1. 62 Max 62 Max Infrared Thermometer Calibration Information Introduction The Fluke 62 MAX and 62 MAX Infrared Thermometers the Product can determine the surface temperature by measurement of the infrared energy radiated from the target surface Note Japanese models indicate Celsius only A Warning To prevent personal injury read all safety information before you use the Product Contact Fluke To contact Fluke call one of the following telephone numbers e Technical Support USA 1 800 44 FLUKE 1 800 443 5853 e Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 e Canada 1 800 36 FLUKE 1 800 363 5853 e Europe 31 402 675 200 e Japan 81 03 6714 3114 e Singapore 65 6799 5566 e Anywhere in the world 1 425 446 5500 Or visit Fluke s website at www fluke com To register your product visit http register fluke com To download manuals or to see print or download the latest manual supplement visit http us fluke com usen support manuals June 2013 2013 Fluke Corporation All rights reserved 1 62 Max 62 Max Calibration Information Safety Information A Warning identifies conditions and procedures that are dangerous to the user A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test The symbols used on the Product and in this manual are shown in Table 1 Top Symbol A A Warning revent eye damage and personal i
2. See Manual Instrumentation product Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Warning Laser Conforms to European Union directives Battery C Conforms to relevant Australian standards Conforms 19 enina oe Conforms to relevant South Korean EMC standards at Metrology Certification Infrared Thermometer Maintenance Maintenance A Caution To prevent damage to the Product do not leave the Product on or near objects of high temperature Clean the Product Use soap and water on a damp sponge or soft cloth to clean the Product case Carefully wipe the lens surface with a moist cotton swab The swab may be moistened with water See Figure 1 i 4 hiy002 eps Figure 1 Clean the Product 62 Max 62 Max Calibration Information Change the Battery To install or change the AA IEC LRO6 battery use a flat head screwdriver to remove the battery compartment door Replace the battery as shown in Figure 2 Be sure to observe the correct polarity hiy001 eps Figure 2 Change the Battery Specifications Temperature Range Infrared Thermometer Specifications 62 MAX 30 C to 500 C 22 F to 932 F 62 MAX 30 C to 650 C 22 F to 1202 F Accuracy Calibration geometry with ambient temperature 23 C 2 C 20 C 1 5 C or 1 5 of reading whichever is greater 232
3. section of this document User replaceable parts are listed in Table 4 Table 4 User Replaceable Parts Description Fluke Part Number Quantity Fluke 62 Max Rear Decal 4134155 1 Fluke 62 Max Rear Decal 4134162 1 Fluke 62 Max Battery Door Assembly 4103199 1 Battery AA Alkaline 1311933 1 Sere Carabineer Solid Aluminum 2 3 in with spring hinge 4149787 A 62 Max 62 Max Users Manual 4060712 1 LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase This warranty does not cover fuses disposable batteries or damage from accident neglect misuse alteration contamination or abnormal conditions of operation or handling Resellers are not authorized to extend any other warranty on Fluke s behalf To obtain service during the warranty period contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages this limitation
4. F 3 F or 1 5 of reading whichever is greater gt 10 C to lt 0 C 2 C 214 F to lt 32 F 4 F lt 10 C 43 C lt 14 F 6 F 20 C 1 C or 1 of reading whichever is greater 232 F 2 F or 1 of reading whichever is greater gt 10 C to lt 0 C 2 C 214 F to lt 32 F 4 F lt 10 C 3 C lt 14 F 6 F Response Time 95 lt 500 ms 95 of reading lt 300 ms 95 of reading Spectral Response 8 to 14 microns Emissivity 0 10 to 1 00 Temperature Coefficient 0 1 C C or 40 1 C of reading whichever is greater Optical Resolution 10 1 calculated at 90 energy 12 1 calculated at 90 energy Display Resolution Repeatability of reading 0 1 C 0 2 F 0 8 of reading or 1 0 C 2 F whichever is greater 0 5 of reading or 0 5 C 1 F whichever is greater Power 1 AA IEC LRO6 Battery Battery Life 10 hours with laser and backlight on 8 hours with laser and backlight on Weight 255 g 8 99 oz Size 175 x 85 x 75 mm 6 88 x 3 34 x 2 95 inches Operating Temperature 0 C to 50 C 32 F to 122 F Storage Temperature 20 C to 60 C 4 F to 140 F without battery Operating Humidity 10 to 90 RH non condensing 30 C 86 F Operating Altitude 2000 meters above mean sea level Storage Altitu
