Home
Siemens OXYMAT 61 man
Contents
1. Figure 7 5 Reference gas channel for internal reference gas supply with ambient pressure 50 hPa Table 7 4 Partno Designation Order no Remarks 04 1 Nozzle complete C79451 A3480 B1 6 mm 3 000 hPa Nozzle complete C79451 A3480 B2 1 4 3 000 hPa 04 3 Capillary C79451 A3480 D518 3 000 hPa 04 4 Reference gas supply A5E00114838 Pump module complete OXYMAT 61 122 Operating Instructions 12 2011 A5E00123067 02 Appendix Technical specifications 8 1 General technical specifications Table 8 1 Technical specifications general technical specifications General technical specifications Measuring ranges 4 internally and externally switchable manual and automatic range switching is possible Smallest possible measuring span 3 2 vol or 5 vol O2 Largest possible span 0 to 100 vol O2 Measuring ranges with suppressed zero point Any zero point can be implemented within 0 100 vol O2 provided that a suitable reference gas is used see also Reference gases Page 126 EMC interference immunity In accordance with standard requirements of electromagnetic compatibility NAMUR NE21 08 98 EN 50081 1 EN 50082 1 Degree of protection IP 20 according to EN 60529 Electrical safety According to EN 61010 1 overvoltage category III Position of use analyzer Front panel vertical D
2. No Message Possible causes Remedy Remarks w8 Temperature of More than 3 C deviation If temperature remains Q measuring head from the setpoint temperature constant no need for outside see also S7 action otherwise contact tolerance service department w9 External Signal from external source Check Function 72 must be configured maintenance accordingly request W10 AUTOCAL Chec Tolerances that are exceeded Perform AUTOCAL again This message disappears only k error in AUTOCAL Check or false when the AUTOCAL has been sample gas assignment to the successfully completed measuring range is wrong Causes of maintenance requests OXYMAT 61 112 Operating Instructions 12 2011 A5E00123067 02 Appendix 6 6 2 Faults The following faults result in a fault message output in the display and are signaled externally if a corresponding relay has been configured using function 71 The measurement capability of the device is restricted if these messages appear Immediate remedial measures should always be carried out here by qualified maintenance personnel Function 87 can be used to switch off disable each fault message individually Each maintenance requests labeled with Q in the left column of the following table must be acknowledged No S1 Parameter memory test failed Message Possible causes EEPROM contains incorrect or incomplete data in the working range Remedy 1 Execute RESET or swi
3. Attenuator Dn Restrictor DS Pressure switch P Pump for reference gas supply Figure 7 2 Sample gas channel with hoses OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 119 Appendix I E E G D e ep Ho A Figure 7 3 Gas connections Table 7 3 Part no Designation Order no Remarks 03 1 Restrictor C79451 A3480 C10 Gas channel hose 03 1 Attenuator A5E00118236 Attenuator with hose 03 2 O ring FPM Viton C74121 Z100 A6 1 item 03 4 Flow meter C79402 Z560 T1 03 5 Pressure switch C79302 Z1210 A2 OXYMAT 61 120 Operating Instructions 12 2011 A5E00123067 02 Appendix Figure 7 4 Reference gas channel for external reference gas supply with 2000 to 4000 hPa OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 121 Appendix 10 Sie 88 me KOs if 4
4. The display of the drift values for the sensitivity calibration depends on the setting of the calibration method If total calibration is selected here only the selected measuring range for the sensitivity calibration is displayed For single calibration all measuring ranges can be individually adjusted and can therefore have different drift values which are shown separately The drift values can reset to 0 0 with the Reset drift values instruction When you reboot the device all measuring ranges have 0 0 as the drift value OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 4 5 4 1 OXYMAT 61 5 4 Measuring ranges Measuring ranges After selecting the measuring range functions in the main menu the following screen appears when the third softkey Measuring ranges is pressed Meas ranges 40 Range selection 41 Define meas ranges Figure 5 21 Measuring ranges submenu The measuring ranges menu contains all functions you need for selecting and setting the measuring ranges This menu is protected by the code of code level 1 Select measuring ranges function 40 40 Select range 0 0 0 0 0 0 0 0 Autorange Figure 5 22 Select measuring ranges You can select a fixed measuring range or switch to automatic measuring range switching All selection options are subject to mutual interlocking Automatic measuring range switching is only possible under the following conditions
5. 5 5 2 Limits function 51 51 Limits MC Limit 1 0 00 v v on relay Alarms at decrease 0 signal Applies to MR agg Oo Limit alarm on off O Limit 2 Figure 5 31 Limits The device monitors up to four limits and these limits can be allocated to any measuring range Each limit can be allocated to any relay using function 71 Ifthis was not configured the indicator appears in the limit screen You can only configure positive limits up to 100 Alarms at decrease signal Here you select whether a limit alarm is to be switched if the entered limit is exceeded or fallen short of Applies to measuring range Here you allocate the limit to the desired measuring range s by pressing the third softkey several times When you do so the pointers move over the measuring range numbers and indicate the ranges in which the limit monitor is to be active The menu screen shown here is for measuring range 3 Limit alarm The limit monitoring of every single limit can be switched off individually using function 52 Resetting the limit alarm If the limit relay is triggered this state is obtained even when the measured value goes back to the permissible range The response of a limit relay is registered in the logbook function 3 As soon as the cause of the limit alarm has been eliminated the limit relay is reset automatically You can change to the next limit using the fifth softkey L mit OXYM
6. e At least two measuring ranges must be available A measuring range is said to be present when the following is true Start of scale value full scale value e The measuring spans must become greater or smaller e The measuring ranges must border on each other or overlap Operating Instructions 12 2011 A5E00123067 02 63 Operation 5 4 Measuring ranges 64 Measuring range types The permissible measuring range constellations result which are shown in the figure below mbe i lt mbe i 1 i measuring range 1 4 mbe i gt mbe i 1 mbe end of measuring range Figure 5 23 Measuring range types Two measuring range types are distinguished Type A The full scale value must be smaller than the full scale value which follows it The top measuring range limit therefore becomes larger with every measuring range mba UU OU mbe SSV Start of scale value FSV Full of scale value LS Low switchover point select smaller measuring range HS High switchover point select larger measuring range Figure 5 24 Measuring range type A OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation OXYMAT 61 5 4 Measuring ranges The following applies to measuring range switching 20 10 of span from full scale value Measuring range 1 A
7. 3 3 Connections 3 3 1 2 Reference gas line The connection stems for the reference gas has a pipe diameter of 6 mm or 1 4 The reference gas line must be as short as possible and have a small cross section If you use N2 or O2as a reference gas you must use a metal pipe as the supply line If air is used as the reference gas it is recommended to install a dryer attachment in the intake to avoid volume errors in comparison to the gas side that can be caused by air humidity If you subsequently convert the device to a different reference gas supply ensure that only trained service personnel perform the replacement work for the connector stems and the reference gas restrictor 3 3 1 3 Pressure sensor The device has an internal pressure sensor for correcting the effect of pressure on the measured value It is firmly mounted in the analyzer unit and directly measures sample gas pressure through the reference gas supply The pressure sensor is therefore not involved in the installation 3 3 1 4 Gas preparation To prevent contamination of parts from contact with the sample gas which can thus influence the measurement the sample gas must be sufficiently prepared before it is introduced in the device The following components are usually located upstream from the sample gas inlet of the device e Gas sampling device e Sample gas cooler e Filter e A gas suction pump Depending on the composition of the sample gas you may need a
8. 3 nti nee ee at net ee a cide eee ties 118 7 4 Gas eanne yee OO 119 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 5 Table of contents 8 Technical specifications enseina kerian aaa EA AAE AEEA EARRA E A AEE ERKE E 123 8 1 General technical specifications oooonnoonncccnnnnnnnncnnoconccnnnnnnnnnnnnnncnnnnnnnnnnn nn nnnn rnrn nenns nn nn nnmnnn eenn 123 8 2 Reference Jasmine 126 8 3 Z ro p inNtErrOr s ee e A A eet 127 8 4 Materials used in the sample gas Channel ooccccoonocccccnnaciccnononcncnnnancnn nano nc nc nnnn nar nnnn rca rra rca 128 9 Dimension drawings 2 000000 iaa 129 9 1 DIMENSION drawings ere e ta d da de edel 129 A APDO toda ATT A TT cad dir s 131 A 1 EC Declaration of ConforMitY oooooncccnnnnoccccnonocccccononcnnncnnncnc nano cnn ccnnn cnn rca nc arrancar nn rnnn anar 131 A 2 Return delivery Rinne od iia 133 B ESD Directives c s sscciasersa dabei stecevceonsa longs p aapa aaaea a tena cevnetdanataecdearedbecd Oaa TAG SOAD Ti ENT pa SEENA 135 B 1 ESD guideline S ironia tatiana 135 e List of abbreviations acronyms uussussssnssnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnsnnnnnsnsnnnnnsnsennnnsnnnnnsnnnnnnn 137 C 1 List of abbreviations comico e dd 137 GIOSSAlY n A Ati ties elec ai ted a Ata eset E AN 141 o A O A NT 149 OXYMAT 61 6 Operating Instructions 12 2011 A5E00123067 02 General information 1 1 1 1 2 1 3 OXYMAT 61 Operating Instructions 12 2011 A
9. Calibration tolerances function 78 Setting the calibration tolerances OXYMAT 61 With this function you define the calibration tolerances 78 Calib tolerance MC Calib tolerance at zero in of smallest MR 10 Calib tolerance of sens in of current MR 10 Signal tolerance oO violation Figure 5 54 Setting the calibration tolerances Operating Instructions 12 2011 A5E00123067 02 89 Operation 3 6 Configuration Example 90 With softkey 3 Signal tolerance violation you activate or deactivate the tolerance monitoring If you previously configured a relay output to Maintenance request using function 71 the device outputs the changes in the zero point or the span with respect to the last calibration as Maintenance request Note In order to be able to use this function Total calibration function 23 must be activated With Calibration tolerance at zero point you define the max deviation proportional to the smallest measuring span in With Calibration tolerance at span you define the max deviation proportional to the current measuring span in The calibration tolerance which can be set from 0 to 99 refers to the measuring range with the smallest measuring span in the case of the zero point and to the measuring range where the total adjustment is carried out marked with m in function 22 in the case of the span Measuring range 1 95
10. External N x maintenance request 1 to 7 Condensate overflow Gas cooler fault see also section Maintenance requests and fault messages Deleting the N X After deletion the device is restored to the initial state logbook entries If the cause of a fault or maintenance request is not eliminated the corresponding message reappears in the logbook Function check N X Relay must be configured for functional check using CTRL 1 to 4 function 71 if for example the function is to be checked by a second device Start AUTOCAL N X Note Effective activation only possible in Measure operating mode AUTOCAL must be configured functions 23 24 and 25 Measuring range N X For measuring range switching 1to4 on Note Automatic measuring range switching function 52 must be turned off OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 83 Operation 3 6 Configuration 84 Function Necessary activation voltage Remark effect Zero gas on Calibration gas on Sample gas on OV 24 V 24 V pulse min 1 s Note Effective activation only possible in Measure operating mode Relay function under 71 must be set to zero gas calibration gas sample gas and the corresponding valves to be connected Applies only to total calibration since only one calibration gas can be used function 22 Zero calibration Span calibration
11. Release of hazardous gases Gases may be released if there are leaks in the sample gas channel For this reason toxic and corrosive gases are not allowed to be be measured with this device Failure to comply with this can result in death personal injury as well as damage to property and the environment CAUTION Incorrect mounting The device can be damaged destroyed or its functionality impaired through improper mounting e Before installing ensure there is no visible damage present on the device e Make sure that process connectors are clean and suitable gaskets and glands are used e Mount the device using suitable tools Refer to the information in Technical data for example installation torques requirements OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 17 Mounting and connecting 3 2 Installation requirements 3 2 Installation requirements Vibration Select an installation location which is as vibration free as possible If you install the device in a cabinet or in desktop housing you must hang it on support rails It is not sufficient to only secure the front side since the intrinsic weight of the device creates too much load on the chassis Temperature Ensure sufficient ventilation between the devices if you install them in control cabinets If the device is installed outdoors it must be protected from direct sunlight Ensure that the permissible ambient temperature range
12. Sample gas pressure O2 oo laptop IN A Maintenance center Local i at ULTRAMAT 6E E esse control 2 channel unit y nd RS485 ELAN Sample gas pressure O2 CO2 O2 sample gas 1 pressure I N OXYMAT6E PT ULTRAMAT OXYMAT 68 ak y Bi i Te oo ee 2 Container C yl en a e ELAN Economical Local Area Network Remote transmission via modem Networking of various analyzers Up to 12 channel addresses interfaces can be used for measured ULTRAMAT 6 OXYMAT 6 OXYMAT 61 ULTRAMAT 23 variables concentrations pressures temperatures etc Centralized maintenance by coupling to SIPROM GA higher level computer Measurement data available in ASCII PROFIBUS format for fast processing Remote control and download of records Technology RS485 9600 baud repetition rate 0 5 s via PC range up to 500m 1 ae 12 Channel addresses 13 and Control addresses e g for PC etc Figure 2 2 Various analyzers networked with 485 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 15 Description 2 4 Principle of operation 2 4 16 Principle of operation Reference gas channel Restrictors Reference gas channels Microflow sensor for measured signal Sample gas inlet Sample chamber Paramagnetic measurement effect Electromagnet with alternating field strength OMAN DOA FWD Sample gas and reference gas outlet OXYMAT 61 principle o
13. automatic calibration and for AUTOCAL Check automatic check that the set calibration tolerances are complied with without calibration The settings for AUTOCAL Check 4th subfunction only refer to the check of the calibration tolerances without calibration OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 55 Operation 5 3 Calibration AUTOCAL Check operating mode With this AUTOCAL function you configure various AUTOCAL operating modes The AUTOCAL Check is only for checking the calibrations As with AUTOCAL the device executes the sequence configured in subfunction AUTOCAL Check sequence In contrast to AUTOCAL no new calibrations are performed only the deviations are checked with respect to selectable calibration tolerances Autocal check mode Autocal check on off O Start autocal check O cyclically Start autocal check via O binary input Trigger autocal once Abort autocal Figure 5 13 AUTOCAL Check operating modes AUTOCAL Check on off During the state 4UTOCAL off shown as 0 the settings on the switch AUTOCAL start cyclically parameter and AUTOCAL start via digital input do not have an effect on the device Trigger AUTOCAL once cannot be selected A previously set cycle time continues to run but without triggering automatic calibration AUTOCAL Check start cyclically You can activate AUTOCAL to be a regularly repeated cycle if you first set
14. ie Digital input 4 24 V 13 to 33 V vo ARS SAI igital input 2 0 0V Oto 4 5 V DONT VA Digital input EP floating via optocoupler me TEYAT Digital input 1 to 4 N ag I Relay 6 lt el A amp _ 10 Relay 5 KY gt HQ d Mm 1 dD 8 _ 11 Relay 4 Contact load SS EEE max 24 V 1 A AC DC es OL a relay contacts shown 2 Say coy vhs ae Relay 3 de energized relay coil gt FT TAVI A gt gt TB Relay 2 AP Note E The wire and connector must SS Ss Relay 1 oh be shielded and be ee pe ys connected to the housing Sa GND potential Figure 3 6 Motherboard pin assignments OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 27 Mounting and connecting 3 3 Connections Connector SUB D 37 F optional Connector SUB D 9F X90 PROFIBUS DP 4 SSS SS CNTR P direction control ie 1 RxD TxD P B 7 RxD TxD N A 4 VP 5V Figure 3 7 Pin assignments of the add on board 28 Digital input 14 P Digital input 13 P Digital input 12 P Floating Digital input 11 P via optocoupler Digital input 10 P 0 OV 0 to 4 5 V Digital input 9 P 4 24V 13 to 33V Digital input 8 P Digital input 7 P Digital input 7 to 14 N Contact load max 24 V 1 A AD DC the relay has zero current for the shown relay contact position Note The wire and connector must be shielded and be connected to the housing
15. not trigger a maintenance request or fault message The logbook contains a max of eight pages with four messages per page lt works according to the cyclic buffer principle i e when all eight pages all 32 locations are occupied the oldest message is overwritten You can delete or block logbook entries function 60 but you can also switch them off individually function 87 Note If an error occurs whose error message is switched off with function 87 there is no reaction at any configured interface This applies to the ELAN interface the analog output and the relay output Display measuring ranges function 4 The measuring ranges defined using function 41 are listed here However you cannot carry out any changes in this function Operating Instructions 12 2011 A5E00123067 02 49 Operation 5 3 Calibration 5 3 50 Calibration After selecting the calibration functions in the main menu by pressing the second softkey Calibration the following menu screen appears with additional choices Calibration 20 Zero calibration 21 Span calibration 22 Setpoints total 23 Total single cal continue Figure 5 6 Calibration submenu This menu is protected by the code of code level 1 You can calibrate the device using the functions available in this menu If the function you are looking for is not included in this overview press the fifth softkey Continue to go to other functions T
16. 15 Pressure sensor 8 Pressure switch in sample gas channel option OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 23 Mounting and connecting 3 3 Connections 3 3 2 Electrical connections 3 3 2 1 Mains connection AR WARNING Dangerous contact voltage Danger of electric shock in case of incorrect connection e For the electrical connection specifications refer to the information in Chapter Technical specifications Page 123 e Atthe mounting location of the device observe the applicable directives and laws for installation of electrical power installations with rated voltages below 1000 V e The device comes with an appliance plug which may only be connected to the mains supply line by qualified personnel The mains supply line must comply with the valid regulations and conditions for the place of installation and be provided with a protective conductor which lies at the potential of the enclosure The cross section of each wire must be gt 1 mm Connect the phase conductor to the marked position in the plug e Install the mains line separately from the signal lines e Provide a mains disconnection device in the direct vicinity of the device for load rating see rating plate lt must be readily accessible and marked e Check whether the existing mains voltage agrees with the mains voltage specified on the rating plate OXYMAT 61 24 Operating Instructions 12 2011 A5E00123067 02 Mount
17. 2 Desi vi iaa aa 13 2 3 Communication interface 24244444444Hanennnnennnnnnnannnnnnennnnnnnnennnnnnnennennnnennnnrnnennnnnnanrnnen 15 2 4 Principle Of Operation tie ua HE Hate bade a ES 16 3 M unting nd conneclinGg 2 2 A ee he ee as 17 3 1 Safety INSTUCIONS ss A A A etch sede an a ERDE Renee 17 3 2 Installation requirements 2 2444000042424nnnannanannnnnnnnnnnnnnnannnnnnnnnnnnnnnannnnnnn nn nannnn nn 18 3 3 A A A 19 3 3 1 Gas connections and internal gas flow diagraM oooonnncccnnniniccnnnnncoccnnnancccnn narco noc cnn rra 20 3 3 1 1 Sample gas Nosice dei 20 3 3 1 2 Reference gas ING cities 21 33 13 SO AS 21 3 31 41 GAS PEPA O E 21 ASE AA eek eier keseehle 22 3 3 2 Electrical Connection Snas liarla area A 24 8 32 IMAINS connection 4 ac eee ee aie en aie ae none sinn 24 3 3 2 2 Connection of the signal CableS oooocnniccinnnnncicinnnoccccncnnonccnnnnnnn cnn nono cnn nro nana rca 25 3323 PINVASSIQNIMEMS ee E 27 3 4 Example of AUTOCAL Connection ccc cccecceceeeeeeeeeeceececeeeeeeeaaeaeceeeeeceseneanaeceeeeeeneeesiaeeeees 29 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 3 Table of contents COMMISSIONING A NN 31 4 1 Safety instructions ee te 31 4 2 Preparation for COMMISSIONING errien iaaa eene nono na nono n cnn nano cnn anno rra rra 32 4 2 1 General informatica tddi listar dict 32 4 2 2 Selection of the reference Gas 4 ek dades
18. 2011 A5E00123067 02 117 Appendix 7 3 Electronics 02 1 1 02 2 Table 7 2 A 2 er T Part no Designation 02 1 Motherboard Order no C79451 A3480 D501 C79451 A3480 D502 C79451 A3480 D503 C79451 A3480 D504 C79451 A3480 D505 Remarks Motherboard and firmware German Motherboard and firmware English Motherboard and firmware French Motherboard and firmware Spanish v Italian 02 1 1 Firmware FlashPROM C79451 A3460 S501 German C79451 A3460 S502 English C79451 A3460 S503 French C79451 A3460 S504 Spanish C79451 A3460 S505 Italian 02 2 Add on board C79451 A3480 D511 Relay A5E00057307 PROFIBUS PA A5E00057312 PROFIBUS DP A5E00057164 Firmware update PROFIBUS 02 3 Plug in filter W75041 E5602 K2 02 4 G type fuse W79054 L1010 T630 W79054 L1011 T100 200 V to 240 V T 0 63 A 250 V 100 V to 120 V T 1 A 250 V 02 5 LC display W75025 B5001 B1 02 6 Adapter board LCD keyboard C79451 A3474 B605 02 7 Front panel A5E00259978 With membrane keyboard OXYMAT 61 Electronics 118 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Appendix 7 4 Gas channel Restrictor inserted into the hose and fastened with a hose clamp 1 2 Sample gas 4 Reference gas
