Endoscopes Inspection Report – Sale ------ date: --- / --

image

Contents

1. 141 OLYMPUS CF Q260DL Videoscope 1600x14 00 Insertion Tube Satisfactory condition some slight kinks at bottom of insertion tube Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition some slight kinks at bottom of insertion tube 142 OLYMPUS CF Q260DL Videoscope 1600x14 0mm Insertion Tube Satisfactory condition with slight kinks at bottom of insertion tube Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition with slight kinks at bottom of insertion tube 143 OLYMPUS GIF H260 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good HG 115 16 v2 Videoscope 1000x10 9mm Objective Lens and Focusing Good optics Remote Control Switches
2. Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 127 OLYMPUS CF Q260DL Videoscope 1600x13 80mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Satisfactory optics as there is a crack to corner of lens which causes a slight blur at one edge of the image Remote Control Switches Fully functional Distal End Good condition Image Quality Satisfactory image as there is a crack to corner of lens which causes a slight blur in at one edge of the image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition however there is a crack to lens which produces a slight blur at edge of image 128 OLYMPUS GIF H260 Videoscope 1000x10 8mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 129 OLYMPUS GIF H260 Videoscope 1000x11 10mm Insertion Tube Good condition Bending mechanism amp Ang
3. 1x C60 5 2 probe 896 Medtronic Lifepack The defibrillator is able to power up and pass through its initial self test 20 Defibrilator with successfully The unit does however display message for maintenance Leads and Paddles due The unit includes 1x set of paddles and 1x AED cable The unit is biphasic The unit showed to be able to give an output but the outputs measurement reading has not been verified The lot includes 1x 3 wire ECG with the units ECG parameter not being tested The battery and printer are both functional The defibrillator is in satisfactory physical condition Manufacture date 2007 HG 115 116 v2 945 Huntleigh SC1000 The patient monitor powers up well and passed its initialization The Patient Monitor on display LCD touchscreen and function buttons are in good working Stand order The patient monitor has not been fully tested but appears to be functional No leads are included but ECG trunk is included Overall physical condition is fair 1176 Zeiss Humphrey 740 The field analyzer s HDD has been completely wiped therefore the Vision Field software programme is not accessible Hard disk is included but Analyser Hard software programme is not installed Chin rest adjustment is functional Drive Wiped Includes event marker The motorized table height adjustment is functional The overall physical condition is fair HG 115 116 v2
4. Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 144 OLYMPUS PCF 240S Videoscope 900x12 20mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 145 OLYMPUS TJF 260V Videoscope 1200x12 80mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Missing distal cap Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism Leak Test Pass Overall Condition Good overall condition 146 OLYMPUS CF Q260DL Videoscope 1600x13 7mm Insertion Tube Satisfactory condition as the outer cover is starting to peel away at the top of the insertion tube Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Object
5. Drager Cato The anesthetic machine powers up well but unable to complete its self Anaesthetic Machine test as the anesthetic machine is not connected to medical gas supply Therefore unable to carry out any further tests Absorber assembly and bellows are included Ventilation modes included are Man Spont IPPV SIMV PCV fIPPV and fIMV Medical gas hoses are not included The overall physical condition is fair due to usage Version 02 04 318 Stryker Trolley with All of the units are able to power up The insufflator has not been Stryker Insufflator tested The SDC pro 2 is unable to complete boot up The lot is in fair and Stryker SDC physical condition Pro2 DVD Recorder 320 Karl Storz Stack The unit does not include any camera heads so unable to test the lot Trolley x 2 with The 2x National display system monitors are able to power up National Display however the Storz flat screen monitor is unable to power up The two Systems Flat Screens Storz processors are able to power up The Storz Tulip vision Line MP x 2 Storz 200902 31 is also unable able to power up No further checks able to be carried Flat Screen x 1 out on lot The lot is in satisfactory physical condition Tricam SC Pal 202220 20 Processor Telecam SL Pal 202120 20 Processor Tulip Vision Line PC and HP Printer 321 Stortz trolley with All units are able to power up The lot does not include a camera head Stortz 202040 20 or a monitor but some were used for test pu
6. ICC200 The diathermy powers up well and passed its initialization The Diathermy and Erbe display functions and function buttons are in good working order The APC300 Diathermy diathermy is tested using an ESU analyzer and the Monopolar cut and with Footswitch coagulation output readings where within the limits However unable 36417 36417 01 to test the bipolar as there is no footswitch included Includes monopolar cut and coagulation footswitch Includes Argon Plasma Coagulation which powers up well but has not been fully tested Overall physical condition is fair Argon Plasma Coagulation Version is V2 20 308 Laserscope Unable to power up the unit as the unit requires 3 phase power supply Greenlight PV Laser The unit is missing its start up key There are no accessories included System in lot The lot does include 1x footswitch The unit is in good physical condition Manufactured 03 05 309 Drager Cato The anesthetic machine powers up well but unable to complete its self Anaesthetic Machine HG 115 116 v2 test as the anesthetic machine is not connected to medical gas supply Therefore unable to carry out any further tests Absorber assembly and bellows are included Ventilation modes included are Man Spont IPPV SIMV PCV fIPPV and fIMV AGSS receiver with hose is not included and medical gas hoses are not included The overall physical condition is fair due to usage Version 02 04 310
7. SIII Heart The heart lung machine powers up well and passed its initial self test Lung Bypass The display indicators and function buttons are in good working order Machine One of the rotary pumps is functional however the four rotary pumps from the left are not functional The heart lung machine has not been fully tested Level sensor is included The overall physical is fair YOM 2002 565 Stockert 3T Heater The heater cooler powers up well and passed its initialization stage Cooler System The display and function buttons are in good working order The heater cooler has not been fully tested but appears to be functional Overall physical condition is fair Version 2 00 and YOM 2002 566 Stockert 3T Heater The heater cooler powers up well and passed its initialization stage Cooler System The display and function buttons are in good working order The heater cooler has not been fully tested but appears to be functional Overall physical condition is fair Version 2 00 and YOM 2003 567 Stockert 3T Heater The heater cooler powers up well and passed its initialization stage Cooler System The display and function buttons are in good working order The heater cooler has not been fully tested but appears to be functional Overall physical condition is fair Version 2 03 and YOM 2003 571 Respironics Bipap Both units are able to power up and pass through their initial self tests Ventilator on Stand x2 successfully The unit serial number
