B700 Bariatric Bed Full-Electric DC Model User Manual
Contents
1. head motor cable into ae motor plug foot motor cable LA into control box RR mE EESTI FEM H plug hi lo motor cable c into control box plug head motor cable FOOT SECTION into control box 1 Cabling The cables shown above will be installed at the factory on the foot section as well as the control box foot motor hi lo motor and head motor If you ever need to replace an electrical component or cabling follow the diagram above NOTICE To avoid damage to cabling all cables must be routed correctly and cable tied to proper parts of the frame as shown m e a cable and pendant cable along frame and under cross brace cable tie to cross brace cable tie cable tie power cable power cable and pendant cable 0 to hole in frame to hole in d IL rum Cu p TC TOOL HEAD BT UT power cable and pendant cable cable tied to frame CT power cable HET IT Tip ie no n frame IIT HET BM f The head section of the bed does not come with any cable routing The power cable and pendant cables will be coiled and cable tied to the foot section in packaging To route the power cable and pendant cable 2 Cut off the cable ties from the coiled sections 3 Run the cables along the inside and underside of the main frame rails with the pendant
2. If any of the pendant functions do not work as expected recheck all electrical connections to ensure all cables are seated properly within the motors and the control box then retry the pen dant function that did not work If the function still does not work contact your Graham Field equipment provider immediately to help you troubleshoot the problem and provide you with assistance on what to do next If the power goes out in your home or facility unplug the bed immediately from the power out let An emergency crank has been included with the bed to use in such an emergency tube end of crank in foot motor To lower the bed platform with the emergency crank insert the crank s flat end into the hi lo motor shaft slot as shown at above left and turn it counterclockwise To lower the head or foot section with the emergency crank insert the crank s tube end into the end of either the head or foot motor as shown at above right and turn counterclockwise 18 B700 INS LAB RevC10 TRENDELENBURG AND REVERSE TRENDELENBURG POSITION The Bariatric Bed has been designed so that both Trendelenburg and Reverse Trendelenburg positions can be achieved by adjusting the drive shaft and using the pendant hand controller hi lo functions Often used with patients who have a history of low blood pressure or shock or in post surgery situations requiring reductions in abdominal anomalies the positions allow the patient to lie prone with either th
3. cut and discard end of head motor tube REMOVE SHIPPING CABLE TIES FROM HEAD SECTION TRUSSES Tools required blade The two truss assemblies are folded and secured to the head section with two cable ties at either end as shown at right Cut and discard the cable ties HEAD cut cable ties at both ends to detach the trusses before as SECTION ana sembly two folded truss assemblies B700 INS LAB RevC10 7 5 ASSEMBLY INSTRUCTIONS A WARNING Because the bed is very heavy bed assembly requires two people ASSEMBLING THE HEAD AND FOOT SECTIONS Tools required two 9 16 wrenches head motor tube FOOT SECTION SECTION 1 Lay the bed s foot section with motors and junction box upside down flat on the floor as shown above Do the same with the head section close by f A SECTION FOOT move the head motor tube SECTION slightly over to left or right for easier assembly 2 Move the head motor tube slightly over to the right or left to get it out of the way Using two people lift the head section so that it stands perpendicular to the foot section as shown above Lift the head section and align the two rivets on either side of the inside frame with the two hooks in the foot section frame as shown above Lower and slowly maneuver the head deck so that the rivets seat in the hooks A WARNING The rivets MUST be properly seated in the hooks otherwise per so
4. ty 999 0775 206 HI LO motor cable ys 20 0 gt 999 082 204 e FOOT motor cable i 25 5 control box plug 4 control box plug control box plug hi lo motor cable with blue o ring 999 0822 204 plugs in here 17 6 BED OPERATION BED OPERATION WITH THE PENDANT HAND CONTROLLER A NOTICE To use the pendant hand controller correctly pause slightly before pressing a second button for a new operation Avoid pressing multiple but tons at the same time or pressing buttons rapidly If the buttons are pressed too rapidly or multiple buttons are pressed at the same time the pendant functions may not operate properly Release the button wait at least a sec ond then proceed The drawing at right illustrates the func tion of each pendant hand controller but ton To operate the bed with the pendant use the following guidelines 1 Toraise only the head section press the top left button f 2 To lower only the head section press the top right button 3 Toraise the entire bed platform press the middle left button 4 To lower the entire bed platform press the middle right button 5 To raise only the foot section press the bottom left button 6 To lower only the foot section press the bottom right button raise the head il lower the head raise the bed lower the bed raise the foot lower the foot pendant pendant hand controller
