Flex Integral FC and FCV User`s manual

image

Contents

1. day rinse Monday warm Tuesday cold Wednesday cold Thursday warm Friday cold Example 3 ofre DO gt 0 O 5 Flex Integral FC and FCV user s manual 1999 01 01 4 Hygiene and care Morning rinse On such units it is possible to provide a 18 minute flush through the sy stem FCV the count down is visible on the screen During the process the two light emitting diodes on the instrument bridge flash When starting one beep is heard When ending 3 beeps are heard To spare the equipment the water is only 90 C every third time or if more than 35 hours have passed since last rinse Amalgam separator Cleansing the cuspidor drain 1 Activate section 3 on the separator display while the suction is on 2 Activate the cuspidor flush 3 Release the button when all the water has passed through Replacement of the receptacle The amalgam receptacle must be changed every 6 9 months 1 Switch the unit off loosen the 2 lock screws 1 and 2 with a coin widen the cover a little at the top and remove it Y Unscrew the lid on the new receptacle Za Put protective gloves on and change the receptacle 3 Change the coarse filter 4 and put the old filter in the filled receptacle al aS Pour the disinfectant supplied with the new receptacle into the filled receptacle and
2. DS EN 46001 Flex Integral FC and FCV unit EN 60601 1 2 Flex operating lamp DS En 1640 Flex Integral patient chair Il Flex Integral FC and FCV user s manual 1999 01 01 Contents Listofversions for YA 8 78 A s 2 Mis AGUNG E 3 E A E E EEE EET 4 PE ET CU 6 231 le Te Be LTE 6 ae Au lte EE 6 Sie USC A A 7 FOO CONTO EE 7 FEW Unit WITT SCENE danita ais 7 KSE e LE 8 SUCIO ira He ea ee ee Mee a eee 8 Guspidor flush sai tie eis cad eve al 8 Assistant Call ai AST ETT 8 Operating lamp EE 8 Stop Watanabe 8 Ek AAM WETTER SES CHE EE Cup OT 9 VarioFlex operators StO0l oooococonnoniconinonoconncocccnnnca ronca 9 ue E ue TE EE 9 32 IASTUMENS is i tai 10 Flex Integral QF 00 A A A AA a el a AA eels 10 EE Flex Integral leia li Wied ae dat SEA pr eg 10 A ai E NE NEARE AEAEE N E AAAA NAE 10 O EE E E 10 MO A A IDA 10 Surge plateren it aaa taa toas 11 VU A a eee 12 Ultrasonic scaler type Odontoson cocccccccccococcnonocononccononanononnnnn cnc nn canaria rcnnnns 13 Composite Curing lanpi ina A a ie ees 14 Fibre Op DOE crucis ta ta tia de 14 d HYQIONG and Care cio ini da indi EEN 15 Hygiene routines and cleansing ooomocccnnnncccnnnonccnnnnonacnnnnnnocnn craneo cnn canon nn nnnnn cnn nnnn cnn cnn nnnn cnn ranma caninas 15 Suction HygiFlex Vac HygiFlex Vac Ultra oooooconncccnnnccnnccononnonocnnnnoccnnnc cnn naar n nn rca rca ranma 15 Cuspidor and cup EE 16 Une E EE 16 QUICKFIEX COU PIINGs s22
3. Outer sleeve complete for Flex 3 and 6 fUNCtION spray cnn SD 510 ISX SYNC TU EE SD 214 Flex Integral motor jacket ECH SD 216 Flex Integral motor jacket INTRA A SD 276 Saler p EE SD 217 Composite curing lamp tester c ccceccceceeeeeeneeceeeeeceaeeeeaaeeeeeeeseaeeesaaeeeeaaeseeeessaeeesaeeseneeenaees UC 665 Standard fibre optic diode 8 mm 70 for Flex Integral composite curing Jamp SD 220 Alternative fibre optic diode 8 mm 90 for Flex Integral composite curing laMp SD 221 Alternative fibre optic diode 13 mm 70 for Flex Integral composite curing lamp SD 222 Protective looking glasses used with the composite curing lamp oocooonnnnccnnnnncicnnnniccccnnnacccnns SD 223 Protective shield for composite curing lamp nn norc cana norccnnnnrrccnnnn SD 205 Jacket for large SUCTION nonai eare ea a aaea aa ee cnn eea ian rA a nena e i ranas MC 188 Jacket for small Suction susce uienn a ed a Da la MC 190 Adaptor pieces 11 7 mm MC 263 Flex Integral FC and FCV user s manual 1999 01 01 8 Spare parts etc Tips and tools for Ultrasonic scaler type Odontoson Standard straight For removal of supra and subgingival tartar labially and lingually on incisors Also used for removal of discoloured plaque E Standard right For removal of supra and subgingival tartar especially in 1st and 3rd quadrant Standard left For removal of supra and subgingival tartar especially
4. gt Increase the amount of water spray this way SS FCV the actual amount of water spray in ml min is visible on the screen lt Reduce the amount of water spray this way Refer to chapter 5 for programmering the air jet volume The unit can be programmed so that brief downward activation of the pedal turns the fibre light on off and a brief pedal activation on the left produces a chip spray commencing with water spray and finishing with air Surgery plant As an accessory the unit can be fitted with a surgery plant making it possible to spray with sterilized salt water when the motor rotates The normal spray cooling and chip blow functions are automatically disconnected Mounting and activating A Mount the frame on the right or left side handle on the bridge 1 B Connect the cable to the socket under the bridge 2 C Place the salt water bag in the rubber sleeve and hang it on the frame 3 D Put a clip on both ends of the motor tube where it feels hard and mount the thick tube onto the clips 4 E Draw the piston back while placing the soft part of the tube in the groove 5 F Connect the tube to the bag 6 and pump the rubber sleeve up 7 G Take the surgery motor H Press the P button under the instrument bridge 8 J Put the motor back The yellow light emitting diode on the instrument bridge 7 flashes to indicate the surgery position Flex Integral FC and FCV user
