Endoscopes Inspection Report – Sale ------ date: --- / --

image

Contents

1. 957 Valley Lab The diathermy powers up well and passed its initial self test The display Force FX functions and function buttons are in good working order The diathermy is Daithermy tested using an ESU analyser and the monopolar cut and coagulation output MA readings where within the limits Unable to verify the bipolar output as there is no footswitch The lot includes monopolar cut and coagulation footswitches however bipolar footswitch is not included Overall physical condition is satisfactory 961 Hill Rom Unit powers up with all of its electrical movements in working order Units Affinity 3 mechanical leg rests are also in working order Unit s wheels and brakes are Birthing Bed functional Unit includes cot sides and a head board Unit s mattress is in fair physical condition Unit is in fair physical condition 962 Hill Rom Unit powers up with all of its electrical movements in working order Units Affinity 3 mechanical leg rests are also in working order Unit s wheels and brake are Birthing Bed functional Unit includes cot sides and a head board Unit s mattress is in fair physical condition Unit is in fair physical condition 965 Marquette Unable to determine if unit is powering up as unit does not include a remote Beta Star control Unit s mechanical movements are in working order Units wheels Operating and brake are functional Unit includes a set of cushions which are in Tecan reasonable physical condition w
2. Image Quality Good image Working Channel s Clear Leak test Leaking Over all condition Satisfactory overall condition one light output is dim insertion tube is stained in colour and scope is leaking 115 OLYMPUS GIF Q230 Videoscope 1000x11 70mm Insertion tube Good physical condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good optics Remote control switches Fully functional Distal end Good physical condition Image Quality Good image Working Channel s Clear Leak test Pass Over all condition Good overall condition 116 OLYMPUS OSF 3 Scope 600x12 90mm Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Satisfactory optics Remote control switches N A Distal end Satisfactory condition Image Quality Satisfactory image 1 fibre is missing Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition one fibre is missing 117 WOLF 7316 Scope Insertion tube Good condition Bending mechanism amp Angulation control Good movement no locks Light guide cable amp Light transmission Good light transmission no light guide
3. modules included with the unit Overall the unit is in reasonable physical condition with minor signs of wear and tear The battery also found to be working at the time of inspection Air O2 and N20 medical gas hoses are included Absorber absorber assembly is not included Y O M 1999 5 Datex Power up and system self test completed successfully Display panel Ohmeda controls and trim knobs are all in good working order All ventilation modes Anaesthetic VCV SIMV PCV are selectable and appear to be functional but has not Machine wath been fully tested The patient monitor powers up and appears to be workin AS 3 Patient y l ee Monitoring the module rack is in place but no modules included with the unit no and S5 Screen further checks could be carried out Overall the unit is in reasonable physical condition with minor signs of wear and tear The battery also found to be working at the time of inspection Air O2 and N20 medical gas hoses are included Absorber absorber assembly is not included The unit comes with a spare S 5 monitor Y O M 1999 HG 116 v2 6 Fibrolux 250 Powers up ok Both lamp 1 and 2 are in good working order and the Light Source brightness control is also functional The unit is in good physical and cosmetic condition 7 Sonosite Unit s battery is flat and mains adapter is not included Unit was tested with TITAN another battery and its powers up ok with system self te
4. working order Unit has not been fully tested Unit does not include any hand Machine pieces Units modes are selectable but have not been tested Units IV pole height adjustment is functional Units button audio voice selection is functional The footswitch is responding but unable to verify if it is fully functional Unit is in satisfactory physical condition Version V3 25 Manufacture date 11 01 HG 116 v2 252 Aloka The ultrasound power up well but does not pass its self test and stays on a Prosound loading screen Therefore unable to carry out any further checks Includes 4000 two probes of models UST 9123 and UST 9124 but have not been checked cn a for its functionality The cable for the probes has physical damage to them 9123 and which has been taped Printer is included but has not been tested Overall UST9124 physical condition is reasonable Y O M 2004 Probes and Sony Printer 254 Drager The ventilator powers up well and passed its initial self test The display and Savina function buttons are in good working order The ventilator flow sensor and Ventilator on expiratory block is missing and therefore unable to fully test the ventilator Stand The back up battery is functional Overall physical condition is reasonable S W 02 01 and working hours is 4556Hrs 255 Alcon Legacy Unit is powering up with its function buttons and touch screen display in 20000 Phaco working order Unit has not been
5. Airshields functions and function buttons are in good working order The set air C2000 temperature is reached and maintained The temperature is verified using a marae thermometer which is within the limits The skin temperature is not tested as there is no skin probe The height adjustment and the scale are in good working order The mattress is intact The humidifier tray is not included HG 116 v2 Oxygen hose is included Overall physical condition is satisfactory 1094 Hill Rom The incubator powers up well and passed its initial self test The display Airshields functions and function buttons are in good working order The set air C2000 temperature is reached and maintained The temperature is verified using a Tacuba or thermometer which is within the limits The skin temperature is not tested as ZJ06183 f there is no skin probe The height adjustment and the scale are in good working order The mattress is intact The humidifier tray is not included Oxygen hose is included Overall physical condition is satisfactory 1095 Drager All three of the vapourisors are physically intact With the vaporisers being Isoflurane in fair physical condition with minor ware and tare and scratches to paint Vapor 2000 work Vapouriser x 3 One Has Damage 1096 Drager Unit has not been tested Unit shows to be physically intact With unit being Babylog 2 in fair physical condition Unit is missing
6. Lot includes 1x battery charger unit unable to determine if with Battery charger unit is powering on as there is no indication light All units are in Charger x1 reasonable physical condition with units having lots of scratching to the displays and covers 1041 Drager Both vaporisers halothane and isoflurane are physically intact With Vapourisers vaporisers being in fair physical condition with minor scratches and ware Halothane x1 and tare to casing and Isoflurane x 1 1042 Stryker 40L Both units are able to power with their function displays functional and their High Flow function buttons showing to respond No further checks have been carried Insufflator x2 out on units Units are in fair physical condition with some dents to display covers 1049 Marquette Unit powers up with units function buttons and display in working order MAC 5000 Unable to test unit as unit does not include any ECG leads Unit s battery is ECG Machine functional but needs charging Unit is in fair physical condition No Leads 46230 1050 Marquette Unit powers up with its function buttons in working order Units display is MAC 5000 functional however it has an area of white lines which run down the displays ECG Machine background image Unable to carry out any further checks as unit does not ys frien j include and ECG leads Unit s battery is functional but needs charging Unit is in fair physical condition 1051 Marquette Unit powers u
7. showed to be within limits The skin temperature mode is was tested with skin temperature and found to be working fine Skin temperature probe not included Oxygen sensor calibrated at 21 and humidifier monitor appears to be working fine Unit s motorised height adjustment and mechanical cot tilt are both functional Unit includes a gel mattress and an O2 hose Unit is in fair physical condition Side door covers missing On Off switch is a bit stiff to operate Version 2 18 396 Hill Rom Air Unit is powering up and self test completed successfully The LCD screen Shields Infant Incubator C2000 power up and function keys are in good working order The unit was tested in the air mode and the set temperature is maintained accordingly Temperature was verified using an external thermometer and temperature showed to be within limits The skin temperature mode is was tested with skin temperature and found to be working fine Oxygen sensor calibrated at 21 and humidifier monitor appears to be working fine Unit s motorised height adjustment and mechanical cot tilt are both functional Unit includes a gel mattress and an O2 hose Unit is in fair physical condition HG 116 v2 Version 2 18 397 SECA CT6 Pi Unit is powering up and LCD screen amp keys pad are working fine Unit was ECG machine tested with patient simulator and found to be working ok with good ECG traces on all leads were displayed an
8. test Units display and function buttons are in working order In air mode unit shows to be able to heat up well and maintain its set temperature Temperature was verified using an external thermometer and temperature showed to be within limits Unable to test units skin mode as unit does not include any skin temperature probes Oxygen and humidifier modes have also not been tested Units electrical height adjustment and mechanical cot tilt are both functional Unit includes a gel mattress and an O2 hose Unit is in satisfactory physical condition Version V2 18 238 Hill Rom Air Shields C2000 Incubator Unit is powering up with unit being able to pass through its initial power up test However unit then displays error messages controller failure 10 and controller failure 14 with a constant alarm Unit is not functional Unit display shows to have pixels missing and the function buttons show to be responding Units electric height adjustment and its mechanical colt tilt are both functional Unit includes a mattress and an O2 hose Unit is in satisfactory physical condition Version 2 18 239 Conmed Sabre 180 Electrosurgic al Unit with Foot Switch The diathermy powers up well and passed its initial self test The display functions and function buttons are in good working order The diathermy is tested using an ESU analyser and the monopolar cut and coagulation output readings where within the limits Unable t
