User Manual SHS800 Series Handheld Digital Oscilloscope V1.3
Contents
1. Compare the trigger conditions of pulse It triggers as soon as matching the conditions Condition 4 Negative pulse width larger than pulse width setting 4 Negative pulse width equal to pulse width setting Selecting this option can turn the Set Width 20 0ns 10 0s arrow keys to set up the pulse width Page 1 2 Ente r the second page of the Page menu Type Mode Pulse Auto aat Up Page 2 7 Figure 2 41 Pulse Trigger Function Menu 2 Table 2 23 Pulse Trigger Function Menu 2 Select the pulse to trigger the pulse match the trigger condition Auto Select the type of triggering Normal mode is best Normal for most Pulse Width trigger applications nm Enter the Trigger setup menu Next Page Page 2 2 Return to the first page of the menu SHS800 Series 27 Pulse Trigger Application Example Operations steps Input a pulse signal Press Trigger to enter trigger menu Press F1 to choose pulse trigger Press F3 to set pulse trigger conditions Press F4 and use arrow keys to set pulse width Move trigger level line with the up and down arrow keys It will trigger as soon as it meets the trigger conditions DELIS m I Notes We can t adjust trigger level line under the SetWidth menu as we use the up and down arrow keys to set both the pulse width and trigger level line If you want to adjust the trigger level please make sure that you have exited the SetW
2. Una 1X 5X 10X 50X Set to match the type of probe you are 100X 500X 1000X using to ensure correct vertical readouts Enter the second page of CH1 CH2 Next Page Page1 2 mani Invert ext Page Off Filter Tozera Page 772 Off Fiter Tozero Page22 Figure 2 3 Channel Menu 2 Table 2 3 CH1 CH2 Function Menu 2 On Off Turn on off invert function Filter Enter the FILTER menu To Zero ei Set waveform vertical position and trigger level to zero Next Page Page 2 2 Return to the first page of CH1 CH2 menu Fitter Type UppLitrit aa E Figure 2 4 Digital Filter Function Menu SHS800 Senes 7 Table 2 4 Digital Filter Function Menu Option Digital Filter Type Up_Limit Introduction Turn on the digital filter Turn off the digital filter Setup as LPF Low Pass Filter Setup as HPF High Pass Filter Setup as BPF Band Pass Filter Setup as BRF Band Reject Filter Use the up and down arrow keys to set Upp_Limit Low Limit Use the up and down arrow keys to set Low Limit Return Digital Filter Application Example Operation steps n ul Return to the CH1 or CH2 menu 1 Inputa signal to CH1 and press AUTO E AN N p Press CH1 to enter CH1 menu Press F5 to enter the second page of the CH1 menu Press F3 to enter the digital Filter function Press F2 to choose a filter type For example input a signal with BW 20M and choose Upp L
3. 10s Support Signal Formats PAL SECAM NTSC Video Trigger Trigger Conditions odd field even field all lines pointed line gt lt Positive slope gt lt Negative slope Time 20ns 10s CH1 trigger type Edge Pulse Video Slope Slope Trigger Alternative Trigger CH2 trigger type Edge Pulse Video Slope X Y Mode X pole Input Y pole Channel 1 CH1 Channel 2 CH2 Input Sampling Frequency 25KSa s 250Msa s 1 2 5 5 order SHS800 Series 67 Measure System Vpp Vmax Vmin Vamp Vtop Vbase Vavg Mean Crms Vrms ROVShoot FOVShoot Auto Measure 32 types RPREShoot FPREShoot Rise Fall Freq Prd Wid Wid Dut Dut BWid Phas FRR FRF FFR FFF LRR LRF LFR LFF Cursor Measure Manual mode Track mode and Auto mode Control Panel Function Auto Set Auto adjusts the vertical horizontal system and trigger position Support 2 groups of referenced Waveforms 20 groups of setups 10 groups of captured waveforms internal storage recall functions and USB flash driver storage function Save Recall Hard Ware Frequency Counter Reading resolution 1Hz Range DC coupling 10Hz to maximal bandwidth Signs Ee to all trigger signals Except pulse width and video trigger Digital Multimeter Specifications Environment Temperature 23 5 C Relative Humidity 75 Max Display 6000 DC voltage AC voltage resistance diode cont
4. Press F1 to choose Replay mode Press F5 to enter the second page of the menu Set replay Start Frame Curr_frame End Frame and return to the first page of the menu Set Replay Mode Interval and press F2 to perform waveform replay Press F1 to choose Off mode to exit waveform record EECORD Mode Source Interval EmdFrame Operate Fecord CHI 1 Posi ps G 1 AGGAAkH RECORD ode Operate Play Mode Interval Nest Page Figure 2 69 Waveform Play Back 42 SHS800 Series Chapter 3 Using the Multimeter About this Chapter This chapter provides a step by step introduction to the multimeter functions of SHS800 series Handheld Digital Oscilloscope The introduction gives basic examples to show how to use the menus and perform basic operations The digital multimeter provides the following functions making DC voltage AC voltage resistance diode continuity capacitance DC current and AC current measurements Notes 1 You should use the multimeter with correct connections as instructions 2 Thekey KRub Stop can hold the screen p DX X coe NS operation type 388 8rriv O I UI oO 200 2mV 100 200 300 40 500 60m Meter Relative BI Ori DEEE v Trend Plat Figure 3 1 Multimeter User Interface the range of the multimeter the indicate of input connection SHS800 Series 43 Making DC and AC Voltage Measurement Table 3 1 DC and AC Function Menu Op
5. the multimeter will not alarm and read value Making Capacitance Measurement i51 o o 122 64F O 1000200 s00 AQ uF Meter Relatve Cap GOff Trend Plot Figure 3 7 Capacitance Measurement Operation steps 1 Press Meter to enter multimeter mode press F1 to choose Cap measurement 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the measured object 3 Turn on off the relative according to the real demand 4 Read measurement value SHS800 Series 47 Making DC and AC Current Measurement 9 o 6 o 244 0mA Q CO D 1000200 s00 400 500 Bn Meter Relatve DCI Crt Auto mA Trend Plot Figure 3 8 DC Current mA Measurement Fed o o O 2 024A 0 2 Meter Relathe DCI Crt Manual A Trend Plot Figure 3 9 DC Current A Measurement Meter Relative AC Crt Auto mA Trend Plot Figure 3 10 AC Current mA Measurement 48 SHS800 Series I9 o o O 1 479A Meter Relative ACI Ort A Trend Plat Figure 3 11 AC Current A Measurement Operation steps 1 Press Meter to enter multimeter mode press F1 to choose DCI ACI measure 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the power or load to be measured 3 Turn on off the relative according to the real demand Choose Manual or Auto a
6. waveform operation Saving waveform to USB Flash Drive Type Swe a Figure 2 52 Saving Waveform to USB Flash Drive Menu Table 2 32 Saving Waveform to USB Flash Drive Function Menu Option Setup Introduction Sae Accomplish tne storage _ Saving Picture Waveform interface image can be saved to USB flash drive but they can t be recalled You can view them on correlative computer software Type Print Key IRE Save Picture Figure 2 53 Saving Picture Menu 34 SHS800 Series Table 2 33 Saving Picture Function Menu Setting Introduction Picture Menu for the Storage Recall waveform interface image Choose Print Picture option and press Type Print Picture Save Recall for 4 seconds to print the picture while the SHS800 connects to the printer Print Key Choose Save Picture option and press Save Recall for 4 seconds to save the picture while you insert an USB flash driver to the SHS800 Go to the Save Recall interface Save Picture Saving CSV Type Cata Depth Fara Save Figure 2 54 Saving CSV Menu Table 2 34 Saving CSV Function Menu option Setting Introduction Maximum Set to store maximum waveform data to CSV file Save Co to tne Save Recallinterface Recall Factory Setups Press Recall you can recall factory setups Table 2 35 Factory Setups Function Menu To view the Factory setup Recall the Factory setup Save Recall Waveform to USB F
7. 1 47T 1 71 SHz Curb 4 ammz IELITTE HORIZON Delayed MembDepth Figure 2 23 Long Memory Reference waveform Press Scope and choose Ref to enter the reference waveform function menu Source Save Figure 2 24 Reference waveform Menu Table 2 11 REF Waveform Function Menu Function Setting Instruction Signal CH1 CH2 Choose the waveform to be saved Ref A Ref B Choose to save or recall the reference position of the waveform Save the source waveform to the pointed Save reference position Ref A Ref B On Display the reference waveform on the screen Off Clean the reference waveform on the screen Reference Waveform Application Example Operation steps Input a waveform to CH1 or CH2 Adjust time base to display the best waveform Press Scope and choose Ref to enter horizontal system Choose the reference waveform need to save and press F4 to save Press F5 to show the reference waveform Press F5 to exit the reference waveform p png p Io SHS800 Series 17 00 01 14 210154 III CHiziBBJ sO SS CHL GI REF V Source REF A Figure 2 25 Reference Waveform Cursor and Measure System Cursor measure Press Cursor Measure once to enter the cursor measure system There are three modes of measure manual track automatic Manual mode horizontal or vertical cursors appear in couple and we use them to measure voltage or time parameters The space between two curs
8. plot measurements over time 58 SHS800 Series KF Time Ge SOL 02mv Lo ist 36 3 8mU a002 80 TrendPlot SaRate Met Page Figure 4 15 Multimeter Trend Plot Recording Curve 3 Press F4 to stop or run recording data 4 Atthe second page of the menu you can choose manual or auto mode to record the data Display the record data 5 Press F3 to choose data display mode Normal mode the screen displays the data up to the minute The recorded data before saved in the memorizer Full view mode the screen display all recorded data in the memorizer 6 Zoom function under full screen mode press time base to zoom in or zoom out 7 Data analysis move cursor analyzing data over time DO EF AMV ao Time 01 24 135 aL uuu 121 3mU ail asa Trend Plot ca Rate Ml Fage gut Figure 4 16 Analysis Trend Plot Data 8 Save waveform save the recorded waveform to exterior storage device to make more detailed analysis 9 Press Return to exit trend plot SHS800 Series 59 Chapter 5 Prompting and Troubleshooting About this Chapter This chapter gives a detailed instruction of every system prompting appears on the screen as well as some basic troubleshooting System Prompting Messages Instruction Trig level at limit Mention you that the trigger level is at a limit when you turn the Trig level knob Horizon position at limit Mention you that the horizont
