Home

REGIUS MODEL 190 (DD-941)

image

Contents

1. 13 Life Expectancy of Parts 14 1 Overview 1 1 Installation Flowchart 1 1 1 2 Name of Parts 1 2 1 2 1 REGIUS 190 1 2 1 2 2 Option CS 2 3 LCD Mounting Table with Arm Table 1 3 1 2 3 Option CS 2 3 LCD Mounting Table with Arm PC Rack 1 4 1 3 Preparation for the Installation 1 5 1 3 1 Requirements for the Installation 1 5 1 3 2 Tools and Jigs for the Installation 1 6 1 4 Specifications
2. 1 7 2 Installation of REGIUS MODEL 190 2 1 Unpacking 2 1 2 1 1 Unpacking of REGIUSMODEL 190 2 1 2 1 2 Removing of Fixing Hardware 2 9 2 1 3 Checking of Fixing Hardware 2 17 2 2 Checking of Input Voltage 2 19 2 3 Checking after Installation 2 20 2 3 1 Power Connection 2 20 2 3 2 Power ON 2 20 2 3 3 Connecting to Network 2 21 2 3 4 Fixing of Adjuster Foot
3. 2 22 2 3 5 Checking of Cassette 2 22 3 Installation of the Optional Equipment 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table 3 1 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack 3 9 3 3 Installation of CS LCD Monitor Fixing Metal 3 14 i REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Forward Direct Digitizer REGIUS MODEL 190 hereafter referred to as REGIUS MODEL 190 is a device for reading an X ray image cassette that uses photostimulable uorescent material as an X ray detector This Installation Manual provides the procedures of installing the REGIUS 190 for a eld ser vice engineer Be sure to carefully read and understand this manual before installing the REGIUS 190 TRADEMARK Microsoft and Windows are trademarks of Microsoft Corporation USA in the U S and other countries Windows 2000 is an abbreviation of Microsoft Windows 2000 Professional operating system The company and product names described in this manual are trademarks of their respective companies Please note that the marks such as and are omitted in this manual ii REGIUS MODE
4. Gloves for work performed around the steel belt 1 7 1 4 Specifications REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 4 Speci cations Device Type Direct Digitizer DD 941 Model Name REGIUS MODEL 190 CODE No 0676 Safety Standard IEC60601 1 IEC60825 Film Size Note 1 14 X17 14 X14 11 X14 10 X12 8 X10 24X30cm 18X24cm Mammography 24X30cm 18X24cm Pantomography 15x30cm Sampling Pitch Note 2 43 75 m 87 5 m and 175 m Maximum Resolution 7080x9480 Digital Gradation Level 4096 steps12bit Cycle Time 90 sh hr 14 x 14 175 m Cassette Feed Load Time Note 3 approx 40sec 14 x 14 175 m Slots Insertion x 1 1 stack Exit x 1 4 cassettes in stack Dimensions 580 W 580 D 1230 H mm Weight approx 170kg excluding the cassette Power Supply AC100 240V 10 50 60Hz 1 1kw Operation Condition 15 30 35 80RH no condensation Storage Transportation Condition 10 40 10 95RH no condensation Noise Level 60dB or less Classi cation Class II 510K Electromagnetic Conformity EMC Domestic Voluntary Control for medical use has been declared IEC06061 1 2 2001 Accessories Power Cable Operation Manual Options REGIUS Cassette REGIUS Console JM JM BOX Note 1 Exept the USA Note 2 A separate license is required for the optional sampling pitch 43 75 m Note 3 Cassette Feed Load Time Time required between
5. The REGIUS MODEL 190 is furnished with air intake and outlet vents to prevent any rise in the internal temperature of the unit ensure that these vents are not blocked Cautions Regarding Storage of the Product when Temporarily out of Use Power supply breakers should be switched off following pre scribed procedures When disconnecting be sure to hold cables properly All accessories and cables should be cleaned and set in order at the place of storage In order to prevent problems the next time the product is used the place of storage should be cleaned Cautions Regarding Usage after an Extended Period of None Use Check the power cables for damage Check that earth terminals are properly connected Check that all cables are properly and rmly connected Check to ensure that the LCD panel is not cracked or fractured Check to ensure that the REGIUS MODEL 190 is operating nor mally and safely Use of the unit with other equipment may result in erroneous diagnosis or pose danger under such circumstances the unit must be operated with utmost caution The REGIUS MODEL 190 is furnished with air intake and outlet vents to prevent any rise in the internal temperature of the unit ensure that these vents are not blocked Other Cautions WARNING If the REGIUS
6. Table of Contents Forward i Features of REGIUS MODEL 190 ii ALERT SYMBOL MARK iii OTHER SYMBOL MARKS iii WARNING NOTICES SIGNAL WORDS iv Cautions relating to Laws 1 Warning Labels 3 Cautions for Handling the REGIUS MODEL 190 8 Cautions for Handling Cassettes and Plates 9 Cautions during Servicing 11 Cautions Relating to Disposal
7. C 6 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Front Door inner shelf Steel Belt Optical Unit Do not place anything on the inner shelft Warning for Electrostatic Warning for Steel Belt Warning for Laser C 7 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Warning for CS 2 3 LCD Mounting Table with Arm Table CS 2 3 LCD Mounting Table with Arm Table Warning for CS 2 3 LCD Mounting Table with Arm Table C 8 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions for Handling the REGIUS MODEL 190 The following information should be read before proceeding with maintenance work Those who are involved in servic ing should become thoroughly familiar with these cautions relating to normal handling of the REGIUS MODEL 190 CAUTION Failure to observe the cautions set out below will cause personal injury and or damage to the equipment itself Cautions prior to Usage of the Equipment Cautions detailed in the user s manual accompanying the prod uct must be carefully read and thoroughly understood The power cable should be checked for damage Check to ensure that the earth terminals are properly connected Check that all cables are properly and rmly connected Check to ensure that the LCD panel is not cracked or fractured Check to ensure that the REGIUS MODEL 190 is operating nor mally and safely
