1220 Infinity LC Mobile Upgrade Kit User Manual
Contents
1. Ete Agilent Technologies Notices Agilent Technologies Inc 2012 No part of this manual may be reproduced in any form or by any means including elec tronic storage and retrieval or translation into a foreign language without prior agree ment and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number G4292 90000 Edition 07 2012 Printed in Germany Agilent Technologies Hewlett Packard Strasse 8 76337 Waldbronn This product may be used as a com ponent of an in vitro diagnostic sys tem if the system is registered with the appropriate authorities and com plies with the relevant regulations Otherwise it is intended only for gen eral laboratory use Warranty The material contained in this docu ment is provided as is and is sub ject to being changed without notice in future editions Further to the max imum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a par ticular purpose Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing use or perfor mance of this document or of any information contained herein Should Agilent and the user have a separate writt
2. 28 Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Installing the Attenuation Unit to the Bench 1 Identify the final position of the instrument on the bench according to the specifications given below 2 Drill the 4 holes in the bench Either use the attenuation unit as a model or prepare the holes according to the figure below SAFETY AREA 4x 08 5 mm drilling holes Agilent 1220 Infinity Mobile Upgrade Kit 29 3 Installation 3 Position the attenuation unit on the bench The snap fasteners must be oriented to the front and the prepared holes in the attenuation unit must stand above the drill holes in the bench 4 Use the chosen screws or bolts to fix the attenuation unit to the bench 30 Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Securing the LC to the attenuation unit When Whenever the instrument is installed onto the attenuation unit in the mobile lab Install the handle plate to the 1220 Infinity LC Instrument see Installing the Handle Plate to the 1220 Infinity LC Instrument on page 20 Install the attenuation unit to the bench see Installing the Attenuation Unit to the Bench on page 28 WARNING Personal injury or damage to the system The system is bulky and heavy Preparations gt Lift the instrument with at least two persons 1 Open the fixing angles 1 and slide the instrument on the handling plate onto the attenuation unit 2 Verify that i
3. 1 This chapter provides information on unpacking checking on completeness and installation of the mobile kit ee Agilent Technologies 3 Installation Cautions and Warnings Personal injury or damage to the system The system is bulky and heavy gt Lift the instrument with at least two persons Sharp metal edges Sharp edged parts of the equipment may cause injuries gt To prevent personal injury be careful when getting in contact with sharp metal areas CAUTION Damage to the system If the handle plate is not fixed precisely the mobile system may tumble or slide off the bench gt Use the handle plate only if it is fixed correctly to the instrument 16 Please note that the design of the 7220 Infinity mobile solution is intended for protecting the instrument from influences during transit Analytical measurement however has to be conducted under stationary conditions Operating the 1220 Infinity LC in a mobile lab environment can require special equipment or facilities to comply with local safety regulations e g special solvent waste and storage containers Please verify that these regulations are met before you operate the 1220 Infinity LC in your mobile lab Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Unpacking the Module Damage to the module If the delivery packaging shows signs of external damage please call your Agilent Technologies sales and service office immediately I
