1. 7 The sample is initially a liquid After several minutes the sample begins to evolve into a clot The analyzer detects this initial clot formation and calculates the time the sample remained a liquid ACT During the next several minutes of analysis the fibrinogen converts into a fibrin gel The analyzer calculates the rate of change in the clot signal Clot Rate 8 When a result has been calculated the analyzer will beep and display the time that the sample remained a liquid ACT and the Clot Rate CR ACT 177 CR 19 Times for results vary from 2 to 20 minutes depending on heparin levels and clinical conditions The numeric result is displayed next to the corresponding text for the appropriate channel NR is displayed if no results were found for the test sample gbACT 9 The analyzer will automatically stop data collection when results for ACT and Clot Rate have been calculated or after 30 minutes 10 Open the head and remove the probe from the hub with a probe extractor Avoid moving the hub sideways Propery discard the probe and the cuvette Lower the head assebly to maintain temperature control of the instrument 2 4 Chapter 2 Running a Whole Blood Sample Chapter 3 Maintenance and Troubleshooting Maintenance Cleaning Clean after use to reduce biohazard risks The analyzer can be sprayed or wiped with a disinfectant approved by your institution Close the heads when spraying to avoi
2. This requires QC kit 900 1303 Monthly and prior to the use of a new shipment validate the activated cuvettes This requires QC kit 900 1318 Complete instructions for running each QC test are included in the kits The kits also include forms for recording QC results To order quality control kits contact Sienco at 800 432 1624 or contact your distributor 1 6 Chapter 1 Installation and Setup Chapter 2 Running a Whole Blood Sample The analyzer monitors mechanical changes that occur during hemostasis This section describes how to prepare and run a test with native whole blood samples Use only test kits available from Sienco Inc Sienco offers different tests for different applications Choose the test best suited for your application Analyzer Preparation 1 Cuvettes should be placed in the convenience wells so they will be ready when the blood is drawn Place probes into the lids of the cuvettes EA iG a Es ae Pee age PBI AS i a A er LG HE Pa II a ne ad probe cuvette 2 Ifthe analyzer has just been recently turned on allow it to warm with head assembly in the down position until the analyzer reaches 37 C If the analyzer is not at the desired temperature it will display an error message and not run the test 3 When the analyzer is ready the display reads i nn gbACT NOTE Results from the previous test may be displayed on the bottom of the LCD display Press the S
3. in the European community CE marking of conformity Biological risk Temperature limitation Humidity limitation Direct current DC Serial number Do not dispose of electrical equipment as municipal waste HES O j gt Y E CONTROL Do not reuse Model Number Lot Number Manufacture date Expiration date Consult instructions for use Contains sufficient for n tests Caution consult accompanying documents Control Decontamination Form This decontamination form must be filled out and returned with the Sonoclot Coagulation Analyzer when shipping the analyzer to Sienco for any reason or a 150 00 biohazard decontamination charge will be assessed Please photocopy this page complete and include with your analyzer We require you to thoroughly clean and decontaminate the analyzer per guidelines followed by your institution Here are some tips to follow when decontaminating the analyzer e Place a clean probe on the probe adapter to protect the sensing mechanism while cleaning Do not spray any cleaning solvent into the head since fluid will damage the sensor e Decontaminate all surfaces of the analyzer using a product certified by your Institution e Remove and discard all probes and cuvettes before packing the analyzer Institution Dpmt Serial Number _________ Model Decontaminated by ___ Date Decontaminate Used Probe removed NE Cuvette removed ME Power cord amp ad