8. REGIUS MODEL 190 The outer cover may be damaged if the device is held with improper forces Don t lift the front cover by holding the bottom of the cover Don t hold the ejection plate Don t lift the device by holding the handle on the rear side Ejection Plate Handle Bottom of the Front Cover 2 8 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 16 Lift the REGIUS MODEL 190 approximately 1 cm high and lower it until the front caster contacts the slope 17 Slowly unload the REGIUS MODEL 190 from the skid lt IMPORTANT gt Handle the REGIUS MODEL 190 as slow as possi ble in order to avoid physical shock when it con tacts the oor 18 Transport the REGIUS MODEL 190 to the place where it should be installed 2 9 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 1 2 Removing of Fixing Hardware Remove the xing hardware which supports the REGIUS MODEL 190 while transporting DANGER Don t touch the steel belt Refer to the procedure 9 The edge of the steel belt is sharp which may seriously cut ngers 1 Remove the xing screws 6pcs to remove the xing hardware B The xing hardware B xes the transport unit 2 Remove the xing screws 2pcs to remove the light shielding cover bottom Fixing Hardware B Fixing Screws 6pcs Light Shielding Cover bottom Fixi
9. 4pcs that secure the xing hardware F on the device Procedure 11 to 15 are for the removal of the xing hardware F 12 Open the front cover Fixing Screws Fixing Screws Fixing Screws Fixing Hardware E Balancing Weight Fixing Screws 2 14 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 13 Remove the handle from its holder on the front cover to t it in the adapter of the transport motor 14 Turn the handle clockwise until the transport unit tilts to its limit Be sure to ensure the direction of the light shielding cover before turning the handle Refer to the Procedure 3 CAUTION If the convex side of the light shielding cover faces up the convex side will be pushed up by the transport unit which resulting in a damage to the light shielding cover 15 Remove the xing screws 4pcs to remove the xing hardware F The xing hardware F secures the cassette magnetizing plate from under side 16 Lift the balance weight until it reaches the top end As the balance weight is lifted up the cassette magnetizing plate will be lowered accordingly Adapter Toward you Toward Home Position Handle Fixing Hardware F Fixing Screws 4pcs Magnetizing Plate 2 15 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 17 Remove the xing hardware G from the cassette mag
10. CS 3 CS 1 Installation Service manual Back up of CF Card Data Back up the data in the CF card mounted on the SCB2 of REGIUS MODEL 190 onto the CS 3 CS 1 Refer to the Checks and Back up of the Setup Data of the REGIUS Console CS 3 CS 1 Installation Service manual lt IMPORTANT gt Network data and other setting data that are unique to each REGIUS 190 are stored in the CF card Availability of the back up data will greatly help set up of the REGIUS 190 when the CF card needs to be replaced Always back up the data in the CF card at the time of installation 2 22 2 3 Checking after Installa REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 3 4 Fixing of Adjuster Foot After con rming with the facility s administrator that the installation site is appropriate pull down the adjuster foot 2pcs at the front bottom of REGIUS MODEL 190 to secure it to the oor 1 Rotate and lower the adjuster foots using 26mm M16 spanner until the front adjuster foot rmly contact the oor and its rubber part is slightly depressed Lower the adjuster foots so that the device does not mover right and left 2 Gently shake the REGIUS MODEL 190 and con rm that it is rmly xed to the oor 2 3 5 Checking of Cassette Ensure that 1 or more hours have passed after the REGIUS plate had been installed to the cassette Ensure that all claws of the cassettes are
11. PC Rack at this stage is heavy Be careful not to drop it when hooking it at the right side of the REGIUS 190 CAUTION Fix the screws immediately after hooking the CS 2 3 PC Rack at the right side of the REGIUS 190 Do not leave it unscrewed 3 13 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 Attach the Edge Saddle to the Fixing Plate Right Bundle and put through the Edge Saddle the cables for mouse key board and barcode reader 3 Connect the cables according the instructions given in the Installa tion Service Manual for CS 1 2 3 Edge saddle 3 14 3 3 Installation of CS LCD Monitor Fixing Metal REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 3 Installation of CS LCD Monitor Fixing Metal How to install the CS LCD Monitor Fixing Metal on the REGIUS 190 is described below CAUTION CS 2 3 LCD Monitor Fixing Metal is to reduce the risk of the monitor falling down by means of xing the CS 2 3 Wide View Monitor to the Upper Top Plate of the REGIUS 190 1 Check that all of following components are ready on your hand 2 Unplug the LAN cable when it is connected to the REGIUS 190 Delivered pcs 1 Option CS 2 3 LCD Mounting Table with Arm Table 1 Collar 1 Large 2 2 Collar 2 Small 2 3 Holding base 1 4 Holding plate 1 5 Truss screw M4x8 2 6 Truss
12. adversely affected by atmospheric pressure temperature humidity drafts light dust or air containing salt or sulfur etc Store the equipment in a stable location Do not store it on an inclined surface and avoid vibration or shock including during transportation Do not store the equipment in a location where chemical reagents are stored or where chemical fumes may be present 4 Keep all accessories cables terminals etc neatly after cleaning them 5 Be sure to keep the equipment clean for next use 6 In the event that the equipment operates incorrectly or breaks down do not try to repair it yourself Have a trained engineer to repair it 7 Never attempt to modify the equipment 8 Maintenance and Inspection 1 Inspect the equipment and its components regularly 2 If the equipment has not been used for an extended period of time clean the equipment and make sure it operates safely before using it 9 Operate the equipment correctly according to the Operation Manual Disposing of the system lt IMPORTANT gt Should this manual become not readable due to any reason replace it with a new one which is available at charged basis When disposing of the following used parts accept the regulations of each local government 1 REGIUS MODEL 190 main body and components 2 REGIUS MODEL 190 packing materials 3 REGIUS MODEL 190 accessories and their packing materials Fluorescent medium which contains toxic su