9. the oscilloscope does not display a waveform until after the first trigger When you want the oscilloscope to acquire a single waveform press the SINGLE button Setup Enter the Trigger setup menu Next Page Page 2 2 Return the first page of Video Trigger menu SHS800 Series 29 Video Trigger Application Example Operation steps 1 Inputa video signal Press Trigger to enter trigger menu Press F1 to choose Video Press F5 to enter the second page of video trigger menu Press F2 to set the video standard PAL SECAM or NTSC meeting with the input signal Press F5 to return to the first page of video trigger menu 7 press F4 to choose the type of Sync If you choose Line Number use the up and down arrow keys to set the number of line 8 Move trigger level line with the up and down arrow keys to set the trigger position oe m o Notes We can t adjust trigger level line under the Line Number menu as we use the up and down arrow keys to set both the pulse width and trigger level line If you want to adjust the trigger lever please make sure that you have exited the Line Number menu MM Pos 0 60 pe ENEA Chin aad ELETC EMEND Ee EDIT TRIGGER Type standard Mode Next Page video Auto Set Up Page 2 2 Figure 2 45 Video Trigger Slope trigger Type Source When iuc Next Page Slope REM Figure 2 46 Slope Trigger 1 30 SHS800 Series Table 2 26
10. 3 X C reading offset 1 of offsetl All Gain settingss200mv div 0 2div 2mv DC Measurement Accuracy 3 X reading offset 1 of offsetl All Gain Settings gt 200mv div 0 2div 100mv lt 1 7ns SHS820 lt 2 3ns SHS815 lt 3 5ns SHS810 lt 5 8ns SHS806 Window Modes Hanning Hamming FFT Blackman Rectangular Bandwidth Rise Time BNC value Sampling points 1024 Bandwidth Limit 20MHz 3dB 66 SHS800 Series Horizontal System Real Time Sampling Single channel below 50ns div 1GSa s double Rate channel 500MSa s Equivalent Sampling lt 50GSa S Measure Display Modes MAIN WINDOW WINDOW ZOOM SCAN X Y Time Base Accuracy 50ppm measured over 1ms interval 2 5nS div 50S div SHS810 SHS815 SHS820 Horizontal Scan Range 5nS div 50S div SHS806 Scan 100mS div 50S div 1 2 5 5 order Rate Trigger System Trigger Types Edge Pulse Width Video Slope Alternative Trigger Source CH1 CH2 Trigger Modes Auto normal Single Trigger Coupling AC DC LF reject HF reject Trigger Level CH1 CH2 6 divisions from the center of the screen Range Trigger Pre trigger Memory depth 2 sampling Displacement Delay Trigger 268 04 div Holdoff Range 100ns 1 5s Edge Trigger Edge type Rising Falling Rising and Falling Trigger Modes gt lt Positive Pulse Width gt lt Pulse Width Trigger Negative Pulse Width Pulse Width Range 20ns
11. 4 Press F5 to return pa 2 04 Figure 2 35 All Measure Function Menu 32 types of parameters measure application example Operation steps 1 Input two sine signals respectively to CH1 and CH2 with the same frequency amplitude and different phases 2 Adjust time base and vertical scale to obtain the best waveform Press Cursor Measure and choose All 4 Turn on Voltage Time Delay and will display 32 types of parameter measure p Figure 2 36 32 Types of Parameter Measure 24 SHS800 Series Trigger System There are 5 kinds of trigger function edge pulse video slope alternative Press Trigger to enter the trigger system Edge Trigger Type cnilrce Slope Mode Edge Es Auto Set Up Figure 2 37 Edge Trigger Function Menu Table 2 20 Edge Trigger Function Menu Trigger on the rising or falling edge of the input Type Edge sera CH1 CH2 Set CH1 or CH2 as a trigger source Trigger on rising edge of the trigger signal Trigger on falling edge of the trigger signal Slope Trigger on rising and falling edge of the trigger E signal Use this mode to let the acquisition free run in the AUS absence of a valid trigger This mode allows an un triggered scanning waveform at 100 ms div or slower time base settings Mode Use this mode when you want to see the valid Normal trigger waveforms only the scope will not acquire waveform until satisfied trigger Sinale The setup detects a trigger
12. Press Meter to enter multimeter mode press F1 to choose Res measurement 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the power or load to be measured 3 Turn on off the relative according to the real demand 4 Choose Manual or Auto according to the real demand 5 Read resistance value Notes When measuring resistant please make sure that the circuit is power off and the capacitance is discharged to avoid damage to the SHS800 SHS800 Series 45 Making Diode Measurement tod O O O 0 554V london O 0 4 0 5 teter Diode Figure 3 5 Diode Measurement Operation steps 1 Press Meter to enter multimeter mode press F1 to choose Diode measurement 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the diode to be measured 3 Read the value Making Continuity Measurement 200 300 400 S00 BO Meter Continuity Figure 3 6 Continuity Measurement 46 SHS800 Series Operation steps 1 Press Meter to enter multimeter mode press F1 to choose Continuity measurement 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the object to be measured 3 When the measured object is under 502 the multimeter will alarm and read value 4 When the measured object is above 509
13. Save Mode Star Figure 4 8 Waveform Recorder Saving Mode Menu Table 4 4 Waveform Recorder Saving Mode Function Menu Option Instruction Begin to record waveform record waveform under 100mS and above scan time base Replay the recorded waveform Copy the waveform saved in the memory to the USB flash disk Start Choose a place to save record including memory USB flash Save mode disk USB flash driver saves only under 2 5s div and above time base Exit the submenu and return to the waveform recorder main menu Figure 4 9 Waveform Recorder Replaying Mode Menu return 54 SHS800 Series Table 4 5 Waveform Recorder Replaying Mode Function Menu Pause or contnue playing waveform automatically you can Stop Continue change the time base to observe the waveform in the memory Replay the waveform Previous Back the waveform and then play Next Speed the playing of the waveform Return Exit the replaying menu Viewer Record Replay Spi Figure 4 10 Waveform Recorder Setting Menu Table 4 6 Waveform Recorder Setting Menu Record and replay channel waveform with full screen Viewer ra mode Record and replay channel waveform with divided Split screen The up half of the screen displays CH1 while the down half displays CH2 Full screen Record circularly when the waveform recorder is continuous Record full the after data will cover the pre data mode ER Stop recording data when
14. Series 73 Appendix C Battery Installation The battery of the SHS800 is separated from it s host please install the battery according to following steps 1 Dismantle the two screws of the battery cap by using screw knife as figure 1 shows 2 Draw back the packing block of the Handheld Digital Oscilloscope and then dismantle the battery cap as figure 2 shows 3 Put battery into the battery bin stably as figure 3 shows 4 Cover the battery cap and tightly twist the two screws which are dismantled from the battery cap forward such as figure 3 shows 5 Successful installation of the battery depends on whether the Handheld Digital Oscilloscope could be normally turned on or not Notes The battery plug is designed to prevent from connecting revesly it preferes operating slightly to powerfully Make a caution at the direction of the battery cap logo of the battery is attached to the external end If The Handheld Digital Oscilloscope can not be turned on normally after correctly installation it is likely that the battery quantity of electricity has been used off please refresh for the battery right away Figure 3 74 SHS800 Series Appendix D Daily Maintaining and Cleaning Daily Maintaining Do not let the LCD exposed in the sun directly for a long period when storing or placing the SHS800 CAUTION To avoid damage to the instrument or probes do not expose them to sprays liquids or solvents Cleaning C
15. Slope Trigger Function Menu 1 Option Setting Instruction Trigger on positive slope of negative slope Type Slope 99 p p g p according to setup time of the oscilloscope Source CH1 CH2 Select trigger source ce Condition l Select trigger conditions She Time gt Use the arrow keys to set slope time Time Set time setup range is 20ns 10s Enter the second page of the slope trigger Next Page Page 1 2 dex ped menu 1 Pos Alps DE ood RS du ea CH1 Sam M 160 ps CH1 5 Bea sms TRIGGER Type erties Mode Mert Page Slope v Auto Set Up Page 2 Figure 2 47 Slope Trigger 2 cee Table 2 27 Slope Trigger Function Menu 2 Setting Instruction Trigger on positive slope of negative Slope slope according to setup time of the oscilloscope Select the trigger level that can be adjusted by LEVEL knob You can adjust LEVEL A LEVEL B or adjust them at the same time Vertical Use this mode to let the acquisition free run in the absence of a valid trigger This mode allows an un triggered scanning waveform at 100 ms div or slower time base settings Normal Use this mode when you want to see only SHS800 Series display a waveform until after the first trigger When you want the oscilloscope to acquire a single waveform press the SINGLE button Setup Enter the Trigger setup menu Next Page Page 2 2 Return to the first page of slope
