Home
nf 1200 / 1200 r multi-purpose bench top centrifuges user`s manual
Contents
1. 14 462 IMBALANCE DETECTION Z aaah 14 77 s aaa 14 SECHON 522522 ob E 15 OPERATING PHINCIPLES ad da amad Dv s au d ui 15 5 15 PREPARATION OF THE ROTOR TO BUN 15 52 ADI AA 15 5 9 TAO TOR ING TALIA TION aa nab He Pete a 16 54 5 oo ca o Pide b b b 17 SECTION 17 CLEANING AND PERIODIC 2222 2 52 17 6 L PERIODIC MAINT ENANG 17 6225 6 17 6 0 CORROSIONINEOR MA TION 17 Dod Chemical b op 18 STRESS CORROSION m AA 18 65 NI CPC n 18 ELECTE C T s nan ss s s n 18 SECTION 7 UNS 19 DISPOSAL MANAGEMENT 2 222 19 SEC HON o a a 19 5 lcm cc 1 05501056 56568 R IR A B 5000002
2. 19 SECTION S 225 6452 025 2024 2280 255 0000000000 MELDE M ME 20 ELECTRICAL CIRCUIT D BGRAMSO a s la b d a d 20 9 1 ELECTRICAL GIRGUIT DIAGRAM OF NF 1200 2 ie 20 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R ui 21 ca Se eet N B Aaaa 22 WARNING LABEL wa AA AAA Ra sil 22 SECTION 1 INTRODUCTION 1 1 USE AND FUNCTION The NF 1200 NF 1200 R centrifuges are designed to separate liquids of different densities by applying centrifugal force Different types of rotors which are svving out and angle rotors can be fitted to the centrifuges The microprocessor control system of the centrifuges enables the user to observe and to control the speed RPM or RCF the acceleration and braking rates run time program number and temperature of the bowl for NF 1200 R The microprocessor control system also displays the RPM or RCF value during the run The ventilation system of the NF 1200 limits temperature in the bowl between the ambient temperature 3 C and ambient temperature 12 C depending on the rotor type The NF 1200 R is equipped with a cooling system which provides to operate between 9 C and 40 C lt ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which does not a
3. 08 062 Puller for rotor removal 50 016 280 swing out rotor 47280 4 100rpm 3 082 x g 51007 Set of 4 buckes for RA 280 G 51 036 Set of 4 buckets with sealing cap 51 008 Set of 4 insert 1x280 ml G 51 009 Setof 4 insert 1x250 ml flat 51010 Setof 4 inserts 1x200 ml flat 51011 Set of 4 inserts 1x100 ml 51 012 Set of 4 inserts 2x50 ml conical 51013 Set of 4 inserts 4x25 ml 51014 Set of 4 inserts 2x15 ml conical 51015 Setof 4 inserts 12x15 ml 51016 Setof 4 inserts 12x5 7 ml 51017 Set of 4 inserts 19x5 ml RIA B 50021 MP 200 microtitre rotor 4 100 rpm 2 349 xg 50 023 RS 600 Angle rotor 6x100 ml 9 000 rpm 8 965 xg G 51018 Set of 6 adaptors 1x50 ml G 51019 Set of 6 adaptors 1x50 ml conical 51020 Set of 6 adaptors 1 30 38 ml 51021 Set of 6 adaptors 1x15 ml conical 50022 RS 100 Angle rotor 10x10 ml 12 000 rom 14 167xg 51022 Set of 6 adaptors 1x6 ml 51023 Set of 6 adaptors 1x1 5 2 ml 50014 RS 24 Angle rotor 24 x1 5 2 ml 14 000 rom 18 407xg A14005 Adaptor 500 800 ul A14006 Adaptor 200 ul PCR A14007 Adaptor 250 400 700 ul 2 3 GENERAL PRESENTATION Figure 6 9 NF I200 On off switch 2 lokpn 88 8080772 3 Gas spring 9 Electrical cables inlet and fuses 4 Gasket 10 Manual opening hole 5 Rotor 11
4. 02 Time Display e Display of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display e Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bovvl temperature during programming and the run 04 Program No Display e Display of the program number from 0 to 9 05 Running indicator leds e The 4 leds light up consecutively during the run to indicate the spinning of the rotor 06 START key e Starts the centrifugation run 10 07 STOP key Stops the run by starting the braking phase manually 08 LID key Opens the lid if the lid indicator lights up 09 LID Indicator Led e This led is on if the opening of the lid is possible 10 START Indicator Led e This led turns on as the start button is pushed and turns off as the Lid Indicator 9 turns on after the rotor has stopped spinning 11 LID OPEN Indicator Led e on if the lid remains open or is not properly closed 12 Imbalance Indicator Led 115 activated if imbalance occurs Brake is applied immediately 13 Acceleration Indicator Led turns on while the acceleration rate is being programmed and while the rotor is accelerating 14 Braking Indicator Led e This indicator turns on while the breaking rate is being programmed and while the rotor is breaking 15 Value Increase Decrease Keys e They increase or decrease values during programming stage a