5. correct radiometric temperature will need to be determined before blackbodies are used to make measurements e Make sure that the required calibration equipment is allowed a warm up period as per manufacturer recommendations See the documentation for the equipment Infrared Thermometer Performance Verification Tests To prepare the Unit Under Test UUT for verification Visually inspect the UUT outer case for cracks or damage Open the battery compartment and inspect the battery and battery compartment for corrosion or damage Install anew AA battery prior to verification Close the UUT battery compartment door Set the UUT temperature units for degrees Celsius C if necessary Push SEL until C alone is shown Change by pushing SET Nw WN a Performance Verification Test Procedures Note The subsequent verification test requires the blackbody emissivity to be set to 0 95 at all temperature points Note For temperatures below ambient to reduce the chance of condensation or ice buildup on the emitter plate of the 4180 it is recommended to install a clean dry air or nitrogen gas purge box on the front of the 4180 Blackbody Although the purge box is recommended it is optional If temperature measurements are made without a purge box refer to the 4180 User Manual for how to make proper temperature measurements below ambient See Figure 3 hiy003 eps Figure 3 Aim the 62 Max for Verification 62 Max 62 Max Calibratio
6. de 12 000 meters above mean sea level Drop Test 3 meters IP Rating IEC 60529 IP 54 Implement Standard Q SXAV37 Laser Safety IEC 60825 1 Class 2 Electromagnetic Environment IEC 61326 1 Portable Electromagnetic Compatibility Applies to use in Korea only Class A Equipment Industrial Broadcasting amp Communication Equipment 1 This product meets requirements for industrial Class A electromagnetic wave equipment and the seller or user should take notice of it This equipment is intended for use in business environments and is not to be used in homes 1 62 Max 62 Max Calibration Information Performance Verification Tests The subsequent performance tests verify the complete operation of the Product and check the accuracy of each function against Product specifications If the Product fails any part of the test calibration adjustment and or repair is necessary Product calibration can only be adjusted at the Fluke factory To return your Product for calibration adjustment call one of the numbers in the Contact Fluke section of this document Read the entire Performance Tests section before you make any adjustment Required Equipment Table 2 shows the required equipment necessary for the performance tests Note Fluke recommends plate blackbodies as verification equipment because cavity blackbodies have small field of view which could increase uncertainty for a thermometer wit
7. h low D S If cavity blackbodies are used to verify the Product make sure that the distance between the cavity blackbody opening and the Product is equal to the diameter of the cavity blackbody opening x 2 Where the temperature of the cavity tube near the outside is lower than the temperature of the cavity readings will be lower Moving the 62Max closer to the cavity opening may be necessary for the 62 Max to only measure the cavity temperature If other than Fluke blackbodies are used the target should be checked for actual radiometric output with a more accurate device such as a Heitronic KT19 82 Table 2 Required Equipment Item 2 1 Blackbody 2 2 Blackbody 2 3 Blackbody Minimum Use Specifications Range 10 C to 100 C Range 100 C to 500 C Range 500 C to 650 C Calibration Equipment Fluke 4180 Fluke 4181 Plate Blackbody with 10 cm to 15 cm target 1 Calibrate with an IR thermometer that has low uncertainty such as KT19 and TRT Before the Performance Verification Tests Before you do the performance verification tests e Performance verification tests are to be done within the temperature of 23 3 C and between 25 and 60 relative humidity e Make sure that all required calibration equipment has a valid calibration certificate and or label e Make sure that any item s substituted for the equipment in Table 2 meets or exceeds the stated minimum use specifications e The
