge extension which can later be accessed from a different CD ROM drive or archived for reference Follow the procedure below to save data to a CD R disc using the Digital Capture Unit s CD R drive If you want to print images we recommend that you do so before saving images to a CD R disc because the UPA P100MD returns to Enter Patients name after the writing procedure t is important not to turn off the unit while writing images to a CD R disc as this co
37. keys to move the cursor to the first character that you want to use Press ENT The blinking character appears at the end of the string at the bottom of the display Repeat this step as necessary The programmable string can be of a maximum length of 16 alphanumerical characters including hyphens If you want to erase all the characters input so far press the 0 key 5 Press the key to return to the previous screen This has the effect of saving the input string in the remote control memory 6 Select another function key and repeat steps 3 to 5 If you wish to finish select End and press the ENT key to exit programming mode 37 Specifications This section contains the hardware specifications of the device Display Display Type STN type LCD Display Field 20 characters x 4 lines Display Area 57 55 x 17 15 mm Character Size 2 45 x 3 95 mm Displayable Alphanumerical 165 Character Type Cursor Numerical input blinking block Uppercase input underscore Lowercase input superimposed blinking block and underscore Controller MPU 8 bit CMOS microprocessor Memory RAM 32 KB CMOS RAM 24 KB for display buffer 32 KB CMOS PROM 256 byte CMOS EEPROM ROM Keyboard Keyboard Keys 28 input keys 8 function keys that can be programmed with up to 16 characters each Default Operating States
38. ments This port is reserved for the RM P110 Remote Control Unit Connecting another device to this port could cause it to malfunction Male connector of the RM P110 Remote Control Unit connection cable Pin number Signal 1 DC 5 V maximum 500 mA RxD TxD DTR SGND DSR RTS CTS Q o oj or AJOJN RI Not supported by the unit at this time This connector outputs 5 V therefore do not connect it to any device other than the RM P110 Remote Control Unit or other control device designed for use with the UPA P100MD If you connect a device other than the remote control to this connector there is a risk of damage to the device The screws of the remote control port of the Digital Capture Unit are threaded in inches Make sure that you have appropriate screws on the connector 29 Troubleshooting This section contains information about how to rectify commonly encountered problems with the UPA P100MD Digital Capture Unit Troubleshooting Below are remedies to some common problems that can be encountered when using the UPA P100MD Digital Capture Unit Problem Remedy The unit does not respond to control commands Restart the device Cannot access the configuration screen or the image screen Verify that the address in your Internet browser is correct If it is not type the correct address Cannot make changes in the configuration
39. n the configuration screen the default name appears allowing you to modify it if necessary If you do not need to modify the name simply press the ENT key to proceed See The Configuration Screen for details 4 Using the remote control keyboard enter the name of the procedure being performed The procedure name has a maximum length of 10 alphanumerical characters Press ENT The function screen appears on the remote control display Patient 1 Grab 2 Print 3 Burn O Exit Tue Jul 31 17 30 f there is no video device connected to the UPA P100MD the No video present message appears on the display of the remote control You are unable to use the UPA P100MD until a video device is properly connected and turned on Note that if you press ENT without entering any information the system assigns the name Procedure to the current procedure However if you have selected Never ask in the configuration screen this screen does not appear and the system goes directly to the next screen 19 Also note that when you specify a default procedure name and select Use default in the configuration screen the default name appears allowing you to modify it if necessary If you do not need to modify the name simply press the ENT key to proceed See The Configuration Screen for details 5 Make sure that your video device is connected to the UPA P100MD and outputting a video signal When you are ready to ca
40. n which case the user will be required to correct the interference at his own expense You are cautioned that any changes or modifications not expressly approved in this manual could void your authority to operate this equipment This device requires shielded interface cables to comply with FCC emission limits Model UPA P100MD is Non Patient Equipment This unit can not be used in the vicinity of patients Patient Vicinity zt For the customers in Canada for UPA P100MD This unit has been certified according to Standard CSA C22 2 NO 601 1 For UPA P100MD Important safeguards notices for use in the medical environments 1 Allthe equipments connected to this unit shall be certified according to Standard IEC60601 1 IEC60950 IEC60065 or other IEC ISO Standards applicable to the equipments 2 When this unit is used together with other equipment in the patient area the equipment shall be either powered by an isolation transformer or connected via an additional protective earth terminal to system ground unless it is certified according to Standard IEC60601 1 Patient Area 3 The leakage current could increase when connected to other equipment 4 This equipment generates uses and can radiate frequency energy If it is not installed and used in accordance with the instruction manual it may cause interference to other equipment If this unit causes interference which can be determined by unplugging the power cord from
