Home
F-100Tv9 Manual - Advanced Research Instruments Corp.
Contents
1. A low bias potential at the entrance also creates a low electrostatic field that is less likely to attract a variety of possible spurious charged particles generated by a field emission thermal emission or photo emission and contribute to the background count rate This becomes particularly important when operating at low intensities where the background is highly significant The positive biasing scheme however has one disadvantage The noise ripple from the H V supply is directly coupled to the preamplifier which determines the lowest detection limit for the signal It should be obvious that if the ripple is specified as 10mV peak to peak value we cannot expect to detect any pulses smaller than that unless we want to count the H V ripple also Caution Some H V supply manufacturers specify the ripple as an RMS value which may be 5 10 times smaller than the peak to peak value of the same ripple Since most of the H V supplies today are the switching regulator type the noise consists mostly of high narrow shape pulses yielding a rather low RMS value It seems to be more attractive to specify a low noise level in RMS value than a much higher one in the peak to peak value A good indication of whether or not the H V supply we are using is too noisy is the specific rate meter reading the switching frequency of the supply without signal present with the H V on and the threshold of the noise discriminator safely above the amplifier dete
2. COMPATIBLE Diens c2 004uF SkU Seccececeeocecceccececccocosecsecsecceccceseceecccscesoeceeeceseseseeeseessesocesereecseeeeseeseee seceeceeeseececoececsceccesesecesecceecesecoccceseccceeeceeeet SMALL BIAS OR GND Fig 1 Detailed H V Biasing Diagram of MTS 100 The H V bias illustrated in Figure 1 requires positive High Voltage for the detector to have its collector at a more positive potential than the front of the electron multiplier This biasing scheme has several advantages over a negative biasing scheme Since the detectors entrance is close to the ground potenitial and it s not committed it may be biased up to several hundred volts negative by an additional supply to attract positive ions or other positively charged particles By changing the polarity of such additional bias supply we simply switch the system to detect negatively charged particles However when using such an additional bias we must remember to change the detector s bias accordingly since both supplies are typically grounded referenced to ground When using a floating uncommitted high voltage supply for the detector switching from positive to negative particle detection is much simpler All we have to do is change the entrance bias Although in this case we don t have to worry about the influence on the 3 potential across the detector itis absolutely necessary to shut the supplies off before changing the polarity of the bias supply
3. noise even less Check the supply ground connection and if there is no obvious power supply problem then the preamplifier needs to be repaired If the noise disappears when you disconnect the input then the noise is originating at the detector or in the way the preamplifier is connected to the detection system To find out the source of the noise in the detection system first turn off the detector s bias supply and the signal source and reconnect the preamplifier to the detector If the noise persists try to locate its level by turning the discriminator trimmer clockwise until the noise disappears Ifthe noise is too high to be eliminated by the trimmer check the input connections for a missing ground or shielding connection The input signal conductor should not be exposed anywhere from the detector to the preamplifier Good R F rules must be observed so as not to turn the input connection into a receiving antenna 13 One of the less obvious missing ground problems in the connection may be found in the SHV connector itself In some cases an intermittent noise is detected especially when the preamplifier is touched or moved Such a symptom usually indicates a poor ground connection at the input The most likely culprit is the SHV connector Check the ground connection at the SHV connector where the ground shroud of the female connector mates with the male The shroud may be collapsed and failing to make solid ground connection Fix or re
