Endoscopes Inspection Report – Sale: 1514 Date: 11 / 08 / 2015

image

Contents

1. Weighing Scales 631 GE I Vent The unit power up ok and The display and function knobs and keys are working The ventilator verification test VVT failed due to battery pack not holding charge Test run with medical air and test lung was carried out and the unit was measuring twice the set tidal volume needs checking amp calibration Unit is in a good physical amp cosmetic condition Medical hose amp boxes of single use patient circuit included 632 Biotek Index 2 Series Powers up and the display panel and function keys are in good SPO2 Simulator working order The SpO2 simulation functions are fully functional tested with patient monitor and the results were as expected Charger cable attachments and carry case are included and the unit is in very good condition Last calibration date April 2014 635 Valleylab Force FX Powers up and completes the system self test successfully Diathermy with Display panel and output selector keys are in good working Monopolar amp Bipolar order Patient plate monitoring circuit is working The unit was Foot Pedal tested with ESU analyser at several different settings and all measured values were well within the limits The footswitches are in good working order Overall in good physical condition 635 Valleylab Force FX Powers up and completes the system self test successfully ESU with footswitches Display panel and output selector keys are in good w
2. Clear Forceps Elevation Mechanism N A Leak Test Fail Overall Condition Good but leaking slightly 102 OLYMPUS Insertion Tube Good condition TJF 260V Bending mechanism amp Angulation Control Good movement amp locks Videoscope Light Guide Cable amp Light Transmission Good 1100x11 9mm Objective Lens and Focusing Good Remote Control Switches Functional Distal End Good condition Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism Good Leak Test Fail Overall Condition Good but leaking 103 OLYMPUS Insertion Tube Very good condition but one small dent GIF X0230 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good but two small dents 1000x10 5mm Objective Lens and Focusing Lens in good condition average focusing Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good 104 OLYMPUS Insertion Tube Very poor condition sleeve has come completely off two thirds of insertion tube small dent in area with laminate still on CYF 5 Bending mechanism amp Angulation Control Down direction movement Direct View eee works but up doesn t locking mechanism is fine Scope 350x5 3mm Ligh
3. physical condition with some wear to the stands 30150 1084 Alaris P7000 Syringe The lot consists of 6 pumps All are in fair physical condition and LED indicators are working 5 of the pumps display the Pumps x 6 message requires service on start up Their selection keys clamps and syringe holders are all working fine but 1 of the 5 has a central horizontal section of the display not working The 6 pump displays the message Error ERS RAM and does not proceed to boot and this pump is also set to a language other than English 1085 Alaris P7000 Syringe Pumps x 6 The lot consists of 6 pumps All of the units are in fair physical condition all power up and LED indicators are working 1 pump is in good working order and 5 of the pumps display Requires service upon start up All 6 of their displays selection keys clamps syringe holders and backup batteries are working fine 1088 Alaris P7000 Syringe Pumps x 8 The lot consists of 8 pumps All power up and are in fair physical condition with LED indicators in working order 7 of the pumps display the message Requires service upon start up 6 of these have displays selection keys clamps syringe holders and backup batteries that are working fine The T one s backup battery does not work The 8 pump powers up but displays the message Error ER5 RAM and does not proceed to boot Its display clamp and syringe ho
4. Fair condition lamination has worn off a little EG 3830UT Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Very good transmission but 1200x13 4mm two small dents in cable Objective Lens and Focusing Processor was not available so unable to test Remote Control Switches Processor was not available so unable to test Distal End Fairly good condition Image Quality Processor was not available so unable to test Working Channel s Clear Forceps Elevation Mechanism Working Leak Test Pass Overall Condition Fair physical condition 108 FUJINON Insertion Tube Very good condition EC 450 WLS Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Very good 1600x14 2mm Objective Lens and Focusing Good Remote Control Switches Fully functional Distal End Very good condition Image Quality Good Working Channel s Clear Leak Test Leaking Overall Condition Good but leaking 109 FUJINON Insertion Tube Very good condition Unknown Bending mechanism amp Angulation Control Very good movement and Model locks Videoscope Light Guide Cable amp Light Transmission Good 1600x13 6mm Objective Lens and Focusing Objective lens is in good condition but no image is produced so focusing cannot be tes
5. Overall Condition Average very slightly leaking and missing fibres 225 FUJINON IPX7 Direct View Scope 500x5 7mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission N A Objective Lens and Focusing Good Remote Control Switches N A Distal End Good condition Image Quality Good but one fibre is out Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Fairly good 226 FUJINON Videoscope Insertion Tube Very good condition Bending mechanism amp Angulation Control Good Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Very good Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good 227 FUJINON EB 530P Videoscope 500x3 9mm Insertion Tube Good Bending mechanism amp Angulation Control Good Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Objective lens condition is fine focusing is fair Remote Control Switches Functional Distal End Good Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good Condition Report S
6. and emergency lamps is working The selection keys for auto Olympus CLV S40 manual toggle and brightness adjustment on the light source do Light Source not work The monitor has fully functioning selection keys and Olympus OTV SP1 displays average quality images The camera connects well to Camera Processor the processor and is in good condition Image quality is good with Olympus OTV but focus is poor The keyboard is working but 1 key cover is SP1H N 12E Camera missing The printers selection keys and LCD display panel are Head Sony Video working and print quality is very good The insufflator has Printer Olympus working LED displays and selection keys but has not been fully MH 317 Foot Switch tested There are plenty of spare ink ribbons and photo paper included The stack system in fair physical condition 807 Sonosite iLook 25 The ultrasound unit powers up and passes the self test The Personal Imaging Tool 29496 display screen and the function keys are working The transducer is producing images The transducer s membrane is in good condition but there is a small chip in the plastic casing around the membrane The ultrasound connects well on the stand and the backup battery is also working The castors of the trolley are moving freely Mfg August 2003 819 Fresenius Volumetric The volumetric pumps power up fine the displays and function Pumps x 2 Ivac 572 keys are wo
7. Lamp on Table with Lightlas 532 Laser with Key and Foot Switch Both the slit lamp and the laser unit are in good working order The optical system of the slit lamp is clear and the lamp and brightness adjustment are also functional The laser output was tested on thermal paper and the results were clearly visible The laser delivery device and laser fibre are in place Overall the unit is in good physical and cosmetic condition 678 Olympus Stack system Processor OTV S7 power up ok the exposure adjustment and white balance are working fine The camera head produces good image quality on the LCD monitor The light source CLV S40 powers up ok with manual and auto brightness adjustment xenon lamp and back up halogen lamp are in good working order Overall unit is in satisfactory physical condition Overall the stack is in good physical condition The lock pin on the coupler has snap off 679 Drager Fabius GS Anaesthetic System The unit power up ok and the firmware self test completed successfully Compliance and Leak Test passed The display and function knobs and keys are working The pressure gauges and flow meters are functional anti hypoxic safety features are working ok Oxygen sensor failed and flow sensor is working fine Back up battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working fine Unit is in a good physical amp cosmetic condition Software
