AutovaxID Operations Manual
Contents
1. Authorized User A person who has the proper permissions to use the AutovaxID Permissions are defined by the user classification administrator vs non administrator and by use of strict mode If user names and passwords are not in use enabled then any walk up user has the permissions of an administrator Bioreactor BRX Unit containing thousands of hollow fibers where media exchange occurs and the cell culture grows BRX Abbreviation for Bioreactor BSC Acronym for Biological Safety Cabinet Circulation Pump Centrifugal pump which circulates media through the IC of the flowpath Conditioned Media Used to describe the media circulating within the IC and EC of the cultureware The media is a homogenous mixture of nutrients gasses waste products and protein Cultureware The disposable component of the system which includes the Bioreactor Gas Exchange Cartridge DH probe temperature probe EC and IC chambers the peristaltic pump head the factor and harvest bags and all tubing see Flowpath Cycling The movement of media across the bioreactor fibers transmembrane between the ICS and the ECS One complete cycle consists of an EC Rise and an EC Fall ETO Ethylene Oxide Expansion Chamber The fluid reservoir in the Expansion Circuit Extracapillary Space The volume in the bioreactor outside of the hollow fibers where cells are ECS grown and product is retained until harvested Part of the EC Extracapillar
2. AutovaxID Operations Manual 4 2 Biovest International Inc Introduction to the User Interface 4 2 User Interface Features The AutovaxID user interface uses several types of elements to assist the operator with identifying on screen information and performing on screen functions 4 2 1 Title Bar A title bar appears at the top of nearly all screens The title bar typically shows the name of the screen to assist with software navigation the current date and the current time in 24 hour format HH MM SS On the Monitor screen the screen title is replaced by the Instrument Name if it has been defined in the Instrument Setup 4 2 2 Information Panels Many screens include information panels that display process control parameters to indicate current status For example the area of the Monitor screen that displays the cycling information Cycle state EC Pressure and IC Pressure is an information panel Information panels may contain headings data Fields and or Buttons see below Some elements that appear to be backgrounds such as the gray background of the cycling panel in Figure 4 1 may function as a button which when touched will cause another screen to appear In the case of the cycling panel touching anywhere in this region will bring up the Cycle Detail screen see Figure 5 8 4 2 3 Buttons Buttons are special touch sensitive areas of the screen that provide access to unique software functions They a
3. 8 Flushing IC 0 07 29 skip Fill Flush in progress Cycle ICPR Media Rate nl hr 400 C02 Sensor 20 00 y Circulation Rate ml min 500 Media Tenperature Control 37 0 Qutflow Rate nl hr 480 Refrigerator Temperature Control 4 0 Hold Exit Help Figure 7 14 Fill Flush Step 8 AutovaxID Operations Manual 7 26 Biovest International Inc Operating Procedures Jan 12 200 Fill Flush 16 33 17 9 IC Flush Cycle Low 0 00 19 skip Fill Flush in progress Cycle Lou Media Rate ml hr 400 C02 Sensor 120 00 y Circulation Rate nl min 500 Media Tenperature Control 37 0 Qutflow Rate nl hr 480 Refrigerator Temperature Control 4 0 Hold Exit Help Figure 7 15 Fill Flush Step 9 dan 12 2007 Pi11 Flush 16 34 15 10 Flushing IC and EC 0 00 59 skip Fill Flush in progress Cycle Fall Media Rate nl hr 400 C02 Sensor 20 00 y Circulation Rate ml min 500 Media Temperature Control 37 0 Outflow Rate nl hr 480 Refrigerator Temperature Control 4 0 Hold Exit Help Figure 7 16 Fill Flush Step 10 AutovaxID Operations Manual 7 27 Biovest International Inc Operating Procedures dan 12 200 Fill Flush 16 34 48 11 Fill Flush Reduced Rates 0 14 21 skip Fill Flush in progress Cycle Fall Media Rate ml hr 100 C02 Sensor 20 00 y Circulation Rate ml min
4. Jan 23 2007 Circulation Controller 17 37 50 Controller Type Manual Controller Setup Hold Exit Help Figure 5 34 Circulation Controller Screen 5 5 13 Circulation Pump Schedule Setup If the circulation controller type is set to Schedule then pressing the Controller Setup button on the Circulation Controller screen will bring up the Circulation Pump Schedule screen see Figure 5 35 This screen allows an authorized user to schedule up to ten elapsed times with variable pump rate values In strict mode and when user names and passwords are in use only administrators may create schedules Normal users may view the schedule but they may not change the values in the Elapsed Time or the Circulation fields nor can they add or delete events Normal users may advance or delay the schedule using the Advance and Back buttons or by entering a manual time shift To add an event press the Add Event button at the bottom of the screen enter the desired Elapsed Time for the start of the event using the on screen keypad and then press Enter Events are in day hour minute format The new event will appear in the schedule all events are automatically sorted chronologically Press the Cir field for the new event and enter the desired rate for the circulation pump in mL min using the on screen keypad The circulation pump has a range of 0 200 to 500 mL min in 1mL min increments To delete an event press the Delete button to th
5. The media or factor bag is empty The calculated volume pumped has reached the alarm point Exchange the empty bag with One containing the correct fluid following the procedure for changing containers see Section 7 17 The container capacity is not set correctly Adjust the container capacity from the pump Controller screen The instrument registered a Startup possible clock failure alarm The battery and or the single board computer have failed Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The instrument registered a Startup down time alarm The instrument was running and was turned off and then back on Train users not to turn the instrument off during a cell culture run A power fluctuation caused the instrument to reboot Use an uninterruptible power supply to prevent power fluctuations The instrument registered an Instrument data recovery failure alarm The battery and or the single board computer have failed Contact Biovest Account Services The instrument logged that it restored from secondary state file The battery and or the single board computer have failed Contact Biovest Account Services The Outflow Media Factor and or Harvest pump has stalled The pump drive failed to sync with the moto
6. e Manual The media pump rate can be entered manually at any time by an authorized user Each manual pump rate change is automatically logged e Schedule The media pump rate is scheduled to change at user defined times to user defined rates All changes to the schedule are automatically logged all scheduled pump rate changes are automatically logged See Section 5 5 3 for details on setting up a schedule e Lactate Sets the media pump to be controlled by the lactate setpoint See Section 5 5 4 for details An authorized user may change the media controller type by touching the Controller Type field and then selecting the desired media controller type from the on screen list Some controllers have additional setup parameters that can be accessed by pressing the Controller Setup button Authorized users may change the controller type during a cell culture run Jan 24 2007 Media Control ler 17 44 05 J J J J Controller Type Manual Container Underflow Alarm Controller Setup Used Z At 10 7 Container Volume 20 00 L a D Figure 5 24 Media Controller Screen AutovaxID Operations Manual 5 28 Biovest International Inc Primary User Interface Screens 5 5 3 Media Pump Schedule Setup If the media controller type is set to Schedule then pressing the Controller Setup button on the Media Controller screen will bring up the Media Pump Schedule screen see Figure 5 25 This screen allows an authorize
7. 5 5 8 Factor Pump Time Cycle Controller Setup If the factor controller type is set to Time Cycle then pressing the Controller Setup button on the Factor Controller screen will bring up the Time Cycle screen see Figure 5 30 This screen allows an authorized user to setup a time based controller for the factor pump The time cycle controller sets the pump to act ina repeating cycle defined by the elapsed time Note The Time Cycle pump controller is NOT related to EC Cycling and does not affect the EC Rise or EC Fall times AutovaxID Operations Manual 5 35 Biovest International Inc Primary User Interface Screens Jan 16 2007 Tine Cycle 14 05 17 Factor Pump Tine Cycle Control ler Auto Rate 0 Delay Until 0 00 00 Total Cycle Time 6 00 00 On Tine 6 00 00 Hold Exit Help Figure 5 30 Factor Pump Time Cycle Controller Screen Auto Rate The desired pump rate to be used when the factor pump is on Delay Until The elapsed time before the time cycle controller will be activated Total Cycle Time The total duration of the cycle which is equal to the on time plus the off time On Time The duration that the pump is on and running at the desired rate during the cycle NOTE The On Time must be less than or equal to Total Cycle Time For example if the following entries are made Auto Rate 100 mL hr Delay Until 3 00 00 days hr min Total Cycle Time 0 02 00 days hr min On Time
8. Date and Time Screen The Date and Time screen is accessed by pressing the Next button on the Network Configuration Screen Monitor gt Main Menu gt Setup gt Configure gt Next gt Next The Date and Time screen see Figure 6 5 allows an authorized user to change the system date and time After a Time or Date field is modified an Apply button will appear the Apply button must be pressed to save the changes NOTE The AutovaxID does NOT automatically adjust its clock for daylight savings time Jun 20 2006 Date and Tine 15 33 58 Tine 14 56 f8 Date Jan 29 2004 Apply Hold Exit Back Next Help Figure 6 5 Date and Time Screen AutovaxID Operations Manual 6 7 Biovest International Inc Software Setup amp Configuration 6 4 6 3 1 6 3 2 Time This is the time of day and is displayed on the title bar of all screens The time is also used to timestamp all log entries The time must be entered in 24 hour format where noon is 12 00 00 Date This is the date that is displayed on the title bar of all screens This date is also used to timestamp all log entries Run Configuration 1 Screen The Run Configuration 1 screen is accessed by pressing the Next button on the Date and Time screen Monitor gt Main Menu gt Setup gt Configure gt Next gt Next gt Next The Run Configuration 1 screen see Figure 6 6 provides the option to select and enable a run settings profile
9. The instrument is experiencing a hardware failure Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The refrigerator is too warm The refrigerator door was recently opened possibly for an extended period of time The fluid in the refrigerator was too warm when it was placed into the refrigerator Allow the refrigerator temperature to stabilize for 24 hours and re check The ambient air temperature is outside of the parameters specified in Section 11 Adjust the ambient air temperature allow the refrigerator temperature to stabilize for 24 hours and re check There is an obstruction preventing the door from closing Check that the factor amp harvest bag tubing lines are properly inserted into the channels on the front of the instrument and that the refrigerator door can close tightly There is a build up of ice on the inner cooling fins of the refrigerator Turn the instrument off and allow it to sit with the door open to defrost see Section 7 27 4 Confirm that there are no obstructions in the door opening Contact Biovest Account Services The refrigerator has lost power Check that the power cord is plugged into the instrument and to an outlet that is proven to supply the necessary voltage Check the power switch on the back of the instrument
10. 500 Media Temperature Control 37 0 Qutflow Rate nl hr 150 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 17 Fill Flush Step 11 Jan 12 2007 Fill Alush 16 35 09 12 Fill Flush Complete kip IN Pre Inoculation starts in 11 seconds Cycle Fall Fill Flush complete Media Rate ml hr 20 C02 Sensor 20 00 y Circulation Rate ml min 500 Media Temperature Control 37 0 Qutflow dein Ap 50 Refrigerator Temperature Control 4 0 t Hold Exit Next Help Figure 7 18 Fill Flush Step 12 AutovaxID Operations Manual 7 28 Biovest International Inc Operating Procedures 7 6 Note If the operator exits FILL FLUSH at step 12 before the countdown to PRE INOCULATION finishes the system will NOT enter PRE INOCULATION The system will continue to run the Media Outflow and Circulation pumps supply a fixed gas blend and maintain the media temperature until PRE INOCULATION begins The operator must return to the FILL FLUSH screen to start PRE INOCULATION The countdown to PRE INOCULATION will resume when the operator returns to the FILL FLUSH screen Pre Inoculation Procedure During PRE INOCULATION the cell culture environment is stabilized and is prepared for inoculation The pumps operate at preset parameters temperature control EC cycling and alarm event logging are active active pH control is turned on after the pH probe has been calibrated
11. 6 8 3 Jan 26 2007 Log Printer 14 36 31 IP Address 192 168 1 100 Pages per Sheet 4 Log Select Current Print Log Print Test Hold Exit Help Figure 6 22 Log Printer Screen Printer IP Address To print a log to a network printer the operator must know the IP address of the target network attached printer Consult your local IT support group for assistance with determining the IP address of your printer Once the IP address is known touch the IP Address field and enter it in dotted decimal format using the on screen keyboard Pages per Sheet The printout can be formatted so that multiple pages can be printed on each sheet of paper To change the formatting touch the Pages per Sheet field and select from the available formats one two four or eight pages per sheet Log Selection Any of the logs currently stored by the AutovaxID can be printed using the network printing function To choose the log to be printed touch the Log Select field and select one of the available logs AutovaxID Operations Manual 6 25 Biovest International Inc Software Setup amp Configuration 6 8 4 Test Print Before printing a log it is suggested that a test print be made to confirm the printer function A printing test will only print the first sheet of the current log formatted four sheets per page To perform a printing test first define the IP address of the printer and then press the Print Test b
12. Press the Exit button at the bottom of the Alarm Settings 2 screen to return to the Main Menu screen Alternatively repeatedly press the Back button to return to the Run Settings 1 screen so that the run settings may be saved to a profile before exiting AutovaxID Operations Manual 5 48 Biovest International Inc Primary User Interface Screens 5 9 Alarm Log Screen The Alarms screen see Figure 5 39 may be accessed by pressing the Alarm Log button on the Main Menu and may be used by any operator to view alarms and access the log Alternatively for quick access to the Alarm screen an operator may touch the title bar on the Monitor screen The top portion of the Alarms screen contains a white viewing window that displays alarms The number of alarms in queue is displayed in the title bar in parentheses next to the screen name Alarms number in queue The Alarms screen allows the user to view all current alarms clear each expired alarm acknowledge active alarms make log entries view a log current or historical and verify that the log is free of errors Press the Exit button to return to the Main Menu screen Jun 20 2006 AS Alarns 1 ACTIVE A 14 58 08 Jun 20 2006 14 57 38 Alarm High pH 7 97 Acknowledge Previous Next Log Entry Verify View Log Hold Exit Help Figure 5 39 Alarms Screen AutovaxID Operations Manual 5 49 Biovest International Inc Primary User Interface Scree
13. Strict Mode enhances proper and successful operation of the AutovaxID by limiting operators to progressing through a cell culture run using the correct sequence of steps and pre defined process control parameters Strict mode is most beneficial when used in conjunction with usernames and passwords When Strict Mode is enabled users that do not have administrative privileges are required to follow a rigid sequence of operation It does not allow non administrative users to perform overrides modify the run settings nor does it allow them to deviate from the defined sequence on the Run Control screen Strict Mode is not enabled by default See Section 6 6 for information on setting up user accounts AutovaxID Operations Manual 6 3 Biovest International Inc Software Setup amp Configuration 6 1 4 Audible Alarm Used This check box enables disables the internal audible alarm The audible alarm will sound whenever an alarm is active and has not yet been acknowledged The audible alarm will be silenced once an alarm is cleared or acknowledged by an authorized user The audible alarm is not enabled by default 6 1 5 Passwords Expire days This field defines how often new passwords must be created by registered users Password expiration only applies if usernames and passwords are in use Setting the password expiration to zero 0 disables this feature e Password Expiration Range 1 to 365 days in increments of one day e Default P
14. 1 1 3 The Intracapillary Circuit ee 1 1 4 The Extracapillary Circuit 0 0 0 0 ccceeeeeeeeeeeeeeeeeneeeeeeeees 12 EC Gycling Theoy tasa di Teel e 122 EE alle 1 3 System DESTA EE 1 3 1 EGO TOuenscreen EE 1 3 2 Refrigerator EE 1 3 3 Peristaltic Pump Module AAA 1 3 4 Cultureware Interface Panel 1 3 5 Disposable Cultureware ssssssneeeeeesseerrrrrrersserrrrrrre 1 3 6 Process Control Software A 1 3 7 AutovaxID Connect 1 3 8 AutovaxlD Distinctions setos EE Ee AT NET iii il 2 1 Space RoquilSMeNiS criada ratico 2 2 Power ee 2 3 Unpacking Instructions ever te SE enege ENKEN Eed geed 24 Moving Instructions eebe EK NESENEEEE ENER AEN ENER KENNEN 2 5 Voltage Selection EE 2 6 PUSS E ee E 2 Gas GOnnectionS Eege 2d LO SUP adi 2 7 2 Offgas Collection EEN 2 8 Ethernet Connection ss cacon ds 2 9 Bar Code Scanner Connechon EEN 210 iPower UP sos 2 11 External Computer Connechon EEN AutovaxID Operations Manual vi Biovest International Inc Table of Contents 3 Introduction to the Process Control Software cccccsesseeeceesseeeecesseeeeneesseeeeeenseeeseeaseeesenans 3 1 Bale SDI GON OL EE 3 2 3 2 Lactate CONTO RE 3 4 3 3 Media Temperature COn TO duos iaa 3 4 3 4 Refrigerator Temperature Control EEN 3 5 A EE A A EEEE E E 3 5 BO Ed EE 3 6 B20 Data ATCOIVO ite ee EE EE 3 10 4 Introduction to the User Interface ooccococcccooocononoccnonccnanononancnaaronannonann conan nona
15. 1 2 EC Cycling Theory EC Cycling Cycling is the process of alternating the pressure differential between the IC and EC to induce fluid flow across the membrane of the hollow fibers transmembrane This alternating flow of media between IC and EC acts to mix the media in the ECS and greatly reduces microenvironments and gradients within the bioreactor that would otherwise be detrimental to the health and productivity of the cell culture Cycling has two phases EC Rise and EC Fall see Figure 1 2 i Circ IC amp Circ Pressure Pressure Only EC Fall EC Rise IC EC IC EC Total IC Pressure gt EC Pressure Total IC Pressure lt EC Pressure Figure 1 2 EC Rise and EC Fall AutovaxID Operations Manual 1 4 Biovest International Inc Introduction 1 2 1 1 2 2 EC Rise During an EC Rise the IC pressure is greater than the EC pressure In the AutovaxID the pressure differential is induced pneumatically using the system s integral air pump and pressure control electronics The air pressure is applied to the IC at the IC reservoir The circulation pump also contributes some net pressure to the IC The IC EC pressure differential causes cell culture media within the ICS to slowly but continuously flow across the semi permeable hollow fiber membrane into the ECS The result is a net transfer of media from the IC to the EC Thus cells are provided fresh media including small molecular weight components As med
16. AutovaxID Operations Manual BIOVEST INTERNATIONAL 8500 Evergreen Boulevard Minneapolis MN 55433 6000 USA www biovest com Telephone 763 786 0302 Toll free in the U S 800 325 1112 After Hours 763 786 0302 voice prompt Fax 763 786 0915 Email accountservices biovest com Biovest International Inc Copyright Notice 2014 Biovest International Inc All rights reserved No part of this document may be reproduced in any form without the prior written consent of Biovest International Inc Biovest International Inc makes no warranties with respect to this documentation and disclaims any implied warranties of merchantability and fitness for a particular purpose Information in this document is subject to change without notice Biovest International Inc assumes no responsibility for any errors that may appear in this document Trademark Acknowledgements AutovaxID is a registered trademark of Biovest International Inc CompactFlash is a registered trademark of SanDisk Corporation MasterFlex L S and Easy Load are registered trademarks of Cole Parmer Instrument Company Luer and Luer Lok are trademarks of Becton Dickinson and Co PostScript is a registered trademark of Adobe Systems Incorporated Tygon is a registered trademark of Norton Performance Plastics Revision History Document Number 700371 000 Revision Date Description A 12 2006 Initial Release B 2 13 2006 Update to Revisi
17. AutovaxID Refrigerator AutovaxID Operations Manual 11 4 Biovest International Inc Specifications AutovaxID Incubator AutovaxID Flowpath Schematic For Legacy Cultureware Flow Temperature Sensor Probe EC Reservoir Permeable hollow fibers permit exchange between IC and EC of small molecules lt nominally 10 KDa Bioreactor Harvest EC Valve Open when Cycling Harvest EC Valve Closed when Cycling and harvesting Factor Valve N ec jormally open to Factor Valve jormally closed to iC lt gt d B O V E S T Intracapillary Circuit INTER Circulation Pump gt Female Luer Check Valve lt Male Luer Pump X Ratchet Clamp SYMBOL CODE T Flow Direction D I Slide Clamp T Y g Za E 0 2 Sterilizing Filter Automated Clamp E 3 A z En H Tubing Connection 3 SR NATI O NAL _ Extracapillary Circuit Factor Bou Set harvest Bag Set AutovaxID Refrigerator Figure 11 2 gt Supernatant Sample Supernatant lt lt AutovaxID Operations Manual 11 5 Biovest International Inc Glossary 12 Glossary Administrator Person who is qualified to supervise the operation of an instrument perform manual overrides and add other users to the system Alarm Notification Optional capability to use a remote computer to display alarms whenever an alarm occurs
18. Biovest International Inc Remote Access 10 2 INSTR NAME Windows Internet Explorer 5 x 6 Je http 127 0 0 1 8080 index html Google Pr we de mem NAME State GROWTH Elapsed Time 0 00 09 Media Temp 37 0 pH 7 10 IC Pressure EC Pressure Ref Temp EC State Air Flow 5 CO2 o Log Remote Lactate 0 KE EE CO2 Sensor Outflow 100 mlhr Medium 20 Factor 0 S Harvest bo Circulation 200 Name INSTR NAME 1 17 07 13 51 57 E memet Figure 10 2 Remote Access Home Page With Alarm From the main remote access screen two additional pages may be accessed a log pages that can be used to view and download the run logs and a remote page that can be used to remotely control the AutovaxID instrument Remote Control Remote control of the AutovaxID is achieved through a clickable web image that shows the actual instrument screen and allows mouse clicks to act as the touchscreen input The remote control screen is primarily useful to perform control changes remotely such as changing a pump rate Note If remote access is used an operator will not typically know if there is a leak or other physical instrument problem In some cases the operator may need to be standing in front of the instrument to make process control changes or perform calibrations during a cell culture run to ensure that the change has the desir
19. Biovest International Inc Software Setup amp Configuration Jun 20 2006 Help 15 37 26 User passwords are not enabled lo start using passuords press NEXT Or press EXIT to cancel Hold Exit Next Figure 6 8 Enable User Accounts amp Passwords Jun 20 2006 Help 15 38 14 A Administrator account is assigned to the first account It is the prinary administrator account This account can only be removed after all other accounts and will turn off password use when removed The account has adminstrator privileges to administrate other accounts It s very important not to forget the Administrator user nane and password Hold Exit Back Next Figure 6 9 Administrator Account Setup AutovaxID Operations Manual 6 13 Biovest International Inc Software Setup amp Configuration a ae Figure 6 10 Administrator Account Password AutovaxID Operations Manual 6 14 Biovest International Inc Software Setup amp Configuration After reviewing the first three screens the first input step is to enter the default password for the software 7860302 see Figure 6 11 Figure 6 11 Default System Password AutovaxID Operations Manual 6 15 Biovest International Inc Software Setup amp Configuration A username must then be entered for the primary administrator account see Figure 6 12 Please note that the username for the primary administrator account is
20. Gas Exchanger Bioreactor 1 Bioreactor 2 Bioreactor 3 Circulation Pump GEX Out Always Air CO2 AutovaxID Incubator Quick Coupler Circulation Pump Connection Media lt gt Outflow t IC Sample gt Factor 1 lt gt gt gt Harvest 1 Factor 2 lt gt gt Harvest 2 lt gt Factor 3 t gt gt Harvest 3 lt gt Factor Prime 1 Factor Prime 2 Factor Prime 3 m D D 2 2 2 a a a D E E E Ki a amp w v e v Check Valve gt Female Luer o 4 gt 2 O 2 H S O 3 2 s Pump lt 4 Male Luer a 2 2 E lt Flow Direction x Ratchet Clamp S Intracapillary Circuit n EE gt 0 24 Sterilizing Filter DI Slide Clamp o Tubing Connection Extracapillary Circuit Figure 8 2 Simplified Cultureware Flowpath Schematic Triple 0 5 m Bioreactor Version 8 5 Simplified Cultureware Flowpath Schematic A quick way to see the differences between simplified and standard cultureware is to compare their flowpath schematics figures 8 1 amp 8 2 versus 8 4 These schematics show the similarity in the Intracapillary Circuit IC components and layout between simplified and standard cultureware Conversely the Extracapillary Circuit EC components and layout are different between the two cultureware types primarily because simplified cultureware lack EC Cycling technology The IC and EC components are described in further detail next AutovaxID Operations M
21. The Cycling State has been overridden Clear any overrides to the Cycling State Contact Biovest Account Services There is no circulation The cultureware is not installed properly Override the valves to the Load Unload position and then place the instrument on Hold Unlatch and then re install the cultureware Resume the pumps from Hold and clear the valve overrides There is no fluid in the IC Make sure the Media clamp is open and the pump cassette is properly installed The fluid level in the IC reservoir is too low Temporarily override the Media pump to fill the IC reservoir 1 2 full There is air in the Circulation pump The Circulation pump is design to clear itself of air Allow 1 hour to elapse and re check circulation The Circulation pump is overridden to off or 0 Clear the override to the Circulation pump Contact Biovest Account Services The Circulation pump is running Contact Biovest Account tog fast Instrument hardware failure Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The COs supply line is disconnected or has become blocked Check that the CO supply line is connected and there is a flow of CO at the instrument There is an override on the gas blend Clear the override to the gas blend Incorrect media or media pH is to
22. The instrument is experiencing a hardware failure Contact Biovest Account Services The refrigerator is too cold The ambient air temperature is outside of the parameters specified in Section 11 Adjust the ambient air temperature allow the refrigerator temperature to stabilize for 24 hours and re check The fluid in the refrigerator was too cold when it was placed into the refrigerator Allow the refrigerator temperature to stabilize for 24 hours and re check The instrument is experiencing a hardware failure Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The instrument cannot raise the Media Temperature to the desired setpoint The ambient air temperature is outside of the parameters specified in Section 11 Adjust the room temperature to within the specified range The Media Temperature setpoint is set too high for the ambient room temperature Either decrease the Media Temperature setpoint or increase the room temperature The basal media supply is too cold The basal supply media should be 18 C to allow adequate media temperature control Loss of power to the heater Check for a Heater current fail alarm see Section 3 6 The instrument is experiencing a hardware failure Contact Biovest Account Services The ins
23. Total 245 Errors 1 Records 155 thru 164 w Mi al wi D Hold Exit Options Help Figure 5 41 Log Screen AutovaxID Operations Manual 5 52 Biovest International Inc Primary User Interface Screens There are three types of log entries alarms events and status Color is used to distinguish the types alarms are red events are green and status lines are black e Status Lines Status lines include sensor readings pH EC pressure IC pressure refrigerator temperature media temperature offgas percent CO and circulation rate gas number lactate and the outflow media factor and harvest pump rates The status lines give a snapshot of the system at periodic intervals The status lines are logged after FILL FLUSH according to the frequency set by the operator see Section 6 1 8 e Events Events include but are not limited to the progression of the run setup steps manual overrides changes to setpoints changes in cycling state from Rise to Fall manual or automatic changes to pump rates changes to the pump controllers user logon logoff and manual log entries e Alarms Alarms are logged as they occur and show the name of the alarm and whether it was out of range on the high or low side The log can be navigated using the on screen buttons e TOP move to the start of the log e PG UP move toward the start of the log by one page e Arrow Up move toward the start of the log by one record e Arrow
24. Touch Cal ibration Diagnostics Netuork Print Log Passwords Hold Exit Help Figure 6 1 Setup Screen Pressing each button will allow an authorized user to perform a unique instrument setup or maintenance function AutovaxID Operations Manual 6 1 Biovest International Inc Software Setup amp Configuration 6 1 System Features Screen Jan 16 200 Systen Features 11 34 17 server Confirmation Required Passwords Expire days 0 x Strict Mode Mininun Password Length 6 PS Audible Alarn Used logoff Tineout nin 60 Status Logging min 5 Om E Instrument Nane Hold Exit Next Hel p Figure 6 2 System Features Screen The System Features screen is accessed by pressing the Configure button on the Setup screen Monitor gt Main Menu gt Setup gt Configure The System Features screen see Figure 6 2 is used to setup several global instrument features including 6 1 1 Instrument Name The instrument name is user defined The instrument name is recorded at the start of a log appears in the title bar of the Monitor screen and is displayed on instrument s remote access home page see Section 10 1 Naming the instrument is optional but is particularly convenient when multiple AutovaxID systems are in use The instrument name may be any combination of letters numbers or punctuation marks that can be typed using the on screen keyboard up to twelve characters l
25. 1 inject the inoculum 2 remove supernatant samples to assay product concentration AutovaxID Operations Manual 8 6 Biovest International Inc Simplified Cultureware Usage Flow Temperature Sensor Probe EC Reservoir Bioreactor Harvest EC Valve hosed when Cychng ond harvesting Circulation AutovaxID Incubator SS Circulation Pump gt Female Luer w Check Valve lt 4 Male Luer a o o Pump X Ratchet Clamp 4 O i a lt Flow Direction P Slide Clamp v a S Za 28 3 Sa 0 24 Sterilizing Filter Gi Automated Clamp 5 3 E wn ev Y o Tubing Connection Y Luer Activated Valve 3 g gt Supernatant lt lt B O V E S T ntracapillary Circuit E N T E R N A T 0 N A t Extracapillary Circuit ee harvest Bag Set AutovaxID Refrigerator Figure 8 4 Standard Cultureware Flowpath Schematic AutovaxID Operations Manual 8 7 Biovest International Inc Simplified Cultureware Usage 8 6 Pumping Factor and Harvest The method of pumping Factor and Harvest differs between the two versions of simplified cultureware When using the single 1 1 m bioreactor version the AutovaxID s built in Factor and Harvest pumps and the pump cassette in which the Factor and Harvest tubing are loaded are used to deliver and collect Factor and Harvest When using the triple 0 5 m bioreactor version the AutovaxID s built in Factor and Harvest
26. 7 11 Mid Run Lactate Calibration A lactate calibration can be performed by an authorized user at any time during a cell culture run to force the AutovaxID system to recalculate the lactate variables The initial lactate variables were automatically calculated after the initial pH calibration during PRE INOCULATION Mid run lactate calibrations are useful when a different lot of media is added to the system If a lactate calibration is to be performed it is strongly recommended that a pH calibration be done first A single IC sample may be used for both tests although the pH should be measured first because the pH of the sample will change relatively rapidly whereas the lactate concentration is relatively stable AutovaxID Operations Manual 7 45 Biovest International Inc Operating Procedures To perform a lactate calibration 1 Press 3 on the Calibration Menu screen Monitor gt pH field gt Calibrate pH gt 3 The lactate calibration assumes that you have already obtained an IC sample and have performed a pH calibration Use the same sample to perform an off line lactate measurement Enter the off line lactate measurement on the Lactate Calibration Entry screen see Figure 7 22 and then press the Exit button If lactate control is not used then lactate calibrations can be ignored or omitted Consult a Biovest Technical Service representative or the Media Control Screen in Flowpath Setup for further information
27. Account Services if the filter is completely wetted Contact Biovest Account Services The IC circuit is empty The media supply is not connected Connect a media container to the Media line The Media clamp is closed Open the Media clamp so that the instrument can draw in media The connection to the media container is loose and the Media pump is drawing in air Tighten the media line connection Proper Q C tests should be implemented as the cultureware could be contaminated The pump cassette is not loaded properly Check for a Pump head not latched alarm see Section 3 6 Contact Biovest Account Services The EC pressure is too low EC pressure leak Visually inspect the cultureware for an EC media leak Review the log to verify that the cultureware Pressure Test was performed and that the cultureware passed The instrument is experiencing a hardware failure Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION EC reservoir is empty The IC is not pressurized Check that the IC vent filter has not become wetted with media The IC pressure is low Check for an IC pressure is too low alarm see Section 3 6 Check for a Cycling has not occurred Alarm see Section 3 6 Contact Biovest Account Ser
28. AutovaxID Operations Manual 7 12 Biovest International Inc Operating Procedures 7 3 5 Cultureware Installation 1 Transfer the entire cultureware including the loaded refrigerator rack the connected media container and the connected outflow container from the hood or BSC to the AutovaxID instrument 2 Place the factor harvest bag rack in the AutovaxID refrigerator compartment CAUTION Be careful of the chiller mounted on the ceiling of the refrigerator compartment to prevent damaging the materials or injuring your hands 3 Place the main cultureware assembly onto the AutovaxID instrument by sliding the incubator enclosure onto the black rails Push the cultureware back until it is mated flush with the interface panel 4 Place the three lines coming out of the refrigerator into the three slots on the left hand side of the refrigerator opening Close the refrigerator door Ensure that no tubing lines are pinched The luer activated valve is designed to stay within the refrigerator after each factor bag filling operation 5 Use the two latches at the top and side of the instrument to secure the cultureware in place Ensure that no external tubing lines are pinched CAUTION Do NOT use the latches to force the cultureware into position The cultureware should slide smoothly and mate easily with the instrument If any resistance is felt remove the cultureware and attempt to install it again If resistance is again fe
29. Contact Biovest Account Services if further assistance is required AutovaxID Operations Manual 7 50 Biovest International Inc Operating Procedures To change a container touch the pump field on the Monitor screen for the container that you would like to change On the pump Detail screen that appears press the Change Container button After pressing this button two screens will provide assistance with the change procedure The first screen the Help screen simply explains the change procedure see Figure 7 38 after reviewing press Next to proceed dan 16 2007 Help 19 53 26 This is used to change a container for the Media pump fou may enter a container identification and an optional log entry before changing the container Press NEXT to continue with the data entry Press EXIT to cancel Hold Exit Next Figure 7 38 Container Change Help Screen AutovaxID Operations Manual 7 51 Biovest International Inc Operating Procedures The next screen provides an opportunity to enter any details about the next container that will be connected and to make an entry in the log if necessary see Figure 7 39 The necessary container information is at the discretion of the operator but it may include the Media lot number the Factor lot number the ID of the Outflow container or the ID of the Harvest container that the Supernatant was transferred into as examples The container ID may be entered by touching th
30. During this phase the Circulation pump runs at 500mL min the Media pump runs at the user defined FILL FLUSH rate 20mL hr default see Section 6 4 4 and the Outflow pump runs automatically at varying rates to ensure proper system volume After an initial required pH calibration PRE INOCULATION continues automatically through factor addition until INOCULATION During PRE INOCULATION the AutovaxID state will be noted as PRE INOC on the Monitor Screen Jan 12 200 Preinoculation 1 16 37 33 1 Pre Inoculation Started MIT Media Rate ml Ar 20 Factor Rate nl ir 0 pH 17 10 C02 20 00 Hold Exit Help Figure 7 19 Pre Inoculation Step 1 AutovaxID Operations Manual 7 29 Biovest International Inc Operating Procedures 7 6 1 Initial pH Calibration PRE INOCULATION relies on operator intervention and will not advance until the pH probe is calibrated If an operator presses the Exit button and returns to the Monitor screen during FILL FLUSH or PRE INOCULATION the PRE INOCULATION procedure will pause at pH Calibration and await operator invention the AutovaxID State will be noted as PRE INOC on the Monitor screen To return to the current PRE INOCULATION screen press Menu gt Run Menu gt Continue Current Run The AutovaxID requires that the operator calibrate the pH probe using an off line analyzer Accurate pH and lactate control depends upon the quality and accuracy of the off line pH measurement
31. Events are in day hour minute format The new event will appear in the schedule all events are automatically sorted chronologically Press the Har field for the new event and enter the desired rate for the harvest pump in mL hr using the on screen keypad The factor pump has a range of 0 to 400 mL hour in 1mL hour increments AutovaxID Operations Manual 5 38 Biovest International Inc Primary User Interface Screens Jan 25 2007 Harvest Pump Schedule 7 33 50 Harvest Pump Schedule Elapsed Time Har Elapsed Tine Har Delete 10 00 00 4 Delete 15 00 00 2 All tines are on schedule 0 00 00 Advance Back Add Event Hold Exit Help Figure 5 32 Harvest Pump Schedule Screen To delete an event press the Delete button to the left of the event Events may be added or deleted during an active run by an authorized user The event that is currently in use is highlighted see the first event in Figure 5 25 If no events are currently active because not enough time has yet elapsed the highlighted area will appear after the last entry see Figure 5 32 An authorized user may enter a new value for any Harvest rate at any time during a run The Elapsed Time may also be changed for any event by an authorized user so long as the schedule has not been advanced or delayed The Advance and Back buttons are used to shift the time of all events backward or forward These buttons are intended to be used if the ce
32. Likewise if the Harvest pump rate is faster than the Factor pump rate it is possible to dilute the factor concentration in the EC In certain applications a lower EC factor concentration has been shown to result in greater productivity and or purer harvests AutovaxID Operations Manual 7 60 Biovest International Inc Operating Procedures 7 26 Optimization Guidelines The following suggestions are offered to users to help improve the quality of cell cultures that are grown using the AutovaxID Because each cell line responds uniquely to the cell culture conditions it is not possible to specify protocols that will work in all situations Biovest provides no guarantees of cell culture success using these methods Instead these methods should be used to gain experience and insight with your cell line and the AutovaxID Contact Biovest Account Services for assistance with optimization issues 7 26 1 Inoculation Guidelines Several factors related to the inoculum influence the growth of the cell culture in the flowpath These include the number of cells in the bioreactor their viability and their stage of growth when the inoculum is prepared A higher inoculum number cells mL may shorten the period between inoculation and the first indication of significant metabolism Although this may seem to be an advantage the user must be aware of the practical concerns in preparing a higher density inoculum e How many static culture vesse
33. NOTE In order to obtain the most accurate results perform the lactate calibration a few hours after a different lot of media has been attached to the cultureware The lactate calibration is specific to the lot of media and is dependent upon the bicarbonate concentration At least two system volumes of the new media should be allowed to dilute the old media in the flowpath before the lactate calibration is performed the slower Media pump rate the longer it will take to sufficiently dilute the IC media Jan 16 2007 lactate Calibration Entry 15 08 29 Sample Jan 16 2007 14 49 15 pH 100 coz 20 00 Enter offline neasurenent Lactate ngyml Hold Exit Help Figure 7 37 Lactate Calibration Entry Screen AutovaxID Operations Manual 7 46 Biovest International Inc Operating Procedures 7 12 Recommended Run Tasks There are various day to day procedures and process parameter changes that authorized users may be required to perform to maintain the cell culture environment The following schedule is a suggested guideline actual maintenance is dependent upon the cell culture conditions 7 12 1 Daily Tasks 7 12 2 Check that both the media and factor levels are sufficient to supply the AutovaxiID for the next twenty four hours Keep in mind that schedule or lactate controllers may change the current media pump rate and the rate of media consumption may increase For an estimation of the volume remaining
34. On the pH Calibration Menu press 1 to bring up the Calibration Sample screen see Figure 7 21 with instructions on how to perform a pH calibration 4 To complete Step 1 on the Calibration Sample screen verify that the off line pH meter has been recently calibrated using the appropriate calibration buffers or their equivalent and that it is prepared to measure a pH sample 5 Use a 10cc syringe to flush the IC Sample line Step 2 on the Calibration Sample screen If a 10cc was left on the IC Sample port during a previous pH calibration open the clamp on the IC Sample line and use the syringe to flush the line Alternatively aseptically connect a new 10cc syringe and perform the flush 6 Use the 5cc syringe to collect the pH sample Step 3 on the Calibration Sample screen Press the Next button The 10cc flush syringe from Step 2 may be discarded according to your facility s requirements As noted in Step 4 on the Calibration Sample screen it is important that the Next button is pressed IMMEDIATEL Y after the sample is taken so that the system can capture the current cell culture conditions for use during the calibration process Erroneous pH and lactate calibration results may occur if Step 4 is performed after the off line pH measurement is complete AutovaxID Operations Manual 7 43 Biovest International Inc Operating Procedures 7 After taking the sample and pressing the Next button you will be returned to the Ca
35. for target ranges of inoculation density volume media formulation and viability From the screen showing PRE INOCULATION Step 11 complete see Figure 7 29 press the Next button to advance to INOCULATION If the operator has already exited to the Monitor screen he she may return to the current PRE INOCULATION screen by pressing Menu gt Run Menu gt Continue Current Run 7 7 1 Inoculation Step 1 Prior to inoculation the system positions the Harvest EC valve so that the ECS is isolated the path from the Inoculate EC Sample port to the bioreactor is opened The Outflow pump speed is set to maximum to remove fluid from the IC reservoir to compensate for the fluid being injected during the inoculation process The operator is instructed by the AutovaxID software to enter the Inoculation Syringe ID see Figure 7 31 An Inoculation ID must be entered before you will be allowed to proceed Press the Syringe ID field and use the on screen keyboard or the barcode scanner to enter any alpha numeric ID AutovaxID Operations Manual 7 39 Biovest International Inc Operating Procedures 7 7 2 Jan 12 2007 Inoculation Step 1 16 45 11 Enter the Syringe ID Barcode Hold Exit Next Help Figure 7 31 Inoculation Step 1 Inoculation Step 2 After the Syringe ID has been entered press the Next button to proceed to the Inoculation Step 2 screen The operator should follow the on screen instructions to connect the sy
36. in the media or factor reservoir press either pump on the Monitor screen to bring up the Detail screen for that pump The Detail screen will give the operator the estimated current volume remaining in the respective reservoir as well as the operator specified capacity and current pump rate Ensure that the pH is within tolerance of the specified setpoint perform a pH calibration if necessary see the calibration schedule in Section 7 10 Check the CO supply to ensure sufficient capacity and that the pressure is 12 1 psi at the source Check the outflow reservoir to ensure that there is sufficient volume remaining to contain the outflow for the next twenty four hours Keep in mind that schedule or lactate controllers may change the current outflow pump rate For an estimation of the volume remaining in the outflow reservoir press the outflow pump on the Monitor screen to bring up the Outflow pump Detail screen The Detail screen will give the operator the estimated current volume remaining in the outflow container as well as the operator specified capacity and current Outflow pump rate Check that the outflow reservoir is properly vented and there are no kinked tubing lines Check the Monitor screen and Alarm Log screen for any alarms Investigate any alarms active or inactive and take appropriate action adding notes to the log as appropriate Check the offgas container empty any condensate from the container as required following
37. media bags with integral tubing lines and others Factor and Harvest containers generally hold small volumes i e 100 s of mLs to a few liters The appropriate size is determined by the volumes of Factor and Harvest needed during the run and the choice to remove batches of harvest during the run versus collecting all of the harvest into a single container Note smaller containers require more frequent changes which increase the amount of aseptic handling and risk of contamination Another determining aspect of container size is will they fit into a refrigerator placed near the AutovaxID Storing Factor and Harvest at 4 C during the run is beneficial but optional Beyond size considerations container choice also is influenced by the desire to reuse the containers over many runs versus using the containers only once This decision is often determined by the application When using a glass or plastic bottle with a ported cap to deliver Factor or collect Harvest the bottle cap must include a port to connect a sterilizing grade ventilation filter to allow the changing liquid volume in the bottle to be accommodated by air entering or exiting the bottle If the ventilation filter becomes wet the bottle will pressurize or experience vacuum These filters can be reused re sterilized but this practice is not recommended The bottle cap s ventilation filter MUST NOT become wet to prevent a hazard to personnel and equipment Because the cu
38. of the Advance button and enter zero 0 using the on screen keypad Press the Next button at the bottom of the Run Settings 2 screen to advance to the Run Settings 3 screen AutovaxID Operations Manual 5 43 Biovest International Inc Primary User Interface Screens 5 6 Run Settings 3 Screen The Run Settings 3 screen allows access to all of the user definable pH and temperature parameters see Figure 5 36 Similar to the previous two Run Settings screens pH and temperature settings are retained from run to run The settings displayed in the Previous Run Settings column indicate the settings that are associated with the current run settings profile The default settings are also presented for comparison Saved pH and temperature settings can be loaded using the Load Settings button on the Run Settings 1 screen See Section 5 1 2 for information about the run states during which each of these settings is used Jan 25 2007 hun Settings 3 9 22 02 Previous Run Settings Default Settings pH Setpoint 7 10 7 10 Media Tenp Setpoint 37 0 37 0 Refrig Temp Setpoint 4 0 4 0 Hold Exit Back Next Help Figure 5 36 Run Settings 3 Screen 5 6 1 pH Setpoint An authorized user may specify the pH setpoint The AutovaxID gassing system automatically adjusts the air and CO flow rates delivered to the GEX to maintain the circulating media at this pH e Default pH 7 10 e pH Range 6 80 to 7 60 in i
39. on position Confirm that both of the cooling fans on the rear of the instrument are functioning The two fans work in opposite directions the inlet fan near the top of the rear panel draws air into the instrument while the outlet fan near the power entry module pushes air out The inlet fan has a dust filter on it When the power switch is turned on the internal control computer will boot automatically and launch the AutovaxID User Interface Application control software The first screen that will be displayed is the touch calibration screen see Figure 2 9 The touchscreen is calibrated at the factory and should not require recalibration The calibration screen will close within a few moments if it is not touched See Section 6 7 Touch Calibration Screen for more information on calibrating the touchscreen AutovaxID Operations Manual 2 9 Biovest International Inc Instrument Setup Figure 2 9 Touch Calibration Screen 2 11 External Computer Connection The use of an external or remote computer is not required for most operations of the AutovaxID and is not supplied by Biovest However certain functions are only available when using an external computer that is connected to the AutovaxID instrument via an Ethernet network LAN No other provisions are provided for connecting a computer to the AutovaxID Refer to Section 10 for more information about using an external computer Refer to the user manual that comes
