Home

Agilent Varian TwisTorr 304 FS AG Rack Controller

image

Contents

1. Figure 3 Controller TwisTorr 304 FS AG Rack Rear panel 1 Logic input signals connectors mating connector supplied with link 2 Logic output signals connector and pump freguency monitor plus programmable analog output Power output connector 24 Vdc for pump cooling fan Vent Valve and Gauge Controller power entry module consisting of mains fuses mains socket and EMC filter Rotary switches for Profibus address setting Profibus Optional Remote control serial connector oo Ml o o B Pump cable bm long 202 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instructions for Use 16 Use Procedure Controller Startup To startup the controller plug the power cable into a suitable power source Starting the Pump To start the pump confirm the interlock input and press the START pushbutton on the controller front panel Pump Shutdown To shutdown the pump press the STOP push button on the controller front panel Maintenance The TwisTorr 304 FS AG Rack series controller does not reguire any maintenance Any work performed on the controller must be carried out by authorized personnel When a fault has occurred it is possible to use the Agilent repair service Replacement controllers are available on an advance exchange basis through Agilent WARNING Before carrying out
2. Controller power supply circuitry is faulty or the Controller received a spike Press the START push button twice to start the pump Should the message still be present call the Agilent service Check connections and shortages between pump and controller then press the STOP RESET pushbutton twice to start the pump TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Orderable Parts Tab 22 Description Technical Information 17 Part number TwisTorr 304 FS AG Rack Controller with RS 232 485 X3506 64002 TwisTorr 304 FS AG Rack Controller with Profibus option X3506 64003 TwisTorr 304 FS AG Air Cooling Kit 0 5 m cable X3500 68011 Turbo V 301 AG Air Cooling Kit extention cable 5 m 969 9940 Vent Valve N 0 0 5 mm orifice 0 5 m cable 969 9844 Vent Valve extention cable 5 m 969 9941 Mains cable NEMA plug 3 m 969 9958 Mains cable european plug 3 m 969 9957 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 295 301 EE Agilent Technologies Vacuum Products Division Dear Customer Thank you for purchasing an Agilent vacuum product At Agilent Vacuum Products Division we make every effort to ensure that you will be satisfied with the product and or service you have purchased As part of our Continuous Improvement effort we ask that y
3. Where NORMAL indicates the operating regime of the pump XXXHz see description Pressing once Measures key I X XXA indicates the current absorbed by the pump V XXX V indicates the pump driving voltage Pressing 3 times the following is obtained P X X X W X X X X H z X X X JE Y Y m B ja R TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 251 301 17 Technical Information P X X X W X X X X JH z Where P XXXW indicates the current power supplied to the pump XXXHz see description Pressing once Measures key X XE YYmBar indicates the pressure currently measured by the gauge ZZZZ indicates the kind of gauge see table of gauge supported in the gauge chapter Other possible indications P IX X X W X X X X H z G A U G JE ER RIO R Where GAUGE ERROR gauge not connected or broken P IX X X W X X X X H z PIR JE JS S JU R JE U N D E R Where PRESS UNDER indicates that the pressure read is below the minimum measurement threshold set 5 E 9 mBar P X X X W X X X X H z PIR JE ISIS JU R JE O IV JE R 252 301 TwisTorr 304 FS AG Rack
4. El OA LJ D EI Slika 3 Ozadnja plo ca Controllera TwisTorr 304 FS AG Rack 1 Vhodni konektor logi amp nih signalov spajalni konektor ima ustrezen zapiralni mosti ek 2 Konekotr logi n h izhodnih signalov in nadzor frekvence rpalke plus analogi ni izhod ki lahko programirate Izhodni napajalni konektor 24 Vdc za ohlajevalni ventil rpalke vent valve in kaliber Vhodni modul napajanje controller a ki vsebuje varovalke napajalno stikalo in filter ENC Rotacijski Switch za nastavljanje Profibusnega naslova Profibus kot opcija Serijski konektor za nadzor iz oddaljenosti 0 J O o B w Konektor rpalke 190 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Priro amp nik za Navodila 15 Postopki uporabe Vklopitev Controllera Za vklopitev Controller a morate napajalni kabel vtakniti v omre no vti nico Zagon rpalke Da bi zagnali rpalko morate usposobiti vhod interlock a in pitisniti tipko START na prednji plo i Zaustavitev rpalke Zaustavite rpalko s tipko STOP na prednji plo i Vzdr evanje Controllerji serije TwisTorr 304 FS AG Rack ne potrebujejo vzdr evanja Samo poobla eno osebje lahko rokuje s to napravo V slu aju okvare obrnite se servisu Agilent ali Agilent advanced exchange service ki Vam bo dal na razpolago obnovljeni controller
5. 5 1 3 2 4 Postava 2 eln panel Konrol ru TwisTorr 304 FS AG Rack 1 Tla tko p ivolan ch parametr na displeji cycle number cycle time a pump life 2 Tla tko na volbu zp sobem LOW SPEED Je aktivn jen p i zvolen m zp sobu ovl d n z eln ho panelu Stiskne li se jen jednou erpadlo pracuje rychlost stand by A stiskne li se je t jednou deaktivuje se zp sob LOW SPEED 3 Tla tko na p ivol n parametr na displeji pump current pump power rotational speed temperatura spodn ho lo iska temperatura vrchn ho lo iska temperatura motoru frekvence indukce Nez visle od zvolen ho provozn ho zp sobu je neust le aktivn Jestli e se tla tka 3 a 1 stisknou spole n alespo na 2 vte iny bude aktivov n program dle kter ho se mohou naprogramovat n kter operativn parametry 4 Tla tko ovl d n START STOP RESET Je aktivn m jen po volb na eln m panelu Stiskne li se jen jednou spou t c f ze bude aktivov na jestli e se znovu stiskne erpadlo se zastav Jestli e se erpadlo automaticky zastav z d vodu n jak poruchy tla tko se mus napoprv stisknout aby se uskute nil reset kontrol ru a po druh aby se erpadlo znovu spustilo 5 Alfanumerick displej na tekut krystaly matice bod 2 linky x 16 charakter TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 165 301 13 N vod k Pou it
6. 2 FR I 7 One Postava 3 Zadn panel Kontrol ru Twis Torr 304 FS AG Rack Vstupn konektor logick ch sign l konektor spojky je dod v n s p slu n m znovu zav raj c m m stkem 2 Konektor v stupn ch logick ch sign l a monitorov n frekvence erpadla plus naprogramovateln analogov v stup 3 Konektor v stupu nap jen 24 Vdc na ochlazovac lopatku erpadla vent valvea a kalibr 4 Vstupn nap jec modul kontrol ru kter zahrnuje i tavn pojistky nap jec z suvku a filtr ENC 5 Rota n switch na stanoven Profibusov adresy 6 Profibus op n 7 Seri lni konektor na d lkovou kontrolu 8 Konektor erpadla 166 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod k Pou it 13 Pou van procedury Zapalov n Kontrol ru Kontrol r se zapne vlo en m kabele do z suvky na s ti Spou t n erpadla Ke spu t n erpadla je t eba opr vnit vstup interlocku a stisknout tla tko START na eln m panelu Zastaven erpadla erpadlo se zastav stisknut m tla tka STOP na eln m panelu Udr ba Kontrol r s rie TwisTorr 304 FS AG Rack nevy aduj dnou dr bu Jak koliv z sah mus b t vykon v n jen opr vn n m person lem V p pad poruchy se m e vyu t oprav sk ho servisu
7. 5 Display alfanumerico a cristalli liquidi matrice di punti 2 linee x 16 caratteri TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 21 301 1 Istruzioni per uso E EE Ef Figura 3 Pannello posteriore dei Controller TwisTorr 304 FS AG Rack Connettore di ingresso dei segnali logici il connettore di accoppiamento viene fornito con l apposito ponticello di richiusura 2 Connettore segnali logici in uscita e monitoraggio freguenza pompa piu uscita analogica programmabile 3 Connettore di uscita alimentazione 24 Vdc per la ventola di raffreddamento pompa vent valve e calibro 4 Modulo di ingresso alimentazione controller che comprende i fusibili la presa di alimentazione ed il filtro ENC 5 Switch a rotazione per impostare l indirizzo del Profibus 6 Profibus opzionale 7 Connettore seriale per controllo da remoto 8 Pompa 22 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruzioni per l uso 1 Procedure di uso Accensione del Controller Per accendere il controller sufficiente inserire il cavo di alimentazione nella presa di rete Avvio della Pompa Per avviare la pompa occorre abilitare l ingresso di interlock e premere il pulsante START del pannello frontale Arresto della Pompa Per arrestare la pompa occorre premere il pulsante STOP del
8. ES Z I os Essas 1 tr I I SERIAL PE gt BAUD RATE I 1677 i seen K Flow 1 6 Figure 38 Mode Menu MODE permits selection see Flow 1 6 of the operating mode of the controller from those indicated below FRONT accepts commands from the front panel REMOTE accepts commands from the remote input connector SERIAL accepts commands from the serial interface or Profibus interface if present PROFIBUS this operating mode is present only if the Profibus interface is installed on the controller TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Pump Setting Menu Technical Information 17 PS 1 PUMPSETTING ELZ p k v HIGH SPEED ADJ Hz E LOW SPEED ADJ DEEE REES Hz lt gt 0000 x 4 MEER AA 9999 SOFT START amp gt YES niec n san A E i y i Only Front Mode H 9999 E Configuration NO i aties Et EXTERNAL FAN CNF gt ALWAYS ON A OFF LIMPOWER SERIAL Y VENT VALVE MODE AUTO TIME A k MANUAL v AUTO PUMP SPEED VENT VALVE DELAY lt gt 0000 VENT OPEN TIME 0000 soc i 0 NO LIMIT LOAD GAS TYPE p Argon E Air v WATER COOLING 4 pi YES N b NO v ACTIVE STOP lt gt YES A No v
9. TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 25 301 1 Istruzioni per l uso 26 301 MESSAGGIO DESCRIZIONE AZIONE CORRETTIVA TOO HIGH La corrente assorbita dalla pompa Verificare che il rotore della LOAD maggiore di quella programmata pompa abbia la possibilit di ruotare liberamente Premere due volte il pulsante START per riavviare la pompa SHORT Durante il funzionamento normale Verificare i collegamenti tra CIRCUIT dopo la fase di avvio la pompa e controller Premere connessione di uscita in corto due volte il pulsante START circuito per riavviare la pompa SYSTEM La pompa stata fermata da un Staccare il cavo di OVERRIDE segnale di emergenza proveniente alimentazione del controller da un contatto remoto e correggere la causa dell emergenza Ricollegare il cavo di alimentazione e premere due volte il pulsante START per riavviare la pompa OVERVOLTAGE Si verificato un guasto nella Premere due volte il pulsante sezione di alimentazione del START per riavviare la controller o il controller ha ricevuto pompa Se il messaggio si un segnale spurio ripresenta rivolgersi in Agilent per la manutenzione POWER FAIL Guasto nella sezione di Rivolgersi alla Agilent per alimentazione della pompa del controller Manutenzione TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual Gebrauchsanleitung Allgemeines
10. Podczas pracy kontrolera nalezy zapewnic nastepujace warunki otoczenia temperatura od 5 C do 45 C a wilgotnosc wzgledna 0 95 nie kondensujaca Sposob polaczenia i instalowania akcesoriow jest opisany w rozdziale Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 151 301 12 Podrecznik Instrukcji Uzytkowanie Paragraf ten zawiera podstawowe procedury operatywne Dla uzyskania dodatkowych bardziej szezegolowych informacji lub dostepu do procedur na temat polaczen lub akcesoriow skonsultowac paragraf Use znajdujacy sie w dodatku do Technical Information Przed przystapieniem do uzytkowania kontrolera wykonac wszystkie polaczenia elektryczne i pneumatyczne podlaczenie pompy konsultowac z odnosna instrukcja Dla unikniecia obrazen personelu obslugujacego urzadzenie jak rowniez uszkodzen samego urzadzenia nalezy upewnic sie szczegolnie w przypadku gdy pompa zostala ustawiona na stole czy stol jest odpowiednio stabilny Pod zadnym pozorem nie uruchamiac pompy jezeli kolnierz wejsciowy nie jest polaczony z systemem lub nie jest zamknieta kolnierzem zamykajacym W przypadku kiedy nie zostanie wykonane zadne polaczenie zewnetrzne lacznik zamykajacy J1 powinien pozostac polaczony ze swoim mostkiem Pompa wstepnej prozni i pompa TwisTorr 304 FS moga byc uruchomione jednoczesnie 152 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Podrecznik
11. RUNNING A STARTING Y SPEED DISPLAY gt KRPM A E Hz v LCD CONTRAST gt o meen i I I 7 Flow 3 6 Figure 40 START MODE Permits selection of one of the three types of functioning of the START STOP and INTERLOCK inputs 268 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 START INTERLOCK The pump is started if the START STOP input is activated and is stopped if the START STOP input is de activated The interlock input must be constantly active INTERLOCK TYPE parameter continuous or may be active only at the time of starting INTERLOCK TYPE parameter impulsive START STOP LEVEL The START STOP input acts as only START command active on the level The INTERLOCK input acts as STOP command active on the level In the case of concurrent activation of the two inputs the STOP command takes priority START STOP EDGE The START STOP input acts as START command active on the edge The INTERLOCK input acts as STOP command active on the edge INTERLOCK TYPE This parameter is displayed only when the previous parameter START MODE is egual to START INTERLOCK and pernits selection of one of the following two operating modes of the INTERLOCK input CONTINUOUS the INTERLOCK input must be active continuously so that the pump can be driven by the controller If with the controller in start the input INTERLOCK is de activated the
12. SVARILO Pred katerikoli posegom izklopite iz controller a napajalni kabel Controller zavrzite kot industrijski odpadek v skladu z dr avnimi predpisi TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 191 301 15 Priro nik za Navodila Odlaganje opadkov 192 301 Pomen znamke WEEE na etiketah Spodaj navedeni simbol je v skladu z direktivo ES znano pod imenom WEEE Ta simbol ki velja samo v dr avah Evropske Skupnosti pomeni da izdelek NE SMETE ODSTRANITI skupaj z ostalimi komunalnimi ali pa industrijskimi odpadki temve morate poskrbeti za njihovo primerno lo evanje Zato pozivamo uporabnike da se ali pri prodajalnem centru ali pa pri prodajalcu seznani o postopku lo evanja in odstranitve odpadkov ele nato ko se je seznanil s pogoji in z merili kupopordajne pogodbe TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Obvestilo o napaki Priro amp nik za Navodila 15 V nekaterih slu ajih kvara poka e elektri no vezje avtodiagnoze controller a nekaj obvestil o napakah kot so opisane v spodnji tabeli Tab 1 OBVESTILO OPIS POPRAVA CHECK CONNECTION TO PUMP Napaka v priklju itvi rpalke in controller a Ali rpalka ima ni jo temperaturo od 0 C Preverite da je priklju amp ni kabel med amp rpalko in controller em na obeh koncih trdno pritrjen in da ni prekinjen Dvakrat pritisnite tipko START za ponoven zagon rpalke WAITING INTERLOCK
13. A szivatty h t ventill tor nak 24 Vdc fesz lts gen t rt n vez rl se a A vent valve vez rl se A szivatty sebess g nek leolvas sa a le ll t si parancsot k vet en le ll si sebess g leolvas sa A nyom s rt k leolvas sa Profibus illeszt opcion lis A bemeneti fesz lts g automatikus be ll t sa Az elk vetkez szakaszok tartalmazz k mindazokat az ismereteket melyek a kezel biztons g t hivatottak garant lni a berendez s haszn lata sor n R szletes inform ci k tal lhat k a Technical Information c m mell kletben 136 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Felhaszn l i K zik nyv 11 E k zik nyv az al bbi megegyez sszer jel l seket alkalmazza VESZ LY A vesz lyt jelz zenetek olyan elj r sra vagy specifikus tev kenys gre h vj k fel a kezel figyelm t melyek ha nem megfelel m don ker lnek v grehajt sra s lyos szem lyi s r l seket okozhatnak 31973831116 A figyelmeztet zenetek azon elj r sok el tt ker lnek megjelen t sre melyek figyelmen k v l hagy sa a berendez s k rosod s hoz vezethet A megjegyz sek fontos a sz vegb l kivonatolt inform ci kat tartalmaznak T rol s A vez rl k sz ll t sa s t rol sa sor n az al bbi k rnyezeti felt teleket kell biztos tani a h m rs klet 20 C s 70 C k z tt relativ nedvess gtartalom 0
14. 98 301 Istruktionsbog MEDDELELSE BESKRIVELSE KONTROL TOO HIGH Under normal drift efter Kontroll r om pumpens rotor LOAD startfasen forbruger pumpen en kan rotere frit st rre effekt end den Tryk to gange p START for at programmerede v rdi starte pumpen igen SHORT Under normal drift efter Kontroll r forbindelserne CIRCUIT startfasen er udgangseffekten mellem pumpe og styreenhed kortsluttet Tryk to gange p START for at Str mmen som optages af starte pumpen igen pumpen overskrider den programmerede v rdi SYSTEM Pumpen er blevet standset af et Frakobl styreenhedens OVERRIDE n dstopsignal fra en fjernkontakt str mkabel og kontroll r rsagen til n dstoppet S t derefter str mkablet i Tryk to gange p START for at starte pumpen igen OVERVOLTAGE Derer opst et fejl i styreenhedens Tryk to gange p START f desp nding eller styreenheden knappen for at starte pumpen har f et et falsk signal igen Hvis meddelelsen kommer igen tag kontakt med Agilent for n dvendig vedligeholdelse POWER FAIL Der er opst et fejl i sektionen for Tag kontakt med Agilent for str mtilf rsel til pumpen styreenheden Vedligeholdelse TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 8 Bruksanvisning Allman Information 100 F rvaring 101 Forberedelser f r installation 102 Installation 103 Anv ndning 104 Indikatorer och Konta
15. Technical Information 17 lt STX gt Start of transmission 0x02 lt ADDR gt Unit address 0x80 for RS 232 lt ADDR gt Unit address 0x80 device number 0 to 31 for RS 485 lt WIN gt Window a string of 3 numeric character indicating the window number from 000 to 999 for the meaning of each window see the relevant paragraph lt COM gt Command 0x30 to read the window 0x31 to write into the window lt DATA gt an alphanumeric ASCII string with the data to be written into the window In case of a reading command this field is not present The field length is variable according to the data type as per the following table Tab 13 Data Type Field Length Valid Characters Logic L 1 0 OFF V Z ON Numeric N 6 O Y right justified with 0 Alphanumeric A 10 from blank to ASCII lt ETX gt End of transmission 0x03 lt CRC gt XOR of all characters subseguent to lt STX gt and including the lt ETX gt terminator The value is hexadecimal coded and indicated by two ASCII character The addressed SLAVE will respond with an ANSWER whose structure depends from the MESSAGE type When the MESSAGE is a reading command the SLAVE will respond transmitting a string with the same structure of the MESSAGE TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 239 301 17 Technical Information Using the RS 485 interface the message structure rema
16. Toll Free 800 882 7426 Option 3 vpl raQagilent com EUROPE Toll Free 00 800 234 234 00 vpt customercare agilent com PACIFIC RIM please visit our website for individual office information http www agilent com Worldwide Web Site Catalog and Order On line www agilent com Representative in most countries 1 12
17. Ohjek sikirja Valmistelut asennusta varten Valvoja toimitetaan erityisess suojaavassa pakkauksessa Mik li havaitsette mahdollisesti kuljetuksen aikana tapahtuneita vaurioita ottakaa yhteys paikalliseen myyntitoimistoon Pakkauksen purkamisen yhteydess huolehtikaa ett valvoja ei p se putoamaan ja v ltt k sen joutumista iskujen kohteeksi lk j tt k pakkausta ymp rist n Materiaali voidaan kokonaisuudessaan kierr tt ja se vastaa EU n 85 399 direktiivi ymp rist n suojelusta Kuva 1 Valvojan pakkaus 126 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Asennus Ohjek sikirja 10 VUL K ytt j n turvallisuuden takaamiseksi laite on tarkoitettu vain sis tiloissa k ytett v ksi ja sit pit sy tt 3 lla johtimella varustetulla s hk johdolla ks tilattavissa olevien vaihto osien taulukosta jonka tulee olla varustettu kansainv lisesti hyv ksytyll s hk t pselill S hk iskuvaaran v ltt miseksi ja annettujen EC m r ysten mukaisesti k yt aina t m nlaista s hk johtoa ja aseta t pseli asianmukaisesti maadoitettuun s hk pisteeseen Valvojan sis ll syntyy korkeaj nnitett joka voi aiheuttaa vakavia vammoja tai jopa kuoleman Ennen mink tahansa valvojan huolto tai asennustoimenpiteen suorittamista irrottakaa valvoja s hk verkosta Valvoja voidaan asentaa p yd n p lle tai siihen tarkoitukseen sopivan hyllyn
18. 100 C See following chart for more details 9 1 Freguency Analog output proportional to Optically isolated Analog Output freguency linear OV 0Hz 10V 1010Hz Rload min 10kr Ripple Vpp max 200mv Precision 1 Resolution 40mv Pin 1negative Pin 9 positive TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 219 301 17 Technical Information R1 R2 Output Diagrams F Hz 963 t sec Mask Time Figure 14 Diagram 1 R1 or R2 related to frequency Mask Time Figure 15 Diagram 2 R1 or R2 related to power 220 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 V Volt R1 R2 24 Command Mask Time and Hyst ignored 0 KZ Value t sec Start Figure 16 Diagram 3 R1 or R2 related to time V Volt R1 R2 Stop i Waiting Ramp Auto Interlock tuning Braking i Normal i Fail t sec Figure 17 Diagram 4 R1 or R2 related to Normal status Pressure Only if gauge 4 Mask Time CAMAS SE O Threshold Value 4 Threshold gt 0 t sec V Volt A 24 t sec Figure 18 Diagram 5 R1 or R2 related to pressure reading if gauge connected TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 221 301 17 Technical Information Programmable Analog Output Diagrams lt a V
19. TO PUMP sszek ttetesi hiba a szivatty s a vez rl k z tt vagy a szivatty h m rs klete 0 C alatt van Ellen rizni hogy a szivatty s a vezerl k zti sszek t k bel mindk t v g n j l van e r gzitve s nincs e megszakadva A szivattyu jraindit s hoz k tszer megnyomni a START nyom gombot WAITING INTERLOCK A P1 csatlakoz reteszel s jelen jele akt v a J1 csatlakoz 3 s 8 pin je k zti r vidz r megszakad sa vagy a k ls reteszel jel nyitott llapota k vetkezteben Vissza llitani a P1 csatlakoz 3 s 8 pin je k zti r vidz rat vagy z rni a k ls reteszel jelet PUMP OVERTEMP A fels csap gy vagy a szivatty h m rs klete meghaladta a 60 C t V rakozni m g a h m rs klet a k sz b rt k al esik A szivatty jraind t s hoz k tszer megnyomni a START nyom gombot CONTROLLER OVERTEMP A vez rl nak h m rs klete t ll pte a 70 C t vagy a vez rl h t s nek h m rs klete t ll pte a 60 C t V rakozni m g a h m rs klet a k sz b rt k al esik A szivatty jraind t s hoz k tszer megnyomni a START nyom gombot TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 145 301 11 146 301 Felhaszn l i K zik nyv ZENET LE R S JAV T SI TEV KENYS G TOO HIGH A szivatty ltal felvett ram Ellen rizni hogy a szivatty LOAD m
20. WEEE Tento symbol platn iba pre t ty Eur pskej nie znamen e v robok s t mto t tkom NESMIE by odstr nen spolu s be n m dom cim alebo priemyseln m odpadom ale sa mus odstr ni ako delen odpad Vyz vame preto kone n ho u vate a aby sa skontaktoval s dod vate om pr stroja i u je to v robca alebo predajca za elom jeho likvid cie pod a zmluvn ch podmienok predaja 180 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Oznamy v d N vod na Obsluhu 14 V niektor ch pripadoch tykaj cich sa v d obvod samodiagnostiky controller signalizuje niekolko oznamov v d uveden ch na al ej strane Tab 1 OZNAM POPIS Z SAH OPRAVY CHECK CONNECTION TO PUMP Nefunguje napojenie medzi erpadlom a controller Alebo erpadlo m nizsiu teplotu ako O C Skontrolujte i je spojovac k bel medzi erpadlom a controller spr vne zasunuty na obidvoch koncoch a i nie je preru en Stla te dvakr t tla idlo START pre znovuna tartovanie erpadla WAITING INTERLOCK je aktivny sign l interlock na spojke P1 z d vodu preru enia skratu medzi pin 3 a pin 8 spojky J1 alebo z d vodu otvorenia sign lu vonkaj ieho interlock Obnovi skrat medzi pin 3 a pin 8 spojky PT alebo zavrie sign l vonkaj ieho interlock PUMP OVERTEMP Teplota horn ho lo iska alebo erpadla prekro ila 60 C Vy ka
21. a sa teplota op vr ti pod hranicu Stla te dvakr t tla idlo START pre znovuna tartovanie erpadla CONTROLLER OVERTEMP Teplota controller prekro ila 70 C Alebo Teplota radi tora controllera presahuje 60 C Po ka k m teplota neklesne pod hranicu Stla te dvakr t tla idlo START pre znovuna tartovanie erpadla TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 181 301 14 N vod na Obsluhu 182 301 OZNAM POPIS Z SAH OPRAVY TOO HIGH Po as be n ho chodu po Skontrolujte i m rotor LOAD na tartovan erpadlom erpadla mo nost to it sa absorbovany pr d prevy uje ten volne Stla te dvakr t tla idlo naprogramovan START pre znovuna tartovanie erpadla SHORT Po as be n ho chodu po Skontrolujte napojenia medzi CIRCUIT na tartovan v chodzie napojenie erpadlom a controller je v skrate Stla te dvakr t tla idlo START pre znovuna tartovanie erpadla SYSTEM N dzov sign l vych dzaj ci z Vytiahnite Snuru z elektrickej OVERRIDE dialkov ho kontaktu zastavil siete controller a ostr te erpadlo d vod alarmu Zasu te nap jaciu Sn ru a stla te dvakr t tla idlo START pre znovuna tartovanie erpadla OVERVOLTAGE Objavila sa vada v sekcii nap jania Stla te dvakrat tla idlo START controller alebo controller pre znovuna tartovanie dostal falo n sign l erpadla Ke sa tento oznam obja
22. ch dielov s jednou medzin rodne schv lenou z str kou Aby sa predi lo elektrick mu z sahu a aby boli splnen po iadavky CE pou vajte st le tento nap jac k bel tak e zasuniete z str ku do elektrickej z suvky s pr slu n m uzemnen m Vo vn tri Controller sa vytv ra vysok nap tie ktor m e sp sobi v ne kody alebo smr Pred ak mko vek konom t kaj ceho sa in tal cie alebo dr by Controller nezabudnite vytiahnu z str ku z elektrickej z suvky Controller m e by in talovan na pracovnom stole alebo vo vn tri track V ka dom pr pade je potrebn aby chladiaci vzduch mohol vo ne kr i okolo pr stroja Controller nepou vajte a nein talujte v prostred vystavenom atmosferick m vplyvom d mr z sneh prachu agres vnym plynom a v prostred s vysok m rizikom po iaru Po as chodu je d le it aby boli re pektovan tieto podmienky prostredia teplota od 5 C do 45 C a relat vna vlhkos 0 95 nekondenzovana Pre ostatn napojenia a in tal ciu zvolite n ho pr slu enstva vi as Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 175 301 14 N vod na Obsluhu Pou itie V tomto odstavci su uveden hlavn operat vne postupy Pre ostatn detaily a pre postupy ktor sa vztahuj na napojenia alebo na zvolen pr slusenstvo postupujte podla asti Use
23. controller geillustreerd Voor meer informatie wordt verwezen naar het hoofdstuk Technical Information Beschrijving frontpaneel Figuur 2 Frontpaneel van de Controllers TwisTorr 304 FS AG Rack 1 Drukknop voor weergave op het display van de parameters cycle number cycle time en pump life 2 Drukknop voor selectie van de bedrijfswijze LOW SPEED Is alleen actief wanneer de bedrijfswijze op het frontpaneel wordt geselecteerd Door de knop een maal in te drukken draait de pomp met stand by snelheid Door de knop opnieuw in te drukken wordt LOW SPEED geinactiveerd 3 Drukknop voor weergave op het display van de parameters pump current pump temperature pump power en rotational speed Is altijd actief onafhankelijk van de gekozen bedrijfswijze Door drukknoppen 3 en 1 gelijktijdig gedurende ten minste 2 seconden in te drukken wordt een programma gestart waarmee enkele bedrijfsparameters kunnen worden geprogrammeerd 4 Drukknop voor het zenden van de sturingen START STOP RESET Is alleen actief wanneer de bedrijfswijze op het frontpaneel wordt geselecteerd Door de knop een maal in te drukken wordt de startfase geactiveerd door de knop opnieuw te bedienen wordt de pomp gestopt Als de pomp automatisch is gestopt ten gevolge van een storing moet deze knop een eerste maal worden bediend om de controller te resetten en een t
24. on indigue les principales proc dures op rationnelles Pour tous autres d tails et pour les proc dures concernant des connexions ou des l ments en option se reporter au paragraphe Use de l appendice Technical Information Avant d utiliser le contr leur effectuer toutes les connexions lectriques et pneumatiques et se r f rer la notice de la pompe connect e AVERTISSEMENT Pour viter tous dommages aux personnes et I appareil si la pompe est plac e sur un plateau d appui s assurer gue ce dernier est stable Ne jamais faire fonctionner la pompe si la bride d entr e n est pas connect e au syst me ou n est pas ferm e l aide de la bride de fermeture Laisser le connecteur de r enclenchement J1 connect sa barrette s il n est proc d aucune connexion ext rieure La pompe pr vide et la pompe TwisTorr 304 FS peuvent tre mises en marche simultan ment 44 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Utilisation 3 Commandes Indicateurs et Connecteurs On pr sente ci dessous le tableau de commande du Contr leur ainsi que les tableaux d interconnexion Pour de plus amples d tails se reporter la section Technical Information Description du Tableau avant kJ O TwisTorr 304 FS AG Figure 2 Tableau avant du Contr