8. e Image Quality Below satisfactory image as image is too dark too see due to dim light output and there appears to be some water damage round image e Working Channel s Clear e Leak Test Pass e Overall Condition Below satisfactory overall condition as the scopes image is too dark due to dim light outputs 1600x14 70mm 117 OLYMPUS e Insertion Tube Below satisfactory condition as it has been ripped around GIF PQ20 the bottom of bending end e Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Satisfactory optics Remote Control Switches N A Distal End broken off and missing Image Quality Satisfactory image 12 fibres missing Working Channel s Clear Leak Test Leaking Overall Condition Below satisfactory overall condition damage to the insertion tube by distal end with the distal ends cap missing 12 fibres Direct View 1000x10 30mm HG 115 116 v2 missing and scope is leaking 118 FUJINON EC 450LP Videoscope 1600x11 40mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Satisfactory condition Image Quality Good image Work
9. for test Series SP O2 purposes and the unit showed to be unable to power up The unit is in Simulator good physical condition 513 Stockert Heart Lung The heart and lung machine is able to power up Its function buttons Machine and displays are in working order The unit has not been fully tested HG 115 116 v2 with no accessories included in lot The units pumps showed to be functional The unit is in satisfactory physical condition 514 Stockert System 3T The heater cooler system is able to power up with its displays and Heater Cooler function buttons in working order the unit has not been tested The units tubing from back of unit is missing The lot is in satisfactory physical condition 518 Dinamap Pro 1000 The Pro 1000 patient monitor powers up successfully with no errors or Patient Monitor on faults Display panel function keys and rotary control knob are all in Stand with good working order and the printer is also functional No further tests Fingerclip carried out Parameters available on the monitor include ECG SpO2 NIBP and Temperature Overall the monitor is in good physical condition Incudes a user manual and SpO2 and Temp probes 519 Hill Rom Air Shields The unit is able to power up The function buttons and displays are in Resuscitaire working order The unit showed to be able to heat up well in its pre warm and manual modes The heater adjustment in manual mode showed to be funct
10. satisfactory physical condition HG 115 116 v2 13 Eschmann TD411RS The diathermy powers up well and passed its initial self test The Diathermy with Foot display indicators rotary knobs and functions are in good working Switches order The diathermy is tested using an ESU analyzer where the monopolar cut and coagulation and bipolar output readings where within the limits Includes coagulation x2 and one bipolar footswitch The overall physical condition is fair due to usage 26 Storz SCB Image I Unable to power on any of the included monitors as none of them 222000 20 Camera include their power adaptors The processor is able to power up and Processor with was tested using a separate monitor not included in lot The processor Image 1 Camera and camera head are able to produce a reasonable image with the Head Medical processor having its focusing and magnification adjustments in Device 19 working order The lot is in satisfactory physical condition Touchscreen Monitors x 3 and Medical Device 15 Touchscreen Monitor x 1 27 Storz SCB Image I Unable to power on any of the included monitors as none of them 222000 20 Camera include their power adaptors The processor is able to power up and Processor with was tested using a separate monitor not included in lot The processor Image 1 Camera and camera head are able to produce a reasonable image with the Head Medical processor having its focusing and ma
11. table height adjustment is functional The slit lamp is in reasonable physical condition 300 Aesculap Stack Trolley with Aesculap System Monitoring NDS Surgical Imaging Monitor Aesculap Axel 300 Light Source Aesculap David 3 Digital All units are able to power up The processor and camera head are able to produce a satisfactory image on the flat screen Aesculp monitor however the image is slightly fuzzy even after focusing The light source is able to produce a good light output with its brightness adjustment in working order The stack system is in good physical condition HG 115 116 v2 Camera Control Unit and Camera Head 32984 32984 01 32984 03 32984 05 301 Strotz Stack Trolley All units are able to power up The processor and camera head are able with Stortz LCD to produce a good image on the flat screen Storz monitor with the Monitor Stortz camera head showing to have its zoom and focusing adjustments in Xenon Nova 300 working order The additional GE flat screen monitor included in lot is Light Source Stortz missing its power adaptor as a result unable to power up the monitor SCB Image 1 The light source is unable to produce a light output The DVD recorder Camera Control is able to power up The lot is in satisfactory physical condition Unit with Camera Head Sony DVD Recorder and GE CDA15R Monitor 37448 37327 03 37327 01 146686 37328 05 33992 302 Erbe
12. 103172 does have its check vent error indication light on Both of the units function buttons and displays are in working order The units have not been fully tested but both show to function with all their vent modes selectable The unit HG 115 116 v2 serial number 103172 however has large cracks and bits of its casing broken off Also its exhalation port has been broken off with the units power button cover missing Both units include an O2 hose Modes CPAP and S T and unit Serial number 103172 also includes PAV T mode Operating hours 11546h and 34273h 574 Merivaara Rapido All of the units mechanical movements are in working order The units Hydraulic Minor wheels are freely movable and the brake locks are functional The unit Ops Trolley includes cot sides The units cushions are in satisfactory physical condition The unit is in satisfactory physical condition 584 Metro Flex Crash Both carts include a suction pump at the sides of units which both Trolley with Leardal showed to power up and produce good suction However one of the Suction Unit x2 suction pumps suction jars is broken The carts are in fair physical condition 613 Laerdal Suction Both suction units power up ok Suction pumps regulators and Unit x 2 Inc pressure gauges are in working order and the containers are in place Transformer Batteries appeared to be flat at the time testing but mains adaptors are incl
13. Endoscopes Inspection Report Sale 1425 Date 11 11 2014 Condition Reports for other items at the end of the Scopes Report These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Condition 115 OLYMPUS e Insertion Tube Below satisfactory as there is a dent at top of insertion HYF XP tube Bending mechanism amp Angulation Control Good movement and lock Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Satisfactory optics Remote Control Switches N A Distal End Good condition Image Quality Below satisfactory image 30 fibres missing Working Channel s No correct size biopsy wire to test with Leak Test Pass Overall Condition Reasonable overall condition 30 fibres are missing to image and there is a dent to insertion tube Direct View 200x3 30mm 116 OLYMPUS Insertion Tube Satisfactory condition CF 1T20L e Bending mechanism amp Angulation Control Good movement and locks e Light Guide Cable amp Light Transmission Below satisfactory light output Direct View as light output is dim e Objective Lens and Focusing Satisfactory optics however image is dark to see due to dim light output e Remote Control Switches N A e Distal End Satisfactory condition
14. F H260 Videoscope 1000x10 8mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 139 OLYMPUS GIF 1T240 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks HG 115 16 v2 Videoscope 1000x12 0mm Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 140 OLYMPUS TJF 260V Videoscope 1200x12 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Missing distal cap Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism Functional Leak Test Pass Overall Condition Good overall condition
