CDN 3083-B100&CDN 3083-B200 User Manual english.indd

image

Contents

1. zc 3 phase mains L10 T ED Eum to supply EUT input i209 7 unt e L30 Sass 1 gt 1004H Ferrites Ce 33 nF Reference ground Decoupling section Coupling section connector Diagramm of connection with 3 phase AC main supply Signal from test generator Nas DCto PR EUT input 1304 1 EN 1 gt 100 Ferrites Ce 33nF Reference ground Decoupling section Coupling section connector Diagramm of connection with DC supply CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks AC 1 phase main supply Signal from test generator I phase to EUT Reference ground T connector gt 100 uH Ferrites Cc 33 nF Decoupling section Coupling section Diagramm of connection with 1 phase AC main supply TASEO Advanced Test Solutions for EMC 13 14 6 OPERATION 6 1 EUT power supply The power is fed in via a 5 core cable usually brought out as Color 3 phase AC 1 phase DC brown L1 phase L positive black L2 black L3 blue N neutral N negative green yellow PE ground PE PE Note For 1 phase or DC application it is possible to use the input channels in parallel thus it is possible to increase the max EUT current prevent the risk of
2. CDN 3083 B100 amp CDN 3083 B200 EFT BURST COUPLING DECOUPLING NETWORKS USER MANUAL 601 336B Advanced Test Solutions for EMC CDN 3083 B100 amp CDN 3083 B200 EFT BURST COUPLING DECOUPLING NETWORKS USER MANUAL CONTENTS 1 Explanation of the symbols used in this manual 5 2 Safety advice 6 2 1 Safety measures 7 3 Introduction 8 4 Installation 10 4 1 Installation of an EUT power switch 10 4 2 Leakage current 11 4 3 EUT supply protection 11 5 Circuit diagram 12 6 Operation 14 6 1 EUT power supply 14 6 1 1 Earth connection 15 6 2 Test rig 15 6 3 Operation of the instruments 16 7 Accessories 17 7 1 Connectors for reduced current application INA 163 max 32 A optional 17 7 2 Pulse measurement adapter INA 3243 optional 17 7 3 Burst EFT Pulse Verification Kit CAS 3025 optional 18 7 4 Isolated allen Key INA 3005 included 18 7 5 EFT cable INA 3008 included 19 7 6 Single SHV plug INA 6546 optional 19 8 Maintenance 20 9 Warranty 21 10 Technical specifications 22 11 Ordering information 23 11 1 Accessories included 23 11 2 Options 23 12 Declaration of conformity CE 24 13 Addresses 26 1 EXPLANATION OF THE SYMBOLS USED IN THIS MANUAL Please take note of the following explanations of the symbols used in order to achieve the optimum benefit from this manual and to ensure safety during operation of the equipment The following symbol draws your attention to a circumstance where
3. B100_2012 LVD CDN3083 B100 2012 Teseq AG CH 4542 Luterbach The purpose of this instrument is the generation of defined interference signals for EMI immunity testing Depending on the arrangement of the test rig the configuration the cabling and the properties of the EUT itself a significant amount of electromagnetic radiation may result that could also affect other equipment and systems The user himself or herself is ultimately responsible for the correct and controlled operation of the rig In case of doubt the tests should be carried out in a Faraday cage European representative Place and Date Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland Luterbach March 19 2012 Johannes Schmid President CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks Declaration of conformity Manufacturer Address Product Options Generic standards Technical file 25 T3SEO Advanced Test Solutions for EMC Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland T 41 32 681 40 40 F 41 32 681 40 48 www teseq com Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland declares that the following product CDN 3083 B200 EFT Burst Coupling Decoupling Network all conforms to the following Directives and Regulations EMC Directive 2004 108 EEC LVD Directive 2006 95 EEC EN61326 1 2005 EN61326 2 1 2005 EN61010 1 2001 The relevant technical file is available for inspection N EM
4. Isolating support Connection to 10 cm ground reference plane Ground reference plane to protective ground Non conductive table Example of a system configuration with the NSG 3040 Please refer to IEC 61000 4 4 and respective product standards for the various test setups 6 3 Operation of the instruments Itis assumed that the test rig has been set up in accordance with the preceding notes and that the device to be tested has been connected taking the relevant safety measures into account the EUT power supply is switched off and the generator s EE Only handle the test rig or the device being tested when test sequence has been halted or aborted It is possible to overstress the CDN 3083 B100 amp B200 for a certain time However attention has to be given to the heat control placed on the CDN 3083 B100 or B200 The displayed temperature on the CDN 3083 B100 amp 200 HE must not exceed 70 However the temperature at the screwed terminals may be higher this as a result of contact resistance between cable and terminal CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 7 ACCESSORIES 17 7 1 Connectors for reduced current application INA 163 max 32 A optional In case an EUT needs to be exchanged frequently the optional safety banana plug set INA 163 gives the possibility to connect the EUT in an easy and safe way With this option the max current consumption of the EUT must no