40. or fields which show system information or provide opportunities to enter data and operational parameters into the system pH and temperature setpoints pump speeds etc Each screen provides a different control and display activity The primary user interface screens are Monitor Displays a summary of the status of the instrument and flowpath Accessible by pressing the Menu button on the Monitor screen Provides access to the Run Menu Run Settings Alarm Log Flowpath Setup and Main Menu Advanced screens Also provides details about the instrument software and current run if applicable Allows an authorized operator to start a new run continue an existing run Run Menu ows alle H b 9 or stop the current run Run Settings A series of screens that allow access to run settings including alarm limits Displays a log current or historical of all run activities and events such as Alarm Log periodic status lines system events and alarms Also displays a queue of active alarms for review or individual action Flowpath Graphically displays the status of the flowpath Access to the system configuration date time IP address default run settings etc diagnostics for sensor calibration and sensor board troubleshooting user and password setup touchscreen calibration and printing the log to a TCP IP network enabled printer Setup Access to the Run Menu for advanced system operation and
41. or it may be controlled manually by an authorized user The controller type is defined on the Factor Controller Screen see Figure 5 27 which can be accessed by pressing the Settings button The factor controller is typically set during run setup see Section 5 5 See Section 5 5 5 for details about each of the factor pump controller types Jan 24 2007 Factor Detail 14 03 33 GG 4 Factor 0 ml hr 3000 ml Capacity 2960 nl Current Volune Change Container Settings Hold Exit Help Figure 5 13 Factor Detail Screen AutovaxID Operations Manual 5 15 Biovest International Inc Primary User Interface Screens If the factor pump is being automatically controlled by the process control software it may be overridden only by an authorized user using an advanced override see Section 5 12 2 Biovest does not recommend manually overriding the factor pump rate if the pump is being automatically controlled and if the system is operating normally Improperly setting the factor pump or neglecting to punctually clear an override may negatively affect the cell culture environment An authorized user may change the factor container by pressing the Change Container button see Section 7 17 An authorized user may also change the capacity of the factor container see Section 7 18 AutovaxID Operations Manual 5 16 Biovest International Inc Primary User Interface Screens 5 1 15 Harvest Detail Screen
42. pumps and the cultureware s pump cassette are NOT used to deliver and collect Factor and Harvest Rather a separate pump s and pump head s are used because the AutovaxID s pump motors and pump cassette do not accommodate three Factor and three Harvest tubing lines The type and number of pump motor s and pump head s to be used will vary depending on your budget and the necessity to vary Factor and Harvest pumping methods between the three bioreactors We suggest the following equipment below that while not the least expensive setup is one that will be fairly simple to operate and provide reasonable flow rates for the size of the bioreactors This minimal configuration will not allow different flow rates between the three bioreactors however For assistance with selecting other equipment that will meet your application please contact your supplier of this equipment or Biovest s Account Services department The three Factor and three Harvest tubing lines are PharMed and Masterflex size 14 Suggested minimum pumping equipment necessary to use simplified cultureware with three 0 5 m bioreactors 1 Masterflex Pump Motor quantity 1 e Cole Parmer part number EW 07523 90 e Masterflex L S Digital Drive e 0 02 100 RPM e 90 260 VAC e Continuous flow rate range using Factor and Harvest s size 14 tubing is 0 25 1 260 mL hr other rates are possible using its batch pumping methods 2 Masterflex Pump Head quantity 3 e Cole P
43. respectively e Low When a sensor value is lower than the Low limit an alarm is issued e High When a sensor value is higher than the High limit an alarm is issued NOTE The Low value must be less than or equal to the High value See Table 3 4 for the High Low alarm limits If enabled alarms are activated on the following schedule e The pH alarm is active during PRE INOCULATION and GROWTH e The Media and Refrigerator Temperature alarms are active after 7 5 hours into the 24 hour FILL FLUSH and remain active until the end of the run e The IC and EC pressure alarms are active when FILL FLUSH starts and remain active until the end of the run Authorized users may enable and disable alarms during active cell culture runs Jan 16 2007 Alarm Settings 1 11 27 07 Alarms g Im a Media Tenperature 35 0 39 0 Refr Tenperature iz 2 0 10 0 IC Pressure Es 1 0 0 275 0 mhg EC Pressure IM 1 0 0 1275 0 mhg Hold Exit Back Next Help Figure 5 37 Alarm Settings 1 Screen AutovaxID Operations Manual 5 46 Biovest International Inc Primary User Interface Screens Press the Next button at the bottom of the Alarm Settings 1 screen to advance to the Alarm Settings 2 screen 5 8 Alarm Settings 2 Screen The Alarm Settings 2 screen see Figure 5 38 allows the user to view and change the high and low alarm limits for Lactate Cabinet Temperature and Cycle Time Pressing a Low or High field wil
44. 47 3 wir EC Fall Tine 15 minutes 54 2 mhg Type Tine Hold Exit Help Figure 5 8 Cycle Detail Screen AutovaxID Operations Manual 5 10 Biovest International Inc Primary User Interface Screens 5 1 10 Media Temperature Screen On the Monitor screen press the Media Temperature field or the surrounding area to bring up the Media Temp screen see Figure 5 9 This screen allows an operator to view the media temperature setpoint the current temperature of the circulating media and the temperature of the air exiting the heating coil An authorized user may change the media temperature setpoint by pressing the Setpoint field and entering a new media temperature setpoint using the on screen keypad The air temperature exiting the heating coil may not be modified and is shown for diagnostic purposes only The operator also may observe if the media temperature control system is on or off checked or unchecked respectively Media temperature control may be manually turned on or off only by an administrator as a diagnostic function If the instrument is functioning normally Biovest does not recommend turning off the media temperature control system doing so will likely cause the incubator temperature to fall and may negatively affect the condition of the cell culture environment Jan 24 2007 Media Temp 11 40 3 Setpoint 37 0 Control NA Current Reading 37 0 Air Tenperature 65 7 Hold
45. 600285 120 600285 230 600285 240 Description AutovaxID Cell Culture System 100VAC AutovaxID Cell Culture System 120VAC AutovaxID Cell Culture System 230VAC AutovaxID Cell Culture System 240VAC Cultureware Options 600307 211 600308 205 600320 211 600320 221 AutovaxID Simplified Cultureware One 1 1m Bioreactor excludes EC Cycling pH Probe Factor and Harvest bags AutovaxID Simplified Cultureware Three 0 5m Bioreactors excludes EC Cycling pH Probe Factor and Harvest bags AutovaxID Standard Cultureware One 1 1m Bioreactor includes EC Cycling pH Probe Factor and Harvest bags AutovaxID Standard Cultureware One 2 1m Bioreactor includes EC Cycling pH Probe Factor and Harvest bags Legacy Cultureware Options 600290 211 600290 221 AutovaxID Legacy Cultureware One 1 1m Bioreactor includes EC Cycling pH Probe Factor and Harvest bags AutovaxID Legacy Cultureware One 2 1m Bioreactor includes EC Cycling pH Probe Factor and Harvest bags Optional Sterile General Cultureware Accessories 600288 000 600295 000 600063 000 600063 001 600100 072 Spare Factor Bag Set Sterile 3L Total Volume Single Use use with 600290 XXX Legacy cultureware only Spare Harvest Bag Set Sterile 3L Total Volume Single Use Mid Volume In Line Media Filters Sterile Single Use 4 per package 1 per sterile pack High Volume In Line Media Filters Sterile Single Use 4 per package 1 per sterile p
46. 7 is intended to be used only by Biovest service personnel It allows them to perform hardware tests and to view diagnostic readings associated with the instrument hardware for diagnostic and calibration purposes These functions can interfere with normal operation of the instrument and should be performed only by trained personnel Descriptions of these screens are beyond the scope of this manual Feb 06 2007 Diagnostic Menu 17 38 01 Boards Cal ibrations Sensors system PN 300172 000 Rey B Hold Exit Hel p Figure 6 7 Diagnostic Menu Screen AutovaxID Operations Manual 6 11 Biovest International Inc Software Setup amp Configuration 6 6 User Accounts Screen The User Accounts screen is accessed by pressing the Passwords button on the Setup screen Monitor gt Main Menu gt Setup gt Passwords The AutovaxID software can provide user security for up to twenty unique user accounts Different access rights can be assigned to Administrators and normal users Users thereby optionally restricting User access to critical functions The log will automatically record logon and logoff events as they occur The security features require that each operator is given a unique username and presumes that their password is kept confidential A user with administrator rights has full access to the AutovaxID system An administrator may manage his her own account and setup other accounts An administrator also ma
47. 7 22 pH Calibration Entry AutovaxID Operations Manual 7 32 Biovest International Inc Operating Procedures After the pH value has been entered press the Exit button The AutovaxID will then automatically perform the initial lactate calibration see Figure 7 23 After the automatic lactate calibration the AutovaxID will report that all calibrations are complete and that it is ready for factor addition Just prior to preparing the scaled up cells for inoculation press the Next button to start factor addition see Figure 7 24 Figure 7 23 Pre Inoculation Step 3 AutovaxID Operations Manual 7 33 Biovest International Inc Operating Procedures 16 39 05 a 0 os Figure 7 24 Pre Inoculation Calibrations Complete 7 6 2 Factor Addition A ter the initial pH calibration and just before preparing the cells for inoculation the operator initiates the Factor Addition procedure as pre defined on the Run Settings 1 screen see Section 5 4 Factor Addition is used to add ECS media factor to either the IC and or EC prior to inoculation Examples of factor could include serum or other high molecular weight protein supplements used during the scale up of cells prior to inoculation Unless protein free media is to be used the addition of factor to the bioreactor is necessary because FILL FLUSH is performed using only basal media If protein supplements are necessary Factor Addition results in the conversion o
48. Account Services Wet or discolored area around Off Gas port on rear cover of instrument Offgas line is not connected Install an offgas line per Section 2 7 2 AutovaxID Operations Manual 9 12 Biovest International Inc Remote Access 10 Remote Access The AutovaxID is designed to optionally allow the use of a remote computer to access various instrument functions Remote access requires that the AutovaxID be connected to a facility s computer network LAN Each facility must consider the impact of remote access relative to their individual security policies Note The remote control feature can be disabled in the Instrument Setup screens see Section 6 if a facility has a concern about security Using Remote Access operators can remotely monitor and control the AutovaxID using a network attached computer and a standard web browser Remote access is described below and is briefly discussed elsewhere throughout the manual To monitor or control an AutovaxID remotely you must know the IP address of the instrument If you are on the same physical and logical network as the AutovaxID instrument that you wish to monitor then point your web browser to http 111 222 333 444 where 111 222 333 444 is the IP address of the instrument that you wish to access Please see the AutovaxID system administrator or your IT personnel to determine this address it will be unique for each instrument I
49. AutovaxID software will automatically step through the fill flush procedure without operator intervention until the first pH calibration is required The entire fill flush process takes approximately 24 hours to complete from start to finish and uses approximately five liters of media 3 Calibrate This step is required to calibrate the pH sensor and to perform the initial lactate calibration The pH probe is dry during sterilization shipping and storage and is re hydrated during FILL FLUSH so this pH calibration is critical to system function and performance The pH calibration must be performed by a trained operator After the pH calibration is complete a lactate calibration is automatically performed by the AutovaxID software NOTE Two additional pH calibrations are required during each cell culture run See Sections 7 9 and 7 10 4 Pre Inoculation During pre inoculation active process controls are used to stabilize and control pH and EC Cycling to user defined setpoints During this step factor may be introduced into the cultureware to pre condition the flowpath 5 Inoculation Cells are manually introduced into the bioreactor by a trained operator during this step The software will walk the operator through the requisite steps Inoculation should take approximately five minutes not including the time to prepare the cell inoculum 6 Growth The growth phase starts immediately after inoculation is complete The AutovaxID sy
50. EC Cycling but may be removed by an operator if required by the facility procedures To maximize the cell removal this procedure should be performed immediately after the end of a Rise cycle This procedure requires Advanced Overrides to be set and may be performed only by an administrative user To remove cells from the EC reservoir 1 Obtain three sterile 60cc syringes with male Luer connectors that can be used to collect the cell mass from the EC reservoir An operator should expect to collect 140mL from the EC reservoir 2 Aseptically remove the sterility protector from and connect one of the syringes to the Cell Removal line of the cultureware This line exits the incubator at the front cultureware near the left side Aseptically retain the sterility protector for re use after the cell removal is complete 3 If usernames and passwords are enabled logon to the AutovaxID as an administrative user AutovaxID Operations Manual 7 55 Biovest International Inc Operating Procedures 4 Watch the Monitor screen and wait for the cycle state to switch from Rise to Fall After the cycle switches press Menu gt Advanced gt Overrides to access the Overrides screen see Figure 5 49 5 Touch the Harvest EC valve field and override the valve to the 3 Brx position 6 Touch the Cycle State field and override the cycle state to Drain 7 Open the clamp on the Cell Removal line and allow the contents of the EC reservoir to be pus
51. EE E E 5 49 5 9 1 Alarm Handling DEE EE 5 50 5 9 2 Og ENY eegene A Aa E LER 5 50 5 9 3 Very LOG EE 5 51 E EE Ee E 5 52 GH Log View OPtOnS osno pod nent 5 54 5 10 Flowpath Screen hoe ae aac Gaeta ca Se 5 56 A A a eaa E E e aa ae aeea eaaa a AEE A A E Eaa AKEE R 5 59 5 12 Advanced En EE 5 60 5 12 1 Advanced RUM Men sacd 5 61 5 12 2 Overrides SCreGn sesir aeei cia odia cod 5 64 5 123 Cycle Overrides 10 As 5 65 5 12 4 Gas OVE res ce te A ee ee eal eas eae 5 66 512 50 ractor Valve Overrides ee ested cok een hat a EE ag 5 66 5 12 6 Harvest Valve Overrides scraps 5 66 5 12 PUMP EE 5 67 AutovaxID Operations Manual viii Biovest International Inc Table of Contents 6 Software Setup amp Configuration ccccccsesseeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeesseeeeneeeeeeeeesesseeeneeeeeeeeenseees 6 1 6 1 System Features Screen eher 6 2 Ce WR Ter n 6 2 6 1 2 Server Confirmation RequirSd iia ala 6 2 6 1 3 Strict de E 6 3 6 1 4 Audible Alarm Used WEEN 6 4 6 1 5 Passwords Expire T lee ege dean dade Radha da Rae EES 6 4 6 1 6 Minimum Password Leng aorta gees oh gege ERC EEeechES deed 6 4 612 Geet dree DEEN nal deeg 6 4 6 1 8 lt Status Bee Le ae Hd EE 6 4 6 1 9 Status LOGGING ON van ti en ege Been Eed Ee ee eet ee 6 5 6 2 Network Configuration Gcreen EEN 6 5 621 A a ae a a eenaa E a eaaa eA AEE EEREN 6 5 6 2 2 Automatic Address Contiguration EE 6 6 6 2 3 Browser Remote Control aros aia 6 6 6 2 4 Network Write C
52. Exit Help Figure 5 9 Media Temp Screen AutovaxID Operations Manual 5 11 Biovest International Inc Primary User Interface Screens 5 1 11 Refrigerator Temperature Screen On the Monitor screen press the Refrigerator Temp field or the surrounding area to bring up the Refrigerator Temp screen see Figure 5 10 This screen allows an operator to view the refrigerator temperature setpoint the current refrigerator temperature and the temperature of the air inside the instrument cabinet An authorized user may change the refrigerator temperature setpoint by pressing the Setpoint field and entering a new refrigerator temperature setpoint using the on screen keypad The cabinet temperature may not be modified and is shown for diagnostic purposes only The operator also may observe if the refrigerator temperature control system is on or off checked or unchecked respectively Refrigerator temperature control may be manually turned on or off only by an administrator as a diagnostic function If the instrument is functioning normally Biovest does not recommend turning off the refrigerator control system doing so will cause the refrigerator to warm to the ambient temperature at minimum and may harm any factor or harvested product stored therein Jan 24 2007 Refrigerator Temp 13 13 57 Setpoint 4 0 Control On EA Current Reading 4 0 t Cabinet Tenperature 30 7 Hold Exit Help Figure 5 10 Refrig
53. FLUSH so it is critical that a pH calibration is performed after FILL FLUSH It takes approximately two to ten minutes to complete a pH calibration depending upon the off line measuring device After the pH calibration is complete then a lactate calibration is automatically performed and factor addition begins see Sections 7 6 1 and 7 6 2 Factor addition is setup by the operator on the Run Settings 1 screen see Section 5 4 The duration of the factor addition step will depend upon the volume of factor to be delivered to the IC and or EC The factor pump operates at maximum rate of 400mL hr to deliver the factor It is recommended that factor addition be done no more than 24 hours prior to inoculation The instrument will remain in the PRE INOCULATION phase until an authorized operator initiates inoculation The media pump operates at a minimum of 20 mL hr for the duration of PRE INOCULATION to compensate for any evaporative loss through the GEX e INOC INOC indicates the instrument is in the inoculation phase of the cell culture run During this phase cells are inoculated into the bioreactor The AutovaxID software will walk the user through the requisite steps Inoculation should take approximately five minutes not including the time to prepare the cell inoculum e GROWTH The GROWTH phase starts immediately after inoculation is complete The AutovaxID system remains in this state until the end of the run Cells will star
54. It is best to remove the air from each new Factor container s tubing rather than allowing it to be pumped into the bioreactor ECS where the cell culture resides The air is removed by pumping it up the cultureware s Factor tubing line and diverting it down the Factor Prime line and into a syringe Another use of Factor Prime is to inject high molecular weight components such as growth supplements into the ECS This optionally includes the initial Factor Addition to the EC if the automated volume to be pumped is set to 0 mL This may be done using a highly concentrated solution of factor to conserve the volume in the Factor container for use during the run after inoculation Additionally if growth supplements are needed after inoculation they may be added via Factor Prime rather than from the Factor container Doing so requires extra aseptic connections which should be minimized to reduce the risk of contamination Procedure to remove air from each new Factor container All steps must be done aseptically 1 When the Factor container in use is nearly empty yet the run will continue the Factor container must be changed Note the change container alarm may be used when the single 1 1 m bioreactor version is in use 2 Change the Factor pump speed to 0 mL hr Note alternatively the change container feature may be used when the single 1 1 m bioreactor version is in use 3 Close the Factor 1 clamp or Factor 2 or Factor 3 as neces
55. Off Gas Port Ethernet Port Power Entry Module ZE Barcode Scanner CO Port Figure 2 1 AutovaxID Rear Panel Connector AutovaxID Operations Manual 2 2 Biovest International Inc Instrument Setup 2 1 Space Requirements The instrument should be placed on a level surface in a room with ambient temperature of 20 to 30 C 68 to 86 F The location of the instrument should provide easy access to a laminar flow hood See Section 11 Specifications for the footprint dimensions of the instrument Additional access is needed for the operators to turn the pump loading knob on the left side of the instrument and to allow the refrigerator door to swing open It is important to keep a minimum of four inches clearance in back of the instrument for facility connections and air flow 2 2 Power Requirements The AutovaxID instrument is factory pre set for the operating voltage requested by the customer The electrical supply circuit must be capable of supplying at least 600 watts of power See Section 11 Specifications for details about power requirements Biovest recommends the use of an uninterruptible power source with an emergency backup to prevent any interruption to an active cell culture run WARNING Do not apply power to the instrument before verifying that the voltage selection and fusing is correct for the local supply voltage Applying incorrect voltage may damage the instrument and cause unsafe conditions See Section 2 5 V
56. Operations Manual Biovest International Inc Software Setup amp Configuration 6 7 Touch Calibration Screen An authorized user may calibrate the touchscreen by pressing the Touch Calibration button on the Setup screen Monitor gt Main Menu gt Setup gt Touch Calibration The Touch Calibration screen see Figure 6 21 prompts the user to touch the screen in five different places in sequence at the center of a crosshairs The accuracy of the calibration is dependent upon the accuracy of the person doing the calibration This screen is also briefly displayed during system startup though it should not be used unless needed During system startup Touch Calibration is accessible to all users Touch Calibration Touch the cross Figure 6 21 Touch Calibration Screen 6 8 Network Log Printing An authorized user can print a current or historical log to a network connected TCP IP printer The printer must be PostScript capable and the operator must know the IP address of the printer To print a log to a network printer press the Network Print Log button on the Setup screen Monitor gt Main Menu gt Setup gt Network Print Log The Log Printer screen see Figure 6 22 allows the operator to specify the printer select the log configure the printing preferences and execute a printing test AutovaxID Operations Manual 6 24 Biovest International Inc Software Setup amp Configuration 6 8 1 6 8 2
57. Service Contact us with any question you have when operating the AutovaxID instrument or cultureware Please be ready to provide as much detailed information as possible including instrument serial number cultureware manufacturing lot number printed or electronic run log process alarms chronology leading to the situation relevant previous run experiences etc Customer Service The Parts List chapter 13 has ordering information for common items Field Service The Parts List chapter 13 has important information about Calibration and Preventive Maintenance services that keep your AutovaxID operating optimally and consistently to ensure its highest production lowest operating cost and minimal unplanned downtime We offer a Full Service Contract to simplify maintenance of the instrument over its many years of usage If it is malfunctioning we provide full repair services AutovaxID is not user serviceable Sales Support We re happy to help you purchase any of Biovest s products or services AutovaxID Operations Manual V Biovest International Inc Table of Contents Table of Contents Symbols and WarningS cccceseeeeeseeeeeeeeeeeeeeeeeeeeeeeeeeeeaeeeeeeneeeeeeeeseneeseeeeees About this Manual comiiioiniaicia aii aida Service and Support ae isirorcoionion iii Table e Y Introduction aaaea diaria 1 1 Hollow Fiber TACO ssl tas 1 1 1 Bioreactor Technology ke 1 1 2 AutovaxID Technology Difference eee
58. allow the AutovaxID to obtain an IP address automatically from a DHCP server check the Automatic Address Configuration box and then press the Apply button The IP address assigned by the server will appear in the IP Address fields Press the Info button for further diagnostic information if required 6 2 3 Browser Remote Control Use this checkbox to enable checked or disable not checked remote access See Section 10 for more information about remote access This feature must be disabled when using the AutovaxID in a Part 11 compliant facility AutovaxID Operations Manual 6 6 Biovest International Inc Software Setup amp Configuration 6 3 6 2 4 Network Write Control Use this checkbox to enable checked or disable not checked write access to the instrument When Network Write Control is enabled commands can be sent to the instrument using the LAN to read and make changes to the instrument configuration and run setup parameters If this feature is disabled then changes are not permitted and a network device can only read from the instrument This feature should be disabled unless needed by a LIMS or MES The proper utilization of this feature must be evaluated when the AutovaxID is used in a Part 11 compliant facility NOTE This feature is independent of the Browser Remote Control Feature Network Write Control is intended for use with a LIMS or MES Disabling Network Write Control does NOT disable Browser Remote Control
59. an alternate Factor Pump Schedule screen see Figure 5 29 This screen allows an authorized user to schedule up to ten elapsed times with variable pump rate values for the factor and harvest pumps The Harvest Pump Schedule screen is not available when Tie to Harvest is enabled In strict mode and when user names and passwords are in use only administrators may create schedules Normal users may view the schedule but they may not change the values in the Elapsed Time or the Factor fields nor can they add or delete events Normal users may advance or delay the schedule using the Advance and Back buttons or by entering a manual time shift To add an event press the Add Event button at the bottom of the screen enter the desired Elapsed Time for the start of the event using the on screen keypad and then press Enter Events are in day hour minute format The new event will appear in the schedule all events are automatically sorted chronologically Press the Fac field for the new event and enter the desired rate for the factor pump in mL hr using the on screen keypad similarly press the Harv field for the new event and enter the desired rate for the harvest pump in mL hr using the on screen keypad Both pumps have a range of 0 to 400 mL hour in 1mL hour increments The pumps may be run at different rates To delete an event press the Delete button to the left of the event Events may be added or deleted during an active run by an author
60. and adherence to the pH calibration procedure Care must be taken to ensure good results To prevent delaying inoculation this initial pH calibration should be performed as soon as possible after the system reaches PRE INOCULATION Active pH control will start after pH calibration is complete Prior to starting the pH calibration the operator must have three sterile syringes one 10cc syringe that will be used for flushing the sample line one 5cc syringe that will be used for drawing the sample and another 10cc syringe that will remain connected to the IC Sample port to protect the sterility of the port It may also be used for flushing the IC Sample line prior to taking the next pH calibration sample All three syringes must have male Luer fittings To perform the pH calibration 1 Press the pH Calibrate button on the Preinoculation 1 screen see Figure 7 20 2 The Calibration Sample screen see Figure 7 21 will instruct the operator how to perform a pH calibration 3 To complete Step 1 Make sure Off line meter is ready verify that the off line pH meter has been recently calibrated using the appropriate calibration buffers or their equivalent and that it is prepared to measure a pH sample Use one of the larger 10cc syringes to complete Step 2 Use the smaller 5cc syringe to collect the pH sample during Step 3 Press the Next button The flush syringe from Step 2 may be discarded according to your facility s requi
61. are defined by the operator the outflow pump rate is automatically controlled by the AutovaxID software All pumps have active feedback so that a stall or race condition can be detected and reported to the operator The pump tubing is put in to the pumps by the operator using a pre loaded pump cassette that is included with the sterile cultureware The pump cassette is installed to the pump module using the knob on the left side of the instrument The pump module can detect the presence of the pump cassette the AutovaxID software will report if the cassette is not loaded properly 1 3 4 Cultureware Interface Panel The cultureware interface panel is the area of the instrument that contains all of the components that mate with the disposable cultureware including e Circulation Pump this is a magnetically coupled motor that mates with a disposable pump head in the sterile cultureware The circulation pump rate is defined by the operator and is controlled by the AutovaxID software e Circulation Flow Sensor this sensor couples with a disposable flow cell in the sterile cultureware It detects the IC media flow rate and provides feedback to the circulation pump control algorithm e Rotary Valves these cam style valves occlude various tubes in the flowpath at different stages throughout a cell culture run The valves are controlled by the AutovaxID software and have active position feedback The tubing is automatically positioned on the val
62. begin pumping factor into the factor bags 12 Stop the pump when the bags contain desired volume determined above Do not exceed 3L Over filling the factor bags may damage them and cause factor to leak 13 Close the five clamps on the Factor Fill tubing line 14 Aseptically disconnect the Factor Fill tubing from the two top ports of the factor bags 15 Use the two supplied sterile Luer plugs to close the top ports of the factor bags 16 Unload the Factor Fill tubing assembly from the pump and discard it per your facility s procedures 17 If you are following the Alternative Procedure Legacy Factor Bag Set Only described in the previous section see Cultureware Preparation above perform the following steps a Refrigerate the filled Factor Bag Set until it is needed for a cell culture run b To start a run disinfect the Factor Bag Set and place it in a sterile laminar hood or BSC c Transfer the Factor Bag Set from the rack the bags are on to the other rack that already has the cultureware s harvest bags mounted on it see step 15 under Cultureware Preparation above d Perform steps 16 18 under Cultureware Preparation above Note Biovest recommends that the filled Factor Bag Set be cooled to approximately 4 C before it is placed in the AutovaxID refrigerator Placing warmer bags in the refrigerator may cause alarms and will take up to twenty four hours to reach the refrigerator temperature setpoint
63. begins If an operator returns to the Monitor screen during SETUP FILL FLUSH will not begin until another pressurization test is performed and the run sequence resumes e FILL FILL indicates that the instrument is performing a fill flush of the cultureware This step will automatically begin after a successful pressurization test unless an operator intervenes FILL FLUSH introduces growth media into the cultureware and brings the incubator up to operating temperature through a series of pre defined automated procedures see Section 7 5 The entire fill flush process takes approximately 24 hours to complete from start to finish and uses approximately five liters of media AutovaxID Operations Manual 5 2 Biovest International Inc Primary User Interface Screens e PRE INOC PRE INOC indicates that the instrument is in the PRE INOCULATION phase of the cell culture run PRE INOCULATION will continue indefinitely without operator intervention During PRE INOCULATION active process controls are used to stabilize and control the pH and EC Cycling to specified setpoints and factor may be introduced into the cultureware to pre condition the flowpath The necessary peristaltic pumps operate as programmed and the circulation pump runs at 500mL min The first step of PRE INOCULATION is to calibrate the pH sensor see Section 7 6 1 The pH probe is dry during sterilization shipping and storage and is re hydrated during FILL
64. capacity of the harvest container see Section 7 18 AutovaxID Operations Manual 5 17 Biovest International Inc Primary User Interface Screens 5 1 16 Circulation Detail Screen On the Monitor screen press the Circulation field or the surrounding area to bring up the Circulation Detail screen see Figure 5 15 This screen shows the target and actual circulation pump rates in mL min Depending upon the instrument state and the condition of the cell culture the sensed flow rate may not always match the target flow rate An alarm will be issued if the difference is excessive If the circulation controller is set to Manual then an authorized user can set the circulation pump rate by touching the Circulation field and entering the desired circulation pump rate using the on screen keypad An authorized user can access the Circulation Controller screen see Figure 5 34 by pressing the Settings button See Section 5 5 12 for details on the Circulation Controller screen Jan 23 2007 Circulation Detail 17 30 42 J 2 Circulation 200 ml min Flow Sensor 190 ml min settings Hold Exit Hel p Figure 5 15 Circulation Detail Screen AutovaxID Operations Manual 5 18 Biovest International Inc Primary User Interface Screens 5 2 5 3 Main Menu From the Monitor screen press the Menu button to access the Main Menu screen see Figure 5 16 This screen displays six buttons that provide access to
65. cells for an End of Production EOP cell bank Supernatant Sample This port is used to withdraw product directly from the harvest bag for quantization Cell Removal This port can be used to remove accumulated cell debris from the EC reservoir The four remaining ports are Media This port is used for connecting the IC media supply It has male and female Luer connectors a coiled tube that can stretch to 60 and a clamp Outflow This port is used for connecting an outflow line or container It has male and female Luer connectors a coiled tube that can stretch to 60 and a clamp Supernatant This port may be used to connect the harvest bags to another system for post processing It has a male luer connector and a clamp 600320 XXX Standard Cultureware fourth port Factor Fill This port is used for filling re filling the Factor Bag Set The port has a female Luer connector to which a sterile Luer Activated Valve and LAV cap are attached after the initial filling operation 600290 XXX Legacy Cultureware fourth port Factor 1 This port is used for connecting the Factor Bag Set to the main cultureware via the pump cassette The port has a male Luer connector AutovaxID Operations Manual 7 6 Biovest International Inc Operating Procedures 7 3 2 Factor Bag Set Description The Factor Bag Set is designed to hold media supplemented with high molecular weight components that are needed for optimal cell growth Thi
66. figure 8 5 is supplied with a threaded electrical cap at the top of the probe and a small bottle with storage solution to keep the tip of the probe wet cap and bottle are not shown Remove these items and do not discard them They will be reused when the probe is autoclaved or stored 3 Connect the threaded end of the pH probe cable figure 8 6 to the reusable pH probe Connect the other end of the pH cable to the cylindrical end of the BNC to Two Pin Adapter figure 8 7 4 Note the AutovaxID s pH jack has a specific polarity The black jack is ground The ground side pin of the BNC to two pin adapter is distinguished by a tab of plastic marked GND Important insert the adapter s ground pin into the black jack Figures 8 7 and 8 8 show GND using green text for clarity 5 Rinse the probe and place it in pH 7 buffer that is in a container that can stand upright by itself Do not hold the pH probe at this time are purchased separately from Figure 8 5 Reusable pH Probe Part Number 103037 000 A Figure 8 6 Cable for Reusable pH Probe Part Number 401111 000 AutovaxID Operations Manual 8 12 Biovest International Inc Simplified Cultureware Usage 6 Sequentially press Menu Setup Diagnostics Sensors When the pH A D value reasonably stabilizes ensure the value is within the range of 1396 to 2702 Ignore the pH Value because the pH probe has not been and need not be calibrated Calibration is done
67. form Main Fuse For 100 VAC One 6 25 Amp 3AG Time Delayed SLO BLO Ya x 17 For 120 VAC One 5 Amp 3AG Time Delayed SLO BLO Y x 1 For 220 and 240 VAC Two 3 15 Amp Type T SLO BLO Metric 5 x 20 mm Heater Fuse 2 5 Amp Type T SLO BLO Metric 5 x 20 mm Not intended for replacement by Operator Battery Type Lithium CR1225 3 Volt Soldered to SBC Lithium CR2032 3 Volt below Serial Number 10040 Not intended for replacement by Operator Operating Temperature External Ambient 20 30 C Internal Incubator 37 C Nominal Relative Humidity 15 to 90 non condensing Storage Temperature Ambient 15 38 C Storage Relative Humidity Ambient 10 95 non condensing Shipping Temperature Ambient 30 55 C Cultureware Storage Temperature Ambient 15 38 C Cultureware Storage Relative Humidity Ambient 10 95 non condensing Cultureware Shipping Temperature Ambient 2 55 C DO NOT FREEZE Cultureware Shipping Relative Humidity Ambient 10 95 non condensing Instrument Dimensions Door Closed 19 H x 20 W x 20 D 48cm H x 51cm W x 51cm D Door Open 19 H x 20 W x 27 5 D 48cm H x 51cm W x 70cm D Instrument Weight 66 lb 30kg uncrated Pollution Degree I Maximum Altitude 2000 meters Installation Overvoltage Category II local level Gas 100 Carbon Dioxide CO2 medical grade regulated to 12
68. from the refrigerator Remove the factor bags from the rack N Oa P Ww If the Supernatant in the harvest bags is to be processed at a later time clamp and disconnect clamp and cut the line from the culturware Remove the bags and filter from the rack and store according to your facility s specifications 8 Place the emtpy rack back inside the refrigertor or store per your facility s requirements for re use 9 Unlock and remove the pump cassette 10 Unlatch the cultureware from the instrument 11 Place the cultureware pump cassette and factor bags into a biohazard bag and dispose of according to your facility s procedures for biohazardous material 12 Prior to the next cell culture run clean the instrument with a disinfectant or mild soap solution Avoid spilling or spraying directly on the touchscreen Refer to the Section 7 27 2 for further cleaning recommendations 7 24 Data Archive The purpose of data archiving is to provide a stored record of the cell culture run This record can be used as a method to review runs print logs and graph data using commercially available programs It is recommended that the log for each production run be saved prior to shutting down the AutovaxID refer to Section 10 3 1 for further information AutovaxID Operations Manual 7 59 Biovest International Inc Operating Procedures 7 25 Production Guidelines The following guidelines may be used as a general cell culture protocol
69. has been stably achieved Once the previous step is complete press the pH Calibrate button on the screen in figure 8 23 Although no pH probe is in use and without removing a sample of cell culture medium perform a calibration using the normal procedure and enter a pH value of 7 0 A warning message appears saying the pH calibration would be excessive Do you still want to use it Press YES A warning message appears on the pH calibration screen saying pH Correction Excessive Press Exit Allow Lactate Calibration to be performed automatically even though Lactate Control does not properly function when a pH probe is not being used It is critical to ensure pH is correct during the Pre inoculation phase before the inoculation procedure is performed A pH between 7 1 and 7 3 is appropriate during the growth phase for routine cell lines Adjust pH during the Pre inoculation phase to be within this range or to an experimentally determined optimal pH Once pH is stable at the desired pH the inoculation procedure may be performed Later in the run when cell densities are high a somewhat lower pH is often desirable To control pH manually the technician periodically removes a sample of medium from the IC Sample line using syringes and measures pH off line using the same techniques that apply to using standard cultureware Whenever medium pH is more than 0 05 pH units different from the desired pH value the technician should adjust either the Au
70. inoculation Once a cell culture has begun to grow efficiently continued factor addition to the IC is unnecessary Jan 12 200 Preinoculation 1 16 39 21 7 Operator Initiated Factor Addition skip Pre inoculation in progress Media Rate ml hr 20 Factor Rate ml hr 0 ph 7 05 02 20 00 Hold Exit Help Figure 7 25 Pre Inoculation Step 7 AutovaxID Operations Manual 7 35 Biovest International Inc Operating Procedures Jan 12 2007 Preinoculation 1 16 39 34 8 Factor IC line purge 0 00 01 skip Pre inoculation in progress Media Rate ml hr 20 Factor Rate ml hr 400 pH 7 05 C02 20 00 Hold Exit Help Figure 7 26 Pre Inoculation Step 8 Jan 12 200 Preinoculation 1 16 40 11 9 Factor Addition to EC 0 00 14 skip Pre inoculation in progress Media Rate ml hr 10 Factor Rate nl hr 400 pi 7 05 C02 20 00 Hold Exit Help Figure 7 27 Pre Inoculation Step 9 AutovaxID Operations Manual 7 36 Biovest International Inc Operating Procedures Jan 12 200 Preinoculation 1 16 40 40 10 Factor Addition to IC 0 00 14 skip Pre inoculation in progress Media Rate ml hr 10 Factor Rate ml hr 400 ph 7 05 C02 20 00 Hold Exit Help Figure 7 28 Pre Inoculation Step 10 Jan 12 2007 Preinoculation 1 16 40 54 11 Factor addition complete skip Press Next for I
71. need to be taken Failure to do so may result in injury or damage to the system The power cord serves as the disconnecting device for the instrument The instrument must be positioned such that access to the power plug is never blocked All equipment is for indoor use only AutovaxID Operations Manual iii Biovest International Inc About this Manual About this Manual This manual is your guide to operating the AutovaxID system lt contains all of the information that you will need for its operation including e How to set up the system e How to set up and inoculate the cultureware e How to control the growth and production conditions for the cell culture e How to set up harvesting e How to complete the production run A thorough understanding of this manual and the system is necessary before using the AutovaxID The first section of this manual describes the overall system and its theory of operation and introduces its components and user interface Section 2 explains how to unpack and setup your AutovaxID instrument for use The third section details the AutovaxID process control capabilities and functions In Sections 4 5 and 6 an operator can learn to navigate use and interpret the AutovaxID user interface screens and configure the instrument The seventh section provides details of operation including information about the cultureware It also provides a walk through of setting up the instrument for a new cell cu
72. of the run or until the cells start to establish themselves The circulation rate should then be increased by 50mL every other day up to 500mL min maximum rate and then kept at that rate for the remainder of the run 7 26 3 Cycling Guidelines The purpose of Cycling is to promote a homogeneous environment within the bioreactor via enhanced nutrient delivery and waste removal Cycling should be initiated when the cells show signs of steady metabolic growth Cycling typically begins during the first week of the cell culture s growth For sensitive cell lines or bioreactors operating in low protein conditions Cycling should be delayed until the second week after inoculation or a longer cycle time should initially be used to minimize the early loss of too many cells to the EC reservoir Questions regarding the initiation of Cycling should be directed to Biovest Account Services Cycling generally begins between the third and fourteenth day after inoculation AutovaxID Operations Manual 7 62 Biovest International Inc Operating Procedures 7 26 4 Harvesting Guidelines Starting the Harvest pump is at the discretion of the facility and or operator For secreted products that show feedback inhibition the user should determine the threshold level to monitor Samples analyzed from the flowpath will provide the necessary information in this case After the Harvest pump has been started product will be collected and the content of th
73. of the quick coupler connectors to the unlocked position 3 Disconnect the pH probe cable from the top of the cultureware 4 Disconnect the pH probe assembly from the cultureware 5 Connect the tubing adapter that was removed and saved during the installation of the pH probe at the beginning of the run to the cultureware s two quick coupler connections 6 Remove the clamps 7 Seal the biohazard bag and continue with the disposal procedure 8 Disconnect the cable from the pH probe and connect the electrical cap 9 Disassemble the pH probe assembly 10 Rinse the pH probe with tap water and blot it dry 11 Fill the pH probe storage bottle with fresh pH storage solution 12 Insert the pH probe in the storage bottle and ensure it does not leak 13 Store the pH probe and cable in a safe location for later use Disposable pH Probe 1 Before removing the cultureware from the AutovaxID leave the disposable pH probe connected to the cultureware but remove it from its reusable holder 2 Store the reusable holder in a safe location for later use 3 Continue with the disposal procedure AutovaxID Operations Manual 8 26 Biovest International Inc Troubleshooting 9 Troubleshooting This section can be used to assist operators with troubleshooting their AutovaxID system and determining if assistance from Biovest Account Services is necessary Be extremely careful when troubleshooting and closely follow the instruc
74. that is automatically loaded during run setup See Section 5 4 It also allows an authorized user to define some initial pump settings that are used during GROWTH and FILL FLUSH Finally the harvest line volume may also be defined by an authorized user Jan 16 2007 hun Conf ig 1 11 37 2 E futo Load Run Settings Initial Circ Rate 1200 Nane Defaults Initial Media Rate 20 2 Fill Flush Media Rate 20 2 Harvest Line Volume 70 nl Hold Exit Back Help Figure 6 6 Run Configuration 1 Screen AutovaxID Operations Manual 6 8 Biovest International Inc Software Setup amp Configuration 6 4 1 Auto Load Run Settings This field allows an authorized user to enable or disable checked or unchecked respectively automatic run settings During run setup a series of run setting screens allow an authorized user to define factor addition volumes cycle times pump controllers pH and temperature setpoints and process alarms see pages 5 21 through 5 47 These run settings can be modified and saved to a recipe for future use or the factory default profile can be used Any of these recipes can be automatically loaded during run setup by enabling this feature To define the run settings profile that should be automatically loaded during run setup touch the Name field and select a recipe from the list of those that have been previously saved Regardless of the run settings profile that is defined in the N
75. the Used for troubleshooting EC air bag is vented ar Aan LE 44 54 Jan mees 54 Override Cycle State Cycle Off Press Cycle Stable ME Low IC Pr Drain Figure 5 50 Cycle States Overrides AutovaxID Operations Manual 5 65 Biovest International Inc Primary User Interface Screens 5 12 4 Gas Overrides During normal operation the pH Control screen allows an operator to set the pH setpoint and perform a calibration of the pH probe see Section 5 1 7 However if the pH needs to be adjusted quickly by increasing or decreasing the amount of CO to the system an authorized user can override the automatic gassing system The percent CO can range from 0 to 20 a higher percent CO will decrease the pH a lower percent CO will increase the pH Alternatively the air flow rate can be increased to remove CO from the system The overrides for CO and air flow rate are mutually exclusive only one can be set at a time To adjust either the operator should touch the appropriate field on the Overrides screen Gassing overrides should only be used in special circumstances such as failure of the pH probe as large fluctuations in pH can be detrimental to a cell cultures When overriding the gassing system the operator should remain at the instrument until the pH is within the desired range Once pH is within range the operator should clear the override to resume normal gassing operation 5 12 5 Factor Valve Overri
76. the harvest pump is on Delay Until The elapsed time before the time cycle controller will be activated Total Cycle Time The total duration of the cycle which is equal to the on time plus the off time On Time The duration that the pump is on and running at the desired rate during the cycle AutovaxID Operations Manual 5 40 Biovest International Inc Primary User Interface Screens NOTE The On Time must be less than or equal to Total Cycle Time For example if the following entries are made Auto Rate 5 mL hr Delay Until 15 00 00 days hr min Total Cycle Time 0 01 30 days hr min On Time 0 00 30 days hr min then the pump will not turn on until the fifteenth day of GROWTH When it does turn on it will run at a rate of 5 mL hr for thirty minutes and then shut off again for one hour the remaining duration of the Total Cycle Time The pump will continue turning on for thirty minutes and then off for one hour for the remainder of the run The Total Cycle Time and the On Time must be greater than zero If either is set to zero then the harvest pump will never turn on even if the elapsed time is surpassed Furthermore setting the Total Cycle Time and the On Time to the same non zero value will result in the harvest pump running continuously after the elapsed time has surpassed 5 5 12 Circulation Controller The circulation pump circulates media in the IC to supply fresh oxygenated media to the cells in the