4. output from the MTS 100 11 Figure 7 shows signal and noise coming from the linear preamplifier just before it enters the discriminator for thresholding When the threshold trimmer is turned fully counterclockwise the threshold is buried in noise input noise plus preamplifier noise and the MTS 100 produces high count rate pulses even if no signal is present see Figure 8 Note the count rate when the threshold is set close to zero Since this is an unacceptable situation the solution is to increase the threshold by turning the trimmer clockwise until the noise disappears Perform this adjustment with the MTS 100 connected to the detector and the rate meter e Be sure that when you perform the adjustment that power is applied to the MTS 100 but without the high voltage applied to the detector Note If there are no pulses even with the discriminator fully counterclockwise the noise level might be below the minimum detectable threshold or additional problems may be present With the signal source photons electrons etc turned off turn on the high voltage and slowly increase to the approximate operating range If no H V adjustment is feasible then just turn it on Noise should increase because the bias background is significantly higher than the noise of the preamplifier Increase turning clockwise the threshold above the noise level A count rate of 1 to 5 cps may be acceptable The background count rate is usually dete
5. since the maximum count rate is limited by the pulse pair resolution For example the F 100E preamplifier coupled with the PRM 100 precision rate meter is a system capable of processing pulses only 5ns wide which results in counting frequencies of up to 75 MHz Even higher count rate up to 400 MHz can be obtained from a F 400E and PRM 400 system The key word here is frequency rather than count rate Frequency implies a periodic wave form or pulses equally spaced which is not usually the case in which intensities are measured ion photon electron counting or X Ray intensity measurements So the pulse overlapping probability that results can be described statistically by the Poisson distribution function The measurement here consists of the number of counts during a fixed period of time regardless of spacing or distribution This gives us an average number of counts during that period This number is usually recalculated to give us a number of counts per second CPS For a given pulse pair resolution the closer we approach the 5ns limit the more likely we are to encounter pulses closer than our pulse pair resolution and the more often we lose them However the loss is statistically predictable and we can correct for it using the Poisson distribution function and push the maximum count rate a bit closer to the maximum frequency How to Automate Your Data Acquisition The PRM 100 coupled with a computer possessing a fast paralle
6. 5 Q resistor and make sure there is enough of signal for the counting system to use Use a 50 Q resistor for a 50 Q cable and 75 Q resistor for 75 Q cable If the 50 Q termination is reducing the pulse amplitude excessively you may have to resort to a 75 Q cable termination system 15 Reducing Power Supply Induced Noise Noise is one of the basic limitations of virtually any signal detection system this is true no matter how advanced the system is However excess noise defined as noise above the theoretical limits can often be virtually eliminated The most common sources of excess noise are poor shielding poor grounding and noisy power supplies Since the noise induced by power supplies is the easiest to spot and remedy the following paragraphs deal with reducing this type of noise Using a Suitable Power Supply to Reduce Noise Those who have system noise often comment that the power supply cannot be responsible They say they have checked the specifications and they are incredibly good However the facts are often otherwise For example a typical high quality high voltage power supply specification may indicate noise less than 001 of the output voltage Sounds good but is it Let s say the high voltage is 2000VDC The 001 represents a 20mV noise level Many detector preamplifier configurations use a positive H V biasing technique directly coupling this noise to the preamplifier and therefore directly determining the min
7. Advanced Research Instruments Corporation 11950 Spruce Canyon Circle Golden CO 80403 USA Phone 303 463 5500 Fax 303 463 5505 Email aricorp aricorp com Web Site www aricorp com USER MANUAL for the MTS 100 PREAMPLIFIER DISCRIMINATOR March 2000 TABLE OF CONTENTS MTS 100 SPECIFICIATIONS INTRODUCTION Figure 1 Detailed H V Biasing Diagram of MTS 100 INSTALLATION Output Consideration Pulse Width Figure 2 Timing Capacitor Selection for Desired Ouput Pulse Width Figure 3 Connector Pin Out Figure 4 Cable Selection Guide Figure 5 Cable Diagram Potential Problems Figure 6 Example of Ringing Reduction by 50 Ohm Termination Horizontal 100ns Vertical 1V div Input Considerations OPERATION Connecting and Disconnecting Your Preamplifier Discriminator Set Up Figure 7 Typical Signal from the Detector and Threshold Setting Figure 8 Relationship Between Threshold Position and resulting Count Rate TROBLE SHOOTING GUIDE No Output Pulses Too Much Noise Figure 9 Missing Ground in the SHV Connector Double Pulsing Reducing Power Supply Induced Noise Using a Suitable Power Supply to Reduce Noise Noise from the Positive H V Supply Noise Reduction by Negative H V Biasing Technique Positive H V Biasing PHOTON COUNTING APPLICATIONS How to Optimize Photon Counting How To Minimize the Background Count Maximizing the Useful Count Rate How to Automate Your Data Acquisition oor or W W PD COND