5. NUVE SANAYI MALZEMELERI IMALAT VE TICARET A S NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL 4 MANUFACTURER NUVE SANAYI MALZEMELERI IMALAT VE TICARET A S Esenbo a yolu 22 km 06750 Akyurt ANKARA TEL FAX 90 312 399 28 30 pbx 90 312 399 21 97 E mail sales nuve com tr WARRANTY CERTIFICATE Nuve warrants that the equipment delivered is free from defects during material and workmanship This warranty is provided for a period of two years The warranty period begins from the delivery date Warranty does not apply to parts normally consumed during operation or general maintenance or any adjustments described in the operating instructions provided with the equipment Nuve does not accept any liability in the case where the goods are not used in accordance with their proper intent The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper handling or use the defects in maintenance negligence bad functioning of auxiliary equipment in the case of force majeure or accident and incorrect power supply In the event of failure NUve shall be under no liability for any injury or any loss or damage as the result of the failure other than the guarantee conditions e BEFORE OPERATING THE INSTRUMENT THIS MANUAL SHOULD BE READ CAREFULLY THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF THE INSTRUCTIONS AND PRECAU
6. e Place the buckets to the rotor 9 Connect the centrifuge to the power supply Figure 4 16 5 4 6 1 DRIVE SYSTEM The rotor is driven by a three phase asynchronous motor The microprocessor control system assures the correct drive speed The force applied to the rotor is directly related to the shape of the rotor the swing out rotor receives more load than the angle rotor does Longer radius and more accessories increase the load of the rotor and decrease the spinning speed The centrifuge does not allow the rotor to spin at a speed which it cannot resist mechanically Please make that sure the correct type of rotor is selected during programming SECTION 6 CLEANING AND PERIODIC MAINTENANCE PERIODIC MAINTENANCE Disconnect the power cable and after the maintenance check the presence of the mains grounding line Rotors should be washed after every use especially if a spillage has occurred in warm water containing a few drops of liquid soap A mild washing liquid is ideal as a cleaner Rotors and other accessories must be clean if any spillage or chemicals occur You may use a nylon brush to clean the buckets and tube inserts of the rotor Do not use metal brushes Dry the rotor with a piece of soft absorber cloth Please make sure that the buckets and inserts are well dried you may use hair dryer The buckets of the swing out rotors should be greased frequently with the oil provided with the centrifuge Please
7. manually make sure that the rotor already stopped spinning Upon opening the lid lift tt by hand and observe the rotor If the rotor is still spinning close the lid and wait approximately 10 minutes before repeating the operation This operation must be carried by someone who is informed of the danger and of the precautions which must be undertaken 14 SECTION 5 OPERATING PRINCIPLES 5 1 PREPARATION OF THE ROTOR TO RUN e Before installation check the rotor for corrosion and cleanliness Chemical corrosion mechanical corrosion may do severe damage to the rotor and the centrifuge Particles which are stuck inside the inserts cause the breakage of tubes and lead to major imbalance please check to make sure that no particles are left on the rotor e The central hole of the rotor and the motor shaft should also be clean and dry before all centrifugal operations and they must be kept in that vvay all the time 5 2 LOADING The most important condition of an efficient centrifugation is to balance the tubes properly Loading must be done by meeting the requisites of static and dynamic balance Static Balance This balance states that diametrically oppositely replaced vveights are almost the same application the liquid level in the tubes should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to
8. remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most of the imbalance problems are mostly raised by the users who do not clean and oil the pins Please do not leave the rotor on a metal surface particularly stainless steel as electrochemical reactions set off easily with the aluminum or magnesium in the rotor Make sure that no deposit remains at the bottom of the bucket because the pressure of a flask or tube from above during centrifugation will certainly increase the chance of corrosion STERILIZATION Apply alcohol for example 70 ethanol or isoprophanol for 10 minutes against bacteria and viruses The rotors and buckets may be autoclaved at 121 C and under 215 kPa pressure for 20 minutes but please do not forget to remove all accessories Do not use formaldehyde for the sterilization process Phenol is a corrosive substance and should never be used Glutaraldehyde is a toxic substance and increases the rate of fatty acid in the body CORROSION INFORMATION Nuve rotors which are made of aluminum are designed to spin at proper RCFs for many years When used properly their resistance to corrosion and their life soan increases and the imbalance problems decrease All accessories should be checked thoroughly and regularly as almost all laboratories already have the conditions which lead to corrosion easily 17 6 3 1 Chemical Corrosion T
9. Cooling 6 Display and panel 2 4 ROTOR SELECTION TABLE SVVING OUT ROTORS AND ACCESSORIES Ja siad otor buckets nominal a bees ea E e e __ Set of 4 inserts 1x200mi flat 2 50 861004 42 25 conical 56 0 4 5565 12xtbm 17 07772 2 a S Y Setof4inserts 19xX5m RIA 1 7727511254 2X3 MP oo Microtitre Plate microtitre 4 100 2 349 Rotor plates ANGLE ROTOR ACCESSORIES Max Tube Max Radius Max Speed Max Rotor Description Capacity mm pm 8059 RS600 Angle Rotor 68 100 38 99 9 00 8965 Setof6adaptors 1 500 39071 _ Setof 6 adaptors 1x50 ml conical 305 Set of 6 adaptors 1x30 088m 26 Set of 6 adaptors 1x15 ml conical 17 o Set of 10 1x6ml 13 adaptors Angle Rotor Adaptor 1x500 800ul 8 Adaptor 15200 65 RS30 Angle Rotor 3502020 11 9 13 500 Adaptor 1x500 800ul 8 Adaptor 15200 65 2 5 PRECAUTIONS AND USAGE LIMITATIONS Do not use the device for any purpose other than the usage purpose Prior to first use the user s manual should be read and the device is only to be used by authorized and trained pers
10. Switch off the centrifuge and switch on again after the motor has stopped If you still see the error code contact to the service PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICAL HELP IF AN ERROR OCCURS 19 SECTION 9 BRAKING RESISTOR ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200 5 SNIGNNOYS 405835 NORE T WOLA JA an 1 H 80193130 0000 8 0708 NIYA 0021 340 0 ep HOLIMS 0 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R 9 2 84 14 3NI1 SNIGNNOYS 9NT002 L 1 5 l Nigog 05 55 ie 1 L 6 4010930304 aS 3ONvW8NI BE d 1111 2 3 God 40021 HOLIMS 440 NO 1 SECTION 10 WARNING LABEL 22 GROUNDED PLUG FixttA 250V NF 1200R FUSES 2x10A F2x4A 250V NF 1200 FUSES 2x4A 23
11. TIONS DESCRIBED IN THIS MANUAL e INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NUVE IT MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT PERMISSION Dear Nuve User We would like to take this opportunity to thank you for preferring this Nuve product Please read the operating instructions carefully and keep them handy for future reference Please detain the packing material until you see that the unit is in good condition and it is operating properly If an external or internal damage is observed contact the transportation company immediately and report the damage According to ICC regulations this responsibility belongs to the customer VVhile you are operating the instrument please obey all the warning labels do not remove the warning labels not operate damaged instrument do not operate the instrument with a damaged cable not move the instrument during operation In case of a problem contact your Nuve agent for an authorized service or maintenance The validity of the guarantee is subject to compliance with the instructions and precautions described in this manual Nuve reserves the right to improve or change the design of its products without any obligation to modify previously manufactured products Information contained in this document is the property of Nuve lt may not be duplicated distributed without its permission PLEASE REGISTER ONLINE TO VALIDATE WARRANTY To register you
