1. x 95 H e Weight 3Kg with battery e Application preset modes Up to 8 different preset modes PROBES e 3 5 and 5 0 MHz radial scan probe 170 degree only Length 16 e 3 5 MHz axial scan probe 85 degree and 170 degree e 5 0 MHz axial scan probe 85 degree and 170 degree e 3 5 and 5 0 MHz split dual axial scan probe 170 degree only Duo Scan 3 manual indd 13 D 15 06 2009 21 38 Duo Scan 3 Main service centres BCF Technology Ltd 3 Tailend Court Starlaw Road Livingston Scotland UK EH54 8TE Tel 44 0 1506 460023 Fax 44 0 1506 460045 Email service bcftechnology com Web www bcftechnology com BCF Technology USA 2625 Highway 14 West Suite K Rochester Minnesota 55901 USA Tel 1 800 210 9665 Fax 1 507 529 8205 Email usa bcftechnology com Web www bcftechnology com BCF Ultrasound Australasia Pty Ltd Unit 10 56 Norcal Road Nunawading 3131 Melbourne Victoria Australia Tel 61 3 9894 8980 Fax 61 3 9894 8991 Email service bcfultrasound com Web www bcfultrasound com Duo Scan 3 manual indd 14 15 06 2009 21 38 e Accessories Rectal Probe Axial Probe Batteries and charger Wrist Display Goggles Monitor Video Cable Carry Sack 15 Duo Scan 3 manual indd 15 D 15 06 2009 21 38 Duo Scan 3 Innovative Imaging www bcftechnology com Duo Scan 3 manual indd 16 15 06 2009 21 38
2. 3 DUO SCAN 3 CONSOLE e Keeping all the connectors clean and dry is especially important use methylated spirits or equivalent alcohol and a small toothbrush to gain access to the contacts flush with alcohol and leave to drain dry The outer case and PVC carrying case can be cleaned using a cloth warm soapy water The front panel can be cleaned with a cleaning agent such as isopropyl alcohol cleaning cloths Note If using soap and water on the front panel ensure that no water enters any of the connectors e Examine all cables power and probe on a weekly basis for signs of damage such as cracks or splits Pay extra attention close to connectors and joints If damage is found it is imperative to have repairs carried out immediately to avoid further damage or risk of electric shock e Any other problems or maintenance queries should be referred to a BCF Service Engineer To ensure optimum performance we recommend that the equipment is serviced once a year by a BCF Service Engineer The Service Engineer will thoroughly check the equipment ensuring that electrical safety is being maintained and advise of any work which appears to be necessary Duo Scan 3 manual indd 12 D 15 06 2009 21 38 Technical specifications SPECIFICATIONS e Operation Image Modes Standard B mode e Grey Shades 256 different shades e Power Battery Up to 5 hours in continuous use e 12V battery attached to the unit e Physical Dimensions mm 280 D x 195 W
3. of the display Each bar on the meter represents approximately 1 hour of run time The battery meter will flash when there is less than 5 minutes of power remaining The current battery voltage is displayed in the lower left of the display this is primarily for factory use Charging The battery must first be removed from the Duo Scan 3 before charging Attach the battery to the charger by securing both latches Once the battery is connected the charger may be plugged in and switched on An indication of battery condition may be observed on the charger when the Push button is pressed and the unit will show the state of battery condition during charging Low Good or Full The battery will require at least eight hours charging from fully discharged but may be left connected for longer without damage Duo Scan 3 manual indd 10 D 15 06 2009 21 38 2 Care and maintenance The Duo Scan 3 was designed for reliability in the animal husbandry marketplace BCF engineers and their appointed agents are available to provide rapid service assistance to ensure optimum performance of BCF systems However there is no substitute for careful usage As specialists in providing ultrasound imaging equipment for animal husbandry applications we understand the importance of having reliable equipment All BCF products are designed and manufactured to withstand the environments in which they are going to be used However should you expe
4. 22cm Frequency 3 5MHz Foetal Ageing Circles Off Mode 7 Range 25cm Frequency 3 5MHz Foetal Ageing Circles On Mode 8 Range 25cm Frequency 3 5MHz Foetal Ageing Circles Off These modes are the default factory settings Your scanner will be preloaded with the settings requested when ordered so the information provided above may vary The shorter ranges are for use in early pregnancy and for improved detail the longer ranges for late pregnancy and better penetration As the range is altered the speed of the probe varies to provide the best quality image The smaller the range the faster the probe rotates The scanner always powers on in Mode 1 To scroll through preset modes press A 1 4 3 Gain and power adjustment Both ultrasound gain and power can be adjusted by the user the menu along the bottom of the screen will display either Gain or Power in the 2nd box To switch between gain and power press the Select button To adjust the Gain or Power values use the Up and Down keys on the front panel 1 4 4 Freeze image To freeze the image press the power button briefly FREEZE Pressing this button briefly once more returns the scanner to live imaging Duo Scan 3 manual indd 9 D 15 06 2009 21 38 Duo Scan 3 1 5 BATTERY LIFE METERING AND CHARGING The Duo Scan 3 runs for approximately 5 hours on a fully charged battery the current charge level of the battery is displayed in the lower right