4. 60001 Mobile solvent compartment including 4 special brackets WARNING Sharp metal edges Sharp edged parts of the equipment may cause injuries gt To prevent personal injury be careful when getting in contact with sharp metal areas 1 Take out the grey solvent bottle container of the instrument s top cover and attach the 4 leak tub brackets to it two on each side Look at the instrument s top and identify the referring spaces where the arrows in the picture of step 2 on page 25 point to This is the dedicated position for middle part of the clamps This information is necessary for a correct positioning of the clamps It is hard to correct their position once they are installed 24 Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Installing the Mobile Solvent Compartment 2 Now take the leak tub and reinstall it to the top cover Verify that the leak tub brackets properly hook in underneath the top cover on the instrument s side Agilent 1220 Infinity Mobile Upgrade Kit 25 3 Installation Installing the Mobile Solvent Compartment 3 Take the mobile solvent compartment and push it into the solvent container The two metal clips will grip into the solvent container s plastic On the rear of the mobile solvent compartment is the grounding clamp which needs to hook into the metal sheet between the board extension and the main assembly Grounding clamp 26 Agilent 1220 Infinity Mobile Upgrade Kit I
5. 8 This chapter gives an introduction to the use of the mobile kit ee Agilent Technologies 1 Introduction Introduction The Agilent 1220 Infinity Mobile LC Solution is an Agilent 1220 Infinity LC with integrated variable wavelength or diode array detector and a mobile upgrade kit that enables transportation The 1220 Infinity LC Mobile Upgrade Kit consists of functional parts that allows the Agilent 1220 Infinity LC to be mounted into a mobile lab so it can be moved to different locations to access remote measurement sites The main component is the attenuation unit which acts as a shock absorber to protect the instrument during transit or from influences of operators moving in the mobile lab A solvent bottle unit keeps the bottles fixed to the instrument A column wire mesh keeps the column safe in the column heater and the mobile solvent compartment secures the solvent bottles during operation of the instrument 6 Agilent 1220 Infinity Mobile Upgrade Kit Figure 1 Agilent 1220 Infinity Instrument with attached mobile kit Agilent 1220 Infinity Mobile Upgrade Kit 1 Introduction Application Examples When samples of interest are locally remote from the lab the lab has to come to the samples GC and GC MS based mobile labs have been around for quite some time but the need for mobile LCs has become a more burning issue recently Mobile LC applications range from analysis of non volatile pesticides in environmental and food a
6. e mobile upgrade kit and a proper solution must be chosen See Site Requirements on page 10 for more details Agilent 1220 Infinity Mobile Upgrade Kit Leak tub brackets Mobile solvent compartment SSV support Handling plate Attenuation unit Figure 4 Mobile Kit Overview Agilent 1220 Infinity Mobile Upgrade Kit Installation Unpacking the Module 19 3 Installation Installing the Handle Plate to the 1220 Infinity LC Instrument WARNING Personal injury or damage to the system The system is bulky and heavy gt Lift the instrument with at least two persons 1 Lay the 1220 Infinity LC Module carefully on its rear plate For the G4290B it is necessary either to position the bottom of the instrument slightly free over the bench s edge or to lay it onto the carton box of the solvent compartment 2 Push on the two lock pins 1 to release the front fixing angles 2 and separate the handling plate 3 from the rest of the attenuation unit 3 Use the four delivered screws to mount the handle plate to the bottom of the 1220 Infinity LC instrument The holes for the screws are already prepared in the bottom of the instrument Pay attention that the side with the concave cut out section is orientated to the instrument front 20 Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Installing the Handle Plate to the 1220 Infinity LC Instrument 4 Verify that the screws are fixed and car
7. e to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated condi tions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indi cated conditions are fully under stood and met Agilent 1220 Infinity Mobile Upgrade Kit Contents Contents 1 Introduction 5 Introduction 6 Application Examples 8 2 Site Requirements and Specifications 9 Site Requirements 10 Specifications 14 3 Installation 15 Cautions and Warnings 16 Unpacking the Module 17 Installing the Handle Plate to the 1220 Infinity LC Instrument 20 Installing the Solvent Selection Valve Support 22 Installing the Mobile Solvent Compartment 24 Installing the Column Wire Mesh 27 Installing the Attenuation Unit to the Bench 28 Securing the LC to the attenuation unit 31 4 General Operation Instructions 33 General Operation Instructions 34 5 Consumables 35 Consumables 36 6 Appendix 37 General Safety Information 38 Agilent Technologies on Internet 39 Agilent 1220 Infinity Mobile Upgrade Kit Contents 4 Agilent 1220 Infinity Mobile Upgrade Kit s Agilent 1220 Infinity Mobile Upgrade Kit 000 ee 1 Introduction e E e j Introduction 6 Application Examples