8. O OO MTS 100 SPECIFICATIONS Input impedance Input range Input charge sensitivity Linear section gain Noise referred to input Input pulse polarity Input over voltage protection Input connector Threshold adjustment range Maximum pulse repetition rate Pulse pair resolution Minimum output pulse width Maximum output pulse width Output signal Output impedance Power connector and signal Second Output Signal Power requirements Physical size Maximum H V Bias 50Q 160uV to 200mV lt 70 fC 60db lt 5uV RMS Negative Standard SHV 100uV to 2 5mV ref to input gt 50MHz lt 20ns 10ns standard 1us 5V TTL 50Q 9 pin mini D type BNC 8V to 24VDC 50mA 1 7 x 2 4 x 7 2 5kV USER MANUAL for the MTS 100 PREAMPLIFIER DISCRIMINATOR V7 INTRODUCTION The MTS 100 Preamplifier amplifies minute pulses from Photomultiplier tubes Electron multipliers and similar detectors to produce standard TTL pulses for counting applications The MTS 100 also includes a decoupling H V network necessary to fulfill the detector bias requirements It features a low pass filter in the H V path to lower any noise there may be on the H V line The filter is especially crucial when the positive H V biasing scheme is used SHU CONNECTORS oseessoecosesseceseeosssoosesocccocoeccessecessoscoeososeeeoss MTS 100 BNC TTL OUT 002uF SkV iM SW te F 100T POWER AND a a o SIGNAL lo 1 F 100T
9. ROUBLE SHOOTING GUIDE No Output Pulses First check the D C supply voltage Unplug the 9 pin cable from the preamplifier and measure the voltage on the cable connector You should see 8V to 24V on pin 3 and 8V to 24V on pin 8 relative to pin 7 power ground Then disconnect the input connector then turn the threshold control fully counterclockwise With a piece of wire touch the center pin of the input connector This should produce plenty of signal on the output If no signal appears at the output the preamplifier may need to be repaired However if the preamplifier appears to be functioning then you should consider the detector its power supply or even the signal source itself Use an oscilloscope or different preamplifier to make sure the detector is producing a useful signal If it is you may have to slightly increase the H V bias on your detector The reason for this is because some slower preamplifiers are even more sensitive then the MTS 100 and it is possible that the signal produced by your detector at this bias setting is below the sensitivity level of the MTS 100 Too Much Noise First determine where the noise originates whether in the amplifier itself or from another source To find out where the noise is coming from disconnect the input to the preamplifier and if the noise persists check the noise level on the positive and negative supply The noise level should be less than 10mVp p 120Hz and a high frequency
10. ctor noise level Although the H V always increases the background count rate when the rate meter reads a stable frequency between 20kHz and 200kHz or Counts Per Second CPS it is a definite indication that we are counting the H V ripple frequency The RC filter in Figure 1 consisting of R3 and C2 will reduce the H V noise level significantly but it will not eliminate it For the best results the lowest possible noise ripple H V supply is absolutely essential The preamplifier follows the decoupling network and its function and operation are described in the next part of this manual USER MANUAL for the MTS 100 PREAMPLIFIER DISCRIMINATOR V7 This manual is written to assist you in the installation and operation of your MTS 100 preampilifier discriminator INSTALLATION Refer to Figure 3 on the following page for the connector pin out The preamplifier requires an unregulated dual power supply of 8V to 24V typically 12V 15V or 24V at 50mA and 30mA This is the least critical connection since built in voltage regulators virtually eliminate any ripple instability or noise on the power lines Output Considerations Output pulses from pin 1 represent the standard TTL CMOS level 5 0V unterminated with respect to ground pin 6 The output impedance of the MTS 100 is 50 ohms To avoid any potential problems related to an improperly terminated transmission line such as ringing or multiple pulsing use a 500 coaxial cable t