77. the system a gas exchange cartridge GEX perfuses an air and CO gas mixture into the media oxygenating it while maintaining the prescribed pH setpoint The process control computer is the brain of the AutovaxID It monitors and controls the various parts of the system ensuring that the environmental conditions are maintained within their optimal prescribed ranges The cell secreted products remain in a circuit separate from the metabolic waste products and may be harvested by activating a separate pump With these general concepts in mind the following detailed description will give you an understanding of the operational aspects of the AutovaxID AutovaxID Operations Manual 1 1 Biovest International Inc Introduction 1 1 1 Bioreactor Technology The core of all hollow fiber based mammalian cell culture systems is the hollow fiber bioreactor BRX The bioreactor is a plastic cylindrical housing containing thousands of hollow fibers that are encapsulated at each end of the tube thereby isolating the interior of the fibers from the exterior fibers expect through their pores The volume within the bioreactor outside of the hollow fibers is known as the extracapillary space ECS the volume inside of the hollow fibers is known as the intracapillary space ICS see Figure 1 1 The ECS and ICS are isolated from one another inside the bioreactor although small molecules can move through the pores of the fibers Cells a
78. these differing bioreactor ECS volumes e The triple bioreactor version can be used with one to three Factor containers and one to three Harvest containers Using fewer than three Factor or Harvest containers requires the user to build sterilize and attach tubing manifolds to deliver Factor and or collect Harvest into only a single container each Manifolding these containers is applicable when all three bioreactors are inoculated with the same cell line e Operating the triple bioreactor version requires the technician to purchase and use a standalone Masterflex pump motor s and pump heads The AutovaxID s built in Factor and Harvest pumps cannot be used to pump the triple bioreactor cultureware s three Factor and three Harvest tubing lines AutovaxID Operations Manual 8 1 Biovest International Inc Simplified Cultureware Usage The remainder of the following information in Section 8 generally does not distinguish between using the single bioreactor and triple bioreactor versions of the simplified cultureware with exceptions where necessary 8 3 Simplified Cultureware Retained Features Simplified cultureware retain the most important features and capabilities of the AutovaxID such as Pre assembled and pre sterilized cultureware for single use disposability pH control Lactate control when a pH probe is being used Incubator temperature control Independently controlled media factor and harvest pumps Minimal tech
79. to the Advanced Overrides screen for advanced overrides to the gassing system cycling system pumps and valves See Section 5 for a detailed description of these screens AutovaxID Operations Manual 4 1 Biovest International Inc Introduction to the User Interface 4 1 Startup When the power switch is turned on the internal computer boots and automatically loads the AutovaxID User Interface Application After the software loads the Monitor screen is displayed see Figure 4 1 Assuming that the instrument is not being turned back on after a power loss the Monitor screen should show that it is in the IDLE state NOTE Before the Monitor screen appears the operating system loads The information that quickly scrolls on the screen while the operating system is loading during startup is not intended to be used by operators Immediately after the boot sequence a Touch Calibration screen appears see Figure 2 9 The touchscreen will be calibrated at the factory and need not be re calibrated at every startup The Touch Calibration screen will automatically disappear if it is not touched See Section 6 7 Touch Calibration Screen for more information dan 23 2007 INSTA NAME 13 13 22 Elapsed Tine 0 00 00 State IDLE OH Cycle Off Air 500 ml min Lactate 0 00 Ges Media Temperature 37 0 Refrigerator Temp 4 0 Harvest Circulatio 100 o o a D Figure 4 1 Monitor Screen Factor E
80. volume the state of the outflow container alarm and the alarm threshold Authorized users can enable and disable the outflow container alarm checked and unchecked respectively and enter an alarm threshold between 0 and 99 percent The alarm threshold defines when an alarm will be issued For example if the threshold value is 10 then when the AutovaxID software has calculated that the outflow container is within 10 of being full an alarm will be issued to notify the operator that the waste container should be changed e Default Outflow Container Volume 20L e Outflow Container Range 0 to 999 99L in increments of 0 1L e Default Alarm Threshold 10 e Default Alarm State Enabled Checked CAUTION The design of the outflow pump requires that the waste container is adequately vented The outflow pump moves both media and air out of the cultureware so the user must vent the waste container either into a bleach waste bottle or via a sterilizing grade vent filter If a filter is used ensure that it does not become wetted WARNING Do not use a glass vessel with an exhaust filter if the filter becomes wetted the vessel will become pressurized and could pose a hazard to the operator The only allowable outflow pump controller type is the built in controller indicated as Outflow in the Controller Type field AutovaxID Operations Manual 5 26 Biovest International Inc Primary User Interface Screens Jan 24 200 Qutfl
81. 0 00 45 days hr min then the pump will not turn on until the 3 day of GROWTH When it does turn on it will run at a rate of 100 mL hr for 45 minutes and then shut off again for 1 hour and 15 minutes the remaining duration of the Total Cycle Time The pump will continue turning on for 45 minutes and then off for 1 hour and 15 minutes for the remainder of the run AutovaxID Operations Manual 5 36 Biovest International Inc Primary User Interface Screens 5 5 9 The Total Cycle Time and the On Time must be greater than zero If either is set to zero then the factor pump will never turn on even if the elapsed time is surpassed Furthermore setting the Total Cycle Time and the On Time to the same non zero value will result in the factor pump running continuously after the elapsed time has surpassed Harvest Controller The harvest pump moves cell secreted products from the flowpath to storage bags in the refrigerator for post run processing Pressing the Harvest button on the Run Settings 2 screen brings up the Harvest Controller screen see Figure 5 31 This screen shows the harvest pump controller type the harvest container volume the state of the harvest container alarm and the alarm threshold Authorized users can enable and disable the harvest container alarm checked and unchecked respectively and enter an alarm threshold between 0 and 99 percent The alarm threshold defines when an alarm will be issued For example
82. 1 psi Requirements CO Consumption 35 mL min 20 GEX blend nominal Air Supply On board air compressor Offgas As needed slight vacuum AutovaxID Operations Manual 11 2 Biovest International Inc Specifications Approximate Standard cultureware with single 1 1m bioreactor 600320 211 Flowpath Total System Volume 600 mL Volumes EC Volume 235 mL mid cycle Standard cultureware with single 2 1m bioreactor 600320 221 Total System Volume 670 mL EC Volume 305 mL mid cycle Legacy cultureware with single 1 1m bioreactor 600290 21 1 Total System Volume 600 mL EC Volume 235 mL mid cycle Legacy cultureware with single 2 1m bioreactor 600290 221 Total System Volume 670 mL EC Volume 305 mL mid cycle Simplified cultureware with single 1 1m bioreactor 600307 211 Total System Volume 400 mL EC Volume 90 mL Simplified cultureware with three 0 5m bioreactors 600308 205 Total System Volume 460 mL EC Volume 55 mL per bioreactor Factor Bag Assembly Liquid Volume Fill to 3L maximum Harvest Bag Assembly Liquid Volume Fill to 3L maximum Peristaltic Pump Rates Outflow Pump 0 480 mL hr 1 mL hr increment Media Pump 0 400 mL hr 1 mL hr increment Factor Pump 0 400 mL hr 1 mL hr increment Harvest Pump 0 400 mL hr 1 mL hr increment Circulation Pump Rate Off 200 500 mL min 1 mL min increment pH Measurement Range 6 00 8 00 theoretical
83. 127 0 0 1 8080 0 x survey 104183 gt E Ry Pace Tools e DB Date Time pH Gas ICP ECP Lact State MediaT RefT AirT CO2 Outflow Media Factor Harvest Circ CabT 17 Jan 2007 13 42 23 Log start SW 300172 000 Rev 56 Instrument INSTR NAME SN 17 Jan 2007 13 23 Elapsed Time is 0 06 10 day hr min 17 Jan 2007 13 123 Process Growth state started 17 Jan 2007 13 23 Elapsed Time reset to 0 00 00 17 Jan 2007 13 42 23 Valve Factor to 3 EC 17 Jan 2007 13 23 Pump Media to 20 ml hr 17 Jan 2007 13 23 Pump Circulation to 200 ml min 17 Jan 2007 13 42 23 7 10 0 0 o o 0 00 0 37 0 17 Jan 2007 13 42 Pump Outflow to 100 ml hr 17 Jan 2007 13 27 Valve HarvestEC to 3 Brx 17 Jan 2007 13 g Alarm Pump head not latched 17 Jan 2007 13 42 Alarm acknowledged 17 Jan 2007 13 42 58 Alarm cleared Pump head not latched 17 Jan 2007 13 47 7 10 0 0 o 0 0 00 0 37 0 17 Jan 2007 13 51 Alarm Pump head not latched 17 Jan 2007 13 52 0 0 o D 0 00 14 37 0 17 Jan 2007 13 57 7 0 0 o o 0 00 14 37 0 17 Jan 2007 13 2 acknowledged 17 Jan 2007 13 59 cleared Pump head not latched 17 Jan 2007 14 02 e 0 0 o 0 0 0 37 17 Jan 2007 14 07 37 17 Jan 2007 14 37 17 Jan 2007 14 17 37 17 Jan 2007 14 37 17 Jan 2007 14 27 37 17 Jan 2007 14 32 37 17 Jan 2007 14 37 37 17 Jan 2007 14 37 17 Jan 2007 37 17 Jan 2007 37 17 Jan 2007 STe 17 Jan 2007 37 17 Jan 2007 37 17 Jan 2007 37 Pone 000000 0000 000000000000000 Ppooecocco
84. 3 screen Note for the Factor Container Hold hit Back Next clamp for simplified cultureware open the Factor 1 clamp and Factor 2 and Figure 8 22 Setup Run 2 Screen AutovaxID Operations Manual 8 22 Biovest International Inc Simplified Cultureware Usage 17 18 19 20 Factor 3 clamps in simplified cultureware with three bioreactors Also open the Harvest container clamp s Harvest 1 Harvest 2 and Harvest 3 as appropriate which is not listed in the on screen instructions Allow the automated Pressure Test to complete which takes only a few minutes Ensure the test passes before leaving the AutovaxID unattended for a long period of time If the test does not pass corrective action is necessary to fix the problem Fill Flush must not begin until the pressure test passes Feb 25 2007 Preinoculation 1 23 05 05 2b Waiting for pH Calibration skip Perfor the pH calibration Media Rate nl hr 20 Factor Rate nl hr 0 mf 002 120 0 gi Hold dit Help Figure 8 23 Setup Run 2 Screen Fill Flush automatically begins after a successful pressure test Allow the Fill Flush procedure to complete and the Preinoculation 1 screen to appear with the instruction Perform the pH calibration figure 8 23 Go to either step 20 or 21 as appropriate When using manual pH control no pH probe is installed e Press Exit until the Monitor screen appears Now press Menu Advanced an
85. 7 34 CAUTION Do not leave the instrument on the Inoculation Step 2 screen indefinitely because at this step process control has not yet been started The cell culture environment will not be controlled by the instrument to the user defined parameters until the instrument is in GROWTH After pressing the Next button the operator will automatically be returned to the Monitor screen The instrument State will indicate GROWTH Jan 12 200 Start Growth 16 46 16 Grouth Mode is about to begin Push NEXT to initiate growth mode Hold Next Help Figure 7 34 Start Growth Screen 7 8 Cycling The Cycling parameters Delay and Cycle Time are dependent upon the characteristics of the cell line and should be decided upon by the facility or operator Cycling is typically initiated once the cells show signs of steady metabolic growth Cycling can be delayed for slow growing cell lines or if operating under low protein conditions Cycling is typically delayed not in use post inoculation to minimize the early loss of cells into the EC reservoir After the start of GROWTH Cycling will be delayed for the operator configured time see AutovaxID Operations Manual 7 42 Biovest International Inc Operating Procedures Sections 5 4 3 5 4 5 During GROWTH the cycling parameters can be adjusted by touching the Cycle field on the Monitor screen On the Cycle Detail Screen see Figure 5 8 the Cycle
86. 7 60 7 26 Optimization EI EE 7 61 726 1 IMOCUIAON GUIDCNINES EE 7 61 7 26 2 Circulation Rate Guidelmes AAA 7 62 7 26 3 Cycling Guideline S es aeea aae ape REEERE E EERE EEA EEE EA A 7 62 AutovaxID Operations Manual x Biovest International Inc Table of Contents 7 26 4 Harvesting Guidelines AAA 7 63 120 9 Factor hoquiremenis cota E ee 7 63 126 0 S r m Free MOI oi api 7 64 721 Mate dice sitas En O OA EOS 7 65 7 271 Cultureware Storage nr cit 7 65 7 27 2 Exterior Cleaning A O OS 7 65 7 27 3 Inlet Fan Filter Cleaning sd a ia ds 7 65 7 27 4 Refrigerator DefroStidO costaria carlinas 7 66 PLASMA MON cris EE 7 66 7 27 6 Preventivo Maintenance cui a 7 66 8 Simplified Gultureware USAqe ooomiocicnaicassi caricia iba 8 1 8 1 Simplified Cultureware Descriotion ENNEN 8 1 8 2 Simplified Cultureware Versions vicio ii 8 1 8 3 Simplified Cultureware Retained Features AAA 8 2 8 4 Simplified Cultureware Removed Features occcccnonocccccccncccnonnnanancnnnnncnnnnnnnnnnnccnnnnnnnnnnnnnnnnno 8 2 8 5 Simplified Cultureware Flowpath Gchemate A 8 4 8 5 1 Intracapillary Circuit Components tum e 8 5 8 5 2 Extracapillary Circuit Component AEN EEN 8 6 8 6 Pumping Factor and Harvest cuco eegen deg EAEe Ee Eeer Ee Ee dee e 8 8 9 2 COMO POMS ed i daa dae ae 8 9 8 7 1 Manual el E nt EE 8 9 8 7 2 Automatic pH Control with Reusable pH Probe 8 9 8 7 3 Automatic pH Control with Pre Sterilized pH Probe ooooooccccnccccccccconcccccncc
87. AA AAAS MTS Ar A4 A Enter ADMIN s password rones ej y jajaj nj el bel lt Cancel Enter d Figure 6 13 Primary Administrator Password This primary administrator account is now the first of twenty available accounts Please record the username for future reference and keep the password confidential AutovaxID Operations Manual 6 17 Biovest International Inc Software Setup amp Configuration 6 6 2 User Account Setup by an Administrator Once the primary administrator account is setup the primary administrator can create additional usernames and passwords for additional users with either normal or administrator rights A regular non administrator account is intended for the typical technician responsible for operating the AutovaxID An administrator account is an account that may be used to manage all accounts and all system settings To assist with keeping all user accounts secure the AutovaxID software requires that all users assign their own password when they log on for the first time To create a new user log in as the primary Administrator and then go to the User Accounts screen see Figure 6 14 by pressing the Passwords button on the Setup screen from the Main Menu Monitor gt Menu gt Setup gt Passwords If you are not logged on as an administrator you will be prompted to do so Jan 26 200 deer Accounts 11 43 19 Admin ADMINISTRATOR Al e Delete Account Chang
88. As prompted when the container change is complete press the Done button to resume the pumps you will be returned to the pump detail screen Press the Exit button to return to the Monitor screen After the container change has been completed the volume indicator on the Monitor screen will revert to full or empty as is appropriate for the changed container The container graphic on the associated pump Detail screen will also change appropriately Note If container volume alarms are enabled see the pump Controller screens do not change containers without using this procedure Changing pump containers without following the container changing procedure will prevent the estimated current container volume from resetting will result in continued container alarms and will cause erroneous future volume estimations by the AutovaxID software AutovaxID Operations Manual 7 54 Biovest International Inc Operating Procedures 7 18 7 19 7 20 Container Capacity Change An authorized user may change the capacity of a pump container Outflow Media Factor or Harvest by pressing the Settings button on the appropriate pump Detail screen to bring up the associated pump Controller screen The pump Detail screen can be accessed by pressing the appropriate pump field on the Monitor screen On the pump Controller screen press the Container Volume field to bring up an on screen keypad Alternatively you may touch the word Capacity on t
89. AutovaxID Operations Manual 7 22 Biovest International Inc Operating Procedures dan 12 2007 Pill Flush 16 22 14 1 Positioning Val ves skip Fill Flush in progress Cycle Off Media Rate ml hr 0 C02 Sensor 20 00 y Circulation Rate nlmin 0 Media Temperature Control Off Outflow Rate nl hr O Refrigerator Temperature Control 4 0 Hold Exit Help Figure 7 7 Fill Flush Step 1 Jan 12 200 Fill Alush 16 22 42 2 Evacuating EC Chamber 0 00 01 skip Fill Flush in progress Cycle Drain Media Rate ml hr 0 C02 Sensor 20 00 y Circulation Rate ml min 0 Media Tenperature Control Off Qutflow Rate nl Ar 0 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 8 Fill Flush Step 2 AutovaxID Operations Manual 7 23 Biovest International Inc Operating Procedures dan 12 200 Fill Flush 16 24 06 3 Evacuating EC Complete IN Cycle Off Media Rate ml hr 0 C02 Sensor 20 00 y Circulation Rate ml min 0 Media Tenperature Control Off Qutflow Rate nl Ar O Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 9 Fill Flush Step 3 Jan 12 200 Fill Alush 16 25 44 4 Initial IC Fill 0 00 11 skip Fill Flush in progress Cycle Off Media Rate ml hr 480 C02 Sensor 20 00 y Circulation Rate nl min 0 Media Temperature Control Of
90. C Sample port Adjust the Harvest pump rate based upon the productivity of the cell line e Sample the harvest bag post filter for accumulated production using the Supernatant Sample port Ensure that the harvest bag is mixed prior to withdrawing a sample for quantization e Withdraw cells from the bioreactor prior to shut down for End of Production Cell Bank if required 7 13 IC Sampling Procedure An IC sample can be taken during a production run to assess the metabolic activity and cell growth To take a sample three sterile syringes will be required two 10cc syringes for flushing the IC Sample line and one 5cc for the IC sample All syringes must have male Luers A larger syringe may be used for the IC sample if required by the facility s assay procedure Aseptically attach the 10cc syringe to the IC Sample port Unclamp the IC Sample line and remove 5 to 10 mL to flush the line Close the IC Sample clamp Ze a Ji Aseptically disconnect the 10cc syringe and replace it with the syringe selected for the IC sample 5cc The 10cc flush syringe may be discarded per your facility s requirements 5 Unclamp the IC Sample line again and remove the necessary sample volume for off line analysis Close the IC Sample clamp 7 Aseptically disconnect the IC Sample syringe and replace it with a new 10cc syringe Leave the IC Sample clamp closed This 10cc syringe will protect the sterility of the IC Sample port and can be used to fl
91. D pH control system is based on bicarbonate buffering chemistry The media must contain 2 2 to 3 5 g L of Sodium bicarbonate Concentrations of Sodium bicarbonate outside of this range may prevent the AutovaxID from properly controlling pH Other buffers such as HEPES or Tris must NOT be used in the media or the factor because they can interfere with the Instrument e ability to control pH using a CO gas blend The concentration of bicarbonate should be increased as the concentration of glucose increases in the media formulation Higher incoming media pH will reduce the attainable initial pH setpoint for the GROWTH phase but will increase the system s ability to maintain pH during the production phase which is often of greater concern AutovaxID Operations Manual 7 3 Biovest International Inc Operating Procedures 7 2 3 7 2 4 7 2 5 NOTE Biovest recommends a high glucose media 3 0 4 5 g L for use in the AutovaxID to minimize the volume of media required while supplying adequate glucose to the cell culture Excessive media usage may result from a glucose level lower than 3 0 g L L glutamine is an especially important metabolite Biovest recommends a minimum concentration of 4 mM of L glutamine Media Sterilization As key components of an aseptic process all cell culture media and supplements must be sterilized before being pumped into the system ECS media is intended to be sterile filtered directly into the provided fa
92. Delay Time EC Rise Time and EC Fall Time can all be independently adjusted by an authorized user 7 9 Mid Run pH Calibrations Authorized users should perform additional pH calibrations during a cell culture run The pH probe will likely drift during the first week after FILL FLUSH A pH calibration should be performed if the difference between the current value and the off line value exceeds 0 05 If using the lactate control the difference should not exceed 0 03 After a pH calibration allow the reading to stabilize and then retest the pH using a new IC sample until the instrument agrees with the off line measurement Prior to starting the pH calibration the operator must have three sterile syringes one 10cc syringe that will be used for flushing the sample line one 5cc syringe that will be used for drawing the sample and another 10cc syringe that will remain connected to the IC Sample port protect the sterility of the port It may also be used for flushing the IC Sample line prior to taking the next pH calibration sample All three syringes must have male Leur fittings If a 10cc syringe was left on the IC Sample port after a previous pH calibration then only one 10cc and one 5cc syringe will be required To perform a pH calibration during a cell culture run 1 Press the pH field on the Monitor screen 2 Press the Calibrate pH button on the pH Control screen see Figure 7 35 to bring up the pH Calibration Menu see Figure 7 36 3
93. Down move toward the end of the log by one record e PG DN move toward the end of the log by one page e END move to the end of the log Press the Exit button to return to the Alarm Log screen AutovaxID Operations Manual 5 53 Biovest International Inc Primary User Interface Screens 5 9 5 Log View Options Any user may perform advanced log functions using the features on the Log View Options screen see Figure 5 42 This screen is accessed by pressing the Options button on the Log screen Jan 25 2007 Log View Options 10 37 15 EA Status lines on YA Event lines on Fa Alarm lines on A Search ALARM y Loy Select Current g urren deity Hold Exit Help Figure 5 42 Log View Options Screen Advanced log functions include e Filtering the displayed log based upon the log entry type e Searching the log based upon a user defined search term e Selecting the log current or historical to be viewed or searched e Verifying the selected log current or historical Any user may select a specific log for viewing by touching the Log Select field and then selecting the appropriate log from the on screen list Up to eight historical logs can be stored and viewed on the AutovaxID Any advanced functions performed using the features on the Log View Options screen will be performed on the log shown in this field Pressing the Verify button on the Log View Options screen will perform a verification of the c
94. Help provides context sensitive help information for each screen See Section 4 4 for details Flowpath Status The remainder of the Monitor screen displays the current conditions of the flowpath including e pH e Gas blend displayed as either mL min of air or percent CO3 e Calculated lactate concentration e Cycle state see Table 5 1 e EC and IC pressures e Circulating media and refrigerator temperatures e The Outflow Media Factor and Harvest pump rates in mL hr e Calculated volume mL in the Outflow Media Factor and Harvest containers e The programmed circulation pump rate in mL min The Outflow Media Factor and Harvest pump fields are listed in the same order left to right as the pumps are mounted in the instrument as viewed from the front The pump tubing also exits the cultureware and is routed through the pump cassette in the same order left to right All pump fields show the current pump rate except the circulation pump the Circulation field shows only the programmed pump rate as defined by the circulation pump controller The actual measured circulation rate can be observed by touching the circulation pump field to bring up the Circulation Detail screen see Figure 5 15 Similarly touching any pump field on the Monitor screen will bring up an associated detail screen AutovaxID Operations Manual 5 7 Biovest International Inc Primary User Interface Screens 5 1 7 pH Control Screen On the Mo
95. If the value moves back within range the corresponding sensor or parameter values reverts to normal black characters The audible and visual alarms will not be silenced until they are acknowledged and or cleared by an authorized user All alarms should be investigated and corrective action taken See Section 5 9 Alarm Log Screen and Section 9 Troubleshooting for more information AutovaxID Operations Manual Biovest International Inc Introduction to the Process Control Software 3 7 Data Archive The AutovaxID software saves log data to the internal control computer s CompactFlash memory In addition to saving the run data for the current on going production run the software also retains eight historical production logs Any of the logs may be viewed on the instrument or they may be downloaded using remote access see Section 10 Remote Access Downloading the log has several advantages e Provides the security of archiving the production run By using the Download Text function see Section 10 3 1 Saving a Log the entire production log from initial setup to run termination or the current elapsed time for on going runs is downloaded from the AutovaxID s CompactFlash storage media An electronic archive is stored in a fraction of the space required by a paper log e Provides a backup record in the event of computer failure If certain aspects of process control miscellaneous log entries etc are not supplementally docume
96. Incubator Temperature Range 15 45 C subject to ambient conditions Refrigerator Temperature Range 4 25 C not exceeding ambient conditions EC Level Sensor Type Non contact convergent photoelectric IC EC Pressure Sensor Type Piezoresistive Silicon IC EC Pressure Sensor Range 0 375 mmHg 0 7 25 psi CO Sensor Range 0 20 AutovaxID Operations Manual 11 3 Biovest International Inc Specifications AutovaxID Flowpath Schematic For Standard Cultureware Temperature Probe EC Reservoir Permeable hollow fibers permit exchange between IC and EC of small molecules lt nominally 10 KDa Media flows into and out of EC Reservoir when Cycling is active A Bioreactor Harvest EC Valve Open when harvesting Harvest EC Valve Open when Cycling Harvest EC Valve Closed when Cycling and harvesting pH Sensor Factor Valve Normally open to EC Factor Valve Normally closed to IC AutovaxID Incubator Circulation Pump uy Check Valve lt 4 Male Luer O O O Pump X Ratchet Clamp gt j T a 1 1 a lt Flow Direction PI Slide Clamp E a a Zo cB 0 24 Sterilizing Filter Automated Clamp n 3 A 2 28 Q Tubing Connection y Luer Activated Valve 3 g Supernatant lt lt Supernatant Sample 3L Extracapillary Circuit Factor Bag Set b BIOVEST nom INTERNATIONAL
97. LE CAUSE ACTION The lactate value too high The maximum allowed value of the Media pump as set in the pump Controller has been reached Increase the maximum Media pump rate on the Media pump Controller screen The pH is out of calibration Perform a pH calibration see Section 7 9 The lactate level is out of calibration Calibrate the pH and lactate Variations in media can result in differences in lactate calibration See Section 7 11 for more detail The instrument is out of media Attach a new container of media to the Media line Contact Biovest Account Services The lactate level is too low The minimum allowed value of the Media pump as set in by the pump Controller is too high Decrease the minimum Media pump rate in the Media pump Controller screen The pH is out of calibration Perform a pH calibration see Section 7 9 The lactate level is out of calibration Calibrate the pH and lactate Variations in media can result in differences in lactate calibration See section 7 11 for more detail Contact Biovest Account Services The outflow or harvest bag is full The calculated volume pumped has reached the alarm point Exchange the full bag with an empty one following the procedure for changing containers see Section 7 17 The container capacity is not set correctly Adjust the container capacity from the pump Controller screen
98. NOT REMOVE ANY PANELS FROM THE INSTRUMENT THERE ARE NO USER SERVICEABLE PARTS INSIDE REFER SERVICING TO QUALIFIED SERVICE PERSONNEL WARNING If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired WARNING A properly grounded main supply must be used for safe operation of the equipment Connect to an appropriately grounded outlet Consult local electrical codes CAUTION Wherever a caution statement is used the documentation needs to be consulted in order to find out the nature of the potential hazard and any action which may need to be taken Failure to do so may result in injury or damage to the system NOTE The power cord serves as the disconnecting device for the instrument The instrument must be positioned such that access to the power plug is never blocked NOTE All equipment is for indoor use only Biovest International services all components of the AutovaxID Facilities should add the AutovaxID to their list of equipment that require routine maintenance AutovaxID Operations Manual 11 1 Biovest International Inc Specifications Electrical 600 Watts Max Requirements 100 VAC 6 Amp Max 60 Hz 120 VAC 5 Amp Max 60 Hz 230 VAC 2 6 AMP Max 50 Hz 240 VAC 2 5 AMP Max 50 Hz Note Supply Voltage not to exceed 10 of nominal voltage Recommended Uninterruptible Power Supply UPS Backup On line 600 VA nominal sine wave
99. NOT set from the factory Instead the primary administrator account username can be any of your choosing You may opt to use ADMINISTRATOR as the username or you may use any other username as you prefer In Figure 6 12 the primary administrator username is typed as ADMIN After you type the new username of your choosing using the on screen keyboard press the Enter button J AA MAAS I 1 ir jA 4 New Usernane ADMIN Figure 6 12 Primary Administrator Username AutovaxID Operations Manual 6 16 Biovest International Inc Software Setup amp Configuration A password for this newly created administrator account must be entered and confirmed see Figure 6 13 Type a valid password that is at least six characters long or the minimum password length as defined on the System Features screen The password may be any combination of the letters numbers and punctuation marks that are available on the on screen keyboard up to 34 characters long As the password is typed it is kept secret by displaying only the asterisk character in the entry field After typing the password press the Enter button You will then be prompted to type the password again as confirmation Press the Enter button when you are finished CAUTION Do not lose the primary administrator username and password If the username and password are lost the AutovaxID software will need to be re installed by Biovest Account Services I
100. Note when the fuse holder is installed in the single fuse position use a screwdriver as shown to gently pry it out when the fuse holder is installed in the dual fuse position it will normally release as soon as the door is opened Figure 2 5 Remove Fuse Holder AutovaxID Operations Manual 2 6 Biovest International Inc Instrument Setup 7 Install one 1 AG fuse or two 2 metric fuses see Figure 2 6 and Figure 2 7 Note Install fuses on one side only do NOT install both AG and metric fuses at the same time Figure 2 7 230 240 Voltage Fuse Arrangement 8 Re install the fuse holder into the power entry module orienting the fuse holder so that the fuses are inserted into the module first 9 Replace the cover snap closed 10 Verify that indicator pin shows the desired voltage 2 7 Gas Connections 2 7 1 CO Supply The AutovaxID is equipped with an internal gas mixing system that requires an external source of CO gas 100 CO medical grade Set the external CO source to 12 1psi Connect the CO supply to the CO port on the rear of the instrument see Figure 2 1 using one of the supplied quick disconnect fittings 1 8 ID non gas permeable tubing such as Tygon is recommended AutovaxID Operations Manual 2 7 Biovest International Inc Instrument Setup 2 7 2 Offgas Collection All of the waste gasses from the flowpath are exhausted through the Off Gas port on the rear of the ins
101. On the Monitor screen press the Harvest field or the surrounding area to bring up the Harvest Detail screen This screen shows the harvest pump position the harvest pump rate in mL hr the harvest container capacity and the estimated volume of fluid currently in the harvest container see Figure 5 14 The pump illustration also shows the direction of pump rotation see the arrow above the pump and the direction of fluid flow see the arrow next to the pump tubing Jan 24 200 Harvest Detail 14 30 04 a DI Ze Harvest 7 ml h 3000 ml Capacity 64 nl Current Volune Change Container settings Hold Exit Help Figure 5 14 Harvest Detail Screen The harvest pump rate may be automatically controlled by the AutovaxID process control software or it may be controlled manually by an authorized user The controller type is defined on the Harvest Controller Screen see Figure 5 31 which can be accessed by pressing the Settings button The harvest controller is typically set during run setup see Section 5 5 See Section 5 5 9 for details about each of the harvest pump controller types If the harvest pump is being automatically controlled by the process control software it may be overridden only by an authorized user using an advanced override see Section 5 12 2 An authorized user may change the harvest container by pressing the Change Container button see Section 7 17 An authorized user may also change the
102. Pump Schedule Screen AutovaxID Operations Manual 5 29 Biovest International Inc Primary User Interface Screens The event that is currently in use is highlighted see the first event in Figure 5 25 If no events are currently active because not enough time has yet elapsed the highlighted area will appear after the last entry see Figure 5 28 An authorized user may enter a new value for any Media rate at any time during a run The Elapsed Time may also be changed for any event by an authorized user So long as the schedule has not been advanced or delayed The Advance and Back buttons are used to shift the time of all events backward or forward These buttons are intended to be used if the cell culture is growing faster or slower than originally expected The time shift is displayed at the bottom of the screen To skip ahead to the next scheduled event press the Advance button to revert to a previous event press the Back button To remove any advance or delay touch the time shift field at the bottom of the screen to the left of the Advance button and enter zero 0 using the on screen keypad 5 5 4 Media Pump Lactate Controller Setup If the media controller type is set to Lactate then pressing the Controller Setup button on the Media Controller screen will bring up the Media Control screen see Figure 5 26 This screen allows the media pump control to be configured based upon a lactate setpoint dan 24 2007 M
103. Setting 1 screen and set the Factor Add IC volume to 0 mL because simplified cultureware do not support this feature Adding factor to the IC can still be done however The technician determines the desired volume of factor to be added to the IC based on the total IC Circuit volume listed in the Specifications section in the Operation Manual This volume is loaded into a syringe and injected into the IC Sample line The AutovaxID will prompt the technician to add factor at the appropriate time Note Observe the volume in the IC Reservoir as the factor is being injected Be sure to inject the factor slowly enough that the IC Reservoir does not overfill Reminder factor addition to the IC is optional and unnecessary for many applications AutovaxID Operations Manual 8 21 Biovest International Inc Simplified Cultureware Usage 10 11 12 13 14 15 16 In the Run Settings 1 screen set the Cycle Delay elapsed time to 0 00 00 and the Cycle Rise and Cycle Fall times to 15 minutes These settings will remain in use and Cycle State will display Rise during the entire production run Although simplified cultureware do not have the EC Cycling components these Cycling settings assist the Fill Flush process and then prevent air accumulation in the IC circuit when using simplified cultureware In the Alarm Settings 2 screen disable the Cycle Time alarm since this alarm is not relevant to using simplified culturewa
104. ack 1 16 ID Sterile Tubing Extension Line Silicone 72 183 cm Single Use Used to extend Outflow Media Factor Harvest IC Sample Inoculate Supernatant Sample and Factor Prime tubing lines to reach the laminar flow hood AutovaxID Operations Manual 13 1 Biovest International Inc Parts List 600101 072 600294 000 600318 000 1 8 ID Sterile Tubing Extension Line Silicone 72 183 cm Single Use Used to 1 extend Cell Removal tubing line to reach the laminar flow hood and 2 refill 600320 XXX cultureware Factor bags Supernatant Tubing Extension Line Sterile 24 61 cm Used to extend Supernatant tubing line to reach laminar flow hood Luer Activated Valve LAV Kit 15 each of Sterile and Individually Packaged LAVs e Polypropylene Luer Dust Covers Sterilizable Reusable pH Probe Items when using simplified cultureware only 103037 000 103047 000 103150 000 401111 000 600298 000 600300 000 600301 000 Reusable pH Probe Autoclave Sterilizable pH Storage Solution BNC to Two Pin Adapter pH Probe Cable pH Probe Case Autoclave Tubing Lines Single Use Set of O rings for 600298 000 Single Use Disposable pH Probe Items when using simplified cultureware only 600305 000 600306 000 pH Probe Assembly Sterile Single Use requires 600306 000 Reusable Holder for 600305 000 Reusable Glass Bottles and Polypropylene Ported Cap Assemblies Bottles are borosilicate with pol
105. al Inc Primary User Interface Screens 5 9 4 Log Screen Any user may view the current log by pressing the View Log button on the Alarm Log screen The Log screen see Figure 5 41 shows the log entries for the selected run Current by default that have accumulated since the start of the run The log screen is also used to optionally display one of eight historical logs The name of the log that is being displayed on the Log screen is displayed in the title bar A new log is automatically started when run setup is initiated see Section 7 4 1 All lines in the log are time stamped Jan 25 200 Log Current 10 27 50 a4 67 ADMINISTRATOR logged out voluntarily Jan 24 2007 pH e ele Lac BF T Med T COZ OUT taal elo CIRC 18 41 52 6 99 187 5 1212 0 00 37 3120 0 480 400 0 500 Jan 24 2007 18 43 25 ADMINISTRATOR logged on successfully Jan 24 2007 18 43 34 Factor scheduler11 rate set 1 Jan 24 2007 18 43 41 Factor schedule 2 rate set 2 Jan 24 2007 18 43 46 ADMINISTRATOR logged out voluntarily Jan 24 2007 pH Saz SIP Lac a T Med T OUT leet so CIRC 18 46 52 6 99 187 5 1212 0 00 37 31 20 0 480 400 500 Jan 24 2007 pH Gas ele Lac vi T Med T cas OUT yd pate bia CIRC 18 51 52 6 99 187 5 1212 0 00 37 3 20 0 480 400 500 Tan 2422082 pH Gas ICP or Lac vi T Med T ons OUT teal ala CIRC 18 56 52 6 99 187 5 1212 0 00 37 3 20 0 480 400 500 Jan 25 2007 6 59 11 Alarm Startup down time 0 12 00
106. all process control and user interface functions The software part number and revision level is displayed at the lower left of the screen If they have been entered the run name cultureware lot number inoculant ID and instrument name are displayed at the lower right see Sections 6 1 1 and 7 4 1 The Exit button will bring the user back to the Monitor screen The Logon button will only be available if user names and passwords are in use see Section 6 6 Feb 05 2007 Main Menu 14 36 12 Run Menu Setup Run Settings Advanced dd Alarm Log Fun RUN NAME Houpath CU CULTUREUARE LOT INDC INDCULANT 1D Software 200472 000 Rev B INSTR wt Hold Exit Logon Help Figure 5 16 Main Menu H Run Menu Pressing the Run Menu button on the Main Menu screen will bring an authorized user to the Run Menu screen This is the primary screen to guide an operator through the various steps of a cell culture run When the instrument is in the IDLE state the Run Menu screen will contain a single button labeled Start New Run see Figure 5 17 that when pressed will allow an authorized user to start a cell culture run see Section 7 4 If a cell culture run is already in progress the Run Menu screen will show two buttons Continue Current Run and Stop Current Run see Figure 5 18 If the Continue Current Run button is pressed the user will be taken to the most current screen for the current run state if the St
107. alysis Remote system access for system monitoring and or control Flowpath and instrument sensor monitoring is continuous and automatic for these parameters pH Lactate indirect Media temperature Refrigerator temperature IC pressure EC pressure Circulation flow rate The AutovaxID software stores all control setpoints e pH media temperature cycling times etc in the CompactFlash media on the single board computer If the AutovaxID experiences a power failure it will automatically resume operation once power is restored The AutovaxID process control software will evaluate the instrument s status and will resume full operation including data logging based upon the process control entries that existed prior to the power failure Once full operation is restored the user may modify process control entries if necessary The data log will record the power failure event and the duration of the power failure 3 1 pH Control The AutovaxID actively controls the pH of the cell culture to maintain it at a user defined setpoint by adjusting the gas mixture that flows through the gas exchange cartridge GEX The pH is continuously measured using a disposable pH probe that is integrated into the sterile cultureware The probe measures the pH of the IC media The IC media flows through the GEX wherein it interacts with a mix of air and CO supplied by the AutovaxID gassing system The pH of the circulating IC media is changed by modifyi
108. ame field the settings will only be automatically loaded if the Auto Load Run Settings box is checked 6 4 2 Initial Circulation Rate The Initial Circ Rate field defines the initial circulation pump rate that is used when the instrument enters the GROWTH state The specified Initial Circ Rate is only used if the Circulation Controller is set to None or Manual If the Schedule controller is used then the initial pump rate defined in the circulation pump schedule takes precedence e Initial Circ Rate Range 200 400 mL min in 1 mL min increments e Default Initial Circ Rate 200 mL min 6 4 3 Initial Media Rate The Initial Media Rate field defines the initial media pump rate that is used when the instrument enters the GROWTH state The specified Initial Media Rate is only used if the Media Controller is set to None or Manual If the Schedule controller is used then the initial pump rate defined in the media pump schedule takes precedence Similarly if the Lactate controller is used then the media pump rate will be defined by the lactate controller e Initial Media Rate Range 20 200 mL hr in 1mL hr increments e Default Initial Media Rate 20 mL hr AutovaxID Operations Manual 6 9 Biovest International Inc Software Setup amp Configuration 6 4 4 Fill Flush Media Rate The Fill Flush Media Rate field defines the media pump rate that is used during the last step of FILL FLUSH during the PRE INOCULATION calibrations whi
109. amp is found closed or tubing is obstructed 11 Confirm all factor bag tubing clamps are open and no tubing is obstructed Contact Biovest Account Services if a clamp is found closed or tubing is obstructed 12 Confirm all cultureware Luer connectors are finger tight Do not over tighten the connectors to prevent cracking them or making them excessively difficult to open them later when necessary 13 Confirm all factor bag Luer connectors are finger tight Do not over tighten the connectors to prevent cracking them or making them excessively difficult to open them later when necessary 14 Hang the harvest bags on the refrigerator rack Do these bags first a Mount one harvest bag on each side of the rack b Use caution to not puncture the bags c Do not kink or twist any tubing d Slide the attached in line filter into the clip in the center of the rack 15 Hang the factor bags on the refrigerator rack Do these bags second a Mount one factor bag on each side of the rack and on the outside of the harvest bags b Use caution to not puncture the bags c Do not kink or twist any tubing Note The harvest bags should be towards the inside of the rack and the factor bags should be towards the outside There should be one harvest bag and one factor bag on each side of the rack AutovaxID Operations Manual 7 9 Biovest International Inc Operating Procedures Note for Legacy Cultureware Only If the factor bags wer
110. anual 8 4 Biovest International Inc Simplified Cultureware Usage 8 5 1 Intracapillary Circuit Components e IC Reservoir is where fresh Media is pumped into the system and Outflow media laden with metabolic wastes but is product free is pumped out of the system These media typically have no growth supplements e Circulation Pump creates flow through the IntraCapillary Circuit e Tubing amp Check Valve ensure unidirectional flow of media _ through the Intracapillary Circuit Detail view showing the porous membrane of the hollow fibers which e Gas Exchange Cartridge GEX 2 open at their ends re oxygenates the cell culture IC media shown as red flows through h A the inside of the fibers deg and raises or lowers its EC media the cell culture and secreted protein all shown as beige surround the fibers The IC and EC compartments are separated by the porous fiber membrane and a seal surrounding the e Temperature Probe ensures fibers at their open ends correct incubator temperature e Flow Sensor ensures correct Circulation Pump flow rate e IC Sample is used with syringes to remove media for off line monitoring of pH and the concentrations of metabolites and dissolved oxygen lt Seal e pH Sensor is optional in simplified cultureware When present it allows automatic pH control When absent pH control is manual e Bioreactor the IntraCapillary Space or ICS is the interna
111. aracteristics of cells in static cultures will give some indication of how fast they will grow in AutovaxID cultures Cells that exhibit relatively slow growth in static cultures will respond similarly in hollow fiber perfusion systems AutovaxID Operations Manual 7 61 Biovest International Inc Operating Procedures If allowed problems associated with microbial contamination may be avoided by using antibiotic supplemented media in static culture and by routinely testing stock cultures for mycoplasma and or unwanted viruses Targeted range for inoculation of a 1 1m or a 2 1m bioreactor with hybridomas is 0 2E8 to 5E8 total viable cells Targeted volume for inoculation of a 1 1m or a 2 1m bioreactor is 30 to 50 mL Using up to 1E9 cells has proven helpful in certain applications when the initial lag time is long Re suspending cells in 100 conditioned media or a 1 1 blend of complete fresh and conditioned media may be beneficial versus 100 fresh complete media Targeted scale up volume of 200 to 1000mL 7 26 2 Circulation Rate Guidelines As a cell culture grows the cells consume more of the oxygen that is dissolved in the surrounding media and therefore need an ever higher delivery to maintain the same level of available oxygen By increasing the circulation pump rate the flowpath is capable of delivering more dissolved oxygen to the cell culture An initial circulation rate of 200mL min is recommended for the first few days
112. are refillable if additional factor is needed but this requires the purchase of an optional sterile refill tubing line part number 600101 072 see chapter 13 Legacy Version part number series 600290 XXX was provided as two sterile assemblies The cultureware is separate from the factor bags These factor bags are NOT refillable so a new Factor Bag Set must be purchased part number 600288 000 see chapter 13 if additional factor is needed Standard cultureware contain e Hollow fiber bioreactor e Gas exchange cartridge GEX e EC reservoir e IC reservoir e Cycling apparatus e pH probe AutovaxID Operations Manual 7 5 Biovest International Inc Operating Procedures Circulation pump head Circulation flow sensor Sealed temperature probe well Factor bag set with tubing fill line Harvest bag set with 0 2um filter Four place pump cassette Most of the components of the cultureware except the harvest and factor bags and the pump cassette are enclosed within a clear polycarbonate cover that serves as the incubator enclosure The cultureware has eight ports Four ports are available for sampling each with a clamp and a female Luer connector IC Sample This is the pre bioreactor port for pH DO Glucose Lactate and other metabolic test samples Inoculate EC Sample This port is used to inoculate cells into the bioreactor to optionally monitor harvest concentration and cell viability and to optionally remove
113. armer part number EW 77202 60 e Two Channel Easy Load II pump head with stainless steel rotor 3 Mounting Hardware quantity 1 e Cole Parmer part number EW 07013 08 e Mounts three Easy Load Il pump heads to one pump motor 4 Note this equipment is not sold by Biovest International Inc Purchase this equipment from your preferred supplier AutovaxID Operations Manual 8 8 Biovest International Inc Simplified Cultureware Usage 8 7 pH Control Options There are three options for controlling pH when using simplified cultureware These options have varying labor requirements and expense summarized below Subsequent pages provide operating procedures for these options 8 7 1 Manual pH Control Advantages Disadvantages fastest run startup requires no use of a pH probe in the cultureware adjust pH based on color or use your bench pH meter least expensive of the three pH control options increased technician time and contamination risk when using a bench pH meter to monitor the AutovaxID s pH because this method requires frequent aseptic removal of samples of cell culture medium adjusting pH based on color of phenol red can lead to sub optimal culture growth or protein secretion infrequent pH adjustment can lead to sub optimal culture growth or protein secretion technicians are responsible for estimating gaining expertise in determining and entering the appropriate AutovaxID air CO2 mixture to raise or lo
114. assword Expiration 0 password expiration disabled NOTE Biovest suggests that each facility review their own regulations regarding periodically changing passwords and implement proper procedures 6 1 6 Minimum Password Length This field defines the minimum valid length for valid operator passwords The minimum password length specification only applies if usernames and passwords are in use Setting the minimum password length to zero 0 disables this feature e Password Length Range 2 to 12 characters e Default Password Length 6 characters NOTE Biovest suggests that each facility review their own regulations regarding passwords length and implement proper procedures 6 1 7 Logoff Timeout min The Logoff Timeout field is duration of inactivity in minutes that will be allowed before the current logged on user is automatically logged off The elapsed inactivity time is from the last interaction by the operator with the touchscreen The software reverts to the Monitor screen whenever the Logoff Timeout duration is met Entering a zero 0 in this field will disable the logoff timeout feature e Logoff Timeout Range 1 to 999 minutes in one minute increments e Default Logoff Timeout Value 60 minutes Due to the length of the procedures users will NOT be automatically logged off during CO calibrations container changes and when the instrument is on Hold 6 1 8 Status Logging min This field sets the frequency i
