Mini Holter Event ECG System - User Manual
Contents
1. 34 8 INSTALLING MINI HOLTER EVENT ECG SYSTEM 35 8 1 Memory card reader 35 8 2 Configure Welch Allyn Mini Holter Event ECG System 35 8 3 Formatting HeartCard memory cards 35 9 TROUBLESHOOTING 37 Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 5 of 42 APPENDIX Using the Mini Holter Event ECG System recorder Instructions for the patient 1 Retrospective test 2 One channel event test 3 Two channel event test Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 6 of 42 Welcome Welcome to Mini Holter Event ECG System a versatile ambulatory ECG recording and analysis system With this system patients can be monitored over a weeks or longer This allows you to capture ECG traces accompanying intermittent or infrequent symptoms not usually captured by conventional Holter systems Intended use Mini Holter Event ECG monitoring is used to capture symptomatic episodes related to rhythm rate conduction and ischemia It may be useful as a diagnostic tool for patients with
2. 26 Apply anti aliasing 26 Bleep level 18 Card overview 34 Chart speed 26 Check recorder time 16 Chest electrode placement 18 Clean recorder 30 Comment 26 Configure 35 Connect recorder to patient 18 Connection quality 22 32 Create new retrospective test 12 Customize Event ECG System 31 DC offset 23 Details tab 24 Event ECG System recorder 8 Event ECG System settings 31 Event ECG System window 7 Event test 11 20 Full disclosure 25 General tab 31 HeartCard formatting 35 Hookup settings
3. 27 3 3 View the test settings 27 4 INTERPRETING A TEST 28 4 1 Edit and confirm an interpretation 28 5 PRINTING A TEST 29 5 1 About printing a test 29 6 MAINTAINING THE RECORDER 30 6 1 About maintaining the recorder 30 6 2 Clean the recorder case 30 7 CUSTOMIZING MINI HOLTER EVENT ECG SYSTEM 31 7 1 About customizing Mini Holter Event ECG System 31 7 2 Event ECG System settings General Tab 31 7 3 Event ECG System settings Hookup tab 32 7 4 Event ECG System Settings Printing tab 33 7 5 Card overview
4. 20 Record event test 20 retrospective test 11 Recorder preparation 15 Recorder time Adjust 16 Check 16 Recording duration 13 Recordings at specific times 14 Reset zoom 25 Retrieve test 20 Retrospective test 10 11 Save battery power 15 Save recording times settings 15 Sensitivity 26 Set chart speed and sensitivity 26 Set ECG recording times 13 Settings General 31 Hookup 32 Printing 33 Start recorder 19 Test Event 11 Interpreting 28 Recording
5. 32 Impedance 23 Installing 35 Instructions for the patient 12 Interpreting a test 28 Maintaining the recorder 30 Make measurements 26 Measurements 26 Memory Card Formatting 35 MemoryCard 10 Mini Holter Event ECG System 6 New test 12 OmniDrive installation 35 One channel test 12 Overview tab 23 Patient connection 18 Patient connection quality 22 32 Patient instructions 12 Power Save 15 Prepare recorder 15 Print settings 33 Print a test 29 Printing tab 33 Read back recordings
6. indicates that the recording was made by the patient indicates that the recording was made automatically at a specific time Impedance and DC offset give information about the patient connection quality 3 You can now view all recordings or one single recording To view all recordings at the same time Click the Overview tab in the workspace To view one single recording at a time 1 Select the recording you want to view 2 Click the Details tab in the workspace The recording is displayed in detail Patient connection quality The Event ECG System recorder automatically measures patient connection quality impedance and DC offset These measurements are made and stored at the end of each recording using a method recommended by the American Heart Association and the Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 23 of 42 Association for the Advancement of Medical Instrumentation When you read back recordings from the Memory Card you can view connection quality for each recording in the Retrieve test from Memory Card dialog box The impedance and DC offset are shown also on printouts DC offset This is the voltage between the two electrodes used to record each channel of ECG It is measured in millivolts mV Ideally DC offset should be zero but small polarization effects due to the half cell potential of each electrode are normal Impedance
7. 8 3 Formatting HeartCard memory cards In the following circumstances it might be necessary to format the HeartCard 1 Card was used with Recollect software 2 03 or earlier 2 Card s back up battery was replaced Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 36 of 42 To format a HeartCard proceed as follows 1 Insert the card into the card reader 2 Start the Recollect format utility which was installed in the Cardio Control group on your Start menu The program s caption will indicate at which drive letter it will expect the HeartCard 3 Enter the card identifier to be programmed onto the card The identifier can be any text you like just make sure the card can be easily recognized with the text used and that all cards you use have a different card identifier 4 Press Start to start the formatting process Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 37 of 42 9 Troubleshooting When I use the Format utility I cannot set the recording time to correspond with the size of the Card The actual size may differ slightly from the size as mentioned on the card e g a 90 minute card may in reality be a 88 53 card or so Check if the header in the format utility displays the correct drive letter The actual drive letter of the card reader can be found with the use of Windows Explorer to set the correct drive letter in the WACP settings m
8. This means that you cannot access that functionality at that particular point either because it is not available or because you are not authorized to use it Tool bar The tool bar contains buttons that give you easy access to the most common tasks Search area The search area on the left hand side contains search and display functionality that lets you easily retrieve patients and tests In the search area you can find a patient see which tests were recorded for that patient and see the kind of tests that were recorded You can also create search patterns an easy way of finding information that you need frequently Workspace The workspace displays the test and the recordings of a test This is where you do the actual recording viewing and interpreting The workspace contains a viewing area with a number of tabs that let you see the test in different ways The tabs that are available are Recordings Overview and Details Menu bar Title bar Toolbar Search area Workspace Status bar Interpretation area Comments area Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 8 of 42 Status bar The status bar at the bottom of the window shows information about the patient and test that are currently selected such as the test and patient name and recording date 1 2 The Event ECG System recorder This section describes the Event ECG recorder 1 Memory Card 2 Slot for Memory