8. Light Transmission Good but slight kinking in one 1600x13 0mm area of the light guide cable Objective Lens and Focusing Good Remote Control Switches Fully Functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Fail Overall Condition Good but leaking Some slight wear to the handle 120 OLYMPUS Insertion Tube Good condition GIF X0230 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 1000x9 5mm Objective Lens and Focusing Objective lens in good condition focusing is average Remote Control Switches Not working not yet programmed Distal End Good condition Image Quality Fairly good Working Channel s Clear Leak Test Pass Overall Condition Good 121 OLYMPUS Insertion Tube Good condition BF 240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good 543x6 3mm Objective Lens and Focusing Objective lens is in good condition focusing is average Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Fail Overall Condition Good but leaking 122 OLYMPUS Insertion Tube Very good condition GIF XQ230 Videoscope 1000x10 2mm Bending
9. The trolley is in fair condition and the backup battery is working fine 960 LEC Medical Fridge All the medical refrigerators are working and the temperature and pharmacy Fridge settings are maintained Fair physical condition consistent to x2 usage 963 Topcon XR1 Retinal The slit lamp power supply comes on but the light beam and Camera bulb is not on The brightness control on top of the unit has broken off which may be preventing the lamp from coming on The still camera adapter and Nikon D70S are working fine and in very good condition There is no memory card to fully test it The attached tonometer light is on and in good condition All the interconnecting cables are attached but there is no PC with the unit Apart for the missing PC and slit lamp not working the system is in very good physical condition Year 2006 966 Trolley with Philips LCD power up ok in good physical condition LCD Display 968 Topcon 501X Retinal The unit power up ok Function keys selector knobs and Camera indicator are responding The camera system is working with fair magnification and focusing Fixation lamp is working The trigger switch is working and Topcon 35mm camera MT 10 is attached There are two other cameras included the Nikon D200 and Mega Plus camera model 1 41 10bit both in good physical condition but not fully tested The control console is in good condition All the interconnecting cables are attached but there is no P
10. Workstation with Olympus OTV S7 Processor with Camera Head Olympus CLV S30 Light Source Sony Colour Video Printers Sony Video Cassette Recorder Sony Trinitron Monitor Sony Video The stack powers up successfully The light source is in good working order producing a good output and all selection keys are functioning The spare lamp is not working The video processor powers up and all selection keys and LED indicators are working but could not be fully tested as the camera head included is not compatible with this processor though it is in very good physical condition The monitor powers up fine and all selection keys are working The keyboard included is also not compatible with this processor so was not tested An interconnecting cable kit is included The VCR powers up and the display panel is working but was not fully tested Both the video graphic printer and the colour video printer power up an Graphic all selection keys are fully functioning but the printers were not fully tested Interconnecting cables or power cables not included The stack system is in average physical condition 48 Valleylab Force FX The Force FX powers up ok The displays and function keys are Diathermy with Foot working fine The patient plate monitoring circuit are Switch functional System was not tested with analyser Unit in good physical condition with footswitches The control console surface has worn ou
11. a compatible video processor and good image was Number 7901032 obtained The connector and cable are intact Lot is in fair physical condition 153 Storz 10Fr Graduated The rigid scope is in good physical and cosmetic condition with Ureteroscope with good light transmission and satisfactory image However there Flexible Eyepiece are signs of broken fibres 154 Storz 9 5Fr The rigid scope is in good physical and cosmetic condition with Ureteroscope good light transmission and satisfactory image However there are signs of broken fibres and the image is slightly blurred 169 Olympus MH 977 Pig The video cable was tested with a compatible stack system It Tail connects fine and is in good physical condition but the image cuts out intermittently indicating damage to the interior of the cable 175 Olympus MH 977 Pig The video cable was tested with a compatible stack system It Tail works well producing good images The cable is in good physical condition 187 Storz 70 Rigid Scope The rigid scope is in good physical and cosmetic condition with with Light Lead good light transmission and very clear images Fibre optic cable is also included 188 Storz 70 Rigid Scope The rigid scope is in good physical and cosmetic condition with satisfactory image the image is slightly blurred Fibre optic with Light Lead cable is also included 201 Olympus Evis OVC 200 Endoscope Camera Head The camera head con
12. is torn a little along the seam under one arm and another vest piece has one shoulder where the Velcro doesn t fasten One of the skirt pieces has the inner lining tearing off one corner The other 25 pieces are fine There are some minor cosmetic markings on some pieces 1010 Alaris P7000 Syringe The lot consists of 6 pumps All are in fair physical condition Pumps x 6 and LED indicators are working 5 of the pumps power up fine but all 5 display requires service after start up Their displays function keys clamps and syringe holders are all working fine The 6 pump has a functional display and the syringe holder and clamp are fine but the messages Error ER8 Hardware and Error ER5 RAM are displayed at start up and do not proceed to boot Backup batteries were not tested 1058 Wardray X Ray Lead The lot consists of 2 vests and 1 thyroid collar The items are Vests x 2 little worn but the lead inside seems to be intact 1061 Braun Compact The trolley is in fair physical condition with the drawers CP EM 2 Trolley compartments and holders in place Mfg 2003 1069 Marquette Capnostat The item is in very good physical condition but was not tested Co2 Sensor 150824 1071 Bird Mark 8 x3 and_ The dials are all adjustable but the respirators were not fully Bird Mark 7A tested and were not tested with medical air O2 hoses are Respirators x 1 included The castors are moving freely The units are in fair M178 M177 M181
13. is worn Objective Lens and Focusing Lens is slightly clouded focusing is fairly good Remote Control Switches N A Distal End Worn Image Quality Slightly cloudy Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Valve is faulty so cannot test Overall Condition Fair very worn 197 OLYMPUS Insertion Tube Good condition OSF 3 Bending mechanism amp Angulation Control Good movement and locks Direct View Light Guide Cable amp Light Transmission Good transmission cable is Scope slightly worn 600x12 6mm Objective Lens and Focusing Very good Remote Control Switches N A No Carry Case Distal End Good condition Image Quality Very good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Valve is faulty so cannot test Overall Condition Fairly good 202 RS MEDICAL Insertion Tube Good condition Demonstration Bending mechanism amp Angulation Control N A Device Direct View 699x2 0mm Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good Remote Control Switches N A Distal End Good Image Quality Fair Working Channel s N A Forceps Elevation Mechanism N A Leak Test N A Overall Condition Good 207 OLYMPUS Insertion Tube Lamination is a little worn BF 10 Bending mechanism amp Angulation Contro
14. mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Objective lens in good condition focusing is fairly good Remote Control Switches One is very stiff and only spontaneously works Other 3 either haven t yet been programmed or do not work Distal End Good condition Image Quality Fair Working Channel s Clear Leak Test Pass Overall Condition Good 123 OLYMPUS GIF XQ240 Videoscope 1000x10 2mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Objective lens in good condition focusing is fair Remote Control Switches Fully Functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Very good 124 OLYMPUS BF 240 Videoscope 545x6 2mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Good Remote Control Switches Fully Functional Distal End Good condition Image Quality Very good Working Channel s Clear Leak Test Pass Overall Condition Very good 125 OLYMPUS I