5. instructions concerning use cleansing disinfection maintenance and installation not being observed The Flex dealer is not responsible for defects should the product be installed or repaired by persons who are not trained by Flex if the parts which are not delivered or approved by Flex for this purpose are mounted in the product or should modifications in construction be made by a third party The Flex dealer is not responsible for loss of profit delays lost earnings or any other indirect loss The Flex dealer s liability for defects is limited to the price agreed on for the defective part of the consignment Any claim for guarantee must be made to the Flex dealer Flex Integral FC and FCV user s manual 1999 01 01 10 Your feedback Do you have questions related to a faulty Flex product or if you have suggestions for improvements we kindly ask you to fill in the following form and send it to us Please state LNT EE Profession EE AMES E id E Which product is concerned Serial NOS mate liinda Date of installation ooonnooconnnnncnnnnnniccnnnaccccnanaccnnnnnarnnrnnnrannrnnnns Dealer subsidiary if ANY oooocccnonncnccnonociconanocnnonanacnnnnnncnnnnnnnns Service report No if ANY comcccnnncnnnnnnnccconccccnnnnnnnnnnanancnancnnns Software version see Start up display ccceeeeseeeeeeeeeees And now your question problem suggestion for improvement FUNCTION ta a ts Flex Integra
6. The unit will beep when beginning once and again when ready for use 3 times If you want to flush with clean water inbetween two patients omit adding disinfectant Never interrupt the HygiFlex process Afterwards 1 Disconnect the tubes from the unit and press the filters out 2 Wash the filters filter holders covers and nipples in a thermo disinfector Autoclave covers and nipples 3 Lubricate the O rings on the filter holders using Flex silicone grease before inserting clean filters Replace the suction tubes at least every 3 months For HygiFlex Vac never use other disinfectants than Flex Vac Clean For Filters and tubes contain mercury that must be handled in a responsible manner HygiFlex Vac Ultra never use other disinfectants than Durr Orotol Ultra or Durr Orotol Flex Vac Clean and Orotol Ultra are aggressive Any waste must be removed immediately Remove Flex Vac Clean with acloth Remove Orotol Ultra e g using the suction Wear gloves and goggles Read the warning on the container Cuspidor and cup filler 1 Remove the gold trap and rinse it 2 Turn the flushing pipe 1 to a side and remove the bowls for cleaning not in a thermo disinfector 3 Lubricate the O rings using Flex silicone grease before placing the bowls again Note Avoid using abrasive cleaners for the bowls The contained material may contain mercury that must be handled in a responsible manner Syringe The syringe ja
7. a i SC 973 Trimming pin set for screen with backlight 2 pre AE 308 Lubrication Tube Flex silicone grease coin dd a a YR 002 Grease JUN turbine cit lidia SA 051 Lubrimed grease for grease gun turbine 6 cartridges ooomnoccinniniciconncoccnnnocannnnnnnannnn nana nnnnnnnn SD 318 O rings QuickElex coUpliNng ee ere ta t SC 740 Coupling piece 0N Motor oi A AA A at SA 024 Bulbs Flex integral turbines ici HE 005 Flex Integral motor with INTRA COUPliNQ coooooccccnnnoccconnnoccconnnoncconananccnn nn cnnn nc crac HE 005 Flex Integral composite curing lamp ooooncccnnnnccccnnnaccconnnonccnnononccn nono c cnn rc cnn WH 004 Flex Integral operating lompe WH 001 Consumer goods Large suction tube without nipple or filter cartridge ooonnnininnnninicnnnnncccnnnococcnnnncccnnnnnnarnnnnnnns AC 279 Small suction tube without nipple and filter Cartridge ooonniociinicnnnnnnnnnnnnnncnnnccnnnrannnccanancccnnns AC 280 RUE Tee TEE SD 400 12 bottles of Flex Vac Clean 12 months suppl YR 035 E Geller CEET SD 401 Flex Make up for treatment of lacquered surfaces oocoocinnccccnnnococinnnoccccnnnorc cnn non ccnnnn rca YR 001 50 tube sets for surgery plants oooccoonocccccnnoccncnanoncncnanoncnn nano nnnncnno cnn rro rca r nana rra rn BA 061 8 x 1 1 litre sterile Salt water visioita AA e A cnn BA 062 Amalgam separator DU UC 664 Accessories spare parts 4 handles for instrument bridge surgical lompe SD 399 INSTUMOENTE Pads asin ains AC 543
8. ce eevee A AAA a Ne Ae ieee ees Gee Bn Lees 17 Motor MOS viana ele 17 Flex Integral Turbine type Band P oon eee ee eeeceeeeeenneeeeeeaeeeeeeaaeeeeeeaaeeeeeeaaeeeseeaaeeeeeeaaeeeeeeaaeeeeeenaeeeeneaas 17 TEUrDINGS Of other MAKES ocio de 17 Ultrasonic scaler type Odontoson cooccooccconoccconcccnonanononannnn corn cnn nora 17 Ultrasonic scaler of other MAKES A 17 Flex Integral Composite Curing lompe 18 Composite Curing lamp of other makes cee eeeeaeeeeeeeseeeeseaeeeeaaeseeeeeseeeesaeeseaeeseeeeees 18 FIDE optie te e 18 INSTUMON pad sitiada 18 Operating lamp siiani u a oda 18 EE 18 HygiFlex Thermo Spray duts isis aii dee da dad 18 au Elle ER Ee EE 19 Flex Integral FC and FCV user s manual 1999 01 01 10 Flex Integral FC and FCV user s manual 1999 01 01 Contents FE programming adjustmment crio icon tonalidad ica dci 20 Patient A o eee e 20 RN 20 RRE 20 Composite Curing laMp ET 20 Time and date FGV sc ail AA ee a ee ae ee ad 21 PAVE ie seeds beatae dt act cation tarda 21 Delete funcion est Zeene DESEN ta 21 NAT 21 Maintenance and epa ia 22 Operating lamp viii A pi a 22 SE Le TEE 22 Ee E ee ln e DEE 22 Motor MES KEE 22 Flex Integral Turbine typ B and P innnccccinnccccnnnoccccnononccnnnnnnncnnnnnnn cnn anna rn nn 22 Ultrasonic scaler type Odontoson coocooccccccccccconiconocanonocononcn ronca ranma 23 Flex Integral Composite Curing lompe 23 Technical Cata coioo ica ii ri tii 24 Spare Parts Blocca secavete secs evs
