1. If changing sampler frequency outside the 1GSPS to 2GSPS range it will be necessary to readjust potentiometer SVG3 as shown in Figure 2 An optional pulser clock input is fed into the SMA jack labeled PULSER CLOCK IN and may be set to any frequency from 1 MHz to 3GHz It should ideally be a 1V p p square wave with rise time under 150pS and duty cycle of approximately 50 a 6dBm to 12dBm sine wave input may be substituted for the square wave for frequencies above 200MHz Each PCLK transition causes the board to produce a pulse on the sampler pulser node which then appears at the 2 4mm sampler pulser TDR input output connector where it can be used to stimulate an external device under test DUT while the board s sampler can concurrently view this pulse and its reflections as well as any additional signal emitted from the external DUT When the pulser is not used the board s negative power supply should be reduced to 5V from its default value of 7V so as not to couple in any noise that is generated by the pulser in the absence of a valid pulser clock FURAXA SAMPLER SPECIFICATIONS Sample rate 1MSPS to 3000MSPS SVG3 must be adjusted for rates significantly different than 1GSPS Sampler clock input AC coupled sine wave of amplitude 12dBm Rise and fall times must be lt 150ps Sampler clock input waveshape Sine wave of 200MHz to 3000MHz or Squarewave with frequency of 1MHz to 3GHz and risetime 10 to 80 of 150ps or less Duty
2. cycle may be 40 60 to 60 40 50 50 preferred Number of channels 1 Full Scale Input range 200mV DO NOT EXCEED 300mV or unit will be damaged 3dB Input Bandwidth 50 GHz for Sampulse50GHz model 70GHz for Sampulse70GHzZAnt model 6dB Input Bandwidth 70 GHz for Sampulse50Ghz model Usable Bandwidth while maintaining 50dB SNR 70 GHz Dynamic Range 85dB typical 2MHz to 50 GHz Does not include coherent Fsampie Noise at output SFDR Approximately 30dB at maximum input level SVG3 and SVG2 may be adjusted to optimize SFDR Input connector type 2 4MM Jack for Sampulse50GHz model antenna for Sampulse70GHzAnt model Clock input connector type SMA Jack Output connector type SMA Jack Output IF signal 1V peak into 50 ohms can produce up to 2 5V when sampler input is fed with strong signal Please note that some external circuitry may be damaged by these voltage levels IF output bandwidth DC to 100MHz PULSER SPECIFICATIONS Pulser clock inputs AC coupled amplitude 500mV p p to 1200mV p p Pulser clock input waveshape Sine wave of 200MHz to 3000MHz or Squarewave with frequency of 1MHz to 3GHz and risetime 10 to 80 of 150ps or less Duty cycle may be 40 60 to 60 40 50 50 preferred Pulse repetition rate 1MPPS to 2000MPPS clock sources under 200MHz must be square wave with lt 150ps rise time slower rise times will result in broader or nonexistent output pulse and or added jitter Pulse amplitude Adjustabl
3. FURAXA SAMPULSES50GHz and SAMPULSE 0GHZANT PULSER SAMPLER DEMO BOARD REV B PRELIMINARY USER MANUAL 2 25 15 Samp 3 Adj 7 Fal Sere ex Samp z S Output Figure 1 Connections to SAMPULSE50GHz left or SAMPULSE70GHZANT right INTRODUCTION The SAMPULSE50GHZz and SAMPULSE70GHZANT are demonstration boards for Furaxa InP single channel sampler pulser ICs The SAMPULSE50GHZz model contains a 2 4mm combined TDR input output connector to the common sampler pulser node in the IC The SAMPULSE70GHZANT is similar except that instead of a single ended 2 4mm connector the sampler IC s differential inputs are fed from a bowtie antenna In both models the IC s differential sampler output is converted using an AD8009 op amp into a single ended ground referenced output that is sent to an SMA jack labeled SAMPLER OUTPUT The sampler clock input is fed into the SMA jack labeled SAMPLER CLOCK and should ideally be a square wave with rise time under 150pS and amplitude of approximately 1V to 1 5V p p It may be set to any frequency from 1 MHz to 83GHz a 12dBm sine wave input with approximately a 50 duty cycle may be substituted for the square wave for frequencies above 200MHz causing the board take samples at rates from 1 million to 3 billion per second Amplitude of either sine or square wave sampler clock input should be approximately 12dBm 850mV RMS The module is calibrated at the factory using a 1GHz sampler clock frequency
4. SVG3 and possibly the SVG2 potentiometer slightly for optimal linearity and bandwidth or even for any sampler operation at all Figure 3 TDR using sampler and pulser concurrently Sampler output displayed shows heterodyne of train of internally generated pulses spaced 1ns apart 990 MPPS generated by on chip pulser Sampler is run at 989 98 MSPS resulting in 20 000 waveforms second 50 microseconds of scope display corresponds to 1ns of real time Smaller pulse to right of each main pulse is reflection from open end of 2 4mm connector The sampler can be used without the internal pulser or can be used together to perform TDR measurements or to provide fiducial marks on the sampled signal For example in Figure 4 a sampler clock of 989 98MHz is used and the pulser clock is 990MHz The sampler output shows a heterodyne of the pulses spaced 1ns apart 990 MPPS generated by on chip pulser Additionally a 19 972GHz external signal is transmitted through air to the attached bowtie antenna on A Sampulse70GHZzAnt model it may similarly be instead inputted via the 2 4GHz connector on a Sampulse50GHz The sampler therefore runs at 989 98 MSPS 20KHz lower than the pulser frequency of 990MHzZ resulting in 20 000 waveforms second 50 microseconds of scope display FURAXA corresponds to 1ns of real time As can be seen in Figure 4 approximately 20 cycles of the 19 972GHz externally inputted sine wave are displayed between each of the 1ns