19. 3 l min and 1 l min OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 35 Commissioning 4 2 Preparation for commissioning Calibrating the zero point The physical zero point is calibrated with function 20 It is applied to all configured measuring ranges Calibrating sensitivity The sensitivity is calibrated with function 21 Compensating the effect of temperature The compensation of the temperature effect is stored permanently in the software firmware for the OXYMAT 61 No further actions are required for this Noise suppression Noise in the measured signal can be suppressed using function 50 This function allows you to configure a low pass filter which is assigned a time constant of maximum 100 s OXYMAT 61 36 Operating Instructions 12 2011 A5E00123067 02 Commissioning 4 2 4 Calibration examples 4 2 Preparation for commissioning The following examples reflect typical applications for the OXYMAT 61 O2 monitoring in gases Measurement task Measurement of oxygen in N2 Measuring range 0 to 5 O2 Reference gas N2 Calibration gas 4 7 O2 Procedure Function Input Remarks no Selection of measuring range start of scale value 41 0 0 gt 0 2 4 mA Selection of measuring range full scale value 41 5 5 gt 20 mA Specification of the setpoint for the physical zero point 22 0 Setpoint for physical zero point Specification of the setpoint for the sensitivity 22 4 7
20. 33 4 2 3 Initial COMMISSIONING 100 A 35 424 Calibration example Sesen ri cita 37 Opera teeth ech e detache eee hag eet eatin ae ae Ses ee aie 39 5 1 General information e eiea nenne nad steadied taeda nade ge ant 39 5 1 1 Measured value display and control panell cccccceeeeeeeececeeeeeeeeeeeeaeeeeeeesesecaaeaeeeeeeeeeeeenaees 39 5 1 2 Device operating Modes aida ita 41 5 1 3 Ecning US e o o 43 5 1 4 Operating SsegUENGE diia 43 STAL Introduction to the man menu ee kette el 44 91 472 Entering a submen ono oi 45 9 1 4 3 Returning to display Mode oca td ge 46 9 1 4 4 East function Selection cti 46 5 1 5 Overview of operating fungtions n n ee len E AS 47 5 2 Analyzer Status dida 48 5 2 1 Analyzer configuration function 1 o ooooocnnnncccnnnnacicinnnancccnnnorccnnnnorr cnn narran rr 48 5 2 2 Diagnostic values function Zj irs aias aa raae a AANE na raaa aaa AARET fia Ea Aa ia ia aA 49 5 2 3 Logbook function IJss niera aniani ic 49 5 2 4 Display measuring ranges function 4 oooonnncccnnnnniciconnncccnnnnonancnn nono nnnnnnnn cnn nnnn narran rca nnnn nn 49 5 3 Calre o R T E A E 50 5 3 1 Calibration of the zero point function 20 uuressnnnnnesnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnennnnnnn nn 51 5 3 2 Span calibration function 21 ooononcccnnnnniccnonocccccnnonnnnnnnnancnc nan ncncnnno nn nr rnnn rca r cnn rca rara 52 5 3 3 Zero span setpoints function 22 o ooccconnnnccccnnnoccccnononccc
21. 5 7 General information The following overview is a list of the device functions This list corresponds to software version 4 Main menu item Function number Name of function Analyzer status 1 Factory data 2 Diagnostics values 3 Logbook 4 Display measuring ranges Calibration code 1 20 Calibrate zero 21 Calibrate span 22 Zero span setpoints 23 Calibration settings Total Single 24 AUTOCAL 25 Drift values Measuring ranges 40 Select measuring ranges code 1 41 Define measuring ranges Parameters code 1 50 Electric time constant 51 Limits 52 On off functions 53 Status signals 54 Graphic measured value display 55 Measured value display 56 LCD contrast 57 Magnetic field frequency 58 Date time 59 Measuring point switchover 60 Logbook settings Configuration code 2 70 Analog output 71 Relay assignment 72 Digital inputs 73 ELAN configuration 74 Reset 75 Save load data 76 Suppress short noise signals 77 Measured value memory analog output 78 Calibration tolerance 79 Change codes 80 Analyzer test 81 Select language 82 Pressure correction 83 Correction of cross interference 84 Phase adjust 85 Switch valves 86 Linear temperature compensation 87 Faults on off 90 PROFIBUS configuration Overview of operating functions Operating Instructions 12 2011 A5E00123067 02 47 Operation 5 2 Analyzer status 5 2 Analyzer status After selecting the diagnostic functions in the main menu by
22. Activation initiates calibration See also Example digital inputs Autorange N X Activate Automatic measuring range switching Autocal check N X Note Effective activation only possible in Measure operating mode Start Autocal check function 24 Measuring protection N X It is possible to define a digital input Measuring protection which has the following effect If the device is in the Measure operating mode functional check deactivated it remains in this state i e e The device can no longer be decoded e The device can no longer be set to Remote The message Measuring protection activated appears in the status line of the measured value display OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration 5 6 4 ELAN configuration function 73 With this function you set the parameters for an ELAN network 73 ELAN config Channel address Measured value telegram ELAN tag Figure 5 48 Configuring ELAN Channel address Set the channel address for the device here Addresses from 1 to 12 can be set In an ELAN network each address may only be used once Addresses of devices that are used to correct the pressure or the influence of interfering gas may not be entered at this location Measured value telegrams on off Here you can switch the cyclic independent transmission of measured values on or off every 50
23. O Figure 5 58 Selecting a language The device is always delivered in the ordered language Usually English is included as a second language if English is the first language Spanish is set as the second language 5 6 13 Pressure correction function 82 Variations in the sample gas pressure can be corrected within the range from 800 to 1200 hPa absolute with the OXYMAT 61 This function offers the possibility for pressure correction using one of the following options e Pressure correction using an internal pressure sensor e Pressure correction using an external pressure sensor via analog input 2 example as shown e Pressure correction using an external pressure sensor via ELAN RS485 82 Pressure corr 0 ID T With ext pressure signal on analog input 2 Analog inp 2 O to 20 mA for range 0 1013 hPa Figure 5 59 Pressure correction via analog output 2 The pressure correction can also be switched off using function 52 On off functions OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 93 Operation 3 6 Configuration Pressure correction using an external pressure sensor via ELAN 82 Pressure corr 0 With ext pressure signal from ELAN Channel 4 Oo NO 994 hPa CTRL Figure 5 60 Pressure correction via ELAN Note Measured value Pressure The measured value Pressure is an internal value in the OXYMAT 61 gas analyzer which can be supplied to ano
24. Setpoint for sensitivity Start zero point calibration 20 Introduce N2 Start sensitivity calibration 21 Introduce calibration gas Room air monitoring Measuring range 15 to 21 Oz Reference gas Air 20 95 Oz Calibration gas 15 3 O2 Procedure Function Input Remarks no Selection of measuring range start of scale value 41 15 15 gt 0 2 4 mA Selection of measuring range full scale value 41 21 21 gt 20 mA Specification of the setpoint for the physical zero point 22 20 95 Setpoint for physical zero point Specification of the setpoint for the sensitivity 22 15 3 Setpoint for sensitivity Start zero point calibration 20 Introduce air Start sensitivity calibration 21 Introduce calibration gas OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 37 Commissioning 4 2 Preparation for commissioning O2 measurement in flue gas Measuring range 0 to 10 Oz Reference gas Air Calibration gas Na NOTICE Reference gas out of range section 5 In this example the O2 content of the reference gas is not within the measuring range of up to 10 O2 Since the measuring span is greater than 5 however an exception can be made in the selection of the reference gas In this case the pressure correction must be necessarily activated see also Function 82 in Procedure Function Input Remarks no Selection of measuring range start of
25. Table 5 2 Device operating modes 5 7 General information Mode Coded display mode Properties e Measured value display is shown e Protected submenus can only be reached by entering a code e The current operating mode of the device except for Measure is displayed in the bottom line e Functional check not active Remarks The device only supplies reliable measured values in this mode as far as it s in the Measure operating mode From the operator control mode you can reach this mode by pressing the MEAS button twice Decoded display mode e Measured value display is shown e The submenus protected by the entered code are accessible e The current operating mode ofthe device except for Measure is displayed in the bottom line e Functional check active e Measured value can be influenced From the operator control mode you can reach this mode by pressing the MEAS button once and confirming or discarding the made entries Operator control mode e Menu or function is displayed e Settings and inputs can be made e Functional check active e Measured value can be influenced From Coded display mode you can reach this mode by entering the code of the corresponding operation level In this mode you can configure calibrate the device OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 41 Operation 5 7 General information Schematic di
26. and commissioning as well as for safety during the operation and maintenance of the described product Only qualified personnel have the required professional knowledge for correctly interpreting the generally valid safety notes and warnings in this manual in each specific case and to act accordingly This manual is an inherent part of the scope of delivery despite the fact that it can be ordered separately for logistic reasons General information 1 4 Special information and warnings Due to the variety of technical details it is not possible to consider every single detail for all versions of the described product and for every conceivable case in the set up operation maintenance and use in systems For further information or in the case of problems which are not covered in enough detail in this document please request the required information from your local or responsible Siemens regional office Note In particular before using the device for new research and development applications we recommend that you first contact us to discuss the application in question 1 4 Special information and warnings This manual provides you with information on using installing operating and maintaining the device Pay particular attention to all special information and warnings Information of this type is set apart from the rest of the text and is marked with the corresponding pictograms This information provides you with useful
27. chamber it can be compensated with this function 86 Lin temp comp MC Aftercompensation of the zero point Aftercompensation of the span Figure 5 67 Linear temperature compensation Operating Instructions 12 2011 A5E00123067 02 99 Operation 5 6 Configuration Temperature compensation at the zero point Example Assuming an average temperature Tm two different correction values can be specified for ranges of increased temperature and decreased temperature If the zero point changes by 0 3 relative in relation to the difference between 100 O2 and the start value of the smallest span due to a temperature increase in the sample chamber changes from Tm to Tw the value determined by the following formula should be entered under A for a temperature increase 0 3 A Tu Tm x 10 10 C A factor can be determined for a temperature decrease in the same manner If only a correction factor is determined the first value of opposite sign is entered for the second correction value Temperature compensation in the measured value Example 100 The procedure is the same as for the zero point however the percentage change refers to the measured value itself If the measured value changes from 70 to 69 with a temperature increase of 4 C the percentage change is Ze x 100 1 42 4 C and therefore A 3 55 10 C Note If the zero point changes to negative
28. control mode Generic term for a number of independent states which the device can assume during operation e g measure standby pause etc Various modes into which the device can be put by intervening from the outside Three modes are distinguished e Coded display mode e Decoded display mode e Operator control mode Generic term for a number of independent states which the device can assume during operation e g measure standby pause etc Certain operating area menus and functions which is either freely accessible or protected by a code Operator control mode Sample gas Sealing OXYMAT 61 Operating mode in which the device is partially or entirely decoded and the functional check is activated The device is configured in this mode A gas which has been extracted from a process for analysis Protection of the materials against escaping of gases Operating Instructions 12 2011 A5E00123067 02 145 Glossary Signaling contact Usually a floating contact contained in electronic components e g relay which signals the occurrence of an event defined as a fault to a control unit Span Difference between defined start of scale value and defined full scale value Span calibration Adjustment of the span using a suitable calibration gas Spike English Thorn pin prong Undesired briefly occurring peaked interference of a measured signal Standard configuration Totality of standard featur
29. following query appears Return to meas mode Accept modifications Figure 5 4 Query before jumping back e Press YES to accept changes permanently into the working memory of the parameter memory or NO to discard the changes The device then changes to Decoded display mode By pressing the ESC brings you back to the most recent function screen Changes are accepted here without a further query Coding the device After jumping back to Decoded display mode with ESC or MEAS if you press MEAS again you put the device back into Coded display mode Only now is the function check once again deactivated and all states invoked by the decoding are released again Fast function selection In order to directly access the desired function when operating frequently from the Coded display mode a Power user operation was created This allows you to directly access the desired function by entering the function number This allows you to skip menu levels However the functions can only be directly called from the measured value display For fast function selection proceed as follows 1 Enter the number of the desired function using the number keys 2 Press the softkey of the component with the arrow If the desired function is protected by a code you will be asked to enter the code OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 1 5 OXYMAT 61 Overview of operating functions
30. frequency back to the factory default of 8 095 Hz NOTICE Calibration of zero point and sensitivity After each frequency change the zero point and sensitivity must be calibrated again Operating Instructions 12 2011 A5E00123067 02 75 Operation 5 5 Parameters 5 5 9 Date time function 58 With this function you can set the date and time 58 Date Time MC New date dd mm yy 24h da 01 03 07 y3 New time 12 53 Set clock e Actual date Actual time 01 03 2007 12 53 Figure 5 40 Setting the date time The device features a system clock that is not buffered against mains failure no a real time clock When starting the device the clock starts with the date 1 1 1995 When the function is called the cursor is placed at the first place of the date display Enter the new settings in the order day month year By pressing ENTER jump to the next input field With a 24 hour based system hours minutes set the time in the same way NOTICE When the device is switched off the clock stops and is not updated The settings are especially important for troubleshooting Errors which are always stored in the logbook function 60 can be allocated more easily with the help of the date and time Press the third softkey Set clock in order to accept the set data These then appear at the bottom edge of the display OXYMAT 61 76 Operating Instructions 12 2011 A5E00123067 0
31. function 41 The start of scale or full scale values are assigned to the 0 2 4 or 20 mA of the analog output If you define multiple measurement ranges we recommend that you set measurement range 1 for the smallest measuring span MS etc In general the following applies then MS1 lt MS2 lt MS3 lt MS4 Setting the physical Zero point If the composition of test and reference gas are identical their O2 difference is therefore zero and there is no measuring signal We refer to this as the physical zero point Depending on the reference gas the physical zero point may therefore have any value between 0 and 100 Oz The setpoint of the physical zero point is specified under function 22 Setting the setpoint of the sensitivity The setpoints of the sensitivity must be as far from the physical zero point as possible at least 60 of the respective measuring span The setpoint or setpoints are entered with function 22 The calibration gases required for this must be made available for the adjustment Selecting the calibration method Set the calibration mode to total or single calibration using either function 23 or 52 e Single calibration means that each measuring range is calibrated using its own calibration gas e Total calibration means that you adjust one leading measuring range and all other measuring ranges are tracked to this using the switching ratio When performing the calibration ensure that the gas flows between 0
32. in normal Measuring range limit in case of fault output mA operation CTRL Start of scale Full scale value Start of scale Full scale value value mA mA value mA mA 0 20 1 21 0 21 2 20 1 21 2 21 4 20 2 21 2 21 4 20 NAMUR 13 8 20 5 3 21 5 Note If the electronics is defective it is possible that the analog output may get stuck at approx 1 mA or approx 24 mA Operating Instructions 12 2011 A5E00123067 02 79 Operation 3 6 Configuration 5 6 2 Relay outputs function 71 80 With this function you can assign each relay functions listed in the table Relay outputs possible functions whereby you may only assign each function once In other words for example Fault may not be assigned to two relays at the same time 71 Relay outputs R 1 Fault R 2 Maint req R 3 Funct cont R 4 not used continue Figure 5 45 Assigning the relays There are six freely configurable relays included in the standard equipment whose switchable output contacts max 24 V 1 A can be used for signaling actuating valves etc If these six relays are insufficient you can retrofit additional electronics with eight further relays add on board The connection assignments for the individual relays in the zero current state can be found in the routing assignments Motherboard pin assignments The relays are preset at delivery Press the fifth softkey continue to access the
33. person coming into contact with the materials indicated These values correspond to IEC 801 2 specifications Voltage in kV kv A 16 O Synthetic material 15 14 N Wool 13 N 8 Antistatic material for example wood 12 N or concrete 11 SO 10 9 N 8 N 7 N N 8 5 N ES 4 N a D x eC m 2 1 DS HHH gt 5110 20 a 40 50 60 70 80 90 10 Relative humidity in Figure B 1 Electrostatic voltages on an operator Basic protective measures against electrostatic discharge e Ensure good equipotential bonding When handling electrostatic sensitive devices ensure that your body the workplace and packaging are grounded This prevents electrostatic charge e Avoid direct contact As a general rule only touch electrostatic sensitive devices when this is unavoidable e g during maintenance work Handle the modules without touching any chip pins or PCB traces In this way the discharged energy can not affect the sensitive devices Discharge your body before you start taking any measurements on a module Do so by touching grounded metallic parts Always use grounded measuring instruments 136 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 List of abbreviations acronyms C 1 List of abbreviations Abbreviation Meaning gt Greater than lt Smaller than gt Greater than or equal to lt S
34. potential Connector SUB D 9M X90 PROFIBUS PA OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting 3 4 Example of AUTOCAL connection 3 4 Example of AUTOCAL connection Function in the device Connection end VO d o Connector SUB D 37 F optional Zero gas M V1 d c Tham E ug ET A Calibration EPA gas 1 a V2 d Cu d c AAA Calibration gas 2 V3 d c Calibration PE V4 d c A ce Calibratio gas 4 AAA 24 V power Sample M supply max gas AL In the relay contact position shown the relays are de energized VM d c d o de energized open d c de energized closed Sample gas inlet Gas preparation elements Figure 3 8 Connector pin assignment and valve diagram of an AUTOCAL connection OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 29 Commissioning 4 4 1 Safety instructions Explosion hazard The analyzer must not be operated in potentially explosive atmospheres The supply of gases with combustible components in concentrations above the lower explosive limit LEL must be clarified in consultation with the relevant hazard experts and is ultimately the responsibility of the operator warn ING Dangerous contact voltage Certain parts of this device are under dangerous voltage In order to avoid contact with live parts t
35. scale value 41 0 0 gt 0 2 4 mA Selection of measuring range full scale value 41 10 10 gt 20mA Specification of the setpoint for the physical zero point 22 20 95 Setpoint for physical zero point Specification of the setpoint for the sensitivity 22 0 Setpoint for sensitivity Start zero point calibration 20 Introduce air Start sensitivity calibration 21 Introduce N2 Purity monitoring of oxygen Measuring range 95 to 100 Oz Reference gas Pure Oz Calibration gas 95 6 O2 Procedure Function Input Remarks no Selection of measuring range start of scale value 41 95 95 gt 0 2 4 mA Selection of measuring range full scale value 41 100 100 20 mA Specification of the setpoint for the physical zero point 22 100 Setpoint for physical zero point Specification of the setpoint for the sensitivity 22 95 6 Setpoint for sensitivity Start zero point calibration 20 Introduce pure O2 100 Start sensitivity calibration 21 Introduce calibration gas Detailed operating instructions for performing all the required testing functions are provided in section 5 Operation 38 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation Operation 5 1 General information 5 1 1 Measured value display and control panel Q e 3 40 72 11 I E d mnnn 6 noes Status display Display of dimension Status line can be programmed under function 53 Display of measured c