5. electrical shock through unused chan nels screwed connectors It is recommended to screw the safety banana plug adapters INA 163 on the unused connectors ub Care has to be taken for 1 phase or DC application to The screwed terminals are made for a maximum strengthen torque of 8 Nm CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 6 1 1 Earth connection The earth connection lug on the power supply input serves to ensure a good connection to the protective earth on the test rig 6 2 Test rig Every test rig must be planned carefully All the instrumentation should be readily accessible and rigidly positioned The preceding safety measures are to be observed The test object may only be handled when the instruments and mains are switched off The whole test setup should be supplied from the same mains connection in order to prevent an uncontrolled flow of pulse current in other parts of the system Instal lation in a Faraday cage ensures that non associated items and equipment are not disrupted by pulses radi ated from the cabling or the device under test Connec tions to the EUT must be of low impedance and be made with high contact pressure otherwise welding or arcing might occur at the contact points TASEO Advanced Test Solutions for EMC AC mains cable L 0 5m CDN 3083 B100 or B200 EUT Protective earth on ground reference plane Mains supply EFT generator
6. non observation of the warning could lead to inconvenience or impairment in the performance Example This connection must not be confused with the main power input The following symbol draws your attention to a circumstance where non observation of the warning could lead to component damage or danger to the operating personnel Example Never connect or disconnect while the system is per forming a test Symbols used on the product A A Danger Protective earth Attention high voltage earth terminal refer to manual TASEO Advanced Test Solutions for EMC 5 2 SAFETY ADVICE mmm The coupling decoupling networks CDN 3083 B100 and CDN 3083 B200 are intended for use with burst generators like Teseq s NSG 3025 NSG 2025 Modula NSG 3040 and NSG 3060 series These CDNs handle high voltage test pulses and AC DC EUT supply voltages dH Improper or careless handling can be fatal Switch off EUT power supply before starting any manipu lations on the screw terminals These operating instructions form an integral part of the equipment and must be available to the operating personnel at all times All the safety instructions and advice notes are to be observed Keeping responsibility to the instructions and recommendations that follow is imperative The safety advice referring to the generator also applies when using the instru ments together with the CON 3083 B100 B200 Neither Teseq AG Switzerland nor an
7. C_CDN3083 B200_2012 LVD_CDN3083 B200_2012 Teseq AG CH 4542 Luterbach The purpose of this instrument is the generation of defined interference signals for EMI immunity testing Depending on the arrangement of the test rig the configuration the cabling and the properties of the EUT itself a significant amount of electromagnetic radiation may result that could also affect other equipment and systems The user himself or herself is ultimately responsible for the correct and controlled operation of the rig In case of doubt the tests should be carried out in a Faraday cage European representative Place and Date Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland Luterbach March 15 2012 Ss Johannes Schmid President TASEO Advanced Test Solutions for EMC Headquarters Teseq AG 4542 Luterbach Switzerland T 41 32 681 40 40 F 41 32 681 40 48 sales teseq com www teseq com China Teseq Company Limited T 86 10 8460 8080 F 86 10 8460 8078 chinasales teseg com Germany Teseq GmbH T 49 30 5659 8835 F 49 30 5659 8834 desales teseq com Singapore Teseq Pte Ltd T 65 6846 2488 F 65 6841 4282 singaporesales teseg com Taiwan Teseq Ltd T 886 2 2917 8080 F 886 2 2917 2626 taiwansales teseq com USA Teseq Inc T 1 732 417 0501 F 1 732 417 0511 Toll free 1 888 417 0501 usasales teseq com To find your local partner within Teseq s global network pl
8. ease go to www teseq com Manufacturer Teseq AG 4542 Luterbach Switzerland T 41 32 681 40 40 F 41 32 681 40 48 sales teseq com France Teseq Sarl T 33 139 47 42 21 F 33 139 47 40 92 francesales teseq com Japan Teseq K K T 4813 5725 9460 F 481357259461 japansales Qt eseq com Switzerland Teseq AG T 41 32 681 40 50 F 41 32 681 40 48 sales teseq com UK Teseq Ltd T 44 845 074 0660 F 44 845 074 0656 uksales Oteseg com O June 2013 Teseq Specifications subject to change withou notice All trademarks recognized Teseq is an ISO registered company Its products are designed and manufactured under the stric quality and environmental requirements o the ISO 9001 This document has been care fully checked However Teseq does no assume any liability for errors inaccuracie sor changes due to technicaldevelopments