11. e eliminated by setting the discriminator to a lower sensitivity region thus reducing the overall performance Bear in mind that the 160 uV input sensitivity of the MTS 100 preamplifier is about the same as most radio and television receivers except with a much larger bandwidth making the situation much more critical Also the preamplifier must operate in an electrically hostile environment without responding to the severe interference of switching power supplies the R F part of mass spectrometers computers SCRs etc These interference sources are a serious threat to an interference free operation of the preamplifier especially if they are located in the vicinity of the detector preamplifier Therefore good R F grounding and shielding rules must be strictly observed Preamplifiers are normally set to accept negative pulses since all PMTs and electron multipliers 9 produce such pulses Although the input of the MTS 100 is compatible with either a positive or a negative biasing scheme of the PMT the details may vary substantially Finally for the input connection itself use a SHV to SHV adaptor or a very short cable Under some circumstances an additional ground strap may be necessary 10 OPERATION Connecting and Disconnecting Your Preamplifier CAUTION Before applying power recheck the low voltage polarity Reversed V with V may destroy the preamplifier Although the input to the MTS 100 is protected against high
12. erminated by a 500 resistor or a ratemeter with 500 input impedance Pulse Width The minimum output pulse width is about 10ns You can extend the pulse width by adding a C8T capacitor inside the MTS 100 The graph in Figure 2 enables you to determine the required capacitor corresponding to the desired pulse width However it is important to use the minimum pulse width that can be tolerated by the rate meter This is because the minimum pulse pair spacing maximum pulse pair resolution is directly related to the output pulse width and pulse pair resolution is limited by the output pulse width Therefore the wider the output pulse width the worse will be the pulse pair resolution and the lower the maximum count rate 10000 1000 100 Capacitance pF 10 100 1000 Pulse W idth ns Figure 2 Timing Capacitor Selection for Desired Output Pulse Width Pin Number Signal TTL Output Not Used V 8 to 24 VDC Not Used Not Used Signal Ground Power Supply Ground V 8 to 24VDC Optional remote threshold control monitor Figure 3 Power connector pin Out BNC TTL Output OANDOAARWDND 1 Cable Model T CUS 6 Optional remote MTS 100 threshold controlimonitor Amplifier Discriminator F 100PS Power Supply MTS 100 Opt Amplifier Discriminator OUTPUTS ECL JTL ect U Andog NIM PRM 100 C ounter T imer L da a Co A Cable Model T P1 6 MTS 100 ee A
13. even triple the rate it is supposed to be This condition is usually the result of a poor connection between the preamplifier and the detector such as Ground connection Mismatched impedance of various connectors Vacuum feedthrough devices Exposed input signal connection This poor connection results in a standing wave condition on the input of the amplifier that may produce one or more extra pulses for each authentic pulse arriving from the detector Raising the threshold of the discriminator usually solves the problem However if good R F practice was not followed when the input connection was done standing waves may be so severe that the preamplifier will enter sustaining oscillations Avoid severe ringing Here is the primary rule that must never be compromised the connection from the detector to the input of the preamplifier must be done with a 50 Q impedance coaxial cable as short as possible without exposing the center conductor at feedthrough or anywhere else Any compromise at all of this primary rule results in the increase of the standing wave condition ringing and an unnecessary high threshold level of the discriminator to avoid the double pulsing Double pulsing can also occur on the output of the preamplifier if a long and unterminated cable connects the preamplifier to a counting system such as a rate meter This is perhaps the easiest problem to rectify Just terminate the connection on the rate meter side with a 50 Q or 7
14. he input of the preamplifier This direct coupling eliminates an H V blocking capacitor and the necessary additional wiring so that a potential source of noise is eliminated 16 DC current can be measured as the signal when the count rate exceeds the maximum count rate of the pulse processing channel e When pulses and a DC component both are present a special dual channel preamplifier such as a COMBO 100 can be used The COMBO 100 processes a low intensity signal as pulses and it processes the high intensity signal as a DC voltage proportional to the intensity A disadvantage of the negative biasing technique is the presence of the high negative potential at the detector s front end This negative potential prevents negatively charged particles from being detected and tends to collect any positive spurious ions thus increasing the background count Because of the higher background count the negative H V bias scheme is not recommended for applications with signal intensities count rates just slightly above the background level It should be used only for those applications in which the processing of higher count rates is more important than a low background Positive H V Biasing This technique can be used to advantage in some situations For example a photo multiplier tube can be operated at lower dark current or background level Also positive H V biasing can be used to bias a charged particle detector so the front end of
