DIGORA® Optime DXR-50 001 DIGORA® Optime DXR

image

Contents

1. SOREDEX dental Imaging Plates IPs pro tective covers hygiene bags and other related imaging plate accessories APC not supplied in which suitable dental imaging software Alocal area network LAN cable not supplied will be required if the system is to be used in a network The unit is a laser scanning device designed to automatically read reusable SOREDEX dental in traoral imaging plate sizes 0 1 2 and 3 The Com fort Occlusal 4C kit also allows two size 3 imag ing plates to be used to take an occlusal projection image After reading the images can be viewed on the PC using the dental imaging sofware The unit can either be connection directly to the PC or to network via the LAN cable User s manual 204253 1 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 1 Introduction The unit can be set up to work with a single PC the single user configuration or with several PCs the multiconnect multiple user configuration With the multiconnect configuration up to eight PCs can be used one at a time with the unit This man ual describes how to use both configurations NOTES Only personnel trained and authorized by the manu facturer of the unit are allowed to install and config ure the unit Only use the imaging p
2. Computer network connection Intell Al PAO TOO0 MTW Network Connection Scanner F address f IF based W Enable changing IF address fi 92 168 211 Send to Scanner B 4 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit Obtain an IP address for the unit from your net work administrator and key it into the IP field in W Enable changing IF addi the Scanner IP address area NOTE The PC and the unit must be in the fi 92 168 2 11 same subnet when setting the IP address of the unit Scanner F address on the unit and then click the Send to Scanner button on the settings window Send to Scanner You will hear a beep which indicates that the PC is now sending the IP address the unit The Startup animation will appear on the unit display O 3 PC Unit Press and hold down the Start key Startup gt animation Name Serial number FW x xx CORE xxx IP XXX XXX XXX XXX 4 PC When the Startup animation clears click OK to connect the PC to the unit 5 Now connect the other PCs in the network to the unit Just enter the IP address into the IP field and then click OK to connect the PC to the unit it is not necessary to hold down the Start key and click the Send to Scanner button with the
3. 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup Preview Time Plate eject mode Drop in plate collector Preview Time o 6 Allows you to select the length of time that an image will remain on the unit display after scanning Minimum 0 seconds no image appears maximum 60 seconds B 14 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit B3 4 Power options 1 From the dental imaging software you are using select Setup Power options page Power options Standby after seconds Beep when entering standby mode Shutdown after minutes Cancel Power options Standby after seconds Allows you to select the length of time the scanner will remain unused be fore it enters the standby mode Display switches off and power LED turns yellow Beep when entering standby mode Select this and audible signal will be heard when the scanner enters the standby mode Shutdown after minutes Allows you to select the length of time the scanner will remain unused be fore the unit automatically switches itself off NOTE If the Automatic UV disinfection option Hygiene is selected the UV disinfection cycle will be auto matically carried out before the unit switches itself
4. Below 33 C 80 RH and shielded from X rays and ultraviolet radiation lf an IP is stored for over 24 hours in a hygiene bag or in a location that is shielded from ambi ent light the IP must be erased to remove any potential fogging before being used to take an exposure User s manual 204253 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 4 Handling and care of imaging plates Imaging plate unit 4 5 Replacement Replace an IP if e the image shows marks spots or dots which still appear on images even after the IP has been properly cleaned e the IP is mechanically damaged scratched or badly bent 4 6 Disposal IPs must be disposed of in accordance with all local national and international regulations regarding the disposal of non environmentally friendly or hazard ous materials Phosphor substance under the top coat on the sensitive side of IPs must not be swallowed User s manual 204253 31 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 5 Unit care and maintenance 5 Unit care and maintenance WARNING Switch the unit off and disconnect it from the main power supply before cleaning or disinfecting the unit Do not allow liquids to enter the unit 5
5. conform to the Russian Safety Standards This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately Please contact an authorized represen tative of the manufacturer for information concern ing the decommissioning of your equipment 34 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 7 Warnings and precautions Imaging plate unit 7 Warnings and precautions THE UNIT IS ACLASS 1 LASER PRODUCT Note When covers are removed the unit is a class 3B laser product avoid expo sure to the laser beam or the UV C light inside the unit CAUTION Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser radiation exposure or excessive exposure to UV C light e When handling imaging plates protective covers and hygiene bags always take the appropriate hygiene measures and precautions to prevent cross con tamination e The imaging plates are harmful if swallowed e Do not move or knock the unit when it is reading an imaging plate e This unit must only be used to read image plates supplied by the manufacturer and must not be used for any other purpose NEVER use imaging plates pro tective cov
6. into the unit NOTE It may take longer to erase IPs using the eras ing mode than the normal read and erase mode This is to ensure that IPs that have not been used recently are erased properly 3 After the IP has been erased it will be ejected from the unit The IP can now be used to take an exposure 4 To exit the erasing mode either wait 15 sec O onds for the unit to automatically exit the mode or press and hold down the Start key until the erase animation disappears 26 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 12 Taking an occlusal projection image Comfort Occlusal 4C images are produced using TWO size 3 IPs the Comfort Occlusal bite protector and the hygiene bag The Comfort Occlusal 4C IP Refer to the separate instructions supplied with the Comfort Occlusal 4C kit IMPORTANT NOTE Due to occlusal imaging geometry and imaging plate positioning accurate distance and angle measurements cannot be taken from occlusal pro jection images 1 PC Select occlusal from Setup gt Scanner gt Occlusal see section B3 6 Occlusal 2 Place two size 3 IPs into their corresponding protective covers 3 Slide the two size 3 IPs and protective covers into the Comfort Occlusal
7. not reconfigure the system ii Ifthe Dental imaging software animation flashing yellow appears it indicates that a patient has not been selected for the scanned image s Select a patient iii If the Insert animation appears on the unit dis play while inserting the IP but the unit does not insert the IP into the unit nor start scanning the reflective sensor in the unit door is faulty refer to the service manual B 20 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit PROBLEM Local area network LAN subnet configuration problem CAUSE SOLUTION i Ifthe Dental imaging software animation flashing yellow appears and cable s router s are known to be okay ask your network adminis trator for assistance li Conflict mismatch P addresses and or sub net mask of the unit PC Use command lt ipconfig gt from the command prompt Start gt gt Run gt gt cmd in Windows to get information about the active network set tings Change either the IP address of the PC or the unit so that they are both in the same subnet If the problem persists ask your network admin istrator for assistance iii Network connection not functioning or not stable due to LAN signal quality problems Install a network switc