4. cuvette Chapter 1 Installation and Setup 1 5 Warnings Cautions and Hazards 1 WARNING As with any laboratory test result diagnosis should not be based solely on the analyzer test results The attending physician is responsible for interpreting the analyzer test results in conjunction with the patient s condition other test results and clinical observations 2 WARNING To safeguard against electrical shock only use a grounded electrical outlet 3 WARNING Use only as specified by Sienco Inc Otherwise equipment protection may be impaired 4 CAUTION Only properly trained laboratory personnel and or other appropriate health care professionals should use the analyzer 5 CAUTION A biomedical engineering department should check the analyzer at least once a year for compliance with leakage standards 6 HAZARD Human blood is a biohazardous material The operator should wear appropriate protective gear when handling blood and or test cuvettes containing BON blood samples Biocontaminated materials should be handled and disposed of YY E properly in accordance with hospital local state and federal regulations 7 CAUTION Do not place the analyzer on counter tops where other vibrating instruments such as centrifuges are located Mechanical vibration may cause erratic results Quality Control Proper performance requires regular quality control QC tests 1 2 Daily run a reference viscosity test for the analyzer
5. 420 1148 3 2 Chapter 3 Maintenance and Troubleshooting Factory Service or Repair Contact Sienco prior to shipping your analyzer for repairs The analyzer should be shipped in its original packing materials if possible If you do not have the original packing materials please contact Sienco We will ship you packing materials or you may use the packaging materials from your loaner We charge a minimum fee for testing the instruments even if you decide not to have it repaired Damages caused by poor packaging are your responsibility To ship follow these steps Remove cuvettes probes power cord and other supplies Photocopy the decontamination form at the end of this manual Clean and decontaminate the analyzer according to your institutional guidelines and fill out the decontamination form Failure to fill out form and include with the analyzer will result in a 150 decontamination fee 4 Carefully slide the spacer tabs between the instrument case and platen Tape the back of the platen and the top of the instrument case as shown Tape the instrument head down as shown Place the analyzer in a large clean plastic bag and fold the open end under the instrument Place the instrument in the molded foam in the bottom of the box with the ribbon running underneath the analyzer as shown If you are also returning your power adaptor f a and cord fold the power cords secure each Tamem a nn m with a rubber band an
6. Sonoclot Coagulation Analyzer Operator s Manual SCI Revision 2 1 SONOCLOT C oagulation Analyzer For in Vitro Diagnostic Use Manufactured for Sienco Inc 7985 Vance Drive Suite 104 Arvada CO 80003 USA 303 420 1148 1 800 432 1624 Fax 303 379 4403 www sienco com e mail sienco sienco com Copyright 1999 2009 Sienco Inc All Rights Reserved Manufactured For Sienco Inc al gt Vance Drive Suite 104 Arvada CO 80003 USA 303 420 1148 303 379 4403 FAX e sienco sienco com e mail http www sienco com Sonoclot is a registered trademark of Sienco Inc Sonoclot Analyzers are protected under U S and foreign patents Ec Rep QNET BV Hommerterweg 286 C 6436 AM Amstenrade The Netherlands Table of Contents Chapter 1 Installation and Setup Principle of Operation and Intended Use Head Assembly Model SC1 Front View Model SC1 Back View Package Contents Installation Good Operating Procedures applicable for all tests Warnings Cautions and Hazards Quality Control NO N gt gt e e e e ah I ll l OO UVUWWND Chapter 2 Running a Whole Blood Sample Analyzer Preparation Cuvette and Probe Setup 2 2 Obtaining the Blood Sample 2 3 Running the Analyzer 2 3 Chapter 3 Maintenance and Troubleshooting 3 1 Maintenance 3 1 Troubleshooting 3 2 Warning and Error Messages 3 2 Factory Service or Repair 3 3 Appendix A 1 References A 1 Warranty A 2 Technical Specifi
7. aptor returned Y NE N AE Contact Person __ o o o o o o Oo Phone Reason for Service We will call with an estimated cost for repairs before any work is done Please make sure to include a contact name and phone number Ship the Sonoclot Coagulation Analyzer to Sienco Inc Attn Service Manager 7985 Vance Drive Suite 104 Arvada CO 80003 USA If you have any questions please contact Service Manager 800 432 1624 303 420 1148 Appendix A 5