36. with temperature change A has a positive sign and vice versa The same applies to a measured value which becomes smaller OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 18 5 6 19 OXYMAT 61 5 6 Configuration Error on off function 87 With this function you switch the message for maintenance requests and faults see Tables Causes of maintenance requests and Causes of error messages off individually so that there is no entry in the logbook no message in the status line of the measured value display and no external signal sent 87 Error On Off S1 Parameter memory 2 3 S4 External fault continue p Figure 5 68 Switching errors on and off Error messages which do not apply to this analyzer are identified by the absence of text following the fault number PROFIBUS configuration function 90 This function can only be called if the device contains additional PROFIBUS electronics add on board This function is used to set the PROFIBUS parameters 90 PROFIBUS config Address 126 TAG OXYMATSIXTYONE Ident number wo gy Relay on PB Software vers 2 0 0 Boot software 0 2 0 Figure 5 69 Configuring PROFIBUS Operating Instructions 12 2011 A5E00123067 02 101 Operation 3 6 Configuration You can set the following specific PROFIBUS parameters e Address PROFIBUS station address The address can be set from 0 to 126 Under TAG you can display t
37. 0 ms Note Independent sending of measured values If you have set up your own communication control system this option offers a simple way of checking an ELAN telegram In order to avoid unnecessarily loading the device and the ELAN network however the function should only be switched on as needed ELAN tag Display of the tag assigned to the device in the ELAN network For more details refer to the ELAN interface description Order no C79000 B5200 C176 German C79000 B5276 C176 English OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 85 Operation 3 6 Configuration 5 6 5 Reset function 74 With this function you can carry out a warm restart 74 Reset Trigger reset e Figure 5 49 Carrying out a reset After calling this function the device switches to the Warm up phase which activates the functional check at the same time The device is only completely ready for operation again after successfully running through this phase 5 6 6 Save load data function 75 You can configure the user data memory with this function 75 Save data MC Save user data e Load user data e Load factory settings e Figure 5 50 User data memory Press the softkey 1 Save user data e g after the system has been successfully commissioned All individual settings are then saved Press the softkey 2 Load user data to load the last user data saved These functions are important when the device is
38. 2 Operation 5 5 10 5 5 11 OXYMAT 61 5 5 Parameters Measuring point switching function 59 With this function you can assign the device up to six measuring points which are switched automatically at the expiration of a configured period 59 Sample selection Rel 2 45 Rel 1 20 M p switching on off Figure 5 41 Measuring point switching The condition is that you configured the measuring point relay beforehand using function 71 Relay outputs which then actuates the corresponding solenoid valves Every measuring point relay is also allocated a time period which you can enter in the respective input field Values between 0 and 60 minutes are possible for this input Press the fifth softkey to activate deactivate the measuring point switching You can allocate a signal relay to each measuring point relay This allows measuring point identification separate from the measuring point relay You also use function 71 to configure the signal relay Logbook settings function 60 With this function you delete or block logbook entries 60 Setup logbook MC Clear logbook e Lock logbook a Figure 5 42 Configuring the logbook Status messages such as maintenance requests or faults cannot be suppressed by blocking They still appear despite the blocked logbook Operating Instructions 12 2011 A5E00123067 02 77 Operation 3 6 Configuration 5 6 Configuration This menu contains all function
39. 4 Drift values 62 E EC Declaration of Conformity 133 Edit mode 43 ELAN 15 85 Electric time constants See Time constants 149 Index Electrical connections AUTOCAL connection 30 Pin assignments 28 Position 20 Signal connections 26 Electronics Spare parts 118 Error messages 101 ESD guidelines 137 F Factory data 48 Fault 110 Message 113 Fault messages 101 Flue gas measurement 38 Functions See Operating functions Fuses 109 124 G Gas connections Position 20 Spare parts 121 Gas flow diagram 23 Gas inlet 125 Gas preparation 22 32 Guidelines ESD guidelines 137 Influencing variables 126 Initial commissioning 35 Input dialogs 43 Installation Installation location 19 Preconditions 19 Reference gas connection 33 Interface ELAN 85 PROFIBUS 101 Interferences 126 Interfering gas 94 128 L Language 93 Language setting 93 150 Leak test 34 107 Limit monitoring 69 Limits 69 List of abbreviations 139 Logbook 49 77 110 M Magnetic field frequency 75 Main menu 44 Mains connection 25 Maintenance Maintenance intervals 103 Safety instructions 103 Maintenance interval 103 Maintenance request 110 Maintenance requests Message 111 Measured value Display 43 74 Graphical representation 72 Negative measured value 79 Phase adjustment 98 Progress display 72 Signal 98 Temperature compensation 99 Measuring characteristics 12
40. 5 Measuring head 105 106 Measuring point switching 77 Measuring principle 11 17 Measuring range 63 Calibration 52 select 63 Specify 66 Switching 63 Measuring range switching 63 Memory modules 87 Messages 110 Motherboard 108 N Nameplate 115 Noise 88 Noise suppression 68 O On off functions 70 Operating language 93 Operating mode 41 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Index Operating sequence 42 Operation 39 Inputs 43 Menu structure 44 Power User 46 Sequence 42 Operator functions Analyzer status 48 AUTOCAL 55 Calibration 50 Configuring the analog output 78 Correction of cross interference 94 Device test 91 Digital inputs 82 Drift values 62 ELAN 85 Electric time constants 68 Graphical measured value representation 72 Language selection 93 Limits 69 Load user data 86 Logbook 77 Measuring point switching 77 Measuring ranges 63 Noise suppression 68 Operating language 93 Overview 47 Phase adjustment 98 Pressure correction 93 PROFIBUS 101 Quick selection 46 Relay assignment 80 Save user data 86 Set contrast 74 Setting the calibration tolerances 89 Setting the code 91 Setting the date time 76 Setting the magnetic field frequency 75 Setting the measured display 74 Show hide error messages 101 Span calibration 52 Status messages 71 Suppressing noise 88 Switch on off functions 70 Switching valves 99 Temperature compen
41. 50021 x X 94 9 EC EN 50281 1 1 A1 x 94 9 EC EN 60079 14 sec 13 X 94 9 EC EN 954 X 94 9 EC EN ISO 13849 1 X 94 9 EC ZH 1 10 sec 1 3 4 2 2 3 1 X X 94 9 EC DIN EN 50104 94 9 EC DIN EN 50271 all enviroments included Certificates Zertifikate Certificates f E31 a A C E a A C E a A C E a A C E a A C E Zertifikate E32 E33 b 0 1 b 0 1 b 0 1 b 6 7 b 2 3 E38 E39 c B c B B c B c B g 30 d A B G d A B G d AB G d A B G d A B G e E40 e E11 e E12 BVS 03 ATEX G013 X X TUV 03 ATEX 2278 X X T V 01 ATEX 1686 X X T V 01 ATEX 1697 X X T V 01 ATEX 1708 X X PTB 00 ATEX 2022 X X Inspection Surveillance Kontrolle berwachung Directive Notified Body Product Quality Assurance No Richtlinie Benannte Stelle Qualit tssicherung Produktion 94 9 EC ATEX T V NORD CERT GmbH Am T V 1 D 30516 Hannover 0044 Page 2 2 OXYMAT 61 132 Operating Instructions 12 2011 A5E00123067 02 Appendix A 2 A 2 Return delivery Return delivery The analyzer or replacement parts should be returned in their original packaging material If the original packing material is no longer available wrap the analyzer in plastic foil and pack in a sufficiently large box lined with additional padding material wood shavings foam rubber etc If you use wood shavings the padding should be at least 15 cm thick on all sides When shipping overseas the analyzer must be additionally sealed air tight i
42. 5E00123067 02 Device model and validity You have purchased the OXYMAT 61 a device that that undergoes continuous development and therefore can vary in functionality and software release in comparison to other devices of the same type This manual describes the OXYMAT 61 gas analyzer with the software release 4 8 3 Information for our customers Before beginning work with this device please read this manual lt contains important information and data whose observation ensures proper device function and saves you servicing costs The manual will help you to operate the device more easily and efficiently allowing you to achieve reliable results General information This device left the factory in a safe and proper condition and has been tested In order to maintain this condition and to ensure safe operation of this product it should only be used in the manner described by the manufacturer Furthermore proper transportation storage installation operation and maintenance of the device are vital for ensuring correct and safe operation This manual contains the information required for the intended use of the described product It is addressed to technically qualified personnel who are specially trained or who have the relevant knowledge of automation technology measuring and control systems Knowledge and technically correct implementation of the safety notes and warnings contained in this manual are required for safe installation
43. 7 xax0c bcdx Z e f xxx 7MB2021 xXXX0 XXXX Z 9 XXX 7MB2027 xxxx0 bxxx Z g xxx 7MB2041 XXXXX XXXX Z XXX a b c d e f g X see tables page 2 siehe Tabellen Blatt 2 The product described above in the form as delivered is in conformity with the provisions of the following European Directives Das bezeichnete Produkt stimmt in der von uns in Verkehr gebrachten Ausf hrung mit den Vorschriften folgender Europ ischer Richtlinien berein 2004 108 EC Directive of the European Parliament and of the Council on the approximation of the laws of the Member States relating to electromagnetic compatibility and repealing Directive 89 336 EEC EMC Richtlinie des Europ ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten ber die elektromagnetische Vertr glichkeit und zur Aufhebung der Richtlinie 89 336 EWG 2006 95 EC Directive of the European Parliament and of the Council on the harmonisation of the laws of Member States relating to electrical equipment designed for use within certain voltage limits LVD Richtlinie des Europ ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten betreffend elektrische Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen 94 9 EC Directive of the European Parliament and the Council on the approximation of the laws of the Member States concerning equipment and protective systems intended for use in potentially explosive atmosph
44. 78 89 Calibration method 54 Calibration tolerances 89 Code 91 Contrast 74 Date 76 Digital inputs 82 Display contrast 74 ELAN parameters 85 Electric time constants 68 Language 93 Limits 69 Magnetic field frequency 75 Measured value display 74 Noise suppression 68 Operating language 93 PROFIBUS parameters 101 Relay function 80 Setpoints 54 Status messages 71 Suppressing noise 88 Time 76 Software 7 Software release 7 48 Spans 35 152 Spare parts 115 Analyzer section 117 Electronics 118 Reference gas channel 123 Sample gas channel 121 Status messages 71 Submenu 45 Analyzer status 48 Calibration 50 Configuration 78 Measuring range 63 Parameters 67 T Technical specifications 124 Associated gases 128 Climatic conditions 127 Gas inlet conditions 125 Inputs 126 Interfering gases 128 Measuring characteristics 125 Outputs 126 Power supply 124 Reference gas 127 Time response 125 Temperature compensation 99 Time 76 Time constants 68 Time response 125 U User data memory 86 W Warm restart 86 Z Zero point Calibration 51 Setpoint 54 Temperature compensation 99 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Siemens AG Subject to changes without prior notice Industry Automation IA A5E00123067 02 Sensors and Communication O Siemens AG 2001 2011 Process Analytics 76181 KARLSRUHE A5E00123067 www siemens com processautomati
45. A5E00123067 02 0 2 4 to 20 mA isolated Max load 750 Q 6 with changeover contacts can be freely configured e g for measuring range identification Max load 24 V AC DC 1 A floating 2 designed for 0 2 4 to 20 mA for external pressure sensor and or correction of associated gas influence cross interference correction 6 designed for 24 V isolated freely configurable e g for measuring range switchover RS485 Supplementary electronics with 8 additional digital inputs and 8 additional relay outputs e g for triggering automatic calibration supplementary electronics for PROFIBUS PA and PROFIBUS DP 125 Appendix Table 8 9 Technical specifications climatic conditions Climatic conditions Permissible ambient temperature Storage and transport 40 to 70 C operation 5 to 45 C Permissible humidity lt 90 RH 1 as annual mean during storage and transport 4 1 RH relative humidity 2 Maximum accuracy is achieved after approximately 3 hours 3 Relative to sample gas pressure 1 000 hPa absolute sample gas flow 0 5 and ambient temperature 25 4 Dew point low limit may not be violated 5 The error of the calibration gas must also be taken into account Based on DIN EN 61207 IEC 1207 8 2 Reference gases Table 8 10 Reference gases for OXYMAT 61 Measuring range Recommended Reference gas connection pressure Remarks reference gas 0 to 100 vol O2 N2 4 6 t
46. AT 61 Operating Instructions 12 2011 A5E00123067 02 69 Operation 5 5 Parameters 5 5 3 70 On off functions function 52 520n off config Auto range Stored function continue Figure 5 32 On off functions With this function you can easily switch other functions on off for example the ones listed in this function screen Thanks to this simplified operation these functions do not have the longer paths through various menus and submenus For better orientation the function numbers have also been specified Activated functions are marked by deactivated ones by Using the fifth softkey continue jump to the next function screen with further functions Function 52 is used to switch the following functions on and off Table 5 3 Functions switched on off by function 52 Description Fct no Remarks Total calibration 23 AUTOCAL 24 Only with supplementary electronics Automatic measuring range switching 40 Limit monitoring 1 51 Limit monitoring 2 51 Limit monitoring 3 51 Limit monitoring 4 51 Suppression the display of negative measured values 55 Blocking of the logbook 60 Suppression the display of negative measured values 70 Fault WA FCTRL acc to NAMUR 72 Measured value memory 77 Signal tolerance violation 78 Temperature recompensation of zero point 86 Temperature recompensation of span 86 OXYMA
47. H Measuring range 2 Figure 5 25 Measuring range switching type A When the upper switchover point OU is exceeded the next larger measuring range available is selected If the lower switchover point UU of the next smaller measuring range available is fallen below this is selected The UU lies at 80 HystS of the measuring range The OU lies at 90 HystE of the measuring range Type B The full scale value must be greater than or equal to the full scale value following it Since the measuring spans must simultaneously become larger the start of scale values of the following measuring ranges become continuously smaller mbaOU uu mbe SSV Start of scale value FSV Full of scale value LS Low switchover point select smaller measuring range HS High switchover point select larger measuring range Figure 5 26 Measuring range type B The following applies to measuring range switching 0 10 of span from start of scale value Measuring range 1 Measuring range 2 Figure 5 27 Measuring range switching type B Operating Instructions 12 2011 A5E00123067 02 65 Operation 5 4 Measuring ranges 5 4 2 66 When the upper switchover point OU is fallen below the next larger measuring range available is selected If the lower switchover point UU of the next smaller measuring range available is exceeded this is selected The UU lies at 10 100 HystE of the measuring range The O
48. OXYMAT 61 Oxygen analyzer for standard applications 7MB2001 Operating instructions 8 OXYMAT 61 Continuous gas analysis SIEMENS SIEMENS Continuous gas analysis Oxygen analyzer for standard applications OXYMAT 61 Operating Instructions 7MB2001 12 2011 A5E00123067 02 General information Description Mounting and connecting Commissioning Operation Maintenance and servicing Spare parts list Technical specifications Dimension drawings Appendix ESD Directives List of abbreviations acronyms O uw gt O N O oa A W N Legal information Warning notice system This manual contains notices you have to observe in order to ensure your personal safety as well as to prevent damage to property The notices referring to your personal safety are highlighted in the manual by a safety alert symbol notices referring only to property damage have no safety alert symbol These notices shown below are graded according to the degree of danger indicates that death or severe personal injury will result if proper precautions are not taken WARNING indicates that death or severe personal injury may result if proper precautions are not taken CAUTION with a safety alert symbol indicates that minor personal injury can result if proper precautions are not taken CAUTION without a safety alert symbol indicates that property dam
49. OXYMAT 61 The reference gas line is a restrictor which decreases the incoming reference gas pressure to allow a continuous flow of 5 to 10 ml min The reference gas line is installed and removed as follows 1 Loosen the reference gas line at the nozzle and on the analyzer section 2 Remove the reference gas line To install carry out the above steps in reverse order CAUTION Clogged reference gas channel It is important to ensure that no liquid or dust can enter the connection nozzles or the reference gas line The measurement function may deteriorate or the device may even fail as a result Principle of operation Page 16 Operating Instructions 12 2011 A5E00123067 02 105 Appendix 6 2 2 Disassembling the analyzer section Removing the measuring head Proceed as follows for this 1 Pull out the measuring head connector cable out of its socket 2 Loosen socket screws Now you can remove the measuring head To install carry out the above steps in reverse order It is important here to ensure that all O rings are re inserted In additional the screen cup should never be removed from the measuring head Cleaning the sample chamber 106 The sample chamber is not particularly susceptible to faults If condensation passes through the analyzer as a result of a fault in the gas conditioning although there is a temporary fault in the measurement highly oscillating values are displayed the d
50. Output signal variation lt 1 0 75 of the smallest possible measuring range according to rating plate with electronic time constant of 1 s corresponds to 0 25 at 20 Zero point drift lt 0 5 month of respective span Measured value drift lt 0 5 month of respective span Repeatability lt 1 of respective span Linearity deviation lt 1 of respective span OXYMAT 61 124 Operating Instructions 12 2011 A5E00123067 02 Appendix Table 8 7 Technical specifications influencing variables Influencing variables 3 Ambient temperature Sample gas pressure Associated gases Sample gas flow Power supply lt 0 5 10 K relative to smallest possible span according to rating plate With disabled pressure compensation lt 2 of the current measuring range 1 pressure change With enabled pressure compensation lt 0 2 of the current measuring range 1 pressure change Deviation from zero point corresponding to paramagnetic or diamagnetic deviation of accompanying gas see table 3 2 lt 1 of smallest possible span according to rating plate with 0 1 change in flow within the permissible flow range lt 0 1 of the current measuring range with rated voltage 10 Table 8 8 Technical specifications electric inputs and outputs Electric inputs and outputs Analog output Relay outputs Analog inputs Digital inputs Serial interface Options Operating Instructions 12 2011
51. Suppressing noise Spikes are created by electromagnetic interferences or sometimes by mechanical shocks These faults can be suppressed by entering a blanking time of O to 5 s The entered time has the effect that the last measured value before the spike occurred is displayed so that the measured result is not influenced Times can be entered in steps of 0 1 s Under Level in of smallest MR enter the threshold value in of the smallest measuring range above which the noise signals are to be suppressed Note If a change in concentration directly follows a fault this may be displayed after a delay When this function is activated the settings of function 50 Electric time constants must be taken into account since these are executed first OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 8 5 6 9 5 6 Configuration Measured value memory function 77 This function is used to define the response of the analog output or the digital interface when the device assumes a specific state 77 Store MC Analog out to meas value M Analog out to 0 2 4 mA oO Analog out to 21 mA oO Store on off O Figure 5 53 Setting the analog output In case of a fault S or CTRL decoding calibration warm up phase one of these values is sent to the analog output e The last registered measured value or e 0 2 4 mA or e 21mA Memory on m activates the respective setting
52. T 61 Operating Instructions 12 2011 A5E00123067 02 Appendix Magnet circuit A magnetic pole shoe is glued into each casing section The cut strip wound core placed upon it is held down by a tension spring This method deflects the forces of the magnetostriction away from the sample chamber Sample chamber The sample chamber consists of a 1 mm thick center plate in which the sample gas channel is punched and two 0 3 mm thick cover plates which contain the openings for the supply of the sample gas and the reference gas Since the sample gas only comes into contact with the chamber plates the OXYMAT 61 can be used for numerous combinations of sample gas The other channels of the measuring system are purged by the reference gas In the fully assembled analyzer section the sample chamber is installed between the two casing sections Measuring head The microflow sensor of the measuring circuit is located in the measuring head This is installed in a thermostated aluminum block The strong magnetic field required to produce a sufficient measurement effect causes interference in the microflow sensor The aluminum block with the preamp electronics are therefore shielded to eliminate this effect This screen is composed of e Screen cup and e Restrictor plate This includes restrictors the function of which is described in section 2 5 Both parts are made of a material with high permeability Reference gas channel See also