25. sis n Joka tapauksessa huolehtikaa siit ett riitt v j hdytysilma p see vapaasti kiert m n laitteen sis ll lk asentako ja tai k ytt k valvojaa tiloissa joissa se joutuu alttiiksi ymp rist tekij ille sade j lumi p lylle sy vytt ville kaasuille r j hdysalttiissa ymp rist ss tai tiloissa joissa paloriski on suuri Toiminnan aikana tulee noudattaa seuraavia ymp rist olosuhteita koskevia s nt j a l mp tila 5 C ja 45 C v lill suhteellinen kosteus 0 95 v lill ilman lauhdetta Muiden kytkent jen ja valinnaisten lis laitteiden asennusten suorittamiseksi katsokaa kappaletta Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 127 301 10 Ohjekasikirja K ytt T h n kappaleeseen on kirjattu t rkeimm t k ytt toimenpiteet Tarkempia lis tietoja sek kytkent j ja valinnaisia lis laitteita koskevien toimenpiteiden suorittamista k sittelevi tietoja l yd tte kappaleesta K ytt joka on Tekniset tiedot kappaleen liitteen Ennen valvojan k ytt suorittakaa kaikki s hk kytkenn t seuraten kytkett v n pumpun k ytt ohjeita VAROITUS Mik li pumppu on asetettu p yd lle varmistakaa ett se on vakaa N in v ltyt n vammoilta ihmisille sek itse koneelle lk my sk n k ytt k pumppua mik li sis ntulon laippaa ei ole kytketty j rjestelm n tai mik li sit ei ol
26. t Emballasjen m ikke kastes p en ulovlig m te Alle materialer er 100 resirkulerbare og er i samsvar med EU direktiv 85 399 om milj beskyttelse Figur 1 Styreenhetens emballasje 114 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instruksjon Manual 9 Installasjon For brukerens sikkerhet er styrenhetenTurbo V konstruert kun for bruk innend rs og skal forsynes med str m ved hjelp av en stramkabel med 3 ledninger se tabellen med reservedelene som du kan bestille som har et stopsel som er godkjent p internasjonal niv For unng risikoen for elektriske st ter og oppfylle CE kravene skal du alltid bruke denne stromkabelen og kople den til et uttak med en korrekt jordekopling Spenningen inne i styreenheten kan n hoye verdier og kan f re til alvorlige skader og dodsfall Kople alltid stromkabelen fra stromnettet for alle installasjons eller vedlikeholdsarbeider som utfores p styreenheten Styreenheten kan installeres p et bord eller inne i et passende stativ Uansett s m kj leluften kunne sirkulere fritt rundt apparatet Ikke installer eller bruk styreenheten i milj er som utsettes for regn sn eller is st v aggressive gasser eksplosjonsfarlige milj er eller i milj er med stor brannfare Under bruk m f lgende forhold respekteres temperatur fra 5 C til 45 C relativ fuktighet 0 95 uten kondens N r det gjelder andre tilkoplinge
27. valvomia kiinteist materiaaleista tehtyj taajuudenmuuntimia jotka kykenev t itsem rittelyyn ja itsesuojaukseen Valvojan ominaisuudet K ytett vyys etupaneelista et k ytt sarjak ytt Pumpun j hdytyssiivikon 24V tasavirtaohjaus Vent valve ohjaus Pumpun nopeuden luku pys ytysk skyn j lkeen pys ytysnopeuden luku Paineen luku Profibus liit nt valinnainen Automaattinen sis ntulon j nnitteen asetus Seuraavilla sivuilla on luettavissa tarpeelliset tiedot laitteen k ytt j n turvallisuuden takaamiseksi laitteen k yt n aikana Yksityiskohtaiset tiedot l ytyv t liitteest Technical Information 124 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Ohjek sikirja 10 T m ohjekirja k ytt seuraavia merkint j ACTUS Vaara merkinn t saavat k ytt j n huomion kiinnittym n erityisiin toimintotapoihin joiden seuraamatta j tt minen voi aiheuttaa vakavia henkil vaurioita Huomio merkinn t varoittavat toiminnoista joiden laiminly nti voi johtaa laitteen vaurioitumiseen Huomiot sis lt v t t rke tekstist otettua tietoa Varastointi Valvojan kuljetuksen ja varastoinnin aikana tulevat seuraavat ymp rist vaatimukset olla t ytettyin a l mp tila 20 C ja 70 C asteen v lill suhteellinen kosteus 0 95 ilman lauhdetta TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 125 301 10
28. 1 N2 B2 Interlock Type 110 0 continuos the interlock must be always present to run the pump 1 at start the interlock must be present only at start command B3 Setpoint R1 Logic 104 0 active close NO 1 active open NC Ba Setpoint R2 Logic 174 0 active close NO 1 active open NC Bs 6 Pressure Unit 163 0 mbar 1 Pascal 2 Torr B7 Reserved 21 1 Bo 2 Setpoint R1 Type 101 0 frequency 1 current 2 time 3 normal operation 4 pressure B3 5 Setpoint R2 Type 171 0 frequency 1 current 2 time 3 normal operation 4 pressure Bs 7 Analog Output 1 Type 111 0 frequency 1 power 2 temperature 3 pressure 284 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Output Buffer Tab 19 Output buffer s za Unit Res Range Description si 0 1 B lt Start Stop 000 0 lt Stop 1 lt Start Bs Low Speed 001 This bit set the pump target speed to low speed value Output Buffer Offset 3 otherwise the target speed is set to high speed value Output Buffer Offset 1 0 0ff 1 On Bs Soft Start 100 This bit activates the soft start function which run up the pump smoothly Please activate this function only if the pump has been stopped for at least3 months 0 On 1 Off B4 Vent Valve Command 122 0 Close 1 Open Valid only if in Par Buff Offset 20 Bo 1 B3 Pu
29. 16 1 10 10 Setpoint 1 Pressure Value Exponent 162 Set the exponent of the exponential value of the R1 setpoint format X XE XX when it is used for pressure monitoring Parameter Buffer Offset 21 Bo 2 4 17 1 mBar Torr P 0 99 Setpoint 2 Pressure Value Mantissa 176 a Set the mantissa of the exponential value of the R2 setpoint format X XE XX when it is used for pressure monitoring Parameter Buffer Offset 21 B3 5 4 The value set here is divided by 10 that means 11 1 1 18 1 10 10 Setpoint 2 Pressure Value Exponent 176 Set the exponent of the exponential value of the R2 setpoint format X XE XX when it is used for pressure monitoring Parameter Buffer Offset 21 B35 4 19 1 Bo Stop Speed Reading 167 This function measures the pump speed after a stop command 0 Off 1 0n B1 Active Stop 107 This function brakes the pump after a stop command by using the motor as a generator 0 Off 1 0n B2 Reserved B34 Start Stop Interlock Mode 151 Change the logic of Start Stop and Interlock inputs Ask to Agilent personnel for this option TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 283 301 17 Technical Information 20 1 Bo Vent amp Purge Auto Man 125 0 Auto the controller drives purge and vent valves automatically 1 Manual the user can command the valves by serial line or profibus Bi Gas Load Type Select here the gas load 157 0 Ar
30. 28 Lagerung 29 Vor der Installation 30 Installation 31 Gebrauch 32 Steuerungen Anzeigen und Verbinder des Controllers 32 Fronttafel der Controller 33 Bedienung 35 Einschalten des Controllers 35 Pumpenstart 35 Pumpenstopp 35 Wartung 35 Entsorgung 36 Fehlermeldungen 37 bersetzung der Originalanleitungen pp Agilent Technologies 27 301 2 Gebrauchsanleitung Allgemeines 28 301 Dieser Apparat ist f r Fachbetriebe bestimmt Vor Gebrauch sollte der Benutzer dieses Handbuch sowie alle weiteren mitgelieferten Zusatzdokumentationen genau lesen Bei Nichtbeachtung auch teilweise der enthaltenen Hinweise unsachgem em Gebrauch durch ungeschultes Personal nicht autorisierten Eingriffen und MiBachtung der einheimischen hier zur Geltung kommenden Bestimmungen ubernimmt die Firma Agilent keinerlei Haftung Die Controller der Serie TwisTorr 304 FS AG Rack sind mikroprozessorgesteuerte Freguenzwandler Sie sind mit Festk rperbauteilen gefertigt und verf gen uber ein Selbstdiagnose und ein Selbstschutzsystem Controllereigenschaften Konsol Fern und serielle Operationen 24V K hlventilator Treiber Vent Valve Treiber Auslesen der Pumpgeschwindigkeit nach Stoppbefehl Lesen der Stoppgeschwindigkeit a Auslesen des Drucks Profibus Schnittstelle Option a automatische Einstellung der Eingangsspannung In den folgenden Abschnitten sind alle erforderlichen Informationen f r die Sicherheit des
31. 301 The following screen page is displayed on reaching the pre established speed NJOJRIM JAJL 9 0 0 H z F R O N T The rotation value set in both NORMAL and LOW SPEED mode can be modified by the user as described in the Programming section of this manual a Press the MEASURES push button the display shows P 5 0 W 9 6 3 H z T 4 5 2 C o O VACUUM TECHNOLOGIES ENON TwisTorr 304 FS AG O O Figure 34 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Where P is the DC power drawn by the pump range 0 to 999 Watt Hz is the rotational speed of the pump as a function of the controller output freguency C is the temperature of the outer ring of the upper bearing range 00 to 99 C Pressing MEASURES key twice the following screen page is displayed NJO RIMJAJL 96 3 H z L j 1 2 0 A V 4 2 V TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 259 301 17 Technical Information Where Iis the current supplied by the controller to the pump at that precise moment V is the voltage at which the controller is powering the pump at that moment P
32. 304 FS AG Rack Controller User Manual Istruzioni per l uso Informazioni Generali 16 Immagazzinamento 17 Preparazione per l installazione 18 Installazione 18 Uso 20 Comandi Indicatori e Connettori 21 Descrizione pannello frontale 21 Procedure di uso 21 Accensione del Controller 23 Avvio della Pompa 23 Arresto della Pompa 23 Manutenzione 23 Smaltimento 23 Messaggi di errore 25 Traduzione delle istruzioni originali pp Agilent Technologies 15 301 1 Istruzioni per uso Informazioni Generali 16 301 Ouesta apparecchiatura destinata ad uso professionale L utilizzatore deve leggere attentamente il presente manuale di istruzioni ed ogni altra informazione addizionale fornita dalla Agilent prima dell utilizzo dell apparecchiatura La Agilent si ritiene sollevata da eventuali responsabilita dovute all inosservanza totale o parziale delle istruzioni ad uso improprio da parte di personale non addestrato ad interventi non autorizzati o ad uso contrario alle normative nazionali specifiche I controller della serie TwisTorr 304 FS AG Rack sono dei convertitori di freguenza controllati da un microprocessore realizzati con componenti a stato solido e con capacit di autodiagnostica e autoprotezione Caratteristiche del controller Operativit frontale remota seriale a Pilotaggio a 24 Vdc della ventola di raffreddamento pompa Pilotaggio vent valve a Lettura velocit pompa in seguito al commando di arresto l
33. 304 FS AG Rack Controller User Manual 87 901 029 01 35 301 2 Gebrauchsanleitung Entsorgung 36 301 Bedeutung des WEEE Logos auf den Etiketten Das folgende Symbol ist in bereinstimmung mit der EU Richtlinie WEEE Waste Electrical and Electronic Equipment angebracht Dieses Symbol nur in den EU L ndern giiltig zeigt an dass das betreffende Produkt nicht zusammen mit Haushaltsm ll entsorgt werden darf sondern einem speziellen Sammelsystem zugef hrt werden muss Der Endabnehmer sollte daher den Lieferanten des Ger ts d h die Muttergesellschaft oder den Wiederverk ufer kontaktieren um den Entsorgungsprozess zu starten nachdem er die Verkaufsbedingungen geprift hat TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Fehlermeldungen Gebrauchsanleitung 2 In einigen St rungsf llen zeigt das Selbstdiagnosesystem des Controllers die in der nachstehenden Tabelle zusammengefa ten Meldungen an Tab 1 MELDUNG BESCHREIBUNG BEHEBUNG CHECK CONNECTION TO PUMP Fehlfunktion der Pumpen Controller Verbindung Oder Die Temperatur der Pumpe ist unter 0 C Sicherstellen da das Verbindungskabel zwischen Pumpe und Controller an beiden Seiten korrekt befestigt ist und keine Unterbrechung vorliegt Die Pumpe durch zweimalige Bet tigung der START Taste neustarten WAITING INTERLOCK Das Interlock Signal auf dem Verbinder P1 ist wegen der Kurzschlu unterbrechung zwischen Pi
34. 95 nem lecsap d TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 137 301 11 Felhaszn l i K zik nyv A telep t sre val el k sz t s A vez rl speci lis v d csomagol sban ker l lesz ll t sra ha ez k rosod s jeleit mutatja melyek a sz ll t s sor n keletkezhettek kapcsolatba kell l pni a helyi forgalmaz irod val A kicsomagol s sor n k l nleges figyelmet kell ford tani arra hogy a vez rl le ne ess k s ne szenvedjen t d seket A csomagol anyagot nem szabad sz tsz rni a k rnyezetben Az anyag teljes m rt kben jrafelhaszn lhat s megfelel a k rnyezetv delemmel kapcsolatos EGK 85 399 es ir nyelvnek bra 1 sz m vez rlok csomagol sa 138 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Felhaszn l i K zik nyv 11 Telepit s A kiz r lag belt ri haszn latra tervezett vez rl t pl l s hoz a felhaszn l biztons ga rdek ben nemzetk zileg elfogadott t pus csatlakoz dug val ell tott h rom er vezet ket kell haszn lni l sd a rendelhet cserealkatr szek t bl zat t Az ram t sek elker l se s a EK k vetelm nyeinek a kiel g t se rdek ben mindig ilyen k belt kell haszn lni a csatlakoz dug t egy megfelel f ldel s csatlakoz aljzatba dugva A vez rl belsej ben nagyfesz lts gek alakulnak ki melyek s lyos k rokat vagy hal lt okozhatnak A vez rl b rmely te
35. AG Rack Controller User Manual 11 Felhaszn l i K zik nyv ltal nos inform ci k 136 T rol s 137 A telep t sre val el k sz t s 138 Telep t s 139 Haszn lat 140 A vez rl parancsai kijelz i s csatlakoz i 141 Az el lap le r sa 141 Haszn lati elj r sok 143 A vez rl bekapcsol sa 143 A szivatty beind t sa 143 A szivatty le ll t sa 143 Karbantart s 143 Megsemmis t s 144 Hiba zenetek 145 Az eredeti utas t s ford t sa Rem Agilent Technologies 135 301 11 Felhaszn l i K zik nyv ltal nos inform ci k Ez a berendez s hivat sos haszn latra tervezett A felhaszn l nak a berendez s haszn lat t megel z en gondosan el kell olvasnia e felhaszn l i k zik nyvet s minden egy b a Agilent ltal szolg ltatott inform ci t A Agilent elh r t mag t l minden esetleges felel ss get mely a felhaszn l i utas t sok teljes vagy r szleges figyelmen k v l hagy s hoz a nem kik pzett szem lyzet ltal t rt n nem megfelel haszn lathoz meghatalmaz s n lk li beavatkoz sokhoz vagy a specifikus nemzeti ir nyelvekkel ellent tes haszn lathoz k t dik A TwisTorr 304 FS AG Rack sorozat vez rl k mikroprocesszoros vez rl s szil rdtest elemekkel realiz lt frekvencia talak t k ndiagnosztikai s saj t v delmi k pess gekkel A vez rl tulajdons gai El lapr l t rt n t v soros vonalon kereszt li m k dtet s
36. Bedieners bei der Anwendung des Ger ts aufgef hrt Detaillierte technische Informationen sind im Anhang Technical Information enthalten TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Gebrauchsanleitung 2 In dieser Gebrauchsanleitung werden Sicherheitshinweise folgenderma en hervorgehoben Die Warnhinweise lenken die Aufmerksamkeit des Bedieners auf eine bestimmte Prozedur oder Praktik die bei unkorrekter Ausf hrung schwere Verletzungen hervorrufen k nnen VORSICHT Die Vorsichtshinweise vor bestimmten Prozeduren machen den Bediener darauf aufmerksam da bei Nichteinhaltung Sch den an der Anlage entstehen k nnen Die Hinweise enthalten wichtige Informationen die aus dem Text hervorgehoben werden Lagerung Beim Transport und bei der Lagerung der Controller m ssen folgende klimatische Verh ltnisse eingehalten werden Temperatur von 20 C bis 70 C a Relative Luftfeuchtigkeit 0 95 nicht kondensierend TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 29 301 2 Gebrauchsanleitung Vor der Installation Der Controller wird mit einer speziellen Schutzverpackung geliefert Eventuelle Transportsch den m ssen der zust ndigen rtlichen Verkaufsstelle gemeldet werden Beim Auspacken vorsichtig vorgehen damit der Controller nicht f llt oder St fsen ausgesetzt wird Das Verpackungsmaterial mu korrekt entsorgt werden Es ist vollst ndig recyclebar und
37. Controller User Manual 87 901 029 01 Technical Information 17 Where PRESS OVER indicates that the pressure measured is above the maximum measurement threshold set 1000 mBar The unit of measurement of the pressure can be selected GAUGE menu Unit from mBar Pa torr Pressing 4 times returns to the following screen page NJO IR IMJA IL X X X X H z X X X X X X XXX X X Where NORMAL indicates the operating regime of the pump XXXHz see description Pressing once Measures key The operating mode of the controller is shown at the bottom in the center This may be FRONT front mode the controller accepts START STOP and LOW SPEED commands from the front panel REMOTE remote mode the controller accepts START STOP and LOW SPEED commands via the rear connector RS232 the controller accepts commands from the serial port according to the RS232 standard RS485 XX the controller accepts commands from the serial port in RS485 mode with address XX XX from 0 to 31 PROFIBUS YYY the controller accepts commands from the PROFIBUS port with address YYYY TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 253 301 17 Technical Information Black rectangles as shown below may be displayed at any time in the first and last box of the bottom line of the display opposite the LS AG R1 R2 captions The meaning of t
38. STOP SPEED READING E NO Flow 2 6 Figure 39 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 265 301 17 Technical Information 266 301 This menu contains all the parameters that have a direct effect on driving of the pump HIGH SPEED Sets in Hz the speed of rotation of the pump when the low speed function is not active LOW SPEED Sets in Hz the speed of rotation of the pump when the low speed function is active SOFT START enables YES or disables NO the soft start function This parameter is visible only if the controller is configured in FRONT mode EXTERNAL FAN CNF Makes it possible to select driving mode of the external cooling fan Possible options ALWAYS ON the cooling fan is always ON OFF LIMPOWER 10 seconds after the pump has reached NORMAL status the fan is switched off and the driving power of the pump is reduced to 50 W SERIAL the cooling fan is controlled directly via the serial port VENT VALVE MODE makes it possible to select driving of the vent valve AUTO TIME the valve is controlled by the controller MANUAL the valve is controlled by the serial port AUTO PUMP SPEED the valve is controlled by the controller its king into account the pump s deceleration VENT VALVE DELAY Sets the delay expressed in seconds between stopping of the pump and opening of the vent valve VENT VALVE OPEN TIME sets t
39. Toute operation doit tre effectu e par un personnel agr En cas de panne il est possible de s adresser au Service de r paration Agilent ou bien au Agilent advance exchange service gui permet d obtenir un contr leur r g n r la place du contr leur d traqu I ESA AEE Avant d effectuer toute operation sur le contr leur d brancher le cable d alimentation En cas de mise au rebut d un contr leur proc der son limination conform ment aux r glementations nationales en la mati re TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 47 301 3 Mode d emploi Mise au rebut 48 301 Signification du logo WEEE figurant sur les tiguettes Le symbole ci dessous est appliqu conform ment la directive CE nomm e WEEE Ce symbole unigue ment valide pour les pays de la Communaut euro p enne indique que le produit sur lequel il est appliqu NE doit PAS tre mis au rebut avec les ordures m na g res ou les d chets industriels ordinaires mais passer par un syst me de collecte s lective Apr s avoir v rifi les termes et conditions du contrat de vente utilisateur final est donc pri de contacter le fournisseur du dispositif maison m re ou revendeur pour mettre en oeuvre le processus de collecte et mise au rebut TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Messages d erreur Messages d erreur 3 Dans certains cas de panne l ensemble de circuits d aut
40. af pumpen Pumpen startes ved at aktivere interlock indgangen og trykke p START knappen p frontpanelet Stop af pumpen Pumpen stopper ved at trykke p STOP trykknappen p frontpanelet Vedligeholdelse Styreenhederne i TwisTorr 304 FS AG Rack serien beh ver ikke nogen vedligeholdelse Ethvert indgreb p pumpen skal foretages af autoriseret personale Hvis pumpen g r i stykker kan man benytte sig af Agilents reparations service eller Agilent udvekslingsservice hvor man kan f en repareret pumpe i bytte for den der er g et i stykker OKB Inden der foretages noget som helst indgreb p styreenheden skal str mmen f rst afbrydes Skrotning af pumpen skal foreg i overensstemmelse med det p g ldende lands s rlige love TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 95 301 7 Istruktionsbog Bortskaffelse 96 301 Betydningen af WEEE logoet p m rkaterne Nedenst ende symbol anvendes i overensstemmelse med det s kaldte EU direktiv WEEE Symbolet kun g ldende for EU landene viser at produktet som det sidder p IKKE m bortskaffes sammen med affald fra private husholdninger eller industriel affald men skal indleveres p en godkendt affaldsstation Vi opfordrer derfor slutbrugeren til at kontakte leverand ren af anordningen enten fabrikken eller en forhandler for igangs ttelse af afhentnings og bortskaffelsesprocessen efter n je at have kontrolleret betingelserne i salgsko
41. ales de da os que podrian haberse producido durante el transporte ponerse en contacto con la oficina de venta m s cercana Durante la operaci n de desembalaje prestar una atenci n especial a no dejar caer el controler y evitarle golpes No dispersar el embalaje en el medio ambiente El material es totalmente reciclable y cumple con la directiva CEE 85 399 para la preservaci n del medio ambiente Figura 1 Embalaje de los Controlers 54 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de istrucciones 4 Instalaci n El controlador est dise ado s lo para su uso en interiores y para mantener la seguridad del usuario debe ser alimentado mediante un cable de 3 conductores v tabla de las piezas de recambio solicitables con un tipo de clavija aprobado a nivel internacional Para evitar el riesgo de descargas el ctricas y cumplir con los reguisitos CE utilizar siempre este cable de alimentaci n conectando la clavija a una toma el ctrica dotada con una adecuada conexi n a tierra Dentro del controlador se desarrollan altas tensiones gue pueden causar graves da os o la muerte Antes de efectuar cualguier operaci n de instalaci n o mantenimiento del controlador desconectarlo del enchufe de alimentaci n EI controler puede instalarse en una mesa o dentro de un rack especifico En cualguier caso es necesario gue el aire de refrigeraci n pueda circular libremente alrededor del apar
42. eliminac o ap s a oportuna verificac o dos termos e condic es do contrato de venda TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Mensagens de erro Manual de Istru es 5 Em alguns casos de defeitos os circuitos de autodiagn stico do controller apresentam mensagens de erro relacionadas na tabela abaixo Tab 1 MENSAGEM DESCRI O AC O CORRECTIVA CHECK CONNECTION TO PUMP Mau funcionamento na liga o entre a bomba e o controller Ou A bomba possui temperatura inferior a O C Verificar se o cabo de liga o entre a bomba e o controller est bem fixado em ambas as extremidades e n o est interrompido Premer duas vezes o bot o START para reactivar a bomba WAITING INTERLOCK activado o sinal de interlock existente no conector P1 devido interrup o do curto circuito entre o pin 3 e 0 pin 8 do conector J1 ou devido abertura do sinal de interlock externo Restabelecer o curto circuito entreo pin3e pin 8 do conector P1 ou fechar o sinal de interlock externo PUMP OVERTEMP A temperatura do rolamento superior ou da bomba superou os 60 C Aguardar at gue a temperatura volte ao limite estabelecido Premer duas vezes o bot o START para reactivar a bomba CONTROLLER OVERTEMP A temperatura do controller superou os 70 C Ou A temperatura do radiador do controller superior a 60 C Aguardar ate gue a temperatura volt
43. er apent Tilbakestill kortslutningen mellom stift 3 og stift 8 pa kontakt P1 eller steng det eksterne lasesignalet PUMP OVERTEMP Temperaturen pa det vre lageret eller pumpen er over 60 C Vent til temperaturen synker under terskelverdien Trykk to ganger p knappen START for starte pumpen CONTROLLER OVERTEMP Temperaturen p styreenhetens er over 70 C Eller Temperaturen p radiatoren p styreenheten er over 60 C Vent til temperaturen synker under terskelverdien Trykk to ganger p knappen START for starte pumpen TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 121 301 9 Instruksjon Manual 122 301 MELDING BESKRIVELSE FORHOLDSREGEL TOO HIGH Den absorberte str mmen av Kontroller om pumpens rotor LOAD pumpen er st rre enn den som er kan rotere fritt programmert Trykk to ganger p knappen START for starte pumpen SHORT Ved normal drift etter startmo Kontroller tilkoplingene CIRCUIT mentet er utgangen kortsluttet mellom pumpe og styreenhet Trykk to ganger p knappen START for starte pumpen SYSTEM Pumpen har stanset av et Kople fra styreenhetens OVERRIDE ngdstoppsignal en fra fjernkontakt str mkabel og finn frem til rsaken til nedstoppen Kople deretter maskinen til stramnettet igjen Trykk to ganger p knappen START for starte pumpen OVERVOLTAGE Det har oppst tt en feil i Trykk to ganger p knappen styreenhetens mated
44. externa A bomba de pr6 v cuo e a bomba TwisTorr 304 FS podem ser ligadas simultaneamente 68 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de Istru es 5 Comandos Indicadores e Conectores A seguir est o ilustrados o painel de comando do Controller e os pain is de interconex o Para maiores detalhes consultar a secc o Technical Information Descric o painel frontal Figura 2 Painel frontal do Controller TwisTorr 304 FS AG Rack 1 Bot o para evidenciar no display os par metros cycle number cycle time e pump life 2 Bot o para a selec o do modo LOW SPEED E activo somente guando for seleccionado o modo de comando do painel frontal Apertando uma vez a bomba roda com velocidade stand by Apertando o novamente desactiva se o modo LOW SPEED 3 Bot o para evidenciar no display os par metros pump current pump temperature pump power e rotational speed E sempre activo independentemente do modo de funcionamento escolhido Apertando os bot es 3 e 1 em conjunto pelo menos 2 segundos 6 activado um programa com o qual poss vel programar alguns par metros operativos 4 Bot o para enviar os comandos de START STOP RESET activo somente quando for selecionado o modo de comando do painel frontal Apertando o uma vez activa se a fase de arranque apertando o novamente p ra a bomba
45. option Seriell kontakt f r fj rrkontroll o M o aj ek w Pumpkontakt 106 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bruksanvisning 8 Instruktioner f r bruk Start av styrenheten Styrenheten startas enkelt genom att s tta str mkabeln i v gguttaget Start av pumpen Vid start av pumpen m ste ing ngen till interlock aktiveras och START knappen tryckas in p frontpanelen Stopp av pumpen Pumpen stoppas genom att trycka p tryckknappen STOPP p frontpanelen Underh ll Styrenheterna i TwisTorr 304 FS AG Rack serien r under h llsfria Allt servicearbete m ste utf ras av auktoriserad personal Om styrenheten havererar kontakta Agilent reparationsverkstad eller Agilent utbytesservice som kan ers tta styrenheten med en renoverad styrenhet Innan n got arbete utf rs p styrenheten m ste dess str mf rs rjning brytas Skrotning av pumpen ska ske enligt g llande lagstiftning TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 107 301 8 Bruksanvisning Bortskaffning 108 301 Betydelse av logotypen WEEE p etiketterna Symbolen som visas nedan har till mpats i enlighet med CD direktivet som har betecknats som WEEE Den h r symbolen g ller endast i de l nder som tillh r den Europeiska Unionen indikerar att produkten p vilken symbolen har applicerats INTE f r skaffas bort tillsammans med vanliga hush lls eller ind
46. pannello frontale Manutenzione I controller della serie TwisTorr 304 FS AG Rack non richiedono alcuna manutenzione Qualsiasi intervento deve essere eseguito da personale autorizzato In caso di guasto possibile usufruire del servizio di riparazione Agilent o del Agilent advanced exchange service che permette di ottenere un controller rigenerato in sostituzione di quello guasto VILIN Prima di effettuare qualsiasi intervento sul controller scollegare il cavo di alimentazione Qualora un controller dovesse essere rottamato procedere alla sua eliminazione nel rispetto delle normative nazionali specifiche TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 23 301 1 Istruzioni per uso Smaltimento 24 301 Significato del logo WEEE presente sulle etichette Il simbolo gui sotto riportato applicato in ottemperanza alla direttiva CE denominata WEEE Ouesto simbolo valido solo per i paesi della Comunit Europea indica che il prodotto sul quale applicato NON deve essere smaltito insieme ai comuni rifiuti domestici o industriali ma deve essere avviato ad un sistema di raccolta differenziata Si invita pertanto l utente finale a contattare il fornitore del dispositivo sia esso la casa madre o un rivenditore per avviare il processo di raccolta e smaltimento dopo opportuna verifica dei termini e condizioni contrattuali di vendita TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Me
47. pins before Imax 60mA measuring voltage Pin 5 negative Pin 12 positive 11 4 R1 Programmable Set Point Out Optically isolated Connect a 10kohm can be related to Frequency Vmax open 24v resistor between the Power Time Status or Vlow max 1 5v 2 pins before Pressure See following Imax 60mA measuring voltage figures for more details All Pin 5 negative the setting can be done by Pin 12 positive front panel or serial line win 101 to 105 and 162 14 7 R2 Programmable Set Point Out Optically isolated Connect a 10kohm Can be related to Freguency Power Time Status or Pressure See following figures for more details All the setting can be done by front panel or serial line win 171 to 176 Vmax open 24v Vlow max 1 5v Imax 60mA Pin negative Pin 14 positive resistor between the 2 pins before measuring voltage 218 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Pins Name Description Electrical Note 2 1 Programmable Programmable analog output Optically isolated Connect a 100kohm Analog Output Can be related to frequency Rload min 10kr resistor between the power pump temperature Ripple Vpp max 2 pins before linear or pressure 200mv measuring voltage logarithmic The setting can Precision 5 be done by front panel or Resolution 40mv serial line win 111 Pin 1negative OV 0Hz OW 0 C Pin 2 positive 10V 1010Hz 150W