9. close it so the markings on the lid and on the receptacle meet If the receptacle is fitted incorrectly an alarm may occur with the orange section flashing and the audio alarm signal active Remember to order a new receptacle NB Amalgam waste is enviromental wast and must be properly discarded in accordance with the demands of the authorities 5 FC programming adjustment 2 pos lt gt zero post 1998 03 18 19 39 48 gt wi a 4 1998 03 18 19 38 04 LIT Patient chair Il You can programme 2 positions and the zero position 1 Set the chair to the position wanted for programming 2 Press the programming button P button on the chair continuously while moving the position selector 2 Await a beep for correct programming Release the P button and position selector Motor 1 Take the motor and press the P button 1 once 2a Choose speed principle by moving the joystick on the foot control linear 2b Turn all chip functions on off by moving the joystick on the foot control The choice of speed principle and chip functions for motors are individually made for each suspension 2c Increase or decrease the air pressure by moving the pedal increase decrease 3 Replace the motor Turbine 1 Take the turbine and press the P button once 2a Turn all chip functions on off my moving the joysti
10. control and if mounted with guidance from a screen Flex Integral FC and FCV user s manual 1999 01 01 3 Use Foot control The foot pedal 4 controls many functions including the instruments on the unit The pedal can be moved in three directions to the left to the right downwards and a brief activation of the pedal Using the joystick you can regulate for example the amount of water spray to the motor The joystick can be moved in four directions north west P east south EHS zf FVC unit with screen CHIP P The screen gives you easy directions an 25000 when all the instruments are in situ and you wish to start an alarm when one instrument is in use and for example you wish to adjust the maximum speed The small arrows in the top corner of the screen squares tell you in which direction to move the foot control pedal and disk Adjust light 1 and contrast 2 at the bottom rear of the screen with the trimming pins or a 2 mm Allen key by sticking it into the holes and turn To avoid early wear the screen enters a standby position automatically f the display disappears when there has been no activity for a certain D time Flex Integral FC and FCV user s manual 1999 01 01 3 Use TTET TTET 3 1 Basic functions Flex SE When all instruments are in position you can control the basic functions in FCV ER SE two different ways 7 Dire
11. upwards last position right working position left working position Il downwards zero position If armrests are mounted one armrest can be lifted and turned 90 degrees to ease entering and leaving of the chair Emergency Stop a programme movement by touching one of the chair switches or the foot control disk If the chair back collides with e g your leg it automatically stops any movement and lifts itself approx 5 cm Adjust height and chair inclination by activating pedal 1 on the chair in the four directions The pedal placement depends on the chair being right or left handed It is not possible to operate the chair while HygiFlex Thermo is connected The headrest can easily be pulled out To push it back press lock button 3 or 4 home To adjust the headrest inclination loosen the lock 5 Cup Filler The cup filler functions automatically If a hot water tank is connected to the unit the cup filler water will be temperate First place the cup in the holder after the unit has been turned on Use only cups of semitransparent plastic or glas A 5 Exposure to electrical light from an angle or sunlight can cause Chair up down ee interference in the automatics Remove the light or choose to control the Al Backrest 39 cup filler by pedal seat depth VarioFlex operators stool Operate the chair by pulling the two handles up and down Amalgam separator As an accessory the unit c
12. Flex Integral FC and FCV User s manual YA 878 vers 4 00 List of versions List of versions for YA 878 Reason Pages to be replaced New version Initials FC V all 3 00 1996 07 13 Correction page 20 3 01 1997 02 18 Correction page 22 and 23 3 02 1997 04 10 Correction page 3 3 03 1997 12 01 Integral 5 new software all 4 00 1999 01 01 Flex Integral FC and FCV user s manual 1999 01 01 1 Guide Dear Flex user This manual relates to both the Flex Integral QF and the Flex Integral 5 Units The manual describes the working of the units If you have installed instruments other than those which are described in this manual please refer to the manuals supplied with those instruments regarding use maintenance repairs and technical data Please read chapter 2 General in this manual before commencing your work Thereafter use the manual as your reference book The manual focuses on how the equipment works See our Ergonomics Report for optimum methods of working Our Clinical Design Guide will tell you more about optimum surgery design We hope you enjoy working with your new equipment Yours faithfully Flex Dental A S 0510 The following equipment is CE marked in The equipment complies with the requirements in compliance with the COUNCIL DIRECTIVE the following standards 93 42 ECC of Fourteen June 1993 DS EN 9001 CONCERNING MEDICAL DEVICES
