Home
InfiniiVision 4000 X-Series Oscilloscopes
Contents
1. The new InfiniScan Zone touch trigger if you can see it you can trigger on it by just drawing a box 5 in 1 instruments redefines the integration experiences oscilloscope channels digital channels serial protocol analysis dual channel WaveGen and DVM All features are fully upgradeable including bandwidth Industry leading coverage of serial protocol including USB 2 0 trigger and decode Industry s first dual channel WaveGen function arbitrary generator now allows you to generate differential clock and data two channel modulation and IQ signals Modulation of any signal is also included 20 The class leading 7 5 GHz upgradeable bandwidth expands your application coverage including USB 2 0 hi speed signal integrity testing Agilent Technologies 1 20 Serial 1 USB Time PID Addr Endp DATARO t 00 FF FC 00 C Eh A DATA1 CO i 00 02 00 00 hex Waveform Generator 1 Modulation Menu Modulation Type FM us em t lt E Both USB keyboard and mouse are supported for additional ease of use InfiniiVision O us cy fat MSO X 4154A 1 5 GHz Mixed Signal Oscilloscope 5 GSa s 50 00us T JAluus ister Menu Data Fr CRC Errors EBSE 7BSE Frequency 167 4 sine High Z Mc FM Freq FM De 1 000kHz 1 000kH E Industry leading 7 million waveform Docking panels with the capacitive per second update rate minimizes the touch screen add a
2. N2750A InfiniiMode differential probe 1 5 GHz 700 fF 200 kQ with AutoProbe Optional N2790A differential active probe 100 MHz 1 4 kV with AutoProbe Optional N2791A differential active probe 25 MHz 700 V Optional N2792A differential active probe 200 MHz 20V Optional N2793A differential active probe 800 MHz 15V Optional 1147B AC DC current probe 50 MHz 15 A with AutoProbe Optional N2893A AC DC current probe 100 MHz 15 A with AutoProbe Optional N2820A 2 channel high sensitivity current probe 50 uA 5A Optional N2821A 1 channel high sensitivity current probe 50 uA 5A Optional Step 4 Add the final touches Recommended accessories 4000 X Series GPIB to LAN external adapter N4865A Rack mount kit N2763A Soft carrying case N2733A Hard copy manual N6455A MET CAL procedures CalLab Solutions www callabsolutions com products Agilent 23 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics DSO and MSO 4000 X Series oscilloscopes 4000 X Series specification overview 4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A Bandwidth 3dB 4 channel real time bandwidth Calculated rise time 10 90 DSOX Input channels MSOX 200 MHz 350 MHz 500 MHz 1GHz 1 5 GHz 200 MHz 350 MHz 500 MHz 1 GHz 1 GHz lt 1 75 ns lt 1ns lt 700 ps lt 450 ps lt 300ps 2 4 2 4 2 4 4 4 2 16 4 16 2 16 4 16 2 16 4 16 4 16 4 16 Maximum sample rate Maximum memory depth Display size and typ
3. transient overvoltage 800 Vpk 100 mV 3 of threshold setting 10 V about threshold Threshold selections User defined threshold range Maximum input voltage Threshold accuracy Maximum input dynamic range Minimum voltage swing 500 mVpp Input impedance 100 kQ 2 at probe tip Input capacitance 8 pF Vertical resolution 1 bit Denotes warranted specifications all others are typical Specifications are valid after a 30 minute warm up period and 10 C from firmware calibration temperature 1 mV div and 2 mV div is a magnification of 4 mV div setting For vertical accuracy calculations use full scale of 32 mV for 1 mV div and 2 mV div sensitivity setting 1 5 GHz real time bandwidth in half channel mode or full channel equivalent time mode 24 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Horizontal system analog channels Time base range Time base accuracy Time base delay time range ele niga Post trigger Channel to channel deskew range A Time accuracy using cursors Modes XY 4022A 4024A 4032A 2 ns div to 50 s div 4034A 4052A 4054A 4104A 4154A 1 ns div to 50 s div 500 ps div to 50 s div 10 ppm Greater of 1 screen width or 200 us 400 us in interleaving mode 1 s to 500 s 100 ns 0 001 of reading 0 16 screen width 30pS Main zoom roll XY On channels 1 and 2 only Z Blanking on Ext Trigger Input 1 4 V threshold Bandwidth
4. Denotes warranted specifications all others are typical Specifications are valid after a 30 minute warm up period and 10 C from firmware calibration temperature 25 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Trigger system Trigger sources Trigger modes Trigger coupling Trigger holdoff range Trigger sensitivity internal Trigger sensitivity external Trigger level range Normal Auto Single Force DC AC HF reject LF reject Noise reject 200 MHz 1 GHz 1 5 GHz 1 6V 8V Any channel External Analog channel 1 4 digital channel D0 D15 line external WaveGen 1 2 or Mod FM FSK Requires trigger event for oscilloscope to trigger Triggers automatically in absence of trigger event Front panel button that triggers only once on a trigger event Press Single button again for oscilloscope to find another trigger event or press Run front panel button to trigger continuously in either auto or normal mode Front panel button that forces a trigger DC coupled trigger AC coupled trigger cutoff frequency lt 10 Hz internal lt 50 Hz external High frequency reject cutoff frequency 50 kHz Low frequency reject cutoff frequency 50 kHz Adds hysteresis to the trigger circuitry Selectable OFF or ON decreases sensitivity 2x 40 ns to 10 00 s lt 10 mV div greater of 1 div or 5 mV 10 mV div 0 6 div DC to 1 GHz l
5. Windows like operation Oscilloscope Experience Redefined Experience the Usability InfintiScan Zone touch trigger One of the biggest challenges of using an oscilloscope is setting up an advanced trigger to isolate a signal of interest While advanced triggers are powerful features InfiniiScan Zone touch trigger provides a turnkey trigger solution You simply observe the signal of interest on the display and draw a zone box around it What used to be hours of work can now take just a few seconds If you want to move your zones to another location just drag them over The 4000 X Series can be set up to easily trigger on one or two zone boxes simultaneously with either must intersect or must not intersect conditions InfiniiScan Zone triggering does not compromise the waveform update rate the 4000 X Series will still maintain an ultra fast 200 000 waveforms per second or more even with additional features turned on In other words the oscilloscope that sees everything can easily trigger on anything Figure 15 When you see anomalies all you have to do is draw a zone box to trigger on them Figure 16 Isolating your signal of interest has never been this easy Oscilloscope Experience Redefined Experience the Integration Investment protection through a fully upgradable 5 in 1 instrument The InfiniiVision 4000 X Series redefines the parao forrreo fe l Serial protocol rE 2 a ee eee oscilloscope experie
6. e ee IntinuVision 4000 X Series e e Oscilloscopes c amp amp ee Ad hd Re Data sheet Test amp Measurement World ul 2013 AWARD WINNER Oscilloscope Experience Redefined Experience the speed usability and integration Agp Agilent Technologies Anticipate ___ Accelerate ___ Achieve Oscilloscope Experience Redefined Experience the speed usability and integration Imagine an oscilloscope that sees everything triggers on anything has the ease of use of a tablet device and grows with your projects The 4000 X Series oscilloscopes are engineered for next generation performance delivering waveform update rates 20 times faster than the competition to display the most signal detail An industry leading 12 1 inch capacitive touch screen with innovative hardware based InfiniiScan Zone touch triggering provides the most intuitive interface to get you answers faster The 4000 X Series provides maximum investment protection with fully upgradable 5 instruments in 1 Experience the Speed Anomalies and elusive events are the toughest to debug The 4000 X Series oscilloscope redefines your debugging experience with MegaZoom IV smart memory technology The industry leading 1 million waveforms per second update rate means you see more of your signal behavior and can feel more confident in your design Experience the Usability You may be surprised just how easy it is to use the InfiniiVisio
7. Easily create and edit the waveform using the built in editor or Agilent s free BenchLink Waveform Builder Basic software With dual channels you can generate differential signals to output arbitrary clock and data signals to simulate serial buses create complex modulations more than the standard modulation feature output IQ signals and more The two channels can be tracked together as well identical frequency amplitude offset and duty cycle 3 digit voltmeter The 4000 X Series offers an integrated 3 digit voltmeter DVM and 5 digit frequency counter inside the oscilloscope DSOXDVM The voltmeter operates through the same probes as the oscilloscope channels However the DVM measurements are de coupled from the oscilloscope triggering system so that both the DVM and triggered oscilloscope waveform capture can be made with the same connection The voltmeter results are always displayed keeping these quick characterization measurements at your fingertips Frequency 9 1260kHz Figure 24 DVM 3 digit voltage and 5 digit frequency measurements 12 always at your fingertips tor 1 Mi dulation M ore Sine High Z Mod AM Sine AMFreq AM Depth YVave m Wweneral Menu R Modulation Type Waveform ZAN FTO AM Sine 11 00kHz 100 Figure 21 WaveGen sine wave output with and without added AM modulation NE ap A Aa A ae A ak Aa AN n a r N e S WY by K Jan l PaA A Aa A A iad d a JA