48. pulsador START para volver a poner en marcha la bomba CONTROLLER OVERTEMP La temperatura del controler ha superado los 70 C O bien La temperatura del radiador del controler es superior a 60 C Esperar a gue la temperatura vuelva por debajo del umbral Apretar dos veces el pulsador START para volver a poner en marcha la bomba TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 61 301 4 Manual de istrucciones MENSAJE DESCRIPCI N ACCI N CORRECTIVA TOO HIGH LOAD La corriente absorbida por la bomba es superior a la programada Comprobar gue el rotor de la bomba tiene la posibilidad de girar libremente Apretar dos veces el pulsador START para volver a poner en marcha la bomba SHORT CIRCUIT Durante el funcionamiento normal tras la fase de puesta en marcha la conexi n de salida est en cortocircuito Comprobar las conexiones entre la bomba y el controler Apretar dos veces el pulsador START para volver a poner en marcha la bomba SYSTEM OVERRIDE La bomba ha sido parada por una se al de emergencia procedente de un contacto remoto Desenchufar el cable de alimentaci n del controler y corregir la causa de la emergencia Volver a conectar el cable de alimentaci n y apretar dos veces el pulsador START para volver a poner en marcha la bomba OVERVOLTAGE Se ha producido una averia en la secci n de alimentaci n del controler o el controler
49. r atk tessel jumper van ell tva 2 Csatlakoz a kimeneti logikai jelek a szivatty frekvencia ellen rz se s programozhat anal g kimenet sz m ra 3 Kimeneti t pcsatlakoz 24 Vdc a szivatty h t ventill tora a vent valve s az idomszer sz m ra 4 Bementei t pell t modul a vez rl sz m ra Mag ban foglalja a biztos t kokat a t pcsatlakoz t s az EMC sz r t Forgat gomb a Profibus c m nek be ll t s hoz Profibus opcion lis Soros csatlakoz t vvez rl shez oo Glo GI Szivatty csatlakoz 142 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Felhaszn l i K zik nyv 11 Haszn lati elj r sok A vez rl bekapcsol sa A vez rl bekapcsol s hoz elegend a h l zati k belt a h l zati aljzatba illeszteni A szivattyu beindit sa A szivatty beindit s hoz enged lyezni kell az interlock bemenetet s megnyomni az el lapon a START nyom gombot A szivatty le llit sa A szivatty le ll t s hoz meg kell nyomni az el lapon tal lhat STOP nyom gombot Karbantart s A TwisTorr 304 FS AG Rack sorozat vez rl k semmif le karbantart st sem ig nyelnek Valamennyi beavatkoz st meghatalmazott szem lyzet hajthat csak v gre Meghib sod s eset n ig nybe vehet a Agilent javit si szolg ltat sa vagy a Agilent advanced exchange service mely a meghib sodott helyett egy javitott vez rl
50. r pumpen tilsluttes ADVARSEL Sorg for at pumpen st r fast hvis den er installeret p et bord Dette er for at forebygge skader p apparatet og personer Start aldrig pumpen hvis pumpetillobet ikke er tilsluttet systemet eller er blokeret Afbryderkontakten J1 skal forblive tilsluttet med aktuel bro n r der ikke udf res eksterne tilslutninger Forvakuumpumpen og TwisTorr 304 FS pumpen skal fungere samtidigt 92 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruktionsbog 7 Indikatorer og Kontakter p Styreenheden Nedenfor vises styreenhedens front og bagpanel For yderligere information henvises til bilag Technical Information Beskrivelse af frontpanel TwisTorr 304 FS AG Figur 2 Frontpanelet p Styreenheden TwisTorr 304 FS AG Rack 1 Trykknap til at hente cycle number cycle time og pump lift frem p displayet 2 Trykknap til installering af LOW SPEED driftvalg Trykknappen fungerer kun n r der v lges fra frontpanelet Tryk p knappen en gang for at f pumpen til at rotere med standby hastighed Tryk p knappen igen for at g ud af LOW SPEED driften 3 Trykknap til at hente pump current pump temperature pump power og rotational speed frem p displayet Trykknappen fungerer altid uanset den valgte styretype N r trykknapperne 3 og 1 holdes trykket ind samtidigt i mindst 2 seku
51. t biztosit VESZ LY A vez rl n val b rmely beavatkoz s v grehajt sa el tt ki kell h zni a h l zati csatlakoz t Amikor egy vez rl t le kell selejtezni megsemmisitesekor a t rgyk rbe tartoz nemzeti norm knak megfelel en kell elj rni TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 143 301 11 Felhaszn l i K zik nyv Megsemmisites A c mk n jelenl v WEEE logo jelent se Az al bb l that szimb lum az EK WEEE elnevez s ir nyelv vel sszhangban ker l alkalmaz sra Ez a szimb lum mely csak az Eur pai K z ss g orsz gaiban rv nyes azt jelzi hogy a term k melyen megtal lhat NEM ker lhet k z ns ges h ztart si vagy ipari hullad kkal egy tt megsemmis t sre hanem azt egy szelekt v hullad kgy jt rendszerbe kell tov bb tani A v gfelhaszn l nak a gy jt si s megsemmis t si elj r s beind t sa c lj b l az ad sv teli szerz d s hat rid inek s felt teleinek alapos ttanulm nyoz sa ut n fel kell vennie a kapcsolatot a berendez s sz ll t j val legyen ez a gy rt vagy egy viszontelad 144 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Hibaiizenetek Felhaszn l i K zik nyv 11 N h ny meghib sod si esetben a vez rl ondiagnosztikai ramk rei hiba zenetet jelen tenek meg l sd az al bbi t bl zatot Tab 1 UZENET LEIR S JAV T SI TEVEKENYSEG CHECK CONNECTION
52. to 9999 command TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 247 301 17 Technical Information 302 R N Pump life in hours zeroed by the reset 0 to 999999 command 320 Reserved to Agilent service to 399 400 R A CRC EPROM QE QE8XXXX where XXXX are variable 402 R A CRC Param PA PA8XXXX where XXXX are variable 404 R A CRC Parameter structure XXXX 500 Reserved to Agilent service 503 R W N RS 485 address 0 to 31 default 0 504 R W L Serial type select 0 RS 232 1 RS 485 default 0 1 Automatic means that when the controller stops the vent valve is opened with a delay defined by window n 126 when the controller starts the vent valve is immediately closed 2 On command means that the vent valve is opened or closed by means of window n 122 248 301 TOO HIGH LOAD SHORT CIRCUIT NO CONNECTION PUMP OVERTEMP OVERVOLTAGE CONTROLL OVERTEMP AUX FAIL POWER FAIL Figure 29 Window N 206 Bit Description TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 How to Use by Front Panel VACUUM TECHNOLOGIES LJ OU TwisTorr 304 FS AG o Figure 30 Start Stop and Reset Fail button Press once for the S
53. v dodatku Technical Information Pred pou it m Controller preve te v etky potrebn elektrick a pneumatick napojenia a ria te sa podla n vodu napojen ho arpadla NEBEZPE E Aby ste sa vyhli osobn mu nebezpe iu a po kodeniu pristroja skontrolujte aby erpadlo umiestnen na pracovnom stole bolo stabiln erpadlo nikdy nena tartujte vtedy ke vstupn priruba nie je napojen na syst m alebo nie je zavret s uz verovou prirubou Uzavieraj ca pripojka J1 sa mus necha napojen na jej most k vtedy ke nie je preveden iadne vonkaj ie napojenie Pred pr zdnov erpadlo a erpadlo TwisTorr 304 FS AG Rack m u by zapnut s asne 176 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod na Obsluhu 14 Povely Ukazovatele a Pripojky Controller Dalej je zobrazen eln a zadny panel controller a Pre al ie detaily konzultuje as Technical Information Popis eln ho panelu TwisTorr 304 FS AG O O Postava 2 eln panel Controller TwisTorr 304 FS AG Rack 1 Tla idlo pre vyvolanie parametrov cycle number cycle time a pump life na displej 2 Tla idlo pre volbu LOW SPEED Je aktivne iba vtedy ke povel vych dza z eln ho panelu Ke sa stla raz cerpadlo sa to r chlos ou stand by Stla ia
54. wird mit einer Wiederverschlie br cke geliefert 2 Stecker f r logische Signale und Pumpenmonitor plus programmierbarer Analogausgang 3 Stromausgangsstecker 24 V f r Pumpenk hlventilator Vent Valve und Messinstrument 4 Netzeingangsmodul des Controllers bestehend aus Netzsicherungen Netzstecker und EMC Filter 5 Drehschalter zum Setzen der Profibus Adresse 6 Profibus Option 7 Serieller Stecker zur Fernsteuerung 8 Pumpenstecker 34 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Gebrauchsanleitung 2 Bedienung Einschalten des Controllers Zum Einschalten des Controllers gen gt es das Netzkabel an die Steckdose anzuschliefsen Pumpenstart Zum Starten der Pumpe den Verrieglungseingang best tigen und die START Taste of der vorderen Konsole des Controllers dr cken Pumpenstopp Zum Stoppen der Pumpe muf die STOP Taste an der Fronttafel gedr ckt werden Wartung Die Controller der Serie TwisTorr 304 FS AG Rack sind wartungsfrei Eventuell erforderliche Eingriffe m ssen von dazu befugtem Fachpersonal ausgef hrt werden Bei einem Defekt kann der Agilent Reparaturdienst bzw der Agilent advanced exchange service in Anspruch genommen werden der f r die Erneuerung defekter Controller sorgt WARNUNG Vor jedem Eingriff am Controller muB der Netzstecker gezogen werden Ein Controller muf unter Einhaltung der einschl gigen landes blichen Vorschriften erfolgen TwisTorr
55. 01 Bledne informacje Podrecznik Instrukcji 12 W niektorych przypadkach uszodzen obwod do autodiagnozy kontrolera pokazuje blendne informacje przyklady sa wymienionee w ponizszej tabeli Tab 1 INFORMACJA OPIS AKCJA KORYGUJACA CHECK CONNECTION TO PUMP Zaklocenia na polaczeniu miedzy pompa i kontrolerem Lub Temperatura pompy jest ponizej 0 C Zweryfikowac czy kabel laczacy pompe z kontrolerem jest dobrze zamocowany na obydwu koncach czy nie jest przerwany Nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy WAITING INTERLOCK Jest aktywny sygnal Interlock na laczniku P1 spowodowany zwarciem pomiedzy pin 3 pin 8 lacznika J1 Lub na skutek otwarcia sygnalu zewnetrznego interlock Dokonac naprawy obwodu elektrycznego pomiedzy pin 3 i pin 8 lacznika P1 lub zamknac sygnal interlock zewnetrzny PUMP OVERTEMP Temperatura gornego lozyska lub pompy przekroczyla 60 C Odczekac az temperatura sie obnizy ponizej ustalonego poziomu Nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy CONTROLLER OVERTEMP Temperatura w kontrolerze przekroczyla 70 C Lub Temperatura grzejnika nie przekracza 60 C Odczekac az temperatura sie obnizy ponizej ustalonego poziomu Nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 157 301 12 158 301 Podrecznik Ins
56. 1 16 Instructions for Use Preparation for Installation The controller is supplied in a special protective packing If this shows signs of damage which may have occurred during transport contact your local sales office When unpacking the controller ensure that it is not dropped or subjected to any form of impact Do not dispose of the packing materials in an unauthorized manner The material is 100 recyclable and complies with EEC Directive 85 399 Figure 1 Controller packing 198 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instructions for Use 16 Installation WARNING The controller is designed for indoor use only and must be powered with 3 wire power cord see orderable parts table and plug internationally approved for user s safety Use this power cord and plug in conjunction with a properly grounded power socket to avoid electrical shock and to satisfy CE reguirements High voltage developed in the controller can cause severe injury or death Before servicing the unit disconnect the input power cable TwisTorr 304 FS AG Rack controller can be used as a bench unit or a rack module but it must be positioned so that free air can flow through the holes Do not install or use the controller in an environment exposed to atmospheric agents rain snow ice dust aggressive gases or in explosive environments or those with a high fire risk During operation the following envi
57. 160 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod k Pou it 13 Tato p ru ka pou v n sleduj c ozna en 4 1743 HUMN Odkazy kter upozor uj na nebezpe a donut oper tora aby se dil dle procedury nebo specifick praktiky kter nebude li dn vykon na m e ohrozit v n m zp sobem jeho zdrav POZOR Tyto zpr vy b vaj vizualizov ny p ed jakoukoliv procedurou je t eba se jimi dit aby nedo lo k po kozen p stroje Pozn mky vyjmut z textu obsahuj velmi d le it informace Uskladn n B hem dopravy a skladov n kontrol r se mus db t na n sleduj c podm nky prost ed temperatura od 20 C do 70 C a relativn vlhkost 0 95 nesm kondenzovat TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 161 301 13 N vod k Pou it P prava k instalaci Kontrol r je dod v n v speci ln m ochrann m balen objev li se n sledkem dopravy n jak po kozen je t eba se okam it obr tit na m stn prodejn servis B hem rozbalov n se obzvl mus d vat pozor aby kontrol r nespadl nebo aby neutrp l n razy Balic materi l se nesm odkl dat v p rod Tento materi l se m e kompletn recyklovat nebo se shoduje se sm rnicemi EU CEE 85 399 kter se t kaj ivotn ho prost ed Postava 1 Obal kontrol ru 162
58. 3 2 mA Hz sec 0 10000 Setpoint 2 value Enter here the threshold value for the R2 output valid only if B3 Bs of Parameter Buffer Offset 21 is egual to 0 1 2 or 3 172 0 30000 Setpoint 2 Mask Time Set here the time measured from the start pump event while the R2 output is masked not activated 173 2 0 2sec bit 0 30000 Vent Valve Opening Delay Set the delay time between the stop command and vent opening Valid only in purge vent auto mode Parameter Buffer Offset 20 Bo 0 0 infinite the valve will never open 126 11 2 0 2sec bit 0 30000 Vent Valve Open Time Set how long the vent will remain open then both purge and vent will be closed Valid only in purge vent auto mode Parameter Buffer Offset 20 Bo 0 0 infinite the valve remain open till next start command 147 13 1 99 Setpoint 1 Hysteresis Set the Setpoint1 hysteresis around the nominal value 105 14 1 99 Setpoint 2 Hysteresis Set the Setpoint2 hysteresis around the nominal value 175 282 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 15 1 mBar Torr P 0 99 Setpoint 1 Pressure Value Mantissa 162 a Set the mantissa of the exponential value of the R1 setpoint format X XE XX when it is used for pressure monitoring Parameter Buffer Offset 21 Bo 2 4 The value set here is divided by 10 that means 11 1 1
59. 3 78 106 4 4 19 914 3 60 o o sia oj a se em 1 e NN increase SIN u o re PM ran TwisTorr 304 FS AG i o 5 Dimensions mm inches Figure 5 Controller outline 212 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Controller Connection J6 Fan The 24 Vdc 150 mA maximum output voltage is present after START pushbutton is pressed and will remain present until a fault condition is displayed on front panel display or the turbopump is stopped This is a dedicated output for the optional TwisTor 304 pump cooling fan see orderable parts table If you already have fan installed use the available adapter cable to connect to the pump J6 FAN PURGE COMMON COMMON PURGE FAN T24VDG 24VDC PURGE FAN Figure 6 Fan Connector TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 213 301 17 Technical Information J7 Vent J7 VENT COMMON ENABLE N C VALVE ENABLE N O VALVE 24VDC MAX 60mA Figure 7 Vent Connector This is dedicated 24 Vdc connector to control the optional vent valve see orderable parts table The vent valve can be driven by controller automatically or by serial line Vent Valve If you already have a vent valve installed 110 Vac powered this valve must be replaced by a new model
60. 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod k Pou it 13 Instalace Ohledn ochrany bezpe nosti u ivatele kontrol r je v hradn ur en k vnit n mu pou v n a mus b t nap jen kabelem s 3 konduktory viz tabulku n hradn ch d l kter Ize objednat s mezin rodn schv lenou z str kou Aby se p ede lo elektrick mu z sahu a aby byly spln ny po adavky CE dan nap jec kabel pou vejte tak e z str ku zasunete do elektrick z suvky s p slu n m uzemn n m Uvnit kontrol ru doch z k vysok mu nap t kter zp sobuje v n kody na zdrav i smrt P ed proveden m jak hokoli instala n ho i dr bov ho konu s kontrol rem je t eba nejprve p eru it elektrick p kon Kontrol r m e b t nainstalov n na stole nebo uvnit p slu n ho rack V ka d m p pad je t eba aby okolo p stroje mohl vzduch voln proudit Kontrol r se nesm instalovat v m stech vystaven ch atmosferick m vliv m d mr z sn h prachu agresivn m plyn m na v bu n ch m stech nebo na m stech kde existuje velk riziko po r B hem provozu je t eba aby se respektovaly n sleduj c podm nky t kaj c se operativn ho prost ed teplota od 5 C do 45 C a Relativn vlhkost 0 95 nesm kondenzovat Pro dal spojen a instalaci op amp niho p slu enstv je t eba se pod vat do sekce Tech
61. 5 ne kondezna Kar se ti e priklju kov in monta e opcionalnih dodatkov si lahko ogledate priponko Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 187 301 15 Priro nik za Navodila Uporaba SVARILO V tem odsatvku so navedeni vsi glavni operativni postopki Za dodatne detalje in postopke ki zadevajo priklju itve ali posamezne opcije si lahko ogledate odstavek Use v priponki Technical Information Predno bi uporabili controller morate priklju iti vse elektri ne in pnevmatske kable pri tem se nanaj ate na priro nik prilo ene rpalke Da bi se izognili po kodbam bodisi oseb kot naprave morate poskrbeti da ste rpalko stabilno polo ili e ste jo vgradili na mizi rpalka ne sme nikoli delovati e vhodna prirobnica ni priklopljena k sistemu ali e ni zaprta s zapiralno prirobnico Zapiralni konektor J1 mora ostati priklju en k svojemu mosti ku e niste izvr ili nikakr ne druge zunanje priklju itvije rpalko predhodnega praznjenja in rpalko TwisTorr 304 FS lahko isto asno vklopite 188 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Priro amp nik za Navodila 15 Komande indikatorji in konektorji Sledi prikaz prednje in ozadnje plo e controllera Za podrobnej e informacije si ogledajte paragraf Technical Information Opis prednje plo e 5 1 VACUUMTEC
62. 5 Fax 82 0 2 3452 3947 Toll free 080 222 2452 Mexico Agilent Technologies Vacuum Products Division Concepcion Beistegui No 109 Col Del Valle C P 03100 Mexico D F Tel 52 5 523 9465 Fax 52 5 523 9472 Singapore Agilent Technologies Singapore Pte Ltd Vacuum Products Division Agilent Technologies Building 1 Yishun Avenue 7 Singapore 768923 Tel 65 6215 8045 Fax 65 6754 0574 Agilent Technologies Inc 2012 Printed in ITALY 01 2013 Publication Number 87 901 029 01 Agilent Technologies Southeast Asia Agilent Technologies Sales Sdn Bhd Vacuum Products Division Unit 201 Level 2 uptown 2 2 Jalan SS21 37 Damansara Uptown 47400 Petaling Jaya Selangor Malaysia Tel 603 7712 6106 Fax 603 6733 8121 Taiwan Agilent Technologies Taiwan Limited Vacuum Products Division 3F 20 Kao Shuang Rd Pin Chen City 32450 Taoyuan Hsien Taiwan R 0 C Tel 886 34959281 Toll Free 0800 051 342 UK and Ireland Agilent Technologies UK Ltd Vacuum Products Division 6 Mead Road Oxford Industrial Park Yarnton Oxford OX5 10U UK Tel 44 0 1865 291570 Fax 44 0 1865 291571 Toll free 00 800 234 234 00 Other Countries Agilent Technologies Italia S p A Vacuum Products Division Via F lli Varian 54 10040 Leini Torino Italy Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Customer Support amp Service NORTH AMERICA
63. 81 Preklad origin lneho n vodu tle Agilent Technologies 171 301 14 N vod na Obsluhu V eobecn inform cie Tento pr stroj slu i k profesionalnemu pou itiu UZivatel si pred pou it m tohto pr stroja mus pozorne pre ta tento n vod k pou itiu a jeho al ie doplnky dodan podnikom Agilent Podnik Agilent nepreber na seba iadnu zodpovednost za pln alebo iasto n nedodr iavanie n vodu za nespr vne pou vanie pristroja nekompetentn mi osobami za nepovolen z sahy a za pou vanie ktor nie je v zhode s prislu nymi n rodn mi normami Controllers s rie TwisTorr 304 FS AG Rack s meni i kmito tu kontrolovan mikroprocesorom vyroben prvkami pevnej f zy a so schopnos opu samodiagnostiky a samokontroly Charakteristiky controller a eln dialkov s riov riadenie 24 Vdc riadenie chladiacej lopatky erpadla Riadenie vent valve tanie r chlosti erpadla po poveli zastavenia tanie r chlosti zastavenia tanie tlaku Medzi l nok Profibus na objedn vku Automatick nastavenie vstupn ho nap tia alej s pop san v etky potrebn inform cie ktor ch lohou je zabezpe i bezpe nos pracovn ka po as pou vania pr stroja Podrobn inform cie s uveden v dodatku Technical Information 172 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod na Obsluhu 14 Tento n vod pou va nasl
64. 901 029 01 85 301 6 Bedrijfshandleiding 86 301 BOODSCHAP OMSCHRIJVING REMEDIE TOO HIGH De stroomopname van de pomp is Controleer of de pomprotor vrij LOAD groter dan de geprogrammeerde kan draaien waarde Bedien twee maal de START knop om de pomp weer op te starten SHORT Tijdens normale werking na Controleer de verbindingen CIRCUIT startfase is de uitgangsverbinding tussen pomp en controller in kortsluiting Bedien twee maal de START knop om de pomp weer op te starten SYSTEM De pomp is stilgelegd door een Koppel de voedingskabel van OVERRIDE noodsignaal afkomstig van een de controller af en elimineer afstandscontact de oorzaak van de noodstop Sluit de voedingskabel weer aan en bedien twee maal de START knop om de pomp weer op te starten OVERVOLTAGE Eris een storing opgetreden in de Bedien twee maal de START voedingssectie van de controller of knop om de pomp weer op te de controller heeft een stoorsignaal starten Als de melding weer ontvangen verschijnt zich voor onderhoud tot Agilent wenden POWER FAIL Defect in het voedingsgedeelte van Neem contact op met Agilent de pomp van de controller voor onderhoud TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 7 Istruktionsbog Generel Information 88 Opbevaring 89 Forberedelser for installation 90 Installation 91 Anvendelse 92 Indikatorer og Kontakter p Styreenheden 93 Beskriv
65. Agilent nebo Agilent advanced exchange service kde po kozen kontrol r bude vym n n za nov A 174234180 Nejd ve p ed jak mkoliv z sahem na kontrol ru se mus odpojit elektrick kabel Bude li zapot eb kontrol r likvidovat mus se postupovat dle platn ch specifick ch st tn ch norem TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 167 301 13 N vod k Pou it Likvidace V znam loga WEEE nach zej c ho se na t tku N e uveden symbol odpov d sm rnic m CE pojmenovanym WEEE Tento symbol platn jen pro st ty Evropsk Unie ur uje e v robek kter je takto ozna en NESM b t likvidov n spole n s ostatn mi dom c mi nebo pr myslov mi odpady ale je nutno ho p edat do pat i n ch provozn ch sb ren kde mus b t t d n a likvidov n odd len od m stsk ho odpadu U ivateli kter m n v robek likvidovat doporu ujeme aby se nakontaktoval p mo s v robcem nebo prodejcem kter se po pat i n m prov en term n a smluvn ch podm nek postar o kompletn likvidaci uveden ho v robku 168 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Chybn zpr vy N vod k Pou it 13 V n kter ch p padn ch poruch ch na autodiagnostick ch obvodech kontrol r p ed v n kolik chybn ch odkaz kter jsou vyjmenov ny v tabulce kter se nach z na n sleduj c str nce Tab 1 ZPR VY POP
66. Agilent which is contaminated with a toxic or hazardous material that was not disclosed the customer will be held responsible for all costs incurred to ensure the safe handling of the product and is liable for any harm or injury to Agilent employees as well as to any third party occurring as a result of exposure to toxic or hazardous materials present in the product Print Name Authorized Signature sssssssssssssssssessse Date 5 FAILURE INFORMATION Failure Mode REOUIRED FIELD See next pag Detailed Description of Malfunction Please provide the error message Application system and model I understand and agree to the terms of Section 6 Page 3 3 Print Name Authorized Signature Date Pg 2 3 Agilent Technologies Vacuum Products Division Reguest for Return Form Health and Safety Certification Please use these Failure Mode to describe the concern about the product on Page 2 TURBO PUMPS and TURBO CONTROLLERS APPARENT DEFECT MALFUNCTION POSITION PARAMETERS Vacuum leak High voltage problem Solenoid failure Does not start Noise Vertical Power Rotational Speed Does not spin freely Vibrations Horizontal Current Inlet Pressure Does not reach full speed Leak Upside down Temp 1 Foreline Pressure Mechanical Contact Overtemperature Temp 2 Purge flow Cooling defective Clogging OPERATING TIME ION
67. Controller User Manual 87 901 029 01 119 301 9 Instruksjon Manual Eliminering 120 301 Betydelsen av symbolet p logo WEEE p etikettene Symbolet nedenunder som finnes er anvendt i henhold til EC direktiv kalt WEEE Dette symbolet som bare gjelder for land i Det europeiske fellesselskap viser at produktet som det sitter p IKKE m behandles som vanlig hus industriavfall men m legges i kildesortering Derfor oppfordrer man den sluttelige brukeren av anordningen henvende seg til leverandgren av anordningen som kan v re et firma eller en forhandler som s rger for oppsamling og eliminering etter ha kontrollert avtal og betingelser i kjopekontrakten TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Feilmeldinger Instruksjon Manual 9 N r det oppst r visse feil viser selvdiagnosekretsen i styreenheten den aktuelle feilmeldingen i displayet De aktuelle feilmeldingen fremg r av tabellen nedenfor Tab 1 MELDING BESKRIVELSE FORHOLDSREGEL CHECK CONNECTION TO PUMP Defekt kopling mellom pumpe og styreenhet Eller Pumpen har en temperatur under 0 C Kontroller at tilkoplingskabelen mellom pumpe og styreenhet er skikkelig montert samt at kabelen ikke er skadet Trykk to ganger pa knappen START for a starte pumpen WAITING INTERLOCK Lasesignalet for kontakt P1 skyldes en kortslutning mellom stift 3 og stift 8 i kontakten J1 eller fordi det eksterne lasesignalet
68. DDR WINDOW WR ON ETX CRC Source Controller Destination PC 02 80 06 03 38 35 STX ADDR ACK ETX CRC Command STOP Source PC Destination Controller 02 80 30 30 30 31 30 03 42 32 STX ADDR WINDOW WR OFF ETX CRC Source Controller Destination PC 02 80 06 03 38 35 STX ADDR ACK ETX CRC TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 241 301 17 Technical Information Command SOFT START ON Source PC Destination Controller 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Controller Destination PC 02 80 06 03 38 35 STX ADDR ACK ETX CRC Command SOFT START OFF Source PC Destination Controller 02 80 31 30 30 31 30 03 42 33 STX ADDR WINDOW WR OFF ETX CRC Source Controller Destination PC 02 80 06 03 38 35 STX ADDR ACK ETX CRC 242 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Command READ PUMP STATUS Source PC Destination Controller with address 3 Technical Information 17 02 83 32 30 35 30 03 38 37 STX ADDR WINDOW RD ETX CRC Source Controller with address 3 in stop status Destinatio
69. E GAUGE CABLE 1 m mA GAUGE GAUGE EXTENTION CABLE 5 m Figure 4 TwisTorr 304 FS AG Rack Controller Connections 210 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Controller Specifications Technical Information 17 Tab 2 Input Voltage 100 240 Vac voltage fluctuation 10 Freguency 50 to 60 Hz Power 450 VA Fuse 2 x T4 A slow blow 250 V Output Voltage 75 Vac Freguency 1010 Hz Power 150 W Environment storage 20 C to 70 C 0 95 Operating 5 45 C 0 90 Applied norr EN 61010 1 EN 55011 Class A EN 61000 3 2 EN 61000 3 3 EN 61000 6 2 EN 61000 4 2 3 4 5 6 11 8 Weight 1 7 Kg 3 2 Ibs Pollution Degree 2 Installation Category Acoustic noise 1807779 LpA lt 70 dB A normal operation operator position The Controller must be positioned in such way that the mains cable can be easily disconnected TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 211 301 17 Technical Information Controller Outline The outline dimensions for the TwisTorr 304 FS AG Rack Controllers are shown in the following figure TwisTorr 304 FS AG Rack controller can be used as a bench unit or as a rack module but it must be positioned so that free air can flow through the holes 9 4 0 37 I Lu 244 0 9 61 Sa 96 1
70. ETUKSEN k skyjen painonappi Se aktivoituu vain silloin kun etupaneelin toiminta moodi on valittu Yhden kerran painettaessa k ynnistysvaihe aktioituu Uudelleen painettaessa pumppu pys htyy Mik li pumppu on toimintah iri n vuoksi pys htynyt automaattisesti painakaa painonappia yhden kerran jolloin valvoja saadaan asetettua uudelleen Toisen kerran painonappia painettaessa pumppu k ynnistyy 5 Kirjainnumerollinen digitaalin ytt pistematriisi 2 rivi x 16 merkki TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 129 301 10 Ohjek sikirja El OA LJ D EI Kuva 3 Valvojien TwisTorr 304 FS AG Rack takapaneli 1 Logiikkasegnaalien tulokytkent pariliittimet toimitetaan niihin sopivilla yhdyskaapelisulkijalla 2 L htevien logiikkasignaalien ja pumpun taajuuden tarkkailun sek ohjelmoitavan analogisen ulostulon liitin Virran ulostuloliitin 24 V tasavirta pumpun j hdytyssiivikolle vent valve ja kaliiperille Valvojan tulovirran moduuli johon sis ltyv t sulakkeet virtapistoke ja suodatin ENC Profibus osoitteen asettamiseen tarkoitettu kiertokytkin Profibus valinnainen Et ohjauksen sarjaliit nt o M o aj ek w Pumpun liitin 130 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Ohjek sikirja 10 K ytt toimenpiteet V
