Oxymax for Windows - Columbus Instruments

image

Contents

1. Group 1 Group 2 m Pump Valve p Sample Pump O C Testin h Dn Ref Cal pa C Off TNs r Pressure Regulator CO On C Off 0 000 r Source Pump H2 Ze On C Off r Expansion Interface Valve Select 12 3 4 5 AIR FLOW E at ee Ir 2 d H An m Pressure Set Default Lon PRESSURE PRESSURE 2 Notice that there are now two columns of gas sensor data in the Oxymax Utility where there was only one before The first column contains current readings from 103 the Oxygen Carbon Dioxide and Methane sensors The second contains identical copies of the Oxygen and Carbon Dioxide data but shows the Hydrogen Sensor in place of the Methane Sensor in this group 3 At this point the Oxygen Carbon Dioxide and Methane sensors have not been calibrated so their readings may be considerably off from room air 4 The first step to calibrating the Hydrogen Sensor is establishing the proper offset point For this we require a gas without Hydrogen in it Systems that employ Paramagnetic Oxygen Sensors are already equipped with a tank of pure Nitrogen which will provide a Hydrogen offset point Systems without Nitrogen tanks can simply use regular room air as Earth s atmosphere around sea level is only 1 ppm parts per million of Hydrogen or less 5 In either case start flushing the Hydrogen Sensor by clicking the N2 option under the Pump Valve headi
2. Warm Up m min Cool Down m min 15 0 IV Repeat Figure 24 Basic Treadmill Settings e Selected Treadmill Controller This drop down box contains a list of all of the Treadmill Controller units in the current hardware configuration Selecting a device address from this list displays the current settings for that Treadmill Controller e Load Save Allows for the saving and reloading of frequently used Treadmill schedules as tread files e Start Schedule Manually When this option is checked Oxymax will not immediately start the specified exercise schedule at the beginning of the experiment but rather wait for the schedule to be started from the Start Treadmill Schedule option in the Experiment menu e Schedule Table This table details the stages of the Treadmill Control schedule to be executed upon the schedule s start Details for specifying treadmill schedules are further explained in 3 5 2 Treadmill Schedules e Repeat When checked this will cause the specified treadmill schedule to continuously repeat for the duration of the experiment Warm Up The speed the treadmill will be set to upon initialization and continue to use until the start of the treadmill schedule Changes to this value will be immediately applied to the selected Treadmill Controller Cool Down The speed the treadmill will be set to at the conclusion of either the treadmill schedule or the experiment it
3. Oxymax for Windows User Manual June 17 2015 0233 123M Ah Columbus Instruments 950 North Hague Avenue Phone 614 276 0861 Columbus Ohio 43204 2121 USA Fax 614 276 0529 Sales sales colinst com Toll Free 1 800 669 5011 Service service colinst com http www colinst com Table of Contents Chapter 1 Tate CU i sss cteicstespsasncnscse tae cawsieee ea neta nantes cates T a i 3 1 1 OV CTV IEW ssiseseiscascacaisien ccsdicteassean sac vccecstess sssascussuddan sassucndyanci tassieuuasueanidedenonayaeussnadtusassssertosaiuiecvaetyiace 3 KENG Oe UE EN 3 1 3 Preparing for Umsta Atrocities ccc costed dacteceeccscencsdecnctececsevieceat ante ccstecadest cute ccrtecencushestecsuseveeeest 3 1 4 TiS ta Ati e EE 5 Chapter 2 Starting Oxymax fOr WindOWS sssssssssssssnunsununnnnunnununnnnunnnnunnnnnnnnnnnnnnnnunnnnnnnnnan 7 2 1 Hardware Configuration cssscssssssssssssssssssseseseeseeeeesesesseeeeeenensnssnseeeesesesenananaeaeeeeeeeseseenanseeeeees 7 2 2 Hardware E EE 9 2 3 Main Window Navigation ssssssssssssssssssssssssseeeeseseseseseeeeeeensnesesceeesesenenauannaeaeeeseseensnaeeneeeaas 10 Chapter 3 Configuring an Experiment ssssssssssssusussnnunuununnnnunnnnunnnnunnnnunnnnnnnnnnnnnnnnnnnnnnnna 13 3 1 Exploring Experiment Settings scssssssssssssssssssseseeseeeeesesesseeeseeeseeneenesesesesenensnannananenneens 13 3 2 Experiment TE Ur een a eS Se Ea aasaran 15 3 3 Measurement
4. 5 1 3 Drinking Measurements CLAMS systems offer three different methods for recording liquid consumption lick counters Volumetric Drinking Monitors and Load Cell Water Bottles Lick counters are simple electrical devices that register each time the subject drinks from a sipper tube The Volumetric Drinking Monitor detects when the subject licks a specially designed sipper tube and delivers a measured volume of liquid to the subject Load Cell drinking monitors the amount of consumed water by mass rather than volume Systems with lick counter devices register the following measurements for each sipper NI w 74 e Licks The number of licks that have occurred in the interval e Licks Acc The total number of licks that have occurred since the beginning of the experiment Systems with Volumetric Drinking Monitors or Load Cell Water Bottles register the following measurements for each sipper units vary by method e Drink The volume mass of liquid the subject consumed in the interval in milliliters grams e Drink Acc The total volume mass of liquid the subject consumed since the beginning of the experiment 5 1 4 Activity Measurements CLAMS systems can record two different types of activity locomotion via an infrared beam system and exercise via either free running wheels or treadmills Infrared beams can monitor any or all of the chambers axes When using infrared beams data is presented in terms of the number of be
5. K gt VCO subject K VCH subject Equation 13 A 2 Analysis of the Oxidation of Mixtures of Carbohydrate and Fat The following is an excerpt from a correction issued by Graham Lusk in 1923 that was published in the journal Animal Calorimetry It contains the table of oxidation figures for carbohydrates and fats and the equation used by Oxymax for calculation of Heat Equation 14 shows the original form of the equation shown as 5 in Figure 60 which simplifies to the previously mentioned Equation 11 In 1901 Zuntz and Schumburgl Zuntz N and Schumburg Studien zu einer Physiologie des Marsches Berlin 1901 361 published a standard table showing the caloric value of a liter of oxygen when used to oxidize mixtures of carbohydrate and fat Figures were given for respiratory quotients varying from 1 00 for pure carbohydrate to 0 707 for pure fat The writer elaborated this table Williams H B Riche J A and Lusk G J Biol Chem 1912 xii 357 Lusk G The elements of the science of nutrition Philadelphia 3rd edition 1917 61 and introduced columns showing the relative quantity in calories of carbohydrate and fat consumed That these figures were in error was pointed out to me by Dr H H Mitchell in a letter dated June 26 1917 and the error has since been called to my attention by my students To one of them Mr A M Michaelis I am indebted for the following corrected table Although the e
6. Figure 13 Feeding Experiment Properties e Feeding Settings These settings are available for systems that feature any type of CLAMS feed monitor device scale strain gauge etc e Feed Sampling Rate In addition to recording food consumption synchronously with metabolic data all feeding devices can be sampled at a higher rate for higher resolution feeding data The high resolution feeding data is stored separately from the metabolic data Accessing and understanding this data is covered more thoroughly in 5 1 2 Feed Measurements e Minimum Idle Time Oxymax for Windows can also record asynchronous event driven feeding data that meets specified conditions Minimum Idle Time indicates the amount of time the animal must stop eating before the software 22 ee registers a completed feeding event Understanding and using Feeding Event data is described in 5 1 2 Feed Measurements e Minimum Valid Weight This is the smallest amount of consumed food that the animal must eat in order to register a feeding event e Feed Control Settings These settings define the operation of any attached feed control systems The use of these settings is reasonably involved and covered completely in 3 4 Feed Control Settings 3 3 3 Drinking Properties Columbus Instruments offers three options for measuring liquid consumption in CLAMS lick counting monitoring by mass and volumetric drinking monitoring The Volumetric Drinking Monitor VDM has a se
7. 120 sec Activity Threshold fi x Counts 4 23 14 17 16 29 4 23 14 17 19 29 00 03 00 4 23 14 17 20 29 4 23 14 17 22 29 00 02 00 4 23 14 17 20 29 4 23 14 17 23 29 00 03 00 215 4 23 14 17 23 29 00 08 00 423 14 17 19 29 4 23 14 17 27 29 4 23 14 17 28 29 4 23 14 17 32 29 4 23 14 17 28 29 4 23 14 17 32 29 Day 4 23 14 17 38 29 4 23 14 17 35 29 sana lem d ben nem 4 23 18 17 24 29 4 23 14 17 35 29 1 i Analysis Window Start Time Duration 3 2 3 8 8 4 4 2 4 23 14 17 36 29 4 23 14 17 33 23 3 Light Dark Cycle 4 23 14 17 34 29 4 23 14 17 33 29 Light Start Light Duration Save configuration 3 4 23 14 17 36 29 4723 14 17 40 29 4 23 14 17 40 29 4 23 14 17 43 29 Light oo f2 Hous TEG r Statistics All Start Time 4 23 14 17 13 29 Experiment Elapsed Time 63 13 00 Analysis Window Elapsed Time 1 48 00 00 End Time 4 26 14 08 26 29 Epochs 3793 Epochs 2880 02 33 00 1476 24 36 00 00 26 00 1913 31 53 00 3389 56 29 00 Include sleep records in CSV Je Select Cages Generate CSV File Figure 58 The Sleep Analysis Utility e Sleeping Bouts Table of calculated sleeping bouts for the currently selected chamber s based on the selected settings e Detection Parameters e Activity Source Allows for the selection of specific infrared sensors and running wheels in the calc
8. Energy concentrations for diets can be specified as any unit as long as the same unit is used consistently Energy limits can be used on systems with multiple feeders per subject to run studies involving diet preference while limiting consumption based on calories The settings pictured in Figure 21 illustrate how to restrict total consumption by a subject while still allowing for it to make choices about its diet Balance Management Feed Control Schedule Cage 1 Add id Clear Copy Faste JV Repeat Se Acton ett Lit Ere a Open 1440 1000 Figure 21 Feeding Preference with Limits In this case the subject in Cage 1 would be permitted to consume a total of 1000 energy every 24 hours Because the balances for this subject have been combined the limit defined in the schedule applies to the sum of both balances The subject could consume 10 grams of food from the first feeder 2 5 grams from the second or any combination of therein When the subject reaches the limit both feeders in the chamber will close Specifying limits by body mass may be used when trying to normalize consumption across subjects of varying size Limits specified by body mass use the subject s pre experiment body mass as specified on the Experiment Setup window see 3 2 Experiment Setup The schedule shown in Figure 22 would dictate that the selected subject is permitted to eat 5 of its body weight every 12 hours Feed Control
9. Removing every step from the Temperature schedule will release the Watlow controller on the enclosure to manual control Oxymax will still record actual and set point temperatures 3 6 2 Lighting Control Scheduling Figure 27 shows the settings available for lighting scheduling on the Environmental Controller Le Experiment Properties Ki x Metabolic Feeding Drinking Activity Environment Other Selected Controller Chambers 0101 to 0108 DI Lighting Control Add Step Remove Step Timer Ze Schedule Light Color Light min Dark min H 0 S 0 0 L 87 5 7 RHUUDDDDD H 0 100 0 L 25 1 Lighting Fade Times h mm ss onf 0 52 o D a ge Lighting Schedule Start Save to Controller Save to All Controllers Figure 27 Environmental Controller Lighting Schedule Lighting schedules can have up to 32 steps that consist of a Light interval and a Dark interval Each interval can be set up to 18 hours and as low as zero minutes The colors that are used on the LED Light Emitting Diode light strips for Light and Dark can be customized in both hue and intensity e Timer Schedule Allows the lighting to be controlled either by the external Timer on the Environmental Enclosure or scheduled explicitly by the Environmental Controller e Add Remove Step Adds Removes a step to from the end of the lighting schedule el A IL e Repeat When the schedule reaches the last step the contro
10. calibrated and used in an experiment 4 2 Gas Sensor Calibration In most cases all of the Oxymax system s gas sensors must be re calibrated before the start of each experiment The brief and simple procedure ensures that the readings from the sensors are as accurate as possible before collecting experimental data Note Exceptions for calibration Calibration is not always necessary before each experiment if the experiments are being run in quick succession If you re performing several short experiments on the same calendar day calibrating in between each experiment is not required as long as atmospheric conditions remain stable The calibration instructions for the featured gas sensors reflect the proper procedure for currently manufactured systems Older legacy systems may still be supported by the current generation of Oxymax for Windows software with slight differences to procedure Consult your original Oxymax documentation if the procedures listed in this manual are not viable given your system s setup The gas sensors can be calibrated before or after creating a new experiment file If calibrating before opening an Experiment Configuration click Tools gt gt Calibration to open the Oxymax Sensor Calibration Window as shown in Figure 30 The same window can be found under Experiment gt gt Calibration if a blank data file has already been created 45 Group 1 caigas O2 peta Ret Ai
11. or having selected a flow higher than is capable given the pump and restrictors in use 89 VDM Water Reservoir Empty e The VDM unit has detected that the water reservoir is empty and needs to be refilled otherwise no subjects will be able to drink VDM Overpump Warning pumping too much water e The indicated channel has pumped more than the set limit within the given interval This usually indicates a fault of the sipper or the wiring where water is being detected as consumed and replaced at a rate too fast for a subject to actually consume VDM Failure to Drink e The rate of water consumption for a particular subject was lower than the set limits Occasionally this can be because the subject really isn t thirsty but it can also indicate an issue with the sipper where the tip becomes occluded and the subject can no longer drink Zirconia Reference Voltage differential exceeds limits e During reference readings the Zirconia Oxygen sensor checks the voltage indicated while the same gas room air is being measured on both sides of the sensor Under stable conditions this voltage remains largely unchanged If the new value is noticeably different than the last value this message will appear It usually just indicates a noticeable change in room concentration as a result of people entering or leaving the room If this item is logged a couple of times a day it does not indicate an issue If it occurs after every reference 30 times a d
12. 95 82 9 17 1 84 0 16 0 4 985 0 69770 0 96 86 3 13 7 87 2 12 8 4 998 0 69877 0 97 89 8 10 2 90 4 9 58 5 010 0 69984 0 98 93 2 6 83 93 6 6 37 5 022 0 70091 0 99 96 6 3 41 96 8 3 18 5 035 0 70197 1 00 100 0 0 100 0 0 5 047 0 70303 Formula for R Q R Column R 0 707 1 100 _ _ a 0 293 1 00 R 0 293 504 7 R 0 707 3 E eh 5 047 R 0 707 4 686 1 00 EI wu o _ 468 6 1 00 R 5 047 R 0 707 4 686 1 00 R R 0 707 ies 4 E ETH 5 Calories 4 686 0 293 X 0 36 6 Logarithm log of Column 5 Figure 60 Oxidation Table Appendix B Warning Log Messages During an experiment non critical message may be logged to the Warning Log tab which should be checked occasionally The following is a list of messages and descriptions for messages current as of Oxymax for Windows v5 10 Lost serial communication on COM e The serial port Oxymax was using for communication was disconnected from Windows This is usually the result of the USB cable or adapter being removed from the computer Serial communication restored on COM e The serial port Oxymax was using before it was disconnected has been reconnected 02 Out outside of normal bounds e The value read for 02 Out exceeds the normal bounds for an animal based on the 02 In value This can happen because the concentration from the chamber is lower than the sensor can read or because the read differential is in
13. A Oxymax Calculations c sssssssssssssesnesssenseseeeesseeeeessenseaeeneseeeneneseneeanenaees 84 EDNWOIEEITOUdeC Us TE LE 84 A 2 Analysis of the Oxidation of Mixtures of Carbohydrate and Fat 87 Appendix B Warning Log MeSSages ssssssssssssseeesssssseseerssseenesssenserenesseeeeneseneeseenneaes 89 Appendix C Diagram for Oxygen Fuel Cell Replacement ccscsseseeeee 91 Appendix D Procedure for Changing Ammonia Traps sssssssssssssseusnununnunnennnnnnnnnnnana 92 Appendix E Replacing Hydroscopic Tubing in Sample Drier cceeeee 93 Appendix F Environmental Enclosure Light Controller Setup 98 Appendix G Calibrating Modular Treadmills sssssssussssunsususnnnununnunnnnununnunnnnnnnnnnnnnnnna 99 Appendix H Oxymax Hydrogen Sensor ssssssssssussnnunnunuennnunnnnunnnnnnnnnnnnnnnnunnnnnnnanannnnanna 100 a ODE e TE 100 H 2 Components and Installation ss sssssssssssssnsnsnunnnnnsnnnnnnnannnnnunnnunnnunnnnnnnnnannnnnnnnnnnnnnnnnnn annn ana nna 100 H 3 Experiment Setup seesESENENENEREREEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEERENEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEEn 102 FS L RE 102 ST Pre EXperime nt aena aeaa Sra A a Sara EA Saa aaoi LATTES 103 H33 Calibra Te niasa a E A E OA E 103 H 4 Reading Data sssssssssssceesessesssersrsrnasenesssessssseeesesssseeneeesronaneneneneesseneeseeeseasaeneneneonanenenenenees 105 Appendix I Quickstart Experiment Checklist ssssssususssnununnunnnsununnnnunn
14. Configuration cssccscsssssssssesessssseseseseeeeeeeeseseeeeeeeeseeesnenenssesesssenennnsnaneesners 18 3 9 1 Metabolic Epos eege spsecessdacesatzes cher ee 19 3 3 2 Feeding Properti S reegt Ee hatte ete idee nt dade anc E a E 3 3 3 Drinking Properties seh 3 3 4 Activity POP GLties isiissccscescecsasciveccascsssssetiucalcessersnceiubvsscousstsecadtexscsasd suecsattbvecesttvsusetiveadcadpersacruncalccpstiecsibivedsatTeteres 3 4 Feed Control Settings n 1 ccccececeneceneseseseectensnenesensessetennenensnesteescsneseensineseeeetenenenesestescvertrnendararees SAA Standard Feeding scsi sccccccsctiviecisntesestive cheat aataleadive aheinantractics 3 4 2 Linked Feeding EEN 3 4 3 Feeding Limits ee 3 4 4 Feed Control Scheduling eee 3 4 5 Duplicating Schedules nee 3 4 6 Saving and Opening Feed Control Settings EEN 3 5 Treadmill Controller S trim gS sicscoccsaticecccscncscssssccsceeveeccesnesces eesuceeeenstzececetncetscnessdessesvestseesterse 3 5 1 Basic SQtuin GS ee eer 3 5 2 Treadmill Schedules 3 5 3 Treadmill Shock Detectors 3 5 4 Performing Treadmill Experiments 3 6 Environmental Controller Settings 3 6 1 Temperature Control Schedultmng EEN 3 6 2 Lighting Control Scheduling Sege AT 3 7 1 Telemetry ne 3 7 2 Urine Chillers Chapter 4 Running an Experime nt ccsccssssecsssesssseeessseeseseenaeaeeeeeseeeseesenaeseenneseeaeaees 44 4 1 Experiment Preparation cscccscsssessseecesssseeeeseeesssnseeeeeeeeeeeseneeeeeses
15. Data and select the desired CDTA file 2 Click File gt gt Generate Subject CSV s to automatically generate CSV files for every subject in the current data file 3 Each subject may generate multiple files depending on the enabled features of the CLAMS system a Experiment Name Chamber ID CSV Contains all synchronous CDTA measurements for the specified chamber b Experiment Name B Chamber ID CSV Contains all asynchronous feeding data from the BDTA file for the specified chamber c Experiment Name F Chamber ID CSV Contains all high frequency feeding data from the FDTA file for the specified chamber d Experiment Name A Chamber ID CSV Contains all high frequency activity data from the ADTA file for the specified chamber To generate CSV s for specific synchronous measurements 1 Open up the desired experiment data file a The data generated from a just finished experiment is still open after stopping it b To opena previously saved experiment instead of clicking File gt gt Open Experiment Configuration after loading a hardware configuration click File gt gt Open Experiment Data and select the desired CDTA file 2 Click File gt gt Parameters to open the Parameter File Utility shown in Figure 57 Instructions for generating parameter CSV files are also displayed in the Instructions frame of this window Parameter File Utility xj Data File
16. Feeders For systems that have food hardware component it may become necessary during the course of a long experiment to refill feeders without stopping the experiment Oxymax for Windows has a pair of utilities for removing feeders from service while they are being refilled depending on the type of feeders in the system 4 4 1 Balance Feeders To re fill one or more balance based feeders while an experiment is running 1 Click Experiment gt gt Refill Feeders to open the Balance Utility shown in Figure 48 e Note This utility can also be accessed before an experiment configuration has been loaded under the Tools gt Utilities gt Balance Utility menu option il l Unit 1 2 Unit Group 1 Cage 3 Unit I Service Status Value Scale Status IV Tare on Return to Service Remove from Return to Clear All Service Service Refilling feeders during an experiment 1 Select the feeder s to be refilled by checking the box es in the list of tt tr Cage 4 Unit cages 2 Click the REMOVE from SERVICE button 3 Refill the feeder s and place back on balances A MANUALLY tare each feeder that was refilled 5 Check the TARE on return to service box 6 Select the feeder s to be returned to service by checking the box es in the list of cages 7 Click RETURN to SERVICE Figure 48 Balance Utility 2 Selecting any balance in the list of th
