1. ECT THE MEASURED VALUES 2 0mm om 2 54mm 0 1 ME VE a ME VE While there is no battery connected to either of the two pin terminals on the right hand side of the IRM the details screen will display No battery Note ONLY USE ONE CELL CONNECTOR AT A TIME DO NOT ATTEMPT TO USE BOTH THE 2 0 AND 2 54MM CONNECTORS SIMULTANEOUSLY On connecting a cell the IRM will begin to perform the three 3 load tests and determine the internal resistance updating the details display for each After all three 3 tests are complete you can commence to set the cell capacity It will take approximately 10 seconds to finish each round of testing a lot of cells may require two or three passes to stabilize The IRM will continue to test the cell indefinitely while in the details display mode IRM Details view 10 Load voltage Open cell voltage 0 50 Load voltage L NN 1kHz 0 50 load resistance 0 50 load resistance 10 load resistance 4 4 Capacity setting After the test values have started to stabilize pressing the left button increases the cell capacity in 50mAh lots Holding down the button will make the IRM auto increment the value until the button is released 4 5 Summary evaluation IRM Human summary Internal resistance Battery capacity Power efficiency Cell voltage drop Equivilent C rating Output current Usable power output Once the cell capacity has been set the IRM will provide a series of fiv
2. PL Daniels IRM User manual vOO5R2 http nqrc com May 15 2009 Contents 1 Introduction 3 2 Introduction to cell internal resistance 3 2 1 Why does it matter 3 Mi bat ee ee ka alone A ne 3 2 2 Testing method ra st UN Di at bass a 3 2 3 Factors effecting resistance readings 4 2 3 1 Contact corrosion 4 gl au LR EC SES AM 4 2 3 2 Cell temperature ais ates ee e a 4 2 3 3 Celleharge AA 4 234r Qeli age St oy O E Ae ed 4 2 3 5 Cell capacitance and impedence 4 2 4 Choosing appropriate tests 5 3 Testing methodology 5 Ss TENES So ada A ee yh ee tae oh Ad PGE a te atte 5 SES e A A EN ER dr ess 5 A ra a ae 5 3 13 0 5 ohm 1kHz test rat dd a Ba RY 5 4 IRM operation 6 4 1 Precatitionss e iris Sue ee ab A cas ben x 6 A ROWE ON taa La E o Ge UA Ar vs 6 4 3 Testing Details display v2 avaig aras Eur 7 4 4 Capacity setting acota deb e A na lA Ai a Len 8 4 5 Summary evaluation 22 2 2 dada AE G E a eH 8 4 6 Power off amp fet S So amp Eg Sm a de 28 SS eo ed 9 At Hints Tips ad FAQ e Eva ee Late eg ee 9 4 8 Errors Faults tora o oo oe PSY ed oP Rae Da 10 5 Specifications 11 5 1 Electronic operational limits 4 5 4 ccd us see oS ee ee GA 11 5 2 Physical specifications as a eat a AAA ee 11 6 IRM Servicing 12 6 1 EII 12 7 Sample readings 12 1 Introduction For a long time lipo pack makers have been content to delive
3. e 5 voltage drop summaries for the cell that has been tested The IRM displays the available current power efficiency and equivalent C rating for voltage drops between 0 1 and 0 5V For lipoly cells the 0 3V drop setting should be considered as a good benchmark for the cell translating to a 0 9V drop on a 3S pack Notice that the efficiency value drops off as the voltage drop increases this is because more power is being burned up as heat within the cell A theoretical perfect cell would have a zero voltage drop at all loads unfortunately such a thing doesn t exist Press the right button to step to the next voltage drop After the 0 5V drop summary the IRM will return to the details display and commence testing the cell again If you wish to test another cell with the same mAh capacity change the cell over before pressing the button again this way the mAh setting in the IRM will be preserved 4 6 Power off When all the testing is complete remove the cell under test from the 2 pin header then remove the power from the 3 pin header 4 7 Hints Tips and FAQ Hints and tips e Measure your cells at a known room temperature e Measure your cells at a consistent voltage charge e Always clean the IRM cell test contacts between tests e Allow the IRM about 10730 seconds to get stable results from the cell FAQ 1 Why use 1 0 and 0 5 ohm loads why not X y ohms a There are several factors involved in why the IRM uses 1 0 and 0 5 oh