71. FS AG Rack Controller User Manual 87 901 029 01 Manual de istrucciones Mandos Indicadores y Conectores A continuaci n se ilustran el panel de mando del controler y los paneles de interconexi n Para m s detalles consultar la secci n Technical Information Descripci n del panel frontal Figura 2 Panel frontal del Controller TwisTorr 304 FS AG Rack 4 Pulsador para gue aparezcan en el display los par metros cycle number cycle time y pump life Pulsador para la selecci n del modo LOW SPEED Est activado s lo cuando est seleccionado el modo de mando del panel frontal Apretando una vez la bomba gira a velocidad stand by Apret ndolo una vez m s se desactiva el modo LOW SPEED Pulsador para gue aparezcan en el display los par metros pump current pump temperature pump power y rotational speed Est siempre activado independientemente del modo de funcionamiento elegido Apretando juntos los pulsadores 3 y 1 durante 2 segundos por lo menos se activa un programa con el cual se pueden programar algunos par metros operativos Pulsador para enviar los mandos de START STOP RESET Est activo s lo cuando se selecciona el modo de mando del panel frontal Apret ndolo una vez se activa la fase de puesta en marcha apret ndolo otra vez se para la bomba Si la bomba se ha parado autom ticamente a causa de un
72. Figure 26 J 2 Connector Tab 5 PIN N SIGNAL NAME 5 V OUT Reserved TX RS232 RX RS232 NC GND A RS485 NC B RS485 1 2 3 4 5 6 7 8 9 RESERVED TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 227 301 17 Technical Information Connector Examples 5 5 2 2 TV 301 AG PG Rack Controller 3 3 Figure 27 RS 232 Connection TURBO CONTROLLER B TURBO CONTROLLER B Figure 28 RS 485 Connection 228 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 The communication port mating connector is supplied with the RS 232 PCB AMP Cannon or equivalent 15 pin D type male connector For example the Transmit data signal from controller pin 2 must be connected to the host computer s receive data line pin 2 and vice versa Consult the host computer s instruction manual for its serial port connections Agilent cannot guarantee compliance with FCC regulations for radiated emissions unless all external wiring is shielded with the shield being terminated to the metal shroud on the 0 subconnector The cable should be secured to the connector with screws TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 229 301 17 Technical Information Serial Communication Descriptions This unit can comunicate by two different protocol protocol
73. G Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 4 Manual de istrucciones Informaci n general 52 Almacenamiento 53 Preparaci n para la instalaci n 54 Instalaci n 55 Uso 56 Mandos Indicadores y Conectores 57 Descripci n del panel frontal 57 Procedimientos de uso 59 Encendido del controler59 Puesta en marcha de la Bomba 59 Parada de la Bomba 59 Mantenimiento 59 Eliminaci n 60 Mensajes de error 61 Traducci n de las instrucciones originales pp Agilent Technologies 51 301 4 Manual de istrucciones Informaci n general 52 301 Este eguipo se ha concebido para un uso profesional El usuario deber leer atentamente el presente manual de instrucciones y cualguier otra informaci n suplementaria facilitada por Agilent antes de utilizar el eguipo Agilent se considera libre de cualguier responsabilidad debida al incumplimiento total o parcial de las instrucciones al uso poco apropiado por parte de personal sin formaci n a las operaciones no autorizadas o al uso gue no cumpla con las normas nacionales espec ficas Los controlers de la serie TwisTorr 304 FS AG Rack son convertidores de frecuencia controlados por un microprocesador realizados con componentes en estado s lido y con capacidad de autodiagnosis y autoprotecci n Caracteristicas del controler a Operatividad frontal remoto serial a Pilotaje de 24 Vdc del ventilador de refrigeraci n bomba Pilotaje ven
74. Gebrauch 32 Bedienung 35 Wartung 35 Entsorgung 36 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 5 301 Contents 6 301 Fehlermeldungen 37 Mode d emploi 39 Indications generales 40 Emmagasinage 41 Preparation pour installation 42 Installation 43 Utilisation 44 Procedures d utilisation 47 Entretien 47 Mise au rebut 48 Messages d erreur 49 Manual de istrucciones 51 Informaci n general 52 Almacenamiento 53 Preparaci n para la instalaci n 54 Instalaci n 55 Uso56 Procedimientos de uso 59 Mantenimiento 59 Eliminaci n 60 Mensajes de error 61 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Contents 5 Manual de Istruc es 63 Informac es gerais 64 Armazenagem 65 Preparac o para a instalac o 66 Instalac o 67 Utilizac o 68 Procedimentos de uso 71 Manuten o 71 Eliminac o 72 Mensagens de erro 73 6 Bedrijfshandleiding 75 Algemene informatie 76 Opslag 71 Voorbereiding voor installatie 78 Installatie 79 Gebruik 80 Gebruiksprocedures 83 Onderhoud 83 Afvalverwerking 84 Foutmeldingen 85 7 Istruktionsbog 87 Generel Information 88 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 7 301 Contents 8 301 Opbevaring 89 Forberedelser f r installation 90 Installation 91 Anvendelse 92 Instruktion 95 Vedligeholdelse 95 Bortskaffelse 96 Fejlmeddelelser 97 Bruksanvisning 99 Allm n Information 100 F rvaring 101 F rbered
75. HNOLOGIES 3 A 2 menu 4 TwisTorr 304 FS AG Slika 2 Prednja plo ca Controllera TwisTorr 304 FS AG Rack 1 Tipka za preklic na zaslonu parametrov cycle number cycle time in pump life 2 Tipka za LOW SPEED modus je aktivna samo nato ko ste izbrali to komando iz prednje plo e Ce gumb enkrat pritisnite rpalka vrti na hitrosti stand by Ce e enkrat pritisnite gumb izklju ite LOW SPEED modus 3 Tipka za preklic na zaslonu parametrov pump current pump power rotational speed temperatura spodnjega le i a temperatura gornjega le i a temperatura motora frekvenca vzburjenja Ta tipka je vedno aktivna neodvisno od modusa delovanja ki ste ga izbrali e skupaj pritisnite gumbe 3 in 1 za vsaj dve sekundi aktivirate program s katerim je mogo e nastaviti nekatere operativne prarmetre 4 Tipka za zagon komand START STOP RESET Aktivna je samo nato ko ste to komando izbrali iz prednje plo e e gumb enkrat pritisnite boste pognali rpalko e ga e enkrat pritisnete boste rpalko zaustavili e se je rpalka sama zaustavila zaradi kvara morate to tipko enkrat pritisniti da bi lahko controller a resetirali in e drugi za ponoven zagon rpalke 5 Alfa tevil ni zaslon iz teko ih kristalov prebijalo bodov 2 linije x 16 znaka TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 189 301 15 Priro nik za Navodila
76. How to Use by Remote 1 0 276 How to Use in Serial Mode 276 Profibus Option 278 General 278 Installation278 Configuration 279 Software Operation and State Machine 279 Message Mapping 281 Parameter Buffer 282 Output Buffer 285 Input Buffer 286 Diagnostic Management 288 Diagnostic Buffer 290 Error Messages 292 Orderable Parts 295 208 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 TwisTorr 304 FS AG Rack Controller Description The controller is available in two models see orderable parts table with RS 232 485 option with Profibus option The controller is a solid state frequency converter with the following features Drives the pumps of the Twis Torr 304 FS family Powers the pump cooling fan Drives the vent valve Provides and acquires the pressure of the wide range gauge Remote I Os compatible with the previous version Navigator default serial compatible with the previous RS 232 and 485 version Speed reading after stop command See the following diagram for the connections and orderable parts TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 209 301 17 Technical Information AIR COOLING KIT VENT VALVE AIR COOLING KIT EXTENTION CABLE CONTROLLER TO POWER SUPPLY MAINS CABL
77. IS KOREKCE CHECK CONNECTION TO PUMP Chybn fungov n spojen mezi erpadlem a konrol rem P31 Nebo erpadlo m ni teplotu ne je 0 C Prov it zda kabelov spojen mezi pumpou a kontrol rem je spr vn upevn n na obou dvou konc ch a e nedo lo k p eru en Stisknout tla tko START dvakr t pro op tn spu t n erpadla WAITING INTERLOCK Aktivn sign l interlock se nach z na konektoru P1 nebo do lo k p eru en zkratu mezi pin 3 a pin 8 del konektoru J1 nebo kv li otev en sign lu vn j ho interlock Znovu obnovit spojen nakr tko mezi pin 3 a pin 8 konektoru P1 nebo uzav t sign l vn j ho interlock PUMP OVERTEMP Temperatura horn ho lo iska nebo erpadla p ev ila 60 C Po kat a se temperatura op t vr t pod mezn hodnotu Stisknout dvakr t tla tko START pro op tn spu t n erpadla CONTROLLER OVERTEMP Temperatura kontrol ru p ev ila 70 C Nebo Teplota radiatoru kontrol ru je vy ne 60 C Vy kat a se temperatura op t vr t pod mezn hodnotu Stisknout dvakr t tla tko START pro op tn spu t n erpadla TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 169 301 13 N vod k Pou it 170 301 ZPR VY POPIS KOREKCE TOO HIGH Absorbovan proud erpadla je Prov it zda rotor erpadla se LOAD vy ne ten n
78. Instrukcji 12 Sterowniki Wskazniki Laczniki Ponizsze rysunki przedstawiaja panel frontalny i tylny kontrolera Dla uzyskania dokladniejszych szezegolow skonsultowac odpowiedni rozdzial Technical Information Opis pulpitu frontalnego Posta 62 Pulpitfrontalny kontrolera TwisTorr 304 FS AG Rack 1 Przycisk sluzacy do przywolania na display parametrow cycle number cycle time I pump life 2 Przycisk do wybierania sposobu LOW SPEED Dziala tylko wtedy kiedy sposob sterowania jest wybierany na pulpicie frontalnym Naciskajac jeden szybkosc obrotu pompy w pozycji stand by Naciskajac jponownie nastepuje wylaczenie sposobu LOW SPEED ponownie 3 Przycisk sluzacy do przywolania na display parametrow pump current pump temperature pump power I rotational speed Jest zawsze aktywny niezaleznie od wybranego sposobu dzialania Naciskajac jednoczesnie przyciski 3 I I przez conajmniej 2 secundy uruchamia sie program przy pomocy ktorego jest mozliwe programowanie niektorych parametrow operatywnych 4 Przycisk do przesylania sygnalu START STOP RESET Dziala tylko wtedy kiedy sposob sterowania jest wybierany na pulpicie frontalnym naciskajac jeden raz uaktywnia sie faza uruchomienia naciskajac ponownie zatrzymuje sie pompe Jezeli pompa zatrzyma sie automatycznie z powodu uszkodzenia nalezy nacisnac przycisk jeden raz zeby zresetowac kontr
79. PANY FUNCTION ADDRESS TEL N FAX N E MAIL PROBLEM SUGGESTION REFERENCE INFORMATION model n serial n ordering information time to failure after installation etc DATE CORRECTIVE ACTION PLAN ACTUATION LOG N by AGILENT VPD XXX Code for dialing Italy from your country es 01139 from USA 00139 from Japan etc HE Agilent Technologies Agilent Technologies Vacuum Products Division Instructions for returning products Dear Customer Please follow these instructions whenever one of our products needs to be retumed 1 2 3 4 5 6 7 Complete the attached Reguest for Return form and send it to Agilent Technologies see below taking particular care to identify all products that have pumped or been exposed to any toxic or hazardous materials After evaluating the information Agilent Technologies will provide you with a Return Authorization RA number via email or fax as reguested Note Depending on the type of return a Purchase Order may be reguired at the time the Reguest for Return is submitted We will guote any necessary services evaluation repair special cleaning eg Important steps for the shipment of returning product e Remove all accessories from the core product e g inlet screens vent valves e Prior to shipment drain any oils or other liquids purge or flush all gasses and wipe off any excess residue e If ordering an Advance Exc
80. PUMPS CONTROLLERS VALVES COMPONENTS Bad feedthrough Poor vacuum Main seal leak Bellows leak Damaged flange SCROLL AND ROTARY VANE PUMPS Pump doesn t start Noisy pump describe Doesn t reach vacuum Over temperature Pump seized Other Section 6 ADDITIONAL TERMS Error code on display Other Damaged sealing area Other LEAK DETECTORS INSTRUMENTS Cannot calibrate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Other DIFFUSION PUMPS Heater failure Doesn t reach vacuum Vacuum leak Electrical problem Cooling coil damage Other Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division Products and Services Terms of Sale e Customer is responsible for the freight charges for the returning product Return shipments must comply with all applicable Shipping Regulations IATA DOT etc and carrier reguirements e Customers receiving an Advance Exchange product agree to return the defective rebuildable part to Agilent Technologies within 15 business days Failure to do so or returning a non rebuildable part crashed will result in an invoice for the non returned non rebuildable part e Returns for credit toward the purc
81. R2 Valve 176 R A R2 Set Point Pressure Valve Valid in win 171 4 Format X X E X X Where X 0 to 9 s 0r 177 R W L Start Output Mode 0 Starting Output ON only with pump Status Starting 1 running Output ON when the pump is running default 246 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 181 R W N Gas type 0 not configured 1 Nitrogen 2 Argon 3 Idrogen 4 other 182 R W N Gas correction 200 R N Pump current in mA de 201 R N Pump voltage in Vdc 202 R N Pump power in W pump current x pump voltage duty cycle 203 R N Driving frequency in Hz 204 R N Pump temperature in C 0 to 70 205 R N Pump status Stop 0 Waiting intlk 1 Starting 2 Auto tuning 3 Braking 4 Normal 5 Fail 6 206 R N Error code Bit description see the following figure 211 R N Controller Heatsink Temperature C 216 R N Controller Air Temperature C 224 R A Pressure reading Format X X E XX 226 R N Rotation Frequency rpm 257 R N Gauge status 0 No gauge connected 1 gauge connected 2 Under range gauge error 3 over range gauge error 4 Rid unknown 267 R W N Gauge power 0 gauge off 1 gauge On 2 programmable power sp1 3 programmable power sp2 4 programmable power sp3 300 R N Cycle time in minutes zeroed by the reset 0 to 999999 command 301 R N Cycle number zeroed by the reset 0
82. Se a bomba parou automaticamente por causa de uma avaria ocorre apertar este bot o uma primeira vez para executar o reset do controller e uma segunda vez para reiniciar a bomba 5 Display alfanum rico com cristais l quidos matriz de pontos 2 linhas x 16 caracteres TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 69 301 5 Manual de Istru es Jo ojo 23 So So So So So o o ONO of Figura 3 Painel posterior dos Controllers TwisTorr 304 FS AG Rack Conector de entrada dos sinais l gicos o conector de acoplamento fornecido com ponte espec fica de fecho 2 Conector sinais l gicos na sa da e monitoragem freg ncia bomba mais sa da anal gica program vel 3 Conector de sa da alimentac o 24 Vdc para a ventoinha resfriamento bomba vent valve e calibragem 4 Modulo de entrada alimentac o controller que compreende os fus veis a tomada de alimentac o e o filtro ENC 5 Switch em rotac o para impostar o endereco do Profibus 6 Profibus opcional 7 Conector serial para controle de remoto 8 Conector bomba 70 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de Istru es 5 Procedimentos de uso Acendimento do Controller Para ligar o controller suficiente inserir o cabo de alimentac o na tomada de rede Activac o da bomba Para li
83. Signal interlocka prisoten na konektorju P1 zaradi prekinitve kratkega stika med pinom 3 in pinom 8 konektorja J1 ali ker se je odprl signal zunanjega interlocka Ponovno nastavite kratki stik med pinom 3 in pinom 8 konektorja P1 ali zaprite signal zunanjega interlocka PUMP OVERTEMP Temperatura gornjega le i a ali rpalke je presegla i 60 C Po akajte da se temperatura vrne pod normalnim vrednostnim pragom Pritisnite dvakrat tipko START za ponoven zagon amp rpalke CONTROLLER OVERTEMP Temperatura controllera je presegla 70 C Ali Temperatura controllerskega radiatorja je vi ja od 60 C Po akajte da se temperatura vrne pod normalnim vrednostnim pragom Dvakrat pritisnite tipko START za ponoven zagon amp rpalke TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 193 301 15 Priro nik za Navodila 194 301 OBVESTILO OPIS POPRAVA TOO HIGH Tok ki ga rpalka prejme je ve ji od Preverite da lahko rotor LOAD programiranega rpalke prosto vrti Dvakrat pritisnite tipko START za ponoven zagon rpalke SHORT Med navadnim delovanjem po Preverite priklju ke rpalke in CIRCUIT zagonu je priklopitev na izhodu v controllera kratkem stiku Dvakrat pritisnite tipko START za ponoven zagon rpalke SYSTEM Delovanje rpalke je zaustavil signal Izklopite napajalni kabel OVERRIDE zasilnosti iz oddaljenega kontakta controllera in popravite napako ki se je po
84. TART command and again for the STOP command In the case of controller s failure press once to reset the fail condition The front panel START STOP function is available only if the controller is set to FRONT mode If it is set to REMOTE mode when the key is pressed an error message is shown indicating COMMAND LOCKED IN REMOTE if it is set to SERIAL mode the COMMAND LOCKED IN SERIAL MODE error message is displayed The Reset Fail function is always available regardless of controller mode Low Speed Press once to activate the LOW SPEED command press again to de activate The front panel LOW SPEED function is available only if the controller is set to FRONT otherwise an error message is returned If it is set to REMOTE mode the COMMAND LOCKED IN REMOTE error message is displayed while if it is in SERIAL mode the COMMAND LOCKED IN SERIAL MODE error message is returned TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 249 301 17 Technical Information 250 301 Counters Pressing the first time on the display the following information is displayed Where CYCLE XXXX Number of start stop cycles TIME XXXXXm Time of last cycle minutes Pressing a second time the following is displayed PJUJM P LIFE XIXIXIXIXIh PUMP LIFE XXXXXX total operating time hours of the pump Pressing a th
85. TING INTERLOCK The interlock signal of P1 connector is activated by an interruption of the link between pin 3 and 8 of J1 connector or because the external interlock signal is open Reset the short circuit between pin 3 and pin 8 of P1 connector or close the external interlock signal PUMP OVERTEMP The upper bearing pump temperature exceeds 60 C Wait until the temperature decrease below threshold value Press the START push button twice to start the pump CONTROLLER OVERTEMP The controller environment temperature exceeds 70 C Or The controller s radiator temperature is above 60 C Wait until the temperature decrease below threshold value Press the START push button twice to start the pump TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 205 301 16 Instructions for Use 206 301 MESSAGE DESCRIPTION REPAIR ACTION TOO HIGH The current drawn by the pump is Check that the pump rotor is LOAD higher than programmed free to rotate Press the START push button twice to start the pump SHORT After the starting phase the output Check connections and CIRCUIT connection is shorted shortages between pump and controller Press the START push button twice to start the pump SYSTEM The pump is stopped by an Remove the controller power OVERRIDE emergency stop signal provided via cable and check the a remote contact emergency condition Then reconnect the p
86. Tab 6 CHARACTER ASCII CONVERTED ASCII FOR CRC CALCULATION A 41h CRC BFh B 42h CRC BEh c 43h CRC BDh D 44h CRC BCh SE 45h CRC BBh F 46h CRC BAh G 47h CRC 89h H 48h 49h CRC B7h J 4Ah CRC B6h 232 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 CHARACTER ASCII CONVERTED ASCII FOR CRC CALCULATION K 4Bh CRC B5h ACK 06h CRC FAh NACK 15h CRC EBh Answer after a reguest from the host the Controller will answer in one of the following ways ACK NACK Message When the Reguest is A B C D F the Controller will Answer the ACK or NACK TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 233 301 17 Technical Information When the Reguest is E the Answer will contain the complete set of the following parameters Tab 7 BYTES MEANING 1 MSB LSB XXX 0 STOP 1 WAITING INTERLOCK 2 STARTING 3 NORMAL OPERATION 4 5 HIGH LOAD 6 FAILURE 7 APPROACHING LOW SPEED 2 5 Cycle Time 6 9 Pump Life 10 11 Pump Temperature 12 Current 13 Voltage 14 17 Freguency 18 19 Cycle 20 R1 Status 21 R2 Status 22 CRC The values for current and voltage are given as numbers scaled from 0 to 255 where 0 corresponds to 0 V and 255 to the full scale voltage 130 V or current 2 5 A 234 301 Twis
87. Tdeal amp vacuum products amp Manuale di istruzioni Bedienungshandbuch Notice de mode d emploi Manual de istrucciones Manual de istruc es Bedrijfshandleiding Istrukstionsbog Bruksanvisning 87 901 029 01 01 2013 E Agilent Technologies TwisTorr 304 FS AG Rack Controller Models X3506 64002 X3506 64003 Instruksjon manual Ohjek sikirja Felhaszn l i k zik nyv Podrecznik instrukcji N vod k pou it N vod na obsluhu Priro nik za navodila User Manual Notices Agilent Technologies Inc 2013 No part of this manual may be reproduced in any form or by any means including electronic storage and retrieval or translation into a foreign language without prior agreement and written consent from Agilent Technologies Inc as governed by United States and international copyright laws Manual Part Number Publication Number 87 901 029 01 Edition Edition 01 2013 Printed in ITALY Agilent Technologies Italia S p A Vacuum Products Division Via F lli Varian 54 10040 Leini TO ITALY Warranty The material contained in this document is provided as is and is subject to being changed without notice in future editions Further to the maximum extent permitted by applicable law Agilent disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fi
88. Torr 304 FS AG Rack Controller User Manual 87 901 029 01 261 301 17 Technical Information Programming Configuration Menu To access the configuration menu press the COUNTERS MEASURES buttons at the same time for at least 2 sec Figure 37 In programming mode the meaning of the 4 buttons changes as follows COUNTERS LOW SPEED speed become INCREASE and DECREASE and make it possible to select the next or previous value of a parameter in the case of compulsory parameters or to increase decrease the individual digit of a parameter in the case of numeric parameters The MEASURES and START STOP buttons become NEXT and PREVIOUS and make it possible to select the next parameter or next digit of a numeric parameter or previous parameter 262 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Switching from one parameter to the next using the NEXT key indicates confirmation of any changes made to the specific parameter If the changes cannot be accepted by the controller the Data out of limits error message is displayed the changes are cancelled and the value of the parameter prior to the modification is reproposed Switching from a parameter to the previous parameter using the PREVIOUS key indicates the intention to abort any c
89. Torr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 When the Reguest is G the Answer will contain a string of 11 characters with the following parameters Tab 8 BYTES MEANING 1 2 Pump cycles number integer coded in 2 bytes 3 Speed threshold 4 7 Run up time in seconds long coded in 4 bytes 8 Deat time 0 NO 1 YES 9 Reserved 10 Soft Start mode 0 NO 1 YES 11 CRC When the configuration parameters have to be changed send a Request string with H characters of the following type H data CRC TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 235 301 17 Technical Information The following parameters can be changed Tab 9 BYTES MEANING 1 Speed threshold 2 5 Run up time in seconds long coded in 4 bytes 6 Deat time 0 NO 1 YES 7 Reserved 8 Soft Start mode 0 YES 1 NO 9 CRC When the Request is I Status readings the Answer will contain a string of 2 characters with the following parameters Tab 10 BYTES MEANING 1 MSB LSB XXXXXX 0 STOP 1 WAITING INTERLOCK 2 STARTING 3 NORMAL OPERATION 4 5 HIGH LOAD 6 FAILURE 7 APPROACHING LOW SPEED R2 status 0 OFF 1 0N R1 status 0 OFF 1 0N 2 CRC 236 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 When the Reguest is J Numerical readings the Answe
90. Tryck tv g nger p knappen START f r att starta om pumpen CONTROLLER Temperaturen p styrenhetens V nta tills temperaturen OVERTEMP verskrider 70 C sjunker under tr skelv rdet eller ocks Temperaturen p styrenhetens kylare verskrider 60 C Tryck tv g nger p knappen START f r att starta om pumpen TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 109 301 110 301 Bruksanvisning MEDDELANDE BESKRIVNING TG RD TOO HIGH Str mmen som absorberas av Kontrollera att pumpens rotor LOAD pumpen verstiger det v rde som kan rotera fritt programmerats Tryck tv g nger p knappen START f r att starta om pumpen SHORT Under normal drift efter Kontrollera anslutningarna CIRCUIT startmomentet har utg ngen mellan pump och styrenhet kortslutits Tryck tv g nger p knappen START f r att starta om pumpen SYSTEM Pumpen har stannats av en Fr nkoppla styrenhetens OVERRIDE n dstoppssignal fr n en str mkabel och kontrollera fj rrkontakt n dstoppets orsak S tt d refter i str mkabeln Tryck tv g nger p knappen START f r att starta om pumpen OVERVOLTAGE Ett fel i styrenhetens matningsdel Tryck tv g nger p knappen har uppst tt eller styrenheten har START f r att starta om f tt en falsk signal pumpen Om meddelandet visas igen ta kontakt med Agilent f r n dv ndigt underh ll POWER FAIL Fel i sektionen f r str mtillf rsel till Ta kontakt me
91. User Manual 87 901 029 01 287 301 17 Technical Information 7 2 mA 0 10000 Current 200 Motor current consumption 9 2 W 0 200 Power 202 Motor power adsorption 11 2 W 0 200 Power Limit Applied 155 Maximum allowable power 13 2 Hz 0 1500 Driving Freguency 203 Read the rotational speed of the pump 15 2 Hz 0 1500 Target Freguency 120 Read the target rotational speed 17 2 0 65536 Cycle Number 301 Number of cycles start and stop done by the pump 19 2 min 0 65536 Last Cycle Time 300 How long the last cycle lasted 21 2 h 0 65536 Pump Life 302 Total pump running time 23 4 mBar Torr Pressure Reading 224 Pa Reports the pressure reading in the pressure unit selected by Par Buff Offset 20 B5 6 288 301 Diagnostic Management In addition to the ProfiBus six byte standard diagnostic the interface provides both user diagnostic and static diagnostics functions The standard diagnostic management is fully compliant with the ProfiBus specification During the start up phase after power on the first byte in the user diagnostic area can assume values different from zero This byte has the following meaning Bit 0 1 controller not responds RS485 link broken a Bit 1 1 controller not identified not compatible with the ProfiBus gateway TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Bit 2 1 first loop of reguests not ended value
92. Volt Output Voltage ipa pr 0 freq max Frequency Hz Figure 19 Diagram 1 Programmable analog output related to frequency freg max 1010Hz V Volt 4 Output Voltage pol Pn pn 0 150 Power W Figure 20 Diagram 2 Programmable analog output related to power V Volt Output Voltage 10 0 100 Pump Temp C Figure 21 Diagram 3 Programmable analog output related to pump temperature 222 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Pump Connector A five meter long cable is available to connect the controller to the pump The following figure shows the controller output connector configuration where pins A F pump temperature sensor B C D 75 Vac 3 phase output to pump motor stator E EARTH PROTECTION PUMP FEM ALE CONNECTOR 3 PHASE e MP TEMPERATURE SENSOR EARTH Figure 22 Controller to Pump Connector J5 Gauge Connector This connector is available to connect the gauges J5 GAUGE IDENTIFICATION 4 SIGNAL SIGNAL COMMON 25VDC P MAX 2W COMMON TA Figure 23 Gauge Connector The Mini IMG gauge is intended for high vacuum use so with pressure above 50 mTorr the signal is misinterpreted and a false pressure is displayed Use a transition gauge to understand if the reading is really under 50 mT
93. W nastepnych paragrafach zawarte sa wszystkie niezbedne informacje dotyczace gwarancji bezpieczenstwa personelu obslugujacego urzadzenie Bardziej szczegolowe informacje sa zawarte w dodatku do Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Podrecznik Instrukcji 12 Podrecznik ten opiera sie na nastepujacych konwencjach Sygnaly ostrzegajace o zagrozeniu powinny zwrocic uwage operatora na odpowiednia specyficzna procedure ktora jezeli nie jest przestrzegana w sposob prawidlowy moze przyczynic sie do powaznych obrazen osoby obslugujacej Sygnaly wzrokowe ukazujace sie przed procedura jezeli zostana zlekcewazone moga spowodowac uszkodzenia urzadzenia Przypisy zawieraja najwazniejsze informacje wyciagniete z tekstu Magazynowanie Kontrolery podczas transportu i przechowywania w magazynie powinny miec nastepujace warunki otoczenia temperatura od 20 C do 70 C a wilgotnosc wzgledna od 0 95 nie kondensujaca TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 149 301 12 Podrecznik Instrukcji Przygotowanie do instalacji Kontroler zostaje dostarczony w specjalnym ochronnym opakowaniu jezeli opakowanie prezentuje jakiekolwiek uszkodzenia ktore moglyby powstac podczas transportu wowczas nalezy niezwlocznie powiadomic najblizszy lokalny punkt sprzedazy Podczas operacji rozpakowywania nalezy zachowac specjalna ostroznosc zeby nie spowodowac u
94. a averia hay gue apretar este pulsador primero una vez para efectuar el reset del controler y la segunda vez para volver a poner en marcha la bomba Display alfanumerico de cristales liguidos matriz de puntos 2 lineas x 16 caracteres TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 57 301 4 Manual de istrucciones CA O m Figura 3 Panel trasero del controler Twis Torr 304 FS AG Rack Conector de entrada de las se ales l gicas el conector de acoplamiento se suministra con el conector puente especifico de cierre 2 Conector se ales l gicas en salida y monitoreo frecuencia bomba y salida anal gica programable 3 Conector de salida alimentaci n 24 Vdc para el ventilador de refrigeraci n bomba vent valve y calibre 4 M dulo de entrada alimentaci n controler con fusibles toma de alimentaci n y filtro ENC 5 Conmutador de rotaci n para configurar la direcci n del Profibus 6 Profibus opcional 7 Conector serial para control remoto 8 Conector bomba 58 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de istrucciones 4 Procedimientos de uso Encendido del controler Para encender el controler es suficiente introducir el cable de alimentaci n en la toma de red Puesta en marcha de la Bomba Para activar la bomb