16. and acquires i es then stop Setting Enter the Trigger Setup Menu Coupling Holdoff Holdoff Dc Reset Figure 2 38 32 Trigger Setting Function Menu SHS800 Series 25 Table 2 21 Trigger Setting Menu DC Passes all components of the signal Blocks DC components and attenuates signals below 170Hz Couplin Attenuates the high frequency components iilis HF Reject be au aZ above 140kHz Blocks the DC component and attenuates the LF Reject low frequency components below 7 kHz Using the arrow keys to adjust hold off time sec the hold off value is displayed AC Hold off n Cu Hold off Reset hold off time to 100ns Reset Return the first page of the menu i i i MPoz 0 G0y Sees eee MG TETERE G 1 580001 MHz CHi LEGY MM Sean EEEE TRIGGER Type Source Slope Node Edge Es Auto Set Up Figure 2 39 Signal Rise Trigger Menu Pulse Trigger Type Source When SANA Next Page Pulse prium Page 12 Sed us Figure 2 40 Pulse Trigger Function Menu 1 26 SHS800 Series Table 2 22 Pulse Trigger Function Menu 1 Select the pulse to trigger the pulse match the trigger condition CH1 CH2 Select input signal source la Positive pulse width less than pulse width set 24 Positive pulse width rger than pulse width setting Positive pulse width equal to pulse width setting Negative pulse width less than pulse width setting I 76M
17. cip colei ate scat ec AI 25 Save and Recall Syste Msiri iaia a oi 33 UUV SVSTOTL o e E E TE E E Et 37 Chapter 3 Using the Multimecter eee Lee eee ee eese eere eene sesresece nice eeee seo seseseeseionineone 43 Making DC and AC Voltage Measurement 44 Making Resistance MeaSUreMENt i 45 Makmng biode Measurementa ale 46 Making Continuity MeasuremMEnt ii 46 Making Capacitance MeaSUureMENt i 47 Making DC and AC Current Measurement iii 48 Chapter 4 Using the Recorder Functions eee eee ee eere ence eee eee eee tees esee e tntea seas 50 Osellloscope Irend Plot T d 51 Wavell On ROCO O ame cee Rte ratito e eta nn eum Ap E D UE A rat ELE 54 Multimeter rend Flot uote E e itu editt eite ien ates ohana bett bonita UR Ed 57 Chapter 5 Prompting and Troubleshooting mrrrrrrrrrrererereszesesseosicsecenecesccesioseceseeneene 60 System Prompting Messages Instruction i 60 iBroSie aieo nro tee e cT 62 SHS800 Series VII Chapter 6 Service and SUpport e seesseseessoccessoesossossocscesscesessoesosscessoccessossoesoessosoeesoesessoesoesoeesorceesse 64 VEDE ES IE E 64 CODntae b SIGB EIN T assiali 64 Appendix A 9pecilicationiS ioa te uero Sena Eco QN her re ETe ENERO NV NEIN RE EN RS EIN SEU QU
18. color TFT LCD It provides maximal bandwidth 200MHZz real time sampling rate 1GSa s memory depth 2Mpts The multimeter display resolution is 6000 points and can measure voltage current resistance capacitance diode continuity Support scope measure parameters trend plot multimeter measure parameter trend plot and scope waveform recorder 3 types of trigger mode auto normal and single 5 types of trigger type edge pulse video slope and alternative 32 types of auto measurement function and 3 types of cursor measure mode 5 kinds of digital filter mode FFT Unique digital filter function and waveform recording function 2 groups of reference waveform 20 groups of common waveform 10 groups of setting inside save recall Support waveform setting CSV and bitmap file save and recall with USB flash driver Standard configuration interface USB Device USB Host Support software update with USB flash driver PC remote control and PicBridge print For its build in Li battery and small volume its convenient to carry and work outside SHS800 Series V Accessories of SHS800 A user manual A product guaranty card A certification Two 1 1 10 1 probes An USB cable A adapter Meter pens for multimeter A Probes calibrated device 9 A CD including EasyScope3 0 computer software system Optional Probe 100MHz high voltage safety probe CAT Il 1000V CAT III 600V VI SHS800 S
19. inches 145mm diagonal Liquid Crystal Display Resolution 320 horizontal pixels 234 vertical pixels Display Color 24 bit Display Contrast typical Backlight Intensity typical Waveform Display Range 8 x 12 div Off 1 min 2 min 5 min 10 min 15 min 30 min 1 hour 2 hour 5 hour Skin Classical Modern Traditional Succinct waveform Interpolation Sin x X Display Mode Screen Saver Color model Normal Invert Simplified Chinese Traditional Chinese English Language Arabic French German Russian Spanish Portuguese Japanese Korean Italian Power Input Voltage 100 240 V 50 60Hz Output Voltage Battery Supply Power 5000mAh 7 4VDC persisting 5 hours Charging time About 4 hours Adapter Supply Power Environments Work 0 40 C Temperature i Storage 20 C 70 C Humidity 85 RH 40C Height 3000m Length 259 5mm Dimension Width 163 2mm Height 53 3mm Weight 1 5 kg SHS800 Series 71 Appendix B Default Setup Menu or EJ Options Knobs or Buttons Default setup system CH1 CH2 FFT FFTZoom Lr NN zz s m ie Gird 0 SAVEIRECALL 72 SHS800 Series Setup No 1 Edge Pulse Video Mode Auto Type Slope s UN Slope ondition 1 00ms Auto Alternative CH1 Edge DC Slope Rise Time Mode Type TRIGGER Source Alternative Mode Coupling d SHS800
20. interval of recorder waveform End Frame e m the max value of recorder frame record Start to record Operate m stop Stop recording Made Operate PlayMode Interval pest Page Play Back e Page 1 2 Figure 2 66 Waveform Play Back Menu 1 Table 2 43 Waveform Play Back Function Menu 1 Play Back Set the Play Back function menu P Rum Press to start playback playing Operate a Press to stop playing Set circular play mode Play Mode sa Set single time play mode Interval Set interval between frames Next Page SPA 1 2 Enter the second page of the menu Start Frame Curr frame End Frame Next Pace Ee m SEE neun Page 22 Figure 2 67 Waveform Play Back Menu 2 Table 2 44 Waveform Play Back Function Menu 2 Start Frame Set start frame Select current frame to be played Set end frame Press to return the waveform recorder main Return menu Return to the first page of Play Back function menu Next Page Page 2 2 SHS800 Series 41 Waveform Record Application Example Operation steps j 2 3 4 5 6 Input a waveform to be recorded Press User to enter utility system Press F5 to enter the third page of the menu and enter waveform recording menu Press F1 to choose Record mode Use up and down arrow keys to set Interval and End Frame Press F5 to perform waveform recording Waveform Play Back Application Example Operation steps 1 2 3
21. scale O 0 Td o ow TrendPlot ParamaA Param B Mest Page CHiVpp CHI Vmax Figure 4 2 Scope Trend plot Function Menu 1 Table 4 1 Scope Trend plot Function Menu 1 Option Setting Instruction Trend Plot Restart Quilt the current data and start to record afresh Darsena Choose the parameter Voltage time and delay to be measured measurement Run Stop DE Stop or continue recording data Next Page Page 1 2 Enter the second page of the menu SHS800 Series 51 Manual ext Page Figure 4 3 Scope Trend plot Function Menu 2 Table 4 2 Scope Trend plot Function Menu 2 Normal Display the data up to the minute Display Mode eee Display all date in a compressing proportion Memory data transfer to exterior storage device Return to the oscilloscope interface Next Page Page 2 2 Return to the first page of the menu Record data automatically Manual dn Record data manually A Record presses a record First choose a measurement in scope or meter mode You can choose the recorder functions from the waveform recorder main menu To open the main menu do the following Press Recorder to open the recorder main menu Scope Scope Meter TrendPlot Recorder TrendFlot Figure 4 4 Recorder Function Main Menu Scope Trend Plot Application Example Operation steps Open trend plot function 1 Input a signal to CH1 or CH2 Press Recorder to enter the recorder main menu Press F1 to choose
22. the DISSE settings NAER on your paper PictBridge compatible printer SHS800 Series 39 Choose Print Picture option and press Save Recall for 4 seconds to print the picture while the SHS800 connects to Print Picture the printer Print Ke Lila Choose Save Picture option and press Save Recall for 4 seconds to save the Save Picture l picture while you insert an USB flash driver to the SHS800 Next Page Page 1 2 Enter the second page of the menu ImageSize PaperType PrintQuality IDPrint Next Page Default IET IET IET Page 2 7 Figure 2 64 Print Setting Menu 2 Table 2 41 Print Setting Function Menu 2 Option Setting Introduction l Set the paper size available to the Image Size Set the type of the paper Photo FastPhoto Print Quality Default Normal Draft Fine DPm pefaunonof Return to the first page of the menu Waveform Record Waveform record record waveform with a designated interval until reaching the end frame When recording CH1 or CH2 waveform users can set the interval between frames The recorder can record 2500 frames waveform iode Source Interval EndFrame Operate Record CHI AIME 1 Amz BE Figure 2 65 Waveform Record Menu 40 SHS800 Series Table 2 42 Waveform Record Menu Option Setting Instruction Record Set recorder function menu Mode Replay Set replay function menu Off Turn off waveform record menu Se
23. the waveform recorder 9 memory is full When replaying the screen waveform updates By Point every dot from left to right Replay mode When replaying the screen waveform updates the By frame whole screen according to the time of sampling every frame data retum Exit the recorder setup interface Waveform recorder application example Startup the waveform recorder function E d E S Under 100mS or above time base Press Recorder to open the main menu Press F2 to choose scope recorder Press F3 to set the waveform recorder Such as Press F5 to return to the waveform recorder main menu Press F1 to enter record interface Press F4 to set storage mode Interior and USB flash disk storage mode SHS800 Series 55 8 Press F1 to start recording data The waveform will not move right and the recorded data saved to memory The recorded time will be different according to the time base You can pause or stop at any time CH Laa TETTE TETTE RECORD Save Mode Figure 4 11Waveform Recorder Interface Waveform replay 9 Press F2 to replay waveform You can replay the recorded waveform for several times and you can Advance or back off at any time 10 Press F5 to exit the waveform recorder 56 SHS800 Series Multimeter Trend Plot l 2 3 Time G86 25 1 11 8v mun 5 10 ABD I a e E 9 iui i n TrendPlot SafRate ext Page mper AECE Figure 4 12 Multim