13. cassette insertion and cassette exit 1 8 1 4 Specifications REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Chap2 Installation of REGIUS MODEL 190 2 1 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 1 Unpacking 2 1 1 Unpacking of REGIUS MODEL 190 Unpack the REGIUS MODEL 190 and move it where it should be installed WARNING Two or more persons are required to unload the REGIUS MODEL170 from the skid in order to avoid a turnover accident CAUTION Unpack the REGIUS MODEL 190 after placing the crate on the level oor Unpacking the REGIUS MODEL 190 on the inclined oor it may slip from the skid lt IMPORTANT gt Unpacking the REGIUS MODEL 190 immediately after it is carried in a facility may cause condensation due to the difference in temperature between the outside and room air In order to avoid the condensation leave the device about 1 hour after it is carried in lt IMPORTANT gt When unpacking the REGIUS MODEL 190 in a place other than the installation site ensure that there is no bump or obstacle in the carrying in route 1 Using a cutter knife cut the xing band with holding it by one hand CAUTION DON T stand in front of the xing band The xing band is strained tightly so that it may hit you at the time when it is cut Cutter Knife Fixing Band 2pcs 2 2 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 Rem
14. disconnection from the mains 5 ON power connection to the mains 6 Stand by ON OFF The main power source will still be ON even when this is turned OFF 7 Dangerous voltage 8 Type B applied part 9 High temperature cautions The part indicated with this symbol becomes very hot Safety Alert Symbol This is the industry Safety Alert Symbol This symbol draws your attention to items and opera tions associated with the use of this equipment that could result in danger to yourself and others Please read all messages next to alert symbol marks and follow the directions very carefully Before assembling or using this equipment it is important that you read the instructions and safety precautions iv REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 WARNING NOTICES SIGNAL WORDS lt IMPORTANT gt Indicates a supplementary note 10 Caution for electrostatic Internal parts of the device may get damaged due to electrostatic 11 Touching the parts indicated with this mark is prohibited It may cause elec trocution or hand may become entangled in the device 1 Signal words provide an indication of the degree of danger concealed within the product 2 This manual uses three different signal words depending on the probability and severity of injury or damage as explained below DANGER Indicates an acute hazard that will result in death or serious injury if not avoided WARNING Indicates a danger that
15. ingested care must be taken to ensure that no part of the body is contaminated and the plate disposed of following prescribed procedures Cautions to Ensure Mainte nance of Image Quality Constant care must be taken to ensure that the reading surface of the plate is not damaged or soiled Any contaminant dust etc that has attached to the reading surface of the plate should be removed by cleaning immediately As for lms gloves must be worn when handling plates Ensure that plates are not exposed to sunlight Any procedures involving exposure of the plate to uorescent light such as cleaning should be nished as quickly as possible and the plate immediately replaced in the cassette upon comple tion of the procedure Care must be taken to ensure that the front plate does not become damaged or soiled When disposing the REGIUS Plate accept the regulations of each local government C 10 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions Regarding Transpor tation and Storage Temperature and humidity levels must be maintained within the following speci ed ranges Do not expose the plate to X or rays If the cassette is transported or stored in its packing ensure that the cassette is kept in horizontal position Storage of cassettes in vertical position for extended period
16. may result in death or serious injury if not avoided CAUTION Indicates a danger that may result in medium level wound or minor injury if not avoided It is also used to indicate anticipation of a danger of physical damage only Probability of damage High Low Bodily injury and damage to equip ment Death or serious injury serious damage DANGER WARNING Medium level wound or minor injury minor dam age WARNING or CAUTION CAUTION Physical damage only CAUTION 3 In this manual in addition to the above warning notices reference notes providing supplementary explana tions are indicated as follows CAUTION The basic cautions to be observed when installing or operating the REGIUS MODEL 190 are described in this chapter Be sure to read the descriptions in this chapter and drive the users home to them C 1 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions relating to Laws Following cautions must be observed when handling the REGIUS MODEL 190 and become thoroughly famil iar with the correct method of usage Cautions relating to Laser Regulations The REGIUS MODEL 190 is a laser class 1 product Be certain to comply with any and all laws and government regulations including electric codes and utility regulations regarding the installation and operation of this equipment Precautions to be observed when using medical electrical equip ment to ensure safety and prevent danger 1 E
17. properly set into the front panel Be sure to give the declaration of conformity to the user Chap3 Installation of the Optional Equipment 3 1 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table How to install the CS 2 3 LCD Mounting Table with Arm Table on the REGIUS 190 and to the CS 1 2 3 Operation Unit s Arm is described below 1 Check that all of following components are ready on your hand Delivered pcs 1 CS 2 3 LCD Mounting Table with Arm Table refer to the dia gram below 1 Table front plate 1 2 Table rear assembly 1 3 Fixing plate left 1 4 Fixing plate right 1 5 Collar 4 6 Edge saddle 1 7 Truss screw M4x8 6 8 Truss screw M4x20 4 9 Operation Manual 1 2 REGIUS 190 3 CS 1 2 3 Opera tion Unit 4 Monitor Arm 1 Monitor Arm 1 2 Arm Cover 2 3 W SW Ni Fe M4x12 4 4 Cap for Arm Mount Fixing Screw Hole 2 5 Operation Manual 1 2 4 3 1 6 5 7 8 3 2 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 Place the Table Front Plate on the Upper Top Cover and Table Rear Plate on the Lower Top Cover of the REGIUS 190 with their two screw holes at the both edges are facing front 3 Place