9. 11 20 Retrospective 10 Test settings 27 Timed recordings 14 Tools menu 28 Two channel test 11 12 View Information about a test 27 Test 22 Test settings 27 View menu 25 Zoom 25
10. 2 3 4 6 and 7 are explained in more detail on the following pages 1 Select a Memory Card of appropriate duration and place it in the card reader that is connected to your computer 2 Select a patient or create a new one 3 Create a new test Note If the patient s symptoms require a large series of recordings you can supply him with more than one Memory Card You then need to store the test settings on all cards which means that you have to create the same test more than once each time with a new card in the card reader 4 Prepare the recorder before connecting it to the patient 5 Place the Memory Card in the recorder note the arrow Push the card fully in 6 Connect the recorder to the patient 7 Start the recorder Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 12 of 42 8 The patient carries the recorder with him for some time To ensure that the recorder will record properly you need to give the patient a sheet with a few instructions See the appendix you can make a copy of it 9 When the patient returns place the Memory Card in the card reader again and read back the recordings Create a new retrospective test Before you can record a retrospective test you first need to create a new test The test settings are stored on the Memory Card that you put in the card reader that is connected to your computer Note If the patient s symptoms require a large series
11. ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 24 of 42 To view one recording in detail Double click the curve of that recording The Details tab is displayed It shows the recording in detail starti ng at the point you double clicked Available actions on the Overview tab Zoom in or out on the recordings see section 0 for details Set chart speed and sensitivity see section 0 for details Apply anti aliasing see section 0 for details Details tab On the Details tab you can view one single recording in different ways You can view a detail of a recording or a whole recording When you view a recording in detail all used channels are shown When you view a whole recording you view the used channels separately Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 25 of 42 To view a whole recording Click the Full Disclosure button at the top of the tab The whole curve of one channel is displayed When two channels are recorded you can choose another in the Lead list on the top right of the tab To view another recording Choose a recording from the Recordings list at the top of the tab Alternatively you can use the arrows on the right of the list to scroll through the recordings of the test To view all recordings at the same time Click the Overview tab Available actions on the Details tab Zoom in or o
12. ECG System recorder Instructions for the patient 1 Retrospective test 2 One channel event test 3 Two channel event test Using the Event ECG System recorder retrospective test Keep the recorder in good working condition You will wear this recorder for some time To ensure that the recorder stays in good working condition please observe the following Do not expose the recorder to temperatures below 0 C or above 45 C Do not expose the recorder to prolonged direct sunlight Protect the recorder against water Do not attempt to remove the covers of the recorder When it is malfunctioning refer to the hospital Make a recording 1 Press down and hold the record button 2 Release the button when the green light comes on If the bleeper is switched on there will be a long bleep When the green light goes out the recording is finished The recorder reverts to displaying the number of recordings available and continues as before ready for the next recording Note If the record button is pressed again too soon the beginning of the next recording will be blank When a card is full the bleeper sounds and the recorder switches off Replace the batteries In this test the recorder constantly records ECG Because of that the batteries do not last very long You have to replace both batteries when the recorder displays this sign
13. This is the total impedance of the two skins to electrode interfaces per channel plus the effect of any discontinuities in the wires or connectors It is expressed in kilohms k and is measured using a 10Hz frequency Good connection quality depends on achieving both low impedance and a low DC offset In the majority of patients impedance should ideally be lower than 50k and DC offset lower than 10mV It is impossible to achieve these low values in some patients Impedance for females is about 50 higher than for males High impedance higher than 200k indicates poor contact between the electrode and the skin Both impedance and DC offset may improve during the first few minutes after electrode application as the interface between the skin and the electrode stabilizes Extremely high impedance higher than 1450k suggests an electrode has come off the skin or that the cable or connector has broken High impedance does not necessarily result in bad ECGs However the poor connection it implies will make the recordings more susceptible to interference due to external electrical sources or patient movement Both impedance and DC offset may vary unpredictably over time influenced by temperature humidity and patient activity Overview tab On the Overview tab you can view all recordings of one test Use the horizontal and vertical scrollbars to see all curves completely Mini Holter Event
14. chest pain palpitations neurologic symptoms shortness of breath or to monitor response to treatment of cardiac disorders Indications The clinical indications for mini Holter Event electrocardiography include assessment of symptoms that may be related to arrhythmias assessment of risk in patients with or without symptoms of arrhythmia assessment of efficacy of antiarrhythmic therapy detection of myocardial ischemia Contraindications There are no known contraindications for the recording of mini Holter Event ECG tests Operator knowledge To use this device extensive Holter knowledge is required that allows proper identification of sinus and paced rhythms abnormal rhythms supraventricular and ventricular arrhythmias artifact ST segment changes and pacemaker failures In addition all instructions assume a working knowledge of computers and specifically Windows software Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 7 of 42 1 Getting started 1 1 The Event ECG System window This section will guide you around the various parts of the Event ECG System software Title bar The title bar displays the name of the program The title bar contains three buttons that you can use to maximize minimize and close the mini Holter Event ECG System window Menu bar Each menu in the menu bar contains a number of tasks that you can perform Sometimes menus are gray