9. Error 4000 message of Disk boot failure insert system disk and press enter therefore Ultrasound unable to carry out any further checks Includes three probes of models C5 Hoa 2 C7 4 and C8 4v which are physically intact but have not been fully Pr ohe g tested Overall physical condition is reasonable 29 GE Dash 4000 All 3 monitors power up ok Display screens controls and trim knobs are in Patient good working order ECG SpO2 NIBP and Temperature parameters are Monitors x2 functional tested with patient simulator Batteries are in place on all 3 aite k monitors and overall the units are reasonable physical condition The light Dash 3000 cover of the handle on of the monitors is missing Accessories not included Patient Monitor 112766 116839 115304 No Leads 30 Olympus The processor powers up with function buttons and indication lights in OTV S4 working order The processor and camera head were tested using a monitor Camera not included The processor and camera head showed to produce a poor ee wah image as the image it produces has bad interference and is badly blurred resulting in there being no proper picture being obtained Unit is in fair HG 116 v2 physical condition 36 SES Little Powers up ok Display panel controls and door locking mechanism are in Sister 3 good working order No further checks carried out The unit is in reasonable Autoclave physical condition
10. M is 2003 52 Zoll M Series Powers up ok Display screen and controls are in good working order Both Monophasic monitoring and defibrillation modes are functional tested with defibrillator Defib with analyser and the energy delivery was within the limits Pacer mode is also Pacing selectable and appears to be working but has not been fully tested Battery Battery and f Te Oe Paddles ECG lead set and paddles are in place The unit is with case and overall in Mains good physical condition Printer is also functional Note the RAM battery Charging appears to be deflated unit shows clock fault 12 after system power up 53 Zoll M Series Powers up ok Display screen and controls are in good working order ECG Monophasic monitoring is working fine but the defibrillation mode is not fully functional Detib with the unit does not charge to the selected energy and shows pacer fault 117 ae ji when the charge button is activated Pacer mode is also selectable and Paddles appears to be working but has not been fully tested Battery ECG lead set Mains and paddles are in place The unit is with case and overall in good physical Charging condition Printer is also functional Note the RAM battery appears to be deflated unit shows clock fault 12 after system power up 54 Gyrus G400 Powers up and completes the system self tests successfully Display screen Diathermy and controls are in good working order
11. The system menus are accessible but with Foot no further checks has been carried out The unit is in good physical and Switch cosmetic condition Footswitch is included Y O M 2006 62 Alcon The phaco machine powers up well and passed its initial self test The Universal display and function buttons are in good working order The phaco machine Phaco has not been fully tested but appears to be in good working order Includes ee footswitch and one handpiece S N IMHP060295IJT Overall physical Handpiece condition is reasonable 65 Eschmann The autoclave power up ok displayed error cleared Display and function little Sister keys are working fine Only cycle 134 deg C wrapped instrument cycle is Quick Vac selectable The cycle was tested and it completed all the phases with no error The temperature and pressure values were maintained as per set cycle programme Water level and printer are working fine Unit has no power cable or trays System is in good physical condition Year of manufacture 2005 68 Drager The vaporizer has not been fully tested but is in average physical condition Enflurane due to usage The dial appears to be functional Vapouriser 71 Welch Allyn All 3 monitors power up ok Display panels and controls are in good Propaq CS working order ECG SpO2 and NIBP parameters are functional on all 3 Monitor units temperature and IBP parameters have not been tested Two of the units NORE contains CO2 module as well but has not been
12. Unit is in fair physical Semen condition 14655 449 Lifepak 12 The Lifepack 12 battery charger does not power up and the power supply Vehicle connector is loose therefore unable to carry out any further tests Includes Chargers x3 three DC power adaptors which have not been fully tested and Lifepak 12 Battery Tester Charge r 452 Olympus Unit powers up with its function buttons and indication lights in working CLV S Light order Unit showed to be able to produce a good light output with its Source brightness adjustment in working order Unit is in satisfactory physical nie condition 455 Resmed S8 The x12 CPAP units are included which have not been tested but appear to CPAP be intact Controller x 12 Resmed S9 CPAP 28683 55851 59252 57322 47950 47955 39551 58083 28242 57471 62084 460 Olympus All units power up ok The processor produces good image quality when Imaging tested with videoscope and video cable pig tail The light source xenon Stack Trolley jamp is working fine Colour monitor is fully functional The system is in With CV 200 fair physical condition and fully functional Unit does not include vid amp CLV U20 physical condition and fully functional Unit does not include videoscope or video cable pig tail 461 GE Exercise Unit power up and system startup with Windows XP ok The CASE programme start up ok The treadmill is detected and access to the function menu and operating screen is fin
13. Water cap is missing Cycle counter 17646 46 Respironics Unit is powering up however the units display is not functional with it being BPAP Vision badly damaged with the LCD liquid being damaged and the display Ventilatory shattered Unable to determine if the function buttons are responding Unit Support does however show to be giving an output but unable to verify if the output System D Sy me is fully functional Unit includes an O2 hose Unit is poor physical condition with is damaged display and damage to its cover around the display 47 Respironics Unit is powering up and is able to pass through its initial self test BPAP Vision successfully Units function buttons and displays are in working order Unit Ventilatory shows to be functional giving an output however the output has not been Support verified Unit includes an O2 hose Unit is in fair physical condition System Operation hours 6299h 49 Olympus Both units are able to power up with function buttons and indication lights OTV SX2 in working order Unit includes 2x camera heads however only one of the Camera camera heads is an OTV SX2 model The second camera head is an OTV S5 E A model which is not compatible with this processor The OTV S5 camera and Olympus head has not been tested The OTV S2 camera head and the processor were CLV S30 tested using a monitor not included The camera head and processor Light Source showed they were able to produce a satisfactory i
14. but the NiBP and Patient SpO2 parameters appear to be in good working order Printer has not been Monitor x 1 fully tested as there is no printer paper Lot includes five NiBP hoses with Pro30V2 cuff four complete SpO2 probes and one incomplete SpO2 trunk Overall Patient physical condition is average Monitors x 2 332 Eschmann All articulating movement are in working order Units hydraulic height MR Hydraulic adjustment is also in working order Units roller wheels and brake are Operating functional Unit does not include any cushions Unit is in fair physical Table No condition Cushions 335 Ohmeda Ohio Unit powers up with its displays in working order The raise bed function Infant button does not show to be able to activate with the raise and lower bed besos function not functional The rest of function buttons show to be in working order In manual mode the unit shows to be able to heat up well Unable to HG 116 v2 test servo mode as unit does not include a skin temperature probe Units overhead light and its timer are functional Unit is in fair physical condition 337 Hill Rom Air Note Unit can be temperamental in powering up and sometimes can flicker Shields C2000 on and off Incubator However unit is able to power up with unit being able to pass through its initial power up test Units display has a large area of pixels missing the function buttons are in worki
15. cable HG 116 v2 600x3 20mm Objective lens and focusing Below satisfactory optic focusing Remote control switches N A Distal end Good condition Image Quality Below satisfactory image Working Channel s Clear Leak test No leak tester Over all condition Good physical below satisfactory image 175 OLYMPUS GIF 1T30 Scope 1000x13 00mm Insertion tube Satisfactory condition there is a slight dent 150mm up the insertion tube Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Satisfactory as one light output is slightly dim Objective lens and focusing Satisfactory optics Remote control switches N A Distal end Satisfactory condition Image Quality Satisfactory image 2 fibres are missing Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition 2 fibres are missing one light output is slightly dim 118 OLYMPUS GIF XQ10 Scope 1000x10 70mm Insertion tube Satisfactory condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission one of the light outputs is dim Objective lens and focusing Satisfactory optics Remote control switches N A Distal end Satisfactory condition Image Quality Below satisfactory image as 50 fibres are missing Working Channel
16. condition 608 Aloka ASU The 4D convex probe has not been fully tested but is physically intact and its 1010 4D overall physical condition is good Frequency range is 1 15Hz and model is Convex Probe asy 1010 609 Esaote TV Probe cable and connector are in good condition There are significant probe EC 123 dropouts in the middle of the scan image Sign of damaged crystals 610 Toshiba probe Cable probe and connector are in good physical condition probe not tested PVG 681S 6MHz 621 Suction Unit Unit powers up Suction motor running fine Good vacuum recorded on the manometer Vacuum regulator in good working order The main On Off switch is loose Unit is in satisfactory physical condition 627 Bausch and Unit is able to power up however units display is only functional and on Lomb when the monitor is tilted all the way down with touchscreen then being Millennium functional Both the anterior and Posterior modes are selectable but unable to PER be tested unit does not include any hand pieces The IV pole height adjustment is functional Unit includes an air hose Unit is in fair physical condition Software version 2005 1 647 Luxo The lot includes five table mounted magnifying examination lamps The Illuminated locks and all movements are in good working order The lamps are fully Magnifiers x 5 functions where both its bulbs are in good working order However one of the lamps does not turn on and the other lamp has one of its
17. is also working The motorised height adjustment is functional Unit is in good physical condition with gel mattress included 816 Philips Six of the defibrillators power up well and passed its self test Its display and Heartstart function buttons are in good working order The defibrillators have not been Defibs x 6 fully tested but appear to be in good working order Six batteries are included but one of the batteries is low Overall physical condition is reasonable 817 Datascope All 5 monitors power up ok Display panels and controls are in good Passport working order ECG and SpO2 parameters are functional and the NIBP Monitors x 5 pumps are also working The units are in reasonable physical condition with minor signs of wear and tear on some of the function keys The ECG HG 116 v2 parameter on one of the monitors does not detect the leads and therefore could not be tested 834 Respironics The vent inop indicator lights up stays on blank screen continuous alarm Bi Pap Vision is sound and does not go into main menu Therefore unable to carry out any oe further tests Includes O2 hose Overall physical condition is reasonable 842 Datascope 4 of 5 monitors power up ok Display panels and controls are in good Passport 2 working order ECG and SpO2 parameters are functional and the NIBP Patient pumps are also working The units are in reasonable physical condition with oe ee
