GM550 User`s Manual(AST+mmol)-P1~49-EN(101
Contents
1. Setting Up Your Meter Setting the Time Volume and Date 4 Time format 12 24H selection ta 4 l he N T 4 With the time format blinking press the Left or Right button _ a a Ss to adjust it Then press the Main button to confirm it and L I J n l p move to hour setting i 5 Hour setting e g laan eo ann os wu Ei With the hour blinking press the Left or Right button until the current hour appears Then press the Main buttonto 5y a confirm it and move to minute setting a Ua Ce S Oaa 6 Minute setting L Us i L With the minute blinking press the Left or Right button to adjust it Then press the Main button to confirm it Then it I er Z i will move to volume setting aan a 7 Volume setting 7a Sb lt N With the volume blinking press the Left or Right button to 04 E D 04 n D turn it on or off Then press the Main button to confirm and finish the setting L L L Ly L G n AEE un J ure 20 Setting Up Your Meter Setting the Time Volume and Date 8 Ending setting After volume settings for confirmation you will hear a beep if volume is turned on All the settings are saved and completed and will return to time screen If Volume is turned off the display will directly return to time screen without a sound of beep Day Month Z 0 armenn I AI 74 NOTE When you do not do any settings of meter for over 2 m
2. y Y Meter Battery Backlight Battery Rightest GM550 Meter 10 mem No AVG CS OK mmol L OO UO mag dt Erec Key Code Error Day Month S FM TITI CITI nooo Cin G Rightest GM550 Meter Mem Indicates a test result stored in memory Tell when to apply the blood sample Indicates a test result not joining into the Average T l NO AVG function of this meter C ee Warns when using the strip not suitable for your AVG ndicates with the average result Code Error region test or strip error Day Month i l cs idieates contol sclunondcetiesult rn A nr Current date under time mode or Testing date aay under memory mode mmol L Indicates if the environmental temperature is mg dL Unit of test result c exceeded during testing nAg Test result Rae Indicates the time in 12H format Warns when the battery is Cee Current time under time mode or Testing time 4 e T NSA Mele eee LILO LILI under memory mode OK EN Check Key Manufacturing use only Lil wale 11 Precaution Before using the Rightest GM550 BGMS to test your blood glucose please read instructions for USE The Rightest GM550 Meter is only compatible with the Rightest GS550 Test Strip The Rightest GM550 Meter must only be used with Rightest GS550 Test Strip and
3. 3 Non Averaging set and canceled You may select the unwanted or doubtful test results to exclude them from the average calculation after your every test After you finished your test in the test mode you can exclude the unwanted result Keep pressing the right butten until you see the symbol NO AVG in the top right cornetr of the screen By keeping the right button pressed and pressing the main button simultaneously to confirm the value as Non average Now this value is flagged by the symbol NO AVG in the up per right corner of the screen If you want to cancel this non averaging setting you have to stay in the test mode First keep pressing the right button until the NO AVG symbol changes to AVG in the up per right corner of the screen To confirm this original status you again simultaneously press the main button to confirm the AVG status Now the value will be stored as a regular value and included in the average values NOTE A The average function is related to the time setting You must set the time right and have enough time intervals on the basis of current inquiry time to make sure that the average test results will show Use the 38 Recalling Test Result and Average 14 day average for example if your current inquiry time is 2008 1 30 then that you must check that you have tested blood glucose in the past 14 days before 1 30 including today If not the 14 day average will show no
4. GM550 BGMS is intended in vitro diagnostic use only outside the body and coding is auto detecting The testing result is calibrated to plasma equivalent with fresh capillary whole blood samples from the fingertip palm or forearm You may consult your healthcare professional for instructions on how to use the system correctly Our customer support staff are also available to assist YOU Table of Contents About Rightest GM550 System Package of Rightest GM550 System Kt Z O O s ssaaaaa 6 Rightest GM550 Meter Oyo 8 Precaution 1 ec Meter activation and Battery change 16 Se ne ere A i a E A EP T E ae A O E E A a a E E ee ee ee ee ee eee Performing a Blood Test sass ae Alternative Site Testing AST 28 View window appearance 30 Table of Contents Quality Control Lo and Hi readings i Quality Control Test g2 Performing a Quality Control Test 34 Recalling Test Result and Average SF Error Messages and Trouble Shooting 40 Specification 44 LIMIEAIOMS sums Customer Service 47 Warranty 48 5 Package of Rightest GM550 System Kit Your Rightest Blood Glucose Monitoring System GM550 consists of several items Please identify each item of your system and learn what they re called and how they re used These items are included in your Rightest Blood Glucose Monitoring System GM550 1 Rightest GS550 Test Strips 0 10 25 pcs Getting Started Guide Rightest GM550 User Manual