8. efully bring the instrument back into the upright position Agilent 1220 Infinity Mobile Upgrade Kit 21 3 Installation Installing the Solvent Selection Valve Support When The Solvent Selection Valve Support needs to be installed if a solvent selection valve SSV is used with the instrument Parts required p n Description G4292 00001 SSV support delivered as part of the G4292 60000 attenuation unit Preparations If SSV is already installed Unplug the SSV Clear the solvent compartment Remove the whole top cover with the SSV still attached to it WARNING Sharp metal edges Sharp edged parts of the equipment may cause injuries gt To prevent personal injury be careful when getting in contact with sharp metal areas When the SSV support is installed the SSV cannot get attached to the top cover without removing the top cover 22 Agilent 1220 Infinity Mobile Upgrade Kit Installation 3 Installing the Solvent Selection Valve Support 1 Take the SSV Support and slide it into the slits in the metal cover above the gradient valve 2 Ifthe SSV was already installed Reinstall the top cover Connect the SSV OR If the SSV was not already installed Remove the top cover Fix the SSV to the top cover Reinstall the top cover Connect the SSV Agilent 1220 Infinity Mobile Upgrade Kit 23 3 Installation Installing the Mobile Solvent Compartment Parts required p n Description G4292
9. en agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the sep arate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accor dance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcon tract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted computer soft ware as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or dis closure of Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Government users will receive no greater than Limited Rights as defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015 b 2 November 1995 as applicable in any technical data Safety Notices CAUTION A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly per formed or adhered to could result in damag
10. n of the holes that must be prepared to the bench 12 Agilent 1220 Infinity Mobile Upgrade Kit Site Requirements and Specifications 2 Site Requirements SAFETY AREA 120 4x 08 5 mm drilling holes Figure 3 Dimensions of the attenuation unit Agilent 1220 Infinity Mobile Upgrade Kit 13 2 Site Requirements and Specifications Specifications Physical Specifications Table 1 Physical Specifications Type Specification Comments Weight all parts 14 2 kg 31 3 lbs Dimensions 73 x 480 x 445 mm See Installing the Attenuation of the attenuation unit 2 9 x 18 9 x 17 5 in Unit to the Bench on page 28 height x width x depth for more information regarding safety space Performance Specifications The performance specifications show details about the performed test patterns Table 2 Performance Specifications Type Specification Comments Vibration Test Outdoor Agilent internal test guidelines 14 Agilent 1220 Infinity Mobile Upgrade Kit Agilent 1220 Infinity Mobile Upgrade Kit 3 Installation Cautions and Warnings 16 Unpacking the Module 17 Damage to the module 17 Delivery Checklist 18 Installing the Handle Plate to the 1220 Infinity LC Instrument 20 Installing the Solvent Selection Valve Support 22 Installing the Mobile Solvent Compartment 24 Installing the Column Wire Mesh 27 Installing the Attenuation Unit to the Bench 28 Securing the LC to the attenuation unit 3
11. nd Specifications 2 Site Requirements 970 SAFETY AREA aa o Figure 2 Required space for a 1220 Infinity LC Instrument installed to a mobile upgrade kit numbers given in mm Agilent 1220 Infinity Mobile Upgrade Kit 11 2 Site Requirements and Specifications Bench Material and Strength proper bench must be chosen under the aspect that it must bear the combined weight of the instrument with solvent bottles and the mobile kit The bench material must be suitable for a proper fastening solution Fastening Solution Because of the expected varieties in bench material and strength the fastening material is not included in the kit suitable fastening solution screws or bolts must be chosen to secure the attenuation unit to the bench of the mobile lab This solution must be strong enough to withstand the forces that will result from the possible accelerations that can occur in a moving vehicle app 2000 N per screw based on a calculation with 1 G acceleration and the weight of a G4294B The head of the chosen screws or bolts must not stand out more than 15 mm on top of the attenuation unit s bottom plate The attenuation unit comes with four 8 5mm 0 33 in diameter holes prepared in the lower plate and four 20 mm 0 79 in diameter holes located directly above them in the top plate This allows the use of a 14 mm 9 16 in nut for tightening the screws or bolts See Figure 3 on page 13 for the layout informatio