8. n Information 1 Set blackbody emissivity and temperature for the first temperature listed in Table 3 Allow adequate time for the blackbody to stabilize See the blackbody documentation 2 Position the UUT at the correct the distance from the blackbody listed in Table 3 Center the UUT laser s on the blackbody emitter plate For the 62 MAX make sure that the midpoint of the two laser spots is centered on the blackbody plate The 62 Max uses one laser Push and hold the UUT trigger for 5 to 6 seconds Release the trigger Record the UUT indication Move to a location that is not centered on the blackbody Make sure the UUT reading is within the limits listed in Table 3 ann n FY Repeat steps 1 through 7 for the remaining temperatures listed in the Temperature column of Table 3 The 62 Max is specified to 500 C 461 C is the last temperature point to verify 9 Set blackbodies to nominal values Table 3 Verification Tests Distance cm Temperature C Emissivity 62 Max Limits C 62 Max Limits C Fluke 4180 points 28 10 0 95 12 0 to 8 0 12 0 to 8 0 28 0 0 95 1 5 to 1 5 1 0 to 1 0 28 65 0 95 63 5 to 66 5 64 0 to 66 Fluke 4181 point 28 250 0 95 246 2 to 253 8 247 5 to 252 5 28 461 0 95 454 1 to 467 9 456 3 to 465 6 Fluke 62 Max only above 500 C 25 5 580 0 95 574 2 to 585 8 1 Distance is calculated based on the target size Typically a spot si
9. njury Read all safety Information before you use the Product Do not use the Product if it operates incorrectly Use the Product only as specified or the protection supplied by the Product can be compromised Before you use the Product inspect the case Do not use the Product if it appears damaged Look for cracks or missing plastic See emissivity information for actual temperatures Reflective objects result in lower than actual temperature measurements These objects pose a burn hazard Do not look directly into the laser with optical tools for example binoculars telescopes microscopes Optical tools can focus the laser and be dangerous to the eye Do not look into the laser Do not point laser directly at persons or animals or indirectly off reflective surfaces Replace the batteries when the low battery indicator shows to prevent incorrect measurements Do not use the Product around explosive gas vapor or in damp or wet environments Use the Product only as specified or hazardous laser radiation exposure can occur Table 1 Symbols This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE Directive Annex this product is classed as category 9 Monitoring and Control Risk of danger Important information
10. of liability may not apply to you Fluke Corporation Fluke Europe B V P O Box 9090 P O Box 1186 Everett WA 98206 9090 5602 BD Eindhoven U S A The Netherlands 11 99 62 Max 62 Max Calibration Information 10
11. ze that is about 1 2 the target size provides the best results Note Fluke recommends plate blackbodies as verification equipment because cavity blackbodies have small field of view which could increase uncertainty for a thermometer with low D S If cavity blackbodies are used to verify the Product make sure that the distance between the cavity blackbody opening and the Product is equal to the diameter of the cavity blackbody opening x 2 Where the temperature of the cavity tube near the outside is lower than the temperature of the cavity readings will be lower Moving the 62Max closer to the cavity opening may be necessary for the 62 Max to only measure the cavity temperature Temperature verification is complete If the Product does not pass the performance verification tests calibration adjustment or repair is necessary Product calibration can only be adjusted by Fluke To return your Product for calibration adjustment call one of the numbers in the Contact Fluke section of this document Blackbody Layout Make sure that the distance between two blackbodies is more than 1 meter Forced air or HVAC vents should not be near the surface of the radiation source Do not have two blackbodies work face to face For more detail about blackbody use see the Fluke 4180 Technical Guide 8 Infrared Thermometer User Replaceable Parts User Replaceable Parts To order replacement parts call one of the numbers in the Contact Fluke