24. trigger Alternative trigger The trigger signal comes from two vertical channels when you use alternative trigger In this mode you can observe two irrelative signals at the same time You can select different trigger types for two vertical signals and selected types cover edge pulse video and slope trigger Trigger information of two channel signals display on the bottom right of the screen Type SOWIE Mode Slope Alternative Edge Es Set Up Figure 2 48 Alternative Trigger Menu Table 2 28 Alternative Trigger Function Menu Option The trigger signal comes from two vertical channels Type Alternative when you use alternative trigger In this mode you can observe two irrelative signals at the same time CHY Set trigger type information for CHY signal Set trigger type of the vertical channel signal ES Triggering on rising edge Slope Triggering on falling edge TL Triggering on rising edge and falling edge Setup Enter the Trigger setup menu 32 SHS800 Series Save and Recall System SHS800 can save 2 groups of reference waveforms 20 groups of setups and 10 groups of waveforms in its internal memory There is an USB Host interface in the front panel of the SHS800 and you can save setup data waveform data waveform interface image CSV file to an USB flash drive The postfix of setup data is SET while waveform data is DAV The waveform data can be recalled to the current SHS800 or SHS800 wi
25. values of the signals in every interval to display the waveform Average The scope acquires several waveforms averages them and displays the final waveform The more average times the smoother of the waveform Equivalent Time Sampling This mode is good for observing repetitive period waveforms The sampling rate is up to 50GSa s Real Time Sampling The scope has the highest real time sampling rate up to 1GSa s Interpolation Sinx x Application Example Operation steps 1 Press Scope and choose Acquire to enter acquiring system 2 Press F4 to choose Ream Time 3 Press F3 to choose Sinx x LL Js MR Xr CHiz 166m BLETE Acquisition Siri Mode Sa Rate Sampling 1000658 Figure 2 9 Sinx Interpolation 10 SHS800 Series M Poe G00 ps 1 84 661 MHz e CH 188m M 2 58ns CHI 76 66 ACQUIRE PECE Sind Mode Sa Rate Sampling RealTime ness Figure 2 10 X Interpolation Display System Press Scope and choose Display to enter display system See Figure 2 6 Type Persist Intensity SIRES Nest Page ZUM Page 12 Figure 2 11 Display Menu 1 Table 2 6 Display System Function Menu Option Setting Introduction Vectors fill the space between adjacent Vectors e l Type sample points in the display Dots display sample points directly Off 1 sec Persist 2 sec 5 sec Infinite Intensity lt Intensity gt Set waveform intensity Brightness lt Brightness gt Set grid brightness
26. 2 34 All Measure Function Menu Table 2 18 All Measure Function Menu CH1 Source Select input signal channel i j Voltag 3 On Off Turn on off the all voltage parameters measure function Time On Off Turn on off the all time parameters measure function Turn on off the all delay parameters measurement Dela y function Return Return to the All Measure main menu Table 2 19 All Measure Function Menu Measure Type Introduction TJ Vmax The positive peak voltage The negative peak voltage The absolute difference between positive peak voltage and negative peak voltage 22 SHS800 Series The arithmetic mean over the entire waveform Virtual value the true Root Mean Square voltage of the first cycle in the waveform The true Root Mean Square voltage over the entire waveform Defined as Vmax Vhig Vamp after the waveform rising Defined as Vmin Vlow Vamp after the waveform falling Defined as Vmin Vlow Vamp before waveform RPREshoot rising Defined as Vmax Vhig Vamp before waveform falling l l The time between the first voltage level rising from 10 to Rise Time 90 l The time between the first voltage level falling from 90 Fall Time to 10 o The duration of a burst over the entire waveform l The time between the first rising edge and the next falling Wid edge of 50 voltage level n The time between the first falling edge and the next risi
27. 2min 5min 10min Screen saver 15min 30min Set the time of screen saver 1hour 2hour 5hour Off Date Time Set the date and time of the SHS800 Next Page Page4 4 Return to the first page of the menu 38 SHS800 Series Self Calibration Self Calibration is operated to calibrate the relative data of SHS800 to decrease the mistake during the measure If the operating temperature changes by or more than 5 C or the instrument runs more than thirty minutes you should do the self calibration When you do the self calibration you should cut off all the probes and leads Then press User button to choose Do self cal to show the self calibration menu and do self calibration program according to the prompts on the screen Disconnect Everything from All Inputs Going CH2 Cal Figeure 2 62 Self Calibration Print Setting The SHS800 supports PictBridge compatible printers You can connect the side USB Device of the SHS800 to the USB Device of the PictBridge compatible printer through USB cable After setting the print settings press Save Recall button for 4 seconds to complete the print operation InkSaver Layout PaperSize _Frintey est Page Figure 2 63 Print Setting Menu 1 Table 2 40 Print Setup Function Menu 1 Option Introduction On Print the screen image on a white Ink Saver Off background when you select On Portrait Landscape The output direction of the printer Paper Size Set the type of
28. Automatic Settings When measuring unknown signals and having no idea about its voltage range frequency trigger and other information you can use the automatic setting function Automatic Setting Application Example Operating steps 1 Inputa signal to CH1 or CH2 and then press Auto 2 The SHS800 adjusts its settings automatically to display the best peak peak average period frequency and other information according to the characteristics of the signals 3 Adjust time base and voltage range manually to obtain waveforms needed if necessary 11 07 15 171547 CHI Upp Oz o l 5 12777 e amp ean hanmW Re Prd 1 bem Freg 1 kHz G 1 GGGGGkH 4 CHie LAAI M 5 Bps CHI FOAM M Po 5 uz Figure 2 1 Automatic Setting CH1 CH2 Channel Functions Press CH1 CH2 to enter CH1 CH2 channel menu Coupling EvwLimit Y oltsiDn Probe ext Page Figure 2 2 Channel Menu 1 6 SHS800 Series Table 2 2 CH1 CH2 function Menu 1 Option Setting Instruction DC DC passes both AC and DC components of the input signals Goubino AC blocks the DC component of the AC input signals and attenuates signals below 10 Hz Limit the bandwidth above 20M to reduce display noise filter the signals to BW Limit l reduce noise and other unwanted high frequency components Change the range of voltage by 1 2 5 Coarse sequence V div Fine Fine changes the resolution by small steps under the coarse settings
29. Enter the second page of DISPLAY Next Page Page 1 2 men Menu Display Sets the length of time each displayed sample point remains displayed Met Page n3 05 Figure 2 12 Display Menu 2 Format Screen Grid SHS800 Senes 11 Table 2 7 Display system function menu 2 Option Setting Introduction YT format displays the vertical voltage in relation to time horizontal scale Format XY format displays a dot each time a XY sample is acquired on channel 1 and channel 2 Set to normal mode Screen Set to invert color display mode Display grids and axes on the screen Grid Turn off the grids L Turn off the grids and axes Menu 2sec 5sec 10sec Display 20sec Infinite Next Page Page 2 2 Return to the first page of DISPLAY menu Set display time of menu on the screen XY Waveform Application Example Observe XY waveform operation steps 1 Input 2 sine signals to the channels with the same frequency range and phase 90 degree difference press the button Auto Press Scope and choose Display to enter the display system Press Next Page to enter the second page of DISPLAY menu Press F1 to choose XY mode Adjust the vertical range of CH 1 and CH 2 to obtain the best XY waveform pr oN Siglent MMI 1 m 11 87 13 11 139142 CHI LOG CH2 1 881 Sa 1258MHz sY mode Figure 2 13 XY Waveform 12 SHS800 Series Math Waveform Press Scope and choose Math to
30. Horizontal position 14 Channel s coupling states 9 Frequency Counter 15 2 is a symbol of channel 2 10 Shows the trigger level 16 1 is a symbol of channel 1 11 Shows the time base 17 T shows the trigger level SHS800 Series 3 Function Check and Probe Compensation Function Check Let s make a quick function check to make sure whether the SHS800 works normally Please do the following steps 1 Power the SHS800 The SHS800 performs all the self check items and makes sure that it passes the self check 2 Connect the probe to the CH1 of the SHS800 Align the slot of the probe connector with the salient on the CH1 BNC push down and twist right to lock the probe Connect the probe tip and reference lead to the Probe Comp connectors 3 Press Auto you will see a square wave with 1 KHz frequency and about 3V peak peak in a few seconds 4 Press CH1 twice to cancel channel 1 then press CH2 to display channel 2 and repeat step 2 and 3 Probe Compensation When you connect the probe to an arbitrary channel for the first time please make the following adjustment to make the probe match with the channels Probe without compensation or compensation warp may lead to imprecise or false measurements You can perform the adjustments manually to match your probe with the input channels 1 Setthe probe option attenuation in the channel menu to 10X and connect the probe to channel 1 on the scope If you use the probe h
31. Scope Plot Choose measured parameter A B and start recording the trend plot The scope Press F5 to pause or continue recording data prope T9 52 SHS800 Series 1 57MHz 25TMHzZ B75 Bm rm i ua iz B Bb54MHz Biagi 238 TrendFlot Parana Param B Met Page CHI pp CHIFreg Page 12 Figure 4 5 Trend Plot record Curve Display recorded data 6 Press F5 to enter the second page of trend plot menu 7 Press F1 to choose data display mode Normal the screen displays the data up to the minute View All the screen displays all data in the memory 8 Zoom function under full screen mode press time base to zoom in or zoom out 9 Data analysis move cursor analyzing data over time e 2 380V TO TOMHz is 810 00rriv 1 57MHz 2 5TMHz z2 388U EGE SoBe B 824MH2 Sei sia TrendPlot Parana Param B Met Page CHI pp CHIFreg Page 12 Figure 4 6 Analyzing Trend Plot 10 Save waveform save the recorded waveform to exterior storage device to make more detailed analysis 11 Press Return to exit trend plot SHS800 Series 53 Waveform Recorder Press Recorder to enter recorder main menu under scan time base then press F2 to choose Scope recorder Figure 4 7 Waveform Recorder Menu Table 4 3 Waveform Recorder Function Menu Record Record waveform without gap Replay Replay the recorded waveform Setup the parameters of waveform recorder Exit waveform recorder function