15. processor light source monitor and printer all power Trolley with Sony up ok and all selection keys and LED indicators are in good CRT Monitor OTVS7 working order Both the main and emergency lamps for the Camera Controller light source are working The system could not be fully tested CLVS40 Light Source as a camera head is not included The keyboard is in good and Colour Video condition but was not tested The stack system is in average Printer physical condition 758 Stainless Steel Trolley Good physical condition 762 Olympus Stack The video processor light source monitor and the digital Trolley with Sony capture unit all power up successfully and all LED indicators CRT Monitor Digital and selection keys are in good working order Both the main Capture Unit OTVS7 and emergency lamps for the light source are working The Camera Controller stack system could not be fully tested as a camera head is not CLVS40 Light Source included The keyboard remote control and the remote control unit were not tested The stack system is in average physical condition 769 Eschmann VP35 Unit power up with good suction power Regulator and vacuum Suction Unit gauge working fine Good overall physical condition 774 Cambridge Heart Unit power ON ok and goes through the regular boot up Exercise System CH sequence alright ECG module is present but ECG leads are 2000 missing and hence cannot perform tests on the treadmill and also cannot test the ECG
16. section Unit is overall in a good physical condition 779 Allen and Hanburys_ Aged but in good condition not tested with battery Ophthalmoscopes x 2 787 Reichert Tono Plen The unit is presented in good physical condition but it is not XL Tonometer powering up Battery needed 793 Datex Tec 7 Isoflurane Label indicating fault and has been replaced average physical Vaporiser condition 794 Datex Tec 7 Isoflurane Label indicating fault and has been replaced average physical Vaporiser condition 795 Datex Tec 7 Label indicating fault and has been replaced average physical Sevoflurane Vaporiser condition 797 Topcon 3D OCT 1000 The unit powers up but shows error 21 during start up Display with Spectroscope No panel function keys joystick control and the motorised chin Table 203028 rest adjustment are all in good working order The optical system is functional and produces clear images However the system does not capture images when the trigger button is activated Overall the unit is in good physical condition The interconnecting cables for the spectroscope are not included 802 Olympus Stack The stack system and all the associated equipment power up System To Include successfully The video processor has fully functioning Olympus OEV 191 selection keys and LED indicators and white balance Monitor Olympus completed successfully The light source with both the main UHI 3 Insufflator
17. source powers up and produces a CV 240 with Pig Tail good output when tested with a compatible endoscope Its Connector Olympus pump and selection keys are also fully functioning The CLV U20 Light keyboard is working fine The Pentax video system powers up Source and Pentax and all selection keys LED indicators and the pump are EPM3500 VCC032 working White balance completes successfully and a good light output is produced The electrosurgical unit powers up and LEDs and all selection keys are in good working order Its footswitch connects fine but was not fully tested An interconnecting cable kit is included but was not tested The video cable and endoscopy footswitch were not tested The stack is in fair physical condition 79 Maquet Servo I The Servo I power up ok and the pre use check was carried out Ventilator with the oxygen and flow sensor failed to calibrated 100 02 not used The LCD touch screen selector switches and function keys are working fine The battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working fine Unit is in a good physical amp cosmetic condition Operating time 42346Hrs system software version V2 00 03 Medical hose amp user manual included 1 Datex Ohmeda The Cardiocap 5 Anaesthesia monitor power up ok self test Cardiocap 5 Patient completed Display function buttons and the rotary knob are in Monitors working order ECG trunk ca
18. testing ARUB0013 Faulty was conducted Error when power on 437 5x Blood Temperature The modules are in good physical and cosmetic condition Part Monitor Modules for number 651992 Fresenius Dialysis Machine 451 Seca Baby Scales x6 The baby scales are in fair physical condition but not fully and Seca Stand on tested Most of them are without batteries The stands scale Scales x 2 comes on and are functional 454 CPL Orralight Laser The unit does not power up Includes Unit with Keys and Handpieces 4x 10mm 450 1200nm Multi handpiece 40x 16mm 650 1200nm HR Epilation handpiece 40x8mm 650 1200nm HR Epilation handpiece 4x Light guides glass inserts for handpieces 2x Safety goggles Key Overall the unit and the accessories are in reasonable condition 475 Philips M3046A All of the monitors power up ok and system self test completed Monitors x11 successfully The LCD and function keys are in good working order The unit includes multi parameter module M3000A with SpO2 NIBP ECG and Temp The units have not been fully tested but are in fair physical condition There is x1 with damaged handle x1 with battery compartment cover missing amp x1 with loose front panel Year of manufacture 2001 No patient lead or probe included 489 Seca CT6 PI ECG Unit passed power up test ECG monitor fully functional tested using an ECG simulator and printer is in good working order Unit also operational on battery Overall the
19. the syringe arm is fully functional No power cable but unit work on batteries Overall the lot is in good physical and cosmetic condition 31 Datex Ohmeda Cardiocap 5 x5 amp Light Monitors x2 with box of accessories The Cardiocap 5 Anaesthesia monitors and Light monitors powers up ok self test completed Display function buttons and the rotary knob are in working order ECG trunk cable x1 NIBP hose x1 and SpO2 lead included The unit was tested with a patient simulator and all the tested parameters NIBP ECG Temp amp SpO2 are working fine The CO2 gas parameter is responding and AA monitoring activated but not tested with anesthetic agent All monitors are in good physical condition The Cardiocap 5s are of year 2001 2003 amp 2005 The Light monitors are year 2003 The following accessories are included ECG trunk cable x 8 NIBP Hose x 11 SpO2 finger probe complete x6 FlexSat Sensor x4 and power cable Back up battery pack are working ok 38 Capricorn Centrifuge The unit power up ok the rotor test spin and indicators are working fine Timer and lid lock are functional and the unit is in good physical condition 43 B Braun Perfusor FM Syringe Pumps x 15 All 15 Perfusor powers up and self test completed Unit is powering on battery and mains as well Lot is in fair physical condition Only one unit from the lot was tested with syringe and all are working fine 44 Olympus Stack
20. up ELI250 ECG Machine not working 3109EW201 995 Philips Avalon FM30 Unit power up but goes off immediately unable to remain on Monitor 997 Laerdal LSU Suction Power supply not included and battery is not present so unit Unit 25184 could not be fully tested The suction unit is in good physical condition with the canisters and tubing in place Mfg March 2000 1000 Maquet Servo I The system powers up fine O2 and air hoses are included but Mobile Ventilator the expiratory block and battery pack are missing LCD display is in good working order Selection knob is working and menus are accessible The item was not fully tested and was not tested with medial air Unit is in fairly good physical condition System version 4 S W version 4 00 05 Operating hours 42 092 1001 Maquet Servo I The unit does not power up Air and O2 hoses are included but Mobile Ventilator the battery pack and expiratory block are missing Unit is in average physical condition with some scuff marks a small chip to a drawer and the covers for the wheel joins missing 1009 Qty of Lead The lot consists of 10 skirt pieces 11 vest pieces 1 full length Aprons Jackets wrap around apron 2 pairs of foot guards and 4 thyroid collars One of the thyroid collars is a little kinked inside One of the foot guards has lead which is badly damaged inside the foot section but the leg part is fine One of the vest pieces
21. working order The stack system is in CLV V40 Light fair physical condition with a number of cosmetic markings Source 305 Mortara ELI 250 Unit passed power up test ECG monitor fully functional tested ECG Recorder using an ECG simulator and printer is in good working order Unit also operational on battery Overall the unit is in good physical condition 309 GE Vingmed Vivid 5 Unit powers on OK but does not go through its boot up process Unit is overall in an average physical condition 331 Wandsworth Both lamps are working fine and in good physical condition Examination Lights x 2 R516612 332 Stack System with All units on the stack power up ok The processor was tested Sony Trinitron with a compatible camera head and good image was produced Monitor Olympus on the Sony monitor The functions keys white balance and OTV S6 Advanced enhance setting are working fine The Keymed light source is Digimulti Processor fully functional Stack system is in satisfactory physical Keymed Light Source condition 337 Maquet Servo I The Servo I power up ok and the pre use check was carried out Ventilator with the oxygen and flow sensor failed to calibrated 100 02 not used The LCD touch screen selector switches and function keys are working fine The battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working fine Unit is in a good physical amp cosmetic cond