8. Maximum bandwidth Phase error at 1 MHz lt 0 5 degree Horizontal system digital channels Minimum detectable pulse width Channel to channel skew 2 ns 2 ns typical 3 ns maximum Acquisition system Maximum analog channels sample rate Analog channels equivalent sample rate Maximum analog channels record length Maximum digital channels sample rate Maximum digital channels record length Modes Normal Peak detect Averaging High resolution Segmented Roll Equivalent time 4022A 4024A 4032A 4034A 4052A 4054A 4104A 5 GSa s half channel interleaved 2 5 GSa s all channels N A 128 Gsa s 4 Mpts half channel interleaved 2 Mpts all channel 1 25 GSa s 2 Mpts with digital channels only 4154A Default mode Capture glitches as narrow as 200 ps at all time base settings Selectable from 2 4 8 16 64 to 65 536 Real time boxcar averaging reduces random noise and effectively increases vertical resolution 12 bits 50 ps div 11 bits gt 20 ps div 10 bits gt 10 us div 9 bits 5 us div Segmented memory optimizes available memory for data streams that have long dead times between activity Maximum segments 1000 Re arm time 1 us minimum time between trigger events Displays the waveform moving across the screen from right to left Available at the time base 50 ms div or slower 1 GHz and 1 5 GHz models only 7 8 ps fine interpolator resolution yields a maximum effective sample rate of 128 GSa s
9. additional math functions and comes standard on the 4000 X Series Additionally math functions can be nested to provide additional insight into your designs You can create up to four math functions with one resultant math function displayed at a time Operators e Add subtract multiply divide Transforms e Differentiate integrate FFT e Ax B e Squared square root e Absolute value e Common logarithm natural logarithm e Exponential base 10 exponential Filters e Low pass filter high pass filter e Averaged value Visualizations e Magnify e Measurement trend e Chart logic bus timing chart logic bus state 35 automatic measurements Automatic measurements are the essential tool of an oscilloscope In order to make quick and efficient measurements the 4000 X Series provides 35 powerful automatic measurements and can display up to 10 at a time Measurements can be gated by auto select main window zoom window or cursors 500 Ons umge 40 00ns Transforms ad dt Differentiate dt ntegrate u FFT AB Ax B Squared Square Root Absolute Value yg Common Logarithm in Natural Logarithm WATER e exponential 17 8000ns Base 10 Exponential Filters Low Pass Filter High Pass Filter Averaged Value Visualizations Ny Magnify aS A Measurement Trend s Thresholds y Source 1 Display an Figure 30 A variety of advanced math functions are standard in the 4
10. additional resolution and insight into the signal without requiring a repetitive signal Using real time boxcar averaging high resolution mode reduces random noise and effectively increases vertical resolution up to 12 bits Some signal anomalies may be visible only in this extra resolution mode Advanced parametric triggering With today s more complex signals you often need to trigger on complex signal conditions to synchronize the oscilloscope s acquisition on specific events The 4000 X Series oscilloscope can trigger on the following conditions edge edge then edge pulse width time qualitied pattern or rise fall time Nth edge burst runt setup and hold video and various serial buses optional Figure 42 Wide array of advanced parametric trigger modes Freeze display Perhaps you need to share with others an infrequent event you found With the freeze display feature you can keep intensity information on the screen while the oscilloscope Is stopped or before saving a screen shot Default Erase Press to v unr save Figure 43 The freeze screen feature keeps the intensity grading information while stopping the waveform acquisition 19 Oscilloscope Experience Redefined Designed for touch Industry s first and largest 72 7 inch capacitive touch screen to redefine your oscilloscope experience The way an oscilloscope was meant to be driven with a designed for touch interface
11. amp lt o g o z o a 8 hits hits 95 54 Mode 267 93 ps 0 hits 99 51 Hits 4 505 khits 20 000 ps s 108 36 ps Peak 32 hits n 284 38 ps in 230 17 ps Mean 271 05 ps Max 338 53 ps Std Dev 12 928 ps in Width 159 fs i 69 66 Median 270 15 ps Figure 39 InfiniiView enables a variety of advanced signal analysis while providing extensive yet intuitive waveform documentation E Agilent Spectrum Visualizer my 3000 File Control Setup Hardware View Help Lo eed Run Stop Single Autoscale Vertical Reference Level Scale Resolution Bandwidth Horizontal Use Start Stop Channel Select Range J Channel 1 Channel 3 2 kHz AutoRange Center Frequency 100 000 MHz wor Span Figure 40 Waterfall view for ASV spectrogram measurements Secure erase The secure erase feature comes standard with all 4000 X Series models At the press of a button internal non volatile memory is clear of all setup reference waveforms and user preferences ensuring the highest level of security in compliance with National Industrial Security Program Operation Manual NISPOM Chapter 8 requirements Oscilloscope Experience Redefined Other Key Productivity Tools High resolution mode for viewing signal details To build more confidence in your designs sometimes you need to look into more signal detail than you can see with the standard 8 bit vertical resolution of the 4000 X Series High resolution mode offers
12. ii A i ea m m mod j is shi at l j p y wi ri Pasur Add Settings Clear Meas Statistics v y ka Measurement Figure 23 Dual channel WaveGen output of differential arbitrary signals Common mode is shown as a math function Oscilloscope Experience Redefined Other Key Productivity Tools Power measurements and analysis When you are working with switching power supplies and power devices the power measurements application DSOX4PWR provides a full suite of power measurements and analysis in the oscilloscope Included with the DSOX4PWR is a license for the U1881A PC based power analysis software package which provides additional offline measurements and report generation See www agilent com find DSOX4PWR for more information HDTV video triggering and analysis Whether you are debugging consumer electronics with HDTV or characterizing a design the HDTV measurement application DSOX4VID provides support for a variety of HDTV standards for triggering and analysis See www agilent com find DSOX4VID for more information USB 2 0 signal quality analysis With the USB 2 0 signal quality test option DSOX4USBSQ designers of systems with USB interfaces can now perform automated signal quality testing This option supports low speed full speed and hi speed applications hi speed tests require 1 5 GHz models The USB 2 0 signal quality test with HTML pass fail report generation includes eye diagram mask te
13. of 10 s e Minimum time setting 2 ns 500 MHz 1 GHz 1 5 GHz 4 ns 350 MHz 6 ns 200 MHz Trigger on setup hold violations Setup time can be set from 7 s to 10 s Hold time can be set from 0 s to 10 ns Minimum window setup time hold time must be 3 ns or greater Trigger on all lines or individual lines odd even or all fields from composite video or broadcast standards NTSC PAL SECAM PAM M Trigger on lines and fields of enhanced and HDTV standards 480p 60 567p 50 720p 50 720p 60 1080p 24 1080p 25 1080p 30 1080p 50 1080p 60 1080i 50 1080 60 Trigger and decode on ARINC429 data Trigger on word start stop label label bits label range error conditions parity word gap word or gap all all bits eye all 0 bits all 1 bits Trigger on CAN controller area network version 2 0A and 2 0B signals Trigger on the start of frame SOF bit remote frame ID RTR data frame ID RTR remote or data frame ID data frame ID and data error frame all errors acknowledge error and overload frame Trigger on frame ID or specific error condition along with cycle base and repetition cycle filtering Can also trigger on specific events such as BSS TSS FES and wake up Trigger at a start stop condition or user defined frame with address and or data values Also trigger on missing acknowledge address with no acq restart EEPROM read and 10 bit write Trigger on 2 s complement data of audio left channel or right