32. User Manual SHS800 Series Handheld Digital Oscilloscope V1 3 SIGLENT TECHNOLOGIES CO LTD Declaration 1 Copyright SIGLENT TECHNOLOGIES CO LTD 2 Information in this manual takes place of all data published before 3 SIGLENT company reserves the rights to change specifications and price 4 Contents in this manual are not allowed to be copied extracted and translated without permission by the company SHS800 Series I Safety Information Carefully read the following safety information before using the SHS800 Specific warning and caution statements where they apply appear throughout the manual A Warning identifies conditions and actions that pose hazard s to the user A Caution identifies conditions and actions that the user should notice The following international symbols are used on the SHS800 and in this manual A A D Hazardous Protective Warning Earth Ground Power Voltage Earth Ground Switch Use only insulated voltage probes test leads and adapters supplied with the SHS800 or accessories appointed by the company Before use inspect voltage probes test leads and accessories for mechanical damage and replace when damaged Always connect the battery charge first to the AC outlet before connecting it to the SHS800 Do not apply voltages that higher than 600 V from earth ground to any input when using scope ports in a CAT environment Do not apply voltages that hi
33. Y eEE 65 Appendix B Default Setup 1 eere eee eee eene eee seen etes en tenete en teen ette sten e ene tete eene senes nete 72 Appendix C Battery Installation 4 cere eee Leere eere eee e eere teet eene etta eene eene en eee en snae eaae 74 Appendix D Daily Maintaining and Cleaning cere Leere eee eee ee e eee eee eere en estne 75 VIII SHS800 Series Chapter 1 Accidence About this Chapter This chapter mainly covers the following contents Geta primary understanding of the front panel and user interface Abrief function check Probe compensation Accidence of the Front Panel and User Interface You d better get an understanding of the front panel before you operate the SHS800 series Handheld Digital Oscilloscope The following contents introduce the function of the front panel With its help you could be familiar with the operations of the SHS800 in a short time SHS810 100MHz 1GSa s 8 SIGLENT Handheld Digital Oscilloscope CHi 1 004 CH2 1 99U M Says Figure 1 1 Front Panel SHS800 Senes 1 Description sg a ODINO N power on off key CH1 vertical range and position key CH1 on off key Scope Meter menu option keys Handle LCD LOGO BW and sample rate Recorder function Notes 10 arrow keys 12 13 14 15 16 ive menu on off key Auto Cursor function keys CH2 on off key Trigger U
34. al position is at a limit when you turn the horizon position knob Volts Div at limit Mention you that the vertical voltage have already touched the Min 2mV div or the Max 100V div Volts position at limit The system would display this information when the vertical position is at a limit SeciDiv at limit Prompts that the Volts Div is at full range while turning the vertical scale knob Holdoff time at limit Use the arrow keys when holdoff time has been to max or min value now the system will clew this information Function isn t useable Under several special modes the some functions could not be running Nosignal The system would clew this information when the signal could not match the auto set condition Using in the auto set Adjust at limit You could adjust the pulse width by the arrow keys till the pulse width has reached min20 0ns or max 10 0s Location Empty If you have no stored waveforms or setups on some location the screen will display this information when you press the Recall button on this location USB Flash Drive Plug In This information will appear when you invert the USB Flash Drive to the USB Host port USB Flash Drive Pull Out This information will appear when you pull out the USB Flash Drive Store Data Success Save setup data waveform data or Figure data to the 60 SHS800 Series internal of the oscilloscope or USB flash s
35. ale Display ext Page Figure 2 18 FFT Function Menu 2 Table 2 10 FFT Window Function Window Characteristic Advantage content The best frequency Symmetric transients or bursts resolution but the worst Equal amplitude sine waves with Rectangular magnitude resolution It fixed frequencies Broadband is essentially the same random noise with a relatively as no window slowly varying spectrum Better frequency Sine periodic and narrow band Hanning poorer magnitude random noise accuracy than rectangular Hamming has a slightly Transients or bursts pulse The Hamming better frequency ranges of the signals have great resolution than difference from before to after Hanning Best magnitude Single frequency waveforms to Blackman resolution but worst find higher order harmonics rr frequency resolution 14 SHS800 Series FFT Zoom zoom in FFT waveform vertically by 1X 2X 5X and 10X Scale choose dBVrms or Vrms as a measure unit Display Spilt or Full Screen FFT waveform display mode FFT Waveform Operation Application Example Operation steps 1 Inputa signal to CH1 and press Auto Press Scope and choose Math to enter waveform operation Press F1 to choose FFT Press F5 to enter the second page of the menu and adjust the setups Adjust the channel vertical scale and complete FFT waveform s vertical position and pro m vertical scale settings 6 Press F5 to exit waveform operat
36. ccording to the real demand 5 Read current value T SHS800 Series 49 Chapter 4 Using the Recorder Functions About this Chapter This chapter provides a step by step introduction to the recorder functions of SHS800 series Handheld Digital Oscilloscope The introduction gives basic examples to show how to use the menus and perform basic operations The recorder mainly includes the following functions Trend Plot Trend plot is to save the measurements in the memory and then plot a graph of Scope or Meter measurements as a function of time Waveform Recorder Record real time waveform without gap or space That is to say every time the SHS800 can save all captured waveform data and then replay them The maximal recording length of waveform recorder is 7M data points 50 SHS800 Series Oscilloscope Trend Plot 9 8 9 7 8 PEMU ag 35 318 Trend Plot Farama Parara B CHI Vipp CHI Yina Page 12 Figure 4 1 Scope trend plot user interface current recorded time the percentage of recorded date take in the whole memory Value of the latest recorded data point A Value of the latest recorded data point B real time sampling time of the cursor point parameter measurement of the cursor point vertical scale vertical
37. elect input signal source for voltage measure Vpp Vmax Vmin Vamp Vtop Press F2 or use the arrow keys to Vbase Vavg Mean Vrms FOV select voltage measure parameter Display the corresponding icon and measure value of your selected Voltage measure parameter Return to the home page of MEASUREMENT menu Return Time parameter measure Period 1 88ms Figure 2 32 Time Measure Function Menu Table 2 16 Time Measure Function Option Setting Instruction Source CH1 CH2 Select input signal source for Time measure Period Freq Wid Wid Rise P F2 th w keys to Time Fall Time BWid Dut CS select time measure parameter Display the corresponding icon and measure value of your selected time measure parameter Return to the home page of MEASURE menu Delay parameter measure Phase E Figure 2 33 Delay Measure Function Menu SHS800 Series 21 Table 2 17 Delay Measure Function Option Setting Instruction Select input signal source for dela Source p 9 y measure Press the Type button or use Phase FRR FRF FFR FFF Type arrow keys to select delay LRR LRF LFR LFF measure parameter PR mmo dla Display the corresponding icon S MLL OREL PRO mo dl and measure value of your aoo Au m selected Delay measure parameter Return to the home page of MEASURE menu Return All parameters measure Source Voltage Time Delay on Retur Figure
38. enter the math waveform function menu Operation Invert Next Page Figure 2 14 MATH Menu Table 2 8 Math Menu Function Option Setting Instruction D CH1 CH2 CH2 CH1 Operation 2 CH1 CH2 CH1 CH2 CH2 CH1 FT fenor Tarom ee Turn off the invert function Next Page Page1 2 Enter the second page of MATH menu Met Page Figure 2 15 MATH Function Menu Table 2 9 Addition operation Use arrow keys to move the waveform upright Use arrow keys to adjust the scale of math waveform Waveform Turn on the math waveform Math Switch Unique key turn off the math waveform Next page Page2 2 Return to the first menu off math waveform Waveform Math Application Example Operation steps of adding two waveforms 1 Input two signals to the channels and press Auto Press Scope and choose Math to enter waveform operation Press F1 to choose operation Press F5 to enter the second page of waveform operation menu Ol de OH dM Use vi and 4 e functions keys and the arrow keys to change the parameter to display the best waveform 6 Press F4 to exit the math waveform operation SHS800 Series 13 MATH Operation Invert Met Page Figure 2 16 Result of Two Waveforms Add About FFT operation Using FFT math operation can translate time field signal to frequency field signal Operation Source Window FFT Zoom Met Page FFT Hanning 1 Page 12 Figure 2 17 FFT Function Menu 1 Sc