5. Rightest GS550 Test Strips Package Insert Rightest GD500 Lancing Device x Clear Cap Disposable Sterile Lancets 10 pcs Instructions for the lancing device Rightest Control Solution 11 CON DO BW PP 12 x Rightest GM550 Meter with 2 CR2032 batteries installed Rightest Control Solution Package Insert not shown Warranty Card not shown 13 Carrying Case not shown Different packages have different bundled items Some of packages might not include items Package of Rightest GM550 System Kit Rightest GM550 GETTING STARTED GUIDE For setestna Instructions for the Righte Rightest GM550 Meter Test Strip Port Insert test strip here until you hear a a Display Window lt 4 gt Show your test result messages and relative information Main Button Q For operating the meter please check details in following sections Left Button S For operating the meter please check details in following sections Right Button For operating the meter please check details in following sections a N BIONIKNGE Rightest cmsso AS E A amp 110 HBA 0021 Battery Cover Slide cover to change batteries Product name Meter Serial Number Rightest GM550 Meter
6. Alternative site testing palm or forearm blood sampling PRECAUTION Choose Clear Cap and follow step 1 5 on page 24 3a 2 Massage the puncture area of palm or forearm for a few seconds 3 Immediately after massaging the puncture area press and D The glucose test results of blood samples taken from different site might be varied in certain conditions such as your glucose is changing rapidly following a drink a meal insulin dose or exercise In these cases only fingertip should be used DO NOT test on the palm or forearm if you are testing for hypoglycemia Low blood lucose hold the lancing device with the Clear Cap against palm or J Ei j j i Fingertip samples can show the rapid change of glucose faster than palm or forearm 5 samples As the blood flow taken from forearm is slower than fingertip or palm we recommend using 5 Continue holding the lancing device against palm or the Rightest GD500 lancing device with Clear Cap for testing sites other than fingertip forearm and gradually increase pressure for a few seconds until the blood sample size is sufficient 6 Follow step 9 18 on page 25 27 to complete the test and discard the used disposable lancet Pressing for afew seconds Releasing button 4 4 Then press the release button 28 29 View Window appearance Lo and Hi readings The Rightest GM550 Mete
7. on screen in this condition Please move the meter to environment with temperature between 10 40 C 50 104 F and repeat the test after 30 minutes 2 Wait for 5 minutes and insert the batteries as described Meter activation and Battery change on page 16 17 The meter should be working normally after finishing above steps If not please contact your local Bionime customer service 42 43 Specification Measurement Technology Measurement Calibration sample Minimum Sample Volume Measuring Range Test Time Memory Capacity Power Saving Operating Temperature Operating Relative Humidity 44 Oxidase Electrochemical Sensor Plasma Capillary whole blood 1 0 microliters 0 6 33 3 mmol L 5 seconds 500 blood glucose test results with date and time Turn off automatically 2 minutes after last user action Press the button for 4 seconds 10 40 C 50 104 F 10 90 Hematocrit Power Supply Meter Battery Life Meter Dimension Meter Weight Monitor Display Area Meter Storage Conditions Test Strip Storage Conditions Specification 30 60 2 CR2032 batteries About 1000 tests 90 6 mm x 46 mm x 16 5 mm 53 0 5g with batteries LCD display 47 mm x 33 5 mm 10 60 C 14 140 F 4 30 C 39 86 F lt 90 relative humidity 45 Limitations Rightest GM550 BGMS is not intended for serum or plasma test Inaccurate test results may be obtained at high
8. Battery inserted Press the main button first to turn on the meter Then press and hold the main button for 7 seconds the meter will first show backlight then be turned off during this period please keep pressing until you hear a beep indicating you have successfully entered Setting Mode The display screen will show setting data NOTE J When you keep pressing the main button for 2 seconds the display on screen will show backlight Keep pressing the main button till display is off When you keep pressing the main button for 4 seconds the display on screen will turn off Please don t care it Keep pressing the main button till entering setting data Quick press the main button allows you to change setting However if you want to return to the time mode you need to go through all the settings first by pressing quickly the main button 18 Setting Up Your Meter Setting the Time Volume and Date 1 Year setting With the year format blinking press the Left or Right button to adjust it Then press the Main button to confirm it Then it will move to Month setting 2 Month setting With the month blinking press the Left or Right button until the current month appears Then press the Main button to confirm it and move to Day setting 3 Day setting With the day blinking press the Left or Right button until the current day appears Then press the Main button to confirm it and move to Time format setting 1 2 3