12. The corrosive material is pushed against the aluminum alloy by the centrifugation g force during the centrifugation This situation causes the stress corrosion to occur more quickly than the chemical corrosion does Microscopic cracks occur under the force of the centrifugation Every centrifugation causes the aluminum rotor to be attacked by the chemical more and more and eventually micro cracks decrease the resistance of the rotor against the centrifugation force Fortunately no crash occurs just after the first micro cracks have appeared as the rotors are manufactured according to the high safety limits The corrosion of the small amount of corrosive materials does not result in severe cracks but weakens the mechanical resistance of the rotor in the course of time The rotor s places buckets buckets edges and the base of the rotor should be checked regularly If needed they must not be used until a specialist has checked them thoroughly 6 5 CLEANING e Disconnect the centrifuge before cleaning There is no need of daily cleaning unless a tube breakage occurs or any liquid spills 6 6 ELECTRICITY Centrifuged at high voltages are present behind the panels These panels are electrically disconnecting Do not open the centrifuge 18 SECTION 7 DISPOSAL MANAGEMENT CONCEPT The currently valid local regulations governing disposal must be observed t is in the responsibility of the user to arrange proper disposal of the i
13. e function key 16 and see that the value on the speed display 1 flashes Enter the run speed or RCF by pushing value increase and value decrease keys 15 Push the function key 16 and see that the value on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the value on the temperature display 3 flashes Choose the acceleration value among 0 1 2 3 4 5 6 7 8 or 9 by pushing the value increase and decrease keys 15 0 slovvest 10 fastest rates Push the function key 16 and see that the break led turns on and the value on the acc break temperature display 3 flashes Choose the break value among 0 1 2 3 4 5 6 7 8 or 9 by pushing value increase and decrease keys 0 slovvest 9 fastest rates Push the function button 16 and see that the cooling indicator led 17 turns on and the value on the acc break temperature display 3 flashes Enter the bowl temperature value This step is not valid for NF 800 model Push the function key 16 the entered values will be saved You may make 10 0 9 different programs 12 4 4 RUNNING THE PROGRAM Switch on the centrifuge see that the on off switch led turns on 9 See that the control panel activates e Load your samples to the rotor and close the lid You can run the centrifuge in two different ways e You can make a new pro
14. gram see section 5 3 e You can select a program from the memory PRE COOLING While operating the NF 1200 R centrifuge at sub ambient temperatures in order to get the maximum performance without loading the rotor run the centrifuge at 1 500 rom at a temperature which is 3 C lower than the required temperature and at hold position until you see that the temperature inside the bowl drops to the required temperature Then stop the centrifuge load your samples and start centrifugation 4 4 1 SELECTING THE MEMORIZED PROGRAM Select the requested program number at stand by position by pushing value increase and decrease keys All values related to the selected program will be shown on the relevant displays Check that all values are correct Load the rotor with samples by paying attention to the dynamic and static balances Close the lid see that the lid open warning led 11 turns off Push START key 6 and see that the start led 10 turns on The rotor starts accelerating according to the set value and the acceleration led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the rotor accelerates according to the set acceleration value up to 2 000 rpm and then continues accelerating with the highest acceleration value 9 to reach the set speed As the speed reaches the set value sign appears on t
