Home
User Manual Defibtech DDU-2300 Semi
Contents
1. Do not spray cleaning solutions directly on the unit or its connectors Do not use abrasive materials or strong solvents such as acetone or acetone based cleaning agents To wipe the DDU 2300 case clean use a soft cloth dampened with one of the following recommended cleaning agents Soapy water Ammonia based cleaners e g Windex Formula 4099 Fantastik Hydrogen peroxide sopropyl alcohol 70 percent solution Chlorine bleach 30 ml liter water Ensure that the connector socket is completely dry before reinstalling the pads cable After cleaning allow the unit to completely dry Before returning it to service always turn the unit on for a few seconds If the unit detects a problem a service code voice prompt will be heard Otherwise turn the unit off Storage The DDU 2300 AED should be placed in a readily accessible location in an orientation where the Active Status Indicator next to the ON OFF button can be easily seen In general the unit should be stored in clean dry and moderate temperature conditions Make sure that the environmental conditions of the storage location are within the ranges detailed in the Environmental section in Chapter 9 of this manual 5 5 DAC 2513EN BC Operator s Checklist The following checklist may be used as the basis for an Operator s Checklist The table should be copied and filled out as recommend by the schedule in the Routine Maintenance section of this chapt
2. Voltage fluctuations flicker emissions N licabl B IEC 61000 3 3 ot applicable attery operated equipment ELECTROMAGNETIC IMMUNITY Electrostatic discharge ESD IEC 60601 4 2 6 kV contact 6 kV contact There are no special requirements 8 kV air 8 kV air with respect to electrostatic discharge 2 kV for power line Electrical fast transient burst supply lines IEC 61000 4 4 1 kV for input output lines 1 kV line s Surge to line s IEC 61000 4 5 2 kV line s to earth Not applicable Battery operated equipment Not applicable Battery operated equipment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Not applicable Not applicable Battery operated equipment Power frequency 50 60 Hz magnetic Power frequency magnetic fields field should not be greater than levels characteristic of a typical location in a commercial or hospital environment IEC 61000 4 8 10 1 DAC 2513EN BC Electromagnetic Immunity continued Portable and mobile RF communications equipment should be used no closer to any part of 10 V m the DDU 2300 including cables 80 MHz to than necessary The recommended Radiated RF 2 5 GHz separation distance calculated IEC 61000 4 3 from the equation applicable to the 80 5Hz AM frequency of the transmitter is shown Modulation in the following table Interference may occur in the vicinity of equipment
3. id Rescue Breathing Options Softkey Icon Information Main Screen Softkey Icon Text Prompts gt Ben EX er T ILE Sample display screen during Rescue Mode Battery Indicator The Battery Indicator indicates the approximate remaining battery capacity Main Screen The Main Screen displays video instructions to guide the user during a rescue Text Prompts The Text Prompt Area displays text prompts to guide the user during a rescue Softkey Buttons not pictured The Softkey Buttons are located to the right of the display screen If a softkey button is active it will have a softkey icon displayed next to it The softkey buttons are used to navigate menus or select actions Rescue Breathing Options Softkey Icon When this icon is present on the screen during a rescue the user may press the corresponding softkey button to select CPR coaching with compressions only no breathing or CPR coaching with compressions and breathing Information Softkey Icon When this icon is present on the screen the user may press the corresponding softkey button for additional information with video instruction The additional information is context dependent topics include preparing the patient and performing CPR To exit press the softkey button again Preparation Checking The DDU 2300 AED Status Visually check the Active Status Indicator ASI The ASI should flash green The ASI flashes green to indicate ready for use statu
4. marked with this symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 26 957 MHz to 27283 MHz and 40 66 MHz to 40 70 MHz Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the DDU 2300 is used exceeds the applicable RF compliance level above the DDU 2300 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the DDU 2300 Separation Distances The DDU 2300 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the DDU 2300 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the DDU 2300 as recommend
5. AED operational The DDU 2300 AED is designed to be stored in a ready state This chapter tells you how to make the device ready so that if and when you need it few steps are required to begin using the device Overview The following components and accessories are included with your DDU 2300 AED Replacement and other accessories are detailed in the DDU 2300 AED Accessories section Before getting started identify each component and ensure that your package is complete DDU 2300 AED Battery Pack Defibtech Data Card DDC card optional Defibrillation Pads Package Defibtech DDU 2300 Semi Automatic External Defibrillator User Manual on Defibtech User CD o Operating Guide Booklet Operating Guide Connecting The Pads The DDU 2300 AED defibrillation pads are supplied sealed in a package with the connector and part of the cable exposed This allows the pads to be stored in a pre connected state for rapid deployment during an emergency are to be used The packaging should be opened only immediately prior to use otherwise the pads may dry WARNING out and become non functional i DO NOT open the sealed pads package until the pads Note The DDU 2300 AED is designed to be stored with the pads connector already installed This simplifies the procedure for deploying and operating the device in an emergency First check to ensure that the pads package has not expired The expiration date is printe
6. Active Status Indicator beep Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Volume level AED Main Menu gt AED options gt Volume level What it does The Volume level option allows the user to set AED audio to high medium or low volume Once the Volume level option is selected use the TOP and BOTTOM softkey buttons to cycle through the different volume settings When the desired volume selection is chosen press the CENTER softkey button to set that volume level The AED will now use that volume setting for all audio except the volume of the Active Status Indicator beep The user may use the TOP or BOTTOM softkey buttons to navigate to other choices in the menu Note The factory default setting of the volume level is high Audio Recording The Audio recording option allows the user to enable or disable recording of event audio data to a Defibtech Data Card DDC card Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Audio recording AED Main Menu gt AED options gt Audio recording What it does The Audio recording option allows the user to enable disable recording of event audio data Once the Audio recording option item is selected use the TOP and BOTTOM softkey buttons to select either enabled or disabled settings When the desired selection is chosen press the CENTER softkey button to set the feature The AED will now use that au
7. Mode To enter Navigate to AED Maintenance AED Main Menu AED maintenance What it does The unit will display the AED Maintenance menu screen This screen will allow the user to navigate further to perform various maintenance tasks Perform AED test Upgrade AED Transfer data to card Format data card Run application from card To exit Use the TOP or BOTTOM softkey buttons to scroll to and highlight the selection Go to main menu Press the CENTER softkey button The unit will exit the AED Maintenance screen and return to the AED Main Menu screen gt Perform AED Test Perform AED test will initiate a system hardware and software self test Note Running manually initiated AED tests will consume approximately one shock s worth of battery life Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Perform AED test AED Main Menu AED maintenance gt Perform AED test What it does When the user selects Perform AED test selection and presses the CENTER softkey button the unit will begin performing the self test procedure The unit says Performing AED test The unit displays Testing AED The unit will then prompt the user to Press flashing shock button Continue to follow the directions until the test is complete Once the AED test is complete the unit will verbally and visually report the status of the AED The information will be displayed in a pop up window The user mus
8. System time 6 5 System date e Volume level DAC 2513EN BC Audio recording To exit Using the TOP or BOTTOM softkey buttons scroll to and highlight the selection Go to main menu Press the CENTER softkey button The unit will exit the AED Options menu screen and return to the AED Main Menu i gt System Time The System time option allows the user to set the time of the internal AED clock Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to System time AED Main Menu gt AED options System time What it does The System time option allows the user to set the time of internal AED clock using the 24 hour clock Once the System time option is selected press the CENTER softkey button to enter set time mode The hours selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the hours to the desired time Press the CENTER softkey button to accept the hours setting The minutes selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the minutes to the desired time Press the CENTER softkey button to accept the minutes setting The seconds selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the seconds to the desired time Press the CENTER softkey button to accept the seconds setting The time is now set and the user may use TOP or BOTTOM softkey buttons to navigate to other choices
9. USB connector port and Defibtech Data Card DDC card slot Unit Serial Number The unit s serial number can be found on the back of the AED above the battery pack opening Battery Pack Opening This opening is where the battery pack is inserted into the unit Battery Pack Eject Release Latch This release latch releases the battery pack from the DDU 2300 AED Pad Storage Area The pad storage area is found on the back of the AED allowing the pads to be stored in a pre connected state for rapid deployment during an emergency Pads Connector This connector attaches the patient pads to the unit at the Pads Connector Socket item E Defibrillation Pads The defibrillation pads are pads that are placed on the patient The pads may be stored in the pad storage area on the back of the unit Defibrillation Pads Expiration Date back side The defibrillation pads expiration date is located on the back side of the pads package Do not use the pads after the printed date has passed Battery Pack Serial Number The battery pack s serial number is located on the label on the battery pack Battery Pack The battery pack provides a replaceable main power source for the DDU 2300 AED Battery Pack Expiration Date The battery pack expiration date is printed on the label on the battery pack Do not use the battery pack after the printed date has past 1 2 DAC 2513EN BC E Pads Connect
10. Using non Defibtech Data Cards A DDC cards may damage the unit CAUTION 4 will void the warranty gt Run Application From Card Run application from card will initiate a card application on the Defibtech Data Card DDC card The most typical application is a software upgrade Before you begin Be sure the unit is in Maintenance Mode Be sure a DDC Card with a card application is installed in the device Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual To enter Navigate to Run application from card AED Main Menu AED maintenance gt Run application from card Do not turn off the unit or remove the battery A pack or data card until the operation is WARNING Complete Note f a data card is not inserted the unit will speak and display Data Card Missing Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual To exit When the unit finishes performing the application follow the displayed and spoken instructions To manually configure AED options such as time date volume and audio recording select the AED options from the AED Main Menu screen Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to AED Options AED Main Menu AED options What it does The unit will display the AED Options menu screen This screen will allow the user to modify the following user changeable parameters
11. as cellular phones and two way radios can cause improper AED operation Normally using a cell phone near the AED should not cause a problem however a distance of 2 meters 6 feet between RF devices and the DDU 2300 AED is recommended e CPR during analysis can cause incorrect or delayed diagnosis by the patient analysis system Handling or transporting the patient during ECG analysis can cause incorrect or delayed diagnosis especially if very low amplitude or low frequency rhythms are present n patients with cardiac pacemakers the DDU 2300 AED may have reduced sensitivity and not detect all shockable rhythms If you know the patient has an implanted pacemaker do not place electrodes directly over an implanted device During defibrillation air pockets between the skin and defibrillation pads can cause patient skin burns To help prevent air pockets make sure self adhesive defibrillation pads completely adhere to the skin Do not use dried out or expired defibrillation pads Userinitiated and automatic self tests are designed to assess the DDU 2300 AED s readiness for use However no degree of testing can assure performance or detect abuse damage or a defect that occurred after the most recent test is completed Use of damaged equipment or accessories may cause the device to perform improperly and or result in injury to the patient or operator A CAUTIONS Conditions hazards or unsafe practices that may result in