39. eries Table of Contents Sarety Informationi aspa crearla NEUEN IRSE DEINER EN EEUU eERTER II Safety Operation Of Batte ly usted e ee Yee IEEE E CEN YER QR ERE GA NEVER NO Un VEDO YN VE REEUE IV Introduction of SHS800 Series eee LLL Lee Lees eese e eese eeessseses esee ese eee eese e esee sees eee eee V FUNCION GhgrdclerisliGS o tee e teta nb e ai aisi V ACCcessores Ol SSC OO cs e eR tie Nem A EE VI Ghiapter 1 ACCIden6e siano 1 Accidence of the Front Panel and User Interface esses eene 1 FUNCION C eG uu oo cR eV ia lana a ae M LE C MER lede 4 Chapter 2 Using the SCODE nodais lA ARI 5 Menu and ontrol BU MOS tie cau rote ied oth tdem e itv ee tu tardis comet eade dieu ep gets 5 Automatic Settings aeo n s o ela 6 CHAT GCH2Z Channel FUNGIONS ucro otra sete ceto rera ee tr sese opem arare Peso ete eru eet elu eee lui 6 SCODE S FUNCION MENU tarinasi iic eatute esi ue Ula te cedem E Clan D UMS nanan 9 Acquire Signals Sys fe M cicer rh recte Cete ee aoa tos CY YE TE CE YN EN ka e ERN TE C 9 DISDIay SVSIOEI s sce ie ta iut dela m dec dicte tto ont detiene 11 Math Waveform eoe e beso vod b mtt a tee e v tete na 13 Horizontal Systeme eoe eis oro RE ene a paene at 15 Reference WavelormM ra cet acu detenta aser isse ns ae ae emi rane reu eda ne ue eed 17 Cursor and Measure System iii 18 Cursor meas uc LIL LL p M M M 18 Parameter Meadsulte odit teen tcn M UE maureen clans 20 TENSES
40. eter Trend Plot User Interface current recorded time the percentage of the current data take in the whole memory the parameter value of the recorded data up to the minute DC AC manual auto the real time the sample time of the cursor point the parameter measurement value of the cursor point vertical scale 0 vertical scale zs fige cL o p cs SY cc TrendPlot SaRate Mest Page Figure 4 13 Multimeter Trend Plot Menu 1 SHS800 Series 57 Table 4 6 Multimeter Trend Plot Function Menu 1 function setting Instruction Quilt the current data and start to record Restart afresh Sa Rate 10Sa 0 005Sa Set sampling rate Display normal Display the recorded data up to the minute mode All view Display all dots mode Manual ext Page Figure 4 14 Multimeter Trend Plot Menu 2 uc 4 7 Multimeter Trend Plot Function Menu 2 Function Data in the memory transferred to the storage exterior storage device Of Record data automatically Record cea Record data manually A Record presses a record Return Return to the multimeter function state NexPage NexPage Page2 2 Return to the first page Multimeter trend plot application example Start plot function Operation steps 1 Input a measured signal correctly See chapter 3 Using the Multimeter 2 Press F5 to enter Trend Plot at the multimeter main menu The SHS800 will record the measurement value of the input port continuously and
41. gher than 1000 V from earth ground to any input when using scope ports in a CAT environment Do not apply input voltages above the rating of the instrument Use caution when using 1 1 test leads because the probe tip voltage will be directly transmitted to the SHS800 Do not apply voltages that higher than 300 V from earth ground to any input when using multimeter ports in a CAT Ill environment Do not apply voltages that higher than 600 V from earth ground to any input when using multimeter ports in a CAT environment Do not apply voltages that higher than 300 V from earth ground to the isolated inputs when using multimeter ports in a CAT Ill environment Do not apply voltages that higher than 600 V from earth ground to the isolated inputs when using multimeter ports in a CAT environment II SHS800 Series Do not insert metal objects into connectors Use of the SHS800 in a manner not specified may impair the protection provided by the equipment Before use inspect the test leads for mechanical damage and replace damaged test leads Whenever it is likely that the safety has been impaired the SHS800 must be turned off and disconnected from the line power The matter should then be referred to qualified personnel Z Warning Standard probe 10 1 supports CAT II 400V Optional probe supports CAT II 1000V and CAT III 600V SHS800 Series III Safety Operation of Battery SHS800 series handheld digital oscillosc
42. heck the instrument and the probes according to daily use situation Please do the following steps to clean the surface of the instrument To clean the exterior surface perform the following steps 1 Removelooseduston the outside of the instrument and probes with a lint free cloth Use care to avoid scratching the clear plastic display filter 2 Usea soft cloth dampened with water to clean the instrument Please disconnect it from all power sources firstly Note 1 To avoid damage to the surface of the instrument or probes do not use any abrasive or chemical cleaning agents 2 To avoid water causing electric short circuit or body hurt Make sure the instrument is completely dry before reconnecting it to a power source SHS800 Series 75
43. idth menu 1 GGGdGkHz TRIGGER Source Wher 2 SRL Next Fage Figure 2 42 Pulse Trigger Video Trigger Type Source Polarity cunc ext Pace Figure 2 43 Video Trigger Menu 1 28 SHS800 Series Table 2 24 Pulse Trigger Function Menu 2 Option Setting Instruction When you select the video type put the couple set Type Video to the AC then you could trigger the NTSC PAL and SECAM video signal CH1 CH2 Select the input source to be the trigger signal Normal triggers on the negative edge of the sync Normal ua Polarity eae Inverted triggers on the positive edge of the sync pulse Inverted All Lines Line Num Odd Field Even Field Next Page Page 1 2 Enter the second page of Video trigger menu Select appropriate video sync Type standard Mode Next Page Figure 2 44 Video Trigger Menu 2 Table 2 25 Pulse Trigger Function Menu 2 Option Setting instruction When you select the video type put the couple set to the AC then you could trigger the NTSC PAL and SECAM video signal Type NTSC Select the video standard for sync and line PAL SECAM number count Use this mode to let the acquisition free run in the absence of a valid trigger This mode Standard allows an un triggered scanning waveform at 100 ms div or slower time base settings Use this mode when you want to see only valid triggered waveforms when you use this mode
44. imit Use up and down arrow keys to set the filter range 7 Press F1 to turn on the filter 8 SHS800 Series E n zB BB ua 1 3 26 G0G4MH d IELET FILTEE Upp_Limit ELE Return Figure 2 5 Before Turn On the Digital Filter TETE FILTER Fitter Type Upp_Limit Figure 2 6 After Turn On the Digital Filter Scope s Function Menu The button Scope includes the following functions Figure 2 7 Scope Function Menu Acquire Signals System Press Scope and choose Acquire to enter acquiring system See Figure 2 8 Sampling Peak Detect ME 25 Gns CHL Femi ACQUIRE Acquistion Sinki Mode Sa Rate IRR 256 RealTime SiWlliie t Figure 2 8 Acquire Signals system function menu SHS800 Series 9 Table 2 5 zE Signals System Function Menu Introduction Sampling Sample and display most waveforms accurately OUS Peak Detect Detect burr and reduce fake wave phenomena Reduce random and irrelative noise 4 16 32 Averages Select the times of averages 64 128 256 sinx Use sin interpolation Equ time Set the Sampling mode to Equ time Real time Set the Sampling mode to Real time Sinx x Mode Sa Rate Display system sampling rate Sampling To construct the waveform the scope samples the signals in equal interval Peak Detect The scope captures the maximum and minimum
45. inuity Measure Function capacitance DC current AC current Ma input Volge AC virtual value 750V AC Frequency 20Hz 1kHz DC 1000V APA Ra em OM DE are Input Impedance 10M 68 SHS800 Series DC Voltage 60mV 10uV 1 15digit s 196 5digi 600V 100mV 1000V 1V AC Voltage 60mV 10uV 1 15digit 600mV 100uV 6V 1mV 60V 10mV 1 5digit 600V 100mV 750V 1V Resistance Range Resolution Definition 6000 0 10 6K 10 60K 100 1 5digit 600K 1000 6M 1KQ 60M 10KO Diode and Continuity Measure Diode O 2V Continuity lt 500 alarm 400nF 100pF p 4 5digit 40uF 10nF 400uF 100nF SHS800 Series 69 DC Current 60mA 10UA onm 600mA 100uA 1mA Du di 1 5 5digit 10A 10mA A range measure periods10 seconds interval period215 minutes AC Current 60mA 10uA 1 5digit 600mA 100uA 1mA DS di 1 5 S5digit A range measure periods10 seconds interval period215 minutes Recorder Specifications In total 7M original points Single channel 7M Double channel each 3 5M Scope Trend Plot Display Mode full view and normal 800K points 18 hours Record Manual support Multimeter Trend Plot Display Mode full view and normal Record Length 1 2 M points Number Cursor and Zoom Support 70 SHS800 Series Generic Specification Display System Color TFT 5 7
46. ion CHI Laau M258ns X Hanning Pos 3606 AMH Figure 2 19 FFT Waveform Horizontal System Press Scope and then choose Horizontal to enter the horizontal function menu Delayed MembDepth Figure 2 20 Horizontal System Menu Delay scan zoom in the waveform being chosen Memory depth normal memory and long memory SHS800 Series 15 Delay Scan Application Example Operation steps pia SY gt Input a waveform to CH1 or CH2 Adjust time base to display the best waveform Press Scope and then choose Horizontal to enter horizontal system Press F1 to turn on delay scan Change time base and choose a window waveform to zoom in and analyze Press F1 to turn off delay scan iglent NAM de m x 06 01 13 23 57 24 CHi LBL M S6Gns CHI 70 amt W Po 5 BA ps Figure 2 21 Delay Scan Waveform Normal Memory and Long Memory under Real Time Sample Application Example Operations steps 1 d 16 Input a sine signal with 100M bandwidth and 4V Vpp to CH1 or CH2 adjust time base to 50nS The common storage sampling rate will be 1GSa s while the long storage is 500MSa s at this time Press Run Stop respectively at common storage and long storage Change time base to let the whole waveform display on the screen a HH H1 14 milia E M Pos ps MEG 1 0 0001 MHz CH 1 861 DECEME CHi FAGAM HORIZON Delayed hiemDepth Figure 2 22 Normal Memory SHS800 Series