17. Hydrogen Sensor s fittings will already include the requisite nuts and plastic ferrules There are two connections that need to be made for the Hydrogen Sensor to function electrical and gas With regard to the tubing connection required for the sample gas the Hydrogen sensor will receive the sample immediately after the sample pump and before it is dried by the Sample Drier as shown below System Sample Pump 101 102 Before installing the Hydrogen Sensor the top most fitting on the front of the System Sample Pump Output to Drier connects directly to the top most fitting on the front of the Sample Drier Sample Gas Input and the bottom fitting on the Sample Drier Sample Gas Output connects back to the System Sample Pump Input from Drier This second connection will remain untouched as long as tubing lengths and device placement allows for it To add the Hydrogen Sensor to the sample line 1 Disconnect the tube from the front of the Sample Drier s Sample Gas Input port 2 Connect this now loose tubing to the back of the Hydrogen Sensor at the port labeled Sensor In If this piece of tubing is not long enough cut a new piece of tubing from the extra provided to the appropriate length and replace the existing tube 3 Cuta piece of 1 8 tubing long enough to connect the Hydrogen Sensor to the Sample Drier 4 Connect this new piece of tubing from the Sensor Out port on t
18. In cases where there is a Treadmill Shock Detection unit in play this value is recorded from that unit instead which will pause the odometer while the animal is being shocked so it doesn t count towards the distance travelled e Shocks For systems that include the Treadmill Shock Detection this is the cumulative number of shocks that have occurred since the beginning of the experiment e Shock Visits For systems that include the Treadmill Shock Detection this is the cumulative number of visits to the shock grid that have occurred since the beginning of the experiment 5 1 5 Urine Measurements Oxymax for Windows monitors urine production using balances or load cells also used to measure food consumption Urine production data is stored in the same way as the synchronous feeding data is e Urine The mass of urine produced by the subject in the given interval in grams e Urine Acc The total mass of urine produced by the subject since the beginning of the experiment 5 1 6 Telemetry Measurements The telemetry component of CLAMS monitors a subject s core temperature and heart rate Columbus Instruments offers two varieties of implantable transmitters that communicate to Oxymax for Windows via a custom antenna affixed to the chamber One model of transmitter monitors subject core temperature and heart rate another just core temperature The following telemetry measurements are recorded on the Temp or Temp HR tabs of t
19. Properties H x Metabolic Feeding Drinking Activity Other Opto Activity Settings Activity Sampling Rate fi min DI Position Sampling Rate f5 Hz v Treadmill Control Settings 220 zl Treadmill Controller jy ue Load Save Settings for Treadmill 220 Add Remove Copy Paste Step Speed m min Ramp 3 mag Warm Up m min a 10 0 60 300 fo 2 200 60 300 Cool Down m min 5 0 IT Repeat Shock Detection Settings fi 0 Number of Visits Limit 100 Total Number of Shocks Limit Figure 15 Activity Experiment Properties e Activity Settings These settings are used in the acquiring and handling of data from the Opto family of beam break activity sensors e Activity Sampling Rate The frequency in which high resolution activity measurements are taken This feature can also be disabled but it is not recommended e Position Sampling Rate The frequency in which the raw beam data from the activity monitors is read and stored by the software This option is only available with systems using the Opto M4 model of activity monitor and can be adjusted between 1 and 10 Hz 26 ee e Treadmill Control Settings These settings define the operation of any attached treadmill control systems The use of these settings is reasonably involved and covered completely in 3 5 Treadmill Controller Settings 3 4 Feed Control Settings Columbus Instruments Feed Control system allows for independen
20. The amount of gas consumed or produced in an interval is added to a running total of all sample volumes Units are adjusted to match the ones specified in the experiment settings Equation 9 Ax n A n 1 Vy subjectt Equation 9 85 86 The respiratory exchange ratio can be calculated from either set of consumption and production rates Equation 10 VCO2 subject VCO2norm RER _ VO2 subject VO2norm Equation 10 Oxymax for Windows uses one of two possible methods for calculating energy expenditure Heat The first uses a calorific value CV that shows the relationship between heat and the volume of consumed Oxygen This value is derived from empirical data from Graham Lusk s The Elements of the Science of Nutrition Equation 11 RER Kilocalories per liter of 02 0 707 4 6862 0 75 4 7387 0 80 4 8008 0 85 4 8605 0 90 4 9226 0 95 4 9847 1 00 5 0468 CV 3 815 1 232 RER Equation 11 The rate of energy expenditure can then be calculated using the subject s rate of Oxygen consumption Equation 12 This equation is standard for small to medium non ruminant mammals Heat CV VO2 subject Equation 12 Oxymax for Windows also allows for the definition of a custom heat calculation equation Coefficients for Oxygen Carbon Dioxide and Methane rates can be specified and used to calculate heat Equation 13 Heat K X VO subject
21. before When the installation finishes shortcuts for the installed software are placed in the Start Menu under Start gt gt Programs gt gt Columbus Instruments Oxymax for Windows also comes with a troubleshooting program called CI Diagnostic that only needs to be used to troubleshoot issues with a Columbus Instruments Technician Chapter 2 Starting Oxymax for Windows 2 1 Hardware Configuration Oxymax for Windows is the software used to setup experiments and collect data from a CLAMS system Each CLAMS system ships with an installation disc that contains Hardware Configuration information specific to that system Assuming the program is installed in the default directory the system s configuration is stored in C Program Files Oxymax Config orC Program Files x86 Oxymax Config depending on the version of Windows in use 32 and 64 bit respectively Oxymax for Windows v5 00 Columbus Instruments Sf Use Default Hardware Configuration 140000 ini Ki Open Hardware Configuration lt Run Oxymax as Data Viewer Only able to review and export existing data E Run CLAX for Data Analysis Download CLAX for fast simple analysis LI But geen Figure 3 Getting Started When Oxymax launches the Startup window shown in Figure 3 appears There are five options for proceeding when Oxymax has been launched 1 Use Default Hardware Configuration Opens the displayed Hardware Configur
22. events relevant to the experiment that will be included with the exported data To add an event to the current experiment 1 Click Add Event from the main Experiment Oxymax menu to display the Add Event prompt as in Figure 45 CREE Event 2 x Dosed subject with Compound Al Chamber Joi 02 Minnie DI Add Figure 45 Add Event Prompt 2 Describe the significant event in the provided text box 3 Select the chamber for which the event pertains 4 Click Add to commit the event to the data file Intervals in which user generated events have occurred will display an asterisk next to the interval number in the data table The event will also appear under the Event Log tab as in Figure 46 G Experiment oj E ing Log Interval DATE TIME Channel Status Description 1 9 21 2009 4 14 59 PM 0102 EVENT Dosed subject with Compound A SE SEET EE SSES a Start 9 21 2009 4 11 25 PM Interval 135 Measuring Reference 02 0 00 CO2 0 000 UI 24s Figure 46 Event Log To stop the currently running experiment click Experiment gt gt Stop from the main menu When Stop is clicked or any other event occurs that could halt the current experiment the prompt in Figure 47 will appear to verify the experiments completion mya eee x This action will stop the experiment Click OK to stop or Cancel to continue EH Figure 47 Experiment Stop Verification et IL 4 4 Refilling
23. excess of what is healthy for the animal CO2 Out outside of normal bounds e The value read for CO2 Out exceeds the normal bounds for an animal based on the CO2 In value This can happen because the concentration from the chamber is lower than the sensor can read or because the read differential is in excess of what is healthy for the animal CO2 sensor reached upper limit e Detected CO2 values are at the maximum detectable level for the sensor DO2 VCO2 readings will not be correct This usually happens when either the sensor has malfunctioned or fresh air has been set far lower than is necessary High Sample Humidity e When asystem has an Environmental Monitor this indicates that the read Sample humidity is too high This usually is the result of the Drier running out of desiccant or the drier tube failing High Reference Humidity e When asystem has an Environmental Monitor this indicates that the read Reference humidity is too high This usually is the result of the Drier running out of desiccant or the drier tube failing This channel is usually only enabled for systems with Zirconia Oxygen sensors High Purge Humidity e This indicates that the desiccant used in the tube drier has been depleted Fresh Air Flow out of tolerance e The measured fresh air for this chamber or group of chambers is more than 10 away from the target fresh air value Depending on the type of flow controller the issue could be the result of a clog a leak
24. feeder in the Feed Control Schedule table as in Figure 16 The system defined in Figure 16 features two feeders in each cage and therefore displays an additional setting labeled Combine 27 Balance Management 010301 Feed Control Schedule 010201 Add ree Clear Copy Paste I JV Repeat Step Action Timelmin Dan E 1 Open 0 Figure 16 Selecting a Balance Each balance has a schedule comprised of one or more steps Each step is comprised of an action a time and a limit The action either Open or Close is performed at the start of the step which lasts until either the time or limit condition is reached A time limit of zero indicates that the feeder will remain at its current state for an indeterminate amount of time A limit of zero denotes that the subject can consume any amount of food during the current step A step with no time or mass limit will last for the duration of the experiment the schedule in Figure 16 allows the animal access to food for the entire experiment Click the Add button above the Feed Control Schedule table to add new steps to the current schedule The Remove button deletes the last step in the schedule while Clear removes all but the first step The Repeat checkbox in the corner of the schedule indicates whether or not the current schedule will automatically return to the first step at the end of the last step A simple day and n
25. is within 0 001 of the target concentration 9 After the gain has been set click Next 10 Turn off the flow of calibration gas es and click FINISH to complete the calibration of the Carbon Dioxide Sensor If the Oxygen sensor hasn t been calibrated yet just click FINISH and proceed immediately to calibrating it 4 3 Starting an Experiment After configuring the experiment and properly calibrating the gas sensors the system is ready to start an experiment Subjects can be loaded into the system chambers at nearly any point during the system setup after Oxymax for Windows has initialized the system Note Subject acclimation Most research protocols require subjects to be given acclimation time before their generated data can be used This amount of time varies from a couple of hours to an entire day For users who require only a couple of hours of acclimation time it s suggested that the animals are placed in the CLAMS chambers as soon as the software is initialized and warming up In this case when the system is done warming up the subjects will have been acclimated When a longer acclimation time is required it s suggested that the experiment be started as soon as the system is warmed up and data recorded during the acclimation period be ignored To start the currently configured experiment select Experiment gt gt Run from Oxymax main menu If there are any unresolved hardware errors
26. on none of these error messages should appear when an experiment starts After initializing all of the CLAMS hardware Oxymax will begin the measurement cycle and display the status of the current experiment as shown in Figure 44 5 IL 60 T lol x File Experiment Tools View Windows Start 4 30 2014 4 54 04 PM Settling Reference 02 0 000 CO2 0 113 OUT Figure 44 A Running Experiment While the experiment is running the tabs along the top of the data file separating the measurement categories can be used to review all of the currently collected data for this experiment The bar at the bottom of the data file displays the current status of the running experiment e Start The date and time at which the experiment started e Interval This is the current interval being measured e Settling Reference Displays the currently running action for the current measurement This pane also displays the current concentrations from all of the system s gas sensors e Progress Bar Provides a visual representation of the time remaining in the current action e 42s Displays the amount of time remaining in the current action While an experiment is running it may become necessary to open the chambers lids to give the animals various compounds or placebos The removal of the chamber lid or the addition of any external stimulus can have an effect on the acquired data Oxymax for Windows can be used to make register
27. still capable of having an Environmental Controller unit installed on them for control of temperature only as they already have incompatible fluorescent lighting installed on them Contact Columbus Instruments for information on adding Temperature only control to existing units 3 6 1 Temperature Control Scheduling Figure 26 shows the options for temperature control on an Environmental Controller Metabolic Feeding Drinking Activity Environment Other Selected Controller Chambers 0101 to 0108 Temperature Control Lighting Control Add Step Remove Step serra ewro Temperature Schedule Stan Jun 01 2014 07 00 AM DI IT Loop Schedule Save to Controller Save to All Controllers Figure 26 Environmental Controller Temperature Schedule Temperature schedules can have up to 32 individual steps with each step comprising of a starting temperature ending temperature and duration to transition between those temperatures The temperatures specified are set point temperatures and may not reflect the immediate temperature in the enclosure The time to achieve target temperatures may vary based on the size of the enclosure and other atmospheric conditions 89 IL e Add Remove Step Adds Removes a step to from the end of the temperature schedule e Start The temperature set point at the start of the step e End The temperature set point at the end of the step If the start and end temperatures are differe
28. the dark pink lines or any of the other positions pictured below 81 Below is an example of the center position for a subject blocking the beams marked in green In the above diagram there are three beams being broken along the Y axis and two along the X axis The dark green cell indicates the location of the subject as calculated by the software The indicated position would be listed as 5 4 82 ee Note Axis notation Most CLAMS systems mark the long axis on the chamber as the X axis When viewing the chamber from the front this makes the X axis run in the direction usually referred to as the Y axis For the sake of consistency this same nomenclature is used in the position data The position in the lower left corner of the chamber is referred to as 0 0 The Y axis runs left to right the X axis runs bottom to top There are some situations where the animal is not breaking any beams along one or both of the axes When this happens the position index is set to 1 When the animal reappears to break beams a new record will be generated but the distance travelled for that record will always be listed as zero since there s no way to discern what the its previous location was 83 84 Appendix A Oxymax Calculations A 1 Metabolic Calculations The metabolic data reported by Oxymax is based on five empirical measurements e Gas concentrations measured by sensors e Re
29. time from the Activity Monitor accessible from the Tools menu both before and during an experiment An example of the Activity Monitor is pictured below in Figure 51 Oxymax for Windows will draw the selected chamber as its defined in the Hardware Configuration file Before an experiment the displayed chamber designation is the channel number While an experiment is running the chambers can be selected by subject ID The currently calculated position is displayed as an index in the corner See 5 4 Position Analysis for details on this index Note Older configurations Position Monitoring has not always existed in Oxymax If your system has X and Y axis activity sensors but the chamber displayed on the Activity Monitor doesn t match your chamber please contact T Current Chamber ot D DI Position 3 3 Figure 51 Activity Monitor Broken beams are displayed as thin green lines Beams broken along the Z axis in the upper plane of the chamber appear as thick green lines As long as beams are broken on both axes the plot will display the currently calculated center position of the subject Use the Activity Monitor before setting up an experiment to make sure that the activity sensors on each chamber are working properly and that there are no outside disruptions breaking beams Note Chambers Each CLAMS system can be configured with a variety of features that sometimes incompa
30. 02 0 040 0 080 CO2 e If possible identify experiment room as off limits limit traffic and other environmental variations Stop experiment Experiment gt gt Stop e Fresh air will continue to be delivered to chambers while software is running Export data to CSV File gt gt Export gt gt Generate Subject CSVs 109
31. 10 Subjects Enabled Chamber Subject ID Mass g Flow LPM 001 Mickey 27 10 0 50 0102 Minnie 25 00 moO 25 00 d AE pp Figure 10 Subject Setup On the right hand side of the Experiment Setup Window are a series of boxes where additional subject information can be stored with an experiment The fields are not used during an experiment but the information is attached to the selected subject and presented with the exported experiment data The left column Label can be used to define a series of properties for a subject gender color age etc and the right column can store the values for those properties male brown 30 weeks etc The Done button commits any changes on this form to the experiment If the currently configured system features telemetry hardware additional columns will appear in the table that allow for the assigning of E Mitter transmitters to each chamber as in Figure 11 17 18 Experiment Setup E Output Data Filename C SkyDrive Projects Oxymax S ource Oxymax for Windows v4 88 Browse Set Default Backup Directory Browse Clear rm Subjects Chamber Subject ID Mass o Flow LPM Tx Serial Temp C Aetio mm Subject 1 25 00 O50 8330 Info 0102 Subject 2 25 00 9767 0103 Subject 3 25 00 9769 0104 Subject 4 25 00 37693 el 0105 Subject 5 25 00 None 0106 Subject 6 25 00 None 0107 Subject 7 25 00 None 0108 Su
32. 53 54 Group 1 Cal Gas 02 Calibration Das Group 02 Zero 1 00 02 Span 20 50 4 Click the CALIBRATE button under the 02 header to open the Oxymax 02 Sensor Calibration prompt The prompt contains abridged reminders of the instructions laid out in this manual The Zirconia Oxygen Sensor is very sensitive to changes in sample flow especially differences in flow across the sample and reference channels For this reason both flows are automatically controlled by a pair of external flow controllers typically a Dual Mass Flow Controller unit and do not need to be adjusted Verify using the display on the Dual Mass Flow Controller that the sample and reference flows read 0 400 LPM Turn on the calibration gas and adjust the pressure so that excess gas is apparent at the exhaust port 5 10 psi Click NEXT to start the purging the sensors with the offset gas The offset gas in this case is actually just room air The progress bar at the bottom of the prompt will indicates the amount of time remaining during the purge period When the purging period is complete the prompt will display the target Oxygen concentration for the calibration gas Use the Offset knob on each Oxygen sensor to adjust the ratio displayed on the Oxymax Sensor Calibration Window until it reads within 0 0002 of 1 0000 The measured value will lag adjustments to the knob by a couple of seconds so wait for
33. Also the subject to which the measurement pertains is referred to by its chamber identification number under the heading Channel 0101 indicates the first chamber in the first group 0205 indicates the fifth chamber in the second group etc 5 1 1 Metabolic Measurements The crux of CLAMS metabolic system is the determining of consumption and production rates of Oxygen and Carbon Dioxide Oxymax makes these measurements by comparing the concentrations of gases being supplied to the chambers to those of the gases leaving the chambers The difference or delta across the cage is the contribution to the system by the animal in the chamber The supply gas room air depends entirely on the stability of the room in which the system is set up For this reason Oxymax makes routine reference measurements of the room air 69 70 Anytime Oxymax switches sample tubes there is an inherent delay between the time a new sample tube is selected and the time the gas from that tube completely flushes the previous sample out of the sensors This time differs based on the flow rate length of the tube and the response time of the sensor This length of time is referred to as a settle time During this settle time Oxymax for Windows queries the gas sensors for their concentrations but does not use the results to calculate the measurement for that sample The interval directly following the settle time is referred to as the measure