53. T 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 5 Parameters Besides the functions listed in the table Functions switched on off by function 52 function 52 can also be used to address other service functions These are restricted to service engineers and are only visible when the service code is entered code level 3 5 5 4 Status messages function 53 53 Status messages Display automat calibration CAL Display stored 0 value Displa Timit YLIM Displa duc orange AR Display control function CTRL Figure 5 33 Status messages With this function you can configure the status message display in the status line of the measured value display You can display a maximum of four different status messages The message CODE m o provides information on the current operating mode Coded display mode Decoded display mode and is therefore always displayed The device activates the functional check softkey 5 in the function display when it determines that the measured value was influenced This is the case e During the warm up phase e After entry of a code i e switch to Operator control mode e During automatic calibration AUTOCAL or remote calibration OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 71 Operation 5 5 Parameters Table 5 4 Status messages Status What appears in the display depends on fun
54. U lies at the start of the measuring range mba Define measuring ranges function 41 41 Define range MR Start End No 0 00 50 0 00 60 1 2 3 0 00 80 4 0 00 Figure 5 28 Define measuring ranges You can define a maximum of four measuring ranges whose start of scale values are allocated to the bottom value 0 2 4 mA and whose full scale values are allocated to the top value 20 mA of the analog output If the message Measuring ranges not plausible appears this means that autoranging is not possible If the start of scale and full scale values are 0 the measuring range is deactivated Note If a start of scale value other than 0 is defined you must read the section Preparations for start up temperature influence OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 5 Parameters 5 5 Parameters Parameters 50 El time constants 51 Limits 52 On off configuration 53 Status messages continue Figure 5 29 Parameters submenu The parameters menu contains all functions which are required for configuring the device After selecting the parameter functions in the main menu by pressing the fourth softkey Parameters the menu screen for the selection of parameter functions 50 to 53 appears Pressing the fifth softkey Continue brings you to additional parameter functions This menu is protected by the code of code level 1 OXYMAT 61 Operating Inst
55. a value for Time from AUTOCAL to AUTOCAL cycle time using subfunction AUWTOCAL Check cycle parameter AUTOCAL Check start via digital input If you previously configured a digital input using function 72 you can initiate AUTOCAL via a digital input The AUTOCAL start via cycle parameter and AUTOCAL start via digital input operating modes can be activated simultaneously in order e g to verify a weekly calibration and control this check via a digital input OXYMAT 61 56 Operating Instructions 12 2011 A5E00123067 02 Operation 5 3 Calibration Trigger AUTOCAL once In the AUTOCAL on state you can start an AUTOCAL sequence at any time using the softkey Trigger AUTOCAL once providing the device is in the Measure operating mode A sequence initiated this way has no influence on the time cycle of an AUTOCAL i e the cycle time continues to run independent of this After the initial trigger the dot disappears until the process is finished Abort AUTOCAL A running automatic calibration process can be exited at any time using the softkey Abort AUTOCAL With this all calibration data determined up to that point are discarded and the calibration data used before AUTOCAL was started are used further zero and span Canceling has no influence on the time cycle All valid calibration processes are retained AUTOCAL sequence The AUTOCAL sequence is configured in this submenu Autocal check sequence Ze
56. age can result if proper precautions are not taken NOTICE indicates that an unintended result or situation can occur if the relevant information is not taken into account If more than one degree of danger is present the warning notice representing the highest degree of danger will be used A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property damage Qualified Personnel The product system described in this documentation may be operated only by personnel qualified for the specific task in accordance with the relevant documentation in particular its warning notices and safety instructions Qualified personnel are those who based on their training and experience are capable of identifying risks and avoiding potential hazards when working with these products systems Proper use of Siemens products Trademarks Note the following WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation If products and components from other manufacturers are used these must be recommended or approved by Siemens Proper transport storage installation assembly commissioning operation and maintenance are required to ensure that the products operate safely and without any problems The permissible ambient conditions must be complied with The information in the relevant documentation must be obse
57. agram of the operating sequence with operating modes Changes are accepted CAL McTRL Main menu Functions menu Functions Diagnosis THC gt E Calibration Configuration gt Oo gt papoo s Jaz jeuy apo Induj apou uonreledo Display mode Query only following changes CE Accept L Measured value modifications lt o gt memory for analog Yes gt Functional check output function 77 No gt relay CTRL function 71 only following changes Figure 5 2 Operating sequence with operating modes OXYMAT 61 42 Operating Instructions 12 2011 A5E00123067 02 Operation 5 1 3 5 1 4 OXYMAT 61 5 7 General information Editing inputs The values in the menus shown in this chapter are meant as examples An active input field is shown with colons e g 10 as a limiter The cursor blinks under the number to be entered By pressing the ENTER button you finish your input and the value is stored If there are several input fields on one function screen the cursor positions itself at the next input field at the same time Note Confirm every entered value even the last of several values in a function before exiting the function with ENTER With the CLEAR button you can clear a number which you have begun to enter The cursor then jumps back to the first position of the input field Graphic symbols activated ON
58. as calibration 15 minutes Calibration with calibration gas 1 10 minutes Purging with sample gas 8 minutes Sample gas intermediate operation 30 minutes Calibration with calibration gas 2 5 minutes Calibration with calibration gas 3 8 minutes Calibration with calibration gas 4 11 minutes Purging with sample gas 8 minutes ONOAAR ODN Short term signaling contact in order to be able to start AUTOCAL on another device The specified AUTOCAL sequence is shown in the following function displays Autocal check sequence Zero gas 1 15 0 min Cal gas 1 10 0 min SG purging 8 0 min Int SG mode 30 0 min continue Figure 5 15 Example of AUTOCAL sequence OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation OXYMAT 61 5 3 Calibration Autocal check sequence 5 Cal gas 2 5 0 min 6 Cal gas 3 8 0 min 7 Cal gas 4 11 0 min 8 Int SG mode 8 0 min continue Figure 5 16 Example of AUTOCAL sequence Autocal check sequence 9 Sig cont I min 10 min 11 min 12 min continue Figure 5 17 Example of AUTOCAL sequence List for the AUTOCAL sequence Step AUTOCAL sequence Zero gas 1 Function code 1 Zero gas 2 Function code 2 Calibration gas 1 Function code 3 Calibration gas 2 Function code 4 Calibration gas 3 Function code 5 Calibration gas 4 Function code 6 Flush sample gas Function code 7 Sample gas in
59. authorized access by codes Commissioning Totality of measures and actions required to make a machine or system capable of running Containment system Gas channel within the gas analyzer Control panel Panel with operating elements used to make inputs on the device OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 141 Glossary Correction of cross interference Computed correction of the measured value falsification caused by the interfering gas Cross interference deflection The falsification of a measurement caused by an interfering gas Cursor Tool insertion mark writing mark input mark for identifying the current processing position of a program Decoded display mode Operating mode in which the measured value is displayed the device is partially or completely decoded and the functional check is active Dialog language Language in which communication between the user and device takes place Display unit Device component which outputs shows device information and visualizes the communication via the control panel Factory functions Function for device maintenance This function is protected by the highest code level and is only accessible to maintenance personnel Factory setting Standard settings of the device at the time of delivery Front panel Front part of a device usually with clear identification features e g device name manufacturer logo etc Function Numbered
60. binet e Nitrogen oxygen air high pressure variant Pay attention to the purity of the gas 4 6 The supply is here comes from a compressed gas cylinder with a pressure setting of 2000 to 4000 hPa over the sample gas pressure A sintered metal frit porous filter is in the coupling to prevent entry of dirt particles into the analyzer OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 33 Commissioning 4 2 Preparation for commissioning Introducing reference gas Always introduce the reference gas before beginning the measurements Reference gas should always continue to flow even if the measurements are temporarily interrupted The increased consumption caused by this is negligible when compared to reference gas line is sealed Compressed gas cylinder If the reference gas is taken from a compressed gas cylinder the reference gas line must be purged prior to commissioning You must then check the line for leaks because leakage losses are often greater than the reference gas consumption To do this close the valve at the compressed gas cylinder If the pressure indicator at the reducer value of the gas cylinder does not drop by more than 1000 hPa min the gas connection is sufficiently sealed The reference gas pressure must always be 2000 hPa or more above the sample gas pressure Checking the reference gas flow To do this proceed as follows 1 Close the sample gas inlet couplings 2 Lay a hose with an internal d
61. cnnoncccnnnoncnc canon cnc nano naan nano nn cc nan nn cc Edenia 54 5 3 4 Calibration setting function 23 rrerak aA E A LRE R 54 5 3 5 AUTOCAL function 24 02 tlf eiii ras aiii 55 5 3 6 Drift valles unction 285 2 234 ltd ernennen 62 5 4 Measuring ranges ici iia A a a 63 5 4 1 Select measuring ranges function 40 assssessrecssrsesrsrnssriransernnnannnnenntnnasnnnanntennnannannanennaananna 63 5 4 2 Define measuring ranges function 41 oonnncccnnnnicinnnnacccnnnnnccnnnnnnno nan nnnnn cnn rnnnn rca rra rara rr 66 5 5 Parameters ive cite ieee ets ciate N el oh es ati at ie ih dn A ae 67 5 5 1 Electric time constants function 50 ooccocnnnnnccinnnoncccnnnonccnnnn nono nnnononn nn EREE arrancar nn rra 68 5 5 2 Limits TUNCION S ae a elle kein ici 69 5 5 3 On offf nctions f nelion S2 ascii een sehen earmunlerrigennenndeenaden 70 5 5 4 Status Messages function incita da la rein 71 5 5 5 Graphical representation of measured values function 54 ccccconnccccninococccononnncnanancnnnnnancnnnnnons 72 5 5 6 Measured value display function 55 444srnnnnnennnnnnnennnnnnnennnnnnnnnnnnnnnnnnnnnnnnnn nn mann 74 5 5 7 LCD contrasti fUnctiON D6 curacion datada 74 5 5 8 Magnetic field frequency function 57 ooooonnnncccnnnnanocinnnonccnnononnnnn nono nnn nano nan ran n rca rro rn rra 75 5 5 9 Date time TUNCION 58 aci sr sia 76 5 5 10 Measuring point switching function 9 oocconnncccnnnnaccccnnnoncccnn
62. composition Operating hours operating date Service report RH no Date received Date returned Technician 134 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 ESD Directives B B 1 ESD guidelines Definition of ESD All electronic modules are equipped with large scale integrated ICs or components Due to their design these electronic elements are highly sensitive to overvoltage and thus to any electrostatic discharge The electrostatic sensitive components modules are commonly referred to as ESD devices This is also the international abbreviation for such devices ESD modules are identified by the following symbol CAUTION ESD devices can be destroyed by voltages well below the threshold of human perception These static voltages develop when you touch a component or electrical connection of a device without having drained the static charges present on your body The electrostatic discharge current may lead to latent failure of a module that is this damage may not be significant immediately but in operation may cause malfunction OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 135 ESD Directives B 1 ESD guidelines Electrostatic charging Anyone who is not connected to the electrical potential of their surroundings can be electrostatically charged The figure below shows the maximum electrostatic voltage which may build up on a
63. ctions 52 and 53 Fet 53 Fet 52 0 Fct 52 m ies Fct 53 Fct 53 Calibration CAL None CAL Oo CAL m CAL calibration running also in Autocal Measured value None STO a STO STO analog output applied to memory STO memory see function 77 Limit LIM None LIM o LIM LIM limit has been violated see function 51 Automatic None AR DAR AR measuring range switching measuring range actively possible see function switching AR 40 Functional check None O CTRL or m CTRL o CTRL m CTRL device is in the Warm CTRL functional check up phase or in Operator control cannot be switched mode or calibration is in progress off using function 52 If an error occurs during operation the message Maintenance request Fault or Measuring protection appears in the status line depending on the severity of the error This message is alternately displayed with the status messages In addition when communication is over the ELAN interface the message Remote is displayed also alternating with display of the device status 5 5 5 Graphical representation of measured values function 54 With this function you can follow the trend of the measured values for the last ten minutes or 24 hours on the display 54 G signal display Period 10 min Period 24 h Figure 5 34 Graphical measured value representation OXYMAT 61 72 Operating Instructions 12 2011 A5E00123067 02 Operation OXYMAT 61 5 5 Para
64. dance with these operating instructions Qualified Personnel Qualified personnel are people who are familiar with the installation mounting commissioning and operation of the product These people have the following qualifications e They are authorized trained or instructed in operating and maintaining devices and systems according to the safety regulations for electrical circuits high pressures and aggressive as well as hazardous media e For explosion proof devices they are authorized trained or instructed in carrying out work on electrical circuits for hazardous systems e They are trained or instructed in maintenance and use of appropriate safety equipment according to the safety regulations Operating Instructions 12 2011 A5E00123067 02 9 General information 1 7 Notes on warranty 1 7 1 8 1 9 10 Notes on warranty The contents of this manual shall not become part of or modify any prior or existing agreement commitment or legal relationship The sales contract contains all obligations on the part of Siemens as well as the complete and solely applicable warranty conditions Any statements regarding device versions described in the manual do not create new warranties or modify the existing warranty The content reflects the technical status at the time of publishing Siemens reserves the right to make technical changes in the course of further development Delivery information The respective
65. dditional equipment such as a wash bottle additional filters and pressure reducers Also provide for the removal of corrosive components or components that interfere with the measurement process by using appropriate absorption filters OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 21 Mounting and connecting 3 3 Connections o Gas sample probe heated if required Sample gas line heated if required Gas cooler Filter trap if required Sample gas pump Control value Flow indicator if device has no sample gas monitoring To sample gas input Condensation drain Figure 3 2 Example of gas preparation 3 3 1 5 Gas flow diagram Figure 3 3 Gas path OXYMAT 61 with installed reference gas pump OXYMAT 61 22 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting 3 3 Connections E49 Figure 3 4 OXYMAT 61 gas channel with reference gas connection 3 000 to 5 000 hPa absolute Key for gas flow diagrams 1 Sample gas inlet 9 Purging gas 2 Sample gas outlet 10 Restrictor in reference gas channel outlet 3 n c 11 Pressure switch for reference gas monitoring 4 Reference gas inlet 12 Pump 5 Restrictor in reference gas channel 13 Filter 6 Analyzer section 14 Flow indicator in sample gas channel optional 7 Restrictor in sample gas channel
66. e e g for measurement range switchover processing of external signals from sample preparation Can be extended by eight additional digital inputs and eight additional relay outputs for automatic calibration with max four calibration gases RS485 standard Optionally available AK interface for the automotive industry with extended functions Converter for RS232 USB and Ethernet Integration in networks via PROFIBUS DP PA interface SIPROM GA software as the service and maintenance tool Operating Instructions 12 2011 A5E00123067 02 13 Description 2 2 Design Status line for display of analyzer state configurable Two code levels according to NAMUR maintenance and specialist levels Display of activated measuring ranges Five softkeys for menu control ESC key for returning to previous input step with application of any changes made INFO key for help functions for the current menu MEAS key for returning to measuring mode CLEAR key to delete numerical inputs ENTER key to accept numerical values 1 2 3 4 5 6 7 8 o 10 Numerical keypad for entering numerical values 11 Display of start of scale and full scale values 12 Display of concentrations as digits and bars Figure 2 1 Membrane keyboard and graphic display OXYMAT 61 14 Operating Instructions 12 2011 A5E00123067 02 Description 2 3 Communication interface 2 3 Communication interface Pa Container A EN j Y nr
67. e causes Remedy S8 Signal of selected pressure Sample gas blocked at outlet Caution Q sensor outside tolerance gt 2 000 hPa with internal pressure The internal pressure sensor may be sensor or gt 3 000 hPa with external destroyed if the system pressure or system pressure too high exceeds 4 000 hPa 1 Perform a suitable action to reduce the flow resistance at the device inlet until the sample gas is back within the tolerance limits 2 Regulate the system pressure accordingly 3 Check for leaks see section 4 2 Preparation for commissioning If leak is found contact service department Sample gas pressure too low Set system pressure gt 500 hPa lt 500 hPa S9 Signal too high Sample gas pressure gt 3 000 hPa or Reduce pressure or O2 concentration Q Oz concentration in range or 2 000 to 3 000 hPa too high contact service department S10 RAM check RAM PROM Replace motherboard or Q contact service department 11 Reference gas supply failed Reference gas line leaks is Check reference gas flow see section interrupted or blocked 4 2 Preparation for commissioning Reference gas source is empty Connect new reference gas source 12 Mains power supply Mains voltage outside tolerance Line voltage must be within the Q limits specified limits specified by the rating plate 13 Hardware line frequency Mains frequency outside tolerance Connect power supply stabilizer Q limits Crystal or external ADC defective Replace motherboard 14 Measur
68. e g a screwdriver to pry them out and removed them 109 Appendix 6 5 Cleaning the device User Interface The front panel is washable We recommend using a damp sponge or cloth containing a mild detergent for this Yu should only apply slight pressure when cleaning especially in the display area to avoid damaging the thin membrane In addition ensure no water penetrates the interior of the device during cleaning Interior If necessary you can gently remove dust and dirt from the interior using a compressed air gun 6 6 Maintenance requests and fault messages The device is able to detect functional irregularities These appear depending on the severity as either Maintenance request or Fault in the status line of the display At the same time they are recorded in the logbook function 3 and can also be called up there Pressing the button next to the corresponding log entry acknowledges it It will appear again however if you do not eliminate the cause Each new message is written to the next free space in the logbook A total of 32 storage spaces are available which means that the oldest of 32 log entries is overwritten by a new entry A power failure deletes all log entries You can disable the log or delete the messages contained in it using function 60 The display of messages is especially distracting during testing They can be disabled under the function 87 You should not do this during normal operation sinc
69. e you need to be able to detect any faults that may occur Maintenance requests If there are indications of changes in internal device parameters Maintenance request appears in the status line of the display Changes of this type do not affect the measurement capability of the device or affect it only minimally the first time they appear In order to continue to guarantee the measuring capability however it may be necessary to carry out remedial measures If the relay output of the device was configured accordingly see also section 5 Function 71 external signaling may also be performed OXYMAT 61 110 Operating Instructions 12 2011 A5E00123067 02 Appendix 6 6 1 Maintenance requests The following error messages result in a maintenance request output in the display and are signaled externally if a corresponding relay has been configured using function 71 Function 87 can be used to switch off disable each maintenance request individually Each maintenance requests labeled with Q in the left column of the following table must be acknowledged No Message Possible causes Remedy Remarks w1 Calibration Calibration gas was changed Repeat calibration Calibration tolerance see also Q tolerance Drift characteristics Check that drift is normal function 78 l exceeded Drift of device according to technical specifications Zero point 1 of FSV week Span gt 1 of FSV week W2 Signal span to Reference gas contains too C