5. Mechanical components two years Electrical components three years Welds five years The Warranty period for the consumer commences on the first date a product is delivered to consumer by seller distributor If the product is rented or leased the warranty period com mences on the invoice date from GF Health Products Inc A copy of the invoice showing date of purchase must be provided when submitting warranty claims When proof of purchase date is not provided warranty coverage shall commence upon GF Health Products Inc s invoice date to the distributor purchaser If within the warranty period the product or component part is proven to GF Health Products Inc s satisfaction to be defective GF Health Products Inc shall provide at its option one of the following 1 repair or replacement of any defective or nonconforming part or product or 2 a credit and or refund of the original selling price GF HEALTH PRODUCTS INC S SOLE OBLIGATION AND YOUR EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR REPLACEMENT CREDIT AND OR REFUND This warranty does not include any labor charges incurred in replacement part s installation or any associated freight or shipping charges to the manufacturer For warranty service please contact the authorized distributor from whom you acquired your GF Health Products Inc product Upon receiving notice of an alleged defect in a product GF Health Products Inc will issue a return aut
6. THE HEAD AND FOOT BED ENDS sens 13 ASSEMBLING AND INSTALLING THE HI LO MOTOR DRIVE SHAFT se 15 ELECTRICAL EQUIPMENT AND CABLES 17 6 BED OPERATION RRRRRRRRRRRRRRRRRRRRRRRRRRRRRRRRRRMMMMMMMMMMMMMMMMMMMMMMM 18 BED OPERATION WITH THE PENDANT HAND CONTROLLER eene tte tette tte tetti 18 TRENDELENBURG AND REVERSE TRENDELENBURG POSITION sense 19 7 ASSEMBLING AND INSTALLING OPTIONAL SWING DOWN QUICK RELEASE HEAD RAILS 20 BED RAIL SAFETY A 21 8 MAINTENANCE AND SAFETY CHECKS csssssscssssssessesssssesssesssssssssssssssessessssesstssssssssetssnesstsosssnssetsstesnesstssnesseseten 22 CHECK ELECTRONICS FREQUENTLY eines 22 CHECK BED FRAME AND SLEEPING SURFACE FOR STABILITY ses 22 CHECK HEAD AND FOOT BED ENDS FOR SAFETY AND WEAR 22 CHECK HI LO DRIVE SHAFT FOR PROPER FUNCTIONALITY sens 23 CHECK CASTERS FOR PROPER OPERATION sn 23 MECHANICAL HARDWARE MAINTENANCE sens 23 CLEANING THE 0 23 A il 00 107 o L A 24 ELECTRICAL 24 LI l d INTELLA 24 WEIGHT CAPACITY 24 CONSTRUCTION eerte treten tette ttes steel ste stesso ste sese bees 24 FINISH EEE E TREE 24 DIMENSIONS HEIGHT LENGTH AND WIDTH entente tentent trente tette tette tto tsd 25 10
7. run ning out the center of the bed and the power cable running off the end of the head section toward the wall outlet 4 Cable tie the power cable and the pendant cable to frame rail holes in areas indicated above to prevent pinching of cables or interference with normal bed functions B700 INS LAB RevC10 11 INSTALLING THE CASTERS Tools required mallet LI 4 T non locking H caster Tes JIL ttt LILILA Hr LH Js 17 r 1 il non locking T caster FOOT END L1 locking caster Info Before installing the two locking and non locking casters ensure they will be easily accessible once the bed ends are attached For best results position the two locking cast ers so that they are diagonally on opposite sides of the bed as shown above 1 To install the casters in the bed end legs insert the square caster posts directly into the bottoms of the legs centering the caster grip rings in the legs as shown above 2 Tap the bottoms of the caster wheels gently with a mallet forcing the caster grip rings to compress and enter the legs Continue tapping gently until the square caster posts are en tirely inside the legs B700 INS LAB RevC10 INSTALLING THE HEAD AND FOOT BED ENDS Removing the Bed End Locking Hardware hair pin clip rivet plate as fender washer ka a clev
8. this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collat eral contract or other assurance except those set out in this document B700 INS LAB RevC10 29 12 Actuators 17 Assembly instructions 8 Assembly preparing for 7 B Bed end locking hardware installing 14 Bed end locking hardware removing 13 Bed ends assembling 13 Bed frame and sleeping surface check 22 Bed length 25 Bed operation 18 Bed operation with pendant hand controller 18 Bed rail safety 21 Cabling instructions 11 Casters check 23 Casters installing 12 Caution labels 4 CAUTION significance 3 Cleaning the bed 23 Cleaning the pendant hand controller 23 D DANGER significance 3 Deck height high position 25 Deck height low position 25 Deck width 25 E Electrical equipment and cables 17 Electronics check 22 Foot rails optional 20 Freight policy GF Health Products Inc 6 H Handling 6 Hardware maintenance 23 Head and foot bed ends check 22 Head and foot bed ends installing 13 Head and foot sections assembling 8 Head rails optional swing down quick release as sembling and Installing 20 Hi Lo drive shaft check 23 Hi Lo motor drive shaft assembling and installing 15 l Important safety precautions 4 Info significance 3 Introduction 3 L Limited warranty 28 30 M Maintenance and safety checks 22 Motor