15. he acc brk temp display 3 The NF 800 R shows the bowl temperature on the acc brk temp display until the centrifugation ends e The elapsed time is counted down as the centrifugation starts e When the time display shows 00 the program ends the breaking led turns on and the set break rate is shovvn at acc brk temp display 3 The rotor brakes vvith the highest break rate 9 until it slovvs down to 2 000 rpm and continues braking according to the set braking value elapsed time until the rotor stops is counted and is shown at the time display This is the time that passes until the rotor stops e When the speed display shows 70000 the start led 10 turns off the end expression appears and the lid led 9 turns on The user is warned by an intermittently sounding alarm e Push the LID button 8 to open the lid You may leave the centrifuge at stand by position NOTE 1 You may display the RPM or RCF value on the speed display during the run by pushing the function button NOTE 2 You may see all set values during the run by pushing the value increase key 13 NOTE 3 At hold position the rotor keeps spinning until the STOP key 7 is pushed After the STOP has been pushed the rotor brakes according to the set brake rate NOTE 4 The centrifuge vvaits at the stand by position with the program values of the last program NOTE 5 Please open the lid after every operation otherwise the centrifuge cannot be r
16. his type of corrosion is caused by chemical reactions The electrolide liguid on the surface of the material is the main cause of the chemical reaction If that electrolide liguid is allowed to stay at the surface corrosion occurs First discoloration appears and then the metal pittens Aluminum easily reacts with the ionic solutions The other causes of corrosion are as follovvs Chemical vapors in the laboratory environment which dissolve in the water on the rotor in refrigerated centrifuges Corrosive liguids which overflow from overfilled and unsealed tubes the liguids which spread out during centrifugation Contaminated and non cleaned buckets tubes and bottles NOTE If the centrifuged samples are corrosive only rinsing with water is not sufficient The residuals dissolve in the water and humidity on the rotor and in the buckets NOTE Some particles may stick to the tubes buckets and adapters These particles crash and do harm to anodized surface during centrifugation and ease the occurrence of corrosion 6 4 Stress Corrosion This type of corrosion is caused by the force of the centrifugation of the corrosive chemical which is already in contact with the alloy As the aluminum alloy contacts with the corrosive chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the course of time
17. llow the rotor to spin if the lid is not closed The lid locking system which does not allow the centrifuge to operate if the lid is open which also does not allow the lid to be opened while the rotor is spinning and the possibility to open the lid by pushing only one key when the program ends provide safe and easy working conditions There are audible and visible alarms to inform the operator when the lid is open when the program ends and when any error conditions occur In case of a power failure the lid could be opened manually by using a manual lid opening tool The NF series centrifuges are manufactured according to the following standards EN 61010 1 EN 61010 2 020 EN 61000 6 3 SECTION 2 TECHNICAL SPECIFICATIONS 2 1 TECHNICAL SPECIFICATIONS TABLE 1 00 0 Speed RA 280 rotor 4100 rpm MP 200 rotor 4100 rpm RS 600 rotor 9000 RS 100 rotor 12000 rpm RS24 rotor 14000 rpm RS 30 rotor 13500 rpm RCF 280 rotor 3082 x g MP 200 rotor 2349 RS 600 rotor 8965 x g RS 100 rotor 14167 RS24 rotor 18405 x g RS 30 rotor 20170 x g Max Capacity RA 280 rotor 4x280 ml MP 200 rotor 2x3 microtitre plates RS 600 rotor 6x100 ml RS 100 rotor 10x10 ml RS 24 rotor 24x1 5 2 ml RS 30 rotor 30x2 m 1 No of Memories 1 8 10 Temp range 1 920 4090 2 Refrigerant Liquid RtfMa 3 Temp Accuracy 110 0 2 2 ACCESSORIES FOR 1200 1200