41. ng can be inserted into the terminal outside of the molding Also note that the screw posts next to the D sub connectors on the rear panel of the unit are threaded in inches When using cables other that the one provided for use with this unit make sure that you use inch threaded Screws Do not connect the RM P110 to the UPA P100MD when it is on Please make sure that the UPA P100MD is off before connecting the RM P110 Connecting the RM P110 to the UPA P100MD when it is on could cause it to malfunction About the CD R Operation and Storage Precautions When using the CD R drive to write information on discs please pay attention to the following Do not use or store the drive where it might be subjected to vibrations or shocks Do not use or store the drive in hot or humid locations or direct sunlight When writing data on a disc do not turn the Digital Capture Unit off or push the Eject button of the CD R drive Avoiding Condensation Problems As much as possible avoid areas prone to sudden temperature changes Do not attempt to use the drive immediately after moving it from a cold to a warm location or raising the room temperature suddenly as condensation can form within the drive If the temperature changes suddenly while you are using the drive stop using it but keep the power on for at least an hour Handling Discs Handle discs only by the edges as illustrated below Never touch the recording surface
42. ns at the end of the writing procedure and you must change the disc in the CD R drive before performing steps 1 and 2 again If you are in the Multi session recording mode the tray does not open at the end of the writing procedure and you can continue saving images for the same patient or a different patient on the same CD R disc Accessing Images on a CD R Disc To access the images on a CD R disc created with the UPA P100MD proceed as explained below You must use a PC equipped with a MultiRead drive to be able to read this CD R disc A MultiRead MR drive is a drive with the ability to read all of the following types of media CD DA CD ROM CD R and CD RW 1 Insert the CD R disc into the CD ROM drive of your computer 2 On the desktop double click My Computer 3 In the My Computer window double click the CD ROM drive The following window opens amp Cdrom D olx File Edit View Go Favorites Help La z i M e 7 0 3 A Back Forward up dio Gat Copy Paste Undo D Address 9 D X size Tye Modified Ca link OOO File Folder 04 06 01 09 56 C link001 File Folder 04 06 01 10 00 2 objects LEJ My Computer Note If you created the CD R disc in the Single session recording mode the window above contains a list of the images on it 4 Double click the folder that contains the images you want to view 5 Double click the image that you want to view When you do
43. o a digital color printer Compatible printers are the UP D50 and UP D70XR Their SCSI default settings are Terminator on D 001 Refer to the appropriate printer user s guide for details RGB output D sub 15 pin This port is used for servicing purposes only S video output Mini DIN 4 pin Currently not supported 9 Video output RCA PIN jack Currently not supported Network port 10Base T 100Base TX RJ 45 This port is used to connect to a network with a LAN cable 11 USB port 2 series A 4 pin Currently not supported IEEE 1394 port 2 Currently not supported Speaker output Currently not supported 10 Parallel port D sub 25 pin This port is used to connect to a custom made external control unit to control the device Refer to the pin assignments in the Specifications and Troubleshooting chapter of this guide PS2 port 2 mini DIN 6 pin These ports are used for servicing purposes only Remote Control This section describes the RM P110 Remote Control Unit sold separately e c Fi F2 F3 F5 F6 F7 L3 N n P s l v D F4 E n FB 7 t 7 a 4 5 6 v w Y ab ABC 0 The LCD display of the remote control displays messages and input commands It can display 4 lines of 20 alphanumerical characters
44. onfiguration IP Address An IP address can be automatically assigned to this computer If your network does not automatically assign IP addresses ask your network administrator for an address and then type it in the space below Obtain an IP address automatically Specify an IP address 192 168 3 201 IP Address 1 ciick OK The TCP IP Properties dialog box closes 1 1 In the Network dialog box click OK 1 2 Restart your computer If you want to use the UPA P100MD through a network you must change its network settings Proceed to Connecting the Digital Capture Unit to a Network If you use the UPA P100MD locally you can configure the UPA P100MD immediately Proceed to Configuring the UPA P100MD 13 Connecting the Digital Capture Unit to a Network The Digital Capture Unit can be connected to a network but before you connect it to a network environment you must configure it appropriately Follow the procedure below to adjust these settings Refer to the next section The Configuration Screen for a more detailed description of the configuration screen s contents Note Please use Windows computers to configure and access the UPA P100MD Other operating environments are not supported by the UPA P100MD The procedures in this guide are based on Windows 98 dialog boxes and windows Screen appearance may differ slightly in other Windows versions 1 Using an RJ 45 cross
45. pture an image press 1 on the remote control keyboard 6 Repeat step 5 as long as you want to capture images from your video device Once you have completed capturing images you can print them see Printing Images save them on CD R disc see Saving Images to CD R discs or finish the operation by pressing 0 on the keyboard of the remote control Pressing 0 returns the display to Enter Patients Name so you can begin a new operation immediately To preview the images that you have captured and saved on the Digital Capture Unit hard disk refer to Viewing Images Viewing Images To be able to see the images that you have captured and saved on the unit s hard disk proceed as outlined below This can be done from a computer connected to the unit directly or through a network 1 Start your Internet browser 2 In the Address bar type the following f using the UPA P100MD through a network http network IP address For example Address http 192 168 159 14 zj f the UPA P100MD is connected directly to your computer http 192 168 3 200 For example Address o 191083209 mg The following screen appears Help 5 9 d aud n 3tm Address bp 7192 168 3 200 z eGo Sony UPA P100MD Images Configuration Manual PDF e Manual HTML a fe dee internet 7 3 Click Images The following screen appears A Son