22. C The motorised table height adjustment is working fine System is in an average physical condition but not fully tested 978 Maquet Servo I The unit power up ok display and functions menus accessible Mobile Ventilator and working The unit was not tested with medical gases The back up battery packs and the exhalation valve are missing Operating hours of the vent 43323hrs Software version V4 00 04 Unit in fair physical condition Air amp O2 hose included 979 Maquet Servo I The unit power up ok display and functions menus accessible Mobile Ventilator and working The unit was not tested with medical gases The back up battery packs and the exhalation valve are missing Operating hours of the vent 42526hrs Software version V4 00 04 Unit in fair physical condition with chipped lower drawer 988 Alaris P7000 Syringe The lot comprises 6 pumps All are in fair physical condition Pumps x 6 The LED indicators are working 5 of the units power up fine four of which display requires service after start up The syringe holders and clamps are moving freely All the function keys and the display are working well The backup batteries are working The 6 pump has a functional display clamp and syringe holder but displays the messages Error ER8 Hardware and Error ER5 RAM at start up and does not proceed to boot 993 Mortara Instrument The unit is in good physical condition but failed to power
23. Endoscopes Inspection Report Sale 1514 Date 11 08 2015 Condition Reports for other items at the end of the Scopes Report These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Condition 100 OLYMPUS Insertion Tube Fair condition lamination has worn off slightly one GIF 0X230 small bubble like area of damage at 500mm markings are slightly discoloured NALPAEEORE Bending mechanism amp Angulation Control Good movement amp locks Mee eau Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Objective lens in good condition average focusing Remote Control Switches Functional Distal End Good condition Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Fair 101 OLYMPUS Insertion Tube Good condition one very small dent CF 240L Bending mechanism amp Angulation Control Good movement amp locks Videoscope Light Guide Cable amp Light Transmission Good 1600x12 8mm Objective Lens and Focusing Good Remote Control Switches Three out of four functional fourth probably not programmed yet Distal End Good condition Image Quality Good image Working Channel s
24. LYMPUS Insertion Tube Good condition CF 240DL Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good but one small dent on 1600x13 3mm light guide cable Objective Lens and Focusing Fairly good No Carry Case Remote Control Switches Fully functional Distal End Good condition Image Quality Very good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Fail Overall Condition Good but leaking 183 OLYMPUS Insertion Tube Good condition but one dent GIF XQ240 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good but one small dent on 1000x10 1mm light guide cable Objective Lens and Focusing Objective lens in good condition focusing is fairly good No Carry Case Remote Control Switches Fully functional Distal End Good condition Image Quality Very good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good 184 OLYMPUS Insertion Tube Good condition CYF 4 Bending mechanism amp Angulation Control Good movement and locks Direct View Light Guide Cable amp Light Transmission Good but one small dent on Scope light guide cable 372x5 7 Objective Lens and Focusing Good Remote Control Switches N A Distal End Goo
25. The electronic part of the unit appears to be working ok The system has not been tested with medical gases The unit is complete with flow sensor and O2 sensor but calibration not checked Vent operating hrs 1145 Battery 90461 1hrs mixer 59868 Software version 4 30 Unit in good physical condition 884 Spacelab Quest ECG and Treadmill are both in full working condition Keyboard monitor and trolley castors are in good condition ECG checks carried out using ECG simulator Mains power cable is missing for both the equipment ECG amp Treadmill There is a slight damage to zip drive reader could not verify the functioning of zip drive due to absence of zip drive discs Overall the unit is in good working condition 905 Assorted Metal Waste Fair overall condition Bins x 9 922 BCI SPO2 Monitor x Both SpO2 monitors are working fine The probes and 2 with Leads measurement are working ok One of the unit s graphic waves Mitsubishi P91 Video is not working but the values are displayed The B7W printer is Printer in good working order 953 Hill Rom Transport The incubator power up ok and the temperature selector keys Incubator T1500 and numerical displays are working fine The air temperature mode was tested and the set value was maintained accordingly The ventilator is slightly dated and in and average physical condition Not tested with medical gas MiniXO I oxygen monitor needs battery replacement
26. Under door is fine The fridge is very clean internally and all shelves counter Fridge and the drawer are in place The exterior has several areas where the paint has scratched off and some dents 1160 GE T2100 Treadmill The treadmill power up ok but not fully tested with a stress No Workstation ECG system The power cable is included but not the 41051 01 communication cable to the Stress ECG system 1170 Stack system with The light source power up ok xenon lamp working and all light Olympus CLV U20 output modes ok The monitor function keys are not working CO2 Insufflator 91 properly The insufflator comes on but not fully tested Lot is and Sony monitor below average physical condition 1199 Fujifilm FCR Capsula The unit power up and system initialization completed system XL II not fully tested Manufacture year 2007 Application software for Hospital V2 3 Model CR IR359 Unit is in good physical condition 1217 Set of Drager The items are as new Brackets Long x 2 Mfg Drager brackets 2012 SN01120611104182 Triumph mounts 2011 and Short Monitor Mounts x 18 G25758 00
27. ale 1514 Date 11 08 2015 These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Report 1 Sonosite iLook Unit powers up OK It is in good physical condition Battery is in working order Touch screen calibrated OK Image quality is OK linear probe is fully functional Overall the unit is in good physical condition 7 Topcon TRC 50X The unit power up ok Function keys selector knobs and Retinal Camera indicator are responding The camera system is working with fair magnification and focusing Fixation lamp is working The trigger switch is working and the attached 35mm camera is activated The knob for the camera selector is cracked and the membrane key for Barrier amp Exciter has worn out The video connectors are there but video camera itself is missing The motorised table height adjustment is working but one of the castor wheels has broken off System is in an average physical condition 8 Olympus Stack The stack and all of the associated items power up successfully Trolley with Sony The video processor is in good working order with its LED Trinitron Monitor indicators and all selection keys functioning The light source Sony Video Graphic produces a good output and its selection keys pump and LED Printer Olympus CV indicators are al
28. ay and function knobs and keys are working The pressure gauges and flow meters are functional anti hypoxic safety features are working ok Oxygen and flow sensor are working fine Back up battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working but the values affected by the leak in the system Unit is in a good physical amp cosmetic condition Software version 1 39 1 Medical hose amp user manual included Date of manufacture 2001 612 Philips Pagewriter The Trim II power up ok system self test completed The Trim Il ECG display and keyboard are in good working order All the menus Recorder with Leads 18241 Serial Number are accessible and good display of ECG traces was observed when tested with an ECG simulator However printout was not USN0716672 possible since the battery is not working System is in good physical condition but will need battery replacement 613 Hitachi Sumi Unit powers up OK It is in good physical condition The screen tracker ball keyboard and TGC sliders are in full working order A little damage on the Freeze button but it is fully functional No mains power cable with the unit Following probe tested OK along with the unit e 3 5MHz good physical condition no crystal dropouts good image quality Overall the unit is in good physical condition 614 Valleylab Force II The ESU power up ok The d