9. 99 0834 908G rail pivot arms hair pin clips 4 p at end of clevis rail mounting tube lanyard assemblies bracket pre welded clevis pins at end of clevis 999 0834 100 to head section lanyard assembly 999 0834 100 1 Align the rails so that the pivot arms face inward toward the bed and the straight ends of the rails face the head end as shown above 2 From the outside insert the clevis end of the quick release clevis lanyard assembly through the rail pivot arm holes leaving the other end with hair pin clip hanging freely 3 Insert the rails with clevis pin into the two holes in the tube mounting brackets on the underside of the head section assembly as shown above 4 From the inside secure the clevis pins to the brackets using the other end of the clevis lan yard assembly Slide the hair pin clips through the small holes in the ends of the clevis pins u Info When ordering Head Rails for the B700 Bariatric Bed order Kit f 48340 Info The hardware bag included in the rail kit contains both the quick release clevis lanyard assembly to attach the rails to the 700 DC Bariatric Bed and the bolts and hard ware needed to attach both rails of the older AC Bariatric Beds Two rail brackets are also included in the hardware bag that are not used on the B700 DC Bariatric Bed Dis card the rail brackets and bolts and any other hardware that you do not use Optional Foot Rails Foot rails are available for the B700 Bari
10. LUNIEX 700 Bariatric Bed Full Electric DC Model User Manual B700 INS LAB RevC10 2010 GF Health Products Inc Read this manual before assembling or operating the B700 Bariatric Bed Save this manual for future reference The most current version of this manual can be found online at www grahamfield com CONTENTS 1 INTRODUCTION A 3 INTENDED USE Sn LL I AD AMD EN ce 3 SIGNAL WORDS seen 3 TOOLS REQUIRED urnes 3 2 IMPORTANT SAFETY PRECAUTIONS siennes 4 SERVICE ON ELECTRONICS csscssessssssesssesssssssssesseessessessessesstessesstsstessesstsssesonsstesaesstssesoessessesesacesaessnssnesaeeasen 5 RADIO FREQUENCY INTERFERENCE sn 5 3 ea ee ee ee eee 6 GF HEALTH PRODUCTS INC FREIGHT POLICY seen 6 UNPACKING du EI 6 INSPECTION riens 6 e PS 6 4 PREPARING FOR ASSEMBLY certet tette tente tte tnnt tte tente tte 7 REMOVE SHIPPING CABLE TIES FROM FOOT SECTION HEAD MOTOR 7 REMOVE SHIPPING CABLE TIES FROM HEAD SECTION TRUSSES tette 7 5 ASSEMBLY INSTRUCTIONS eee tenente tte tette sels LSU 8 ASSEMBLING THE HEAD AND FOOT SECTIONS eee tette trente tente tette stets testet told 8 CABLING INSTRUCTIONS A E E 11 INSTALLING 07 L A 12 INSTALLING
11. NG Do not allow small children on or near the bed during operation Do not allow small children to operate the bed A WARNING When operating any of the bed s movement functions always en sure the patient is positioned properly within the confines of the bed Do not let any extremities head arms or legs protrude over the side or between the bed rails when operating the bed A WARNING B700 Bariatric Bed maximum weight capacity 700 Ib evenly dis tributed including patient mattress accessories rails etc A WARNING Maximum patient weight 600 Ib 4 B700 INS LAB RevC10 N WARNING Do not use the B700 Bariatric Bed with bed mount trapeze or bed mount traction devices A WARNING Do not use the bed as a patient transport A WARNING Do not use any unauthorized rails parts accessories or adapt ers other than those approved by GF Health Products Inc NOTICE Position and secure all bed cables with cable ties to prevent them from being stepped on tripped over or otherwise subjected to damage or stress A NOTICE The pendant hand controller cable must be routed and secured properly to ensure the cable does not become entangled and eventually sev ered during use A NOTICE Ensure cables do not get tangled around the bed rails or legs dur ing bed operation A NOTICE When using nasal or masked type administered equipment oxygen or air tubing must be routed and secured properly to ensure the tubing does not becom
12. SERVICE PARTS RR S 26 SERVICE PART DIAGRAM een 26 SERVICE PART REFERENCE LIST seen 27 11 LIMITED WARRANT Y sscsscssssssssssssssssssscssssssessssssssessessuessessssssessssssessessssssesatssnessesassstssstsatsssesessansssssetsanesnssasssneeesease 28 7 ET EN esses assesses ees sass sees 30 13 NOTES teases eee 31 Lumex and Graham Field are registered trademarks of GF Health Products Inc Packaging warranties products and specifications are subject to change without notice GF Health Products Inc is not responsible for typographical errors 2 B700 INS LAB RevC10 1 INTRODUCTION All information and specifications in this manual are current at the time of printing Please read this manual including the following Important Safety Precautions section in full before using the B700 Bariatric Bed Always consult your professional healthcare advisors for their recom mendations about safety and never hesitate to ask for their assistance INTENDED USE The 700 Bariatric Bed Full Electric DC Model is designed for bariatric patients in home or institutional environments Contraindications WARNING B700 Bariatric Bed maximum weight capacity 700 Ib evenly distributed total weight including patient mattress accesso ries rails etc 600 Ib maximum patient weight SIGNAL WORDS Please heed the significance of the following special statements used throughout this manual A DANGER Ind