46. uble click an image depending on the File Types settings in the Folder Options of Windows Explorer a program starts automatically and opens the image 23 Finishing Examinations To avoid problems mistakes or lost files it is very important to conclude a patient s examination properly Therefore when you are done capturing printing or saving files on a CD R disc for a patient it is vital that you press 0 on the remote control keyboard This way the unit is ready for a new patient Managing UPA P100MD Hard Disk Space Because the UPA P100MD has a limited hard disk capacity it is necessary to manage the images on it carefully When the hard disk becomes full you cannot capture any further images The RM P110 Remote Control Unit displays the following messages serve to help you manage the UPA P100MD hard disk space These are Hard Disk 8046 used See user manual Hard Disk FULL See user manual When the first of the two messages above appears you should back up the oldest images on the UPA P100MD hard disk Follow one of the procedures below to clear space on the UPA P100MD hard disk for new images Saving Images to Your Computer 1 Follow the procedure outlined in Viewing Images to access the images of specific doctors and patients on the UPA P100MD 2 Click the thumbnail of the first image that you want to save The image appears on the right Be sure to select the oldest images on the hard
47. uld damage the unit and or the media It is also important to make sure that the CD R discs that you use support a recording speed of 12X Some inexpensive brands of CD R discs do not support 12X recording T Insert a blank CD R disc in the UPA P100MD CD R drive Alternately you can use a CD R that was recorded in the UPA PIOOMD Multi session recording mode You cannot use a CD R disc that was previously recorded in the UPA P100MD Single session mode 2 Proceed with the capture of images as outlined in Grabbing Images 3 When you are ready to save the images captured during the current examination on a CD R disc press 3 on the keyboard of the remote control All the images captured from the beginning of the current examination up to the present are saved on the CD R disc When the operation finishes Enter Patients name appears on the remote control display and the unit is ready to begin a new operation Note If you forget to insert a CD R disc in the CD R drive before pressing 3 on the remote control the warning message An error occurred writing CD appears on the remote control display Open the tray insert a CD R disc in the tray and then close the tray 22 While writing images on a CD R disc the unit cannot be used You must wait until the message Burning a CD disappears and the remote control display returns to the function screen If you are in the Single session recording mode the tray ope
48. work IP address f the UPA P100MD is connected directly to your computer http 192 168 3 200 This is the default address It may change when you change the UPA P100MD configuration The Sony UPA P100MD screen opens 4 Click Configuration 5 Configure the UPA P100MD referring to the next section The Configuration Screen 15 The Configuration Screen The UPA P100MD configuration screen contains settings that control how the unit operates These settings are explained below The procedure for accessing the configuration screen is explained in Configuring the UPA P100MD 44 SONY UPA P100MD configuration Microsoft Internet Explorer lej xj Ele Edi View Favorites Tools Help Ea 04 4 8 x 3 B 3 Back Forward Stop Refresh Home Search Favorites History Mail Print Address http 192168 159 143 cai bin loadconfig cgi Go Links al Video NTSC C PAL Zolor Color Mono Remote Control Unit Support IV D Recor mv C Single session Multi session Video Printer Support C 2X2 2X3 IP 4 192 168 159 143 Netmask 255 255 255 0 Gateway 192 168 159 80 Dor e Always ask C Never ask Use default Default Nz doctor Pro Name Always ask C Never ask C Use default Defaul m procedure Time Zone Americe New York E Date Time yyyyimmidd om Aoz 43 8 4 Set Date r Language English x Delay Capture Delay value in milliseconds 250 Debounce value in millisecond
49. y OnSite Mi ft Internet Expl lolx File Edi View Favorites Jools Help gt OA aas di Ow Address hup 192 168 3 200 cobin user coi z Go fel C D Internet 20 4 In the Choose a doctor screen click the name of the doctor whose images you want to view The following screen appears Sony OnSite Microsoft Internet Explorer x Bile Edt View Favorites Tools Help ES e 2 Qid amd 5 3m Address 27 nip 7182 168 3 200 cgibin user cai DoctoredoctortPalientetlmage t Session SNlonl eUSE dg z eG B5 Sony Choose a patient patient 001 H 2 Done ERO Inemet z 5 In the Choose a patient screen click the name of the patient whose images you want to view The following screen appears Sony OnSite Microsoft Internet Explorer oj x Be Edt View Favortes Tools He J e gt 5 O AOE G S S m Address hitp 7 192 168 3 200 c9 bin user cgi D actor doctortPatientepatient 001kImage tSession SNlonl eUSEdg I eco Select an image on the left dee nenet 4 Sample images 6 Thumbnails of the patient images appear on the left Click the desired thumbnail and the image appears on the right as illustrated below Sony OnSite Microsoft Internet Explorer loj x Be Edt View Favores Tools Hep gt 5 Ob gO dn 3m Addtess 27 hntp 77192 168 3 200 eg bin user cg 2D a