34. CH4 You can also select the data to be presented in either kilojoules or kilocalories e Normalize Heat With this selected the rate of heat expenditure be calculated normalized by initial animal body mass energy mass time using the same units as the gas exchange rate measurements e Calculation Parameters These are the units that all of the metabolic calculations will be displayed in e Environmental Sensor Options When enabled logs humidity faults from the Environmental Sensor Humidity readings from either the sample or the drier can indicate the drier is unable to dry the sample gas out of descant failing drier tubing leak etc 3 3 2 Feeding Properties Columbus Instruments offers a variety of options in regards to feed monitoring and control Settings for those components appear on the Feeding tab of the Experiment Properties Window as in Figure 13 21 Le Experiment Properties ki X Scan Feeding Drinking Activity Other Feeding Settings Feed Control Settings Sampling Rate Mode Ze Standard Linked Minimum Idle Time s 10 m Im Ze Mass g C Energy C Body Start Time C Experiment Start 07 00 Minimum Valid Weight a 0 01 Feed Control Settings File Open Save Feeder Management Feed Control Schedule 010101 Add Remove Clear Copy Paste Je Repeat E ea AT SE 1 Open 2 Close 720 0 On Experiment Conclusion Open Feeders DI
35. H and proceed immediately to calibrating it 4 2 5 Manual Carbon Dioxide Sensor Power Carbon Dioxide Sensor O Unlock Columbus Instruments Offset amp 950 NORTH HAGUE AVENUE COLUMBUS OHIO 43204 USA Figure 41 Manual Carbon Dioxide Sensor The Manual Carbon Dioxide Sensor is another variation of infrared CO2 sensor It calibrates and performs similarly to the Automatic Sensor but the offset and gain are adjusted by hand with the software s assistance 1 Click the CALIBRATE button under the C02 header to open the Oxymax C02 Sensor Calibration prompt The prompt contains abridged reminders of the instructions laid out in this manual 2 Ifthe sample flow s for the system is controlled manually on the front of the sample pump adjust the sample flow to 400 mL min If you ve already adjusted the flow before calibrating another sensor do not adjust it again 3 Turn on the calibration gas and adjust the pressure so that excess gas is apparent at the exhaust port 5 10 psi If the system uses Nitrogen for an offset gas instead of Soda Lime turn that tank on as well Otherwise check to see that the column of white granular Soda Lime is secured to requisite tubing A Click NEXT to start the purging the sensors with the offset gas The progress bar at the bottom of the prompt will indicates the amount of time remaining during the purge period During this in
36. Method a DI Samples Heat Calculation Standard 3 815 1 232 x RER x O2 DI kcal kjoule K1 K2 vol IT Normalize Heat Calculation Parameters Volume Mass Time Volume Rate Units mi ko dl d Accumulated Gas Units fi DI Environmental Sensor Options JV Log drier relative humidty errors Figure 12 Metabolic Experiment Properties e Measurement Settings These are all of the metabolic schedule and timing settings e Intervals The number of intervals to be recorded for the experiment When set to 0 zero the experiment will run indefinitely until stopped by the user e Cage Settle The amount of time the system waits for the previously sampled gas to purge from the sensors when switching to a sample chamber e Cage Measure The amount of time the system measures the current sample The gas sensors are sampled once every second for the specified duration and the average is recorded as the measurement For non metabolic CLAMS 20 ee systems this value is used as the time in between samples for activity drinking urine feeding and telemetry data e Reference Settle The amount of time the system waits for the previously sampled gas to purge from the sensors when switching to reference air e Reference Measure The amount of time the system measures reference air The gas sensors are sampled once a second for the specified duration and the average is recorded as the In concentrations for me
37. Schedule 010101 Sc Clear ll Copy Faste JV Repeat Step acion Time Iir Lint Bo Was GEES Figure 22 Limiting Consumption by Body Mass Linked experiments have an additional setting related to consumption limits called Multiplier By default this value is 1 indicating that consumption is multiplied by one before used in master slave calculations For example setting the multiplier to 0 75 for a slave would mean that the subject using the slave feeder would be permitted to eat 75 of its masters average before the feeder would close Conversely raising or lowering the multiplier for a master feeder effectively weights that feeder s contributions to its group s 32 ee average If a master feeder has 2 grams of consumption and a multiplier of 1 5 its contribution to the group s master average is 3 grams 3 4 4 Feed Control Scheduling Feed Control schedules can be set to start at the beginning of an experiment or at a specific time of day Selecting Experiment Start from the Start Time options starts all enabled feeders in the experiment to Step 1 when the experiment starts Selecting the Start Time option allows for the selection of a particular time in which to activate the first step of each feeder s schedule The selected time can either be in the future or the past relative to the time the experiment is started Selecting a start time that is earlier than the current
38. am breaks For the locomotion axes X and Y results are quantified in two different ways total and ambulatory For the rearing axis Z only total counts are recorded The differences between these two metrics are e Total The total number of times infrared beams were broken during an interval e Ambulatory The number of times different beams were broken during an interval If a subject stands still and begins a repetitive act such as grooming it will repeatedly break the same beam The ambulatory count is only incremented when the currently broken beam is different than the last broken beam Running wheels are equipped with a sensor to detect rotation in either direction e Wheel The number of complete rotations by the running wheel in an interval e Wheel Acc The total number of complete rotations by the running wheel since the beginning of the experiment Systems equipped with current generation circa 2010 Treadmill Controller units have the option of recording treadmill speed and odometer values during an experiment The Treadmill Shock Detection unit can be added to record shocks and visits to the shock grid as well The data stored for these products are e Treadmill Speed The current speed of the Treadmill Controller at the time of the sample in meters per minute e Treadmill Odometer The current odometer reading from the Treadmill Controller unit The odometer is reset to zero at the start of the Oxymax experiment
39. aptop Users Laptop users should always perform experiments with the laptop lid open and plugged into an AC outlet The options Turn off hard disks System Standby and System Hibernate should all be set to Never under the Plugged In column Windows Vista 7 8 Users All of the power settings that would need to be changed under Windows XP also need changed for Vista 7 but are found in different places Additionally Oxymax needs to be able to run by a user with administrative privileges without restriction in order to properly save data e Power Options The power settings for Windows Vista 7 can be found in Control Panel gt gt Power Options The option for the Control Panel can usually be found in the Start Menu When the Power Options screen has loaded select Change Plan Settings next to the currently selected power plan It s recommended that the first plan Recommended Settings be used Desktop Users Set the Put the computer to sleep option to Never Laptop Users Laptop users should always perform experiments with the laptop lid open and plugged into an AC outlet Set the option Put the computer to sleep to Never e User Account Control UAC Starting in Windows Vista Microsoft implemented a number of security features in a suite called User Account Control designed to prevent programs from making changes to sensitive files without the user s kno
40. asurements that follow the current reference more on this in 5 1 1 Metabolic Measurements e Reference Method This specifies the frequency of reference measurements during an experiment A value of 4 four specifies that reference air will be measured after every four subjects It is also possible to take only one reference measurement at the start of an experiment this is not recommended and is only possible when the system is in a very environmentally stable room e Measurement Timeline The Measurement Timeline is the table on the right side of the form that indicates the order and schedule for measurements under the current experiment settings The times listed are presented in minutes since the start of the experiment or since the end of the last line of the schedule The timeline will show multiple references or cage measurements up until the point the schedule repeats itself The example in Figure 12 shows a system with four cages being referenced every four cages e Heat Calculation By default Oxymax for Windows uses a generally excepted equation for calculating energy expenditure based on metabolic measurements Information on how this equation was empirically determined can be found in Appendix A Oxymax Calculations If you d like to use a different equation for heat calculation select Custom from the drop down menu and specify coefficients for the volumetric figures VO2 VCOz and in the case of ruminant animals V
41. ation file without prompting 2 Open Hardware Configuration Brings open the selection dialog shown in Figure 4 to select a Hardware Configuration Selecting Yes to the prompt reading Make this the default hardware configuration will replace the current default with the most recently selected file 3 Run Oxymax as Data Viewer Does not setup any of the hardware communication components of Oxymax and allows for the opening of existing data sets only 4 Run CLAX for Data Analysis If installed this will launch the CLAMS Examination Tool CLAX for data analysis If not installed there s an option for internet connected PC s to download CLAX for free from the Columbus Instruments website 5 Quit Oxymax Open Hardware Configuration 2 x Look in e Config DI e Dirr ES My Recent Documents Desktop o Mu Documents mm b 5 File name fo30000 ini DI Ee Files of type Hardware Configurations UNI DI Cancel Figure 4 Select Hardware Configuration Note Do not edit the Hardware Configuration file Hardware Configuration files are written explicitly for the system as it is installed and configured Should the system ever be changed or upgraded a Columbus Instruments representative will alter and or replace the Hardware Configuration file After selecting a Hardware Configuration Oxymax will detect and initialize the Columbus Instruments hardware specified in that configu
42. ay then there may be an issue with sample flow or the sensor itself Zirconia Reference Voltage out of bounds e The reference voltage has risen or fallen to the limits of the sensors ability This would usually occur if the actual calibrated position for the sensor was set incorrectly and the reference voltage moved just a little more out of bounds 02 In value changed significantly between references e The 02 In value changed more than 0 02 between readings If this happens frequently the room could be too unstable or poorly ventilated Increase referencing frequency to make sure the changes are smaller CO2 In value changed significantly between references e The CO2 In value changed more than 0 005 between readings If this happens frequently the room could be too unstable or poorly ventilated Increase referencing frequency to make sure the changes are smaller Appendix C Diagram for Oxygen Fuel Cell Replacement Figure 61 illustrates the proper assembly of an Oxygen Fuel Cell in a standard cell housing It is essential that the provided O ring is secured to the replacement Fuel Cell before inserting it into the housing THUMB SCREWS CAP WIRE LEADS SENSOR POCKET SHAPED TO FIT SENSOR LEADS O RING SENSOR SEAT ELBOW CONNECTOR FOR SAMPLE AIR TUBE Figure 61 Fuel Cell Assembly 91 ER Appendix D Procedure for Changing Ammonia Traps Components inside the Gas Sample Drier and Dual Tube Dri
43. bject 8 25 00 None 0109 Subject 9 25 00 None 0110 Subject 10 25 00 None e 0111 Subject 11 25 00 None Copy to All Subjects Done Figure 11 E Mitter Serial Selection E Mitter transmitters are serialized and have different factory calibration settings Oxymax for Windows needs to know which transmitter is in each cage in order to properly interpret transmitted data Each cell in this column is a drop down list of E Mitter Transmitter serial numbers After implanting the transmitter in the subject or allotting a transmitter to a particular subject click the cell in the row for the chamber that subject will be tested in and select the serial number of the transmitter with which they were implanted The columns to follow will begin to update with measurements after a serial number has been selected if the transmitter is within range or the receiver The Heart Rate HR column will only appear if the system supports and is using transmitters equipped with heart rate leads 3 3 Measurement Configuration After the instance specific settings for the experiment have been defined you can use the Experiment Properties Window to customize the way in which the experiment is executed The Experiment Properties Window is accessible from the Experiment gt gt Properties menu option Not all options will be available to all users settings make themselves available for systems that support the
44. c Oxymax Data Experiment CDTA Available Selected C CO2IN 2 V O2 ml kghr C Activity L020UT gt gt K ml kg hr DC02 v RER SS wi aca C FLOW Ipm v Select All Clear All Instructions There will be a file created for each entry in the SELECTED list The filename format is DATAFILENAME PARAMETER CSV EX Test VO2 CSV The data file folder is the default Use the BROWSE PATH button to specify or create a different folder Hit FINISH to create files Parameter File Export Path f Oxypmax Data Experiment xxx CSV Browse Path Back Next Finish Figure 57 Parameter File Utility 3 Use the Parameter Types list to view the different categories of measurements 78 u 4 Check the box in the Available list next to the desired measurements and click gt gt to move them to the Selected list 5 Click Next to advance the on screen instructions 6 The utility will automatically select the directory that contains the CDTA file as the directory in which to create the generated CSV s To select another directory click Browse Path and select a different one 7 Click Finish to generate the parameter CSV files The filename format for parameter CSV s is Experiment Name Parameter CSV 5 3 Sleep Analysis CLAMS systems equipped with infrared activity monitors and running wheels can be use
45. ce calibration Carbon Dioxide 0 040 0 110 This concentration depends highly on the air quality in the region in which the system is installed It also depends on the turnover rate for the ventilation system and the volume of the room compared to other CO contributors like people Values slightly below 0 040 are acceptable but no environment should read 0 000 unless a chemical CO2 scrubber is being used e OUT The measurement of chamber air taken for the sample Chamber concentrations will vary depending on room air concentration subject contributions and fresh air flow e Dx The difference delta in concentration between the xOUT and xIN values Controlling the fresh air flow to the chambers can be used to move these values into the optimal ranges for the sensors If the difference between the input and output concentrations is too small it s difficult to measure with any degree of certainty If the difference is too high the air will be too thin for the subject to breathe regularly and will have an impact on its respiration e ACCx The total volume in liters or milliliters of gas consumed or produced by the subject e RER Respiratory Exchange Ratio The unit less ratio of the rate of Carbon Dioxide production to the rate of Oxygen consumption When the subject is metabolizing pure fat for energy its RER is close to 0 70 When the subject is metabolizing pure carbohydrates for energy its RER is close to 1 0 The
46. d to analyze sleep behavior based on locomotion For sleep analysis the high frequency activity sampling interval is referred to as an epoch Sampling intervals of once every ten seconds are recommended for utilizable data The subject is understood to be sleeping if it fails to break a particular number beams or turn the running wheel a particular number of revolutions over the course of several epochs The sleeping bout continues until the animal breaks that activity threshold Research has suggested that four consecutive ten second epochs with a threshold of zero activity counts is a sufficient indicator of sleep Note Sleep Experiment Setup Before beginning an experiment in which sleep is an area of interest be sure to enable high frequency activity monitoring on the Experiment Properties Window After loading an experiment template click Experiment gt gt Properties and select the Activity tab Set the Activity Sampling Rate to 10 sec for the best sleep analysis results Sleeping statistics are generated from high frequency activity data by the Sleep Analysis Utility To access the utility shown in Figure 58 select File gt gt Export gt gt Sleep Analysis when Oxymax for Windows has an open data file Display Select fal e r Sleeping Bouts m Detection Parameters Activity Source Sleep Threshold Intervall Cage Duration Epochs Iw XAxis Epochs JM Y axis 2 DI 60 sec
47. data in several files that are generated at the start of an experiment Data is always written to the files as it s acquired so experiment data is preserved in the event of a computer or system power failure Data files generated by Oxymax for Windows for a particular experiment always have the same filename but different extensions Files with the same name should always be kept together if moved on the hard drive or to a portable storage device Each of the file types stores a different kind of CLAMS data not all users should expect to see every one of the file types described below e CDTA The primary CLAMS data file format that contains all of the synchronous measurements and settings for an experiment Every possible measurable device associated with a particular chamber is measured at the same its metabolic measurements Non metabolic measurements are taken in addition to other periodic or event driven measurements of the same devices e BDTA Asynchronous feeding event data e FDTA High frequency feeding data e ADTA High frequency activity infrared beams and running wheels only data e PDTA Position data e WDTA Animal mass data e MDTA Environmental data e ELOG Warning log information Note CLAX support Depending on the version of CLAX in use some of the files above may not import for analysis in CLAX Most of the tables in the CLAMS data files include interval and time information
48. dically at a higher rate than the metabolic sampling frequency This rate is adjustable in the experiment settings This data is only viewable when it has been exported and is stored in the FDTA file Both the synchronous and high frequency measurements consist of the same basic measurements e Date The date and time the measurement was taken e Status If the balance is being disturbed at the time of the measurement it cannot be read and Oxymax for Windows flags the measurement as Unstable The software will then record the difference between the last measurement and the last Stable reading from the balance If the balance is Stable the value can be read and recorded immediately e Feed The amount of food that has been consumed since the last measurement e Feed Acc The total amount of food that has been consumed since the beginning of the experiment The Feeding Events measurements contain additional information about the subjects feeding behavior e Count The number of feeding events a balance has recorded e Interval The metabolic sampling interval in which the event took place e Date The date of the feeding event e Start The time at which the feeding event started e End The time at which the feeding event ended e Duration The length of time the event lasted e Weight The amount of food consumed during the feeding event e Accum The total amount of food consumed during the experiment
49. e four six or twelve samples It s not impossible to use a referencing frequency that does not divide evenly into the number of enabled cages but as Figure 56 shows the result of doing so is that chamber sample intervals will vary in length 04 00 08 00 17 Interval 4 Minutes 08 00 14 00 2nd Interval 6 Minutes Figure 56 Unequal Sample Intervals Figure 56 shows an experiment with two enabled chambers set to reference every three samples In this case the sample interval for any either chamber alternates between four and six minutes For systems that contain only one set Oxygen and Carbon Dioxide of gas sensors only one chamber can be sampled at atime But systems with multiple sets of sensors can sample one chamber for every set of sensors Chambers in multi sensor systems are placed into groups based on the pair of sensors that samples them Chamber identification numbers are prefixed by the group number the fifth chamber of the third group would be referred to as 0305 Every chamber with the same index in every group is sampled at the same time for example 0101 0201 0301 etc would be sampled simultaneously If one chamber in a particular group 0204 for example were disabled for an experiment the sensors for Group 2 would not record data while chambers 0104 and 0304 were being sampled If all the 04 s in each group were all disabled then Oxymax would skip channel four and proceed to the ne