9. altitude more than about 3048 meters 10 000 feet above sea level Severe dehydration and excessive water loss may cause inaccurately low results Rightest GM550 BGMS has not been validated for use on neonates The glucose test may be interfered under abnormal concentration of Uric acid gt 0 96 mmol L NOTE Suggest not to use this meter close to source of strong electromagnetic radiation to avoid interference with proper operation Suggest to keep meter free of dust water or any liquid 46 Customer Service We sincerely like to provide complete considerate services to our customers Please review all the instructions to make sure you are performing the steps correctly If you have any questions or in case of problems with the Rightest GM550 products please contact your local Bionime customer service Description of used symbols For in vitro diagnostic use wa Manufacturer Lot number 2 Expiry date Temperature limitation Q For single use only Method of sterilization using irradiation C CE mark with No of notified body EC Representive pH Consult the instruction for use 47 Warranty The manufacturer warrants that your Rightest GM550 Meter will be free from defects in materials and workmanship for five years from the date of purchase This warranty does not apply to the performance of a Rightest GM550 Meter that has been altered misused tampered with or abused in any way This warrant
10. appears between the cap and hub pull them off in opposite directions 8 Wash your hands with warm soapy water and dry thoroughly 9 Take one strip from the vial Re cap the vial cap immediately 10 Insert the strip into the strip port of the meter with the indication symbol facing up The meter confirms the insertion of the strip with a beep if volume is turned on 2 Pull off the depth adjustable cap 3 Insert a new disposable lancet firmly into lancet carrier 4 Twist off and set aside the protective cover of the NOTE disposable lancet GY The meter will automatically detect Code number on strip You don t have to check the Code number on the 5 Replace the depth adjustable cap meter display and strip vial bNinne _ GMs50 24 25 Performing a Blood Test Performing a Blood Test 11 A blood drop symbol appears and is accompanied by a further beep if volume is turned on Now you can apply the blood sample within 2 minutes 15 Pull off the depth adjustable cap Without touching the used disposable lancet stick the lancet tip into the protective cover 16 Holding the release button in one hand see picture 16 step 1 and pulling on the plunger in the other hand see picture 16 step 2 will safely eject the used disposable lancet 17 Discard the used disposable lancet into an appropriate opuncture proof or biohazard container 18 R
11. figure The Non averaging re averaging function only works in the test mode and when you get the value In the MEM mode you can not perform this action 4 Quick Searching If you want to see all values automatically displayed in sequence first enter the memory mode Then keep pressing the right or left button for 2 seconds right button is for searching the tests the latest value to the oldest one by using the left button you can scroll from the oldest to the latest one Anytime you want to stop just release the pressed button It will stop and display the current displayed value You can use this way to find a particular reading you want to check 5 Reading right after test If you just finish the test press the main button to enter the MEM memory screen and review the latest test result NOTE Y Your Control Solution results will be automatically non averaged without selecting but still can be recalled The Control Solution Test results will show with CS symbol under memory screen PRECAUTION You have to set the time and date to activate the average function The Hi results the Control Solution results and the test result made out of normal temperature range lt 10 C gt 40 C are not calculated in the average 39 Error messages and Trouble shooting Error messages and Trouble shooting Eri he inserted test strip has been used or damaged Please use a r r r