12. contact professional emergency services If another person is available the user should direct that person to call for help and then continue the rescue without delay Preparing The Patient Prepare the patient by removing any clothing from the patient s chest Wipe away moisture from the chest if necessary the defibrillation pads will stick better on dry skin If necessary shave excessive chest hair which can prevent effective patient electrode contact To ensure that defibrillation pads fully contact the patient s skin check that no jewelry or other objects are directly underneath where the pads will be placed Connecting Defibrillation Pads To The DDU 2300 AED Connect pads to unit if not already connected Follow AED voice and display instructions The DDU 2300 AED is designed to be stored with the defibrillation pads connector attached to the unit while the pads themselves remain sealed in their package This reduces the time needed to setup and start treatment in an emergency The Defibtech AED should be stored with the pads connector attached to the unit However if pads were damaged or not properly connected you may need to substitute a new set of pads during an emergency The pads connector socket is on the top left corner of the AED To detach a set of pads from the unit pull firmly on the pad connector Do not reuse used pads Insert the connector for the new pads as shown above The connector will only fit in one way
13. equipment in contact with the patient Do not touch equipment connected to the patient during defibrillation Disconnect other electrical equipment from the patient before defibrillation Do not shock with defibrillation pads touching each other Do not shock with gel surface exposed Do not allow defibrillation pads to touch each other or to touch other ECG electrodes lead wires dressings transdermal patches etc Such contact can cause electrical arcing and patient skin burns during defibrillation and may divert defibrillating energy away from the heart The defibrillation pads are intended for one time use only and must be discarded after use Reuse can lead to potential cross infection improper performance of the device inadequate delivery of therapy and or injury to the patient or operator Avoid contact between parts of the patient s body and conductive fluids such as water gel blood or saline and metal objects which may provide unwanted pathways for defibrillating current 2 1 DAC 2513EN BC WARNINGS continued Disconnect all non defibrillator proof equipment from the patient before defibrillation to prevent electrical shock haz ard and potential damage to that equipment Aggressive or prolonged CPR to a patient with defibrillation pads attached can cause damage to the pads Replace the defibrillation pads if they become damaged during use Possible Radio frequency RF interference from RF devices such
14. off scroll to and select Turn AED off or simply turn the unit off by 6 1 pressing the ON OFF button DAC 2513EN BC Entering Maintenance Mode Before You Begin Make certain the DDU 2300 is turned off and a battery pack is installed Jan 19 2011 17 43 13 STEP 1 Press and release the CENTER softkey button AED status i Result The unit will turn on and display the AED Status Screen for short period of time Battery status h Expires 01 2014 Pads status Adult If the unit does not turn on check to make sure a good battery pack Expires 06 2014 is installed Refer to the Troubleshooting section in Chapter 5 AED S N 099999999 Battery S N 099999999 TF of this manual Software version V13 STEP 2 Press the BOTTOM softkey button to the right of the tool icon Note f the bottom softkey button is not pressed within a short period of time the unit will automatically turn off Result The unit will enter Maintenance Mode and display the AED Main Menu screen The AED Main Menu Screen allows the user to view the status of the AED perform maintenance functions change AED options and access help topics All maintenance functions are accessed through the AED Main Menu Screen The user may select from the following options using the softkey buttons Rescue now Puts device in Rescue Mode AED status AED status Displays current AED status information AED maintenance AED maintenance Displays AED Maintenanc
15. rescue breaths and to give breaths Purpose This instructs the user to give rescue breaths Each time the instruction is given the user should give the patient a rescue breath Operational Environment The Defibtech AED is designed to operate in a wide range of environmental conditions To ensure the reliability and safety of the AED in a given environment refer to the Environmental section in Chapter 9 of this manual for a detailed list of approved environmental conditions 4 10 DAC 2513EN BC 5 Maintenance And Troubleshooting This chapter describes the maintenance and troubleshooting procedures for the DDU 2300 AED The self tests that are performed by the device are described along with the frequency and nature of routine maintenance for which the owner operator is responsible A troubleshooting guide is provided to help diagnose user serviceable problems The DDU 2300 AED contains no user serviceable parts Routine Unit Maintenance Although the DDU 2300 AED is designed to be very low maintenance simple maintenance tasks must be performed by the owner operator on a regular basis to ensure the unit s dependability Check that the Active Status Indicator ASI is flashing green Check the condition of the unit and accessories Run manually initiated self test Replace pads Check pad and battery pack expiration dates Check the DDC card if one was installed Note f the unit has be
16. the unit is turned off Provides maintenance personnel with additional parameter information to assist in troubleshooting a device suspected of malfunctioning DefibView Il is a stand alone software application DefibView Il cannot be used while the AED is in operation and its function is solely to support post event review A Not intended for clinical use Information presented by DefibView should not be used for making clinical decisions CAUTION Defibtech Data Cards DDC cards If a DDC card is installed in the unit every time the DDU 2300 AED is turned on the following information is recorded on a new file on the card The time the AED was turned on Other data such as ECG data time data audio data audio enabled cards only event milestones such as motion detection shock advice shock delivery information This information can be reviewed using the DefibView ll application Using non Defibtech Data Cards DDC cards may damage the unit CAUTION 2 4 will void the warranty Downloading The Internal Data Log Regardless of whether a DDC card is installed in the unit select information is recorded internally within the DDU 2300 AED The information recorded is limited to he time the AED was turned on Other data such as event milestones motion detection shock advice shock delivery information etc Important ECG information Note Audio data is not logged internally Downloading The Interna
17. 3 427 5 601 612 5 607454 5 611 815 5 617853 5 620 470 5 662 690 5 735 879 5 749 904 5 749 905 5 776 166 5 800 460 5 803 927 5 836 978 5 836 993 5 879 374 6 016 059 6 047 212 6 075 369 6 438 415 6 441 582 12 1 DAC 2513EN BC
18. 300 AED Status section in this chapter 3 4 DAC 2513EN BC A Using The DDU 2300 AED In Rescue Mode This chapter describes how to use the DDU 2300 AED in rescue mode The DDU 2300 AED was designed for simple operation allowing the operator to focus on the patient There are two primary control buttons and a display screen Concise and easily understandable voice messages and text and video prompts guide the operator through the use of the unit The following sections describe in detail how to use the DDU 2300 AED The basic steps for use are Turn the DDU 2300 AED ON by pressing the ON OFF button Plug in pads connector into Pad Connector Socket on AED if not yet plugged in Place pads on patient follow instructions on pads package Follow voice and display prompts Press SHOCK button if instructed by the AED Overview Pads Connect or Socket Socket for pads connector Active Status Indicator ASI MM Indicates the current status of the AED ON OFF Button Turns AED on and off Softkey Buttons Top Center Bottom Buttons used to navigate Display Screen menus or select options Displays video text prompts and information Speaker Creates audio output from device SHOCK Button Enabled disabled by software to allow the user to discharge a shock to the patient 4 1 DAC 2513EN BC Overview continued Unit Display Screen During Rescue Mode Battery Indicator
19. 300 AED and or defibrillation training as required by local state provincial or national regulations Any additional training as required by the authorizing physician Thorough knowledge and understanding of the material presented in this User Manual 1 4 DAC 2513EN BC 2 Dangers Warnings And Cautions This chapter includes a list of danger warning and caution messages that relate to the Defibtech DDU 2300 AED and its accessories Many of these messages are repeated elsewhere in this User Manual and on the DDU 2300 AED or accessories The entire list is presented here for convenience A DANGERS Immediate hazards that will result in serious personal injury or death e Hazardous electrical output This equipment is for use only by qualified personnel The DDU 2300 AED is not suitable for use in the presence of a flammable anesthetic mixture Not suitable for use in oxygen enriched atmosphere e The DDU 2300 AED has not been evaluated or approved for use in hazardous locations as defined in the National Electric Code standard In compliance with IEC classification the DDU 2300 AED is not to be used in the presence of flammable substance air mixtures A WARNINGS Conditions hazards or unsafe practices that may result in serious personal injury or death Improper use can cause injury Use the DDU 2300 AED only as instructed in the User Manual The DDU 2300 AED delivers electrical energy that can potentially cau
20. 300 AED contains no user serviceable parts If the unit needs servicing call Defibtech Refer to Chapter 12 of this manual for contact information DAC 2513EN BC 5 8 6 Maintenance Mode Overview Maintenance Mode for the Defibtech DDU 2300 permits the user to perform maintenance related actions such as viewing unit information initiating unit self tests changing unit parameters downloading rescue data and upgrading software Maintenance Mode is navigated through a series of screens menus and menu options In Maintenance Mode the softkey buttons located directly to the right of the display screen are used to scroll through and select menu options When a softkey icon for example an arrow appears on the display screen directly to the left of a softkey button the softkey button is functional for that screen If a softkey icon is not displayed on the screen then the corresponding softkey button has no functionality for that screen Note While the unit is in Maintenance Mode it cannot perform a rescue Maintenance Mode allows the user to go directly to Rescue Mode by selecting the Rescue now option The Rescue now option appears at the top of every screen menu when the unit is in Maintenance Mode The user can also exit Maintenance Mode at any time and go to Rescue Mode by pressing the ON OFF button to turn the unit off and then immediately pressing the ON OFF button again to turn the unit back on The Display Screen During M
21. AED is OFF and ready for use Solid Green DDU 2300 AED is ON and ready for use Flashing or Solid Red DDU 2300 AED needs immediate service Refer to the Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Active No Flashing Light DDU 2300 AED needs immediate service Refer to the Status Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Indicator 5 1 DAC 2513EN BC Checking The AED Status Using The AED Status Screen You may also check the status of the unit when it is off by pressing the center softkey button to enter Maintenance Mode and display the AED Status Screen The AED Status Screen is used to provide a quick overview of the DDU 2300 AED s status and to display select information without turning the unit on in Rescue Mode AED status With the AED off press and release the CENTER softkey button Battery status to display the AED Status Screen The AED Status Screen will be Expires 01 2014 displayed for a short period of time Pads status Adult Expires 06 2014 If the unit does not turn on check to make sure a good battery pack AED S N 099999999 is installed Refer to the Troubleshooting section in this chapter Battery S N 099999999 T Software version V13 ei From the AED Status Screen you can enter Maintenance Mode by pressing the softkey button to the right of the tool i Jan 19 2011 17 43 13 Note f the unit requi
22. DC card is not inserted the unit will speak and display Data Card Missing Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual Press any softkey button to acknowledge the message and then install a Defibtech Data Card DDC card To exit When the unit finishes performing the AED upgrade follow the displayed and spoken instructions gt Transfer Data To Card Transfer data to card will initiate a data transfer from the DDU 2300 AED to a Defibtech Data Card DDC card inserted in the device Internal event data and device history is written to the DDC card Before you begin Be sure the unit is in Maintenance Mode Be sure a DDC Card is installed in the device Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual To enter Navigate to Transfer data to card AED Main Menu gt AED maintenance gt Transfer data to card What it does The unit begins transferring rescue data to the card The unit says Transferring data to data card The unit displays Transferring Data The unit will finish transferring data to the data card and will speak and display data transfer complete pack or data card until the operation is WARNING Complete j Do not turn off the unit or remove the battery Note f a data card is not inserted the unit will speak and display Data Card Missing Refer to the Installing The Defibtech Data Card DDC