18. 2004 12 Collecting the Fixing Hardware Please collect and return the xing hardware using the xing hardware recycle bag Fixing Hardware G Refer to the procedure 17 in page 24 2 19 2 2 Checking of Input Volt REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 2 Checking of Input Voltage Remove the rear cover to ensure that the wiring color of the TAP connec tor suits to the power voltage provided from the facility The wiring color is associated with each power voltage AC Fol lowing list shows the correspondence between the wiring color and power voltage Wire Color Brown Red Orange Yellow Pink Blue Power Voltage 100V 110V 120V 200V 220V 240V TAP Power voltage switching board TAP Connector 2 20 2 3 Checking after Installa REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 3 Checking after Installation Ensure that the REGIUS MODEL 190 has not been damaged while transporting 2 3 1 Power Connection Connect between the AC inlet of REGIUS MODEL 190 and power outlet of the facility with the power cable 2 3 2 Power ON 1 Turn ON the power of the facility 2 Turn ON the power breaker of the REGIUS MODEL 190 3 Push the operation switch 4 Ensure the items below The LCD panel will show the message INITIALIZING and the counter The counter will increment while initializing When the cou
19. 2pcs from bottom of the REGIUS MODEL 190 10 Cut the bottom of dust proof bag to remove it CAUTION Be careful not to damage the device with cutter knife Cushioning Materials Front Side Rear Side Cut arround the dust proof bag Dust Proof Bag 2 6 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 11 Place the ramp in front of the skid Insert the cramp in the holes on both of the ramp and skid to joint each other 12 Open the front cover 13 Remove the xing screws 2pcs to remove the xing hardware A The xing hardware A is mounted in bottom front of the REGIUS MODEL 190 in order to press against the cushioning material The x ing hardware has to be removed in advance to avoid hitting an obsta cles during displacement 14 Close the front cover Ramp Clamp Hole on the ramp Hole on the skid Front Cover Fixing Parts A Fixing Screws 2pcs 2 7 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 15 Support the REGIUS MODEL 190 from both sides with 2 or more persons WARNING Two or more persons are required to unload the REGIUS MODEL 190 from the skid in order to avoid a turnover accident REGIUS MODEL 190 weighs approximately 170kg In order to moderate the speed of the REGIUS MODEL 190 sliding on the slope 3 or more persons may be required CAUTION Be sure to follow the items below while unloading the
20. L 190 INSTALLATION MANUAL Ver 1 01 2004 12 Features of REGIUS MODEL 190 1 With the application of photostimulable uorescent material the X ray images can be digitized directly 2 Wider dynamic range than lm allowing a large amount of diagnostic data acquisition 3 Wide latitude with respect to the emitted X ray minimizing uctuations in exposures 4 Approx 40 seconds of the cassette feed load time for reading 14 x 14 cassette at 175 m saves a patient from waiting 5 Displaying images on the LCD of Console just after exposure allows the doctor to check images on the spot 6 Sampling pitch can be selected from 43 75 m 87 5 m and 175 m depending on diagnostic purposes 7 New scanning method using a semiconductor laser allows space saving design Installation space 58cm x 58cm 8 Capability to transfer the digital image data to a host computer allows the system to be used as a PACS ded icated X ray image acquisition device 9 Up to 16 units of REGIUS 190 can be connected to the REGIUS Console CS 3 iii REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 ALERT SYMBOL MARK OTHER SYMBOL MARKS The meaning of following symbol marks described in this equipment and this manual is as the following table No Symbol Meaning of Symbol 1 Danger Warning or Note Read the description that bears this symbol 2 Alternating current 3 Protective earth ground 4 OFF power
21. MODEL 190 produces any unusual noise odor or smoke power should be switched off immediately Power should not be restored until the cause has been detected and the prob lem resolved The Ethernet terminal should not be used for any other purpose for example connection to telephone lines than Ethernet connection C 9 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions for Handling Cassettes and Plates The user should become thoroughly familiar with the following cautions before handling REGIUS cassettes RC 110 Series and REGIUS Plates RP 3 Series In this manual REGIUS cassettes are referred to as cassettes and REGIUS plates as plates Cautions Regarding Photo stimu lated Fluorescent Coating WARNING The surface of the plate incorporates a photo stimulated uores cent coating which is potentially harmful if ingested In the event of leakage of the uorescent coating milky white due to sub stantial damage to the protective layer on the plate the following measures should be taken If the coating is swallowed a doctor should be consulted imme diately If the eyes are exposed to the coating the eyes should be rinsed with clean water and a doctor consulted immediately If the coating comes into contact with the skin the affected area should be immediately rinsed with clean water Even if the coating has not been
22. MPORTANT gt For the step 7 and 8 work can be facilitated if the Operation Unit is placed on a at table with its touch panel surface facing downward 3 6 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 9 Fix the Monitor Arm to the rear of the Operation Unit with four screws of W SW Ni Fe M4 x 12 3 7 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 10 Place the Monitor Arm on the Table Rear Plate and x with four truss screws M4x29 CAUTION The Monitor Arm with the Operation Unit installed is very unstable Due care should be taken not to drop or fall it When two persons are available for the work one holds the Operation Unit while the other is xing the screws 11 Af x the Screw Hole Concealing Sticker so that the screws attached in the step 10 are not visible 12 When CS 2 3 LCD Mounting Table with Arm PC Rack proceed to 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack 3 8 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 13 Attach the Edge Saddle to the Fixing Plate Right Bundle and put through the Edge Saddle the cables for mouse key board and barcode reader 14 Connect the cables according the instructions given in