12. BIONINNE Blood Glucose Monitoring System gal BIONIME CORPORATION 4 694 Renhua Road Dali Dist Taichung City 412 Taiwan SONINN GM550 ad Tel 886 4 24951268 Fax 886 4 24952568 http www bionime com E mail info bionime com BIONIME GmbH Tramstrasse 16 9442 Berneck Switzerland swisscesigng 5197 USER S MANUAL A N 101 3GM550 0K1 EN Rightest cms550 Preface Thank you for choosing the Rightest Blood Glucose Monitoring System GM550 Please read this manual thoroughly before testing It provides all information you need to use the product The Rightest Blood Glucose Meter GM550 must only be used with Rightest Blood Glucose Test Strip GS550 and Control Solution The application of other Test Strips or control solutions can lead to incorrect results We hereafter call Rightest Blood Glucose Monitoring System GM550 as Rightest GM550 BGMS Rightest Blood Glucose Meter GM550 as Rightest GM550 Meter and Rightest Blood Glucose Test Strip GS550 as Rightest GS550 Test Strip for short It is recommended to monitor blood glucose regularly Rightest GM550 BGMS is accurate and easy to use and is thus your reliable assitant for diabetes management Rightest GM550 BGMS is manufactured and supported by Bionime Corporation If you have any question or concern please contact your local Bionime Customer Service or call us at 886 4 24951268 GMT 08 00 08 30 17 30 Intended Use Rightest
13. Battery Error new test strip from vial cry creo 1 The a symbol is blinking when the battery power is low Please change the m meter battery as soon as you can You can still do the test Er2 Meter has malfunctioned Do the Quality Control Test or reinstall L J Eas the batteries to check if the meter works properly 2 The Ca and Error symbols are blinking when the battery is too low Meter can not do the strip test Please change the meter battery immediately Er3 Signal transmission is disrupted repeat the test z Ce 2 Cu Li Jj CI Er4 Applied blood volume is insufficient please repeat the test with a gt new test strip L J Above if error screen still appears contact your local Bionime Customer Service NOTE When meter backlight has no function you cannot see the backilght please change your backlight battery When the backlight battery is runnign out of power your meter still works properly but without backlight function 40 Error 41 Error messages and Trouble shooting Error messages and Trouble shooting Temperature Error Sampling Error In order to get accurate test result perform testing between 10 40 C 50 104 F Please don t apply the blood drop to the sample entry of the strip before the meter displays If you do so the meter will display Error and accompanied by beeps if volume is turned on Please discard the test strip and repeat the test wi
14. Rightest Control Solution The application of other Test Strips or control solutions can lead to incorrect results Rightest GM550 BGMS is intended for self testing It should not be used to diagnose diabetes mellitus Rightest GM550 BGMS has not been validated for use on neonates Therefore it s not intended for use of neonates Rightest GM550 BGMS is not intended for arterial blood testing Do the test at least 30 minutes after moving into a different location with significant change in temperature 12 Precaution Dispose of used batteries properly Please note the System kit contains small parts like test strios which could result in a choking hazard for children Prevent water from entering the meter Never immerse the meter or hold it under running water The minimum blood sample size for testing is 1 0UL e 1 0uL 1 5uL 2 0uL FOUL 4 0uL Sample Size Example P p o Please take a minimum of 1 0uL to do the test on blood glucose monitoring system Blood sample size above 3 0uL might contaminate the meter A blood sample size below 1 0uL may give an inaccurate test result In this case repeat the test with a new test strip PRECAUTION The Rightest Blood Glucose Meter GM550 will be set to mmol L by default when sold in your country When your display shows mg dL during setting or test please contact customer service Use of the wrong unit of measure may cause you to incorrect
15. ain button for over 3 seconds until the CS symbol appears 4 You will see blinking w symbol and CS symbol symbol on the screen prompting you to apply Control Solution 5 Shake the bottle of Control Solution well before opening the cap Then open the bottle and put the cap on the table 6 Drip a drop of Control Solution on the top of the cap 7 Gently touch sample entry of the strip with the Control Solution on the top of the cap 8 When you hear a beep if volume is turned on please wait for the test result The screen will display the countdown number from 5 34 Performing a Quality Control Test 9 Tightly replace the cap on the Control Solution bottle 10 The Control Solution result appears Compare your Quality Control Test result to the Control Solution Range printed on the test strip vial label PRECAUTION Your Control Solution results will not be calculated for average reading but still can be recalled The Control Solution Test result will be shown with CS symbol on the screen Our suggestion environment for Control Solution Test is 15 40 C Before w and CS appears please don t touch the Control Solution to the sample entry on strip because the meter is still in an internal check If you do so the meter will show jw and Error Accompanied by beeps if volume is turned on Don t drip the Control Solution t