8. number B2 1 Direct connection method uses the unit s n IMPORTANT NOTE It may not be possible to connect the unit to the PC using the direct connection method if another device is already connected to the PC using direct con nection If the direct connection field is not active greyed out or the system does not work correctly after the unit has been connected reconnect the unit using the IP connection method 1 After positioning the unit connect it to the PC s in the local area network using the Ethernet cable s not supplied 2 Switch the unit on The dental imaging software animation will ap pear This indicates that the unit is not communi cating with the PC s in the network 3 PC Install the dental imaging software to be used in the PC s B 2 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit 4 PC Open the dental imaging software and se lect the Scanner setup window SCANniel Settings Driver Info Status Scanner Oleconmected VW ErSIOrT Serial No 5 PC From the Scanner setup window select the Settings tab to open the Scanner Connection page Scanner settings Driver Info Scanner Connection Direct Connection settings Scanner senal number Computer network connec
9. off Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup B3 5 Hygiene 1 From the dental imaging software you are using select Setup Hygiene page Automatic UY disinfection Ultraviolet U5 light inactivates common harmful viruses and bacteria from the intemal parts of the unit Disinfection process may take several minutes Disinfection frequency Cancel disinfection with During idle and shutdown Proximity sensor or START key Manual UY disinfection contral Disinfect Automatic UV disinfection The internal ultraviolet UV light allow you to disin fect the plate carrier when it is inside the scanner Disinfection frequency There are three options During idle and shutdown UV disinfection is automatically performed when the unit is in the idle mode and before shutdown There is no user indication of this taking place After each patient UV disinfection is automati cally carried out after the patient card is closed The remaining UV disinfection time appears on the unit display Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation
10. or yellow CAUSE SOLUTION e Green LED defective display or its cabling Replace e Yellow LED unit is in standby energy save mode B 22 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182
11. other PCs once the unit has already got an IP address B 5 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup B2 3 Multi Connect 1 PC If the unit is to be used with several PCs select the Use Multi Connect check box and select a unique Workstation identifier number between 1 and 8 for the PC being configured from the drop down list Addition workstation in formation for example user name location etc can be entered into the field next to the work Station identifier number IMPORTANT NOTE If only one PC is connected to the unit do not select the Use Multi Connect check box MMult Connect i Use Multi Connect Workstation identifier i wy FIw2251 B16 Scanner Sutorelease timeout seconds pn e m Z 4 4 p b i g Cancel The Scanner Autorelease timeout is the length of time that the unit will remain reserved and unused by a PC before the PC automatically re leased the unit so that it can be used by another PC in the system the scanner can be reserved in advance from another PC The default setting is 40 seconds This can be changed by keying in a new value B 6 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Gro
12. power on off status light and display are off CAUSE SOLUTION The main power supply to the unit is off or the unit is not switched on I Ifthe LED on the PSU is not on it is not receiv ing power from the mains Switch the mains power on ii Ifthe LED on the PSU is on switch the unit on lil If status light and display still do not come on check the unit s membrane control panel and cabling Replace if faulty PROBLEM Animation displaying either Unit connection or Dental imaging software flashing yellow appear on the unit display and the unit will not work CAUSE SOLUTION e Defective RJ45 cable Replace Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup PROBLEM The IP connection between the unit and the PC does not work CAUSE SOLUTION Check that the Direct connection method was configured correctly If all the setting are correct but the connection still does not work use the IP connection method li The unit has been configured to work with one PC only and does not recognize any other PCs Reconfigure the system for multiple users PROBLEM The unit is connected to the PC but cannot scan images CAUSE SOLUTION i Reserve the unit and check that it is configured for multiple users If
13. setup B3 6 Occusal From the dental imaging software you are using select Setup Occlusal page Occlusal Size 3 The system can ether process size 3 imaging plates as individual images or stitch two sequential size 3 plates together to create a single occlusal image f Oeclusal 4C image from two sequential size 3 plates f Size 3 image from each size 3 plate Occlusal Size 3 Occlusal 4C image from two sequential size 3 plates allows the two size 3 IPs used in the Com fort Occlusal 4C IP to be read separately and then stitched together to form a single occlusal image Refer also to 3 12 Taking an occlusal projection image and the Occlusal quick guide NOTE When you are in the Occlusal 4C mode it is possible to temporarily override the mode and read a single I I O size 3 IP Insert the size 3 IP into the unit so that it can be read When the insert second plate symbol appears on the unit display press the start key This will cancel the Occlusal 4C mode for this operation and produce a single size 3 image Size 3 image from each size 3 plate allows size 3 IPs to read as individual IPs Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit B4 Troubleshooting PROBLEM The unit does not come on The units
14. size 3 IPs cannot be read Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit Touchless operation Touchless operation Scanner will wake up automatically from standby mode if touchless operation sensor is in use and activated M Use touchless operation sensor lf Use touchless operation sensor is selected the unit will automatically wake up from the standby mode as soon as the sensor detects a movement within a distance of 20 cm If itis NOT selected the unit will wake up when the Start is pressed Plate eject mode Plate eject mode Preview Time Drop in plate collector 0 0 6 The options are Drop in plate collector the IP is ejected into the plate collector after the IP has been scanned Leave in plate carrier the IP remains in the plate carrier after the IP has been scanned The Leave in plate carrier option is recom mended for users who want to handle the IPs with more care and reduce wear and tear on them This option extends IP service life and al lows greater hygiene standards to be observed B 13 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out
15. system Imaging plate unit 3 PC To release the unit click the GREEN multi connect icon and then click the Free the scan ner unit button Free the scanner 3 Removing IPs from the plate collector When removing IPs from the plate collector hold them by their edges Alternatively pull the plate col lector out of the unit and tip the IPs out onto a flat clean surface If they are not to be reused immedi ately or within a short period of time store them in their storage box 3 8 Standby mode If no IP is inserted into the unit within a certain pe riod of time the unit will beep several times and the status light will start to flash When status light turns yellow and the unit display turns off the unit is in the standby mode The unit will automatically exit the standby mode as soon as a new imaging plate is inserted into the unit or when the Start key is pressed After four hours the unit will exit the standby mode and automatically shut down User s manual 204253 23 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system 3 9 Retrieve last image If the last image read is not transferred to the PC because of a network communication PC or soft ware failure the last image read can be retrieved IMPORTANT NOTE The LAS
16. the unit that are not part of the supplied system not supplied by the manufacturer of the unit not recommended by the manufacturer of the unit User s manual 204253 3 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 2 Intraoral imaging plate unit 2 Intraoral imaging plate unit 2 1 Main parts and controls Power on off key and status light O green on flashing entering the standby mode yellow in the standby mode ae mo a IL O Display H P Start key Touchless operation sensor Unit door and plate carrier Imaging plate collector Rear of unit Power supply connector for the power supply unit PSU N BH CAUTION Only use the PSU supplied with the unit or an approved spare PSU supplied by an authorized distributor S Refer to technical specifications RJ 45 connector for Ethernet cable direct connection to PC network LAN connection Ethernet cable not included Refer to technical specifications PC not included with digital imaging software that conforms to the MDD 4 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this d