23. ORTANT gt Horizontal streaks 4 5 lines interval with 20 60steps may appear when the REGIUS 190 is installed nearby the high frequency therapy equip ment Widen the distance from the high frequency therapy equipment or direct the antenna of the equipment not against the REGIUS 190 Where is wide enough to go through when the cables are connected Where the exhaust duct is not blocked Installation Space Ensure in advance that there is enough space shown below for installa tion The value in shows the space required for the installation or main tenance work At least 100 600 mm At least 100 600 mm At leaset 100 600 mm At least 700mm Rear Front 1 6 1 3 Preparation for the Instal REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Carrying in Route Ensure that there is no bump or gradient that may cause damage to the device in the carrying in route In addition ensure that the entrance of the installation site is wide enough for the device to come through Dimension and Weights of Device Width 580mm Depth 580mm Height 1230mm Weight Approximately 170kg 1 3 2 Tools and Jigs for the Installation Cutter knife Phillips Screwdriver 26mm Spanner for xing the adjust foot Long nose plier
24. TANT gt Align the rear face to the rear edge of the Rack Base Assembly 3 11 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 5 Attach the Rack Top Assembly to the Control Unit as shown in the diagram 6 Fix the Rack Top Assembly to the Rack Base Assembly with four truss screws M4x8 lt IMPORTANT gt Attach the screws to the holes near the center Fixing the Rack Top Assembly to the Rack Base Assembly allows the Control Unit to be xed to the Rack Base Assembly 3 12 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Installing the CS 2 3 PC Rack to the CS 2 3 LCD Mounting Table with Arm Table 1 Engage the crows of the CS 2 3 LCD Mounting Table with Arm PC Rack with the right edge of the REGIUS 190 and x it with two truss screws M4x8 lt IMPORTANT gt Insert the CS 2 3 PC Rack to the CS 2 3 LCD Mounting Table with Arm PC Rack into the gaps at the side of the REGIUS 190 and slightly slide it toward the rear CAUTION Make sure that the CS 2 3 LCD Mounting Table with Arm PC Rack is rmly hooked on the REGIUS 190 Failure in engagement may cause fall of the CS 2 3 LCD Mounting Table with Arm Table while the table is being secured with screws Due care should be taken CAUTION CS 2 3 LCD Mounting Table with Arm
25. bstance is used on the REGIUS Plate Replacement and disposal of the REGIUS Plate should be carried out only by quali ed service engineer C 3 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Warning Labels Warning and Caution labels stuck on the REGIUS 190 are as follows CAUTION To prevent accident during servicing never peel or soil the labels on the REGIUS MODEL 190 CAUTION If the label gets soiled or falls off and the caution is no Longer legible replace it with a new one Warning Labels Warning for LCD fall Warning for Cassette Ejection Slot Warning for Cassette Magnetizing Plate Warning for High Temperature Warning for Steel Belt Warning for High Temperature C 4 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Be careful of static electricity it may damage the electronics Warning for Foreign Object Warning for Laser Warning for Electrostatic Warning for CS 2 3 LCD Mounting Table with Arm Table Warning for Laser External Do not place anything on the inner shelf C 5 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Location of Warning Labels Exterior Transport Unit Cassette Magne tizing Plate Warning for Ejection Slot Warning for LCD fall Warning for Laser Exterior Warning for Magnetizing Plate Warning for High Temperature Warning for Foreign Object Warning for High Temprerature erasure
26. due to any reason replace it with a new one which is available at charged basis The reader device uses a laser light and high voltage NEVER remove the external panel If you remove the external panel you may be exposed to laser light or receive an electric shock Speci cations of laser unit on the REGIUS MODEL 190 are Class IIIb Medium Semiconductor laser Wavelength 660 nm Maximum output 80 mW CW WARNING C 2 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 4 When using the equipment observe the following precautions 1 Be careful not to exceed the time period and the amount of X rays required for diagnosis or treatment 2 Always monitor the entire equipment and a patient for safety 3 If an abnormality occurs on the equipment or a patient take appropriate measures i e turn off the equipment ensuring the safety for patient etc to ensure patient protection 4 Do not allow a patient to touch the equipment 5 After using the equipment observe the following precautions 1 Return the switches and dials to their original positions in the speci ed procedure and turn the equipment OFF 2 When disconnecting the cables do not pull the cables or apply an exces sive force to them 3 Observe the following precautions concerning the storage location Store the equipment in a location where it is not exposed to water Store the equipment in a location where there is no likelihood of being
27. g be careful to handle the plate so that none of the photo stimulated uorescent coating comes into contact with the body Lead foil is incorporated into all plates RP 3S except plates for mammograms RP 3M Disposal of such plates must be carried out in accordance with the appropriate waste disposal regula tions WARNING When disposing the REGIUS Plate accept the regulations of each local government Before unsealing the plate After unsealing the plate Temperature 10 40 10 30 Humidity 80 RH or lower without condensation C 11 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions during Servicing The following cautions must be observed during servicing Cautions Relating to Personal Safety DANGER The REGIUS MODEL 190 incorporates a laser generator class IIIb Direct exposure of the body or the eyes to the laser will result in serious injury Special purpose protective goggles must be worn while carrying out repairs or servicing Only REGIUS MODEL 190 units with the multiple interlocks dis engaged may be treated as class I laser products WARNING The REGIUS MODEL 190 incorporates high voltage parts which pose the risk of electrocution if touched WARNING To minimize risk only properly trained service personnel are allowed to remove covers or handle internal components etc WARNING When in the proximity of moving parts such as motors care must be exercised t