95. a es necesario habilitar la entrada de interlock y pulsar el bot n START situado en el panel frontal Parada de la Bomba Para detener la bomba hay que apretar el pulsador STOP del panel frontal Mantenimiento Los controlers de la serie TwisTorr 304 FS AG Rack no necesitan ningun mantenimiento Cualguier operaci n ha de ser efectuada por personal autorizado En caso de averia es posible utilizar el servicio de reparaci n Agilent o del Agilent advance exchange service que permite obtener un controler regenerado en vez del averiado LILLI Antes de efectuar cualquier operaci n en el controler desenchufar el cable de alimentaci n En caso de gue un controler se tenga gue desguazar efectuar su eliminaci n respetando las normas nacionales espec ficas TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 59 301 4 Manual de istrucciones Eliminaci n 60 301 Significado del logotipo WEEE presente en las etiguetas EI simbolo gue se indica a continuaci n es aplicado en observancia de la directiva CE denominada WEEE Este simbolo v lido s lo para los paises miembros de la Comunidad Europea indica que el producto sobre el cual ha sido aplicado NO debe ser eliminado junto con los residuos comunes sean stos dom sticos o industriales y gue por el contrario deber ser sometido a un procedimiento de recogida diferenciada Por lo tanto se invita al usuario final a ponerse en contacto con el provee
96. alvojan k ynnistys Valvoja k ynnistyy asetettaessa virtakaapeli pistorasiaan Pumpun k ynnistys K ynnist ksenne pumpun pit ottaa k ytt n interlock sis ntulo ja painaa etupaneelin painiketta START Pumpun pys ytt minen Huolto Pumppu pys htyy painettaessa etupaneelissa olevaa STOP painonappia TwisTorr 304 FS AG Rack sarjan valvojat eiv t kaipaa mink nlaista huoltoa Mahdolliset valvojaan teht v t toimenpiteet tulee j tt aina valtuutetun henkil n teht viksi Toimintah iri n sattuessa on mahdollista k ytt Agilentin korjauspalvelua tai Agilent advance exchange service palvelua jolloin on mahdollista vaihtaa rikkoontunut valvoja ladattuun valvojaan VAROITUS Ennen mink tahansa valvojaan teht v n toimenpiteen suorittamista irrottakaa s hk kaapeli pistorasiasta Mik li valvoja t ytyy romuttaa toimikaa sen h vitt misess kansallisten s d sten ja normien m r m ll tavalla TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 131 301 10 Ohjek sikirja H vitt minen Pakkausmerkinn iss olevan WEEE logon merkitys Alla n kyv merkki on lis tty pakkaukseen EY n ns WEEE direktiivin mukaisesti Merkki koskee ainoastaan Euroopan Unionin j senmaita tarkoittaa ett tuotetta EI saa h vitt tavallisen kotitalous tai teollisuusj tteen mukana vaan se on toimitettava erilliseen ker yspisteeseen Loppuk ytt j kehotetaan sen vu
97. anel More details are contained in the appendix Technical Information Front Panel Description 5 1 3 2 4 Figure 2 Controller TwisTorr 304 FS AG Rack Front Panel 1 Keyboard push button to recall on the display the cycle number cycle time and pump life 2 Keyboard push button for LOW SPEED mode selection It is active only when the front panel operation has been selected Pressed once the pump runs at stand by speed To unselect the mode press the push button again 3 Keyboard push button to recall on the display the pump current pump temperature pump power and rotational speed It is always active regardless of the operating mode selected Push buttons 3 and 1 if pressed together for at least 2 seconds put the controller in a routine where it is possible to program some operation parameters 4 Keyboard push button for START STOP RESET mode selection It is active only when the front panel operation has been selected By pressing once the starting phase begins if pressed again it stops the pump If the pump has been stopped automatically by a fault this push button must be pressed once to reset the controller and a second time to restart the pump 5 LCD back lighted alphanumeric display dot matrix 2 lines x 16 characters TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 201 301 16 Instructions for Use
98. anual 87 901 029 01 49 301 3 Mode d emploi MESSAGE DESCRIPTION INTERVENTION TOO HIGH LOAD Pendant le fonctionnement normal apres la phase de mise en marche le courant absorb par la pompe est plus grand gue celui gui a t programm S assurer que le rotor de la pompe a la possibilit de tourner librement Presser deux fois l interrupteur START pour remettre la pompe en marche SHORT CIRCUIT Pendant le fonctionnement normal apr s la phase de mise en marche la connexion de sortie est en court circuit Verifier les connexions entre la pompe et le contr leur Presser deux fois l interrupteur START pour remettre la pompe en marche SYSTEM OVERRIDE La pompe a t arr t e par un signal d alerte provenant d un contact loign D brancher le c ble d alimentation du contr leur et corriger la cause de Valerte Reconnecter le c ble d alimentation et presser deux fois l interrupteur START pour remettre la pompe en marche OVERVOLTAGE Il s est produit une panne de la section d alimentation du contr leur ou bien le contr leur a recu un faux signal Presser deux fois l interrupteur START pour remettre la pompe en marche Si le message se pr sente nouveau s adresser Agilent pour l entretien POWER FAIL Panne dans la section d alimentation de la pompe du contr leur S adresser Agilent pour Maintenance 50 301 TwisTorr 304 FS A
99. any work on the controller disconnect it from the supply If a pump is to be scrapped it must be disposed off in accordance with the specific national standards TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 203 301 16 Instructions for Use Disposal 204 301 Meaning of the WEEE logo found in labels The following symbol is applied in accordance with the EC WEEE Waste Electrical and Electronic Equipment Directive This symbol valid only in countries of the European Community indicates that the product it applies to must NOT be disposed of together with ordinary domestic or industrial waste but must be sent to a differentiated waste collection system The end user is therefore invited to contact the supplier of the device whether the Parent Company or a retailer to initiate the collection and disposal process after checking the contractual terms and conditions of sale TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Error Messages Instructions for Use 16 For a certain type of failure the controller will self diagnose the error and the messages described in the following table are displayed Tab 1 MESSAGE DESCRIPTION REPAIR ACTION CHECK CONNECTION TO PUMP Wrong connection between the pump and the controller Or The pump s temperature is below 0 ec Check connection between controller and pump Press the START push button twice to start the pump WAI
100. ao Yang District Beijing 100102 China Tel 86 0 10 64397888 Fax 86 0 10 64391318 Toll free 800 820 3278 France Agilent Technologies France Vacuum Products Division 7 Avenue des Tropigues Z A de Courtaboeuf B P 12 91941 Les Ulis cedex France Tel 33 0 1 69 86 38 84 Fax 33 0 1 69 86 29 88 Toll free 00 800 234 234 00 Germany and Austria Agilent Technologies Sales amp Services GmbH amp Co KG Vacuum Products Division Lyoner Str 20 60 528 Frankfurt am Main GERMANY Tel 49 69 6773 43 2230 Fax 49 69 6773 43 2250 Sales and Service Offices India Agilent Technologies India Pvt Ltd Vacuum Products Division G01 Prime corporate Park 230 231 Sahar Road Opp Blue Dart Centre Andheri East Mumbai 400 099 India Tel 91 22 30648287 8200 Fax 91 22 30648250 Toll Free 1800 113037 Italy Agilent Technologies Italia S p A Vacuum Products Division Via F Ili Varian 54 10040 Leini Torino Italy Tel 39 011 997 9111 Fax 39 011 997 9350 Toll Free 00 800 234 234 00 Japan Agilent Technologies Japan Ltd Vacuum Products Division 8th Floor Sumitomo Shibaura Building 4 16 36 Shibaura Minato ku Tokyo 108 0023 Japan Tel 81 3 5232 1253 Fax 81 3 5232 1710 Toll Free 0120 655 040 Korea Agilent Technologies Korea Ltd Vacuum Products Division Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 0 2 3452 245
101. aprogramovan m e voln ot et Stisknout dvakr t tla tko START pro op tn spu t n erpadla SHORT B hem norm ln ho provozu po Prov it spojen mezi CIRCUIT spou t c f zi v stupn spojen je erpadlem a kontrol rem ve zkratu Stisknout dvakr t tla tko START pro op tn spu t n erpadla SYSTEM Pumpa byla zastavena nouzov m Odpojit nap jec kabel OVERRIDE sign lem prost ednictv m d lkov ho kontrol ru a napravit p inu kontaktu poplachu Znovu p ipojit nap jec kabel a stisknout dvakr t tla tko START pro op tn spu t n erpadla OVERVOLTAGE Vznikla porucha v nap jec sekci Stisknout dvakr t tla tko kontrol ru nebo kontrol r dostal START pro op tn spu t n fale n sign l erpadla Po op tn zpr v ohledn dr by je nutno se obr tit na Agilent POWER FAIL Porucha v nap jec sekci erpadla Ohledn dr by je t eba se kontrol ru obr tit na Agilent TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 14 N vod na Obsluhu V eobecn inform cie 172 Usklad ovanie 173 Priprava pre in talaciu 174 In talacia 175 Pou itie 176 Povely Ukazovatele a Pripojky Controller 177 Popis elneho panelu 177 Postup pri pou iti 179 Zapnutie Controller 179 Na tartovanie erpadla 179 Zastavenie erpadla 179 dr ba 179 Likvid cia 180 Oznamy v d 1
102. ato No instalar y o utilizar el controler en ambientes expuestos a agentes atmosf ricos lluvia hielo y nieve polvos gases agresivos en ambientes explosivos o con alto riesgo de incendio Durante el funcionamiento es necesario que se respeten las condiciones ambientales siguientes temperatura de 5 C a 45 C humedad relativa O 95 no condensadora Para otras conexiones y la instalaci n de los accesorios opcionales v ase la secci n Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 55 301 4 Manual de istrucciones Uso En este apartado se citan los procedimientos operativos principales Para m s detalles y para procedimientos gue impliguen conexiones u opcionales especiales les remitimos al apartado Use del anexo Technical Informations Antes de usar el controler efectuar todas las conexiones el ctricas y neum ticas y consultar el manual de la bomba conectada iADVERTENCIA Para evitar lesiones a las personas y al aparato si la bomba est apoyada sobre una mesa cerciorarse gue es estable No poner en marcha nunca la bomba si la brida de entrada no est conectada al sistema o no est cerrada con la brida de cierre EI conector di cierre J1 ha de dejarse conectado con su conector puente si no se efect a ninguna conexi n exterior La bomba pre vacio y la bomba TwisTorr 304 FS pueden encenderse simult neamente 56 301 TwisTorr 304
103. b 2 November 1995 as applicable in any technical data Trademarks Windows and MS Windows are U S registered trademarks of Microsoft Corporation Safety Notices A CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly performed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller TwisTorr 304 FS AG Rack Controller TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 3 301 TwisTorr 304 FS AG Rack Controller 4 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Contents Contents 1 Istruzioni per uso 15 Informazioni Generali 16 Immagazzinamento 17 Preparazione per l installazione 18 Installazione 19 Uso20 Procedure di uso 23 Manutenzione 23 Smaltimento 24 Messaggi di errore 25 2 Gebrauchsanleitung 27 Allgemeines28 Lagerung 29 Vor der Installation 30 Installation 3
104. c ho druh raz sa vypne LOW SPEED 3 Tla idlo pre vyvolanie parametrov pump current pump temperature pump power rotational speed Je stale akt vne nez visle od vo by funkcie Stla iac spolu tla idl 3 a 1 na aspo 2 sekundy sa spust program s ktor m ke mo n zvoli niektor operat vne parameter 4 Tla idlo povelov TART STOP RESET Je akt vne iba vtedy ke povel vych dza z eln ho panelu Ke sa stla raz spust sa f za na tartovania stla iac ho znovu sa zastav erpadlo Ke sa erpadlo v pr pade poruchy zastav automaticky treba stla i tla idlo prv kr t pre resetovanie controller a druh kr t pre znovuna tarovanie erpadla 5 Alfanumerick displej na tekut kry t l bodov matica 2 l nie x 16 p smen TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 177 301 14 N vod na Obsluhu El OA LJ D EI Postava 3 Zadny panel controller TwisTorr 304 FS AG Rack 1 Vstupn spojka logick ch sign lov spojovacia spojka je dogan s prislusnym uzavierajucim mostikom 2 Spojka v chodz ch logick ch sign lov a monitorovanie frekvencie erpadla a navy e programovateln analogick v stup V stup spojka nap jania 24 Vdc pre chladiacu lopatku erpadla vent valve a kaliber Vstupn modul nap jania controller u obsahuj ci
105. chine Figure 46 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 279 301 17 Technical Information Tab 16 Profibus state machine Status Meaning POWER ON Interface set up WAIT PRM Wait for parameterisation from the master WAIT CFG Wait for configuration from the master DATA EX Regular data exchange Internal link state machine SRC ADDR Figure 47 Tab 17 Internal link state machine Status Meaning POWER ON Interface set up SRC ADDR Search for the controller serial address IDENTIFY Identify controller model and protocol DATA EX Regular data exchange 280 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 The two machines continuously share I O data with the relative bus and interact each other only in following conditions a ProfiBus Data Exchange is suspended static diagnosis activated if the internal serial link in not established Internal serial link Data Exchange is suspended only for output data if the ProfiBus State machine is not in Data Exchange mode As a general rule unrecoverable communication errors make each machine to jump to the state immediately following the power on state WAIT PRM or SRC ADDR Message Mapping The communication is based on the continuous data exchange capability of the ProfiBus standard where the master sends continuously the entire slave configuration and reads back
106. chnologies Vacuum Products Division Reguest for Return Form Health and Safety Certification Please read important policy information on Page 3 that applies to all returns 1 CUSTOMER INFORMATION Company Name Contact Name Tel Email Fax Customer Ship To Customer Bill To Reference 3 TYPE OF RETURN Choose one from each row and supply Purchase Order if reguesting a billable service 3A O Non Billable Billable gt New PO hard copy must be submitted with this form 3B exchange Repair upgrade consignment Demo calibration evaluation Return for Credit 4 HEALTH and SAFETY CERTIFICATION AGILENT TECHNOLOGIES CANNOT ACCEPT ANY PRODUCTS CONTAMINATED WITH BIOLOGICAL OR EXPLOSIVE HAZARDS RADIOACTIVE MATERIAL OR MERCURY AT ITS FACILITY Call Agilent Technologies to discuss alternatives if this reguirement presents a problem The eguipment listed above check one O HAS NOT pumped or been exposed to any toxic or hazardous materials OR HAS pumped or been exposed to the following toxic or hazardous materials If this box is checked the following information must also be filled out Check boxes for all materials to which product s pumped or was exposed toxic O Corrosive 0 Reactive O Flammable O Explosive O Biological O Radioactive List all toxic hazardous materials Include product name chemical name and chemical symbol or formula NOTE If a product is received at
107. ck Controller User Manual 6 Bedrijfshandleiding Algemene informatie 76 Opslag 77 Voorbereiding voor installatie 78 Installatie 79 Gebruik 80 Bedieningsorganen Controlelampjes en Connectoren van 81 Beschrijving frontpaneel 81 Gebruiksprocedures 81 Inschakelen van de controller 83 Starten van de pomp 83 Stoppen van de pomp 83 Onderhoud 83 Afvalverwerking 84 Foutmeldingen 85 Vertaling van de originele instructies tie Agilent Technologies 75 301 Bedrijfshandleiding Algemene informatie 76 301 Deze apparatuur is bestemd voor beroepsmatig gebruik De gebruiker wordt verzocht aandachtig deze handleiding en alle overige door Agilent verstrekte informatie door te lezen alvorens het apparaat in gebruik te nemen Agilent acht zich niet aansprakelijk voor de gevolgen van het niet of gedeeltelijk in acht nemen van de aanwijzingen onoordeelkundig gebruik door niet hiervoor opgeleid personeel reparaties waarvoor geen toestemming is verkregen of gebruik in strijd met de specifieke nationale wetgeving De controllers van de TwisTorr 304 FS AG Rack serie zijn freguentieomzetters die gestuurd worden door een microprocessor zijn gemaakt van halfgeleiderele menten en zijn in staat om zelfdiagnose en zelfbescherming uit te voeren Kenmerken van de controller a Bediening frontaal op afstand serieel Sturing met 24 Vdc van de koelventilator van de pomp Sturing vent valve a Lezen snelheid pomp na stopcommando lezen stops
108. connected the display will be as follows CJH E C K CIO N N E C T I 0 N TJO PJUJM a Pressing the START button the pump is started and the display shows the following sereen page S TJA R T I N G X X X X Hz F R O N T VARIAN VACUUMTECHNOLOGIES COUNTERS Increase i 0 decrease previous LOW SPEED START STOP TwisTorr 304 FS AG Figure 32 256 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 This screen page will be visible throughout the acceleration phase the pump The increasing value of the rotation freguency will be displayed in the boxes marked X the user may decide to express this value in Hz default or in KRMP Once operating speed has been reached the following screen page is displayed N O R M A L 9 6 3 H z F R O N T If the LOW SPEED key is pressed the above screen page changes to TwisTorr 304 FS AG Figure 33 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 257 301 17 Technical Information 258
109. controller switches to WAITING INTERLOCK status and interrupts driving of the pump IMPULSIVE it is sufficient for the input to be active when the START command is sent to the pump Subseguently the input can also be de activated If the interlock is not active when the pump is started the controller switches too WAITING INTERLOCK status TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 269 301 17 Technical Information ANALOG OUT TYPE permits selection of the type of signal to be sent to the programmable analogue output One of the following signals can be selected FREQUENCY the drive frequency of the pump is shown POWER the power supplied to the pump is shown TEMPERATURE the temperature of the pump is shown PRESSURE the pressure read by the gauge is shown START OUT MODE Selects the operating mode of the START output Possible options RUNNING the output is active when the pump is driven STARTING the output is active only during acceleration ramp up SPEED OUT MODE KRPM Hz permits selection of the unit of measurement of the speed of rotation shown on the display of the controller This setting acts only on the measurement shown on the display Possible options KRPM Hz LCD CONTRAST 270 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Setpoint R1 and Setpoint R2 Submenus This submenu contains the settings relating to the o
110. d Agilent f r pumpen styrenheten Underh ll TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 9 Instruksjon Manual Generell informasjon 112 Lagring 113 Forberede installasjonen 114 Installasjon 115 Bruk 116 Indikatorer og Kontakter pa Styreenheten 117 Beskrivelse av frontpanelet 117 Instruksjoner for bruk 119 Starte styreenheten 119 Starte pumpen 119 Stoppe pumpen 119 Vedlikehold 119 Eliminering 120 Feilmeldinger 121 Oversetting av den opprinnelige samsvarserkl ringen pp Agilent Technologies 111 301 Instruksjon Manual Generell informasjon 112 301 Dette utstyret er beregnet til bruk av profesjonelle brukere Brukeren b r lese denne brukerveiledningen og all annen informasjon fra Agilent for utstyret tas i bruk Agilent kan ikke holdes ansvarlig for hendelser som skjer p grunn av manglende oppfelging selv delvis av disse instruksjonene feilaktig bruk av utrenet personell ikke godkjente endringer av utstyret eller handlinger som p noen m te er i strid med nasjonale bestemmelser Styreenhetene i TwisTorr 304 FS AG Rack serien er mikroprosessor styrte frekvensomvendere som best r av komponenter med fast tilstand Styreenhetene har funksjoner for selvdiagnose og selvbeskyttelse Styreenhetens karakteristikker a Frontal virksomhet fjern seriell Igangsetting av kj leviften p 24 Vdc p pumpen a gangsetting av vent valve A
111. d the gas type 4 Other you can set the correction factor of the pressure measurement for this kind of gas The pressure indicated by the instrument is egual to P indicated P measured x Gas correction 10 274 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Serial Menu This menu contains the settings for management of the RS232 485 serial line ks AT 1 SERIAL le BAUD RATE l 60 E177 I i AS _b 1200 4800 9600 SERIAL MODE RS232 RS485 SERIAL ADDRESS amp 00 01 to I leest I Flow 6 6 Figure 43 BAUD RATE permits selection of the communication speed selecting this from 600 1200 4800 9600 baud SERIAL MODE permits setting of the type of interface to RS232 point to point or RS485 multidrop SERIAL ADDRESS sets the address of the node in the case of the RS485 interface This parameter is not visible if the RS232 is selected TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 275 301 17 Technical Information How to Use by Remote 1 0 GB O N Power ON the controller Press the COUNTERS and MEASURES button toghether Select MODE REMOTE from the INPUT OUTPUT menu Exit from configuration menu Connection example How to Use in Serial Mode 276 301 Power on the controller Connect the serial cable Select by front panel if you want to work
112. de frequ ncia controlados por um microprocessador realizados com componentes em estado s lido e com capacidade de autodiagn stico e autoprotec o Caracter sticas do controller a Operatividade frontal remota serial Pilotagem em 24 Vdc da ventoinha de resfriamento bomba Pilotagem vent valve a Leitura velocidade bomba seguindo ao comando de paragem leitura velocidade de paragem Leitura da press o a Interface Profibus opcional Imposta o autom tica tens o de entrada Nos par grafos seguintes est o descritas todas as informa es necess rias para garantir a seguran a do operador durante o uso da aparelhagem Informa es pormenorizadas s o fornecidas no ap ndice Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de Istru es 5 Este manual utiliza as seguintes conven es ATEN AO As mensagens de aten o chamam a aten o do operador para um procedimento ou uma pr tica especifica que se n o efectuada correctamente pode provocar graves les es pessoais CUIDADO As mensagens de cuidado s o visualizadas antes de procedimentos que se n o observados podem causar danos aparelhagem As notas cont m informa es importantes destacadas do texto Armazenagem Durante o transporte e a armazenagem dos controllers devem ser satisfeitas as seguintes condi es ambientais temperatura de 20 C a 70 C humidade r
113. dor del dispositivo tanto si ste es la casa fabricante o un distribuidor para poder proveer a la recogida y eliminaci n del producto despu s de haber efectuado una verificaci n de los t rminos y condiciones contractuales de venta TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Mensajes de error Manual de istrucciones 4 En algunos casos de aver a los circuitos de autodiagnosis del controler presenta algunos mensajes de error detallados en la tabla siguiente Tab 1 MENSAJE DESCRIPCI N ACCI N CORRECTIVA CHECK CONNECTION TO PUMP Mal funcionamiento en la conexi n entre la bomba y el Controler 0 bien La bomba tiene una temperatura inferior a 0 C Comprobar gue el cable de conexi n entra en la bomba y el controler est bien fijado por ambos extremos y no est interrumpido Apretar dos veces el pulsador START para volver a poner en marcha la bomba WAITING INTERLOCK Est activa la se al de interlock presente en el conector P1 a causa de la interrupci n del cortocircuito entre el pin 3 y el pin 8 del conector J1 o a causa de la apertura de la se al de interlock externo Eliminar el cortocircuito entre el pin 3 y el pin 8 del conector P1 o cerrar la se al de interlock exterior PUMP OVERTEMP La temperatura del rodamiento superior o de la bomba ha superado los 60 C Esperar a que la temperatura vuelva por debajo del umbral Apretar dos veces el
114. e ao limite estabelecido Premer duas vezes o bot o START para reactivar a bomba TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 73 301 5 Manual de Istru es 74 301 MENSAGEM DESCRI O AC O CORRECTIVA TOO HIGH Durante o funcionamento normal Verificar se o rotor da LOAD ap s a fase de accionamento a bomba pode rodar corrente absorvida pela bomba livremente Premer duas maior do que a programada vezes 0 bot o START para reactivar a bomba SHORT Durante o funcionamento normal Verificar as liga es entre a CIRCUIT ap s a fase de accionamento a bomba e o controller conex o de sa da est em curto Premer duas vezes o bot o circuito START para reactivar a bomba SYSTEM A bomba parou por um sinal de Remover o cabo de OVERRIDE emerg ncia proveniente de um alimenta o do controller e contacto remoto corrigir a causa da emerg ncia Ligar novamente o cabo de alimenta o e premer duas vezes o bot o START para reactivar a bomba OVERVOLTAGE Verificou se um defeito na sec o Premer duas vezes o bot o de alimenta o do controller ou o START para reactivar a controller recebeu um sinal falso bomba Se a mensagem se reapresentar dirigir se Agilent para a manuten o POWER FAIL Avaria na se o de alimenta o da Dirigir se a Agilent para bomba do controller Manuten o TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Ra
115. e pump is increasing the speed 3 Auto tuning the speed has been reduced because the gas load is higher than the Power Limit Applied 4 Braking the pump is reducing the speed using the motor 5 Normal operation the pump is at target speed 6 Fail see error code byte Input Buffer Byte 27 286 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 5 1 Error Code bit field 206 B7 Too High Load the gas load is too high for the pump Bs Short Circuit a short circuit happened between the two of the three motor phases Bs Over Voltage too high bus voltage during braking Ba Verify Line Select Not used B3 Power Fail internal circuitry failure B2 Controller Over Temperature Controller Power Section Temperature or Controller CPU Section Temperature are over the limit Bi Pump Over Temperature Pump Bearing Temperature or Pump Body Temperature are over the limit Bo Check Connection To Pump the connection between controller and pump missing 6 1 Bo Set Point R1 Status 221 0 open 1 closed Bi Set Point R2 Status 225 0 open 1 closed B gt Vent Valve Status 122 0 closed 1 open B3 Purge Valve Status 145 0 close 1 open B4 Not Used Bs Not Used Bs Not Used B7 Not Used TwisTorr 304 FS AG Rack Controller
116. e suljettu laippasulkijalla Sulkimen J1 liitin tulee j tt yhdyskaapelilla kytketyksi mik li ulkoisia kytkent j ei suoriteta Esityhji pumppu ja TwisTorr 304 FS pumppu voidaan k ynnist samanaikaisesti 128 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Ohjek sikirja 10 Valvojan S timet Osoittimet ja Liittimet Seuraavassa on kuvattu valvojan etu ja takapaneelit Tarkempia lis tietoja saatte kappaleesta Technical Information Etupaneelin kuvaus 5 1 3 1 menu A TwisTorr 304 FS AG O Kuva 2 Valvojien TwisTorr 304 FS AG Rack etupanelit 1 Painonappi jonka avulla n ytt n saadaan kiertonumero kiertoaika ja pumpun toiminta 2 LOW SPEED moodin valintaan k ytett v painonappi Se aktivoituu vain silloin kun etupaneelin toiminta moodi on valittu Yhden kerran painettaessa pumppu kiert valmiustila nopeudella Painonappia uudelleen painettaessa LOW SPEED moodi disaktivoituu 3 Painonappi jonka avulla n ytt n saadaan pumpun s hk virran pumpun l mp tilan pumpun tehon ja kiertonopeuden parametrit T m on aina aktiivi riippumatta valitusta toimintamoodista Painamalla painonappeja 3 ja 1 v hint n kahden sekunnin ajan aktivoituu ohjelma jonka avulla on mahdollista ohjelmoida joitakin toimintaparametrej 4 START STOP RESET K YNNISTYKSEN PYS YTYKSEN UUDELLEN AS
117. eduj ce sign ly 3132418138 Signal nebezpe ie upozor uje pracovn ka na ur it pracovn postup ktor ke nie je preveden spr vne mo e zapri init v ne osobne poranenia POZOR Sign ly pozor s pred t mi postupmi ktor v pr pade ich nedodr ania m u po kodit pr stroj Pozn mky zahr uj d le it inform cie vytiahnut z textu Usklad ovanie Po as prepravy a usklad ovania Controllers musia by dodr an tieto podmienky prostredia teplota od 20 C do 70 C relat vna vlhkos 0 95 bez kondenz tu TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 173 301 14 N vod na Obsluhu Priprava pre in tal ciu Controller je dodan v peci lnom ochrannom obale keby ste objavili kodu sp soben prepravou skontaktujte sa s kancel riou miestneho predaja Pri rozbal ovani d vajte pozor aby v m Controller nespadol alebo niekde nenarazil Obal nezahadzujte Je vyroben z recyklovateln ho materi lu a zodpoved smernici CEE 85 399 pre ochranu ivotn ho prostredia Postava 1 Obal Controller 174 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod na Obsluhu 14 In tal cia Na ochranu bezpe nosti u vate a controller je ur en v hradne na vn torn pou itie a mus by napojen cez nap jac k bel na 3 konduktory vi tabu ku objednate n ch n hradn