115. at the start of GROWTH cycling is not active When the run s elapsed time exceeds the delay time cycling will begin Biovest recommends waiting three to ten days after inoculation before turning on cycling to ensure that the cells are retained within the bioreactor e Default Cycle Delay 7 days 7 00 00 EC Rise Time An authorized user may specify the length of time in minutes for each EC rise e Default EC Rise Time 15 minutes e EC Rise Time Range 10 to 120 minutes in increments of 1 minute AutovaxID Operations Manual 5 24 Biovest International Inc Primary User Interface Screens 5 4 5 EC Fall Time An authorized user may specify the length of time in minutes for each EC fall e Default EC Fall Time 15 minutes e EC Fall Time Range 10 to 120 minutes in increments of 1 minute Press the Next button at the bottom of the Run Settings 1 screen to advance to the Run Settings 2 screen 5 5 Run Settings 2 Screen The Run Settings 2 screen see Figure 5 22 allows access to all of the user definable pump controller parameters that are used during GROWTH Pressing the button for each pump will display the associated pump Controller screen see the following sections for a description of each For convenience each pump Controller screen is also accessible from the pump Detail screen via the Monitor screen Similar to the Run Settings 1 screen the pump settings are retained from run to run The settings displayed in the Previo
116. authorized user may press the Run Menu button on the Advanced Menu screen to display the Advanced Run Menu see Figure 5 48 This screen shows the seven step sequence that the software follows for a complete cell culture run when running automatically This sequence is NOT enforced when the Advanced Run Menu is used by an administrator or if Strict Mode is disabled so operators must use these steps carefully An authorized user may use the Advanced Run Menu to repeat or skip steps in the cell culture sequence Any actions taken using the Advanced Run Menu are automatically recorded in the log WARNING Failure to complete the proper cell culture sequence can result in a failed cell culture run Jan 25 2007 Advanced Run Menu 13 56 50 EM Setup EJ Inoculate il Pill Flush GE Growth EN Calibrate EN Stop A Pre Inoculation Hold Exit Help Figure 5 48 Advanced Run Menu AutovaxID Operations Manual 5 61 Biovest International Inc Primary User Interface Screens The seven steps are 1 Setup During Setup the operator performs all of the required tasks prior to fill flush These tasks include entering information about the cultureware and fluid containers setting up the run parameters pH cycling controller pump controllers and alarms and performing a pressurization test 2 Fill Flush The fill flush step introduces media into the cultureware and brings the incubator up to operating temperature The
117. barcode scanner and may be up to 51 characters long including letters numbers spaces and punctuation Operators are encouraged to use this feature to document events or make comments during a run Multiple log entries can be used to enter long comments Log entries are considered events and are automatically time stamped AutovaxID Operations Manual 5 50 Biovest International Inc Primary User Interface Screens 5 9 3 Verify Log Any user may perform a verification of the current log data by pressing the Verify button on the Alarm Log screen The AutovaxID software will analyze the log and report on the Verify Log screen if any data corruption is found see Figure 5 40 Jan 25 200 Verify Log 10 24 29 Log Verified 1 Bad Records Record 2281 Sequence Error expected 2281 Found 2211 Total 1 Bad Records 1 thru 9 Hold Exit Help Figure 5 40 Log Verification Report The log verification report may contain five types of error lines e Error Header The first entry giving a summary of errors e End of File Error This entry indicates problems with the file ending before it is expected to e Unknown Record Type The record is of unknown type e Sequence Error The record entries are in the wrong order e Checksum Error The data has been corrupted or changed Contact Biovest Account Services for assistance with log verification errors AutovaxID Operations Manual 5 51 Biovest Internation
118. bioreactor and to remove metabolic waste from the bioreactor Pressing the Circulation button on the Run Settings 2 screen brings up the Circulation Controller screen see Figure 5 34 This screen shows the circulation controller type and provides access to the controller setup The Circulation Controller Types are e None No control is performed or allowed the pump is used in override strict mode only and its rate is defined on the Run Configuration 1 screen see Section 6 4 e Manual The circulation pump rate can be entered manually at any time by an authorized user Each manual pump rate change is automatically logged e Schedule The circulation pump rate is scheduled to change at user defined times to user defined rates All changes to the schedule are automatically logged all scheduled pump rate changes are automatically logged See Section 5 5 13 for details on setting up a schedule See Section 7 26 2 for recommendations on setting the circulation pump rate An authorized user may change the circulation controller type by touching the Controller Type field and then selecting the desired circulation controller type from the on screen list The Schedule controller has additional setup parameters that can be accessed by pressing the Controller Setup button Authorized users may change the controller type during a cell culture run AutovaxID Operations Manual 5 41 Biovest International Inc Primary User Interface Screens
119. bonate Bottle with Ported Cap Assembly Non Sterile one each Vent Filter 2um Non Sterile Single Use 50 per package use 1 per ported bottle cap assembly when that is reused Tygon Off Gas Line Tubing 1 8 ID 60 long Non Sterile Cabinet Fan Filter Refrigerator Rack Calibration recommend annually Installation Qualification Calibration Operation Qualification Service Contract renewable one year coverage Preventive Maintenance recommend every five years AutovaxID Operations Manual 13 3
120. button on the Factor Controller screen will bring up the Factor Pump Schedule screen see Figure 5 28 This screen allows an authorized user to schedule up to ten elapsed times with variable pump rate values In strict mode and when user names and passwords are in use only administrators may create schedules Normal users may view the schedule but they may not change the values in the Elapsed Time or the Factor fields nor can they add or delete events Normal users may advance or delay the schedule using the Advance and Back buttons or by entering a manual time shift To add an event press the Add Event button at the bottom of the screen enter the desired Elapsed Time for the start of the event using the on screen keypad and then press Enter Events are in day hour minute format The new event will appear in the schedule all events are automatically sorted chronologically Press the Fac field for the new event and enter the desired rate for the factor pump in mL hr using the on screen keypad The factor pump has a range of 0 to 400 mL hour in 1mL hour increments To delete an event press the Delete button to the left of the event Events may be added or deleted during an active run by an authorized user Jan 16 2007 Factor Pump Schedule 14 07 02 Factor Pump Schedule Elapsed Tine Fac Elapsed Tine Fac Delete um 1 Delete 10 00 00 2 All tines are on schedule 0 00 00 Advance Back Add Event Hold Ex
121. by the walls of the hollow fibers Intracapillary Space ICS The volume inside of the fibers of the bioreactor Lactate Control The instrument can automatically regulate the media flow rates and CO gassing to control lactate levels inside the Bioreactor Manual Override When an operator enters a software setting or changes a value manually rather than relying on the software to automatically control the instrument Overrides take precedence over process control MWCO Molecular Weight Cutoff Outflow Effluent from the IC conditioned media containing metabolic by products which are small enough to pass through the pores of the fibers Sometimes referred to as Waste Oxygenator See Gas Exchange Cartridge GEX Peristaltic Pump The Media Factor Harvest and Outflow pumps which move fluid into and out of the cultureware flowpath in a pulse like flow Production Phase The phase of a cell culture run when cell growth has leveled off and target protein production is maximized Pump Cassette The plastic guide for the peristaltic pumps which securely holds the tubing and locks into the pump Remote Computer A computer connected through the LAN or directly to the Ethernet port in the back panel of the AutovaxID instrument Run Sequence The defined steps that are taken to prepare and carry out a cell culture run Sample A small portion of media aseptically removed from
122. cbon 2 Get the set of three o rings see figure 8 10 and set of quick coupler latch d two autoclave tubing lines 5 see figure 8 11 Dn 3 Remove the o rings from PA their bag They do not need Figure 8 9 Reusable pH Probe to remain sterile now Assembly Components Inspect the o rings for damage O O O Disassemble the probe case figure 8 12 and ensure it was cleaned at Figure 8 10 the end of the previous run Discard Replacement pH Probe Case O rings the three old o rings if present Set of Three shown in figure 8 1 3 Install the new Shown in red for clarity Their actual color is clear o rings and loosely thread together Part Number 600301 000 the lower and upper sections of the pH probe case ie 5 Disconnect the pH cable and BNC Pe to two pin adapter from the pH probe and connect its electrical cap Figure 8 11 Autoclave Tubing Lines making it only finger tight Part Number 600300 000 6 Rinse the probe and blot it dry 7 Insert the pH probe into its probe case and thread them together making them only finger tight Do not use tools to tighten them together 8 Finish tightening the lower sections of the probe case but just finger tight 9 Rotate the white locking rings of both of the autoclave tubing lines to the unlocked position AutovaxID Operations Manual 8 14 Biovest International Inc Simplified Cultureware Usage 10 Press and hold the latch of the pH IN LO tubing lin
123. cnannnannnncos 8 9 8 8 Using Manual pH Control no pH Probe occcccccccnnncconcccccnoccconananannncncnnncnnnnnnnnnnnnnnnnncnnnnnnns 8 10 8 9 Using a Reusable pH Proba di d Sue 8 12 8 9 1 Reusable pH Probe Validation kee 8 12 8 9 2 Reusable pH Probe Assembly atte 8 14 8 9 3 Reusable pH Probe Autoclave Gierltzaton 8 15 8 9 4 Reusable pH Probe Installation into the Cultureware eee 8 16 8 10 Using a Disposable pH ProDe oici A is 8 18 8 11 Factor and Harvest Container Options oooccccncncccnnocoooccccconccnnnnnnanonononnnnncnnnnnnnoncnnnnnnnncnnnnns 8 20 8 12 Starting a Run Using Simplified Cultureware oooononconcccccnnnccccnnncanonncnnnncnonannnnnnncccnnnncnnnnnnns 8 21 8 137 Using Factor Practicas 8 25 8 14 Removal and Storage of pH Probe Hemes AAA 8 26 AutovaxID Operations Manual xi Biovest International Inc Table of Contents 10 11 12 13 Troubleshooting iira aeaa aaaea araa aaa A aAa KEA E aa eiA a aaar Aa iaa aadA 9 1 Ee E 10 1 HOT enee eebe 10 1 10 2 Remote Tt EE 10 3 103 LOG WebPages nta di Ads 10 5 E E ER EE 10 6 UE Viewing a bLOg WEE 10 8 10 4 Alternative Remote Access ENNEN 10 9 Seene Ee ee Ee Ee 11 1 GIGSSANY EEN 12 1 Paris dis is 13 1 AutovaxID Operations Manual xii Biovest International Inc Introduction 1 Introduction The AutovaxID is an automated mammalian cell culture system that has been designed to produce cell secreted products quickly economically and in qua
124. cocooccae Poocoecoocoeoeeceocso yo 000 2 gt 200000000000000 3 OO OO OO OO OO OO OO h eeoeeoeoeoeoeoeoeoeee e OO oO Oo oO OO oO oO oO oO oO oO oO h eeoeoeoeoeoeoeoeoeoeoeoeoeoe Oo oOoOooOooOoOoOoO o oo eo Tj bb bb bb bb be wb wb bb S Figure 10 6 Text Log in Web Browser 10 3 2 Viewing a Log Any current or historical log can be viewed two ways by using the Alarm Log screen via Remote Control or by viewing as a web page from the Log Menu page To view the log as a web page first select the log from the pulldown box on the Log Menu page Current or History 1 through 8 Then select the log items that you would like to view Events Alarms and or Status lines the number of lines that you would like to view on each page and if you would like to view the last log entries first Tail see Figure 10 4 Finally when all options are set click the View button Figure 10 7 shows a log data page with ten lines displayed lines 11 20 of 111 Notice that events are shaded yellow alarms are shaded red and status lines are shaded green for easy recognition AutovaxID Operations Manual 10 8 Biovest International Inc Remote Access BER 4 Y JE http 127 0 0 1 8080 logfpd htmi test1 show Sev 18al 1 amp st 18hist Current amp lines 108r_ 4 Google Dis ve di GrimLog A E ab 1 Page O Tos ei 3 Log Data Current Date Time pH Gas ICP ECP Lact State MediaT RefT AT CO2 Outflow Media Factor Harvest C
125. computer that is integrated into the AutovaxID The operating system user interface application and all recorded data are stored on a CompactFlash card Several other micro processor based boards integrated throughout the instrument cooperatively control the sub systems the peristaltic pump module the gassing module the rotary valves the incubator heater etc The AutovaxID process control software allows operators to perform the following primary functions e Configure the system hardware e Establish operating parameters e Identify operators using usernames and passwords e f desired restrict the sequence of operation that the operator must follow e Enter raw materials information into the log media lot factor lot etc e Start manage and stop production runs e Monitor system functions and generate alarms for out of tolerance parameters e View backup and print log data via remote computer e Calibrate integral instrument sensors service personnel only e Diagnose system electronics Service personnel only AutovaxID Operations Manual 1 12 Biovest International Inc Introduction 1 3 7 AutovaxlD Connectivity The most basic configuration of the AutovaxID is a stand alone instrument that is not connected to a network As a stand alone instrument it is fully capable of all functions necessary for a successful cell culture production run However to enable remote access the AutovaxID must be connected to a
126. count Services The instrument has registered a Board corrupt ROM alarm The instrument is experiencing a hardware failure Do not use instrument and contact Biovest Account Services The instrument has registered a Pump head not latched alarm Pump cassette is not in place Place the pump cassette in the pump and lock in place using the knob on the side of the instrument The pump cassette has been damaged Check the pump cassette for damage and retry The instrument is experiencing a hardware failure Do not use instrument and contact Biovest Account Services Contact Biovest Account Services The instrument has registered a The cultureware is not in place Place a cultureware into the instrument and lock in place using the latches on the top and side of the instrument Cultureware not latched alarm The cultureware has been damaged Check the cultureware for damage and retry The instrument is experiencing a hardware failure Do not use instrument and contact Biovest Account Services The instrument has registered a CO inlet pressure low alarm The CO inlet pressure is low Increase the CO inlet pressure to the pressure specified in Section 11 The COs supply line is disconnected or has become blocked Check that the CO supply line is connected and there is a flow of CO at the instrument The CO filt
127. ctor bag set see Section 7 3 1 ICS media can be purchased sterile and ready for use or it can be produced by each facility and then filter sterilized into sterile bags or containers as preferred by the facility Media Quality Control The quality of the media is one of the keys to a successful production run and ensuring media sterility is critical Always confirm media sterility before pumping it into the system If the media has been prepared and stored for over six weeks prior to use it should be tested to be sure that it is still usable Certain nutrients in the media may degrade during long periods of storage This degradation could result in either the concentration of certain nutrients being too low or as in the case of L glutamine the degradation converts nutrients to cytotoxic compounds If multiple batches of media are prepared for a particular cell culture run be certain to maintain the media chemistry so that the Sodium bicarbonate concentration remains constant If there are any questions regarding media quality please refer to the manufacturer s recommendations or contact Biovest Account Services Scale Up of Cells for Inoculation Refer to Section 7 24 for inoculation guidelines regarding cell density cell viability and inoculum volume requirements AutovaxID Operations Manual 7 4 Biovest International Inc Operating Procedures 7 3 Cultureware Setup Procedures This section provides descriptions o
128. d Overrides Set the Gas CO2 value to 15 0 It will appear as 75 0 This suggested value will not necessarily result in the desired pH but it should prevent the AutovaxID from adjusting pH too drastically from reasonable values because of the lack of a pH probe e Note Further adjustments later on to the Gas CO2 value likely will be necessary before inoculation should occur to ensure the desired pH has been stably achieved the pH calibration value From the Overrides screen press Exit twice to reach the Main Menu screen Press Run Menu then Continue Current Run to return to the current duty screen Press the pH Calibrate button on the screen in figure 8 23 Without removing a sample of cell culture medium press Next and enter 7 0 as e A warning message appears saying the pH calibration would be excessive Do you still want to use it Press YES Press Exit e Allow Lactate Calibration to be performed automatically 21 When using automatic pH control reusable or disposable pH probe e Press the pH Calibrate button on the screen in figure 8 23 Perform a standard pH calibration using the information provided in the Operating Procedures section of the manual e Allow Lactate Calibration to be performed automatically AutovaxID Operations Manual 8 23 Biovest International Inc Simplified Cultureware Usage 22 23 24 Proceed with Factor Addition as suggested in the Operating Procedures section of the ma
129. d rates The factor pump schedule may optionally be tied to the harvest pump schedule Both the factor and harvest pumps must be under schedule control in order to implement this option All changes to the schedule are automatically logged all scheduled pump rate changes are automatically logged See Section 5 5 6 for details on setting up a schedule e Time Cycle The factor pump can be set to turn on and off based upon a time based schedule where an authorized user may specify the pump rate delay until time total cycle time and on time See Section 5 5 8 for details All pump rate changes are automatically logged An authorized user may change the factor controller type by touching the Controller Type field and then selecting the desired factor controller type from the on screen list Some controllers have additional setup parameters that can be accessed by pressing the Controller Setup button Authorized users may change the controller type during a cell culture run Jan 24 2007 Factor Controller 18 41 17 Controller Tupe Schedule Container Underflow Alarn Controller Setup Used At 10 7 Tie to Harvest 3 Container Volume 3000 nl Hold Exit Hel p Figure 5 27 Factor Controller Screen AutovaxID Operations Manual 5 32 Biovest International Inc Primary User Interface Screens 5 5 6 Factor Pump Schedule Setup If the factor controller type is set to Schedule then pressing the Controller Setup
130. d setpoint The media temperature is measured by a thermo electric temperature sensor that is integrated into the instrument see Section 1 3 4 Cultureware Interface Panel The temperature sensor mates with a sealed disposable temperature well that is integrated into the sterile cultureware The temperature sensor is never in direct contact with the circulating media it is not wetted The temperature sensor provides feedback to the heater control algorithm The heater is integrated into the instrument and heats the air inside the cultureware enclosure incubator The heated air is circulated through the incubator via two ducts in the Cultureware Interface Panel see Figure 1 3 The media temperature control algorithm switches the thermo electric incubator heater on or off based upon the temperature of the IC media The circulation fan runs continuously during an active cell culture run AutovaxID Operations Manual 3 4 Biovest International Inc Introduction to the Process Control Software 3 4 Refrigerator Temperature Control The refrigerator temperature is continuously monitored and controlled by the AutovaxID to maintain it at a user defined setpoint The refrigerator temperature is measured by a thermo electric temperature sensor that is mounted in a heat sink on the back wall of the refrigerator The temperature sensor provides feedback to the refrigerator control algorithm The refrigerator is cooled by a thermo electric chill
131. d user to schedule up to ten elapsed times with variable pump rate values In strict mode and when user names and passwords are in use only administrators may create schedules Normal users may view the schedule but they may not change the values in the Elapsed Time or the Media fields nor can they add or delete events Normal users may advance or delay the schedule using the Advance and Back buttons or by entering a manual time shift To add an event press the Add Event button at the bottom of the screen enter the desired Elapsed Time for the start of the event using the on screen keypad and then press Enter Events are in day hour minute format The new event will appear in the schedule all events are automatically sorted chronologically Press the Med field for the new event and enter the desired rate for the media pump in mL hr using the on screen keypad The media pump has a range of 0 to 400 mL hour in 1mL hour increments The minimum recommended media pump rate is 20 mL hr to compensate for any evaporative loss through the GEX To delete an event press the Delete button to the left of the event Events may be added or deleted during an active run by an authorized user Jan 24 2007 edia Pump Schedule 18 00 54 Media Pump Schedule Elapsed Tine Med Elapsed Tine Med Delete 2 00 00 40 Delete 4 00 00 80 All tines are on schedule 0 00 00 Advance Back Add Event Hold Exit Help Figure 5 25 Media
132. des The factor valve position may be changed only by an authorized user via an override The Factor Valve should only be overridden if directed by a Biovest Service Technician Improperly setting the valve position can damage the cultureware and may be detrimental to the cell culture environment 5 12 6 Harvest Valve Overrides The harvest valve position may be changed only by an authorized user via an override The Harvest Valve should only be overridden if directed by a Biovest Service Technician Improperly setting the valve position can damage the cultureware and may be detrimental to the cell culture environment AutovaxID Operations Manual 5 66 Biovest International Inc Primary User Interface Screens 5 12 7 Pump Overrides Each of the five pumps can be overridden by an authorized user Extreme caution should be used for pump overrides For example if the Outflow pump flow rate is overridden and set at less than the total inflow rate more media will be flowing into the cultureware than is being removed This condition will eventually cause a buildup of pressure within the cultureware and will likely result in irreparable damage to the cultureware Sterility of the cultureware also may be sacrificed and depending on the nature of the cultureware failure media and or cells may leak into the outside environment CAUTION If the operator needs to override the normal pump flow rate during a run the operator should remain at
133. dify Administrator and Operator accounts and passwords Touch Calibration Used to calibrate the touch screen Network Print Log Used to print the log to a network connected TCP IP enabled printer on the same logical network as the AutovaxID See Section 6 Software Setup amp Configuration for details about these features Jan 25 2007 setup 12 49 24 Conf igure Touch Calibration Diagnostics Netuork Print Log Passwords Hold Exit Help Figure 5 46 Setup Screen AutovaxID Operations Manual 5 59 Biovest International Inc Primary User Interface Screens 5 12 Advanced Menu An authorized user can access the Advanced Menu screen by pressing Advanced button on the Main Menu screen The Advanced Menu provides access to the Advanced Run Menu and the Overrides screen These areas of the software do not need to be accessed by regular users operators during a normal automated cell culture run and are accessible only by authorized users Any screens accessed from the Advanced Menu should be used with caution and should only be used by an experienced operator or under the direction of a Biovest Service Technician Jan 25 2007 Advanced Menu 13 04 10 Advanced Controls Use with caution Run Menu Overrides Hold Exit Help Figure 5 47 Advanced Menu AutovaxID Operations Manual 5 60 Biovest International Inc Primary User Interface Screens 5 12 1 Advanced Run Menu An
134. during the run If additional factor is needed to complete the run a new Factor Bag Set must be purchased and then filled as described The Factor Bag Set is intended for a single use and should be disposed of with the cultureware when the run is complete per your facility s requirements 7 3 3 Cultureware Preparation 1 Obtain an empty AutovaxID harvest factor refrigerator rack disinfect it and place it in the laminar flow hood or biological safety cabinet BSC 2 Carefully open the cultureware shipping box and remove the bagged cultureware assembly a For Legacy cultureware only carefully open the Factor Bag Set shipping box and remove the bagged Factor Bag Set Alternative Procedure Legacy Factor Bag Set Only if you prefer to prepare fill and store the Legacy cultureware s Factor Bag Set in advance of preparing the cultureware perform steps 2a 3a 4a 5a 6a 8 11 13 and 15 as part of Fill Factor Bags from 600290 XXX Legacy Cultureware process detailed in the next section In this case perform these steps before the first step of the filling procedure Note Do NOT lift or carry the cultureware and Factor Bag Set by their sterility bag 3 Inspect the cultureware s sterility bag for any punctures or damage Contact Biovest Account Services if any damage is found a For Legacy cultureware only inspect the factor bag set s sterility bag for any punctures or damage Contact Biovest Account Services if any damag
135. e title bar to the left of the screen title when a user is logged on The type of character displayed will depend upon the user s rights e Administrators an asterisk will be displayed e Regular users a hyphen will be displayed AutovaxID Operations Manual 6 20 Biovest International Inc Software Setup amp Configuration Figure 6 18 Monitor Screen with Logoff Button AutovaxID Operations Manual 6 21 Biovest International Inc Software Setup amp Configuration 6 6 4 If a user is logged on and no activity is detected for a length of time greater than the Logoff Timeout as setup by the system administrator the instrument will automatically log the user off and display the Monitor screen Logon and logoff events are automatically entered into the instrument s log WARNING The system must not be left unattended when a user is logged on Any actions taken on the instrument are attributed to the user who is actively logged on to the system Initial Logon Procedure for New Users The first time that a new user attempts to log on to the AutovaxID he she will be required to change their password To log on to the AutovaxID 1 From the Monitor screen Main Menu or Logon screen press the Logon button 2 Type your assigned username as provided by an administrator using the on screen keyboard press the Enter button 3 When prompted for your password press the Enter button Your initial pa
136. e 10 5 The text format will display the log in the web browser thereby allowing you to save it to a text file see Figure 10 6 For example in Internet Explorer click File gt Save As then save the file in txt format The log file will be space delimited and the top row will contain the column headings The raw format will save the file as a xlg file This is the file format used by the AutovaxID software to save the log data and is proprietary to Biovest While this file can be saved as a method of archiving the log a conversion utility is required to read its contents Please contact Biovest Account Services if you wish to make use of this file format AutovaxID Operations Manual 10 6 Biovest International Inc Remote Access Download Log Menu Windows Internet Explorer 09 X showSev 18al 18st 18hist 49 X code ps Download Log Menu x Survey 104183 gt E gt db page 2 Toos O A E http 127 0 0 1 8080 logdl html test1 Download Log Menu for Current Y07M01D17 H16M07 Download Text Download Raw po ll IL ttre 100 y Figure 10 5 Download Log Menu AutovaxID Operations Manual 10 7 Biovest International Inc Remote Access http 127 0 0 1 8080 logfp cur YO7M01D17 H16M07 bxt lines 9999999 amp hist Cur Windows Internet Explorer Go E httv 127 0 0 1 8080 10gfp cur Y07M01D17 H16M07 txtiines 99999998hist Cur Google we de 35 Entp
137. e Container ID field and using either the on screen keyboard or the barcode scanner to enter the necessary information When the Container ID information is complete press the Next button to temporarily stop the appropriate pumps Jan 16 2007 Media Change 19 57 59 Do NOT change the container yet Enter the information below Once you are ready to aseptically nake the change press Next to stop the pump Media container ID MEDIA CONTAINER ID Select to make a log entry Log Entry Figure 7 39 Container Change Details Screen AutovaxID Operations Manual 7 52 Biovest International Inc Operating Procedures Before the AutovaxID stops the pumps it confirms the container change once again If everything is ready the operator should press the Yes button to proceed Otherwise the operator may press No to return to the previous screen See Figure 7 40 Figure 7 40 Container Change Confirmation Screen AutovaxID Operations Manual 7 53 Biovest International Inc Operating Procedures Immediately after the Yes button is pressed the AutovaxID software will stop the appropriate pumps and prompt the operator to aseptically change the selected container see Figure 7 41 Jan 16 2007 Media Change 20 05 10 a JI eb Media pump is stopped Aseptically change the container When conplete press Done to resune Done Hold Help Figure 7 41 Container Change In Process
138. e Password Add Account Number of Users 2 Hold Exit Hel p Figure 6 14 User Accounts Screen On the User Accounts screen press the Add Account button using the on screen keyboard type the desired new username and then press the Enter button A pop up screen will ask if the new user should have Administrator privileges See Figure 6 15 answer appropriately AutovaxID Operations Manual 6 18 Biovest International Inc Software Setup amp Configuration d D AAA An am E i P Give USERI Admin Privileges Figure 6 15 Assigning Administrator Privileges The new user has been created and should be listed on the User Accounts screen The first time that a new user logs in they will be required to enter and confirm a unique password The Administrator cannot read the passwords that have been created by other users If the password is forgotten the Administrator can only clear reset the password at which time the user must enter a new password Up to twenty user accounts can be created If the instrument does not have any user accounts setup then the logon logoff system is disabled The AutovaxID software requires at least the primary administrator account to remain established if the logon logoff feature is to remain in use Do NOT delete the primary administrator account unless you wish to disable the username and password features REMEMBER Valid usernames may only be created by an admin
139. e and push the connector on to the lower right port of the lower section of the pH probe case Release the latch and rotate the locking ring to the locked position Ensure the latch faces upward see figure 8 14 11 Repeat the previous step for the pH OUT HI tubing line connecting it to the upper left port of the lower section of the pH probe case 8 9 3 Reusable pH Probe Autoclave Sterilization 1 Disconnect the Luer fittings at the ends of the autoclave tubing lines Use a syringe to inject sterile or non sterile 1X phosphate buffered saline until the probe case chamber and several inches of each tubing line are filled Figure 8 12 Reusable pH Probe Case 2 Reconnect the Luer fittings that were Part Number 600298 000 disconnected in the previous step 3 Unscrew the lower section of the probe case by turn Loosening the probe case s lower section allows for thermal expansion during the autoclave cycle 4 Use a rubber band to loosely hold the tubing lines up against the upper section of the probe case so they loop smoothly upward preventing the PBS from leaking out Ensure the tubing lines are not kinked or pinched by the rubber band Figure 8 13 Installing o rings in the probe 5 Ensure the protective electrical cap is f E case s lower section connected to the probe Failure to do so may result in moisture damage in the autoclave and will void its warranty 6 Place the pH probe assembly without
140. e cultureware includes the following components e Hollow fiber bioreactor e Gas exchange cartridge e In line media temperature well e Media flow sensor adapter e In line pH probe e Circulation pump head e IC and EC bags in cycling fixture e Pre loaded pump cassette e Harvest Bags and Filter e Incubator Enclosure e Tubing amp fittings NOTE The pH probe and temperature well are integrated in the disposable cultureware They are not re usable and must be discarded with the cultureware All wetted components have been certified USP VI or have been tested to ensure that they are not cytotoxic The cultureware is completely enclosed The enclosure panels are constructed of clear high impact plastic so that the flowpath can be easily observed by the operator The enclosure also serves as the incubator and confines the heated cell culture environment AutovaxID Operations Manual 1 11 Biovest International Inc Introduction 1 3 6 Process Control Software The AutovaxID control software enables operators to monitor and control all system functions including flowpath pH media temperature cycling time and pressure calculated lactate concentration peristaltic pump rates and circulation pump rate In addition the process control software allows operators to start and stop cell culture runs calibrate sensors control the refrigerator temperature and record events The process control software operates on a single board
141. e hollow fibers within the bioreactor are kept wetted with a stabilizing agent during storage and before use to maintain their physical integrity This stabilizing agent is cytotoxic and must be flushed out of the cultureware before inoculation The FILL FLUSH procedure achieves two necessary results flushes the stabilizing agent from the inside and outside of the hollow fibers and adds media to the cultureware Approximately five liters of media is required to perform a complete FILL FLUSH The operator is not required to manipulate any clamps during FILL FLUSH The FILL FLUSH procedure is fully automated and requires approximately twenty four hours If an operator presses the Exit button and returns to the Monitor screen FILL FLUSH will continue in the background and the AutovaxID State will be noted as FILL To return to the current FILL FLUSH screen press Menu gt Run Menu gt Continue Current Run During FILL FLUSH e Rapid cycling is performed to aid flushing the hollow fibers e The air and CO flow rates are fixed to stabilize the pH of the media e Automatic media temperature control is turned on e Media temperature alarms are turned on based upon operator settings e Refrigerator temperature alarms are turned on based upon operator settings e Pumps are turned on and off as necessary AutovaxID Operations Manual 7 21 Biovest International Inc Operating Procedures While operators do not need to understand the details of
142. e is found 4 Disinfect the exterior of the cultureware s sterility bag with 70 Isopropanol a For Legacy cultureware only disinfect the exterior of the factor bag set s sterility bag with 70 Isopropanol 5 Place the bagged cultureware into the sterile laminar hood or BSC next to the refrigerator rack a For Legacy cultureware only place the bagged Factor Bag Set into the sterile laminar hood or BSC next to the refrigerator rack AutovaxID Operations Manual 7 8 Biovest International Inc Operating Procedures 6 Remove the cultureware from the sterility bag The green tape on the cultureware should have visible brown red stripes to indicate proper ETO sterilization Thoroughly inspect the cultureware to confirm that it is undamaged Contact Biovest Account Services if any damage is found a For Legacy cultureware only remove the Factor Bag Set from the sterility bag The green tape on the Factor Bag Set should have visible brown red stripes to indicate proper ETO sterilization Thoroughly inspect the Factor Bag Set to confirm that it is undamaged Contact Biovest Account Services if any damage is found Remove all foam tape and other packing materials from the cultureware Remove all foam tape and other packing materials from the factor bags Leave the cultureware s pump cassette in place 10 Confirm all cultureware tubing clamps are open and no tubing is obstructed Contact Biovest Account Services if a cl
143. e is not detected CO2 inlet pressure low Logged if the instrument is at or past FILL FLUSH and the COs inlet pressure is low Air inlet pressure low Logged if the instrument is at or past FILL FLUSH and the air pressure from the integral air pump is low Heater current fail The heater current is sensed If the current is sensed to be on when it should be off or vice versa for longer than a 1 10 of a second the alarm is generated VALVENAME valve failed Logged if the valve is detected in the wrong position failed VALVENAME will be Factor or Harvest EC Pumps on hold too long Logged if the instrument is at or past PRE INOCULATION and the pumps are on hold for longer than five minutes Missing calibrations Logged at startup if any calibrations have not been performed where is a comma delimited list that may include CO2 air pressure and or temperature For each process control alarm the user can specify if it is enabled or disabled and the acceptable low and high limits for each value see Sections 5 7 and 5 8 However as with other fields while one may specify values for the low and high alarm limits the AutovaxID software restricts the accepted values to a specified range see Section 4 3 Data Entry Table 3 4 shows the allowed range for each alarm as well as the default values AutovaxID Operations Manual Biovest International Inc Introduction t
144. e left of the event Events may be added or deleted during an active run by an authorized user AutovaxID Operations Manual 5 42 Biovest International Inc Primary User Interface Screens Jan 25 2007 Circulation Pump Schedule 9 16 02 Circulation Pump Schedule Elapsed Tine Cir Elapsed Tine Cir Delete 0 00 00 1200 Delete 3 00 00 Do Delete 7 00 00 bm All tines are on schedule 0 00 00 Advance Back Add Event Hold Exit Help Figure 5 35 Circulation Pump Schedule Screen The event that is currently in use is highlighted see the first event in Figure 5 35 If no events are currently active because not enough time has yet elapsed the highlighted area will appear after the last entry see Figure 5 32 An authorized user may enter a new value for any Circulation rate at any time during a run The Elapsed Time may also be changed for any event by an authorized user so long as the schedule has not been advanced or delayed The Advance and Back buttons are used to shift the time of all events backward or forward These buttons are intended to be used if the cell culture is growing faster or slower than originally expected The time shift is displayed at the bottom of the screen To skip ahead to the next scheduled event press the Advance button to revert to a previous event press the Back button To remove any advance or delay touch the time shift field at the bottom of the screen to the left
145. e part of the flowpath that coincides with the inside of the fibers in the bioreactor The IC includes e The IC Reservoir where media and factors are added to the system and waste is removed e The circulation pump e The gas exchange cartridge GEX used to control pH and re oxygenate the cell culture media e The IC flowpath tubing e The ICS of the bioreactor e The probes used for monitoring the pH and temperature of the system When the circulation pump is running the majority of the IC media circulates within the IC but a small portion of the IC media also passes through the hollow fiber wall and enters the Extracapillary Circuit This half of the flowpath the IC is basically the same as in a conventional hollow fiber system If left to circulate this way the Starling Effect would create nutrient and waste gradients which would not permit optimized production AutovaxID Operations Manual 1 3 Biovest International Inc Introduction 1 1 4 The Extracapillary Circuit While the IC in the AutovaxID systems is similar to those in conventional hollow fiber systems the AutovaxID system incorporates a unique Extracapillary Circuit EC The EC is necessary to enable the EC Cycling process The EC is connected to the ECS where the cells grow and produce cell secreted products The EC includes e The EC Reservoir e Two check valves which direct the flow within the EC e The ECS of the bioreactor e The EC flowpath tubing
146. e pre filled and frozen they must be thawed before they are connected to the cultureware hung on the bag rack and placed in the refrigerator Frozen bags are more susceptible to puncture may not fit inside the refrigerator and may drop the refrigerator temperature below the user defined setpoint thereby causing an alarm 16 Legacy Cultureware Only Aseptically connect the Factor 1 line from the cultureware to the Factor 1 from the Factor Bag Set using their mating Luer fittings 17 Aseptically connect an outflow container to the cultureware s Outflow line using its Luer connectors Ensure the outflow container is vented using a 0 2u hydrophobic filter whether it is a collapsible bag or is rigid WARNING Do not use a glass vessel with an exhaust filter if the filter becomes wetted the vessel will become pressurized and could pose an explosion hazard to the operator 18 Aseptically connect a media container to the cultureware s Media line using its Luer connectors If the media container is a collapsible bag it does not need to be vented If the media container is rigid it must be vented using a 0 2u hydrophobic filter 7 3 4 Filling the Factor Bags 1 Factor is your preferred basal cell culture medium with added growth supplements such as fetal bovine serum at a variety of concentrations or one of the many commercially available serum free media 2 Prepare a little more than 3L of factor in advance Or prepa
147. e target protein should be measured Depending upon the type of product secreted by the cells this measurement may be in units of concentration or activity The Harvest pump rate may be gradually increased until the product levels reach a satisfactory concentration activity until the daily yield reaches a maximum amount or until the harvest volume becomes difficult to process The previous discussion on feedback inhibition should also be considered when pump rates are changed A comparison of secreted product from the harvest bags versus from the EC Sample port will undoubtedly show different levels of harvested material concentration or activity This happens because the harvest bags contain a composite of product material from the EC throughout a specified period while the EC Sample originates from a single point in time In addition product levels in the sample will vary with the cycling phase Rise versus Fall Therefore to measure accurately from the flowpath samples should always be removed at the same time in the Cycling process at the operators discretion Cooling the harvest bags is a good way to preserve product integrity by preventing thermal degradation It also has the advantage of stabilizing the changes in pH due to harvested cells as well as reducing the activity of proteases released by cell lysis If cell buildup in the harvest is a concern the in line filter should be used Initiate harvest collection between three a
148. ed result and does not cause any damage to the cell culture or flowpath AutovaxID Operations Manual 10 3 Biovest International Inc Remote Access It is important to remember that remote access interacts directly with the AutovaxID instrument The web browser is simply providing a means of accessing the instrument from a remote location such as outside of a cleanroom or from a home computer The AutovaxID user interface as viewed in web browser functions exactly as it does on the AutovaxID itself The only difference is that the mouse must be used instead of touch input In fact if an operator is standing in front of an AutovaxID instrument while another operator is accessing and commanding it remotely the on site operator will witness the screen changes and inputs that are being performed by the remote operator Note Depending upon the quality of the network connection an operator using remote access may need to wait briefly for the screen to refresh after a mouse click or alphanumeric input Note Just as with direct access individual key presses that are initiated remotely cannot be witnessed on site For example entering a password remotely will still show as asterisks on the screen both locally and remotely and anyone standing in front of the instrument during the remote access session will not be able to see what keys are being pressed while the password is being input Click on the Remote button on the home pa
149. edia Control 18 16 51 Media Pump Rate Controller Lactate Setpoint 1 000 Minimum Rate nl he 20 Maximum Rate nl she 400 PID Tine 0 00 30 PID Proportional 1000 PID Integral 3000 Hold Exit Help Figure 5 26 Media Pump Lactate Controller Screen AutovaxID Operations Manual 5 30 Biovest International Inc Primary User Interface Screens 5 5 5 The lactate concentration is calculated from the initial calculated effective bicarbonate concentration the CO concentration in the offgas and the pH reading The media pump speed is automated by the controller to keep the lactate concentration at the user defined lactate setpoint The minimum and maximum media pump rates allow the user to set boundaries for the lactate controller The minimum recommended media pump rate is 20 mL hr to compensate for any evaporative loss through the GEX e Default Lactate Setpoint 1 000 mg mL e Lactate Setpoint Range 0 000 5 000 mg mL in 0 001 mg mL increments e Default Minimum Media Pump Rate 20 mL hr e Default Maximum Media Pump Rate 400 mL hr The PID settings in the lower right corner are coefficients used by the search algorithms Do NOT change these values without first contacting Biovest Account Services Improperly changing the PID settings could result in adverse nutrient delivery lactate removal or pH control Factor Controller The factor pump delivers fresh growth factor to the flowpath to pre c
150. emoval of the cultureware with the valve in the wrong position may split or damage the tubing of the cultureware being removed or the next cultureware being loaded AE nann DI AA D I 33 H Jan STOP Stop the run and place in idle state Figure 7 42 End of Run Confirmation AutovaxID Operations Manual 7 57 Biovest International Inc Operating Procedures A Figure 7 43 Stop Run 1 Screen Figure 7 44 Stop Run 2 Screen AutovaxID Operations Manual 7 58 Biovest International Inc Operating Procedures 7 23 Cultureware Removal and Disposal Procedure WARNING Used cultureware should be considered a biohazard and handled according to your site s procedures for handling treatment and proper disposal The AutovaxID cultureware is intended for single use only and must be properly disposed of after the cell culture run is completed The refrigerator rack for the harvest and factor bags is the only reusable component be certain to save it for future cell culture runs To remove the cultureware 1 Following the prompts on the Stop Run 1 screen see Figure 7 43 close the Media and Factor clamps Press the Next button 2 Wait for the instrument to show that it is safe to remove the cultureware see Figure 7 44 Press the Next button on the Stop Run 2 screen Close the Outflow clamp Disconnect the Media and Outflow containers Remove the rack with the harvest and factor bags
151. ending upon the position of the valves so as to properly show the fluid routing The two grey blocks on the right hand side of the screen are mechanically the same valve but they are represented as two for flowpath clarity Touching any valve block will bring up the appropriate Valve Detail screen see Figure 5 44 and Figure 5 45 Each Valve Detail screen shows the valve name the currently commanded valve position Position field the actual valve position from the integral valve position sensor and a graphical representation of the valve Nothing on these screens is user configurable users cannot set the valve positions except using Diagnostic Overrides under the guidance of Biovest Account Services see Section 6 5 AutovaxID Operations Manual 5 57 Biovest International Inc Primary User Interface Screens Figure 5 44 Harvest EC Valve Detail Screen Figure 5 45 Factor Valve Detail Screen AutovaxID Operations Manual 5 58 Biovest International Inc Primary User Interface Screens 5 11 Setup Screen Pressing the Setup button on the Main Menu screen brings an authorized user to the Setup screen see Figure 5 46 The five buttons on this screen provides access to Configure Used to configure the instrument settings such as System Features Network Settings and Date amp Time Diagnostics Used by Biovest Technical Service personnel to diagnose or test instrument hardware Passwords Used to create and mo
152. er Peltier device that is integrated into the AutovaxID cabinet It does not contain any liquid refrigerant and does not require any user servicing The cooling fins of the chiller can be seen in the top of the refrigerator The refrigerator control algorithm switches the chiller on or off based upon the temperature inside the refrigerator The heat extracted from the refrigerator is dissipated by the fans on the rear cover of the AutovaxID When the AutovaxID is turned on the refrigerator runs continuously 3 5 Cycling Control The cycling state and pressure are controlled by the AutovaxID gassing system During normal operation when cycling is enabled in the GROWTH phase of a cell culture run the cycle state will be either Rise or Fall The cycling limits EC High and EC Low are detected by two non contact cycling sensors that are integrated into the instrument see Figure 1 3 and Section 1 3 4 Cultureware Interface Panel The sensors detect two edges of a mechanical flag that is integrated into the IC EC reservoir assembly of the disposable cultureware Depending upon the cycle state the gassing system will pressurize either the EC or the IC reservoir see Section 1 2 EC Cycling Theory The magnitude of the pressure is determined by the cycling algorithm During run setup the operator defines the desired cycle times for EC Rise and EC Fall The cycling algorithm measures and tracks the elapsed time for each cycle state If the cycle take