16. at 15 2513 BH The Hague The Netherlands 49
17. e to the Memory screen First mem mon you will see the MEM symbol on the up per left corner of the display When re you press the right button the latest result will be displayed By pressing the CE E right button sequentially former test results will be shown in historical order You will see the sequence number on the lower right corner and the year on the lower left corner of the display followed by date and time of the measurement T Ii I ER LI Note The right button is for reviewing the tests with sequence number increasing and the left button is for reviewing tests with sequence number decreasing The sequence no 1 is the latest result while sequence no 500 is the oldest test result 2 To finish reviewing memory tests press main button again and you will enter the Average screen By pressing the right button the display shows AVG on the upper right corner of the display and the average value for the actual day By sequentially pressing the right or left button you will get to the 1 day 7 days 14 days 30 days and 90 days average of your blood glucose values You will see the number of the calculated days on the lower left corner and the number of the calculated readings on the lower right corner 37 Recalling Test Result and Average AVG f aG AVG AVG AVG apne gure ane gl mo o Lit Lu Lu L aS LI Lh LI aS l aS y L I I 8 7 8 4B wj g X J
18. eplace the depth adjustable cap after finishing the test PRECAUTION Do not apply your blood drop to the sample entry on the strip until you see the appear The meter is performing an internal test and will display and Error if you apply blood too soon Please repeat the test with a new test strip Record the date of opening a new test strip vial for the first time Discard the vial of test strips after 3 months from opening Always keep the metal contact points of the test strip entry point clean If any dust or impurities are present please clean with a small soft brush otherwise the meter may not work correctly when you insert a test strip 12 Place the lancing device against your fingertip and press the release button 13 Touch and hold the drop to the edge of sample entry until you hear a beep if volume is turned on and the View Window is totally filled with blood If the View Window is not totally filled with blood or the test does not start please discard the test strip and repeat the test with a new test strip 14 You will see the countdown mode on the screen After 5 seconds the test result appears Test Strips Discard 4 months after first opening Discard date __2009 10 01 Exp Date Control Solution Range E F F e in 26 27 Alternative Site Testing AST Alternative Site Testing AST
19. inutes the meter will leave setting mode and power off automatically 21 Turning On Off the Meter 1 How to turn on the Meter Press the Main button or Insert a test strip 2 Manual Power off If you want to turn off the meter please keep pressing the Main button for 4 seconds 3 Auto Power off The meter will power off automatically if you don t operate it over 2 minutes 22 Turning on off the Backlight 1 How to turn on the Backlight Keep pressing the Main button for 2 seconds 2 Manual Backlight off If you want to turn off the backlight please keep pressing the Main button for another 2 seconds 3 Auto Backlight off The meter backlight will turn off automatically if you don t operate it over 10 seconds 23 Performing a Blood Test Performing a Blood Test 6 Choose a depth of penetration by rotating the top portion of the depth adjustable cap until the setting depth matches the window Settings are based on skin type ar gt for soft or thin skin anD for average skin amm for thick or calloused skin Depth adjustable cap Release button 7 Hold the hub in one hand and pull on the plunger in the other hand The device will be cocked Release the plunger it will automatically move back to its original position near the hub Hold the depth adjustable cap in one hand and hold the hub in the other hand Bend the cap towards the down side When a gap
20. ion and Battery change Meter activation and Battery Change Your meter comes with two CR2032 3volt batteries installed Two new batteries will provide power to perform about 1000 tests under normal use You can press the main button or insert a strip to activate your meter 1 2 Install the batteries Be sure to 3 Slide the battery cover and push battery cover to put batteries in correct back until it snaps into open direction place 4 The meter performs a self test 5 Press any button to exit the self test and enter Setting Mode L NOTE 6 Set the time and date when the batteries are replaced See Chapter Setting up your meter When meter backlight has no function you cannot see the backlight please change your setting the time volume and date Test results are still stored in the memory backlight battery When the backlight battery is running out of power your meter still works properly but PRECAUTION without backlight function Please follow the local regulation and discard a used battery properly 16 17 Setting Up Your Meter Setting the Time Volume and Date You can enter Setting Mode by below two ways 1 Reload battery After removing the battery please press the main button for several times until no signal on screen then follow the battery installation steps to load batteries The meter will do self test Press the main button to exit the self test and enter Setting Mode 2 With