13. actically possible to avoid unnecessary wear of the instrument Always empty the hand piece for any water before applying a new instru ment Wipe off any drops of water before applying an instrument Be careful not to expose the instrument to any bumps Be careful that the tip of the scaler does not come into contact with the patient s soft parts lips tongue etc as the tip may be warm If necessar se a mirror to hold the lips aside or apply a lip protector see chapter 8 Always make sure that the following parts are tightly secured 1 The black ferrite rod on instruments Use special tongue and pin see figure 2 Files in ENDO instrument Carefully use special key 3 Plastic hut on CEM instrument Use your fingers and tighten hard See list of type Odontoson scalers in chapter 8 Composite Curing lamp Take the composite curing lamp and activate this way It is then on for 40 seconds A beep is heard from the unit at start mid way and at the end Prolong the exposure by 20 seconds by activating the pedal while the lamp is on Prolong the exposure by 5 seconds this way Shorten the exposure by 5 seconds this way Never look directly or indirectly into the intense halogen light Fibre optic probe The fibre optic probe switches on when lifted off the instrument bridge Avoid activating the foot control to the side Flex Integral FC and FCV user s m
14. an be fitted with a D rr Amalgam separator The present level of waste content in the amalgam receptacle is only measured when the unit is switched on i e switch on the unit every morning even if it has not been switched off the previous evening During normal operation the green section lights 1 When the receptacle is 95 full the orange section lights 2 the yellow section lights 3 and an alarm signal is heard Switch off the alarm using 3 if you do not want to change the receptacle now The yellow section will continue to remind you that the receptacle needs changing soon When the receptacle is 100 full the alarm can no longer be suspended The receptacle must be changed Flex Integral FC and FCV user s manual 1999 01 01 ES 3 2 Instruments Flex Integral QF Instruments with Quick Flex coupling are disconnected by pressing the two lock buttons 1 and 2 Connect an instrument by pressing it straight over the pointed end of the coupling The QuickFlex coupling must be dry when an instrument is connected You can either connect a QuickFlex coupling or a motor with Flex 4 4 coupling or instruments of other make on the instrument suspensions 3 4 and 5 unless the unit has fixed motors Flex Integral 5 The instruments on the Flex Integral 5 are fixed Fixed instruments can be change by a service engineer Function The foot control activates the first instrument taken off
15. anual 1999 01 01 4 Hygiene and care In the morning After each patient Evenings after the last patient 1 Rinse the spray ducts Rinse the suction with clean Rinse the suction with Flex Vac water Clean Orotol Ultra 2 Insert gold trap and suction filters Clean the equipment when Remove the suction filters and necessary gold trap 3 Disinfect the equipment Disinfect the equipment Cleanse the equipment and treat with Flex Make Up 4 Mount sterile accessories and Mount sterile accessories and Mount HygiFlex Thermo prepare sterile instruments prepare sterile instruments accessories 5 Place a new plastic cup Place a new plastic cup Cleanse disinfect and sterilize loose parts Hygiene routines and cleansing Before closing for holidays or longer periods where the equipment is not used the motors and QuickFlex couplings must be removed by unscrewing from the suspensions and then cleansed and dried When referring to a thermo disinfector in the text it is a thermo disinfector for surgical use operating at a temperature of 90 C Only thermo disinfect the parts that have been specifically approved for thermo disinfection as mentioned in the manual When referring to disinfection in the text use isopropylic alcohol solution or surgical spirits denaturalized with isopropylic alcohol or disinfectant D rr FD 320 Disinfectants containing acid phenols halogens or sulpho compounds may cause damage to the equipment su
16. ccenedeccesvettvancceuenecescerevates se Aoa Sina sicrdstedescatavguendescartacteesuecus s 25 Guarantee conditions eege eege veces EE deg eer 28 Your GE Le EE 29 2 General Switch on the equipment pressing 1 The chair can be switched off separately pressing 2 When the mains are switched on a green button in the switch lights The unit gives 3 beeps when it is operational and the chair gives 2 beeps When the unit is in use water and air must always be connected This is essential for the unit s function especially in connection with the Combi Separator 2 1 Audio signals The unit gives different types of audio signals 1 Operational signal 1 3 high pitched notes beep meaning ready or understood 2 Operational error signal Low note given in the case of operational or technical error 3 Warning signal Alternation between high and low notes May mean overheating If the signal does not stop when the equipment has cooled call your service technician 2 2 Terminology Automatic double chip blow Chip blow An air spray with max air pressure controlled with your foot Spray chip Activation of spray water the instrument is not active followed by a chip blow The following section gives a description of how the equipment functions when leaving the factory See chapter 5 FC programming adjustment for the numerous possibilities of changing functions The equipment is operated with the foot