14. on a selected channel whose time duration is less than a value greater than a value or inside a time range e Minimum duration setting 2 ns 500 MHz 1 GHz 1 5 GHz 4 ns 350 MHz 6 ns 200 MHz e Maximum duration setting 10 s e Range minimum 10 ns Trigger when a specified pattern of high low and don t care levels on any combination of analog digital or trigger channels is entered exited Pattern must have stabilized for a minimum of 2 ns to qualify as a valid trigger condition e Minimum duration setting 2 ns 500 MHz 1 GHz 1 5 GHz 4 ns 350 MHz 6 ns 200 MHz e Maximum duration setting 10 s Trigger on any selected edges from available sources analog and digital channels only Bandwidth is 500 MHz Trigger on rise time or fall time edge speed violations lt or gt based on user selectable threshold Select from lt or gt and time settings range between e Minimum 1 ns 500 MHz 1 GHz 1 5 GHz model 2 ns 350 MHz model 3 ns 200 MHz model e Maximum 10 s Trigger on the Nth 1 to 65535 edge of a pulse burst Specify idle time 10 ns to 10 s for framing Trigger on a position runt pulse that fails to exceed a high level threshold Trigger on a negative runt pulse that fails to exceed a low level threshold Trigger on either polarity runt pulse based on two threshold settings Runt triggering can also be time qualified lt or gt with a minimum time setting of 2 6 ns and maximum time setting
15. serial protocols including 12C SPI USB 2 0 RS232 UART CAN LIN FlexRay MIL STD 1553 ARINC 429 and 12S see page 22 Acquisition High Resolution A dph o Pii OSS Fe FR Fo TR 4G HS RS BH 155MSa Turn on Tum off Thresholds Bus B D15 D8 D7 DO Figure 18 Digital channels are captured and displayed time correlated with analog channels in MSOs or upgraded DSOs D 85 o pe CEC CC 1 SSC CC aC or wea ee ovo Figure 19 Dual serial bus CAN and LIN decode and interleaved lister display Segment 1 300 Channels a mawi Ti i wli PENA P marani ay AE I D pA a en Can Oe ui j L Search for All Errors Figure 20 USB 2 0 trigger decode and lister display 11 Oscilloscope Experience Redefined Experience the Integration Dual channel WaveGen 20 MHz function arbitrary waveform generator Industry exclusive The 4000 X Series offers the industry s only dual channel integrated 20 MHz function arbitrary waveform generator DSOX4WAVEGEN2 The integrated generator provides stimulus output of sine square ramp pulse DC noise sine cardinal sinc exponential rise exponential fall cardiac Gaussian pulse and arbitrary waveforms AWG to your device under test Signal modulation capability is also available With AWG functionality you can store waveforms from analog channels or reference memory to the arbitrary memory and output from WaveGen
16. stored waveform data to find specific events of interest can be slow and cumbersome With automatic search and navigation capability you can easily set up specific search criteria and then quickly navigate to found and marked events Available search criteria include edges pulse width time qualified rise fall times time qualified runt pulses time and level qualified and serial bus frames packets and errors Oe ee eee ee ee ee EEE D ES i Figure 11 Using the error condition search the 4000 X Series quickly found 5 places with a missing acknowledgment in an 1 C serial bus The navigation feature moves between the errors and zooms automatically to show the error packet Current Mask Eye Test of Tests 22 4200M Status Failing 00 02 01 3 Channel 2 186 474k 0 8317 On Error Source Lock Source Test All Stop Figure 9 Mask testing evaluated gt 22 M waveforms in just 2 minutes Figure 10 The 4000 X Series was set up to capture data signals with various rise time edges Using the search and navigation capability the oscilloscope was able to find mark white triangles and quickly navigate to 16 occurrences of out of compliance rise time edges Oscilloscope Experience Redefined Experience the Usability Industry s largest 12 1 inch display From the start of product development we designed every aspect of this oscilloscope for a touch interface Large easily
17. your favorite PC Web browser the 4000 X Series supports remote oscilloscope control from your tablet devices and smart phones with enough resolution The tablet virtual front panel is identical to the 4000 X Series touch GUI so you can touch icons draw InfiniiScan Zone Touch trigger zones and drag slide panels as if you are sitting in front of the actual oscilloscope Documentation and e mail Annotation becomes a simple task Bring up the annotation menu and start editing it using the keypad and then drag it to the desired location 100 0ns 200 0ns Annotation Menu Annotation Edit Annotation Text Color A non monotonic Channel 1 t Figure 38 a Annotation and keypad HOST DEVICE Auto MDIX 7 a E OA J y i f SVGA VIDEO oS OUT LAN Figure 36 Standard VGA output USB host and device and LAN Figure 37 Tablet virtual front panel control Quick e mail allows you to e mail the data you want instantly to your inbox Send out the screenshot waveform data or even a USB signal quality test report This removes the hassle of connecting your PC to your oscilloscope Setup To contact_usi aqilent com From contact_usigagilent com Sener subject Infinivisian 4000 is great summary Format PNG 24 bit image png Invert Grat Off Palette Color setup Info Off Figure 38 b E mail configuration screen 17 Oscilloscope Experience Redefined Other Key Productivity To
18. 0 uV or 3 digits Accuracy waveform modes 1 5 of offset setting 1 of amplitude 1 mV Accuracy DC mode 1 5 of offset setting 3 mV Main output Impedance 50 0 typical Isolation Not available main output BNC is grounded Protection Overload automatically disables output Trigger output Trigger output available on trig out BNC Digital voltmeter Source Analog channels only 1 4 Functions ACrms DC DCrms frequency Resolution ACV DCV 3 digits Counter frequency 5 5 digits Measuring rate 100 times second Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements Range meter Graphical display of most recent measurement plus extrema over the previous 3 seconds General and environmental characteristics Power line consumption Temperature Operating Nonoperating Humidity Operating Nonoperating Altitude Operating Nonoperating Electromagnetic compatibility Maximum 120 watts 0 to 55 C 30 to 70 C 50 to 95 RH at 40 C for 5 days 90 RH at 65 C for 24 hr up to 3 000 m 15 000 ft up to 15 300 m 50 000 ft Meets EMC Directive 2004 108 EC meets or exceeds IEC 61326 1 2005 EN 61326 1 2006 Group 1 Class A requirement CISPR 11 EN 55011 IEC 61000 4 2 EN 61000 4 2 IEC 61000 4 3 EN 61000 4 3 IEC 61000 4 4 EN 61000 4 4 IEC 61000 4 5 EN 61000 4 5 IEC 61000 4 6 EN 61000 4 6 IEC 61000 4 11 EN 61000 4 11 Canada ICES 001 2004 Australia New Z
19. 000 X Series Display Math Function Math 1 Math 3 Math 4 Scal e 50 0 p sf Bandwidth 4 OMHz Source 1 Math 1 Operator Low Pass Filter Display Math Math 2 Figure 31 Four math functions can be created and nested with one resultant math function Freq 1 xt No ed ges Pk Pk 1 Rise 1 54 6ns Counter 1 555 05kHz Width 1 x 400 2ns Auto Select Main lt s Gated by Cursors Thresholds Measurement Menu WOES Snider CaN uy Cursul Source Type Add Settings Clear Meas Statistics T Measurement w eB ee Figure 32 Up to 10 automated measurements displayed simultaneously Measurements can be gated by cursors Meas Window ata 15 Oscilloscope Experience Redefined Other Key Productivity Tools Reference waveforms Store up to four waveforms in the scope s non volatile reference waveform memory Compare reference waveforms with live waveforms and perform post analysis and measurements on stored data You can also store waveforms on a removable USB memory device in h5 format and recall them back into oscilloscope s reference waveform memory later Save and or transfer waveforms to a PC as XY data pairs in a comma separated values format csv or store bitmap images and transfer them to a PC for documentation purposes in a variety of image formats Powerful probe solutions and compatibility Get the most out of your 4000 X Series scope by using Agil
20. C vertical gain accuracy 0 42 full scale 35 measurements maximum of 10 displayed at a time Measurements continuously updated with statistics Cursors track last selected measurement Select up to four measurements from the list below Voltage Peak to peak maximum minimum amplitude top base overshoot pre shoot average N cycles average full screen DC RMS N cycles DC RMS full screen AC RMS N cycles AC RMS full screen standard deviation ratio RMS1 RMS82 Time Period frequency counter width width burst width duty cycle rise time fall time delay phase X at min Y X at max Y Count Positive pulse count negative pulse count rising edge count falling edge count Mixed Area N cycles area full screen Built in frequency counter Source Any analog or digital channel Resolution 5 digits Up to 8 digits with an external reference 10 MHz input Max frequency Bandwidth of oscilloscope Standard mask limit test capability that provides easy fast pass fail comparison of a signal under test to a predefined mask template or auto mask template Predefined mask templates or edits to an auto mask template can be made via a simple text editor gt 270 000 mask tests per second waveform update rate Number of math functions Arithmetic FFT Four display one at a time Add subtract multiply FFT differentiate integrate square root divide Ax B square absolute common log natural log exponent