23. DU 2300 AED is charging and the operator and others should stand clear of the patient Analysis will continue during this phase and analyzing prompts will continue to be displayed A tone will sound to indicate charging progress If the unit detects a rhythm change to a non shockable one charging will abort and the user will be prompted to begin CPR Press flashing SHOCK button Press SHOCK Button Purpose The DDU 2300 AED has fully charged the heart rhythm analysis algorithm still indicates that a shock is recommended and the unit is ready to deliver a shock The operator should press the SHOCK button to deliver the shock The SHOCK button will flash during this phase and will cancel after 30 seconds Important The DDU 2300 AED will not automatically deliver a shock the user must press the SHOCK button Shock x delivered Shock x Delivered Purpose The DDU 2300 AED has delivered the shock The x indicates the number of shocks that have been delivered since the unit was turned on After each shock the AED will enter Post Shock CPR mode AHA ERC 2010 Protocol Shock cancelled Shock Cancelled Purpose The DDU 2300 AED has aborted shock mode If while waiting for the SHOCK button to be pressed the unit detects a rhythm change to a non shockable rhythm the unit will cancel the shock Also if the SHOCK button is not pressed within 30 seconds of the initial press flashing SHOCK button prompt the unit will a
24. DU 2300 AED is on The speaker also emits a beep when the unit is off and has detected a condition that requires attention from the user or needs servicing SHOCK Button This button will flash when a shock is recommended Pressing this button will deliver a shock when the button is flashing This button is disabled at all other times Display Screen Color display panel used to display text and video prompts messages indicators for rescue unit status and maintenance operations The display screen provides visual prompts including CPR coaching to assist rescuers with step by step instruction ON OFF Button This button is used to turn the DDU 2300 AED on and off Pads Connector Socket The Pads Connector item O is inserted into this socket Active Status Indicator ASI The ASI indicates the current status of the AED This indicator flashes green to indicate the unit is ready for use and flashes red to indicate unit needs attention from the user or needs servicing Softkey Buttons Three context sensitive softkey buttons are used to navigate menus or select actions USB Port The USB port is provided to perform data recovery and maintenance Not to be used during rescue operation Defibtech Data Card DDC card This optional plug in card provides enhanced storage capabilities to the DDU 2300 AED USB and Defibtech Data Card DDC card Access Door Behind the access door is the
25. Defibtech DDU 2300 Semi Automatic External Defibrillator User Manual PRESS ON BUTTON APPLY PADS FOLLOW AED INSTRUCTIONS IF INSTRUCTED PRESS SHOCK BUTTON ii DAC 2513EN BC Notices Defibtech shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Information in this document is subject to change without notice Names and data used in the examples are fictitious unless otherwise noted Limited Warranty The Limited Warranty shipped with Defibtech AED products serves as the sole and exclusive warranty provided by Defibtech LLC with respect to the products contained herein Copyright Copyright 2011 Defibtech LLC All rights reserved Copyright questions should be directed to Defibtech For contact information refer to Chapter 12 of this manual Federal Law USA restricts this device to sale by or on the order of a physician CAUTION iii DAC 2513EN BC Contents 1 Introduction To The DDU 2300 AED 0cceceeeeeeeeceeeceeeeeeeeeeceeeeeeeeeeeeeenees 1 1 ec 1 1 The Defibtech DDU2300 AED intecntisse tont etna Lob taaddgenbbandddiietnnsndbintehd dave fag dd 1 2 INCH GENE OMS T 1 4 ContrandiCaU INE seaeteeses Seanacess seule eae 1 4 Operator Training Requirements ssssssssssssseeeeeene enne enne eere nen
26. F button actively indicates unit status If the unit is ready for use the Active Status Indicator ASI will flash green Ready for use means that the DDU 2300 has passed the most recent self test scheduled or user initiated If the unit needs service the ASI will flash red Anytime the ASI flashs red the unit will also beep periodically to call attention to itself The ASI also uses a distinct flash pattern to assist people with color blindness green will flash a single flash and red will flash a double flash The ASI is powered by the battery pack If the battery pack has been completely discharged or is not installed in the unit the active status indication will be off In this case immediately replace the battery pack or reinsert it into the unit to restore active status indication Flashing Green DDU 2300 AED is OFF and ready for use Solid Green DDU 2300 AED is ON and ready for use Flashing or Solid Red DDU 2300 AED needs immediate service Refer to the Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Active No Flashing Light DDU 2300 AED needs immediate service Refer to the Status Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Indicator 3 3 DAC 2513EN BC Checking The DDU 2300 AED Status continued AED Status Screen Jan 19 2011 AED status Battery status Pads status AED S N Battery S N Software version 17 43 13
27. Refer to the DefibView ll section in Chapter 8 of this manual Before installing the DDC card ensure the AED is turned OFF Locate the data card USB port access door on the right hand side of the unit Open the data card USB port access door by slightly pushing and then sliding the door down to release the latch The door will spring open Insert the DDC card into the thin slot in the side of the AED centered above the USB port opening notched end first label side up until it clicks into place The card should be flush with the surface of the slot If the card does not push in all the way it may have been inserted upside down In that case remove the card flip it over and try inserting it again To remove the DDC card press the card as far as it will go and then release Upon release the DDC card will be partially ejected and can be removed by pulling the DDC card the rest of the way out Close the data card USB port access door by closing and then pushing the door up until the door latch engages Using non Defibtech Data Cards DDC cards may damage the unit CAUTION and will void the warranty 3 2 DAC 2513EN BC Installing And Removing The Battery Pack The battery pack provides power to the DDU 2300 AED Do not install the battery pack after the expiration date printed on the label The battery pack is non rechargeable Before inserting the battery pack into the DDU 2300 AED ensure that the battery pack opening in th
28. To check the status of the unit when the unit is off press the center softkey button The display screen will show unit status Expires 01 2014 battery pack status and pad status After a short period of time Adult the display screen and the unit will turn off Expires 06 2014 099999999 099999999 V1 3 TP Completing The Installation Once you have com pleted the previous steps to set up your DDU 2300 AED follow this procedure 1 Turn the unit on by pressing the power ON OFF button 2 Listen for the Call for Help voice prompt 3 Turn unit off by pressing and holding the ON OFF button 4 Listen for the Powering Off voice prompt 5 Check Active Status Indicator to verify that it is flashing green Refer to the Se f Tests section in Chapter 5 of this manual for instructions on how to run a manually initiated self test StoringThe DDU 2300 AED Store the DDU 2300 AED with pads attached in environmental conditions within range of the specifications Refer to the Environmental section in Chapter 9 of this manual The unit should also be stored so that the Active Status Indicator can be readily seen The Active Status Indicator should periodically flash with a green light If it flashes with a red light or does not flash at all the DDU 2300 A for more information Defibtech recommends storing your AED in an easily accessible location ED needs servicing Refer to the Checking The DDU 2
29. aNg IEC 60601 2 4 requirements charge after 6 discharges Charge time measured from initially switching power on Meets or exceeds AAMI DF80 and to charge ready on the charge after 6 discharges IEC 60601 2 4 requirements years Cid Standby life installed in unit 4 years typical new battery at 25 C Self Adhesive Defibrillation Pads Use only Defibtech Pads with your DDU 2300 AED Defibtech self adhesive defibrillation pads have the following characteristics lt 55 Ibs 25kgs Note n the event of a suspected pad defect the pads should be clearly marked Not for Use and returned to Defibtech LLC for analysis Refer to Chapter 12 of this manual for contact information regarding returns 9 5 DAC 2513EN BC Event Documentation Internal Event Record Critical ECG segments and rescue event parameters are recorded greater than 60 minutes and can be downloaded to a removable data card Removable Storage optional Up to 30 hours of ECG and event data storage no audio option or up to 3 hours of audio audio option ECG and event storage on a removable data card Actual length of storage is dependent on card capacity Defibtech Event Viewer DefibView ll is a PC based application program that allows review of ECG data and other patient and device performance parameters after an emergency event DefibView II runs on various Windows platforms including Windows XP and newer versions Minimum sy
30. ad Contact Press pads firmly Press Pads Firmly Purpose The pads are not making proper contact with the patient and the impedance is out of range for proper ECG analysis and shock delivery Check that the pads are properly placed and fully adhering to the patient and that there are no air bubbles between the pads and the patient If the pads are not sticking due to moisture dry the patient If the pads are not sticking due to excessive hair shave or clip excessive chest hair If the prompts continue try replacing the pads with a new set Check pads Check Pads Purpose The pads are making improper contact with the patient or touching each other and the impedance is out of range for proper ECG analysis and shock delivery Check that the pads are not touching each other and that the patient is dry If the prompts continue try replacing the pads with a new set Pausing for CPR Pausing for CPR Purpose If too long a period of time has passed the user should stop attempting to resolve problems with the pads and assess the condition of the patient The user will be prompted to begin CPR Replace pads Replace Pads Purpose The pads are making improper contact with the patient or touching each other and the impedance is out of range for proper ECG analysis and shock delivery If another set of pads is available replace the pads otherwise check that the pads are properly placed and fully adhering to the patient Make sure that t
31. ad from the blue liner before placing it as shown on the picture on the pad Peel the pad off the blue liner only when the pad is ready to be placed on the patient Place the pads with the sticky side of the pad on the patient s skin Pad placement on infants and children under 8 years or less than 55 pounds 25kgs is different than placement for adults and children 8 years or older or over 55 pounds 25kgs If you are unsure of a child s age or weight or do not have child infant pads do not delay treatment Place the pads on the patients bare chest exactly as shown in the picture on the pad See diagrams below Sample Adult Pads Sample Child Infant Pads Vus E uiid For adults and children 8 years or older or For infants and children under 8 years or less over 55 pounds 25kgs use adult pads than 55 pounds 25kgs use child infant pads Place one pad just below the patient s right collar bone Note Child infant pads can be identified by as shown in the picture Place the second pad over the ribs on the patient s left side below the left breast Use picture on pad to determine individual pad placement their blue connector and pads package Place one pad in the center of chest and one pad on the center of the back as shown Use picture on pad to determine individual pad placement 4 4 DAC 2513EN BC Follow DDU 2300 AED Instructions At this point the DDU 2300 AED will check to make sure that the pads are well connected t
32. affected by absorption and reflection from structures objects and people Regulatory Compliance Changes or modifications of this product not expressly approved by Defibtech may void the user s authority to operate the equipment This device complies with part 15 of the FCC Rules and Industry Canada Radio Standard RSS 210 Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept any interference received including interference that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the
33. aintenance Mode jo AED status Mus AED maintenance ptions IN AED options Center Softkey Rescue options Buttons Help topics Turn AED off 7 O Bottom Title The name of the menu appears in a banner across the top of the screen Softkey Buttons Softkey buttons are used to scroll through menus and to select menu options Softkey Button Icons The presence of an on screen icon located directly to the left of a softkey button indicates that the softkey is functional Menu Options The list of options available from a particular menu Use the top or bottom softkey buttons to scroll through the list of options Select an option by pressing the center softkey button once an option is highlighted with a box Navigation in Maintenance Mode The three softkey buttons located to the right of the display screen are used to navigate in Maintenance Mode Typical functions of the softkey buttons are the following Top softkey button Scroll up Center softkey button Select highlighted option Bottom softkey button Scroll down When a menu option is highlighted and then selected typically by pressing the CENTER softkey button either another screen will be displayed with additional menu options or an action will be performed Exiting Maintenance Mode To exit Maintenance Mode and return to Rescue Mode scroll to and select Rescue now or simply turn the unit off and then back on To exit Maintenance Mode and turn the unit