4. load is recorded eg 4 00V Knowing the difference of the voltages and the load resistance the cell internal resistance can hence be derived 2 3 Factors effecting resistance readings Unfortunately due to the nature of batteries cells exhibit characteristics of re sistance capacitance and inductance of which all contribute to the measured internal resistance Internal resistance can vary considerably even within the same cell this makes the task of determining the internal resistance a lot harder however by keeping other variable factors constant one can produce results that allow comparisons 2 3 1 Contact corrosion While not actually part of the cell itself having dirty contacts to the cells can make a noticable difference in the resistance Wiping clean the contacts can improve contact resistance by 10 or more milliohms 2 3 2 Cell temperature For many cell chemistries the resistance of the pack decreases initially from room temperature to a certain level perhaps 45 C after which the resistance of the pack will increase again Lithium cells in particular suffer poor cell resistance near 0 C 2 3 3 Cell charge Generally the closer to a full charge a cell is the lower the internal resistance 2 3 4 Cell age As cells become older their internal resistance tends to increase 2 3 5 Cell capacitance and impedence Under constant current DC loads the cell capacitance and impedence effects tend to be negligable however with o
5. m loads i Availability of components with sufficient precision reasonable costs and ample power dissipation ii Simplicity of the maths involved in deriving the internal resistance iii 0 5 ohm load is simply two 1 0 ohm loads in parallel iv Sufficiently large load 4A and 8A without being too weak or too severe 2 The IRM results don t match the results obtained from my charger why a A multitude of possible reasons for this i Chargers are putting power into a pack not depleting ii Chargers use a different current setting varies for the duration of the charge too iii Chargers may not always account for resistance the leads being used iv Chargers are likely using an averaging algorithm for the resistance measured during the entire charge cycle 3 The 1 0 ohm test produces an internal resistance that is higher than the 0 5 ohm test a Due to the way the voltage sensing occurs in the electronics of the IRM the 0 5 ohm test actually produces a more accurate reading with twice the precision of the 1 0 ohm test subsequently the 1 0 ohm test may exceed the 0 5 ohm test by a couple of milliohms in some situations Often after a 30 second test period both the 1 0 and 0 5 ohm test results will converge to read very similar 4 8 Errors Faults 1 No display Check that the supply voltage to the IRM is between 7 5V and 15V check also that the center pin is positive and one of the outer pins is ground 2 Battery no
6. po This test typically will produce the lowest internal resistance reading 4 IRM operation 4 1 Precautions 1 Do not leave a cell connected to the IRM for more than one or two minutes 2 Be aware that although the IRM only takes small samples from the battery pack the voltage will be reduced Make sure you check your pack balance 3 Avoid plugging the cell in backwards reverse polarity The IRM can cope with a reversed connection for a few seconds but it s still better not to do it in the first place 4 Do not connect more than 5V to the cell inputs again the IRM can cope for a few seconds but it s best not to do it 4 2 Power on Connect either a 2S lipo or voltage source between 8 0V and 15V 12V is the preferred maximum but the IRM can cope with up to 15V to the IRM via the 3 pin header on the left hand center side with the pin closest to the top of the screen being positive and the bottom pin being negative ground The center pin is disconnected and is used purely as a placement holder The IRM is reverse polarity protected so if you connect the battery the wrong way around there will be no damage The IRM will display a short message on the screen namely giving the current firmware version before commencing to the Testing screen VE 8 15V Ground 0V 4 3 Testing Details display Note ALWAYS CLEAN THE TEST CONTACTS BEFORE TESTING A NEW CELL EVEN THE SLIGHTEST AMOUNT OF GRIME WILL CONSIDER ABLY EFF