17. 1 Cleaning the unit Use a non abrasive cloth moistened with either cool or lukewarm water soapy water mild detergent butylalcohol or ethanol ethyl alcohol 70 96 to clean the unit After cleaning wipe the unit with a non abrasive cloth moistened with water Never use solvents or abrasive cleaners to clean the unit Never use unfamiliar or untested cleaning agents If you are not sure what the cleaning agent contains DO NOT use it lf you use a spray cleaning agent DO NOT spray it directly into the unit door 5 2 Disinfecting the unit CAUTION Wear gloves and other protective clothing when disinfecting the unit Wipe the unit with a cloth dampened with a suitable disinfectant solution such as ethanol 96 Never use abrasive corrosive or solvent disinfectants All surfaces must be dried before the unit is used WARNING Do not use any disinfecting sprays as the vapor could ignite and cause injury Disinfecting techniques for both the unit and the room where the unit is used must comply with all local and national regulations and laws concerning such equipment and its location 32 User manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 5 Unit care and maintenance Imaging plate unit 5 3 Maintenance The unit does not require an
18. 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 A Technical Specifications Imaging plate unit A 3 Main dimensions 190 User s manual 204253 A 3 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit A Technical Specifications Guidance and manufacturer s declaration electromagnetic emissions The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions The DXR 50 uses RF energy only for its internal CISPR 11 function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The DXR 50 is suitable for use in all establishments CISPR 11 including domestic establishments and those directly connected to the public low voltage power supply MARONIE Class A network that supplies buildings used for domestic emissions purposes IEC 61000 3 2 l Voltage Complies fluctuations flicker emissions IEC 61000 3 3 A 4 User s manual 204253 Reviewed Taskine
19. ENGLISH DIGORA Optime DXR 50 001 Digital intraoral imaging plate system User s Manual Document no 204253 rev 6 2012 05 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 DIGORA Optime DXR 50 001 Digital intraoral imaging plate system User s Manual CE 0537 This unit is marked according to the Medical Device Directive 93 42 EEC if the unit contains the CE mark Document number 204253 rev 6 2012 05 Original approved English language version Manufactured by SOREDEX P O BOX 148 FIN 04301 TUUSULA FINLAND Tel 358 0 10 270 2000 Fax 358 9 701 5261 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit SOREDEX endeavours to produce product documentation that is accurate and up to date However our policy of continual product development may result in changes to products that are not reflected in the product documenta tion Therefore this document should not be regard
20. Imaging plate unit 3 Using the system 3 6 Reading an imaging plate Single user configuration NOTE If the unit is operated in a multiconnect configura tion please refer to Multiconnect configuration first to reserve the scanner NOTE If you are taking an occlusal projection exposure refer to section 3 12 Taking an occlusal projection image and to the separate instructions supplied with the Comfort Occlusal 4C kit CAUTION For optimum image quality the exposed IP must be read as soon as possible after exposure If the ex posed IP is not read immediately after exposure but left unread for a period of time it must be protected from ambient light and X rays as both may degrade the image quality 1 Pull the flap to open the hygiene bag 2 Keep the IP in the protective cover so that you do not touch the IP or allow it to be exposed to ambient light and then slide the protective cover and IP out of the hygiene bag 18 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 Hold the protective cover and IP so that the white side of the protective cover is on the right The metal disk on the IP will also be on the right Insert the protective cover and IP into the unit door A magnet will hold th
21. ODUONS sca cecaecocenseinysnccanandanuadindensssdeawuennniesctwitstseueedensietecenisaieaeecons 14 FP OW EF ODOM S seaccasbrnsanteanestezaneciataanteententoncianctaewarertentanit dnceaapiaesanerp ares 14 Boo rI eee oE E eee ee ee ee eee cee eee eer ees 15 Automatic UV GISIMIOCHON aie sacanecsrsceunenichamoecnatenivacerenrrnedeveurtaeeeteseennns 15 Manual UV disinfection control scsecurseericetraccecasomesenmaatene enisunmammadacinanaen 16 Bie hols 6 6 0 e E AE E E EEE EE 17 CGS A ZE S a EA E E A E EEE 17 B4 TE OUDIS SIMO OWING so cendxceenctrietecnntesoncsavawastanescanseseensaseceucqsacesecseeeceacee 18 User s manual 204253 VII Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit VIII User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 1 Introduction Imaging plate unit 1 The DIGORA Optime 1 1 Introduction This manual describes how to use the DIGORA Optime digital imaging plate unit the unit which is part of the digital intraoral imaging plate system the system The complete system comprises the fol lowing the DIGORA Optime digital imaging plate unit the unit
22. PC Switch on the PC connected to the unit 2 PC Open the dental imaging software and a new or existing patient card where you wish to store the intraoral images O 3 Press the on off key to switch the unit on User s manual 204253 11 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system The startup animation will appear on the display and the unit will carry out self test during which the IP carrier will slide out of the unit door Startup gt animation Name Serial number FW x xx CORE xxx IP XXX XXX XXX XXX When the status light turns green and ready ani mation indicating IP insertion appears on the unit display the unit is ready to use in the ready state NOTE If the ready animation does not appear check the system setup described in the installation instructions 12 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 3 Preparing an IP for exposure IMPORTANT NOTE If the IP is being used for the very first time or if it has not been used within the last 24 hours it must be erased befo
23. T read image can only be retrieved if the unit is left on If the unit is switched off the image will be lost To retrieve the last scanned image 1 Correct the problem that caused the communi cation failure When the connection between the unit and the PC is re established the last read image will automatically be transferred to the PC 2 PC If the image is not automatically transferred to the PC select the Setup gt Scanner page from the dental imaging software your are using 3 PC In the Last Image field click the Retrieve now button to retrieve the last read image NOTE If required you can select different parameters e g resolution show image preview etc for the image to be retrieved 4 PC Click OK to close the Setup window The last read image will be transferred to the PC 24 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 10 Shutting down the unit O 1 Press and hold down the power on off key The UV disinfection cycle when active refer to section Installation and set up hygiene will be automatically carried out before the unit switches itself off If you wish to switch the unit off before the UV disinfection cycle is complete press and hold down the power on off
24. aloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system Multiconnect configuration 1 PC To reserve the unit click the GREEN multi connect icon which is in the bottom right hand corner of the PC display The Connect window will appear 2 PC In the Connect window click the Reserve the Scanner button to reserve scanner Reserve the scanner The window will disappear and the GREEN multi connect icon will start to flash This indi cates that you have now reserved the unit and it can now be used to read an image NOTE If the multiconnect icon is YELLOW it indicates that the unit has been reserved by someone else Wait until the unit is free If you wish to know who has reserved the unit place the cursor on the multiconnect icon while yellow and the message Scanner reserved by XXXX where XXXX is the multiconnect ID of the PC that has reserved the unit will appear NOTE If the multiconnect icon is GRAY it indicates that the unit has not been switched on or is just starting up and is not yet ready for use Switch the unit on and or wait until the unit is ready for use 22 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the