118. eghaladja a programozott forg r sze szabadon foroghat rt ket e A szivatty jraind t s hoz k tszer megnyomni a START nyom gombot SHORT Norm l zem k zben az ind t si Ellen rizni az CIRCUIT f zist k vet en a kimeneti sszek ttet seket a szivatty sszek ttet s z rlatos s a vez rl k z tt A szivatty jraind t s hoz k tszer megnyomni a START nyom gombot SYSTEM A szivatty le llt egy Kih zni a vez rl h l zati OVERRIDE t v rintkez t l sz rmaz v szjel csatlakoz j t s megsz ntetni hat s ra a v szjelz s ok t jra csatlakoztatni a h l zati k belt s k tszer megnyomni a START nyom gombot a szivatty jraind t s hoz OVERVOLTAGE Meghibasodas t rt nt a vez rl A szivatty jraind t s hoz t pell t si r szegys g ben vagya k tszer megnyomni a START vez rl egy hamis jelet kapott nyom gombot Ha az iizenet ism tl dik a Agilent t l karbantart si beavatkoz st k rni POWER FAIL Meghib sod s a szivatty vagy a A Agilent t l karbantart si vez rl t pell t si r szegys g ben beavatkoz st k rni TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 12 Podrecznik Instrukcji Informacje ogolne 148 Magazynowanie 149 Przygotowanie do instalacji 150 Instalacja 151 Uzytkowanie 152 Sterowniki Wskazniki Laczniki 153 Opis pulpitu frontalnego 153 Procedure uzytkowan
119. eige der Parameter cycle number cycle time und pump life 2 Taste f r die Einstellung des Modus LOW SPEED Sie ist nur aktiv wenn der Steuermodus auf Fronttafel eingestellt ist Bei einmaligem Dr cken l uft die Pumpe in Bereitschafts Geschwindigkeit Bei nochmaligem Dr cken wird der Modus LOW SPEED deaktiviert 3 Taste fur die Anzeige der Parameter pump current pump temperature pump power und rational speed Sie ist immer aktiv unabh ngig vom gew hlten Betriebsmodus Bei gemeinsamer mindestens 2 Sekunden langer Bet tigung der Tasten 3 und 1 wird ein Programm aktiviert mit dem einige Betriebsparamter programmiert werden k nnen 4 START STOP RESET Taste Sie ist nur aktiv wenn der Steuermodus auf der Fronttafel eingestellt ist Bei einmaligem Dr cken wird die Startphase aktiviert Bei nochmaligem Dr cken stoppt die Pumpe Bei automatischem Pumpenstopp durch St rung mu diese Taste ein erstes Mal zur Controller R cksetzung und dann ein zweites Mal zum Neustarten der Pumpe gedr ckt werden 5 Alphanumerisches Fl ssigkristall Display Punkt matrix 2 Zeilen mit 16 Stellen TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 33 301 2 Gebrauchsanleitung 8 6 2 Abbildung 3 R cktafel der Controller Twis Torr 304 FS AG Rack 1 Eingangsverbinder der logischen Signale der Kupplungsverbinder
120. el eller s har START for starte pumpen styreenheten f tt et falskt signal igjen Vises feilmeldingen om igjen m du ta kontakt med Agilent for n dvendig vedlikehold POWER FAIL Feil p matesektoren p pumpen Henvend deg til Agilent for p styreenheten vedlikehold TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 10 Ohjek sikirja Yleisi tietoja 124 Varastointi 125 Valmistelut asennusta varten 126 Asennus 127 K ytto 128 Valvojan S timet Osoittimet ja Liittimet 129 Etupaneelin kuvaus 129 K ytt toimenpiteet 131 Valvojan k ynnistys 131 Pumpun k ynnistys 131 Pumpun pys ytt minen 131 Huolto 131 H vitt minen 132 Vianetsint 133 Alkuper isen ohjeiden k nn s tie Agilent Technologies 123 301 10 Ohjek sikirja Yleisi tietoja T m laite on tarkoitettu ammattimaiseen k ytt n Ennen laitteen k ytt nottoa k ytt j n tulee lukea huolellisesti mukana seuraava k ytt ohje sek kaikki muut Agilentin toimittamat lis tiedot Agilent ei vastaa seurauksista jotka johtuvat laitteen k ytt ohjeiden t ydellisest tai osittaisesta laiminly misest ammattitaidottomien henkil iden suorittamasta laitteen virheellisest k yt st valtuuttamattomista toimenpiteist tai maakohtaisten s d sten ja normien vastaisesta k yt st Sarjan TwisTorr 304 FS AG Rack valvojat ovat mikroprosessoreiden
121. elativa 0 95 n o condensante TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 65 301 5 Manual de Istru es Preparac o para a instalac o O controller fornecido numa embalagem protectora especial se apresentar sinais de danos gue poderiam verificar se durante o transporte entrar em contacto com o escrit rio de vendas local Durante a retirada da embalagem tomar muito cuidado para n o deixar cair o controller e para n o submet lo a choques N o depositar a embalagem no meio ambiente O material completamente recicl vel e responde directriz CEE 85 399 para a protec o do meio ambiente Figura 1 Embalagem dos controllers 66 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de Istru es 5 Instalac o 0 controller foi concebido para utilizac o no interior e deve ser ligado com um cabo de alimentac o de 3 fios ver tabela de pecas gue pode encomendar e ficha aprovada internacionalmente para garantir a seguranca do utilizador Use este cabo de alimentac o e ficha em conjunto com uma tomada com a devida ligac o a terra para evitar chogues el ctricos e para satisfazer os reguisitos da CE A alta voltagem gerada no controller pode provocar ferimentos graves ou mesmo a morte Antes de efectuar gualguer operac o de instalac o ou manutenc o do controller desligar a tomada de alimentac o O controller pode ser instalado numa mesa ou n
122. elet p styreenheten TwisTorr 304 FS AG Rack 1 Trykknapp for vise parametrene cycle number cycle time og pump life p displayet 2 Trykknapp for innstilling av LOW SPEED drift Trykknappen fungerer kun n r det er valgt styrem te fra frontpanelet Ved trykke p knappen n gang roterer pumpen i stand by hastighet Trykk p knappen igjen for g ut av LOW SPEED drift 3 Trykknapp for vise parametrene pump current pump temperature pump power og rotational speed p displayet Trykknappen fungerer alltid uavhengig av styrem ten N r trykknappene 3 og 1 trykkes ned samtidig i minst to sekunder startes et program som tillater programmering av visse driftsparametre 4 Trykknapp for overfgring av START STOP RESET Trykknappen fungerer kun n r det er valgt styrem te fra frontpanelet Trykk p knappen en gang for starte pumpen trykk p knappen igjen for stanse pumpen Har pumpen stanset automatisk i forbindelse med en feil trykk en gang p denne knappen for tilbakestille styreenheten en gang til for startet pumpen 5 LCD display med siffer og bokstaver punkt matrise 2 rader med 16 tegn TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 117 301 9 Instruksjon Manual 8 6 2 Naj D el Figur 3 Bakpanelet p styreenheten TwisTorr 304 FS AG Rack 1 Inngang
123. else af frontpanel 93 Instruktion 95 Start af styreenheden 95 Start af pumpen 95 Stop af pumpen 95 Vedligeholdelse 95 Bortskaffelse 96 Fejlmeddelelser 97 Oversattelse af originalinstruktionerne dg Agilent Technologies 37 1 7 Istruktionsbog Generel Information 88 301 Dette materiel er beregnet til professionel anvendelse Brugeren b r l se denne brugsanvisning og anden yderligere information fra Agilent f r udstyret anvendes Agilent tager ikke ansvar for skader helt eller delvis som f lge af tilsides ttelse af disse instruktioner fejlagtig brug af personer uden tilstr kkelig kendskab ukorrekt anvendelse af udstyret eller h ndtering der strider imod g ldende lokale regler Styreenhederne i TwisTorr 304 FS AG Rack serien er mikroprocessorstyrede frekvens omformere der best r af komponenter med fast tilstand Styreenhed specifikationer Frontal fjern seriel styring 24 Vdc styring af pumpens bl ser Vent valve styring a Afl sning af pumpens hastighed som f lge af tryk p stop afl sning af stophastighed Afleesning af tryk Profibus brugerflade ekstra Automatisk indstilling af indgangsspendingen De f lgende afsnit indeholder al information der beh ves for at garantere operat rens sikkerhed under anvendelsen Detaljeret information findes i bilaget Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruktionsbog 7 I brugsanvisningen anve
124. elser f r installation 102 Installation 103 Anvandning 104 Instruktioner f r bruk 107 Underh ll 107 Bortskaffning 108 Felmeddelanden 109 Instruksjon Manual 111 Generell informasjon112 Lagring 113 Forberede installasjonen 114 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Contents Installasjon 115 Bruk 116 Instruksjoner for bruk 119 Vedlikehold 119 Eliminering 120 Feilmeldinger 121 10 Ohjek sikirja 123 Yleisi tietoja 124 Varastointi 125 Valmistelut asennusta varten126 Asennus 127 K ytt 128 K ytt toimenpiteet 131 Huolto 131 H vitt minen 132 Vianetsint 133 11 Felhaszn l i Kezikonyv 135 ltal nos inform ci k 136 T rol s 137 A telepitesre val el k sz t s138 Telep t s 139 Haszn lat 140 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 9 301 Contents 10 301 Haszn lati elj r sok 143 Karbantart s 143 Megsemmisit s 144 Hiba zenetek 145 Podrecznik Instrukcji 147 Informacje ogolne 148 Magazynowanie 149 Przygotowanie do instalacji 150 Instalacja 151 Uzytkowanie 152 Procedure uzytkowania 155 Konserwacja 155 Przetworstwo odpadow 156 Bledne informacje 157 N vod k Pouziti 159 V eobecn informace 160 Uskladn n 161 P prava k instalaci 162 Instalace 163 Pouziti 164 Pou van procedury167 dr ba 167 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Contents Likvidace 168 Chybn zpr v
125. en cas d inobservation totale ou partielle des instructions donn es d utilisation incorrecte de la part d un personnel non form d operations non autoris es ou d un emploi contraire aux r glementations nationales sp cifigues Les contr leurs de la s rie TwisTorr 304 FS AG Racksont des convertisseurs de fr guence contr l s par un microprocesseur r alis s avec des l ments l tat solide et ayant des capacit s d autodiagnostic et d autoprotection Caract ristiques du contr leur Caract re op rationnel de front distance s riel a Pilotage 24 Vdc du ventilateur de refroidissement pompe Pilotage vent valve a Lecture vitesse pompe suite la commande d arr t lecture vitesse d arr t Lecture de la pression Interface Profibus facultatif Attribution automatigue tension d entr e Les paragraphes suivants donnent toutes les indications n cessaires garantir la s curit de l op rateur pendant l utilisation de l appareillage Des renseignements plus d taill s se trouvent dans l appendice Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Emmagasinage 3 Cette notice utilise les signes conventionnels suivants NEA AN Les messages d avertissement attirent l attention de l op rateur sur une proc dure ou une manoeuvre sp ciale qui si elle n est pas effectu e correctement risque de provoquer de graves l sions Ward ir 5 Les messages d att
126. ention apparaissent avant certaines proc dures qui si elles ne sont pas observ es pourraient endommager s rieusement l appareillage Les notes contiennent des renseignements importants isol s du texte Emmagasinage Pendant le transport et l emmagasinage des contr leurs il faudra veiller respecter les conditions environnementales suivantes temp rature de 20 C 70 C a humidit relative de 0 95 non condensante TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 41 301 3 Mode d emploi Preparation pour l installation Le contr leur est fourni dans un emballage de protec tion sp cial si l on constate des marques de dommages pouvant s tre produits pendant le transport contacter aussit t le bureau de vente local Pendant l op ration d ouverture de l emballage veiller tout particulierement a ne pas laisser tomber le contr leur et ne lui faire subir aucun choc Ne pas jeter emballage dans la nature Le mat riel est enti rement recyclable et il est conforme aux directives CEE 85 399 en mati re de protection de l environnement Figure 1 Emballage du Contr leur 42 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Installation 3 Installation Pour la s curit de l utilisateur le contr leur ne doit tre utilis qu en int rieur et doit tre branch au moyen d un c ble d alimentation 3 fils cf tableau des pieces de rec
127. entspricht der EG Richtlinie 85 399 f r Umweltschutz Abbildung 1 Verpackung der Controller 30 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Gebrauchsanleitung 2 Installation WARNUNG Zum Schutz des Bedieners darf der Controller nur im Geb ude zum Einsatz kommen und nicht au erhalb Er muss mit einem dreiadrigen Netzkabel siehe Tabelle bestellbares Zubeh r und dem international zugelassenen Stecker angeschlossen werden Es sollte immer dieses Netzkabel benutzt werden das an eine korrekt geerdete Steckdose anzuschlie en ist um den CE Richtlinien zu entsprechen und Stromschl ge zu vermeiden Im Inneren des Controllers entstehen hohe Spannungen die schwere Sch den verursachen und zum Teil lebensgef hrlich sein k nnen Vor jedem Montage bzw Wartungseingriff mu deshalb der Netzstecker gezogen werden Der Controller kann auf einen Tisch oder ein Gestell montiert werden In beiden F llen mu auf die ungehinderte Zirkulation der K hlluft im Bereich des Ger ts geachtet werden Der Controller darf nicht in Umgebungen installiert u o benutzt werden die Witterungseinfl ssen Regen Frost Schnee Staub und aggressiven Gasen ausgesetzt sind und in denen Explosions und erh hte Brandgefahr besteht Beim Betrieb m ssen folgende Umgebungsbedingungen eingehalten werden a Temperatur von 5 C bis 45 C a Relative Luftfeuchtigkeit O 95 nicht kondensierend TwisTor
128. esplosivi o con elevato rischio di incendio Durante il funzionamento necessario che siano rispettate le seguenti condizioni ambientali temperatura da 5 C a 45 C a umidit relativa 0 95 non condensante Per gli altri collegamenti e l installazione degli accessori opzionali vedere la sezione Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 19 301 1 Istruzioni per uso Uso In guesto paragrafo sono riportate le principali procedure operative Per ulteriori dettagli e per procedure che coinvolgono collegamenti o particolari opzionali fare riferimento al paragrafo Use dell appendice Technical Information Prima di usare il controller effettuare tutti i collegamenti elettrici e pneumatici e fare riferimento al manuale della pompa collegata AVVERTENZA Per evitare danni alle persone ed all apparato nel caso in cui la pompa sia appoggiata su di un tavolo assicurarsi che sia stabile Non fare funzionare mai la pompa se la flangia di ingresso non collegata al sistema o non chiusa con la flangia di chiusura Il connettore di richiusura J1 deve essere lasciato collegato con il suo ponticello se non viene effettuato alcun collegamento esterno La pompa di pre vuoto e la pompa TwisTorr 304 FS possono essere accese contemporaneamente 20 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruzioni per l uso 1 Comandi Indicatori e Con
129. et Point Mask sec default 0 139 R W L Gauge Set Point Segnal Activation Type high level active default 1 low level active 140 R W N Gauge Set front Mysteresis in of R2 default 2 Valve 143 R W N External Fan Configuration 0 ON 1 automatic 2 serial default 2 144 R W L External Fan Activation 0 OFF TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 245 301 17 Technical Information 1 ON default 0 147 R W N Vent open time 0 infinite See vent connector paragraph 1 bit 0 2 sec 155 R N Power limit applied watt Read the maximun allowable power 157 R W N Gas load type 0 N2 Select the gas load to the pump 1 Av 161 R W N Pressure Reading Connection Factor 0 to 10 See GAUGE CONNECTOR for more details 0 0 10 1 N2 162 R W A R1 Set Point Pressure Valve Valid if min 101 lt 4 Format X X E XX Where X 0 to 9 s or 163 R W N Pressure unit of measure 0 mBar 1 Pa 2 Torr 167 R W A Stop speed reading 0 disable Activates deactivates the pump speed 1 enable reading after Stop command 171 R W N R2 Set Point Type 0 Freq 1 Power 2 Time 3 Normal default 3 4 Pressure available only if the gauge is connected 172 R W N R2 Set Point Value Hz W s 173 R W N R2 Set Point Mask sec 174 R W L R2 Set Point Segnal Activation Type high level active 1 low level active 175 R W N R2 Set front Mysteresis in of
130. ettura velocita di arresto Lettura della pressione Interfaccia Profibus opzionale Impostazione automatica tensione di ingresso Nei paragrafi seguenti sono riportate tutte le informazioni necessarie a garantire la sicurezza dell operatore durante l utilizzo dell apparecchiatura Informazioni dettagliate sono fornite nell appendice Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruzioni per l uso 1 Questo manuale utilizza le seguenti convenzioni NAJ FANN messaggi di avvertenza attirano l attenzione dell operatore su una procedura o una pratica specifica che se non eseguita in modo corretto potrebbe provocare gravi lesioni personali aaia Aoa messaggi di attenzione sono visualizzati prima di procedure che se non osservate potrebbero causare danni all apparecchiatura Le note contengono informazioni importanti estrapolate dal testo immagazzinamento Durante il trasporto e Vimmagazzinamento dei controller devono essere soddisfatte le seguenti condizioni ambientali temperatura da 20 C a 70 C umidit relativa 0 95 non condensante TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 17 301 1 Istruzioni per uso Preparazione per l installazione Il controller viene fornito in un imballo protettivo speciale se si presentano segni di danni che potrebbero essersi verificati durante il trasporto contattare l ufficio ve
131. formation Innan styrenheten anv nds b r samtliga elektriska och pneumatiska anslutningar utf ras L s bruksanvisningen f r den anslutna pumpen F rs kra dig om att pumpen st r stadigt om den r installerad p ett bord detta f r att f rebygga skador p apparaten och personer S tt aldrig ig ng pumpen om intagsfl nsen varken r kopplad till systemet eller r blockerad p plats med l sfl nsen St ngningskontakten J1 m ste l mnas ansluten med aktuell brygga om ingen extern anslutning utf rs F rvakuumpumpen och TwisTorr 304 FS pumpen kan fungera samtidigt 104 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bruksanvisning 8 Indikatorer och Kontakterp Styrenheten Nedan visas styrenhetens front och bakpanel F r ytterligare information h nvisas till bilaga Technical Information Beskrivning av frontpanel Figur 2 Frontpanelen p Styrenheten TwisTorr 304 FS AG Rack 1 Tryckknapp f r att h mta parametrarna cycle number cycle time och pump life p displayen 2 Tryckknapp f r inst llning av LOW SPEED drifts tt Tryckknappen fungerar endast n r styrs tt fr n frontpanelen har valts Genom att trycka p knappen en g ng kommer pumpen att rotera vid standby hastighet Tryck p knappen igen f r att g ur LOW SPEED drifts ttet 3 Tryckknapp f r att h mta parame
132. gar a bomba ocorre habilitar a entrada de interlock e apertar o bot o START do painel frontal Paragem da bomba Para parar a bomba necess rio premer o bot o STOP do painel frontal Manuten o Os controllers da s rie TwisTorr 304 FS AG Rack n o requerem qualquer manuten o Todas as opera es devem ser efectuadas por pessoal autorizado Em caso de defeito poss vel utilizar o servi o de repara o Agilent ou o Agilent advanced exchange service que permite obter um controller regenerado que substitua o controller com defeito UV SVEN Antes de efectuar qualquer opera o no controller desligar o cabo de alimenta o Caso um controller deva ser destru do proceder sua elimina o respeitando as normas nacionais espec ficas TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 71 301 5 Manual de Istru es Eliminac o 72 301 Significado do log tipo WEEE presente nos r tulos O s mbolo abaixo indicado aplicado de acordo com a directiva CE denominada WEEE Este s mbolo v lido apenas para os pa ses da Comunidade Europeia indica gue o produto no gual est aplicado N O deve ser eliminado juntamente com os residuos dom sticos ou industriais comuns mas deve ser dirigido a um sistema de recolha diferenciada Portanto convidamos o utilizador final a contactar o fornecedor do dispositivo seja este o fabricante ou um revendedor para encaminhar o processo de recolha e
133. ha recibido una se al espurio Apretar dos veces el pulsador START para volver a poner en marcha la bomba Si el mensaje se vuelve a presentar dirigirse a Agilent para el mantenimiento POWER FAIL Averia en la secci n de alimentaci n de la bomba del controler Ponerse en contacto con Agilent para el Mantenimiento 62 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual Manual de Istruc es Informa es gerais 64 Armazenagem 65 Prepara o para a instala o 66 Instalac o 67 Utilizac o 68 Comandos Indicadores e Conectores 69 Descric o painel frontal 69 Procedimentos de uso 71 Acendimento do Controller 71 Activac o da bomba 71 Paragem da bomba71 Manutenc o 71 Eliminac o 72 Mensagens de erro 73 Traduc o das instruc es originais oft Agilent Technologies 63 301 5 Manual de Istru es Informa es gerais 64 301 Esta aparelhagem destina se ao uso profissional O utilizador deve ler atentamente o presente manual de instru es e todas as informa es adicionais fornecidas pela Agilent antes de utilizar a aparelhagem A Agilent n o se responsabiliza pela inobserv ncia total ou parcial das instru es pelo uso indevido por parte de pessoas n o treinadas por opera es n o autorizadas ou pelo uso contr rio s normas nacionais espec ficas Os controllers da s rie TwisTorr 304 FS AG Rack s o conversores
134. hange gui peuvent tre command es avec une fiche du type approuv au niveau international Afin d viter toute d charge lectrigue et satisfaire aux conditions reguises CE il faut toujours utiliser ce c ble d alimentation en introduisant la fiche dans une prise lectrigue pourvue d un branchement appropri la terre A l int rieur du contr leur se d veloppent de hautes tensions gui peuvent causer de graves dommages et m me la mort Avant d effectuer toute operation d installation ou d entretien du contr leur le d brancher de la prise d alimentation Le contr leur peut tre install sur une table ou l int rieur d un rack pr vu cet effet II est en tout cas n cessaire que l air de refroidissement puisse circuler librement I int rieur de I appareil Ne pas installer et ou utiliser le contr leur dans des milieux expos s des agents atmospherigues pluie gel neige des poussieres des gaz de combat ainsi que dans des milieux explosifs ou risgue lev d incendie Pendant le fonctionnement il est n cessaire de respecter les conditions environnementales suivantes temp rature de 5 C 45 C a humidit relative de 0 a 95 non condensante Pour les autres connexions et pour l installation des accessoires en option voir la section Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 43 301 3 Mode d emploi Utilisation Dans ce paragraphe
135. hange product please use the packaging from the Advance Exchange to return the defective product e Sealthe product in a plastic bag and package product carefully to avoid damage in transit You are responsible for loss or damage in transit e Agilent Technologies is not responsible for returning customer provided packaging or containers e Clearly label package with RA number Using the shipping label provided will ensure the proper address and RA number are on the package Packages shipped to Agilent without a RA clearly written on the outside cannot be accepted and will be returned Return only products for which the RA was issued Product being returned under a RA must be received within 15 business days Ship to the location specified on the printable label which will be sent along with the RA number as soon as we have received all of the reguired information Customer is responsible for freight charges on returning product Return shipments must comply with all applicable Shipping Regulations IATA DOT etc and carrier reguirements RETURN THE COMPLETED REOUEST FOR RETURN FORM TO YOUR NEAREST LOCATION EUROPE NORTH AMERICA PACIFIC RIM Fax 00 39 011 9979 330 Fax Free 00800 345 345 00 Fax 1 781 860 9252 please visit our website for individual Toll Free 00800234 234 00 Toll Free 800 882 7426 Option 3 office information vpt customercare agilent com vpl ra agilent com http www agilent com Pg 1 3 v Agilent Te
136. hanges made to the specific parameter Abort is indicated by the caption Changes aborted that is displayed for approx 1 second before moving to the previous parameter The configuration environment has a tree like structure and permits circular navigation on reaching the end ofa branch the user is returned automatically to the start of this Navigation can proceed in a forwards NEXT button or backwards PREVIOUS button direction It is possible to enter a sub branch by pressing the INCREASE button If no buttons are pressed for more than 1 minute the controller automatically quits the programming environment canceling any modifications made to the current parameter The figure below shows the first level of the configuration menu and the links to the respective second level menus indicated below TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 263 301 17 Technical Information 264 301 MODE lt gt FRONT A t REMOTE SERIAL EN a v I I PROFIBUS PUMP SETTING HIGH SPEEDADJ x A HZ 1 wj 1 Y PE sz JE I INPUT OUTPUT lt gt STARTMODE I M __ i Rs Z i li Ir Y nasse a SETPOINT RI SETTING lt x PYRE TRESHOLD I 1 maz y a SETPOINT e R2 SETTING R ESSE P S j PRESSURE UNIT GAUGE gt
137. hase of new or refurbished Products are subject to prior Agilent approval and may incur a restocking fee Please reference the original purchase order number e Units returned for evaluation will be evaluated and a quote for repair will be issued If you choose to have the unit repaired the cost of the evaluation will be deducted from the final repair pricing A Purchase Order for the final repair price should be issued within 3 weeks of guotation date Units without a Purchase Order for repair will be returned to the customer and the evaluation fee will be invoiced e A Special Cleaning fee will apply to all exposed products per Section 4 of this document e If requesting a calibration service units must be functionally capable of being calibrated Pg 3 3 United States Agilent Technologies Vacuum Products Division 121 Hartwell Avenue Lexington MA 02421 USA Tel 1 781 861 7200 Fax 1 781 860 5437 Toll Free 1 800 882 7426 Benelux Agilent Technologies Netherlands B V Vacuum Products Division Herculesweg 8 4338 PL Middelburg The Netherlands Tel 31 118 671570 Fax 31 118 671569 Toll Free 00 800 234 234 00 Canada Central coordination through Agilent Technologies Vacuum Products Division 121 Hartwell Avenue Lexington MA 02421 USA Tel 1 781 861 7200 Fax 1 781 860 5437 Toll Free 1 800 882 7426 China Agilent Technologies China Co Ltd Vacuum Products Division No 3 Wang Jing Bei Lu Ch
138. he signal O high level active can be high level active or low level 1 low level active active default 0 105 R W N Set point hysteresis in of value 0 to 100 default 2 106 R W Water cooling 0 NO 1 YES 107 R W L Active Stop write only in stop 0 NO 1 YES 108 R W N Baud rate 600 0 1200 1 2400 2 4800 3 9600 4 default 4 244 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 109 W L Pump life cycle time cycle number To reset write 1 reset 110 R W L Interlock type default 0 Impulse 0 Continuous 1 111 R W L Analog output type output voltage signal 0 frequency i power proportional to freguency or power default 1 2 Pump temperature 3 Pressure 117 R W N Low Speed frequency Hz 000 to 1010 default 800 120 R W N Rotational frequency setting Hz Low speed frequency win 117 to 1010 default 1010 122 R W L Set vent valve on off on closed On 1 Off 0 default 1 123 Reserved to Agilent service 124 125 R W L Set the vent valve operation Automatic 0 see note 1 On command 1 see note 2 126 R W N Vent valve opening delay expressed in 0 2 0 to 65535 corresponding to sec 0 to 13107 sec 130 Reserved to Agilent service 136 R W N Gauge Set Point Type 0 Freq 1 Power 2 Time 3 Normal default 137 R W N Gauge Set Point Value Hz W s default 867 138 R W N Gauge S
139. he time expressed in seconds of opening of the vent valve Enter 0 if the valve is to be kept open indefinitely until the next start LOAD GAS TYPE selects the type of gas used in the process Therefore the controller limits the power supplied to the pump according to the setting of this parameter TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 WATER COOLING indicates whether the pump is cooled with water or not The controller limits the power supplied to the pump according to this setting ACTIVE STOP enables YES or disables NO active braking STOP SPEED READING enables YES or disables NO the pump rotational freguency reading function also during the slowing phase that follows a stop command TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 267 301 17 Technical Information Input Output Menu This menu contains parameters that determine the operating mode of the inputs and outputs of the controller I I i IMPUT OPTPUT Resi gt STARTMODE ke gt START INTERLOCK I I FIZ A AA I START STOP LEVEL START STOP EDGE v INTERLOCK TYPE gt CONTINUOUS A IMPULSIVE y ANALOG OUT TYPE lt gt FREQUENCY A POWER CURRENT PRESSURE v START OUT MODE 4