29. ble x1 NIBP hose x1 and SpO2 lead included The unit was tested with a patient simulator and all the tested parameters NIBP ECG Temp amp SpO2 are working fine The CO2 gas parameter is responding but the AA monitoring not activated Back up battery pack ok The monitor is in good physical condition Year 2001 83 Sonosite 180 Plus The ultrasound powers up and passes the self test however the Hand Held screen quickly progresses to an error screen displaying Ultrasound with CW P and C60 5 2MHz C 0x0000010 1296 Transducer and D 0x00000000 0 L38 10 5MHz Linear preventing any further examination Transducer The display is in good working order There are two transducers included The linear one Mfg Sep 2002 is in good physical condition but the membrane and its surrounding area is cosmetically stained with ink and is also slightly discoloured The convex transducer Mfg Nov 2001 is in good physical and cosmetic condition Both the main and spare batteries are working Mfg August 2002 87 Graseby Omnifuse Power up LCD selector knob and function keys are working Syringe Pumps x 4 fine Not fully tested but in good physical condition 99 Micro Spirometers This lot contains 20 micro spirometers without any turbines Overall they are all in a good physical condition 152 Olympus OTV S6 The camera head OTV S6 with coupler AR T12SA was tested Camera Head Serial with
30. d Focusing Objective lens in good condition focusing is average Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Fairly good 115 OLYMPUS PCF 240S Videoscope Insertion Tube Good condition but markings are slightly discoloured Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good transmission one dent 900x12 8mm on light guide cable Objective Lens and Focusing Objective lens in good condition average focusing Remote Control Switches Two out of four working well one is very stiff but working the last has probably not yet been programmed Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good 116 OLYMPUS GIF Q260 Videoscope 1000x10 6mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing No Image Remote Control Switches Two out of four working one not working and one probably not programmed yet Distal End Good condition Image Quality No image produced Working Channel s Clear Leak Test Pass Overall Condition Good physical c
31. d condition Image Quality Very good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good 194 OLYMPUS Insertion Tube Good condition OSF 3 Bending mechanism amp Angulation Control Very good movement and Direct View Scope 600x12 4mm No Carry Case locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good Remote Control Switches N A Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good but slightly physically worn 195 OLYMPUS Insertion Tube Very good condition OSF 3 Bending mechanism amp Angulation Control Very good movement and Direct View bi Scope Light Guide Cable amp Light Transmission Very good 600sx12 6mm Objective Lens and Focusing Good Remote Control Switches N A No Carry case Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Valve is faulty so unable to test Overall Condition Good 196 OLYMPUS Insertion Tube Laminate has completely worn off GIF XQ20 Bending mechanism amp Angulation Control Good movement and locks Direct View Scope 1000x10 3mm No Carry Case Light Guide Cable amp Light Transmission Good transmission but cable
32. ection for it CV 240 VMA061 has been removed from the stack trolley The video cable was not tested The stack system is in fair physical condition with some minor cosmetic markings and scratches 63 5 x Drager Vamos Lot of 5 items All lot passed Power On Self test All the adapters are in good working order Only 2 adapters present The screen keypad and rotary knob are fully functional None of the parameters were tested No power cable and equipment does not operate on battery Overall the lot is in good physical and cosmetic condition 67 Olympus UES30 The ESU power up ok indicators numerical displays and Diathermy with function keys are working fine The modes selector is Olympus MH 551 functional The system has not been tested with an ESU Foot Switch analyser but appears to be in good overall condition Footswitch included No accessories with lot 70 Valleylab Force 2 The ESU power up ok The displays and function keys are Diathermy with working fine The patient plate monitoring circuit are Bipolar Foot Switches functional System was not tested with analyser Unit in good physical condition with bipolar footswitch 78 Procedures Evis The stack system powers up successfully but was not fully Trolley TC E1 with tested as a monitor in not included The video processor is in Olympus PSD 20 with good working order with the selection keys and LEDs all Foot Switch Olympus functioning fine The light
33. erall Condition Average leaking and dysfunctional lock 112 OLYMPUS Insertion Tube Good condition GIF H260 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 1000x11 0mm Objective Lens and Focusing Objective lens is in good condition focusing is fairly good Remote Control Switches Three out of four switches are responding the fourth probably hasn t yet been programmed Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good 113 OLYMPUS GIF Q260 Videoscope 1000x10 2mm Insertion Tube Good overall two small dents Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Fail Overall Condition Good but leaking 114 OLYMPUS CF Q260DL Videoscope 1600x13 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good but up direction movement does not work Locking mechanisms are good Light Guide Cable amp Light Transmission Small area of damage to surface of light guide cable good transmission Objective Lens an
34. he unit is in good physical and cosmetic condition 267 AMS Greenlight HPS The unit is in good physical and cosmetic condition with key Laser System with and footswitch in place Function check could not be carried Key and Goggles out as the unit requires a specific power socket Includes 8x safety goggles 268 Drager Cato Unit passed power on self test Display keypad and rotary Anaesthetic Machine knob are in good working order Unit was not tested for its ventilation or anaesthetic functions Overall the unit is in good physical and cosmetic condition 278 Mortara ELI 250 Unit passed power up test ECG monitor fully functional tested ECG Recorder using an ECG simulator and printer is in good working order Unit also operational on battery Overall the unit is in good physical condition 279 Mortara ELI 250 Unit passed power up test ECG monitor fully functional tested ECG Recorder using an ECG simulator and printer is in good working order Unit also operational on battery Overall the unit is in good physical condition 286 Kodak Dry View 6800 The printer powers up and the LCD display and touchscreen is Laser Imager in good working order The menus are accessible Two of three trays have cartridges and were calibrated however one of these did not complete successfully displaying the message Dmax Outside Target A test print was conducted which completed fine The alarm system is also working A safety manual is included The unit i
35. ht Transmission Cable is in fair condition transmission is good Objective Lens and Focusing Good Remote Control Switches Fully Functional Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good 137 PENTAX ECY 1570 Videoscope 400x5 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Cable in fair condition transmission is good Objective Lens and Focusing Lens is in good condition focusing is fair Remote Control Switches Fully functional Distal End Fairly good condition Image Quality Fair Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good 181 OLYMPUS TJF 240 Insertion Tube Very good Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Very good 1200x12 8mm Objective Lens and Focusing Good Remote Control Switches Fully functional No Carry Case Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism Working Leak Test Fail but only marginally leaking Overall Condition Very good but leaking very slightly 182 O
36. isplays and function keys are Diathermy working fine The patient plate monitoring circuit are functional System was not tested with analyser Unit in good physical condition but with a crack on the front panel no footswitch 620 Blease Isoflurane All in fair physical condition and serviced on the 01 2013 Vapouriser x 4 01 2015 08 1009 amp 07 2010 622 Fluke UW4 Powers up and the display panel and function keys are in good Ultrasound working order The sensor is also functional and the Wattmeter measurement readings are present on the screen the measurements have not been verified Power adaptor is not included but the battery is working Overall the unit is in good condition 628 Microlab Spirometer The Graseby anaesthesia pump is in good working order but the Serial Number 17377 display has a patch of pixel out The spirometer power up but Graseby 3500 the display remains blank No further test was carried out The Anaesthesia Pump Dinamap power up ok The numerical display and function 70111 Dinamap Vital keys are working fine The NIBP pump comes and tested ok Signs Monitor 8100 All in average physical condition 810154291 629 Box of Welch Allyn Braun tympanic thermometers below average physical Thermometers x 5 condition with missing batteries and some covers Not all checked 630 Qty of Assorted Seca All in fair physical condition no battery and not fully tested and Tanita Baby Model 724 amp 835