47. lash Drive Application Operation steps 1 Press Save Recall andpress F1 Type to choose Waveforms 2 Insert USB flash drive to USB host you will get the message USB flash driver connects successfully 3 Press F2 Save to to choose File SHS800 Series 35 4 Press F4 Save to enter save recall interface 5 Press F1 Modify to choose File 6 Press F2 New File and input the mane of the file according to the prompts to create a new file Then press Confirm Figure 2 55 Input the Name of the File 7 File saves successfully tacit ext Page Figure 2 56 File Saves Successfully Recalling a file Choose the file to be recalled and press Recall at the memory interface which you complete the recalling operation Notes The picture in the SHS800 with a BMP postfix can t be recalled but you can open it in computer with relative software 36 SHS800 Series Utility System Press User to enter utility system menu See figure 2 57 ee Sound Counter Language Net Page ESS 4X ee JHA Page 14 Figure 2 57 Utility System Menu 1 Table 2 36 Utility System Function Menu 1 System Displays the main information Status of the SHS800 Sound qu Open the key press voice qx Close the key press voice T ff On Off urn on off the frequency counter Simplified Chinese English Traditional Chinese Arabic Language French German Russian Select the interface language Spanish Por
48. lay 2 Maybe the display Type is set to Vectors You could set it to Dots mode to improve the display 7 The multimeter measurements aren t correct 1 Check that if the range of the SHS800 matches with the measured item 2 Make sure that if the multimeter is beyond the calibration date if the measurements and the real values are beyond the relevant precision please contact with the calibration site warranted by SIGLENT company to calibrate the SHS800 3 If you can t use the SHS800 normally all the same please contact with SIGLENT servicing center we will provide service for you 8 The other kind of trouble please contact with SIGLENT servicing center For more details please see service and support Z N Warning Person without warranty by SIGLENT Company shouldn t disconnect the machine for inspection or you will lose the quality guarantee SHS800 Series 63 Chapter 6 Service and Support About this Chapter This chapter covers basic maintain procedures that can be performed by the user You should have a detailed understanding of the content below to use and maintain you legal rights Maintain Summary Each SIGLENT product is warranted to be free from defects in material and workmanship under normal use and service The warranty period is three years and begins on the data of shipment This warranty extends only to the original buyer or ender user c of a SIGLENT authorized reseller If a product or CRT pr
49. ng id edge of 50 voltage level Ar The ratio between the first positive pulse width and the Duty period The ratio between the first negative pulse width and the DUM period Phase The phase difference between two waveforms EPREshoot di t n The time between the first rising edge of source 1 and the Da first rising edge of source 2 The time between the first rising edge of source 1 and the RF first falling edge of source 2 The time between the first falling edge of source 1 and FR ti the first rising edge of source 2 FF The time between the first falling edge of source 1 and the first falling edge of source 2 EE The time between the first rising edge of source 1 and the last rising edge of source 2 The time between the first rising edge of source 1 and the RF last falling edge of source 2 The time between the first falling edge of source 1 and BIS the last rising edge of source 2 LJL F HOLD Ju F MA LJL F MA SL F HL rusi L HL ma L DLL SHALL SL a The time between the first falling edge of source 1 and eee eae the last falling edge of source 2 SHS800 Series 23 Parameter Measure Supplication Example Operation steps 1 Press Cursor Measure twice and any key of F1 F5 to enter parameter measurement system 2 Choose any key out of F1 F5 to choose measure type For example Voltage Press F2 to choose measure parameter For example Vpp
50. ook tip make sure that the hook tip is fixed on the probe firmly 2 Attach the probe tip to the Probe Comp 3V connector and the reference lead to the Probe Comp Ground connector Display the channel and then push button Auto 3 Check the shape of the waveform displaying on the screen Under Compensated Correctly Compensated Over Compensated 4 Please adjust your probe or repeat all the operations above if necessary Multimeter meter pen To avoid obtaining no measurements or unnecessary damage to the SHS800 you should use the right jack when measuring current voltage and other measure 4 SHS800 Series Chapter 2 Using the Scope About this Chapter This chapter provides a step by step introduction to the scope functions of SHS800 series The introduction gives basic examples to show how to use the menus and perform basic operations without d covering all of the capabilities of the scope functions In order to use the SHS800 effectively we need to know the functions of the SHS800 below Menu and control buttons connector and control auto settings Scope measurement system trigger system storage system and utility system Menu and Control Buttons Table 2 1 Function Menu CH1 CH2 channel menu Display display menu automatic setting control menu Run Stop sample Stop button Cursor cursor menu Measure automatic measurement menu Trigger trigger menu Save Recall save Recall menu SHS800 Series 5
51. ope can be used to test float signal when power supplied by battery When using the double channels to test float signal the two channels should be connected to the same earth ground because the earth ground of the two channels is connected ZN waming Do not connect the ground spring to voltages higher than 42 V peak or 30Vrms from earth ground L waming Do not use USB line to connect SHS800 to any instruments such as computer printer and so on which connected to the earth ground or the SHS800 and the instruments connected with will be burned IV SHS800 Series Introduction of SHS800 Series This manual mainly introduces SHS800 series Handheld Digital Oscilloscope The SHS800 series is a high performance handheld oscilloscope with great range of dynamic input scope It has small volume which convenient to carry compact interface and etc It satisfies the most needs of outside measurement and improves working efficiency greatly Function Characteristics 9 The SHS800 combines the functions of oscilloscope multimeter and recorder including trend plot and waveform recorder along with double channels Oscilloscope channels input voltage grade voltage inputs directly through a BNC probe is as high as CAT I 300V and CAT III 150v Standard probe 10X CAT II 400 Optional probe 10X CAT II 1000V and 10X CAT III 600V Oscilloscope and multimeter safety grade is CAT II 600V and CAT III 300V 5 7 inches
52. or not 4 Check the tested goods produce the signal or not 5 Gather the signal again The value of the tested voltage is 10 times higher lower than the real one please do as following steps Check the attenuation quotient whether match the probe attenuation proportion or not Display the wave but not steady please do as following steps 1 Check the signal source on the trigger interface whether or not matches the signal channel 2 Check the trigger mode normal signal should use the edge trigger mode The video signal should use the Video Trigger mode The signal would display steady only using the matching trigger mode 3 Attempt to change the coupling into HF Reject or LF Reject display so that the High low frequency noise disturb the trigger should be filtrated Press RUN STOP button but no display Check the trigger mode on the trigger interface whether or not in the normal or single and check the trigger level is whether or not over the wave range If yes please put the trigger level to the middle position or set the trigger mode to the Auto position In another hand you could choose the Auto button to set up automatically 62 SHS800 Series 6 The signal is displayed as ladder like waveform 1 This phenomenon is normal The time base maybe is too slow you should turn the horizontal SCALE knob to increase horizontal resolution to improve the disp
53. ors can be adjusted Mode Type SOuUrCe Cur A ur Manual CHI Figure 2 26 Manual Cursor Measurement Table 2 12 Manual Mode Function Menu Option Setting Instruction In this menu set the manual cursor Cursor Mode Manual measure Use cursors to measure voltage Voltage i pe l parameters y Time i Use cursors to measure time parameters CH1 CH2 Choose the signal to be measured by Source MATH Sore REFAREFB eee Cur A Use arrow keys to adjust t position of cursor gt A oe Use arrow keys to adjust t position of cursor gt a B 18 SHS800 Series Track mode In this mode the screen displays two cross cursors The cross cursor sets the position on the waveform automatically You could adjust cursor s horizontal position on the waveform by turning the arrow kyes The oscilloscope displays the values on the top of the right screen ode Cursor A Cursor B Cur A Track CHI CHI Figure 2 27 Cursor Trace Menu Table 2 13 Trace Mode Function Menu Cursor Mode Track In this mode set track cursor measure CH1 CH2 Set the input signal channel that the Cursor A will Cursor A NONE measure CH1 CH2 Set the input signal channel that the Cursor B will Cursor B NONE measure Cur A i a Use arrow keys to adjust the position of cursor A Cur B ui n Use arrow keys to adjust the position of cursor B Track Mode Application Example Operation steps 1 Press Cursor Measure once to ente
54. oves defective within the warranty period SIGLENT will provide repair or replacement as described in the complete warranty statement To arrange for service or obtain a copy of the complete warranty statement please contact with the nearest SIGLENT sale and service office Except this summary or the applicable warranty statement SIGLENT makes no warranty of any kind of express or implied including without limitation the implied warranties of merchantability and fitness for a particular purpose In no event shall SIGLENT be liable for indirect special or consequential damages Contact SIGLENT SIGLENT TECHNOLOYIES CO LTD Address 3 F No 4 BUILDING 3rd LIUXIAN Rd ANTONGDA INDUSTRY GARDEN BAO AN DISTRICT SHENZHEN 518101 CHINA Tel 0086 755 36615186 occ ARRON lA i CEP http www siglent com 64 SHS800 Series Appendix A Specifications All specification applies to 10X attenuation switch setup probe and SHS800 series Handheld Digital Oscilloscope To verify if the SHS800 meets with the specifications it must meet the following conditions first The oscilloscope must have been operating continuously for thirty minutes within the specified operating temperature You must perform the Do Self Cal operation accessible through the User menu if the operating temperature changes by or more than 5 C The SHS800 must be within the factory calibration interval The SHS800 should be calibrated once ever