15. Legacy 20000 Phaco Machine 25280 The phaco machine is able to power up with its touch screen display in working order The unit does not include any hand pieces or accessories the unit has not been fully tested The units IV pole height adjustment is in working order The units audio voice is functional The 1x included foot control appears to be responding the 1x remote is functional The unit is in satisfactory physical condition Version V3 24 Manufacture date 07 01 405 Dinamap Pro Care Ausculatory x 2 and Pro 1000 x 1 802334 800760 800768 The Pro 1000 patient monitor powers up successfully with no errors or faults Display panel function keys and rotary control knob are all in good working order and the printer is also functional No further tests carried out Parameters available on the monitor include ECG SpO2 HG 115 116 v2 NIBP and Temperature The two ProCare monitors have not been tested as the mains adaptors chargers are not included No accessories included with the lot apart from a Temp probe for Pro 1000 monitor Overall the monitors are in good physical condition 410 Datasonic System 98XT Baccoon Pump 500694 The balloon pump is temperamental in power on When the unit is able to power on it displayed an error message for system failure and gave a constant alarm The unit is not functional The unit does not include any leads or accessories The unit does include it
16. are all able to power up The ultrasound is in satisfactory physical condition Manufacture date January 2000 Includes 1x CL10 5 probe 1x L12 5 probe 730 Metro Flex Crash The crash cart is in reasonable physical condition Includes two suction Trolley with 2 x units LSU which are intact and in good working order The cart Laerdal Suction contains 3 drawers Units 757 Philips Pagewriter The ECG machine is able to power up The function buttons and Trim II ECG with display are in working order The lot includes 12 wire ECG with the Leads Serial cable and rubber grommet below the module damaged exposing inner Number US00402341 cabling The unit was unable to detect the ECG module and as a result the unit was not functional The units printer is functional the battery is functional but it is low The ECG machine is in satisfactory physical condition Release A 01 01 761 Acuson Sequoia The ultrasound is able to power up The unit is missing a lot of its 50012 Ultrasound with 15L8W Probe Keys Missing and Hard Drive Removed Serial function button covers and a TGC adjustment knob is also missing The function buttons and display are in working order the trackball functions however it is temperamental with its movement The ultrasound showed to be functional with its modes 2D D M and PW functioning The ultrasound was tested using a phantom and all of the HG 115 116 v2 Number 56225 r
17. cal condition 647 Bambi Air 75 150 The compressor is average physical condition due to usage The Dental Air compressor has not been fully tested but appears to be functional Compressor HG 115 116 v2 687 Zeiss FF450 Plus The Fundus camera powers up well and passed its initialization The Fundus Camera with optics and function buttons are in good working order The workstation PC Workstation and is included however it is password protected therefore unable to access Effekta UPS the main menu functions Magnifications of x20 x30 and x50 are included Movements and locks are in good working order Camera head x10 Magnification eyepiece two keyboards x2 UPS S N 02C66013 is not functional and mouse are included The motorized table height adjustment is in good working order The overall physical condition is fair YOM 2002 707 ATL HDI5000 The ultrasound is able to power up however it shows to be Ultrasound with 2 x temperamental with it sometimes displaying a blank white screen that Probes CL10 5 and is unable to be passed or during boot up it can display an error message L12 5 Serial An error has been detected and this message also is unable to be 01936G passed In either situation the ultrasound showed not to be functional Unable to perform any further checks on unit The units included probes are in satisfactory physical condition The Mitsubishi CP700 the video graphic printer and the VCR
18. ction pins at the back of unit as a result unable to power up the unit The unit is in poor physical condition with the printer being broken with its cover missing one of the connections for lead AP R is missing and the display cover is badly damaged 434 Datascope System 98 The balloon pump is able to power up and pass its initial system test XT with Cariosync 2 Ok The units function buttons and display appear to be in working Balloon Pump order The unit does not include any leads or accessories so the unit has not been tested The unit includes its adult condensate removal safety disk and a helium cylinder The unit is missing its Doppler The printer is functional The balloon pump is in satisfactory physical condition HG 115 116 v2 System timer 5265 2h 456 Valley Lab Force FX The diathermy powers up well and passed its initialization Its display Diathermy on functions and indicators are in good working order The diathermy is Trolley with tested using an ESU analyzer and the monopolar cut and coagulation Footswitches and bipolar output readings taken where within the limits Includes monopolar cut and coagulation and bipolar footswitches Overall physical condition is fair 477 GE Vivid Five The ultrasound is able to power on however the unit does not appear to Ultrasound with 1 x be able to boot up With the monitor displaying a blank screen which FPA 2 5 mhz Probe states that it is
19. e ECG 282 Eschmann MR Hydraulic Operating Table All of the operating tables articulating movements are in working order The hydraulic height adjustment is functional The tables caster wheels are freely moveable and the brake lock is functional The table includes a complete set of cushions which appear to be in good physical condition The operating table is in satisfactory physical condition 292 Erbe ICC200 and APC300 Diathermys 40194 40194 01 The diathermy powers up well and passed its initialization The display functions and function buttons are in good working order The diathermy is tested using an ESU analyzer and the Monopolar cut and coagulation output readings where within the limits However unable to test the bipolar as there is no footswitch included Includes monopolar cut and coagulation footswitch Includes Argon Plasma Coagulation which powers up well but has not been fully tested Overall physical condition is fair Argon Plasma Coagulation Version is V2 20 294 Haag Streit Slit Lamp on Stand The slit lamp is unable to power up and as a result there is not light output able to be produced by the unit The units joystick movement is freely movable The chin rest is in place with its chin rest adjustment in working order The fixation light is incomplete with no bulb The unts optical system is able to produce a clear image with its magnification adjustment in working order The mechanical
20. eadings taken showed to be within limits The probe showed to have good resolution and good grey scale The color video printer is able to power up The ultrasound is in fair physical condition with its front cover panel and power buttons button cover missing Includes 1x ISL8w probe 776 Medtronic Stealth The Stealth Station surgical navigation system powers up but does not Station and Treon complete its initialization as an error message is displayed of no Plus Computer bootable device insert boot disk and press any key Therefore unable Assisted Surgery to carry out any further tests Includes surgeon monitor Overall System physical condition is fair due to usage 794 Olympus CKX41 The microscope powers up well and the light source is in good working Inverted Microscope order The optics is functional and gives a good output image However with 1 x Set of Eye the stage plate is damaged and a crack is visible Attachments are not Pieces Serial included Magnifications of x10 x20 and x40 are included Two Number 2F11166 eyepieces of x10 are included Overall physical condition is fair due to usage 806 Acuson Sequioa The ultrasound is unable to power up The lot does not include any C256 Ultrasound probes Two of the TGC button covers are missing The lot is in Not Powering Up satisfactory physical condition no Probes Manufacture date 2002 807 Datascope System 98 The balloon pump is able to power up and pass its init