37. ition Operating time 41830Hrs system software version V2 01 01 Medical hose included 339 Valleylab Force 2 The diathermy powers up and the display and all selection keys Diathermy No Foot are in good working order No leads or footswitches are Switch included so the unit was not fully tested Item is in fair physical condition with two of the rubber legs missing 355 GE Dash 4000 patient All 4 monitor powers up ok self test completed Display monitors x4 function buttons and the rotary knob are in working order The unit was tested with a patient simulator and all the tested parameters NIBP ECG amp SpO2 are working fine All monitors are in good physical condition with some scratches on the casing Back up battery pack are working ok x6 battery packs working x1 missing and 1 with error The Siemens patient monitor power up ok but not fully tested 374 Olympus Stack The stack and all of the associated items power up successfully Trolley with Sony The video processor s selection keys and the LED indicators Monitor Olympus are all working and white balance completes successfully The Evis Procedures Cart light source emits a good output and again the selection keys with Olympus CV 240 pump and LEDs are all in good working order The monitor Camera Controller produces clear images with all of the selection keys Olympus CLV U40 functioning The electrosurgical unit has fully functional Light Source selec
38. l Good movement and locks Direct View Light Guide Cable amp Light Transmission Fair Scope Objective Lens and Focusing Good 540x5 0mm Remote Control Switches N A Distal End Good condition No Carry Case Image Quality Fair but two fibres are out Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Fail Overall Condition Fairly good but leaking 210 OLYMPUS Insertion Tube Very good condition LF TP Bending mechanism amp Angulation Control Very good movement Direct View Light Guide Cable amp Light Transmission No light guide cable Scope transmission is fine 595x5 1mm Objective Lens and Focusing Good Remote Control Switches N A Distal End Very good condition Image Quality Many fibres are out Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good if optical fibres weren t out 211 OLYMPUS Insertion Tube Good condition CYF 2 Bending mechanism amp Angulation Control Good movement and locks Direct View Light Guide Cable amp Light Transmission Very good Scope 373x5 3mm Objective Lens and Focusing Objective lens is in good condition but focus does not adjust Remote Control Switches N A Distal End Good condition Image Quality Good but 10 20 fibres are out Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Fail
39. lder are working fine 1092 Drager Oxylog 2000 Ventilator The ventilator powers up and passes the self test fine The LCD panel alarms and backup battery are all in good working order O2 and inspiratory hoses are included but flow sensor is not The item was not fully tested and was not tested with medial air The item is in good physical condition with some very minor scratches to the paintwork at the top S W version 3 01 1098 Drager Oxylog 2000 Ventilator The ventilator powers up and passes the self test fine The LCD panel alarms and backup battery are all in good working order O2 and inspiratory hoses are included but flow sensor is not The item was not fully tested and was not tested with medial air The item is in good physical condition S W version 3 01 1107 Head lamp The power supply is working and the lamp as well The head gear is in fair condition with the mirror adjustment knob missing on one side 1110 Syringe pumps x6 All the pumps power up ok displays and functions keys ok There are different errors that appear after switching on such as service required Battery error and hardware error Only one of the units is without error Otherwise in reasonable physical condition 1127 LEC Medical The fridge powers up and the LED temperature display and Pharmacy selection keys are working The magnetic mechanism of the Refrigerator
40. nding mechanism amp Angulation Control Good movement and locks Direct View Light Guide Cable amp Light Transmission Good Scope Objective Lens and Focusing Good 1600x14 0mm Remote Control Switches N A Distal End Good Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good 129 OLYMPUS Insertion Tube Very good condition but distal third remains slightly angled JF 1T20 Direct View Scope 1100x11 4mm Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good Remote Control Switches N A Distal End Good condition Image Quality Good but many fibres are out Working Channel s Clear Forceps Elevation Mechanism Working Leak Test Pass Overall Condition Good 130 OLYMPUS Insertion Tube Good GIF K20 Bending mechanism amp Angulation Control Good movement and locks Direct View Light Guide Cable amp Light Transmission Good Scope Objective Lens and Focusing Good 1000x12 4mm Remote Control Switches N A Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism Working Leak Test Pass Overall Condition Good 131 PENTAX Insertion Tube Some laminate has come off the distal quarter of the insertion tube FNL 10RP3 Direc
41. nects well and produces a fair quality image The remote control switches are in good working order The item is in good physical condition 212 213 214 215 216 217 218 219 220 221 222 223 405 406 407 409 410 411 531 532 533 534 781 782 783 784 Various Rigid Scopes All of the rigid scopes are in good physical and cosmetic condition with good light transmission and very clear images 263 GE Vingmed System 5 Power on test passed Software version 1 7 3 Unit is in good physical condition monitor and little display LCD are in good working condition Tracker ball keyboard and TGC sliders fully functional 2D PW M CW CFM modes functional tested ok Footswitch is working intermittently Lamp on keyboard is not functional ECG is in good working order and was tested with a simulator Following probes were tested along with the unit e IC good physical condition no crystal dropouts good image quality e 5A good physical condition no crystal dropouts good image quality e 2 0 MHz CWD good physical condition good working order Overall the unit is in good physical condition 264 Drager Fabius Unit passed power on self test Display keypad and rotary Anaesthetic Machine knob are in good working order Select function for the rotary knob is working intermittently Unit was not tested for its ventilation or anaesthetic functions Overall t
42. nsertion Tube Very good condition CYF 3 Bending mechanism amp Angulation Control Up movement is good down Direct View is a little restricted locking mechanism is fine Scope Light Guide Cable amp Light Transmission Good 370x5 4mm Objective Lens and Focusing Good Remote Control Switches N A Distal End Good condition Image Quality Good but two fibres are out Working Channel s Blocked Leak Test Pass Overall Condition Good but blocked channel 126 OLYMPUS Insertion Tube Good condition BF 1T240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good 540x6 5mm Objective Lens and Focusing Objective lens in good condition focusing is average Remote Control Switches Fully functional Distal End Good condition Image Quality Fairly good Working Channel s Clear Leak Test Pass Overall Condition Good 127 Unknown Insertion Tube Fair Condition Make and Model Bending mechanism amp Angulation Control Good movement Direct View Light Guide Cable amp Light Transmission N A Scope Objective Lens and Focusing Good 304x3 8 S Remote Control Switches N A Distal End Sleeve has come off the end 5mm or so Image Quality Good Working Channel s N A Leak Test Not tested Overall Condition Fair 128 OLYMPUS Insertion Tube Good CF 10L Be
43. ondition but no image 117 OLYMPUS CF H260DL Videoscope 1600x15 1mm Insertion Tube Good condition except the last inch or so of the distal end see below Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Objective lens condition is fine focusing is average Remote Control Switches three out of four are working the fourth probably hasn t yet been programmed Distal End Good condition but the sleeve has come off completely at the very end inch Image Quality A little grainy Working Channel s Clear Leak Test Fail Overall Condition Average 118 OLYMPUS Insertion Tube Good condition CF H260DL Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 1600x13 9mm Objective Lens and Focusing Objective lens condition is fine focusing is fair Remote Control Switches Three out of four are functioning the fourth probably hasn t been programmed yet Distal End Good condition Image Quality Good but slightly grainy Working Channel s Clear Leak Test Pass Overall Condition Fair 119 OLYMPUS Insertion Tube Good condition CF Q260DL Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp
44. orking order Patient plate monitoring circuit is working The unit was tested with ESU analyser at several different settings and all measured values were well within the limits The footswitches are in good working order Unit is in good physical condition 649 Stryker Patient The hydraulic lift trendelenburg and back rest adjustments in Trolley with Mattress good working order Cushions are in good condition The and Attachment casters and brakes are functioning fine 662 EDL Industries Lamp in good working order and fair physical condition Examination Lamp 664 Hanaulux Exam Lamp in good working order and fair physical condition Lamp wall mounted 674 Drager Fabius GS The unit power up ok and the firmware self test completed Anaesthetic System successfully System leak test failed The display and function knobs and keys are working The pressure gauges and flow meters are functional anti hypoxic safety features are working ok Oxygen sensor failed and flow sensor is working fine Back up battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working fine Unit is in a good physical amp cosmetic condition Software version 3 11 Medical hose included Date of manufacture 2001 675 Aloka Prosound 5000 Unit power on OK but does not go through its boot up process Unit is overall in good physical condition 676 Haag Streit Bern 900 Slit
45. ponsive Distal End Fair condition Image Quality Fair Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Fairly good 134 PENTAX Insertion Tube Good condition ECY 1570 2 0 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 400x5 5mm Objective Lens and Focusing Lens is in good condition focusing is fairly good Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good 135 PENTAX Insertion Tube Fair condition slightly worn ECY 1570 2 0 Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Cable is in fair condition one 400x5 6mm small dent Transmission is good Objective Lens and Focusing Good Remote Control Switches Fully functional Distal End Good condition Image Quality Fairly good Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Fairly good 136 PENTAX ECY 1570 2 0 Videoscope 400x5 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Lig
46. rking ok in good condition but not fully tested Volumtric Pumps x 1 VAC 572 pump is not working with error message when Astotherm Plus Fluid power up Fluid warmer power up ok Warmer 824 Cardio Smart ECG The unit power up and the display and control console are Machine No Leads working fine The printer is functional but no leads to fully test 13861 it 827 Sonosite iLook Unit powers up OK It is in good physical condition Battery is in working order Touch screen calibrated OK Image quality is OK linear probe is fully functional but slightly damaged Overall the unit is in good physical condition 830 HP Pagewriter XLI Power up but with battery error 6503 with ECG Leads 831 Maquet Birthing The unit is in average physical condition The cushions are fine Couch but the frame has some rust patches on the base Not fully tested for the electrical function 837 Syringe Pumps amp The syringe pumps are in good condition and power up ok Nebuliser Nebuliser missing battery and cover 855 Penlon Wall mounted The PRIMA 100 is in fair condition and intact Not tested with Anaesthetic unit medical gas 857 Marsden Digital Baby Needs repair Scale 877 Hare Paediatric Fair condition consistent with usage Traction Splint Dynamed Paediatric Immobilisation and Transport System 881 X Ray lead apron x7 Aged but intact 882 Drager Primus The unit power up ok display and function menus are working Anaesthetic Machine fine
47. s in good physical condition 287 Daray Mobile Lamps Only one of the lights is working in average physical x2 condition 289 Acuson Sequoia 512 Power on test passed FPP ver 4 4 Unit is in good physical condition monitor and little display LCD are in good working condition Tracker ball keyboard and TGC sliders fully functional 2D M CD PW modes functional tested ok Following probes were tested along with the unit e 6L3 good physical condition no crystal dropouts good image quality e 6C2 tape on the probe cable possible damage cut image quality is good but there are a few crystal dropouts e 4V1 good physical condition no crystal dropouts good image quality Overall the unit is in good physical condition 291 Olympus WM 60 The stack and all of the associated items power up successfully Mobile Workstation The video processor s selection keys and the LED indicators with Sony Trinitron are all working and white balance completes successfully The Monitor and Sony light source emits a good output and again the selection keys Video Printer pump and LEDs are all in good working order The monitor Olympus WM 60 produces clear images with only some of the selection keys Mobile Workstation functioning The printer has fully functioning selection keys with OlympusCV 240 and produces good quality prints An interconnecting cable kit Processor Olympus is included and is in good
48. so all working The monitor and its selection 240 Camera keys are working and it produces clear images when tested with Processor Olympus a compatible endoscope The printer is in good working order CLV U40 Light and produces good quality prints The keyboard is functional Source Olympus EV but its touchpad and trackball were not tested The ultrasound M30 Endoscopic centre powers up ok but a transducer was not included so could Ultrasound Centre not be fully tested A footswitch is also included but was not tested The stack system is in fair physical condition 10 4x Drager Vamos Lot of 4 items All lot passed Power On Self Test All the adapters are in good working order The screen keypad and rotary knob are fully functional None of the parameters were tested No power cable and equipment does not operate on battery Overall the lot is in good physical and cosmetic condition 12 Eschmann TD411RS Diathermy with Foot The ESU power up ok The displays and function keys are working fine The patient plate monitoring circuit are Switch functional System was not tested with analyser Unit in good physical condition with footswitches 15 4x B Braun Perfuson Compact Syringe Driver Lot of 4 items All lot passed Power On Self test Display and keypad are in good working order except for 1 lot on which the keypad is working intermittently Plunger assembly is in good working order on all of them and
49. t 53 2 x Drager Isoflurane lot Dial is not locking on 0 The other lot is in good working order Average physical condition Dial not working on Ohmeda Average physical condition 1 x Ohmeda Enfluratec 4 55 Olympus CLK 4 Light Unit powers on OK Bulb is in good working order The Source brightness control knob works fine Insufflator works well too Overall the unit is in good physical and cosmetic condition 56 GE MAC 1200ST Unit passed power up test ECG monitor fully functional tested ECG using an ECG simulator printer is in good working order Unit also operational on battery Mains power lead missing Overall the unit is in good physical condition 58 B Braun Perfusor FM All 15 Perfusor powers up and self test completed Unit is Syringe Pumps x 15 powering on battery and mains as well Lot is in fair physical condition Only one unit from the lot was tested with syringe and all are working fine 62 Olympus WM 30 The stack and all of the associated items power up successfully Workstation Sony The video processor s selection keys and LED indicators are all Trinitron Monitor working and white balance completes successfully The light PVM 20M Olympus source emits a good output and again the selection keys pump CLV U40 Light and LEDs are all in good working order The monitor produces Source Pig Tail clear images with all the selection keys functioning An Connector Olympus interconnecting cable kit is included but the conn
50. t Guide Cable amp Light Transmission N A Objective Lens and Focusing Very poor condition objective lens and its casing has broken off rest of insulating layer no image Remote Control Switches N A Distal End Very poor condition Image Quality No image Working Channel s Blocked Leak Test Fail Overall Condition Poor 105 OLYMPUS Insertion Tube Good condition CF 240AL Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 1600x13 4mm Objective Lens and Focusing Good Remote Control Switches Fully functional Distal End Good condition Image Quality Good Working Channel s Clear Leak Test Pass Overall Condition Good cosmetic wear to the handle 106 PENTAX Insertion Tube Good condition E 3630UR Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good 1200x13 0mm Objective Lens and Focusing Processor was not available so unable to test Remote Control Switches Processor was not available so unable to test Distal End Good condition Image Quality Processor was not available so unable to test Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Fail Overall Condition Good physical condition but leaking slightly 107 PENTAX Insertion Tube