153. er has become blocked Contact Biovest Account Services AutovaxID Operations Manual 9 11 Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The inst EE The air filter has become Contact Biovest Account Roche VE blocked Services alari H The instrument is experiencing Contact Biovest Account a hardware failure Services The over temperature switch Allow the instrument to cool and has been tripped retry 3 E The thermal fuse has been Contact Biovest Account The instrument has registered a d Heater current failure low Zen Services pag bats Contact Biovest Account alarm The heater fuse has blown Servi ervices The instrument is experiencing Contact Biovest Account a hardware failure Services The instrument has registered a Valve failure alarm The valve drive failed to sync with the motor Retry moving the valve If failure continues contact Biovest Account Services The instrument is experiencing a hardware failure Contact Biovest Account Services The instrument has registered an alarm for the pump being on hold too long The pumps have been on hold too long Resume the run The instrument has registered a Missing calibrations alarm A calibration has not been done on the instrument Conduct a calibration on the instrument The calibration data file has become corrupt Contact Biovest
154. erant Instead the refrigerator compartment is kept cool using a solid state thermo electric Peltier chiller The refrigerator temperature may be set by the operator and is monitored and controlled by the AutovaxID control software Periodic calibration is required WARNING Setting the refrigerator temperature below 4 C can cause any fluid contained inside to freeze and may cause the Harvest and or Factor lines to rupture subsequently compromising sterility AutovaxID Operations Manual 1 8 Biovest International Inc Introduction 1 3 3 Peristaltic Pump Module The peristaltic pump module consists of four independent peristaltic pump motors and their control electronics The module is mounted inside the instrument cabinet on the lower left side beneath the cultureware platform see Figure 1 3 The four pumps are dedicated to specific flowpath functions and are controlled by the AutovaxID software The four pumps are from left to right e Outflow pumps waste media out of the cultureware to a customer supplied waste container e Media pumps fresh media into the IC from a customer supplied media container e Factor pumps growth factor into the cultureware IC and or EC from its integrated factor bags located in the refrigerator Use is optional though recommended e Harvest pumps product out of the cultureware into its integrated harvest bags located in the refrigerator The media factor and harvest pump rates
155. erator Temp Screen AutovaxID Operations Manual 5 12 Biovest International Inc Primary User Interface Screens 5 1 12 Outflow Detail Screen On the Monitor screen press the Outflow field or the surrounding area to bring up the Outflow Detail screen This screen shows the outflow pump position the outflow pump rate in mL hr the outflow container capacity and the estimated volume of fluid currently in the outflow container see Figure 5 11 The pump illustration also shows the direction of pump rotation see the arrow above the pump and the direction of fluid flow see the arrow next to the pump tubing The outflow pump rate is automatically set by the AutovaxID process control software and can be overridden only by an authorized user using an advanced override see Section 5 12 2 Biovest does not recommend manually overriding the outflow pump rate if the system is operating normally Improperly setting the outflow pump or neglecting to punctually clear an override can cause irreparable damage to the cultureware and may cause a cell culture run to fail An authorized user may change the outflow container by pressing the Change Container button see Section 7 17 An authorized user may also change the capacity of the outflow container see Section 7 18 By pressing the Settings button an authorized user can access the Outflow Controller screen see Figure 5 23 See Section 5 5 1 for details on the Outflow Controller scree
156. erriding the media pump rate if the pump is being automatically controlled and if the system is operating normally Improperly setting the media pump or neglecting to punctually clear an override may negatively affect the cell culture environment The minimum recommended media pump rate is 20 mL hr to compensate for any evaporative loss through the GEX Jan 24 2007 Media Detail 13 51 32 YJ 33 Media 20 mlyhe a 20 00 L Capacity 19 89 L Current Volune Change Container Settings Hold Exit Help Figure 5 12 Media Detail Screen AutovaxID Operations Manual 5 14 Biovest International Inc Primary User Interface Screens An authorized user may change the media container by pressing the Change Container button see Section 7 17 An authorized user may also change the capacity of the media container see Section 7 18 5 1 14 Factor Detail Screen On the Monitor screen press the Factor field or the surrounding area to bring up the Factor Detail screen This screen shows the factor pump position the factor pump rate in mL hr the factor container capacity and the estimated volume of fluid currently in the factor container see Figure 5 13 The pump illustration also shows the direction of pump rotation see the arrow above the pump and the direction of fluid flow see the arrow next to the pump tubing The factor pump rate may be automatically controlled by the AutovaxID process control software
157. es that the instrument is functioning normally with no alarms AutovaxID Operations Manual 5 1 Biovest International Inc Primary User Interface Screens 5 1 1 Elapsed Time The Elapsed Time is the accumulated running time real time since the start of the current instrument state see Section 5 1 2 The Elapsed Time is automatically reset at the start of SETUP and GROWTH and when the instrument returns to IDLE The time is displayed in days hours minutes format D HH MM The elapsed time resets to 0 00 00 as each new state begins State The State field shows the current status of a run The State will automatically change to reflect the current phase of the cell culture run as the operator progresses through the run control steps Pressing this field will bring up the Run Menu see Section 5 3 States include e IDLE The instrument is in the IDLE state from the end of one run to the start of the next When the instrument is first powered up IDLE will be the default state Incubator temperature control is disabled gassing defaults to 500 mL min air and all pumps are off During the IDLE state process control alarms are not issued but system alarms are Diagnostic functions are accessible by authorized users e SETUP The instrument will be in the SETUP state when a new run is initiated and until the start of FILL FLUSH This state will only be observed if the operator returns to the Monitor screen before FILL FLUSH
158. eturn to the Log View Options screen and then press the arrow for the desired search direction This iterative search process can be repeated indefinitely by the operator until the desired entry is located Press the Exit button to return to the Log screen AutovaxID Operations Manual 5 55 Biovest International Inc Primary User Interface Screens 5 10 Flowpath Screen The Flowpath screen see Figure 5 43 may be accessed by pressing the Flowpath button on the Main Menu It provides a graphical diagram of the cultureware fluid path provides operators with a convenient overview of system operation and allows access to process control and maintenance tasks such as reservoir changes and pH calibrations Any user may access the Flowpath screen by pressing the Flowpath button on the Main Menu screen Only authorized users may perform process changes and system maintenance Jan 16 2007 14 04 06 bie CE Ss Ut Figure 5 43 Flowpath Screen On the Flowpath screen fluid tubing appears as a thin line with animated pink dots that show the direction of flow Gas tubing appears as thin green lines The bioreactor appears near the middle of the screen as a pink box with vertical white lines running through it The top and bottom fluid lines connected to the bioreactor are the IC ports the two lines coming out of the right side are the EC ports The IC ports are connected to the ICS where media is circulated
159. f Qutflow Rate nl Ar 0 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 10 Fill Flush Step 4 AutovaxID Operations Manual 7 24 Biovest International Inc Operating Procedures dan 12 200 Fill Flush 16 26 21 5 Pre filling Prine Circulation 0 00 12 skip Fill Flush in progress Cycle ICPR Media Rate ml hr 480 C02 Sensor 120 00 y Circulation Rate ml min 0 Media Tenperature Control Off Qutflow Rate nl hr 0 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 11 Fill Flush Step 5 Jan 12 200 Fill Alush 16 26 57 6 Running Circulation Pump skip Fill Flush in progress Cycle ICPR Media Rate nl hr 480 C02 Sensor 20 00 y Circulation Rate ml min 186 Media Temperature Control Off Qutflow Rate nl hr 0 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 12 Fill Flush Step 6 AutovaxID Operations Manual 7 25 Biovest International Inc Operating Procedures dan 12 200 Fill Flush 16 29 58 7 IC Media Addition 0 00 17 skip Fill Flush in progress Cycle ICPR Media Rate ml hr 480 C02 Sensor 20 00 y Circulation Rate ml min 0 Media Tenperature Control Off Qutflow Rate nl Ar 480 Refrigerator Tenperature Control 4 0 Hold Exit Help Figure 7 13 Fill Flush Step 7 Jan 12 200 Fill Alush 16 32 29
160. f the ECS media from basal to complete supplemented media Factor must be added before inoculation to ensure that it mixes thoroughly throughout the entire ECS To minimize the shock during the seeding process do not rely on inoculation to switch the ECS media from basal to complete During Factor Addition the factor line is first purged of air into the IC see Figure 7 26 and then the defined volumes of factor are pumped into the EC and IC circuits see Figure 7 27 and Figure 7 28 It is recommended that Factor Addition not be done more than twenty four hours prior to inoculation to minimize factor dilution AutovaxID Operations Manual 7 34 Biovest International Inc Operating Procedures NOTE The hollow fiber bioreactor supplied with the AutovaxID cultureware has a nominal molecular weight cutoff MWCO of 10 000 Daltons 10 kDa so small molecular weight components such as peptides hormones cytokines and growth factors may be lost in the IC If any component in the media is larger than 30 000 Daltons molecular weight it may not cross the hollow fiber membrane The operator should determine whether a component can pass through the hollow fiber wall and if the component is critical for cell growth or production Factor Addition to the IC is optional and when setup by the operator occurs only once Factor in the IC effects cell growth only during the first few days of the run and can minimize the initial cell growth lag after
161. f the various types of cultureware and the procedures necessary for using them that vary depending on the type of cultureware in use During the cultureware setup and preparation procedure the cultureware is unpackaged and inspected a full media container is aseptically connected an empty outflow container is aseptically connected the factor bags are connected to the cultureware Legacy version only and the factor bags are filled The cultureware set is then ready to be installed to the AutovaxID instrument for a cell culture run 7 3 1 Cultureware Description The AutovaxID cultureware is the sterile single use wetted disposable component of the system All cultureware are provided sterile and ready for use They are intended for single use only and must be disposed of per your facility s requirements after the cell culture run is complete There are several models of cultureware that have various components and features to fit a variety of applications and uses Cultureware models fall into two types standard and simplified Simplified cultureware are described in chapter 8 Standard cultureware are described here There are two versions of standard cultureware that differ only in how the factor bags are provided and prepared Factor bags are described in the next section Current Version part number series 600320 XXX is provided as a single sterile assembly The cultureware is connected to the factor bags These factor bags
162. f you wish to access the AutovaxID from off site or from a different physical or logical network please contact you IT department for assistance Typically a secure virtual private network VPN is required for off site access though your IT department may have other preferred connection methods 10 1 Home Page The first web page that appears after pointing your web browser to the IP address of the instrument see above can be used to monitor the instrument lt provides a summary of the status of the instrument and displays the same information as the Monitor screen see Figure 10 1 Nothing on this screen can be modified The web page will automatically refresh every sixty seconds one minute Note At this refresh rate some sensor behavior may be overlooked If an alarm has occurred but the remote instrument status appears normal do not simply clear the alarm without any further investigation All alarms should be investigated and cleared directly at the instrument If there are no active alarms the State field top left will be displayed with a White background see Figure 10 1 Conversely if there is at least one active alarm the State field will have a red background see Figure 10 2 AutovaxID Operations Manual 10 1 Biovest International Inc Remote Access INSTR NAME Windo et Explorer 10 x A E Figure 10 1 Remote Access Home Page No Alarms AutovaxID Operations Manual 10 2
163. fault Settings are also presented for comparison The values in the fields in the Previous Run Settings column may be changed by an authorized user and will be used for subsequent cell culture runs Unless the instrument is running in strict mode these parameters may be changed by an authorized user during run setup As an added convenience authorized users may program up to twenty customized run setting profiles on each AutovaxID These profiles are stored on the AutovaxID and are not accessible from any other AutovaxID instrument These profiles may be loaded by any operator during run setup Profiles are especially useful when the instrument is operated in strict mode to ensure that the run parameters are set properly See Section 6 1 3 for more information on enabling strict mode AutovaxID Operations Manual 5 21 Biovest International Inc Primary User Interface Screens The run settings that are saved in each profile include e IC factor addition volume e EC factor addition volume e Cycling delay e EC Rise time e EC Fall time e Outflow pump settings e Media pump settings e Factor pump settings e Harvest pump settings e Outflow container volume e Media container volume e Factor container volume e Harvest container volume e Circulation pump settings e pH setpoint e Media temperature setpoint e Refrigerator temperature setpoint e Process control alarm settings enabled disabled high low limits To save a pro
164. file for future use an operator should first make any desired changes to the settings on this and the following Run Settings screens Run Settings 1 2 and 3 and Alarm Settings 1 and 2 and then press the Save Settings button on the Run Settings 1 screen The operator may save the settings over an existing profile or to create a New Entry see Figure 5 20 To save over an existing profile press the appropriately labeled button and then press Enter To create a new profile press the New Entry button press Enter and then use the on screen keyboard to enter a new name to uniquely identify the run settings The new profile will appear as a button on the Save Settings screen The default run settings are not configurable and the Default profile cannot be deleted To use the default run settings or the run settings from an existing profile press the Load Settings button and then select the appropriate profile To delete a profile press the Remove Settings button press the appropriately labeled button on the Remove Settings screen see Figure 5 21 and then press Enter AutovaxID Operations Manual 5 22 Biovest International Inc Primary User Interface Screens e SAVED SETTINGS 1 Figure 5 20 Save Settings Screen s pam SETTINGS 1 ae m Figure 5 21 Remove Settings Screen AutovaxID Operations Manual 5 23 Biovest International Inc Primary User Interface Screens 5 4 1 5 4 2 5 4 3 5 4 4 Facto
165. fter each clamp is opened in order to proceed to the next screen From the Setup Run 3 screen press Next to continue Jan 12 2007 Setup Run 3 15 57 51 Check that the container clamps are open Outflou Container Clanp Media Container Clanp Factor Container Clamp Check each box then select Next to continue Hold Exit Back Next Figure 7 4 Setup Run 3 Screen AutovaxID Operations Manual 7 18 Biovest International Inc Operating Procedures 7 4 5 Instrument Cultureware Pressure Test An automatic pressure test is performed to confirm the integrity of the cultureware before initiating a new cell culture run Pressure tests can also be performed for periodic servicing or to assess equipment functionality The pressure test will begin automatically after the operator presses Next on the Setup Run 3 screen After the pressure test is complete a pass or fail result will be given Jan 12 2007 Pressure Test 16 00 30 Waiting 2 seconds Pressure Testing IC Pressure mmHg 189 0 EC Pressure mhg 1189 0 Hold Exit Help Figure 7 5 Pressure Test in Process The pressure test sequence is 1 Pressurizing The cultureware is pressurized to approximately 200mmHg during 120 seconds 2 Stabilizing The cultureware is allowed to equilibrate for twenty seconds 3 Waiting The IC and EC will be monitored for pressure loss for sixty seconds During the pressure test t
166. ge to bring up the AutovaxID user interface see Figure 10 3 Using this Remote Control View an operator can interact with the instrument as if they were using the touchscreen Access to all instrument functions is the same as if you were using the AutovaxID software on the instrument itself Refer to other sections of this manual for assistance Likewise all events are automatically logged though there is no indication that any particular action was performed remotely The Remote Control View does NOT automatically refresh It will only refresh upon user input such as clicking a field or typing an alphanumeric character or by clicking the Refresh button For active monitoring it is recommended that the home page be used instead of the Remote Control View CAUTION If user names and passwords are in use be sure to log off before closing the web browser or walking away from your computer An operator using the touchscreen on the instrument may make changes to the instrument settings using your user permissions if you do not log off Likewise you should log off from the touchscreen to prevent other users from making changes remotely while using your permissions If you have forgotten to log off re open the web browser return to the remote control view and click the Logoff button Alternatively you will be logged off automatically by the AutovaxID system if a logoff timeout has been programmed However for security reasons this is no
167. glucose and lactate levels e product concentration and yield e cell viability These established values and other information are used to obtain target values for the process control strategy all cell lines have different requirements Before beginning a culture run a full characterization of the cell line should be completed For further information on establishing process parameters contact Biovest Account Services AutovaxID Operations Manual 7 2 Biovest International Inc Operating Procedures 7 2 2 Media Preparation Approximately two weeks before the bioreactor is planned to be inoculated media preparation for the run should begin unless the facility plans to purchase pre made media The media must contain all of the necessary nutrients adequate sodium bicarbonate buffer and any additional supplements required by the cell line The choice of media is usually determined by the cell requirements and is often established empirically Depending upon the cell line approximately 100 150 liters of ICS media is needed per month during a cell culture run ICS media is pumped into the system as required by the individual cell line from media containers that typically contain only enough media to be consumed by the cell line over seven to ten days As each container is consumed it is aseptically changed by an operator see Section 7 17 Twenty to fifty liter media storage bags are commonly used media containers smaller than
168. goff Timeout setting see Sections 6 1 7 the operator may be automatically logged off during cell removal If this occurs simply continue the cell removal and log back in as necessary The cells and product that have been collected in the syringe s may be retained for processing per your facility s requirements and procedures AutovaxID Operations Manual 7 56 Biovest International Inc Operating Procedures 7 21 7 22 Cell Removal for End of Production EOP Cell Bank Prior to shutting down the cell culture run cells may be withdrawn for an End of Production Cell Bank and or as test samples if required Cells may be drawn from the Inoculate EC Sample port Follow the EC Sampling Procedure see Section 7 14 but draw a larger EC Sample volume as required by your facility The cells should be removed and processed for storage according to your facility s requirements End of Run At the end of a cell culture run two screens instruct the operator how to stop the run and remove the cultureware From the Monitor Screen press Menu gt Run Menu gt Stop Current Run Press Yes to confirm that you would like to stop the run see Figure 7 42 then follow the on screen instructions to place the instrument into the IDLE state and safely remove the cultreware see Figure 7 43 and Figure 7 44 WARNING Failure to use the stop sequence to end a cell culture run may prevent the rotary valves from returning to the load position R
169. he actual pressure readings for both the IC and EC are displayed at the bottom of the screen in mmHg see Figure 7 5 The pressure test passes if the EC and IC lose no more than 5mmHg during the duration of the Waiting portion of the procedure conversely the pressure test fails if the pressure drops more than 5mmHg in either the IC or the EC AutovaxID Operations Manual 7 19 Biovest International Inc Operating Procedures Sixty seconds after a successful pressure test see Figure 7 6 the system will automatically advance to the first FILL FLUSH step Alternatively the operator may press the Next button on the Pressure Test screen to proceed immediately Jan 12 200 16 19 43 ee Pressure Test Passed cw je Figure 7 6 Pressure Test Passed If the pressure test fails the operator should attempt to fix the cause of the failure and retest with the assistance of Biovest Account Services personnel Alternatively the operator may press Exit to stop the current run and replace the cultureware if so instructed by Biovest Account Services personnel If the pressure test fails the following procedure may be used to attempt to locate and correct the cause of the failure 1 Repeat the clamping instructions see Setup Run 2 and Setup Run 3 screens to ensure the clamps are open or closed appropriately 2 Ensure that all tubing is centered in its respective clamps and their entire diameter is clamped Even a small por
170. he pump Detail screen to bring up the on screen keypad Use the on screen keypad to enter the Container Volume If no container is used for example if the media supply is unlimited or the outflow is not pumped into a container the operator may enter zero 0 for the Container Volume If no container is used no container alarms will be issued From the pump Controller screen press Exit to return to the pump Detail screen From the pump Detail screen press Exit to return to the Monitor screen Pump Rate Change When a pump is operated in manual mode an authorized user may change the pump rate by touching the rate field on the appropriate pump Detail screen The pump Detail screen can be accessed by pressing the appropriate pump field on the Monitor screen The pump rate may be entered using the on screen keypad and will be checked by the AutovaxID software before it is implemented to confirm that is within the acceptable range for that particular pump Valid pump rates will be implemented immediately if the flowpath is active not IDLE All changes to the pump rates will be automatically logged as events Advanced manual overrides to the pump rates entered by administrators on the Advanced Overrides screen will supersede any pump rates that are set on the pump Detail screens Cell Removal from the EC Reservoir During a cell culture run cell mass may collect in the EC reservoir This cell mass typically does not affect the function of
171. he unlocked position There are two in the pH probe assembly and two in the cultureware s tubing adapter between the pH IN LO and pH OUT HI labels 11 Press and hold the latch of the short pH IN LO tubing line connected to the disposable pH probe assembly Pull the quick coupler fitting off of the probe case and discard it 12 Press and hold the latch of the cultureware s E pH IN LO quick coupler fitting Disconnect this junction Figure 8 19 Hollow and connect the cultureware s pH IN LO quick coupler Fiber Bioreactor Fluid fitting to the pH probe s exposed port from the Dynamics previous step Ensure the latch faces upward 13 Release the latch and rotate the locking ring to the locked position Simplified Depiction Showing Only Some 14 Repeat the previous three steps for the E Cultureware Tubing pH OUT HI connection 15 Proceed to Factor and Harvest Container Options to continue with run startup Figure 8 20 Cultureware s External Tubing pH IN LO and pH OUT HI AutovaxID Operations Manual 8 19 Biovest International Inc Simplified Cultureware Usage 8 11 Factor and Harvest Container Options CAUTION CAUTION Simplified cultureware do not include integral Factor and Harvest containers These containers must be purchased and prepared separately from the cultureware There are options for the type of container to use Examples include glass or plastic bottles with ported caps
172. hed into the syringe If the cell mass is compacted or is not being pushed out the operator may use the syringe to agitate or extract the contents of the EC reservoir 8 Multiple collection syringes may be required Close the clamp on the Cell Removal line before aseptically switching the syringes 9 When the EC reservoir has been sufficiently drained close the clamp on the Cell Removal line The collection container may remain in place to protect sterility Alternatively the collection container may be aseptically removed and the Cell Removal line plugged using a new syringe the original sterility protector or a sterile male Luer plug 10 Touch the Cycle State field on the Overrides screen and override the cycle state to Low 11 Touch the Harvest EC valve field and remove the valve override by pressing the Clear button on the on screen keypad The valve will revert to 2 Brx amp Cycle 12 The EC reservoir will begin to re fill using fluid from the IC Leave the Cycle State overridden to Low for five to ten minutes or until the EC reservoir is sufficiently re filled and the IC reservoir is approximately half full 13 Touch the Cycle State field on the Overrides screen and remove the cycling override by pressing the Clear button The Cycle State will revert to Rise or Fall 14 Repeatedly press the Exit button to return to the Monitor screen 15 Press the Logoff button Depending upon the duration of this procedure and the Lo
173. ia is transferred from the IC to the EC the media level in the IC reservoir decreases and the media level in the EC reservoir increases hence it is referred to as EC Rise EC Fall When the fluid in the EC reservoir reaches a defined upper limit the AutovaxID process control system reduces the IC pressure and increases the EC pressure Similar to EC Rise the EC pressure is increased pneumatically at the EC reservoir When the EC pressure exceeds the IC pressure media flows from the ECS into the ICS taking with it small molecular weight wastes such as lactate and ammonia Since the media level in the EC reservoir decreases during this portion of the cycle it is referred to as EC Fall When the level of media in the EC reservoir reaches its defined minimum level the process starts over with another EC Rise During both EC Rise and EC Fall the IC and EC pressures are continuously monitored and controlled by the AutovaxID system to achieve the EC Cycling parameters that have been setup by the operator see Section 5 4 The AutovaxID process control software allows an operator to program two independent parameters of EC Cycling 1 the EC Rise time and 2 the EC Fall time The transfer volume between the IC and EC cannot be programmed and is approximately 70mL To attain these cycling parameters the system uses optical sensors and a limit flag to monitor the high and low fluid level in the EC reservoir thereby triggering the swi
174. if the threshold value is 10 then when the AutovaxID software has calculated that the harvest container is within 10 of being full an alarm will be issued to notify the operator that the harvest container should be changed Jan 25 2007 Harvest Control ler 7 24 16 JJ JJ Controller Type Tine Cycle Container Overflow Alarm Controller Setup Used 7 At 10 7 Container Volume 3000 nl Hold Exit Help Figure 5 31 Harvest Controller Screen AutovaxID Operations Manual 5 37 Biovest International Inc Primary User Interface Screens 5 5 10 e Default Harvest Container Volume 3000mL e Harvest Container Range 0 99999mL in increments of 1mL e Default Alarm Threshold 10 e Default Alarm State Enabled Checked The Harvest Controller Types are e None No control is performed or allowed the pump is used in override strict mode only e Manual The harvest pump rate can be entered manually at any time by an authorized user Each manual pump rate change is automatically logged e Schedule The harvest pump rate is scheduled to change at user defined times to user defined rates All changes to the schedule are automatically logged all scheduled pump rate changes are automatically logged See Section 5 5 10 for details on setting up a schedule e Time Cycle The harvest pump can be set to turn on and off based upon a time based schedule where an authorized user may specify the pump rate delay
175. igure 2 8 The connector end with the ferrite is attached to the instrument s Ethernet port Figure 2 8 Ethernet Cable with Ferrite AutovaxID Operations Manual 2 8 Biovest International Inc Instrument Setup Biovest recommends connecting the AutovaxID to a closed network only not to a company wide network nor to the public internet We further recommend that access to this network require a secure login and that physical access to the network is controlled As these issues are not relevant to the actual operation of the AutovaxID it is the responsibility of each facility to assess and implement the level of security necessary for their own operations and compliance requirements Biovest does not provide technical assistance for remote computer hardware and operating systems computer networks or computer security For information related to security as it may relate to the operation of the AutovaxID refer to FDA s 21 CFR Part 11 2 9 Bar Code Scanner Connection The AutovaxID is supplied with a handheld barcode scanner that can be used to enter information into the software The barcode scanner plugs into a nine pin connector on the rear of the instrument beneath the Ethernet port see Figure 2 1 2 10 Power Up After all accessories and gassing lines have been connected plug the power cord into the power entry module Turn on the AutovaxID by flipping the black switch on the power entry module to the I
176. increasing medium pH If CO is higher than 0 0 and the Air flow rate is 120 mL min reduce the CO wait 20 minutes re check pH and repeat until the desired pH is achieved Once CO is 0 0 the minimum rate and pH is still too low increase the Air flow rate from 120 mL min wait 20 minutes re check pH and repeat until the desired medium pH is achieved Proceed to the section below Factor and Harvest Container Options to continue with run startup AutovaxID Operations Manual 8 11 Biovest International Inc Simplified Cultureware Usage 8 9 Using a Reusable pH Probe The following procedures are performed for the reusable pH probe for each production run validation assembly autoclave sterilization and installation into the cultureware Note all pH probe items see also the Parts List section the AutovaxID and cultureware 8 9 1 Reusable pH Probe Validation 1 The reusable pH probe can be used for numerous production runs and is discarded when it no longer functions Its functional status is determined by validation tests that must be done before each run The pH probe validation tests determine the probe s inherent functionality and are not affected by calibration accuracy It is unnecessary to calibrate the pH probe before validation pH probe validation requires fresh pH buffers 4 0 and 7 0 to test the probe s response Check the buffers expiration dates 2 The reusable pH probe
177. ing materials from the pH probe assembly 3 Tighten the probe case s lower section that was loosened prior to autoclave sterilization 4 Insert the pH probe assembly into the black clip mounted on the front of the cultureware Slide the pH probe assembly downward until it rests on the screw post Rotate the pH probe assembly to the right so the pH IN LO port points toward the right see figure 8 15 5 The pH probe assembly will replace the cultureware s tubing adapter between its quick coupler fittings Simplified Depiction labeled pH IN LO and Showing No pH OUT HI This adapter is colored Cultureware Tubing dark green for clarity in figure 8 16 The pH IN LO tubing emerges from the lower right front edge of the cultureware enclosure The pH OUT HI tubing enters the lower left hand side of the cultureware enclosure Figure 8 15 Insert the reusable pH probe assembly into the cultureware s black clip AutovaxID Operations Manual 8 16 Biovest International Inc Simplified Cultureware Usage 6 Rotate the locking rings of the four quick coupler fittings to the unlocked position There are two in the pH probe assembly and two in the cultureware s tubing adapter between the pH IN LO and pH OUT HI labels 7 Press and hold the latch of the autoclave tubing line that is attached to the pH probe case s pH IN LO port Remove the entire autoclave tubing line from the probe case and discard it 8 Press and hold the la
178. instrument can no longer remove CO via the GEX In these cases a separate means of pH control may be required e g base addition via factor feed Note Manually overriding the pH control may adversely affect lactate control AutovaxID Operations Manual 3 3 Biovest International Inc Introduction to the Process Control Software 3 2 Lactate Control Glucose is typically used as the primary energy source for in vitro maintenance of mammalian cells As a result anaerobic metabolism or glycolysis leads to the build up of lactic acid which at high levels can limit growth and production In perfusion culture an expanding cell population combined with continual glucose delivery leads to steadily increasing rates of lactic acid production In response to this it is important to increase the feed rate which consequently increases the rate at which lactic acid is removed from the bioreactor Since lactate is the primary waste product created by mammalian cells a lactate control system was designed to automatically optimize the cell culture environment and continuously monitor the metabolic activity of the cells To accomplish this the ability to monitor lactate has been coupled with automatic adjustment of media feed rate to maintain the culture at the predetermined setpoint Since lactate concentration is an indicator of metabolic activity this can be predicted by measuring the concentration of CO in the offgas from the GEX and the
179. irc CabT 17 Jan 2007 13 42 50 Alarm Pump head not latched 17 Jan 2007 13 42 58 Alarm acknowledged 17 Jan 2007 13 42 58 Alarm cleared Pump head not latched 17 Jan 2007 13 47 23 7 10 0 0 0 0 0 00 0 37 0 4 0 65 7 20 0 100 20 0 0 200 30 7 17 Jan 2007 13 51 54 Alarm Pump head not latched 17 Jan 2007 13 52 23 7 10 0 0 0 0 0 00 14 37 0 4 0 65 7 20 0 100 20 0 0 200 30 7 17 Jan 2007 13 57 23 7 10 0 0 0 0 0 00 14 37 0 4 0 65 7 20 0 100 20 0 0 200 30 7 17 Jan 2007 13 59 15 Alarm acknowledged 17 Jan 2007 13 59 15 Alarm cleared Pump head not latched 17 Jan 2007 14 02 23 7 10 0 0 0 0 0 00 0 37 0 4 0 65 7 20 0 100 20 0 0 200 30 7 lt lt Top lt Back 11 to 20 0 111 Nep End gt gt Home Figure 10 7 Log Data Web Page The log can be navigated by clicking on the appropriate buttons on the web page lt lt Top lt Back Next gt End gt gt The lt lt Top and lt Back buttons do not appear if you are at the beginning of the log likewise the Next gt and End gt gt buttons do not appear if you are at the end of the log 10 4 Alternative Remote Access There are other ways to remotely access the AutovaxID instrument using other systems LIMS MES etc or programs Please contact Biovest International Account Services for assistance AutovaxID Operations Manual 10 9 Biovest International Inc Specifications 11 Specifications WARNING TO REDUCE THE RISK OF ELECTRIC SHOCK DO
180. is 3cc syringe will protect the sterility of the Inoculate EC Sample port and can be used to flush the Inoculate EC Sample line prior to taking the next EC Sample AutovaxID Operations Manual 7 49 Biovest International Inc Operating Procedures 7 15 7 16 7 17 Harvest Supernatant Sampling Procedure Post Filter After the Harvest pump has been started product Supernatant will be collected in the harvest bag and should occasionally be measured for content Depending upon the type of product secreted by the cells this measurement may be in units of concentration or activity The supernatant may be filtered or unfiltered depending upon whether or not the Cell Filter Bypass clamp has been closed or opened respectively Three 3cc syringes with male Luers are required to take a Supernatant sample To take a Supernatant sample from the harvest bags 1 Aseptically connect a 3cc syringe to the Supernatant Sample port 2 Unclamp the Supernatant Sample line and fill the syringe to flush the line 3 Close the Supernatant Sample clamp 4 Aseptically disconnect the 3cc flush syringe and replace it with a new 3cc syringe The 3cc flush syringe may be discarded per your facility s requirements 5 Unclamp the Supernatant Sample line again and remove the volume required for the off line assays Close the Supernatant Sample clamp 7 Aseptically disconnect the 3cc sample syringe and replace it with a new 3cc syringe Leave
181. istrator To logon an individual operator must remember their own username For security reasons the list of valid usernames is not accessible to regular users Also for security reasons user passwords are not accessible and may only be reset not read by an administrator Users may change their own password Usernames may not be changed AutovaxID Operations Manual 6 19 Biovest International Inc Software Setup amp Configuration 6 6 3 Logon Logoff Status There are screen features that can be used to identify if a user is logged on and what privileges that user has These features are only in effect if users have been setup First a Logon button will appear at the bottom of the Monitor and Menu screens if no one is logged on see Figure 6 16 If a screen requires that an authorized user is logged on before it can be accessed a window will appear with a single Logon button see Figure 6 17 Jan 26 2007 INSTR NAME 13 47 08 Elapsed Time 0 02 39 State GROWTH pH 7 09 Cycle Off EC Pressure 0E ml 71 lactate fp 10 7 actate 0 00 IC Pressure 0g Media Temperature 87 30 Refrigerator Temp 3 8 Gage Harvest 0 ml 93 ME Hold Menu Logon Help Figure 6 16 Monitor Screen with Logon Button Factor E hr If a user is logged on a Logoff button will appear at the bottom of the Monitor and Menu screens see Figure 6 18 Also a special character will appear in th
182. it Help Figure 5 28 Factor Pump Schedule Screen AutovaxID Operations Manual 5 33 Biovest International Inc Primary User Interface Screens The event that is currently in use is highlighted see the first event in Figure 5 25 If no events are currently active because not enough time has yet elapsed the highlighted area will appear after the last entry see Figure 5 28 An authorized user may enter a new value for any Factor rate at any time during a run The Elapsed Time may also be changed for any event by an authorized user so long as the schedule has not been advanced or delayed The Advance and Back buttons are used to shift the time of all events backward or forward These buttons are intended to be used if the cell culture is growing faster or slower than originally expected The time shift is displayed at the bottom of the screen To skip ahead to the next scheduled event press the Advance button to revert to a previous event press the Back button To remove any advance or delay touch the time shift field at the bottom of the screen to the left of the Advance button and enter zero 0 using the on screen keypad 5 5 7 Factor Pump Schedule Tied to Harvest Pump Schedule If both the factor and harvest controller types are set to Schedule and the Tie to Harvest box is checked on the Factor Controller screen see Figure 5 27 then pressing the Controller Setup button on the Factor Controller screen will bring up
183. ized user The event that is currently in use is highlighted see the first event in Figure 5 25 If no events are currently active because not enough time has yet elapsed the highlighted area will appear after the last entry see Figure 5 29 An authorized user may enter a new value for any Factor or Harvest rate at any time during a run The Elapsed Time may also be changed for any event by an authorized user so long as the schedule has not been advanced or delayed AutovaxID Operations Manual 5 34 Biovest International Inc Primary User Interface Screens Jan 25 2007 Factor Pump Schedule 8 53 14 Factor Punp Schedule Elapsed Tine Fac Haru Elapsed Tine Fac Haru Delete 3 00 00 1 1 Delete 10 00 00 2 2 q All tines are on schedule 0 00 00 Advance Back Add Event Hold Exit Help Figure 5 29 Factor Pump Schedule Screen w Tie to Harvest Enabled The Advance and Back buttons are used to shift the time of all events backward or forward These buttons are intended to be used if the cell culture is growing faster or slower than originally expected The time shift is displayed at the bottom of the screen To skip ahead to the next scheduled event press the Advance button to revert to a previous event press the Back button To remove any advance or delay touch the time shift field at the bottom of the screen to the left of the Advance button and enter zero 0 using the on screen keypad
184. known as the Starling Effect and can be detrimental to the culture as it can push growth factors and cells to the distal end of the bioreactor AutovaxID Operations Manual 1 2 Biovest International Inc Introduction 1 1 2 AutovaxID Technology Difference To overcome the problems of the Starling Effect in conventional hollow fiber systems Biovest employs patented EC Cycling technology EC Cycling re engineers fluid flow dynamics within the hollow fiber bioreactor EC Cycling achieves a uniform perfusion of the high density cell culture The result is a more homogeneous culture environment with minimal gradients and anoxic micro environments The benefits of EC Cycling include optimized culture growth increased longevity and improved productivity Think of the AutovaxID system as having two separate fluid circuits one for the cells and cell components the Extracapillary Circuit EC and one for the growth media and oxygen transfer the Intracapillary Circuit IC The Intracapillary Circuit connects to the ICS In and ICS Out ports of the bioreactor Similarly the Extracapillary Circuit connects to the ECS In and ECS Out ports of the bioreactor The Intracapillary and Extracapillary Circuits intersect only at the fiber membrane of the bioreactor These two fluid circuits with their fluid reservoirs and other components constitute the flowpath of the system 1 1 3 The Intracapillary Circuit The Intracapillary Circuit IC is th
185. l to bring up an on screen numeric keypad for data entry Each alarm can also be enabled or disabled by checking or un checking the Used checkbox respectively e Low When a sensor value is lower than the Low limit an alarm is issued e High When a sensor value is higher than the High limit an alarm is issued NOTE The Low value must be less than or equal to the High value See Table 3 4 for the High Low alarm limits Jan 16 2007 Alarm Settings 2 11 28 09 Alarns Used Low High Lactate MI 0 000 5 000 mal Cabinet Temperature 15 0 5 0 Cycle Tine E 30 ninutes Hold Exit Back Help Figure 5 38 Alarm Settings 2 Screen AutovaxID Operations Manual 5 47 Biovest International Inc Primary User Interface Screens If enabled alarms are activated on the following schedule e The Lactate alarm will be active at the factor addition step of PRE INOCULATION and will remain active until the end of the run e The Cabinet Temperature alarm is always active including when the instrument is in IDLE state e The Cycle Time alarm is active whenever cycling is on A Cycle Time alarm will be issued if an EC Rise or Fall takes longer than the desired cycle time plus High Cycle Time programmed here For example if the EC Rise time is set for fifteen minutes and the High Cycle Time alarm limit it set to thirty minutes as shown in Figure 5 37 an alarm will occur if an EC Rise takes longer than 45 minutes
186. l volume of the hollow fibers within the bioreactor see figure 8 2 The ICS provides exchange of small Figure 8 3 Hollow molecules such as O CO nutrient and Fiber Bioreactor metabolic wastes with the cell culture which is Fluid Dynamics separated from the ICS by the hollow fiber membrane wall AutovaxID Operations Manual 8 5 Biovest International Inc Simplified Cultureware Usage 8 5 2 Extracapillary Circuit Components Bioreactor the ExtraCapillary Space or ECS is the volume surrounding the outside of the hollow fibers within the bioreactor see figure 8 2 The ECS is where the cell culture resides The molecular weights of added growth factors and secreted proteins are most often too large to pass through the hollow fiber membrane so they are retained within the ECS Factor is fresh media that often is supplemented with growth factors and in the cultureware the location where it is pumped into the bioreactor s ECS Factor Prime is used with syringes for two reasons 1 with each connection of a new full container of Factor to clear the Factor tubing line of any air that may be present to prevent it from entering the ECS 2 to inject high molecular weight growth supplements if necessary For further information see Using Factor Prime at the end of this section Harvest is used to pump supernatant out of the ECS to capture the secreted protein of interest Inoculate EC Sample is used with syringes to
187. later during the Pre inoculation phase Figure 8 7 7 Rinse the probe and place it in pH 4 buffer BNC to Two Pin Adapter that is in a container that can stand upright Part Number 103150 000 by itself Do not hold the pH probe at this time 8 When the pH A D value reasonably stabilizes ensure the value is gt 4000 Ignore the pH Value because the pH probe has not been and need not be calibrated 9 If the A D value for either pH buffer is out of range ensure the buffers have not Figure 8 8 AutovaxID s Two Pin expired and try other bottles of buffer If pH Probe Connection the A D value still is out of range the probe may be faulty Cleaning instructions were provided with the probe which may resolve the problem Also ensure the pH probe cable and BNC to two pin adapter are firmly connected and that the pH cable s ground pin is inserted in the instrument s ground jack AutovaxID Operations Manual 8 13 Biovest International Inc Simplified Cultureware Usage 8 9 2 Reusable pH Probe Assembly a electrical cap 1 The pH probe assembly includes the pH probe amp cap a probe case amp its three cit o rings and a set of pH probe autoclave tubing lines see figure 8 9 The probe case is very durable and t lasts for many runs The probe case o rings and autoclave tubing lines should be replaced for each run threaded connechon probe case upper section arge o ring probe case i lower se
188. le waiting for the operator to initiate INOCULATION and during INOCULATION e Fill Flush Media Rate Range 20 50 mL hr in 1mL hr increments e Default Fill Flush Media Rate 20 mL hr 6 4 5 Harvest Line Volume This field is a correction factor to the calculated accumulated harvest volume to account for the volume of air in the harvest line and filter This value is applied to the harvest volume calculation at the start of SETUP changes made to this value after SETUP has begun are not applied to the current run The Harvest Line Volume should only be changed from the factory setting of 70mL if the harvest container configuration is changed e Harvest Line Volume Range 0 999 mL in 1mL increments e Default Harvest Line Volume 70 mL NOTE The Harvest Current Volume on the Harvest Detail screen will be a negative value equal to the Harvest Line Volume at the start of a new cell culture run AutovaxID Operations Manual 6 10 Biovest International Inc Software Setup amp Configuration 6 5 Diagnostic Menu Screen WARNING System function can be severely affected by changing values in any diagnostic screen or clearing the database Operators should not use these screens except under the direction of a Biovest Account Services representative The Diagnostic Menu is accessed by pressing the Diagnostics button on the Setup screen Monitor gt Main Menu gt Setup gt Diagnostics The Diagnostic Menu see Figure 6
189. libration Menu 8 Aseptically disconnect the 5cc syringe from the IC Sample port and connect a new 10cc syringe Leave the IC Sample line clamp closed 9 Take an off line measurement of the pH sample and then press 2 on the Calibration Menu 10 Enter the off line measurement on the pH Calibration Entry screen see Figure 7 22 and then press Exit Jan 16 2007 pH Control WAB Control On pH Setpoint 7 10 Gas Exchange Cartridge Current pH 7 10 002 15 0 y Air 120 ml min A I Combined Gas Number 87 5 Calibrate ph pH Calibration Offset 0 00 Hold Exit Help Figure 7 35 pH Control Screen AutovaxID Operations Manual 7 44 Biovest International Inc Operating Procedures dan 16 2007 Calibration Menu 14 26 04 SS Take a Sanple d Enter ph Measurenent ER Enter Lactate Measurenent Sample Jan 16 2007 10 29 27 pH 7 05 AD 1920 Corrected to pH 10 Hold Exit Help Figure 7 36 pH Calibration Menu 7 10 pH Calibration Schedule Use the following schedule to perform pH calibrations e Perform the initial required pH calibration after FILL FLUSH e Calibrate pH 3 days after the initial pH calibration e Calibrate pH 7 days after the initial pH calibration After these first three pH calibrations additional calibrations may be performed at the discretion of the operator or facility Off line measurements may be used to evaluate the necessity of pH calibrations