18. powers up Unit is able to produce a good suction with its regulator in Dominant 50 working order Unit includes 3 suction jars its bacteria filter is broken Unit Suction Unit ig in fair physical condition 391 Drager Unit powers up with its function buttons and displays in working order In Incubator air mode unit is able to heat up well with unit being able to reach set 8000 SC temperature and maintain it Units temperature was verified using an external thermometer and temperature showed to be within limits Units humidifier mode has not been checked Units electrical height adjustment is functional units mechanical cot tilt is not fully functional with one side of the cot tilt getting stuck when it is being lowered Some of the units cable access grommets are missing Unit includes a mattress Unit is in satisfactory condition 394 Ohmeda Unit powers up Units mechanical movements are in working order Unit is Phototherpy able to produce a strong light output The fan makes a loud rattling sound Light Unit is in fair physical condition 395 Hill Rom Air Unit is powering up and self test completed successfully The LCD screen Shields Infant power up ok but there are a few pixels metrix missing but display is Incubator readable The function keys are in good working order The unit was tested C2000 in the air mode and the set temperature is maintained accordingly Temperature was verified using an external thermometer and temperature
19. s Clear Leak test Pass Over all condition Below satisfactory overall condition as 50 fibres are missing and one light output is dim 119 OLYMPUS GIF XQ30 Scope 1000x10 25mm Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good optics Remote control switches N A Distal end Good condition Image Quality Satisfactory image as 2 fibres are missing Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition 2 fibres are missing 120 HG 116 v2 OLYMPUS GIF XQ30 Scope Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good optics Remote control switches N A Distal end Good condition 1000x10 90mm Image Quality Satisfactory image as 4 fibres are missing Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition 4 fibres are missing 121 OLYMPUS SIF 10 Scope 1600x12 70mm Insertion tube Satisfactory condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good op
20. tested Chargers are included Monitor and overall the units are in reasonable physical condition Includes 1x SpO2 Propaq CS probe 2x ECG lead sets 1x NIBP hose and a few IBP trunk cables Vital Signs HG 116 v2 Monitor and Leads 73 Spacelabs All 3 monitors power up ok Display panels and touch screen interface are in 91369 Patient good working order The module is in place on each monitor and the ECG Monitors x3 p02 and NIBP parameters are functional tested with patient simulator With Box ot Mains adaptors are included batteri also in place and overall the unit ann ptors are included es are also in place e e S Leads are in good physical condition Includes a large box of accessories 74 GE Logiq The ultrasound powers up well and passed its initial self test The display 100 Portable functions and function buttons are in good working order The portable Ultrasound ultrasound is tested using a phantom and set measurements taken are within ied the limits The grey scale is good but the resolution is poor Ultrasound modes included are B and M which are functional The probe shows some dropouts but does not affect the image The probe has some slight interference at times The overall physical condition is average due to usage Y O M 2000 80 Haag Streit Unit powers up Unit is able to produce a good light output Unit does not Slit Lamp on include a fixation light The optical system i
21. the bed are working Afirmity fine All the motorised positions are working ok CPR release function is Birth Bed functional Brakes and castors are in good working condition Good overall physical condition 303 Hill Rom The bed power up ok and control panels on both side of the bed are working Afirmity fine All the motorised positions are working ok CPR release function is Birth Bed functional Brakes and castors are in good working condition Good overall physical condition 304 Hill Rom The bed power up ok and control panels on both side of the bed are working Afirmity fine All the motorised positions are working ok CPR release function is Birth Bed functional Brakes and castors are in good working condition Good overall physical condition with tears at the foot strap bracket rubber cover 305 Hill Rom The bed power up ok and control panels on both side of the bed are working Afirmity fine All the motorised positions are working ok CPR release function is Birth Bed functional Brakes and castors are in good working condition Good overall physical condition with tears at the foot strap bracket rubber cover 329 Dinamap Pro All five vital signs monitor power up well but one of the monitor freezes and 300 Patient produces a continuous alarm The other four monitor s powers up well and Monitors x 2 passed its initial self test Its display and function buttons are in good Pro 400 working order The monitors have not been fully tested
22. 1 the crystal causing some small areas of dropout to the image Probes showed x 7 5MHz to produce reasonable resolution and greyscale The printer is functional Probes Unit is in fair physical condition Mitsubishi P91 Printer Sony UP D895MD Digital Graphic Printer and Agfa Linx PA Xport 437 Stack Trolley All units are powering up The Lemke processor and camera head are able to with Sony produce a satisfactory image on the Sony monitor with the camera head also LCD Monitor showing to have satisfactory focusing The Stryker processor powers up but il does not include a camera head so unable to test if the processor is C mera functional The printer is functional Lot is in satisfactory physical condition Control Unit Camera Head Stryker HG 116 v2 988i 3 Chip ECG CASE and Treadmill camera Control Unit and Some Assorted Cables 439 Vesamed The ventilator powers up well and passed its initial self test The display and Vent function buttons are in good working order The ventilator however has not Ventilator been fully tested but appears to be functional O2 hose is included Overall with Bax OF hysical condition is average Consumables P a 447 Respironics Unit powers up with it successfully passing through its initial self test Oxypleth successfully Units function buttons and display are in working order Pulse Unable to test unit as unit does not include any leads
23. Endoscopes Condition Report Sale 1339 Date 11 03 2014 Condition Reports for other items are at the end of the Scopes report These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Condition 105 OLYMPUS e Insertion tube Good condition LF TP e Bending mechanism amp Angulation control Good movement and no locks e Light guide cable amp Light transmission No light guide cable good light Scope transmission e Objective lens and focusing Good optics Remote control switches N a Distal end Very good condition Image Quality Satisfactory image 3 fibres missing Working Channel s Clear Leak test Slight leak Over all condition Satisfactory overall condition 3 fibres missing and hand control badly scratched 550x5 60mm 106 OLYMPUS OSF 3 Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good optics Remote control switches N a Distal end Good Image Quality Satisfactory image as fibre is missing Working Channel s Clear Leak test Leaking Over all condition Satisfactory overall condition fibre missing and scope is leaking Scop
24. Lot is in fair physical condition 571 Haag Streit The slit lamp has its bulb assembly missing and therefore unable to test the 900 BM light source output The fixation lamp is functional The optics is good and Ambulift produces a good output image The magnifications included are x1 and x1 6 which are in good working order Lot includes two x10 and x16 eyepieces The overall physical condition is satisfactory 574 Jostra Heater The unit powers up well its display and function buttons are in good Cooler Unit working order The unit however has not been fully tested Overall physical HCU 30 600 condition is reasonable 3499 575 Jostra The unit powers up well its display and function buttons are in good Stainless Steel working order The unit has not been fully tested Overall physical condition Heater is reasonable The version is V1 1 and the working hours is 8090Hrs Cooler Unit 576 Jostra The unit powers up well its display and function buttons are in good Stainless Steel working order The unit has not been fully tested Overall physical condition Heater is reasonable The version is V1 0 and the working hours is 1807Hrs Cooler Unit 577 Jostra The unit powers up well its display and function buttons are in good Stainless Steel working order The unit has not been fully tested Overall physical condition Heater is reasonable The version is V1 1 and the working hours is 2898Hrs Cooler Unit 578 Stockert S3 The he
25. able 492 Fisher and Unit powers up with function buttons and indication lights in working order Paykel In manual mode unit is able to heat up well with its heater control in Neopuft working order Pre warm also showed to be functional Unit does not include ee a a skin temperature probe Unit s timer and overhead light are both in working order Unit includes a mattress and an O2 hose Unit is missing its plastic foot board and one of the plastic cot sides is damaged Unit is in satisfactory physical condition 497 Philips Six of the defibrillators power up well and passed its self test Its display and Heartstream function buttons are in good working order The defibrillators have not been Defibs x 6 fully tested but appear to be in good working order Seven batteries are included but one of the batteries is low Overall physical condition is reasonable 501 Datex The patient monitor powers up well and passed its initial self test The Ohmeda S 5 display and function buttons appear to be in good working order The Patient monitor displays SRAM error The M NESTPR module is included but the Monitor parameters SpO2 ECG and NiBP have some physical damage to them on the module The monitor has not been fully tested but the NiBP pump appears to be functional Includes Philips ECG lead NiBP Hose and cuff and SpO2 probe all of which are not compatible with this monitor Overall physical condition is average 505 Eschmann The autoc
26. able to produce a strong light output and Camera with its brightness adjustment functional The VCR and the medilive primo Head with are missing their correct mains cables Unit is in fair physical condition VCR and Zeiss Medilive Camera Control Unit 1015 Drager The Oxylog 2000 powers up ok with medical air and mains battery supply Oxylog 2000 The self test is completed successfully All function keys and LCD are in HG 116 v2 portable good working order All parameters setting and alarms are in good working ventilator order Test run using test lung and all the set parameters are working ok with the MV was verified with the set Vt amp bpm values were within limits Ventilator modes SIMV CPAP IPPV amp SIPPV The oxygen hose and battery charger are included The systems are in good physical condition Battery in good condition 1020 Aesculap Unit powers up with its display functional and its function buttons showing GN60 High to respond Unit has not been fully tested Unit does not include any Frequency accessories Unit s footswitch appears to be responding Unit is in fair Surgery Unit physical condition 115479 with Foot Switch 1028 Drager Vaporiser is physically intact with the vaporiser being in fair physical Sevorane condition Sevoflurane Vapouriser 1034 Schiller Fred None of the five units include any batteries so units are unable to power up Defibs x 5 or to be tested