15. imum detection limit see Figure 5a Noise from the Positive H V Supply When a positive H V supply is used the noise from the supply enters the preamplifier with almost the same amplitude as at the power supply Note that we are trying to detect pulses of amplitude of 0 2mV in the presence of almost 20mV noise Yet the detector preamplifier i e an MTS 100 preamplifier system may be capable of detecting and processing signals as low as 100uV The 20mV noise degrades the detection level from 100uV to 20mV This problem can be remedied by obtaining an H V power supply with noise less than 100uVp p such as the Advanced Research Instruments model PMT2000 2k Series Noise Reduction by Negative H V Biasing Technique If a negative H V biasing technique is used the 20 mV noise does not appear directly at the preamplifier s input The residual noise reaching the preamplifier is only 20uV well below the detection capability of the system The H V is applied to the front end of the detector while the output from the detector is at virtual ground When the preamplifier is electrically at the ground side of the biasing system the preamplifier will see only a small fraction of the power supply induced noise in other words a negligible noise level Even higher noise levels than this may be quite acceptable Additional benefits of negative H V biasing are The output of the detector is at virtual ground potential and can be DC coupled to t
16. l interface becomes an intelligent easy to use fully automated system when attached to virtually any spectrometer To control the spectrometer to change the wave length etc the parallel interface provides a driver for a low power stepping motor and input for a set of limit switches The computer can be programmed to provide flexible intelligence to the system to adapt specific requirements for a particular measurement to improve accuracy to store data for subsequent analyses or to simply archive many spectra on a mass storage media Since most spectrometers are driven by a synchronous motor which produces a constant speed movement of the wave length selector shaft a strip chart recorder can be driven at the same speed to produce a wave length vs intensity chart Although the PRM 100 produces an analog output compatible with most strip chart recorders there is an easier more accurate and less expensive way to produce a wave length vs intensity chart Just use any personal computer to do the data acquisition and store and plot the charts on a dot matrix printer Also to automate the acquisition of X Ray diffraction patterns and have them readily available for further processing Advanced Research Instrumentation Inc provides software which is specifically designed for an X Ray diffractometer This software can be easily tailored to particular spectrometers typically without any additional expense 19
17. might consider an unterminated signal connection but only if the physical length of the coax cable is less than approximately one meter The shorter the connection the safer it is to use the unterminated cable The decision as to whether to use a terminated cable must be based on the input specifications of the your counter timer rate meter A fast sensitive rate meter gt 20 MHz with sensitivity lt 1V always requires proper termination A relatively slow rate meter usually has an input sensitivity control or a threshold adjustment that can be used for input signals of 1V to 25V and high input impedance In this case although external termination for a short cable may not be absolutely necessary it is nevertheless recommended The 3 0V signal level with a properly adjusted input threshold to about 1V is the best solution Along cable longer than 15 feet gt 5 meters must almost always be terminated The following examples Figures 6a and 6b show significant improvement of reflectance for a terminated cable compared with an unterminated one The most significant value is the positive excursion of the ringing value since the tripping threshold for the TTL signals is about 0 7V minimum The 1 0V glitch on the unterminated cable will very likely produce a double pulsing The problem is most severe when a high frequency and high sensitivity counter is used When selecting the counter timer rate meter make sure that only the signal pulse