18. nd to select the program number during stand by position By pushing value increase key it is possible to see the set values during centrifugation 16 Function key t is pushed to see the RPM or RCF value on the speed display during the run to pass among the displays and to store the programmed values 17 Cooling Indicator Led lt indicates that the cooling system functions During programming it indicates that the temperature value is being set on Acc Br Temp Display 3 It is not valid for NF 1200 model 4 3 MAKING A NEW PROGRAM Determine the rotor type the speed and time the acceleration and braking rates e Push the function key 16 once and see that the value on the program no display 4 flashes Enter the program number by pushing the value increase and value decrease keys 15 e Push the function key 16 again and see that the value on the speed display 1 flashes Select the proper rotor tyoe by pushing value increase and value decrease keys 15 11 ROTOR TYPE DISPLAY MAXIMUM MAXIMUM SPEED SPEED RPM RCF RA 280 Swing Out MP 200 Swing Out RS 600 Angle RS 24 Angle Angle Rotor Swing out Rotor NOTE correct rotor type should be selected in order to see the correct RCF values during the run Push the function key 16 and see that the RPM or RCF option on the speed display 1 flashes Choose the RPM or RCF option by pushing value increase and value decrease keys 15 Push th
19. ndividual components Applicable local regulations for disposal have to be carefully observed The instruments and electronic accessories without batteries power packs etc must be disposed off according to the regulations for the disposal of electronic components Batteries power packs and similar power source have to be dismounted from electric electronic parts and disposed off in accordance with applicable local regulations SECTION 8 TROUBLESHOOTING If the centrifuge fails to operate Check that e The on off switch is on fuses are sound e The plug is not defective e The centrifuge is well connected to the supply e The electricity installation is not defective e Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for at least 20 minutes Err 3 Speed failure The optical sensor could have failed Please observe whether the running indicator leds are on or not If it is not on the optical sensor is defective If it is on the motor could have failed Contact to your service if either occurs e This error code occurs when the mains voltage drops below the lower limit Err 4 e Lid failure This failure occurs when lid is opened during the centrifugation Err 5 e The temperature sensor endings are broken This error code appears only on NF 1200R Err 6 Motor failure The motor driver unit inverter is defective
20. onnel Only authorized technical personnel handle the product in case of any failure e The working bench should be durable to the device weight and vibration isolated e Ensure that the rotor is placed correctly prior to usage e According to the standard IEC 61010 2 020 anyone and any hazardous materials should not be in the 300 mm safety zone while centrifuge is running e Do not move the device while it is running e Do not open the lid while rotor is spinning e Apply the manual lid opening procedure in the case of power cut or in the case of any error e Use only the the spare parts rotors and accessories which are supplied by NUVE Load the rotor according to the explanations in the user s manual e Start the device after ensuring the rotor is loaded correctly 9 Do not use the centrifuge in areas which are in explosive danger Do not centrifuge the explosive flammable radioactive corrosive materials and the materials which may react with each other e The centrifuge and the rotor are not microbiologically leak proof Use tubes with leak proof covers if hazardous toxic and pathogenic microorganisms are centrifuged Do not use corrosive materials which may be harmful for the device integrity rotor and accessories Do not use rotors and accessories with corrosion and mechanical damages e Mains supply should be appropriate to power of the device and grounded e Use tubes whose sizes are suitable to the rotor and accesso
21. r warranty online please visit our web page www nuve com tr and fill in WARRANTY REGISTRATION FORM TABLE OF CONTENTS SECTION 6 05 tar oor ne men EMT 4 INTRODUC TION eb 4 UN EET 4 SECTION AA mo AAA 5 TECHNICAL SPECIFICATIONS AAA 5 2 TECHNICAL SPECIFICATIONS 4 a 5 ACCESSORIES FOE NE 20012 n 5 2 9 GENERALE PRESENTATION wa sea nb bl 6 24 BOTIOHSELEGTION TABLE 7 25 PRECAUTIONS AND USAGE LIMITATIONS N 8 208 11611 5 8 isindi 9 SYMBOLS 252 2001 Dv Ned UTE 9 SEC TION M 9 OPERATING UL deett Wa a itc tal ab 9 41 AR 9 4 2 DISPLAYS AND CONTROLAPAMNE sola b aa 10 4 2 1 Introduction to Control Display Panel Components and Functions 10 AS MAKINGANEW PROGRAM 11 AA MUNNING TRE PROGRAM mp 13 441 SELECTING THE MEMORIZED PROGRAM 13 45