34. card section in Chapter 3 of this manual To exit When the unit finishes transferring the data to the data card it will automatically return to the AED Maintenance menu screen f gt Format Data Card Format data card is a maintenance tool used to repair corrupted cards It is unnecessary to perform this step on cards purchased with your DDU 2300 AED A This step will erase all data on the data card WARNING Before you begin Be sure the unit is in Maintenance Mode Be sure a Defibtech Data Card DDC card is currently installed in the device Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual 6 4 DAC 2513EN BC To enter Navigate to Format data card AED Main Menu gt AED maintenance gt Format data card What it does The unit will format the DDC card that is inserted in the AED The unit says Formatting data card The unit displays Formatting Data Card When unit finishes formatting the DDC card the unit will return to the menu Do not turn off the unit or remove the battery A pack or data card until the operation is WARNING Complete Note f the data card is not inserted the unit will speak and display Data Card Missing Refer to the nstalling The Defibtech Data Card DDC card section in Chapter 3 of this manual To exit When the unit finishes formatting the data card it will automatically return to the AED Maintenance menu screen
35. d Apply pads to patient s bare chest as shown prompt Pads not connected to patient Place pads on patient Pads not making good connection to patient Check pad connection to patient Pads or pad cable damaged Replace pads Poor pad contact to patient or Press pads firmly prompt Dry pads Partial pad connection Replace pads Check that pads are placed securely on patient Check pads prompt Pads touching Separate pads and place correctly on patient Stop motion prompt Stop interference prompt Patient motion has been detected External interference has been detected Stop patient motion Stop external interference Analyzing interrupted prompt Motion or interference detected Stop motion or interference Shock cancelled prompt Patient s ECG rhythm changed No action necessary Shock button not pressed within 30 seconds Press shock button within 30 seconds Low battery insufficient to charge Replace battery pack Bad pad to patient connection Check that pads are placed securely on patient Dry pads Replace pads Replace data card prompt DDC card is full Replace DDC card with a card that is not full DDC card has failed Replace DDC card Pads missing prompt Repair Pads not connected to unit Make sure pads connector is oriented correctly and fully inserted into unit The DDU 2
36. d on the pad pouch and is also reported in the AED status screen Do not use pads past expiration date Discard expired pads 3 1 DAC 2513EN BC Connecting The Pads continued Insert the connector end of the defibrillation pad cable into the pads connector socket on the top left corner of the DDU 2300 AED as shown Insert pads connector firmly until it is fully seated in the unit The connector will only fit in one way if the connector does not fit rotate the connector before trying again The connected pads package can then be stored in the pad storage slot in the back of the DDU 2300 AED After connecting the pads connector to the unit push the pads package rounded end first with the pictures on the package facing out into the pad holder compartment on the back of the AED When the pads package is fully inserted press the pad cable into the groove in the back of the unit to hold the cable in place and tuck any excess cable behind the pads package The pads are intended for one time use only and must be discarded after use or if the package has been opened CAUTION damaged Installing The Defibtech Data Card DDC card Optional The Defibtech Data Card DDC card is used to store event and audio information collected by the AED All DDU 2300 AEDs will operate without DDC cards and will still store critical event information internally DDC cards may be reviewed with a separate Defibtech PC based software package
37. dio recording setting The user may use the TOP or BOTTOM softkey buttons to navigate to other choices in the menu Note The factory default setting of the audio recording is disabled To manually configure rescue options such as Rescue protocol and CPR breathing select the Rescue options from the AED Main Menu screen Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Rescue Options AED Main Menu gt Rescue options What it does The unit will display the Rescue Options menu screen This screen will allow the user to modify certain user changeable parameters CPR breathing Rescue protocol Settings To exit Using the TOP or BOTTOM softkey buttons scroll to and highlight the selection Go to main menu Press the CENTER softkey button The unit will exit the Rescue Options menu screen and return to the AED Main Menu 6 7 DAC 2513EN BC gt CPR Breathing The CPR breathing option allows the user to enable or disable the CPR breathing coaching prompts during CPR Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to CPR breathing AED Main Menu gt Rescue options gt CPR breathing What it does CPR breathing allows the user to enable disable CPR breathing coaching prompts Use the TOP and BOTTOM softkey buttons to select the desired mode When the desired selection is chosen press the CENTER softkey button to set the feature The AED will now
38. e This indicates that the battery pack capacity is low and that the battery pack should be replaced soon The AED will still be able to deliver at least six defibrillation shocks the first time this message is spoken Replace battery now Replace Battery Now Purpose This indicates that the battery pack is almost discharged and that the AED may not be able to deliver defibrillation shocks Replace the battery pack immediately Pads missing Pads Missing Purpose This indicates that the unit did not detect connected pads during a self test Pads expired Pads Expired Purpose This indicates that pads have expired Replace the pads immediately 5 2 DAC 2513EN BC Checking The Condition Of The Unit And Accessories Inspect the unit for dirt and contamination especially in the pads connector socket and around the battery pack opening Refer to the Cleaning section in this chapter of this manual for guidance on cleaning your AED Inspect the unit display screen for damage Look for cracks or other signs of damage on the case especially near the connector socket If any cracks or other signs of damage are visible remove the AED from service and contact an authorized service center Replacing Pads The Defibtech defibrillation pads are intended for one time use only The pads must be replaced after each use or if the package has been damaged The DDU 2300 AED defibrillation pads are supplied in a sealed pouch
39. e Menu screen AED options AED options Displays AED Options Menu screen Rescue options R i Displays R Opti M Help topics escue options Displays Rescue Options enu screen Turn AED off Help topics Displays Help Topics screen Turn AED off Turns unit off When the user selects Rescue now the unit will exit Maintenance Mode and proceed directly to Rescue Mode The other menu options will perform different functions and are described in detail below The AED Status Screen displays unit specific data such as current status battery pack charge state battery pack expiration date pads expiration date unit serial number battery pack serial number and software version number Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to AED status AED Main Menu gt AED status Note When the unit is turned off the AED Status Screen can also be accessed by pressing the center softkey button What it does The unit will display the AED Status Screen This is an informational screen only no action is taken by the AED To exit To exit the AED Status Screen press and release the BOTTOM softkey button The unit will exit the AED Status Screen and return to the AED Main Menu screen 6 2 DAC 2513EN BC The AED Maintenance screen allows the user to select such options as AED tests software upgrades data backups and data card functions Before you begin Be sure the unit is in Maintenance
40. e back of the AED is clean and clear of any foreign objects Insert the battery pack into the opening in the back of the AED Push the pack all the way in until the latch clicks The battery pack will only fit in one way if the battery pack does not fit rotate the battery pack before trying again Once fully inserted the battery pack surface should be flush with the back of the AED Within moments of insertion the DDU 2300 AED will turn on and run a battery pack insertion self test When the test is completed the unit will report the status of the battery pack and shut down Afterwards the Active Status Indicator adjacent to the ON OFF button of the DDU 2300 AED will periodically flash If the indicator flashes green the AED and battery pack are ready for use If the indicator flashes red is solid red or there is no flashing light the AED requires service Refer to the Checking The DDU 2300 AED Status section below for more details on the meaning of the indicator Note The battery pack must be removed from the unit for more than 10 seconds for the battery pack self test to be performed automatically To remove the battery pack push the battery pack eject release latch After the battery pack is partially ejected pull the battery pack out Checking The DDU 2300 AED Status Active Status Indicator ASI Once a fully functional battery pack is installed in the DDU 2300 AED an LED indicator located to the right of the ON OF
41. ed below according to the maximum output of the communications equipment Recommended separation distances between portable and mobile RF communications equipment and the DDU 2300 Separation distance according to frequency of transmitter m Rated maximum 80 MHz to 800 MHz 800 MHz to 2 5 GHz output power of transmitter W d 1 2 P d 2 3JP 10 2 DAC 2513EN BC Separation Distances continued For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer Note 1 As 80 MHz and 800 MHz the separation distance for the higher frequency range applies Note 2 The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz Note 3 An additional factor of 10 3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2 5 GHz to decrease the likelihood that mobile portable communications equipment could cause interference if it is inadvertently brought into patient areas Note 4 These guidelines may not apply in all situations Electromagnetic propagation is
42. edance compensated biphasic truncated exponential waveform The device delivers 150 Joules of defibrillation energy into a 50 ohm load when using adult pads and 50J of defibrillation energy into a 50 ohm load when using attenuated child infant pads also known as pediatric pads Energy delivered does not change significantly with patient impedance although the duration of the generated waveform will vary Defibrillation and AED operating power is supplied by a replaceable non rechargeable battery pack that provides for long standby life and low maintenance operation Battery packs are available in several configurations optimized for use in specific applications Each battery pack is marked with an expiration date The DDU 2300 AED records event documentation internally and optionally on Defibtech Data Cards DDC cards The optional DDC card plugs into a slot in the AED and enables the AED to record event documentation and optionally audio data onto the card Audio recording is selectable through configuration settings Event documentation stored internally can be downloaded onto a DDC card for review A USB port is provided to perform maintenance and data recovery The USB interface allows connection to a personal computer Defibtech PC maintenance software helps support event downloading and unit maintenance operations 1 1 DAC 2513EN BC The Defibtech DDU 2300 AED A Speaker The speaker projects the voice prompts when the D
43. en dropped mishandled or abused a manually initiated self test should be performed Checking The Active Status Indicator The Active Status Indicator ASI is located to the right of the ON OFF button of the DDU 2300 AED and indicates the operational readiness state of the unit The ASI will periodically flash green to indicate that the unit is ready for use Ready for use means that the DDU 2300 has passed the most recent self test scheduled or user initiated If it is flashing red solid red or not flashing at all the AED needs service Anytime the ASI is flashing red the unit will periodically emit two beeps to call attention to it If the ASI is not flashing at all the most likely cause is that the battery pack needs to be replaced Refer to the Installing And Removing The Battery Pack section in Chapter 3 of this manual Once the battery pack has been replaced with a fresh battery pack the ASI should once again flash green If it still does not flash green after inserting a new battery pack the DDU 2300 AED is non operational and may need servicing Call Defibtech for service Refer to the Contacts section in Chapter 12 of this manual If the ASI is flashing red turn the DDU 2300 AED on If the unit does not turn on or does not speak the AED is non operational and requires servicing If the unit does turn on then turn the unit off and the voice prompts will indicate the nature of the problem Flashing Green DDU 2300
44. enseeeesennse 6 9 DDU 2300 AED Accessories 21c2cee cece cece eee eee eee nnmnnn nnmnnn nnmnnn mnnn 7 1 Detibrillation Pads tesserati e tanned Rete ght ot Fate eb ea Sete Ett adde ud tid 7 1 studii 7 1 BEiLXC ra 7 1 UAE o E E E E T E A E TT I E E AT T E bens 7 1 Ia nedum E 8 1 BIO gg m 8 1 Defibtech Data Cards DDC cards eene 8 1 Downloading The Internal Data Log eere 8 1 Technical Specifications LLLeLeeeeeeeeeeeeeeeeeeeeeeeeee eene nnne nnn 9 1 De fibtech DDUS2300 AED antici nc cuero Rc teta get dt 9 1 stunde 9 5 Self Adhesive Defibrillation Pads ssssssssssseeee eem nennen 9 5 Event Documentation sss rennen ee seen nennen e irn nr nnne nnne nnn 9 6 Detibtech EVENT VIGWGOL acero eto eerta eniinn statuto ai i naiaiae fins 9 6 RECYCLING Informations cce emt oer nc Ree detenta ER ee o pta a le 9 6 Notice To European Union GCUSEOITIGES 2c erect rg Fence ves Eget ARETAS 9 6 Electromagnetic Conformity LlLeeeeeeeeeeeeeeeueleeee eren enne nnne 10 1 Guidance And Manufacturer s Declaration sssssssssssee 10 1 Glossary Of Symbols er erecreiirer ere eren rosea n Rn unn RR ERAEN 11 1 CONTA CES me 12 1 Click any of the section headings above to be aut
45. equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help CE Marking And European Union Compliance Radio Transmitter Defibtech LLC declares that the DDU 2300 AED radio transmitter is in compliance with the essential requirements and other relevant provisions of Directive 1999 5 EC Applicable standards are listed in the Environmental section in Chapter 9 of this manual 10 3 DAC 2513EN BC 11 Glossary Of Symbols High voltage present SHOCK Button Delivers defibrillation shock to the patient when the device is ready to shock ON OFF Button Turns the device ON when it is OFF Turns the device OFF when it is ON Caution consult accompanying documents Do not expose to high heat or open flame Do not incinerate Recyclable Consult operating instructions Do not damage or crush Follow proper disposal procedures Meets the requirements of the European Medical Device Directive Note XXXX represents the identification number of the notified body Temperature limitation Use by yyyy mm Defibrillation proof Can withstand the effects of an externally applied defibrillation shock Internally powered with defibrillator proof BF type patient applied parts per EN 60601 1 Manufacturer Date of manufacture Do not reuse For USA users only id 11 1 DAC 2513EN BC Glossary Of Symbo
46. er As each item is completed it should be checked off Defibtech DDU 2300 Operator s Checklist Defibtech DDU 2300 Serial Number Defibtech DDU 2300 Location Check unit and accessories for damage dirt and contamination Clean or replace as necessary Check that spare battery pack and pads are available Check that battery pack and pads are not past expiration dates Check that the ASI is flashing green Comments Inspection by initials or signature 5 6 DAC 2513EN BC Troubleshooting The following table lists the symptoms the possible causes and the possible corrective actions for common problems Refer to the other sections of the user manual for detailed explanations on how to implement the corrective actions If the unit continues to be non functional refer the unit for servicing Refer to Chapter 12 of this manual for contact information Unit will not turn on Battery pack not inserted Insert battery pack Battery pack depleted or needs servicing Replace battery pack or call for service Unit needs servicing Call for service Unit immediately turns off Battery pack depleted Replace battery pack Unit needs servicing Call for service ASI flashes red and or unit makes periodic beep sound ASI does not flash at all Power on test failed service code xxxx Unit needs servicing Go to AED Status Screen by pressing
47. eral Prompts Call for help Call For Help Purpose As soon as the DDU 2300 is turned on the user will be prompted to call for help This indicates that the first step in a rescue should always be to contact professional emergency services If another person is available the user should direct that person to call for help and then continue the rescue without delay Pediatric mode Pediatric Mode Purpose This informs the user that child infant pads are attached to the unit Child infant pads should only be used if the patient is an infant or a child under the age of 8 or less than 55 pounds 25kgs For children 8 years or older or over 55 pounds 25kgs and for adults adult pads should be used Do not delay therapy to determine exact age or weight Training pads Training Pads Purpose This informs the user that training pads are attached to the unit Training pads are used for training purposes only and will not deliver a shock In a rescue immediately replace the training pads with defibrillation pads Purpose This informs the user that the unit is turning off Pad Connection Pad Application Related Prompts Follow instructions to apply pads Follow Instructions Purpose This instructs the user to follow the AED prompts in order to apply the pads to the patient Remove clothing from patient s chest Remove Clothing Purpose This instructs the user to remove all clothing from patient s chest Pads mus
48. f which could pose serious risk for the environment and human health It is for this reason that local administrations provide regulations which encourage reuse and recycling and prohibit the disposal of WEEE as unsorted municipal waste and require the collection of such WEEE separately at specifically authorized treatment facilities Manufacturer and authorized distributors are required to supply information about a safe treatment and disposition of the specific device You may also return this equipment to your distributor when purchasing a new one As for reuse and recycling notwithstanding the limits imposed by the nature and the use of this device the manufacturer will do its best to develop recovery processes Please contact the local distributor for information 9 6 DAC 2513EN BC 10 Electromagnetic Conformity Guidance And Manufacturer s Declaration The DDU 2300 is intended for use in the electromagnetic environment specified below The customer or the user of the DDU 2300 should assure that it is used in such an environment ELECTROMAGNETIC EMISSIONS RF emissions The DDU 2300 uses RF energy only for its internal Group 1 e CISPR 11 Class B function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment CISPR 22 Class B eae Part 15 Class B Harmonic emissions Harmonic emissions IEC 61000 3 2 61000 3 2 Not applicable Battery operated equipment
49. hat a shock is required and has completed charging the SHOCK button will flash and the user will be instructed to press the flashing SHOCK button The user should follow the AED instructions and press the SHOCK button Important The DDU 2300 AED will not automatically deliver a shock the user must press the flashing SHOCK button The user can abort charging or shock delivery at any time by pressing and holding the ON OFF button for approximately two seconds to turn the unit off CPR Period The operator will be prompted to begin CPR The unit will not monitor the patient s ECG rhythm during the CPR period During the CPR period the AED will not advise the user to stop motion even if motion is present The user should follow the AED instructions during this time Once the CPR period is complete the unit will continue in Heart Rhythm Analysis mode CPR coaching is provided through a series of voice and visual prompts and audible tones The factory default setting provides prompts for chest compressions only no breathing However breathing prompts can be enabled disabled by pressing the softkey button next to the Rescue Breathing Options icon displayed on the screen during rescue Refer to the Rescue Breathing Options Softkey Icon section below Breathing prompts can also be enabled disabled by setting the menu option in Maintenance Mode Refer to the CPR Breathing section in Chapter 6 of this manual Rescue Breathing Optio
50. he pads are not touching each other If the pads are not sticking due to moisture dry the patient If the pads are not sticking due to excessive hair shave or clip excessive chest hair If the prompts continue try replacing the pads with a new set Motion Interference Prompts Stop motion Stop Motion Purpose The DDU 2300 AED has detected possible motion in the patient Stop all patient motion including CPR in response to this prompt Stop interference Stop Interference Purpose The DDU 2300 AED has detected interference on the ECG signal Eliminate any radio or electrical sources of interference Check the pads to make sure they are adhering properly to the patient If the environment is very dry minimize movement around the patient to reduce static discharges Pausing for CPR Pausing for CPR Purpose The user should stop attempting to resolve motion and or interference problems and assess the condition of the patient The user will be prompted to begin CPR 4 7 DAC 2513EN BC Heart Rhythm Analysis Prompts Analyzing heart rhythm Analyzing Rhythm Analyzing Analyzing Purpose The DDU 2300 AED is actively analyzing the patient s ECG signal The AED will continue analyzing until it has determined whether a rhythm is shockable or non shockable or if analyzing Is interrupted for some reason Do not touch the patient Do Not Touch Patient Purpose The DDU 2300 AED is trying to analyze
51. ider A new DDC card should be installed before the next use To remove the DDC card ensure the AED is OFF Locate the data card USB port access door on the right hand side of the unit Open the data card USB port access door by slightly pushing and then sliding the door down to release the latch The door will spring open To remove the DDC card press the card in as fas as it will go and then release Upon release the DDC card will be partially ejected and can be removed by pulling the DDC card the rest of the way out To install a new DDC card insert the DDC card into the thin slot in the side of the AED centered above the USB port opening notched end first label side up until it clicks into place The card should be flush with the surface of the slot If the card does not push in all the way it may have been inserted upside down In that case remove the card flip it over and try inserting it again Close the data card USB port access door by closing and then pushing up the door until the door latch engages Note DDC card is not required for the DDU 2300 AED to operate Even if a DDC card is not installed the unit will still record basic essential information internally The AED will still operate properly even after a replace data card message 5 4 DAC 2513EN BC Self Tests The DDU 2300 AED provides for both automatic and manually initiated self tests These self tests test various components of the AED including the
52. if the connector does not fit rotate the connector before trying again Insert connector firmly until it is completely seated in the unit When this Information Softkey Icon is present on the screen the user may press the corresponding softkey button for additional information with video instruction To exit press the softkey button again Opening The Pads Package Remove the pads package from the pad storage slot at the back of the AED Open the pads package by tearing along the dotted line starting at the black arrow follow directions on the package Check that the pads are Free from obvious signs of damage Clean of excessive debris for example dirt if the pad was dropped Not dried out and that the gel is sticky and will adhere to the patient Not expired Do not use pads after the expiration date printed on the package If any of these conditions are found use a new set of pads if possible 4 3 DAC 2513EN BC Applying Pads To The Patient Apply pads correctly to patient Follow AED voice and text prompts Correct pad placement is essential for effective analysis of the patient s cardiac rhythm and subsequent shock delivery if required Remove the pads from the pads package by tearing the package along the dotted line near the top of the package Follow the directions and diagram showing proper defibrillation pad placement located on the defibrillation pads package and on the pads Peel off each p
53. in CPR compressions immediately The unit will emit a beep at the rate that compressions should be given Z Continue E Continue xx Seconds Continue for 1 minute xx seconds Purpose This indicates that the user should continue performing CPR This phrase is spoken to let the user know that the unit is still operating normally The unit will not be monitoring the patient s ECG rhythm during this mandatory two minute CPR period AHA ERC 2010 Protocol Ending in 5 4 3 2 1 Ending in xx Seconds Purpose This indicates that the user should prepare to finish performing CPR This phrase is spoken during the last several seconds of the CPR period to let the operator know that the unit is still operating normally and that the CPR period is ending Stop CPR Stop CPR Stop Now Stop Now Purpose This indicates that the CPR period has ended and the user should stop CPR The unit will enter Analyze Mode Do not touch the patient Do Not Touch Patient Analyzing heart rhythm Analyzing Rhythm Purpose This indicates the unit has entered Analyze Mode and is performing an ECG analysis The user should not touch the patient during the ECG analysis 4 9 DAC 2513EN BC CPR Coaching Help Prompts Purpose This reminds the user of the correct placement of hands for CPR Purpose This reminds the user to perform CPR compressions Purpose This guides the user to prepare the patient for
54. in the menu Note The factory default setting of the internal AED clock is Universal Time GMT gt System Date The System date option allows the user to set the date of the internal AED clock Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to System date AED Main Menu gt AED options System date What it does The System date option allows the user to set the date of internal AED clock Once the System date option is selected press the CENTER softkey button to enter set date mode The year selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the year Press the CENTER softkey button to accept the year setting The month selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the month Press the CENTER softkey button to accept the month setting The day selection will be highlighted green Press the TOP or BOTTOM softkey buttons to adjust the day Press the CENTER softkey button to accept the day setting The date is now set and the user may use TOP or BOTTOM softkey buttons to navigate to other choices in the menu Note The factory default setting of the internal AED clock is Universal Time GMT 6 6 DAC 2513EN BC gt Volume Level The Volume level option allows the user to set AED audio output to high medium or low volume Changing the volume level will not change the volume of the