140. hese rectangles is as follows LS the controller is at LOW SPEED AG the Gauge is connected to the controller R1 relay set point RI is active R2 gt relay set point R2 is active cm LIFT TT mir AG R2 Figure 31 254 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Use General Make all vacuum manifold and electrical connections and refer to TwisTorr 304 FS pump instruction manual before to operating the TwisTorr 304 FS AG Rack controller WARNING To avoid injury to personnel and damage to the eguipment if the pump is laying on a table make sure it is steady Never operate the TwisTorr 304 FS pump if the pump inlet is not connected to the system or blanked off The input signal P1 connector should be left in position including the shipping links if no external connections are made The forepump and the TwisTorr 304 FS pump can be switched on at the same time TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 255 301 17 Technical Information Startup Plug the controller power cable into a suitable power source The display lights up for approx 2 seconds and shows A U T 0 T E S T After 2 seconds the display shows the following screen page SITIO P IH z If the pump is not
141. ia 155 Zaswiecenie kontrolera 155 Uruchomienie pompy 155 Zatrzymanie pompy155 Konserwacja 155 Przetworstwo odpadow 156 Bledne informacje 157 Az eredeti utasit s fordit sa tie Agilent Technologies 147 301 12 Podrecznik Instrukcji Informacje ogolne 148 301 Urzadzenie to jest przeznaczone do uzytku zawodowego Przed przystapieniem do korzystania uzytkownik powinien wnikliwie przesledzic zarowno ten podrecznik zawierajacy instrukcje jak rowniez kazda inna dostarczona przez Agilent informacje Agilent uchyla sie od jakiejkolwiek odpowiedzialnosci w przypadku calkowitego lub czesciowego nie przestrzegania swoich instrukcji przeznaczenia urzadzenia niezgodnie z jego przeznaczeniem i jego obslugi przez personel uprzednio nie przeszkolony naprawy urzadzenia bez odpowiedniej autoryzacji lub przeznaczenia niezgodnego z obowiazujacymi specyficznymi normami krajowymi Kontrolery z serii TwisTorr 304 FS AG Rack sa przetwornikami czestotliwosci kontrolowanymi przez mikroprocesor zrealizowanymi z komponentow w stanie stalym i posiadajacymi zdolnosci samodiagnostyczne i samoochronne Dane kontrolera a Operatywny frontalnie w izolacji seryjnie Sterowanie turbiny chlodzenia pompy z szybkoscia 24 Vdc Sterowanie vent valve Odczyt szybkosci pracy pompy w momencie nakazu zatrzymania odczyt szybkosci w momencie zatrzymania a Odczyt cisnienia a Posrednik Profibus dodatkowy a Automatyczne ustawienie napiecia wejsciowego
142. in 232 or 485 and if necessary the serial address Select SERIAL from the INPUT OUTPUT menu Open the T software see orderable parts table Ready to operate by SERIAL connections TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Open Stop Controller Cloed Start 1 Start Stop N 2 Low Speed 3 Interlock Imput 4 Connector 5 6 7 8 9 Output Connector Fault 24 VDC 15 Fault 8 Start 24 VDC 13 Start 6 Low Speed 24 VDC 12 Low Speed 5 R2 R1 24 VDC 14 11 R2 R1 7 4 Pump Power 2 Monitor 10V 150W Monitor Com 1 Pump Freguency Monitor 10V 81 KRPM a Open Normal Start Open High Speed Cloed Low Speed Closed Stop Fra Open Allow Start Cloed Soft Start 4K7 M S LA OFF OK ON Fail 4K7 M S AI OFF Stop Normal Operation ON Ramp 4K7 M S L OFF High speed ON Low speed setted 4K7 M OFF Pump under speed threshold ON Pump over speed threshold 7 v Power value 7 V Speed value Figure 44 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 277 301 17 Technical Information Profibus Option General The ProfiBus option allow to communicate with the controller as ProfiBus DP slave device The option works as a bi directional bridge between the ProfiBus network and the Agilent
143. ingscondities aanwezig Zijn temperatuur van 5 C tot 45 C relatieve vochtigheid 0 95 niet condenserend Voor de overige aansluitingen en de installatie van de accessoires wordt verwezen naar het hoofdstuk Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 79 301 6 Bedrijfshandleiding Gebruik In deze paragraaf worden de voornaamste bedieningswijzen uitgelegd Voor meer informatie of procedures die aansluitingen of speciale opties betreffen wordt verwezen naar de paragraaf Use van de bijlage Technical Informations Breng alvorens de controller in gebruik te nemen alle elektrische en pneumatische aansluitingen tot stand en raadpleeg hiervoor de handleiding van de aan te sluiten pomp WAARSCHUWING Indien de pomp op een tafel is geplaatst controleren of deze stabiel staat om letsel aan personen en schade aan het apparaat te voorkomen Laat de pomp nooit werken zonder dat de ingangsflens aan het systeem is gekoppeld of de afsluitflens is gesloten De connector J1 moet met zijn jumper aangesloten blijven als geen externe aansluiting tot stand wordt gebracht De pre vacu mpomp en de TwisTorr 304 FS pomp mogen beide gelijktijdig ingeschakeld zijn 80 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bedrijfshandleiding 6 Bedieningsorganen Controlelampjes en Connectoren van Hieronder worden het frontpaneel en het achterpaneel van de
144. ins identical to the one used for the RS 232 interface the only difference being that the value assigned to the ADDRESS lt ADDR gt The controller can answers with the following response types Tab 14 Response Type Response Response Description Length Value Logic 1 byte after a read instruction of a logic window Numeric 6 bytes after a read instruction of a numeric window Alphanumeric 10 bytes after a read instruction of an alphanumeric window ACK 1 byte 0x6 the command execution has been successfully completed NACK 1 byte 0x15 the command execution has been failed Unknown Window 1 byte 0x32 the specified window in the command is not a valid window Data Type Error 1 byte 0x33 the data type specified in the command Logic Numeric or Alphanumeric is not accorded with the specified Window Out of Range 1 byte 0x34 the value expressed during a write command is out of the range value of the specified window Win Disabled 1 byte 0x35 the specified window is Read Only or temporarily disabled for example you can t write the Soft Start when the Pump is running 240 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Examples Command START Source PC Destination Controller Technical Information 17 02 80 30 30 30 31 31 03 42 33 STX A
145. ird time the following message Is displayed MING XIXIXIXIXIXIXIX X S N YIY Y Y VY Y Y VY Y Y MN Model Number of the controller SN Serial Number On pressing the key again the main screen page is redisplayed Measures Pressing once the following is obtained P X X X W X XIX X H z T X X C TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Where P XXX W is the current power supplied to the pump T XX C is the current temperature of the pump XXXX Hz is the current frequency of rotation of the pump This value is set by default to 963 Hz but can be modified by the user according to needs proceeding as described in the PROGRAMMING paragraph The unit of measurement HZ Krpm default Hz is selected by the user from the INPUT OUTPUT gt SPED menu If the SOFT START function PUMP SETTING menu SOFTSTART is active and the pump is ramping indication of the speed of rotation is replaced with a bar graph that indicates the soft start step in course and the total number of steps to be performed On completion of the soft start ramp the bar graph is replaced with indication of the frequency of rotation of the pump Pressing twice the following is obtained N JO R M A IL X X X X H z I X X X A V lt IX X X
146. javila s signalom zasile Ponovno priklju ite napajalni kabel in pritisnite dvakrat tipko START za ponoven zagon rpalke OVERVOLTAGE Kvar v oddelku za napajanje Dvakrat pritisnite tipko START controllera ali je controller dobil za ponoven zagon rpalke e nepristen znak se to obvestilo spet pojavi obrnite se Agilent za pravilno vzdr evanje POWER FAIL Kvar v oddelku napajanja rpalke Obrnite se Agilent za pravilno controllera vzdr evanje TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 16 Instructions for Use General Information 196 Storage 197 Preparation for Installation 198 Installation 199 Use 200 Controls Indicators and Connectors 201 Front Panel Description 201 Use Procedure 203 Controller Startup 203 Starting the Pump 203 Pump Shutdown 203 Maintenance 203 Disposal 204 Error Messages 205 Original Instructions it Agilent Technologies 195 301 16 Instructions for Use General Information 196 301 This eguipment is destined for use by professionals The user should read this instruction manual and any other additional information supplied by Agilent before operating the eguipment Agilent will not be held responsible for any events occurring due to non compliance even partial with these instructions improper use by untrained persons non authorised interference with the eguipment or any action contrary to that provided for by
147. ki lett v lasztva Egyszer megnyomva a szivatty stand by sebess ggel forog M g egyszer megnyomva a LOW SPEED m d kikapcsol sra ker l 3 Nyom gomb a param tereknek szivatty ram a szivatty h m rs klete szivatty teljes tm ny s a forg si sebess g kijelz re val leh v s ra Mindig akt v f ggetlen l a kiv lasztott zemm dt l A 3 as s 1 es nyom gombot egy tt legal bb 2 m sodpercig nyomva tartva egy program ker l aktiv l sra melynek seg ts g vel n h ny zemi param ter programozhat 4 Nyom gomb a START STOP RESET parancsok ind t s ra Csak akkor akt v amikor az el lapr l a parancsm d ki lett v lasztva Egyszer megnyomva az ind t si f zis ker l aktiv l sra m g egyszer megnyomva a szivatty le ll t sra ker l Ha a szivatty valamely hiba k vetkezt ben automatikusan le ll ezt a nyom gombot kell megnyomni az els megnyom sra a vez rl null z dik reset a m sodikra a szivatty jraindul 5 Folyad kkrist lyos alfanumerikus kijelz pontm trixos 2 sor x 16 karakter TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 141 301 11 Felhaszn l i K zik nyv O El OA LJ D EI Abra 3 A TwisTorr 304 FS AG Rack AG sz m vezerlok h tlapja 1 Logikai jelek bemeneti csatlakoz ja az sszek t csatlakoz megfelel lez
148. kommendes Notsignal gestoppt die St rungsursache beheben Das Netzkabel wieder anschlie en und die Pumpe durch zweimalige Bet tigung der START Taste neustarten OVERVOLTAGE Defekt im Versorgungsbereich des Die Pumpe durch zweimalige Controllers bzw der Controller hat Bet tigung der START Taste ein falsches Signal erhalten neustar ten Erscheint die Meldung wieder sollte der Agilent Wartungs dienst gerufen werden POWER FAIL Defekt im Pumpenstromkreis des Agilent f r Wartung Controllers kontaktieren TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual Mode d emploi Indications generales 40 Emmagasinage 41 Preparation pour l installation 42 Installation 43 Utilisation 44 Commandes Indicateurs et Connecteurs 45 Description du Tableau avant 45 Procedures d utilisation 47 Allumage du Contr leur 47 Mise en marche de la Pompe 47 Arr t de la Pompe 47 Entretien 47 Mise au rebut 48 Messages d erreur 49 Traduction de la mode d emploi originale pp Agilent Technologies 39 301 3 Mode d emploi Indications generales 40 301 Cet appareillage a t concu en vue d une utilisation professionnelle Il est conseill l utilisateur de lire attentivement cette notice d instructions ainsi gue toute autre indication suppl mentaire fournie par Agilent avant l utilisation de l appareil Agilent d cline par cons guent toute responsabilit
149. kterp Styrenheten 105 Beskrivning av frontpanel 105 Instruktioner f r bruk 107 Start av styrenheten 107 Start av pumpen 107 Stopp av pumpen 107 Underh ll 107 Bortskaffning 108 Felmeddelanden 109 vers ttning av originalinstruktionerna Rem Agilent Technologies 99 301 Bruksanvisning Allm n Information 100 301 Utrustningen r avsedd f r yrkesm ssig anv ndning Anv ndaren bor l sa denna bruksanvisning samt vrig dokumentation fr n Agilent f re anv ndning av utrustningen Agilent tar inget ansvar f r skador som helt eller delvis orsakats av sidos ttande av instruktionerna ol mplig anv ndning av person utan tillr cklig kunskap obeh rigt bruk av utrustningen eller hantering som strider mot g llande lokala f reskrifter Styrenheterna i TwisTorr 304 FS AG Rack serien r mikroprocessor styrda frekvensomvandlare som best r av komponenter med fast tillst nd Styrenheterna r f rsedda med sj lvdiagnos och sj lvskyddsfunktion Styrenhetens egenskaper a Front fj rr seriell styrning 24 Vdc styrning av pumpens kylfl kt a Vent valve styrning a Avl sning av pumphastighet efter kommando f r stopp avl sning av stopphastighet a Avl sning av tryck a Profibus gr nssnitt option Automatisk inst llning av insp nning De f ljande avsnitten inneh ller all information som beh vs f r att garantera operat rens s kerhet under driften Detaljerade uppgifter finns i bilagan Technical i
150. kynnysarvon Tai alapuolelle Painakaa Valvojan l mmittimen l mp tila on painonappia START kaksi korkeampi kuin 60 C kartaa jolloin pumppu k ynnistyy TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 133 301 10 Ohjek sikirja VIESTI VIKA KORJAUSTOIMENPITEET TOO HIGH Pumpun absorptiovirta ylitt Tarkistakaa ett pumpun LOAD ohjelmoidun arvon roottori py rii vapaasti Painakaa painonappia START kaksi kertaa jolloin pumppu k ynnistyy SHORT Normaality skentelyn aikana Tarkistakaa pumpun ja valvojan CIRCUIT k ynni styksen j lkeen v linen liitos Painakaa poistoliitin on oikosulussa painonappia START kaksi kertaa jolloin pumppu k ynnistyy SYSTEM Kaukokytkimest tuleva Irroittakaa valvojan OVERRIDE h lytyssignaali on pys ytt nyt s hk kaapeli verkkovirrasta ja pumpun korjatkaa h lytyksen aiheuttaja Kytkek s hk kaapeli uudelleen ja painakaa painonappia START kaksi kertaa jolloin pumppu k ynnistyy OVERVOLTAGE Valvojan s hk virran sy t ss on Painakaa painonappia START ilmennyt ongelma tai valvojan kaksi kertaa jolloin pumppu saama signaali on v r k ynnistyy Mik li viesti n kyy uudelleen k ntyk Agilent huoltopalvelun puoleen POWER FAIL Vika pumpun valvojan K ntyk Agilent yhti n virransy tt osassa puoleen Huoltoa tarvittaessa 134 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS
151. ldingen Bedrijfshandleiding 6 In geval van storingen wekt het zelfdiagnose circuit van de controller enkele foutmeldingen op die in de volgende tabel zijn omschreven Tab 1 BOODSCHAP OMSCHRIJVING REMEDIE CHECK CONNECTION TO PUMP Foutieve verbinding tussen pomp en controller Of De pomp heeft een lagere temperatuur dan 0 C Controleren of de verbindingskabel tussen pomp en controller aan beide uiteinden goed bevestigd is en geen onderbrekingen vertoond Bedien twee maal de START knop om de pomp weer op te starten WAITING INTERLOCK Het interlock signaal op connector P1 is actief wegens onderbreking van de kortsluiting tussen pin 3 en pin 8 van connector J1 of wegens het openen van het externe interlock signaal Herstel de kortsluiting tussen pin 3 en pin 8 van connector P1 of sluit het externe interlock signaal PUMP OVERTEMP De temperatuur van het bovenste pomplager bedraagt meer dan 60 C Wacht tot de temperatuur weer onder de drempelwaarde is gezakt Bedien twee maal de START knop om de pomp weer op te starten CONTROLLER OVERTEMP De temperatuur van de controller bedraagt meer dan 70 C Of De temperatuur van de radiator van de controller bedraagt meer dan 60 o C Wacht tot de temperatuur weer onder de drempelwaarde is gezakt Bedien twee maal de START knop om de pomp weer op te starten TwisTorr 304 FS AG Rack Controller User Manual 87
152. lep t si vagy karbantart si m velet nek megkezd se el tt ki kell h zni a h l zati csatlakoz t A vez rl telep t sre ker lhet egy asztalon vagy egy megfelel rack belsej ben Minden esetben sz ks ges azonban hogy a h t leveg szabadon ramolhasson a berendez s belsej ben A vez rl t nem szabad l gk ri hat soknak es fagy h kitett poros mar hat s g zt tartalmaz robban s vagy fokozottan t zvesz lyes k rnyezetben telep teni vagy haszn lni zem k zben az al bbi k rnyezeti felt teleket kell biztos tani h m rs klet 5 C s 45 C k z tt relativ nedvess gtartalom 0 95 nem lecsap d A t bbi sszek t ssel s az opcion lis kieg sz t felszerel sek telep t s vel kapcsolatosan l sd a Technical Information c m alfejezetet TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 139 301 11 Felhaszn l i K zik nyv Haszn lat Ebben a szakaszban a fontosabb m k d si elj r sok ker lnek ismertet sre Tov bbi r szletek s az sszek t seket vagy az opcion lis elemeket rint elj r sok tal lhat k a Technical Information mell klet Use c m szakasz ban A vez rl haszn lata el tt valamennyi elektromos s pneumatikus sszek ttet st l tre kell hozni s figyelembe kell venni a csatlakoztatott szivatty k zik nyv t VESZ LY Abban az esetben ha a szivatty egy asztalon ker lt elhel
153. ler User Manual 87 901 029 01 Technical Information AI the logic output from the controller must be connected at P2 mating connector The following table describes the signals available on the connector The following figure shows a typical logic output connection relay coil but any other device may be connected e g a LED a computer etc and the related simplified circuit of the controller CONTROLLER Figure 12 Digital Output Connection Example 1 gt 100 KO 2 Figure 13 Analog Output Connection Example 17 217 301 17 Technical Information Tab 4 Pins Name Description Electrical Note 15 8 Fault OV no fail Optically isolated Connect a 10kohm 24V fail Vmax open 24v resistor between the Vlow max 1 5v 2 pins before Imax 60mA measuring voltage Pin 8 negative A STOP command Pin 15 positive by front Panel or serial or Start stop input reset the FAIL status 13 6 Start Configurable output Optically isolated Connect a 10kohm Default setting win 177 0 Vmax open 24v resistor between the 24V pump in RAMP state OV Vlow max 1 5v 2 pins before other state If win 177 1 Imax 60mA measuring voltage OV pump in stop or FAIL Pin 6 negative status 24V other state Pin 13 positive 12 5 Low Speed 24V low speed active Optically isolated Connect a 10kohm OV low speed not active Vmax open 24v resistor between the Vlow max 1 5v 2
154. ler ne pade na tla ali da ne podlegne udarcem Ne zavrzite embala e v okolico Material lahko v celoti reciklirate in ustreza direktivi CEE 85 399 za za ito okolja Slika 1 Embala a controllera 186 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Priro amp nik za Navodila 15 Monta a Za bolj o varnost uporabnika iController je namenjen le za notranjo uporabo in napaja napajalni kabel s 3 dovodi glej tudi razpredelnico za nadomestne dele ki se lahko naro ijoli z vtika em z medanrodno oznako Da bi se izognili elektri nim stresljajem in da bi ravnali v skladu s predpisi ES je potrebna uporaba takega napajalnega kabla potrebno je vtkaniti vtika v stikalo z ustrezno ozemljitvijo Znotraj controller jev se stvori visoka napetost ki lahko loveku povzro i hude po kodbe ali celo smrt Preden bi nadaljevali s katerokoli dejavnostjo monta e ali vzdr evanja controllera izklopite controller iz vti nice Controller lahko montirate na mizo ali vgradite v notranjost ustreznega rack a V vsakem slu aju morate poskrbeti da voda za hlajenje prosto kro i skozi napravo Controller a ne smete montirati ali uporabiti v okolici ki je izpostavljena klimatskim dejavnikom de ju ledu snegu prahu agresivn m plinom v okolici kjer obsatja nevarnost vnetja ali po ara Med delovanjem morate zagotoviti naslednje pogoje okolice a temperatura od 5 C do 45 C a relativna vlaga 0 9
155. leurs TwisTorr 304 FS AG Rack 1 Interrupteur rappelant sur I afficheur les parametres de cycle number cycle time et pump life 2 Interrupteur de selection du mode LOW SPEED II n est actif gue lorsgue le mode de commande est selectionne depuis le tableau frontal En le pressant une fois la pompe tourne vitesse stand by En le pressant une deuxieme fois on d sactive le mode LOW SPEED 3 Interrupteur rappelant sur l afficheur les param tres de pump current pump temperature pump power et rotational speed II est toujours actif ind pendamment du mode de fonctionnement choisi En pressant simultanement les interrupteurs 3 et 1 pendant 2 secondes au moins on active un programme avec leguel il est possible de programmer certains param tres op rationnels 4 Interrupteur envoyant les commandes de START STOP RESET II n est actif que lorsgue le mode de commande est s lectionn depuis le tableau frontal Une premiere pression de l interrupteur active la phase de mise en marche une deuxi me pression provoque l arr t de la pompe Si la pompe s est arr t e automatiguement cause d une panne il faut presser cet interrupteur une premiere fois pour effectuer la mise zero du contr leur et une deuxieme fois pour remettre la pompe en marche 5 Ecran alphanumerigue cristaux liguides matrice de points 2 lignes x 16 caract res TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 45 301 3 Mode d e
156. ljanje vent valve Branje hitrosti rpalke po ukazu zaustavitve branje zaustavitvene hitrosti a Branje pritiska Vmesnik Profibus opcijski Avtomati na postavitev vhodne napetosti V naslednjih odstavkih so navedene vse informacije potrebne za zagotavljanje sigurnosti uporabnika med rokovanjem s to napravo Podrobnej e informacije dobite v priponki Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Priro amp nik za Navodila 15 V tem priro niku so varnostne informacije razvr ene v dva razreda SVARILO Znak svarila opozarja uporabnika da mora pri dolo enem postopku ali pri posebnem delovanju paziti Kr itev svarila lahko privede do la jih ali hudih telesnih po kodb POZOR Znak pozor se pojavi pred postopke ki e jih uporabnik zanemari lahko napravo po koduje Opombe vsebujejo vse najbolj pomembne informacije iz besedila Skladi enje Med prevozom in skladi enjem controllerjev morate zagotoviti naslednje pogoje okolice temperatura od 20 C do 70 C relativna vlaga 0 95 bez kondenzacije TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 185 301 15 Priro nik za Navodila Priprava za monta o Controller dobite v posebni za itni embala i e so na embala i znaki po kodbe ki naj bi nastale med transportom obrnite se krejavnemu prodajalnemu biroju Pri odstranitvi embala e morate e posebno paziti da control
157. lt configuration data doesn t match B3 ext diag slave has external diagnostic data B4 not supported slave doesn t support requested function Bs invalid slave response slave sets permanent 0 Be prm fault wrong parameter assignment B7 master lock slave is parameterised by another master 1 Bo prm reg slave has to be re parameterised B1 stat diag static diagnosis 8 Ba fixed 1 gt B3 wd on response monitoring active 5 B Freeze mode received freeze command S Bs Synch mode received synch command S Be reserved a B7 diag deactivated slave is parameterised by another master 2 Reserved 3 Master Add master address after parameterisation FF without parameter 4 Ident number high 5 Ident number low 6 External diagnosis Header length indication including header TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 External diagnostic OFF SET MEANING 7 Interface Status Bo Controller not responds B1 Controller not identified B2 Input data not fully coherent yet B3 Not used B47 Not used 8 B74 Start Stop diag status B30 Low Speed diag status 9 B74 Soft Start diag status B30 Vent Valve 10 B74 Purge Valve B3 0 Mode 11 B74 High speed B30 Low Speed TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 291 301 17 Technical Information Error Messages For a certain type of failure the cont
158. mp Check that pump rotor is free to rotate then press the STOP RESET pushbutton twice to start the pump If the upper bearing pump temperature exceeds 60 C the pump is shut off and the display will be as shown in the following figure P U M P O V E R T E M P The message will stay on until the temperature decreases below threshold value Press the STOP RESET pushbutton twice to start the pump If the controller transformer temperature exceeds 90 C the pump is shut off and the display will be as shown in the following figure TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 293 301 17 Technical Information The message will stay on until the temperature decreases below threshold value Press the STOP RESET pushbutton twice to start the pump PJOJWJEJR FIA TIL Indicates a fault in the circuit of the controller used to power the pump If the pump is stopped by an emergency stop signal provided via a remote contact the display will be as shown in the following figure 294 301 S Y IS T E M O V E R R I D E Remove the controller power cable and check the emergency condition Then reconnect the power cable and press the START pushbutton to start the pump
159. mploi Jo ojo 23 So So So So So o o ONO of Figure 3 Tableau arri re du Contr leur TwisTorr 304 FS AG Rack Connecteur d entr e des signaux logigues le connecteur d enclenchement est dot de la barrette de r enclenchement sp ciale 2 Connecteur signaux logigues en sortie et monitorage fr guence pompe plus sortie analogigue programmables 3 Connecteur de sortie alimentation 24 Vdc pour le ventilateur de refroidissement pompe vent valve et calibre 4 Module d entr e alimentation contr leur gui comprend les fusibles la prise d alimentation et le filtre ENC 5 Interrupteur rotation pour tablir l adresse du Profibus 6 Profibus facultatif 7 Connecteur s riel pour contr le a distance 8 Connecteur pompe 46 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Procedures d utilisation 3 Procedures d utilisation Allumage du Contr leur Pour allumer le contr leur il suffit d introduire le c ble d alimentation dans la prise du r seau Mise en marche de la Pompe Pour faire d marrer la pompe il faut habiliter Ventr e de interlock et appuyer sur la touche START du panneau avant Arr t de la Pompe Pour arr ter la pompe presser l interrupteur STOP du tableau frontal Entretien Les contr leurs de la s rie TwisTorr 304 FS AG Rack n exigent aucun entretien
160. n PC 02 83 3 13 3 3 3 3 3 3 3 03 3 2 0 5 10 10 0 O 0 0 1 ST ADD WINDOW DATA STATUS ET CRC X R X Command REAL SERIAL TYPE Source PC Destination Controller with address 3 in 485 mode 02 8 35 30 34 30 03 38 31 STX ADDR WINDOW RD ETX CRC Source Controller Destination PC 02 83 35 30 34 30 31 03 42 30 STX ADDR WINDOW RD DATA ETX CRC TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 243 301 17 Technical Information Window Meanings Tab 15 Accessories and Spare Parts N Read Data Description Admitted Values Write Type 000 R W L Start Stop in remote Front mode the Start 1 window is a read only Stop 0 001 R W L Low Speed Activation No 0 Yes 1 default 0 008 R W L Remote default or Serial configuration Remote 1 Serial 0 default 1 100 R W L Soft Start write only in Stop condition YES 1 NO 0 101 R W N R1 Set Point type 0 Frequency 1 Power 2 Time 3 Normal default 3 4 Pressure available only if the gauge is connected 102 R W R1 Set Point valve expressed in Hz Wors default 900 103 R W Set Point delay time between the pump 0 to 99999 default 0 start and the set point check seconds 104 R W L Set Point signal activation type t
161. n 3 und Pin 8 des Verbinders K1 oder wegen der ffnung des externen Interlock Signals aktiv Den Kurzschlu zwischen Pin 3 und Pin 8 des Verbinders P1 r cksetzen oder das externe Interlock Signal schlieRen PUMP OVERTEMP Die Temperatur des oberen Lagers bzw der Pumpe hat 60 C berschritten Warten bis die Temperatur unter den Schwellenwert gesunken ist Die Pumpe durch zweimalige Bet tigung der START Taste neustarten CONTROLLER OVERTEMP Die Temperatur des Controllers hat 70 C berschritten Oder Die Temperatur des Kontrollerk hlk rpers ist ber 60 C Warten bis die Temperatur unter den Schwellenwert gesunken ist Die Pumpe durch zweimalige Bet tigung der START Taste neustarten TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 37 301 2 38 301 Gebrauchsanleitung MELDUNG BESCHREIBUNG BEHEBUNG TOO HIGH Die Pumpen ist stromaufnahme Sicherstellen dal der LOAD gr Rer als die vorgesehene Pumpenrotor ungehindert drehen kann Die Pumpe durch zweimalige Bet tigung der START Taste neustarten SHORT Wahrend des Normalbetriebs nach Die Verbindung zwischen CIRCUIT der Startphase erfolgt ein Pumpe und Controller Kurzschlu der pr fen Die Pumpe durch Ausgangsverbindung zweimalige Bet tigung der START Taste neustarten SYSTEM Die Pumpe wurde durch ein von Das Netzkabel des OVERRIDE einem entfernten Kontakt Controllers ausstecken und
162. n voor dat de controller niet kan vallen of stoten te verduren krijgt Laat de verpakking niet ergens buiten achter Het verpakkingsmateriaal is volledig recyclebaar en voldoet aan de EEG milieurichtlijn 85 399 Figuur 1 Verpakking van de controllers 78 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bedrijfshandleiding 6 Installatie De controller is uitsluitend voor gebruik binnenshuis ontworpen en moet van stroom worden voorzien met een 3 geleiderkabel zie tabel met verkrijgbare onderdelen voorzien van een internationaal goedgekeurde stekker met het oog op de veiligheid Ter preventie van elektroshock en in overeenstemming met de CE voorschriften dient altijd van deze voedingskabel gebruik te worden gemaakt steek de stekker in een goed geaard stopcontact In de controller ontwikkelen zich hoge spanningen die zware beschadigingen of de dood kunnen veroorzaken Alvorens installatie of onderhouds werkzaamheden uit te voeren de controller van de contactstop afkoppelen De controller kan op een tafel of in een speciaal rack worden geinstalleerd In leder geval moet de koellucht vrij rondom het apparaat kunnen circuleren De controller mag niet geinstalleerd en of gebruikt worden in ruimten die blootgesteld zijn aan de weersomstandigheden regen vorst sneeuw stof agressieve gassen of in ruimten met ex plosiegevaar of zeer hoog brandgevaar Tijdens de werking moeten de volgende omgev