28. h Arm PC Rack lt IMPORTANT gt Please note that the CS LCD Monitor Fixing Metal and the CS 2 3 LCD Mounting Table with Arm Table are different in shape and purpose The CS LCD Monitor Fixing Metal cannot be installed in combination with the CS 2 3 LCD Mounting Table with Arm Table CS LCD Monitor Fixing Metal 1 5 1 3 Preparation for the Instal REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 3 Preparation for the Installation 1 3 1 Requirements for the Installation Ensure that the installation site meets the condition below Installation Site Where is not subject to any water or wax Where doesn t receive a direct sunlight lt IMPORTANT gt If the insertion slot is subject to the direct sun light the accuracy of the bar code reader may be lost Where is not affected or damaged by air pressure temperature humidity wind dust salt content ammable gas sulfur Where is not affected by the tilt vibration or shock Where is not on the OA oor or the bit for electrical piping lt IMPORTANT gt When the caster or adjuster foot of the device is placed on these surfaces artifact such as streak may appear on the image Where no chemical agent is stored Where no gas is generated Where is away from a source of noise lt IMP
29. imulated uorescent coating Plate Lead foil Plate except for mammo and long object Lithium battery SCB2 System Control Board Fluorescent tube LCD panel C 14 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Life Expectancy of Parts In order to ensure safe operation of the REGIUS MODEL 190 the part listed below should be replaced or serviced before its life expires Regarding the Above Life Expectancy The above gure is valid on condition that the part is used in accordance with prescribed cautions relating to usage and is properly maintained and inspected The above gure should be used as a guide to estimate the tim ing of replacement Period of Warranty In accordance with the warranty parts will be supplied at cost after expiry of the period of warranty one year Part Life Expectancy Eraser lamp approx 10000 shots Chap1 Overview Overview for the installation of REGIUS MODEL 190 is described in this chapter 1 1 1 1 Installation Flowchart REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 1 Installation Flow chart Follow the owchart below in order to facilitate the installation of the REGIUS MODEL 190 CAUTION Be certain to comply with any and all laws and govern ment regulations including electric codes and utility regulations regarding the installation and operation of this equipment Refer to Unpacking 2 1 1 Unpacking
30. move the xing screws 10pcs to remove the xing hardware C The xing hardware C secures the sub scanning unit Fixing Screws 4pcs Rear Cover Handles Fixing Screws 10pcs Fixing Hardware C 2 12 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 9 Remove the xing hardware D lt IMPORTANT gt When the xing hardware D is removed check that the steel belt is aligned to the center of the pulley If it is not gently pushing the tension pulley to check the belt become loose and then bring the steel belt to the center while paying attention to the belt edge Fixing hardware D Styrofoam parts to protect the steel belt DANGER Don t touch the steel belt Refer to the procedure 9 The edge of the steel belt is sharp which may seriously cut ngers CAUTION Wear gloves when working around the steel belt and be careful not to cut your nger with the steel belt CAUTION If the sebum is stuck to the steel belt the rust may be generated on the belt or the sub scanning unit will not work normally Steel Belt Fixing Hardware D 2 13 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 10 Remove the xing screws 14pcs to remove the xing hardware E while lifting up the balance weight The xing hardware E secures the balance weight 11 Remove the xing screws
31. netizing plate The xing hardware G protects the cassette magnetizing plate from top side magnetically sticking to it Fixing Hardware G Access the front of device for work Right Cover is removed The device is viewed from the right side Cassette Magnetizing Plate 2 16 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 18 Turn the handle counterclockwise until the right gear cover blue reaches the home position mark 19 Return the handle to the handle holder to close the front cover 20 Return the rear cover that has been removed in the procedure 7 to the original position Gear Cover right Home Position Mark 2 17 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 2 1 3 Checking of Fixing Hardware Ensure that all 7 xing hardwares are removed CAUTION If the REGIUS MODEL 190 is started up with any of x ing hardwares left inside the device will be damaged Fixing Hardware A Refer to the procedure 13 in page 16 Fixing Hardware B Refer to the procedure 1 in page 19 Fixing Hardware C Refer to the procedure 8 in page 21 Fixing Hardware D Refer to the procedure 9 in page 22 Fixing Hardware E Refer to the procedure 10 in page 22 Fixing Hardware F Refer to the procedure 15 in page 24 2 18 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01
32. ng Screws 2pcs 2 10 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 Insert the light shielding cover with its arrow printed side down to the slit of the REGIUS MODEL 190 There is a slit in the bottom of the REGIUS MODEL 190 in which the bottom cover should be inserted The position of the slit is dif cult to be located Touch it by hand to locate its precise position before inserting the bottom cover 4 Push the bottom cover from its bottom in order to ensure that the bottom cover is rmly inserted and secured in the slit If there is rattle the bottom cover hasn t been secured in the slit Insert it again 5 Secure the bottom cover with 2 screws removed in the procedure 2 CAUTION When transporting the REGIUS MODEL 190 be sure to ensure that there is no unevenness on the oor If the light shielding cover may hit an obstacle in the trans port route which resulting in a damage to the device 1 4 3 2 Light Shielding Cover Arrow Mark Slit Fixing Screws 2pcs GOOD NG Light Shielding Cover Slit Bottom of the Device 2 11 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 6 If the bundle of cable has been removed in the procedure 2 replace it to the original position 7 Remove the xing screws 4pcs to remove the rear cover Hold the handle on the rear cover to lift and remove it 8 Re