27. art and lung bypass system does not power up and therefore unable to Heart Lung carry out any further tests Overall physical condition is average due to Bypass usage System 3583 579 Stockert S3 The heart and lung bypass system does not power up and therefore unable to Heart Lung carry out any further tests Overall physical condition is average due to Bypass usage System 3582 585 Drager The Oxylog 1000 powers up ok with medical air supply All function Oxylog 1000 switches are in good working order All parameters setting and alarms are in HG 116 v2 portable good working order Test run using test lung and all the set parameters are ventilator working ok Oxygen hose is not included Battery for audible alarm is in good condition The systems are in good physical condition 594 Drager The three Oxylog 2000 powers up ok with medical air and mains battery Oxylog 2000 supply The self test is completed successfully All function keys and LCD portable are in good working order All parameters setting and alarms are in good ventilator x3 working order Test run using test lung and all the set parameters are working ok with the MV was verified with the set Vt amp bpm values were within limits Ventilator modes SIMV CPAP amp IPPV one with SIPPV Only two oxygen hoses are included and one battery charger with wire exposed The systems are in fair physical condition Battery in good
28. bulbs which turn s on but flickers and the other bulb does not turn on 658 ALM Electric The operating table does not power up doesn t react to the hand held remote Operating function buttons and the battery is not functional Therefore unable to carry Table with out any further checks Its upholstery is intact and in average physical Ga condition Incudes charger handheld remote head and leg rest Overall physical condition is average due to the cover being loose 667 ERBE The ERBE footswitch has not been tested but appears to be intact Overall Diathermy HG 116 v2 Foot Switch physical condition is reasonable 668 Philips Probe head cable and connector are intact and in good physical condition Ultrasound Good image scan and function in CW mode Good working order Probe S12 669 Aloka UST The convex probe has not been fully tested but is physically intact and its 9123 Convex overall physical condition is good Frequency range is 1 15 and its model is Probe UST 9123 679 Olympus TI All units power up Lot does not include a camera head so unable to test the 1900 Imaging processor Units light source is able to produce a strong light output with its Trolley With brightness control functional Unit s spare lamp is also functional The See insufflator has not been fully tested but shows to have its function buttons Monitor responding and its displays functional Both the colour video pr
29. d and therefore unable to fully test Includes footswitch Plus with Foot Overall physical condition is average due to usage Switch 537 Marquette The operating table powers up well The operating table appears to be intact Electric and the upholstery is in average condition due to usage All its articulating Operating movements are in good working order The battery and the brakes are in Table With Pads 00362 good working order Includes head rest leg rests and a remote Overall physical condition is average 538 Penlon SP The anaesthetic trolley powers up well its display and function buttons are Anaesthetic in good working order The anaesthetic machine has not been fully tested Machine With with medical gas but appears to be in good working order Gas failure alarm Penlon O2 Monitor 9605 functional Includes hoses O2 and N20 Scavenging receiver and vacuum regulator with hose Overall physical condition is average 539 Penlon SP The anaesthetic trolley powers up well its display and function buttons are Anaesthetic in good working order The anaesthetic machine has not been fully tested Machine With with medical gas but appears to be in good working order Gas failure alarm Pe G 599 is functional Includes hoses O2 and N20 Scavenging receiver and eas vacuum regulator with hose Overall physical condition is average 541 Stille Scandia The operating table powers up well its upholstery are intact and in good RC Electric phys
30. d printed Unit is in good physical condition Back up battery is in good condition 405 Curapuls 419 The shortwave diathermy unit powers up well its display and function Therapy Unit buttons appear to be functional The output has not been verified but appears to be functional Includes x5 capacitor heads Overall physical condition is average 418 Hitachi EUB Unit powers up and self test completed ok The display keyboard and 525 control console function keys are in good working order The unit comes Diagnostic with three probes convex 3 5MHz Linear array large 7 55MHz amp Linear saa gt array small 1OMHz all in good working condition with the transducer scan probes image clear and visible The convex probe has a small tear at one side of the membrane but working fine The function modes PW Colour M mode and B mode are working ok The track ball caliper measurement amp gain control are all functional The connected B amp W printer is fully functional A spare B amp W printer is also included Also included is an AGFA LINX PAXPORT PACS unit System is in fair physical condition Year 1998 418 Hitachi EUB Unit powers up with units function buttons trackball and display in working 525 order Unit showed to be functional with modes B M color and PW Ultrasound functional Unit was tested using a phantom and all readings taken showed to See ae be within set limits However probe EUP L33 has small areas of damage to 10MHz ae d
31. duced a good output image The overall physical condition is reasonable Ultrasound Transducer 163 Hysteroscope The three hysteroscopes are physically intact but have not been fully tested CIS Sheath with 3 x Taps Continuous Flow Type Hysteroscope 10mm Sheath x 2 with 3 x Taps No Scopes Inc 164 Agilent Probe head cable and connector are intact and in good physical condition Technologies The angulation control of flexible end is working fine Good image scan Transducer with some interference on the scan image in CW mode T6210 21369A 183 Philips Linear Fair overall physical condition but not tested array probe L7 4 184 Toshiba Probe is in an average physical condition with some damage to the head Convex Probe membrane and discoloration Not tested PVT 375BT 3 5MHz 189 Karl Storz The instruments have not been fully tested but appear to be physically intact Telescope Protection HG 116 v2 Dies geanes 210 Philips Five of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included but one of the batteries is low Overall physical condition is reasonable 211 Philips Five of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in
32. e 600x12 60mm 107 OLYMPUS e Insertion tube Satisfactory condition GIF Q30 e Bending mechanism amp Angulation control Good movement and locks e Light guide cable amp Light transmission Good Scope e Objective lens and focusing Good optics e Remote control switches N A e Distal end Good condition e Image Quality Satisfactory image as 4 fibres are missing 1000x12 0mm Working Channel s Clear e Leak test Pass e Over all condition Satisfactory overall condition as there are 4 fibres missing 108 OLYMPUS e Insertion tube Satisfactory condition as slight dent 200mm up insertion GIF Q30 tube e Bending mechanism amp Angulation control Good movement locks Scope e Light guide cable amp Light transmission Good e Objective lens and focusing Good optics e Remote control switches N A e Distal end Good condition HG 116 v2 e Image Quality Satisfactory image as 3 fibres missing 1000x11 90mm e Working Channel s Clear e Leak test Leaking e Over all condition Overall scope is in Satisfactory physical condition with a slight dent to insertion tube and 3 fibres missing 109 OLYMPUS e Insertion tube Good condition CYF 3 e Bending mechanism amp Angulation control Good movement and lock e Light guide cable amp Light transmission Good Scope e Objective lens and focusing Good optics e Remote control s
33. e However the unit is not starting up since HG 116 v2 the acquisition module and patient ECG leads is not included The unit is in good physical condition but not fully tested Software CD ROM is included 464 Wisap HLX Unit powers up with its function button and indication lights in working 250 Endo Hi order Unit showed to be able to produce a good light output Unit is in Lite Light satisfactory physical condition Source 480 Philips Five of the defibrillators power up well and passed its self test Its display Heartstream and function buttons are in good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included but one of the batteries is low Overall physical condition is reasonable 481 Philips Four of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not Defib x 4 been fully tested but appear to be in good working order Three batteries are included Overall physical condition is reasonable 487 Endomed 582 The Endomed 582 is not powering up The Sonopuls 492 powers up well the Therapy Unit output is not been verified but appears to be in good working order The Sonopuls 492 probe is functional but the cable has some physical damage to it Overall Therapy Unit f RE physical condition is reason
34. ead and the cable sheath Probe Serial has also been damaged Overall the unit is in satisfactory condition aie i 67 System version 8 0 797 HP The defibrillator powers up well and passed its initial self test The display Codemaster and function buttons are in good working order The defibrillator however XL Defib does not detect the pads and therefore unable to verify the defibrillator S output readings The ECG is tested using a simulator and produces good strong traces of ECG signals Printer appears to be functional but has not been tested Includes AED therapy cable and two ECG cables Overall physical condition is reasonable 809 Hitachi EUB The system powers up but the display screen does not display any 405 Sumi information CRT powers up and the indicators on the control panel also Portable light up Unable to carry out any further checks The printer powers up but ripe the door des not close Overall the unit is in satisfactory physical condition alg ob avi a One of the hinge covers is missing and the probe head has signs of wear and Sony Video tear Graphic Printer 815 Hill Rom Air Unit power up ok system self test completed ok The display and function Shield Infant keys are in good working order The set temperatures are maintained Incubator correctly and the oxygen sensor is calibrated at 21 oxygen Both air amp skin temperature mode are functional but the skin temperature probe is not included Humidifier monitor