21. GA for each debug pin on the outside of the FPGA in seconds all while automatically mapping internal signal names to the oscilloscope channel labels The DSOX4FPGAX supports the Virtex 6 series Virtex 5 series Virtex 4 series Virtex Il Pro series Virtex Il series and Spartan 3 series devices Educator s oscillocope training kit Teach your students what an oscilloscope is and how to perform basic measurements with the Educator s Oscilloscope Training Kit DSOXEDK This kit includes training tools created specifically for electrical engineering and physics undergraduate students and professors It contains an array of built in training signals a comprehensive oscilloscope lab guide and tutorial written specifically for undergraduate students and an oscilloscope fundamentals PowerPoint slide set for professors and lab assistants Also available is an advanced triggering guide to help even the most experienced oscilloscope users to get the most out of their 4000 X Series oscilloscope See www agilent com find EDK for more information Figure 28 FPGA dynamic probe routing out internal FPGA signals to be captured by the 4000 X Series correlated with this external analog sine wave in the design Figure 29 DSOXEDK gets both students and experienced users efficiently using the 4000 X Series Oscilloscope Experience Redefined Other Key Productivity Tools Advanced math analysis Advanced math analysis provides a variety of
22. Hz 4 ch Service center partner amp Agilent service CCC iy center DSOX4B2T52U 200 MHz 500 MHz 2 ch Service center DSOX4B2T54U 200 MHz 500 MHz 4 ch Service center Customer Upgraded scope DSOX4B2T104U 200 MHz 1 GHz 4 ch Service center shipped back _ SOS KERT a DSOX4B2T154U 200 MHz 1 5 GHz 4 ch Service center DSOX4B3T104U 350 MHz 1 GHz 4 ch Service center aaa DSOX4B3T154U 350 MHz 1 5 GHz 4 ch Service center DSOX4B5T154U 500 MHz 1 5 GHz 4 ch Service center 34 Process description Place order for a return to Agilent service center bandwidth upgrade product to an Agilent sales partner Service center installation cost is in addition to bandwidth upgrade product price Service center calibration is included in the bandwidth upgrade product price 2 Agilent Business Center will contact you regarding process and timing of the service center installation Continue to use oscilloscope until contacted again later when parts are available at service center Ship the oscilloscope per provided instructions freight paid by Agilent to service center 4 Service center ships back upgraded oscilloscope with stick on labels applied to front and rear panels indicating upgraded bandwidth specification Model number and serial number of the oscilloscope do not change a Agilent Technologies Oscilloscopes Multiple form factors from 20 MHz to gt 90 GHz Industry leading specs Powerful applic
23. S232 UART DSOX4B2T32U 200 MHz 350 MHz 2ch Service Center DSOX4USBFL USB 2 0 Full Low Speed serial decode and triggering DSOX4B2T34U 200 MHz 350 MHz 4ch Service Center DSOX4USBH USB 2 0 Hi Speed serial decode and triggering DSOX4B3T52U 350 MHz 500 MHz 2ch License Only DSOX4USBSQ USB 2 0 signal quality test DSOX4B3T54U 350 MHz 500 MHz 4ch License Only DSOX4AUTO Automotive serial triggering and analysis CAN LIN DSOX4B5T104U 500 MHz 1 GHz 4 ch Service Center DSOX4FLEX FlexRay serial triggering and analysis DS0X4B1T154U 1 GHz 1 5 GHz 4 ch Service Center DSOX4AUDIO Audio serial triggering and analysis I S DSOX4B2T52U 200 MHz 500MHz 2ch Service Center DSOX4AERO Aerospace and defence serial triggering and decode MIL STD DSOX4B2T54U 200 MHz 500 MHz 4ch Service Center 1553 and ARINC 429 DSOX4B2T104U 200 MHz 1 GHz 4 ch Service Center DSOX4WAVEGEN2 Dual channel WaveGen 20 MHz arbitrary function generator DSOX4B2T154U 200 MHz 1 5 GHz 4 ch Service Center DSOXDVM 3 digit voltmeter DVM DSOX4B3T104U 350 MHz 1 GHz 4 ch Service Center DSOX4PWR Power analysis application DSOX4B3T154U 350 MHz 1 5 GHz 4 ch Service Center DSOX4MASK Mask Limit Testing DSOX4B5T154U 500 MHz 1 5 GHz 4 ch Service Center DSOX4VID Enhanced video TV application package DSOX4FPGAX FPGA dynamic probe option for Xilinx DSOXEDK Education and training kit DSOX4USBH is available only for 1 GHz and 1 5 GHz models Process description 1 Place order for a lic
24. ask limit testing Enhanced video TV application package FPGA dynamic probe option for Xilinx USB 2 0 signal quality test option Productivity tools Education training kit InfiniiView oscilloscope analysis software Agilent spectrum visualizer ASV Vector signal analyzer software version 16 and higher BenchLink waveform builder pro and basic DSOX4USBH is only available for 1 GHz and 1 5 GHz models USB 2 0 hi speed tests require 1 5 GHz models See page 33 for more detailed upgradability and installation process information Model Number DSOX4EMBD EMB DSOX4COMP CMP DSOX4USBFL USF DSOX4USBH U2H DSOX4AUTO AMS DSOX4FLEX FLX DSOX4AUDIO SND DSOX4AERO AER DSOX4WAVEGEN2 WAV DSOXDVM DVM DSOX4PWR PWR DSOX4MASK MSK DSOX4VID VID DSOX4FPGAX FPX DSOX4USBSQ U2Q DSOXEDK EDK N8900A 64997A 89601B 33503A Oscilloscope Experience Redefined Configuring your InfiniiVision 4000 X Series Oscilloscope Step 3 Choose your probes For a complete list of compatible probes visit www agilent com find scope_ probes Probes 4000 X Series N2894A passive probe 700 MHz 10 1 10 MQ Included standard 1 per channel N6450 60001 16 digital channel MSO cable Included on MSOX models and DSOXPERFMSO 10076B high voltage passive probe 250 MHz 4 kV Optional N2795A active single ended probe 1 GHz 1 pF 1 MQ with AutoProbe Optional N2796A active single ended probe 2 GHz 1 pF 1 MQ with AutoProbe Optional
25. ations 35 myAgilent www agilent com find myagilent A personalized view into the information most relevant to you P AKle www axiestandard org AdvancedTCA Extensions for Instrumentation and Test AXle is an open standard that extends the AdvancedTCA for general purpose and semiconductor test Agilent is a founding member of the AXle consortium LX www l xistandard org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems Agilent is a founding member of the LXI consortium mn Hi PAI www pxisa org PCI eXtensions for Instrumentation PXI modular instrumentation delivers a rugged PC based high performance measurement and automation system Agilent Channel Partners www agilent com find channelpartners Get the best of both worlds Agilent s measurement expertise and product breadth combined with channel partner convenience 3 to tea Warranty WARRANTY www agilent com find ThreeYearWarranty Agilent s combination of product reliability and three year warranty coverage is another way we help you achieve your business goals increased confidence in uptime reduced cost of ownership and greater convenience ad Agilent Advantage Services www agilent com find AdvantageServices Accurate measurements throughout the life of your instruments DEKRA Ceftitied 1 0 9001 2008 uality Management System www agilent com quality www agilent com www a
26. channel 4 lt gt gt lt lt gt increasing value or decreasing value Trigger on LIN local interconnect network sync break sync frame ID or frame ID and data 27 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Trigger type selections MIL STD1553 Option SPI Option UART RS232 422 485 Option USB Option Waveform measurements Trigger on MIL STD 1553 signals on data word start stop command status start stop RTA RTA 11 bits and error conditions parity sync Manchester Trigger on SPI serial peripheral interface data pattern during a specific framing period Supports positive and negative chip select framing as well as clock Idle framing and user specified number of bits per frame Supports MOSI and MISO data Trigger on Rx or Ix start bit stop bit data content or parity error KKK Trigger on start of packet SOP end of packet EOP suspend resume reset packets token data handshake or special and errors PID CRC5 CRC16 glitch bit stuff SE1 Supports USB 2 0 low speed full speed and hi speed hi speed is supported on 1 GHz and 1 5 GHz models only DC vertical accuracy cursors Number of measurements Automatic measurements Counter Mask limit test option Waveform math Single cursor accuracy DC vertical gain accuracy DC vertical offset accuracy 0 21 full scale Dual cursor accuracy D