190. ll culture is growing faster or slower than originally expected The time shift is displayed at the bottom of the screen To skip ahead to the next scheduled event press the Advance button to revert to a previous event press the Back button To remove any advance or delay touch the time shift field at the bottom of the screen to the left of the Advance button and enter zero 0 using the on screen keypad See Section 5 5 7 for details on setting the Harvest schedule when it is tied to the Factor schedule AutovaxID Operations Manual 5 39 Biovest International Inc Primary User Interface Screens 5 5 11 Harvest Pump Time Cycle Controller Setup If the harvest controller type is set to Time Cycle then pressing the Controller Setup button on the Harvest Controller screen will bring up the Time Cycle screen see Figure 5 33 This screen allows an authorized user to setup a time based controller for the harvest pump The time cycle controller sets the pump to act in a repeating cycle defined by the elapsed time Note The Time Cycle pump controller is NOT related to EC Cycling and does not affect the EC Rise or EC Fall times Jan 25 200 Tine Cycle 8 27 15 Harvest Pump Tine Cycle Controller Auto Rate 1 Delay Until 0 00 00 Total Cycle Time 10 00 00 Un Time 10 00 00 Hold Exit Help Figure 5 33 Harvest Pump Time Cycle Controller Screen Auto Rate The desired pump rate to be used when
191. local area network LAN Remote access will permit an operator to monitor the AutovaxID status and make process control changes from outside the laboratory using a standard web browser A LAN connection and external computer are also required for archiving production logs Please refer to Section 10 for details about accessing the system remotely CAUTION Security is the responsibility of the end user To minimize security concerns the AutovaxID should be connected to an isolated network that only allows secure logons Biovest does not recommend connecting the AutovaxID to the company wide network or to a network that is accessible via the public internet 1 3 8 AutovaxID Distinctions The AutovaxID system offers the advantage of compliance with FDA s 21 CFR Part 11 guidelines for a closed system and continuous production while retaining ease of use both in terms of operating procedures and operator training AutovaxID Operations Manual 1 13 Biovest International Inc Instrument Setup 2 Instrument Setup The initial setup of the AutovaxID system requires electrical power a CO supply and an optional LAN connection Routinely these connections are made only once at the time of initial installation disconnection is not required between production runs Biovest recommends that the AutovaxID installation and operation be qualified when used in a regulated environment Equipment qualification IQ OQ is an optional se
192. ls are needed e What volume of cell suspension can be processed safely e How much time is required to achieve an adequate seeding density Anchorage dependent cells may require a higher inoculum density than suspension cells Re suspending them in conditioned media rather than fresh media may improve the initial growth period by providing the cell specific growth factors secreted by the cells Re suspension volumes for the inoculum may also affect the initial lag time after inoculation Best results are often obtained by preparing the inoculum from cells that are in the mid log phase of growth A standard growth curve is often useful to track the progress of cells in static culture to coincide with the day of inoculating the flowpath Seeding a bioreactor more than once may be considered an option in the following situations low inoculum density low inoculum viability or poor initial growth Ideally the first two situations should be avoided altogether but time constraints or other factors may take precedence over preparing a suitable inoculum In the third case poor initial growth in the AutovaxID cultureware may be attributed to many conditions that can be avoided prior to inoculation Cytotoxicity could be assessed before cells are added to the bioreactor if repeatable significant lag times occur However the specified flush volume is known to make the flowpath non cytotoxic Confirm that the proper protocols are followed The ch
193. lt contact Biovest Account Services 6 Remove the pump cassette from its shipping location in the cultureware cover Rotate it downward and install it into the pump panel in the front of the instrument beneath the cultureware shelf Insert the pump cassette into the panel until stiff resistance is felt Then while gently pushing the pump cassette towards the instrument rotate the knob on the left side of the instrument towards the front of the instrument clockwise when looking at the knob from the left 1 4 turn to draw and lock the pump cassette into the pump The cultureware is now installed and is ready for use AutovaxID Operations Manual 7 13 Biovest International Inc Operating Procedures 7 4 Run Settings The following sections assume that you have read and understand Sections 5 and 6 at minimum They also assume that you have a sufficient understanding of hollow fiber cell culture theory how to operate the AutovaxID user interface and that the instrument has been setup properly Please review the previous sections as necessary before attempting to start a cell culture run 7 4 1 Setup Run From the Monitor Screen navigate to the Setup Run screen see Figure 7 1 by pressing Menu gt Run Menu gt Start New Run Touch the Run Name field and enter a description for the Run Name This can be the cell line name batch number lot number instrument number or any unique set of numbers and or letters to ide
194. lture run Section 8 provides specific information to operate simplified cultureware which are useful when small scale productions or development runs are necessary Finally the last sections provide details about troubleshooting remote access and instrument specifications We have also provided a list of consumable items that you may need to re order from Biovest or your local supplier We hope that this manual will assist you in understanding how an AutovaxID cell culture run is managed from start to finish As you become familiar with the system you may choose to modify procedures and to apply more details to the operation of the system Please contact Biovest Account Services with any questions or concerns that you may have that are not covered by this manual AutovaxID Operations Manual iv Biovest International Inc Service and Support Service and Support Contact Biovest International s Account Services department for e Technical Service e Field Service e Customer Service e Sales Information We are always willing to discuss your questions suggestions or problems with the AutovaxID cell culture system Account Services is available weekdays from 8 00 5 00 Central Time GMT 6 Our contact information is on the manual s cover page After hours telephone callers hear recorded instructions for emergency technical assistance Keep a copy of our telephone number posted near the AutovaxID for quick reference Technical
195. ltureware s Factor and Harvest lines terminate in female Luer fittings the Factor and Harvest containers must be configured to include at least one male Luer connection When using disposable media bags at least one of their integral tubing line s must terminate in a male Luer fitting When using glass or plastic bottles their corresponding bottle caps liquid In port for Harvest or Out port for Factor must have an attached tubing line that terminates in a male Luer fitting When using media bags as Factor and Harvest containers their integral tubing line s do NOT need an internal diameter larger than 1 16 inch although such small tubing is not commonly available Tubing with an internal diameter of inch often is available and compatible for use with the AutovaxID Contact Biovest International for additional technical and purchasing information about specific container options configurations and requirements The Parts List section at the end of the manual includes some of the options that are available from Biovest Factor and Harvest containers and their connected tubing and fittings must be sterile before connection to the cultureware AutovaxID Operations Manual 8 20 Biovest International Inc Simplified Cultureware Usage 8 12 Starting a Run Using Simplified Cultureware 1 Reminder choose the methods and equipment for pH control Factor and Harvest containers and pumping equipment relevant when using the th
196. n Jan 24 2007 Qutflow Detail 13 19 30 T M 2 Qutflow 100 vr 20 00 L Capacity 0 14 L Current Volune Change Container Settings Hold Exit Help Figure 5 11 Outflow Detail Screen AutovaxID Operations Manual 5 13 Biovest International Inc Primary User Interface Screens 5 1 13 Media Detail Screen On the Monitor screen press the Media field or the surrounding area to bring up the Media Detail screen see Figure 5 12 This screen shows the media pump position the media pump rate in mL hr the media container capacity and the estimated volume of fluid currently in the media container The pump illustration also shows the direction of pump rotation see the arrow above the pump and the direction of fluid flow see the arrow next to the pump tubing The media pump rate may be automatically controlled by the AutovaxID process control software or it may be controlled manually by an authorized user The controller type is defined on the Media Controller Screen see Figure 5 24 which can be accessed by pressing the Settings button The media controller is typically set during run setup see Section 5 5 See Section 5 5 2 for details about each of the media pump controller types If the media pump is being automatically controlled by the process control software it may be overridden only by an authorized user using an advanced override see Section 5 12 2 Biovest does not recommend manually ov
197. n minutes at which status information will be AutovaxID Operations Manual 6 4 Biovest International Inc Software Setup amp Configuration automatically recorded to the current log during a cell culture run This setting only affects status lines status lines give a snapshot of the system state Status logging begins with FILL FLUSH e Status Logging Range 1 to 60 minutes in one minute increments e Default Status Logging Value 5 minutes NOTE Do not set the Status Logging interval to zero 0 Setting the interval to zero will disable status logging even if the Status Logging On box is checked enabled 6 1 9 Status Logging On This check box is used to indicate whether status logging is turned on or off An authorized user can check or uncheck this box to turn status logging on or off respectively Status logging is automatically enabled during FILL FLUSH PRE INOCULATION INOCULATION and GROWTH and the status lines are logged at the user defined interval To disable status logging for the duration of a cell culture run set the Status Logging interval to zero see 6 1 8 6 2 Network Configuration Screen The Network Configuration screen is accessed by pressing Next on the System Features screen Monitor gt Main Menu gt Setup gt Configure gt Next The Network Configuration screen see Figure 6 4 allows an authorized user to setup the IP address so that the AutovaxID can be connected to a LAN enable or disable
198. ncluding those described on the following pages 7 27 6 Preventive Maintenance Biovest International also specifies a three year interval for preventive maintenance on all AutovaxID instruments The purpose of preventive maintenance is to minimize the risk that the system will fail or prematurely end a production run Preventive maintenance replaces all wearable components of the system and includes calibration Contact Biovest Internationals Account Services department for information about calibration preventive maintenance and other service options AutovaxID Operations Manual 7 66 Biovest International Inc Simplified Cultureware Usage 8 Simplified Cultureware Usage 8 1 8 2 Simplified Cultureware Description Biovest International recognizes the need for lower cost cultureware that are tailored for small scale productions R amp D or early process development For these needs we have engineered simplified cultureware options that work without modification of the AutovaxID equipment you already own Simplified cultureware eliminate components that are unnecessary for the intended uses yet retain the most important features They are pre assembled and pre sterilized for fast turn around time and support pH control and other parameters listed below If and when production requirements scale beyond simplified cultureware you can immediately switch to using standard cultureware which allow larger culture volumes Aut
199. ncnnnnrnnnancnnaninana 4 1 Bede SALUD E 4 2 4 2 User lntertaceteabres arinaa e aa a a i a i a aa 4 3 GB Ee DATE CT F PE E EAE A E oboe EOE EAE E N AN E NE 4 3 4 2 2 Information Panels ib 4 3 Se ULO AIEEE EEEE A E A A EE OA 4 3 ADA A E E E 4 4 43 Data EMI eege deeg 4 5 GE rd Den Biet Relata a nT Os o ei 4 7 43 2 Date e o A E E E Ee ebe AA 4 7 4 3 3 Selection Eeler ein dente nod e elenco 4 7 4 3 4 Schedule Elei t desse et ce cov dak Cece Dee ere eels heh Gata Sen det ere eet 4 7 A AOD E 4 8 4 5 Remote Computer ACCESS ccc hese vests circ notire ces cece en EE 4 9 4 6 Access Restrictions Eine er aE A ETE E E a A a 4 9 5 Primary User Interface Screens oioiionrrrria nnmnnn nenna 5 1 S b MONTO SC CON ita A EE EES 5 1 GL Elapsed TIMO iii de 5 2 NS A E EE E 5 2 aa TOI H Eu ti EE 5 4 ET MEN BA cias 5 7 5 1 5 Help BUON saeta ist 5 7 56 een EE 5 7 5 1 7 PH Control SEEM aa a aE eE E E RE EEE EE A E E 5 8 bel Ee A 5 9 5 1 9 Cycle Detail Sereno cece catdecs sec panencus cacteeceaters cqugaeoswcngnertaqencessgccnierecquheesegenanes 5 9 5 1 10 Media Temperature Screen ett eege Eeer eg 5 11 5 1 11 Refrigerator Temperature Screen ENNEN 5 12 5 112 Outllow Detail Sr ocio n wv NEE EENS 5 13 5 1 13 Media Detail Screen iii ceda 5 14 51 14 Factor Detall Screen E A A A e See 5 15 5 1 15 Harvest Detail Screen a e aE EE conan co na nc cnc rc nana 5 17 5 1 16 Circulation Detail Gcreen conc nc ona nnn narco na nannn 5 18 52 Mal Mellad
200. ncrements of 0 01 AutovaxID Operations Manual 5 44 Biovest International Inc Primary User Interface Screens 5 6 2 5 6 3 Media Temperature Setpoint An authorized user may specify the desired temperature setpoint for the circulating media e Default Media Temperature 37 0 C e Media Temperature Range 15 to 45 C in increments of 0 1 C The ability for the instrument to control the media temperature to the full specified temperature range depends upon the ambient temperature The incubator can only heat the media the AutovaxID cannot cool the incubator or the circulating media Refrigerator Temperature Setpoint An authorized user may specify the desired temperature setpoint for the refrigerator e Default Refrigerator Temperature 4 0 C e Refrigerator Temperature Range 2 to 25 C in increments of 0 1 C Press the Next button at the bottom of the Run Settings 3 screen to advance to the Alarm Settings 1 screen AutovaxID Operations Manual 5 45 Biovest International Inc Primary User Interface Screens 5 7 Alarm Settings 1 Screen The Alarm Settings 1 screen see Figure 5 37 allows an authorized user to view and change the high and low alarm limits for pH Media Temperature Refrigerator Temperature IC Pressure and EC Pressure Pressing a Low or High field will bring up an on screen numeric keypad for data entry Each alarm can also be enabled or disabled by checking or un checking the Used checkbox
201. nd fourteen days after inoculation The harvest bag set has a total capacity of three liters and is typically kept at 4 C throughout the cell culture run 7 26 5 Factor Requirements Pre treatment of the bioreactor ECS with factor is an important step in conditioning the flowpath before inoculation Although the inoculum should contain the same amount of factor as in static culture lack of factor in the EC would dilute the inoculum to an undesirable level Therefore factor should be added to the EC prior to inoculation This procedure is described in Section 7 6 2 Starting the Factor pump is at the discretion of the facility and or operator Some components of factor are consumed while others notably enzymes are used as tools for biochemical reactions The Factor pump should be started when either of these functions components is below optimum level Monitoring growth and productivity parameters will give an indication of the need to add factor Cultures that show a drop in metabolic activity may require additional factor to AutovaxID Operations Manual 7 63 Biovest International Inc Operating Procedures 7 26 6 recover A syringe can be used to inject a quantity of factor into the EC or the Factor pump can be used to deliver additional factor from the factor bags As acell culture matures it may become less dependent on the factor To reduce the factor content in this event the Factor pump rate can be reduced while main
202. ng AutovaxID Operations Manual 3 6 Biovest International Inc Introduction to the Process Control Software System alarms include Table 3 3 System Alarms Alarm Description Startup possible clock failure HHH MM SS Logged on program startup if the current time is found to be less than the previously recorded time Time printed is recorded time minus the current time Possible battery or single board computer hardware failure Startup down time Day Hr Min Logged on Time printed is current time minus the previously recorded time Instrument data recovery failed Logged on power up if the database cannot be restored Outflow pump stalled Media pump stalled Factor pump stalled Harvest pump stalled Outflow pump racing Media pump racing Factor pump racing Harvest pump racing The peristaltic pump stall and race alarms will be logged when the pumps are running and the condition is detected The condition is detected at the pump I O board based upon sensors that detect movement on the pump output shaft Circulation pump stalled The pump I O board has detected a stall condition The I O board will report a stall condition if the detected flow rate is under the setpoint for longer than four seconds after the pump reaches maximum voltage The alarm is reset when the setpoint is reset to zero ora flow exceeding the setpoint is detected The circulation pump will restar
203. ng the concentration of the gasses dissolved therein Specifically the gassing system controls the pH by modifying the CO concentration supplied to the IC media through the GEX Similar to the in vivo environment an increase or decrease in dissolved CO concentration causes a corresponding shift in bicarbonate equilibrium to maintain a steady pH For example increasing the CO flow rate increases the dissolved CO in the media which in turn increases acidity thereby decreasing the pH Conversely by increasing the flow rate of air and reducing the CO flow rate to zero the GEX removes CO from the cell culture media thereby increasing the pH The gassing system provides gas mixtures represented by a percent CO in the mixture OR by the flow rate of air in mL min 500 mL min of air is the most potent for increasing pH twenty percent CO is the most potent for decreasing pH Air flow is adjustable from 120 to 500 mL min CO concentration is adjustable from zero to twenty percent with a fixed air flow rate of 120 mL min NOTE CO and air overrides are mutually exclusive that is you can override CO concentration OR airflow rate but not both at the same time AutovaxID Operations Manual 3 2 Biovest International Inc Introduction to the Process Control Software A gas number may be used to describe the mixture of air and CO The gas number represents a combined flow of air and CO Table 3 1 shows several gas numbers and
204. ng on firmware that are separate from the internal control computer Instrument control functions include flowpath control and sensor monitoring The primary function of AutovaxID software is to control the cell culture environment It is important that process control maintain a cell culture under constant conditions when required but that it also has the ability to manage varying conditions during the different phases of a production run The AutovaxID software provides the following features e Automatic pH control of the flowpath to user specified pH by mixed gas e Automatic or user specified controls for the media factor harvest and circulation pumps e Automatic control of the outflow pump e Automatic refrigerator temperature control to user specified temperature e Automatic media temperature control to user specified temperature e EC and IC pressure control per user specified cycling parameters e Automatic media pump control based upon a lactate concentration algorithm optional e User modifiable alarms for pH media temperature refrigerator temperature IC pressure EC pressure lactate cabinet temperature and cycle time e Operator identification e Automatic event logging e Electronic data storage and retrieval e Full system state recovery from power failures AutovaxID Operations Manual 3 1 Biovest International Inc Introduction to the Process Control Software Remote system access to export log data for an
205. ng will not appear Once INOCULATION has begun the AutovaxID will show the State as INOC on the Monitor screen Tine fron Factor Addition has exceeded 10 hrs Push Wl to repeat Factor Addition Step Push NEXT to continue without repeating Tine since factor addition 0 00 00 Figure 7 30 Inoculation Warning AutovaxID Operations Manual 7 38 Biovest International Inc Operating Procedures 7 7 Inoculation Procedure The following steps are recommended just prior to inoculation Inspect all culture vessels microscopically before pooling to verify the presence of healthy culture in mid log phase of growth The most important factors in inoculation are the health viability and density of the inoculum Variability in the inoculum has a lasting effect on the outcome of the cell culture run If applicable pool all culture vessels and withdraw a sample for counting purposes The initial lag time for growth in the bioreactor will be less if they are in mid log phase of growth Pull any required samples for testing as defined by the facility or operator for example Sterility Mycoplasma etc If required concentrate the cell inoculum by centrifugation and then pool all cell pellets after re suspension in conditioned or fresh media No more than 50 ml may be used for inoculation A larger volume may cause flooding of the IC and wetting of the IC bag vent filter Refer to the Optimization Procedure Section 7 26
206. nician interaction 8 4 Simplified Cultureware Removed Features Missing from the simplified cultureware are the EC Cycling technology components because EC Cycling is unnecessary for the intended uses Simplified cultureware also do nat include the built in pH probe integral Factor and Harvest containers automated Factor Addition to the Intracapillary Circuit prior to inoculation and in line harvest filtration The following list describes what a user needs to do regarding these missing features EC Cycling although simplified cultureware do not have EC Cycling its absence does not need to be replaced by another mechanism pH Control Operators must choose one of three options for controlling pH These options are described below in pH Control Options Factor amp Harvest Containers Operators must supply sterile containers to deliver Factor and collect Harvest These options are described below in Factor and Harvest Container Options Automated Factor Addition to the IC Pre inoculation this is optional and is easily replicated using manual methods At the time when Factor Addition is normally performed as instructed by the on screen prompt use a syringe to inject the necessary volume of growth supplement into the IC Sample line For further information contact Biovest s Account Services department In line Harvest Filtration this is optional and can be replicated if desired Operators must select a filter s that is compatible
207. nitor screen press the pH field the CO2 or Air field depending upon the gassing status or the surrounding area to bring up the pH Control screen see Figure 5 6 This screen allows an authorized user to enter the pH setpoint perform a pH calibration observe the CO and air flow rates and determine if the pH control system is on or off checked or unchecked respectively An authorized user may change the pH setpoint by pressing the pH Setpoint field and entering a new target pH using the on screen keypad Similarly an authorized user may perform a pH calibration by pressing the Calibrate pH button See Section 7 9 for details on performing a pH calibration pH control may be manually turned on or off only by an administrator as a diagnostic function Similarly the gas mixture may be overridden only by an authorized user using an advanced override see Section 5 12 2 Biovest does not recommend turning off pH control or modifying the gas mixture if the instrument is functioning normally changing these settings may negatively affect the condition of the cell culture environment Jan 24 2007 pH Control 10 05 53 Control On 7 pH Setpoint 7 10 Gas Exchange Cartridge Current pH 7 10 0021 15 0 y Air 120 ml min A SS Combined Gas Number 87 5 Calibrate pH pH Calibration Offset 0 01 Hold Exit Help Figure 5 6 pH Control Screen AutovaxID Operations Manual 5 8 Biovest International Inc Prima
208. noculation or Exit to Menu Media Rate nl hr 20 Factor Rate ml hr 0 pH 7 05 C02 20 00 Hold Exit Next Help Figure 7 29 Pre Inoculation Step 11 AutovaxID Operations Manual 7 37 Biovest International Inc Operating Procedures At the end of the PRE INOCULATION stage the instrument will wait indefinitely until the user starts the inoculation procedure see Figure 7 29 During this time the Media pump will continue to run at a minimum rate 20 mL hour to ensure that there is sufficient media in the flowpath though the factor in the cultureware will become continually more diluted For best results the AutovaxID cultureware should be inoculated within ten hours of the completion of Factor Addition If more than ten hours has elapsed since Factor Addition was performed a warning will be issued prior to INOCULATION see Figure 7 30 On the Inoculation Warning screen the operator is given the option of repeating the Factor Addition or continuing on to INOCULATION Though the recommended procedure is to repeat the Factor Addition it is ultimately at the discretion of the operator or the facility s operating procedures Always wait until the cells are ready for inoculation before advancing the AutovaxID into INOCULATION NOTE If the user enters INOCULATION by pressing Next after Step 11 of PRE INOCULATION see Figure 7 29 but then waits more than ten hours before inoculating the bioreactor the warni
209. ns 5 9 1 Alarm Handling Users should always review each alarm and take the necessary corrective action to correct the alarm condition Alarms should be selectively cleared once a parameter resumes operating within its normal range Clearing the alarm restores a particular alarm s ability to recur should the parameter go out of range in the future To clear an expired alarm use the Previous and Next arrow buttons until the desired alarm is displayed in the alarm window and then press the Clear Alarm button Active alarms cannot be cleared until the condition causing the alarm is corrected Except for startup alarms if an alarm remains in the queue active inactive or acknowledged future instances of the parameter going out of range will not cause a second alarm To silence the audible alarm if enabled and reset the red title bar to a non flashing state while an alarm is active an authorized user must press the Acknowledge button Pressing the Acknowledge button silences all currently active alarms Any new alarms that occur will cause the audible alarm to sound again if enabled and will start the title bar flashing again Any active alarms that are acknowledged will remain active and cannot be cleared until corrective action is taken to correct the alarm condition 5 9 2 Log Entry An authorized user may make a manual log entry by pressing the Log Entry button The log entry may be created using the on screen keyboard or the
210. nted on paper batch records having a backup electronic record of the on going run ensures a complete production record e Provides the user with the ability to export production data to standard programs for inclusion in status reports for analysis in custom spreadsheets for printing and attachment to batch records etc See Section 10 3 1 Saving a Log for more information about using the Download Text function of remote access More information about the data logs is provided in Section 5 9 Alarm Log Screen Note The CompactFlash card may not be removed by the operator to backup run logs The log data must be downloaded using a network attached computer as described AutovaxID Operations Manual 3 10 Biovest International Inc Introduction to the User Interface 4 Introduction to the User Interface The AutovaxID user interface allows the operator to setup and control the AutovaxID system The user interface hardware and software is controlled by the internal single board computer and includes e the LCD touchscreen e the alarm system e archiving retrieval and display of flowpath data and run logs current and historical e remote access interface e user account creation and validation e data entry validation e power failure recovery All user interface functions required for instrument operation are accessible through the LCD touchscreen The software screens provide a selection of related display and input cells
211. ntify the particular run up to twenty five characters When in Strict Mode the Run Name is required Touch the Cultureware Lot Number field and enter the unique Cultureware Lot Number The cultureware lot number is provided in three locations on the outside of the shipping carton on the certificate of compliance and on an adhesive label on the cultureware cover When the on screen keyboard is displayed the lot number can be manually typed in or it may be scanned from any of the labels using the barcode scanner When in Strict Mode the Cultureware Lot Number is required Use the Run Information field to optionally enter any other miscellaneous information about the cell culture run Run Name Cultureware Lot Number and Run Information are automatically entered into the log Press the Run Settings button to review the run settings to load the default or previously saved run settings or to save the current settings see Sections 5 4 5 8 From the Setup Run screen press the Next button to continue AutovaxID Operations Manual 7 14 Biovest International Inc Operating Procedures Setup Rim NN geg ae A Figure 7 1 Setup Run Screen AutovaxID Operations Manual 7 15 Biovest International Inc Operating Procedures 7 4 2 Setup Run 1 On the Setup Run 1 screen see Figure 7 2 sequentially select the Outflow Media and Factor Container ID fields and enter their identifying information When the on
212. ntities appropriate for mid scale production The AutovaxID is managed by the AutovaxID User Interface Application its built in process control software The AutovaxID User Interface Application is an automated system that provides the means to create an optimal environment for cell growth and production The AutovaxID combines system control with system flexibility The AutovaxID system consists of e The AutovaxiID cell culture instrument e A disposable single use sterile flowpath 1 1 Hollow Fiber Theory The AutovaxiID cell culture system is based on hollow fiber technology Its sterile disposable flowpath or cultureware integrates a hollow fiber bioreactor wherein both suspension and anchorage dependent cells can be grown The AutovaxID system monitors and controls the flowpath environment mimicking a mammalian body so that the cells maintain optimal growth and production In fact using the analogy of a mammalian body we can describe how the system functions The hollow fibers inside the bioreactor function like capillaries cells are grown in the space surrounding the outside of the hollow fibers cell culture growth media the blood of the system flows through the hollow fibers The media carries fresh nutrients and oxygen to the cells while carrying away cell wastes such as ammonia and CO Media is circulated through the bioreactor using a circulation pump that acts as the heart of the system The lungs of
213. nual Factor Addition to the EC can be used as when using standard cultureware with the difference that the factor is drawn from the user supplied factor container At this time Factor Addition to the IC should also be performed if desired however it is usually unnecessary When desired inject factor via the IC Sample port using a syringe To determine the amount of factor volume to be used during Factor Addition use the volumes below into which the factor will be diluted e Approximate EC volume is 85 mL single 1 1 m bioreactor or 50 mL for each of the three 0 5 m bioreactors e Approximate IC volume is 315 mL single 1 1 m bioreactor version or 310 mL three 0 5 m bioreactor version Proceed with Inoculation as suggested in the Operating Procedures section of the manual When the on screen instruction appears to inject the inoculum inject each of the three bioreactors when relevant The remainder of the run will proceed similarly to what is described in the Operating Procedures section of the manual However there is one new aspect to using simplified cultureware that is not described there Factor Primer usage because standard cultureware do not have this feature see below AutovaxID Operations Manual 8 24 Biovest International Inc Simplified Cultureware Usage 8 13 Using Factor Prime Factor Prime is used primarily to remove the air that is present in each new Factor container s connected or integral tubing
214. o high preparation Review the media specification and compare it to the recommended media preparation outlined in Section 22 The instrument is out of media Attach a new container of media to the Media line Erroneous pH calibration Perform a pH calibration see Section 7 9 Contact Biovest Account Services There is an override on the gas blend Clear the override to the gas blend Incorrect media or media preparation Review the media specification and compare it to the recommended media preparation outlined in Section 7 2 2 pH is too low Erroneous pH calibration Perform a pH calibration see Section 7 9 The offgas line has become obstructed Inspect the offgas line and remove any obstruction or replace the line Contact Biovest Account Services Faulty cable or connections pH is not stable Override the valves to the Load Unload position and then place the instrument on Hold Unlatch and then re install the cultureware Resume the pumps from Hold and clear the valve overrides Verify that the cultureware latches are closed tightly Malfunctioning pH probe Confirm the pH of the media using an offline method Use Gas overrides to manually control the pH through rest of run Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIB
215. o odiada EE 5 19 53 IRUAUMGNU ta ais A e ia iia 5 19 AutovaxID Operations Manual vii Biovest International Inc Table of Contents p4 R n Settings T Screen 5 21 54h Factor ee HEH 5 24 54 2 Factor Add RE 5 24 5 4 3 Cycle Delay Time 00000 See be Sege Sei 5 24 By AAs EG ISS IAS ance cg ee cece Aare ae eg E ac erties eng a 5 24 BAS EG Fall Tie sats cease teat oth a uaa ei coats ee Tea bce CO 5 25 5 95 RUM Setings 2 cord ais 5 25 5 5 1 e inte en ee EE 5 26 5 5 2 Media Controller resseta as 5 27 5 5 3 Media Pump Schedule Setups acute cele d Set da 5 29 5 5 4 Media Pump Lactate Controller Setup ooooooooocccccnnnccccnncanoncccnoncccnnnnnananncnnnnnnnnnnnnns 5 30 50 59 eene TE 5 31 5 5 6 Factor Pump Schedule EE eege Ee 5 33 5 5 7 Factor Pump Schedule Tied to Harvest Pump Gchedule 5 34 5 5 8 Factor Pump Time Cycle Controller Setup ooooonccccnnncccocccccnnncconanannanncnnnncncnnnnnns 5 35 5 5 9 Harvest age UE 5 37 5 5 10 Harvest Pump Schedule Setup kee ENEE 5 38 5 5 11 Harvest Pump Time Cycle Controller Setup ooocccconnccoocccconnccconnnanaaanccnnncncnannnos 5 40 53 12 CirullatiGty Controlan 5 41 5 5 13 Circulation Pump Schedule Getup Au 5 42 5 6 RUM SeMINGS 3 Screen EE 5 44 Eech JEE ee a EE 5 44 5 6 2 Media Temperature Sopot erte eters tats te et aed tet acter Een 5 45 5 6 3 Refrigerator Temperature Getpoomt AEN 5 45 5ST Alarm Settings 1 O EES 5 46 5 8 Alarm Settings Z Scr en eme SENNENG 5 47 O EE ooh ee aaa ee
216. o the Process Control Software Table 3 4 Alarm Limits Alarm Range Default Alarm Values Alarm Low High Low High pH 6 0 8 0 6 80 7 40 Media Temperature 15 0 C 45 0 C 35 0 C 39 0 C Refrigerator Temperature 0 C 25 0 C 2 0 C 10 0 C IC Pressure 0 mmHg 300 mmHg 0 mmHg 275 mmHg EC Pressure 0 mmHg 300 mmHg 0 mmHg 275 mmHg Lactate 0 mg mL 5 001 mg mL 0 mg mL 5 000 mg mL Cabinet Temperature 0 C 80 1 C 15 0 C 55 0 C Cycle Time 0 minutes 999 minutes NA 30 minutes Process control alarms are issued only when the flowpath is active not if the system is in the IDLE state System alarms are issued no matter the system state but the type of log entry alarm or event may vary based upon the system state If an alarm is enabled its corresponding sensor value is continuously checked to determine if it falls within the range specified by the user If the value is beyond the limits of the user specified range e The value in the corresponding field in the Main Monitor screen is changed to red characters e The title bar is displayed in blinking red e A message is logged using red characters showing the sensor reading that was out of range and whether it was too low or too high e The audible alarm is turned on if enabled see Section 6 The alarm is turned off by pressing the Clear Alarm button or the Acknowledge button on the Alarms screen see Section 5 9 Alarm Log Screen
217. oltage Selection for fuse and voltage selection The AutovaxID is equipped with an IEC 60320 power connector to facilitate easily switching the power cord to comply with local plug configurations Each facility must provide the appropriate power cord if the cord provided with the AutovaxID is incompatible with their local power connection The AC power connection power switch AC voltage selection and AC fuse s are contained in the power entry module see Section 2 5 Voltage Selection 2 3 Unpacking Instructions The AutovaxID is very heavy Two people are required for unpacking Handle the AutovaxID with care DO NOT DROP Failure to follow these instructions could result in injury to the person s lifting or carrying the instrument and or damage to the instrument 1 Place the crate on the floor and remove the top cover 2 Remove the top foam and center tube section AutovaxID Operations Manual 2 3 Biovest International Inc Instrument Setup 2 4 2 5 3 Two people working together should lift the AutovaxID to the desired location Do NOT lift the instrument by grasping the refrigerator door Grip the sides and the back of the instrument only Remove and discard the anti static wrapping from the instrument Inspect the instrument for any shipping damage immediately report any damage to Biovest Account Services or your local supplier The barcode scanner power cord harvest factor bag rack and gas connecti
218. on into the cultureware with minimal preparation Unlike the reusable pH probe the disposable pH probe figure 8 17 requires the use of a non sterile reusable holder figure 8 18 to firmly hold the disposable pH probe assembly in the black clip that is mounted on the front of the cultureware The reusable holder is necessary because the diameter of the disposable pH probe assembly is much smaller than the diameter of the black clip Note both pH probe items see also the Parts List section are purchased separately from the AutovaxID and cultureware Do NOT lift the cultureware from the shipping box by its sterile bag Lift it out by supporting the cultureware from its bottom surface 1 Prepare the simplified cultureware using the instructions in the section Cultureware Preparation under Operating Procedures Steps related to the Factor and Harvest Bag Sets do not apply and the refrigerator rack isn t needed because simplified cultureware do not have integral Factor and Harvest bags The technician Fiqure 8 17 must provide and prepare sterile Factor and Harvest g containers separately described below Disposable pH Probe Assembly Sterile bag and packaging not shown Disinfect the reusable pH probe holder and place it in Part Number 600305 000 the sterile laminar flow hood Inspect the disposable pH probe s sterile bag for holes or tears Disinfect the outside of the bag and place it in a sterile hood Remove and di
219. on B Software C 7 2010 Add Instructions to Use Simplified Cultureware U 4 2014 Add ferrite instructions section 2 8 update factor bag info in section 7 AutovaxID Operations Manual Biovest International Inc Symbols and Warnings Symbols and Warnings This symbol on the instrument indicates high voltage is present It is dangerous to make any kind of contact with any inside part of this product This symbol on the instrument indicates a warning to observe the maximum CO inlet pressure allowed Pressures above this value will cause the pressure relief valve to open venting CO into the room and possible damage to the instrument AutovaxID Operations Manual Biovest International Inc Symbols and Warnings WARNING WARNING WARNING CAUTION NOTE NOTE TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT REMOVE ANY PANELS FROM THE INSTRUMENT THERE ARE NO USER SERVICEABLE PARTS INSIDE REFER SERVICING TO QUALIFIED SERVICE PERSONNEL If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A properly grounded main supply must be used for safe operation of the equipment Connect to an appropriately grounded outlet Consult local electrical codes Wherever a caution statement is used the documentation needs to be consulted in order to find out the nature of the potential hazard and any action which may
220. on EE 7 7 7 3 3 Cultureware Preparacion cima ca a oo oda 7 8 7 3 4 Filling the Factor Bags resol ida io 7 10 7 3 5 Cultureware Installation EE 7 13 T4 e TRU SO MINS EE 7 14 TAN Setup Op et il lcd 7 14 Ke E eg Ti EE 7 16 7 4 3 Setup Run EEN 7 17 TAA SSUP Ti EE 7 18 7 4 5 Instrument Cultureware Pressure Test ENNEN 7 19 FER GU E ee 7 21 ZE Pre Inoculation Procedures 7 29 7 6 1 Initial PAI a A LS A ea 7 30 116 23 Fator Addition as a a ds 7 34 TT Ioculati n Procedure its 7 39 A A O 7 39 T2 WNOGUIATOM Step 2 a 7 40 78 een EE 7 42 7 9 Mid Run pH Calibrations tac aida eet ln ate 7 43 740 pH Calibration Schedules E 7 45 7 11 Mid Run Lactate Calibration cts A Es atada 7 45 7 12 Recommended UMA ASKS sia vcarabaitiehs gege Deg 7 47 7 121 Daily EE 7 47 TAa A eege Tage gees A 7 47 7 13 IC Sampling Procedure EE 7 48 7 14 EG Sampling Procedure Pre Filter cion Eesen 7 49 7 15 Harvest Supernatant Sampling Procedure Post Eiter 7 50 7416 Gell Filler Bypass Clamp EE 7 50 7 17 Replacing Pump Contenu 20gekegeEh EENS a ci 7 50 7 18 Container Capacity Change E 7 55 7 19 Pump Rate CMAN cite Meneame in oar Manes ae Banas ads 7 55 7 20 Cell Removal from the EC Reservoir ANNE EEN 7 55 7 21 Cell Removal for End of Production EOP Cell Bank 7 57 1 22 ENG Of RUM EE 7 57 7 23 Cultureware Removal and Disposal Procedure ooooooooccccccccccononcnoooncnnnncccnnnnannnncnnnnncnannnnns 7 59 PAO DAA ACA a a e ten all cats ee 7 59 2 25 e e ee TEE
221. on card see Figure 2 3 flip open Figure 2 3 Lift and Pivot to Open the Cover Pull the voltage selector card straight out of the power entry module Orient the voltage selector card so that the desired supply voltage is readable at the bottom see Figure 2 4 8 Move the voltage indicator pin to point up see Figure 2 4 230V 240V Figure 2 4 Voltage Selector Card amp Indicator Pin Orientation 2 5 AutovaxID Operations Manual Biovest International Inc Instrument Setup 9 Re insert the voltage selector card into the power entry module with the printed side of the card facing forward toward the power cord socket The edge containing the desired voltage should be inserted first 10 Replace the cover snap closed 11 Verify that indicator pin shows the desired voltage 2 6 Fuse Installation The fuse selection must match the voltage selection See the Main Fuse section in Specifications page 11 1 1 Turn off the power switch 2 Disconnect the power cord 3 Use a small flat blade screwdriver or similar tool to pry open the cover below the power cord socket see Figure 2 2 Pry at X Gently lift UP on the door until it lifts up approximately 1 4 see Figure 2 3 Once lifted the door will pivot on its hinges and expose the fuse holder and voltage selection card see Figure 2 3 6 Pull the fuse holder straight out of the power entry module see Figure 2 5
222. on fittings are all packaged inside the refrigerator Remove these accessories from the refrigerator and set aside Moving Instructions WARNING The AutovaxID weighs 66 Ibs 30 kg Handle with care to avoid personal injury to and or damage to the instrument See Section 2 3 for lifting instructions Prior to moving the AutovaxID if it was previously installed in another location disconnect the power cord network cable CO supply off gas line and barcode scanner from the instrument Remove the bag rack from the refrigerator Ensure that there is adequate clearance in the doorways hallways through which the instrument will travel A cart is recommended to move the instrument on smooth level surfaces Contact Biovest Account Services for additional instructions if the instrument is being moved under rugged conditions Voltage Selection The power entry module must be set to match the supply voltage To change the voltage setting 1 Turn off the power switch 2 Disconnect the power cord 3 Use a small flat blade screwdriver or similar tool to pry open the cover below the power cord socket see Figure 2 2 Pry at X Figure 2 2 Pry Open the Cover AutovaxID Operations Manual 2 4 Biovest International Inc Instrument Setup 4 Gently lift UP on the door until it lifts up approximately 1 4 see Figure 2 3 5 Once lifted the door will pivot on its hinges and expose the fuse holder and voltage selecti
223. ondition the bioreactor and to maintain cell viability as required by the cell line and defined by the operator Pressing the Factor button on the Run Settings 2 screen brings up the Factor Controller screen see Figure 5 27 This screen shows the factor pump controller type the factor container volume the state of the factor container alarm and the alarm threshold Authorized users can enable and disable the factor container alarm checked and unchecked respectively and enter an alarm threshold between 0 and 99 percent The alarm threshold defines when an alarm will be issued For example if the threshold value is 10 then when the AutovaxID software has calculated that the factor container is within 10 of being empty an alarm will be issued to notify the operator that the factor container should be changed e Default Factor Container Volume 3000mL e Factor Container Range 0 99999mL in increments of 1mL e Default Alarm Threshold 10 e Default Alarm State Enabled Checked AutovaxID Operations Manual 5 31 Biovest International Inc Primary User Interface Screens The Factor Controller Types are e None No control is performed or allowed the pump is used in override strict mode only e Manual The factor pump rate can be entered manually at any time by an authorized user Each manual pump rate change is automatically logged e Schedule The factor pump rate is scheduled to change at user defined times to user define
224. ong By default the instrument name is left blank Server Confirmation Required This mode is intended for use with a LIMS or MES system Checking this box adds confirmation requirements to the run sequence The server must then validate the entered data such as cultureware lot number media lot number inoculum ID etc before the operator is allowed to proceed Please contact Biovest Technical Service Personnel for additional information AutovaxID Operations Manual 6 2 Biovest International Inc Software Setup amp Configuration If Server Confirmation is enabled then a Server Confirmation screen see Figure 6 3 will be presented to the operator after data is entered in specific fields The server request confirmation screen is used to allow a LIMS system to check the values entered before an operator is allowed to continue This screen will pop up during run setup inoculation pump container change and run termination The LIMS system will then receive a flag in the normal status check operation and it will have to return an acknowledgement packet before the user is allowed to advance An authorized user is allowed to skip this screen by pressing the Next button or may return to the previous data entry screen by pressing the Exit button Jan 25 2007 Confirmation 16 39 58 Waiting for confirmation Inoculant identification Hold Exit Next Help Figure 6 3 Server Confirmation Screen 6 1 3 Strict Mode
225. ontrol tedio 6 7 6 3 Date and Time Seems e al a 6 7 e ie 6 8 e EE 6 8 6 4 Run Configuration 1 Mr a A etal coats 6 8 6 4 1 Auto Load Run SOUS conan di 6 9 6 4 2 Initial Circulation Rate ovni eee ae E E EEE EER E 6 9 6 4 3 Intial Media E 6 9 6 4 4 Fill Flush Media Rate un 6 10 6 4 5 Harvest Line e un 6 10 6 5 Diagnostic Menu SCLC EE 6 11 6 6 User ACCOUNTS Screenin See id 6 12 6 6 1 Administrator Account Setup coincida caseeceetedadees dee lcd 6 12 6 6 2 User Account Setup by an Administrator 6 18 6 6 3 LOGO Re Le 6 20 6 6 4 Initial Logon Procedure for New Ueers 6 22 6 7 Touch Cabra to ocre e AE EAEE EE EE ARE AEE 6 24 6 8 Network Log PriDliNd coda ia sda o teca 6 24 6 31 Printer IP Ggl iregeg eecht S e ged is 6 25 6 8 2 Pages per Ee ee act ea eh zee east a race east Ee 6 25 LE ON e ET EE 6 25 6 04 TESEPIN eege ebe tee en a 6 26 6 8 5 E EE 6 26 AutovaxID Operations Manual ix Biovest International Inc Table of Contents T Operating Procedures iaaa aara aa adaa Eaa aad eE a aeaa Naaa aiaa 7 1 Tel Outline of Thee EE 7 1 1 2 Biological Preparation EE 7 2 7 2 1 Cell Line CharacionzalO Macael ali 7 2 7 2 2 Media Preparation iia EREECHEN 7 3 7 2 3 Media Sterilization EE 7 4 7 2 4 Media Quality Control cid da dn Rae EES 7 4 7 2 5 Scale Up of Cells for Inoculation 00 2 2 ENNEN 7 4 7 3 Cultureware Setup Procedures EE 7 5 7 3 1 Cultureware DescrplilO viii 7 5 7 3 2 Factor Bag Set Descripti
226. op Current Run button is pressed the user will be taken through the steps required to terminate the run see Section 7 22 AutovaxID Operations Manual 5 19 Biovest International Inc Primary User Interface Screens BUS ES Figure 5 17 Run Menu Start New Run EJES ES Figure 5 18 Run Menu Continue Stop Current Run AutovaxID Operations Manual 5 20 Biovest International Inc Primary User Interface Screens 5 4 Run Settings 1 Screen Pressing the Run Settings button on the Main Menu screen will bring the operator to the Run Settings 1 screen see Figure 5 19 This screen allows access to a number of user definable run settings that can be modified by an authorized user These run settings may be reviewed by all users during run setup see Section 7 4 1 Only authorized users may make changes to these settings See the following sections for a description of each setting Jan 24 2007 hun Settings 4 16 41 06 Previous Run Settings Default Settings Factor Add Oil 100 100 Factor Add EC nl 100 100 Load Settings ld Settings _ Sue Settings Jee Settings Cycle Delay day hr min 7 00 00 7 00 00 Cycle Rise ninutes 15 15 Cycle Fall ninutes 15 15 Renove Settings Hold Exit Next Help Figure 5 19 Run Settings 1 Screen save Settings E For convenience the run settings are retained from run to run and are listed in the Previous Run Settings column The De