35. elds Disconnect is displayed but the incubator is verified with a working sensor C2000 and the same alarm appears Therefore there is a fault with the incubator Ba ae itself The height adjustment is in good working order Humidifier tray is not included Oxygen gas hose is included Overall physical condition is average version 2 11 866 Maico MV10 All units are able to power up with only six out of the seven units being able Otoscopes x7 to produce a light output All units are complete with a hand piece Lot is in with Qty of good physical condition Accessories 867 Ultramedic Infusion device analyser does not power up and therefore unable to carry out IDA 2 Plus any further checks Overall physical condition is poor as there is a crack on Infusion the casing which has been glued together Device Analyser OP5835 872 Sonopuls 490 Both the Sonopuls units power up well its display functions and function Therapy buttons are in good working order The therapeutic ultrasound is functional Units x 2 with but the output has not been verified However one of the probes S N Prones 24450 is not functional but the three remaining probes are functional Overall physical condition is reasonable 875 GE Treadmill Lot is in good condition unable to fully check since the stress ECG machine T2100 is not included Comes with Tango NIBP unit ECG leads CAM USBA T and power adapters amp cables 880 X Ray Power up ok and lamp working fine viewi
36. fully tested Unit does not include any hand Machine pieces Units modes are selectable but have not been tested Units IV pole Operators height adjustment is functional Units buttons audio voice selection is Manual and functional The footswitch is responding but unable to verify if it is fully Remote functional Unit is in satisfactory physical condition Control Version V3 00 Manufacture date 02 01 258 Stack Trolley All units are powering up ok on the stack trolley Processor with good with Stortz quality image when camera head is connected The light source xenon lamp Tricam SL is working fine Good image produced on the colour monitor The Sony PAL 202220 DVCAM and Digital Still Picture Recorder are working fine Lot is in good 20 Camera physical condition Control Unit Sony DVCam Digital Video Cassette Recorder Stortz Xenon Nova Light Source Sony Digital Still Recorder Stortz Tricam 20221030 PAL Camera Head and Sony Monitor HG 116 v2 259 Olympus Unit powers up Units mechanical movements and locks are in working OB300 order Unit is able to produce a strong light output with its brightness control Colposcope functional Units optical system is able to produce a good image with unit also showing to have good focusing and magnification Unit is in satisfactory physical condition 1x set of eyepieces 12 5 302 Hill Rom The bed power up ok and control panels on both side of
37. g order The overall physical condition is satisfactory but the control panel is loose where its screws are missing to hold the control panel in place 92 Spacelabs All 4 monitors power up ok Display panels and touch screen interface are in 90369 Patient HG 116 v2 Monitor x4 good working order the touch screen response on one of the monitors is with Box of slow The module is in place on each monitor and the ECG SpO2 and Assorted NIBP parameters are functional tested with patient simulator Overall the aa units are in reasonable physical condition Includes a box of accessories which contains IEC leads mains adaptors ECG lead sets and SpO2 probes 156 Philips Probe head cable and connector are intact with slight discoloring of the Ultrasound membrane Scan image is with good resolution and grey scale Good Probe L9 3 working order 157 Philips TV Probe and connector intact the cable is slightly discolored Unable to test the probe C8 4V probe Unsupported probe displayed when connected to a test unit 159 Hitachi Good overall physical condition but not tested Convex Probe EUP C715 5 1 MHz 159 Hitachi EUP The probe is tested using a compatible ultrasound with a phantom and C715 produced a good output image The overall physical condition is reasonable Ultrasound Transducer 160 Hitachi EUP The probe is tested using a compatible ultrasound with a phantom and V53W pro
38. good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included Overall physical condition is reasonable 223 Airox Both the units have not been fully tested but appear to be intact Legendair Ventilators x 2 with Airox Open Pack x 1 228 Richard Wolf All units are powering up ok with function buttons in working order The Stack Trolley processor and both of the camera heads were able to produce a good image with Olympus on the Olympus monitor with both camera head showing to have good OEVIOIH focusing One of th heads is sensitive to the light which at times can Monitor Wolf ocusing ne o aS comers eads is sensitive e g c 1 s ca 5132 Light affect the image giving a glare around the obtained image The light source Source Wolf Was able to produce a strong light output Lot is in good physical condition 5520 Camera Endocam and 2 x Camera Heads 229 Datex Ohmea The anaesthesia trolley power up ok and the medical gas pressure gauges and Aestiva 5 with flowmeters are in good working order The basel flow and O2 amp N20 Smart Vent proportional link 25 system is functional The ventilator 7100 power up 7100 and the start test completed successfully All function keys and LCD are in good working order The absorber bellows and ventilator are working fine in both volume and pressure control modes with alarm indicating Insp reverse Flow Test run using test l
39. ic and the tube loading cover does not open on one of the pumps The batteries Pumps as appear to be flat otherwise the units are in satisfactory physical condition Y O M 2002 2005 726 Draper AC Unit powers on with its function buttons and display in working order Unit 9000 Air shows to be functioning but has not been fully tested to verify Unit is in fair Conditioner physical condition 728 Penlon Prima The anaesthetic trolley with ventilator powers up well its display and SP function buttons are in good working order The anaesthetic machine has not Anaesthetic been fully tested with medical gas but appears to be functional Includes Machine yen hoses O2 Air and N2O Scavenging receiver and vacuum regulator with AV900 Ventilator hose Overall physical condition is average and Datex Ohmeda S 5 Monitor 6183 736 Dinamap Pro The one of the vital signs monitor powers up and the other monitor does not 1000 x 1 and Pro 400 x 1 with Leads and Finger Clips power up The monitor that powers up has not been fully tested but appears to be functional NiBP pump is in working order but has not been fully tested The printer appears to be functional The lot includes two NiBP hoses ECG leads SPO2 probe and two temperature probes Overall physical HG 116 v2 condition is average as the rotary knobs are missing 737 ALM The operating table power up well its upholstery is intact and appears
40. ical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Aabe Description Report 2 Stack Trolley All units power up The processor and both if the camera heads are able to With produce a satisfactory image on the Panasonic monitor with both camera Panasonic heads showing to have satisfactory focusing The white set function is D temperamental to function which does affect the colour of the image The HD900 light source is able to produce a good light output with its brightness Camera adjustment in working functional Bulb hours at 68 The Dyonics Power is Control Unit powering up but has not been tested The unit does not include any with 2x accessories Dyonics power version v3 0 The colour video printer is Camera functional Overall lot is in satisfactory physical condition Heads 300 XL Xenon Light Source Dyonics Power Unit and Sony UP 21MD Colour Video Printer 4 Datex Power up and system self test completed successfully Display panel Ohmeda controls and trim knobs are all in good working order All ventilation modes Anaesthetic VCV SIMV PCV are selectable and appear to be functional but has not Machine with been fully tested The patient monitor does not power up therefore no AS 3 Patient r te Monitoring further checks could be carried out the module rack is in place but no
41. ical condition but on two cushions there is some physical damage to a them All the articulating movements are functional but there is not enough charge to test fully The manual trendelenburg is in good working order No AC adaptor is included but handheld remote is incuded Overall physical condition is average Stille Reison AB make Model Scandia RC HG 116 v2 543 Philips The mobile X Ray powers up but does not go into main menu and stays on Practix 2000 displayed message line cord connected The kV and mAs are not Mobile X Ray selectable Therefore unable to carry out any further checks Includes a key with Key Overall physical condition is average due to usage Y O M is 1997 and software version is 3 0 5 AEC 545 Schiller AT The ECG recorder does not power up and therefore unable to carry out any 102 ECG further checks AC power indicator turns on but the recorder does not switch pane with on ECG leads are included Overall physical condition is reasonable 567 Enraf Nonius The unit is not powering up and therefore unable to carry out any further Sonoplus 591 tests Includes two transducers Overall physical condition is average Therapy Unit 570 Eschmann Both units are powering up however neither unit appear to be able to VP45 Suction produce any suction as one the units regulator knob is broken off and the Pump x 2 other units regulator knob is not functional
42. inter and the Olympus VCR are powering up Overall lot is in fair physical condition UHI 2 Insufflator Olympus CLV S30 Light Source Olympus OTV SX2 Processor Sony DV Cam Recorder and Sony Colour Video Printer 681 Toshiba Unit powers up with function buttons trackball and display in working Powervision order Unit shows to be functional with modes B M PW color and angio 6000 functional Unit was tested using a phantom and all readings taken showed to Saas be within set limits However probe PVM 375AT has damage to the crystals Probe to one side of the scan head which produces bad drop out to one side and PVM375AT bad interference which all does affect the overall image Both probes show to 3 75MHz have satisfactory resolution and satisfactory greyscale Probe PVM 375AT Probe also has one of its rubber grommets damaged and loose The printer and Damaged VCR power up Overall unit is in fair physical condition Lead Mitsubishi Version V4 00ERO01 P91 Printer and Sony Manufacture date December 2000 Video Cassette Recorder 682 Drager The anaesthetic machine with ventilator powers up well and passed its initial Fabius GS self test The display and function buttons are in good working order The Anaesthetic anaesthetic machine has not been fully tested with medical gas but appears to Machine be functional The bellows and absorber assembly is included which are HG 116 v2 intact Its system check was comp