27. e Waveform update rate 5 GSa s half channel 2 5 GSa s all channel Standard 4 Mpts standard segmented memory 12 1 inch high definition capacitive touch display gt 1 million waveforms per second System bandwidth with N2894A 200 MHz 200 MHz 350 MHz 350MHz 500MHz 500MHz 700 MHz 700 MHz standard passive probe Vertical system analog channels Hardware bandwidth limits Approximately 20 MHz selectable Input coupling AC DC Selectable 1 MQ 1 16 pF 50 Q 1 5 200 MHz 500 MHz models 1 mV div to 5 V div 1 MQ and 50 ohm 1 GHz and 1 5 GHz models 1 mV div to 5 V div 1 MQ 1 mV div to 1 V div 50 ohm Vertical resolution 8 bits CAT 300 Vrms 400 Vpk transient overvoltage 1 6 kVpk CAT II 300 Vrms 400 Vpk 2 0 full scale 0 1div 2 mV 1 of offset setting 200 MHz 1 GHz gt 40 dB from DC to maximum specified bandwidth of each model 1 5 GHz gt 40 dB from DC to 1 GHz 35 dB from 1 GHz to 1 5 GHz 5 V lt 10 mV div 20 V 10 to 200 mV div 75 V gt 200 mV div Input impedance Input sensitivity range Maximum input voltage DC vertical gain accuracy DC vertical offset accuracy Channel to channel isolation Offset range Vertical system digital channels Digital input channels 16 digital DO to D15 Pod 1 D7 DO Pod 2 D15 D8 Thresholds Threshold per pod TTL 1 4 V 5 V CMOS 2 5 V ECL 1 3 V user defined selectable by pod 8 0 V in 10 mV steps 40 V peak CAT
28. e fall Frequency range 0 1 Hz to 5 0 MHz Cardiac Frequency range 0 1 Hz to 200 0 kHz Gaussian pulse Frequency range 0 1 Hz to 5 0 MHz Arbitrary Waveform length 1 to 8 192 points Amplitude resolution 10 bits including sign bit Repetition rate 0 1 Hz to 12 MHz Sample rate 100 MSa s Filter bandwidth 20 MHz Frequency Sine wave and ramp accuracy 130 ppm frequency lt 10 kHz 50 ppm frequency gt 10 kHz Square wave and pulse accuracy 50 frequency 200 ppm frequency lt 25 kHz 50 ppm frequency 25 kHz Resolution 0 1 Hz or 4 digits whichever is larger Amplitude Range minimum 20 mVpp if offset lt 0 5 Vpp into Hi Z 10 mVpp if offset lt 0 5 Vpp into 500 Range maximum 10 Vpp except 9 Vpp if Sinc or Cardiac 7 5 Vpp if Gaussian pulse into Hi Z 5 Vpp 4 5 Vpp into 50 0 Resolution 100 uV or 3 digits whichever is higher Accuracy 1 5 frequency 1 kHz Sinc cardiac and Gaussian pulse 1 25 V into Hi Z 625 mV into 50 Q Full resolution is not available at output due to internal attenuator stepping 30 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Dual channel WaveGen built in function arbitary waveform generator DC offset Range 5 V into Hi Z except 4 V if sine wave 2 5 V if sinc cardiac or Gaussian pulse into Hi Z 2 5 V into Hi Z except 2 V if sine wave 1 25 V if sinc cardiac or Gaussian pulse into 50 Q Resolution Larger of 25
29. ealand AS NZS Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics General and environmental characteristics Safety Vibration Shock Dimensions Weight Kensington style lock UL61010 1 2nd edition CAN CSA22 2 No 61010 1 04 Meets IEC60068 2 6 and MIL PRF 28800 class 3 random Meets IEC 60068 2 27 and MIL PRF 28800 class 3 random operating 30 g sine 11 ms duration 3 shocks axis along major axis total of 18 shocks 454 mm W x 275 mm H x 156 mm D Net 6 3 kg 13 9 Ibs shipping 11 4 kg 25 Ibs Rear panel security slot connects to standard Kensington style lock Nonvolatile storage Reference waveform display Save formats Max USB flash drive size Set ups without USB flash drive Set ups with USB flash drive 2 internal waveforms or USB thumb drive Setup scp 8 or 24 bit bitmap image bmp PNG 24 bit image png CSV data csv ASCII XY data csv binary data bin lister data csv reference waveform data h5 mask mask arbitrary waveform data csv power harmonics data csv Supports industry standard flash drives 10 internal setups Limited by size of USB drive Included standard with oscilloscope Factory warranty Calibration Probes Localized interface Power cord Front panel protection Documentation 3 year warranty Certificate of Calibration 2 year calibration interval One per channel N2894A 700 MHz passive pr
30. ense only upgrade or measurement application product with an Agilent sales partner 2 For measurement applications you will receive a paper or electronic pdf entitlement certificate For bandwidth upgrades only you will receive a stick on label document indicating upgraded bandwidth specification in addition to a paper entitlement certificate 3 Use entitlement certificate containing instructions and certificate number needed to generate a license file for a particular 4000 X Series oscilloscope model number and serial number unit 4 Receive the licensed file and installation instructions via email Copy license file lic extension from email to a USB drive and follow instructions in email to install the purchased bandwidth upgrade Or measurement application on the oscilloscope 6 For bandwidth upgrades only attach bandwidth upgraded stick on labels to front and rear panels of the oscilloscope Model number and serial number of the oscilloscope do not change 33 Oscilloscope Experience Redefined Return to Agilent Service Center Bandwidth Upgrades a a 4000 X Series bandwidth upgrade models groans Model Number Description Type j DS0X4B2T32U 200 MHz 350 MHz 2 ch Service center DSOX4B2T34U 200 MHz 350 MHz 4 ch Service center DSOX4B3T52U 350 MHz 500 MHz 2 ch License only amp DSOX4B3T54U 350 MHz 500 MHz 4 ch License only A DSOX4B5T104U 500 MHz 1 GHz 4 ch Service center Agilent sales DSOX4B1T154U 1 GHz 1 5 G
31. ent s complete family of innovative probes and accessories for your application The 4000 X Series supports up to four active probes simultaneously with its full AutoProbe interface All 4000 X Series scopes come standard with a 700 MHz bandwidth 10 MQ input passive probe per each channel and gives you 00 MHz system bandwidth when used in conjunction with the 4000 X Series 1 GHz 1 5 GHz models Also available is the N2750A InfiniiMode differential probe and N2 795A 96A single ended active probe for high signal fidelity measurements without the high price For ultra low current measurements the N2820A Series high sensitivity current probes are the best solution in the industry For the most up to date and complete information about Agilent s probes and accessories visit our Web site at www agilent com find scope_probes or refer to the InfiniiVision Probes and Accessories data sheet with the Agilent literature number 5968 8153EN Some restriction may apply Contact Agilent for more details Localized front panel GUI and help Operate the oscilloscope in the language most familiar to you The graphical user interface built in help system front panel overlays and user s manual are available in 11 languages During operation access the built in help system just by pressing and holding any button Source Save to Clear Skew Options Figure 33 Store and recall up to four reference waveforms Figure 34 The N2750A Infini
32. finiiVision 4000 X Series gives you the highest probability of capturing random and infrequent events that you would miss on an oscilloscope with a lower waveform update rate Powered by MegaZoom IV smart memory technology the InfiniiVision 4000 X Series not only lets you see more waveforms but it has the uncompromised ability to find the most difficult problems in your design Unlike other oscilloscopes uncompromised ability means e Always fast responsive operation No slowdown with logic channels ou Figure 3 The 4000 X Series captures a glitch occurring once in e No slowdown with protocol decoding on a million waveform cycles e No slowdown with math functions turned on e No slowdown with measurements turned on What is waveform update rate and why is it important As oscilloscopes acquire data process It and plot it to the screen there Is inevitable dead time or the time oscilloscopes miss signals completely In general the faster the waveform update rate the shorter the dead time The shorter the dead time the more likely an oscilloscope is to capture anomalies and infrequent events This is why it is critical to select an oscilloscope with a fast waveform update rate Q Figure 4 Other vendor s oscilloscope with 50 000 waveforms second A long dead time decreases your chances of capturing infrequent events EEAAIEI gt gt gt 4 4 te Figure 5 InfiniiVision 4000 X Series with 1 000 000 wave