70. ed value greater than Sample gas pressure exceeds the Check the sample gas pressure and operating full scale value 5 pressure correction range of 2 000 or possibly reduce it or switch to an 3 000 hPa external pressure sensor with an appropriate measuring range Incorrect calibration of the measuring Repeat calibration and possibly check range the sample test S 15 Calibration aborted Fault if the device is in AUTOCAL Remedy the cause Q mode Fault in calibration via digital input S16 Sample gas flow is too low Ensure sufficient flow Causes of fault messages 114 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Appendix Spare parts list F 7 1 General information This spare parts list corresponds to the technical status of February 2011 The rating plate shows the year of construction of the gas analyzer coded in the production no SIEMENS OXYMAT 61 C MLFB no 7MB2001 0FA00 0AAO F Nr N1 JD 352 Prod no 230 V48 63Hz 70VA min 5 O2 A Siemens AG Industry Sector 76181 Karlsruhe German An order for spare parts must contain the following information Quantity Designation of the spare part Order no of the spare part Device name order number and production no of the gas analyzer for which the spare part is intended Example for ordering 2 measuring heads C79000 A3460 B525 for OXYMAT 61 MLFB 7MB2001 0FAO00 0AAO Prod no N1 JD 352 OXYMAT 61 Operating Instructio
71. ement one distinguishes between an input signal and an output signal Measured value The measured value is an output value which reflects a determined variable Measured value display Totality of the displayed information in Coded display mode and Decoded display mode The following are displayed for example measured value status line with status messages footer measuring ranges components etc Measuring point Location where a measured value is recorded Measuring point relay Relay to which a measuring point is allocated Measuring range Range within which the result of a measurement moves The measuring range has a reference number e g 1 It is characterized by a start of scale value and a full scale value Measuring range switching Automatic switching over of measuring ranges Also referred to as autorange in the software Membrane keyboard Keyboard whose keys are protected against atmospheric influences by a membrane Motherboard Board which contains the basic data and firmware for the device OXYMAT 61 144 Operating Instructions 12 2011 A5E00123067 02 Glossary Noise Operating mode Operating mode Operation level Totality of all phenomena which could disturb the transmission and or recording of information Various modes into which the device can be put by intervening from the outside Three modes are distinguished e Coded display mode e Decoded display mode e Operator
72. er of digits and the maximum number of decimal places Note that a maximum of five digits can be displayed decimal point also counts as one digit 5 5 7 LCD contrast function 56 74 With this function you can make the display contrast brighter or darker 56 LCD contrast MC brighter e darker e Basic setting e Test O Figure 5 38 Setting the LCD contrast OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 5 8 OXYMAT 61 5 5 Parameters If the contrast is misadjusted you can restore the factory settings using the third softkey Basic settings With the fourth softkey Zest you can carry out an LCD test Various test screens are then displayed in succession From Coded decoded display mode you can restore the basic setting by entering 8 8 8 8 ENTER Magnetic field frequency function 57 You can use this function for example to minimize vibration dependent frequencies superimposed on the analog output beat or even eliminate them in the ideal case 57 Magn field freq Frequency 8 086 Hz Default value 8 095 Hz E Figure 5 39 Magnetic field frequency To do this enter the desired frequency value in the Frequency input field by calling function 57 The permitted values for this are between 7 to 11 Hz Ifthe change to a specific frequency does not have the desired effect enter different frequency values Pressing the fifth softkey resets clock
73. eres ATEX Richtlinie des Europ ischen Parlaments und des Rates zur Angleichung der Rechtsvorschriften der Mitgliedstaaten f r Ger te und Schutzsysteme zur bestimmungsgem en Verwendung in explosionsgef hrdeten Bereichen Karlsruhe 22 06 2009 Siemens AG Markus Nj klas Wilhelm Eckhard Research amp Development Entwicklung Quality Qualit t Namg function 1 Funktion Name function Funktion Annex A is integral part of this declaration Anhang A ist integraler Bestandteil dieser Erkl rung This declaration certifies the conformity to the specified directives but contains no assurance of properties The safety documentation accompanying the product shall be considered in detail Diese Erkl rung bescheinigt die bereinstimmung mit den genannten Richtlinien ist jedoch keine Beschaffenheits oder Haltbarkeitsgarantie nach 443 BGB Die Sicherheitshinweise der mitgelieferten Produktdokumentation sind zu beachten Siemens Aktiengesellschaft Chairman of the Supervisory Board Gerhard Cromme Managing Board Peter Loescher Chairman President and Chief Executive Officer Wolfgang Dehen Heinrich Hiesinger Joe Kaeser Jim Reid Anderson Hermann Requarat Siegfried Russwurm Peter Y Solmssen Registered offices Berlin and Munich Commercial registries Berlin Charlottenburg HRB 12300 Munich HRB 6684 WEEE Reg Nr DE 23691322 Page 1 2 Operating Instructions 12 2011 A5E00123067 02 131 Appendix A T EC Declaration of Co
74. es contained in the device at the time of delivery without optional extensions Start up state Settable operating mode to which the device changes after the warm up phase has been successfully run through Status message Selectable output of various messages in the status line of the measured value display Subfunction Independent functional unit within a function Submenu Menu which is listed under a higher order menu item Time constant System parameter which determines the way a system relevant variable changes with time OXYMAT 61 146 Operating Instructions 12 2011 A5E00123067 02 Glossary Warm restart Restarting the device from its warm state with the last set parameters Warm up phase Time which the analyzer needs to reach operating temperature The warm up phase counts as one of the operating modes Weighting Evaluation of single factors of a solution approach with regard to their importance It has the effect that more relevant factors have a greater influence on the result Zero calibration Calibration of the zero point with a suitable zero gas Zero gas Gas used to calibrate the zero point Zero point Smallest possible point of a measuring range usually the start of the measuring range OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 147 Glossary OXYMAT 61 148 Operating Instructions 12 2011 A5E00123067 02 Index A Access code 91 Add on board 108 Ana
75. evice is ready for measurements again once the sample chamber has dried out However one of the inlets for the reference gas may become clogged and measurement may fail if there is extremely high contamination This will be apparent from extremely strong fluctuations in the display of measured values If such situations you need to clean the sample chamber as follows 1 Remove the measuring head as described in the section Removing the measuring head 2 Blow compressed gas into the sample chamber to clean it The compressed air passes through the sample gas outlet and through the reference gas channels in the upper part of the casing You can also flush the sample chamber with trichloroethylene or alcohol In such situations you need to dry it afterwards with a stream of dry air 3 Install the measuring head again Ifthe cleaning procedure described above does not produce the desired result the sample chamber must be completely removed and cleaned in an ultrasonic bath If this measure does not succeed you may need to replace the sample chamber OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Appendix Removing the analyzer section Proceed as follows for this 1 Pull out the measuring head connector cable out of its socket on the magnetic field terminal board 2 Remove the reference gas inlet hose from the analyzer section 3 Disconnect the hose at a suitable location 4 Remove the analyzer section t
76. f cross interference By pressing the first softkey you now select the type of cross interference correction The following distinctions are possible e No correction of cross interference e Correction of cross interference with constant influence of interfering gas e Correction of cross interference with variable influence of interfering gas via analog input e Correction of cross interference with variable influence of interfering gas via ELAN Note The correction of cross interference is deactivated for the duration of a calibration process zero or span It is active again when calibration has been completed and the device has returned to the Measure operating mode Correction of cross interference with constant influence of interfering gas 83 Interf correction With const interference correction Valid for range Interference Figure 5 62 Constant influence of interfering gas OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 95 Operation 3 6 Configuration With constant residual gas composition and low concentration of O2 a residual gas effect occurs which varies only by fluctuations in O2 content and can therefore be regarded as approximately constant The device must be informed of the value of the zero point offset interfering gas swing see Example 1 Example 1 The residual gas sample gas without O2 zero gas is 50 propane remainder is N2 N2 is used a
77. f operation In contrast to almost all other gases oxygen is paramagnetic This property is utilized as a measuring effect by the OXYMAT 61 channel Due to their paramagnetism oxygen molecules in an inhomogeneous magnetic field always move in the direction of increased field strength When two gases with different oxygen contents meet in a magnetic field a pressure difference is produced between them In the case of OXYMAT 61 one gas 1 is a reference gas N2 O2 or air the other is the sample gas 5 The reference gas is introduced into the sample chamber 6 through two channels 3 One of these reference gas streams meets the sample gas within the area of a magnetic field 7 Because the two channels are connected the generated pressure difference creates a flow This flow is converted into an electric signal by a microflow sensor 4 The pressure difference is proportional to the difference in the oxygen content of sample and reference gas The microflow sensor consists of two nickel plated grids which are heated to approximately 120 C Together with two supplementary resistors they form a Wheatstone bridge The pulsating flow results in a change in the resistance of the Ni grids This leads to an offset in the bridge which is proportional to the oxygen content of the sample gas Because the microflow sensor is located in the reference gas stream the measurement is not influenced by the thermal conductivity the specific heat or
78. gure 5 10 Setting the setpoints The function display shows the setpoint input with total calibration The third measuring range is chosen here as the leading measuring range With single calibration there is no choice for the leading measuring range 5 3 4 Calibration setting function 23 23 Total single cal Total range calibration Figure 5 11 Select calibration method With this function select between total and single calibration of the measuring ranges Total calibration means that you adjust one leading measuring range and all other measuring ranges are calibrated using the switching ratio If the total calibration is not activated as shown here on the function display you must calibrate every measuring range individually OXYMAT 61 54 Operating Instructions 12 2011 A5E00123067 02 Operation 5 3 Calibration 5 3 5 AUTOCAL function 24 Note You can only make use of automatic calibration AUTOCAL if your device contains additional electronics add on board If it does not a corresponding message will appear on the display when the AUTOCAL function is called Automatic calibration can only be started if the device is in the Measure operating mode 24 Autocal check Autocal check mode Autocal check sequence Autocal check cyclic parameters Autocal check Figure 5 12 AUTOCAL function Note The settings for Acal Check subfunctions 1 to 3 are valid both for AUTOCAL
79. hPa corresponds to one millibar mbar HS High switchover point HU Height unit 1 HU correspondence to 1 3 4 inch i e 44 45 millimeter HystS Start point of the hysteresis HystE End point of the hysteresis Hz Hertz i e In other words IEC International Electrotechnical Commission IP Ingress Protection Degree of protection K Kelvin kg Kilogram NP No parity I Liter LCD Liquid Crystal Display LED Light Emitting Diode LEL Lower explosion limit LS Low switchover point OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 List of abbreviations acronyms OXYMAT 61 C 7 List of abbreviations Abbreviation Meaning m Meter mA Milliampere Mat no Material number max Maximum min Minute min Minimum MC Measured component MLFB Machine readable product designation mm Millimeter MR Measuring range MR Maintenance request MS Span N2 Nitrogen NAMUR Standards working committee for measuring and control technology in the chemical industry NC No connection NO Nitrogen oxide No Number O2 Oxygen OP Odd parity PA Process Automation PA Polyamide a plastic commercial name e g nylon PC Personal computer PROFIBUS Process fieldbus PTFE Polytetrafluoroethylene a plastic commercial name e g Teflon Q Acknowledgment required RAM Random access memory memory with ra
80. hat the sample gas pressure in the analyzer does not accumulate When several devices are connected in series you must ensure that there are no restrictions in the gas channels of the downstream devices free exhaust gas flow You must remove any sample gas restrictors installed in the device The only restrictor that can be retained is one between the sample gas line and the first gas analyzer unit After removing the sample gas restrictors the sample gas monitoring pressure switches of the downstream analyzers will be inoperable To avoid error messages you must have disable the corresponding error messages Gas flow rate is too low in the configuration function of the software section 5 6 18 Function 87 Page 101 Error S16 You should also note in this case that the assignment of a relay to the Sample gas flow error message has no function If the sample gas flows to an exhaust gas line you must note the following e The flow resistance in the exhaust gas line should be kept as low as possible with a line as short as possible or with a large diameter junction e The exhaust gas line must be free of rapid pressure fluctuations If this is not the case either a special exhaust gas line must be installed or a damping vessel gt 1 I with restrictor must be installed between the device and exhaust gas line pneumatic low pass OXYMAT 61 20 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting
81. he device allows you to perform calibration manually or automatically function 24 The latter is only possible with an add on board which contains eight additional digital inputs and eight relay outputs The setpoints for the zero and span calibrations are entered using function 22 The drift values are set with function 25 the single or total calibration with function 23 Following the selection of functions 20 and 21 the required gases must be introduced manually OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 3 Calibration 5 3 1 Calibration of the zero point function 20 20 Zero calib Setpoint 0 00 viv Act value 15 60 v v Start calibration CANCEL Zero calibration In order to calibrate the zero the device must be in the Measure operating mode If this is not the case calibration is not possible The display will show the message Analyzer is not measuring The zero point is calibrated simultaneously for all measuring ranges even when the span of the measuring ranges is calibrated individually Only initiate calibration when the measured value actual value has stabilized after applying the zero gas Otherwise the calibration may be imprecise If there is a lot of noise or the measured value is unstable over time increase the time constant before calibrating function 50 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 51 Operation 5 3 Calibrat
82. he housing must be closed and grounded upon completion of all assembly and installation work Failure to comply with this can result in death personal injury and or damage to property Awarn ING Release of hazardous gases Gases may be released if there are leaks in the sample gas channel For this reason toxic and corrosive gases are not allowed to be be measured with this device Failure to comply with this can result in death personal injury as well as damage to property and the environment OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 31 Commissioning 4 2 Preparation for commissioning 4 2 Preparation for commissioning 4 2 1 General information Gas preparation Make all devices in the gas channel upstream from the analyzer ready for operation e g gas sampling devices gas cooling device condensate vessels filters and any connected controllers recorders or indicators Read carefully the instructions and information in the operation manuals of the devices Operation Before switching on the device familiarize yourself with its operation see section 5 of this manual Interfaces Make sure that all interfaces are properly assigned and configured Electrical connections Make sure that all electrical connections have been correctly made Read also the information in the Electric connection section OXYMAT 61 32 Operating Instructions 12 2011 A5E00123067 02 Commissi
83. he tag assigned to a device in the ELAN network or the first 16 characters thereof e The configuration behavior of the device is configured with the dent number softkey 2 Parameters 0 1 and 3 can be selected and have the following meaning Parameter Meaning 0 Only the PROFILE ID number is positively acknowledged 1 Only the device specific ID number is positively acknowledged 3 Only the PROFILE ID number for multivariable devices complex analyzers is positively acknowledged Note In order to work with the provided GSD and DD the value 1 must be entered for Ident number e Relay by PB The 8 relays of the additional electronics add on board can be controlled over PROFIBUS using this To allow activation none of these relays must be occupied by a device internal function Note The parameter Relay by PB is only available with software version 2 0 0 and higher of the PROFIBUS additional electronics add on board OXYMAT 61 102 Operating Instructions 12 2011 A5E00123067 02 Maintenance and servicing 6 1 Safety instructions A WARNING Work on an open device Disconnect the gas and power supply beforeopening the device Only carry out adjustments using appropriate tools for example to avoid short circuits on the electronic PCBs Leaks in the gas channel caused by incorrect installation or calibration may result in the release of gas which in turn could result in
84. heck reference gas Zero gas and reference gas Q high with zero much oxygen point adjustment Zero gas contains too much oxygen Check zero gas should be identical Signal voltage Q too low with Zero gas contains too little oxygen Check calibration gas span adjustment Reference gas flow is too low Check reference gas flow and correct if necessary An incorrect measuring range has been selected for the calibration Select the correct measuring range W4 Set clock The device was switched off Reenter the date and time See function 58 Q W6 Temperature of Ambient temperature outside Make sure that the ambient Q LCD is too high tolerances specified in the temperature is in the range or too low technical specifications 5 to from 5 to 45 C 45 C W7 Temperature of Ambient temperature too high Check ambient temperature Q analyzer section 245 C max 45 C especially with built in analyzers Measuring head temperature Contact service department too high 278 C only applies to non heated model With a heated sample No error chamber selected setpoint Wait until the analyzer temperature is too low or section has cooled down to heating is switched off the new setpoint Message W7 appears until the temperature new setpoint temperature has been reached OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 111 Appendix
85. iameter of 4 mm from the sample gas outlet nozzle into a beaker filled with water 3 Observe the glass The reference gas must be slow with 1 to 2 bubbles per second coming through the water Checking for leaks in the sample gas channel with hosed devices To do this proceed as follows 1 Close the reference gas connection 2 Establish a pressure of approx 100 hPa in the sample gas channel 3 Wait about 1 minute During this time the inflowing sample gas will adapt to the ambient temperature 4 Note the pressure 5 Wait another 15 minutes 6 Note the pressure again The sample gas channel is sufficiently tight when the pressure has changed by no more than 1 hPa 1 mbar within these 15 minutes OXYMAT 61 34 Operating Instructions 12 2011 A5E00123067 02 Commissioning 4 2 Preparation for commissioning 4 2 3 Initial commissioning Switch on the power supply After you have ensured that all connections have been made and there are no leaks switch on the device After a short period the measured value display appears on the operator display The status display is located above this in the top row for details on the operator display see section 5 1 The measuring head is in the warm up phase during the first five minutes During this time the display signals CTRL function check Setting the measuring span Specify the desired spans measuring range full scale value measuring range start of scale value with