8. cations A 2 Environmental Conditions for Transportation and Storage A 2 Environmental Conditions for Use A 3 Electrical Classification A 3 Disposal Instructions A 3 Glossary of International Symbols A 4 Decontamination Form A 5 Chapter 1 Installation and Setup Principle of Operation and Intended Use The Sonoclot Coagulation Analyzer calculates onset of clot formation and clot rate for in vitro coagulation studies by monitoring mechanical changes that occur in blood samples during hemostasis The mechanism is a tubular probe that moves up and down within a blood sample As the sample progresses through various stages of clotting the electronic circuitry detects increasing resistance This produces a series of electronic signals that are processed by a microcomputer Results are reported on the LCD display panel Head Assembly analyzer head probe mount hub convenience wells cuvette holder platen head alignment pin 1 1 Model SC1 Front View J analyzer head O Par tine era vt convenience b well platen LCD display Start Stop button LOT Coagulation Anal y zer For in Vitro Diagnostic Use Select Test button Model SC1 Back View serial number AS labe a re ili The Neiherands j a een electrical information Arvada Colorado 20003 MADE IN USA s power cord pol receptacle 1 2 Chapter 1 Installation and Setup Package Contents e Sonoclot Coagulatio
9. d contaminating the transducers Avoid excess wetting Place a clean probe on the probe adaptor to protect the sensing mechanism while cleaning Gently wipe around the cuvette holders or inside the heads Do not clean the probe mount hubs unless absolutely necessary The transducers attached to the probe mount hubs can easily be damaged by debris or liquids NOTE Do not use isopropyl alcohol or other solvents on the front panel or LCD cover Calibration amp Service The user cannot calibrate or service the analyzer Running a reference viscosity QC test PN 900 1302 will verify if the analyzer is operating within calibration standards Sienco recommends performing the reference viscosity QC test daily prior to use of the analyzer If the analyzer requires calibration or service contact Sienco Inc or your distributor 3 1 Troubleshooting If you have a problem with the analyzer follow these steps 1 Review Warnings Cautions and Hazards on page 1 6 2 Review Good Operating Procedures on page 1 5 3 Review the operating instructions that come with the test you are running 4 If problems persist contact the Sienco service department Phone 303 420 1148 Toll Free 800 432 1624 Fax 303 379 4403 E mail sienco sienco com Mail Sienco Service Department 7985 Vance Drive Suite 104 Arvada CO 80003 USA Warning and Error Messages The Sonoclot Analyzer will occasionally display messages that may not be fami
10. d place in the spaces on either side of the analyzer Sonoctor Place the top of the molded foam over the salvar MENE sure sa is securely packed around the analyzer head Make sure the decontamination form is complete Be sure to include a brief description of the problem along with a contact name phone number and purchase order Place the completed form on top of the molded foam Tape the box shut and ship the package to Sienco Inc Service Department 7985 Vance Drive Suite 104 Arvada CO 80003 USA Chapter 3 Maintenance and Troubleshooting 3 3 Appendix References Sonoclot Analysis is used in many clinical and research applications and consequently is referenced in numerous studies and articles For acomplete list of references please visit our web site at www sienco com 1 Nilsson CU and Engstr m Monitoring fondaprinux with the Sonoclot Blood Coagulation and Fibrinolysis 2007 18 619 622 2 Ganter MT Monn A Tavakoli R et al Monitoring activated clotting time for combined heparin and aprotinin application in vivo evaluation of a new aprotinin insensitive test using Sonoclot Eur J Cardiothorac Surg 2006 30 2 278 84 3 Shibata T Sasaki Y Hattori K et al Sonoclot analysis in cardiac surgery in dialysis dependent patients Ann Thorac Surg 2004 77 1 220 05 4 Liszka Hackzell JJ and Ekback G Analysis of the information content in Sonoclot data and reconstruction of coagulation test