21. gnification adjustments in Device 19 working order The lot is in satisfactory physical condition Touchscreen Monitors x 3 and Medical Device 15 Touchscreen Monitor x 1 28 Alcon Phaco The unit is able to power up Unable to test the unit as an advisory machine on trolley message please adjust supply source pressure for standard pressure with Footswitch operation which is unable to be passed The unit does not include any hand pieces or accessories No further checks are able to be carried out The unit includes 1x foot control The unit is in satisfactory physical condition 29 Zeiss Colposcope The colposcope is able to power up The units mechanical movements Opni IFC are in working order The unit is able to produce a good light output with its brightness adjustment in working order The optical system is able to produce a good image however there are dirt particles from inside of the optical system visible on the image The magnification adjustment is in working order The colposcope is in satisfactory physical condition HG 115 116 v2 Includes 1x set of eyepieces 10x 22B 1x objective lens 300 31 Olympus UES 40 The diathermy is able to power up with its function buttons and Diathermy Olympus displays in working order The units output has not been tested The UHI 2 Insufflator units patient plate alarm appears to be functional The lot includes a and Footswitch x2 cu
22. good light output with its brightness Printer Osteo Power adjustment in working order The printer showed to function but has 2 Linvatec C7100A_ its front panel broken off The Osteopower 2 was able to power up with irrigation and its function buttons in working order The unit includes 2x hand Autoclav C3282 pieces however unable to test the units functionality as it is missing its Camera Head x2 foot control Unit includes A O chuck 2x 1 1 straight attachment 3010501 3010505 1x sagittal attachment 1x and W P driver body 1x The APEX 2047004 3010502 unit is also able to power up but has not been able to be tested as there 2053203 27910 301 are no cassettes included The lot is in satisfactory physical condition 89 Storz Xenon 300 The processor is able to power up The processor was tested using a Light Source and camera head and a monitor which are both not included in lot The Image Control Unit processor showed to help in producing a satisfactory image The light source is able to power up with its function buttons in working order The light source is able to produce a good light output with its brightness adjustment in working order The units air pump is functioning Units bulb hours at 128h The lot is in satisfactory physical condition 98 Storz Image 1 Both processors are able to power up The processors were tested using Camera Control x 2 a camera head and a monitor which are both not included in lot Bo
23. has been Convex Probe For conducted on this probe The probe appears to be in satisfactory Model X 300 Serial physical condition There are some kinks to the probes cable but no Number other signs of obvious damage The probe is loose and not in a case SCX5CD14501 192 Olympus MH 977 The pigtail is in good physical condition and functional Tested with a Pigtail compatible endoscopy stack system and the images were clear with no signs of interference visible on the screen 257 Olympus WM30 All units are able to power up with their function buttons in working Workstation to order The Fujinon processor and light source were tested using a video Include Fujinon scope and monitor which are not included The processor was then able E400 Light Source to help in producing a good image The light source showed to be able Fujinon E400 to produce a good light output The Olympus processor and light Processor Olympus source were tested using a video scope monitor and pig tail which are CV 240 Processor not included The processor then showed to be able to help in Olympus CLV 240 producing a good image The light source showed to be able to produce Light Source and a good light output with its light output modes OES Evis 100 and Keyboard Evis 200 all functioning The units air pump showing to function The lot is in satisfactory physical condition 258 Eschmann Hydraulic All of the operating tables articulating movements are in working Operat
24. he diathermy is tested using an ESU analyzer and the monopolar cut and coagulation and bipolar output readings taken where within the limits Includes monopolar cut and coagulation and bipolar footswitches Overall physical condition is fair 361 Valleylab Force EZ Diathermy no Footswitch 40512 The diathermy powers up well and passed its initialization Its display functions and indicators are in good working order The diathermy is tested using an ESU analyzer and the monopolar cut and coagulation output readings taken where within the limits However as bipolar footswitch is not included therefore unable to verify the output readings No footswitches are included Overall physical condition is fair HG 115 116 v2 370 Drager Babytherm 8010 Resuscitaire The unit is able to power up with its function buttons and displays in working order The unit is able to heat up well in manual mode with its heater adjustment in working order The unit does not include any skin temperature probes so unable to test skin mode In mattress mode the unit showed to be able to heat up well with the unit reaching and maintaining the set temperature The overhead spot lights are functional The units electrical height adjustment is in working order The unit does not include a mattress The unit is in satisfactory physical condition 374 Fisher amp Paykel Resuscitaire The unit is able to power up with it
25. ial system test Balloon Pump ok The units function buttons and display appear to be in working order The unit does not include any leads or accessories so the unit has not been tested The unit includes its adult condensate removal safety disk and helium cylinder The units Doppler is missing The printer is functional The balloon pump is in satisfactory physical condition System timer 16098 8h 808 Datascope System 98 The balloon pump is able to power up and pass its initial system test XT with Cariosyne 2 ok The units function buttons and display appear to be in working Balloon Pump order The unit does not include any leads or accessories so the unit has not been tested The unit includes its adult condensate removal safety disk and helium cylinder The units Doppler is missing The printer is functional The balloon pump is in satisfactory physical condition System timer 4626 3h 839 Valleylab Force FX The diathermy powers up well and passed its initialization Its display Diathermy with functions and indicators are in good working order The diathermy is Footswitch 34999 tested using an ESU analyzer and the monopolar cut and coagulation HG 115 116 v2 and bipolar output readings taken where within the limits Includes monopolar cut and coagulation and bipolar footswitches Overall physical condition is fair 843 Keeler Pulsair The pulsair is able to power up with its function button
26. in power save mode the displays on the control panel 1 x GE 2 0 mhz are not loading up The unit is not functional and unable to perform any Probe and a Sony further checks on the unit The included probe and Doppler are in Video Graphics satisfactory physical condition Both the VCR and video graphic Printer printer are able to power up The ultrasound is in satisfactory physical condition Manufactured 2001 Includes 1x 1C probe 1x 2 0MHz doppler 478 Stack Trolley with All of the units are able to power up The processor and both of the Radiance NDS camera heads are able to produce a satisfactory image on the flat screen Monitor Conmed_ monitor However both camera heads produce interference to the 3CCD Camera image with movement at their connection ends to the processor and Controller Conmed also one of the camera heads is unable to complete its white balance Xenon Light Source with its image being off colour The light source is able to produce a Conmed 24k good light output with its brightness adjustment in working order The Insufflator Sony conmed 24K is able to power up however unable to check the unit as Colour Video there are no accessories included in lot The colour video printer is Printer 2 x Linvatec functional The lot is in good physical condition IM331 Camera Head and Rigid Scopes 14372610 14372602 14397802 143726 148497 2922301 497 Bio Tek Index 2 The unit is missing its power adaptor but one was used