33. nsure that only a trained or approved operator uses the equip ment 2 When installing the equipment observe the following precau tions 1 Install the equipment in a location where it is not exposed to water 2 Install the equipment in a location where there is no likelihood of being adversely affected by atmospheric pressure temperature humidity drafts light dust or air containing salt or sulfur etc 3 Install the equipment in a stable location Do not install it on an inclined surface and avoid vibration or shock including during transportation 4 Do not install the equipment in a location where chemical reagents are stored or where chemical fumes may be present 5 Check the frequency voltage and allowable current or power consump tion of the power source 6 Con rm the conditions of the batteries how exposed they are or their polarity 7 Ground the equipment well 3 Before using the equipment observe the following precautions 1 Check the contact state of the switches the polarity dial setting and the various meters and con rm that the equipment operates normally 2 Make sure that the equipment is well grounded 3 Make sure that all the cables are connected correctly and safely 4 Using two X ray devices at the same time may result in an incorrect diag nosis or hazards to people So extreme care should be taken 5 Check the batteries lt IMPORTANT gt Should this manual become not readable
34. nter reaches the right end a message Network Error 24000 will appear The LCD panel will alternately show the default MAC address and IP address LCD Panel operation Switch READY Lamp Power Breaker Network Error 24000 MAC XXXXXXXXX Network Error 24000 IP XXX XXX XXX XXX 2 21 2 3 Checking after Installa REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 5 After ensuring that the REGIUS MODEL 190 is normal condition push the operation switch for at least 5 seconds The back light of LCD panel will extinguish 2 3 3 Connecting to Network Checks of REGIUS 190 operation shall be performed by the CS 3 CS 1 that is connected with the REGIUS 190 For details of the network settings refer to the REGIUS Console CS 3 CS 1 Installation Service Manual Setting the Network Connect the REGIUS 190 with the CS 3 CS 1 in 1 to 1 con guration and set the network condition Refer to the Network Setting of the REGIUS Console CS 3 CS 1 Installation Service manual Connecting to the Network After completing the network setup connect to the facility s network Refer to the Connecting the System of the REGIUS Console CS 3 CS 1 Installation Service manual Checking the Output Image Read the halftone density and check the image quality Refer to the Checks and Back up of the Setup Data of the REGIUS Console
35. o ensure that parts of the body or clothing do become tangled in the mechanism WARNING Before removing or replacing internal circuit boards or discon necting connectors or cables the REGIUS MODEL 190 power supply must be switched off Carrying out such procedures with the power on will result in serious injury and must always be avoided Work procedures should be carried out properly and safely with due regard the descriptions of the warning labels C 12 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions Relating to Power Supply Voltage WARNING The voltage supply to the REGIUS MODEL 190 is AC110V or higher To avoid electrocution the following cautions must be observed Check that the REGIUS MODEL 190 earth is properly connected Ensure that all cables are properly connected and check for damage Before carrying out work which may involve touching power sup ply lines the power supply must be switched off In cases where the power supply must be left on to carry out for example voltage measurement every precaution must be taken to avoid contact with power supply lines Cautions Relating to Lithium Batteries The SCB2 System Control Board in the REGIUS MODEL 190 incorpo rates a lithium battery Unless given speci c instructions the lithium battery should not be removed In cases where the lithium ba
36. of REGIUSMODEL 190 Move where the REGIUS MODEL 190 is installed Removing the xing hardware 2 1 2 Removing of Fixing Hardware Connecting the power cable 2 3 1 Power Connection Checking the stand alone operation 2 3 2 Power ON Network connection Refer to Installation Service Manual for CS 3 CS 1 Checking the output image Refer to Installation Service Manual for CS 3 CS 1 Backing up the CF data Refer to Installation Service Manual for CS 3 CS 1 Fixing the main unit with adjuster foot 2 3 4 Fixing of Adjuster Foot 1 2 1 2 Name of Parts REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 2 Name of Parts 1 2 1 REGIUS 190 LCD Panel operation Switch erase Switch operation Lamp Green erase Lamp READY Lamp Insertion Ejection Unit Front Cover Ejection Plate Front Cover Cassette Ejection Lamp Top Cover Front Top Cover Insertion Guide Right Cover eject Switch Ethernet Port AC Inlet Rear Cover Power Breaker Left Cover Exhaust Duct Exhaust Duct 1 3 1 2 Name of Parts REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 2 2 Option CS 2 3 LCD Mounting Table with Arm Table CS 2 3 LCD Mounting Table CS 2 3 LCD Arm CS 2 3 LCD Mounting Table with Arm PC Rack OK ERROR 1 4 1 2 Name of Parts REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 1 2 3 Option CS 2 3 LCD Mounting Table wit
37. ove the lid from the container box 3 Bring out the packing styrofoam 2pcs and parts box 1pc on top of the device Following items are packed in the parts box Warranty Check List Operation Manual Declaration of conformity Be sure to give these 4 items above to the user after installation Accessory List Power Cable Collection Bag for Fixing Hardware 4 Cut the tape xing the bottom corners of the outer carton 4 cor ners Lid Parts Box Packing Material 2 3 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 5 Turn left the knobs 3pcs in the Lockpiece and remove them Use a long nose plier to easily turn the knob of the lockpiece 6 Pull out all 3 lockpieces from the outer carton Lockpiece 3pcs Knob 2 4 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 7 Open the outer carton of the crate and remove it 8 Cut the tape to remove the slope with supporting it so that it doesn t fall over CAUTION Cutting the tape without supporting the ramp may cause fall over trouble resulting in serious damage to you The clamp is put on the back side of the ramp which will be used in the procedure 11 Crate Ramp Tape 2 5 2 1 Unpacking REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 9 Remove the cushioning styrofoam