55. l Data Log Using The DDC Card To download the internally logged information perform the following procedure Insert a DDC card into the unit Turn the unit on in Maintenance Mode by pressing the center softkey button Press the tool icon to enter the AED Maintenance screen From the AED Maintenance screen select the Transfer Data To Card option Allow the unit to write the contents of the internal log to the DDC card The DDU 2300 AED will write the contents of the internal log onto the DDC card This information can then be reviewed using DefibView ll software Downloading The Internal Data Log Using The USB Port To download the internal data log using the USB port in the unit connect a USB cable between the unit and a PC Launch the DefibView Il software and follow USB download directions 8 1 A Do not operate the DDU 2300 AED in Rescue sieneeee Mode while a USB cable is plugged into the unit WARNING 9 Technical Specifications Defibtech DDU 2300 AED General 18 5 x 24 x 5 8 cm Meets applicable requirements of EC 60601 1 UL 60601 1 Design standards CAN CSA C22 2 No 601 1 M90 e IEC 60601 1 2 e IEC 60601 2 4 AAMI DF80 Patient safety All patient connections are electrically isolated Defibrillator Impedance Compensated Biphasic Truncated Exponential Adult 150 Joules nominal 15 delivered into a 50 ohm load Child Infant 50 Joules nominal 15 delivered into a 50 ohm load
56. ls continued Catalogue number Keep dry Handle with care Transportation and storage requirements See environmental requirements Authorized European Representative pem Does not contain latex Ger fmm 4 IP55 Dust protected Protected against water jets S Classified by TUV Rheinland of NA with respect to electric shock fire and mechanical hazard only A in accordance with UL 60601 1 CAN CSA C22 2 No 601 1 M90 IEC 60601 1 and IEC 60601 2 4 ch Lithium Manganese Dioxide Battery Product is not sterile d v Conforms to UL Standard UL 60601 1 Certified to CAN CSA Standard C22 2 No 601 1 M90 11 2 DAC 2513EN BC 12 Contacts Manufacturer Defibtech LLC 741 Boston Post Road Guilford CT 06437 Tel 866 333 4241 Toll free within North America 203 453 4507 Fax 203 453 6657 Emails sales defibtech com Sales reporting defibtech com Medical Device Reporting service defibtech com Service and Repair European Authorized Representative EC REP Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Tel 31 70 345 8570 Fax 31 70 346 7299 Patents pending This product and its accessories are manufactured and sold under one or more of the following United States patents D523 393 D548 346 D551 628 This product and its accessories are manufactured and sold under license to at least one or more of the following United States patents 5 591 213 5 59
57. lt 4 seconds Charge time from shock advised Charge time may increase at the end of battery life and for temperatures below 10 C AN e SHOCK button flashing Charge complete indication l 2s e Press flashing shock button voice prompt Energy Shock delivery Shock is delivered by a single SHOCK button e f Patient Analysis System decides rhythm is no longer shockable or e Within 30 seconds after charge complete if operator has not pressed Automatic SHOCK button DISARM or e f defibrillation pads are removed from patient or unplugged from unit f operator presses and holds the ON OFF button for approximately two Manual isa seconds device will disarm and turn off typical new battery at 25 C 9 1 DAC 2513EN BC Defibtech DDU 2300 AED continued Waveform Specifications The DDU 2300 AED delivers a 150J Biphasic Truncated Exponential waveform to patients with impedances ranging from 25 to 180 ohms V max v1 cA B t V min va The waveform is adjusted to compensate for measured patient impedance Nominal phase times and energy delivered are shown in the tables below ADULT PEDIATRIC If impedance is out of range for proper analysis and shock delivery voice and or visual prompts will inform the user 9 2 DAC 2513EN BC Defibtech DDU 2300 AED continued Environmental 0 50 C 32 122 F o 5 95 no
58. minor personal injury damage to the DDU 2300 AED or loss of data Follow all battery pack labeling instructions Do not install battery packs after the expiration date Follow all defibrillation pad label instructions Use defibrillation pads prior to their expiration date Do not re use defibrillation pads Discard defibrillation pads after use in the event of suspected pad malfunction return pads to Defibtech for testing Recycle or dispose of lithium battery packs in accordance with federal state and or local laws To avoid fire and explosion hazard do not burn or incinerate the battery pack Do not crush Use and store the DDU 2300 AED only within the range of environmental conditions specified in the technical specifications f possible disconnect the DDU 2300 AED from the patient prior to use of other defibrillators Do not connect the DDU 2300 to a PC or other device using the USB port while the unit s electrodes are still connected to the patient Using non Defibtech Data Cards DDC cards may damage the unit and will void the warranty Although the DDU 2300 AED is designed for a wide variety of field use conditions rough handling beyond specifications may result in damage to the unit e Federal Law USA restricts this device to sale by or on the order of a physician 2 2 DAC 2513EN BC 3 Setting Up The DDU 2300 AED This chapter describes the steps required to make your Defibtech DDU 2300
59. n condensing Maintenance 700 to 1060 hPa 21 to 31 inHg 0 50 C 32 122 F e 5 95 non condensing Transport 500 to 1060 hPa 15 to 31 inHg l 150 to 4500 meters 500 to 15 000 feet per Altitude MILSTD 810F 500 4 Procedure Il Shock Drop Abuse Tolerance MILSTD 810F 516 5 Procedure IV 48 in 1 2 meters any edge corner or surface in standby mode MILSTD 810F 514 5 Category 20 Ground Vibration RTCA DO 160D Section 8 8 2 Cat R Zone 2 Curve G Helicopter RTCA DO 160D Section 8 Cat H Zone 2 Curves B amp R Jet Aircraft IEC60529 class IP55 Dust Protected Protected against water jets Battery pack installed ESD and EMI radiated and immunity Refer to Chapter 10 for details R amp TTE Directive 1999 5 EC Radio Frequency Emissions Applicable ETSI EN 300 220 2 V2 1 2 2007 06 Directive and Standards ERC RECOMMENDATION 70 03 ETSI EN 301 489 3 V1 4 1 2002 08 Sealing Water Resistance Patient Analysis System The DDU 2300 Patient Analysis System ensures that the pad patient impedance is within the proper range and analyzes the patient s ECG rhythm to determine whether a shock is required On detection of a non shockable rhythm the user is prompted to perform CPR For shockable rhythms the AED automatically charges in preparation for shock delivery The patient analysis system will detect electrical noise or artifact in the ECG signal that may interfere with accurate rhythm analysis Thi
60. ne enne 1 4 Dangers Warnings And Cautions Leseeeeeeeeeeeeeeeee eene 2 1 BEIM Ion 2 1 MV EMU DRE m 2 1 SDN C 2 2 Setting Up The DDU 2300 AED eeeeeeeeeee eee nene n nennen nen 3 1 uu 3 1 Connecting Wie Pads cscen secca dite decedit dese E EEEE ETEN AER 3 1 Installing The Defibtech Data Card DDC card Optional sssssssse 3 2 Installing And Removing the Battery Pack ssssssssssssssseeee meme 3 3 Checking The DDU 2300 AED Status en eene eene 3 3 Completing Th InstallatiQris oir etr rerit ott Lio e tienda no Lcx re Lio age ds 3 4 Stoning The DDU 2300 AED ic ictu dont deae dean kan rd esta decks EEAS bau d oca 3 4 Using The DDU 2300 AED In Rescue Mode 4 1 ee c r 4 1 mue M M 4 2 Heart Rhythm Analysts iine oa nah i Fe e a eR Ee m d darte eig e ERR Pn ee eae 4 5 Delivering The Shock sssssssssssssssseeneene mee ener iren enne 4 5 CPR FENO sepre dass ssa 4 5 Post USS PIOCSdUEeS crues tT ERU detected eda nas aderat re tme Deni uemaCel ga 4 6 Rescue Mode Voice And Text Prompts sss 4 6 jersiufenilsin rose ems 4 10 Maintenance And Troubleshooting e eeeeeeeee eere 5 1 Routine Uni
61. ns Softkey Icon During a rescue when this icon is present on the Screen the user may press the corresponding softkey button to select CPR coaching with compressions only no breathing or CPR coaching with compressions and breathing Note Refer to the CPR Breathing section of Chapter 6 for instructions on how to change the factory default setting When this Information Softkey Icon is present on the screen the user may press the corresponding softkey button for additional information with video instruction To exit press the softkey button again 4 5 DAC 2513EN BC Post Use Procedures After the DDU 2300 AED has been used on a patient the unit should be cleaned following procedures in the Cleaning section in Chapter 5 of this manual and prepared for the next use The following steps should be performed 1 Connect a new pads package check to make sure the package is not expired and package is not damaged 2 Perform a self test manually Unit will report status at the end of the self test Refer to the Se f Tests section in Chapter 5 of this manual for instructions on how to run a manually initiated self test 3 Turn off the unit by pressing the ON OFF button 4 Check to make sure that the Active Status Indicator is flashing green Rescue Mode Voice AndText Prompts The following section provides brief descriptions of some of the voice and text prompts that the user will hear and see in Rescue Mode Gen
62. o the patient and that an adequate ECG signal is being received Do not touch the patient Eliminate any patient movement and cease CPR at this time If there is a problem with the pad connection socket connection patient motion or other interference the AED will guide the operator with both audible and displayed instructions Text prompts are identical to or are an ab breviated form of the audio prompts Video prompts reinforce the audio and text prompts and aid in high ambient noise environments Heart Rhythm Analysis Once the DDU 2300 AED has determined that the pads are making a good connection to the patient the AED will start the ECG rhythm analysis The unit analyzes the ECG signal and determines whether a shockable or non shockable rhythm is present While analyzing the AED will continue to monitor signal and pad conditions and will reassess analysis and prompt the user if additional user action is needed Delivering The Shock If the DDU 2300 AED ECG analysis algorithm has determined that a shock is required the unit will automatically charge in preparation for shock delivery While the AED charges the unit may emit a charging tone and will continue to analyze the patient s heart rhythm If the unit detects that the heart rhythm has changed to one that does not require a shock the unit will prompt the user to begin CPR While analyzing the AED will continue to monitor signal and pad conditions If the unit has determined t
63. omatically re directed to that page See page references noted in green throughout this book for other interactive links V DAC 2513EN BC 1 Introduction To The DDU 2300 AED This User Manual provides information to guide trained operators in the use and maintenance of the Defibtech DDU 2300 Semi Automatic External Defibrillator AED and its accessories The DDU 2300 refers to all AEDs in the DDU 2300 series of AEDs This chapter includes intended use an overview of the AED a discussion of when it should and should not be used and information on operator training Overview The DDU 2300 AED is a Semi Automatic External Defibrillator AED that is designed to be easy to use portable and battery powered It has two primary user controls the ON OFF and SHOCK buttons Voice prompts text prompts and a display screen with visual prompts provide a simple interface for the operator The DDU 2300 AED is capable of recording event information including ECG audio data optional and SHOCK NO SHOCK recommendations When connected to a patient who is unconscious and not breathing the DDU 2300 AED performs the following tasks Prompts the operator through audio text and video prompts to prepare the patient for treatment Automatically analyzes the patient s ECG e Determines whether a shockable rhythm is present Charges the defibrillation capacitor and arms the SHOCK button if the AED detects a shockable rhythm Pr
64. ommendation of Specificity gt 99 Meets the AAMI DF80 requirement and the AHA recommendation of Specificity gt 95 Meets the AAMI DF80 requirement and the AHA recommendation of Specificity gt 95 Shockable Rhythm Ventricular Fibrillation Shockable Rhythm Ventricular Tachycardia Non Shockable Rhythm Normal Sinus Rhythm Non Shockable Rhythm Asystole Non Shockable Rhythm All other non shockable rhythms 1 From Defibtech ECG Rhythm Databases 2 Automatic External Defibrillators for Public Access Defibrillation Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance Incorporating New Waveforms and Enhancing Safety American Heart Association AHA Task Force on Automatic External Defibrillation Subcommittee on AED Safety and Efficacy Circulation 1997 95 1677 1682 3 Schneider T Martens PR Paschen H et al Multicenter randomized controlled trial of 150J biphasic shocks compared with 200 to 360 J monophasic shocks in the resuscitation of out of hospital cardiac arrest victims Circulation 2000 102 1780 1787 Note Additional information available upon request 9 4 DAC 2513EN BC Battery Packs Use only Defibtech battery packs in the DDU 2300 AED DBP 2003 Battery Pack Disposable recyclable non rechargeable Charge time from the initiation of rhythm analysis to readiness for discharge with a fresh battery and on the Mec te Bree AAMI PrOD
65. ompts the operator to press the SHOCK button when the device is ready and a shock is recommended Delivers a shock once the device has determined a shock is required and the SHOCK button has been pressed Provides instructions to perform CPR Repeats the process if additional shocks are required The Defibtech AED will NOT shock a patient automatically it will only advise the operator The SHOCK button is enabled only when a shockable rhythm is detected and the device is charged and ready to shock Charging occurs automatically when the device detects a shockable rhythm The operator must press the SHOCK button to initiate defibrillation The DDU 2300 AED uses two self adhesive defibrillation pads also known as electrode pads electrodes monitoring pads or pads to monitor ECG signals and if necessary to deliver defibrillation energy to the patient These pads are provided in a single use disposable package The DDU 2300 AED determines proper pad to patient contact by monitoring the impedance between the two pads impedance varies with the electrical resistance of the patient s body The DDU 2300 AED user interface is clear and concise It has two primary push button controls and a display screen Easily understandable voice messages and text and video prompts guide the operator through the use of the unit The device communicates the status of the AED and of the patient to the operator Defibrillation energy is delivered as an imp