12. nform your service representative that the instrument may have been damaged during shipment CAUTION Defective on arrival problems If there are signs of damage please do not attempt to install the module Inspection by Agilent is required to evaluate if the instrument is in good condition or damaged gt Notify your Agilent sales and service office about the damage gt An Agilent service representative will inspect the instrument at your site and initiate appropriate actions Agilent 1220 Infinity Mobile Upgrade Kit 17 3 18 Installation Delivery Checklist Ensure all parts and materials have been delivered with the mobile upgrade kit The delivery checklist is shown below Please report missing or damaged parts to your local Agilent Technologies sales and service office p n G4292 60001 G4292 60000 G4292 60003 G4292 40000 G4292 00001 Description Mobile solvent compartment Includes 4 prefixed bottle holders and 4 Leak Tub brackets to fix it to the module s solvent compartment Attenuation unit Shock absorbing unit to mount on the Mobile Lab s bench Handling plate Pre connected to the Attenuation unit see Installing the Handle Plate to the 1220 Infinity LC Instrument on page 20 for further instructions Column wire mesh To protect the column SSV support Needs to be installed when a solvent selection valve is installed The required fastening bolts or screws are not delivered with th
13. nstallation Installing the Column Wire Mesh p n Description G4292 40000 Column wire mesh Parts required 1 Put the column wire mesh with its open side facing to the top into the column heater assembly 2 Position the column in the Column wire mesh so that it is fixed properly Agilent 1220 Infinity Mobile Upgrade Kit 27 3 Installation Installing the Attenuation Unit to the Bench The bench holding the instrument requires 4 screws or bolts to secure the attenuation unit of the 1220 Infinity Mobile LC Kit to the bench This fastening material is not provided from Agilent because of the unpredictable situation of bench material and dimension Tools required Description Wrench size depending on chosen fastening material Parts required p n Description G4292 60000 Attenuation unit Preparations Prepare the appropriate fastening material 4 proper screws or bolts and corresponding tools for attaching the attenuation unit onto the desired surface Four 8 5 mm 0 33 in holes are prepared on the lower plate of the attenuation unit and four equivalent 20 mm 0 79 in holes in the top plate above allowing the use of a 14mm nut The screws or bolts must withstand a maximum drag force of 2000 N and their head should not stand out more than 15 mm out of the lower plate See figure in step 2 on page 29 for the layout location of these holes that must be provided in the bench top The fastening material is not part of this kit
14. pplications monitoring of drugs or testing of counterfeit drugs It can also be used in teaching when a mobile lab goes to students residing in remote areas The diode array detector integrated in the Agilent 1220 Infinity LC enables spectra analysis for purity identification and comparison with self created library data 8 Agilent 1220 Infinity Mobile Upgrade Kit s Agilent 1220 Infinity Mobile Upgrade Kit 000 ee 2 Site Requirements and Specifications e i e j Site Requirements 10 Specifications 14 This chapter provides information on environmental requirements physical and performance specifications gg Agilent Technologies 9 2 Site Requirements and Specifications Site Requirements 10 Bench Space The attenuation unit has a footprint of 480 mm 18 9 in in width and 445 mm 17 5 in in depth The overall room specified for operation width x depth x height of 560 mm 22 1 in x 665 mm 26 2 in x 970 mm 38 2 in already includes the space for a complete 1220 Infinity LC Instrument mounted onto the attenuation unit and considering the necessary room for air circulation electrical connections and safety area for the possible movement of the instrument on the attenuation unit For a comfortable handling of the solvent bottles additionally 20 cm 7 9 in should be kept free above the bottles See Figure 2 on page 11 for more details Agilent 1220 Infinity Mobile Upgrade Kit Site Requirements a