50. e Activity Monitor Real time display of activity beams for all chambers e Treadmill Controllers Start Treadmill Schedule In systems with treadmill controllers this will start the user specified speed schedule Stop Treadmill Schedule In systems with treadmill controllers this will stop the user specified speed schedule and set the treadmill speed to the specified cool down speed Release Treadmill Control Enables manual control of the treadmill from the unit itself e Warning The treadmill will immediately change speeds to the value set by the knob on the front make sure the knob is in a reasonable position or the treadmill is empty before returning control of the unit e Urine Chillers Set Chillers to Chill Begins chilling samples in all Urine Chillers Set Chillers to Thaw Begins thawing samples in all Urine Chillers 14 Turn Chillers Off Disables all detected Urine Chillers e View e Hardware Configuration Opens the current Hardware Configuration file for review e Experiment Configuration Opens the current Experiment Configuration file for review e Window Contains a list of all currently open windows The tabs on the Experiment Window separate the different kinds of acquired CLAMS data Viewing and understanding acquired data is discussed later in 5 1 Understanding Data 3 2 Experiment Setup Some of the settings contained in the Experiment Configuration like measurement time or feeding bout th
51. e left will show its current status on the right e Service Status Indicates whether the balance is currently In Service and being measured or Out of Service and being refilled e Value Last stable measurement from the balance e Scale Status Indicates whether the balance is stable unstable or experiencing an error 3 Check the boxes of any balances that need to be refilled in the table on the left side of the window 62 4 Click Remove from Service to prevent Oxymax for Windows from sampling those balances while they are being refilled 5 Remove the chamber and replace the feeder for all of the balances selected in the utility 6 When all of the selected feeders have been refilled and placed back on the balances click Return to Service to allow the software to record measurements again 4 4 2 Load Cell Feeders To re fill one or more load cell feeders while an experiment is running 1 Click Experiment gt gt Refill Feeders to open the Load Cell Utility shown in Figure 49 e Note This utility can also be accessed before an experiment configuration has been loaded as pictured below under the Tools gt Utilities gt Load Cell Utility menu option Cage Enable Feed1 Feed2 Drink 1 Drink2 Body Urine mo i i 0 00g 002 0 00g 0103 0 00g 004 0 00g 0 00g 0 00g 0 00g 0 00g 0 00g Enable All Disable All 0105 0106 0107 0108 0109 OOOOO0O0000 Fig
52. e out This procedure is easiest performed by doing one side at a time Typically the red side first then rotating the tube assembly as pictured in Figure 7 outside the cabinet to more easily access the blue tube fittings Quick Connect Fittings To release the colored nylon tube push in on the fitting s collar indicated by the black arrows below While holding the collar in you can pull the tube out of the fitting The tube will not release unless the collar is completely pushed in 95 Figure 10 7 Remove the new replacement tube drier from the packaging Reverse the preceding steps to install the new tube First attach the colored tubes to the fittings then secure the tube drier in place with the white clamps The screw fittings need only be finger tight no tools should be used To insert the tube into the quick connect fitting simply push the tube into the hole The collar does not need to be pushed in Please note that the white fittings circled below should neither be tightened nor loosened Loosening will cause a leak tightening it will crack the Nafion material thus causing a leak When securing the tube drier and placing it into the white clamps it is critical to not twist or kink the tube This will damage the Nafion tubing and possibly create a leak It s easiest to first secure the tube in the vertically mounted white clamp in the back The tube assembly must be lifted slightl
53. e this chamber from service before the next experiment This is useful when experiment groups are smaller than the CLAMS system or when it s necessary to shorten the time in between samples for each chamber Subjects cannot be added or removed while an experiment is running e Subject ID This is the animal specific identifier associated with the subject that s going to be used in this chamber It s a good idea to use unique Subject ID s for each subject though Oxymax does not enforce this as a rule e Mass This should be the subject s body mass taken immediately preceding an experiment Metabolic rates are automatically normalized to body mass and this measurement is used in their calculation more on that in Appendix A Oxymax Calculations e Flow This is the fresh air will be delivered to the chamber Depending on the allocation of flow controllers in a system some chambers share flow controllers and therefore will always have the same rates In those cases the cell for fresh air will span the row of every chamber connected to that controller Common flow rates are between 0 5 and 0 6 LPM liters per minute for mice and 2 0 to 3 5 LPM for rats To edit any property within the subject table simply click the field you wish to edit and type into it as you would a regular spreadsheet program Pressing the Enter key will commit your changes to the table and advance the cursor to the next cell in the table as shown in Figure
54. ea and may not allow for access to that data after an experiment Select another location My Documents Desktop etc on the local computer to save experiment data A dedicated directory on the root drive C is a good place to keep data where all users of the CLAMS system can store data without running into permissions issues If a Backup Directory is specified Oxymax for Windows will make a duplicate of all experiment data and save it in that location The Clear option resets the Backup Directory and disables data backup It s recommended that a hard drive on a remote computer or server be used as a backup location in the event of a computer failure The roster of subjects to be included in the current experiment appears as a table on the left hand side of the window Each cage in the system appears as a row in this table with the following properties e Chamber This numerical identifier is comprised of the group to which the chamber belongs as well as its position within that group A group is defined as a collection of chambers that share gas sensors Systems with multiple sets of gas sensors can take samples from multiple chambers at once but each chamber is always sampled from only one set of sensors In the example system presented in Figure 9 the system is comprised of eight chamber and one set of gas sensors Oxygen and Carbon Dioxide e Enabled Checking or unchecking this box will add or remov
55. ed without halting an experiment with the Stop Experiment Schedule menu option under the Experiment menu Whether the treadmill schedule is stopped manually or with the conclusion of the experiment the treadmill will proceed to the cool down speed specified in the settings When the Oxymax for Windows program is closed the Treadmill Controller will immediately stop the motor At this point the controls on the front of the Treadmill Controller will remain disabled until the black Reset button in the lower left corner of the front panel is pushed 3 6 Environmental Controller Settings Newer Environmental Enclosures may be equipped with Columbus Instruments Environmental Controllers which allow for the control and recording of the lighting and temperature in the enclosure The color and intensity of the lighting can be set for both Light and Dark intervals with customizable sunrise and sunset periods The temperature set point can be adjusted between 2 and 50 degrees Celsius with gradual adjustments that can span days or weeks Both temperature and lighting schedules can be set on a by system or by controller basis The Selected Controller dropdown box lists all Environmental Controllers and the chambers they contain Selecting an Environmental Controller will update the window with that unit s current settings Note Environmental Controller Retrofitting Some Environmental Enclosures that shipped before 2014 are
56. een detected and initialized the main Oxymax for Windows window will appear as it does in Figure 7 The Oxymax Hardware Status window can remain minimized in the corner of the application for the duration of the systems use unless there is a perceived problem with the system s communication Figure 7 Main Oxymax for Windows Window The Main Window s workspace is a container for all Oxymax for Windows features and settings These options are accessible from the menu pictured at the top of Figure 7 The menu options change depending on the feature that s in use After a successful system initialization the Main Window s menu options are e File e Open Experiment Configuration Creates a blank data CDTA file from the selected Experiment Configuration e Open Experiment Data Opens a previously created CLAMS Data File The acquired data can then be reviewed graphed or exported e View Hardware Configuration Displays the currently loaded Hardware Configuration for review e Last Used EXP If available automatically loads the more recently used Experiment Configuration file e Tools e Calorimetry All of the utilities relevant to the Oxymax calorimeter Oxymax Utility Tool used to test the system s valves sensors and flow controllers Calibration Opens the Oxymax Calibration Utility for calibrating and testing the system s gas sensors e Feeding Balance Utility Form used to refill feeder balances during an
57. eesasaeseseeneneseneseseeneases 44 4 2 Gas Sensor Calibration sescsessesssseesesesseseeessseeeeeeeeeesessereesnonaeenenesessneeeececeeeneneeesronnaneneneeseees 4 2 1 Fuel Cell Oxygen Sensor 42 2 ParaMagnetic OXY RE A 2 3 AAT COMMA OXY E E 4 2 4 Automatic Carbon Dioxide Sensor 4 2 5 Manual Carbon Dioxide Sensor sss 4 3 Starting an Experime nt ccscccscecsssssseesseseseesnseeeeeseneesseseeeeseseeeesnsaeenseseeeesesaesenesneneseneenseeneaees 44A Refilling Feeders eege ee 4 4 1 Balance Feeders 4 4 2 Load Cell Feeders 4 5 Monitoring Real Time POSitiOn scscscscscssesssssssseseseensnenssesseeeeseeesananansnseereeeeeeeeseanseseeeenes 4 6 Graphing the Current Experiment ccssssssseessesseeseerersneeneneeesssseeeeeeeseneneneeesnonaneneneeees 66 Chapter 5 Data P E E E E 69 5 1 Understanding Data cssssssssssssssssesesssssssseneeenseseeeeeeeononaneneneeeeseeneeceseseaeneneeesnonananeneneseeeeeeeeeees 69 5 1 1 M ta olic M as reMmentS eege ten EES 69 S12 Feed Measure E 72 5 1 3 Drinking Measurement eege 73 5 1 4 Activity Measurements ENEE 74 5 1 5 Urine Measurements criameniaanamnn naaa aa aa A 75 5 1 6 Telemetry Measurementt S euer 75 5 1 7 Environmental Measurementzs EEN 75 5 2 Exporting Data essssssccessessseeserersnseesenesssesseeececesesseeneeeseononaceneneeesseneesesesenenenesrononanenenenseseeeseeeeees 76 KEE ICH EE 78 5 4 POSITION ER EI UC EE 80 SAL Ee et 81 Appendix
58. eing turned on make sure each balance has its associated feeder or collector on it before turning it on If it s not already make sure the PC used to manage the CLAMS system is powered on and logged in If the PC is connected to the internet it s advisable to disable Microsoft s Automatic Updates as these updates can possible restart Windows without user intervention during an active experiment It s also advised that all energy saving settings be disabled so that the computer will not automatically turn off its hard disk or power down after long periods without user interaction Run Oxymax for Windows and select the appropriate Hardware Configuration If the current PC controls a single fixed configuration CLAMS system then select Yes when Oxymax asks Use the current hardware configuration After selecting a Hardware Configuration Oxymax will attempt to initialize the hardware defined therein If all of the hardware is detected properly the CLAMS Hardware Status window will automatically minimize itself If the CLAMS Hardware Status window lists any instruments as having errors check the specified hardware to make sure that it is powered on and setup properly before clicking Refresh The system cannot warm up properly if initialization errors are detected After the system has been properly initialized the gas sensors and controllers require approximately two hours of warm up time before the system can be
59. ended e Hard drive with at least 100MB of free space for installation e Space for acquired data may vary at least 1GB recommended e CD ROM drive for installation e Monitor running minimum resolution of 1024 x 768 e Atleast one free RS 232 Communications Port or a registered USB to Serial Adapter e Most new systems include the USB to Serial Adapter used to test and setup the system at Columbus Instruments 1 3 Preparing for Installation To ensure that Oxymax for Windows performs reliably there are a small number of Windows settings that should be configured before installation e Windows XP Depending on the manufacturer of the PC sometimes power saving settings are in place to put Windows to sleep after long periods of inactivity Sleeping will interrupt and halt any currently running Oxymax experiments and should be disabled on computers running Oxymax experiments Note Windows XP End of Life Microsoft has ended support for Windows XP as of April of 2014 Some newer releases of Oxymax for Windows may cease to function under Windows XP and it s recommended that users upgrade to at least Windows 7 e Power Options The power settings for Windows XP can be found in Control Panel gt gt Power Options The option for the Control Panel can usually be found in the Start Menu Desktop Users If the PC is a desktop the action that is required is setting the option for Turn off hard disks to Never L
60. er products are very sensitive to the ammonia in urine Ammonia needs properly scrubbed from any sample passed through the driers to prevent damage The ammonia is removed from incoming samples via traps packed with copper sulfate like those pictured below The copper sulfate will turn dark blue and or aqua green depending on the concentration of ammonia passing through it The best practice for attaching and maintaining ammonia filters is as follows e Ammonia traps should be connected two at a time in series right before a particulate filter as pictured above e The location of this filter assembly will vary by system configuration but is usually the last component connected to the vacuum side of the sample pump The drier is usually the first component on the sample pump s pressure side e When the system is in use the first ammonia trap will change color depending on the ammonia saturation The exact color of the expended portion of the trap is not important A completely expended ammonia trap is pictured below e When the second ammonia trap begins to exhibit any color change it means that the first trap is expended and needs to be replaced e Itis best to replace the ammonia trap in between experiments but it can be exchanged during an experiment in an emergency If the trap is being changed while an experiment is running try to perform the exchange during a reference reading and be sure to mark an Event e To replace the expe
61. etup exe When the installer finishes loading it will pause on the window in Figure 1 Y Oxy max for Windows v4 40 Setup xi 8 Welcome to the Oxymax For Windows v4 40 installation program ee Setup cannot install system Files or update shared Files if they are in use Before proceeding we recommend that you close any applications you may be running Exit Setup Figure 1 Installation Welcome Screen Selecting OK will proceed to the next step of the setup picking an installation directory The default directory is C Program Files Oxymax but it can be changed if necessary Clicking the large button in Figure 2 is the last step in installing Oxymax for Windows G Oxy max for Windows v4 40 Setup x Begin the installation by clicking the button below Click this button to install Oxymax for Windows v4 40 software to the specified destination directory Directory C Program Files Oxymax Change Directory Exit Setup Figure 2 Directory Selection Window Before the end of the installation the installer may prompt to Overwrite Configuration if this is an update This will copy the contents instruction manuals etc of the included Config folder to the Oxymax directory Also the installer may prompt to install communication drivers in the event that the system uses an included USB to Serial Adapter Installing these drivers is recommended if it has not been done
62. experiment Load Cell Utility Form used to refill overhead load cell feeders during an experiment Feed Control el UU IL e Open Feeders Allows all animals access to all feeders e Close Feeders Denies all animals access to all feeders e Stop Feed Control Stops the set feeding algorithm for the running experiment Activity Monitor Real time display of activity beams for all chambers Drinking Purge Purges the intake line Pump Status Displays the current status of the system s Volumetric Drinking Monitors Urine Balance Utility Form used to monitor urine balances during an experiment Load Cell Utility Form used to monitor load cell urine monitors during an experiment Urine Chillers e Chill Sets all Urine Chillers to chill their samples e Thaw Sets all Urine Chillers to thaw their samples e Off Turns off all detected Urine Chillers Weight Utility Tool for testing animal weight fixtures Iso Thermex Temperature Utility Diagnostic utility for reading temperatures from all Iso Thermex probes Window Contains a list of all currently open windows Help Documents Lists PDF manuals and instructions for various procedures Video Help Contains instructional videos About Displays the Oxymax for Windows About Window Chapter 3 Configuring an Experiment 3 1 Exploring Experiment Settings To start or setup an experiment Oxymax for Windows must have an Experiment Configuration file CLAMS sy
63. ference Oxygen Concentration OZ e Reference Carbon Dioxide Concentration C02 e Sample Oxygen Concentration 02 e Sample Carbon Dioxide Concentration CO2 e Fresh Air Flow e The fresh air flow measured by Oxymax for Windows depends on the ventilation system used Positive ventilation systems throttle and measure the fresh air before delivering it to the chambers Vi Negative ventilation systems draw fresh air into the chamber from the atmosphere around it The measured flow in this case is not the rate of air entering the cage but exiting it Vo The calculation of Oxygen consumption VOz and Carbon Dioxide production VCOz values requires the use of both the input Vi and output Vo flows to the chamber Despite the fact that the chamber is not pressurized and does not allow gas to build up as an experiment progresses the chamber cannot be modeled as a perfect steady state system because of the subject s unequal effect on the respiratory gases the subject does not exhale the same volume of gas as it inhales Gases in an Oxymax system are broken into three categories e Consumed Oxygen e Produced Carbon Dioxide Methane e Inert All other gases not involved in respiration Nitrogen Argon and other noble gases Since inert gases are not used during respiration their rates of flow at the output VNo and input VNj are the same Equation 1 VN VN Equation 1 The flows of inert gases are products of the flow
64. from when Oxymax was first started a reminder prompt will appear as in Figure 42 P Hardware errors were found on program initialization Some NO7 devices may not respond Do you with to continue Yes No Figure 42 Hardware Error Prompt It s recommended that all hardware errors be resolved before starting an experiment but the experiment can proceed without the functionality of the failed units 58 Oxymax will program any necessary devices with the experiment settings and make a final check that all equipment has been initialized properly During this time counters are re initialized and scales are tared before the experiment If Oxymax fails to initialize any of these CLAMS devices a window labeled Initialization Error Log detailing device errors will appear along with a prompt as in Figure 43 GER Date Time 9 21 2009 11 30 39 AM Tare Error Error zeroing balance Address 128 9 21 2009 11 30 39 AM Tare Error Error zeroing balance Address 129 9 21 2009 11 30 39 AM Tare Error Error zeroing balance Address 130 9 21 2009 11 30 39 AM Tare Error Error zeroing balance Address 131 9 21 2009 11 30 39 AM Clear Error Error clearing Opto M3 counters Address 160 9 21 2009 11 30 39 AM Clear Error Error clearing Opto M3 counters Address 161 Oxy mar hl Initialization errors were detected Click OK to continue or Cancel to quit Figure 43 Initialization Error Log If all CLAMS hardware is plugged and powered
65. he back of the Hydrogen Sensor to the Sample Gas Input on the front of the Sample Drier To connect the 25 pin serial cable that functions both a power cable and communication line 1 Power off the entire system 2 On the back of the Carbon Dioxide Sensor there are two 25 pin serial ports with cables attached One of these cables leads to the Oxygen Sensor disconnect that cable from the back of the Carbon Dioxide Sensor 3 Connect one end of the provided 25 pin serial cable to the now vacant port on the back of the Carbon Dioxide Sensor and connect the other end to either of the free ports on the back of the Hydrogen Sensor 4 Connect the loose end of the Oxygen Sensor s serial cable to the remaining open port on the back of the Hydrogen Sensor 5 Tighten the thumb screws on the ends of each cable to ensure solid connections 6 Turn the power to the system back on H 3 Experiment Setup H 3 1 Overview It s worth noting at this time that the Oxymax calorimeter for animals was designed to only natively support three gases and it is only through its scalability and a few tweaks that additional gases are supported It is because of this that Hydrogen Sensor calibration and data collection require a few additional steps as the software does not support it natively Understanding the data collected from a four gas system is covered in H 4 Reading Data It must also be noted that the Hydrogen Sensor is only supported by Oxyma