27. ing Channel s Clear Leak Test Leaking Overall Condition Satisfactory overall condition scope is leaking 119 OLYMPUS URF P5 Direct View 650x3 0mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and lock Light Guide Cable amp Light Transmission Poor transmission no light guide cable Objective Lens and Focusing Below satisfactory image is hard to see due too poor light output Remote Control Switches N A Distal End Good condition Image Quality Poor image is hard to see due to poor light output and 100 fibres missing Working Channel s No correct sized biopsy wire to test with Leak Test Pass Overall Condition Below poor image is hard to see due to poor light output and 100 fibres missing scope is in good physical condition 120 FUJINON Videoscope 1000x10 0mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 121 FUJINON Videoscope 1600x11 40mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good move
28. ing Table order The tables hydraulic height adjustment is functional The table s caster wheels are freely movable and the brake lock is functional The table includes a complete set of cushions which have tape covering up rips in the cushions The operating table is in satisfactory physical condition 260 Valley Lab Force EZ The diathermy powers up well and passed its initial self test The Diathermy on display functions and function buttons are in good working order The Trolley with diathermy is tested using an ESU but when testing for the monopolar Footswitches cut mode the diathermy cuts out and alarms Therefore unable to verify the monopolar cut output readings The monopolar coagulation is tested and the output readings where within the range Overall physical condition is fair HG 115 116 v2 273 Olympus WM 30 All units are able to power up Unable to test the processor as the lot Workstation with does not include a camera head The light source is able to produce a Olympus OTV 55 good light output with its brightness adjustment in working order The Processor Olympus units spare lamp is functional The insufflator has its function buttons UH1 3 Insufflator and displays in working order however the unit has not been tested Olympus CLV 530 The insufflator includes 1x footswitch The color video printer is able Light Source and to power up The lot is in satisfactory physical conditio
29. ional The lot does not include a temperature probe so unable to check baby mode The units over head spot light and the timer are functional The electrical height adjustment and mechanical cot tilt adjustment are both functional The unit includes a mattress and an O2 hose Unit is missing its suction jar The lot is in fair physical condition 531 Philips Pagewriter The ECG machine is able to power up The function buttons and Trim II ECG display in working order however two of the key board keys are Recorder with Leads missing there function covers The lot includes 12 wire ECG with the unit being unable to detect the ECG module and as a result the unit was not functional The units printer is unable to function as the battery is not functional The lot is in satisfactory physical condition Release A 01 01 536 Eschmann MR All of the operating tables articulating movements are in working Hydraulic Operating order The hydraulic height adjustment is functional The tables caster Table No Cushions wheels are freely movable and the brake lock is function The table does not include any cushions The operating table is in satisfactory physical condition 547 Toshiba Nemio 10 The ultrasound is able to power up The function buttons trackball and Ultrasound Machne with 3 75mhz Probe x 1 and Mitsubishi Graphic Printer display are in working order The ultrasound showed to be functional with its included modes 2D and M functioni
30. ive Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition HG 115 116 v2 Condition Report Sale 1425 Date 11 11 2014 These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Report Number 1 Eschmann RX500 All motorised positioning adjustments are in good working order and Electric Operating the remote control is also functional The battery is working and the Table with table base lock is operational All cushions are in place and overall the Controller unit is in satisfactory physical condition 2 Sonosite iLook The ultrasound is able to power up The function buttons and Ultrasound touchscreen display are in working order The ultrasound showed to be functional with its modes 2D and CPD functioning The unit has been tested using a phantom and all of the readings taken showed to be within set limits The probe showed to have satisfactory resolution and satisfactory greyscale The units battery is not functional The ultrasound is in satisfactory physical condition with the unit missing its stylus pen The included stand has its tilt adjustment k
31. l the monitors are in reasonable physical condition The rotary control knob on one of the monitors have an intermittent fault does allow access to full system menus settings 415 Valleylab Force EZ The diathermy powers up well and passed its initialization Its display Diathermy no functions and indicators are in good working order The diathermy is Footswitch 40511 tested using an ESU analyzer and the monopolar cut and coagulation output readings taken where within the limits However as bipolar footswitch is not included therefore unable to verify the output readings No footswitches are included Overall physical condition is fair 416 Valleylab Force E The diathermy powers up well and passed its initialization Its display Diathermy no functions and indicators are in good working order The diathermy is Footswitch 33041 tested using an ESU analyzer and the monopolar cut and coagulation output readings taken where within the limits However as bipolar footswitch is not included therefore unable to verify the output readings No footswitches are included Overall physical condition is fair 421 Bio Tek 601 Pro The SPo2 simulator is able to power up with its function buttons and Analyser Bio Tek display in working order The unit has not been tested The unit Index 2 SPO2 includes its power adaptor The unit is in satisfactory physical Simulator condition with its screen scratched The 601 pro is missing its power cord conne
32. ment and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Satisfactory condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition HG 115 16 v2 122 FUJINON Videoscope 1600x14 00 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good condition Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 123 FUJINON EC 450WLS5 Videoscope 1600x13 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 124 OLYMPUS CF 100HI Videoscope 1300x14 5mm Insertion Tube Below satisfactory condition kinks at bottom of insertion tube and outer cover peeling off in place
33. n Sony Colour Video Printer and Sony CRT Monitor 274 Datex Ohmeda The unit powers up successfully Ventilator display screen and control Excel 210SE panel are in working order 7900 V3 5 However some pixels are Anaesthetic Trolley missing from the lower section of the LCD screen but the information oe is still readable No further checks carried out Breathing circuit Vapourizer absorber ad bellows are in place Air O2 and N20 hoses are included Suction regulator with hose and AGSS receiver with hose are also included Overall the anaesthetic machine is in reasonable physical condition Includes a sevoflurane vaporiser 275 Stryker Stack All of the units are able to power up The processor and camera head Trolley with 888i showed to be able to produce a satisfactory image on the flat screen Digital Camera with Stryker monitor with the camera heads focusing showing to be in Camera Head working order The light source is able to produce a good light output Stryker 6000 Light with its brightness adjustment in working order Bulb hours 318 The Source Stryker lot is in satisfactory physical condition Vision Flat Screen Monitor 278 Valleylab Ligasure The Ligasure unit powers up well and passed its initialization The Diathermy On display and function buttons are in good working order The unit has Trolly with not been fully tested but appears to be functional Lot includes Footswitch footswitch Overall physical condition is fair 280 Ol