15. entered If you wish you can add some comments to the test To view information about a test Click the Info button on the toolbar The Information dialog box is displayed To add a comment to a test 1 Click the Info button on the toolbar The Information dialog box is displayed 2 Click in the Comment field and start typing 3 Click OK to save you comment 3 3 View the test settings You can view the settings of the test which were entered when the test was created such as the type of test retrospective or event the number of channels and the duration of the recordings To view the test settings In the View menu click Card The Card Settings window is displayed Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 28 of 42 4 Interpreting a test 4 1 Edit and confirm an interpretation You can edit an interpretation in the Interpretation editor In this editor you can enter your own texts Alternatively you can use the statement tree to select and enter common interpretation statements Once you edit an interpretation you must confirm it Otherwise your changes will not be saved 1 In the Tools menu click Interpretation The Interpretation editor is displayed 2 In the right hand pane at the pointer start typing 3 Click the Confirm button to save your comments and to return to the test To edit and confirm an interpretation with the statement
16. good The recorder will display You have to the patient symbol continuously the heartbeat symbol flashing at each heartbeat Continue with step 2 If the connection is bad The recorder will display You have to the patient symbol flashing Solve connection problems To solve connection problems Check the plug leads electrodes skin preparation When the patient symbol stops flashing the problem is solved Note The recorder displays the symbol if the fault is on channel two 2 Press the record button again briefly to complete the quality test and start retrospective monitoring The recorder will now display the following symbols Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 20 of 42 heartbeat symbol or looping symbol retrospective test with one or two channels number of recordings available 2 3 Recording an event test To record a test in event mode you can follow the steps of recording a retrospective test Please follow the instructions on page11 and further Instead of selecting the Retrospective mode you select the Event mode 2 4 Read back recordings When the patient has returned with the recorder you need to read back the recordings from the Memory Card and transfer them to your computer Once you have transferred them to your computer you can view them with Mini Holter Event ECG System To read back
17. of recordings you can supply him with two Memory Cards You then need to store the test settings on two cards which means that you have to create the same test twice To create a new retrospective test you need to follow these global steps Step 5 6 and 7 are explained below in more detail 10 In the File menu point to New and click Recollect test The New Recollect Test dialog box is displayed 11 Select the physician that ordered the test and a specialty Click Next The Recording tab is displayed Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 13 of 42 12 Select Retrospective Mode as recorder mode 13 Select the number of channels you want to use 14 Set the ECG Recording Times The Possible Recordings field shows how many recordings of the selected duration can be stored on the Memory Card and the Utilization fi eld shows how much of the card capacity will be used 15 On the Times tab you can set timed recordings if you also want the recorder to make automatic recordings 16 On the Times tab you can activate the Power Save function if you want to extend battery life 17 Click OK The test settings are now stored on the Memory Card Tip for creating a new Event ECG test If you want to use the test settings again for another test you can save them on your computer You can use test settings you have saved by choosing the proper settings from the Schem
18. reader Before you can start recording Recollect tests you need to Install the memory card reader if it is not a built in part of the computer Configure the software Configure the memory card Insert the card into the Event ECG System recorder 8 1 Memory card reader On most laptop systems the built in PC Card slots can be used for the Mini Holter Event ECG System HeartCard memory cards If you are using a desktop machine it may be necessary to install an external PC Card reader Welch Allyn Cardio Control BV actively supports the OmniDrive external parallel port card reader and the OmniDrive PRO USB card reader Please refer to the documentation that came with the drive for instructions on how to install After installing the card reader please make a note of the drive letter that the operating system has assigned to this card drive This information will be necessary for Event ECG System software configuration 8 2 Configure Welch Allyn Mini Holter Event ECG System After making sure the card reader is present you need to configure Welch Allyn Mini Holter Event ECG System To configure Welch Allyn Mini Holter Event ECG System for use with the card reader 1 Start Welch Allyn Mini Holter Event ECG System 2 In the File menu click Settings and click Recollect 3 Set the PC Card drive letter to the letter assigned to the card reader by the operating system 4 Click OK to save the settings
19. recordings 1 Remove the Memory Card from the recorder and put it in the card reader that is connected to your computer 2 Select the patient who made the recordings 3 From the test list select the test the recordings belong to Event ECG System shows that the test is in progress 4 Click Actions gt Retrieve Data The Retrieve test from Memory Card dialog b ox is displayed 5 If you want to preview the recordings before you transfer them to your computer click the Preview button 6 Click the Retrieve button to transfer the recordings to your computer Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 21 of 42 Note If you click Cancel the recordings are not transferred but left on the card If you click Clear the recordings not the test settings are deleted from the card The recordings are not transferred to the computer 7 Now you can view the recordings Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 22 of 42 3 Viewing a Event ECG System test 3 1 View a Event ECG System test To view an Event ECG System test 1 Select a patient The tests that belong to the patient are shown in the Test list 2 Select the Event ECG System test you want to view Event ECG System tests are indicated with Event ECG System is started automatically and the recordings are displayed in the workspace on the Recordings tab