86. imensions analyzer See Section Dimensional drawing Weight analyzer Approx 13 kg Table 8 2 Technical specifications power supply Power supply Auxiliary power 100 to 120 V AC rated range of use 90 V to see rating plate 132 V 48 to 63 Hz or 200 to 240 V AC rated range of use 180 V to 264 V 48 to 63 Hz Power consumption analyzer Approx 37 VA Fuse ratings 100 120 V 1T 250 200 240 V 0 63T 250 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 123 Appendix Table 8 3 Technical specifications gas inlet conditions Gas inlet conditions Permissible sample gas pressure 800 to 1 200 hPa absolute with external reference gas supply atm pressure range with built in pump Sample gas flow 18 to 60 I h 0 3 to 1 l min Sample gas temperature 0 to 50 C Sample gas moisture content lt 90 RH Table 8 4 Technical specifications time response Time response Warmup time At room temperature lt 30 min 2 Display response time too 3 58 Damping electronic time constant 0 to 100 s can be configured Dead time purging time of the gas channel inthe Approx 0 5 to 2 5 s depending on model device at 1 l min Time for signal processing in the device lt 1s Table 8 5 Technical specifications pressure correction range Pressure correction range Pressure sensor 500 to 2 000 hPa abs Table 8 6 Technical specifications measuring characteristics Measuring characteristics
87. ing and connecting 3 3 2 2 OXYMAT 61 3 3 Connections Connection of the signal cables AR WARNING Inadequate signal lines If signals e g 4 to 20 mA analog output are to be introduced into a hazardous area of zone 1 they must be intrinsically safe Additional upgrading or retrofitting of the device with energy limiting modules is also required in this case Failure to comply with this can result in death serious personal injury and or damage to property The Ex marking of these energy limiting modules must be clearly visibly on the device AL WARNING Only connect the signal lines to devices which have reliable electric isolation from their power supply e The connection lines to the relay outputs the digital inputs the analog inputs and the analog outputs must be shielded e The reference ground of the analog inputs is the potential of the housing e The analog output is floating Operating Instructions 12 2011 A5E00123067 02 25 Mounting and connecting 3 3 Connections e Asa measure to suppress sparking across the relay contacts e g limit relays RC elements must be connected as shown in the following figure Note that the RC element results in a drop out delay for an inductive component e g solenoid valve The RC element should be sized according to the following rule of thumb R Q 02xRL Q C pF I A You must also use a non polarized capacitor for the RC eleme
88. injury to personnel e g poisoning symptoms as well as damage to the equipment itself e g danger of corrosion Maintenance The device must be subjected to annual maintenance to verify its electrical safety and correct operation in particular the tightness of the sample gas channel containment system The operator can extend the maintenance interval in individual cases if no negative effects are expected with respect to chemical corrosion to the seals that come in contact with the sample gas To allow maintenance work the top cover can be removed and the front panel swung forwards If maintenance work is interrupted for more than two hours the analyzer must be closed again Service All modules within the housing can be replaced in the event of a defect Repairs to these modules are not envisaged OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 103 Appendix 6 2 Analyzer section 6 2 1 Design of the analyzer section Analyzer section 104 The analyzer section consists of the functional parts magnetic circuit sample chamber and the measuring head Measuring head Restrictor plate O ring measuring head Pressure sensor Electrical terminal board Retaining plate Cut strip wound core Tension spring OLISI Coil Mounting screws Bottom part of casing Sample chamber O ring sample chamber Top part of casing Reference gas connection OXYMA
89. ion 5 3 2 Span calibration function 21 A single or total calibration is carried out depending on the setting made with function 23 This function is specific for the component Single calibration Measure operating mode required 21 Span calib Calibrate Calibrate Calibrate Calibrate Figure 5 7 Carry out single calibration Single calibration is only possible if the 7ofa calibration subfunction has been deactivated in function 23 The display will show then number of measuring ranges that were set using function 41 The function screen above is therefore an example of the single calibration of four measuring ranges If you would like to calibrate measuring range 3 for example press the corresponding softkey The setpoint and current actual value of the selected measuring range then appear in the display 21 Sp cal MR 3 Setpoint 80 00 viv Act value 79 79 viv Start calibration CANCEL Figure 5 8 Calibrating the measuring range single calibration OXYMAT 61 52 Operating Instructions 12 2011 A5E00123067 02 Operation OXYMAT 61 5 3 Calibration When the actual value has stabilized you can initiate the calibration process by pressing the fourth softkey The actual value is now made to agree with the setpoint If you make a mistake during the calibration e g wrong calibration gas reload the original calibration by pressing the Cancel calibration softkey Total calibra
90. log output 78 89 Analyzer section 104 107 Design 104 Measuring head 105 Sample chamber 105 Spare parts 117 Analyzer status 48 Application examples 37 Area of application 11 Associated gas 94 128 AUTOCAL Check 56 Connection 30 Cycle 60 Function 55 Operating mode 56 Sequence 57 B Board Replacing 108 C Calibration 107 AUTOCAL 55 Automatic calibration 55 Calibration tolerances 89 Measuring range 52 Sensitivity 52 Setpoints 54 Single calibration 52 54 Total calibration 53 54 Zero point 51 Cleaning Front panel 110 Interior 110 Sample chamber 106 Climatic conditions 127 Code 91 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Commissioning Initial commissioning 35 Preparations 32 Safety instructions 18 31 Communication interface 15 Configuration 78 Analog output 78 Correction of cross interference 94 Device test 91 Digital inputs 82 ELAN 85 Pressure correction 93 Relay assignment 80 Temperature compensation 99 User data memory 86 Connection Electrical connections 20 Gas connections 20 Mains connection 25 Reference gas 22 33 Sample gas 21 Contrast 74 Control keys 14 40 Control panel 39 D Date 76 Device parameters 48 Device test 91 Dialog language 93 Digital inputs 82 See Digital inputs Dimension drawings 130 Display Contrast 74 Measured value 39 43 74 Display field 14 39 Display unit 14 39 72 Contrast 7
91. lp of this function 84 Phase adjust 0 144349 9 231 2 Meas value 20 95 Phase adjust Figure 5 65 Phase adjustment With the maximum possible signal sample gas e g air the gain is automatically adjusted so that V assumes a value of approximately 500 000 The phase angle is then calculated using this signal value and saved whereby V is maximized and W minimized This angle was determined at the factory and should only be readjusted when the magnetic field frequency is changed OXYMAT 61 98 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 16 5 6 17 OXYMAT 61 5 6 Configuration Switch valves function 85 With this function you can manually switch up to six valves This is done via the relays which are allocated to the individual valves and are available on the motherboard and add on board The precondition is that the corresponding relays have been previously configured with function 71 Relay outputs 85 Switch valves 01 Zero gas Figure 5 66 Switching valves The Switch valves function only applies to the relay functions Zero gas Calibration gas 1 to 4 and Sample gas Only one of the six valves can be switched at a time since the valves are mutually interlocked Linear temperature compensation function 86 The OXYMAT 61 is temperature compensated for sensitivity If an additional temperature error occurs during operation for instance due to light contamination of the sample
92. maller than or equal to Equal to Not equal to Congruent Plus minus Approximately equal ES Inch 1 inch corresponds to 25 4 mm Percent 1 percent corresponds to the hundredth part of the whole vol Volume percent R Degrees angular C Degrees Celsius F Degrees Fahrenheit A Amperes AC Alternating current AC Automobile channel Approx Approximately ASCII American Standard Code for Information Interchange AUTOCAL Automatic calibration CHa Methane cm Centimeter co Carbon monoxide CO2 Carbon dioxide D Slow blow fuse values DB Data bits DC Direct current DD Device Description DIN Deutsche Industrie Norm German industrial standard DP Decentralized periphery a PROFIBUS version DS Pressure switch OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 137 List of abbreviations acronyms C 7 List of abbreviations Abbreviation Meaning e g For example EEPROM Electrically erasable programmable read only memory ESD Electrostatic sensitive device ELAN Economical Local Area Network EN European standard EP Even parity EPROM Erasable Programmable Read only Memory ETX End of text Fet Function FCTRL Function check FKM Fluoroelastomer name according to DIN ISO 1629 FSV Full of scale value GA Gas analyzers GND Ground GSD Device master data h Hour from lat hora hPa Hectopascal 1
93. meters Select the desired time period with softkey 1 or 2 The device now graphs the measured value vs time display 10 min MC viv Parameter min0 1234567890 Figure 5 35 Measured value trend The most recent measured value is on the time axis at the left att 0 Select softkey 1 Parameters Here assign a certain measuring range to the measured value axis Meas value disp par Optimum meas val dis Range 1 Range 2 Range 3 Range 4 Figure 5 36 Parameters for measured value representation If you select Optimum measured value display the software automatically scales the measured value axis The device adapts the scale to the measured value scatter Operating Instructions 12 2011 A5E00123067 02 73 Operation 5 5 Parameters 5 5 6 Measured value display function 55 With this function you can configure the display of measured values 55 Select digits Suppress negative measured values Automatic Total digits Digits after 3 decimal point The decimal point counts as a digit Figure 5 37 Configuring the measured value display You have the following options for this e With softkey 1 you suppress the display of negative measured values e With softkey 2 Automatic you can activate the automatic display of the measured value with 5 digits The number of decimal places depends on the size of the measured value e With the softkeys 3 and 4 you can choose the total numb
94. n polyethylene foil at least 0 2 mm thick and a drying agent e g silica gel should be added In addition the transport container should be lined with a layer of kraft paper Photocopy the return delivery form fill it in and enclose it with the returned device In case of guarantee claim please enclose your guarantee card Return delivery addresses OXYMAT 61 Spare parts service Send your orders for spare parts to the following address SIEMENS SPA CSC Tel 00333 69066677 Fax 00333 69066688 1 chemin de la Sandlach F 67506 Haguenau France DP order form recipient 0011E Repairs To enable us to quickly identify and correct faults return the analyzers to the following address applicable until further notice SIEMENS SPA CSC Tel 00333 69066677 Fax 00333 69066688 1 chemin de la Sandlach F 67506 Haguenau France DP order form recipient 0011E Operating Instructions 12 2011 A5E00123067 02 133 Appendix A 2 Return delivery Returned deliveries form Repair Guarantee Customer name Person responsible Delivery address Phone Fax E mail Return address if different from address above Customer original Order number Device name MLFB no Production no Description of returned part Fault description Process data at location of use Operating temperature Operating pressure Sample gas
95. nancccnnnnncc canon cnc naar n cnc naar RE T 77 5 5 11 Logbook settings function GO iesire a a RE 77 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Table of contents 5 6 Configuration een dei onen dia ida 78 5 6 1 Analog output function TOV iaa en a E E E 78 5 6 2 Relay outp lts MICHON TT raae A E E EA E O 80 5 6 3 Binaryinputs fUNGUON 72 area EE E TAS 82 5 6 4 ELAN configuration function Z3 jeo a E naar nennen 85 5 6 5 Reset unction TDi A TERO ER 86 5 6 6 Save load data function 9 irna A a E A 86 5 6 7 Suppression of short noise signals function 76 ooo nncccinnnniccnnnnococcnnnaccccnnnarccnnn ronca cnn 88 5 6 8 Measured value memory function 77 oooooccccnnnnccccnonaccccnonancccnnnanccc nano ncnnnnnn cnc r nana nc cc naar nccr nana nncnnns 89 5 6 9 Calibration tolerances function 78 oooooocnnnncccnnnnanoccnnnoncccnnnorccnnnnrrr anea E AE m i a aaki E RE 89 5 6 10 Change codes function 79 u 2444snnnnnnnennnnnnnennnnnnnennnnnnnennnnnnnennnnnnnennnnnnnnnnnnnnnnnnnnnnnnnn nn 91 SoTL Device test funetion GO iia 91 5 6 12 Language selection function 81 uusersnnesennnnnnennnnnnnennnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nenn 93 5 6 13 Pressure correction function 82 uuersnneennnnnnnnennnnnnnnnnnnnnnennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnn nn 93 5 6 14 Correction of cross interference function 83 24surnnnennennnnnnnennnnnnnnnnnnnnnennnnnnnn
96. ndom direct access RC R resistor C capacitor RD Received data R H RH Relative humidity RS232 Serial interface for data communication RS485 Serial interface for high speed data transfer s Second SB Stop bit s SG Sample gas SO2 Sulfur dioxide SSV Start of scale value STX Start of text SUB D Subminiature D form a plug system for data connections SW Software Operating Instructions 12 2011 A5E00123067 02 139 List of abbreviations acronyms C 7 List of abbreviations 140 Abbreviation t Meaning Time from lat tempus TD Transmitted data USB Universal Serial Bus V Volts V Version VA Volt ampere A Difference pA Microampere yF Microfarad 0 Phase angle Q Ohm OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Glossary Additional electronics Generic term for all add on boards Back wall of housing Part for sealing the rear housing opening of the device Calibration Elimination of deviations between the setpoint and actual value of certain measured variables Calibration gas Gas used for calibration procedures Code A selectable sequence of characters for enabling a protected submenu Code level Area of protected functions or states which are enabled after entering a certain code Coded display mode Operating mode in which the measured value is displayed and the device is protected against un
97. next function screen thus branching to further relays NOTICE Changing the configuration Every change to the configuration of the relay assignments should always be saved in the user data memory by means of function 75 If you neglect to do this there is a danger that a previous unwanted configuration will be loaded by Load user data function 75 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration Table 5 6 Relay outputs possible functions Function Relay is passive when Relay conducts Remark current Not assigned Relay is permanently passive zero current Fault Fault Also shown in the status line of the measured value display Maintenance request Maintenance request Calibration Calibration running For identification purposes Measuring range 1 to 4 Measuring range 1 Measuring range to 4 on identification Limit 1 to 4 Limit 1 to 4 triggered Limit monitoring Functional check Functional check Decoding Signal when CTRL active e g when warm up phase e Device is decoded changing to Operator Autocal running e Warm up phase control mode in the approx 30 min warm up phase when e Calibration running Autocal is running Autocal when calibration is e Remote running Sample gas Supply of sample gas Valve actuation for Autocal Zero gas 1 Supply of zero gas Calibration gas 1 to 4 Supply of calibration gas Meas
98. nformity IENS Annex A to the EC Declaration of Conformity Anhang A zur EG Konformit tserkl rung No A5E00484554 03 Product description Gas analyzer Gasanalysator Produktbezeichnung OXYMAT 61 OXYMAT 6 E F OXYMAT 64 E Type Typ 7MB2001 xxA00 xxxx Z xxx 7MB2011 xax0x bcdx Z e f xxx 7MB2017 xax0c bcdx Z e f xxx 7MB2021 xxxx0 xxxx Z g xxx 7MB2027 xxxx0 bxxx Z g xxX 7MB2041 xxxxx Xxxx Z HXXX a b c d e f g x see tables page 2 siehe Tabellen Blatt 2 Conformity to the Directives indicated on page 1 is assured through the application of the following standards depending on versions Die Konformit t mit den auf Blatt 1 angef hrten Richtlinien wird nachgewiesen durch die Einhaltung folgender Normen variantenabh ngig Directive Standard Reference number Edition f E31 a A C E a A C E a A C E a A C E a A C E Richtlinie Norm Referenznummer Aus x 0 9 E32 E33 b 0 1 b 0 1 b 0 1 b 6 7 b 2 3 gabe or oder E38 E39 c B c B c B c B c B datum A Z g E30 d A B G d A B G d A B G d A B G d A B G e E40 e E11 e E12 2004 108 EC_ EN 61326 1 x X X X X x X 2006 95 EC__ EN 61010 1 2001 X X X X X X X 94 9 EC EN 50014 1997 X X 94 9 EC EN 60079 0 2006 x 94 9 EC EN 50016 1995 x x 94 9 EC EN 60079 2 2004 x 94 9 EC EN 50020 1994 X X 94 9 EC EN 60079 11 X 94 9 EC EN
99. nnnnnnn mann 94 5 6 15 Phase adjustment function 84 sesters a EE E A A RARR 98 5 6 16 Switch valves function 85 ya ai E AS E 99 5 6 17 Linear temperature compensation function 86 oooconncccinonncccccnnnonacnnononancnn nono nan n nono ncn rca nara 99 9 6 18 gt Error On Off FUNCIONO as 101 5 6 19 PROFIBUS configuration function 9 ooonnnccccnnnociccnnnocccccononcncnnnoncncnnno cnn c nano nn nr nnno rca r rana mann 101 6 Maintenance and ServiciOQ ooooonccnnccnocononanonononoonnncnnnnnnnnnnn ccoo cnn nnnnn nono nono nnnnn nn corran nn n nn nnmnnn 103 6 1 Safely MASCAR 103 6 2 Analyzer section ci tdo td 104 6 2 1 Design of the analyzer SECtION ooooonnncccnnnnnniccnnnoccccnononcccnn narco ca ronca rca rr rr 104 6 2 2 Disassembling the analyzer section ooooonnncccnnnniciccnnnanccnnnoncccnn corn ccnn acc rra 106 6 2 3 Removing the sample gas restrictor oooooonnccccnnncicinonoccccnnnorccannnorr cnn nn cnnn nr rca 108 6 3 Replacement of motherboard and add on board 44usrsssssnennennnnennnnnnnnnnennennnnen 108 6 4 Replacing USGS coccion cir ii data 109 6 5 Cleaning the device sucre ti rbd 110 6 6 Maintenance requests and fault messages oocoocccccnnocicicononoccnnnnnncccnn non n cana no noc nn rra 110 6 6 1 MEE 111 6 6 2 Eaults ri Aa tad 113 7 Spareparts list aoia A A E E AAE E EA AE AAE aE E 115 7 1 General information nenn anr denen 115 7 2 Analyzer SOCUOM il laicidad 117 7 3 Electro
100. ns 12 2011 A5E00123067 02 115 Appendix Overview of the modules Analyzer section 01 Electronics 02 Sample gas channel 03 Reference gas channel 04 OXYMAT 61 modules Figure 7 1 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 116 Appendix 7 2 Analyzer section Table 7 1 Part no Designation Order no Remarks 01 1 Analyzer section complete material no C79451 A3460 B31 1 4571 01 2 Measuring head C79451 A3460 B525 01 2 1 O ring C79121 Z100 A32 1 item 01 3 Sample chamber material no 1 4571 C79451 A3277 B535 01 4 Magnet terminal board C79451 A3474 B606 01 5 O ring FPM Viton C71121 Z100 A159 1 item OXYMAT 61 analyzer section OXYMAT 61 Operating Instructions 12
101. nt Connector SUB D 25F Power supply 24 V max vay R 0 0 2 x R 0 C HF 1 A Da a a O e O O 6 a 25 o pS Figure 3 5 Example of spark suppression on a relay contact When operated with direct current a spark suppression diode can be installed instead of the RC element e Connect the signal lines to the SUB D plug connectors at the rear of the device e Refer to the ELAN interface description Order No C79000 B5200 C176 German C79000 B5276 C176 English for details on the interface cable OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting 3 3 2 3 Pin assignments Connector SUB D 9F RS 485 GND GND R_Level N NC RD TD N RD TD P 5V R_Level P NC NC Connector SUB D 15F NC NC Digital inputs Analog inputs analog outputs z Analog output 1 P Analog output 1 N Analog input 2 P Analog input 2 N 3 3 Connections Terminating resistors for the bus can be connected to pins 7 and 9 Analog outputs floating mutually isolated as well R lt 750 Q Correction of compressed or interfering gas Analog inputs non floating O to Analog input 1 P i 20 mA 500 Q or Analog input 1 N Correction of 0 to 10 V low Digital input 6 P interfering gas impedance Digital input 5 P Digital input 5 to 6 N GND Connector SUB D 25F A A GND AAA Digital input 4 P Digital inputs
102. o 100 vol O2 suppressed The reference gas flow is set zero point with measuring O2 2 000 to 4 000 hPa above sample gas automatically to 5 to 10 range pressure ml min full scale value 100 vol max 5 000 hPa absolute O2 Around 21 O2 suppressed zero with 21 Air Atmospheric pressure with internal vol O2 within the span reference gas pump OXYMAT 61 126 Operating Instructions 12 2011 A5E00123067 02 Appendix 8 3 Zero point error Table 8 11 Zero point errors of different accompanying gases Associated gas concentration 100 vol Zero point Associated gas Zero point deviation in concentration 100 vol deviation in vol vol absolute absolute Organic gases Inert gases Ethyne acetylene C2H2 0 29 Argon Ar 0 25 1 2 butadiene C4H6 0 65 Helium He 0 33 1 3 butadiene CaHe 0 49 Krypton Kr 0 55 n butane CaH10 1 26 Neon Ne 0 17 Isobutane CaH10 1 30 Xenon Xe 1 05 1 butene C4Hs 0 96 Inorganic gases Isobutene C4H8 1 06 Ammonia NH3 0 20 Dichlorodifluoromethane CCl2F2 1 32 Hydrogen bromide HBr 0 76 R12 Ethanoic acid CH3COOH 0 64 Chlorine Cle 0 94 n heptane C7Hi6 2 40 Hydrogen chloride HCl 0 35 n hexane CeH14 2 02 Dinitrogen monoxide N20 0 23 Cyclo hexane C H12 1 84 Hydrogen fluoride HF 0 10 Methane CHa 0 18 Hydrogen iodide HI 1 19 Methanol CH30H 0 31 Carbon dioxide CO2 0 30 n
103. octane CeH18 2 78 Carbon monoxide co 0 07 n pentane CsH12 1 68 Oxygen O2 100 Isopentane CsH12 1 49 Nitric oxide NO 42 94 Propane C3H8 0 87 Nitrogen N2 0 00 Propylene C3He 0 64 Nitrogen dioxide NO2 20 00 Trichlorofluoromethane R11 CClaF 1 63 Sulfur dioxide SO2 0 20 Vinyl chloride C2H3Cl 0 77 Sulfur hexafluoride SFe 1 05 Vinyl fluoride C2H3F 0 55 Hydrogen sulfide H2S 0 44 1 1 vinylidene chloride C2H2Cle 1 22 Water H20 0 03 Hydrogen H2 0 26 Zero point error due to diamagnetism or paramagnetism of some residual gases with nitrogen as the reference gas at 60 C and 1 000 hPa absolute according to IEC 1207 3 Conversion to other temperatures For this multiply the deviations from the zero point listed in Table 3 2 by a correction factor k e For diamagnetic gases k 333 K 0 C 273 K e For paramagnetic gases k 333 K 8 C 273 K All diamagnetic gases have a negative deviation from zero point OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 127 Appendix 8 4 128 Materials used in the sample gas channel Standard Gas channel 19 rack unit With hoses Procedure 1 4571 Hose FPM Viton Sample chamber 1 4571 Nozzles for sample chamber 1 4571 Restrictor PTFE Teflon O rings FPM Viton Hose coupling Polyamide 6 Options Flow indicator Measuring tube Duran glass Floa