21. o sample entry of the strip directly The reagent on strip might be sucked into the bottle of Control Solution and might cause the degeneration of Control Solution Doing this might contaminate the meter via the test strip port as well Don t touch the tip of the control solution bottle If you have touched it please clean up with water 35 Performing a Quality Control Test Maintenance Keep your meter and test strip free of dust water or any other liquid Store the meter in the carrying case if you do not use it If meter is dropped or damaged perform a Quality Control Test before doing a blood glucose test Cleaning Meter Clean the outside of the meter with a damp cloth and mild soap detergent Protect the test strip port from getting wet Please always keep the two metal contact points of the test strip port clean If you find it s covered by dust or impurities please clean it with a little soft brush The meter might not be activated when you insert a test strip but the contact points are heavily covered by dust or impurities 36 Recalling Test Result and Average The Rightest GM550 Meter is able to store 500 test results with time and date automatically If your meter has stored 500 results which is the maximum memory of the meter the newest test result will replace the oldest one To recall your test memory start the meter without test strip inserted 1 Press the main button to switch from the time mod
22. r displays results between 0 6 and 33 3 mmol L If your mmol L test result is below 0 6 mmol L Lo will appear on the screen Please repeat your f 0 test with by a new test strip If you still get Lo result you should immediately oo contact your healthcare professional O O If your test result is above the high end of the system s detection range ie 33 3 mmol L Hi will appear on the screen Please repeat your test with a new test Y Insufficient blood sample Enough blood sample strip If you still get Hi result you should immediately contact your healthcare Mi professional Make sure your blood sample covers the J whole area of the View Window to get an accurate test result An insufficient blood sample will result in an error message Er4 If this occurs repeat the test with a new test strip E I ER COD mM ZI a mmol L N PRECAUTION Check the expiration date printed on the strip vial every time you use a test strip Do not use a li at C Lu expired test strips Use each test strip immediately after removing from the vial L NOTE Do not reuse test strips If your blood glucose result is unusually high or low or if you question your testing results If the Rightest GM550 meter and strips are exposed to a high temperature difference please repeat the test with a new test strip You can also run a Quality Control Test to check your wait 30 minutes befo
23. re measurement Apply the blood drop only on the sample entry of the test strip Please don t drip or inject the blood sample directly by syringe to the sample entry of test strip Doing this might contaminate the meter or cause damages and is not recommended meter and test strip If the test result still remains unusually high or low contact your healthcare professional immediately If you are experiencing symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions of this manual contact your healthcare professional immediately 30 31 Quality Control Test Please use Control Solution tested with Rightest GM550 BGMS under Control Solution Mode If the test result is within the Control Solution Range printed on the test strip vial label the Rightest GM550 BGMS passes Quality Control Test That means your Rightest GM550 BGMS is working correctly Control Solution Range Control Solution Range 83 113 mg dL Example of Control Solution Range 4 6 6 3 mmol L printed on your test strip vial label When should a Quality Control Test be performed Whenever you want to check if your system is working properly or not Whenever you want to practice testing and check correct procedure To prepare for your initial blood glucose test To check the Rightest Blood Glucose Test Strip GS550 when you open a new vial of strips To check your Righte
