1. HARDWARE amp INSTALLATION INTEGRITY TESTS If a problem is suspected with the T4 equipment the Integrity Tests provide a means for checking whether or not the equipment is functioning correctly NOTE Always perform these tests before calling the help line to allow T4 to determine if there is an error with the equipment The Integrity Tests can only be selected from the Welcome screen If there is a problem accessing the Welcome Screen from the current display turn T4 OFF then back ON again SWEL0000 Diagnostic System 4 0 24 April 2001 DRG0012B Technical Diagnostic Ea E Help Config Ea E Desk uration At the Welcome Screen the Integrity Test will appear as an option button on the Control Panel Integrity Test 2 Select the Integrity Test button to view the the following Integrity Test menu 64 SRITO100 Integrity Tests Tests TestBook Integrity Roving Probe LAN Connection CSI Cable Loopback Serial Port Loopback Test v O U m a Use the up and down scroll arrows at the right hand side to highlight the required Test selection 4 Select Continue from the Control Panel to run the selected test A new display will be shown dependent on specific selection and the selected test will begin automatically NOTE When performing Integrity Tests it is important to remember that the results o
2. btained can be affected by adverse transient conditions such as extremes of temperature or voltage spikes Always repeat the tests several times to confirm any test failures