66. he data file e Body Temp The subject s core temperature at the time of the sample e Heart Rate The subject s heart rate at the time of the sample in beats per minute In addition to the subject s core temperature CLAMS can monitor the ambient temperature of the chamber using Columbus Instruments Iso Thermex e Cage Temp The ambient temperature of the chamber at the time of the sample 5 1 7 Environmental Measurements In addition to monitoring subject behavior and health CLAMS systems with the Columbus Instruments Environmental Monitor can record conditions in the room into the MDTA file The Environmental Monitor data records the following information e RH Samp The relative humidity of the sample gas before drying e RH Purge The relative humidity of the dry air used to remove moisture from the sample e RH Amb The relative humidity of the environment e Temp Amb The ambient temperature e Baro Press The barometric pressure of the room in millimeters of Mercury e Room Light Whether or not the lights were on in the room at the time of the sample 75 Contained in the same data file as the Environmental Monitor data is data from the Environmental Controller product if the system has one e Enclosure Temp The measured temperature inside the Environmental Enclosure in degrees Celsius e Enclosure Setpoint The set point temperature of the Environmental Controller e LED Hue The hue of the LED cont
67. hillers JV Auto Thaw Iv Thaw Time min 3 Figure 29 Urine Chiller Settings 43 44 Chapter 4 Running an Experiment 4 1 Experiment Preparation The order and timing in the execution of the first steps to running a CLAMS experiment are crucial Many components have to be powered on in a particular order and allowed to warm up for a period of time before setting up and running an experiment These instructions assume that the CLAMS system has been powered down and dormant for at least a day before beginning the next experiment L When CLAMS systems are installed most components are usually wired such that they are controlled by one or two mains power switches These switches are usually found on bus strips and surge protectors near the system Turn on all bus strips and surge protectors associated with the CLAMS system Most of the system s blue and silver cabinets have a light that indicates that the device is powered check all visible devices for power before proceeding Note CLAMS systems with balances If the system contains any balances for monitoring food or urine they initialize into Standby mode and need to be turned on manually If the balance is not turned on manually before starting an experiment no data will be collected from that balance To turn them on press the On Off button on the front of all system balances Also all balances must have a minimum load on them before b
68. ight feeding schedule would consist of alternating 12 hour feeding and 12 hour fasting periods Figure 17 shows a limitless repeating feeding cycle for a single balance Feed Control Schedule 010101 Add Remove Clear Copy Paste I JV Repeat z Tmo um Figure 17 Simple Day Night schedule e Step 1 Opens the feeder and leaves it open for 720 minutes 12 hours e Step 2 Closes the feeder and leaves it closed for 720 minutes e Repeat Returns the feeder to Step 1 28 ee A limit can be placed on an Open step that causes either the feeder to close or the controller to advance to the next step when reached If the step has a time limit anda consumption limit the feeder will close for the remainder of the specified time when the consumption limit is reached If a consumption limit is specified without a time limit the controller will advance to the next step when the subject reaches the specified limit Figure 18 shows a modified version of the day night schedule from Figure 17 with the addition of limits r Feed Control Schedule 010101 Add Remove Clear I Copy Time min Limit g JV Repeat Figure 18 Day Night Schedule with Limits e Step 1 Opens the feeder and leaves it open for 12 hours or until the subject consumes 3 grams of food The controller remains in this step for the entire 12 hours regardless if the consumption limit is met or not e Step 2 Closes the feede
69. ings to file click Save under the Feed Control Settings File heading The dialog window like one pictured in Figure 23 Save Feed Control Settings will appear where the name and location of the FCTL Feed Control file can be specified Save in o Contig D EI rz ES ie My Recent Documents Desktop e CH My Documents WE File name Day Night fctl Meme Save as type Feed Control Settings fctl DI Cancel Always Open fetl KE Daily Preference fctl Figure 23 Save Feed Control Settings It s important for users with access to multiple CLAMS systems to note that Feed Control Settings files can only be used with the system for which they were created 3 5 Treadmill Controller Settings Columbus Instruments Treadmill Controller allows for the scheduled control of one or more lanes of a metabolic treadmill during an Oxymax experiment Oxymax can be used to set customized speed schedules for treadmills while simultaneously collecting metabolic data The Treadmill Control settings pictured in Figure 15 only appear if the hardware configuration file in use contains settings for a Treadmill Controller device Many systems that make use of metabolic treadmills are also equipped with regular chambers and usually ship with separate hardware configurations for the two setups Note Check the Treadmill Controller for Oxymax Compatibility Not all Columbus Instruments Treadmill Co
70. ion Saved message will be displayed as a further reassurance that the most recent calibration was Saved automatically To complete the process and easily return to the main menu simply power the unit off and then back on using the switch located on the rear of the controller 99 100 Appendix H Oxymax Hydrogen Sensor H 1 Overview The Columbus Instruments Hydrogen Sensor for metabolic applications is a fuel cell type measurement device operating under many of the same principles as the Fuel Cell Oxygen Sensor A chemical reaction in the cell produces electricity that varies with the amount of Hydrogen present in the cell e Operating Range 0 to 2000ppm e Expected Operating Life 2 Years e Temperature Range 20 C to 50 C e Relative Humidity Range 15 to 90 non condensing Additionally it should be noted that the chemical fuel cell will dry out and expire if kept in exceptionally dry environments for extended periods of time This is counter to the best practices regarding other gas sensors which function best in perfectly dry environments It is for this reason that the Hydrogen sensor is plumbed in after the sample pump and before the drier unit H 2 Components and Installation The only components required for installing a Hydrogen Sensor into a functioning Oxymax system are e Hydrogen Sensor e A few feet of 1 8 tubing e Extra nuts and ferrules e 25 pin serial cable In fact it s likely that the
71. ion displayed on the Oxymax Sensor Calibration Window until it matches the target calibration concentration The measured concentration will lag adjustments to the knobs by a couple of seconds so wait for the value to settle before continuing The displayed value s will turn green when the measured value is within 0 01 of the target concentration After the measured value matches the target concentration click NEXT to display the finalization instructions If all gas sensors have been calibrated turn off the flow of calibration gas and click FINISH Otherwise just click FINISH and proceed immediately to calibrating other sensors 4 2 2 Paramagnetic Oxygen Sensor Power Paramax 101 S Sensor Usable Output Range e Span Offset Paramagnetic Oxygen Sensor Coarse Span Offset Offset Increment from Low Gas Point Low Gas Point Low Gas Point Leg W Instruments Calibration Adjustments LI LJ Figure 36 Original Paramagnetic Oxygen Sensor PARAMAGNETIC GAS ANALYZER cate Output Range O Pwr Es Sur mans e our IN DUT A Coarse Fino Span Offset Offset Increment from BS Columbus Instruments Low Gas Point Low Gas Point Low Gas Pom Ceres COLUMBUS OHIO 43204 USA Calibration Adjustmonts TEL61 4 275 0661 POWER e E ee 0232 0020 Figure 37 Cl Bus Paramagnetic Oxygen Sensor The C
72. le Add Adds a new step the current treadmill schedule Remove Removes the last step from the current treadmill schedule Copy Copies the current treadmill schedule to the clipboard for pasting on other Treadmill Controllers Paste Pastes the settings from the Treadmill Controller on the clipboard to the currently selected controller m Treadmill Control Settings 220 zl Treadmill Controller fv jc Load Save Settings for Treadmill 220 Hodis Warm Up m min 5A 60 60 Cool Down m min 60 fi 5 0 V Repeat Figure 25 Treadmill Schedule To make changes to the treadmill schedule select the cell in the table and use the keyboard to change the desired value Press the Enter key to commit the altered value to the schedule The current schedule is depicted in the timeline beneath the schedule shown in Figure 25 The red line indicates the speed of the treadmill during the warm up period the line in blue represents the cool down period The plot will automatically adjust the scale to contain the entire length of the schedule as well as the schedule s highest speed 3 5 3 Treadmill Shock Detectors Also depicted in Figure 25 are settings for the optional Treadmill Shock Detection product When combined with a Treadmill Controller the Treadmill Shock Detection device can record shocks and shock events as well as disable the shock grids when the subject exceeds customizable limits e Numbe
73. lick NEXT to move to the next step of the calibration After making sure the span gas tank is turned on and providing sufficient flow click NEXT to start purging the sensors with span gas When the purging period is complete the prompt will display the target Oxygen concentration for the span gas Use the Span knob on each Oxygen sensor to adjust the concentration displayed on the Oxymax Sensor Calibration Window until it matches the target calibration concentration The measured concentration will lag adjustments to the knob by a couple of seconds so wait for the value to settle before continuing The displayed value s will turn green when the measured value is within 0 01 of the target concentration After the measured value matches the target concentration click NEXT to display the finalization instructions If all gas sensors have been calibrated turn off the flow of calibration gas and click FINISH Otherwise just click FINISH and proceed immediately to calibrating other sensors 4 2 3 Zirconia Oxygen Sensor ZIRCONIA OXYGEN SENSOR Sensor Temperature in deg C Unlock Si emp sg Thermal Fautt over Power Fault Columbus Instruments 950 NORTH HAGUE AVENUE COLUMBUS OHIO 43204 USA TEL 614 276 0861 FAX 614 276 0529 www colinst com Figure 38 Zirconia Oxygen Sensor The Columbus Instruments Zirconia Oxygen Sens
74. ll Controllers Saves the current displayed Lighting settings to all Environmental Controllers 3 7 Other Settings Settings for smaller sub systems of CLAMS can be found under the Other tab in Experiment Properties 3 7 1 Telemetry Oxymax for Windows allows for a filter to be applied to heart rate data acquired through the optional telemetry system as seen in Figure 28 d 2 x Metabolic Feeding Drinking Activity Other Telemetry Heart Rate Filter Enable Filter Jv Maximum BPM 1000 Minimum BPM 100 Figure 28 Telemetry Settings When enabled heart rates outside the specified range will not be included in the averages presented in the data Be careful when using this filter as data outside the specified range cannot be recovered 3 7 2 Urine Chillers Systems equipped with the Waste Management bases have an available Urine Chiller option The chillers can be configured to automatically freeze and thaw urine and at the start and end of an experiment with the settings shown in Figure 29 Selecting Enable Chillers will give control of all units over to the software which will begin freezing the sample vial at the start of an experiment Auto Thaw indicates that the chillers will automatically warm the sample for the indicated amount of time selectable between 1 and 5 minutes at the conclusion of an experiment e 2 x Scan Feeding Drinking Activity Other Chillers Enable C
75. ller will loop back around to the step marked Repeat This allows for an acclimation period of one or more steps followed then by a repetitive pattern of one or more steps e Light Dark min The intervals in which the lights will be on and off in each step Customizable between zero minutes and 18 hours e Light Dark Color Clicking either color panel will bring open the Windows Color Selector prompt where a color for each state can be selected Colors on the LED s will not appear exactly as in the software it s suggested that both colors are verified before running an experiment to make sure they match expectations Information on each color is presented in standard HSL Hue Saturation Lightness format e Lighting Fade Times The LED lighting can be set to fade from Light to Dark and Dark to Light over customized periods up to 18 hours This can be used to simulate sunrise and sunset in the enclosures e Lighting Schedule Start If the Lighting Controller is set to Schedule mode this is the time that the schedule will begin its first Light step As with the Temperature Schedule if the time is set in the past the controller will automatically shift to the appropriate step the current schedule If set in the future the lights will remain at their current state until the schedule begins e Save to Controller Saves the currently displayed Lighting settings to the selected Environmental Controller e Save to A
76. m Clicking OK at the bottom of the form commits any settings changes to the experiment By default most altered settings are saved back onto the original Experiment Template EXP But if you make changes to your experiment configuration that you wish to re use you can write them to the template so they become the default settings To overwrite an existing set of Experiment Properties just click File gt gt Save Experiment Configuration A few products store their settings in the devices themselves rather than any file and will persist between experiments regardless Note Multiple Experiment Configurations If you are consistently performing experiments with the same sets of different settings you can use the File gt gt Save Experiment Configuration As menu option to save the current settings to a new file When you wish to run that new experiment type later just be sure to select it from the Open Experiment Configuration dialog when creating a new experiment 3 3 1 Metabolic Properties When the Experiment Properties Window first opens it displays the settings for the metabolic components of the CLAMS system as in Figure 12 OE IL ccs Experiment Properties zl Metabolic Feeding Drinking Activity Other Measurement Settings Intervals 0 Cage Settle sl 30 Cage Measure s 30 Reference Settle s 30 Reference Measure s 30 Reference
77. mbus Instruments Automatic Carbon Dioxide Sensor uses a non dispersive infrared measurement technique to measure Carbon Dioxide It has a short warm up time with a front mounted Ready light that indicates when the sensor is ready to be used Once the system is warmed up and the calibration gases are turned on calibration of the Automatic Carbon Dioxide Sensor is controlled entirely by the software and requires no user intervention Note Ready Light The infrared bulb used by the Automatic Carbon Dioxide Sensor is similar to a conventional light bulb in that it has a filament After years of use it s possible for this filament to burn out at which point the sensor will need to be serviced The unit comes equipped with an LED that indicates if the sensor is warmed up for use In most cases 56 the sensor warms up in about twenty minutes If several hours after powering the unit on the Ready indicator light on the front of the unit is not lit it s possible that the infrared bulb has burned out and requires replacing Click the CALIBRATE button under the C02 header to open the Oxymax C02 Sensor Calibration prompt The prompt contains abridged reminders of the instructions laid out in this manual If the sample flow s for the system is controlled manually on the front of the sample pump adjust the sample flow to 400 mL min If you ve already adjusted the flow before calibrating another sens
78. ment time During this time the sensors are queried once a second and the average reading over the measurement time is recorded as the measurement value Oxymax for Windows repeatedly takes reference measurements after a specified number of chamber measurements Figure 55 shows the first four cages of an example experiment with both settle times set to 90 seconds and both measure times set to 30 seconds 00 00 01 30 02 00 03 30 04 00 05 30 06 00 07 30 08 00 09 30 Ref Settle Cage 1 Settle Cage 2 Settle Cage 3 Settle Cage 4 Settle O U UOU O LU 01 30 02 00 03 30 04 00 05 30 06 00 07 30 08 00 09 30 10 00 Ref Measure Cage 1 Measure Cage 2 Measure Cage 3 Measure Cage 4 Measure Figure 55 Example Experiment Timeline If the example system had only four chambers and was set to make a reference reading every four chambers the timeline in Figure 55 would repeat until the experiment was stopped The referencing frequency in samples is usually calculated such that a reference measurement takes place at least every 20 30 minutes Research environments with especially stable ambient conditions can afford to go longer periods in between references while noisy environments should reference more frequently It is also recommended that the number of samples in between references be evenly divisible by the number of enabled cages in the system For example a sixteen chamber system with twelve enabled chambers could reference every thre
79. nded trap remove it from the assembly and shift the old second trap into the first trap position e Put the brand new ammonia trap into the second trap position Appendix E Replacing Hydroscopic Tubing in Sample Drier Remove the two screws from the lid as in Figure 1 and slide the lid off to reveal the inside of the cabinet as shown in Figure 2 The Tube Drier is held in place by three white clamps indicated in Figure 3 Two are located on the side in plain view and the third is mounted vertically on the rear panel beneath the stainless steel fittings for the desiccant columns E SA l To unlock the clamps push down on the tab then pull it away slightly from the round clamp The round clamp will naturally loosen itself partially you may need to pull it free of the tab in order to completely open the clamp Figures 4 and 5 show the directions in which to manipulate the clamp in order to open it completely Carefully remove the tube drier from the clamps so that it lays loosely inside the cabinet as pictured in Figure 6 94 Disconnect the tube drier at the four fittings shown below There are two types of fittings on each side a red 1 8 inch quick connect fitting a red 1 4 inch screw fitting and a duplicate set on the blue side The screw fittings simply unscrew to disconnect To disconnect from the quick connect fitting compress the ringed collar against the fitting itself holding it in while pulling the tub
80. ndetectable or uninitialized hardware will be listed as Fail The reason for the device failure is listed in the Description column of each table Common error descriptions include e No Response The device did not respond to communication at all This is likely because it is powered off or disconnected from power communication or both e Power Fail Usually pertaining to Balance devices this indicates a failure to respond from the balance This is usually the result of the balance being powered off if the display is blank or reads OFF just tap the blue Tare button to turn the balance on If this does not turn the unit on check the power cable on the back to make sure it s secure e Checksum Error Response Error Some serial RS 232 adapters are not robust enough to provide consistent communication with a CLAMS system Try updating the drivers for the unit in use or replacing with a different model After the software is able to successfully detect all of the system s specified hardware the Oxymax Hardware Status window can be minimized Hardware Statuses Devices that are listed as Fail on the status window will not be polled during an experiment which will mean missing measurements and or incorrect system operation It s advised that all initialization errors be addressed and refreshed before proceeding to an experiment 2 3 Main Window Navigation After all of the hardware has b
81. ndow 1 Opto Activity Monitor Measures beam break counts and gives position information about 0 5 1Hz The actual sampling rate varies depending on the size and complexity of the system For stand alone models this product is usually branded as the Opto M3 2 Opto Activity Position Monitor Measures beam break counts at around the same frequency as the Opto Activity Monitor but can acquire position information as fast as 10Hz The actual scanning frequency is adjustable within the experiment setup per 3 3 4 Activity Properties For stand alone models this product is usually branded as the Opto M4 At the conclusion of an experiment position data can be viewed and exported via the Position Analysis utility accessible from the File gt gt Export gt gt Position Analysis r Analysis Options Points toAverage _ 2 Points Fence Sie _ 0375 Fence Refresh 1 see Ierwe Recalculate Subject Records Distance in Brown and Whit 101874 91565 84 Brown 107771 96288 33 Black and White 89559 84990 59 Butterscotch 103008 98358 34 Big Rat 26418 24649 90 Small Rat 53567 66836 54 r Results Brown and White Record 0 101873 Save Raw Save Analysis 19 12 4 21 2014 13 59 07 800 Calculations complete Figure 59 The Position Export Utility 80 S The position data acquired from the sensor