5. Duo Scan 3 User manual Innovative Imaging Duo Scan 3 manual indd 1 D 15 06 2009 21 38 Duo Scan 3 Contents page Introduction 3 Operator safety 4 1 Operating instructions 5 1 1 Preliminary switch on 5 1 1 2 Imaging display device connection 5 1 2 Power on off 6 1 3 Display layout T 1 4 Scanning 8 1 41 Standard operation 8 1 4 2 Mode selection 9 1 4 3 Gain and power adjustment 9 1 4 4 Freeze image 9 1 5 Battery life metering and charging 10 2 Care amp maintenance 1 2 1 Axial and radial probe 11 2 2 Duo Scan 3 battery 12 2 3 Duo Scan 3 console 12 Technical specifications 13 Main service centres 14 Accessories 15 Innovative Imaging 2 Duo Scan 3 manual indd 2 D 15 06 2009 21 38 Introduction BCF Technology Ltd is a market leader in the development and production of real time ultrasound pregnancy scanners for most animal species We have over 26 years experience of developing manufacturing and the distribution and support of animal imaging equipment BCF in consultation with leading experts in animal husbandry applications have combined the best features of the Duo Scan 2 with the latest state of the art technology The result is Duo Scan 3 an ultrasound scanner which is small light robust portable and easy to work with The Duo Scan 3 is designed and manufactured in Scotland by BCF Technology Ltd 3 Tailend Court Starlaw Road Livingston West Lothian EH54 8TE Scotland UK Telephon
6. e 44 0 1506 460023 Facsimile 44 0 1506 460045 Email info bcftechnology com Website www bcftechnology com As specialists in providing ultrasound imaging equipment for animal husbandry applications we understand the importance of having reliable equipment All BCF products are designed and manufactured to withstand the environments in which they are going to be used However should you experience any problems with your Duo Scan 3 BCF Technology Service Engineers will be available to provide rapid service support and ensure optimum performance of your BCF system The Duo Scan 3 has been designed to conform with the requirements of European Directive 89 336 EEC relating to electromagnetic compatibility EMC and also 2002 95 EC relating to Reduction of Hazardous Substances RoHS The CE mark on the rear panel is a symbol of this conformance If you require any further information regarding these directives please contact us directly Note All BCF Technology products are to be used for animal applications only Duo Scan 3 manual indd 3 D 15 06 2009 21 38 Duo Scan 3 Operator safety The Duo Scan 3 is designed to operate only from a 12V DC source The use of a 12V lead acid battery ensures safety from electrical hazard When charging batteries use only the BCF supplied charger This charger is matched to the battery capacity and will ensure long life and continued safety Higher charge rates from unsuitable chargers may lead t
7. ner of the screen 2 The grey bar contains 32 of the 256 shades of grey This helps with the adjustment of brightness and contrast and gives a comparison of grey intensity for the ultrasound picture 3 The current function of the soft keys is displayed along the bottom of the screen The menu disappears after approximately 6 seconds leaving only the scan voltage and battery capacity displayed To switch the menu back on press any button on the scanner 4 A battery capacity and voltage reading are displayed in the lower corners of the display Duo Scan 3 manual indd 7 D 15 06 2009 21 38 Duo Scan 3 1 4 SCANNING 1 4 1 Standard operation All functions of the console are set using the front panel keys shown below A e SS A SETTING SELECT Scrolls through Switches between Power or Gain down Power or Gain up preset Modes Power or Gain I O FREEZE Short press freezes image Long press switches on amp off Duo Scan 3 manual indd 8 D 15 06 2009 21 38 1 4 2 Mode selection The Duo Scan 3 comes with 8 default Preset modes for cattle pregnancy scanning as listed below Mode 1 Range 16cm Frequency 5MHz Foetal Ageing Circles On Mode 2 Range 16cm Frequency 5MHz Foetal Ageing Circles Off Mode 3 Range 19cm Frequency 3 5MHz Foetal Ageing Circles On Mode 4 Range 19cm Frequency 3 5MHz Foetal Ageing Circles Off Mode 5 Range 22cm Frequency 3 5MHz Foetal Ageing Circles On Mode 6 Range
8. nsion tube If a window or its seal leaks then the probe must be returned to BCF Further use of the probe with a window leakage will further damage other internal parts of the probe due to dirt from outside contaminating the internal gears e Do not forcibly bend or pull the cable connecting the probe to the module e Examine the probe regularly for any damage especially the window The chamber behind the window is filled with a silicon fluid and it is normal for a small bubble about 1mm diameter to be present A large bubble or signs of leaking fluid should be referred to a BCF Service Engineer e Any other problems or maintenance queries should be referred to a BCF Service Engineer Duo Scan 3 manual indd 11 D 15 06 2009 21 38 Duo Scan 3 2 2 DUO SCAN 3 BATTERY The battery unit is of sealed lead acid type and will give many hundreds of recharge cycles if the supplied charger is used While the battery is in good condition and is fully charged it will run the Duo Scan 3 for approximately three hours Electrical connection to the battery is through the latches on either side of the battery Battery output is protected with a resetting fuse but care should be taken to prevent short circuiting between latches The battery unit may be cleaned by wiping down with a damp cloth Dry the battery unit out at the earliest opportunity if it gets wet Warning The battery unit must not be stored damp severe corrosion will result 2