51. t View Scope 296x3 5mm Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good transmission but the sheath has come off the distal end of the light guide cable exposing the fibre optic cable Objective Lens and Focusing Good Remote Control Switches N A Distal End Fair condition Image Quality Good Working Channel s N A Leak Test Fail Overall Condition Fair but leaking 132 PENTAX ECY 1570 2 0 400x5 7mm Videoscope Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Objective lens is in good condition focusing is fairly good Remote Control Switches Fully functional Distal End Good condition Image Quality Fair Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Fair 133 PENTAX ECY 1570 2 0 Videoscope 400x5 6mm Insertion Tube Fair condition markings are slightly worn Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Cable is in fair condition transmission is very good Objective Lens and Focusing Lens is in good condition focusing is average Remote Control Switches Three out of four are res
52. ted Remote Control Switches Cannot be tested Distal End Good condition Image Quality No image is produced Working Channel s Clear Leak Test Pass Overall Condition Good physical condition but no image 110 PENTAX Insertion Tube Very good condition Unknown Bending mechanism amp Angulation Control Very good movement and Model locks Videoscope Light Guide Cable amp Light Transmission Cable is in very good 1000x11 0mm condition unable to test transmission as a compatible processor is unavailable Objective Lens and Focusing Processor was not available so unable to test Remote Control Switches Processor was not available so unable to test Distal End Very good condition Image Quality Processor was not available so unable to test Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good physical condition 111 FUJINON Insertion Tube Good condition EC0 60WL5 Bending mechanism amp Angulation Control Movement is good but Videuseope locking mechanism for left right movement does not work 1600x13 4mm Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Very good Remote Control Switches Two of three are working third probably hasn t been programmed Distal End Good condition Image Quality Very good Working Channel s Clear Leak Test Leaking Ov
53. tested Astothermplus Fluid They are in fair physical condition Warmer x 2 The fluid warmers both power up fine and displays and selection keys are working Not fully tested They are in average poor physical condition with some cosmetic markings and a little rust 598 MIE Vapamasta 5 Vaporiser in an average physical condition and no indication if Isoflurane Vapouriser it s been serviced lately 606 Burtons Series 3 The autoclave power and the indicators and displays are Instaclave Vacuum working fine The internal of the chamber is in fair condition 2545B Serial with instrument trays in place The unit has not been fully Number 015884 tested but appears to be in fair overall physical condition 610 Drager Fabius GS The unit power up ok and the firmware self test completed Anaesthetic System successfully The display and function knobs and keys are working System leak test passed The pressure gauges and flow meters are functional anti hypoxic safety features are working ok Oxygen and flow sensor are working fine Back up battery pack in good condition Test run with medical air and test lung was carried out and the unit seems to be working fine Unit is in a good physical amp cosmetic condition Software version 1 39 1 Medical hose amp user manual included Date of manufacture 2001 611 Drager Fabius GS The unit power up ok and the firmware self test completed Anaesthetic System successfully System leak test failed The displ
54. tion keys and all LEDs are working but was not fully Olympus PSD 10 tested Its footswitch was also not tested The printer has fully Diathermy with Foot functioning selection keys and printing was successful The HP Switch and Olympus printer is in good physical condition but a power supply is not OVC 200 Teaching included so could not be tested The keyboard is working but Adaptor has a crack in the lower left corner of the casing The endoscopy footswitch was not tested An interconnecting cable kit is included and is working fine Various accessories such as syringes and tubing are also included The stack system is in average physical condition with some cosmetic markings scratches and wear to some of the items 377 Eschmann M Series All the hydraulic and mechanical adjustment and positioning Operating Table No are working fine The unit has aged but still functional one Cushions handle screw missing 399 Bosch ST20 Unit powers up OK It is in full working order along with the Audiometer headphones The headphone has a round 6 pin connector There is a slight damage to the headphone cable near ear piece Overall the unit is in good physical condition 427 Drager Cato Edition The unit powers up and the displays are all in good working Anaesthetic Machine order but the message Fault Nr 9 is displayed though it is Serial Number possible to continue on to the main screen No further
55. unit is in good physical condition 491 Surgical clippers and All in fair condition the charged clippers are working fine chargers Clipper x8 amp charger x14 498 Philips Pagewriter Unit passed power up test ECG monitor fully functional tested Trim Il ECG using an ECG simulator and printer is in good working order Recorder Unit also operational on battery Overall the unit is in good physical condition 504 Philips Avalon FM30 Unit failed to power up Monitor 506 Seca Baby Scales Good physical condition no batteries not fully tested 520 Stainless Steel Trolley Good condition retaining bracket with nuts missing 525 Seca CT3000I ECG The unit power up ok system self test completed The display with Leads and keyboard are in good working order A good ECG traces printout was obtained when tested with an ECG simulator Battery was not working at the time of test System is in good physical condition 538 Stack Trolley with Processor power up ok the exposure adjustment and white Radience HD Display balance are working fine Colour bar is displayed on the LCD Storz Xenon Nova 300 monitor with no camera connected The light source powers up Light Source and ok with manual and auto brightness adjustment and xenon lamp Storz SCB Image 1 are in good working order Overall unit is in good physical Camera Control Unit condition 539 Philips HDI 5000 Unit does not power
56. up Overall in average physical condition 543 Valleylab Force 2 The diathermy powers up and the display and all selection keys Diathermy on Trolley are in good working order No leads or footswitches are included so the unit was not fully tested Item is in fair physical condition 544 HP Pagewriter 200 The unit power up but battery error unable further testing of the ECG Machine No system leads 548 Valleylab Force 2 The diathermy powers up and the display and all selection keys Diathermy on Trolley are in good working order No leads or footswitches are included so the unit was not fully tested Item is in fair physical condition 572 Penlon PPVe Vapouriser in an average physical condition and no indication Isoflurane Vapouriser if it s been serviced lately 592 Drager Vamos AA All four units power up and the system self test completed monitors x4 successfully The displays and rotary selector knobs are functioning The system was tested for detection of CO2 only and the graph and values were displayed correctly Lot is in very good physical amp cosmetic condition Year of manufacture June 2003 x2 amp 2007 x2 Software version 02 02 and 02 15 593 Braun Perfusor For the syringe drivers all power up and LCD displays and Compact S Syringe selection keys are working All backup batteries are working Driver x 3 and The clamp is missing from one Units were not fully
57. version 1 39 1 Medical hose included Date of manufacture 2001 692 Cambridge Heart Exercise System CH 2000 Unit does not switch On at all keeps beeping with nothing on the display screen Unit is overall in a good physical condition 719 Datex Ohmeda 3800 SPO2 Monitor Unit power ok display LCD and function keys are working fine The SpO2 probe was tested and it is fully functional Good condition 732 Welsh Allyn Schiller Both ECG machines power up ok indicators and key pad are ECG machine AT 1 working fine The units was tested with an ECG simulator good ECG traces were obtained on the printout on one unit only The other unit has issue with the print head All in good physical condition Only one ECG lead available 742 Eschmann VP35 Unit power up but the suction power is low needs servicing Suction Unit 753 Olympus Stack The video processor light source monitor and printer all power Trolley with Sony up fine and all selection keys are working Both the main and CRT Monitor OTVS7 emergency lamps for the light source are in good working Camera Controller order The stack system could not be fully tested as a camera CLVS40 Light Source head is not included The keyboard and remote control unit and Colour Video were not tested The keyboard has 3 key covers missing else the Printer stack system is in average physical condition 754 Olympus Stack The video