18. mplifier Discriminator SUB MNID 9 PINCONNECTOR Cable Moda T P4 6 PRM 400 Counter Timer MTS 100 i Amplifier Discriminator SUB MINID 8 PIN CONNECTOR Figure 4 Cable Selection Guide Cable Model T P1 6 MTS 100 to PRM 100 or to F 100PS I IL REIU a T i e ia POWER V awcwerED FOWERY AWG 6 BLACK POWER GND AWG 24 GREEN 9 DET THRESH MONITOR R REIU Cable Model T CUS 6 MTS 100 to Open Ended Cable RGIWU Connectors Spin Submin D F emale AMP No 207752 1 745002 61831 1 Pins AMP No 1 66505 0 Figure 5 Cable Diagrams Potential Problems Two potential problems may be associated with the 500 termination These are discussed as follows Some rate meters or counters may have only a high impedance input available therefore an external terminator must be used If your rate meter does not have 50 Q input impedance you can easily modify it by attaching a metal film or carbon resistor across the input Do not use wire wound resistors because they tend to have high inductance If modifying the input is not feasible a BNC tee adaptor may be used with a BNC 50 Q terminator The 50 terminators are available from Advanced Research Instruments Corp or any major electronic supplier The 500 terminator reduces the pulse height to 3 0V which may not be sufficient for some rate meters to detect If the 3 0V pulse height is insufficient for your rate meter or counter you
19. place the female connector Figure 9 Missing Ground in the SHV Connector The ground shroud in the SHV connector may be responsible for poor ground connection When you successfully locate the noise level with the discriminator trimmer the next step is to lower it to an acceptable level via improved grounding and shielding When the noise level is below the minimum threshold setting turn on the detector s bias and note any changes of the noise level A slight increase is normal A drastic increase of noise points to either a noisy bias supply or an H V insulations breakdown Check the noise level of the supply Since even the best supplies may have a 2 50 mV high frequency ripple that may be directly coupled to the input of the preamplifier you may need to geta better H V supply such as ARI s model PMT 2000 If you are using a negative high voltage supply and your preamplifier is connected to the ground side of the supply The ripple is not critical because only a small portion of it reaches the preamplifier Attention should instead be focused on the H V breakdown effect where the most likely cause is contamination or device break down Note In many experimental set ups the researcher forgets the high frequency grounding and shielding rules Then any high and unpredictable noise is usually 14 blamed on the preamplifier since that is where the noise shows up Double Pulsing The count rate is consistently about twice or
20. rmined by the quality of the detector or the entire experimental set up Now apply the signal to your detector PMT or electron multiplier The count rate should increase to the correct level With the signal applied increase the threshold by turning the threshold trimmer further clockwise and note the change in the count rate There should be minimal change within certain ranges i e the best setting as shown in Figure 8 When you reach a threshold setting that is too high the count rate drastically drops off The best setting is approximately in the middle between the noise level and loss of the signal If you cannot locate the point of losing the signal this indicates unnecessarily high voltage to the detector Reduce this voltage to the level where you can observe the output change with respect to the threshold shown in Figure 8 The best setting range should not be greater than 2 5 turns of the threshold trimmer If you cannot reduce the detector voltage as described you may operate at the higher voltage without any damage to the preamplifier The reason for minimizing the bias on the detector is e t Extends the life of the Detector e The detector produces narrower pulses which improves the pulse pair resolution e Reduces the background count rate Check the best setting at least once a year to assure that detector aging or changes in the high voltage bias do not shift the threshold out of the correct setting range 12 T
21. s are being counted Measurements on a shorter cable indicate only minor ringing amplitude So it is possible to use even unterminated cables but only if an input threshold is or can be adjusted safely above the ringing level Or you could use high threshold integrated circuits such as 7414 74LS14 74514 74HC series etc on the input of your counter timer or rate meter Figure 6a 16 feet of unterminated 50 cable Figure 6b The same 16 foot 500 terminated Note the 1 0V glitch that may be detected by cable Note the absence of glitches but the the counter as a second pulse pulse height is reduced to 2 8V Figure 6 Example of Ringing Reduction by 509 Termination Horizontal 25 ns Vertical 1V div To conclude most TTL signals do not need termination unless long coaxial cables are involved If you terminate the cable ringing is reduced significantly unfortunately the signal is thereby reduced to about half its original amplitude If this is a problem an intermediate solution may be the answer A 1000 terminator on a 500 cable produces only slightly larger ringing than a 50 Q terminator but the pulse amplitude is only slightly less than the original Input Considerations The input connection is far more critical than the output connection High sensitivity input requires the shortest possible connection to the PMT or to an electron multiplier Also the noise in the input signal or in the input ground connection must b