20. tree 1 In the Tools menu click Interpretation The Interpretation editor is displayed 2 From the statement tree click on one of the categories to display all possible statements for that category 3 From a category click on the statement that you want to include in the interpretation The statement is added 4 To delete a statement from the interpretation select the statement text and press BACKSPACE to delete it 5 Click the Confirm button to save your comments and to return to the test Tips for editing and confirming an interpretation You can automatically insert the current date and time by clicking the Date time button You can clear the interpretation editor by clicking the Clear button The statements in the statement tree can be changed Contact your system administrator or local dealer for new or changed statements Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 29 of 42 5 Printing a test 5 1 About printing a test With Mini Holter Event ECG System you can print the ECG recordings of a test on plain paper with a regular inkjet or laser printer Note You can print one recording at a time it is not possible to print the whole test at the same time Print formats Mini Holter Event ECG System can print the following reports Details Complete details Full disclosure 25 mm s Full disclosure 10 mm s Full dis
21. 17 of 42 6 Change the value of the selected parameter by pressing the record button 7 Repeat this for all parameters you want to change Note If you change the minute first you can only change the hour directly after that not the date If you also want to change the date you need to step through the sequence twice 8 Press the clock set button repeatedly until the time is displayed again The changes you have made are now stored 9 Replace the battery compartment cover Tips The seconds value is automatically set to zero whenever the time or date is changed The recorder switches off if more than ten seconds has elapsed since the last button press To adjust the time and date by using the Memory Card and the Event ECG System software 1 Place the Memory Card in the card reader 2 In the Tools menu select Set recorder Date time The Configure the Recollect recording device dialog box is displayed 3 Insert an empty Memory Card into the reader Click Next 4 Set a date and a time in the near future Click Next 5 Put the card in the recorder Click Next 6 Start the recorder at the specified time see section 0 Start the recorder The recorder now displays the correct time and date Click Next 7 Click OK Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 18 of 42 Adjust the bleep level The recorder has a bleeper to indicate its function when
22. 5 0 Page 31 of 42 7 Customizing Mini Holter Event ECG System 7 1 About customizing Mini Holter Event ECG System You can customize Event ECG System to your own preferences The features that you can customize are located in File gt Settings gt Recollect Important When you change the settings in the General tab they are only changed on the computer that you are currently working on Other computers remain unchanged 7 2 Event ECG System settings General Tab Use the General tab to customize general settings of Event ECG Sys tem Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 32 of 42 The General tab contains the following information PC Card Drive From this list you can select the drive to which the Memory Card reader is connected Appearance Here you can specify individual colors for the background grid lines grid dots trace and the font that is used in displaying the ECG Click on the color to select another color Color schemes This list lets you choose from a number of predefined color formats Gridstyle From this list you can select the style of the grid in which the ECG recordings are displayed You can either leave out the grid use lines to represent the grid or display a full grid Overview With this option you can determine how many rows of ECG trace are shown on the Overview tab The more rows are shown the less you have to scroll horizontally Detai
23. Card 3 Warning bleeper 4 Display 5 Recording indicator 6 Patient cable connection 7 Record button 8 Patient cables 9 Chest electrodes 10 Wrist electrodes Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 9 of 42 11 Battery compartment 12 Clock setting button 13 Bleep setting button 14 Instruction label 15 Battery compartment label Note the meaning of the following symbols Attention consult accompanying documents Defibrillator proof type BF electrocardiograph Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 10 of 42 2 Recording a mini Holter Event ECG test 2 1 About recording a test With Welch Allyn Mini Holter Event ECG System you can easily record ambulatory ECG tests You can choose between two test modes retrospective or event In both modes you can record one or two channels of ECG A mini Holter Event ECG test consists of a number of ECG recordings made by the patient These recordings are stored on the Memory Card To start a test you place this card in the card reader that is connected to your computer You create a new test and prepare the recorder for use After placing the Memory Card in the recorder you connect the recorder to the patient except when you record an event test with one channel of ECG In that case the patient carries the recorder with him and attaches t
24. Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Welch Allyn Cardio Control BV Staalweg 50 2612 KK Delft The Netherlands Welch Allyn Inc 4341 State Street Road Skaneateles Falls NY 13153 0220 USA www welchallyn com 0297 Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 2 of 42 Instructions for use The information contained in this manual is subject to change without notice All changes will be in compliance with regulations governing manufacture of medical equipment Document WACPW MHO ENG Rev 02 Copyright Welch Allyn Cardio Control BV 2006 No part of this publication may be reproduced transmitted transcribed stored in a retrieval system or translated into any human or computer language in any form by any means without the consent of the copyright holder Unauthorized copying of this publication may not only infringe copyright but also reduce the ability of Welch Allyn Cardio Control BV to provide accurate and up to date information to users and operators alike User responsibility This product is designed to perform in conformity with the description thereof contained in this operation manual and accompanying labels and inserts when assembled operated maintained and repaired in accordance with the instructions provided A defective product should not be used Parts that are broken plainly worn missing or incomplete distor