34. nding mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 133 OLYMPUS GIF H260 Videoscope 1000x11 0mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Slight leak Overall Condition Good overall condition however there is a slight leak 134 HG 115 OLYMPUS PCF 204S Videoscope 900x12 0mm 16 v2 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 135 OLYMPUS PCF 240S Videoscope 900x12 3mm Insertion Tube Good c
35. ng The included probe showed to have large area of dropout on one side of the scan The ultrasound has been tested using a phantom and all of the readings taken showed to be within set limits The probe showed to have satisfactory resolution and satisfactory greyscale The probe head also has some damage with the outer insulation of cable coming away from HG 115 116 v2 the rubber grommet at the probe connection end The printer is broken and not able to power up The unit is in satisfactory physical condition Includes 1x PVM 375AT probe 557 Sager Emergency Fair physical condition Traction Splint x 5 562 Eschmann J3 All of the mechanical articulating movements are in working order Hydraulic Operating The hydraulic height adjustment is functional The caster wheels are Table freely movable and the brake lock is functional The unit includes a complete set of cushions which appear to be in good physical condition The operating table is in satisfactory physical condition 563 Stockert SIII Heart The heart lung machine powers up well and passed its initial self test Lung Bypass The display indicators and function buttons are in good working order Machine The rotary pumps are in good working order however middle rotary pump does not rotate The heart lung machine has not been fully tested but appears to be functional Level sensor is included The overall physical is fair YOM 2003 564 Stockert
36. nob missing Manufacture date 2004 02 Includes 1x probe 3 Zonare Medical The super cart is able to power on with its power indication lights on System Inc However the unit is not complete with it missing its scan engine As a Diagnostic result the unit is not functional and unable to power the unit up The Ultrasound System super cart is in good physical condition with Ultra Cart 4 B K Medical Viking The ultrasound is able to power on with its power indication lights on 2400 Ultrasound however nothing is being displayed on the units display The with 5 12MHz ultrasound is not functional no further checks carried out The units Linear Probe and included probes have not been tested but are in fair physical condition 2 5MHz Convex with probe 8811 having bad chips to probe head around the rubber Probe membrane The video graphic printer powers on The lot is in satisfactory physical condition 9 Olympus UHI 3 The processor and camera head were tested using a monitor which is Insufflator with Foot no included in the lot The processor was able to power up The Switch Olympus processor and camera head were able to produce a good image with Visera OTV S7 the camera head having its focusing adjustment in working order The Digital Processor insufflator is able to power up with its function buttons and displays in Olympus OTV S7 working order however the unit has not been tested The unit includes Camera Head 1x footswitch The lot is in
37. ondition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Only freeze is functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Leaking Overall Condition Good overall condition but scope is leaking 136 OLYMPUS CF Q260DL Videoscope 1600x13 60mm Insertion Tube Satisfactory condition as slight kinks to tube of insertion tube Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition slight kinks at top of insertion tube 137 OLYMPUS PCF Q260AL Videoscope 1600x11 90 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 138 OLYMPUS GI
38. rder The units light output is temperamental in functioning When the light output does come on the unit showed to produce a good light output with its brightness adjustment showing to be in working order The units air pump is functional Bulb hours 120h Xenon 175 Powers up with its function buttons and display in working order The units light output showed to take some time before coming on but then is able to produce a good light output The units brightness adjustment is in working order The units air pump is functional Bulb hours 440h The lot is in satisfactory physical condition HG 115 116 v2 52 Drager Babytherm 8010 Resuscitaire The unit is able to power up with its function buttons and displays in working order In manual mode the unit showed to be able to heat up well however unable to test skin mode as the unit does not include any skin temperature probes In mattress mode the unit is also able to heat up well with it able to reach the set temperature and then maintain it The overhead spot lights are functional The unit includes a gel mattress The electrical height adjustment is functional The unit is in satisfactory physical condition 53 Drager Babytherm 8010 Resuscitaire The unit is able to power up with its function buttons and display in working order In manual mode the unit showed to be able to heat up well however unable to test skin mode as the unit does not include an
39. rposes The processor SCB aida DVD showed to help in producing a satisfactory image The light source is Recorder Stortz able to produce a good light output with its brightness adjustment in working order Bulb hours 225 The light sources air pump is functioning The 2024020 image management unit powers up but is unable to complete boot up with the unit displaying Reboot or select proper boot device or insert boot media in selected boot device The lot is in satisfactory physical condition 325 Haag Streit Slit The slit lamp is able to power up The unit is able to give a good light Lamp on Stand output The joystick movement is freely movable The fixation light is functional The chin rest is in place with its chin rest adjustment in working order The units optical system is able to produce a clear image with its magnification adjustment in working order The table s mechanical height adjustment is functional however the table is missing one of its wheels making the slit lamp off balance The slit lamp is in reasonable physical condition Includes HG 115 116 v2 1x set of eyepieces 10x 1x set of eyepieces 16x 326 Drager Airshields Isolette C2000 Incubator The incubator is able to power up and go through its intial power up test successfully The function buttons and display are in working order In air mode the unit is able to heat up well with the unit able to reach the required tempera
40. s Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good transmission kinks to light guide cable Objective Lens and Focusing No pig tail to test with Remote Control Switches No pig tail to test with Distal End Satisfactory physical condition Image Quality No pig tail to test with Working Channel s Clear Leak Test Leaking Overall Condition Fair overall physical condition with kinks to insertion tube and scope is leaking unable to test the scopes working condition 125 126 OLYMPUS TJF 260V Videoscope 1200x12 50mm OLYMPUS GIF H260 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Missing distal cap Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism Functional Leak Test Pass Overall Condition Good overall condition Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks HG 115 16 v2 Videoscope 1000x10 80mm Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition
41. s adult condensate removal safety disk it is however missing its helium cylinder and Doppler The balloon pump is in satisfactory physical condition System timer 24072 2h 411 Datasonic System 98XT Baccoon Pump 500695 The balloon pump is able to power up and pass through ts initial system test ok The units function buttons and display appear to be in working order The unit does not include any leads or accessories the unit has not been tested The unit includes its adult condensate removal safety disk but however is missing its helium cylinder The unit includes its Doppler which is not functional and its transducer is badly damaged and broken The printer is functional The balloon pump is in satisfactory physical condition System timer 18570 2h 412 Stack Trolley with Sony Trinitron Monitor Linvatec C3140 Lightsource C3110 Camera Controller Osteo Power 2 Linvatec C7100A Irrigation Console Sony Colour Video Printer and Autoclav C3282 Camera Head 2053202 20532 2053201 32615 30105 24627 HG 115 116 v2 All of the units are able to power up The camera head and processor are able to produce a satisfactory image on the Sony monitor however interference can be produced to the image with movement at the connection end of the camera head to the processor The camera head showed to have its focusing adjustment in working order The lots included riged scope showed to have a clear image and is in
42. s function buttons and indication lights in working order The unit showed to be able to heat up well in prewarm and manual modes with the heater adjustment in manual mode in working order There is no skin temperature probe included in lot The overhead spot light and the timer are functional The electrical height adjustment is in working order The unit includes a mattress The unit is in satisfactory physical condition 380 Valleylab Force FX Diathermy with Footswitch 35001 The diathermy powers up well and passed its initialization Its display functions and indicators are in good working order The diathermy is tested using an ESU analyzer and the monopolar cut and coagulation and bipolar output readings taken where within the limits However the monopolar coagulation Fulgurate and Spray modes do not function Includes monopolar cut and coagulation and bipolar footswitches Overall physical condition is fair 402 Drager Air Shields Resuscitaire The unit is able to power up with its function buttons and displays in working order The unit showed to be able to heat up well in pre warm and manual modes with the temperature adjustment in manual mode functional Unable to test baby mode as the unit does not include a skin temperature probe The units overhead spot light is functional as is the timer The unit includes a mattress and suction jar The unit is in satisfactory physical condition 404 Alcon Surgical
43. s in working Easyeye and order The lot includes its handpiece however its output has not been Warmflow Fluid Warming System tested ata aiid The warmflow is also able to power up with its function buttons and display in working order The unit has not been tested The lot is in good physical condition 866 Zeiss IOL Master The IOL Master powers up but does not complete its initialization as it Retinal Camera displays Operating Software Not Found Therefore unable to go into the main menu options and carry out further checks The hard drive is missing The motorized height adjustment is in good working order The overall physical condition is average due to usage YOM 2001 889 Sonosite 180 Plus The ultrasound is able to power up The function buttons and display Ultrasound with are in working order The ultrasound is functional with its modes 2D Probe x 2 C60 5 2 M and CPD functioning The probe L38 10 5 shows to have damage and L38 10 5 to its crystals with bad areas of drop out to its image Also the probes cable has tried to be repaired with visible damage to it and there are chips to the probe head plastic casing The ultrasound was tested using a phantom and all of the readings taken showed to be within limits Both probes showed to have satisfactory resolution and satisfactory greyscale The units battery is not functional The ultrasound is in reasonable physical condition Manufacture date Includes 1x L38 10 5 probe
44. satisfactory physical condition The light source is able to produce a good light output with its brightness adjustment in working order The Linvatec Universal Irrigation Console is able to power up with its display in working order No cassettes included so unable to test units Version 3 3 The Osteopower 2 is also able to power up with its function buttons and display in working order The unit includes 1x hand piece however there is not footswitch included and as a result unable to check the units functionality The color video printer is able to power up The lot is in satisfactory physical condition Includes 1x Rigid scope 30 1x W P driver module piece 1x sagittal saw piece 1x W P driver body piece 1x A O chuck piece and 1x 1 1 straight piece 413 GE Dash 3000 All four of the units are able to power up with their function buttons Patient Monitors and displays in working order The lot does not include any leads or with Multiparameter accessories as a result the units have not been test One of the units Modules x 4 batteries was functional The lot is in satisfactory physical condition 413 GE Dash 3000 All four monitors power up ok with no errors or faults Display panels Patient Monitors function keys and rotary control knobs are all in good working order with i No further checks carried out Parameters available on the monitors a N include ECG SpO2 NIBP IBP and Temperature Overal
45. t coag footswitch and a bipolar footswitch The insufflator is also able to power up with its function buttons and displays in working order The unit has not been tested The lot is in satisfactory physical condition with the bipolar footswitch having it plastic grommet missing with the cable slightly damaged 34 Erbe ICC200 The diathermy powers up and passed its initial self test The display Diathermy No indicators and function buttons are in good working order The Footswitch diathermy however displays error message of Error 13 therefore unable to carry out any further checks Overall physical condition is fair 35 GE Dash 3000 All of the units are able to power up however one unit does not load Patient Monitors x 3 with its initial GE screen on a constant loop Also one of the units and 1 x Marquette displays is smashed and therefore is not functional The other 2 units Dash 3000 Monitor have their function buttons and displays in working order however For Spares Only their rotary knobs button covers are missing The lot does not include any leads or accessories so unable to check the two functioning units however their NiBP pumps did show to function None of the batteries were functional The units are in poor physical condition with parts and covers missing and broken with paint work coming away and chipped 46 Storz Xenon 300 and Xenon 300 175 Light Sources Powers up with its function buttons and display in working o
46. th processors showed to help in producing a satisfactory image The processors are in satisfactory physical condition 99 Valleylab Ligasure Lot includes two Ligasure units which both power up well and passed Vessel Sealing its initialization The display and function buttons are in good working System with 1 x Foot order The units have not been fully tested but appear to be functional Switch Lot includes footswitch Overall physical condition is fair 171 Oculus SDI Ie The lot appears to be physically intact but has not been tested Lot is in Camera Attachment in Carry Case satisfactory physical condition HG 115 116 v2 176 Siemens Probes x2 Neither probe has been tested only a physical check has been VFX 13 5 P4 1 undertaken on probes The Px4 1 probe showed to be in satisfactory Condition Unknown physical condition with no signs of any obvious damage The VFX13 5 probe is in fair physical condition with it however having a few small marks and a rip to its outer insulation of its cable Also there is a large hole to the probes heads rubber membrane 187 Siemens C5 2 The probe has not been tested only a physical check has been Convex Probe conducted on this probe The probe appears to be in satisfactory Serial Number physical condition with no obvious signs of damage The probe is B2622714 loose and not in a case 188 Siemens CX 5 2 The probe has not been tested only a physical check