17. ck on the foot control Off on 2b The choice of chip functions is individually made for each turbine suspension Increase or decrease the air pressure by moving the pedal A CG increase decrease 3 Replace the turbine Composite Curing lamp 1 Take the composite lamp and press the P button 1 once 2a Adjust the duration of exposure A E ep Flex Integral FC and FCV user s manual 1999 01 01 5 FC programming adjustment 2b Adjust the cooling air flow by moving the joystick on the foot control 10 per activation 3 Replace the composite curing lamp The air stream is set at maximum at the factory The choice of duration of the curing lamp exposure and of the flow of cooling air is identical on all suspensions Time and date FCV j 1 When all instruments are in position press the P button 1 once D position p Sen 2a Choose year month or day Adjust CES 2b Choose time or minutes 3 Push the P button once Alarm 1 Push the P button once when all instruments are i Siu 2 Choose minutes or seconds 3 Push the P button once Delete function When pressing the P button 1 while switching the unit on you replace all your personal programming with the original values Extra The following functions can be programmed adjusted by a service technician Extra functions Spray chip by moving the foot control pedal shortly to the left Pedal contr
18. cket can be autoclaved at max 121 C Press the lock button 1 and slide off the jacket Remove the tip of the cover by loosening the union 2 Flex Integral FC and FCV user s manual 1999 01 01 Flex Integral FC and FCV user s manual 1999 01 01 4 Hygiene and care QuickFlex coupling The QuickFlex couplings must not be autoclaved but only surface disinfected Lubricate the couplings when required using Lubrimed grease also used for the turbine Do not lubricate the O rings with Silicone grease Motor MC3 The motor jacket 1 can easily be removed for autoclaving Do not pull the tube The motor itself must only be surface disinfected Lubricate the O rings with Lubrimed grease Flex Integral turbine type B and P The turbine is supplied with a nozzle cleaner and grease gun 1 Cleanse the exterior of the tube using a toothbrush dipped in disinfectant Y Cleanse the spray ducts using the nozzle cleaner and then blow dry them with the syringe amp Wrench the grease gun until the grease is visible at the tip Insert the tip in the burr aperture and wrench the grease gun a half turn al aS Insert the burr in the turbine and activate the turbine without spray for approx 10 sec 6 Remove the burr and remove excess grease The completely dry turbine and grease gun can be autoclaved After the autoclaving remove the turbine from t
19. ct operation stated under 1 8 7 Foot contact stated under 2 di due Suction Activate the suction by taking a tube out of its holder Note Pull out the suction hose holder when working without an assistant Cuspidor Start stop the cuspidor flush for 30 seconds by 1 awaiting automatic activation when the chair reaches its zero position or when the cup is lifted from the cup filler or 2 all instruments in position The amount of water can be altered by your service technician Assistant call a Call your assistant by all instruments in position Connect door opener bell or the like Operating lamp Switch the operating lamp on off by Kei or gt all instruments in position gt Note The lamp switches on automatically when the chair is moved into a programme position When the chair moves into zero position the lamp switches off Alternate between the 3 light intensities by When switching on the lamp it lights with the intensity last used Stopwatch only FCV units with SEN The stopwatch activates this way lt b At the second activation the watch stops and at the third activation the main menu returns Alarm only FCV units with screen The alarm is activated this way See chapter 5 for alarm installation ES Flex Integral FC and FCV user s manual 1999 01 01 3 Use Patient chair Il Select a programme position by 1 activating pedal 2 on the chair
20. e light emitting diode being damaged Bulb Do not touch the bulb or reflector with unprotected fingers Use gloves or a cloth 1 Remove the composite curing lamp from the unit and pull the light emitting diode 1 from the handpiece 2 Unscrew the bolt 2 and remove the cover 3 3 Release the reflector and bulb 4 by pressing down and outward 4 Hold the printed circuit board and pull the reflector off Insert the new reflector 5 Fasten the reflector behind the cams Assemble the lamp and mount it on a suspension 6 Hold the tip of the light emitting diode against a thick layer of paper and activate the composite curing lamp 7 Ifthe powerful light is not concentrated in the middle of the field adjust the position of the bulb Light filter Open the lamp as described above and remove the filter 5 Cleanse the lamp with a dry cloth When you assemble the lamp again make sure that the reflecting surface of the filter faces the bulb Flex Integral FC and FCV user s manual 1999 01 01 7 Technical data Flex Integral QF or 5 Rated voltage 220 230 VAC 10 50 Hz Rated power 2 200 VA Size of group fuse 10A Water pressure Min 2 5 bar max 5 bar Air pressure Min 5 5 bar max 6 bar Compressed air consumption 40 l min 5 bar Indoor temperature 15 C 35 C Weight Max 90 kg incl operating lamp Max load on instrument bridge 5 kg Flex Integral chair II Max outer measures Heig