25. al The recorder will switch off automatically approximately eight hours after the warning starts Do not remove the Memory Card 1 Hold the recorder w ith its underside facing you Remove the battery cover by pressing down on the finger grip and sliding it sideways 2 Lift out the old batteries Replace with two AA MN1500 alkaline batteries Note the correct polarity when inserting them 3 Replace the battery cover by sliding it until the catch engages 4 Press the record button once when the bleep starts Replace the Memory Card If you need a large series of recordings you have received more than one Memory Card When a card is full you have to replace it Note Do not remove the full Memory Card directly after the last recording is made The recorder needs a few seconds to store the recording 1 Grip the exposed part of the Memory Card and pull firmly 2 Place the new card in the recorder note the arrow Push the card fully in Using the Event ECG System recorder event test one channel Keep the recorder in good working condition You will wear this recorder for some time To ensure that the recorder stays in good working condition please observe the following Do not expose the recorder to temperatures below 0 C or above 45 C Do not expose the recorder to prolonged direct sunlight Protect the recorder against water Do no
26. asurements When you view a single recording on the Details tab you can make measurements between two points To make a measurement between two points 1 Move the crosshair pointer to the starting point of your measurement 2 Double click on the starting point Event ECG System inserts a mark 3 Move the crosshair pointer to the end point of your measurement 4 Double click on the end point Event ECG System draws a line between the two points and displays the measured values Add a comment to a recording On the Details tab you can enter a comment for a single recording To add a comment Start typing in the Recording Comment area at the bottom of the workspace Your comment is saved automatically Apply anti aliasing When you view an ECG on a monitor the signal can look ragged especially when you zoom in on a complex You can use anti aliasing to make the signal look smoother Anti aliasing does not change the signal itself It only affects how the signal is displayed on the monitor To apply anti aliasing In the View menu click Anti aliasing Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 27 of 42 3 2 View information about a test The Information window gives you all kinds of information about the ECG that you are currently viewing You can see the name of the referring physician the name of the technician that recorded the ECG and any comments that were
27. closure 5 mm s Recordings Overview Each report contains the view and a header that displays additional information For more information about printing recordings please refer to the Workstation manual Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 30 of 42 6 Maintaining the recorder 6 1 About maintaining the recorder To keep the recorder in good working condition please observe the following Remove the battery when the recorder is not used for a longer period of time When the recorder is being used it should not be exposed to temperatures below 0 C or above 45 C and to prolonged direct sunlight When you store the recorder make sure it is not exposed to temperatures below than 20 C or above 65 C Protect the recorder against water Do not attempt to remove the covers of the recorder In the event of a malfunction refer servicing to the manufacturer or an approved distributor Clean the recorder case regularly 6 2 Clean the recorder case To minimize the possibility of cross infection the case should be cleaned each time the recorder is returned by a patient To clean the recorder case Clean the exterior surface of the recorder with a cloth dampened with a non abrasive detergent Note Do not immerse the instrument in any liquid Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1
28. corder with its underside facing you Remove the battery cover by pressing down on the finger grip and sliding it sideways Put in two AA MN1500 alkaline batteries Duracell or Procell Note the correct polarity when inserting the batteries Replace the battery cover by sliding it until the catch engages Press the record button once when the bleep starts Make a recording 1 Press down and hold the record button 2 Release the button when the green light comes on When the green light goes out the recording is finished Note A subsequent recording cannot be made until the recorder display goes blank Replace the Memory Card If you need a large series of recordings you have received more than one Memory Card When a card is full you have to replace it Note Do not remove a full Memory Card directly after the last recording is made The recorder needs a few seconds to store the recording 1 Grip the exposed part of the Memory Card and pull firmly 2 Place the new card in the recorder note the arrow Push the card fully in Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 01 Version 1 5 0 Page 42 of 42 Index Add Comment to a recording 26 Comment to a test 27 Adjust Bleep level 18 Recorder time 16 Anti aliased
29. e available recording time and using only two electrodes Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 11 of 42 Recordings at specific times In retrospective mode you can also set Event ECG System to make automatic recordings at specific times during day and night Event mode You can use event mode if the patient has rather infrequent and irregular symptoms you do not need to record the onset of an event With an event test patients can be monitored over several weeks or even months Event tests can be conducted with one or two channels You can freely adjust the amount of ECG that is stored It is only limited by the capacity of the Memory Card that is used One c hannel event test In a one channel event test the patient carries the recorder with him during the test period If he feels an irregularity in his heartbeat he attaches a pair of electrodes to the wrists and activates the recorder After the ECG recording is made the electrodes are removed and the recorder is put away Two channel event test In a two channel event test the recorder is constantly connected to the patient During the whole test period the electrodes are attached to the patient s chest If the patient feels an irregularity he activates the recorder and ECG is recorded 2 2 Recording a retrospective test To record a test in retrospective mode you need to follow these global steps Steps