25. and setup Imaging plate unit Disable the user can switch off UV disinfection so that it does not operate Also all warning messages related to UV lamp status are dis carded NOTE lf any automatic UV disinfection mode is select ed a UV disinfection cycle will ALWAYS be au tomatically carried out when the unit is switched off and prior to final unit shutdown During this shutdown UV disinfection cycle the remaining disinfection time will appear on the unit display If you wish to stop the UV shutdown disinfection cycle off key the unit will immediately shutdown or O Press and hold down 4 seconds the start key Press and hold down 4 seconds the power on the unit will return to the normal operation mode Manual UV disinfection control Manual UY disinfection contral Disinfect Click the Disinfect key to manually start and stop Cancel UV disinfection The elapsed disinfection time will appear on the unit display UV disinfection will automatically stop after 15 minutes If you wish to stop manual UV disinfection before the complete cycle is finished press the Manual UV disinfection control Cancel key or press and O hold down 4 seconds the start key Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and
26. bite protector Insert the Comfort Occlusal bite protector and IPs into the Comfort Occlusal hygiene bag and seal the bag 4 Place the sealed Comfort Occlusal 4C IP into the patient s mouth and take an exposure 5 Remove the sealed Comfort Occlusal 4C IP from the patient s mouth 6 Open the sealed Comfort Occlusal 4C IP Re move the two size 3 IPs and scan them User s manual 204253 27 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system 3 13 Errors and warnings e lf there is a problem with the unit the error state symbol and an error number will appear on the A N unit display Error Restart the unit The error should clear If not contact your authorized distributor for assist ance e Warning UV lamp blown NOTE The unit can still to be used if the UV disinfect O ing lamp has blown Press the start key to con tinue using the unit If you wish to disable UV disinfection completely see section B 3 5 Hygiene Automatic UV dis infection Disinfection frequency e PC If the message Degraded image quality appears on the PC display during or after an IP is read it indicates that the image may not be within the factory set quality limits If the image looks okay no action is necessary If the message appears f
27. cturers IP storage box For storing IPs safely and conveniently Microfiber cloth For cleaning IPs IP holders For bitewing periapical and endodontic exposures optional not included See section Imaging plate holders for more infor mation 6 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 2 Intraoral imaging plate unit Imaging plate unit 2 3 Display symbols and what they mean During use symbols and animations will appear on the unit display These indicate the status of the unit help you to operate the unit correctly show user mistakes and corrective actions display error codes display a preview image The main symbols are Startup During startup the unit serial number IP address Startup and other information will appear on the unit display animation Name Serial number FW x xx CORE xxx IP XXX XXX XXX XXX Patient name Indicates a patient card is open The patient s name is shown on the unit display Multiconnect wait Multiconnect configuration The unit is not reserved by any PC in the system Multiconnect reservation Multiconnect configuration The unit has been re served by a PC e g PC number 2 UV disinfection in progress The system is carrying out automatic UV disinfec tion cycle Remaini
28. e IP in the correct position Slide the protective cover off of the IP and leave the IP in position half way in the unit door As soon as the protective cover is removed the 4 4 unit detects that the IP is in the unit door and will switch from the standby state to the ready state ready to use and automatically slide the IP into the unit CAUTION If the IP does not slide into the unit after the protective cover has been removed the IP is misaligned or has been placed in the unit the wrong way round Reposition the IP and insert it into the unit again into the unit press the Start key to manually switch the unit from the standby to the ready state and the IP will then slide into the unit when the protective cover is removed O If after repositioning the IP it still does not slide User s manual 204253 19 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system The busy animation will appear on the display Progress which indicates that the imaging plate is being r a read animation After few seconds a preview image will appear on the unit display PC Aread out progress window will appear on the PC display After a few seconds the image will appear in the dental imaging software The image can now be saved Refer to
29. e black side of the hygiene bag i am 5 Peel off the cover paper from the sealing tape and then fold the flap along the pre formed line over and onto the sealing tape 6 Press and slide your thumb along the tape to ensure that the flap is properly sealed 7 Turn the sealed hygiene bag over and check that the IP and protective cover are in the cor rect position You must be able to see the light side of the pro tective cover and the metal disk on the IP 14 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 4 Imaging plate holders It is recommended that imaging plate holders be used to ensure accurate IP positing and consistently good images quality Using imaging plate holders improves image quality because the IP is positioned correctly in relation to the tooth there is no positioning guesswork the IP is not bent and thus distortion is eliminated the IP cannot move in relation to the X ray unit images are standardized and reproducible there is no overlapping nor cone cut off IP wear and tear is minimized image quality can be maintained irrespective of who takes the image time is saved and profitability increased Problems caused by manually positioning the IP incl
30. ed as an infallible guide to current product specifications SOREDEX maintains the right to make changes and alterations without prior notice IV User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit Contents 1 The DIGORA Optime 2 0 0 0 cece eecccceeeeceseeceeeeteeeeeeeeseneneeseneetaneeeas 1 Mee PEE OCU WONG as se A EE EA E E ES success T E 1 1 2 System MsStalallON opesanescosccecateneesacaaevss saeecaceresnqesesoiet EEEE n 2 FPOSIUONING ME UNE aoigo castes eoine geceisseceencees esas eE EREET EE 2 FPOSIMOMMNG Me PU serer era E E E E E 3 AMET OC CSS era a E E E E 3 2 Intraoral imaging plate unit ee saceee sees etederddatceute sca aces euesessseatsase 4 2 1 Main parts ANG controls scicncisavseevecsvenendonnaonnnntenswantiondsnsnarsennleteetonebiepsyeeedosvswents 4 ZZ ACCC SS OMNES sasaccsicecacnaetosais rE NE EEEE E EREA EEEN E OR Ee E OEO 5 2 3 Display symbols and what they mean cccecceecceeeeeeeeeeeeeeseeeseeseeeeeeees T 3 Using the system ccccccceccseecaeeceeceeesaeeceeceeecaeeceeseuecaeeseesaeesaeeeas 11 Se 1 CCI A PE E A E 11 3 2 Preparing Ne SyS lOi sesssr rreren T EAEE ERER ES EEE EERTE SE 11 3 3 Preparing an IP for 4 96 0 n oe ee 13 3 4 imaging piate TIONG CIS srssescssrriroriresis dr
31. ers or hygiene bags from other manufacturers e This unit or its accessories must not be modified altered or remanufactured in any way e Only the manufacturer s authorized service personnel are authorized to carry out maintenance and repair of the unit There are no user serviceable parts inside the unit e Infection control procedures must be observed when using accessories such as film holders x ray tube guides and imaging plates When using accessories always follow the manufacturer s instructions on how to use the accessory and prevent cross contamination from one patient to another e This unit can interfere with other devices due to its EMC characteristics e Other devices can interfere with this unit due to their EMC characteristics e This unit complies with IEC 60601 1 standard Accessory equipment connect ed to this device must be in compliance with the related nationally harmonized IEC standards User s manual 204253 35 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 7 Warnings and precautions e Unit not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide e For ethernet connections use an unshielded CAT6 LAN cable so that multiple chassis must not be connected The PC Ethernet sw