38. s or in a high temperature environment may result in warping of the back plate Once the plate is unpacked store it in the cassette Do not store the front and back plates separately In particular storage of the back plate separately in angled position may result in warping Be careful not to drop or bump cassettes As for previous cassettes cassettes in use should be stored ver tically in a rack Cautions Before Inserting the Cassette In order to avoid problems with cassette conveying operation inside the REGIUS MODEL 190 the following cautions should be observed Ensure that the locking mechanism is functioning properly and check for bending or other damage Do not insert damaged cas settes into the REGIUS MODEL 190 Make sure that there is no paper clipped into the memo clip on the back plate Make sure that there are no foreign objects such as clips between the front and back plates In order to ensure that the back plate properly adheres to the absorption plate check that there is nothing e g tape adhered to the back plate Ensure that the bar code has not become soiled Cautions Regarding Disposal Disposal must be carried out by following the local regulations in each country If the protective layer on the plate is damaged and the photo stimulated uorescent coating leakin
39. screw M4x30 2 7 Operation Manual 1 2 REGIUS 190 3 CS 1 2 3 Opera tion Unit 3 2 1 4 6 5 3 15 3 3 Installation of CS LCD Monitor Fixing Metal REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 Remove two screws from the upper rear of the REGIUS 190 4 Sandwich the Holding Plate Base with Collar 1s Large and Collar 2s Small and x the Holding Plate Base with two truss screws M4x30 to the REGIUS 190 5 With two truss screws M4x8 temporary x the Holding Plate to the Holding Plate Base that has been xed in the step 4 Holding base Collar 2 Small Collar 1 Large Truss screw 3 16 3 3 Installation of CS LCD Monitor Fixing Metal REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 6 Place the Control Unit on the REGIUS 190 lt IMPORTANT gt Align the rear edge of the stand to the rear edge of the REGIUS 190 CAUTION Please note that the CS LCD Monitor Fixing Metal and the CS 2 3 LCD Mounting Table with Arm Table are different in shape and purpose The CS LCD Monitor Fixing Metal cannot be installed in combination with the CS 2 3 LCD Mounting Table with Arm Table 7 Adjust the crow of the Holding Plate attached in the step 3 so that the stand of the Control Unit ts to the crow Then tighten the screws 8 Connect the LAN cable that has been unplugged in the step 2
40. the Fixing Plates Right and Left on the Table Front Plate and Table Rear Plate as shown in the diagram and x them with six truss screws M4x8 Use one truss screw M4x8 each to x the Fixing Plates to the Table Front Plate Use two truss screws M4x8 each to x the Fixing Plates to the Table Rear Plate Note The screw hole of the Fixing Plate Right has a play Do not tighten the screw before the nal adjustment is com pleted Table rear assembly Table front plate Fixing plate left Fixing plate right 3 3 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 4 Hold and lift the table assembled in the step 3 and reverse it and place on the Top Cover of the REGIUS 190 5 Remove four screws from the upper edges of the right and left Side Covers of the REGIUS 190 3 4 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 6 Using four screws M4x20 with collars x the Fixing Plates to the REGIUS 7 Unplug all cables when they are connected to the CS 1 2 3 Opera tion Unit 3 5 3 1 Installation of CS 2 3 LCD Mounting Table with Arm Table REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 8 Remove two W SW Ni Fe M4 x 12 and remove the mount when the mount is attached to the CS 1 2 3 lt I
41. the Installa tion Service Manual for CS 1 2 3 Edge saddle 3 9 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack How to install the CS 2 3 LCD Mounting Table with Arm PC Rack on the REGIUS 190 is described below lt IMPORTANT gt Before installing the CS 2 3 PC Rack on the REGIUS 190 it is necessary that the CS 2 3 LCD Mounting Table with Arm Table has been installed CS 2 3 LCD Mounting Table with Arm PC Rack cannot be installed by itself alone 1 Check that all of following components are ready on your hand Assembling the CS 2 3 LCD Mounting Table with Arm PC Rack 1 Install the CS 2 3 LCD Mounting Table with Arm Table 2 Turn off the power of CS 1 2 3 and unplug all cables from the Con trol Unit Delivered pcs Option CS 2 3 LCD Mounting Table with Arm Table 1 Rack top assembly 1 2 Rack base assembly 1 3 Truss screw M4x8 6 4 Operation Manual 1 5 Edge saddle 1 1 2 5 3 3 10 3 2 Installation of CS 2 3 LCD Mounting Table with Arm PC Rack REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 3 Place the Rack Base Assembly on a at table as shown in the dia gram 4 Place the Control Unit PC on the Rack Base Assembly with the direction as shown in the diagram lt IMPOR
42. ttery must be removed care must be exercised to avoid exposure to re or water Such exposure may cause the battery to explode Cautions Relating to Damage to the Equipment CAUTION The electronic circuits in the REGIUS MODEL 190 may be damaged by static electricity The main body and any electrical parts removed should be handled with due care during installation and or repairs CAUTION An earthing strap should always be worn when han dling circuit boards CAUTION Never loosen the screws painted in white Loosening these screws may disable the restoration of the normal status in the eld Cautions Relating to the REGIUS Consoles Do not install the REGIUS Console directly on the oor In the case that a vibration may be applied to the REGIUS Con sole when servicing the REGIUS Console always shut down the REGIUS Console rst and wait for 30sec or more before starting the work C 13 REGIUS MODEL 190 INSTALLATION MANUAL Ver 1 01 2004 12 Cautions Relating to Disposal Main unit and packing material for optional equipments should also be disposed of observing the local regulations For the purposes of disposal this product must be handled in accordance with the appropriate waste disposal regulations Disposal of the product must be carried out in accordance with local ordinances and regulations Part that should be disposed of observing the local regula tions Location Photo st

Download Pdf Manuals

image

Related Search

Related Contents

39/41B - Advanced Test Equipment Rentals  WX06A取扱説明書 - 株式会社エイビット  Moxa Technologies VPORT 351 User's Manual  4 - BMW Motorrad Authorities  Harbor Freight Tools 1000 lb. Capacity Bi_Fold Aluminum Ramp Product manual  Integral IN3T16GRYGGX4 memory module  Sistema digitale    BC71866_ X-QAM Twin 3 user manual  CD-CHANGER CH-X11 MANUAL DE INSTALAÇÃO  