22. ries e Tubes which are used in the centrifuge should not be deformed by the effect of the centrifuge force e Use glass tubes to balance if glass tubes are used Use plastics tubes to balance if plastics tubes are used e Do not start the device unless tubes are in balance imbalance loading may cause mixing the samples broken tubes and damages on the rotor and motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED NUVE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol refers to safety relevant warnings and indicates possibly dangerous situations The non adherence to these warnings can lead to material damage and injury to personal cge oymbol in the operating instructions This symbol refers to important circumstances SECTION 4 OPERATING UNIT 4 1 OPERATING Open the lid and check that there is no sample in the instrument Turn the centrifuges on by using On Off switch oee that the display and control panel activates Learn the functions of the control panel See Part 4 2 1 Set the values and start the operation See Part 4 3 4 2 DISPLAYS AND CONTROL PANEL 11 12 16 14 15 3 9 Sekil 3 4 2 1 Introduction to Control Display Panel Components and Functions 01 Speed Display e Display of speed 500 14 000 rom or display of RCF 0 18 405 g during the run and rotor type during programming
23. spinning axis of the rotor e Although in hospitals the samples having almost the same densities are centrifuged in industry samples having different densities may be centrifuged In this case the dynamic balance becomes more important factor than the static balance is e f the number of tubes to be centrifuged is less than the capacity of the rotor the tubes must be placed oppositely If an odd number of tubes is centrifuged a water filled tube at the same weight should be used for balancing Imbalance of the rotor may cause major damage to the rotor and centrifuge Never attempt introduce liquids into the tube inserts Balance the rotor with glass tubes if you use glass tubes for centrifugation Balance the rotor with plastic tubes if you use plastic tubes for centrifugation Always use tubes which can withstand to the set speeds Examples of the proper and improper loading are shown below 15 RIGHT LOADING WRONG LOADING Figure 3 NOTE Distribution of the load is more important than the load itself 5 3 ROTOR INSTALLATION Put some light oil on the drive head to prevent sticking Be careful that the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socket wrench while screwing e For swing out rotors lubricate the pins where the buckets are hanged
24. un again 4 5 SAFETY INTERLOCK SYSTEM The safety interlock system prevents opening of the lid when the rotor is spinning The centrifuge does not operate until the lid is closed and the lid remains locked until the rotor stops spinning The Lid Open indicator turns on to warn the user if the lid is not closed properly NOTE f power failure occurs access to the samples is possible by opening the lid with a special tool Please see the manual lid opening section 3 6 for further information 4 6 IMBALANCE DETECTION SYSTEM e The imbalance detection system of the NF 1200 NF 1200 R centrifuges operates electronically when an unacceptable imbalance occurs In this case the brake is applied immediately e The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid cannot be opened until the rotor stops spinning The centrifuge can only be started after the lid has been opened and the rotor has been re loaded correctly e To avoid facing imbalance problem please make sure to insert the tubes correctly 4 7 MANUAL LID OPENING In case of power cut or any breakdown the centrifuge be opened manually to access the samples To open the lid manually 9 Switch off the instrument e Insert the manual lid opening tool into the hole on the left side of the instrument see the general view drawing e Push the tool keeping it horizontal until the lid opens WARNING Before opening the lid
Download Pdf Manuals
Related Search
Related Contents
2100 [U2.05.01] Notice d`utilisation des opérateurs de mécanique de la Model 008 - Daisy Outdoor Products Lewmar Pro-Sport/Series/Fish Windlass Buffalo AirStation N600 BOTTIN DE LA VIE SOCIALE Smile on Line - N°19 - Mai 2009 Sony CMT-CP2W/CP2WA User's Manual Copyright © All rights reserved.
Failed to retrieve file