82. ng in the upper right corner of the Oxymax Utility This will either flush the sensor with Nitrogen or with CO2 free air from the Soda Lime column 6 All the system to move air through the sensor for at least three minutes it may be helpful to have a timer present 7 Atthe end of the three minute period observe the Hydrogen reading Slowly adjust the Offset knob on the front of the sensor itself to move the H2 reading to 0 ppm If the display read 0 ppm to begin with it s advisable to slowly turn the knob clockwise until the value reads above 0 ppm Then slowly turn the knob counter clockwise until the value just barely switches back to 0 ppm 8 Use the black switch on the perimeter of the Offset knob to lock the current value into place 9 Under the Pump Valve heading in the Oxymax Utility now select the Ref Cal option 10 If your system has multiple tanks of Calibration Gas be sure that the Hydrogen Calibration Gas is connected to the calibration line 11 Turn the Calibration Gas tank on and check the overflow line at the base of the calibration line for overflow A small but perceivable excess of gas should be noticeable from this tube 12 Once the gas is on wait at least three minutes for the system to flush the old gas out of the Hydrogen Sensor and replace it with Calibration Gas 13 At the end of the three minute period slowly adjust the Gain knob on the front of the unit until the value for Hydr
83. nnunnnnnnnnnnnnnnnnna 108 Chapter 1 Introduction 1 1 Overview Columbus Instruments Comprehensive Lab Animal Monitoring System CLAMS brings together a diverse range of physiological and behavioral instrumentation into a simply controlled system capable of monitoring up to 32 lab animals CLAMS is based on Columbus Instruments original Oxymax open circuit calorimeter Oxymax serves as the central platform onto which clients may add activity feed monitoring feed control telemetry and more These elements are tied together using the CI Bus communication system that allows easy integration of all CLAMS components This manual contains instructions for the operation of the Oxymax for Windows software The CLAMS Hardware manual contains the specifications and function of the hardware components in the CLAMS family Since many CLAMS installations are considered custom not all sections of either manual will be relevant to every customer This manual when viewed completely covers material applicable to nearly every CLAMS user Note Screenshots and diagrams presented in this manual may be resized or edited to fit the format of the page Do not expect all figures to be verbatim representations of the actual Oxymax for Windows software 1 2 System Requirements The minimum system requirements for a computer to run Oxymax for Windows are e Windows based PC running Windows XP Vista 7 8 e 512MB RAM 1GB Recomm
84. nt the controller will adjust the temperature linearly across the duration of the step e Duration If the set point temperatures are the same this is the amount of time the controller will hold a constant temperature If they are different this is the period in which the temperature will change from Start to End This value is expressed in minutes and can set to any value between 1 minute and 60 days 86400 minutes e Temperature Schedule Start The date and time that the defined temperature schedule begins If the specified time is in the past the controller will immediately shift to the appropriate step in the schedule upon saving the schedule to the unit If the specified start time is set in the future the controller will remain at the current set point until the schedule begins e Loop Schedule When checked the specified schedule will loop back to the first step when the last step is completed When unchecked the controller will remain at the End temperature of the last step e Save to Controller Saves the specified settings to the selected Environmental Controller e Save to All Controllers Saves the current settings to all enabled Environmental Controllers The bottom of the Temperature Control settings pane contains a timeline that show the currently listed schedule with bright red representing higher temperatures and blue representing lower temperatures Each step in the schedule is bordered in white Manual Control
85. ntrollers support Oxymax integration out of the box Older Treadmill Control units may require a firmware update and modifications to the associated hardware configuration file Contact Columbus Instruments to discuss Treadmill Controller options 34 3 5 1 Basic Settings A custom treadmill schedule consists of three parts the warm up period the schedule itself and the cool down period When Oxymax initializes the specified hardware upon startup it sets the treadmill speed to the last used warm up speed which will be maintained until the start of the experiment schedule Note Computer Control of the Treadmill Controller Please note that in order for Oxymax to adjust the treadmill belt speed the Treadmill Controller must be set into either the Run or Accel position on the front panel Switching the Treadmill Controller to Stop at any point immediately stops the Treadmill After the software has assumed control of the unit the Treadmill Controller s controls for speed and electric stimulus will be disabled until the controller is Reset with the button in the lower left corner of the front panel Figure 24 shows an example of a treadmill experiment with some standard settings applied Treadmill Control Settings 220 sl Treadmill Controller Le SE at Load Save r Settings for Treadmill 220 Add Remove Copy Paste Step Speed m min Ramp s Hold s 1 10 0 30 60
86. ogen on the screen matches the concentration of Hydrogen in the Calibration Gas This value can either be found on a label on the tank or with a separate certificate 14 Use the black switch on the perimeter of the Gain knob to lock the current value into place 15 Turn off the calibration gas 16 Close the Oxymax Utility Now continue setting up and preparing for an experiment as you normally would The calibration procedures for the Oxygen and Carbon Dioxide Sensors will prompt you to calibrate two sets of sensors just remember that both displays are actually tied to the same set of sensors When calibrating the Methane sensor the Methane calibration procedure should work properly for the Group 1 sensor but will list calibration errors for Group 2 Ignore the errors regarding Group 2 The knobs on the Hydrogen Sensor should not be adjusted at this time regardless of the software s prompts When setting up Subject properties on the Experiment gt gt Setup screen notice that twice as many subjects appear as there are channels in the system For every subject example Subject 1 it appears in both channel 0101 and 0201 Both channels must be enabled in order to collect samples from all sensors properly Additionally the calculation of production and consumption rates to work out properly any changes in flow must also be made to both entries for a subject as well For example when adjusting the flow in Chamber 2
87. olumbus Instruments Paramagnetic Oxygen Sensor uses Oxygen s relatively high magnetic susceptibility to measure its concentration in a sample The sample passes through a magnetic field over a glass dumbbell filled with an inert gas The magnetically responsive oxygen causes the dumbbell to rotate and the degree of this rotation is used to measure the concentration of Oxygen in the sample Because of this mechanical implementation the sensor is very susceptible to motion and it s recommended when possible to place this cabinet on a static shelf or to lock the castors on any mobile shelving The Paramagnetic Oxygen Sensor requires two points for calibration and therefore requires two different tanks of calibration gas Usually pure Nitrogen is used as the offset gas and standard calibration gas 20 50 02 is used for the span gas Depending on the specifications provided at the time of purchase some systems use a higher concentration gas 19 00 02 for the offset gas Note Nitrogen replaces Soda Lime All of Columbus Instruments Carbon Dioxide Sensors require a two point calibration using a CO2 free gas as a zero gas In most cases that zero gas is created by using Soda Lime to remove Carbon Dioxide from room air Systems with Paramagnetic Oxygen Sensors usually use the same Nitrogen tank for zeroing both types of sensors 1 Click the CALIBRATE button under the O2 header to open
88. or do not adjust it again Turn on the calibration gas and adjust the pressure so that excess gas is apparent at the exhaust port 5 10 psi If the system uses Nitrogen for an offset gas instead of Soda Lime turn that tank on as well Otherwise check to see that the column of white granular Soda Lime is secured to requisite tubing Click NEXT to start the purging the sensors with the offset gas The progress bar at the bottom of the prompt will indicates the amount of time remaining during the purge period During this interval the same lines and gas sensors are being purged of their previous sample and replaced with CO2 free gas When the purging period is complete the system will measure the zero point of the Carbon Dioxide sensor During this time the Ready light on the unit will go dark for a few seconds this is normal After the zero point of the sensor has been established the system will begin purging the sensors with span gas When the purging period is complete the system will measure the gain necessary to calibrate the sensor After the gain has been calculated Oxymax will display a prompt listing the calculated offset and gain values necessary to the Carbon Dioxide Sensor s Turn off the flow of calibration gas es and click FINISH to commit this calibration to memory and implement it for the next experiment If the Oxygen sensor hasn t been calibrated yet just click FINIS
89. or exporting collected data Full use of this feature is discussed in 5 2 Exporting Data e Close Closes the currently open Experiment Configuration e Experiment e Setup Opens the Experiment Setup window for setting and configuring cage and data options This feature is discussed further in 3 2 e Properties Opens the Experiment Properties window used in configuring measurement options for the specified CLAMS hardware This feature is discussed further in 3 3 Measurement Configuration e Calibrate Opens the dialog for calibrating the system s gas sensors The calibration procedure is discussed in 4 2 Gas Sensor Calibration e Run Starts the currently open experiment This option is unavailable until a save location for experiment data has been selected e Stop Stops the currently running experiment e Add Event While an experiment is running this option allows the user to make a note of something that occurred in the experiment for later review e Tools e Feeding Refill Feeder Balances Opens the utility for refilling balance based animal feeders while an experiment is running Refill Feeder Load Cells Opens the utility for refilling load cell feeders while an experiment is running Feed Control e Open Feeders In systems with feed access control equipment this opens all of the feeders in the system e Close Feeders In systems with feed access control equipment this closes all of the feeders in the system
90. or is a differential gas sensor that compares a sample gas to a known assumedly fixed concentration reference gas The sensor only returns a ratio that represents the concentration of the sample gas to that of the reference gas The calibration procedure is designed to adjust the sensor such that when the same gas is used for the sample and reference the ratio returned by the sensor is one The Zirconia Oxygen Sensor has a small heating element that needs to warm up to about 725 degrees Celsius before calibrating if the Temp STB LED is lit the sensor is warmed up The calibration portion of the procedure does not actually use calibration gas at all Rather room air is supplied to both channels of the sensor and then adjusted to return a ratio of one Then calibration gas ata known concentration is supplied to the sample channel and the system uses the measured ratio to calculate the concentration of room air If the calculated value of room air falls into the normal range 20 85 20 98 then the sensor is considered calibrated Note CalGas settings for Zirconia Sensors The Zirconia Oxygen Sensor requires a specific O2 Zero parameter in the CalGas frame before starting a calibration Instead of specifying the zero gas concentration you have to specify the target ratio for comparing room air to room air This ideal ratio is 1 00 and should appear in the CalGas settings as depicted here
91. r C Sample 2 0 9 5 Nitrogen 0 0 4 0 Calibration Gas Source Pump On C Off Expansion Interface Yalye Select i 4 eee AIR FLOW EE WEE Air SAMPIT GE 02 c02 CALIBRATE CALIBRATE Figure 30 Calibration Window Most of the controls on the Oxymax Sensor Calibration Window are for display purposes only and cannot be controlled Systems with multiple sets of gas sensors will see multiple columns of Groups as depicted in Figure 31 The only Calibration Window components that are critical for calibration are the black and white gas concentration displays and the Calibrate buttons 46 S GT max Sensor Calibration Group 1 Cal Gas Group 2 Cal Gas Group 02 02 O2 20 95 20 95 CO2 CO2 CO2 0 040 ti EAZ AIR FLOW AIR FLOW AIR FLC ro E W Figure 31 Calibration with Multiple Sensors All of the gas sensors use one or more control calibration gases to calibrate or validate their measurements Either automatically or manually the sensors are tweaked to ensure proper measurement of one or two calibration gases Usually the main calibration gas consists of 20 50 Oxygen and 0 50 Carbon Dioxide Sensors with two gas calibrations usually also require Nitrogen Note Definition of a Calibration Gas The purpose of a Calibration Gas is to act as a known control for measurement by a sensor Calibration Gases are presented with two tolerances a mixture
92. r for 12 hours e Step 3 Opens the feeder and leaves it open for 12 hours or until the subject consumes 2 grams of food The controller remains in this step for the entire 12 hours regardless if the consumption limit is met or not e Step 4 Closes the feeder for 12 hours e Repeat Returns the feeder to Step 1 This schedule allows the animal to eat up to 3 grams for 12 hours during the first day and 2 grams during the same 12 hours on the next day This pattern repeats for the duration of the experiment Systems with multiple feeders in each chamber have the option of combining those balances together and are treated like one feeder Combined balances share feeding schedules and the limits in those schedules are evaluated using the sum of consumption from all combined balances in the same chamber 3 4 2 Linked Feeding The Linked feeding mode allows for subjects access to food to be limited by the consumption of other subjects Linked feeding can be set up to perform paired feeding experiments Within Linked feeding experiments feeders are placed into different groups and are assigned as masters or slaves within that group One or more masters within a group control the amount of food the slaves within that group can consume 29 Clicking the Linked option button under the Feed Control Settings frame inserts additional settings to the Balance Management table Group Master and Multiplier Balance
93. r of Visits Limit This value represents the total number of visits to the shock grid a subject can make before the grid is disabled A shock visit begins on a first shock and concludes when the animal leaves the shock grid This value is adjustable from between 1 and 999 visits 38 e Total Number of Shocks Limit This value represents the total number of shocks a subject will receive before the shock grid is disabled This vale is adjustable from between 1 and 999 shocks 3 5 4 Performing Treadmill Experiments After schedules have been specified for all connected Treadmill Controllers the rest of the Oxymax experiment can be configured as it normally would When the experiment is started Oxymax will again check that the switch on the front of the Treadmill Controller to Run or Accel so that it can take control of the treadmill motor If the Start Schedule Manually option was not selected before the experiment begins the Treadmill Controller will immediately proceed to the first step of the schedule Ifthe manual start option was selected the treadmill will continue at the warm up pace until the schedule is started by the user The option to start the treadmill schedule appears in the Experiment menu as Start Experiment Schedule Schedules set to repeat will continue to repeat until the experiment is stopped or the treadmill schedule specifically is stopped Any running treadmill schedule can be stopp
94. ration and present a loading screen as in Figure 5 Oxymax for Windows will automatically detect the serial communications COM port the CLAMS system is connected to Oxymax Initialization Initializing devices Please Wait Figure 5 Oxymax Initialization 2 2 Har tus While Oxymax for Windows is searching for CLAMS hardware it will display the Oxymax Hardware Status window that will remain suppressed unless the software fails to detect a required component If an error is detected the status window pictured in Figure 6 will appear The Oxymax Hardware Status window is divided into two tables that display a complete listing of the system s Oxymax and CLAMS hardware The refresh buttons in each frame are used to attempt to communicate with the listed hardware Oxymax Hardware contains a listing of all calorimetry related hardware CLAMS Hardware lists all hardware relating to feeding drinking activity and other features Oxymax Hardware Status CISI 2 System Sample Pump CISI OK SSP OK FlowM ax Device initialized successfully SampleM ax Device initialized successfully Address Status Description Opto Activity Monitor 160 OK ai devices initialized successfully Figure 6 Oxymax Hardware Status el ee 10 In Figure 6 Oxymax for Windows attempted to connect to the hardware of a simple four channel CLAMS system Successfully detected hardware will be listed as OK whereas u
95. reen Ready light should take 20 30 min e Feeding Fill feeders e Fill water bottles Volumetric Drinking Monitor Check liquid reservoir and make sure detection clips are secure to the sippers e Place subjects in applicable chambers Lids should clasp firmly and evenly e Check nut and ferrule fittings on lids to ensure a tight fit e Ifthe system adapts to both mice and rats check plastic flow restrictors in fresh air lines All restrictors should be present regardless if the chamber will be used in the current experiment All restrictors should be the same color e Ifthe system s sample flow is controlled manually set to 0 40 LPM as shown After warm up period calibrate gas sensors e Tools gt gt Calibrate e Set CalGas span values to and e Turn on all tanks of calibration gas to 5 10psi Excess flow should be apparent at T fitting Paramagnetic 02 Sensor Turn on Nitrogen tank to 5 10psi e If applicable check connections on the column of white granular Soda Lime e Click CO2 CALIBRATE to calibrate Carbon Dioxide Sensor Automatic CO2 Sensor If Power light is lit but Ready light is not call Columbus Instruments support Manual CO2 Sensor Adjust knobs to match displayed target values after purging lock calibration knobs when finished e Click O2 CALIBRATE to calibrate Oxygen Sensor Fuel Cell 02 Sensor If the Coarse Gain knob needs to be turned more than half of a complete turn in be
96. resholds are persistent settings that remain the same for any instance of the same experiment However some settings like subject identifiers and output file locations need to change each time the experiment is run These settings are presented on the Experiment Setup Window The Experiment Setup Window can be reached from the Experiment gt gt Setup menu after a blank data file has been created An example of an Experiment Setup Window appears in Figure 9 Output Data Filename 4C 0xymax Data 2012 11 30 CDTA Browse Set Default Backup Directory Browse Clear Subject ID Flow LPM Current Subject Subject 1 0 50 Label Ino Subject 2 Subject 3 Subject 4 Subject 5 Subject 6 Subject 7 Subject 8 Subject 9 Subject 10 Subject 11 Copy to All Subjects Figure 9 Experiment Setup Window Many of the settings required to start an experiment persist from experiment to experiment The only setting that the user must set before starting an experiment is the 1 5 Data Filename setting The Data Filename indicates the name and location that the data from the experiment will be saved Clicking Browse opens a dialog window to select a save location for experiment data Note Save Location It is recommended that experiment data be saved outside the Oxymax directory and out of Program Files entirely Windows considers Program Files a sensitive ar
97. ries of safety settings to warn the user of potential dehydration or malfunctioning sipper conditions These warnings are referred to as Under Pump and Over Pump warnings Under Pump errors are issued when the subject fails to drink a minimum amount of water over a given period of time This can either happen because the animal is unable to access the sipper tube or the drinking system has malfunctioned in some way clogged water line empty water reservoir etc The warning is logged either when a subject fails to drink at all for a certain period of time Under Delay or the subject drinks less than a certain amount Under Threshold over a period of time Under Period Over Pump errors are issued if the drinking monitor pumps more water than is realistic for a subject to drink Dirty sipper tubes can cause the system continuously pump water into the cage when the subject is not drinking If the system pumps more than a certain number of droplets of water into the cage Over Threshold during a given interval Over Period then Oxymax for Windows can log an error and possibly disable the water pump for that channel Over Disable The available drinking settings are depicted in Figure 14 G Experiment Properties Metabolic Feeding Drinking Activity Other Volumetric Drinking Error Settings Edit OverPump OverPump Under Pump Under Pump Under Pump Threshold Period Delay Threshold Period Cer af er Sr a IT Di