7. r to us a single rating for performance of their packs the C rating Unfortunately for us this C rating is near worthless due to each maker choosing their own basis on which to define what C is The PLDaniels IRM has been created to give us the ability to make an informed deteminiation of what the true cell and pack ratings by providing a far more useful measure the cell internal resistance Let it be stated that there is no perfect test for cell internal resistance The IRM has been built to allow cells to be tested under a known set of loads which will facilitate comparisons DO NOT short circuit ANY of the pins Use these connectors Use this connector to test your cell to power the IRM 005R2 2541011 PLD IRM v1 1 GND pee 2 00 Sue DO NOT power the IRM with the same battery that you are testing 2 Introduction to cell internal resistance 2 1 Why does it matter Once we have an idea of the internal resistance of a cell we can then use that to determine how efficient our power setup is as well as knowing how much power is being turned to heat in the battery pack and additionally know how much the voltage can potentially drop under load 2 2 Testing method To determine the internal resistance of a cell the cell is connected to a purely resistive load of known value eg 1 0 ohms The voltage of the cell before the load is attached is recorded eg 4 20V then the voltage of the cell when attached to the
8. scillating loads eg 1kHz ESC switching the cell can appear to have different resistances 2 4 Choosing appropriate tests Is there a perfect test for the cell internal resistance No The IRM was not built to persue the holy grail of testing The IRM has been built to provide a convenient way of applying three 3 common tests that will allow you to compare cells and give insight into cell performance 3 Testing methodology In order for a measurement to be meaningful the testing methodology needs to be clearly stated to allow others to perform the same tests Each person may have a preferred testing method that they feel more correctly reflects real world usage however this will not be discussed here rather this section is about detailing the tests already chosen The IRM has been preprogrammed to perform three 3 different tests Each test is performed after a two 2 second interval 3 1 IRM tests 3 1 1 1 ohm test Load 1 0 ohm 0 01 ohms 0 99 to 1 01 Duration Less than 1 second Typical current 4A for a fully charged lipo 3 1 2 0 5 ohm test Load 0 5 ohms 0 01 ohms 0 49 to 0 51 Duration Less than 1 second Typical current 8A for a fully charged lipo This test should produce the highest internal resistance reading towards the end of the test 3 1 3 0 5 ohm 1kHz test Load 0 5 ohms 0 01 ohms 0 49 to 0 51 Duration 1 0 seconds 0 1 seconds Duty 50 Typical current 4A average for a fully charged li
9. t detected Check that the polarity of the cell is the correct way around The positive pin should be towards the top of the IRM 3 Open cell voltage doesn t match voltage measured with multimeter This can happen if the voltage is below 7 5V or above 15V to the IRM Also it s quite possible the multimeter is not correctly calibrated 10 4 Milliohms readings fluctuate dramatically Contacts for the cell connection are likely dirty or frayed Clean with a 3M microcloth or similar 5 Smoke escaping from IRM Disconnect everything from the IRM visually inspect the back of the board to see where the smoke came from If the IRM is no longer operational contact PLDaniels NQRC Possible causes are excessive over voltage on the input 20V short cir cuit on the board or outright bad luck 5 Specifications 5 1 Electronic operational limits Supply voltage 8 15V 8 12V for optimal accuracy Cell compatibility Lithium polymer Lithium lon LiFeO4 NiMH NiCd Cell test voltage minimum 1 0V Cell test voltage maximum 5 0V Cell resistance minimum 4 milliohms Cell resistance maximum 999 milliohms Cell resistance precision 2milliohms Minimum temperature 0 C 32 F Maximum temperature 40 C 104 F 5 2 Physical specifications Size 85 x 40 x 40mm Weight 150g Construction Single 85 x 40 PCB with modular LCD panel attachment shrink wrapped Cell test input connectors 2 pin 2 54mm 0 1 and 2 pin 2 00mm Supply inp
10. ut connector 3 pin 2 54mm 0 1 pitch connector center pin positive two outer pins ground negative 11 6 IRM Servicing 6 1 Reflashing The IRM can be flashed with a new firmware via the 6 pin ISP head located in the middle of the board adjacent to the microcontroller Atmel AVR mega48 7 Sample readings 12

Download Pdf Manuals

image

Related Search

Related Contents

macchina carica bomboline interne con * bilancia elettronica user  nstrucciones top 420 n eurotop 420 n eurotop 421 n  ToxiRAEI- Détecteur NH3 O2 - Détection Gaz - Gamme K  Samsung C3110 User's Manual  Konformitätserklärung des Herstellers  Manual del usuario  2. Où trouve-t-on le module Médiathèque  