43. ith some damage at corners of cushions Unit comes with both of its cot sides Overall unit is in fair physical condition 989 Hillrom The birthing bed powers up well All the articulating movements are in good Affinity 3 working order Both the sides control panels are functional The manual Birthing Bed trendelenburg is functional along with the brakes The mattresses are intact physically and include a lower section Overall physical condition is reasonable HG 116 v2 990 Hillrom The birthing bed powers up well All the articulating movements are in good Affinity 3 working order Both the sides control panels are functional The manual Birthing Bed trendelenburg is functional along with the brakes The mattresses are intact physically and include a lower section Overall physical condition is reasonable 991 Hillrom The birthing bed powers up well All the articulating movements are in good Affinity 3 working order Both the sides control panels are functional The manual Birthing Bed trendelenburg is functional along with the brakes The mattresses are intact physically and include a lower section Overall physical condition is reasonable 998 Marquette Unable to determine if unit is powering up as unit does not come with ithe Electric correct remote controller The roller wheels and brake are functional Unit is Operating complete with a set of cushions which are in poor condition with rip
44. its O2 and air hoses Ventilator 1145 Drager CF800 Unit has not been tested Unit is in fair physical condition an looks to be Mixer intact with a small portion of the cover missing at one corner and some cracks at the back of cover Unit includes O2 and air hoses 1155 Hitachi EUB Unit is able to power up with units function buttons and display in working 525 order Unit has not been fully tested but unit does show to be functional Ultrasound Units modes B M color and PW show to be functional The EUP V53W reg probe is in fair physical condition Overall unit is in fair physical condition Probe 1164 Datex All five units power up however the datex 3800 does not show to be Normocap displaying any information on its display The rest of the units displays show 200 Patient to be functional with their function buttons responding Units have not been ee fully tested The datex 3800 has its power button cover missing and the Micromon Picco plus has its printer missing Lot includes a various selection of 7141 ECG accessories but the units are not complete with accessories Lot is in fair Monitor physical condition Datex 3800 Pulse Oximeter With Picco Plus ECG Monitor and Satelite Plus 1170 Siemens Unit is able to power up Units function buttons trackball and display are in Acuson working order Function buttons for M and C modes are missing their button Antares covers Unable to fully test unit as unit does not include any p
45. lave power up ok displayed error cleared Display and function little Sister keys are working fine Only cycle 134 deg C wrapped instrument cycle is Quick Vac selectable The cycle was tested but failed to complete error 30 vacuum fail HG 116 v2 Water level and printer are working fine Unit has no power cable or trays System is in good physical condition 511 Atlantech The processor powers up giving a constant beep with its indication lights Stack Trolley coming on before them then turning off with the processor showing to be with A1 Video unresponsive Unable to determine if the processor is functional as unit does pre not include a camera head The light source is able to power up with its Source and function button and display in working order Unit is able to produce a Light Guide strong light output from both the halogen and xenon light outputs with both Adaptor their brightness adjustments functional Xenon time 289h Unit is in fair physical condition 514 Neoblue LED Only one unit is powering up The unit which is powering up shows to be Phototherapy functioning but has not been fully tested with both its high and low Lights x 2 functions showing to be able to be selected Both units are in fair physical condition Operation hours 3746 8h for the working unit and 5234 5h 521 Conmed The Hyfrecator powers up well and its display is in good working order No Hyfrecator handpice is include
46. leted successfully A set of medical gas hoses are included Overall physical condition is average S W Version 1 39 1 703 Hill Rom Air Unit powers up with its function buttons and displays in working order In Shields both pre warm and manual modes unit is able to heat up well with its Radiant heater control in working order Unable to test baby mode as unit does not Maan Ne include a skin temperature probe Unit s timer and its over head light both show to be functional Unit includes a mattress and O2 and air hoses Unit is in fair physical condition with its heater cover being slightly bent 704 Eschmann All units articulating movements show to be in working order Unit s Hydraulic hydraulic height adjustment is functional Unit s roller wheels and brake are Operating also functional Unit includes a full set of cushions with the head rest are with cushion having a large rip to its cover and some pieces of its inside padding as missing The rest of the cushion are in fair physical condition Overall lot is in fair physical condition 710 Drager Both the vaporizers have not been fully tested but are in average physical Vapour 19 3 condition due to usage The dial appears to be functional Isoflurane Vapourisers x2 717 Baxter 4 of 5 units power up ok Display panels controls and the pump modules are Colleague in good working order not fully tested One of the units does not power up Volumetr
47. mage with the camera head also showing to have satisfactory focusing The light source is missing its main bulb However its spare lamp was functional with its brightness adjustment functional Lot is in satisfactory physical condition 50 Lifepak 20 The defibrillator powers up well and passed its initial self test The display Defib with functions and function buttons are in good working order The defibrillator is ECG Leads tested using defibrillator analyser and output energies produced are within No Eadddles the limits Both the manual and AED mode are in good working order ECG output is tested using a simulator and produces good strong traces of ECG signals Includes ECG cable and AED therapy cable Printer is in good working order Overall physical condition is satisfactory Y O M is 2002 51 Lifepak 20 The defibrillator powers up well and passed its initial self test The display Defib with functions and function buttons are in good working order The service ECG Leads indicator light stays on and the sound is not detected The defibrillator is No Padddles tested using defibrillator analyser and output energies produced are within the limits Both the manual and AED mode are in good working order ECG output is tested using a simulator and produces good strong traces of ECG signals Includes ECG cable and AED therapy cable Printer is in good HG 116 v2 working order Overall physical condition is satisfactory Y O
48. ng box double 894 BIPAP Vision The BiPAP powers up well and passed its initial self test The unit has not Ventilatory been fully tested but appears to be functional O2 hose is included Overall Support physical condition is reasonable Operating hours is 26427 System 895 Stockert The heater cooler unit powers up but has not been fully tested The computer Shiley perfusion unit also shows to be powering up with its displays functional and pened its function buttons looking to be responding All four of the pump modules HG 116 v2 Perfusion are showing to be power up All of the pumps appear to be able to rotate but System with have not been fully tested Unit has not been fully tested Overall lot is in Cooler reasonable physical condition 896 Stockert The heater cooler unit powers up but has not been tested The computer Shiley aided perfusion unit also shows to be powering up with its displays Computer functional and its function buttons looking to be responding One of the ae pump modules constantly alarms and displays error message ERROR Sistem Kith PUMP B The other three pump modules show to be selectable displaying Cooler their operation hours 11573h 13099h 11994h Unit has not been fully Newer tested Lot is in fair physical condition Model No Option 953 Hill Rom Unit not tested In fair physical condition with only one mattress No remote Birthing Bed control x2 Avanguard
49. ng order In air mode unit shows to be able to heat up well and maintain its set temperature Temperature was verified using an external thermometer and temperature showed to be within limits Unable to test skin mode as unit does not include any skin temperature probes Oxygen and humidifier modes have also not been tested Units electrical height adjustment and mechanical cot tilt are both functional Unit includes an O2 hose Unit is in satisfactory physical condition Version 2 18 338 Scimed Unit is powering up with unit being able to pass through its initial power up Logidop 3 test Units function buttons and display are in working order In air mode with 2MHz unit shows to be able to heat up well and maintain its set temperature Doppler and Temperature was verified using an external thermometer and temperature Foot Switch showed to be within limits Unable to test skin mode as unit does not include any skin temperature probes Oxygen and humidifier modes have also not been tested Units electrical height adjustment and mechanical cot tilt are both functional Unit includes a gel mattress and an O2 hose Unit is in satisfactory physical condition Version 2 19 366 Bair Hugger The three patient warmers powers up well and its display indicator are in Patient good working order However one of the heaters is not functional and the Warmers x3 temperature range is not selectable The other two patient warmers are functional and the different tempera
50. o verify the bipolar output as there is no footswitch The lot includes monopolar cut and coagulation footswitch however bipolar footswitch is not included Overall physical condition is satisfactory 240 Olympus TI 1900 Imaging Trolley with Dyonics 300XL Light Source Smith and Nephew All units are powering up The processor and camera head were tested on a monitor no monitor in lot The processor and camera head were able to produce a satisfactory image with the camera head also showing to have satisfactory focusing Slight interference can be made from to much movement to the connection end of the camera head The light source is able to produce a strong light output with its brightness adjustment functional HG 116 v2 360 3CCD The insufflator powers up with its displays functional and its function Camera buttons showing to respond unit has not been fully tested Overall lot is in Control satisfactory physical condition Camera Head and Smith amp Nephew 400 LPM Insufflator 241 Drager Unit power up with function buttons and displays in working order In Babytherm manual mode only one of the two heaters is able to heat up with the second 8004 114074 heater non functional The radiant heater adjustment functioning Unable to test skin mode as unit does not include a skin temperature probe Units spot light is functional Units electrical height adjustment is functional Unit
51. or to be tested Lot includes 1x battery charger unit unable to determine if with Battery the charger unit is powering up as there is no indication light All units are in Charger x1 reasonable physical condition with units having lots of scratching to the displays and covers 1091 Hill Rom The incubator powers up well and passed its initial self test The display Airshields functions and function buttons are in good working order The set air C2000 temperature is reached and maintained The temperature is verified using a ee thermometer which is within the limits The skin temperature is not tested as there is no skin probe The height adjustment and the scale are in good working order The mattress is intact The humidifier tray is not included Oxygen hose is included Overall physical condition is satisfactory 1092 Hill Rom The incubator powers up well and passed its initial self test The display Airshields functions and function buttons are in good working order The set air C2000 temperature is reached and maintained The temperature is verified using a ear thermometer which is within the limits The skin temperature is not tested as there is no skin probe The height adjustment and the scale are in good working order The mattress is intact The humidifier tray is not included Oxygen hose is included Overall physical condition is satisfactory 1093 Hill Rom The incubator powers up well and passed its initial self test The display