33. forms second A short dead time increases your chances of capturing infrequent events Oscilloscope Experience Redefined Experience the Speed Agilent achieves this industry leading waveform update rate with MegaZoom IV smart memory technology Traditionally CPU processing was the major bottleneck for The InfiniiVision 4000 X Series requires minimum support oscilloscope waveform update rate and responsiveness from a CPU as most core operations are handled by Agilent Typically the CPU handles interpolations logic channel proprietary technology the MegaZoom IV smart memory plotting serial bus decoding measurements and more ASIC MegaZoom includes hardware serial decoders and and the waveform update rate drops dramatically as these hardware mask limit testing capability plots analog and features are turned on digital data directly to the display supports GUI operation and integrates additional instruments like the dual channel WaveGen function arbitrary waveform generator MegaZoom IV ASIC Hardware Anal Accelerated n aa Measurements Signals Search noo Acquisition Memory Manager Segmented Memory Math 1 0 GUI gt gt DRAM Signals i Serial Decoders O Mask Trigger WaveGen infin amp Synthesis Figure 6 The 4000 X Series oscilloscopes uncompromised responsiveness speed and waveform update rate is enabled by the MegaZoom IV smart memory ASIC The CPU is not used for c
34. gilent com tind 4000X Series For more information on Agilent Technologies products applications or services please contact your local Agilent office The complete list is available at www agilent com tind contactus Americas a Canada 877 894 4414 Brazil 11 4197 3600 Mexico 01800 5064 800 United States 800 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 421 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries 65 375 8100 Europe amp Middle East Belgium 32 0 2 404 93 40 Denmark 45 45 80 12 15 Finland 358 0 10 855 2100 France 0825 010 700 0 125 minute Germany 49 0 7031 464 6333 lreland 1890 924 204 Israel 972 3 9288 504 544 Italy 39 02 92 60 8484 Netherlands 31 0 20 547 2111 Spain 34 91 631 3300 Sweden 0200 88 22 55 United Kingdom 44 0 118 927 6201 For other unlisted countries www agilent com tind contactus BP2 19 13 Product specifications and descriptions in this document subject to change without notice Agilent Technologies Inc 2012 2013 Published in USA March 24 2013 5991 1103EN Eee Agilent Technologies
35. he speed usability and integration Oscilloscope Experience Redefined Experience the speed usability and integration Overview of Agilent InfiniiVision X Series oscilloscopes InfiniiVision InfiniiVision InfiniiVision 4000 X Series 3000 X Series 2000 X Series Analog channels 2 and 4 2 and 4 2 and 4 Bandwidth upgradable 200 350 500 MHz 1 GHz 1 5 GHz 100 200 350 500 MHz 1 GHz 70 100 200 MHz Digital channels 16 MSO models or upgrade 16 MSO models or upgrade 8 MSO models or upgrade Maximum sample rate 5 GSa s 5 GSa s 1 GHz models 2 GSa s 4 GSa s 100 500 MHz models Maximum memory depth 4 Mpts standard 2 Mpts standard 100 kpts standard 4 Mpts option 1 Mpts option Waveform update rate gt 1 000 000 waveforms per second gt 1 000 000 waveforms per gt 50 000 waveforms per second second Display 12 1 inch capacitive touch display 8 5 inch display 8 5 inch display InfiniiScan Zone touch trigger Standard No No WaveGen 20 MHz function Dual channel AWG option Single channel AWG option Single channel function arbitrary waveform generator option Integrated digital voltmeter Yes option Yes option Yes option Search and navigate Yes Yes No Serial protocol analysis Yes optional ARINC 429 CAN Yes optional ARINC 429 CAN Yes optional CAN 1 C LIN FlexRay I2C I4S LIN MIL STD 1553 FlexRay I C 12S LIN MIL SPI UART RS232 SPI UART RS232 USB 2 0 STD 1553 SPI VART RS232 Segmented memory Standa
36. iMode probe allows convenient measurement of differential single ended and common mode signals with a single probe tip without reconnecting the probe to change the connection Language Enalist BIER Chinese Simplified Feta PX Chinese Traditional Francais French Deutsch German Italiano Italian BAS Japanese StH Korean Portugu s Portuguese Pycckwi Russian Espanol Spanish Figure 35 Operate the oscilloscope in your choice of 11 languages Oscilloscope Experience Redefined Other Key Productivity Tools Connectivity and LXI compatibility Standard USB 2 0 hi speed host two on front one on back and device one on back ports make PC connectivity easy Operate the scope from your PC and save recall stored waveforms and setup files via standard LAN LXI IPv6 Extended Function Connect your projector or external monitor through VGA output standard with the 4000 X Series when sharing and presenting screen information An optional external GPIB to LAN adapter is also available N4865A IntuiLink toolbars and data capture let you quickly move screen shots and data into Microsoft Word and Excel software You can download and install these toolbars from www agilent com find intuilink Virtual front panel The 4000 X Series innovative capacitive touch screen matches perfectly with the latest tablet technologies In addition to the traditional virtual front panel remote operation through
37. ial base 10 exponential LP filter HP filter averaged value magnify measurement trend chart logic bus timing or state Up to 64 kpts resolution FFT window types hanning flat top rectangular Blackman Harris Denotes warranted specifications all others are typical Specifications are valid after a 30 minute warm up period and 10 C from firmware calibration temperature 1 mV div and 2 mV div is a magnification of 4 mV div setting For vertical accuracy calculations use full scale of 32 mV for 1 mV div and 2 mV div sensitivity setting Suspend resume reset bit stuff error and SE1 error are USB 2 0 low and full speed only Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Display characteristics Display 12 1 inch high definition capacitive touch display Resolution 800 H x 600 V pixel format screen area Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls Format YT XY and Roll Maximum waveform update rate gt 1 000 000 wfm s Persistence Off infinite variable persistence 100 ms 60 s Intensity gradation 64 intensity levels Connectivity USB 2 0 hi speed host port USB 2 0 hi speed host ports x3 two front and one real panel Supports memory devices and printers USB 2 0 hi speed device port One USB 2 0 hi speed device port on rear panel USB Test amp Measurement Class USBTMC compatible LAN port 10 100Base T port on rear pane
38. ion 1 Hz to Carrier Frequency or Ze Carrier Frequency whichever is smaller FSK Modulation 50 duty cycle square wave FSK Rate 1Hz to 20 kHz Hop Frequency 2 x FSK Rate to 10 MHz Only the following combination of wave shapes can be frequency tracked or completely tracked 1 sine ramp sine cardinal cardiac and Gaussian pulse 2 square wave and pulse 3 exponential rise and exponetial fall 4 arbitrary 29 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Dual channel WaveGen built in function arbitary waveform generator Sine Frequency range 0 1 Hz to 10 MHz Amplitude flatness 0 5 dB relative to 1 kHz Harmonic distortion 40 dBc Spurious non harmonics 40 dBc Total harmonic distortion 1 SNR 50 ohm load 500 MHz BW 40 dB Vpp 0 1 V 30 dB Vpp lt 0 1V Square wave pulse Frequency range 0 1 Hz to 20 MHz Duty cycle 20 to 80 Duty cycle resolution Larger of 1 or 10 ns Rise fall time 19 ns 10 to 90 Overshoot lt 2 Asymmetry at 50 DC 1 dns Jitter TIE RMS 500 ps Ramp triangle wave Frequency range 0 1 Hz to 200 kHz Linearity 1 Variable symmetry 0 to 100 Symmetry resolution 1 Pulse Frequency range 0 1 Hz to 10 MHz Pulse width 20 ns minimum Pulse with resolution 10 ns Edge time Fixed at 19 ns not variable Overshoot lt 2 Noise Bandwidth 20 MHz typical Sine cardinal Sinc Frequency range 0 1 Hz to 1 0 MHz Exponential ris
39. l LXI IPv6 extended function Web remote control VNC Web interface via major Web browsers Video out port SVGA out on rear panel Connect oscilloscope display to an external monitor or projector GPIB port N4865A GPIB to LAN adapter optional 10 MHz out in BNC connector on the rear panel Supported modes Off 10 MHz out or reference signal mode 10 MHz in Trigger out BNC connector on the rear panel Supported modes triggers mask waveform generator 1 sync pulse and waveform generator 2 sync pulse Dual channel WaveGen built in function arbitary waveform generator WaveGen outputs Two front panel BNC connectors Both waveform generator outputs can be frequency tracked amplitude tracked or completely tracked A generator s output can be inverted to create a differential signal Waveforms Sine square ramp pulse DC noise sine cardinal sinc exponential rise exponential fall cardiac Gaussian pulse and arbitrary Modulation Modulation is available on channel 1 only Modulation is not available when tracking mode is enabled Modulation Types AM FM FSK Carrier Waveforms Sine Ramp Sine Cardinal Exponential Rise Exponential Fall and Cardiac Modulation Source Internal no external modulation capability AM Modulation Sine Square Ramp Modulation Frequency 1Hz to 20 kHz Depth 0 to 100 FM Modulation Sine Square Ramp Modulation Frequency 1Hz to 20 kHz Minimum Carrier Frequency 10Hz Deviat