13. ST The bed diagram on the previous page and the following list of service parts have been pro vided for your convenience Keep this list in a secure location and use it as a reference when ordering new or replacement parts Keep the bed s serial number and manufacture date handy for reference for your convenience page 31 has a place to record them 1 1 2 6 999 0775 206 2 control box power cable 1 18 650 3002 052__ Head motor 8a 999 0778 204 8 0 head motor cable yellow o ring 9 650 3002053 JFootmotor 999 0808 204 25 5 foot motor cable red o ring 18e 1 each 3 WARNING The use of non Lumex replacement parts could change the struc ture of the bed void the warranty and create a hazardous condition which could result in serious personal injury 1 600 2001 912 Foot end gearbox right angle elbow 1 1 1 B700 INS LAB RevC10 27 11 LIMITED WARRANTY GF Health Products Inc warrants this product to be free from manufacturing defects in mate rial and workmanship for its period of warranty that is in accordance with industry standards This warranty is extended only to the original purchaser consumer or distributor non con sumer of this new product and to no other purchaser or transferee GF Health Products Inc warrants a new 700 Bariatric Bed components to be free from defects in workmanship and materials according to the following conditions
14. a 2 prong plug Do not modify the plug in any way WARNING Do not use the bed if the power cable or plug is damaged or de fective Defective cables or plugs could result in fire or electric shock WARNING Do not use potentially explosive gases near the bed Fire hazard is possible with use of oxygen administering equipment other than nasal or masked devices A WARNING When using liquid or around the bed ensure liquid of any kind is not spilled If liquid is spilled in or around the bed unplug the bed from the electric outlet immediately Clean up the spill and allow the bed and area around the bed to dry thoroughly before using the electric controls again A WARNING Ensure the patient s body weight is evenly distributed over the entire sleeping surface of the bed and that the patient does not lie sit or lean in such a manner that their entire body weight is placed entirely on the raised head or foot section of the bed This also applies to anyone providing assistance when repositioning or transferring a patient WARNING Do not move or push the bed while it is in an elevated position Always lower the bed to its lowest position before moving N WARNING Ensure the casters are unlocked and free from restriction before moving the bed A WARNING Keep hands and feet clear of all moving parts WARNING Never permit anyone under the bed at any time WARNING Do not leave the bed unattended when plugged in A WARNI
15. atric Bed They require bolting two mounting brackets to the foot section The rails brackets hardware and quick release lanyards are included in the kit Info When ordering Foot Rails for the B700 Bariatric Bed order Kit A8510 20 B700 INS LAB RevC10 BED RAIL SAFETY WARNING If bed rails are not properly installed and adjusted personal in jury and or damage to the bed rail could result The bed frame and its components including the mattress bed rails head and foot board bedding and any accessories added to the bed can all af fect the risk of entrapment Thorough patient assessment and monitor ing are necessary to reduce the risk of entrapment including establishing whether the use of a bed rail is in the best interest of the patient Patients most vulnerable to entrapment are those who are frail confused restless or who have uncontrolled body movement For information on Bed Rail safe ty please see the following brochure A Guide to Bed Safety Bed Rails in Hospitals Nursing Homes and Home Health Care The Facts on the FDA website at www fda gov MedicalDevices ProductsandMedicalProcedures GeneralHospitalDevicesandSupplies HospitalBeds WARNING The use of non Lumex replacement parts and or bed rails will void the warranty and could create a hazardous condition resulting in seri ous personal injury WARNING GF Health Products Inc assumes no responsibility for any dam age or injury ca
16. cables for cuts or damage and all plugs for damage 3 Inspect the junction box for any damage to the connectors 4 Ensure the motor plugs are in good working order and fit properly into the control junction box receptacles STORAGE If the bed will not be used immediately upon receipt repackage the bed after inspection in the original carton reusing the packing materials Store the re packaged bed in a dry area until needed A NOTICE DO NOT place anything on top of the re packaged bed or damage could occur B700 INS LAB RevC10 4 PREPARING FOR ASSEMBLY WARNING Shipping cable ties must be removed before assembly REMOVE SHIPPING CABLE TIES FROM FOOT SECTION HEAD MOTOR TUBE Tools required blade 1 The head motor tube is secured to the foot section with two cable ties for added security during shipping as shown at right Cut the two cable ties and discard them 2 The end of the head motor tube is secured to the foot section with a hook and loop strap as shown at right to secure it during moving and storage when the bed is not in use The hook and loop strap is attached to the head motor tube with a cable tie Do not remove this cable tie When setting up the bed SECTION simply unstrap the hook and loop strap from around the 4 deck tube and wrap it tightly with cable tie do not cut around the head motor tube then reinstall as shown at right for moving or storage two shipping cable ties