11. eep blood dirt and other substances away from the probe mount hub to avoid contaminating the electromechanical transducer e Never reuse a probe or cuvette to prevent thrombin contamination For consistent results the cuvettes should be warmed prior to analysis To warm place the cuvette in its holder at least 30 seconds prior to running a test Always insert and remove the probe by moving it vertically over the probe mount hub Never move the hub horizontally Make sure the probe is fully seated to avoid interference with the stir bar Always remove the cap from the cuvette prior to placing it into the cuvette holder Failure to do so can damage the transducer The cuvette must be fully seated in the cuvette holder to avoid interference between the probe and stir bar Do not overfill the cuvette The proper fill level is slightly below the inner rim of the cuvette Native whole blood must be analyzed within 2 minutes or less of collection Perform QC testing to verify proper operation of the analyzer and activation cuvettes See page 1 6 Use proper biohazard handling techniques to dispose of probes and cuvettes On rare occasions mechanical disturbances may cause incorrect results Always inspect results to ensure that they are consistent High viscosity blood samples immersion response gt 25 clot signal units as displayed on the instrument can stratify Use an external device to mix the blood sample before placing it in the
12. elect Test button on the front panel until the desired test is displayed on the LCD 2 1 Cuvette and Probe Setup 1 Open the head by tilting it backwards The display will read Open 2 Sharply tap the cuvette on a hard surface to dislodge activation powder from the sides and lid 3 With a slight twisting motion use the cuvette to seat the probe on the mount hub This motion should result in the probe sliding straight over the probe mount hub The hub should not move sideways The probe must be fully seated for proper operation 4 Remove the cuvette leaving the probe on the hub 5 Remove the lid from the cuvette by placing it in a convenience well and popping the lid off with your thumb 2 2 Chapter 2 Running a Whole Blood Sample 6 Insert the cuvette into the cuvette holder Make sure it is fully seated Close the head Allow the cuvette to warm in the cuvette holder for at least 30 seconds DO NOT close the head on a lidded cuvette It will damage the instrument Different tests have different set up requirements Please refer to the product insert for detailed ve lt a ae Lbs AR Instructions Obtaining the Blood Sample Native whole blood must be analyzed within 2 minutes of collection When drawing blood observe the following precautions 1 Carefully decide where to draw the sample Contamination from a heparinized line a heparin impregnated catheter or a surgery p
13. liar Below are some examples of these messages and procedures to follow if necessary NOISE indicates that the clot signal is being disturbed by either mechanical noise such as bumping the analyzer or by interference between the probe and the cuvette Check to ensure that the probe is tight against the mount hub and the cuvette is in contact with the bottom of the cuvette holder The message should disappear within a few seconds ERROR indicates that the self test routine has detected a problem Normally the message will disappear after several seconds You can resume testing then If the error message remains make a note of the error number and call Sienco s service department TEST NOT RUN NOT AT TEMP indicates that the analyzer is not to temperature and the test was not run The analyzer must be at 37 C to run a test NR indicates that no results were found for the last test analysis This can occur due to operator error i e failure to close instrument head within 60 seconds of Close Head message instrument error or an unusual clot that cannot be analyzed NUMBERED ERROR CODES The analyzer is programed to display various numbered error codes Should your analyzer display a numbered error code record the number displayed and contact your distributor or Sienco s service department If you have any questions in regards to a display message call Sienco s service department at 800 432 1624 or 303
14. n Analyzer Model SC1 e nower adaptor and cord e probe extractor e operator s manual e USB cord e SonoView installation CD Installation 1 Unpack the contents of the shipping container and check you have all the above Items NOTE Sienco recommends that you retain the original shipping box in case you need to return your instrument for service 2 Pull the ribbon to remove the packaging from the top of the instrument Then pull the ribbon to lift the instrument out of the packaging foam in the bottom of the box E 3 Remove the analyzer from the plastic bag 4 Carefully remove the tape from the top and back of the instrument 5 Remove the spacer tabs from between the analyzer case and the head assembly platen Keep these tabs with the original packaging 6 Plug the power cord into the analyzer and an easily accessible grounded wall outlet Do not use an ungrounded extension cord or adaptor Do not obstruct access to the outlet 7 The analyzer will automatically turn on First the LCD display will illuminate Next the analyzer will display the version number product name and copyright 8 The analyzer will begin warming While warming the display reads gbACT O Heating 36 3 9 When the analyzer has reached 37 0 C the display indicates the default test The analyzer is now ready to run a test 10 Review the remainder of this chapter Chapter 1 Installation and Setup 1 3 Installing am