40. n 4000 X Series A 12 1 inch capacitive touch screen the industry s largest works just like your favorite tablet or smart phone so debugging your devices is faster than ever before Innovative InfiniiScan Zone touch triggering makes triggering on anything a snap Just draw a box around signals of interest and the oscilloscope triggers on them So if you can see It you can trigger on it Experience the Integration The 4000 X Series further redefines your oscilloscope experience by integrating the capabilities of five instruments in one oscilloscope channels logic channels digital voltmeter DVM dual channel WaveGen function arbitrary waveform generator and serial protocol analyzer including USB All are upgradable including bandwidth for the ultimate investment protection 2 2013 FINALIST 4000 X Series Oscilloscope experience redefined Key features Experience the speed One million waveforms per second update rate MegaZoom IV smart memory technology Standard segmented memory e Experience the usability Industry s first capacitive touch screen Industry s largest 12 inch display InfiniiScan Zone touch trigger Experience the integration Industry s first 5 instruments in 1 Industry s first fully upgradable including bandwidth to 1 5 GHz Industry s leading application solutions 72989 os o Resm AOE Figure 1 MegaZoom IV smart memory technology enables t
41. nce with unprecedented analysis LL integration This 5 in 1 instrument provides aoe ee __ a CE eC CO e Oscilloscope im aa Oscilloscope e 16 digital channels analog e Serial protocol analyzer channels Diaital e Dual channel WaveGen 20 MHz ol aiar function arbitrary Logic channels waveform generator iaeiae e 3 digit voltmeter WaveGen Figure 17 The 4000 X Series provides the capabilities of five instruments seamlessly integrated into one Fully upgradable oscilloscope Add at the time of your purchase or upgrade later e Bandwidth up to best in class 1 5 GHz Digital channels MSO Dual channel WaveGen 20 MHz function arbitrary waveform generator 3 digit voltmeter Project needs change and now your oscilloscope can too With the 4000 X Series your investment is protected If you need more bandwidth up to the best in class 1 5 GHz digital channels dual channel WaveGen DVM or measurement applications in the future you can easily add them at any time Measurement applications See pages 33 and 34 for more detailed information on Serial protocol analysis available upgrades o Power measurement analysis o HDTV video triggering and analysis o Mask testing o Educators training kit Oscilloscope Experience Redefined Experience the Integration Mixed signal oscilloscope MSO Integrated 16 digital channels With an additional 16 integrated digital channels you now have up to 20 channels of time correla
42. new dimension dead time for maximum probability of usability See setup summary of capturing infrequent events and automatic measurements cursor info anomalies DVM and navigation pane in any combination anywhere on the screen meca Z om 1M wims s Em Horizontal e Run Control r 0 ae Single Standard advanced math and four ao cascade able math functions enable uto i gi i lt gt Setup Sethe even the most sophisticated signal Navioate lt m gt vine analysis Freq 1 m Trigger mume Measure 188 OkHz Rise 1 Trigger EA a ror OIS buns Fone bewel Cursors Display up to 70 measurements Fal ets __ O PA UO 7 simultaneously without compromising a0ns Waveform mum File 180ns mm Waveform ps other key info 35 automatic Pulse Count 1 a Anatyze Acquire Display ES Recai Print Print F measurements can be gated by cursors 41 y Tools Quick Um Action Pk Pk 1 Freq 2 Not a touch screen fan Turn off the touch screen from a front panel button if desired Rise i Fa if Pulse Count Pk Pk 2 Independent knobs per channel for fast operation All front panel knobs are push able for access to common controls Symmetry IMN 16pF y Standard segmented memory powered by MegaZoom V smart memory technology provides intelligent capture of just the signal of interest Four AutoP
43. obe 10 1 attenuation N6450 60001 16 digital channel MSO cable 1 per oscilloscope included on all MSO models and DSOXPERFMSO English Chinese simplfied and traditional French German Italian Japanese Korean Portuguese Russian and Spanish localized front panel overlays interface and built in help system Localized power cord Front panel cover CD containing localized user s guide service guide and programmer s manual Related literature Publication title Publication type Publication number Triggering on Infrequent Anomalies and Complex Signals Using InfiniiScan Zone Trigger Application note 5991 1107EN Characterizing Hi Speed USB 2 0 Designs in Embedded Designs Application note 5991 1148EN Oscilloscope Memory Architectures Why All Acquisition Memory is not Created Equal Application note 5991 1024EN Switch Mode Power Supply Measurements Application note 5991 1117EN 32 Oscilloscope Experience Redefined License only Bandwidth Upgrades and Measurement Applications Entitlement certificate Stick On label sheet Agilent sales for bandwidth 2 l partner upgrades Agilent software i licensing website Customer Email containing USB drive containing _ attached license file license file ae 4000 X Series bandwidth upgrade models Model Number Description Model Description Type DSOX4EMBD Embedded serial triggering and analysis I C SPI Number DSOX4COMP Computer serial triggering and analysis R
44. ols InfiniiView oscilloscope analysis software Agilent s InfiniiView PC based oscilloscope analysis software N8900A allows you to do additional signal viewing analysis and documentation tasks away from your oscilloscope Capture waveforms save to a file and recall the waveforms into InfiniiView The application supports a variety of popular waveform formats from multiple oscilloscope vendors and includes the following features navigate view measurements analyze view windows documentation and optional analysis upgrades For more information go to www agilent com find InfiniiView Agilent spectrum visualizer ASV software ASV PC based software package 64997A connects to the oscilloscope via USB or Ethernet connection ASV provides affordable advanced FFT frequency domain analysis and spectrum and spectrogram analysis with an intuitive user interface familiar to RF engineers For more information go to www agilent com find ASV_InfintiVision 3 Year standard warranty 2 Year calibration interval Through improved quality processes and rigorous testing the Agilent InfiniiVision 4000 X Series oscilloscopes are able to perform at guaranteed specification for two years without calibration thereby reducing your cost of ownership These oscilloscopes come with a standard factory 3 year warranty 18 Lo File Control Setup Display Trigger Measure Analyze Utilities Help fon fon tee Vv A E 3 D F
45. ore waveform operations Oscilloscope Experience Redefined Experience the Speed Segmented memory A smart and efficient way to capture waveforms Acquisition memory size is an essential oscilloscope specification because it determines the amount of data you can capture in a single acquisition In general longer memory is better However no memory is always long enough to capture all the signals you need especially when capturing infrequent anomalies data bursts or multiple serial bus packets Segmented memory acquisition lets you selectively capture and store important signal activity without capturing unimportant signal idle time with the time stamp of each segment relative to the first trigger event Segmented memory comes standard in the 4000 X Series Segment 90 1000 7 28752ms Figure 7 shows segmented memory successfully capturing 1 000 events in 3 27274 seconds Traditional memory architecture would require 2 Gpts of memory to accomplish the same result This memory is not available on any scope in the market Segment 50 1000 4 01147ms A gt Segment 10 1000 735 408us Figure 7 Segmented memory efficiently manages the memory to capture up to 1 000 segments of interest to you making it an effective ultra deep memory oscilloscope that can easily capture infrequent events and anomalies Segmented memory serial decode se Be 1 486500 0 0s Segmented memory works in conjunction with se
46. rd Yes option Yes option Mask limit testing Yes option Yes option Yes option Power analysis Yes option Yes option No USB 2 0 signal quality test Yes option No No HDTV analysis Yes option Yes option No Advanced waveform math Standard Yes option No Connectivity Standard USB2 0 LAN video out Standard USB2 0 LAN Video Standard USB2 0 LAN Video GPIB option option GPIB option option GPIB option Need more memory bandwidth analysis or bits Ho Consider the Infiniium 9000A and 9000 H Series 600 MHz 1 GHz 2 5 GHz 4 GHz A Series e 250 MHz 500 MHz 1 GHz 2 GHz H Series Up to 20 GSa s A Series up to 10 GSa s H Series e 20 Mpts A Series 50 Mpts H Series per channel standard e Up to 1 Gpts optional A Series up to 500 Mpts optional H Series e 4 channel 16 digital channel MSO or upgrade e Up to 12 bits of vertical resolution H Series Industry s largest 15 inch touch display Industry s only serial protocol viewer with multi tab viewing e Widest range of applications including serial compliance jitter analysis and more See www agilent com find 9000 and www agilent com find 9000H Figure 2 Infiniium 9000 Series oscilloscope for details Oscilloscope Experience Redefined Experience the Speed One million waveforms per second update rate If you can t see the problem it is hard to troubleshoot it With an industry leading one million waveforms per second update rate the In