227. ote that online help is not a substitute for this manual It provides general information and does not discuss the finer points of system operation It is intended as a refresher to remind you of the purpose of screens fields valid entry ranges etc dan 23 2007 Help 14 29 53 The nonitor screen gives an overview of the current instrunent state The top of the screen includes the tine date and title bar The title bar flashes red when alarms are unacknouledged If the alarm is acknowledged and remains in the queue then the background of the title bar is light red The Elapsed time is a running tine since the start of a state change to Setup Growth or Idle Current sensor readings include pH EC Pressure IC Pressure uj AE Figure 4 4 Help Screen for Monitor screen AutovaxID Operations Manual 4 8 Biovest International Inc Introduction to the User Interface 4 5 4 6 Remote Computer Access Control and monitoring of a production run are typically done primarily using the integral LCD touchscreen but many aspects of run management may optionally be performed from a remote computer system using a standard web browser Remote access requires that the AutovaxID be connected to a facility s computer network LAN The remote computer can be either in the production room with the AutovaxID or at a remote location that is outside the clean room but tied to the same local network Such a setup facilitate
228. ovaxID instruments support standard and simplified cultureware without modification Simplified Cultureware Versions There are two versions of simplified cultureware One version contains a single 1 1 m bioreactor that has an approximate ECS volume of 110 mL The other version contains three 0 5 m bioreactors each of which has an approximate ECS volume of 50 mL These three bioreactors share a common IC Circuit but their ECS volumes are entirely independent of one another meaning that there are three Factor tubing lines labeled F1 F2 and F3 to deliver growth supplemented medium to each bioreactor individually There also are three Harvest tubing lines labeled H1 H2 and H3 to collect supernatant from each bioreactor individually Because the triple bioreactor version keeps the bioreactors ECS volumes segregated from one another it can be inoculated with one to three cell lines When multiple cell lines are cultured because the bioreactors share the IC media the IC media should be appropriate for all cell lines being cultured Alternatively the same cell line can be cultured in all bioreactors while the ECS media is varied to optimize growth and production parameters Considerations for using these two simplified cultureware versions e Inoculum cell number and density should be taken into consideration given their differing bioreactor ECS volumes e EC Factor Addition that occurs before inoculation also should take into account
229. ow Controller 13 49 00 Controller Type Outflow Container Overflow Alarm Used Z At 10 7 Container Volume 20 00 L Hold Exit Help Figure 5 23 Outflow Controller Screen 5 5 2 Media Controller The media pump delivers fresh cell culture media to the IC Pressing the Media button on the Run Settings 2 screen brings up the Media Controller screen see Figure 5 24 This screen shows the media pump controller type the media container volume the state of the media container alarm and the alarm threshold Authorized users can enable and disable the media container alarm checked and unchecked respectively and enter an alarm threshold between 0 and 99 percent The alarm threshold defines when an alarm will be issued For example if the threshold value is 10 then when the AutovaxID software has calculated that the media container is within 10 of being empty an alarm will be issued to notify the operator that the media container should be changed e Default Media Container Volume 20L e Media Container Range 0 to 999 99L in increments of 0 1L e Default Alarm Threshold 10 e Default Alarm State Enabled Checked AutovaxID Operations Manual 5 27 Biovest International Inc Primary User Interface Screens The Media Controller Types are e None No control is performed or allowed the pump is used in override strict mode only and its rate is defined on the Run Configuration 1 screen see Section 6 4 1
230. pH of the circulating IC media In other words the lactate concentration can be controlled by modifying the media feed rate the media pump adds fresh lactate free media to the flowpath while the outflow pump continuously removes lactate rich media Lactate control is therefore tied to the media pump control The target lactate concentration is defined by the operator during run setup If lactate control is in use then the lactate controller will continuously adjust the media pump rate to maintain the target lactate concentration In addition media consumption is optimized by using precisely the right amount needed to support the expanding culture and prevent the accumulation of metabolic waste The CO concentration in the offgas is measured by an electronic CO sensor that is integrated into the instrument and requires no user servicing This sensor is not wetted and is protected from offgas condensate by a condensate trap The pH is continuously measured using a disposable pH probe that is integrated into the sterile cultureware Both sensors provide feedback to the lactate controller All pH and lactate calibrations must be performed on schedule to ensure the accuracy of the lactate controller Lactate control may be initiated by an authorized operator at any time during an active cell culture run 3 3 Media Temperature Control The IC media temperature is continuously monitored and controlled by the AutovaxID to maintain it at a user define
231. ported by the AutovaxID software for operator reference The pH may be used to automatically control the gas blend through the gas exchange cartridge It also provides feedback to the optional lactate controller algorithm The connector mates automatically when the cultureware is loaded to the instrument by the operator e Cycling Sensors these two photoelectric sensors interact with the cycling mechanism in the cultureware to detect the high and low fluid level limits of the EC reservoir They are non contact sensors and are housed in the instrument cabinet behind a clear polycarbonate window The cycling sensors provide feedback to the cycling algorithm None of the components on the instrument side of the cultureware interface panel are wetted all of them stay with the instrument The offgas port one of the four gassing ports may occasionally come in contact with condensate from the gas exchange cartridge Condensate in the offgas is normal and will not harm the instrument AutovaxID Operations Manual 1 10 Biovest International Inc Introduction 1 3 5 Disposable Cultureware The AutovaxID cultureware is shipped pre assembled and pre sterilized by Biovest International Each cultureware set is for single use only and is disposable WARNING The cultureware is not intended to be reused Reusing a cultureware may pose a hazard to the operator and may cause undesired results and or damage to the instrument Each disposabl
232. proper biohazard procedures Re fill the container with bleach or other sanitary solution per your facility s requirements so that the end of the offgas line is submerged Periodic Tasks Perform pH calibrations per the recommended schedule see Section 7 10 Perform a lactate calibration if applicable due to media container change Change the Outflow and Media containers as required If in manual mode adjust the Media pump rate to maintain cell growth NOTE If utilizing the lactate or schedule control feature then manual AutovaxID Operations Manual 7 47 Biovest International Inc Operating Procedures adjustments to the Media pump are not required e If in manual mode adjust the circulation pump rate See Section 7 26 2 for recommended Circulation pump rates e As desired sample the ICS for metabolic parameters pH dissolved oxygen DO Glucose Lactate Glutamine Glutamate and Ammonia These data may be useful for cell line optimization but are not required They may be used to improve cell growth and productivity during a cell culture run The metabolic indicators will vary depending upon the individual cell line inoculation density doubling time etc If using manual pump controllers sampling the IC at least two to three times per week is recommended until the Media pump and Circulation pump rates are at their maximum levels e Sample the EC Bioreactor pre filter for current productivity using the Inoculate E
233. pumps and tiners Resune End Cell Culture Tine system has been on hold 0 00 09 Help From the Emergency Hold screen when the Resume button is pressed a pop up screen will request confirmation that the operator wants to resume see Figure 5 4 On the Resume Confirmation screen pressing the Yes button will return the operator to the screen that was in use prior to the instrument being put on hold pressing the No button will return the operator to the Emergency Hold screen Figure 5 3 Emergency Hold Screen Similarly from the Emergency Hold screen when the End Cell Culture button is pressed a pop up screen will request confirmation that the operator wants to abandon the cell culture see Figure 5 5 On the End Cell Culture Confirmation screen pressing the No button will return the operator to the Emergency Hold screen pressing the Yes button will initiate the End of Run sequence see Section 7 22 AutovaxID Operations Manual 5 5 Biovest International Inc Primary User Interface Screens A Ready to resume run Yes No Figure 5 4 Resume Confirmation STOP Are you sure you want to abandon this culture Yes No Figure 5 5 End Cell Culture Confirmation AutovaxID Operations Manual 5 6 Biovest International Inc Primary User Interface Screens 5 1 4 Menu Button The Menu provides access to the Main Menu screen See Section 0 for details Help Button The
234. quired Values outside the allowable ranges are not permitted Selection Fields Fields whose values are displayed in text form e g OFF LOW IDLE do not require the actual text to be entered Instead simply select one of the preset choices that are presented and then press Enter Schedule Field A schedule field is commonly used for scheduling events or cycling delays This field is a combination of a date and time field and uses the format DD HH MM where DD Days range 0 999 HH Hours range 0 99 MM Minutes range 0 99 AutovaxID Operations Manual 4 7 Biovest International Inc Introduction to the User Interface Enter the desired event time and then press Enter Entering 999 for days will result in a conversion to the maximum number of days allowed which are 365 Entering 99 for the Hours or Minutes will result in a conversion to the equivalent values for these units Entries of less than seven digits are allowed and will be interpreted according to character positions For example Data Entry Interpretation 9999999 365 00 00 999900 103 03 00 12300 001 23 00 210 000 02 10 10 000 00 10 1 000 00 01 4 4 Help Context sensitive help is available for most software screens Help information describes the screen s available functions buttons and restrictions To scroll through the information given in the help window use the available arrow buttons see Figure 4 4 Please n
235. r Place the pumps on Hold and then Resume the pumps A foreign object is in the pump and or the pump cassette Place the pumps on Hold remove the pump cassette inspect the pump and cassette for foreign objects reinstall and then Resume the pumps The pump tubing is not seated properly Place the pumps on Hold remove the pump cassette make sure the pump tubing lies in the pump cassette correctly reinstall and then Resume the pumps The instrument is experiencing Contact Biovest Account a hardware failure Services The Outflow Media Factor The instrument is experiencing Contact Biovest Account and or Harvest pump is racing a hardware failure Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION The instrument has registered a Board offline alarm A power disturbance has occurred Clear the alarm and monitor If alarm recurs contact Biovest Account Services A communication cable has Contact Biovest Account come loose Services The instrument is experiencing Contact Biovest Account a hardware failure Services The instrument has registered a Board reset alarm A power disturbance has occurred Clear the alarm and monitor If alarm recurs contact Biovest Account Services The instrument is experiencing a hardware failure Contact Biovest Ac
236. r Add IC An authorized user may specify the volume in mL of factor typically ECS feed media or other supplement to deliver to the IC during PRE INOCULATION e Default IC Factor Volume 100mL e IC Factor Volume Range 0 to 9999 mL in increments of 1mL Note the factor line is purged to the IC before factor addition begins The purge volume is not user configurable and it does not need to be accounted for in the IC factor volume To determine the IC Factor Volume consider that at EC Low the IC volume is approximately 370mL Factor Add EC An authorized user may specify the volume in mL of factor typically ECS feed media or other supplement to deliver to the EC during PRE INOCULATION e Default EC Factor Volume 100mL e EC Factor Volume Range 0 to 9999 mL in increments of 1mL Note the factor line is purged to the IC before factor addition begins The purge volume is not user configurable and it does not need to be accounted for in the EC factor volume To determine the EC Factor Volume consider that at EC Low cultureware with a 1 1m bioreactor has an EC volume of approximately 230mL a cultureware with a 2 1m bioreactor has an EC volume of approximately 300mL Cycle Delay Time An authorized user may specify the initial time delay before cycling is turned on during the GROWTH phase of the cell culture run The data entry format is days hours minutes day hr min If a delay is specified cycling is placed in the OFF state
237. re Start a new run and enter information for Run Name Cultureware Lot Number and Run Information on the Setup Run screen using the same procedures as for standard cultureware Enter information for Outflow Container ID Media Container ID and Factor Container ID on the Setup Run 1 screen using the same procedures as for standard cultureware However note that the Factor Container is user supplied when using simplified cultureware Using Container IDs is optional The user must determine if container IDs are useful and apply them to these containers when their use is desired Simplified cultureware differ in the name and number of tubing clamps from standard cultureware Close the following clamps rather than the list specified in the Setup Run 2 screen figure 8 21 e IC Sample e Inoculate EC Sample e Factor Prime Note simplified cultureware do not have told Bit Back the Supernatant Supernatant Sample Cell Removal and Cell Filter Bypass Figure 8 21 Setup Run 2 Screen clamps so these clamp instructions are not relevant when using simplified cultureware When the Setup Run 3 screen is reached figure 8 22 ensure the Media Outflow Factor and Harvest containers are connected to the simplified cultureware before proceeding with the clamping instructions The AutovaxID s on screen instructions do not prompt the technician to connect these containers Open the clamps listed on the Setup Run
238. re lt 3 L if the production run will be short i e approximately lt 45 days 3 Transfer the sterile factor into a sterile container s that has a cap or tubing line that is compatible with male or female Luer connectors 4 If the Factor container s is rigid it needs to be vented using a 0 2u hydrophobic filter If the Factor container s is a collapsible bag do not use a vent filter 5 Plug the container s Luer fittings 6 Store the factor solution at 4 C for at least 24 hours so it will be cold during the next step in the process Filling the Factor Bags described next In this manner the cell culture run can begin as soon as possible If the factor solution is not pre cooled it will take the AutovaxID refrigerator up to 24 hours to cool the factor which may affect the factor s potency re Fill the factor bags using the procedure relevant for the style of factor bags in use either the new version 600320 XXX or the Legacy version 600290 XXX 8 Fill the factor bags with minimal delays to minimize the amount the factor warms during the process of filling the bags AutovaxID Operations Manual 7 10 Biovest International Inc Operating Procedures Fill Factor Bags from 600320 XXX Standard Cultureware 1 Load a portion of the factor bags 72 inch 183 cm long size 16 silicone Factor Fill tubing line into a compatible size 16 Masterflex peristaltic pump 2 Verify that the tubing is loaded in the correc
239. re grown in the ECS growth media is circulated through the ICS The fibers perform two primary functions they provide a substrate upon which the cells can grow and they are semi permeable thereby permitting the exchange of nutrients and wastes The pores have a nominal molecular weight cut off MWCO of 10 000 Daltons 10kDa These pores allow small molecular weight molecules such as glucose and ammonia to move freely across the fiber while the cells added growth factors and large molecular weight products produced by the cells remain in the ECS Figure 1 1 Bioreactor ECS amp ICS As the circulation pump moves growth media into and through the hollow fiber a high relative pressure is created within the ICS This high relative pressure forces media through the pores of the hollow fibers and into the ECS near the proximal end entrance of the bioreactor As media flows into the ECS the pressure in the ICS drops and continues to do so along the length of the fiber Near the distal end exit of the bioreactor high relative pressure in the ECS forces media to flow back into the ICS carrying with it the low molecular weight metabolic waste from the cell culture see Figure 1 1 Due to the ECS flow along the length of the bioreactor the cells at the entrance of the bioreactor are exposed to fresh media with greater levels of oxygen and nutrients than the cells at the outlet end which are exposed to conditioned media This process is
240. re preparation e Cultureware installation e Run settings e Instrument Cultureware pressure test 4 Fill Flush 5 Pre Inoculation e pH and Lactate calibrations e Supplement pre treatment e Cycling initiation 6 Inoculation AutovaxID Operations Manual 7 1 Biovest International Inc Operating Procedures 7 Growth e IC Sampling e EC Sampling pre filter BRX e Harvest Sampling post filter e pH and Lactate calibrations e Replacing media feed and waste bags e Pump rate change e Pump capacity change e Pump setting change e Cell removal for EOP cell bank 8 End of Run Cultureware Removal and Disposal 7 2 Biological Preparation 7 2 1 Cell Line Characterization Successful mammalian cell culture production requires precise process control The AutovaxID is equipped with a process control system that provides the user with complete control of the cell culture environment A good process control strategy quickly stimulates high density culture growth then shifts control parameters to keep the culture in a stable environment while maintaining production at its peak It also allows the operator to vary conditions if required to maximize production yield A successful process control strategy depends on a thorough knowledge of the characteristics of the cell line Regardless of the method used for cell characterization specific values for the following should be established e optimum pH e optimum temperature e
241. re shaded to be distinct from the surrounding area and typically contain a text or graphic label that indicates what type of screen or information will be displayed when the respective button is touched There are three buttons on the Monitor screen in Figure 4 1 Hold Menu and Help AutovaxID Operations Manual 4 3 Biovest International Inc Introduction to the User Interface 4 2 4 Fields Fields are boxes that display process control parameters Fields are contextual in terms of data meaning accessibility and validation A field s value may apply to the particular run or to the instrument Additionally the color used to display the text or number may indicate the state of the process control parameter Values displayed in black indicate normal operation for the process parameter Values displayed in red indicate the given process parameter is operating out of the allowed user defined range is in the alarm state Values prefixed with an asterisk indicate that the process parameter is under manual override A field s shading may be either white or light gray Except for check boxes gray shading typically indicates that the field s value is strictly for monitoring and may not be modified by the operator Conversely white shading typically indicates that the field s value may be changed When an operator touches a field either its value toggles between ON and OFF or a data entry screen will appear The data en
242. ree bioreactor cultureware version Prepare the items as necessary Remove the cultureware from its shipping box inspect the bag for damage disinfect the outside of the bag and place the cultureware inside a sterile laminar flow hood Inspect the cultureware as is normally done when using standard cultureware all clamps are open the Luer fittings are snug and no tubing is kinked When using a pH probe aseptically connect the sterile probe to the cultureware Connect the Factor and Harvest containers Luer fittings to their respective cultureware tubing lines Remove the cultureware from the laminar flow hood Connect the cultureware to the instrument using the same procedures as for standard cultureware When using a pH probe connect the pH cable to the probe relevant when using a reusable pH probe and connect the opposite end to the BNC jack mounted to the top of the cultureware relevant for both pH probe options When relevant load the three Factor and three Harvest pump segments the light yellow PharMed tubing into the pump heads Near each pump segment is a label with an arrow that shows the required direction of liquid flow Ensure Factor flows from the container toward the cultureware and Harvest flows from the cultureware toward the container Consult the user manuals provided with the pumping equipment for instructions on how to load the tubing and set the rotational direction of the pump motor Access the Run
243. reen keyboard without saving an entry It is not necessary to clear a field prior to entering new information AAAG Figure 4 3 On Screen Numeric Keypad To accept the typed value press Enter After Enter is pressed the AutovaxID software performs an input validation to determine if the input value is within an acceptable range for that parameter If the desired value is within the allowed range the AutovaxID software accepts the value and puts it into operation If the desired value is not within the allowed range the AutovaxID software rejects the value and notifies the operator Sometimes if the AutovaxID software interprets the value as out of range it may automatically enter the lowest or highest possible value as appropriate AutovaxID Operations Manual 4 6 Biovest International Inc Introduction to the User Interface 4 3 1 4 3 2 4 3 3 4 3 4 Time Field A time filed is used for entering clock time Use the format HH MM SS where HH Hours range 0 99 MM Minutes range 0 99 SS Seconds range 0 59 Leading zeroes are not required Values outside the allowable ranges are not permitted Time is displayed on a 24 hour clock where midnight 0 00 and noon 12 00 Date Field A date field is used for entering the calendar date Use the format MM DD YY where MM Month range 1 12 DD Day range 1 31 depending upon the month YY Year range 2000 2099 Leading zeroes are not re
244. rements As noted in Step 4 it is important that the Next button is pressed IMMEDIATELY after the sample is taken so that the system can capture the current cell culture conditions for use during the calibration process Erroneous pH and lactate calibration results may occur if Step 4 is performed after the off line pH measurement is complete 6 Aseptically replace the 5cc sample syringe with the other clean 10cc AutovaxID Operations Manual 7 30 Biovest International Inc Operating Procedures syringe Leave the clamp on the IC Sample line closed 7 Measure the pH of the sample in an off line analyzer and then enter the off line measurement on the pH Calibration Entry screen see Figure 7 22 Figure 7 20 Pre Inoculation Step 2b AutovaxID Operations Manual 7 31 Biovest International Inc Operating Procedures Jan 12 2007 Calibration Sample 16 37 5 EM Make sure Off line neter is ready W Ka Aseptically connect a flush syringe open IC Sample clanp Take flush and close IC Sample clamp Gd Aseptically connect sample syringe open IC Sample da Take sample and close IC Sample clamp _ EN Press Next immediately after the sample is taken Hold Exit Next Help Figure 7 21 pH Calibration Sample Jan 12 2007 pH Calibration Entry 16 38 11 Sample Jan 12 2007 16 38 08 pH 7 10 AD 1920 Enter offline neasurenent pH Hold Exit Help Figure
245. remote access and enable or disable remote write access to the instrument CAUTION It is the responsibility of the end user to ensure the security of the LAN to which the AutovaxID is connected Access to the LAN must be controlled 6 2 1 IP Address An authorized user may assign a static IP address to the AutovaxID or a dynamically assigned address can be accepted from a DHCP server To set a static IP address enter a valid number in each of the four IP Address fields then press the Apply button The Apply button must be pressed to save the network configuration changes The IP address must be unique to this device and cannot match the IP address of any other device on the LAN Consult your local IT support group for assistance with determining the correct IP address configuration e Default IP Address 192 168 9 233 NOTE The default IP address may not be valid for your facility and may not provide network connectivity even if an Ethernet cable is connected to the AutovaxID AutovaxID Operations Manual 6 5 Biovest International Inc Software Setup amp Configuration Jan 26 2007 Network Conf ig 9 14 41 z futonatic Address Configuration IP Address 192 168 9 124 State negotiated 100baseTx FD flow control 1ink ok Info Fal Browser Renote Control E Network Write Control Hold Exit Back Next Help Figure 6 4 Network Configuration Screen 6 2 2 Automatic Address Configuration To
246. ringe containing the inoculum to the cultureware and inject the cells into inoculate the bioreactor see Figure 7 32 CAUTION Do not force the inoculum into the system too rapidly to prevent damage to the bioreactor fibers To assist the operator each step may be checked on the Inoculation Step 2 screen as it is completed by touching the appropriate line see Figure 7 33 When the bioreactor has been inoculated press the Next button AutovaxID Operations Manual 7 40 Biovest International Inc Operating Procedures Jan 12 2007 Inoculation Step 2 16 45 54 Aseptically connect the inoculation syringe to the Inoculate EC Sample port Open the Inoculate EC Sample am Inject the cells Close the Inoculate EC Sample clanp Press Next to continue Hold Back Next Figure 7 32 Inoculation Step 2 Jan 16 2007 Inoculation Step 2 16 30 04 Aseptically connect the inoculation syringe to the InoculateyEC Sample port Open the InoculateyEC Sample clamp Inject the cells Close the Inoculate EC Sample da Press Next to continue Hold Back Next Figure 7 33 Inoculation Step 2 Checklist AutovaxID Operations Manual 7 41 Biovest International Inc Operating Procedures The Start Growth screen will be presented to the operator As indicated the instrument will enter GROWTH mode after the operator presses the Next button see Figure
247. rvice provided by Biovest A trained Biovest International Accounts Services Representative can perform operator training and provide information necessary for the safe operation of the equipment Under certain circumstances e g when the instrument is physically moved reinstallation and equipment re qualification may be necessary WARNING WARNING WARNING CAUTION NOTE NOTE TO REDUCE THE RISK OF ELECTRIC SHOCK DO NOT REMOVE ANY PANELS FROM THE INSTRUMENT THERE ARE NO USER SERVICEABLE PARTS INSIDE REFER SERVICING TO QUALIFIED SERVICE PERSONNEL If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired A properly grounded main supply must be used for safe operation of the equipment Connect to an appropriately grounded outlet Consult local electrical codes Wherever a caution statement is used the documentation needs to be consulted in order to find out the nature of the potential hazard and any action which may need to be taken Failure to do so may result in injury or damage to the system The power cord serves as the disconnecting device for the instrument The instrument must be positioned such that access to the power plugs is never blocked All equipment is for indoor use only AutovaxID Operations Manual 2 1 Biovest International Inc Instrument Setup Exhaust Fan Inlet Fan and Air Filter Assembly _ H
248. ry User Interface Screens 5 1 8 Lactate Screen On the Monitor screen press the Lactate field or the surrounding area to bring up the Lactate screen see Figure 5 7 This screen allows an operator to view the calculated lactate concentration in mg mL the percent CO in the offgas and the bicarbonate constant of the supply media The bicarbonate constant is calculated during the lactate calibration step of the PRE INOCULATION procedure though an authorized user may manually change it Biovest does not recommend modifying the bicarbonate constant if the instrument is functioning normally changing this value may negatively affect the condition of the cell culture environment dan 24 2007 Lactate 11 04 23 Lactatelmg n1 0 000 UK Sensor 120 00 y Bicarb Constant nM 40 934 Hold Exit Hel p Figure 5 7 Lactate Screen 5 1 9 Cycle Detail Screen On the Monitor screen press any cycling field Cycle State EC Pressure or IC Pressure or the surrounding area to bring up the Cycle Detail screen see Figure 5 8 This screen allows an operator to view the current cycling state the user defined cycling parameters Cycle Delay Time EC Rise Time and EC Fall Time and the cycling pressures that will be utilized by the cycling algorithm during the next cycle An authorized user may change the Cycle Delay Time EC Rise Time and EC Fall Time by pressing their respective fields These parameters are normally set on the R
249. s are refilled disconnect and discard the 600101 072 assembly disinfect the LAV and cover it with the LAV cap If the factor bags need to be refilled again purchase another 600101 072 tubing assembly and repeat the previous steps For the 600290 XXX Legacy Cultureware The Factor Bag Set is packaged separately from legacy cultureware upon arrival The Factor Bag Set consists of e Two 1 5L bags connected together to create a single 3L volume e Fill Tubing Line Masterflex size 24 silicone tubing e One 0 2u media filter in the Factor Fill tubing line e Pre assembled Luer fittings tubing and clamps e Two individually packaged sterile male Luer plugs The Factor Bag Set is aseptically filled via their single use Factor Fill tubing line which requires using a Masterflex compatible peristaltic pump A 0 2u inline filter is provided as a precaution to ensure the sterile factor is not contaminated during AutovaxID Operations Manual 7 7 Biovest International Inc Operating Procedures the process of filling the Factor Bag Set Once the factor bags are full aseptically disconnect and discard the Factor Fill tubing line from the two female Luer ports and connect the two male Luer plugs to seal these ports During a cell culture run the factor bags hang from a rack inside the AutovaxID refrigerator Depending upon the length of the cell culture run and the rate of the Factor pump the Factor Bag Set may not need to be replaced
250. s factor is primarily added to the EC Thoroughly inspect the factor bag set before use to confirm that it is sterile and undamaged Contact Biovest Account Services if any damage is discovered For the 600320 XXX Standard Cultureware The factor bags consist of e Two 1 5L bags connected together to create a single 3L volume e Fill Tubing tubing line Masterflex size 16 silicone tubing e Pre assembled Luer fittings tubing and clamps e Luer Activated Valve LAV one e Luer Activated Valve Cap one The factor bags are aseptically filled via their single use Factor Fill tubing Line which requires using a Masterflex compatible peristaltic pump Once the factor bags are full aseptically disconnect and discard the Fill Tubing Line from the female Luer port connect the LAV to the female luer port to keep it closed and sterile and connect the luer activated valve cap on the LAV to keep it clean During a cell culture run the factor bags hang from a rack inside the AutovaxID refrigerator Depending upon the length of the cell culture run and the rate of the Factor pump the factor bags may need to be refilled during the run If additional factor is needed to complete the run they may be refilled To refill the factor bags purchase the optional 600101 072 tubing assembly remove and save the LAV cap disinfect the LAV and connect the male luer of the new 600101 072 to the LAV Fill the factor bags as described below Once the factor bag
251. s grown in serum free media may not remain in the bioreactor if the protein content of the serum free media is too low In this case cycling and harvesting conditions should be modified to avoid excessive loss of cells AutovaxID Operations Manual 7 64 Biovest International Inc Operating Procedures 7 27 Maintenance 7 27 1 7 27 2 7 27 3 Cultureware Storage The cultureware should be stored in a location maintained at 15 38 C 10 95 relative humidity non condensing see Section 11 Exterior Cleaning The cabinet is painted molded plastic that may be cleaned with a mild detergent or alcohol cleaner 70 isopropyl alcohol Major cleaning of the instrument should only be performed when the instrument is turned off in order to minimize the amount of cleaning solution that may get inside the instrument With the power off use a non abrasive lint free dry wipe to clean the LCD touchscreen Do not spray cleaning solution directly on the touchscreen Dampen a cloth with water and gently wipe the display clean Minimize spraying cleaning solution directly into the peristaltic pump drive and lock mechanism located on the lower left side of the instrument as this can strip the lubricant from these components Also do not spray cleaning solution on the electrical connections or into the fans on the rear panel of the instrument Inlet Fan Filter Cleaning Every six months or more often in environments with high particula
252. s in a clean laminar flow hood or biological safety cabinet and do not open the packaging yet 4 Remove the Factor Bag Set from the sterility bag in the hood or BSC The green tape on the Factor Bag Set should have visible brown red stripes to indicate proper ETO sterilization Remove all foam tape and other packing materials 5 Thoroughly inspect the Factor Bag Set to confirm it is undamaged Contact Biovest Account Services if any damage is found AutovaxID Operations Manual 7 11 Biovest International Inc Operating Procedures 6 Confirm all tubing clamps are open and no tubing is obstructed Contact Biovest Account Services if any clamps are found closed or if any tubing is obstructed 7 Identify the Factor Bag Set s 0 2u inline filter s inlet tubing This tubing is opposite from the two factor bags and is Masterflex size 24 Loada portion of this inlet tubing into a compatible size 24 Masterflex peristaltic pump Ensure the tubing is loaded in the correct orientation so that media is pumped into the factor bags and not out of them 8 Aseptically connect the inlet end of the Factor Fill tubing to the container with the EC media factor using the Luer fittings on the end of the Factor Fill tubing line 9 Close the Factor 1 clamp 10 Confirm the bleed screw in the 0 2u inline filter housing is closed and all five tubing clamps on the Factor Fill tubing line are open 11 Turn on the pump in the correct direction to
253. s routine run management tasks and saves time when gowning is unnecessary Remote computer access is described in detail in Section 10 Biovest designed the AutovaxID system to support compliance with FDA guidelines for 21 CFR Part 11 for a closed system Remote access must be prohibited disabled when the AutovaxID is used in a compliant environment Biovest recommends that facilities operate the AutovaxID in accordance with these guidelines At a minimum Biovest recommends using an external computer to periodically download logs from the internal control computer Access Restrictions The AutovaxID software provides user security with the ability to assign different access rights to Administrators and Users operators These rights can restrict User access to critical functions The AutovaxID software allows up to twenty unique accounts to be created The log will record events and attribute them to the logged on user as they occur All security features presume that each operator is given a unique user account and that the user account and password information is kept confidential See Section 6 6 for more information on setting up and using user accounts If user accounts are in use the AutovaxID security features permit operators to view certain screens even if they are logged off Operators are prompted to logon when attempting to access restricted functions This is useful for monitoring the flowpath status especially when the instr
254. s too long the applied pressure will be increased during the next cycle Likewise if the cycle is completed too quickly the applied pressure will be decreased during the next cycle These incremental adjustments will continue with increasing precision until the elapsed cycle time matches the user defined target When cycling is first initiated it may take several complete cycles before the cycling algorithm determines the proper cycling pressures and the cycle times stabilize The cycling control system automatically accounts for any increased transmembrane flow restriction that may be caused by the cell growth in the bioreactor As the transmembrane flow is reduced the pressure is automatically increased to maintain the target cycle times until the maximum system pressure is reached The control system can also account for any changes in the circulation pump rate that affect the net IC pressure Additional cycle states can be manually overridden by an authorized user for diagnostic or maintenance purposes see Table 5 1 Biovest does not recommend manually controlling the cycle state when the system is operating normally AutovaxID Operations Manual 3 5 Biovest International Inc Introduction to the Process Control Software 3 6 Alarms The AutovaxID can issue two types of alarms process control alarms and system alarms Process control alarms are based upon user defined alarm criteria system alarms are not user configurable The t
255. sary 4 Replace the empty factor container with a full factor container 5 Connect a 10 mL syringe to the Factor Prime port if one is not already attached that connects into the Factor tubing line from the bottle that was just replaced 6 Open the Factor 1 2 or 3 and respective Factor Prime clamps 7 Change the Factor pump speed to 200 mL hr 8 Use the syringe connected to the Factor Prime port to pull the incoming air from the Factor container away from entering the bioreactor ECS Only pull on this syringe at the rate the air is being pumped by the Factor pump Once all air has been diverted away from entering the bioreactor ECS reduce the Factor pump speed to the normally desired speed 9 Close the Factor Prime clamp 10 Aseptically replace the syringe on the Factor Prime port with a new 10 mL syringe to keep the port covered and sterile and ready for use the next time this Factor container is changed AutovaxID Operations Manual 8 25 Biovest International Inc Simplified Cultureware Usage 8 14 Removal and Storage of pH Probe Items When the run has ended follow the Cultureware Removal and Disposal Procedure in section 7 but include the following extra steps Reusable pH Probe 1 After the cultureware is in the biohazard bag but before it is sealed shut place clamps on the pH IN LO and pH OUT HI tubing lines near the pH probe near the quick coupler connectors Close the clamps 2 Rotate the locking rings
256. scard the packaging materials from the disposable pH probe assembly Press the disposable pH probe case into the reusable pH probe holder s small black clip Slide the disposable pH probe assembly upward until its upper end is firmly held within the reusable pH probe holder see figure 8 19 Press the combined reusable pH probe holder and disposable pH probe assembly into the black clip bel mounted on the front of the cultureware Slide it all downward until it rests on the screw post Rotate this Figure 8 18 assembly to the right so the pH IN LO port points Reusable pH Probe Holder toward the right see figure 8 19 Part Number 600306 000 AutovaxID Operations Manual 8 18 Biovest International Inc Simplified Cultureware Usage 8 Connect the pH cable to the round connector that is mounted in the top of the cultureware enclosure Press the cable s connector on to the mating connector and twist it clockwise to lock it in place 9 The pH probe assembly will replace the cultureware s tubing adapter between its quick coupler fittings labeled pH IN LO and pH OUT HI The adapter is colored dark green for clarity in figure 8 20 The pH IN LO tubing emerges from the lower right front edge of the cultureware enclosure The pH OUT HI tubing enters the lower left hand side of the cultureware enclosure Simplified Depiction Showing No Cultureware Tubing 10 Rotate the locking rings of the four quick coupler fittings to t
257. screen keyboard is displayed the container IDs can be manually typed in or they can be scanned from a barcode label using the barcode scanner The container IDs may be up to thirty characters long When in Strict Mode the Container IDs are required From the Setup Run 1 screen press Next to continue Figure 7 2 Setup Run 1 Screen AutovaxID Operations Manual 7 16 Biovest International Inc Operating Procedures 7 4 3 Setup Run 2 Following the sequence shown on the Setup Run 2 screen see Figure 7 3 close the indicated clamps In Strict Mode the box next to each instruction must be checked after each clamp is closed in order to proceed to the next screen From the Setup Run 2 screen press Next to continue Jan 12 2007 Setup Run Z 15 55 A The cultureware should be in place and ready to pressure test Close the 1C Sample clanp Close the Inoculate EC Sample dat Close the Supernatant Sample dag Close the Supernatant dat Close the Cell Renoval clanp Close the Cell Filter Bypass dag Check each box then select Next to continue Hold Exit Back Next Figure 7 3 Setup Run 2 Screen AutovaxID Operations Manual 7 17 Biovest International Inc Operating Procedures 7 4 4 Setup Run 3 Following the sequence shown on the Setup Run 3 screen see Figure 7 4 open the indicated clamps In Strict Mode the box next to each instruction must be checked a
258. see Section 5 3 the AutovaxID positions the valves automatically AutovaxID Operations Manual 5 63 Biovest International Inc Primary User Interface Screens 5 12 2 Overrides Screen As the screen warns the Overrides Screen should be used with caution as it will override the normal functions of the instrument such as cycling gassing valve positions and pump flow rates An authorized user may access the Overrides Screen see Figure 5 49 by pressing the Overrides button on the Advanced Menu screen Overrides should only be used to temporarily correct an unexpected situation with the cell culture run and should remain in effect for a minimum amount of time When in override mode the overridden function can only be cleared with operator intervention In other words the process control software will not clear or change a value once it is overridden The exception is at the end of a run when any pump overrides will be cleared any other overrides will remain until they are cleared by an operator Jan 25 2007 Overrides 14 42 12 Dverride Controls Use with caution Cycle State Rise Gas 002 15 0 Air 120 ml min Values Factor 3 EC Haruest EC 2 Brx amp Cycle Pump Rate Overrides Qutflow Media Factor Harvest Circulation Hold Exit Help Figure 5 49 Overrides Screen If a parameter is overridden the data in the field will be preceded by an asterisk see the Factor and Harvest fields in Fig
259. ssword is blank so you need not type anything for this step 4 Ascreen will appear that warns the user that their password has been reset and that they need to create a new one see Figure 6 19 press the OK button 5 The next screen will prompt you to enter a new password see Figure 6 20 Type a valid password that is at least six characters long or the minimum password length as defined by the system administrator After typing the password press the Enter button If you attempt to enter a password that is too short the AutovaxID software will notify you and asks you to reenter a new longer password Your password may be any combination of the letters numbers and punctuation marks that are available on the on screen keyboard up to 34 characters As the password is typed it is kept secret by displaying only the asterisk character in the entry field 6 If the password is acceptable then you will be prompted to type it again as confirmation Press the Enter button when you are finished If this second entry of your password does not match the first the AutovaxID software will notify you and asks you to re enter the new password from the beginning 7 Your new password has now been created Please keep it confidential AutovaxID Operations Manual 6 22 Software Setup amp Configuration Biovest International Inc Figure 6 19 Password Warning Figure 6 20 Create Your New Password 6 23 AutovaxID
260. stem typically remains in this state until the end of the run During this step cells will grow and fill the ECS and cell secreted products will be harvested and collected The operator is responsible for ensuring that the process control setpoints that are in use during this phase of the cell culture run are sufficient to support cell growth and productivity 7 Stop This step is used to end the cell culture production run and remove the cultureware After the run is stopped the instrument is placed in the IDLE state pH and incubator temperature controls are stopped the pumps are stopped cycling is stopped and periodic status logging is stopped AutovaxID Operations Manual 5 62 Biovest International Inc Primary User Interface Screens Completed steps will be indicated with a checkmark the current step will be indicated in green and steps that are incomplete or have been skipped will be neither checked nor green NOTE When running an instrument in Strict Mode the user is not allowed to deviate from the normal sequence of events as described above WARNING When using the Advanced Run Menu the user should go through the stop sequence in order to place the rotary valves in the load position for proper removal of the cultureware Failure to do so may split or damage the tubing of the cultureware being removed or the next cultureware being loaded When a cell culture run is terminated normally using the Run Menu
261. t a recommended logoff method AutovaxID Operations Manual 10 4 Biovest International Inc Remote Access remote Windows Internet Explorer ID x Go KL le http 127 0 0 1 8080 remote html ES Sen we d r remote DE v e Page O Tools ei 3 Remote Control View Figure 10 3 Remote Control View 10 3 Log Web Pages From the Home Page or the Remote Control View see Figure 10 1 or Figure 10 3 click the Log button to access the Log Menu see Figure 10 4 An operator can use this page to view or save the instrument logs current or historical AutovaxID Operations Manual 10 5 Biovest International Inc Remote Access Menu Windows Internet Explorer oj sl Go yy le http 127 0 0 1 8080 l0gpg html y Zeta Google Pi aa Sven gt El deb Page AS Log Menu Events Alarms Y Status V Current y Lines 10 Tait iy View Download All Home S Done PITT TT e internet E 100 4 Figure 10 4 Log Menu Page 10 3 1 Saving a Log Any current or historical log can be downloaded saved in one of two formats Raw or Text To download a complete log first select the log from the pull down box on the Log Menu page Current or History 1 through 8 Then click the Download All button to download the entire contents of the selected log The next page the Download Log Menu will allow you to select the file format that you prefer see Figur
262. t orientation so that media will be pumped into the factor bags and not out of them 3 Aseptically connect the inlet end of the Factor Fill tubing line to the container with the cold 4 C EC media factor using the Luer fittings on the end of the Factor Fill tubing line Close the Factor 1 clamp Ensure the factor Fill clamp is open Turn on the pump in the correct direction to begin pumping the factor into the factor bags 7 Stop the pump when the bags contain desired volume determined above Do not exceed 3L Over filling the factor bags may damage them and cause factor to leak Close the factor Fill clamp Aseptically disconnect the Factor Fill tubing line from the single female luer port near the factor Fill clamp 10 Connect the supplied sterile Luer Activated Valve LAV to close the open Factor Fill port Do not over tighten the LAV 11 Use 70 Isopropanol to disinfect the supplied non sterile LAV Cap and connect it to the LAV 12 Unload the tubing from the pump and discard it per your facility s procedures Fill Factor Bags from 600290 XXX Legacy Cultureware 1 Carefully open the Factor Bag Set shipping box and remove the bagged Factor Bag Set 2 Inspect the sterility bag for any punctures or damage Contact Biovest Account Services if any damage is found 3 Clean the exterior of the sterility bag and the exterior of the male Luer plug packages with 70 Isopropanol Place the pre packaged male Luer plug
263. t periodically to prime itself when detecting a low flow condition For a stall the desired rate must be greater than or equal to 10mL min for the detection to be active Circulation pump is running too fast The circulation pump running too fast alarm race will be logged when the pump is on and the instrument is at or past FILL FLUSH The race condition is asserted if the rate falls outside of the circulation setpoint by 20 for longer than 30 seconds For a race condition the desired rate must be greater than or equal to 20mL min for the detection to be active IOBOARD board offline Logged if one of the I O boards has a communication failure The board is placed in offline status IOBOARD will be Gas Pump Valve or Chassis board AutovaxID Operations Manual Biovest International Inc Introduction to the Process Control Software Alarm Description IOBOARD board has been reset Logged if an unexpected I O board reset has been detected on one of the boards IOBOARD will be Gas Pump Valve or Chassis board IOBOARD board has corrupted ROM Logged if an I O board fails its power on Read Only Memory checksum test IOBOARD will be Gas Pump Valve or Chassis board Pump head not latched Logged if the instrument is at or past FILL FLUSH and the pump cassette is not detected Cultureware not latched Logged if the instrument is at or past FILL FLUSH and the culturewar
264. t to grow and fill the EC space of the bioreactor user defined process parameters will be automatically controlled by the system and product will be harvested and collected It is the responsibility of the operator to ensure that the process parameters support both cell growth and productivity AutovaxID Operations Manual 5 3 Biovest International Inc Primary User Interface Screens 5 1 3 Hold Button The Hold button allows an authorized operator to quickly halt the pumps in case of an emergency The Hold button is accessible on nearly all screens When the instrument is placed on hold all pumps will be stopped an event will be logged and the operator will be warned that the instrument is on hold When the Hold button is pressed a pop up screen will request confirmation that the operator wants to put the instrument into hold see Figure 5 2 The only two options available once the system is in the Hold state is either to Resume or End Cell Culture see Figure 5 3 The elapsed time the system is on hold will be displayed for the operator in HH MM SS The instrument will log the elapsed time that the system was on hold if the run is resumed An AAA Jan STOP Are you sure you want to hold instrument e Figure 5 2 Hold Confirmation AutovaxID Operations Manual 5 4 Biovest International Inc Primary User Interface Screens Jan 23 200 Emergency Hold 16 53 11 This screen stops all