32. for 5 sec Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If user of the DXR 50 requires continued operation during power mains interruptions it is recommended that the DXR 50 be powered from an uninterruptible power supply or a battery Power frequency magnetic field should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE Uzis the AC mains voltage prior to application of the test level User s manual 204253 A 5 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit A Technical Specifications Guidance and manufacturer s declaration electromagnetic immunity The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Immunity IEC 60601 Compliance Electromagnetic environment guidance test test level level Co
33. h between the unit and PC or network socket PROBLEM Ethernet link not active The Ethernet connection consists of correctly in stalled cabling and any switches and or hubs used basically the complete Ethernet link If the link is active it does not necessarily mean that the unit is physically connected to PC It only means that the unit is physically connected to something that is Ethernet comp atible hub switch another PC etc B 21 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup CAUSE SOLUTION First remove and then reconnect the Ethernet cable between the unit and PC to see if this solves the problem Physical connection is okay link active but the the Dental imaging software animation flash ing yellow appears on the unit display The unit is not configured correctly Reconfigure li Cabling not correct link not active and the Unit connection animation appears on the unit display The unit is not connected to the PC Connect the unit to the PC ili The PCs Ethernet connector s is faulty There is usually a green link LED near the PCs Ether net connector s on the NIC If the LED is not on replace the NIC PROBLEM The unit display is not active but the power on off status light is on green
34. isted above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people User s manual 204253 A 7 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit A Technical Specifications A 8 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit Appendix B Installation and setup Installation and setup must only be carried out by service personnel trained and approved by the manufacturer of the unit B1 System installation B1 1 Positioning the unit Position the unit on a stable flat surface that is not effected by vibrations that could degrade the image quality The unit can also be attached to a
35. itch to which unit is con nected to should be approved appropriately e g EN 60950 IEC 60950 UL 60950 After installation check that the IEC 60601 1 leakage current levels are not exceeded e In order to maintain safe and correct functioning of the unit only the power supply unit PSU delivered with the unit or distributed by authorized dealers Please refer to the unit technical specifications for a list of the PSUs e If this device will be used with 3rd party imaging application software not sup plied by the manufacturer the 3rd party imaging application software must comply with all local laws on patient information software This includes for example the Medical Device Directive 93 42 EEC and or FDA if applicable e Medical electrical equipment needs special precautions regarding EMC and needs to be installed according to EMC information 36 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 A Technical Specifications Imaging plate unit Appendix A Technical Specifications A 1 Unit Intraoral imaging plate unit DXR 50 Classification Class 1 or 2 equipment depending on the classification IEC60601 1 of the PSU No applied part Continuous operation IPXO enclosed equipment without protection against ingress of liquids Laser Safety Clas
36. key again 4 seconds and the unit will shutdown If you wish to return to the normal operation mode press and hold down 4 seconds the start key NOTE If there is an untransferred image in the unit s memory the unit cannot be shut down The im age must be transferred first Refer to section 3 9 Retrieve last image 3 11 IP erasing mode Initial erasing of the IPs If an IP is being used for the very first time or has not been used for 24 hours and has not been ex posed to ambient light the erasing procedure must be carried out before the IP is used The erasing procedure removes any fogging due to background radiation The dedicated erasing mode erases the IP but does not send the image blank to the dental imaging software User s manual 204253 25 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system Press and hold down the Start key for several seconds until the erase mode animation ap pears on the unit display The unit is now in the erase mode Hold the IP by its edges or use a protective cover and position it so that the side with the metal disk the back is on the right Insert the IP into the unit door remove the pro tective cover if used A magnet will hold the IP in position and then automatically slide it
37. lates protective covers and hygiene bags supplied by the manufacturer of the unit Please read the section 7 Warnings and preca tions before using the unit 1 2 System installation Positioning the unit Do not position the unit in direct sunlight or near bright light Sunlight or bright light must not be al lowed to shine directly on the unit door into which the IPs are inserted Position the unit on a stable flat surface so that vi brations will not degrade the image quality The unit can also be attached to a wall under or on a shelf with the optional mounting kit The unit must not be positioned so that it it touching other equipment It must not be placed on top of or under other equipment The unit can be positioned within the environment in which the patient is examined and treated patient environment 2 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 1 Introduction Imaging plate unit Positioning the PC The PC connected to the unit should not be used in the patient environment The minimum horizontal distance between the pa tient and the PC is 1 5 m 4 5 ft The minimum vertical distance between the patient and the PC is 2 5 m 6 5 ft Other devices DO NOT connect any other devices to the unit or the PC connected to
38. mination maintain image quality maximize IP service life User s manual 204253 29 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 4 3 Cleaning 4 Handling and care of imaging plates If IPs are handled and stored correctly cleaning should not be necessary or can be kept to an abso lute minimum Clean the IP if there you see any visible marks or stains which are not obvious scratches on the im age or if there is a reason to believe that the IP is contaminated 4 4 Storage 30 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 Use the microfibre cloth supplied First very gently wipe the IP in a backwards forwards movement in both the widthwise and lengthwise directions and then finish with a cir cular wiping movement For stubborn marks and stains that cannot be be removed using the microfibre cloth use 70 96 ethanol 70 96 EtOH 30 4 DI WATER or ethanol anhydride Wipe the IP as described above and then wipe dry NEVER USE abrasive chemicals or materials to clean IPs Unsuitable cleaning solutions methods may damage or destroy the IPs or leave residues on the sensitive surface that may appear on the images Unpacked exposed to ambient light in the dedi cated storage box
39. n Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 A Technical Specifications Imaging plate unit Guidance and manufacturer s declaration electromagnetic immunity The DXR 50 is intended for use in the electromagnetic environment specified below The customer or the user of the DXR 50 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 Electrical fast transients bursts IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply lines IEC 61000 4 11 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode lt 5 Ur gt 95 dip in U for 0 5 cycle 40 Ur 60 dip in U 7 for 5 cycles 70 Ur 30 dip in Us for 25 cycles lt 5 U1 gt 95 dip in U for 5 sec 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode lt 5 Ur gt 95 dip in Us for 0 5 cycle 40 Ur 60 dip in U7 for 5 cycles 70 Ur 30 dip in Us for 25 cycles lt 5 U1 gt 95 dip in U
40. nducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 MHz to 2 5 GHz Portable and mobile RF communications equipment should be used no closer to any part of the DXR 50 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol R NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic
41. ng cycle time will be displayed Comfort Occlusal 4C mode system in occlusal projection imaging mode User s manual 204253 7 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 2 Intraoral imaging plate unit Comfort Occlusal 4C mode System in occlusal projection imaging mode insert second size 3 IP Unit door Protective cover and imaging plate Yellow remove protective cover J Imaging plate Yellow wrong way round rotate a ee ee Protective cover Remove disconnect aja Insert connect User s manual 204253 0 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 2 Intraoral imaging plate unit Imaging plate unit Busy Progress Unit in operation animation UV disinfection Pro aie Unit in UV disinfection mode the remaining elapsed animation disinfection time is shown blue Unit in erasing mode Check Something wrong or take alternative action Pa gt Dental imaging software Software not open not ready or waiting for user ac tion Unit connection Not connected or connection
42. not working User s manual 204253 9 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 2 Intraoral imaging plate unit Rotate Error state and error number gt 2 Error Check documentation supplied with the unit Unit in service mode Service technicians only 10 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit 3 Using the system 3 1 General For optimum performance only use IPs protective covers and hygiene bags designed for this unit and supplied by authorized distributors The manufacturer of this unit will not be held re sponsible for any problems caused by using acces sories from other manufacturers Proper handling cleaning and storage of the IPs ensures the best image quality and maximum serv ice life of the IPs Refer to section 4 Handling and care of imaging plates IMPORTANT NOTE If you are using the system for the first time check the setup options to ensure that they are suitable for your work methods and dental environment See section Appendix B3 Setup options 3 2 Preparing the system 1
43. ocument Printed out 28 08 12 13 43 18 D504182 2 Intraoral imaging plate unit Imaging plate unit 2 2 Accessories For additional information about the accessories listed below contact your authorized dealer Not all accessories are available separately for all units Imaging plates IPs Equivalent to film sizes 0 1 LN IDOT version IDOT IPs optional have individual identification codes that appear on the images IDOT aids quality assurance by making it possible to identify the plates used to take individual images Standard STD version The standard version has no identification mark on the active side of the plate Protective covers For IPs 0 1 2 and 3 5 3 Hygiene bags For IPs 0 1 2 3 and Comfort Oc clusal 4C User s manual 204253 5 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 2 Intraoral imaging plate unit Comfort Occlusal 4C bite protector For holding two size 3 IPs that are used for taking an occlusal projection image CAUTION For optimum performance only use IPs protective covers and hygiene bags supplied by the manufac turer of the unit or the manufacturer s authorized distributors The manufacturer will not be held responsible for problems caused by using accessories from other manufa
44. operation sensor Preview Time fo 0 60 Plate eject mode Drop in plate collector Restore factory settings Resets user configurable settings to their factory set values Restore factory settings B 11 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup Readout start Readout start Automatic Start after Start delay Plate insert Y Shor E Manual Start readout by pressing unit s START key Select Automatic if you want the scanner to start automatically The Start after options allow to to select when the scanner will start after Plate insert the scanner will automatically start when it detects and IP in the plate carrier after Cover removal after the IP and protective cover have been inserted into the plate carrier the unit will automatically start when the protec tive cover is removed The Start delay options allow the the start delay time to be selected Short approximately 0 2 seconds Medium approximately 0 4 seconds Long approximately 0 6 seconds Select Manual if you want the unit to start reading an IP only when the Start key is pressed O NOTE If Manual is selected when the unit is also set up to use the occlusal mode See B3 6 Occlusal single
45. pproved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B1 System installato wxccaaus onsectewensdeanetnmeonaceeunaseeidpeunsentueeneapatmestaceenon 1 B1 1 Positioning Ne UN ssiru ne ee ee eer EEEE 1 B1 2 Positioning the PC s not supplied icc ccccwecceatcecedeesssase ene venice oueteestaeeedes 1 Bko ONE dov COS oea ee ee eee 2 B2 Connecting the unit to a PC LAN cece eee eeeeeeseeeeeeteeeeeeees 2 B2 1 Direct connection method uses the unit S N cece eecceeeeeeeeeeeeeeees 2 B2 2 IP method using the unit IP addreSs cc ccccceccceccsececeeeseeeseeeeaeeesaees 4 B2 MUC SOMME CE sax ecccacatennieucncreeasannonentiancasisaadenpeasiuaaasncayiaganaeepnaendmeeraaneenonen 6 B3 Setup Options ccccceccseeeeeeeeeeeseeeeeteeeeeeeeeeeeeteeeeueteeteeeseeeeeteeees 8 BEMESSEN EE E E P E A scenes A E E E E 8 SLUG ee a E e E E EEEE EEE EE EE EA E e AEE A 9 ae oC aN O eea EE A EREE EEE EE 9 Using the CSN AIG AL essiri aa a 9 BS ONO eaer A E E T anes 10 MaS FTO SS O eea SEEE 10 EASL 0612 6 e eee ee eee ee eae mere eee eee eee 10 Scanner Unit Serial NUMDET ceeraceceeesfucicavicaredeeeeensnaraansentenetdonstaneinesaatinedes 10 Bo Zoo e E sta EEEE E E 11 gt VV OF MOV eirean E E E AE E E E e AREE 11 PREGQCOUl SIGINT eera EE E R 12 OUGIMN CSS OPETI eeraa RE E E 13 ae CA IM OGG E 13 Bate OWED
46. re use to remove any fogging caused by background radiation See section 3 11 IP eras ing mode Initial erasing of the IPs CAUTION When handling IPs protective covers and hygiene bags take all appropriate measures and precautions to prevent cross contamination NOTE If you are taking an occlusal projection exposure refer to section 3 12 Taking an occlusal projection image and to the separate instructions supplied with the Comfort Occlusal 4C kit Place the IP you wish to use onto a protective cover The light blue side sensitive of the IP must face and be placed on the half of the pro tective cover that is the same shape as the IP Fold the half of the protective cover with the semicircular cutout over onto the IP The metal disk on the back black side of the IP must ap pear in the semicircular cut out Turn the protective cover and IP over so that the black side of the protective cover is uppermost This makes it easier to slide the protective cover and IP into the hygiene bag User s manual 204253 13 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system 4 Slide the protective cover and IP into the hy giene bag as far as they will go Make sure that the black side of the protective cover is on the same side as th
47. requently contact your authorized distributor for assistance 28 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 4 Handling and care of imaging plates Imaging plate unit 4 Handling and care of imaging plates The correct use handling cleaning and storage of imaging plates guarantees the best image quality and maximum service life of the imaging plates 4 1 General e X ray exposures DO NOT cause IPs to age e The light coloured side of the IP is the side which stores the x ray image It is the SENSI TIVE side e Image information stored on an IP after an exposure is sensitive to light 4 2 Handling e Handle IPs in the same way as you would han dle CDs or DVDs e Hold IPs by their edges using your finger tips e Protect the sensitive side of the IP from scratches wear and mechanical damage Stains dirt dust and fingerprints NOTE Stains fingerprints on the sensitive side can degrade the diagnostic value of the image Dust wear on the sensitive side can appear as white grey spots on the image e Unprotected IPs must not come in contact with the patient the patient s saliva or any other bod ily fluids Always use hygiene bags and protective covers with IPs when taking exposures to eliminate cross conta