21. etc using benzine Operating lamp Cleanse the operating lamp parabola using alcohol Cleanse the transpa rent cover using an antistatic detergent Surfaces Cleanse the surfaces of the equipment using soap suds of either soft soap or soap flakes Treat lacquered surfaces with Flex Make Up Cleanse rubber parts using benzine Disinfect surfaces using a cloth HygiFlex Thermo Spray ducts The HygiFlex Thermo system secures that the bacteria content in the unit water system is kept at an acceptable level Cannot be used together with the HygiFlex Vac Note lf the unit has not been fitted with HygiFlex Thermo rinse the spray ducts in the morning by activating the spray Night rinse 1 Turn the cuspidor flushing pipe 1 to the side and place the instrument holder with the intermediate couplings 2 in the bowl 2 Connect the extension tube to the cup filler tap 3 3 Remove instruments contra angles and motor and syringe jackets 4 Position all suspensions vertically for the screen to enter the HygiFlex mode Lock the suspensions using the lock button 4 under the instrument bridge 5 Connect the suspensions to the instrument holder The unit rinses using cold water for approx 5 min with 3 hours interval The unit must be switched on The air compressor and water supply must be connected Return to normal position be repeating the procedure in reverse Flex Integral FC and FCV user s manual 1999 01 01
22. gular control of the ferrite rod is recommended it must be tightened carefully Life span expectancy of the instrument As the instrument tip wears the effectiveness drops Change the tip when this happens Several factors determine the life span number of patients per day type of deposit amount of coolant used etc The average life of an instrument tip is estimated at 3 4 months in normal working circumstances Note The form and shape of the instrument tips are vital for their function and longevity Therefore one must not bend polish or in any other way change the shape of the tips Flex Integral Composite Curing lamp Curing test With the supplied tester you can measure the curing ability of the lamp using different plastics 1 Put the tester on a white piece of paper with the small opening downwards fill it with plastic and cover the plastic with a matrix X Place the tip of the light probe in close contact with the matrix and expose the sample for 40 seconds 3 After 5 minutes the plastic is pushed out of the mould Remove the soft material Measure the Polymerization depth using a slide gauge Flex Integral FC and FCV user s manual 1999 01 01 6 Maintenance and repairs Test the curing ability of the composite lamp regularly to avoid slow deterioration If the curing ability is reduced it may be due to the bulb loosing its power the light filter being dirty or th
23. he autoclave immediately Lubricate the turbine at least twice a day and before and after each autoclaving Turbines of other makes See separate handbook Ultrasonic scaler type Odontoson Cleanse the scaler exteriorly using a toothbrush dipped in disinfectant The completely dry scaler can be autoclaved Take the instrument off the handle and protect it with the sterilization cap before autoclaving Sterilize the handle at a temperature of up to 134 C and the instrument at up to 134 C Ultrasonic scalers of other makes See separate handbook 4 Hygiene and care BEA ex TAaB2 08 0 1 Tiida HABITET EN 18 00 LHEBHO O Flex Integral Composite Curing lamp The composite curing lamp is exteriorly cleansed using a cloth dampened with disinfectant Remaining composite material is removed immediately using surgical spirits Autoclave the completely dry light probe After autoclaving remove the light probe from the autoclave immediately Composite Curing lamps of other makes See separate handbook Fibre optic probe Cleanse the fibre optic probe exteriorly using a cloth dampened with disinfectant Autoclave the completely dry fibre optic probe After autoclaving remove the fibre optic probe from the autoclave immediately Instrument pad Cleanse the instrument pad of the instrument bridge in a thermo disinfector and then autoclave it Remove grease residue
24. ht 152 cm length 182 cm width 62 cm Max movement vertical 40 cm horisontal 90 cm Weight 105 kg Max lifting capacity 135 kg Flex operating lamp Light intensity 22 000 15 000 and 8 000 Lux Flex Integral motor MC3 Speed 100 40 000 r p m Speed control principle non linear or linear Torque max 1 5 Ncm contra angle 1 1 Max absorbed power 55 W Flex Integral turbine type B and P Speed turbine type B 240 000 295 000 r p m unloaded turbine type P 270 000 420 000 r p m unloaded Speed control principle linear or one step Flex Integral scaler type Odontoson Frequency 42 kHz Power control principle three steps Max absorbed power 10 W Flex Integral Composite Curing lamp Wavelength 400 500 nm Max absorbed power 25 W Polymerization time 15 60 sec Durr amalgam separator Total capacity max 5 l min Flex Integral FC and FCV user s manual 1999 01 01 8 Spare parts etc This chapter contains a list of tools lubrication and service agents spare parts and accessories that can be purchased Parts that are supplied with the equipment are marked with an asterisk Tools Order No Fork Spanner 13 22 mm MC 453 Fork spanner T3 MM ic a ee eal Welded eels MC 500 Aller koy o MI e ae aAA RRAS AA AE Ea YA 050 Allen KOV AMIN arios i A A ee ee YA 004 Fork wrench for headrest ssni teea ea Rae Aea p EAA A AA EE EAA Aa AAAA Ea REE TANS SD 388 Turbine NOZZle cleaner oamenii eaa i aeaea ati L