52. p with units function buttons and display in working order MAC 5000 Unable to test unit as unit does not include any ECG leads Unit s battery is ECG Machine functional but needs charging Unit is in fair physical condition No Leads 46232 HG 116 v2 1063 Drager Baby Unit is powering up with unit showing to be able to start to heat up but test Therm 4200 not fully carried out Unit has not been fully tested The mechanical cot tilt is Incubator in working order Unit includes a mattress Unit is reasonable physical condition 1076 Echmann J3 All units articulating movements are in working order However the table is Hydraulic at a constant slight forward tilt Unit s hydraulic height adjustment is Operating functional Unit s roller wheels and brake are functional Unit includes a full aie 2078 set of cushions which are in fair physical condition Overall unit is in fair physical condition 1084 Schiller Fred None of the four units include any batteries so units are unable to power up Defibs x 4 or to be tested Lot includes 1x battery charger unit unable to determine if with Battery the charger unit is powering up as there is no indication light All units are in Charger x1 reasonable physical condition with units having lots of scratching to the displays and covers 1085 Schiller Fred None of the four units include any batteries so units are unable to power up Defibs x 4
53. r and mains battery supply Oxylog 2000 The self test is completed successfully All function keys and LCD are in portable good working order All parameters setting and alarms are in good working ventilator x2 order Test run using test lung and all the set parameters are working ok with the MV was verified with the set Vt amp bpm values were within limits Ventilator modes SIMV CPAP amp IPPV one with SIPPV The oxygen hoses are not included but battery chargers are included The systems are in fair physical condition with scuffs and scratch on casing consistent with use Battery in good condition 14 Alcon Both units power up and the initial probe test are displayed Only one probe AcuScan and a footswitch are included The probe test is completed on one unit only Ophthalmic The display and function keys are working fine on that unit The printers on Scanner both units are not working The second unit does not proceed with the probe test and the soft keys are not responding when depressed The system is in fair physical condition Year 1999 amp 1992 Software version 3 02 29 05 98 15 Philips Five of the defibrillators power up well and passed its self test Its display HG 116 v2 Heartstart and function buttons are in good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included Overall physical condi
54. robes Also Ultrasound unable to verify any functional modes Unit is in fair physical condition HG 116 v2 Software 4 0 105A Product version 3 5 Manufacture date 2005 1173 Respironics Unit is able to pass through its initial self test successfully Unit is missing BiPAP Vision two lines of pixels across the top of units display units display is still 129191 functional Unit function buttons are in working order Unit shows to be functioning but unit has not been fully tested Unit is missing its O2 hose Unit is in fair physical condition Operation hours 2871h 1174 Respironics Unit is able to pass through its initial self test successfully Units function BiPAP Vision buttons and displays are in working order Unit shows to be functioning but 129203 unit has not been fully tested Unit is missing its O2 hose Unit is in fair physical condition Operation hours 3521h HG 116 v2
55. s battery is not functional with unit displaying a constant error message battery discharged Unit is in satisfactory physical condition 244 Olympus Ocs_ Unit powers up with unit able to produce a light output from both A and B Colposcope lamp outputs The light outputs glass prism has a large chip to it the light 104042 guide cable is missing all of its outer insulation and the brightness control is loose resulting in the brightness adjustment not being fully functional Unit s optical system is able to produce a satisfactory image with satisfactory focusing There are dirt particles from inside of the colposcope visible on the image The magnification is functional but is stiff to rotate Unit is in fair physical condition 246 Alcon Legacy Unit is powering up with its function buttons and touch screen display in Everest Phaco working order the bottle height LED segmental display has the bottom half Machine of the LED segments not functioning Unit has not been fully tested Unit does not include any hand pieces Units modes are selectable but have not been tested Units IV pole height adjustment is functional Units button audio voice selection is functional The footswitch is responding but unable to verify if it is fully functional Unit is in satisfactory physical condition Version V3 25 Manufacture date 02 02 247 Alcon Legacy Unit is powering up with its function buttons and touch screen display in Everest Phaco
56. s able to produce a good clear Trolley image units magnification settings 1x and 1 6x both producing a good image Unit is also showing to focus well The chin rest is in place with its chin rest height adjustment in working order The table s electric height adjustment is functional Unit is in good physical condition 1 Zeiss OPMI1 Unit powers up through its external light source power supply with the main H Operating units stand not power in on Units mechanical movements show to be in Microscope working order Unit is able to produce a strong light output with its brightness adjustment in working order Units optical system is able to produce a good image however there are visible dirt particles from inside of the microscope visible on the image Unit showed to have good focusing and magnification Unit is in reasonable physical condition Eyepieces 12 5x Objective lens 300 82 Hitachi EUB The ultrasound powers up well and passed its initial self test The display 6500 functions and function buttons are in good working order No probes are Ultrasound included with the ultrasound but tested using another compatible probe for on t testing The ultrasound is tested using a phantom and set caliper Brinter N measurements taken are within the limits The grey scale and resolution is Probes good The ultrasound modes included are B Mode Colour M Mode and PW However the CW mode is not activated The printer is in good workin
57. s and sane damage to them Overall lot is in fair physical condition Unit includes to spare head rests 1007 Atmos Atmos Atmos Systema 2 S N KF27382 01 Unit power up and the suction pump Type System is working fine The light source is missing both its halogen bulbs There are 2 ENT Station barely any attachment or accessories with the unit The structure is in fair physical condition 1010 Marquette Unable to determine if unit is powering up as unit does not come with the Electric correct remote control All of the mechanical movements are in working Operating order The roller wheels and brake are functional Unit includes a complete Table set of cushions Cushions are in reasonable physical condition with them having tape covering areas of the cushions Overall lot is in fair physical condition 1011 Stack Trolley The camera head and the processor are both powering up The processor and With Sony camera head are then able to produce a satisfactory image on the Sony Monitor monitor with the camera head also showing to have satisfactory focusing at DX The monitor produces a faded strip across the top of its display but this does Pal Camera not affect the overall image The camera heads connection chip is badly Control Unit Worn away at the edges with some of its copper plates also being worn and Xenon Nova damaged However this does not appear to have any effect on the image Light Source which is produced The light source is
58. ssor has not been fully tested but appears to be functional The printer is in good 2232 working order Overall physical condition is satisfactory Light source Insufflator working hours is 212hrs 5131 Light Source Sony Video Printer and Wolf 5506 Camera Head 791 Atlantech All units power up ok Processor and camera head produce a good quality Stack System image on colour monitor Comes with 2 camera heads but one of the camera with heads is not working The light source is in good working order xenon lamp Atlantech hours 211 and the printer is also functional Overall the units are in good Monitor AL1 a P 8 Light Source physical condition A1 Video Processor Sony Video Printer Atlantech Camera Head and M25 Camera Head 793 Welch Allyn Powers up ok and the display panel and controls are functional ECG CP200 ECG monitoring functions and the acquisition module are in good working order HG 116 v2 Recorder The printer appears to be working but could not be fully tested out of With ECG paper Overall the unit and the acquisition module are in good physical and Leads cosmetic condition 794 Aloka SSD Powers up ok and the display panel controls are in good working order Both 500 B and M modes are functional and the system produces clear images The Ultrasound probe is working fine but a small section of the crystals appear to be eee damaged on one side and the casing of the scan h
59. st completed Portable successfully The LCD screen and control console s function keys are in Ultrasound good working order Both transducers linear array L38 10 5MHz amp Convex Unit with C15 4 2MHz display good quality images All modes 2D M mode Colour transducer x2 amp Doppler are working fine System is in good working order The unit does not include mains adapter charger Unit is in good physical condition Year 11 2004 software Boot amp ARM version 21 80 203 019 DSP version 20 80 202 73 8 Datex AS 3 One of two monitors power up and complete the system self test Patient successfully Display screen controls and trim knob are in good working Monitors x 2 order ECG SpO2 NIBP and CO2 parameters are functional other eee and parameters are not tested One of the monitors does not power up Both Se cable monitors contains a M NESTPR and a M CAiO module Overall the monitors are in reasonable physical condition No accessories included with the lot Y O M 1998 1999 10 Medtronic Both units power up and pass the self tests successfully Display screens and Lifepak 500 controls are in good working order Tested with defibrillator analyser both Defibs x 2 units are functional and the energy delivery is also within the limits Both units are with cases and in reasonable physical condition One of the units has a battery and the battery is working Y O M 1999 2002 11 Drager Both Oxylog 2000 powers up ok with medical ai