30. e field Set ECG recording times When you create a retrospective Event ECG test you need to set these ECG recording times You can do this in the New Recollect Test dialog box 1 In the Recording Duration field enter the total time ECG is stored when the patient makes a recording You can enter the time value in two ways the number of seconds or the number of minutes Example The total recording duration is one minute If you want to enter the value in Type seconds 60 minutes 1 00 Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 14 of 42 2 In the Pre Trigger Time field enter the time ECG is stored before the patient presses the record button You can also set the pre and post trigger time by using the slide bar Note The total amount of time before and after the button press always equals the total recording duration When you have entered the pre trigger time the post trigger time is automatically determined Set timed recordings Note Unless you only want to record timed recordings ensure that the ECG duration and timed recordings that you choose will not fill the Memory Card prematurely For example if the ECG duration single channel is set to 3 minutes and 15 timed recordings per day are selected a 90 minute Memory Card will be filled in 48 hours Either use a card with more capacity or shorter ECG duration 1 In the New Recollect Test dialog box click the Tim
31. e made until the recorder display goes blank Replace the electrodes after 10 applications or sooner if they do no stick well any longer Replace the Memory Card If you need a large series of recordings you have received more than one Memory Card When a card is full you have to replace it Note Do not remove the full Memory Card directly after the last recording is made The recorder needs a few seconds to store the recording 1 Grip the exposed part of the Memory Card and pull firmly 2 Place the new card in the recorder note the arrow Push the card fully in Using the Event ECG System recorder event test two channels Keep the recorder in good working condition You will wear this recorder for some time To ensure that the recorder stays in good working condition please observe the following Do not expose the recorder to temperatures below 0 C or above 45 C Do not expose the recorder to prolonged direct sunlight Protect the recorder against water Do not attempt to remove the covers of the recorder When it is malfunctioning refer to the hospital Fit the batteries In this test you do not use the recorder constantly Therefore it is better that you do not store the batteries in t he recorder Take them out after a recording and fit them again when you want to make a recording Do not remove the Memory Card Hold the re
32. ecord button should be visible through the cut out in the case Note Do not press the record button when inserting the recorder 2 If the two channel cable is being used fit the neck cord around the patient s neck to support the cable yoke This prevents signal artifact which can result from the leads pulling on the electrodes 3 Prepare the patient s skin Make sure it is thoroughly clean and dry so that the electrodes will adhere well for a long period without causing irritation 4 Place the electrodes and attach the snap connectors on the lead wires at the positions indicated by the illustrations Note Use Medicotest VLC 00 S electrodes and other soft gel types These electrodes give good results particularly when the patient is active Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 19 of 42 One channel Two channels 5 Hook the neck cord into the cable yoke stress relief molding to support the cable This helps avoid the need for adhesive tape 6 You can now start the recorder Start the recorder During a retrospective test the patient wears the recorder constantly Therefore you have connected the recorder to the patient Now you have to start the recorder so that a continuous loop of ECG will be recorded To start the recorder 1 Press the record button briefly The recorder automatically tests the quality of the connection to the patient If the connection is
33. enu go to the Recollect settings General tab Is formatting the card with the Recollect Format utility the same as formatting with the Windows formatting program The Windows formatting program really puts a new data structure on the card normally this is not required and should be avoided In some cases especially with Windows XP this can make the card unsuitable for use with the Recollect recorder The Recollect Format utility creates a file called Recollect bin on the card in this file the ECG recordings are stored This allow s you to reserve a certain amount of space on the card The patient symbol flashes almost each time a measurement is performed telling me the electrode quality is not good although at the initialization the connections were tested O K If the electrodes are attached for a longer time the connection may become poor due to transpiration If this happens often check whether proper long lasting electrodes are used if the skin is prepared properly and if the threshold values for DC offset and Impedance are not set too low If the patient symbol flashes already at the initialization check the following Was the skin well prepared Is the patient cable in good condition visible inspection Is the cable connector pushed into the recorder correctly it has to click when inserting it Are the threshold values for DC offset and Impedance not set too low Appendix Using the Mini Holter Event
34. es tab 2 Select the Timed Recordings Mode box 3 You can now set the timed recordings in two ways by entering all specific times or by setting a recording cycle To enter specific times 1 In the Time field enter a time in the format hh mm 2 Click the Set button to set the entered time 3 Repeat these steps for all times you want to set To set a cycle 1 In the Every hours field enter a number of hours in the format hh mm 2 In the Starting at field enter the first time of the day a recording must be made 3 Click the Set button to set the cycle Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 15 of 42 4 The recorder will bleep when an automatic recording is made To mute this bleep select the Mute Bleep box Activate Power Save You can extend battery life by switching off the loop recording and the display functions while the patient is asleep 1 In the New Recollect Test dialog box click the Times tab 2 Select the Power Save Mode box 3 Enter the time you want the recorder to turn off and the time you want the recorder to turn on again Save test settings After setting the test settings you can save them The settings are saved on your computer so that you can use them again for another test To save the test settings 1 In the New Recollect Test dialog box click the Save As button The Save Scheme dialog box is displayed 2 E