17. e brace end slot large ends over the protruding welded shoulder bolts near the ends of both the head and foot section frames then slide the small slot ends over the shoulder bolts so that the bolts are trapped in the small slot ends as shown above motor guard truss assembly vertical link bolt truss assembly vertical link locking nut truss assembly vertical link 5 Attach the truss assembly vertical link to the main frame Remove the nut from the bolt at the end of the vertical link as shown above From the outside of the main frame insert the bolt with link through the central hole From the inside of the frame secure the bolt using the same locking nut you removed earlier Tighten both the bolt and the nut using two 9 16 wrenches Repeat for other side 6 Once the truss assemblies are securely installed carefully turn the bed over so that it rests on the motor guards 10 B700 INS LAB RevC10 CABLING INSTRUCTIONS plug power cable cable tie all three motor into control box cables at control box cable tie corner of control cable tie hi lo and foot cable tie box to hole in main frame motor cables to underside power cable and with large cable tie of motor pendant cable 1 2 to hole in frame cable tie control box to main frame rail using two cable ties into foot motor Mw T a gt hi lo motor cable 2 into hillo motor es
18. e entangled and eventually crimped or severed during bed opera tion A NOTICE Keep all moving parts free of obstruction i e blankets sheets heating blankets pads tubing wiring and any other objects A NOTICE To use the pendant correctly pause slightly before pressing a sec ond button for a new operation avoid pressing multiple buttons at the same time or pressing buttons rapidly If the buttons are pressed too quickly or multiple buttons are pressed at the same time the pendant functions may not operate properly Release the button s wait a second or two then pro ceed SERVICE ON ELECTRONICS WARNING The bed s electronic system contains no serviceable parts Do not open the motors cables pendant hand controller or junction box Only an authorized equipment provider or factory trained personnel may perform service or adjustments on these components RADIO FREQUENCY INTERFERENCE A WARNING Most electronic equipment is influenced by Radio Frequency Interference RFI ALWAYS exercise caution in regard to the operation of any portable communication devices within the area of the bed s electronic equipment B700 INS LAB RevC10 5 3 HANDLING GF HEALTH PRODUCTS INC FREIGHT POLICY 1 For Your Protection Read Carefully The carrier accepted this merchandise in good condition and is responsible for safe deliv ery Before signing any freight bill inspect the shipment carefully for any damage or missing pa
19. e feet raised on angle or head end raised without knees bent Physician approval is recommended before using the feature Trendelenburg Position 1 Lower the bed platform to its lowest height Disengage the hi lo drive shaft by compress ing the spring fitting at the head end and removing it from the head gear shaft and foot hi lo motor perform step 3 on page 15 2 With the bed now in its lowest position use the pendant middle button hi lo function to raise the foot end to its highest position The head end will remain in the low position Reverse Trendelenburg Position 1 Raise the bed platform to its highest height Disengage the hi lo drive shaft by compress ing the spring fitting at the head end and removing it from the head gear shaft and foot hi lo motor perform step 3 on page 15 2 With the bed now in its highest position use the pendant middle button hi lo function to lower the foot bed end to its lowest position The head end will remain in the high position B700 INS LAB RevC10 19 7 ASSEMBLING AND INSTALLING OPTIONAL SWING DOWN QUICK RELEASE HEAD RAILS The Bariatric Bed has the option of half rail installation on the head section The rails mount on brackets which have been pre welded onto the head section A quick release system will allow the caregiver to quickly remove or attach the rails as needed HEAD SECTION angled end of right half rail 999 0834 907G J straight end of left half rail 9
20. e small spring loaded inner drive shaft and place the shaft end notch on the head end gearbox shaft cross pin as shown at right ud ND gearbox FOOT SECTION cross pin hi lo motor with coupler HEAD SECTION outer drive shaft gearbox FOSS pm notch inner div shaft spring loaded y B700 INS LAB RevC10 ELECTRICAL EQUIPMENT AND CABLES The following diagrams show the electronic components and proper cable plug in locations pendant cable attached to pendant 650 3002 305 controller power cable 650 3002 305 X AW receptacle CONTROL BOX Fg d 999 0822 204 blue o ri 650 3002 300 ue o ring FOOT motor cable receptacle 999 0808 204 red o ring not used pendant cable receptacle Motors Actuators HEAD MOTOR 650 3002 052 idi head motor cable with yellow o ring 999 0778 204 plugs in here FOOT MOTOR 650 3002 053 Pad foot motor cable with red o ring 999 0808 204 plugs in here B700 INS LAB RevC10 HEAD motor cable receptacle 999 0778 204 yellow o ring power outlet HI LO motor cable receptacle attached to pendant 650 3002 305 HI LO MOTOR with coupler 650 3002 914 motor plug HEAD motor cable cR 8 0 a aL 2 999 0778 204 trol b motor plug Su lu g 12 E E uu motor plug s 2 C 999 0808 204 power cable T 126 0