15. t is completely fixed by the two rear retaining brackets 3 Agilent 1220 Infinity Mobile Upgrade Kit 31 3 Installation Securing the LC to the attenuation unit 2 Close the two front retaining brackets NOTE The two front retaining brackets must always get closed and secured by pressing the arrestor button They will not secure the instrument if not properly closed 32 Agilent 1220 Infinity Mobile Upgrade Kit j Agilent 1220 Infinity Mobile Upgrade Kit ss d e e 0e eve a Operation Instructions a e j General Operation Instructions 34 e This chapter provides general operation information Agilent Technologies 33 4 General Operation Instructions General Operation Instructions 34 General Operation Instructions Please note that the design of the 7220 Infinity mobile solution is intended for protecting the instrument from influences during transit Analytical measurement however has to be conducted under stationary conditions Operating the 1220 Infinity LC in a mobile lab environment can require special equipment or facilities to comply with local safety regulations e g special solvent waste and storage containers Please verify that these regulations are met before you operate the 1220 Infinity LC in your mobile lab Preparing transit and relocation transit 1220 Infinity Instrument installed on the attenuation unit while the mobile lab is moving relocation Moving
16. the 1220 Infinity Instrument to or from the mobile lab e Park the arm of the autosampler before relocation or transit e Remove the solvent bottles from the solvent cabinet and store them in a proper safety storage e Verify that the retaining brackets on the attenuation unit are properly locked when the instrument is on the attenuation unit see Securing the LC to the attenuation unit on page 31 e Store the sample vials in a safe place during relocation and transit The sample tray is not a proper storage place General Operation Use sealed solvent bottle caps and waste containers to prevent dangerous vaporization More information can be found in the chapter Consumables on page 36 in this manual and also on the internet www agilent com chem safetycaps Agilent 1220 Infinity Mobile Upgrade Kit Agilent 1220 Infinity Mobile Upgrade Kit a 36 i e Consumables ee e0 6e eve omnii J This chapter gives an overview of the recommended consumables Agilent Technologies 35 5 Consumables Consumables Safety Cap with 1 shutoff valve GL45 5043 0225 Comes with Venting valve 5043 0232 Venting valve 5043 0232 with 1 um PTFE membrane PEHD waste can 5043 0243 Comes with a S60 safety cap 5043 0227 including 4 ports and 1 leak port The Charcoal filter 5043 0230 must be ordered seperately Charcoal filter 5043 0230 lt 6 e 36 Agilent 1220 Infini
17. ty Mobile Upgrade Kit j Agilent 1220 Infinity Mobile Upgrade Kit eve Appendix e i e i General Safety Information 38 Agilent Technologies on Internet 39 This chapter provides addition information on safety and web ee Agilent Technologies 37 6 Appendix General Safety Information EURE es alerts you to situations that could cause physical injury or death gt Do not proceed beyond a warning until you have fully understood and met the indicated conditions CAUTION A CAUTION alerts you to situations that could cause loss of data or damage of equipment gt Do not proceed beyond a caution until you have fully understood and met the indicated conditions 38 Agilent 1220 Infinity Mobile Upgrade Kit Appendix 6 Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at http www agilent com Select Products Chemical Analysis It will provide also the latest firmware of the modules for download Agilent 1220 Infinity Mobile Upgrade Kit 39 www agilent com In This Book This manual contains information about the Agilent 1220 Infinity LC Mobile Upgrade Kit G4292 The manual describes the following e introduction e site requirements and specifications installation e general operating intstructions Agilent Technologies 2012 Printed in Germany 07 2012 G4292 90000 EE Agilent Technologies
Download Pdf Manuals
Related Search
Related Contents
LASER SAFETY MANUAL - Case Western Reserve University Operator Manual Installation Instruction Introducción Sí No (o no estoy seguro) ¿Tiene el router WiFi MDA-Win8086 MANUAL K-Junior Smart EVO User manual - K Rack Systems etiqueta - Dow AgroSciences CEDIA® Cyclosporine PLUS Assay www.boombotix.com ABUS TVVR35011 digital video recorder Copyright © All rights reserved.