98. rogen for defining a sensor s zero point will always uses 0 00 as the concentration of both Oxygen and Carbon Dioxide The specified concentrations of the calibration span gas should be entered into the appropriate fields as they appear on the Certificate of Analysis that should accompany the calibration gas tank Click Apply to commit the new gas values to memory and then click Close to remove the Calibration Gas frame Some of the steps in the calibration procedure are common to all sensor types Before calibrating any sensor systems that use manual controls to adjust sample flow should be set to about 400 mL min The middle of the red indicator ball should float on top of the line that marks the 4 L min mark as in Figure 33 After adjusting the flow to 400 mL min do not re adjust the flow before starting an experiment Altering the sample flow requires the system be re calibrated Some CLAMS systems have digital flow controllers and do not need to be manually adjusted at all When turning on any tank of calibration gas the outlet pressure as indicated on the tank s pressure gauge should read between 5 10 psi for most systems Systems with multiple sets of sensors also have multiple sample pumps and will require a higher pressure to ensure a sufficient flow of calibration gas At the end of the calibration gas tubing is a T fitting Figure 33 with a small length of tubing on it that s used to exhaust exce
99. rols at that time on the standard Hue scale value between 0 and 255 e LED Saturation Saturation level of the LED color expressed as a percentage e LED Lightness Measure of LED intensity expressed as a percentage This is a measure of delivered power to the LED s which may not scale evenly compared to actual brightness 5 2 Exporting Data By default Oxymax for Windows stores experiment data in a proprietary format that cannot be read by third party software In order to investigate data in other applications the data must be exported into a format that is compatible with other spreadsheet software Microsoft Excel OpenOffice Apple Numbers Google Sheets etc All data recorded by Oxymax for Windows can be exported as Comma Separated Values CSV file which can be read by all major spreadsheet programs Oxymax for Windows can make CSV s based on either subject or measurement parameter Subject based CSV s contain every measurement recorded for a given subject parameter based CSV s contain every subject s data for a given parameter To automatically generate all of an experiment s subject based CSV s 1 Open up the desired experiment data file a The data generated from a just finished experiment is still open after stopping it b To open a previously saved experiment instead of clicking File gt gt Open Experiment Configuration after loading a hardware configuration click File gt gt Open Experiment
100. rror is not great yet it is worthy of note and of record The table serves to illuminate the charts recently prepared by Dr E F Du Bois which he presents in the article immediately following this weas yes zk i 0 293 i Equation 14 CV 3 815 1 232 RER 87 TABLE I Analysis of the Oxidation of Mixtures of Carbohydrate and Fat Porc ntago id gel ee he hed i heat Calories per liter Os R Q ee Ne irel Fat Poe ee Fat Number Logarithm O 2 3 4 5 6 0 707 0 100 0 0 100 0 4 686 0 67080 0 71 1 02 99 0 1 10 98 9 4 690 0 67114 0 72 4 44 95 6 4 76 95 2 4 702 0 67228 0 73 7 85 92 2 8 40 91 6 4 714 0 67342 0 74 11 3 88 7 12 0 88 0 4 727 0 67456 0 75 14 7 85 3 15 6 84 4 4 739 0 67569 0 76 18 1 81 9 19 2 80 8 4 751 0 67682 0 77 21 5 78 5 22 8 77 2 4 764 0 67794 0 78 24 9 75 1 26 3 73 7 4 776 0 67906 0 79 28 3 71 7 29 9 70 1 4 788 0 68018 0 80 31 7 68 3 33 4 66 6 4 801 0 68129 0 81 35 2 64 8 36 9 63 1 4 813 0 68241 0 82 38 6 61 4 40 3 59 7 4 825 0 68352 0 83 42 0 58 0 43 8 56 2 4 838 0 68463 0 84 45 4 54 6 47 2 52 8 4 850 0 68573 0 85 48 8 51 2 50 7 49 3 4 862 0 68683 0 86 52 2 47 8 54 1 45 9 4 875 0 68793 0 87 55 6 44 4 57 5 42 5 4 887 0 68903 0 88 59 0 41 0 60 8 39 2 4 899 0 69012 0 89 62 5 37 5 64 2 35 8 4 911 0 69121 0 90 65 9 34 1 67 5 32 5 4 924 0 69230 0 91 69 3 30 7 70 8 29 2 4 936 0 69339 0 92 72 7 27 3 74 1 25 9 4 948 0 69447 0 93 76 1 23 9 77 4 22 6 4 961 0 69555 0 94 79 5 20 5 80 7 19 3 4 973 0 69663 0
101. s can be assigned to any group between 1 and 128 There is no limit on the number of slaves or masters in any given group Figure 19 shows an example experiment where balances are placed into groups of four with two masters and two slaves in each group Balance Management 010601 010701 010801 010901 QO NM MN M HUUBRUURR Feed Control Schedule Group 1 Add zl Clear Copy Paste JV Repeat Step o Tiet ang Soom o o Figure 19 Linked Feeding Experiment When performing a linked feeding experiment schedules are specified for by all group masters rather than each individual balance All masters within a group share the same schedule According to Figure 19 the masters for Group 1 are permanently open for the duration of the experiment At any given moment the slaves from Group 1 are allowed to have consumed as much food as the average consumption of the masters from Group 1 If a slave s consumption equals or exceeds the masters average consumption its feeder is closed and remains closed until the masters average increases For example Table 1 represents the consumption and feeder statuses of all subjects in Group 1 as defined in Figure 19 a few hours after the start of an experiment Balance Consumed Controller 010101 Open 010201 1 00 Open 010301 2 10 Closed 010401 8 Closed Table 1 Group 1 Consumption Example 30 ee In this case the masters from Group 1 ha
102. s can be exported directly in a raw format which consists of every new position record collected during the course of the entire experiment at the selected sampling rate This raw data includes very small movements that may include rotating and grooming In general this raw data is still useful for analyzing time spent in particular zones in the chamber but is less useful for estimating distance travelled To get a better sense of ambulation the raw data can be filtered to remove some of the stereotypic data There are a handful of settings that can be adjusted before exporting the analyzed position data e Points to Average This is the number of successive beam break patterns that are added together to help remove some of the small rapid movements that don t contribute to distance travelled Any movements that are repetitive will return as the average position over these points e Fence Size Movements centered around a small area usually constitute stereotypic activity As part of the filtering process an adjustable fence is drawn around the subject s last location Any movement inside of the fence is not counted against the distance traveled and the animal s location is not updated Once the subject steps outside of the fence each new position is considered part of the ambulation until the animal remains still for a particular period of time the Fence Refresh Interval e Fence Refresh Interval Any time the subjec
103. s to and from the chamber and the concentration of inert gases N and N in each flow Equation 2 The concentrations can be calculated and one of the flows is measured so the other V for positive ventilation Vi for negative ventilation can be calculated ViNi bah Equation 2 The concentration of inert gases in a sample is assumed to be all gases that are not involved in respiration Equation 3 N 100 0 CO Equation 3 Assuming a positive flow system combining the previous equations yields the V for the system Equation 4 100 03 CO V V o i100 035 C030 Equation 4 The volumetric flow rates for consumption and production can then both be calculated The equations are arranged such that both consumption and production are presented as positive values Equation 5 Equation 6 Equation 7 HO subject V 02 V 020 Equation 5 VCO2 subject Kater ViC Ozi Equation 6 HCH subject VCH Mea Equation 7 Oxymax for Windows always presents consumption and production rates normalized to the subject s body mass in kilograms To convert the whole body rates into mass normalized rates just divide the whole body rate by the mass m of the subject in kilograms Equation 8 V sub ject m Veenorm Equation 8 The accumulated respiration figures A are calculated using the whole body rate and the amount of time that has elapsed since the last measurement t
104. sable Channel on Overpump Help Apply Figure 14 Drinking Experiment Properties e Volumetric Drinking Error Settings This is the table of settings that control error handling for the Volumetric Drinking Monitor The Channel column refers to a specific VDM sipper e Edit To make changes to the settings table select one or more VDM channels from the table and use the drop down controls to change the settings for those channels Hit Apply to commit those changes to the selected channels 3 3 4 Activity Properties Activity measurements are presented as the number of times the subject breaks an infrared beam during the course of a given interval Chambers can monitor these beam breaks in up 24 to two axes just above the chamber floor and another axis near the ceiling used to measure rearing activity Animal activity can be recorded in a number of different ways by Oxymax By default Oxymax for Windows takes beam break activity measurements at two frequencies in sync with the metabolic measurements and at a higher independent rate The high frequency rate can be set by the user while the low frequency rate is determined by the metabolic measurement time and number of enabled cages For systems that can measure beams breaks in both axes on the lower plane the X and Y axes Oxymax can also monitor animal position over the course of the The activity settings for an experiment are presented in Figure 15 Le Experiment
105. self Usually this is the same speed as the warm up speed Schedule Distance The distance the treadmill will travel by the end of one complete iteration of the defined treadmill schedule This value is updated when any of the schedule parameters are changed 3 5 2 Treadmill Schedules The treadmill schedule itself is configured much in the same way as a feed control schedule When the schedule starts Oxymax progresses each treadmill through the series of defined steps each of which consist of three properties Speed m min The target speed that the treadmill will reach by the end of the step The selected speed is range checked against the limitations of the treadmill motor defined in the Hardware Configuration file Ramp sec The period of time during which the treadmill will accelerate to the specified speed The acceleration rate is dependent on the ramp time as well as the previous speed of the treadmill When the ramp time is set to zero the Treadmill Controller will proceed to the requested speed as fast as mechanically possible Hold sec The amount of time the treadmill will remain at the requested speed before proceeding to the next step in the schedule When the hold time is set to zero the schedule will proceed to the next step as soon as the treadmill has reached the target speed The buttons above the schedule table as shown in Figure 25 can be used to make changes to the currently selected treadmill s schedu
106. ss gas When Sample Flow the calibration tank is turned on a noticeable flow of gas should be apparent from the exhaust tubing It s always safer to adjust the calibration gas toa pressure that s too high than too low If the calibration gas pressure is too low the system can t use pure calibration gas and the sensors will not calibrate properly After the system has warmed up for the appropriate length of time clicking the CALIBRATE buttons under the O2 CO2 or OTHER headers will initiate the appropriate calibration procedure Exceptions to the aforementioned rules will appear in the following sections regarding the standard gas sensors if your gas sensor is not listed in this manual follow Oxymax s on screen instructions or refer to additionally provided literature to calibrate your sensor 4 2 1 Fuel Cell Oxygen Sensor Power Oxygen Sensor O Extended Range Calibration l amp Columbus Instruments Standard Range Coarse Gain Fine Gain COLUMBUS OHIO 43204 USA Calibration LI LI Figure 34 Original Fuel Cell Oxygen Sensor 02 GAS ANALYZER O Pm O DATA A O our DATA IN Fine Gain O O di er SAMPLE SAMPLE DUT IN PWR s Columbus Instruments E VENUE SS cou Figure 35 CI Bus Fuel Cell Oxygen Sensor The Columbus Instruments Fuel Cell Oxygen Sensor uses a fuel cell battery to genera
107. stems ship with a default configuration for each Hardware Configuration The Experiment Configuration is named to match the Hardware Configuration with which it is meant to be used The Experiment Configuration contains measurement interval lengths units subject information and other settings for an experiment To open an Experiment Configuration file click Open Experiment Configuration from the menu and select the desired file Oxymax creates a blank Experiment Window configured with the options specified in the selected Experiment Configuration as in Figure 8 Note Every CLAMS system is different The blank data file you see with your configuration will most likely not look exactly like Figure 8 The example configuration may contain more or fewer measurement options that your system T File Experiment Tools view Windows Event Li Figure 8 Blank Data File After opening an Experiment Configuration the menu contains options for the configuration and control of the currently opened experiment Not all of the options listed here need to be customized in order to properly run an experiment Additionally some options will only make themselves available if the CLAMS system supports them e File e Save Experiment Saves any changes to the currently open Experiment Configuration e Save Experiment As Save the current Experiment Configuration under the chosen filename T3 e Export Contains options f
108. subject s RER can exceed 1 0 during intense exercise or while synthesizing fat from carbohydrates e Heat The measure of energy expenditure for a subject This is presented either as a whole subject or mass normalized rate either in kilocalories or kilojoules e Flow The rate of fresh air delivery to the chamber in liters per minute LPM The target fresh air flow rate is based mostly on subject mass but typical values are e Mice 0 50 0 60 LPM e Rats 2 0 3 5 LPM e Pressure This is the pressure in the sample loop as reported by some models of gas sampler Not all products contain or require the requisite components to record this value Itis not used in the computation of metabolic figures but indicates the state of the sampler during the experiment 5 1 2 Feed Measurements Oxymax for Windows makes feeding measurements in several methods to suit a variety of research needs Feeding data is recorded in three different methods e Synchronously with the metabolic data Consumption data for a chamber is measured and logged at the same time as the metabolic data for that chamber This data appears under the Feeding tab and is stored in the CDTA file e Asynchronously as the animal interacts with its food Each time the balance is disturbed the software keeps track of how long the animal disturbs the feeder and how much food it removes This data appears under the Feeding Events tab and is stored in the BDTA file e Perio
109. t or dependent control of access to every feeder in the system Subjects or particular diets within a subject s chamber can be set to follow unique schedules or cyclical patterns The system can be run in one of two modes Standard or Linked Consumption limits for subjects can be specified in grams energy consumed or consumption as a percentage of animal body mass The Feed Control settings pictured in Figure 13 only appear if the system is equipped with a Columbus Instruments FeederMax control system The Feed Control Settings frame in the corner of the Feeding tab allows for the selection of basic control settings Custom control schedules can be saved and loaded from the Feed Control Settings File frame The Balance Management table displays all balances for every enabled chamber in the system along with mode specific settings The Feed Control Schedule table displays the schedule for the currently selected balance or group of balances from the Balance Management table 3 4 1 Standard Feeding The Standard feeding mode allows for each feeder to be controlled independent of the other feeders in the system The exception to this independence is in cases where subjects have access to multiple feeders Feeders in the same chamber have the option of being combined to share limits and access conditions Under the Balance Management table clicking the ID in the left most column for a feeder displays the schedule for that
110. t remains still for this period of time the fence is redrawn around the animal Once settings have been selected for the export clicking Recalculate will re analyze the position data for each subject Chambers with empty data files or chambers without both X and Y axis sensors will not be analyzed After analysis both the raw position data acquired and the analyzed results can be exported as CSV files The CSV files are generated for all eligible chambers and are automatically saved in the same directory as the rest of the experiment data After the data has been analyzed each position record can be plotted in the space on the right Select a subject in the table and then use the slider to move through the experiment The animals location at the printed time appears as a circle If that record was made as a result of motion a line indicating the animal s previous position is drawn The larger circle for the animal s position indicates it was rearing at the time 5 4 1 Position Indexes Position data is presented as a set of coordinates in a unit less index system The subject s center position on any axis can be calculated to any beam on that axis or the space in between beams if an even number of contiguous beams are broken In this example there are eight beams as indicated by pink lines on each axis The software can calculate the subject s center position to be any of the intersections of the beams as indicated by
111. t to the list of displayed measurements on the right The Up and Down buttons can be used to re order the displayed measurements 4 The list on the right contains every chamber in the system Check the box next to display their data on the graph 5 Click Apply to generate a plot featuring the selected subjects and measurements Figure 53 shows a plot based on the parameters selected in Figure 52 BEE one vents Metabolic Feeding CTIE ole Ef p Activity Temp HR Environment Event Log Warning Log YO2 MLIKGIMIN GRP 1 CH1 GRP 1 CH2 Start 9 21 2009 4 11 25 PM Interval 135 Measuring Reference 02 0 00 CO2 0 000 DI Figure 53 Example Data Graph The legend for the graph is on the right side of the window and lists each of the selected chambers and the color that chamber s data appears in The colors for both the subjects and the background can be altered to improve readability On the Graph Setup frame right clicking any of the colored bars on the subject list displayed the color selection window in Figure 54 CN 21x Basic colors Zl HETEN a0 HHHH ET HEHH E HEHEH Define Custom Colors gt gt Figure 54 Color Selection Window 68 S Chapter 5 Data 5 1 Understanding Data This section outlines and explains the measurements made by Oxymax and how to interpret the results Oxymax for Windows stores experiment
112. te small amounts of electricity using Oxygen The rest of the sensor circuitry determines the Oxygen concentration from the signal the fuel cell generates Without the presence of Oxygen the cell doesn t generate electricity and the system interprets that signal as 0 0 As a result the Fuel Cell Oxygen Sensor only requires one fixed point to properly calibrate the sensor 49 50 Click the CALIBRATE button under the 02 header to open the Oxymax 02 Sensor Calibration prompt The prompt contains abridged reminders of the instructions laid out in this manual If the sample flow s for the system is controlled manually on the front of the sample pump adjust the sample flow to 400 mL min If you ve already adjusted the flow before calibrating another sensor do not adjust it again Turn on the calibration gas and adjust the pressure so that excess gas is apparent at the exhaust port 5 10 psi Click NEXT to start the purging the sensors with the calibration gas The progress bar at the bottom of the prompt will indicates the amount of time remaining during the purge period During this interval the same lines and gas sensors are being purged of their previous sample and replaced with pure calibration gas When the purging period is complete the prompt will display the target Oxygen concentration for the calibration gas Use the Coarse Gain and Fine Gain knobs on each Oxygen sensor to adjust the concentrat
113. terval the same lines and gas sensors are being purged of their previous sample and replaced with CO gt free gas 5 When the purging period is complete the prompt will display the target Carbon Dioxide concentration for the offset gas Use the Offset knob on each Carbon Dioxide sensor to adjust the concentration displayed on the Oxymax Sensor Calibration Window until it matches the target concentration The measured concentration will lag adjustments to the knobs by a couple of seconds so wait for the value to settle before continuing The displayed value s will turn green when the measured value is within 0 001 of the target concentration 6 After the zero point of the sensor has been established click NEXT to move to the next step of the calibration 7 As the prompt instructs make sure the calibration gas is turned on and providing sufficient flow Click NEXT to start purging the sensor with the calibration gas 57 8 When the purging period is complete the prompt will display the target Carbon Dioxide concentration for the span gas Use the Gain knob on each Carbon Dioxide sensor to adjust the concentration displayed on the Oxymax Sensor Calibration Window until it matches the target concentration The measured concentration will lag adjustments to the knobs by a couple of seconds so wait for the value to settle before continuing The displayed value s will turn green when the measured value