9. g term and high current tests need to be kept under observation The displayed temperature on the CDN 3083 B100 amp 200 must not exceed 70 However the temperature at the screwed terminals may be higher this as a result of contact resistance between cable and terminal TASEO Advanced Test Solutions for EMC 8 3 INTRODUCTION mmm High frequency pulse signals can result in cross talk between conductors running in parallel even over short distances This behaviour is taken into account in the latest and currently valid edition of the EFT burst standard EN IEC 61000 4 4 2012 Only the single coupling mode of all conductors together against reference ground is called for in interference immunity testing with fast transients The units allow for test pulses of up to 8 8 kV The CDN 3083 B100 and the CDN 3083 B200 are designed to be used with an UT supply in AC mode up to 690 V in line to line or line to ground method DC voltage can be taken up to 1000 V line to line or line to ground m The CDN 3083 B100 amp B200 are 3 phase burst pulse coupling networks for large EUT currents Although designed for continuous duty at up to 100 and 200 A per phase the units can be subjected to considerably higher currents for short durations As a result high inrush currents or pulse shaped peak currents can be handled with ease The current limitation is largely a matter of thermal loading and ambient temperature A built in t
10. her solvents are not permitted The thermometer can be removed and may be replaced in case of damage It can be pulled out of his fixture by hand The CDN 3083 B100 amp B200 are fully potted so there are no serviceable parts internally CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 9 WARRANTY Teseq grants a warranty of 24 months against material and manufacturing defects Excepted from this warranty are damages incurred during transporta tion damages caused through use of the equipment for unforeseen purposes and the effects of excessive physical force TASEO Advanced Test Solutions for EMC 21 22 10 TECHNICAL SPECIFICATIONS EUT supply voltage EUT current AC or DC Burst pulse peak voltage Terminals Temperature monitoring Grounding Size Weight EFT burst connector Temperatur range Humidity Air pressure 690 VAC line to line and line to ground 1000 VDC line to line and line to ground 2100 A continuous for B100 and gt 200 A continuous for B200 The maximum permissible current is given by the heat dissipated in the coupler As the tem perature is monitored heavy overloads until the temperature reaches 70 C are possible 8800 V Screw terminals rated for 200 A Bimetal thermometer 0 to 120 C 4 mm combined banana and screw terminal safety grounding Via housing bottom plate and brackets HF ground to test setu
11. hermometer enables the heating effect to be monitored The type of construction inside a metal housing for use directly on the ground reference plane ensures excellent HF earthing The EUT supply connections take the form of generously dimensioned screw terminals The coupling network CDN 3083 B100 amp B200 serves to inject fast transients burst up to 8 8 kV 5 50 ns 50 O into the supply to the device under test CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 3 1 CDN 3083 B100 3 2 CDN 3083 B200 TASEO Advanced Test Solutions for EMC 10 4 INSTALLATION mmm The instruments conform to the safety requirements but with an increased leakage current given by the decoupling filter Operation without a protective earth connection is for E bidden Operate the equipment only in dry and clean surroundings Any condensation that occurs must be allowed to evaporate before putting the equipment into operation Do not exceed the permissible ambient temperature or humidity given in the IEC specification the EUT power supply is switched off and the generator s m Only handle the test rig or the device being tested when test sequence has been halted or aborted 4 1 Installation of an EUT power switch It is recommended to connect the CDN 3083 B100 amp B200 through a properly rated power switch device which should be located close to the test setup In order to ensure an easy and quick acces to