104. of 5 to 45 C is maintained during operation see Technical Specification section Magnetic fields Magnetically sensitive devices should not be installed in close proximity of OXYMAT 61 since it is inherently emits stray magnetic fields Depending on the sensitivity distances up to 50 cm are required see also Function 57 OXYMAT 61 18 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting 3 3 Connections 3 3 Connections 15 pin connector Digital 9 pin inputs and analog inputs connector and outputs RS485 Reference gas inlet Sample gas outlet Sample gas inlet Purge gas inlet 9 pin interface connector optional e g PROFIBUS Power supply onnection and miniature fuses 37 pin connector Digital inputs and relay outputs tional 25 pin connector Digital optional Gas connections Nozzle 6 mm or 1 4 inputs and relay outputs Figure 3 1 Back side with gas and electrical connections OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 19 Mounting and connecting 3 3 Connections 3 3 1 Gas connections and internal gas flow diagram 3 3 1 1 Sample gas line Connector stems with a pipe diameter of 6 mm or 1 4 are provided for gas connections For the gas supply and discharge from the device you need to select a suitable material for the sample gas CAUTION Dynamic pressure in the sample gas channel The device must be operated in such a way t
105. ogether with the retaining plate and lift both out of the analyzer 5 Remove the analyzer section from the retaining plate 6 Remove the sample gas lines from the analyzer section Removing the sample chamber Calibration Leak test OXYMAT 61 Proceed as follows for this 1 Remove the measuring head as described in the section Removing the measuring head 2 Insert a suitable tool e g screwdriver between the retaining plate and the cut strip wound core U core and pry out the retaining plate until the tension spring is released 3 Remove cut strip wound core and retaining plate 4 Loosen the four mounting screws and pull the two parts of the casing apart The sample chamber is now accessible and can be removed To install carry out the above steps in reverse order It is important to note the following in this regard e Check all O rings to ensure they are OK and replace them if they are damaged e Tighten the mounting screws diagonally opposed with a torque of 6 Nm Each time a measuring head is replaced or the analyzer section reassembled you need to recalibrate the device This procedure is described in section Calibration Page 50 A leak test must be carried out following every maintenance measure that affects the analyzer section or the gas channel This procedure is described in section Sealing Page 35 If the device fails the leak test you must replace all seals and hoses or pipes Operating In
106. omponent Buttons with varying function softkeys Display of activated measuring ranges with the current range marked Limit marker on bargraph Start of scale and full scale values OMAN DOA FWD Analog measured value display bar chart graph of the measured value with the start of scale and full scale value of the current measuring range 10 Measured value Figure 5 1 Measured value display and control panel OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 39 Operation 5 7 General information 40 Table 5 1 Button meanings Button Meaning function CLEAR Clears the commenced number input ENTER Every number entered except fast activation of a function must be confirmed with ENTER ESC Jumps back one step in the operating structure Changes are accepted without need for confirmation INFO Information about current menu current function MEAS Jump back from every position in the operating structure to the Decoded display mode you may be asked to confirm the entered data Pressing the MEAS button again changes to Coded display mode i e another change in the Operator control mode requires the entry of the corresponding code Softkey Varying function possible here are e Submenu selection function selection e Selection of a subfunction e ON OFF switch function OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 1 2 Device operating modes
107. on
108. oning 4 2 Preparation for commissioning 4 2 2 Selection of the reference gas When you select the reference gas you must ensure that the various measurement spans have at least one common point This point is then defined as physical zero point This applies to all measuring ranges Once this point is known the reference gas can be selected This is explained by the following example There are four measuring ranges e 17 22 02 e 15 25 O2 e 0 25 02 e 0 100 O2 These four measurement ranged have the spans from 17 to 22 O2 in common The physical zero point should lie within this range Air 20 95 O2 is suitable as the reference gas in this case An exception to this principle is possible if the smallest measuring span 25 O2 and the distance to the reference gas is not more than 20 Oz Then the physical zero point may be located outside the measuring range In this case however the pressure correction see Function 82 in section 5 must be activated since there a pressure relationship due to the large zero point offset The purity of the reference gas has to be appropriate for the measuring task Installing the reference gas connection The reference gas connection is made depending on the order e Air low pressure variant Air as the reference gas is sucked in with the internal pump When the analyzer is mounted in a cabinet you must therefore ensure that the air sucked in is supplied from outside the ca
109. pressed this is displayed in the bottom line of the display After pressing the INFO key a message is displayed in plain text The MEAS and ESC keys retain their jump back function OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 91 Operation 3 6 Configuration Relay and binary test Analog test 92 Relay and Binary Relay Binary continue Figure 5 57 Test relay outputs and digital inputs The first function screen shows the state of the six relay and digital inputs of the standard device With an add on board there are another eight relay digital inputs on a second page With the relay test individual relays can be activated This is done via the input field With 1 the relay is on with 0 it returns to its release condition Numbers other than 0 and 1 are not accepted by the input field After leaving function 80 the relays are back in the state they had before calling the relay and binary test The current state of the digital inputs is displayed in the Binary column With the analog test the analog output is set to a constant current of 0 to 24 000 pA for testing purposes The analog input displays the input currents in pA OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration 5 6 12 Language selection function 81 With this function you set the device to a second dialog language 81 Select language MC English E Deutsch
110. pressing the first top softkey Analyzer status the following menu screen appears with additional choices Analyzer status 1 Analyzer config 2 Diagnostic values 3 Logbook 4 Display meas ranges Figure 5 5 Analyzer status submenu The analyzer status functions are freely accessible There is therefore no code query and no change in the operating mode The analyzer status submenu offers you various functions for displaying device parameters and stored data 5 2 1 Analyzer configuration function 1 When this function is selected important device manufacturing data can be viewed e Firmware No Order No of the software in the EPROM e Order No Information on device ordering data e Production No Indication of date of manufacture and device serial number e Object version Indication of the hardware design of the device e Software version and date Indication of the functional scope of the device OXYMAT 61 48 Operating Instructions 12 2011 A5E00123067 02 Operation 5 2 2 5 2 3 5 2 4 OXYMAT 61 5 2 Analyzer status Diagnostic values function 2 The most important diagnostic values may be called using function 2 They may allow conclusions to be drawn for evaluation of errors or setting work Logbook function 3 In the logbook all errors which led to a maintenance request W or a fault S are listed The limit alarm LIM and functional check CTRL are also registered However these do
111. ration tolerances to be checked by AUTOCAL Check Using functions 71 and 72 you can also select the relay output or digital input for AUTOCAL Check When a calibration limit is exceeded the maintenance request W10 is set and also if configured the relay Acal Chk Dif Both will be reset again after an error free AUTOCAL Check W10 is retained in the logbook When starting via cycle initiate AUTOCAL Check If this function is activated the device executes an AUTOCAL Check when AUTOCAL is started via a cycle In other words it only checks that the calibration tolerances are observed but does not carry out a calibration Operating Instructions 12 2011 A5E00123067 02 61 Operation 5 3 Calibration 5 3 6 62 Sequence 1 Start the AUTOCAL Check By means of the softkey Trigger AUTOCAL Check once in the subfunction Aca Check sequence Via digital input Via cycle 2 The device executes the sequence as configured in the subfunction Autocal Check sequence Drift values function 25 This function shows deviations occurring for the calibrations and AUTOCAL as well actual value setpoint as a sum parameter All rated zero point and sensitivity calibrations of each range are calculated here for the selected measured component Every new deviation is added to the existing drift value 25 Drift values Zero 0 00 vpm Span MR4 36 00 vpm Drift values reset o Figure 5 20 Drift values
112. remove it Remove the data plug from the rear of the device Unscrew the three M3 screws located between the plugs Pullout the plugs of the ribbon cables from the motherboard oa kw nN gt Remove the motherboard from the device OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Appendix Removing the add on board Mounting 6 4 Fuses F3 F4 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 The procedure is the same as for the motherboard In contrast to the motherboard the add on board is only attached to the back of the housing with two screws Proceed in reverse order to install the both board Replacing fuses AR warn ING Dangerous contact voltage You need to open the enclosure to do this work This makes live parts are accessible which may cause death or injury due to electric shock when touched For this reason you must always disconnect the device from the mains before replacing the board Improper handling The fuses may be only be replaced by trained or instructed personnel You may only use fuses of the same type The device is protected by multiple fuses the values of which depend on the device model e g supply voltage heating etc Power Fuse ratings 200 to 240 V 0 63 T 250 100 to 120 V 1 T 250 Fuse ratings in Ampere These fuses are located in a drawer of the mains connection box To replace the fuses use a suitable tool
113. ressing the associated softkey the code of the corresponding operation level is queried once exception the submenu Analyzer status is freely accessible selecting this does not change the operating mode Decoding level 2 also decodes level 1 By entering a corresponding code the device switches to the operator control mode whereby the functional check is activated The functional check CTRL shown in the status line of the measured value display is always activated by the device when an intervention endangers correct measurement e g when the code is entered If you have configured a corresponding relay using function 71 with Functional check an extra signal is sent out via the relay contact together with the decoding This relay contact then signals every functional check activation e g even the warm up phases and calibration states of the device Decoding also activates the measured value memory providing you have configured this using function 77 The coding status of the device can be read in the status line of the measured value display as a symbol m CODE for coded or 2 CODE for decoded Operating Instructions 12 2011 A5E00123067 02 45 Operation 5 7 General information 5 1 4 3 5 1 4 4 46 Returning to display mode You can use the MEAS key to jump back to Decoded display mode from any location in Operator control mode Any input started is aborted Before jumping back the
114. ro gas 1 1 0 min Zero gas 2 1 min Cal gas 1 2 min Cal gas 2 1 3 min continue Figure 5 14 AUTOCAL Check sequence A AUTOCAL can be composed of up to 12 individual steps Besides the supply of one zero gas and up to four calibration gases per component you can also program purging with sample gas sample gas intermediate operation as well as a signaling contact The signaling contact is only available if you have previously assigned it to a relay output using function 71 Sample gas intermediate operation Sample gas intermediate operation can be necessary if the system may only leave measuring mode for a certain time If the sum of the required purge times are greater than the permissible downtime you must return to the measuring mode between the individual calibrations OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 57 Operation 5 3 Calibration 58 Signaling contact Use the signaling contact to initiate an automatic calibration process of a second device for example or to signal the beginning or end of an AUTOCAL Relay outputs If you have defined relay outputs for the sample gas zero gas calibration gases and or measuring calibration function 71 these are used to activate the corresponding external solenoid valves The same applies to the signaling contact AUTOCAL This is closed for 1 s when the command is executed Example You want to program the following sequence Zero g
115. ructions 12 2011 A5E00123067 02 67 Operation 5 5 Parameters 5 5 1 68 Electric time constants function 50 50 Electr time con MC Effective bandwidth in of smallest MR 6 0 Time constant within bandwidth ti 10 Time constant outside bandwidth ta 1 Actual measured value 0 00 viv Figure 5 30 Electric time constants With this function you set various time constants which reduce the underlying noise in an message value signal The noise reduction is approximately equivalent to a low pass filter with the corresponding time constant The display delay is given as the 90 time The time constant ti acts within an configurable effective interval which is defined in of the smallest span It attenuates small changes in the measurement e g noise on the one hand but becomes ineffective immediately when the measured value exceeds the effective interval In this case the outer time constant ta attenuates the measured value You can configure values up to 100 for the effective interval and values up to 100 s for the time constants ti and ta By cleverly combining these three parameters you can achieve a low display delay 90 time despite high noise suppression The effect on the configured attenuation parameters can be observed in the bottom line The live measured value is displayed here OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 5 Parameters
116. rved All names identified by are registered trademarks of Siemens AG The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner Disclaimer of Liability We have reviewed the contents of this publication to ensure consistency with the hardware and software described Since variance cannot be precluded entirely we cannot guarantee full consistency However the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions Siemens AG Order number A5E00123067 Copyright Siemens AG 2001 Industry Sector 02 2012 Technical data subject to change 2011 Postfach 48 48 All rights reserved 90026 N RNBERG GERMANY Table of contents 1 General informati n ii iia 7 1 1 Device model and Validity narnii annae aaa nano cnn aaa rca nn 7 1 2 Information for our customers a e aa aae a a conan nn eaa nn nena nn nn aE Aaa a eaaa aea 7 1 3 Gone MINO MAIN cito it A A A a aa a a 7 1 4 Special information and Warnings stisne rnea a E ncnc nano nn nc nano nn EAA nc rcnnnancnnnns 8 1 5 Pr per USO iii oia 9 1 6 Qualified Personineliiisu u 2m reihe iii Pane Seyi add ceded 9 1 7 Netes on Walanty 2 2 22 A AA Ai ee eae 10 1 8 Deliveryinformation AAN 10 1 9 Standards and regulations sasani iiuair ENE rca 10 2 Description A em AA AE AE E AE aid 11 2 1 Area 0l applieation nee 11 2
117. s a reference gas 1 Determine the influence of interfering gas The diamagnetic zero point offset O2 equivalent of the propane is 0 86 O2 With content of 50 the interfering gas swing would be 0 43 Oz 2 Enter the interfering gas swing her 0 43 Oz Correction of cross interference with variable influence of interfering gas 83 Interference corr 0 With var interference o influence on an input 1 v 3 Valid for range 121314 O Interfering gas 100 causes an interference 42 94 Analog in 1 4 20 mA e for range 0 00 5 00 Figure 5 63 Variable influence of interfering gas Example of a variable cross interference correction The concentration of the interfering component must be measured with a suitable analyzer and the corresponding analog signal must be transmitted to the OXYMAT 61 for inclusion in calculations A pure residual gas should be specified as the interfering gas swing O2 equivalent By specifying the measuring range of the residual gas analyzer as a as well as its electricity output an internal calculation of the actual O2 offset can be made Example 2 A sample gas is composed of 4 NO and 96 Na It should be monitored for Oz The O2 equivalent of 100 NO amounts to 42 94 Oz The NO analyzer has a measuring range of 5 NO and an analog output from 4 to 20 mA OXYMAT 61 96 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Config
118. s required for configuring the device Configuration 70 Analog output 71 Relay outputs 72 Binary inputs 73 ELAN configuration continue Figure 5 43 Configuration submenu All functions of this menu are only accessible via the code of level 2 Press the fifth softkey continue to branch to further configuration functions 5 6 1 Analog output function 70 With this function you can configure the analog output 70 Analog output MC 4 20 mA o Output inverted a Suppress negative a measured values Figure 5 44 Configuring the analog output With softkey 1 define the start of scale value of the measuring range The following settings see also table Configurations of the analog output are possible e 0 20mA e 2 20mA e 4 20mA e NAMUR 4 20 mA with limit at 3 8 mA OXYMAT 61 78 Operating Instructions 12 2011 A5E00123067 02 Operation OXYMAT 61 5 6 Configuration You can invert the analog output with softkey 2 e g 0 20 O2 0 20 mA gt 0 20 O2 20 0 mA With softkey 3 you can suppress the negative measured values If a negative value has an unfavorable effect on further processing the negative measured values at the analog output can be limited to 0 or 2 4 3 8 NAMUR mA when this function is activated The display continues to show the actual measured value Table 5 5 Configurations of the analog output Defined analog Measuring range limit
119. sation 99 Zero calibration 51 Order no 115 Order No 48 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 P Parameters 67 Analog output 78 Code 91 Date 76 Digital inputs 82 Display contrast 74 ELAN 85 Graphical measured value representation 72 Limits 69 Logbook 77 Magnetic field frequency 75 Measured value display 74 Measuring point switching 77 On off functions 70 PROFIBUS 101 Relay assignment 80 Status messages 71 Time 76 Time constants 68 Pin assignments 28 AUTOCAL 30 Power supply 124 Pressure correction 93 Principle of operation 17 Process tags 77 Production No 48 PROFIBUS 101 Proper use 9 Purity monitoring 38 Q Qualified personnel 9 R Reference gas Connection 22 33 Line 105 Selection 33 Spare parts 123 Technical specifications 127 Relay assignment 80 Relay function 80 Reset 86 Return delivery 135 Room air monitoring 37 RS485 15 151 Index S Safety instructions Commissioning 18 31 General 7 Mains connection 25 Maintenance 103 Servicing 103 Signal connections 26 Sample chamber 105 106 107 Cleaning 106 Sample gas Associated gases 94 Conditioning 22 Connection 21 Correction of cross interference 94 Discharge 21 Interfering gases 94 Leak test 34 Line 21 Pressure 93 Spare parts 121 Sample gas restrictor 108 Self test 91 Sensitivity Calibration 52 Setpoint 54 Setting Analog output
120. scope of delivery is listed on the shipping documents enclosed with the delivery in accordance with the valid sales contract When opening the packaging please observe the corresponding information on the packaging material Check the delivery for completeness and undamaged condition In particular you should compare the Order No on the rating plates with the ordering data if available If possible please retain the packaging material since you can use it again in case of return deliveries Standards and regulations As far as possible the harmonized European standards were the basis for the specification and production of this device If no harmonized European standards have been applied the standards and regulations for the Federal Republic of Germany are valid When this product is used beyond the scope of these standards and regulations the valid standards and regulations of the country of the operating company apply OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Description Description 2 2 1 Area of application The OXYMAT 61 gas analyzers use a measuring principle based on the paramagnetic alternating pressure method and are used to measure oxygen in gases NOTICE The OXYMAT 61 is not suited for measuring flammable toxic and corrosive gases Application examples Measurement of oxygen e For boiler control in incineration plants e For quality monitoring in ultra pure gas e For en
121. software function of a device Functions are listed in submenus Function display Screen display depending on the called function OXYMAT 61 142 Operating Instructions 12 2011 A5E00123067 02 Glossary Functional check Identification activated by the device if it determines that the measured value was influenced e g when changing to operator control mode by decoding the device Gas analyzer Device for quantitative analysis of gases and gas mixtures Gas inlet Defined point for connecting a gas to the analyzer Gas outlet Defined point for directing a gas out of the gas analyzer Input field Single or multi line area for entering data Interfering gas A gas which interferes with the measurement and which may be contained in the sample gas Limit alarm Signaling of violation of a high or low limit Limit monitoring Function which monitors adherence to adjustable limits and signals if high or low limit is violated Limit relay Relay to which a certain limit is assigned and which is switched when there is a limit alarm Main menu Menu of the highest hierarchical level lt contains the submenus OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 143 Glossary Measured signal Representation of measured variables in the signal path by an allocated physical quantity of the same or different type Depending on the position of the measured signal at the input or output of the considered el