48. ri Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit If the exposure time is too short images will be noisy Such images may still usable for some diagnostic purposes lf the exposure time is too long images will be too dark or will show patient movement These images will not be good enough for accurate diagnostic examination The exposure values table below should only be used as a guide Recommended exposure values s for DC x ray units 60 kV 7 mA 70 kV 7 mA short cone short cone long cone Adult Child Adult Child Adult Child Adult Child Bitewing Maxillary incisor Maxillary cuspid 0 20 Maxillary molar Occlusal Mandibular incisor 0 20 Mandibular cuspid 0 25 Mandibular molar 0 25 For AC x ray units increase the exposure times by about 30 3 Protect yourself from radiation and take the exposure 4 After the exposure remove the sealed hygiene bag from the patient s mouth CAUTION If there is a risk of cross contamination wash disinfect and dry the hygiene bag before open ing it User s manual 204253 17 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182
49. sification CLASS 1 LASER PRODUCT EN 60825 1 2007 Dimensions H x W x D 190 mm x 200 mm x 383mm 7 5in x 7 8in x 15 1in 9 8 10 3 kg 21 6 22 7 Ib Power supply unit PSU POWERBOX EMX 805121 24 VDC PSU 100 240 VAC 50 60 Hz Less than 1 5 A 35 um Super 64 um High 14 bits grayscale Interface cable For Ethernet connections use an unshielded CAT6 LAN cable so that multiple chassis must not be connected The PC Ethernet switch to which unit is connected to should be approved appropriately e g EN 60950 IEC 60950 UL 60950 Operating environment 10 C 40 C 30 90 RH 700 1060 mbar Storage transportation 10 C 50 C 0 90 RH 500 1080 mbar environment User s manual 204253 A 1 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit A Technical Specifications A 2 Imaging Plates and hygiene bags Imaging plates IP D imensions mm 22 x 31 24 x 40 31 x 41 27x54 48 x 54 nominal Image size 628 x 885 685 x 1143 886 x 1171 771 x 1542 1372 x 1542 pixels 35 ym nominal Image size KB 1085 1529 2026 2322 4000 35 um nominal Image size 484 x 344 625 x 375 641 x484 844x422 750 x 843 pixels 64 ym nominal Image size KB 325 458 695 1150 64 um nominal Storage Imaging pla
50. site survey should be considered If the measured field strength in the location in which the DXR 50 is used exceeds the applicable RF compliance level above the DXR 50 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting of relocating the DXR 50 P Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m A 6 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 A Technical Specifications Imaging plate unit Recommended separation distances between portable and mobile RF communications equipment and the DXR 50 The DXR 50 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the DXR 50 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the DXR 50 as recommended below according to the maximum output power of the communications equipment Rated maximum Separation distance according to frequency of transmitter m output power of transmitter W 150 kHz to 80 MHz 800 MHz to 2 5 GHz d 1 2 P d 1 2 P d 2 3 P For transmitters rated at a maximum output power not l
51. solution but more memory is required as the image files are larger High default gives a pixel size of 64u This results in images with less noise especially if short expo sure times are used Image Processing Image Processing Noise Filtering normally selected makes images smoother when they are taken at low radiation doses Last image Last Image Retrieve latest image from the scanner unit memory Aetieve now Refer to section 3 9 Retrieve last image Scanner Unit Serial number Scanner Unit Serial Humber Add serial number to new images Adds the unit serial number to all new images B 10 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit B3 2 Settings See B2 Connecting the unit to a PC LAN for information on Settings B3 3 Workflow 1 From the dental imaging software you are using select Setup Workflow page Scanner Settings Workdlow Power options Hygiene Occlusal Driver Info Readout start i Automatic Start after Start delay Plate insert T Short C Manual Start readout by pressing unit s START key Touchless operation Scanner will wake up automatically from standby mode if touchless operation sensor is in use and activated M Use touchless
52. tes must be stored in their protective box below 33 C environment The box must be kept closed to remain dust free Photo stimulable phosphorous material uniformly coated ona support plastic material Shielded with a top coat layer on the active surface and encapsulated with lacquer around edges Imaging plates do not include phosphor no The typical service life for an imaging plate is several hundreds of cycles provided that the imaging plate is handled with care and according to the supplied instructions The use of genuine hygiene accessories protective covers and hygiene bags will extend the service life of the imaging plates Disposal Imaging plates are industrial waste and must be disposed of in accordance with local and national regulations concerning the disposal of such material Never use damaged imaging plates Hygiene bags Material Food grade polyethylene Latex free Packaging Supplied in boxes For the best performance it is recommended the hygiene bags are used within two years from the date of manufacture The date of manufacture is printed on the bottom of the box containing the hygiene bags DDMMYYxXxX Extended storage time or exceeding the specified storage conditions may compromise the performance of the adhesive tape and or the plastic material from which the hygiene bags are made Disposal Observe relevant national requirements A 2 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12
53. the documentation supplied with the dental imaging software you are using CAUTION If the metal disk on the rear of the IP can be seen on the digital image it indicates that the IP was exposed from the wrong side CAUTION RETRIEVING IMAGES If the image is not transferred to the PC because of a network PC or software failure the image can be retrieved from the units memory as long as the unit is NOT switched off For information on how to do this see section 3 9 Retrieve last image 5 After the IP has been read it will be automati cally erased and then ejected from the unit into the plate collector 20 User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 3 Using the system Imaging plate unit NOTE The unit can be configured so that the IP re mains in the plate carrier it is not ejected into the plate collector When the IP is in the plate carrier the IP is easier to handle and stricter hygiene measures can be observed Refer to Appendix B 3 3 Workflow Plate ejec tion mode After the IP has been ejected or removed the ready animation will reappear on the display The unit is now ready to read the next IP User s manual 204253 21 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the P
54. tion C Direct Conmectior Intell Al PAO O00 MTW Network Connection 6 PC Select Direct Connection Scanner Settings Driver Info Scanner Connection Direct Connection settings Scanner senal number g nannan Computer network connection Intel R PAO TOO0 MT Network Connection f Direct Connector B 3 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup Key the serial number of the unit into the Scan ner serial number field The serial number of the unit will appear on the unit display when the unit is switched on It can also be found on the type label on the back of the unit Make sure that the Computer network connec tion that provides the LAN network connection is selected B2 2 IP method using the unit IP address If your system does not allow the direct connection method to be used to connect the PC s connection can be done using an IP address 1 Follow steps 1 to 5 from the previous section B2 1 Direct connection method uses the unit s n 2 PC From the Settings tab select IP based and then select the Enable changing IP address box Scanner Settings Driver Info Scanner Connection Direct Connection settings C Direct Connectior Scanner senal number