21. er shaft also has a small spring loaded inner shaft on the opposite end for ease of installation 2 Depress the spring button and slide the inner shaft into the outer shaft until the spring but ton engages in the 3rd hole of the outer shaft as shown Unlocking the Hi Lo Coupler Info Connecting the drive shaft to the bed is easier if you first insert the hi lo motor shaft coupler into the foot bed end gearbox 3 See picture at right The hi lo motor comes pre assembled on the foot section of the bed The hi lo motor shaft includes a coupler located toward the foot end that connects to the FOOT SECTION foot bed end gearbox and is shipped in the locked position gearbox To unlock the coupler and insert it into the gearbox turn o it clockwise to release it from Vi the locking pin li hi lo motor B700 INS LAB RevC10 15 4 Once the coupler is unlocked the inner spring will force it toward the foot end gearbox located at the bot tom of the bed end assembly The hi lo motor shaft cross pin should now rest at the end of the long groove in the coupler and the end of the coupler should rest inside the gearbox and engage the gear shaft Attaching the Drive Shaft to the Bed 5 To attach the drive shaft assembly to the bed at the foot end place the inner drive shaft end notch on the hi lo motor shaft cross pin as shown at right inner drive shaft 6 At the head end depress th
22. horization The defective product or part s must then be returned at the purchaser s cost for warranty inspection using the serial number as identification or if the product is not serialized lot number and date code within thirty 30 days of return authorization issue date In the event you do not receive satisfactory warranty service please contact GF Health Products Inc at the address below DO NOT return products to our factory without prior authorization LIMITATIONS AND EXCLUSIONS The foregoing warranty excludes consumables such as cast ers as these require routine maintenance The foregoing warranty shall not apply to serial numbered products if the serial number has been removed or defaced Products subjected to negligence abuse misuse improper operation improper maintenance improper cleaning improper storage or damages beyond GF Health Products Inc s control are not covered by this warranty and that evaluation shall be solely determined by GF Health Products Inc This warranty shall not apply to problems arising from normal wear and tear or failure to follow instructions The warranty shall also not apply to products modified without GF Health Prod ucts Inc s express written consent nor shall it apply if parts not manufactured by GF Health Products Inc or if parts not complying with original equipment specifications are added to GF Health Products Inc products or if the product or part is serviced by an ent
23. icates a potential hazard situation or unsafe practice that if not avoided will result in death or serious personal injury WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious personal injury CAUTION Indicates a potential hazard situation or unsafe practice that if not avoided could result in minor or moderate personal injury A NOTICE Indicates a potential hazard situation or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other useful information to ensure you get the most from your product TOOLS REQUIRED The following tools are required to perform assembly Blade to cut cable ties Mallet Two 9 16 wrenches B700 INS LAB RevC10 2 IMPORTANT SAFETY PRECAUTIONS WARNING Warning Caution labels applied to the B700 Bariatric Bed de scribe hazards or unsafe practices that could result in personal injury and or property damage Do not remove such labels WARNING The B700 Bariatric Bed engineered to provide you with years of reliable operation is intended for household and commercial use only WARNING The B700 Bariatric Bed is intended for indoor use only WARNING Unplug the bed from the electrical outlet when not in use and be fore servicing or cleaning A WARNING The B700 Bariatric Bed is equipped with a double insulated elec trical system using
24. is pin 1 Vou must first remove the bed end locking hardware clevis pins washers and hair pin clips total 4 each from the frame rivet plates on both sides of the bed before vou can at tach the bed ends The bed end locking hardware has been pre installed at the factorv for ease of shipment and will be reinstalled after installing the bed ends to act as latches when the bed ends are attached to head and foot sections Installing the Bed Ends on the Bed A WARNING Because the bed heavy bed end assembly requires two people foot section end rivet bed end hook 2 Position the bed on the floor resting on the motor guards with deck on top as shown above Position the foot bed end assembly near the bed so that the hooks face toward the foot section as shown With a person on each side carefully lift the foot section at an angle and ease the two foot section end rivets into the bed end hooks on both sides of the foot bed end You may need to tilt the bed end backward slightly away from the foot section as shown above in order to maneuver the rivets into both hooks B700 INS LAB RevC10 13 3 Once the rivets are aligned with the hooks tilt the bed end forward toward the foot section as shown above then lower the bed into the hooks to lock into place Do the same for the head end of the bed repeating steps 2 and 3 Installing the Bed End Locking Hardware A WARNING Ensure the clevis pins washers and hair pin clips