47. ts function buttons and display in working order The units output has not been tested The lot is in satisfactory physical condition 86 Graseby 3400 Syringe Pumps x 10 HG 115 116 v2 All of the units are able to power on The units have not been tested The units look to be physically intact The lot is in fair physical condition with some dents the displays covers 87 Datasonic System The balloon pump is able to power up and pass its initial system test 98XT Balloon Pump Ok The units function buttons and display appear to be in working 500698 order The unit does not include any leads or accessories so the unit has not been tested The unit includes its adult condensate removal safety disk it is however missing its helium cylinder and Doppler The printer is functional The balloon pump is in satisfactory physical condition System timer 16537 9h 88 Stack Trolley with All units are able to power up The processor and both camera heads Sony Trinitron showed to be able to produce a satisfactory image on the Sony monitor Monitor Linvatec however one of the camera heads showed to have slight interference to C3140 Light Source the image with movement of the connection end Both of the camera C3110 Cambra heads showed to have good focusing Both of the camera heads showed Controller Sony to have some ware at the corners of the connection chips The light Colour Video source is able to produce a
48. ture and maintain it The units temperature has not been verified There are no skin temperature probes included in lot so unable to test skin mode The oxygen and humidifier modes are selectable but have not been tested The units electrical height adjustment is functional as are the mechanical cot tilts The lot does not include a mattress 1x O2 hose is included The incubator is in satisfactory physical condition Version 2 19 Manufacture date 2006 327 Drager Airshields Isolette C2000 Incubator The incubator is able to power up and go through its initial power up test successfully The function buttons and display are in working order In air mode the unit is able to heat up well with the unit able to reach the required temperature and maintain it The units temperature has not been verified There are no skin temperature probes included in lot so unable to test skin mode The oxygen and humidifier modes are selectable but have not been tested The units electrical height adjustment is functional as are the mechanical cot tilts The lot does not include a mattress 1x O2 hose is included in lot The incubator is in satisfactory physical condition with however two of its rubber gaskets for the back access doors missing Version 2 18 355 Valleylab Force FX Diathermy with Footswitch 41562 The diathermy powers up well and passed its initialization Its display functions and indicators are in good working order T
49. uded Overall the units are in reasonable physical condition 621 Olympus TI 1900 The stack system powers up successfully Both video processor and Trolley with Sony light source are in good working order and the monitor is also Trinitron Monitor functional Tested with a videoendoscope and the system produced Olympus CV 240 eee Processor and good quality images with all relevant functions such as white balance Olympus CLV U20 light adjustment and colour settings all in good working order The Light Source interconnecting cable between the processor and the light source is also included Overall the unit is in reasonable physical condition 623 Metro Flex Crash The crash carts are in reasonable physical condition Both carts contain Trolley with Leardal a suction unit LSU which are intact and in good working order Each Suction Unit x 2 cart contains 3 drawers 627 Stryker Head Neck The patient trolley is in average physical condition its upholstery Hydraulic Surgical appears to be fair condition The height adjustment and articulating Bed movements are in good working order 640 Eschmann MR All of the operating table s mechanical articulating movements are in Hydraulic Operating working order The table s hydraulic height adjustment is functional Table No Cushions The caster wheels are freely movable and the brake lock is functional The unit is missing all of its cushions The operating table is in satisfactory physi
50. ulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 130 OLYMPUS GIF H260 Videoscope Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional HG 115 16 v2 1000x11 0mm Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 131 OLYMPUS TJF 260V Videoscope 1200x12 60mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good Objective Lens and Focusing Good optics Remote Control Switches Fully functional Distal End Missing distal cap Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism Functional Leak Test Pass Overall Condition Good overall condition 132 OLYMPUS GIF H260 Videoscope 1000x10 8mm Insertion Tube Good condition Be
51. y skin temperature probes In mattress mode the unit is also able to heat up well with it able to reach the set temperature and maintain it The overhead spot lights are functional The unit includes a gel mattress The electrical height adjustment is functional The unit is in satisfactory physical condition 57 Aesculap TB50 BiPolar Coagulator No Footswitch The unit is able to power up with its function buttons and displays in working order The lot does not include a footswitch or accessories Unable to test the unit or verify if it is able to give an output The unit is in satisfactory physical condition 58 Valleylab Force EZ Diathermy with Footswitches x 2 Both units are able to power up One units function buttons and rotary knobs are not responding the unit does not look to be functional with it unable to switch between bipolar and monopolar modes The second unit has its function buttons and displays in working order Neither of the units outputs have been checked or verified Both the units patient plate alarms are functional The lot includes 1x bipolar and 1x monopolar footswitches which looked to be responding The lot is in satisfactory physical condition 78 Zeiss GDX VCC Laser Diagnostic Unit The unit is able to power up and complete its initial warm up However an initial D2P warning message is displayed which is not able to be removed The unit still showed to be functioning with i
52. ympus WM 60 The stack system powers up successfully Both video processor and Workstation with light source are in good working order and the monitor is also Sony Trinitron functional Tested with a videoendoscope and the system produced Monitor PVM a f 20M2MDE good gulity images with all relevant funicuyns such a white balance Olympus CV 240 light adjustment and colour settings all in good working order The and CLV 40 Light interconnecting cable between the processor and the light source is also Source included Overall the unit is in reasonable physical condition 281 HP Sonos 4500 The ultrasound is able to power up however the ultrasound is unable to Ultrasound with 1x S4 Probe 1x 1 9MHz Probe ECG Leads and Panasonic complete boot up as the unit displays message for Back end has detected missing or defective boards required for imaging Unit also quickly displays a number of error numbers The ultrasounds image memory board and DSR SCSI board have been removed No further HG 115 116 v2 MD830 VCR Image Memory and DSR SCSI Boards Removed checks able to be carried on unit The units included probe appears to be in satisfactory physical condition with its cable having some damage to it covered up by tape and its rubber grip around bottom of probe head being worn away The lot is in satisfactory physical condition Manufacture date 04 04 00 Includes 1x s4 probe 1x 1 9Mhz Doppler 1x 3 wir