22. the detector is at ground potential or at a slight positive or negative bias to attract charged particles of given polarity A disadvantage of this technique is the need for an H V supply with a noise level below the detection capability of the detector Summary As we can see both positive and negative biasing techniques present advantages and disadvantages So if you want to improve the sensitivity and detection limits of your detector you will want to consider the constraints of a given application plus the limitations described above Then you can better evaluate and choose a particular biasing arrangement and define the necessary specifications for the H V power supplies 17 PHOTON COUNTING APPLICATIONS Two applications are discussed how to optimize photon counting and how to automate your data acquisition How to Optimize Photon Counting The best way to perform very low light level measurements is by counting individual photons with a photo multiplier tube PMT and associated counting electronics When we use the PMT for counting individual photons we are usually concerned with the background level dark count as well as the accuracy at the maximum count rates The background count rate creates the low light level limit the point at which the signal becomes smaller than the background fluctuation We can use high pulse pair resolution to improve accuracy at high count rates thus minimizing the error caused by pulse pile
23. up and the resulting loss of counts So the best PMT systems feature the lowest possible background count coupled with the highest pulse pair resolution The following two sections explain how to minimize the background count and to maximize the useful count rate How To Minimize the Background Count Select a PMT with the certified lowest dark count rate usually specified at a given temperature and bias voltage Lower the operating temperature by cooling the PMT to the lowest level that is practical CAUTION Too low a temperature may result in moisture condensation on the tube and cause problems with the high voltage bias Lower the high voltage bias so that the signal incoming photons can be comfortably detected i e the threshold is safely above the noise level but not too high To do this you need a good preamplifier The MTS 100 preamplifier is probably the best choice because it combines high sensitivity with high pulse pair resolution Minimize light leaks This suggestion sounds so obvious it seems ridiculous to even mention it But experience shows that a light leak is the most common culprit when it comes to minimizing background count For example just as a simple test if you think there is a light leak in the system try turning the room lights off and on and note the change of the background count rate 18 Maximizing the Useful Count Rate You ll need to get the minimum pulse width your system can handle
24. voltage spikes do not connect or disconnect the input with high voltage bias applied to the detector Use the following procedure to ensure virtually trouble free performance in your preamplifier e To connect the preamplifier to the detector first make the connection then turn on the high voltage To disconnect the preamplifier from the detector first turn off the high voltage then wait 5 10 seconds Then you may safely disconnect the preamplifier Discriminator Set Up The discriminator is factory set to the position that works well for the Advanced Research Instruments PMT test system However this setting may need to be changed for different systems and different bias settings Consider Figures 7 and 8 1V a CW 5 B Threshold too high region 6 Missing most of the signal 5 O ao A Top ofthe ke 2 signal pulses 2 zi t a a J 0A I g a I f l 1 im L Threshold J ae Best Setting Peak to ie peak noise Noise region level n A OV Ons Time 50ns Count rate True Count Output from the Comparator rate TI Miia Figure 7 Typical signal from the Figure 8 Relationship between the detector and the threshold setting The threshold position and resulting count output from the comparator and the rate final TTL
Download Pdf Manuals
Related Search
Related Contents
Manuel Utilisateur Tanaka QCP-121 User's Manual Touquet Infos Juillet Delta RP23601 Installation Guide Master`s Thesis campeonato sudamericano de clubes Samsung GT-C3200 Kullanıcı Klavuzu E18 Primary Mouse Cortical Cells, Cat. # N200201 INTRODUCTION Modifications for the Kenwood TS-450 Chicago Electric 67845 User's Manual Copyright © All rights reserved.
Failed to retrieve file