122. state also in status message in the status line deactivated OFF state also in status message in the status line gt access a submenu subfunction e trigger a function subfunction e g Start calibration Operating sequence CAL OCTRL 63 28 Vol 0 0 Figure 5 3 Measured value display Operating Instructions 12 2011 A5E00123067 02 43 Operation 5 7 General information 5 1 4 1 Introduction to the main menu The device must be in measuring mode The right side of the display field shows the measured component which is indicated with MC in the following illustrations of the display A right arrow gt appears for it This arrow indicates a softkey The main menu can be opened by pressing this softkey The main menu consists of the following commands with the associated code level on the right Function group Code level Analyzer status Not coded Calibration Code of code level 1 Measuring ranges Code of code level 1 Parameter Code of code level 1 Configuration Code of code level 2 The code of level 1 is factory set to 111 that of level 2 is factory set to 222 Main menu Analyzer status Calibration Measuring ranges Parameters Configuration OXYMAT 61 44 Operating Instructions 12 2011 A5E00123067 02 Operation 5 1 4 2 OXYMAT 61 5 7 General information Entering a submenu If you select a submenu by p
123. structions 12 2011 A5E00123067 02 107 Appendix 6 2 3 6 3 Removing the sample gas restrictor Ifthe sample gas restrictor is clogged from condensate and therefore needs to be cleaned or if it a hindrance to the measurement task for other reasons e g due to the use of OXYMAT 61 with gas analyzers of the type ULTRAMAT 6 in certain measuring configuration you must remove the restrictor If there is no a flow indicator optional in the device the sample gas restrictor is located in the sample gas hose between the pipe inlet nozzle and the analyzer section If there is a flow indicator the sample gas restrictor is located between it and the analyzer section The restrictor is secured with a hose clamp Remove the restrictor as follows 1 Loosen the hose section in which the restrictor is located 2 Remove the hose clamp over the restrictor 3 Use a suitable object stick etc to push the restrictor out of the hose Replacement of motherboard and add on board AL WARNING Dangerous contact voltage You need to open the enclosure to do this work This makes live parts are accessible which may cause death or injury due to electric shock when touched For this reason you must always disconnect the device from the mains before replacing the board Replacing the motherboard 108 To do this proceed as follows Disconnect the device from the mains Loosen the screws on the cover of the housing and
124. t type meters Duran glass Suspension boundary PTFE Teflon Angle pieces FPM Viton Sample gas Membrane FPM Viton pressure switch Enclosure PA 6 3T OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Appendix Dimension drawings 9 9 1 Dimension drawings OO OULU OXYMAT 61 465 483 440 AAA Aa O Aa cA Ad L 384 5 Figure 9 1 Dimensions for preparing installation in mm front view and plan view OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 129 Appendix 4 3949 gt p 351 5 a 27 5 pq 156 Figure 9 2 Dimensions for preparing installation in mm front view and plan view OXYMAT 61 130 Operating Instructions 12 2011 A5E00123067 02 Appendix Appendix EC Declaration of Conformity OXYMAT 61 SIEMENS Industry EC Declaration of Conformity EG Konformit tserkl rung No A5E00484554 03 Manufacturer Siemens AG Hersteller Industry Sector Industry Automation Division Address 76181 Karlsruhe Germany Anschrift Product description Gas analyzer Gasanalysator Produktbezeichnung OXYMAT 61 OXYMAT 6 E F OXYMAT 64 E Type Typ 7MB2001 xxA00 xxxx Z Xxx 7MB2011 xax0x bcdx Z e f xxx 7MB201
125. tch the device off and on again If the error message S1 appears again 2 Load user data function 75 3 contact service department Make sure you leave the device in operation in order to assist the fault diagnostics of the service staff 2 Magnetic field supply faulty Ribbon cable connection interrupted Check connection Motherboard faulty contact service department 3 Microflow sensor faulty One half of the grid has been Replace measuring head or contact Q destroyed service department S4 External fault message Signal from external source Check Function 72 must be configured accordingly S5 Temperature of analyzer section Ambient temperature outside Make sure that the ambient Q is too high or too low tolerances specified in the technical temperature is in the permissible range specifications 5 to 45 C from 5 to 45 C Measuring head temperature too Perform restart RESET if this is not high gt 80 C or too low lt 10 C successful Temperature sensor faulty contact service department gt Temperature rises above setpoint temperature S7 Temperature of measuring head More than 5 C deviation from the Replace measuring head or contact Q outside tolerance setpoint temperature 75 C service department OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 113 Appendix No Message Possibl
126. termediate operation Function code 8 Signaling contact Function code 9 59 Operating Instructions 12 2011 A5E00123067 02 Operation 5 3 Calibration AUTOCAL Check cycle parameters With this subfunction you can configure various time constants for activating a cyclically repeated AUTOCAL sequence Autocal check cycle Time from autocal to auto cal cycle time 0 h Time up to next autocal 15 min Carry out span calibration at any 1st cycle Total range calibration Cal gas Figure 5 18 AUTOCAL Check cycle parameters Time from AUTOCAL to AUTOCAL cycle time Any setting between 0 and 1 000 hours is accepted by the device 0 corresponds to no AUTOCAL cycle Time to first AUTOCAL after the time of setting If you enter 0 here and AUTOCAL is activated with AUTOCAL on off subfunction the device begins immediately with the AUTOCAL sequence If AUTOCAL is switched off the device only starts an AUTOCAL sequence if AUTOCAL is switched on within one minute after entering 0 If not the total time between two AUTOCAL cycles passes by starting with the input of 0 Note The clock inside the device also runs when AUTOCAL is deactivated It starts the first time the device is switched on with the factory set start time and must be set to the current time using function 58 Carry out span calibration for each xth cycle Here you set the number of cycles after which a calibration wi
127. th the calibration gas is to be carried out If you would like to save on calibration gas for example and not calibrate the span along with every zero calibration enter in the line Calibrate with calibration gas at every cycle an integer gt 1 OXYMAT 61 60 Operating Instructions 12 2011 A5E00123067 02 Operation 5 3 Calibration Information in the last two lines The information in the last bottom lines indicates that specified parameters relate to total calibration with calibration gas for measuring range 3 This measuring range was selected beforehand with function 22 Note As long as AUTOCAL is activated AUTOCAL 5 access to functions 20 and 21 is blocked If you activate these functions anyway a corresponding message appears on the display Settings for AUTOCAL Check OXYMAT 61 The AUTOCAL Check subfunction is only used for checking the calibrations Similar to AUTOCAL sequence set in the AUTOCAL Sequence menu is performed In contrast to AUTOCAL no new calibrations are performed only the deviations are checked with respect to selectable calibration tolerances Autocal check Calib tolerance at zero in of smallest MR 6 Calib tolerance at sens in of current Start only Autocal check o cyclically Trigger acal check once o Abort autocal check 0 Figure 5 19 AUTOCAL Check Functions of the AUTOCAL Check Inthe AUTOCAL Check subfunction enter the desired calib
128. the internal friction of the sample gas This also provides a high degree of corrosion resistance because the microflow sensor is not exposed to the direct influence of the sample gas A magnetic field with alternating strength 8 prevents detection of the background flow in the microflow sensor enabling measurement to be performed independent of the sample chamber position and thus also independent of the location where the gas analyzer is used The sample chamber is directly in the sample channel and has a small volume and the microflow sensor is a low lag sensor This results in a very fast response time for the OXYMAT 61 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Mounting and connecting Mounting and connecting 3 1 Safety instructions DA R Explosion hazard The analyzer must not be operated in potentially explosive atmospheres The supply of gases with combustible components in concentrations above the lower explosive limit LEL must be clarified in consultation with the relevant hazard experts and is ultimately the responsibility of the operator AR WARNING Dangerous contact voltage Certain parts of this device are under dangerous voltage In order to avoid contact with live parts the housing must be closed and grounded upon completion of all assembly and installation work Failure to comply with this can result in death personal injury and or damage to property AR WARNING
129. ther gas analyzer via ELAN It is basically possible to use other pressure measuring instruments or analyzers with pressure measuring function and ELAN capability for the pressure measurement On the device supplying the pressure data the parameter Measured value telegrams must be set to On using the function 73 5 6 14 Correction of cross interference function 83 94 Associated gases contained in the sample gas interfering gases can affect the measurement This function is used to reduce such interference or eliminate it altogether in the ideal case Note A correction for interfering gas is usually only useful if the corrected O2 equivalent is not greater than the smallest span When the reference gas and residual gas have different compositions sample gas without Oz content a zero point offset interfering gas swing occurs triggered by the paramagnetic or diamagnetic difference of the two gases The value of this zero point offset can be specified to compensate for this offset Note A differentiation must always be made with the correction of cross interference as to whether the interfering gas has a constant or variable concentration OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration Press the first softkey The following function screen appears on the display 83 Interf correct 0 No interference correct 0 Figure 5 61 Carrying out correction o
130. tion Measure operating mode required Total calibration is only possible if the 7ofa calibration subfunction has been activated using function 23 With a total calibration you calibrate all measuring ranges together Define the leading measuring range in function 22 We recommend that you use the largest measuring range for this After selecting this function the display shows the setpoint and the current value of the leading measuring range 21 Sp cal all MRs Setpoint 100 00 v v Act value 97 79 viv Start calibration 0 CANCEL e Figure 5 9 Calibrating the measuring range total calibration When the actual value has stabilized you can initiate the calibration process by pressing the fourth softkey The actual value is now made to agree with the setpoint If you make a mistake during the calibration e g wrong calibration gas reload the original calibration by pressing the CANCEL softkey Note Measuring accuracy With a switching ratio of the spans of more than 1 10 an individual calibration should be carried out for each range to achieve a higher measuring accuracy Operating Instructions 12 2011 A5E00123067 02 53 Operation 5 3 Calibration 5 3 3 Zero span setpoints function 22 22 Setpoints total setpoint for zero 0 00 viv setpoint for MR 1 50 00 viv Setpoint for MR 2 60 00 v v Setpoint for MR 3 80 00 viv Setpoint for MR 4 100 00 viv Fi
131. tips and helps avoid maloperations OXYMAT 61 8 Operating Instructions 12 2011 A5E00123067 02 General information 1 5 1 6 OXYMAT 61 1 5 Proper use Proper use Proper use within the context of this manual means that the product may be used only for the applications described in the catalog or the technical description and only in combination with the equipment components and devices of other manufacturers recommended or permitted by Siemens The product described in this manual has been developed manufactured tested and documented in compliance with relevant safety standards When the handling rules described for the configuration installation proper operation and maintenance as well at the safety guidelines are adhered to therefore there is normally no risk to the health of persons or in respect to damage to property This device was designed to ensure safe isolation of the primary and secondary circuits Low voltages that are connected must therefore also be generated with safe isolation AR warn ING Dangerous contact voltage After removing the housing or protection against direct contact or after opening the system cabinet certain parts of of this device system will be exposed that can carry hazardous voltage Therefore only appropriately qualified persons are permitted to perform work within this device These persons must be thoroughly familiar with all sources of danger and service activities in accor
132. to 100 O2 Measuring range 2 90 to 100 O2 Smallest span 100 95 O2 5 O2 Measuring range in which calibration takes place Measuring range 2 Selected calibration tolerance 6 Response threshold for zero point 5 O2 x 0 06 0 3 O2 Response threshold for span 10 O2 x 0 06 0 6 O2 If the zero span deviates by more than the configured value compared to the last calibration and if Signal tolerance violation is activated the corresponding relay signals a maintenance request OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration 5 6 10 Change codes function 79 With this function you replace the factory set codes 111 for level 1 222 for level 2 by your own 79 Codes program Code 1 2111 Code 2 222 Figure 5 55 Change codes If you enter the value 000 for a code code blocking is canceled and you have immediate access to the corresponding operation level 5 6 11 Device test function 80 You can check the functionality of the device by performing the following three tests 80 Analyzer test Keyboard Relay and Binary test Analog test Figure 5 56 Testing the device Keyboard test You can test all keys on the control panel by means of the keyboard test The five softkeys on the right edge can be used to make the associated item appear disappear If a number key or the sign key is
133. undergoing repairs or maintenance work or for example if new parameter settings are to be tested OXYMAT 61 86 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration Press softkey 3 to restore the factory settings A confirmation query is set up in this function In order to actually load the respective data in the memory you must confirm with yes If you select no this is cancelled NOTICE Make sure you save every change in the configuration of the relay outputs in the user data memory using function 75 If you neglect to do this there is a danger that a previous unwanted configuration will be loaded by Load user data function 75 The following figure shows an overview of the interaction between the various memory modules Program memory Main memory Parameter memory EPROM Flash PROM RAM EEPROM Factory data Fct 99 Program Running only operations service engineers User data Working area Factory basic data change Fct 99 only service engineers Figure 5 51 Memory modules OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 87 Operation 3 6 Configuration 5 6 7 88 Suppression of short noise signals function 76 With function 76 you can eliminate undesired spikes which exceed a settable threshold 76 Suppress fault Suppress noise signals with a ee ar up to 0 0 s Level in of smallest MR Figure 5 52
134. uration Correction of cross interference via ELAN 83 Interf correction With var interference influence on ELAN Valid for range Interfering gas causes an interference Channel 03 NO 5 Figure 5 64 Correction of cross interference with variable influence via ELAN If the cross interference correction is performed via the serial RS485 interface ELAN the same specification must be made as for the cross interference correction with a variable interfering gas influence Also enter the channel number and component number of the interfering gas analyzer The type of gas allocated to the channel and component the measuring range and possibly the device status appear in the display see also function 82 Pressure compensation With softkey 2 you can set whether this correction of cross interference should only apply to certain measuring ranges Note On the device supplying the correction data the parameter Measured value telegrams function 73 must be set to On OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 97 Operation 3 6 Configuration 5 6 15 Phase adjustment function 84 The physical principle of the measurement procedure and the mechanical design of the detector result in a delayed reaction phase offset of the analog measured value signal compared to the clock signal of the magnet controller This phase shift can be adjusted if necessary with the he
135. uring point 1 to 6 Measuring point 1 to For taking a gas sample via 6 selected solenoid valves at various measuring points Measuring point signal Measuring point 1 to For identifying the 1to6 6 selected measuring point runs parallel to measuring point Signaling contact When there s a e g in the case of signal the relay AUTOCAL conducts current for actuation of a second a short time device Sample gas flow Sample gas flow too For identification purposes low Reference gas pressure Reference gas For identification purposes pressure too low AUTOCAL Check Autocal difference too large function 24 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 81 Operation 3 6 Configuration 5 6 3 Binary inputs function 72 Digital input functions 82 Six floating digital inputs are included in the standard equipment 0 0 V 0 to 4 5 V 1 24 V 13 to 33 V If these six inputs are insufficient you can retrofit additional electronics with eight further digital inputs add on board 72 Binary inputs MC Fault Maint r CTRL NAMUR O Binary inputs define Figure 5 46 Digital input function The operation of the digital inputs are defined with softkey 1 If you activate the NAMUR operating mode the digital inputs behave as in table Digital input activation functions marked with N If you deactivate the NAMUR 0 operating mode the digital inputs behave compatibl
136. uring ranges depending on version 2 0 or 5 0 O2 Easy device replacement since electric connections can be simply disconnected from the device 19 rack unit with 4 height units U for installation in swivel frame 19 rack unit with 4 U for installation in cabinets with or without telescopic rails Front plate can be swung down for servicing purposes laptop connection Internal gas channels Hose made of FKM Viton Connections for sample gas Pipe diameter 6 mm or 1 4 Sample chamber of stainless steel mat no 1 4571 OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Description 2 2 Design 2 2 Design Display and control panel Inputs and outputs Communication OXYMAT 61 Large LCD for simultaneous display of Measured value digital and analog displays Status line Measuring ranges Contrast of the LCD adjustable via the menu Permanent backlighting with LED 5 digit measured value display decimal point counts as digit Washable membrane keyboard front panel Menu guided operation for parameter assignment configuration and calibration Operating help in plain text Graphic display of concentration trend programmable time intervals Two analog inputs configurable e g interference correction or external pressure sensor Six relay outputs freely configurable e g faults maintenance request calibration limit alarm external solenoid valves Six digital inputs freely configurabl
137. vironmental protection e For monitoring of process exhaust gas e Forprocess optimization OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 11 Description 2 1 Area of application Important features 12 Four freely configurable measuring ranges also with suppressed zero point all measuring ranges are linear Electrically isolated measured value output 0 2 4 to 20 mA Selectable automatic or manual measuring range switching remote switching is also possible Measured values can be stored during calibration Wide range of selectable time constants static dynamic noise suppression i e the response time of the device can be adapted to the respective measuring task Simple handling through menu guided operation according to NAMUR recommendation Fast response time Low long term drift Reference gas supply either externally N2 O2 or air approx 2000 to 4000 hPa via sample gas pressure or via built in reference gas pump ambient air Two control levels with separate authorization codes to prevent unintentional and unauthorized operator intervention External pressure sensor can be connected for correction of variations in the process gas pressure in the range 800 to 1200 hPa absolute Automatic measuring range calibration can be configured Fieldbus connection optional Monitoring of sample gas optional Monitoring of reference gas only with version with internal reference gas pump Variety of narrow meas
138. y with the software outputs of the older version V 4 3 0 in the table Digital input activation functions marked with X Press softkey 2 The following function screen appears on the display 72 Binary inputs D 1 Autocal Check D2 Vacant D 3 Vacant D4 Vacant continue Figure 5 47 Define digital inputs OXYMAT 61 Operating Instructions 12 2011 A5E00123067 02 Operation 5 6 Configuration Assign one of the activation functions listed below to each input as you like whereby every function may only be assigned once The connection assignments for the individual inputs can be found in the section Electrical connection No digital input has been preassigned in the factory Up to four digital inputs can be configured in a menu screen Press the fifth softkey continue to access the next function screen thus branching to further digital inputs NOTICE Save changes Make sure you save every change in the configuration of the digital inputs in the user data memory using function 75 If you neglect to do this there is a danger that a previous unwanted configuration will be loaded by Load user data function 75 Table 5 7 Digital input activation functions Function Necessary activation Remark effect voltage OV 24 24V pulse V min 1 s Not assigned No effect when activated External fault 1 N X e g due to to 7 e Signal from a gas preparation system
Download Pdf Manuals
Related Search
Related Contents
JEOL 7500F Quick User Guide LOREX Technology ACC-DRW User's Manual JobExchanger32 - MPL Automation DeepC User Manual Fonctionnement de l`application Diese Anleitung ist ein Vorausgabe der gedruckten Version Manual de usuario - Serie Vita II Grizzly G0703 11 Dust Collector User Manual 1. Monter un projet artistique et culturel - CRDP Basse Manuel d`utilisation et d`applications Copyright © All rights reserved.
Failed to retrieve file