25. in 2nd and 4th quadrant Perio Removes subgingival tartar in deep pockets of max 14 mm Only one version exists today which is considered universal Crown remover Permits removal of crowns and bridges without ruining these Dissolves silicate based cement whereupon the restoration can be removed The cement must be a non plastic type Endo instrument Used for endodontic treatments Gives smooth canals and easy injection of gutta percha Used with Odontoson files in sizes 15 25 and 40 ODONTOSON Tool for Endo instrument D p x Flex Integral FC and FCV user s manual 1999 01 01 8 Spare parts etc Tips and tools for Ultrasonic scaler type Odontoson Instrument standard Straight ccccccceceeeeeeeeaeee ce ea aaaea aa aa aaaea aeaa a aaa aaia FH 106 Instrument standard obt FH 107 Instrument standard NC EE FH 105 Instrument Perio standard FH 119 instrument crown Ee dE SA 048 Instrument Endodontal FH 113 Files for ENDO instrument 10 X 3 pce 0 cece eee eceeeeeeeeeeeeeee eae eeeeaeeeeeeeseeeesaeeeseaeseeaeeseeneess FH 115 Instrument kit ENDO incl nes FH 116 TOO ENDGOgelru ment 0 akut EEN iii edad SC 811 Instrument universal FH 099 Instrument Thin line straight recta datada trabada FH 123 Instrument thin line right cccceeceeeeseeeeceeeeeeeeeeeaeeeceeeeeseeaaeeeeeeeeesgeaaeaeeeeeeeseseeeeeaeeeeeseeseees FH 124 ASTrUMENE THINNING lefa a ido ii dadas FH 125 Instrumen
26. ion with 3 beeps QuickFlex coupling In case of leaks between the QuickFlex coupling and the instrument change the three O rings Motor MC3 Fibre optic bulb Motors with INTRA coupling have built in halogen bulbs Do not touch the bulb with unprotected fingers Use gloves or a cloth 1 Pull the motor jacket 1 off 2 Change the bulb 2 O rings In case of leaks between the motor and the angle piece change the 3 O rings 3 on the connection tube Flex Integral turbine type B and P Fibre optic bulb Do not touch the bulb with unprotected fingers Use gloves or a cloth 1 Unscrew the rear part 1 from the turbine 2 Change the bulb Flex Integral FC and FCV user s manual 1999 01 01 6 Maintenance and repairs Ultrasonic scaler type Odontoson Hand piece The black ferrite rod may break in the hand piece if it is exposed to impacts In such cases it is important to remove the broken ferrite pieces Ss Se Replacement of ferrite rod ieee If the oscillations are not transmitted properly to the instrument it may be caused by a worn or broken ferrite rod Y 14 EU DOY L 1 Hold the instrument with the tongs hg Sae T 2 Stick the pin through the small hole close to the ferrite rod 3 Unscrew the ferrite rod and replace it with a new one Tighten properly Empty the hand piece for water before applying an instrument Check that there are no broken ferrite pieces in the hand piece Note Re
27. l FC and FCV user s manual 1999 01 01 10 Your feedback Date And SIQGMAture uc a e daa e o e daa o paa a Flex Integral FC and FCV user s manual 1999 01 01
28. olled cup filler Only FCV with the joystick facing west Only FC short push to the right A short activation of the pedal downwards switches the fibre optic light on off Removable functions Auto chip blow Pedal controlled chip blow time delay between the activation of the foot control and the activation of the instrument FC Pedal foot control of operating lamp Patient chair control of operating lamp and cuspidor flush Automatic cup filler see above Water heating Changeable Adjustable functions Cup filler and cuspidor bowl water flow Syringe air and water flow Flex Integral FC and FCV user s manual 1999 01 01 6 Maintenance and repairs Here we describe minor repairs that you may want to do yourself Operating lamp Bulb Do not touch the bulb or the reflector with unprotected fingers Use gloves or acloth 1 Switch the lamp off 2 Loosen the cover screw 1 and remove the cover 3 Press the spring 2 turn anti clockwise and remove it 4 Pull the bulb incl wire out and replace it Patient chair Il If the chair gives a constant beeping sound it has to be synchronized 1 Switch the chair mains off 1 2 Switch the chair mains on while pressing the position selector 2 When you let go of the buttons the chair will first be set in zero posi tion and then move to semi reclined position Finally the chair confirms the synchronizat
29. rfaces When referring to autoclaving it is an autoclave which operates using steam at max 135 C and 2 2 bar Only autoclave parts that have been specifically approved for such treatment as mentioned in the manual Frequent autoclaving may reduce the instrument life time Suction HygiFlex Vac HygiFlex Vac Ultra The HygiFlex Vac system rinses the suction using either clean water or a dilution of Flex Vac Clean Orotol Ultra and water Cannot be used at the same time as HygiFlex Thermo If the unit is not fitted with HygiFlex Vac or HygiFlex Vac Ultra the tubes are rinsed using a separate tank 1 Remove the covers on the 2 suction hose nipples 2 Connect the tubes to the 2 connectors 1 and pull slightly downwards to activate the HygiFlex mode Open door 2 HygiFlex Vac Pump 2 portions of Flex Vac Clean into the mixing tank at 3 When the bottle with Flex Vac Clean is empty unscrew the pump and use it on the new bottle HygiFlex Vac Ultra Load one spoonful of Orotol Ultra into the funnel behind the door If necessary tap lightly on the funnel to ensure that all the powder reaches the container 3 4 ea HygiFlex Vac Ultra Flex Integral FC and FCV user s manual 1999 01 01 4 Hygiene and care 5 Start the process by pressing the P button under the instrument bridge The processes last approx 5 and 5 3 4 min respectively the screen clock counts down