15. p Using SonoView Data Management Software SonoView The analyzer indicates real time results on the LCD display It also collects and stores data much like a digital camera This data can be accessed by connecting the analyzer to a computer running SonoView To install SonoView on your computer 1 Read the installation instructions and hardware specifications on the CD sleeve Instructions are also located in PDF format on the CD 2 Plug a USB cord into the back of the analyzer and a USB port on your computer NOTE The analyzer requires a 1 1 USB Port or higher 3 Open SonoView on your computer For software instructions please see the SonoView operator s manual PDF file on the installation disk 1 4 Chapter 1 Installation and Setup Good Operating Procedures applicable for all tests The quality of test results depends on proper technique Carefully follow these precautions 1 10 11 XA oe 12 13 Ner 14 Ner Operator should read this operator s manual in its entirely before attempting to operate the analyzer Only properly trained lab personnel and healthcare professionals should operate the analyzer Diagnosis should be based on test results in combination with clinical observations Handle materials with care e When taking a blood sample avoid heparin contamination from catheters e Never use the first sample from a new line to avoid sample contamination with tissue thromboplastin e K
16. ranty is limited to the replacement or repair of defective parts and components The warranty shall be voided if the product is 1 Used in any manner or subjected to any condition that is inconsistent with its intended purpose or accepted industry practice or 2 Modified in any way without prior written approval by Sienco or 3 Repaired in any manner so as to adversely affect its operation or reliability by persons other than Sienco or its agents Warranty claims shall be submitted in writing to Sienco during the 2 year warranty period The claim shall state the nature and details of the analyzer defect and the serial number of the analyzer The claim must accompany any defective product submitted for repair or replacement Purchasers must decontaminate and package the analyzer per instructions on pages 3 1 and 3 3 Such products shall at Sienco s election be either repaired replaced or returned Technical Specifications Width 4 25 10 8 cm Depth 5 75 14 6 cm Height 4 75 12 1 cm Weight 2 5 lbs 1 13 kg Electrical voltage requirement 100 to 240V Electrical power requirement 9V 2A Frequency 50 to 60 Hz Temperature regulation of platen 37 C 0 5 C Viscosity range for test sample lt 80 cP Environmental Conditions for Transportation and Storage Transport and store within the ambient temperature range of 25 C and 70 C 95 Transport and store to a maximum relative ambient humidity of 95 A 2 Appendi
17. rep line will cause inaccurate results 2 Withdraw blood in a smooth slow and non traumatic manner Do not use force 3 When drawing from an anesthesia or pump port use a 2 syringe technique Discard the first syringe of 2 to 3 ml Use the second syringe for the sample You MUST use plastic syringes to avoid uncontrolled glass activation Running the Analyzer 1 Transfer the whole blood sample from the syringe into the warmed cuvette Use a blunt cannula tip for a clean and controlled fill Fill the warmed cuvette with the blood sample until the fluid level is slightly below the inner rim of the cuvette This volume is approximately 330 to 360 ul Fill slightly below here Dg N Bes 4 E x 2 Leaving the head open immediately press the START STOP button QbACT 3 The magnetic stirrer will rotate and the display Mixing will read 4 When mixing is completed the analyzer QbACT will beep and the display will read Close Close Head Head Chapter 2 Running a Whole Blood Sample 2 3 5 Close the analyzer head Make sure it is fully seated over the head alignment pin 6 During data collection the LCD display will show the test being run the clot signal reading 9 bACT 13 Ll CS the time in seconds and 222 for each JACT CR result When the analyzer is calculating a result the display will flash the number the result is greater than