9. o hazards from fire or battery leakage in addition to reducing battery life Although the Duo Scan 3 runs from a low voltage source dangerous voltages are generated within the unit For this reason no attempt should be made to disassemble the scanner Should you require any assistance please contact a BCF Service Engineer Duo Scan 3 manual indd 4 D 15 06 2009 21 38 1 Operating Instructions 1 1 PRELIMINARY SWITCH ON Before connecting or disconnecting the probe ensure that the Duo Scan 3 power is switched off and that the probe connector is free from dirt Connect the probe to the front connector marked G Rotate the probe connector until the notches on the probe and scanner align and then screw the retaining ring clockwise to lock the connector in position 1 1 2 Imaging display device connection The Duo Scan 3 accessories allow the ultrasound image to be viewed on the display headset monitor or wrist mounted display options Before connecting an imaging display device be sure the connector is free from dirt Displays are connected to the Display connector H Make sure the connector is correctly oriented before pushing home and the external threaded metal lock ring is engaged carefully before tightening gently If you are using a wireless monitor you do not need to connect it to the console and the display connector cover plug may be left in place If the monitor uses a video cable follow the instruction
10. rience any problems with your Duo Scan 3 BCF Service Engineers will be available to provide rapid service support and ensure optimum performance of your BCF system The Duo Scan 3 is designed to be robust and reliable and to withstand the environment in which it will be used Should you experience any problems BCF Service Engineers are available to provide rapid service support and ensure optimum performance 2 1 AXIAL AND RADIAL PROBE e Always clean the probe immediately after use and before any contamination can dry in The probe head can be immersed in warm not hot water to clean e The probe connector carries critical pulses from the console to the probe head and back It is imperative to keep these contacts clean and free of contamination To maintain a good contact clean the connectors using methylated spirits or equivalent alcohol and a small toothbrush to gain access to the contacts flush with alcohol and leave to drain hanging downwards e Never force a connector if it will not go together easily Debris may collect between the pins and can be carefully removed with a small screwdriver e Be very careful to avoid anything getting into the gold plated connector inserts this may cause damage and some of these are expensive to replace e Do not knock or scrape the transducer window this may cause the window to burst Do not push against the probe window the sealing may be damaged or fluid may be forced out of the internal expa
11. s above D OBBO SETTING SELECT DUO SCAN DISPLAY POWER G H CE y BCF veth Imaging WWW BCFTECHNOLOGY COM Duo Scan 3 manual indd 5 D 15 06 2009 21 38 Duo Scan 3 1 2 POWER ON OFF Duo Scan 3 is designed to operate only from a 12V DC battery pack Use only the BCF supplied battery pack Ensure the charged battery pack is securely latched to the Duo Scan 3 Press and hold the POWER button on the front panel until the system beeps FREEZE The green DC POWER light located above the power button should be illuminated and the scanner will now perform a system initialisation If the probe is connected it will start to rotate If the imaging display device is connected its screen will be illuminated The sector scan and settings will be displayed on the screen The buzzer will sound twice the second indicates that the system initialisation is complete and the system is ready to function If the supply voltage is too low less than 9 volts the system will display a Low Battery message for 6 seconds before powering down To switch the scanner off press and hold the power button until the scanner shuts down FREEZE D OGBO SETTING SELECT DUO SCAN DISPLAY o POWER G H WWW BCFTECHNOLOGY COM J Duo Scan 3 manual indd 6 D 15 06 2009 21 38 1 3 DISPLAY LAYOUT 1 The current range operating frequency power and gain settings are displayed at the top left hand cor

Download Pdf Manuals

image

Related Search

Related Contents

Amarra Symphony 3.0 Read Me Overview  7186.801 7186.811 Mesures relatives à l`installation    Hardware Installation Guide  L20, L40, and L80 series of tape libraries  OPERATOR`S MANUAL MANUEL de L`UTILISATEUR  Laderegler - bei der Westech  6 種類の核種の参照放出源 種類の核種の参照放出源 計測表  thesis-public.  Mode d` emploi  