66. on Multiple events may be recorded on a single DDC card If the DDC card becomes full the AED will stop recording to the card however the most critical event documentation for the current session will still be recorded internally Internally recorded event information can be downloaded for external review by inserting a blank DDC card into the unit and following the Data Download procedure Refer to the Downloading The Internal Data Log section in Chapter 8 of this manual Using non Defibtech Data Cards DDC cards may damage the unit CAUTION and will void the warranty USB Cable An optional USB cable may be used with the DDU 2300 AED for connecting the AED to a personal computer running Defibtech maintenance software The AED has a mini USB connector located on the right side of the unit behind the data card USB port access door A Do not have a USB cable attached to the unit during a rescue WARNING 74 DAC 2513EN BC 8 Event Viewing This chapter includes information about DefibView Il Defibtech Data Cards DDC cards and downloading internal data logs DefibView II DefibView Il is a Windows based software application that reads data stored on a DDC card or downloaded through the USB port and displays the data on a personal computer DefibView ll serves the following primary functions Enables emergency care personnel to review a cardiac episode from the time the AED was turned on and connected to the patient until
67. or Socket D ON OFF Button F Active Status Indicator ASI C Display Screen G Softkey Buttons Top Center Bottom B SHOCK Button H USB Port A Speaker l Defibtech Data Card DDC Card optional J USB and Defibtech Data Card DDC Card Access Door FRONT OF AED K Unit Serial Number L Battery Pack Opening M Battery Pack Eject Release Latch O Pads Connector N Pad Storage Area P Defibrillation Pads Q Defibrillation Pads BACK OF AED Expiration Date on back of package ADULT ELECTRODE PADS S Battery Pack R Battery Pack Serial Number T Battery Pack Expiration Date DEFIBRILLATION PADS PACKAGE 1 3 BATTERY PACK DAC 2513EN BC Indications The DDU 2300 Semi Automatic External Defibrillator AED is indicated for use on victims of sudden cardiac arrest SCA who are e Unconscious and unresponsive Not breathing For patients under 8 years old or less than 55 pounds 25kg use child infant electrode pads Do not delay therapy to determine exact age or weight The DDU 2300 AED must be used by or on the order of a physician Contraindications The DDU 2300 AED should not be used if the patient shows any of the following signs Conscious and or responsive Breathing Has a detectable pulse Operator Training Requirements In order to safely and effectively operate the DDU 2300 AED a person shall have met the following requirements DDU 2
68. res service the AED Status Screen will present information about the problem to the user The user should follow the text prompts to address the condition requiring attention Card Application Softkey Icon If an application is present on an inserted Defibtech Data Card DDC card a card icon will also appear next to the center softkey button Pressing this button will load and run the application contained on the card Maintenance Related Prompts Power on test failed Power On Test Failed Service code xxxx Service Code xxxx Purpose This indicates that the DDU 2300 AED has failed the poweron self test and is non operational and needs servicing The code number will indicate to service personnel the type of problem the unit is experiencing Battery test failed Battery Test Failed Service code xxxx Service Code xxxx Purpose This indicates that the DDU 2300 AED s battery pack is non operational and needs servicing The code number will indicate to the service personnel the type of problem that the unit is experiencing Service code xxxx Service Code xxxx Purpose The DDU 2300 AED will report this message when it powers off indicating a service code that was previously detected Service required Service Required Purpose This indicates that the DDU 2300 AED has detected an internal error is non operational and needs servicing Battery low Battery Low Purpos
69. s The ASI flashes red solid red or is not lit at all to indicate that service is required The ASI is powered by the battery pack If the battery pack has been completely discharged or is not installed in the unit the active status indication will not be available In this case immediately replace the battery pack or reinsert it into the unit to restore active status indication Flashing Green DDU 2300 AED is OFF and ready for use Solid Green DDU 2300 AED is ON and ready for use Flashing or Solid Red DDU 2300 AED needs immediate service Refer to the Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Active No Flashing Light DDU 2300 AED needs immediate service Refer to the Status Troubleshooting section in Chapter 5 of this manual or call Defibtech for service Indicator 4 2 DAC 2513EN BC Turning On The DDU 2300 AED Press the green ON OFF button to turn the DDU 2300 AED on The unit will emit a beep and the display screen will turn on The ASI indicator next to the ON OFF button will illuminate green anytime the AED is on To turn the unit off press AND HOLD the ON OFF button for approximately two seconds the unit will emit a beep and power off Call For Help Call professional emergency services for help As soon as the AED is turned on the unit will prompt the user to Call for help This is a reminder that the first step in a rescue should always be to
70. s artifact may be caused by excessive motion to the patient or by external electrical noise When this artifact is present the AED will prompt the user to Stop motion or Stop Interference until the ECG signal is free of noise and then proceed to analysis 9 3 DAC 2513EN BC Defibtech DDU 2300 AED continued Shockable Rhythm Criteria When placed on a patient meeting the indications for use criteria the DDU 2300 AED is designed to recommend a defibrillation shock when it detects proper pad impedance and one of the following Ventricular Fibrillation Peak to peak amplitude at least 200 uVolts A Some very low amplitude or low frequency VF rhythms may not be interpreted as shockable WARNING Ventricular Tachycardia Including ventricular flutter and polymorphic VT Cardiac rhythm rate of at least 180 bpm and peak to peak amplitude at least 200 uVolts A Some very low amplitude or low frequency VT rhythms may not be interpreted as shockable WARNING The DDU 2300 AED is designed to recommend no shock for all other rhythms including Normal Sinus Rhythms fine Ventricular Fibrillation 200 Volts and some slow Ventricular Tachycardia and Asystole Patient Analysis System Performance Meets the AAMI DF80 requirement and AHA recommendation of Sensitivity gt 90 Meets the AAMI DF80 requirement and AHA recommendation of Sensitivity gt 75 Meets the AAMI DF80 requirement of Specificity gt 95 and the AHA rec
71. se death or injury if it is used or discharged improperly Improper maintenance can cause the DDU 2300 AED not to function Maintain the DDU 2300 AED only as described in the User Manual The AED contains no user serviceable parts do not take the unit apart Electrical Shock Hazard Dangerous high voltages and currents are present Do not open unit remove cover or back or attempt repair There are no user serviceable components in the DDU 2300 AED Refer servicing to qualified service personnel e Lithium battery packs are not rechargeable Any attempt to recharge a lithium battery pack may result in fire or explosion Do not immerse battery pack in water or other liquids Immersion in fluids may result in fire or explosion e Do not let fluids get into the DDU 2300 AED Avoid spilling fluids on the AED or its accessories Spilling fluids into the DDU 2300 AED may damage it or cause a fire or shock hazard Do not sterilize the DDU 2300 AED or its accessories Use only Defibtech disposable self adhesive defibrillation pads battery packs and other accessories supplied by Defibtech or its authorized distributors Substitution of non Defibtech approved accessories may cause the device to perform improperly Do not open sealed pads package until pads are to be used Do not touch the patient during defibrillation Defibrillation current can cause operator or bystander injury Do not allow pads to touch metal objects or
72. stem requirements for adequate performance are as follows Pentium 4 Processor 512 MByte System Memory GByte free space on hard disk USB 1 0 connectivity Recycling Information At the end of useful life recycle the defibrillator and its accessories Recycling Assistance For recycling assistance contact your local Defibtech distributor Recycle in accordance with local and national regulations Preparation For Recycling Items should be clean and contaminant free prior to being recycled When recycling used disposable electrodes follow local clinical procedures Packaging For Recycling Packaging should be recycled in accordance with local and national requirements Notice To European Union Customers The crossed out wheeled bin symbol on this device indicates that this equipment has been put on the market after 13 August 2005 and is included in the scope of the directive 2002 96 EEC on Waste Electrical and Electronic Equipment WEEE and of the national decree s which transpose provisions of such directive At the end of its lifetime this device can only be disposed of in compliance with the provisions of the above mentioned European directive and following possible revisions as well as with the corresponding national regulation Severe penalties are possible for unauthorized disposal Electrical and Electronic Equipment EEE may contain polluting components and hazardous substances the accumulation o
73. system controls battery pack condition charge discharge functions and measurement and signal acquisition functions Automatic Unit Self Tests Every time the unit is turned on poweron self tests are performed to test the basic operation of the unit The unit also performs daily weekly monthly and quarterly self tests automatically without any intervention from the operator to check the integrity of the unit s hardware and software The unit will also perform a battery insert self test when the battery pack is inserted Manual Self Tests Manually initiated self tests may be run at any time by the user to test the DDU 2300 AED s systems including the charging and shocking functions the shock is internally dissipated i e no voltage will be present at the pads To run a manually initiated AED test the unit must be put into Maintenance Mode of operation Refer to the AED Maintenance Screen section in Chapter 6 of this manual for detailed information on performing these self tests Note Running a manually initiated self test will consume approximately one shock s worth of energy from the battery Cleaning After each use clean the DDU 2300 AED of any dirt or contaminants on the case and connector socket The following are important guidelines to be adhered to when cleaning the device he battery pack should be installed when cleaning the DDU 2300 Do not immerse the DDU 2300 in fluids or allow fluids to enter the unit
74. t Malntenarice eoi oer EE e Mri umida rias de Exon ee aaa 5 1 SIBI E E A A A A A E T 5 5 SE E E EEE ATENE N A A TAE E E E ten 5 5 E e T EN E A E T E S E E E E E EEE E A E E 5 5 Operator s CheckllStz in sch Ira AE EE A EAE AEEA E RUD e RERO 5 6 TroublesHlioOtlng iis i orte iria Ene aar Eetedaes erit te irr ire d MuR 5 7 ml 5 8 Click any of the section headings above to be automatically re directed to that page See page references noted in green throughout this book for other interactive links iv DAC 2513EN BC 10 11 12 Maintenance Mode cece ccteedccii sen npa nEREnRRrERRRRRRRRRESERRREERORERRERERREERARSRRRRRRNSS 6 1 e m EE 6 1 Navigation in Maintenance Mode ener nennen enne 6 1 Entering Maintenance Mode ssssssssssssseeeeee ene nennen en EEEE nennen nnn 6 2 AED Main Menu SCIGGrn ce trito rre tete noraen ra deans e eie ea a s 6 2 AED Status SOEI visisteceatunsonecesundsansnetionieeeatinoledneaihontaetabiodianantundletcetummontseaatigntesealundlaxes dan 6 2 AED Maintenance Sore secas oc i Rd Pe vua Unos Fac uic pew LR n Un Dea d e 6 3 AED OPtlOnS Sele m En E AE E AEEA AA A 6 5 Rescue Options Screen 2 0 ceeeecccceceeeececcseeeeececseneeeececeeeeeeeeccsesseeescneesseeecesessseesenetsseeesentaaees 6 7 Help TOPICS SCION ssssccscsesssescsseeessncssseeseseassesessasenseensoneesseesnenseeeeesesaaaseassansnseeeas
75. t be applied to the patient s bare chest Locate pads package in back of AED Locate Pads Purpose This helps the user locate the pads in the pad storage area which is located on the back of the unit Plug in pads connector Plug In Pads Purpose The DDU 2300 is unable to detect that the pads are plugged in Check that the connector is fully inserted into the unit If the pads are properly plugged in continue to follow audio and visual instructions Tear open pads package Open Pads Package Purpose This instructs the user to tear open the pads package on the dotted line on the top of the package Once the package is open the user will be able to remove the pads from inside the package 4 6 DAC 2513EN BC Pad Connection Pad Application Related Prompts continued Peel pads from blue liner Peel Pads Purpose This instructs the user to peel each pad from the blue liner before placing the pads on the patient Peel the pads from the blue liner only when the pad is ready to be placed Place the pads with the sticky side of the pad on the patient s bare skin Apply pads to patient s bare chest as shown Apply Pads to Patient Purpose The DDU 2300 AED has determined that the pads are not placed on the patient or not properly applied Place pads on the patient following instructions on the pads package If the prompts continue try replacing the pads with a new set Poor pad contact to patient Poor P