5. e from approximately 30mV to 300mV Pulser may be disabled by ensuring that PCLK is not driven and reducing negative power supply to 5V to prevent residual noise Pulse width at 2 4mm input output connector Approximately 12ps viewed on HP 54750 54752A scope approx 7ps internal to IC when optimally adjusted using pot PVG3 and viewed directly by sampler on IC Demo Board Power Requirements 4 6V may use 5V at 80mA typical 5V at 150mA typ when using sampler only no pulser Negative supply must be set to 7V and valid pulser clock applied if using pulser Cooling Requirements Light Fan cooling recommended 20LFPM minimum SETUP DIRECTIONS 1 Verify that voltages on external power supply are set to 4 6V and 5V 7V if using pulser Turn off external power supply and connect both Orange and Red power input cable wires to grounds common on external power supply Connect 4 6 to 5V output to 5V wire on power cable Connect 5V or 7V output to wire labeled 5V 7v if pulser If not using pulser then this wire should instead be fed with 5V 2 Using a high quality flexible 50 GHz microwave cable with 2 4MM connectors connect the 2 4mm jack to the desired RF signal source to be sampled or studied using the TDR capabilities of the Sampulse50GHz Board is fragile do not apply excessive torque when tightening cables on connectors on the SAMPULSE50GHz a light finger tight connection is adequate on all cable connectio
6. just the SVG3 potentiometer and possibly SVG2 slightly for optimal linearity and bandwidth For sampling rates significantly below 1GSPS it may be necessary to adjust SVG3 for the sampler to function OME 3 we r 4 _ A Figure 2 Adjusting sampling aperture for optimal bandwidth using SVG3 potentiometer Necessary when using SCLK frequencies significantly different than 1GHz FURAXA USING THE SAMPLER The IC that generates the pulse inside the SAMPULSE50GHz has a sampler that can take 7ps snapshots of the internally generated pulse and its reflections from the device being driven as well as any external signals fed into the 2 4mm jack To use the sampler connect to the Sampler Clock Input SCLK a continuous square wave having a frequency that differs from that of the continuous pulser clock PCLK input by approximately 20KHz or so The sampler output may then be viewed by an A D converter or on a low speed oscilloscope or ADC board with approximately 20 MHz bandwidth limiting filter The waveform so displayed will be the heterodyned downconverted picture of the pulse and its reflections as shown in figure 3 Note that in this figure the sampler output shows the downconverted pulse and its reflection from the unconnected 2 4mm connector IMPORTANT The sampler has been adjusted at the factory for optimal operation at 1000 MSPS When using different sampling clock frequencies input to SCLK you may need to adjust the
7. ns 2 4MM CONNECTOR MUST NEVER BE CONNECTED TO ANY INPUT THAT CONTAINS A VOLTAGE OF OVER 300mV OR THE UNIT MAY BE DESTROYED A FAILURE NOT COVERED BY WARRANTY 3 Using flexible cables connect a square wave clock source of 500 1200mV p p with rise time lt 150ps to the SCLK and optionally PCLK connectors For initial testing as a TDR use a 12dBm sine wave or square wave with a frequency of 1000 MHz for PCLK and a frequency of 999 98MHz for SCLK As the output pulses are narrow 7ps and may be difficult to find it is most convenient to use as high a clock frequency as possible for initial testing If using SCLK frequencies significantly different than 1GHz it will be necessary to readjust SVG3 and possibly SVG2 as shown in Figure 2 Similarly if using PCLK frequencies significantly different than 1GHZz it will be necessary to readjust PVG3 4 Connect the SMA SAMPLER OUTPUT signal to an oscilloscope set for 10us per division 50 ohm input impedance and a vertical scale of 500mV or 1V per division 5 Turn on the power supply and clock sources 6 If you are using both the sampler and pulser you should see a pulse with some ringing on the oscilloscope as shown in figure 4 Please refer to section USING THE SAMPLER for more details IMPORTANT The samplers have been adjusted at the factory for optimal operation at approximately 1GHz When using significantly different sampling clock SCLK frequencies you may need to ad
8. separated pulses internally generated by the sampler pulser IC Figure 4 Sampler output showing heterodyne of pulse train of pulses spaced 1ns apart 990 MPPS generated by on chip pulser Additionally a superimposed 19 972GHz signal is transmitted through air to the attached bowtie antenna on A Sampulse70GHzAnt model Sampler is run at 989 98 MSPS resulting in 20 000 waveforms second 50 microseconds of scope display corresponds to 1ns of real time KNOWN BUGS 1 DC control voltages SVG3 and sometimes SVG2 need to be readjusted when changing sampling frequency outside the 1GSPS 2GSPS frequency range 2 DC control voltages PVG3 and sometimes PVG2 need to be readjusted when changing pulse repetition frequency from factory calibrated 1GPPS frequency