25. ity not authorized by GF Health Products Inc THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRAN TIES AND IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRAN TIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND SHALL NOT 28 B700 INS LAB RevC10 EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE LIMITED TO THE REPAIR REPLACEMENT CREDIT AND OR REFUND OF THE DEFECTIVE PRODUCT OR PART PURSUANT TO THE TERMS CONTAINED HEREIN GF HEALTH PRODUCTS INC SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state province to province Some states provinces do not allow the exclu sion or limitation of incidental or consequential damage or limitation on how long an implied warranty lasts so the above exclusion and limitations may not apply to you This warranty contains the entire agreement between the parties and supersedes any prior contrary or ad ditional representations or understandings whether oral or written concerning our warranty policy The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The recipient of
26. nal injury and or property damage could occur 8 B700 INS LAB RevC10 head motor tube clevis pin circle cotter SECTION P g mounting bracket FOOT 3 Carefully lower the head section to the floor ensuring the head section rivets remain en gaged in the foot section hooks Remove the clevis pin and circle cotter shown above from the head section s mounting brackets Lift up the head motor tube and place it so that it rests between the two brackets and the first hole lines up with the bracket holes Insert the clevis pin and secure with the circle cotter as shown above slots on both ends of truss assemblies attach to shoulder bolts on frame bolt and nut to attach vertical truss link to main frame comes RIGHT AND LEFT pre assembled on truss assembly TRUSS ASSEMBLIES The two truss assemblies shown above span the decks and give them more stability much like the trusses on a bridge The trusses attach to the main frame via welded shoulder bolts on either side of the head and foot sections B700 INS LAB RevC10 side brace vertical FOOT SECTION a position side brace end slot large ends over bed frame shoulder bolts then slide small slot ends up to trap bolts on both sides of the bed 4 Position the truss assemblies on both sides of the bed so that the vertical truss links face the middle of the bed and the side braces are on the outside Position the sid
27. on 3 Check to ensure all casters roll freely without restriction when unlocked MECHANICAL HARDWARE MAINTENANCE Check all securing hardware i e bolts nuts clevis pins circle cotters to ensure they are se cured and or tightened properly CLEANING THE BED WARNING Unplug the bed from the electrical outlet before servicing or cleaning Cleaning the Metal Components All metal parts of the bed have been covered with a baked on epoxy coating 1 To clean coated metal parts Wipe with a clean soft cloth dampened with mild detergent and warm water 2 To clean the mattress deck Remove the mattress then wipe mattress deck with a clean soft cloth dampened with mild detergent and warm water Allow to air dry before replacing mat tress 3 To remove dust from the frame Carefully raise the head and foot sections of the bed and wipe the frame with a clean soft cloth Cleaning the Pendant Hand Controller Wipe the pendant hand controller with a clean soft cloth dampened with mild detergent and warm water Air dry A NOTICE Do not use corrosive or powdered cleansers to clean the pendant hand controller A NOTICE Do not immerse or soak the pendant hand controller B700 INS LAB RevC10 23 9 SPECIFICATIONS Info Specifications are subject to change without notice ELECTRICAL e Full electric three function operation e Motors 120V 50 60Hz Max 3 5 amps SHIPPING WEIGHT e 295 Ib WEIGHT CAPACITY e Maxim
28. perly to the bed frame with cable ties so that they do not interfere with any moving parts 4 Ensure all plugs are fully inserted and secured in the proper receptacles 5 Ensure all hardware used to secure the control box is properly tightened CHECK BED FRAME AND SLEEPING SURFACE FOR STABILITY 1 Visually check all the welds ensuring that there are no stress cracks e check the head section e check the foot section e check the main frame 2 Check all joints between the sleeping surface sections for loose fasteners 3 Inspect all center mounting rivets and latches to ensure they not sheared off or bent 4 Inspect all head and foot section surfaces for any bends warping or stress damage CHECK HEAD AND FOOT BED ENDS FOR SAFETY AND WEAR 1 Inspect the end panels and trim for any wear holes and or sharp edges 2 Inspect the corner locks and locking hardware for any wear cracks or stress damage 22 B700 INS LAB RevC10 CHECK HI LO DRIVE SHAFT FOR PROPER FUNCTIONALITY 1 Ensure the center spring button functions properly and locks both sections 2 Inspect the connections at both head and foot ends for any wear or bends 3 Inspect the entire length of the hi lo drive shaft for any bends or stress damage CHECK CASTERS FOR PROPER OPERATION 1 Ensure the locking casters are positioned at opposite ends diagonally 2 Checkthe locking function on both locking casters ensuring they work well and lock prop erly without restricti
29. rts Apparent Loss or Damage Should visual inspection show loss or damage it MUST be noted on the freight bill and signed by the carrier s agent Failure to do so may result in the carrier failing to honor the claim Please contact the carrier to obtain the paperwork necessary to file the claim Concealed Loss or Damage If damage is discovered AFTER delivery is made a concealed damage claim MUST BE CORDED with the freight carrier When this occurs send a written request to the carrier for inspection This request must be made WITHIN 15 DAYS OF DELIVERY The carrier should provide all paperwork necessary to file a concealed damage or loss claim since such dam age or loss is the carrier s responsibility UNPACKING m o Info Unless the bed is to be used immediately retain all containers and packing materi als for storage until the bed is required for use Check for any obvious damage to the bed carton or its contents If damage is evident please notify the carrier and your Graham Field equipment provider immediately Remove all loose packing material from the carton and inspect to ensure there are no parts enclosed in the material Carefully remove all components from the packing carton and lay them on a stable surface INSPECTION WARNING Do not operate bed controls before fully assembling the bed 1 Examine each item carefully for nicks dents scratches or any other damage 2 all