76. t press any softkey button to confirm the test status and return to the AED Maintenance screen If self test passes The unit will speak and display AED OK If self test fails The unit will display an error screen with text prompts providing instructions to address the condition Note f self test fails the user should follow the text prompts to address the condition requiring attention or refer to the Troubleshooting section in Chapter 5 of this manual To exit Press any softkey button The self test status pop up window will clear and the display will return to the AED Maintenance menu screen gt Upgrade AED The Upgrade AED menu selection is a method of performing a unit upgrade and will activate the system upgrade procedure from a Defibtech Data Card DDC card containing an upgrade application Note Upgrades can also be performed directly from the AED Status Screen if an upgrade card is present when the AED Status Screen mode is launched Before you begin Be sure the unit is in Maintenance Mode 6 3 DAC 2513EN BC To enter Navigate to Upgrade AED AED Main Menu gt AED maintenance Upgrade AED What it does f an upgrade data card is present the unit will start performing the upgrade process Follow any prompts and instructions that the upgrade application provides pack or data card until the operation is WARNING Complete Do not turn off the unit or remove the battery Note f the D
77. ted press the pad cable into the groove in the back of the unit to hold it in place and tuck any excess cable behind the pads package A The pads are intended for one time use only and must be discarded after use or if the package has been opened CAUTION 5 3 DAC 2513EN BC Checking Pad And Battery Pack Expiration Dates It is important to check the expiration dates of the pads and battery pack The expiration date of the pads package is printed on the outside of the sealed package The expiration date of the battery pack is printed on the label on the battery pack Once an accessory is past its expiration date it should be taken out of service and replaced as soon as possible Follow the instructions in the nstalling And Removing The Battery Pack and Connecting The Pads sections of Chapter 3 in this manual to replace an expired part with an un expired part Pads should be discarded Battery packs should be appropriately recycled You may also check the status of pads and battery pack when the unit is off by pressing the center softkey button to display the AED Status Screen and enter Maintenance Mode Refer to the AED Status Screen section in Chapter 6 of this manual Checking The DDC Card If One Is Installed Each time the DDU 2300 AED is used an event file is created on the DDC card if installed If the unit was used to treat a patient the DDC card in the unit should be removed and provided to the patient s care prov
78. the CENTER softkey button or call for service Battery pack non functional Replace battery pack Defibrillation pads are not pre connected to unit Connect defibrillation pads to unit Defibrillation pads or battery pack expired Battery pack not inserted Replace expired component Insert battery pack Battery pack is low or needs servicing Replace battery pack or call for service Unit needs servicing Unit needs servicing Call for service Record code number and call for service Battery test failed service code xxxx Battery pack needs servicing Record code number and call for service Service required Unit needs servicing Call for service Replace battery now prompt Battery pack capacity is critically low Unit may not deliver a shock replace battery pack immediately Battery low prompt Battery pack capacity is getting low Replace battery pack as soon as possible Display screen does not work Battery pack depleted Replace battery pack Battery pack not inserted properly Make sure battery pack is oriented correctly and fully inserted Unit needs servicing Call for service Plug in pads connector prompt Pads connector not plugged in Plug in pads connector Pads connector broken Replace pads Unit s connector broken Call for service DAC 2513EN BC 5 7 Troubleshooting continue
79. the package remove the pads from the blue liner apply the pads to the patient and administer care The AED has a storage area in the back of the unit that allows for storage of a single sealed adult pads package Battery Packs The DDU 2300 AED uses a lithium battery pack to provide the AED with a long shelf and standby life The battery pack is inserted into the battery pack opening on the back of the AED and latches into place Battery packs are not rechargeable Data Cards DDU 2300 AED is designed to optionally use Defibtech Data Cards DDC cards The AED will operate with or without a DDC card but if a DDC card is installed additional event storage capacity is available The DDU 2300 AED accepts DDC cards capable of recording an assortment of data for a given period of time The DDU 2300 allows the user to enable or disable recording of audio data Refer to the AED Options Screen section in Chapter 6 of this manual The DDC card is inserted into a slot behind the data card USB port access door found on the side of the AED Refer to the Installing The Defibtech Data Card DDC card section in Chapter 3 of this manual A new event file is created on the DDC card each time the AED is turned on and the following information is recorded e The time the AED was turned on Other data such as ECG data time data audio data audio enabled card only event milestones such as motion detection shock advice shock delivery informati
80. the patient s heart rhythm The operator should not touch the patient This prompt will be spoken at the beginning of the analysis period and also if motion or interference has been detected Analyzing interrupted Analyzing Interrupted Purpose The DDU 2300 AED has determined that accurate ECG analysis is not possible and has ceased analyzing The operator is prompted to resolve the problem Refer to the Motion Interference Prompts and the Pad Connection Pad Application Related Prompts section in this chapter Once the problem is resolved the unit will enter analysis mode again No shock advised No Shock Advised Purpose The DDU 2300 AED has determined that a shock is not required The unit will NOT charge and the SHOCK button will NOT be enabled The user will be prompted to begin CPR Shock advised Shock Advised Purpose The DDU 2300 AED has determined that a shock is recommended and the unit will begin charging in anticipation of delivering a defibrillation shock Shock Related Prompts Charging Charging Purpose The DDU 2300 AED has determined that a shock is recommended and is charging the unit in anticipation of a defibrillation shock Analysis will continue during this phase A tone will sound to indicate charging progress If the unit detects a rhythm change to a non shockable one charging will abort and the user will be prompted to begin CPR Stand clear Stand Clear Purpose The D
81. tient Analyzing and shocking e Performing CPR Replacing the battery Replacing the pads Checking the AED status Replacing the data card Use the TOP and BOTTOM softkey buttons to scroll through the different Help topics items When the desired help selection is highlighted with a box press the CENTER softkey button to get more information To exit Using the TOP or BOTTOM softkey buttons scroll and highlight the selection Go to main menu Press the CENTER softkey button The unit will exit the Help Topics menu screen and return to the AED Main Menu 6 9 DAC 2513EN BC 7 DDU 2300 AED Accessories This chapter describes the component parts and the accessories that can be used with the Defibtech DDU 2300 AED For contact information on obtaining replacement component parts and accessories refer to Chapter 12 in this manual Defibrillation Pads The DDU 2300 AED must be used with Defibtech self adhesive defibrillation pads for adults or with child infant pads for infants and children These pads serve two functions Allow the unit to read the patient s electrocardiograph ECG rhythm Deliver defibrillation energy to the patient if needed The Defibtech self adhesive defibrillation pad assembly comes in a leads out sealed package that allows the device to be stored with the pads connected to the AED When the DDU 2300 AED is used the operator needs only to turn the device on remove the pad package tear open
82. to Settings AED Main Menu gt Rescue options gt Settings What it does The Settings option allows the user to change the currently enabled protocol by entering a special protocol code This code is a special code that encodes all the important information regarding the protocol This code is custom generated by Defibtech If the code is not entered correctly the protocol will not be changed Based on the protocol code entered the currently selected protocol will be changed to that described by the special protocol code You can obtain this code from your medical director or from Defibtech For Defibtech contact information please refer to the Contacts section in Chapter 12 of this manual Once the code has been entered the settings will have been changed To enter the code use the TOP softkey button to scroll through numbers letters Once the correct number letter appears use the CENTER softkey button to move to the next space Once all of the numbers letters have been entered press the CENTER softkey button The settings will have been changed based on the code entered 6 8 DAC 2513EN BC The Help Topics option on the AED Main Menu provides a list of available help topics Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Help topics AED Main Menu gt Help topics What it does The Help topics provides a list of available help topics The Help topics are the following e Preparing the pa
83. use that coaching setting Note The factory default setting of CPR breathing is set to disabled gt Rescue Protcol The Rescue protocol option allows the user to select a rescue protocol Rescue protocol options include the AHA ERC 2010 protocol or Custom Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Rescue protocol AED Main Menu gt Rescue options gt Rescue protocol What it does The rescue protocol option allows the user to select between up to two rescue protocols that have been enabled in the unit The factory default setting of the rescue protocol is AHA ERC 2010 To change the protocol press the CENTER softkey button to highlight the protocol The user will be prompted to enter a password to proceed The password may be obtained from your medical director or from Defibtech For Defibtech contact information please refer to the Contacts section in Chapter 12 of this manual Once the password has been entered the user may select between the two protocols To enter the password use the TOP softkey button to scroll through numbers Once the correct number appears use the CENTER softkey button to move to the next space Once all of the numbers have been entered press the CENTER softkey button The user will now be able to choose a different rescue protocol gt Settings Before you begin Be sure the unit is in Maintenance Mode To enter Navigate
84. utomatically cancel the shock 4 8 DAC 2513EN BC Shock Related Prompts continued SHOCK button not pressed Button Not Pressed Purpose After shock is advised the DDU 2300 AED will prompt user to press the flashing shock button If after 30 seconds the shock button is not pressed the DDU 2300 AED will give this prompt and immediately go to CPR mode No Shock Required Prompts No shock advised No Shock Advised It is safe to touch the patient OK to Touch Patient Purpose The DDU 2300 AED has determined that a shock is not required The unit will not charge and the SHOCK button will not be enabled The user will be prompted to begin CPR CPR Prompts Note CPR breathing coaching prompts can be set through the Rescue options menu option listed on the AED Main Menu screen The factory default setting provides prompts for chest compressions only no breathing Breathing prompts can be included either by changing the menu option refer to the CPR Breathing section in Chapter 6 of this manual or by pressing a softkey button during rescue Refer to Rescue Breathing Options Softkey Icon section of this chapter Begin CPR now Begin CPR Now Purpose This indicates that the user should begin performing CPR immediately The unit will not monitor the patient s ECG rhythm during this CPR period Give Compressions Give xx Compressions Purpose This indicates that the user should beg
85. with the connector and part of the cable exposed The DDU 2300 AED is designed to be stored with the electrode cable already installed This allows the pads to be stored in a pre connected state for rapid deployment during an emergency package until the pads are to be used The packaging should be opened only immediately prior to use otherwise CAUTION ithe pads may dry out and become non functional f DO NOT remove the defibrillation pads from the sealed STEP 1 Inspect The Pads First check to ensure that the pads package has not expired Do not use pads past their expiration date Discard expired pads Next check to ensure that the pads package has not been torn opened or damaged Dispose of the pads if the package is open or damaged Inspect the pads cable and replace pads if any nicks cuts or broken cables are found STEP 2 Connect The Pads To The Unit Insert the connector end of the defibrillation pad cable into the pads connector socket on the top left corner of the DDU 2300 AED as shown Press the pads connector in firmly until itis fully seated in the unit STEP 3 Store The Pads In Back Of Unit The pads package can then be stored in the pad storage slot in the back of the DDU 2300 AED After connecting the pads connector to the unit push the pads package rounded end first with the pictures on the package facing up and out into the pad holder compartment on the back of the AED When the pads package is fully inser