12. p ground plane 410 x 170 x 190 mm B100 430 x 310 x 190 mm B200 5 kg B100 11 kg B200 approx SHV Operation at 10 to 40 C Storage at 10 to 60 C 30 to 7896 non condensing 860 to 1060 hPa CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 11 ORDERING INFORMATION 23 11 1 Accessories included Type INA 3005 INA 3008 User manual Calibration certificate 11 2 Options INA 163 INA 3243 CAS 3025 INA 6546 Description Isolated Allen Key Coax connection cable 0 8 m SHV SHV to burst generator Safety banana plug set 10 pcs 6 to 4 mm Pulse measurement adapter Burst EFT Pulse Verification Kit Single SHV plug for custom made cables TASEO Advanced Test Solutions for EMC 24 12 DECLARATION OF CONFORMITY CE Declaration of conformity Manufacturer Address Product Options Generic standards Technical file THSEO Advanced Test Solutions for EMC Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland T 41 32 681 40 40 F 41 32 681 40 48 www teseq com Teseq AG Nordstrasse 11F 4542 Luterbach Switzerland declares that the following product CDN 3083 B100 EFT Burst Coupling Decoupling Network all conforms to the following Directives and Regulations EMC Directive 2004 108 EEC LVD Directive 2006 95 EEC EN61326 1 2005 EN61326 2 1 2005 EN61010 1 2001 The relevant technical file is available for inspection N EMC_CDN3083
13. t exceed 32 A INA 163 safety banana plug adapter set 7 2 Pulse measurement adapter INA 3243 optional The measurement adapter INA 3241 is necessary to measure the Burst EFT pulse at output terminals INA 324 3 Art Nr 1001106 INA 3243 measuring adapter TASEO Advanced Test Solutions for EMC be done with a switched off and disconnected power The measurement or the verification of the has to supply Pulses can then be observed at the output connectors by means of an oscil loscope 7 3 Burst EFT Pulse Verification Kit CAS 3025 optional To fulfil the EN IEC 61000 4 4 standard recommendation each line needs to be measured separately terminated with 50 For this purpose the termi nator attenuator INA 265A which is part of the CAS 3025 calibration set is recommended 7 4 Isolated allen Key INA 3005 included CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 7 5 EFT Cable INA 3008 included 19 7 6 Single SHV plug INA 6546 optional Single SHV plug for custom made cables to connect with any EFT generator TASEO Advanced Test Solutions for EMC 20 8 MAINTENANCE mmm Maintenance is limited to cleaning and eventually removal of any foreign par ticles or objects from the contact surfaces since such matter can adversely affect the test parameters The housing can be cleaned with a moist cloth with optionaly just a trace of liquid detergent Ot
14. the EUT power the switch should be clearly and visibly labelled as a device for EUT power on off switching The inhouse power distribution needs to be equipped with a proper circuit breaker and an emergency off button as per IEC 61010 1 CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks The test setup should only be accessible to trained Hn persons 4 2 Leakage current Local installation regulations must be respected to ensure the safe flow of eakage currents Use only nationally approved connectors and accessory items Ensure that a reliable return path for the interference current is provided between the EUT and the coupling network The reference ground plane and the earth connections to the instruments as described in the relevant test standards serve this purpose well 4 3 EUT supply protection Dimensioning of EUT supply and rating of fuse protection AC or DC must conform with National prescriptions and EUT requirements An inappropriate arrangement mounting cabling and or handling of the device under test or the protective elements can make the protective features that are incorporated in the concept of the device worthless TASEO Advanced Test Solutions for EMC 11 12 5 CIRCUIT DIAGRAM Block diagram of the CDN 3083 B100 amp 200 Signal from test generator Cc PEO 4 4 mms mE OPE Rus D i Lc Phase qp n EL
15. y of the associated sales organisations accept any responsibility for personal injury or for material or consequential damage that may result from irresponsible or negligent operation of this equipment The coupling devices are built in conformity with the IEC 61010 1 2001 safety standard and provide all the amenities for safe and reliable operation CDN 3083 B100 amp CDN 3083 B200 EFT burst coupling decoupling networks 2 1 Safety measures These operating instructions form an integral part of the instrument and must be studied carefully before putting the device into use They must be available at all times to the operating personnel The instrument must only be used by trained personnel Persons fitted with a heart pacemaker must not operate the instrument and should not be in the vicinity of the test rig when pulses are triggered Arrange the working area in compliance with appropriate safety standards Connect the additional earth plug on the mains supply input side first The test rig must provide adequate insulation protection up to max pulse voltage evel of the generator up to 8 8 kV Particular care should be given to the con nections between the CDN 3083 B100 amp 200 and the equipment under test EUT The pulse voltage must not be able to find its way to unearthed metal objects in the event of the EUT failing Never leave the instrument unattended when it is connected Do not open the instrument Lon