24. st Blood Glucose Meter GM550 after it has been dropped damaged or exposed to liquids When you suspect that your test results are inaccurate or if your test results are not consistent with the way you feel 32 Quality Control Test The possible reasons your Control Solution results are out of the range Your Control Solution has expired or has been opened 3 months ago Your test strip has expired You left the cap of the tst strip vial opened or the control solution off for a long time You didn t perform the test procedure correctly Malfunction of the meter or the test strip If Control Solution results are out of range your Bionime BGMS may not be working properly Repeat the Quality Control Test If your Control Solution results outside the range still exist do not use the BGMS to test your blood glucose and contact your Bionime Customer service Example of Expiration Date Exp Date 2012 10 PRECAUTION Each time you open a new bottle of Control Solution write the discard date on the label Control Solution is good for 3 months after opening the bottle or until the expiration date printed on the label whichever comes first 33 Performing a Quality Control Test Take one test strip from vial and Re cap the vial cap immediately 2 Insert the test strip with view window facing up into test strip port 3 While the blood drop symbol flashing press and hold the m
25. th a new teststrip 1 When the ambient temperature is O 9 C 32 48 F or 41 50 C 106 122 F a the warning symbol will be blinking 1a you still can do the test but the test result is only for reference because the test result under these ranges of 1 Error S temperature might not be correct In the memory screen this value will be flagged Te Code Error with symbol 1b Repeat the test at an area with temperature between _ It means the strip is not inserted right or to the proper position Please reinsert the operating range 10 40 C or 50 104 F strip following the instruction If Code Error repeats again it means you use wrong ef Ibmem ow strip or strip not suitable for your region test After you make sure you use the right sean 2 When you move from the area with temperature outside the operating range of T C strip and the strip is inserted right but error screen still appears please contact test strip to another area with temperature inside operating range of test strip Lh l your local Bionime Customer Service lease wait for 30 minutes before you do the test ae p y an T i Meter Malfunction i please follow th low 3 When the temperature is below O C 32 F or over 50 C 122 F Meter can not pe Ber Neem OR ee iea eee onl tac 3 1 Open the battery cover remove the batteries do the test and the Error symbol will blink with thermograph symbol displaying
26. treatment 13 Rightest GS550 Test Strip The Rightest GM550 Meter must only be used with Rightest GS550 Test Strips and Rightest Control Solution The application of other Test Strips or control solutions can lead to incorrect results Sample Entry Apply a drop of blood or control solution here The test request is only 1 0 uL of blood Hand Bar Grip here to insert Noble Metal Electrodes Electrochemical sensor Indication Symbol View Window Insert strip with indication symbol This window is yellow before applying up and toward meter blood sample It gradually turns red when filled with blood 14 test strip into meter Rightest GS550 Test Strip DC ODC Electrode Contacts Sensing signal output terminals PRECAUTION Re cap the test strip vial immediately after taking out a test strip Do not reuse test strips Do not use expired test strips Record the date of opening a new test strip vial for the first time Discard the vial of test strips after 3 month from opening Store the test strips in a cool 4 30 C and dry location lt 90 relative humidity Do not expose to direct sunlight or heat For detailed information please refer to the Rightest GS550 Test Strips Package Insert If the Rightest GM550 meters and strips are exposed to a high temperature difference please wait 30 minutes before measurement 15 Meter activat
27. y applies only to the original purchaser of the meter Please complete and return the enclosed warranty card to your local Bionime affiliate If the Rightest GM550 Meter and strip are exposed to a high temperature difference please wait 30 minutes before measurement NOTE During blood glucose measurement the Rightest GM550 Meter itself may come into contact with blood Used Rightest GM550 Meter therefore carry a risk of infection Please dispose of your used Rightest GM550 Meter after removing the battery according to the regulations applicable in your country For information about correct disposal please contact your local council and authority 48 Main Parts Blood Glucose Meter Test Strip Control Solution Manufacturer Bionime Corporation 694 Renhua Road Dali Dist Taichung City 412 Taiwan Product complied with In Vitro Diagnostic Medical Device Directive 98 79 EC CE0197 EC Rep Bionime GmbH Tramstrasse 16 9442 Berneck Switzerland Lancing device Manufacturer Bionime Corporation 694 Renhua Road Dali Dist Taichung City 412 Taiwan Product complied with Medical Device Directive 93 42 EEC EC Rep Bionime GmbH Tramstrasse 16 9442 Berneck Switzerland Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LilangHe RD XiangCheng SuZhou JiangSu 215133 PR China Product complied with Medical Device Directive 93 42 EEC CE0197 EC Rep EMERGO EUROPE Molenstra