265. taining the current Harvest pump rate This change should be performed gradually to avoid sudden changes in environmental conditions Metabolism and productivity should be monitored to determine the effect of lowering the factor content A small amount of factor one to three percent may be added to the IC media if a culture proves difficult to start A typical symptom is slow initial growth Factor in the IC balances the concentration of low molecular weight components between the EC and the IC and prevents a severe deprivation of these components from the EC After suitable metabolism is attained factor can be eliminated from the IC typically after one to seven days In most cases the one time IC factor addition prior to inoculation is sufficient The factor bag set has a total capacity of three liters and is typically kept at 4 C throughout the run The factor should be between 2 and 10 C when placed into the refrigerator The initial factor addition volume can range from 100 to 250 mL for the IC and EC volumes Serum Free Media Experience with serum free media varies widely A cell culture can be started with serum supplemented media and then weaned onto serum free media or it can be started with serum free media Regardless of the technique the important point to consider is that the large molecular weight components of serum free media must be added directly to the EC for the cells to benefit from these supplements Hybridoma
266. tch between cycle phases Depending upon the elapsed time for the cycle phase the system adaptively regulates the pressures applied to the IC and EC reservoirs to meet the desired cycle times AutovaxID Operations Manual 1 5 Biovest International Inc Introduction 1 3 System Description In addition to Biovest s proprietary EC Cycling technology the AutovaxID provides sophisticated process control capabilities Process control is managed by a proprietary user interface application developed by Biovest International The application operates on a single board computer that is integrated into the AutovaxID all user interface functions can be performed on the integrated LCD touchscreen The AutovaxID software can monitor and or control a variety of process parameters including but not limited to pH Lactate Concentration Incubator Temperature EC Cycling Rates EC Rise amp EC Fall Circulation Pump Rate Peristaltic Pump Rates Outflow Media Factor Harvest Refrigerator Temperature The AutovaxID software can display the current culture status and the collected run data as well as monitor and analyze ongoing changes to the cell culture environment The software can use that information to control the cell culture environment for the duration of the culture run AutovaxID Operations Manual 1 6 Biovest International Inc Introduction Circulation Flow Sensor Gassing Connections Heater Duct Cold Air Re
267. tch of the cultureware s pH IN LO quick coupler fitting Disconnect this junction and connect the cultureware s pH IN LO quick coupler fitting to the pH probe s exposed port from the previous step Ensure the latch faces upward 9 Release the latch and rotate the locking ring to the locked position 10 Repeat the previous three steps for the pH OUT HI connection 11 Save the removed cultureware tubing adapter for use during the Cultureware Removal and Disposal Procedure 12 Grasp the top of the pH probe to prevent it from rotating in the probe case Use your other hand to remove the electrical cap Do not discard this cap It will be used during the life of the pH probe 13 Screw the threaded end of the pH cable to the pH probe s electrical connector Press the pH cable s other connector on to the round connector that is mounted in the top of the cultureware enclosure Twist the connector clockwise to lock it in place Note the BNC to two pin adapter is not used during the run 14 Proceed to the section below Factor and Harvest Container Options to continue with run startup Simplified Depiction Showing Only Some Cultureware Tubing Figure 8 16 Cultureware s External Tubing pH IN LO and pH OUT HI AutovaxID Operations Manual 8 17 Biovest International Inc Simplified Cultureware Usage 8 10 Using a Disposable pH Probe CAUTION The disposable pH probe is ready for use and connecti
268. te levels clean the filter on the upper cooling fan on the AutovaxID rear cover To clean the filter 1 Snap the outer bracket and filter off of the fan housing 2 Rinse the filter in tap water and gently dry 3 Replace the filter if is torn or stretched see Section 13 4 Once dry snap the filter and bracket back into place over the fan Verify that the filter covers all exposed areas of the fan inlet AutovaxID Operations Manual 7 65 Biovest International Inc Operating Procedures 7 27 4 Refrigerator Defrosting After a run any frost accumulation in the refrigerator can be removed by 1 Turn off the AutovaxID 2 Remove the harvest factor bag rack from the refrigerator 3 Leave the refrigerator door open and allow time for the frost to melt The amount of time required will depend upon how much frost has accumulated Use clean cloths or towels per your facility s requirements to absorb any liquid from the melting frost 4 After the frost has melted clean and dry the fins at top of refrigerator with a dry soft cloth Put the harvest factor bag rack back in the refrigerator The AutovaxID may be turned back on or left off until the next cell culture run per your facility s requirements 7 27 5 Calibration Biovest International recommends annual calibration of all AutovaxID instruments Adherence to the schedule not only facilitates c GMP manufacturing but it helps prevent various problems from occurring i
269. that is often successful with murine hybridomas Due to the varying characteristics of cell lines these guidelines may or may not result in a successful cell culture run in all applications Biovest claims no responsibility for the use of these settings It is the user s own responsibility to develop a successful process control strategy and production protocols Contact Biovest Account Services to discuss your application Start the Media pump at 20mL hr on day zero It is critical to keep a minimum amount of media flowing through the instrument to ensure that the water vapor evaporated through the GEX does not cause metabolites to become concentrated Start the Factor pump at a rate of 1 to 3 mL hr two to five days after inoculation depending upon the cell line requirements The Factor pump rate may need to be increased over the course of the production run to best maintain a healthy cell population An increasing Factor pump rate often results in greater overall productivity Start the Harvest pump at 1 to 3 mL hr two to ten days after inoculation The Harvest pump should often be run at the same rate as the Factor pump Similar to the Factor pump depending upon the specific cell line requirements the Harvest pump rate may need to be increased over the course of the cell culture run for optimal productivity NOTE if the Harvest pump rate is slower than the Factor pump rate the concentration of factor in the EC will be higher than expected
270. the EC ports are connected to the ECS where cells are grown The pink block on the upper right hand side of the screen with only a single fluid tubing port represents the EC reservoir Touching it will bring the user to the Cycle Detail screen see Section 5 1 9 AutovaxID Operations Manual 5 56 Biovest International Inc Primary User Interface Screens The pink block in the top left hand corner of the screen with four tubing ports represents the IC reservoir Touching it will bring the user to the Media Temp screen see Section 5 1 10 The yellow block at the middle of the screen represents the pH probe The current pH is noted above the probe Touching the pH probe allows access to the pH Control screen see Section 5 1 7 The green block to the left of the pH probe represents the GEX noted above it the current gassing number rounded to the nearest integer The five dotted circles represent the five fluid pumps Outflow Media Factor Harvest and Circulation The Outflow Media Factor and Harvest pumps are shown in the same order left to right as the pumps are mounted in the instrument as viewed from the front The speed of each pump if it is running is noted next to its graphic Touching any of the five pump graphics will bring the user to that pump s detail screen see Sections 5 1 12 5 1 13 5 1 14 5 1 15 and 5 1 16 Valves appear on the Flowpath screen as grey blocks The orientation of the blocks changes dep
271. the Supernatant Sample clamp closed This 3cc syringe will protect the sterility of the Supernatant Sample port and can be used to flush the Supernatant Sample line prior to taking the next Supernatant Sample Cell Filter Bypass Clamp The harvest bag assembly contains a 0 2um cellulose acetate filter This filter can be used to filter the harvested product Supernatant as it is pumped into the harvest bags Alternatively if filtering is not required preferred or if the filter becomes clogged a Cell Filter Bypass line is also available This line is normally clamped during run setup see Figure 7 3 but it may be opened at any time during a cell culture run at the operator s or the facility s discretion No process control changes are required if the filter is bypassed but the calculated harvest volume will not compensate for the change in harvest routing Replacing Pump Containers The Outflow Media and Factor containers may all be changed during a cell culture run The Harvest bags are pre assembled to the cultureware and do not have any disconnect provisions However the Supernatant in the Harvest bags may be moved into another container during a cell culture run using the container changing procedure The Supernatant may be transferred using a pump or siphon as preferred by the operator or facility The Supernatant port has a male Luer and may be used to aseptically connect the harvest bags to the preferred Supernatant storage container
272. the automated FILL FLUSH procedure for successful operation of the AutovaxID the steps of the procedure are listed below for informational purposes These steps coincide with those shown on the following FILL FLUSH screens e Position the valves for EC reservoir evacuation see Figure 7 7 e Evacuate all of the air from the EC reservoir see Figure 7 8 e Position the valves for IC reservoir fill see Figure 7 9 e Pre fill the IC reservoir for twenty five minutes see Figure 7 10 and Figure 7 11 e Fill the GEX and the IC side of the bioreactor ICS see Figure 7 12 e Refill the IC reservoir approximately twenty minutes see Figure 7 13 e Enable temperature control and begin IC flush continues for 71 2 hours see Figure 7 14 e Position the valves to allow EC flush using cycling The system is held in the LOW state for twenty minutes while media moves into the EC and the IC refills see Figure 7 15 e Cycling begins the IC and EC are flushed for the remainder of the twenty four hours see Figure 7 16 and Figure 7 17 Note The values shown in the fields of the following FILL FLUSH screens may not match the values observed on your AutovaxID instrument and should not be cause for concern The following screens are intended to show the FILL FLUSH sequence not typical FILL FLUSH sensor data or process control parameters At the end of FILL FLUSH the system will be ready for PRE INOCULATION see Figure 7 18
273. the flowpath for off line analysis Sample Port A controlled opening in the flowpath which permits removal of fluid for analysis SBC Single Board Computer Schedule Controller User defined pump flow rate changes that occur according to a timed schedule defined by the operator Setpoint The desired value of a given entry which is maintained by the software AutovaxID Operations Manual 12 2 Biovest International Inc Glossary Supernatant Harvested cell products Time Cycle Controller User defined pump flow rate changes that occur in a repeating time based pattern according to a cycle defined by the operator Tubing Slots Grooves in the instrument panel to route tubing into and out of the refrigerator compartment Transmembrane Across the fibers of the bioreactor Vent Filters The in line filters in the flowpath in the IC reservoir and at the top of the GEX that allow gases but not media to enter and exit the flowpath and prevent pressure buildup in the flowpath Waste See Outflow AutovaxID Operations Manual 12 3 Biovest International Inc Parts List 13 Parts List The following list provides ordering information for disposable cultureware sterile disposables and miscellaneous parts for running the instrument If you need an item that is not on the list Biovest may be able to supply it please contact your sales representative Part Number Instruments 600285 100
274. the instrument until the override is cleared While the override is active the operator should carefully observe the cultureware The operator should never walk away from the instrument while using pump overrides To perform a pump override an authorized operator must simply touch the pump rate field for the pump to be overridden and enter the desired pump override rate using the on screen keypad Similarly to clear a pump override touch the pump rate field for the overridden pump and press the Clear button on the on screen keypad AutovaxID Operations Manual 5 67 Biovest International Inc Software Setup amp Configuration 6 Software Setup amp Configuration An authorized user can access the AutovaxID software setup and configuration functions by pressing the Setup button on the Main Menu screen see Figure 5 16 The Setup screen see Figure 6 1 contains the following five buttons Configure Used to configure the instrument settings such as System Features Network Settings and Date amp Time Diagnostics Used by Biovest Technical Service personnel to diagnose or test instrument hardware Passwords Used to create and modify Administrator and Operator accounts and passwords Touch Calibration Used to calibrate the touch screen Network Print Log Used to print the log to a network connected TCP IP enabled printer on the same logical network as the AutovaxID Jan 25 200 setup 12 49 24 Conf igure
275. the pH cable into an appropriate autoclavable bag or packaging 7 Place the pH probe assembly in an autoclavable container to prevent it from lying on its side while inside the autoclave Ensure the probe case tubing is not kinked Figure 8 14 Connect the Autoclave Tubing Lines AutovaxID Operations Manual 8 15 Biovest International Inc Simplified Cultureware Usage 8 Autoclave the pH probe assembly at 121 C 15 psi maximum for 55 minutes followed by a slow exhaust phase lasting at least 15 minutes 9 Remove the pH probe assembly from the autoclave and bring it to a sterile laminar flow hood Disinfect the outside of the autoclave packaging before placing it in the laminar flow hood to continue cooling 8 9 4 Reusable pH Probe Installation into the Cultureware CAUTION Do NOT lift the cultureware from the shipping box by its sterile bag Lift it out by supporting the cultureware from its bottom surface 1 Prepare the simplified cultureware using the instructions in the section Cultureware Preparation under Operating Procedures Steps related to the Factor and Harvest Bag Sets do not apply and the refrigerator rack isn t needed because simplified cultureware do not have integral Factor and Harvest bags The technician must provide and prepare sterile Factor and Harvest containers separately described below 2 Once the pH probe assembly is cool enough to safely handle remove and discard the autoclaved packag
276. their corresponding air and CO flow rates A gas number of zero 0 will drive the pH up no CO maximum air a gas number of 100 will drive the pH down maximum CO minimum air approximately 20 CO The air flow is always between 120 mL min gas numbers 51 to 100 and 500 mL min gas number 0 The CO is always between zero any number 50 or less and twenty percent gas number 100 of the combined flow or zero to thirty mL min Table 3 1 Air Flow Rate CO Flow Rate Approximate sas Number mL min mL min Percent CO 0 500 0 0 25 285 0 0 50 120 0 0 75 120 13 3 10 100 120 30 20 The user defines the pH setpoint during run setup An authorized user may also modify the pH setpoint during an active cell culture run The AutovaxlD process control software will continuously monitor the pH and adjust the gas mixture to maintain pH at the setpoint Alternatively an authorized user may disable the automatic pH control system completely and manually override the gassing system by specifying a gas blend COz Biovest strongly recommends using automatic pH control until the user fully understands this critical control function There are times when manual pH control may be necessary or desirable though Biovest does not recommend attempting to manually control pH when the system is operating normally However in some instances pH may fall below the user defined setpoint when the gas number reaches 0 and the
277. tion of unclamped tubing can allow a leak 3 Ensure that all fittings are reasonably tight Use caution to not over tighten fittings damage may occur 4 Ensure that the pump cassette is properly installed to the pump module and that all pump tubing is properly secured 5 Monitor the IC and EC pressure during the pressure test sequence to help determine the source of the failure AutovaxID Operations Manual 7 20 Biovest International Inc Operating Procedures If the pressure test continues to fail after a couple of attempts and the source of the leak failure cannot be located or corrected please contact Biovest Account Services Do NOT proceed to FILL FLUSH WARNING If Strict Mode is NOT in use the operator can continue to FILL FLUSH even if the pressure test has failed Biovest does NOT recommend or support advancing to FILL FLUSH after a failed pressure test Proceeding with FILL FLUSH after a failed pressure test may damage the cultureware or result in an unsuccessful cell culture run Note Pressing the Exit button before the pressure test is complete or pressing the Back button before the start of FILL FLUSH will return the operator the Setup Run screens The Run Name Cultureware Lot Number Run Information Run Settings and Container IDs will be retained the clamp states are not retained and will need to be reconfirmed before starting a new pressure test if in Strict Mode 7 5 Fill Flush Procedure Th
278. tions It is easy to forget to restore valves to their starting positions remove overrides or reconnect tubing To avoid most problems with your system be sure to perform periodic recommended maintenance and to examine the cultureware carefully AutovaxID Operations Manual 9 1 Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION No display on touch screen Loss of power Confirm that the power cord is securely plugged into the back of the instrument and to an outlet that is proven to supply the necessary voltage Ensure that the power switch on the back of the instrument is turned on Unplug the instrument and check the fuse s in the power entry module see Section 2 6 Unplug the instrument and check that the voltage selection card in the power entry module is set to the correct supply voltage see Section 2 5 The instrument is experiencing a hardware failure Note if the red amber amp green LEDs are illuminated through the clear window on the front of the instrument contact Biovest Account Services Touch screen not responding or irrationally responding Touch screen is out of calibration If possible press Menu gt Setup gt Calibration If not possible turn the instrument off allow 30 seconds to elapse and then turn the instrument back on Upon power up the Touch Calibration button will appear in the center of the screen
279. tovaxID s CO or Air flow rate described below Wait approximately 20 minutes for pH to change and stabilize Repeat the process for sampling IC medium measuring pH off line and adjusting CO or Air flow rate until the desired pH is achieved During the growth phase of the culture pH tends to change more frequently than later in the run when the Medium pump rate often remains consistent over time The technician must determine the frequency of sampling IC medium to monitor and control the pH of the cell culture medium Refer to pH Control section 3 and Gas Overrides section 5 for a description of how the AutovaxID adjusts CO or Air flow rate to control medium pH In summary When medium pH is too high decreasing Air flow rate and increasing CO labeled on screen as Gas CO2 each result in decreasing medium pH If CO is 0 0 and the Air flow rate is higher than 120 mL min reduce the Air flow rate wait 20 minutes re check pH and repeat until the desired pH is achieved Once Air flow rate is 120 mL min the minimum rate and pH is still too high increase the CO value from 0 0 wait 20 minutes re check pH and repeat until the desired medium pH is achieved The allowed range for CO is 0 0 to 20 0 AutovaxID Operations Manual 8 10 Biovest International Inc Simplified Cultureware Usage When medium pH is too low decreasing CO labeled on screen as Gas CO2 and increasing Air flow rate each result in
280. trument see Figure 2 1 The waste gasses offgas will typically contain water condensate from the Gas Exchange Cartridge GEX approximately 15 to 20 mL per day The Off Gas port is provided so that the waste gasses may be captured filtered or incinerated as necessary per your facility requirements A quick disconnect is provided 1 8 ID tubing is recommended If an open container is used for offgas condensate collection submerge the end of the offgas line in bleach or other sanitary solution per your facility s requirements Be certain that the offgas container has sufficient volume to collect the condensate and that your maintenance schedule includes checking this container WARNING Do NOT use a glass vessel with an exhaust filter to collect the offgas condensate If the filter becomes wetted the vessel will become pressurized and could pose an explosion hazard to the operator 2 8 Ethernet Connection The AutovaxID has an Ethernet port on its rear panel see Figure 2 1 This port is used to connect the AutovaxID to a LAN to allow access by remote computers Use a standard Ethernet cable not supplied to connect the AutovaxID to a LAN To comply with EN 61326 1 one clamp on ferrite provided with the instrument must be added to the Ethernet cable as shown in figure 2 8 Failure to add the ferrite may make the system non compliant Place the ferrite approximately 2 inches 5 cm from the end of the connector as shown in f
281. trument exceeds the desired Media Temperature setpoint The ambient air temperature is outside of the parameters specified in Section 11 Adjust the room temperature to be within the specified range The basal media supply is too warm The basal supply media should be 18 C to allow adequate media temperature control The Media Temperature setpoint is set too low for the ambient room temperature Either increase the Media Temperature setpoint or decrease the room temperature The instrument is experiencing a hardware failure Contact Biovest Account Services The IC pressure is too low There is an IC leak Visually inspect the cultureware for an IC media leak Review the instrument run log to verify that the cultureware Pressure Test was performed and that the cultureware passed The instrument is experiencing Contact Biovest Account a hardware failure Services AutovaxID Operations Manual Biovest International Inc Troubleshooting Clogged tubing or wetted IC vent filter SYMPTOM POSSIBLE CAUSE ACTION Loose Luer connection or fitting Inspect all Luer and barbed connections Ensure that the appropriate A clamp has come loose clamps are closed Check the Outflow pump for Outflow pump tubing leak worn or damaged tubing IC fluid leak Visually inspect the IC vent filter on top of the IC reservoir for signs of wetting contact Biovest
282. try screen may give several input options or allow for text and or numeric input Field input commonly involves changing a control setpoint or applying a manual override to an automated process control Some field s shading changes between gray and white depending upon the context of the screen or the state of process control AutovaxID Operations Manual 4 4 Biovest International Inc Introduction to the User Interface 4 3 Data Entry Configuring and maintaining a production run requires selecting process control options and entering text and or numeric information into the AutovaxID software The data entry options include an on screen QWERTY keyboard see Figure 4 2 an on screen numeric keypad see Figure 4 3 and an integral handheld barcode scanner Figure 4 2 On Screen QWERTY Keyboard AutovaxID Operations Manual 4 5 Biovest International Inc Introduction to the User Interface AAA E af Aan ada 6 Logging Interval minutes Jan HRS SS Cancel Del Enter E If a field requires alphanumeric data entry the appropriate on screen entry pad keyboard or keypad will automatically appear To enter values using either simply touch the appropriate key to type the desired text or numeric value s The typed values will appear on the screen as you touch their corresponding key s If you make a mistake simply press Delete Del or Backspace lt Pressing Cancel will exit the on sc
283. turn Rotary Valves Cycling Sensors pH Probe Connector Circulation Pump Pump Loading Knob Peristaltic Pumps Figure 1 3 AutovaxID Instrument Front View Heater Duct Warm Air LCD Touchscreen Media Temperature Probe Refrigerator Door AutovaxID Operations Manual Biovest International Inc Introduction 1 3 1 1 3 2 LCD Touchscreen The Autovax D utilizes an LCD touchscreen as its primary method of user interface It is located on the upper right of the instrument above the refrigerator see Figure 1 3 The touchscreen is used to both monitor and operate the system it is the primary method for operators to interact with the AutovaxID software The display is sensitive to touch and can be used with either bare or gloved hands Rather than using a traditional keyboard or mouse operators merely touch on screen keys or buttons to navigate the software and provide input A light touch is all that is needed for the system to sense excessive force or the use of hard or sharp objects can damage the touchscreen CAUTION Do not spray the touchscreen to clean it Dampen a cloth with water and gently wipe the display clean Do not use alcohol other cleaners solvents as they may damage the screen Refrigerator The refrigerator provides a cool temperature controlled environment for storing growth factors and harvested product The refrigerator does not use a traditional compressor or refrig
284. twenty liters may be impractical The small media containers are often stored at room temperature while they are connected to the cultureware Conversely media containers larger than fifty liters may not be consumed quickly enough before media components begin to degrade and should be stored cold Depending upon the media formulation and container size it may be necessary to store the media in the dark during the run The cultureware is equipped with male and female Luer connectors on the Media port The media container s that will be used during the cell culture run must be equipped with Luer connectors or the facility must acquire the proper Luer adapters Similarly depending upon the cell line approximately two liters of ECS media is needed per month during a cell culture run The ECS media will be stored in the Autovax D refrigerator during the run If the lactate controller will be used to automatically adjust the Media pump rate then glucose should be the primary carbohydrate in the media Cultured mammalian cells typically consume glucose anaerobically despite the presence of ample oxygen Serum and media may contribute other sources of carbohydrate but in vitro cells primarily utilize glucose which results in the formation of lactic acid Genetically engineered or transformed cell lines that do not metabolize glucose and generate lactate as the principle metabolic waste product should not be used with lactate control The AutovaxI
285. ument is running normally and no maintenance or troubleshooting tasks are required AutovaxID Operations Manual 4 9 Biovest International Inc Primary User Interface Screens 5 Primary User Interface Screens There are eight primary screens from which all user interface functions required for operating the AutovaxID are accessible They are 5 1 Monitor Screen The Monitor screen is the starting point to view or control all instrument functions Many of the fields on this screen can be selected to bring up an associated detail screen where process control changes can be made by authorized operators The Monitor screen comprises three main areas the title bar along the top screen an information panel to display the current instrument status and a row of buttons along the bottom of the screen see Figure 5 1 mee ne SR D 10 0 0 00 MI D E 0 EX Figure 5 1 Monitor Screen In Figure 5 1 the title bar displays the screen name Monitor Alternatively the title bar on the Monitor screen may show the instrument name if it has been defined on the System Features screen see Section 6 The calendar date and time of day are also displayed and are setup on the Date and Time screen see Section 6 The title bar will flash red if there is an unacknowledged alarm If the alarm has been acknowledged but remains in the queue the title bar will stop flashing but will remain red A blue title bar indicat
286. un Settings 1 screen see Figure 5 19 AutovaxID Operations Manual 5 9 Biovest International Inc Primary User Interface Screens The cycle state can be overridden only by an authorized user using an advanced override see Section 5 12 2 Biovest does not recommend manually overriding the cycle state if the system is operating normally Improperly setting the cycling state can cause irreparable damage to the cultureware and may cause a cell culture run to fail Similarly the cycling pressures can be overridden only by an authorized user as a diagnostic function Biovest does not recommend manually overriding the cycle pressures if the system is operating normally The cycling algorithm will automatically adjust these pressures to meet the desired cycle time The algorithm will normally meet the cycle time setpoint within three full cycles though it may occasionally take longer Overriding these values may prevent cycling from functioning properly and subsequently cause the cell culture environment to degrade The cycle Type may be indicated as either Time or Pressure Press though pressure based cycling is only used during FILL FLUSH Time based cycling is used during all other run states The cycle Type may not be modified and is shown for diagnostic purposes only Jan 23 2007 Cycle Detail 13 37 39 Cycle Run Settings Mi Current State Off Cycle Delay Tine 7 00 00 day hr min of Growth EC Rise Tine 15 minutes
287. until time total cycle time and on time See Section 5 5 11 for details All pump rate changes are automatically logged An authorized user may change the harvest controller type by touching the Controller Type field and then selecting the desired harvest controller type from the on screen list Some controllers have additional setup parameters that can be accessed by pressing the Controller Setup button Authorized users may change the controller type during a cell culture run Harvest Pump Schedule Setup If the harvest controller type is set to Schedule then pressing the Controller Setup button on the Harvest Controller screen will bring up the Harvest Pump Schedule screen see Figure 5 32 unless the Tie to Harvest box is checked on the Factor Controller screen see Figure 5 27 The Harvest Controller screen allows an authorized user to schedule up to ten elapsed times with variable pump rate values In strict mode and when user names and passwords are in use only administrators may create schedules Normal users may view the schedule but they may not change the values in the Elapsed Time or the Factor fields nor can they add or delete events Normal users may advance or delay the schedule using the Advance and Back buttons or by entering a manual time shift To add an event press the Add Event button at the bottom of the screen enter the desired Elapsed Time for the start of the event using the on screen keypad and then press Enter
288. ure 5 49 To clear an override and revert to automatic process control the operator should touch the overridden field and press the Clear button on the on screen keypad CAUTION If an operator needs to make an override during a cell culture run the operator should remain at the instrument until the override is cleared AutovaxID Operations Manual 5 64 Biovest International Inc Primary User Interface Screens 5 12 3 Cycle Overrides On the Overrides screen the cycle state can be overridden to seven possible states Table 5 1 Cycle States Cycle SE State Description Typical Use Off Cycling is turned off The IC reservoir and the EC air Used during the first 3 10 bag are vented to atmosphere days of GROWTH Cycle Cycling is on IC and EC pressure are controlled by the Used during GROWTH after cycling algorithm to meet the user defined cycle times any user defined delay Low The fluid level in the EC reservoir is held near EC low Used during FILL FLUSH Applies maximum system pressure to the EC while the E prain IC is vented thereby draining the EC reservoir esedouing reo P Applies maximum system pressure to both the IC and Used during automated ress SEN EC pressurization testing Stable Turns off the valves to both the IC and EC and holds Used during automated pressure at a constant value pressurization testing IC Pr Applies maximum system pressure to the IC while
289. urrently selected log see the Log Select field See Section 5 9 3 for details on the verification report AutovaxID Operations Manual 5 54 Biovest International Inc Primary User Interface Screens The log may be filtered for status lines event lines and or alarm lines by checking the appropriate box es on the Log View Options screen A checked box will display those types of entries on the Log screen for the currently selected log see the Log Select field For example if an operator only wishes to view the alarm lines for the current log then the Status lines on and Events line on checkboxes should be cleared unchecked and the current log should be selected on the Log Select field To view the filtered log the operator must press the Exit button to return to the Log screen To change or reset the filtering press the Options button on the Log screen to again return to the Log View Options screen The log may also be searched for a specific term by using the search tools on the Log View Options screen To perform a search first select the appropriate log then enter the desired search term into the Search field using the on screen keyboard Finally press the up or down arrow next to the Search field to choose the search direction The first entry that is found to contain the search term will be displayed at the top of the Log screen To search for the next instance press the Options button on the Log screen to r
290. us Settings column indicate the settings that are associated with the current profile The Default Settings are also presented for comparison Saved pump controller settings can be loaded using the Load Settings button on the Run Settings 1 screen Jan 24 200 Run Settings 2 11 26 33 Pump Settings Previous Settings Default Settings Outflow Outflow 100 E 100 Outflow Melia Manual HE 20 E Manual bm Manual E 0 Manual bet Manual z 0 Manual Circulation Manal CP 200 Es Manual Hold Exit Back Next Help Figure 5 22 Run Settings 2 Screen AutovaxID Operations Manual 5 25 Biovest International Inc Primary User Interface Screens 5 5 1 Outflow Controller The purpose of the outflow pump is to maintain an appropriate volume of media in the cultureware The AutovaxID software automates the speed of the outflow pump and generally maintains the outflow pump at a fixed rate that is faster than the combined delivery of the other pumps The outflow pump operates constantly to remove media from the IC circuit Users need not determine the rate for this pump nor enter a speed in process control The outflow rate is set to m edi Rump Rat e Fact o1Pump Rat d Harves Pump Rat g 1 2 or 100mL hr whichever is greater Pressing the Outflow button on the Run Settings 2 screen brings up the Outflow Controller screen see Figure 5 23 This screen shows the outflow pump controller type the outflow container
291. ush the IC Sample line prior to taking the next IC Sample AutovaxID Operations Manual 7 48 Biovest International Inc Operating Procedures 7 14 EC Sampling Procedure Pre Filter The cell culture and its secreted protein product are retained in the EC Knowing the product concentration can aid the operator in establishing and adjusting the Harvest pump rate Comparison of the product from the harvest bag versus directly from the bioreactor may be useful to determine cell culture activity When comparing the product from these two locations keep in mind that the harvest bag contains a composite of all product that has been collected whenever the Harvest pump has been run throughout the cell culture run Three 3cc syringes with male Luers are required to take an EC sample To take an EC sample from the bioreactor 1 2 3 4 Aseptically disconnect the 3cc flush syringe and replace it with a new 3cc Aseptically connect a 3cc syringe to the Inoculate EC Sample port Unclamp the Inoculate EC Sample line and fill the syringe to flush the line Close the Inoculate EC Sample clamp syringe The 3cc flush syringe may be discarded per your facility s requirements Unclamp the Inoculate EC Sample line again and remove the volume required for the off line assays Close the Inoculate EC Sample clamp Aseptically disconnect the 3cc sample syringe and replace it with a new 3cc syringe Leave the Inoculate EC Sample clamp closed Th
292. utton If the printer is compatible and the IP address specified properly on the Log Printer screen the first four pages of the current log will be printed on one sheet 6 8 5 Print Log Once the log formatting and printer IP address are defined press the Print Log button to start printing AutovaxID Operations Manual 6 26 Biovest International Inc Operating Procedures 7 Operating Procedures This section of the manual describes the basic procedures for operating the AutovaxID instrument It is important that the user completes these procedures accurately for every production run in order to ensure success WARNING Proper training in the handling of biohazardous materials and the use of protective devices is the responsibility of your facility It is important to follow your facility s procedures for safety and gowning and the preparation of handling of treatment of and proper disposal of biohazardous materials to minimize the risk of injury 7 1 Outline of a Cell Culture Run The following is a chronological list of the major events of a typical cell culture run Many of these activities are performed while interacting with various AutovaxID software screens especially selections made in Run Control 1 Biological Preparation e Cell line characterization e Media preparation e Scale up of cells for inoculation see Section 7 24 2 Instrument Setup see previous sections 3 Cultureware Setup e Culturewa
293. ves when the cultureware is loaded to the instrument by an operator AutovaxID Operations Manual 1 9 Biovest International Inc Introduction e Gassing Connections these four ports interact with two different systems in the cultureware Two of the ports provide pressure to the IC and EC reservoirs for cycling one port provides a mixture of air and CO to the gas exchange cartridge the remaining port is the exhaust port for the gasses expelled from the gas exchange cartridge All ports interact with the AutovaxID gassing system and are mated to the cultureware automatically when it is loaded to the instrument by an operator e Heater Ducts these circulate air from the incubator through a thermoelectric heater that is housed in the instrument The incubator is the cultureware enclosure The incubator media temperature is defined by the operator and is controlled by the AutovaxID software e Media Temperature Probe this sensor couples with a disposable temperature well in the sterile cultureware This probe is not wetted The probe temperature provides feedback to the heater controller and is reported in the AutovaxID software for operator reference e pH Connector this connector mates with a two pin connector on the disposable pH probe that is included in the sterile cultureware The pH probe is not re usable and requires no operator intervention prior to use The pH is used to measure the condition of the cell culture and is re
294. vices EC reservoir is overfilled The EC is not pressurized Check that IC vent filter has not become wetted with media Check for a Cycling has not occurred Alarm see Section 3 6 Contact Biovest Account Services Cycling has not occurred The cycle state is overridden Verify the Cycle Controller settings and review the log for cycling overrides Clear any overrides to the Cycle State The level sensors have not detected the cycling flag Verify that there are no objects obstructing the cycling sensors Verify that the cycling sensors are on by looking thru the window on the front of the instrument to inspect the red amber amp green LED lights The green LEDs should be illuminated The IC pressure is low Check that the IC vent filter has not become wetted with media The EC pressure is low Check that the IC vent filter has not become wetted with media The Harvest EC Valve is overridden with the EC return valve closed Clear any overrides to the Harvest EC Valve Contact Biovest Account Services AutovaxID Operations Manual Biovest International Inc Troubleshooting SYMPTOM POSSIBLE CAUSE ACTION Cycling is taking too long The cultureware has become plugged with cells Perform a Cell Removal see Section 7 20 The Harvest EC Valve is overridden Clear any overrides to the Harvest EC Valve
295. wer pH Lactate control does not function 8 7 2 Automatic pH Control with Reusable pH Probe Advantages Disadvantages automatic pH control means infrequent sampling burden reusable pH probe reduces operating costs compared to using the pre sterilized pH probe technicians must spend time and correctly perform testing assembly autoclave sterilization of the pH probe and aseptic insertion into the cultureware requires aseptic insertion into the cultureware 8 7 3 Automatic pH Control with Pre Sterilized pH Probe Advantages Disadvantages fast run startup this option is ready for immediate use eliminates complicated and time consuming testing assembly and sterilization of reusable pH probe option automatic pH control means infrequent sampling burden most expensive of the three pH control options requires a 24 hour period for Pre Inoculation phase requires aseptic insertion into the cultureware AutovaxID Operations Manual 8 9 Biovest International Inc Simplified Cultureware Usage 8 8 Using Manual pH Control no pH Probe When no pH probe is being used in the Overrides screen set the Gas CO2 value to 15 0 This suggested value will not necessarily result in the desired pH but it should prevent the AutovaxID from adjusting pH too drastically from reasonable values Further adjustments to the Gas CO2 value likely will be necessary before inoculation should occur to ensure the desired pH
296. with the application AutovaxID Operations Manual 8 2 Biovest International Inc Simplified Cultureware Usage AutovaxID Flowpath Schematic For Simplified Cultureware with One 1 1 m2 Bioreactor GEX In Air or Air CO2 Flow Temperature ca Sensor Probe Permeable hollow fibers permit exchange between IC and EC of small molecules lt nominally 10 KDa Gas Exchanger Bioreactor GEX Out Always Air CO2 AutovaxID Incubator pH Sensor optional Factor lt gt Factor Prime gt Quick Coupler Connection Media lt gt Outflow lt gt IC Sample gt gt Harvest lt gt amp Circulation Pump E Check Valve gt Female Luer x D Inoculate EC Sam ple Male Luer lt Flow Direction Ratchet Clamp SYMBOL CODE Intracapillary Circuit EE 0 24 Sterilizing Filter Slide Clamp o Tubing Connection Extracapillary Circuit Figure 8 1 Simplified Cultureware Flowpath Schematic Single 1 1 m Bioreactor Version AutovaxID Operations Manual 8 3 Biovest International Inc Simplified Cultureware Usage AutovaxID Flowpath Schematic For Simplified Cultureware with Three 0 5 m2 Bioreactors Permeable hollow fibers permit exchange between IC and EC of small molecules lt nominally 10 KDa ATT pH Sensor optional Air C02 Flow Temperature CE Sensor Probe Ic Reservoir
297. with the remote computer for instructions on its setup and requirements AutovaxID Operations Manual 2 10 Biovest International Inc Introduction to the Process Control Software 3 Introduction to the Process Control Software The ability to monitor and control environmental conditions based on a predetermined protocol is referred to as process control Process control in the AutovaxID system is monitored and directed by the internal control computer proprietary software and instrument firmware The AutovaxID process control software has been designed by Biovest International to provide complete process control capabilities Specifically the AutovaxID software enables an operator to e Customize the process control strategies according to each facility s unique needs e Monitor the conditions of the cell culture e Control the operation of the AutovaxID with automatic or override capabilities The AutovaxID process control system is a set of coordinated software programs which together operate the entire AutovaxID instrument Some of the programs run on the internal single board computer while others are algorithms that are built into firmware that in turn run on microprocessor based printed circuit boards This separation of software increases the overall robustness of the system and leads to the AutovaxID s two distinct functions instrument control and user interface Instrument control functions derive from programs runni
298. wo types of alarms are NOT distinguishable in the log Process control alarms include Table 3 2 Process Control Alarms Alarm Description f Logged if the instrument is at or past PRE ae E INOCULATION after calibration and the ES alarm condition exists High Media Temp Logged if the instrument is at or past Low Media Temp FILL FLUSH and the alarm condition exists High Ref Temp Logged if the instrument is at or past Low Ref Temp FILL FLUSH and the alarm condition exists High IC Pressure Logged if the instrument is at or past Low IC Pressure FILL FLUSH and the alarm condition exists High EC Pressure Logged if the instrument is at or past Low EC Pressure FILL FLUSH and the alarm condition exists Logged if the instrument is at or past PRE High Lactate 4 INOCULATION and the alarm condition Low Lactate exists Logged if the container volume alarms are PUMPNAME pump bag is full used and the volume exceeds the alarm Current Volume ml setting PUMPNAME will be Outflow or Harvest Logged if the container volume alarms are used and the volume is below the alarm setting PUMPNAME will be Media or Factor Logged if cycling is active the cycle too long alarm is enabled and the cycle is taking longer than the configured cycle time plus the cycle too long alarm time out value PUMPNAME pump bag is empty Current Volume ml Cycle taking too lo
299. y Circuit EC The portion of the flowpath that includes the outside of the fibers in the bioreactor the EC reservoir the factor bags the harvest bags and their interconnecting tubing The EC is separated from the IC by the walls of the hollow fibers External Computer See Remote Computer Factor A supplement or additive to a growth media This may be serum insulin transferrin or other components typically added to the ECS media Flowpath All components of the EC and IC see Cultureware Gas Exchange Cartridge GEX The cartridge in the cultureware which is responsible for the introduction of Oz via air flow and the introduction or removal of CO to maintain pH AutovaxID Operations Manual 12 1 Biovest International Inc Glossary Growth Phase The early stage of cell growth when growth is exponential Harvest The process of collecting media containing cell secreted product Or the media which contains the cell product or the protein of interest Hollow Fibers The thousands of semi permeable tubes within the bioreactor They function as a membrane with a nominal molecular weight cutoff of 10kDa They permit the exchange of nutrients and wastes and can serve as the growth substrate for cultured cells Intracapillary Circuit IC The portion of the flowpath which includes the inside of the fibers the IC reservoir and their interconnecting tubing The IC is separated from the EC
300. y configure all of the system settings A regular user a user without administrator rights has restricted access to the AutovaxID Regular users may not modify the AutovaxID system settings and they may only manage their own account When strict mode is enabled regular users cannot change run settings pH and temperature setpoints pump controller types cycle times or alarm settings but they can perform regular run maintenance tasks such as pH calibrations pump rate changes depending upon the controller type and container changes Regular users are also allowed to step through the normal run sequence though they may not deviate from it With usernames and passwords in use access is limited to the following screens until an operator logs on e Monitor e Main Menu e Run Settings view only e Alarm Log view only e Flowpath If any operator attempts an activity that requires a logged on user a Logon button will be presented on the screen 6 6 1 Administrator Account Setup An administrator will establish user accounts on the AutovaxID and configure the AutovaxID The first time that an operator presses the Passwords button on the Setup screen the setup of usernames and passwords will be initiated The first three screens will explain that an administrator account must be setup first see Figure 6 8 Figure 6 9 and Figure 6 10 This first account will be the primary administrator AutovaxID Operations Manual 6 12
301. ypropylene pouring ring GL45 thread and volume markings Caps are polypropylene with two liquid ports accept male luer slip connection one vent port 1825 000 filter purchased separately and they include internal silicone tubing to deliver or collect fluid Each part number is a package of four items 600013 000 600014 000 600015 000 600016 000 600017 000 600018 000 600019 000 600020 000 600021 000 600022 000 600023 000 600061 000 Media Bottle 0 5 liter Media Bottle 1 0 liter Media Bottle 2 0 liters Media Bottle 5 0 liters Media Bottle 10 0 liters Delivery Cap Assembly 0 5 liter Delivery Cap Assembly 1 0 liter Delivery Cap Assembly 2 0 liters Delivery Cap Assembly 5 0 liters Delivery Cap Assembly 10 0 liters Receiving Cap Assembly use on any volume bottle Bottle amp Cap Startup Package includes e Six 1 liter bottles e Four 10 liter bottles e Two 10 liter delivery cap assemblies e Four 1 liter delivery cap assemblies AutovaxID Operations Manual 13 2 Biovest International Inc Parts List Four receiving cap assemblies fit any size bottle e One box of vent filters Plastic Bottles and Ported Cap Assemblies 400836 000 400840 000 400842 000 Miscellaneous 1825 000 2380 600 102778 001 102787 000 Services CALIB SERV SERV CON 401199 000 Ported Cap Assembly for 2L PETG Bottle Sterile one each 2L PETG Bottle with Ported Cap Assembly Non Sterile one each 250 mL Polycar
Download Pdf Manuals
Related Search
Related Contents
Bedienungsanleitung de Operating instructions en Mode d Bedienungsanleitung 0 89 – 29 16 00 80 Anwenderhandbuch UM EN FL SWITCH 7000 BC 410 II BC 4125 II - AL-KO El presente manual debe considerarse parte permanente TK-7302/ TK-8302 TK-7302H/ TK-8302H TK-7302HV/ TK User`s Manual Mavis Beacon Teaches Typing Teacher Guide Copyright © All rights reserved.