55. r cursor system 2 Press F1 to choose Tracke mode 3 Adjust cursor A and cursor B to trace waveform with arrow keys II iT Mode Cursor A Cursor Cur A Track CHI CEE Figure 2 28 Cursor Trace Menu Automatic measure mode This mode will take effect with automatic measurements The instruments will display cursors while measuring parameters automatically These cursors demonstrate the physical meanings of these measurements SHS800 Series 19 Automatic Measure Application Example Operation steps 1 Press Cursor Measure once to enter cursor system 2 Press F1 to choose Auto mode 3 Press Cursor Measure again and choose parameter types to be measured CH LBB 00 FEX MEASURE Source Type Figure 2 29 Auto Measure Parameter Measure Press I Cursor Measure twice and any key of F1 F5 to enter parameter measurement system Figure 2 30 Parameter Measure Function Menu Table 2 14 Parameter Measure Function Voltage Press this button to enter the Voltage measure menu Time Press this button to enter the Time measure menu Delay Press this button to enter the Delay measure menu All Mea Press this button to enter the All Measurement menu Return Return to the home page of MEASURE menu Voltage parameter measure Source Type p an Jis Ter Figure 2 31 Parameter Measure Function Menu 20 SHS800 Series Table 2 15 Voltage Measure Function Setting Instruction S
56. ser Save Recall function keys CH2 vertical range and position keys time base and horizon position keys multimeter input ports Run Stop The arrow keys include these functions direction keys moving trigger level setting the trigger level to zero choosing menu setting horizontal position to Zero moving cursor Figure 1 2 Side Panel Description 1 2 3 USB Device USB Host power input port SHS800 Series CH 2060 CHT 200A OM ETT CHI Coupling YY Limit WoltsiDiv Probe MextPage CNN 7 cone NES 14 13 12 H 10 Figure 1 3 User Interface Description 1 Trigger state Armed The scope is acquiring data of pre trigger please ignore all triggers under this state Ready The scope has sampled all pre triggers data and is ready to accept trigger Trig d The scope has found a trigger and is sampling data after trigger Stop The scope stops sampling data Auto The scope is sampling waveform without trigger under automatic mode Scan The scope samples and displays waveform under scan mode Shows location of current waveform in the memory Shows the trigger position in the memory Co N 4 9 Print Key option chooses to print figure 8 Print Key option chooses to save figure 5 E USB Device option chooses to connect computer 5 USB Device option chooses to connect printer 6 Shows power state 12 B shows the BW limited is on 7 Shows current time 13 Channel s vertical range 8
57. th the same the model Figure data and CSV file can t be recalled to the SHS800 but it can be opened on the computer through correlative software CSV can be opened by EXCEL software on computer Saving Setups Saving Setups to Device All setups are stored in nonvolatile memory When recall the setups the SHS800 will be under the setup save mode Type Save To Setup Setups m Figure 2 49 Saving Setups to Device Menu Table 2 29 Saving Setups to Device Function Menu Option Setting introduefion Saw __ Accomplish he operation of saving setups Recall Recal te saved setups Saving Setups to USB Flash Drive Type Save To Figure 2 50 Saving Setups to USB Flash Drive Menu Table 2 30 Saving Setups to USB Flash Drive Function Menu Setting Instruction Setup Used to save recall the SHS800 s setup menu Save to File Save the setup data of the SHS800 to USB flash drive Save Enter the waveform save recall interface SHS800 Series 33 Saving waveform Saving waveform to device Type SaveTo VVaveforrm Figure 2 51 Saving Waveform to Device Menu Table 2 31 Saving Waveform to Device Function Menu Waveforms Tvoe Menu for the Storage Recall waveforms in the dd Scope Save waveforms to the SHS800 s internal memory No 1 to Choose the position number to save recall setups O Sae Accomplish the storage Recall Recall the storage in the
58. tion Setting Instruction Save the current input value as a reference and record again Real value equals relative value plus measurement value Relative Value Real value equals measurement value Choose the best measurement scale automatically Manual Choose measurement scale manually Choose the best measurement scale Auto automatically according to the measurement value Choose measurement scale manually and there will be a warring when over the scale Tendency Plot Plot with the measurements according to time Oo O O 99 9mV 0 100 200 SUD 400 S00 60m UTO Meter Relative BI Off AUTO W Trend Plot Figure 3 2 DC Voltage Measurement O O O 196 5mV Q 100 200 300 A S00 er Relative Crt Auto n Trend Plat Figure 3 3 AC Voltage Measurement 44 SHS800 Series Operation Steps 1 Press Meter to enter multimeter mode press F1 to choose DCV ACV measurement 2 Insert the red probe to the V Q C banana jack input and the black probe to the COM Connect the other end of probes to the power or load to be measured 3 Turn on off the relative according to the real demand Choose Manual or Auto according to the real demand 5 Read voltage value T Making Resistance Measurement _ n t x O O O 99 60 Im 1 1 100 200 300 400 500 600 Meter Relative Ries Off hanual Trend Plot Figure 3 4 Resistance Measurement Operation steps 1
59. tuguese Japanese Korean Italian Page 1 4 Enter the next page of the Page menu Table 2 37 Utility System Function Menu 2 Option Introduction Do self Cal Do a self calibration to calibrate the channels Screen Test Run the screen detect pro Do Self Test dico Keyboard Test Run the keyboard detect program Print Setup Enter the print setup menu to set print options SHS800 connects to the printer through USB cable When you execute print function Printer mE i please select Printer At this time the print USB Device icon displays on the top of the screen SHS800 connects to the computer through USB cable When you execute EasyScope Computer software please select Computer At this time the computer icon displays on the screen Next Page Page 2 4 Enter the third page of the menu SHS800 Series 37 USE Device Net Page Da Da Page 24 Figure 2 58 Utility System Menu 2 USB Device Met Page Do cella saffet Setup Pinter Page 24 Figure 2 59 Utility System Print Setup Next Page Page 34 Figure 2 60 Utility System Menu 3 Table 2 38 Es System eee Menu 3 Introduction Update You can update the SHS800 by using USB flash Firmware a driver About two minutes Record _ Press this button to enter the waveform record menu Screen SalJer Mex Page 14min DateiTi Page 4 4 Figure 2 61 Utility System Menu 4 Table 2 39 Utility System Function Menu 4 1min
60. uccessful Ready Data Success Read setup data or waveform data from the internal of the oscilloscope or USB flash successful Please set USB Device to printer Press the S div knob will appear this information on the screen when the Print Key option is set to Print Figure and the USB Device option is set to Computer USB Flash Drive isn t connected When the Save To option is set to File or the Print Key option is set to Save Figure in Save Recall menu Press the Save option button or the S div knob before inverting the USB Flash Drive to the USB Host port will appear this information on the screen Record Wave Success This message will appear when you finish recording waveforms SHS800 Series 61 Troubleshooting 1 After the SHS800 is powered on if the screen remains dark please do as following steps 1 Check the power cable s connection 2 Ensure the power switch is turned on 3 After the inspections above restart the Handheld Digital Oscilloscope 4 If the Handheld Digital Oscilloscope is still not used after the checking please connect with MY company If there is no signal wave in the screen after gathering the signal please do as following steps 1 Check the probe connecting with the signal cable or not 2 Check the signal cable connecting with the BNC connector or not 3 Check the probe whether connect with the goods tested
61. y year All specifications are guaranteed except those noted typical Oscilloscope Specifications Inputs System Input Coupling AC DC GND Input Impedance 1MO 2 18pft3pf Probe Attenuator 1X 10X Voltage Probe Attenuator Factors 1X 5X 10X 50X 100X 500X 1000X BNC Maximal Input Voltages CAT Il 300Vrms refer to BNC crust CAT III 150Vrms Standard Probe 10X CAT Il 400Vrms Optional Probe 10X CAT III 600Vrms l i CAT Il 600Vrms Multimeter Floating Voltages CAT III 300Vrms h Mod C anne Common ode 100 1 50MHz Rejection Isolation Degree between Channels gt 35dB SHS800 Series 65 Acquisition System Sample Types Real time Equivalent time SHS820 Single Channel 32K double channels 16K Channel Sampling Mode Rate Single Don t Memory SHS815 1Gsa s 40kpts Channel support mae Singl 500MSa SHS806 oe i 20kpts 2Mpts Channel or below Double 500MSa s Channels or below Common Long 20kpts 1Mpts Sample Mode Sample Peak Measure Average Averages 4 16 32 64 128 256 Vertical System Vertical Sensitivity 2mV div 100V div 1 2 5 order 2mV 200mV 1 6V Channel Voltage Offset Range 206mV 10V 40V 10 2V 100V 400V Channels 2 60 MHz SHS806 100 MHz SHS810 150 MHz SHS815 200 MHz SHS820 Lower Frequency Limit AC 3dB lt 10Hz BNC input DC Gain Accuracy 5mv div 100v div st3 2mv div s 4 DC Measurement Accuracy
Download Pdf Manuals
Related Search
Related Contents
USER INSTRUCTIONS THE CLEAR™440 SERIES Mode de lecture des épaisseurs des deux verres d`un vitrage isolant STAND FAN - Bionaire link 1 - Yuneec Dell OpenManage Integration for VMware vCenter 2.0 Quick Installation Guide PDF-Bedienungsanleitung electronic locking for high voltage grounding system ProdSoft AB Roll Shop Computer System Lubrificazione e lubrificanti 形 V680S-HMD6D-EIP Copyright © All rights reserved.