35. he electrodes when he wants to make a recording When the patient returns you take the Memory Card from the recorder place it in the card reader again and read back the stored measurements Memory Card A Memory Card is a removable card which is used to store a patient s ECG recordings The Memory Card is available in two capacities 90 and 180 minutes If a patient needs a very large series of recordings he can be supplied with more than one card Retrospective mode You can use retrospective mode if the patient has rather frequent symptoms you want to record the onset of an event Monitoring can last from a few days up to a few weeks You can freely adjust the amount of ECG recorded before and after activation It is only limited by the capacity of the Memory Card that is used In a retrospective test the patient wears the recorder for a certain period The recorder is connected to the patient with special long term electrodes It continuously records ECG into a loop of memory until the patient feels an irregularity in his heartbeat and chooses to make a recording When the patient activates the recorder by pressing the button ECG is stored from both before and after activation One or two channel retrospective tests Retrospective mode is commonly used to record two channels of ECG enabling better discrimination of cardiac rhythm and conduction The recorder can however operate on one channel only giving twice th
36. ls With this option you can determine how many rows of ECG trace are shown on the Details tab The more rows are shown the less you have to scroll horizontally 7 3 Event ECG System settings Hookup tab Use the Hookup tab to set the thresholds the recorder uses to test patient connection quality When the recorder is connected to the patient it tests the quality of the connection The recorder checks if the impedance and the DC offset are acceptable To determine this the recorder needs a threshold You can set this threshold on the Hookup tab The threshold values are then stored on the Memory Card and rea d by the recorder when it checks the connection quality Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 33 of 42 As a default the impedance threshold is 800 kOhm and the DC offset threshold 80 mV for both retrospective and event mode These are the recommended values You can change these values by choosing other values from the lists The next time a new test is created the changed settings are stored on the Memory Card 7 4 Event ECG System Settings Printing tab Use the Printing tab to select the reports that you want to use for the Print selected formats function The Printing tab contains the following information Default report templates Here you can select the formats that you want to print when you use the Print selected formats function You can select multiple format
37. nter a name for the settings 3 Click OK Prepare the recorder Before you can place the Memory Card with the stored settings in the recorder you have to prepare the recorder provide the recorder with a fresh set of batteries attach the patient cable check if the recorder displays the correct time and date adjust the bleep level if necessary Fit the batteries 1 Remove the Memory Card if fitted 2 Hold the recorder with its underside facing you Remove the battery cover by pressing down on the finger grip and sliding it sideways 3 Put in two AA MN1500 alkaline batteries Duracell or Procell Note the correct polarity when inserting the batteries Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 16 of 42 4 Replace the battery cover by sliding it until the catch engages Attach the patient cable 1 Select the appropriate chest cable and the neck cord The cable you choose must match the number of channels you selected when you created the test 2 Align the marks on the cable grip and recorder and push the plug firmly into the recorder socket until it clicks into place If the plug is difficult to insert try rotating it slightly as you insert it The plug cannot be removed by pulling on the cable only by pulling back the metal sleeve so it is unlikely that the patient disconnects the cable accidentally Note Do not attach any other type
38. oblem and to check all accessories to ensure that they are not the cause of the problem When calling please be prepared to provide Product name and model number and complete description of the problem The serial number of your product if applicable The complete name address and phone number of your facility For out of warranty repairs or spare parts orders a purchase order or credit card number For parts order the required spare or replacement part number s Repairs If your product requires warranty extended warranty or non warranty repair service please call first the nearest Welch Allyn Technical Support Center A representative will assist you troubleshooting the problem and will make every effort to solve it over the phone avoiding potential unnecessary return In case the return cannot be avoided the representative will record all necessary information and will provide a Return Material Authorization RMA number as well as the appropriate return address A Return Material Authorization RMA number must be obtained prior to any return Packing Instructions If you have to return goods for service follow these recommended packing instructions Remove all hoses cables sensors power cords and ancillary products as appropriate before packing unless you suspect they are associated with the problem Wherever possible use the original shipping carton and packing materials Include a packing lis
39. of cable to the recorder Do not attach the patient cable electrodes to any other electrical device Check the time and date The recorder has a built in calendar and clock that are used to record the time and date on the Memory Card each time a recording is made Therefore make sure the recorder is set at the correct time and date 1 Remove the Memory Card if fitted 2 Press the record button briefly and check the time The display switches off automatically after a few seconds 3 Adjust the time if necessary Adjust the time and date You can adjust the calendar and clock in two ways by using the clock button located within the battery compartment by storing the correct time and date on the Memory Card with the Event ECG System software To adjust the time and date using the clock button 1 Remove the Memory Card if fitted 2 Hold the recorder with its underside facing you 3 Remove the battery compartment cover by pressing down on the finger grip and sliding it sideways The clock set button is the triangular button next to the batteries 4 Press the record button The time is now displayed 5 Press the clock set button to scroll to the parameter you want to change To change the Press the clock set button minute once hour twice day three times month four times year five times Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page