47. rial protocol decode For example by setting the trigger 4 i as Data 4 92 26 GE 42 7717 condition to CAN serial bus error segmented memory 14 718 pera 2 captures and stores only CAN error packets and stitches ac a oe E together each segment for easy viewing You can quickly Biso pr peep or compare time tags to discover time intervals between errors K O Segment 150 150 Segment 150 of 150 15 0755s Segment 1 Time 2012 08 24 12 36 31 t Segmented Current Seg of Segs Z 150 150 Figure 8 Segmented memory being used in conjunction with serial decode resulting in maximum insight into serial bus Oscilloscope Experience Redefined Experience the Speed Mask limit testing option Whether you are performing pass fail tests to specified standards in manufacturing or testing for infrequent signal anomalies mask limit testing can be a valuable productivity tool DSOX4MASK The 4000 X Series features powerful hardware based mask testing and can perform up to 2 0 000 tests per second You can select multiple test criteria including the ability to run tests for a specific number of acquisitions a specified time or until detection of a failure Search and navigation The parametric and serial bus search and navigation feature comes standard on the 4000 X Series oscilloscopes When you are capturing long complex waveforms using an oscilloscope s deep acquisition memory manually scrolling through
48. robe active or current Simultaneous 1GHz bandwidth probes are supported simultaneously for Industry s only integrated DVM across all 4 channels demanding applications Asynchronous from the 4 analog triggered waveforms 21 Oscilloscope Experience Redefined Configuring your InfiniiVision 4000 X Series Oscilloscope Step 1 Choose your bandwidth and number of channels InfintiVision 4000 X Series scopes oscilloscopes 4022A 4024A 4032A 4034A Bandwidth 3dB 200 MHz 350 MHz Calculated rise time 10 90 lt 1 75 ns lt ins DSOX 2 4 2 4 Input channels MSOX 2 16 4 16 2 16 4 16 For example if you chose 1 GHz 4 16 channels the model number will be MSOX4104A 4054A 4104A 4154A 1 GHz 1 5 GHz lt 450ps lt 300ps 4 4 4 4 16 4 16 4 16 Step 2 Tailor your oscilloscope with integrated capabilities and measurement applications to save time and money Description Serial protocols Embedded serial triggering and decode 1 C SPI Computer serial triggering and decode RS232 UART USB 2 0 Full Low Speed serial triggering and decode USB 2 0 Hi Speed serial triggering and decode Automotive serial triggering and decode CAN LIN FlexRay serial triggering and decode Audio serial triggering and decode S Aerospace and defence serial triggering and decode MIL STD 1553 ARINC 429 Measurement applications Dual channel WaveGen 20 MHz arbitrary function generator 3 digit voltmeter DVM Power analysis application M
49. sting jitter analysis EOP bit width signal ing rate edge monotonicity and rise fall times all based on official USB IF algorithms embedded in the oscilloscope See www agilent com find DSOX4USBSO for more information Power Factor 1 2 658m Real M1 3 41W Apparent i 2 5 18VA Reactive 1 2 3 90VAR Crest 1 hee 1 484 Crest 2 2 861 Phase Angle i 2 48 95 DC RMS Cyc 1 16 0V Measure Current Mean Min Power A C A A 658m 660 72m 649m 47 90 16 036V 16 0V 16 2V 60 119Hz 60 036Hz 59 639Hz 60 524Hz Features Ana Signals Type Statistics Power Power Qualit Be Power Factor B Apply Figure 25 Power quality measurement one of many in the power measurements application Clear Display lan D cicto AD Clear Persistence Capture Waveforms 3 Features Source Auto Set Video 1080p santan 200m7 Analyze Menu Features Test Signals Aa f Display Results USB Signal Quality Dev Hi Speed SQ g Auto Setup Run Test Figure 27 Perform automatic signal quality testing on USB 2 0 low speed full speed and hi speed signals 13 Oscilloscope Experience Redefined Other Key Productivity Tools The Xilinx FPGA dynamic probe The Xilinx FPGA Dynamic Probe for the 4000 X Series DSOX4FPGAX allows internal FPGA signal activity to be correlated to external signal activity Running on an external PC it allows the user to switch between up to 64 signals inside the FP
50. t 10 mV div greater of 1 div or 5 mV 10 mV div 0 6 div 1 GHz to 1 5 GHz lt 10 mV div greater of 1 5 div or 5 mV 10 mV div 1 0 div 40 mVpp DC to 100 MHz 70 mVpp 100 MHz to 200 MHz 200 mVpp DC to 100 MHz 350 mVpp 100 MHz to 200 MHz 6 div from center screen 8 V range 8V 1 6 Vrange 1 6V Denotes warranted specifications all others are typical Specifications are valid after a 30 minute warm up period and 10 C from firmware calibration temperature 26 Oscilloscope Experience Redefined InfiniiVision 4000 X Series Performance Characteristics Trigger type selections InfiniiScan Zone HW zone qualifier Edge Edge then edge B trigger Pulse width Pattern Or Rise fall time Nth edge burst Runt Setup and hold Video Enhanced video HDTV Option ARINC429 Option CAN Option FlexRay Option I C Option I S Option LIN Option Trigger on user defined zones drawn on the display Applies to one analog channel at a time Specify zones as either must intersect or must not intersect Up to two zones gt 200 000 wfm sec update rate Supported modes normal peak detect high resolution Also works simultaneously with the serial decodes and mask limit test Trigger on a rising falling alternating or either edge of any source Arm on a selected edge wait a specified time then trigger on a specified count of another selected edge Minimum 4 ns Trigger on a pulse
51. ted triggering acquisition and viewing on the same instrument This is especially important in today s embedded designs with sophisticated digital control circuitry Unlike other oscilloscopes in this class you can buy a 2 or 4 channel DSO and enable the 16 digital channels already in the instrument at any time to make it an MSO DSOXPERFMSO Serial protocol analysis Hardware based serial protocol decode and triggering Agilent InfiniiVision Series including the new 4000 X Series are the only oscilloscopes to use hardware based serial protocol decoding Other vendors oscilloscopes use software post processing techniques to decode serial packets frames and therefore have slow waveform and decode capture rates and could miss critical events and errors due to a long dead time Faster decoding with hardware based technology enhances the probability of capturing infrequent serial communication errors After capturing serial bus communication you can easily perform a search operation based on specific criteria and then quickly navigate to bytes frames of serial data that satisfy that search criteria The 4000 X Series can decode two serial buses simultaneously using hardware based decoding and display the captured data in a time interleaved lister display Serial protocol decoding can be used simultaneously with segmented memory and InfiniiScan Zone touch triggering The 4000 X Series has 8 options supporting 10 different
52. touchable targets on the industry s largest 12 1 inch display with capacitive touch screen technology mean operation is quick and natural just like your favorite tablet devices Capacitive touch screen technology Capacitive touch screen technology provides enhanced productivity Use the alphanumeric pad for quick annotation 7 place waveforms or cursors in exact positions and drag Figure 12 The industry s largest 12 1 inch display and capacitive docking panels across the screen to see more measurement uch screen technology with large touchable targets information The 4000 X Series offers three ways to access key menus and features touch GUI for those that prefer tablet or smart phone touch interfaces front panel keypads for the traditional oscilloscope users and Agilent Spark pull down menu for users who prefer Windows like operations The 4000 X Series also offers a touch off button as well as USB mouse and keyboard support 78 125000us 81 250000us 12 308kHz 2 79150V Redefine your remote Web control oscilloscope experience The 4000 X Series not only supports traditional control via a PC Web browser but also supports remote control through popular tablet devices Cursors x1 y2 X2 78 125000us Figure 13 See 10 measurements cursor information and the DVM simultaneously by dragging the desired docking panel to any open area Agilent Technologies Figure 14 Use the Agilent Spark pull down menu for