114. the Oxymax 02 Sensor Calibration prompt The prompt contains abridged reminders of the instructions laid out in this manual 2 Ifthe sample flow s for the system is controlled manually on the front of the sample pump adjust the sample flow to 400 mL min If you ve already adjusted the flow before calibrating another sensor do not adjust it again 3 Turn on both tanks of calibration gas and adjust the pressure so that excess gas is apparent at their exhaust ports 5 10 psi 51 10 Click NEXT to start the purging the sensors with offset gas The progress bar at the bottom of the prompt will indicates the amount of time remaining during the purge period During this interval the same lines and gas sensors are being purged of their previous sample and replaced with pure Nitrogen When the purging period is complete the prompt will display the target Oxygen concentration for the offset gas Use the Coarse Offset and Fine Offset knobs on each Oxygen sensor to adjust the concentration displayed on the Oxymax Sensor Calibration Window until it matches the target calibration concentration 0 00 for Nitrogen The measured concentration will lag adjustments to the knobs by a couple of seconds so wait for the value to settle before continuing The displayed value s will turn green when the measured value is within 0 01 of the target concentration After the measured value matches the target concentration c
115. the value to settle before continuing The displayed value s will turn green when the measured value is within 0 0002 of the target ratio After the measured ratio matches the target ratio click NEXT to begin purging the sensors with calibration gas At the conclusion of the purging period Oxymax will calculate and display the concentration of room air based on the calibration gas reading If the measured concentration falls into the acceptable range for well mixed room air 20 85 20 98 the calibration is considered successful The measured value for room air will be used as the O2 In value for all samples during an experiment for more on what this means see 5 1 1 Metabolic Measurements Click NEXT to display the finalization instructions If all gas sensors have been calibrated turn off the flow of calibration gas and click FINISH Otherwise just click FINISH and proceed immediately to calibrating other sensors 4 2 4 Automatic Carbon Dioxide Sensor Power Ready Carbon Dioxide Sensor O O Columbus Instruments amp r 950 NORTH HAGUE AVENUE COLUMBUS OHIO 43204 USA Figure 39 Original Automatic Carbon Dioxide Sensor CO2 GAS ANALYZER e O PWR o IN O READY DATA O N A DATA Vv O ou PWR Columbus Instruments 930 NORTH HAGUE AVENUE Figure 40 CI Bus Automatic Carbon Dioxide Sensor The Colu
116. tible with animal position monitoring Features like Running Wheels have the 65 66 potential to block most if not all of the beams on an axis in the chamber Position monitoring is only available when both X and Y axis sensors are largely unobstructed 4 6 Graphing the Current Experiment Oxymax for Windows comes with a rudimentary graphing utility for quickly observing trends in up to two measurements while an experiment is running The same utility can be used to generate graphs of data from completed experiments To create a graph 1 Click the Graph tab at the top of the current data file Because the graph has not yet been setup the frame will be empty except for the message No Data Available 2 To select data for display click the left most icon in the toolbar above the blank frame When the mouse cursor rests above the icon a tool tip will appear that reads Select Data The Graph Setup panel will appear as in Figure 52 CLECE HLE gt Graph Setup GRP 1 CH 2 GRP 1 CH 3 GRP 1 CH 4 Stripe START Duration SelectALL ClearALL 7 NONE VW NONE z Colors mmm be Help Apply Close Figure 52 Graph Setup Panel 3 The available measurements in the current experiment appear in the left table under the same categorical headings as the data file s tabs Select a measurement in the list and click gt gt to copy i
117. time will cause Oxymax to calculate what step the feeder would be on if the experiment had actually been started at the specified time For example if the day and night schedule specified by the settings in Figure 17 was set to start at 07 00 7 00 am and the experiment was started at 1 00 pm the feeders would open at the start of the experiment and remain open until 7 00 pm Even though the feeders would only have actually been open for 6 hours the schedule proceeds as if it had actually been started at the specified time If the selected schedule start time is later than the experiment start time all feeders remain in the current state until the specified time 3 4 5 Duplicating Schedules Complicated or frequently used schedules can be copied from one feeder and applied to others Clicking the Copy button above the currently selected Feed Control Schedule will put that current schedule onto the clipboard for use on other feeders Select another feeder and click Paste to replace that feeder s current schedule with the one on the clipboard When configuring a linked experiment schedules cannot be pasted onto slave feeders 3 4 6 Saving and Opening Feed Control Settings After a set of feeding schedules for an experiment has been created those settings can be saved to the hard disk for use on other experiments This is recommended for any group of feed control settings that will be used frequently To save the current sett
118. to something like 5 LPM both 0102 and 0202 must read 5 Due to the comparatively long response time of the Hydrogen Fuel Cell the settle times for both reference and cage measurements needs to be set to at least three minutes instead of the usual one to two minutes H 4 Reading Data In order to add support for a fourth gas sensor to Oxymax the hardware configuration files have been modified to show two chambers for every one chamber that actually exists in the system Each actual chamber then has two entries for every measurement For each chamber XX there will be two sets of measurements listed for channels 01XX and 02XX Many of the listed measurements will be duplicated with very small variances the following table indicates the locations in the dataset where each gas is recorded BCEE Methane Jm The other base metabolic measurements should all be taken from the Group 1 01XX channel for each chamber 105 106 107 108 Appendix I Quickstart Experiment Checklist Power on the CLAMS system e Check Power LED s for all boxes e Feeding Urine Turn on all balances Feeding Put feeders on balances before powering them on Run Oxymax for Windows to start warm up e Select hardware configuration e Allow software to initialize hardware Allow system to warm up e About one to two hours e Zirconia O2 Sensor Wait for temperature to reach 725 degrees C e Automatic CO2 Sensor Wait for blue or g
119. tolerance and an analysis tolerance The mixture tolerance implies how precisely the gas was mixed compared to the requested concentration The analysis tolerance is a measure of how accurate the label on the tank is compared to the actual concentration Columbus Instruments recommends the use of Calibration Gas that is mixed to a tolerance of 2 of the requested concentration and analyzed to within 0 02 absolute or 1 of reading whichever is smaller Some gas suppliers refer to these specifications as a Primary Standard gas Before calibrating make sure that Oxymax has the proper calibration gas concentrations registered for the gases that will be used during calibration To set the calibration gas concentrations click on the Cal Gas button next to the sensor group label to display the calibration gas settings frame as in Figure 32 Oxymax assumes the same tanks of calibration gas will be used on all sets of gas sensors so the values do not need to be defined for each group 47 Group lr 02 m Calibration Gas Group 1 02 Zero ooo 02 Span 20 50 4 C02 Zero 0 000 Z CO2 Span 0 500 4 CH4 Zero 0 ppm CH4 Span 0 ppm Close Apply Figure 32 Standard Calibration Gas Definitions The Calibration Gas frame shows two concentrations for three different gases Oxygen Carbon Dioxide and Methane Systems without Methane sensors will be unable to edit Methane values Applications that use Nit
120. tween each calibration the cell is near the end of its life call Columbus Instruments to order a replacement Paramagnetic 02 Sensor Adjust knobs to match displayed target values after purging lock calibration knobs when finished Zirconia 02 Sensor Adjust knob to achieve a ratio of 1 000 lock knob when finished e Turn off all calibration gas Configure experiment e Select an experiment configuration template File gt gt Open Experiment Configuration e Fill out subject roster Experiment gt gt Setup Check the boxes for enabled rows Enter in animal weights Set fresh air flow subjects sharing a flow controller will have identical flows e Mice 0 50 LPM for 20g subjects 0 60 LPM for 40g subjects e Rats 2 0 LPM for 200g subjects 3 0 LPM for 500g subjects e Choose a location on the hard disk to store experiment data Data Filename e Choose a location on a network drive to backup experiment data Backup Directory e Telemetry Register E Mitter transmitters to chambers Assign E Mitters e Click Apply when finished e Verify experiment settings Experiment gt gt Properties Choose reference method based on number of enabled cages 5 1 1 Metabolic Measurements Feed Control Setup feeding schedule 3 4 4 Feed Control Scheduling Run experiment Experiment gt gt Run e Monitor first few readings to verify system is working properly e Verify reasonable In gas concentrations 20 86 20 96
121. ulation of sleeping bouts e Sleep Threshold The conditions for qualifying inactivity as a sleeping bout Epochs The number of consecutive samples that have fewer than the specified number of beam breaks to qualify as a bout Activity Threshold The most beam breaks that can occur in an epoch and still qualify as a sleeping bout e Analysis Window Specifies a period of time within the experiment for analysis e Light Dark Cycle Specifies the pattern for qualifying bouts as having occurred during the day or at night e Save Configuration Saves the current sleep detection parameters for use with other data e Update Recalculates sleeping bouts based on new settings e Statistics Calculations related to the sleeping patterns of the selected subjects e Select Cages Allows for the selection of specific cages for statistical analysis e Generate CSV File Creates a CSV file contains statistics for each selected subject Include sleep records in CSV When checked the sleeping bouts table will be included in the generated CSV 79 5 4 Position Analysis CLAMS systems that are equipped with both X and Y axis activity sensors have the option of collecting animal position data as well as the standard beam break data Columbus Instruments has a range of beam based activity monitors in different forms but they can essentially be broken down into two product types on the Oxymax Hardware Status 2 2 Hardware Status wi
122. ure 49 Load Cell Utility 2 Each of the load cell feeders in the system are displayed in the table at the top of the form e Each cell in the table contains the chamber ID the corresponding load cell belongs to and the current reading from that load cell e Load cells that are not currently enabled will have the word Disabled appended to the end of the reading e The background color of the cell indicates the load cell s current stability Green cells are stable red are unstable e Urine collecting cells will also be reset when the controller is re enabled it is wise to empty or replace urine receptacles when the feeders are refilled 3 Ifthe system is also equipped with optional feed control hardware it might be necessary for the control system to be a particular state while refilling feeders If 63 64 this is the case select the state the controller should be in while refilling from the During Refill dropdown option Feed Control Options During Refill Do Nothing v Figure 50 Feed Control Refill Options e The controller will return to the state determined by the currently running feed control schedule at the conclusion of the refill 4 Feeders can only be successfully refilled if taken out of service while they are not being disturbed Once the subject is away from the feeder being refilled wait for the corresponding cell in the table to turn green and then uncheck the corresponding Enable checkbo
123. use P01 and press OK to select Select Auto The light cycle should begin at this point sm means Appendix G Calibrating Modular Treadmills Current models of the Columbus Instruments Metabolic Treadmill can be recalibrated in the field by use of the included speed sensor Calibration of the treadmill belt speed should be performed every three months It is also helpful to calibrate the unit whenever switching between species 1 Ensure the Treadmill Controller front panel toggle is set to STOP 2 Place speed sensor on treadmill belt and fasten it with the screw nut as shown 3 Using the dial on the Treadmill Controller adjust the speed to 88 7 m min 4 Set the electric stimulus frequency to 2 Hz on Treadmill Controller front panel 5 Push the ODOMETER RESET button to initialize calibration procedure The LCD on the front panel will read AutoCal Speed as shown 6 Push the ODOMETER RESET button a second time to advance to belt calibration start The LCD on the front panel will read Run to Start as shown 7 Set toggle to RUN to begin the series of belt speed and acceleration tests 8 Upon belt calibration completion the most recent calibration is automatically saved and speed sensor counts OD counts are displayed Obtaining count values between 4000 5000 indicates calibration was successful 9 To exit the calibration press the ODOMETER RESET button A Calibrat
124. ve eaten an average of 2 00 grams each 3 00 grams 1 00 grams 2 masters The slaves both ate to that limit and slightly above The system cannot immediately restrict access to food when a slave passes a limit because the balance has to stabilize for a second before Oxymax can determine if the limit has been reached If the subject using Balance 010201 consumed another quarter of a gram of food the average for the masters would rise to 2 125 grams The consumption on Balance 010301 would then be below the masters average and the feeder would open as illustrated in Table 2 Balance Consumed Controller 010101 010201 010301 010401 Table 2 Group 1 Consumption Example continued 3 4 3 Feeding Limits Consumption limits are specified in grams by default but can also be stipulated by energy consumption or consumption as a percent of subject body mass Selecting Energy from the Limit options as in Figure 20 adds an additional setting to the Balance Management table Energy g PUTTIN ED Pe Te Lei J Limit Minimum Valid Weight g 0 01 Feed Control Settings File Start Time Open Save Experiment Stat Cf 00 00 Balance Management 010501 Figure 20 Feeders with Two Different Diets Figure 20 shows that the feeders in this experiment are filled with two diets with different calorific content Note Energy has no units The term energy is used instead of a specific unit for energy
125. wledge These security features also inadvertently disrupt intended behavior in legacy software like Oxymax for Windows To ensure that Oxymax for Windows behaves as expected UAC should be disabled for computers that communicate with CLAMS hardware Windows Vista Go to Control Panel gt gt User Accounts and select Turn User Account Control on or off Click Continue if prompted by Windows Uncheck the box that says Use User Account Control UAC to help protect your computer and click OK Windows 7 8 Go to Control Panel gt gt User Accounts and select Change User Account Control settings Click Continue if prompted by Windows Move the vertical slider down to Never notify and click OR 1 4 Installation The Oxymax for Windows installation disc contains configuration information that is specific to the system or systems for which it was created If the disc is misplaced Columbus Instruments keeps all customer Hardware Configurations on file contact Columbus Instruments for a new installation disc If installing from a CD the installation disc may or may not automatically load the Oxymax for Windows installer upon insertion depending on the computer s settings If it does not click Start gt gt Run and select E setup exe where E is the letter of the computer s CD ROM drive If installing from a downloaded package unzip the contents and then run s
126. x for Windows version 4 51 or later If you have received an updated installation disc or configuration files with the Hydrogen Sensor be sure to install the latest version of each before proceeding H 3 2 Pre Experiment Unless otherwise instructed typical warm up times for an Oxymax system from power on to calibration range from about one to three hours However the Hydrogen Sensor requires eight hours of power before it can be calibrated It is therefore advises that the system be powered on and the software initialized the evening before the start of a scheduled experiment After the software has been initialized and the sample pump is turned on it s important to check the sample flow on the front of the System Sample Pump The red ball in the flow meter that marks the sample flow should hover directly between the marks indicating 0 4 and 0 5 LPM This indicates a sample flow of about 0 45 LPM H 3 3 Calibration After the system has been turned on and the software has been initialized for at least eight hours the Hydrogen Sensor is ready to be calibrated Since there is no built in calibration routine for the Hydrogen Sensor it must be done manually using the Oxymax Utility The Oxymax Utility can only be accessed before opening an experiment configuration so it must be done first To calibrate the Hydrogen Sensor 1 Select Oxymax Utility from the Tools gt gt Utilities to open the window shown oxymax Utility 21x
127. x in the table The word Disabled will appear next to the corresponding feeder value in the table and if present the feed control system will move to the selected state e Given the overhead involved in removing and refilling load cell feeders it s advisable to refill them completely one at a time However the utility does offer the option to Disable All and Enable All feeders at once should it be necessary to do so 5 It s at this time the feeder can be removed and refilled The procedure for removing and replacing the feeder in the chamber will vary by chamber design Note that depending on the size and design of the feeder it may be necessary to disregard metabolic data collected while the feeder is out of service due to the opening created in the chamber by the feeder s removal e The status of out of service feeders will be logged as part of the data but it is also advised that an event is registered manually for clarity 6 Once the feeder has be reassembled and reattached to the chamber wait for the load cell to stabilize and then check the Enable box to tare and enable it for service 4 5 Monitoring Real Time Position Position data can be recorded at a high sampling rate on systems with X and Y Axis activity sensors Selecting the sampling rate for the compatible activity monitors is explained in 3 3 4 Activity Properties The calculated center position of an animal in its chamber can be observed in real
128. xt channel in which at least one group had an enabled chamber For this reason the most efficient way to organize a small quantity of subjects across a multi group system would be to spread them evenly over the lowest numbered channels 0101 0201 0301 0102 0202 0302 etc All of the metabolic measurements are presented by Oxymax for Windows in shorthand form The following list details all of the measurements taken by Oxymax The equations behind the calculations of derived measurements are presented in 5 4 Position Analysis e Gas Measurements Oxymax for Windows records the same type of measurements for each gas Oxymax monitors For simplicity s sake rates for gases that are consumed Oxygen and those that are produced Carbon Dioxide Methane etc are both presented as positive measurements In the explanation of measurements below x takes the place of O2 Oxygen CO2 Carbon Dioxide or CH4 Methane e Vx The rate of gas consumption production normalized to animal body mass The units for this measurement are selectable but are typically displayed in mL kg hr 71 e IN The measured concentration of room air taken at the last reference The generally excepted values for room air concentrations are Oxygen 20 88 20 96 Readings greater than 20 97 are usually the result of an insufficiently long warm up period or in the case of the Fuel Cell Sensor a significant increase in ambient temperature sin
129. y and moved toward the front of the cabinet in order to slide the tube drier under the stainless steel fittings Once the rear clamp is secured the two sides will lay almost naturally into the remaining clamps Secure these and place the lid back onto the cabinet securing it with the two screws removed at the beginning of these instructions 97 98 Appendix F Environmental Enclosure Light Controller Setup The following are the instructions for the first time setup of the Intermatic Incubator Light Controller 1 2 E Hi BP Ww 8 9 Press the Reset button Set the time format to 24 Hour or AM PM by using the buttons and press OK to select Set the current hour using the buttons and press OK to select Set the current minute using the buttons and press OK to select Select current day of the week using the buttons and press OK to select Check the failsafe dials in the refrigeration cabinet a The Low Failsafe should be at 0 b The High Failsafe should be at 50 Set P01 referenced on Page 10 of the Intermatic Instruction Manual a Depress the reset button P01 should blink Press OK Select P01 and press OK Use to select the hour to turn on and press OK to select Use to select the minute to turn on and press OK to select Use to select the hour to turn off and press OK to select Use to select the minute to turn off and press OK to select i Use to select what days to