30. s 17 N Non Lumex bed rails 21 Non Lumex replacement parts 21 Notes 31 NOTICE significance 3 5 Service electronics 5 Service part diagram 26 Service part reference list 27 Service parts 26 Shipping cable ties remove 7 Signal words 3 Specifications 24 T Tools required to perform assembly 3 Trendelenburg position 19 Trendelenburg position reverse 19 U Unpacking 6 W Warning labels 4 WARNING significance 3 Weight capacity 24 Weight shipping 24 B700 INS LAB RevC10 13 5 Serial number Manufacture date B700 INS LAB RevC10 31 LUNIEX A Graham Field Brand Corporate Headquarters GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 U S A 110 368 4700 GRAHAM FIELD www grahamfield com Made and Printed in USA
31. um weight 700 Ib evenly distributed including patient mattress accessories rails etc e Maximum patient weight 600 Ib CONSTRUCTION e Frame 1875 steel angle e Sleeping surface 11 GA 120 hot rolled steel e Self wrenching nut assembled for easy part replacement e Split frame with removable ends for easy transport e Truss reinforced frame e Ball screw bed ends e Heavy duty grid style sleeping surface e Optional swing down style trigger rail FINISH e Features an antimicrobial powder coat process which has been proven to inhibit the growth of damaging and disease causing microbes bacteria mold fungus algae etc on the sur face of coated items Recorded result reduction of bacteria has been 99 9 24 B700 INS LAB RevC10 DIMENSIONS HEIGHT LENGTH AND WIDTH The following diagrams depict the various positions and dimensions of the 700 Bariatric Bed Deck Height High Position deck height in fully raised position top of deck to floor 27 00 Deck Height Low Position deck height in fully lowered position _ E s I ee Je So E LE AS topofdeck dl to floor 17 00 amp en Bed Length full bed length head gear box to foot gear box 88 00 Deck Width full bed width outside of bed ends 42 00 B700 INS LAB RevC10 25 10 SERVICE PARTS 26 B700 INS LAB RevC10 SERVICE PART REFERENCE LI
32. used by improper installation of bed rails or unauthorized bed rail use Only Lumex bed rails are recommended for use with Lumex beds Please consult the instructions that accompany the Lumex bed rails for complete rail assembly and installation instructions B700 INS LAB RevC10 21 8 MAINTENANCE AND SAFETY CHECKS A WARNING Perform the following maintenance program to assure that the B700 Bariatric Bed continues to operate correctly Perform these checks at least once a year and between patient placements A WARNING Do not try to modify the Bariatric bed in any way Unauthorized modification could change the structure of the bed void the warranty and create a hazardous condition which could result in serious personal injury A WARNING If you experience a problem with the Bariatric Bed and un able to resolve it yourself please contact your Graham Field equipment pro vider CHECK ELECTRONICS FREQUENTLY 1 Using the pendant ensure all functions work properly and that each button performs the appropriate function see Bed Operation with Pendant Hand Controller 2 Check all plugs and cables for damage wear and frayed wires including e power cable e head motor cable e hi lo motor cable foot motor cable e pendant cable WARNING Do not use the bed if cables or plugs are damaged or defective Defective cables or plugs could result in fire or electric shock 3 Ensure all wires are routed and attached pro
33. which act as latches are properly installed to avoid personal injury or property damage fender washer clevis pin Once you have the head and foot bed ends installed on the bed with the rivets resting securely in the hooks as shown at right reinstall the locking hardware on all four corners to lock them into place Slip a fender washer onto the exposed end of each clevis pin From the outside insert the two clevis pins with washers that you removed in step 1 through the holes on both sides of the foot section s rivet plates as shown at right The fender washers will rest on the outside of the bed end s hook plates as shown From the inside insert a hair pin clip into the small hole at the end of each clevis pin Follow the same procedure for the head bed end section foot section rivet plate bed end hook plate B700 INS LAB RevC10 ASSEMBLING AND INSTALLING THE HI LO MOTOR DRIVE SHAFT The telescoping hi lo motor drive shaft assembly consists of an inner shaft with a positioning spring button and an outer shaft with several positioning holes drive shaft assembled inner drive shaft with spring button outer drive shaft with holes spring button seats in 3rd hole 1 Lay the drive shaft parts on a flat surface so that the inner shaft is to your left and the outer shaft is to your right with the spring button end of the inner shaft facing the holed end of the outer shaft as shown above The out