55. ude incorrect vertical alignment distortion cone cut off poor projection standardization inferior image quality For more information on imaging plate holders and systems contact you dealer User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 15 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 3 Using the system 3 5 Taking an exposure 1 Place the IP in its sealed hygiene bag into the appropriate imaging plate holder and then insert it into the patient s mouth in the position for the image you wish to take Note that the back of the sealed hygiene bag the black side must face the X ray source Wu 2 Select exposure values appropriate for the ex posure you are taking The system will produce excellent images even if the exposure values differ considerably from the optimum values In most cases the same exposure values can be used for virtually all imaging purposes For normal everyday use select the Adult Bitewing exposure time from the following table If required the exposure time can be increased for very large patients and reduced for children The optimum exposure values also depends on the performance of the X ray unit being used and may vary by 1 step from the values in the following table 16 User s manual 204253 Reviewed Taskinen Ja
56. un nn eE AE EEr R EENE AEE NEE REES 15 OLD WARING aN EXPOS U O serer E EEEE 16 3 6 Reading an imaging plate c ec ccecceecceeeseeceeeceeeseeeeeeeeeseeeseeseeeseeeaees 18 Single user configuration cc cece eecceececeeeceeeceeeceuecaueceeeeseeseeseeesaeeseas 18 Multiconnect configuration nnannannannnnnennnnnnrnnnrnnnnnnnnrrnnrnnrrnrerrrnrrnrenni 22 3 7 Removing IPs from the plate COleCtOr ccc cceeccceceseeeceeeceeeeeeeaeeeseees 23 20 Andoy MOI ence eE E EE E S 23 3 9 FROLIC VG lasi WAG O erasini tene Erea EE EEEa EA EEE EEE EENE EN EEEE 24 3 10 SMUG down Ne UNIL ccsncscocseeeespscecacnee lt sccenctenecaacnesxeetcaaemesensaceceasmentenene 25 3 11 IP erasing mode Initial erasing of the IPS cece eecceeeeeeeeaeeeeeees 25 3 12 Taking an occlusal projection IMAGE cccccceceseeceeceeeeeeeaeeeeeeseeeeeeeaes 2 3 19 Errors and WaininNgS ctecisee aac sete necceegeas eiu annn SEE EEREN E ETE ETEEN 28 4 Handling and care of imaging plates cc ccecc cece eeeceeeeeeeeeeees 29 4 1 CMON AA sarc ncucoseenimennescenseneineieenuneademaie pelganiiatentauonesnanonsiaoy a meetnnnudimespan ensue uicontdnsreveteelen 29 AZ TVA VON e O E SE EE E E A E T 29 4 3 Cleaning ecieces snes secrsenty airetnnna dienes signee toneusensweeatea atnesieasansnaeeonatcwesnerennaneacesaets 30 Ah As STO FAS ct enc A EA coe mnbe ces vaste E A 30 AD REDIACEMENMI scacszxccsadebonsnanancexs cdnananonetansiesesecentya
57. unanssandss acnezeusaotnnaneana EEEE 31 4 O 11S DOSE renare aE R EEE E AE EEEE 31 User s manual 204253 V Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 5 Unit care and TYEINIGINAINC ccs cei deriecees tieeinent cnccawnesectdencsdidadduerscacteasens 32 SES ARAN UV Nota cease ee E E ae siete a eee tase ate eee 32 9 2 Disinecting Me 9 srren ER E AERE EE eee eee 32 5 3 IVI CIN MN Octane ect cheer eeraea e ek ante bane mies ae EEEa EEE cade EN akr Eni 33 DARO ee ee E T eee eee 33 ISO SAM asacece seca sc ase oto edges E pieeeeuasanerrecssesces teeta 33 6 Symbols that appear on or in the UNit ec ceeeseeeeeeeeeeeeeeees 34 7 Warnings and PreCautiOns ccccccceecceeceeeceeceeeeeeseeseeeeeeseeeeeesees 35 Appendix A Technical Specifications cccccecceeceeceeeseeeeeeeeeeeeeaees 1 PN MTN ces cs sty ese cee otc sete dete tse Pe etc este pose eee ce ee nee pede ented eee 1 A 2 Imaging Plates and hygiene DAO Sissies ots cc erareceteteen Dasht vento cacea eietiasuerene 2 A 3 MaN CUS ING ONS toc ccc cece ates cd rentn satan scab ade sate bese Siasendoeeds ycubaeced EAEE 3 Appendix B Installation And setup ccccccccseeeeeeeeeeeeeeeeeeeeseeteeeaess 1 VI User s manual 204253 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 A
58. up Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit 2 Click OK to connect the PC to the unit NOTE An automatic technique will automatically locate the unit within the local area network and con nect the PC 3 Repeat the above process for all the other PCs in the network Make sure that you give each PC a different Workstation identifier 4 Check the installation by starting image capture using the imaging software If the Use Multi Connect was selected the Workstation identifier of the PC 1 8 being used will appear on the unit display B Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup B3 Setup options The scanner setup options allow you to configure the scanner to the user s clinical requirements B3 1 Scanner 1 From the dental imaging software you are using select Setup Scanner page Cannel Settings Workflow Power options Hygiene Occlusal Driver Info Status Scanner DIGORA Optime Y ETSIOn AU Seral No SJOS00022 Image Scanning Show Image Preview and Dental Chart Resolution f Super High Image Processing Last Image Retrieve latest image trom the scanner unit memory Retie
59. ve now Scanner Unit Serial Number Add serial number to new images B 8 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 B Installation and setup Imaging plate unit Status Status Scanner DOlGORA Optime VW ersiore Ald Senallo SJOS00022 Shows the scanner type version and serial number Image Scanning Image Scanning Show Image Preview and Dental Chart Resolution Super f High lf Show Image Preview and Dental Chart is se lected a preview image with a dental chart for tooth numbering will appear before the image is saved Using the dental chart 1 After an IP has been read a window will open that shows a preview image and a dental chart 2 Click the tooth teeth on the chart that cor respond to the tooth teeth in the image Tooth numbers will be assigned to the selected teeth The tools at the top of the window allow the im age to be manipulated 3 Click OK to save the image and tooth numbers Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup Resolution Super gives a pixel size of 35u This results in im ages with better re
60. wall under or on a shelf using the optional scanner mounting kit pt no 204138 Do not position the unit in direct sunlight or near bright light Sunlight or bright light must not be al lowed to shine directly on the unit door into which the IPs are inserted The unit must not be positioned so that it touching other equipment The unit must not be placed on top of or under other equipment The unit can be positioned within the environment in which the patient is examined and treated patient environment B1 2 Positioning the PC s not supplied The PC s connected to the unit should not be used in the patient environment The minimum horizontal distance between the pa tient and the PC s is 2 5 m 6 5 ft B 1 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved Aarynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit B Installation and setup B1 3 Other devices DO NOT connect any other devices to the unit or the PC s connected to the unit that are not part of the supplied system not supplied by the manufacturer of the unit not recommended by the manufacturer of the unit B2 Connecting the unit to a PC LAN The procedure for connecting the unit to a single PC or several PCs in a local area network LAN is exactly the same except that every PC in the LAN needs to be given a unique ID
61. y maintenance 5 4 Repair If the unit is damaged or malfunctions in any way it must only be repaired by service personnel author ized by the manufacturer of the unit 5 5 Disposal At the end of the useful working life of the unit and or its accessories make sure that you follow national and local regulations regarding the disposal of the unit its accessories parts and materials The unit includes some or all of the following parts that are made of or include materials that are non environ mentally friendly or hazardous electronic circuit boards electronic components UV lamp Hg User s manual 204253 33 Reviewed Taskinen Jari Pekka 29 05 12 11 37 01 Approved A rynen Teemu 29 05 12 14 52 44 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 28 08 12 13 43 18 D504182 Imaging plate unit 6 Symbols that appear on or in the unit 6 Symbols that appear on or in the unit DANGEROUS VOLTAGE LASER RADIATION ULTRAVIOLET RADIATION IMPORTANT INFORMATION Refer to user s manual P Dee Direct current Eternet connector RJ45 straight cable l C 4 CE 0537 Symbol MDD 93 42 EEC This unit is marked according to the Medical Device 0537 Directive 93 42 EEC if the unit contains the CE mark Cll ETL symbol C US GOST R certificate Certificate of Conformity T asserts that the goods delivered in Russia and or production line where they were manufactured

Download Pdf Manuals

image

Related Search

Related Contents

Salamandra Eléctrica Manual de Usuario  série economy 5040 manuel d`utilisateur  Grindmaster AT-2W User's Manual    Bose CD Player CD Player User Manual  Lirio by Philips Wall light 36922/32/LH  Shindaiwa 68002-94310 Owner's Manual  LC-16-1 E/V/G LC-20-1 E/V/G  Samsung GT-I9001 用戶手冊  Digital Blood Pressure Monitor Model UA  