30. s manual 1999 01 01 Arranging for sterile operation K Fit sterilized motor cover instrument pad and hand and angle piece L Connect the thin tube to the external duct of the angle piece 8 M Connect the thin and the thick tube 9 N Regulate the water flow at 10 Water cooling on off P The green light emitting diode on the instrument bridge indicates when salt water is connected Change to normal motor function by repeating items G J Turbine Take the turbine and activate the foot control pedal Choose a spray combination by keeping the foot control pedal pressed down until the desired combination is shown on the light emitting diodes of the instrument bridge Green water yellow air Two combinations are possible spray nothing Use of water is automaticallv followed by a double chip blow TTA Activate a chip blow briefly Increase maximum rotation speed this way lt FCV the speed is visible on the screen Reduce maximum rotation speed this way FCV the speed is visible on the screen Increase the amount of water spray this way FCV the actual amount of spray water in ml min is visible on the screen lt Decrease the amount of the water spray this way SS FCV the actual amount of spray water in ml min is visible on the screen Refer to chapter 5 for programmering the volume of air The unit can be programmed so that brief downward ac
31. t CEM id FH 126 Plastic hut CEM instrument 10 PCE vcrius nirien TE AERE TE E UC 759 Lip protect EE FH 127 Universal Used for removal of supra and subgingival tartar in all areas and surfaces FB es Thin line straigth Used for smoothing after rough depuration Also gives access to the furcation area and is used for root smoothing Thin line right Gives good access to the branches in 1st and 3rd quadrant Same tactile feeling as an explorer Used after rough depuration with another Odontoson instrument Thin line left Especially used in 2nd and 4th quadrant Identical use to thin line right O AS IS CEM instrument Used for cementation of porcelain filling Activates the cement and ensures a uniform division of the cement around the filling Flex Integral FC and FCV user s manual 1999 01 01 9 Guarantee conditions The Flex dealer assumes towards the buyer the responsibility for correct functioning faultless material and faultless workmanship for a period of 12 months from the day of delivery For ball bearings and rotors for turbines as well as fibre optic light diodes the guarantee period is 6 months from the day of delivery The Flex dealer guarantees that nondurable articles such as electrical bulbs rubber parts tips for scalers etc have no defects on delivery Nondurable articles are not covered by guarantee hereafter The Flex dealer holds no responsibility for defects due to ordinary tear or in case of Flex
32. the suspension not the syringe All adjustments are memorized until you re adjust The adjustments of motors and turbines are memorized individually for each suspension Do not activate the foot control pedal while changing an instrument unless the foot control is presently controlling another instrument Syringe The left button supplies air 1 and the right button water 2 ates Es As an optional extra the syringe can be supplied with a heating element A Be green light glows when the element is on It is possible to alternate between cold and warm water and air by rotating a switch on the connector Motor Take the motor off the suspension 2 Activate the motor with the pedal of the foot control a gt SS motor rotates clockwise motor rotates anticlockwise indicated by a beep Select one of the spray combinations by keeping the pedal pressed down until the desired combination is shown on the light emitting diodes on the instrument bridge see 7 on figure Green water yellow air Four combinations are possible spray air water nothing Flex Integral FC and FCV user s manual 1999 01 01 3 Use An automatic double chip blow follows after use with water Activate one chip blow ak y Increase the maximum rotation speed this way FCV the speed is visible on the screen Reduce the maximum rotation speed this way FCV the speed is visible on the screen gt Aa
33. tivation of the pedal turns the fiber light on off and a brief pedal activation to the left produce a chip spray Note The turbine must not rotate without a burr Use only burrs and diamonds with a diameter of 1 59 1 6 mm anda maximum length of 26 mm The burr must not be mounted on the turbine when out of use fora longer period of time Flex Integral FC and FCV user s manual 1999 01 01 Flex Integral FC and FCV user s manual 1999 01 01 3 Use Ultrasonic scaler type Odontoson Take the scaler Choose power level Keep the pedal pressed down till the desired power level is shown on the light emitting diodes of the instrument bridge FCV refer to the screen for power levels No diode on minimum One diodeon medium Both diodes on maximum Activate the scaler this way Increase the power level this way Reduce the power level this way Increase the amount of water coolant this way Reduce the amount of water coolant this way Use Trace the instrument tip parallel to the tooth and use only the side of thetip Work using as low a power level and contact pressure as possible Only use the tip on teeth Avoid contact with ceramics and gold Always use small brushing movements When using Thin line instruments with especially thin tips Flex Dental recommend working with max 50 of the maximum power level Always use as large a water flow as pr