18. required to correct the interference Disposal Instructions Sienco Inc expects end users to dispose of the Sonoclot Coagulation Analyzer in an environmentally friendly way Electrical and electronic equipment is labeled with the following crossed out wheeled bin symbol indicating that the equipment should be disposed of by the end user separate from other types of waste End users should contact their dealer distributor or Sienco Inc for disposal collection and recycling options and terms and conditions in your country In 2002 the European Union introduced the Directive on Waste Electrical and Electronic Equipment WEEE The main aim of the Directive is to ensure that WEEE is collected and Appendix A 3 treated separately WEEE may contain hazardous substances that should not end up in the human environment and can have adverse effects on it if they do WEEE is a vast source of raw materials With the ever rising worldwide demand for new equipment and the ever decreasing volume of natural raw materials letting this potential source go to waste is unacceptable If equipment is collected separately the equipment can be recycled and up to 85 to 90 of the equipment can be re used as new material saving the use of virgin raw materials and energy of producing these Glossary of International Symbols se lt e RARE A 4 Appendix Manufacturer In Vitro Diagnostic medical device Authorized representative
19. variables Journal of Medical Systems 2002 26 1 1 8 5 Pivalizza EG Pivalizza PJ Kee S Gottschalk LI Szmuk P Abramson DC Sonoclot analysis in healthy children Anesth Analg 2001 92 4 904 6 6 Santucci RA Erlich J Labriola J et al Measurement of tissue factor activity in whole blood Thromb Haemost 2000 83 3 445 54 7 Ekback G Carlsson O Schott U Sonoclot coagulation analysis a study of test variability J Cardiothorac Vasc Anesth 1999 13 4 393 7 8 Hett DA Walker D Pilkington SN Smith DC Sonoclot analysis Br J Anaesth 1995 75 6 771 6 9 Steer PL Krantz HB Thromboelastography and Sonoclot analysis in the healthy patient J Clin Anesth 1993 5 5 419 24 10 Shenaq SA Saleem A Viscoelastic measurement of clot formation the Sonoclot In Ellison N Jobes DR eds Effective Hemostasis in Cardiac Surgery Philadelphia PA Harcourt Brace Jovanovich Inc 1988 183 93 11 Newland MC Chapin JW Hurlbert BJ Kennedy EM Newland JR Thrombelastograph and Sonoclot Signature monitoring of changes in blood coagulation following cardiopulmonary bypass Anesthesiology 1987 67 3A A199 12 Newlin F Ens G Leppke L and Hamstra R Heparin Control with the SONOCLOT American Journal of Medical Technology 44 508 1978 A 1 Warranty The Sonoclot Coagulation Analyzer has a 2 year warranty that it conforms to specifications and is free of defects in material and workmanship The war
20. x Environmental Conditions for Use Indoor use only Use at a maximum altitude of 2000m K Use within the ambient temperature range of 18 C and 27 C Use at a maximum relative humidity of 80 for temperatures up to 31 C 7 decreasing linearly to 50 relative humidity at 40 C Mains supply voltage fluctuation shall not exceed 10 The unit is intended for INSTALLATION CATEGORY II Installation Category Il Local level appliances Portable equipment etc with smaller transient overvoltages than Installation Category II The unit is intended for use in a POLLUTION DEGREE 2 ENVIRONMENT Pollution Degree 2 is nonconductive pollution of the sort where occasionally a temporary conductivity caused by condensation must be expected This is the usual pollution degree used for equipment being evaluated to 60950 and is suitable for equipment employed in an office environment Electrical Classification This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the US FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in its installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications If this equipment does cause harmful interference the user will be

Download Pdf Manuals

image

Related Search

Related Contents

Recurring Contract Billing Importer  平成15年那審第12号 プレジャーボートトッチャンボーヤ2号同乗者死亡  - Magnetrol  Bedienungsanleitung-Scansonic DA88  Element 302 User Manual  Manual de usuario en español Modelo ZMRK700    HP SureStore E Disk Array 12H System Administrator`s  HomComfort HCPHSSQ Instructions / Assembly  Endpoint Protector - Mobile Device Management - User Manual  