40. s Grid From the list you can select the grid style in which the selected templates should be printed Full Disclosure Lead From the list you can select the default lead that appears in the Full disclosure view and print Print after retrieving test data from card If this option is selected the results are automatically printed when the test data is retrieved from the memory card The software prints the templates that you selected in the Default report template s area Print after interpretation confirmed If this option is selected the results are automatically printed when the interpretation of the test has been confirmed The software prints the templates that you selected in the Default report templates area Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 34 of 42 7 5 Card overview The card overview displays the Memory Cards that are currently in use the patients to whom they are assigned and the times when recordings have started To view the card overview In the Tools menu click Card overview The Card overview is displayed Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 35 of 42 8 Installing Mini Holter Event ECG System Mini Holter Event ECG System is made up of four elements the Event ECG System recorder the software that runs on the computer the SRAM memory card on which ECG data is recorded and the memory card
41. t and the Welch Allyn Return Material Authorization RMA number It is recommended that all returned goods be insured Claims for loss or damage to the product must be initiated by the sender Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 4 of 42 Table of Contents 1 GETTING STARTED 7 1 1 The Event ECG System window 7 1 2 The Event ECG System recorder 8 2 RECORDING A MINI HOLTER EVENT ECG TEST 10 2 1 About recording a test 10 2 2 Recording a retrospective test 11 2 3 Recording an event test 20 2 4 Read back recordings 20 3 VIEWING A EVENT ECG SYSTEM TEST 22 3 1 View a Event ECG System test 22 3 2 View information about a test
42. t attempt to remove the covers of the recorder When it is malfunctioning refer to the hospital Fit the batteries In this test you do not use the recorder very often Therefore it is better that you do not leave the batteries in the recorder Take them out after a recording and put them in again when you want to make a recording Do not remove the Memory Card 1 Hold the recorder with its underside facing you Remove the battery cover by pressing down on the finger grip and sliding it sideways 2 Put in two AA MN1500 alkaline batteries Duracell or Procell Note the correct polarity when inserting the batteries 3 Replace the battery cover by sliding it until the catch engages 4 Press the record button once when the bleep starts Make a recording 1 Remove the electrodes from the backing strip and firmly press an electrode onto the inside of each wrist black on the left white on the right 2 Press down and hold the record button 3 Release the record button when the green light comes on Try to relax with your hands and the recorder resting on the lap 4 When the green light goes out the recording is complete Remove the electrodes and replace them onto the backing strip Remove the batteries from the recorder Note You do not have to unplug the cable from the recorder However do not wrap it around the recorder this can cause the cable to break A subsequent recording cannot b
43. ted or contaminated should be replaced immediately Should any repair or replacement become necessary we recommend that service be performed at the nearest factory approved service center The user of the product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone other than WelchAllyn Cardio Control BV or their authorized service personnel Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 3 of 42 Warranty Service and Spare Parts Warranty All repairs on products un der warranty must be performed or approved by Welch Allyn Unauthorized repairs will void the warranty In addition whether or not covered under warranty any product repair shall exclusively be performed by Welch Allyn certified service personnel Assistance and Parts If the product fails to function properly or if assistance service or spare parts are required contact the nearest Welch Allyn Technical Support Center USA 1 800 535 6663 Canada 1 800 561 8797 Latin America 1 305 669 9591 South Africa 27 11 777 7509 European Call Center 353 469 067 790 Australia 61 2 9638 3000 United Kingdom 0 207 365 6780 Singapore 65 6291 0882 France 33 1 60 09 33 66 Japan 81 3 5212 7391 Germany 49 7477 927 173 China 86 21 6327 9631 Before contacting Welch Allyn it is helpful to attempt to duplicate the pr
44. the display is not visible for instance when the recorder is worn under clothing It only operates in retrospective mode When a recording is initiated by the patient a long bleep is emitted If the recorder switches off because of a fault e g low battery level a series of short bleeps is emitted The bleeper can be set to one of four levels Level 0 is off levels 1 to 3 are different types of sound To adjust the bleep level 1 Remove the Memory Card 2 Hold the recorder with its underside facing you 3 Remove the battery compartment cover by pressing down on the finger grip and sliding it sideways The bleep set button is the semicircular one next to the batteries 4 Press the record button 5 Press the bleep set button once The display shows the current bleep level 6 Press the record button repeatedly until the correct level is displayed Each time the record b utton is pressed the recorder bleeps once at the new level except at level 0 7 Press the bleep set button The time will be displayed 8 Replace the battery compartment cover Note The recorder switches off if more than ten seconds have elapsed since th e last button press Connect the recorder to the patient In a retrospective test the patient wears the recorder constantly during a certain period Therefore you have to connect the recorder to the patient To connect the recorder to the patient 1 Put recorder into the carrying case The r
45. ut on the recordings see section 0 for details Set chart speed and sensitivity see section 0 for details Make measurements see section 0 for details Add a comment to a recording see section 0 for details Apply anti aliasing see section 0 for details Zoom in or out on a ECG recording If you want to take a closer look at the recording you can adjust size of the curve by zooming in and out To see the recording in more detail Click the Zoom In button on the toolbar To see a larger part of the recording Click the Zoom Out button on the toolbar To reset the zoom function and return to the standard curve size In the View menu click Reset zoom Mini Holter Event ECG System User Manual WACPW MHO ENG Rev 02 Version 1 5 0 Page 26 of 42 Set chart speed and sensitivity Mini Holter Event ECG System offers a wide range of speed and sensitivity You can adjust the scale of the ECG by adjusting the chart speed and the sensitivity To set chart speed 1 In the View menu point to Chart speed and select a value The higher the chart speed the more curve will be visible or 2 Choose the chart speed from the list mm sec in the toolbar To set sensitivity 1 In the View menu point to Sensitivity and select a value The lower the value the lower the peaks in the curve will be or 2 Choose the chart speed from the list mm mV in the toolbar Make me