60. tics Remote control switches N A Distal end Satisfactory condition Image Quality Good image Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition 122 OLYMPUS GIF XQ20 Scope 1000x10 50mm Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good Remote control switches N A Distal end Very good condition Image Quality Satisfactory as one fibre is missing Working Channel s Clear Leak test Pass Over all condition Satisfactory overall condition one fibre is missing 123 OLYMPUS GIF Q Scope 1000x12 40mm Insertion tube Good condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good light transmission there is a dent to light guide cable Objective lens and focusing Good optics Remote control switches N A Distal end Good condition Image Quality Satisfactory image 1 fibre is missing Working Channel s Clear Leak test No leak tester to test with Over all condition Satisfactory overall condition fibre is missing and there a dent to the light guide cable HG 116 v2 Condition Reports Sale 1339 Date 11 03 2014 These reports are compiled by our Med
61. tion is reasonable 16 Philips Five of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included Overall physical condition is reasonable 17 Philips Five of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not Defibs x 5 been fully tested but appear to be in good working order Five batteries are included Overall physical condition is reasonable 24 Pentax EPK Powers up ok Control panel and indicators are in good working order The 700 xenon lamp is working fine and the air pump is also operational The unit is Lightsource in good physical and cosmetic condition No further checks could be carried out due to a compatible endoscope not being available Y O M 1998 26 Philips M4 All units power up with their function buttons touchpad function buttons Monitors x3 and display in working order Unable to fully test the units an lot does not and Agilent include any leads All of the units NiBP pumps show to be functioning but i g have not been fully tested None of the units batteries are functional Lot is in Monitors with fair physical condition Modules on Stands 28 ATL HDI The ultrasound powers up but does not go into the main menu option
62. to be Tansferis in good physical condition All its articulating movements are in good Electric working order The manual and remote function buttons are functional a ae Includes handheld remote and AC adaptor Brakes are in good working Pads 205 order Overall physical condition is reasonable 759 Zeiss 10x 22B The binoculars are intact but have not been fully tested The optics appears Microscope to be reasonable but not verified Includes two x10 22B eyepieces Overall Eye Piece physical condition is satisfactory Attachment 785 Drager Unit is able to power up The display for the baby mode is not functional Airshields the rest of the displays are functional All function buttons are in working Resuscitaire order Unit is able to heat up in both its pre warm and manual modes Unable to check baby mode as unit does not include a skin temperature probe Units overhead light and timer are both functional Unit includes a mattress Unit is in fair physical condition 788 Wolf Stack The stack system powers up well the unit s functions and function buttons System Inc are in good working order The stack system produces a good output image Sony The processer is in good working order The camera head is intact produced Trinitron i ee i Monitor Wolf a good output image aog its pOCHSING is minena TE light Source produces 5506 Video a good ray of light and its brightness control is functional The insufflator Proce
63. trol Good movement and locks e Light guide cable amp Light transmission Good Videoscope Objective lens and focusing No processor to test scope on e Remote control switches No processor to test scope on e Distal end Satisfactory condition e Image Quality No processor to test scope on e Working Channel s Clear 1000x10 80mm s dekel Leaking e Overall condition Satisfactory physical condition scope is leaking HG 116 v2 unable to test scopes working condition 113 OLYMPUS CYF 4 Scope 350x5 70mm Insertion tube Good physical condition Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Good Objective lens and focusing Good optics Remote control switches N A Distal end Good physical condition Image Quality Good image Working Channel s Clear Leak test Pass the valve is damaged but still functional Over all condition Good Overall condition 114 OLYMPUS CF 1T200L Videoscope 1600x15 10mm Insertion tube Satisfactory physical condition but is stained in colour Bending mechanism amp Angulation control Good movement and locks Light guide cable amp Light transmission Satisfactory transmission as one light output is dim Objective lens and focusing Good optics Remote control switches Fully functional Distal end Satisfactory physical condition
64. ts functional Also unit s lamp Therapy 4000 output is functioning Unit is in fair physical condition Light Unit No Stand 15680 854 Philips Four of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not FR2 Defibs x been fully tested but appear to be in good working order No batteries are 4 included Overall physical condition is reasonable 855 Philips Four of the defibrillators power up well and passed its self test Its display Heartstart and function buttons are in good working order The defibrillators have not FR2 Defibs x been fully tested but appear to be in good working order No batteries are HG 116 v2 4 included Overall physical condition is reasonable 859 Hill Rom The incubator powers up well and passed its initial self test The display Airshields functions and function buttons are in good working order The set air C2000 temperature is reached and maintained The temperature is verified using a Fie thermometer which is within the limits The skin temperature is not tested as there is no skin probe The height adjustment and the scale are in good working order The mattress is intact The humidifier tray is not included Oxygen hose is included Overall physical condition is satisfactory 860 Hill Rom The incubator powers up well but a continuous alarm of Sensor Airshi
65. ture ranges are selectable All three patient warmers have tubing s and overall the patient warmers are physically intact 367 Dinamap Pro Both the Propag Encore and the Pro 1000 monitors power up ok Display 1000 Vital panels and controls are in good working order ECG SpO2 and NIBP Signs Monitor parameters are functional on both monitors but the speaker appears to be not Mier sae working on the Prol000 monitor QRS tone is not present Overall the Paq Encore monitors are in good physical condition Includes ECG lead set and SpO2 Patient probe for Pro 1000 ECG trunk cable SpO2 probe and NIBP hose for Propaq Monitor with Encore and user manual for Propaq Encore Leads 385 Gambro The four dialysis machine powers up well and passed its initial self test The AK95 Dialysis display and function buttons are in good working order The dialysis HG 116 v2 Machines x 4 with 2 x Boxes machine has not been fully tested but appears to be in good working order The dialysis machines includes heparin and air detector modules but have of not been fully tested The dialysis machine all have single blood pump Consumables which are all physically intact The lot includes a number of Hemodialysis tubing set The dialysis machines have water and drain hoses The overall physical condition is satisfactory Y O M is 2000 for all four dialysis machines Programme Version on all dialysis machines is 12 20 388 Medela Unit
66. ung and all the set parameters are working ok The MV was verified with the set Vt amp bpm value within limits The oxygen amp flow sensors are functional A leak was observed with the bellow as the system was tested The system is in fair physical condition with front lid of the rebreathing circuit assembly missing Backup battery in good condition 230 Respironics Unit powers up with function buttons and display in working order Unit BPAP Vision showed to be functioning however the output has not been fully tested Unit Ventilatory includes an O2 hose Unit is in fair physical condition Support System Operation hours 17580h HG 116 v2 233 Siemens Servo ventilator 900C The 900C power up ok and system indicators numerical displays and function knobs are in good working order All parameters setting and alarms are in good working order Test run using test lung and all the set parameters are working ok with the MV was verified with the set Tidal volume Vt amp Frequency breath min value within limits All ventilator modes SIMV CPAP Volume and pressure controlled are working fine The medical gases hose are included A small leak can be heard on the Air Oxygen mixer Oxygen sensor is in good working order The system is in good physical condition Operations counter at 11413hrs 237 Hill Rom Air Shields C2000 Incubator Unit is powering up with unit being able to pass through its initial power up
67. witches N A e Distal end Satisfactory condition e Image Quality Satisfactory image as 4 fibres are missing 350x5 70mm Working Channel s Clear e Leak test Pass e Over all condition Satisfactory overall condition with 4 fibres missing from image 110 PENTAX e Insertion tube Satisfactory condition ES 3840 e Bending mechanism amp Angulation control Good movement and locks e Light guide cable amp Light transmission Good Video scope Objective lens and focusing No processor to test scope on e Remote control switches No processor to test scope on e Distal end Missing distal cap e Image Quality No processor to test scope on 600x13 50mm Working Channel s Clear e Leak test Pass e Over all condition Satisfactory physical condition unable to test scopes working condition 111 OLYMPUS e Insertion tube Satisfactory condition slight nocks to tube ENF P4 e Bending mechanism amp Angulation control Good movement no locks e Light guide cable amp Light transmission Good Scope e Objective lens and focusing Good optics e Remote control switches N A e Distal end Good condition e Image Quality Satisfactory image as one fibre is missing 250x4 05mm Working Channel s N A e Leak test Pass e Over all condition Satisfactory overall condition 112 PENTAX e Insertion tube Satisfactory condition EG 2940 e Bending mechanism amp Angulation con
68. xa minor signs of wear and tear on some of the function keys One of the ra ds SS monitors power up but the LCD remains blank Lot includes 3x ECG lead sets and 5x NIBP hoses 843 ERBE ICC The diathermy powers up well and passed its initial self test The display 200 functions and function buttons are in good working order The diathermy is Diathermy tested using an ESU analyser and its monopolar cut and coag output eey ee d readings where within the range Unable to verify the bipolar readings as re ee ds there is no footswitches included Includes monopolar cut and coag footswitch but bipolar footswitch is not included Overall physical condition is reasonable 844 ERBE ICC The diathermy powers up well and passed its initial self test The display and 200 function buttons are in working order However the patient plate indicator Diathermy alarm stays and does not change Therefore unable to carry out any further e Pie d tests Includes monopolar cut and coag footswitch Overall physical ia n ds condition is reasonable 845 ERBE ICC The diathermy powers up well and passed its initial self test The display and 200 function buttons are in working order However the patient plate indicator Diathermy alarm stays and does not change Therefore unable to carry out any further am nee d tests Includes monopolar cut and coag footswitch Overall physical aaa E ds condition is reasonable 852 Drager Photo Unit powers up with the phototherapy ligh