163. nder starter et program der tillader at programmere visse driftparametre 4 Trykknap til overf ring af START STOP RESET Trykknappen fungerer kun n r der v lges fra frontpanelet Tryk p knappen en gang for at starte pumpen tryk p knappen igen for at standse pumpen Hvis pumpen standser automatisk ved en fejl tryk p denne knap en gang for at tilbagestille styreenheden og yderligere en gang for at starte pumpen igen 5 LCD display med tal og bogstaver punktmatrice 2 rader med 16 tegn TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 93 301 7 Istruktionsbog Figur 3 Bagpanelet p Styreenheden TwisTorr 304 FS AG Rack 1 Indgangskontakt til logiske signaler tilslutningskontakten leveres med speciel forbindelsesbro 2 Udgangskontakt til logiske signaler og kontrol af pumpens frekvens samt programmerbar analog udgang Udtag 24 Vdc til pumpens bl ser vent valve og manometer Indgangsmodul til styreenhed omfattende sikringer str mudtag og ENC filter Drejekontakt til indstilling af Profibus adressen Profibus ekstra Seriel kontakt til fjernkontrol ol anion w Pumpekontakt 94 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruktionsbog 7 Instruktion Start af styreenheden Styreenheden startes ved at s tte stromkablet i vegudtaget Start
164. ndes folgende standard rubrikker VEN SIR Advarselsmeddelelserne informerer operat ren om at en speciel procedure eller en vis type arbejde skal udf res pr cist efter anvisningerne modsat fald er der risiko for sv re personskader illa Denne advarselsmeddelelse vises f r procedurer der skal f lges neje for ikke at risikere maskinskader Dette g r opm rksom p vigtig information i teksten Opbevaring F lgende krav til omgivelsesforholdene g lder ved transport og opbevaring af styreenheden temperatur fra 20 C til 70 C relativ luftfugtighed 0 95 ikke kondenserende TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 89 301 7 Istruktionsbog Forberedelser fer installation 90 301 Styreenheden leveres i en speciel beskyttende emballage Kontakt den lokale forhandler hvis emballagen viser tegn p skader der kan v re opst et under transporten S rg for at styreenheden ikke tabes eller uds ttes for st d ved udpakningen Smid ikke emballagen ud Materialet kan genbruges 100 og opfylder EU direktiv 85 399 om milj beskyttelse U LS JI BAL ule a Mid TA Figur 1 Styreenhedens emballage TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruktionsbog 7 Installation Af hensyn til sikkerheden er styreenhed kun bregnet til idend rs brug og skal drives af et el kabel med 3 ledere Se tabel over reservedele med et s
165. ndite locale Durante l operazione di disimballaggio prestare particolare attenzione a non lasciar cadere il controller e a non sottoporlo ad urti Non disperdere l imballo nell ambiente Il materiale completamente riciclabile e risponde alla direttiva CEE 85 399 per la tutela dell ambiente Figura 1 Imballo dei controller 18 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Istruzioni per l uso 1 Installazione Il controller progettato solo per uso interno e deve essere alimentato mediante un cavo di alimentazione a tre fili vedere tabella delle parti ordinabili con una spina di tipo approvato a livello internazionale ai fini della sicurezza dell utente Utilizzare sempre questo cavo di alimentazione ed inserire la spina in una presa con un adeguato collegamento di terra onde evitare scariche elettriche e per rispettare le specifiche CE All interno del controller si sviluppano alte tensioni che possono recare gravi danni o la morte Prima di eseguire qualsiasi operazione di installazione o manutenzione del controller scollegarlo dalla presa di alimentazione Il controller pu essere installato su di un tavolo o all interno di un apposito rack In ogni caso occorre che l aria di raffreddamento possa circolare liberamente intorno all apparato Non installare n utilizzare il controller in ambienti esposti ad agenti atmosferici pioggia gelo neve polveri gas aggressivi in ambienti
166. neel bedienen Onderhoud De controllers van de serie TwisTorr 304 FS AG Rack zijn onderhoudsvrij Eventuele werkzaamheden moeten door bevoegd personeel worden uitgevoerd In geval van storing is het mogelijk om de reparatiedienst van Agilent of de Agilent advanced exchange service in te schakelen zo krijgt men een ruilcontroller ter vervanging van de defecte controller MH OVINOS Alvorens werkzaamheden aan de controller uit te voeren de voedingskabel afkoppelen Mocht de controller gesloopt worden ga dan overeenkomstig de specifieke nationale wetgeving te werk TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 83 301 6 Bedrijfshandleiding Afvalverwerking 84 301 Betekenis van het logo WEEE op de etiketten Het onderstaande symbool wordt aangebracht in overeenstemming met de EG richtlijn WEEE Dit symbool alleen geldig voor de landen van de Europese Gemeenschap geeft aan dat het product waarop het is aangebracht NIET mag worden afgevoerd samen met mormaal huisvuil of industrieel afval maar gescheiden moet worden ingezameld De eindgebruiker wordt dus verzocht contact op te nemen met de leverancier van het apparaat zij het de fabrikant of een wederverkoper om het proces van gescheiden inzameling en verwerking in gang te zetten na de van toepassing zijnde termen en voorwaarden van het verkoopcontract te hebben gecontroleerd TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Foutme
167. nelheid Lezen van de druk Interface Profibus optie Automatische instelling ingangsspanning In de volgende paragrafen is alle informatie vermeld om de veiligheid van de operator tijdens het gebruik van de apparatuur te verzekeren Gedetailleerde informatie is te vinden in de bijlage Technical information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bedrijfshandleiding 6 Deze handleiding hanteert de volgende symbolen EVE INNEN Bij dit symbool staat tekst die de aandacht van de operator vestigt op een speciale procedure of methode die indien niet correct uitgevoerd ernstig lichamelijk letsel kan veroorzaken Mode tue Bij dit symbool staat tekst met procedures die indien niet opgevolgd schade aan apparatuur kunnen veroorzaken De opmerkingen bevatten belangrijke informatie die uit de tekst is gelicht Opslag Tijdens het transport en de opslag van de controllers moeten de volgende omgevingscondities aanwezig zijn temperatuur van 20 C tot 70 C relatieve vochtigheid 0 95 niet condenserend TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 71 301 6 Bedrijfshandleiding Voorbereiding voor installatie De controller wordt in een speciale beschermende verpakking geleverd als er schade wordt geconstateerd die tijdens het transport veroorzaakt zou kunnen zijn meteen contact opnemen met het plaatselijke verkoopkantoor Zorg er bij het uitpakke
168. nettori Di seguito sono illustrati il pannello di comando del Controller ed i pannelli di interconnessione Per maggiori dettagli fare riferimento alla sezione Technical Information Descrizione pannello frontale Figura 2 Pannello frontale del Controller TwisTorr 304 FS AG Rack 1 Pulsante per richiamare sul display i parametri cycle number cycle time e pump life 2 Pulsante per la selezione del modo LOW SPEED attivo solo quando selezionato il modo di comando dal pannello frontale Premendolo una volta la pompa ruota a velocit stand by Premendolo ancora si disattiva il modo LOW SPEED 3 Pulsante per richiamare sul display i parametri pump current pump temperature pump power e rotational speed sempre attivo indipendentemente dal modo di funzionamento scelto Premendo assieme i pulsanti 3 e 1 per almeno 2 secondi viene attivato un programma con il quale possibile programmare alcuni parametri operativi 4 Pulsante per inviare i comandi di START STOP RESET attivo solo quando selezionato il modo di comando dal pannello frontale Premendolo una volta si attiva la fase di avvio premendolo nuovamente si arresta la pompa Se la pompa si fermata automaticamente a causa di un guasto occorre premere questo pulsante una prima volta per eseguire il reset del controller ed una seconda volta per riavviare la pompa
169. nformation TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bruksanvisning 8 I bruksanvisningen anv nds f ljande standard rubriker Varningsmeddelandena informerar operat ren om att en speciell procedur eller en viss typ av arbete m ste utf ras exakt enligt anvisningarna annat fall finns risk f r sv ra personskador Nida Detta varningsmeddelande visas framf r procedurer som m ste f ljas exakt f r att undvika skador p maskinen Detta visar p viktig information i texten F rvaring F ljande krav p omgivningsf rh llanden g ller vid transport och f rvaring av styrenheten temperatur fran 20 C till 70 C relativ luftfuktighet 0 95 utan kondens TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 101 301 8 Bruksanvisning F rberedelser f r installation Styrenheten levereras i ett s rskilt skyddande emballage Kontakta det lokala f rs ljningskontoret om emballaget visar tecken p skador som kan ha uppst tt under transporten Se till att styrenheten inte tappas eller uts tts f r st tar vid uppackningen Kasta inte packmaterialet i soporna Materialet r tervinningsbart till 100 och uppfyller EU direktiv 85 399 om milj skydd Figur 1 Styrenhetens f rpackning 102 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bruksanvisning 8 Installation VARNING F r anv ndarens s kerhet ska s
170. nical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 163 301 13 N vod k Pou it Pou it V tomto odstavci jsou uvedeny hlavn operativn postupy Pro dal detajly i pro procedury t kaj c se spojen nebo dal ch op n ch detajl je t eba se obracet na odstavec Use v p loze Technical Information Ve ker elektrick a pneumatick spojen se mus vykonat p ed pou it m kontrol ru p i em je t eba se obracet na p slu nou p ru ku napojen ho erpadla NEBEZPE Aby se vyhnulo nebezpe kter m e ohrozit osoby i p stroj mus se zkontrolovat jeho stabilita v p pad e erpadlo je polo en na stole Jestli e vstupn p ruba nen napojen na syst m nebo nen uzav en se zav raj c p rubou erpadlo se nesm spustit do provozu Uzav raj c konektor JT jestli e nebude uskute n no dn vn j spojen mus z stat propojen se sv m m stkem erpadla pre pr zdna a TwisTorr 304 FS se mohou zapnout sou asn 164 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod k Pou it 13 Ovl d n indik tory a konektory V n sleduj c m je uk z n eln a zadn panel kontrol ru Pro ir detajly je t eba se obr tit na sekci Technical Information Popis eln ho panelu
171. ntrakten TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Fejlmeddelelser Istruktionsbog 7 N r visse fejl opst r viser styreenheden ved selvdiagnose aktuelle fejl p displayet De mulige meddelelser listes i f lgende tabel Tab 1 MEDDELELSE BESKRIVELSE KONTROL CHECK CONNECTION TO PUMP Tilslutning mellem pumpe og styreenhed er defekt eller Temperaturen p pumpen ligger under 0 C Kontroller at tilslutningskabel mellem pumpe og styreenhed er korrekt monteret samt at ingen afbrydninger forekommer Tryk to gange p START for at starte pumpen igen WAITING INTERLOCK Interlocksignalet findes p kontakt P1 p g a kortslutning mellem stift 3 og stift 8 i kontakten J1 eller p g a at det eksterne interlocksignal er bent Tilbagestil kortslutningen mellem stift 3 og stift 8 p kontakt P1 eller sluk for det eksterne interlocksignal PUMP OVERTEMP Temperaturen p det verste leje eller pumpen overskrider 60 C Vent p at temperaturen falder til under t rskelv rdi Tryk to gange p START for at starte pumpen igen CONTROLLER OVERTEMP Temperaturen p styreenhedens overskrider 70 C eller Temperaturen p styreenhedens k ler overskrider 60 C Vent p at temperaturen falder til under t rskelv rdi Tryk to gange p START for at starte pumpen igen TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 97 301 7
172. o 1E3 1 9 to 10 6 PCG 750 752 5E 5 to 1 5E3 0 61 to 10 23 7 FRG 720 730 5E 10 to 1E3 0 774 to 10 8 IM 500 Eyesys Mini IMG 1 33E 9 to 1 33E 1 1to9 9 CDG 500 T1000 2 66E 1 to 1 33E3 0 2 to 10 10 CDG 500 T0100 2 66E 0 to 1 33E2 0 2 to 10 11 CDG 500 T0010 2 66E 1 to 1 33E1 0 2 to 10 12 CDG 500 T0001 2 66E 2 to 1 33 0 2 to 10 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 When in programmable analog signal you read a voltage below the voltage range in above table it means Error Over range if you read a voltage below the voltage range it means Error Under range in these cases you can read the manual of the gauge for further details With the serial communication you can use the following windows Win 224 Pressure reading with the format X X E XX Win 257 Gauge status read only 0 gt when a gauge is not connected 1 gt when a gauge is connected 2 gt when the pressure read is under range or when the gauge give an internal error 3 gt when the pressure read is under range or when the gauge give an internal error 4 gt when the identification is present but not known by the controller Win 267 Gauge power read write If you need to turn off the gauge at certain condition you can use this window 0 gt gauge off 1 gt gauge on default 2 gt gauge normally off It turns on when the set point 1 is activated If the set point 1 type win 101 is
173. o interior de um rack especifico Em todo caso 6 necess rio gue o ar de refrigerac o possa circular livremente ao redor da aparelhagem N o instalar e ou utilizar o controller em ambientes expostos a agentes atmosfericos chuva gelo neve poeiras gases agressivos ou em ambientes com perigo de explos o ou com elevado risco de inc ndio Durante o funcionamento 6 necess rio gue sejam respeitadas as seguintes condic es ambientais temperatura de 5 C a 45 C humidade relativa 0 95 n o condensante Para as outras liga es e a instala o dos acess rios opcionais ver a sec o Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 67 301 5 Manual de Istru es Utilizac o Neste par grafo s o descritos os principais procedimentos operativos Para maiores detalhes e para procedimentos gue envolvem ligac es ou pecas opcionais consultar o par grafo Use do ap ndice Technical Information Antes de usar o controller efectuar todas as ligac es el ctricas e pneum ticas e consultar o manual da bomba ligada ATENCAO Para evitar danos s pessoas e a aparelhagem caso a bomba esteja apoiada numa mesa certificar se que esteja est vel Nunca activar a bomba se o flange de entrada n o estiver ligado ao sistema ou n o estiver fechado com o flange de fecho O conector de fecho J1 deve permanecer ligado sua ponte se n o 6 efectuada nenhuma ligac o
174. odiagnostic du contr leur pr sente certains mes sages d erreur indiqu s dans le tableau ci dessous Tab 1 MESSAGE DESCRIPTION INTERVENTION CHECK CONNECTION TO PUMP Dysfonctionnement de la connexion entre la pompe et le contr leur Ou bien La pompe a une temp rature inf rieure a a 0 C S assurer gue le c ble de connexion entre la pompe et le contr leur et le contr leur est bien fix aux deux extr mit s et gu il n est pas coup Presser deux fois l interrupteur START pour r activer la pompe WAITING INTERLOCK Le signal d interlock situ sur le connecteur P1 est actif cause de la coupure du court circuit entre le pin 3 et le pin 8 du connecteur J1 ou cause de l ouverture du signal d interlock ext rieur R tablir le court circuit entre le pin 3 et le pin 8 du connecteur P1 ou fermer le signal d interlock ext rieur PUMP OVERTEMP La temp rature du palier sup rieur ou de la a depasse 60 C Attendre gue la temp rature retourne au dessous du seuil Presser deux fois l interrupteur START pour remettre la pompe en marche CONTROLLER OVERTEMP La temp rature du contr leur a d pass 70 C Ou bien La temp rature du radiateur du contr leur est sup rieure 60 C Attendre que la temp rature retourne au dessous du seuil Presser deux fois l interrupteur START pour remettre la pompe en marche TwisTorr 304 FS AG Rack Controller User M
175. oksi ottamaan ker ys ja h vitt misprosessia varten yhteytt laitteen toimittajaan olipa se sitten laitteen valmistaja tai j lleenmyyj tarkastettuaan ensin kaupan sopimusehdot 132 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Ohjek sikirja 10 Vianetsint Joidenkin toimintah iri iden yhteydess valvojan itsem rittelypiiri analysoi virheen joka n kyy viestein jotka on kuvailtu seuraavassa taulukossa Tab 1 VIESTI VIKA KORJAUSTOIMENPITEET CHECK Toimintah iri pumpun ja valvojan Tarkistakaa ett pumpun ja CONNECTION liit nn ss valvojan v linen yhteyskaapeli TO PUMP Tai on hyvin kiinnitetty p ist n Pumpun l mp tila on alempi kuin eik sen varrella ole esteit 0 C Painakaa kaksi kertaa painonappia START jolloin pumppu k ynnistyy WAITING Lukitussignaali interlock Palauttakaa liittimen P1 INTERLOCK liittimess P on aktiivinen neulojen 3 ja 8 v linen johtuen liittimen J1 neulojen 3 ja 8 oikosulku tai sulkekaa v lill tapahtuneen oikosulun ulkopuolinen lukitussignaali keskeytyksest tai ulkoisen interlock lukitussignaalin interlock avautumisesta PUMP Ylemm n laakerin tai pumpun Odottakaa ett l mp tila OVERTEMP l mp tila on ylitt nyt 60 C putoaa kynnysarvon alapuolelle Painakaa painonappia START kaksi kertaa jolloin pumppu k ynnistyy CONTROLLER Valvojan muuntajan l mp tila on Odottakaa ett l mp tila OVERTEMP 70 C putoaa
176. old system letter Window protocol new system These two protocols can be used as well with 232 or 485 media Please use Window protocol for new development 230 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Letter Protocol Description The default protocol is Window Communication format 8 data bit no parity 1 stop bit The baud rate is programmable via front panel from 600 to 9600 baud The controller is factory set for 9600 baud operation Communication protocol Host Master Controller Slave The communication is performed in the following way Host Controller Request CRC Answer CRC Reguest is an ASCII character identifying the action that must be performed by the controller or the reguested information Allowed ASCII characters A START B STOP a C Low Speed ON D Low Speed OFF E Request for operational parameters F Pump times zeroing TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 231 301 17 Technical Information G H Parameters writing pr J K Parameters reading Reguest for operating status Reguest for numerical reading Reguest for counters reading CRC corresponds to the sum with inverted sign of all the preceding bytes e g the START command A in ASCII code 41 inverted it will be FF 1 41 BF
177. olera poraz drugi zeby uruchomic pompe 5 Display alfanumeryczny na cieklych krysztalach matryca punktowana 2 linie x 16 czcionek TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 153 301 12 Podrecznik Instrukcji 8 6 2 Naj D el Posta 63 Pulpit tylny kontrolerow TwisTorr 304 FS AG Rack 1 Lacznik wejsciowy sygnalow logicznych lacznik blizniaczy jest dostarczany z specialnym mostkiem zamykajacym 2 Lacznik wyjsciowy sygnalow logicznych i nadzor nad czestotlowoscia na wyjsciu podczas zaprogramowanej wielokrotnej pracy pompy 3 Lacznik zasilania na wyjsciu 24 Vdc dla turbiny chlodzacej pompy vent valve i kaliber 4 Modul wejsciowy zasilania kontrolera zawierajacy bezpieczniki topikowe gniazdko zasilajace i filtr ENC 5 Switch obrotowy do ustawiania kierunku Profibus 6 Profibus dodatkowy 7 Lacznik seryjny do kontroli na odleglosc 8 Lacznik pompy 154 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Podrecznik Instrukcji 12 Procedure uzytkowania Zaswiecenie kontrolera Do zaswiecenia kontrolera wystarczy wprowadzic kabel zasilajacy do gniazdka sieci Uruchomienie pompy Zeby uruchomic pompe nalezy odblokowac wejscie interlock i nacisnac znajdujacy sie na pulpicie frontalnym przycik START Zatrzymanie pompy Do zatrzymania pompy wystarczy nacisnac prz
178. ontrol ru 167 Spou t n erpadla 167 Zastaven erpadla 167 dr ba 167 Likvidace 168 Chybn zpr vy 169 Preklad origin ln ho n vodu kai Agilent Technologies 159 301 13 N vod k Pou it V eobecn informace Tento p stroj je ur en pro profesion ln pot ebu U ivatel si mus p ed pou it m tuto p ru ku velmi pozorn p e st ale i ka dou dal aktualizaci kterou podnik bude p ed vat V d sledku nedostate n ho i ste n nespr vn ho pou it nebo v p pad chyb nevy kolen ho person lu i neopr vn n ch z sah a neodpov daj c ch specifick m st tn m norm m Agilent nep ij m a nenese dnou zodpov dnost Kontrol ry s rie TwisTorr 304 FS AG Rack jsou frekven n mi konvertory jsou kontrolovan mikroprocesorem a skl daj se z pevn ch komponent s autodiagnostickou a samoochrannou schopnost Charakteristiky kontrol ru eln operativnost d lkov seri ln a Pilota o 24 Vdc ochlazovac lopatky erpadla Pilot vent valve ten rychlosti erpadla po p kazu zastaven ten rychlosti zastaven ten tlaku Rozhran Profibus op n Automatick nastaven vstupn ho nap t V n sleduj c ch odstavc ch jsou uvedeny ve ker nutn informace t kaj c se oper tora a bezpe nostn ch podm nek b hem provozu p stroje Detailn informace se nach zej v dodatku Technical Information
179. orr or not TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 223 301 17 Technical Information The maximum power provided by controller to drive a gauge is 2W if your gauge reguire more power you need an external power supply The controller recognizes automatically the gauge connected by reading the signal identification in the connector Following a table with the gauge supported Gauge model Name showed in the display FRG 700 702 FRGa IMG 500 IMG PVG 500 502 PVGa FRG 720 730 FRGb IM 500 IM CDG 500 T1000 CDG3 CDG 500 T0100 CDG2 CDG 500 T0010 CDG1 CDG 500 T0001 CDGO The controller give the error gauge error if the gauge connected is not supported There are some gauges for example the CDG family gauge that have the same identification In this case the controller recognizes the family of the gauge but the user should select the right gauge by front panel The pressure can be read by display by serial line win 224 and by Remote I O Programmable Analog Out but the controller doesn t replicate the signal from the gauge The Programmable analog signal is according with the following 224 301 table Pressure range Voltage range on ID Gauge Model mbar 3 ai a V 1 CT 100 1 33E 2 to 1 01E3 1 11 to 8 25 2 FRG 700 702 BE 9 to 1E3 1 82 to 8 6 3 IMG 500 2E 9 to 1E 2 1 8 to 8 5 4 PVG 500 502 5E 4 t
180. ou report to us any problem you may have had with the purchase or operation of our products On the back side you find a Corrective Action request form that you may fill out in the first part and return to us This form is intended to supplement normal lines of communications and to resolve problems that existing systems are not addressing in an adequate or timely manner Upon receipt of your Corrective Action Request we will determine the Root Cause of the problem and take the necessary actions to eliminate it You will be contacted by one of our employees who will review the problem with you and update you with the second part of the same form on our actions Your business is very important to us Please take the time and let us know how we can improve Sincerely Ez J LT Giampaolo LEVI Vice President and General Manager Agilent Vacuum Products Division Note Fax or mail the Customer Request for Action see backside page to Agilent Vacuum Products Division Torino Quality Assurance or to your nearest Agilent representative for onward transmission to the same address CUSTOMER REQUEST FOR CORRECTIVE PREVENTIVE IMPROVEMENT ACTION TO AGILENT VACUUM PRODUCTS DIVISION TORINO OUALITY ASSURANCE FAX N XXXX 011 9979350 ADDRESS AGILENT TECHNOLOGIES ITALIA S p A Vacuum Products Division Via F Ili Varian 54 10040 Lein TO Italy E MAIL vpd qualityassurance_pdl ext agilent com NAME COM
181. ower cable and press the START push button twice to start the pump OVERVOLTAGE Controller power supply circuitryis Press the START push button faulty or the Controller received a twice to start the pump spike Should the message still be present call the Agilent service POWER FAIL Failure in the controller s pump Contact Agilent for power supply section Maintenance TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual ete 000 Ge 17 0 Technical Information a e TwisTorr 304 FS AG Rack Controller Description 209 z Controller Specifications 211 Controller Outline 212 Controller Connection 213 J6 Fan 213 J7 Vent 214 P1 Input 215 Output 216 R2 Output Diagrams 220 Programmable Analog Output Diagrams 222 Pump Connector 223 J5 Gauge Connector 223 Stop Speed Reading and Active Stop 226 J2 Serial Connector 227 Connector Examples 228 Serial Communication Descriptions 230 Letter Protocol Description 231 Window Protocol 238 Description 238 Communication Protocol 238 Examples 241 Window Meanings 244 How to Use by Front Panel 249 Use 255 General 255 Ra 256 OEE s Agilent Technologies 207 301 17 Technical Information Programming 262 Configuration Menu 262 Mode Menu 264 Pump Setting Menu 265 Input Output Menu 268 Setpoint R1 and Setpoint R2 Submenus 271 Gauge Menu 273 Serial Menu 275
182. padku kontrolera lub nie narazic na uderzenia Nie wyrzucac opakowania poniewaz jest ono calkowicie rycyklowalne i odpowiada normom CEE 85 399 dotyczacym ochrony srodowiska Posta 61 Opakowanie kontrolerow 150 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Podrecznik Instrukcji 12 Instalacja ZAGROZENIE W celu zapewnienia bezpieczenstwa osobie obslugujacej kontroler przeznaczony jest wylacznie do uzytku wewnetrznego i musi byc podlaczony za pomoca kabla zasilajacego z 3 przewodami patrz tabela czesci zamiennych ktore mozna zamowic z wtyczka majaca standardy miedzynarodowe Aby zapobiec ryzuku porazenia pradem oraz aby zapewnic standardy UE nalezy uzywac zawsze ten kabel zasilajacy podlaczajac wtyczke do kontaktu majacego odpowiednie uziemienie We wnetrzu kontrolera powstaja wysokie napiecia ktore moga byc przyczyna powaznych obrazen lub wypadkow smiertelnych Przed przystapieniem do jakiejkolwiek operacji instalowania lub konserwacji nalezy odlaczyc go od zrodla zasilania Kontroler moze byc zainstalowany na stole lub wewnatrz odpowiedniego Rack W obydwu przypadkach nalezy pamietac aby wokol kontrolera zapewnic swobodny przeplyw powietrza dla prawidlowego chlodzenia Nie instalowac i lub uzytkowac kontrolerow w miejscach narazonych na dzialanie warunkow atmosferycznych jak deszcz mroz snieg pylow gazow agresywnych w srodowisku wybuchowym lub o duzym stopniu zagrozenia pozarowego
183. poistky nap jaciu z suvku a filter ENC Rota n switch pre nastavenie adresy Profibusu Profibus na po iadavku S riov spojka pre dialkov kontrolu ol o aj ek w Spojka erpadla 178 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 N vod na Obsluhu 14 Postup pri pou it Zapnutie Controller Pre zapnutie controller sta zapoji k bel na elektrick sie a vyp na siete da do polohy 1 Na tartovanie erpadla Pre na tartovanie erpadla je potrebn spr stupni vstup Iinterlock a stla i tla idlo START na elnom paneli Zastavenie erpadla Pre zastavenie erpadla treba stla i tla idlo STOP na elnom paneli Udr ba Controllers s rie TwisTorr 304 FS AG Rack nevyZaduj iadnu dr bu Ak ko vek z sah musi previes poveren person l V pr pade vady je mo n vyu i opravn slu bu Agilent alebo Agilent advanced exchange service ktor vymen vadn controller 4 74J2H 48 Pred ak mko vek z sahom na controller vytiahnite n ru z elektrickej siete V pr pade e controller mus s do zberu ria te sa platn mi pr slu n mi n rodn mi normami TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 179 301 14 N vod na Obsluhu Likvid cia V znam loga WEEE nach dzaj ceho sa na t tkoch Aplikovanie doluozzna en ho symbolu dodr uje smernicu EU s n zvom
184. r 304 FS AG Rack Controller User Manual 87 901 029 01 31 301 2 Gebrauchsanleitung Gebrauch In diesem Kapitel sind die wichtigsten Betriebsvorg nge aufgef hrt Fur weitere Hinweise bez glich Anschl sse und Montage des bestellbaren Zubeh rs siehe Kapitel Use im Anhang zu Technical Information Vor Benutzung des Controllers s mtliche elektrischen und pneumatischen Anschl sse ausf hren und die Betriebsanleitung der angeschlossenen Pumpe durchlesen Steht die Pumpe auf einem Tisch muB auf den stabilen Stand geachtet werden da sonst die Gefahr von Personen und Ger tesch den besteht Die Pumpe nie einschalten wenn der Eingangsflansch nicht am System angeschlossen bzw nicht mit dem SchlieBflansch abgedeckt ist Der WiederverschlieR Verbinder J1 mu mit seiner Br cke angeschlossen bleiben wenn kein externer Anschlu erfolgt Die Vorvakuumpumpe und die TwisTorr 304 FS Pumpe k nnen gleichzeitig eingeschaltet werden 32 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Gebrauchsanleitung 2 Steuerungen Anzeigen und Verbinder des Controllers Der folgende Abschnitt beschreibt die vordere und hintere Konsole des Controllers F r weitere Einzelheiten siehe Technical Information Fronttafel der Controller 5 1 3 2 4 Abbildung 2 Fronttafel der Controller TwisTorr 304 FS AG Rack 1 Taste fur die Anz
185. r og installasjon av ekstrautstyr vises det til avsnittet Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 115 301 9 Instruksjon Manual Bruk Dette avsnittet beskriver de viktigste driftsmomentene For en detaljert beskrivelse samt moment som omfatter tilkoplinger eller ekstrautstyr vises det til avsnittet Use i vedlegget Technical Information F r styreenheten tas i bruk b r samtlige elektriske og pneumatiske tilkoplinger gj res Les brukerveiledningen for pumpen som er tilkoplet Dersom pumpen er installert p et bord m du kontrollere at pumpen st r st tt Dette er viktig for forhindre skader p apparatet og p personer Dersom inngangsflensen hverken er tilkoplet systemet eller dersom den er blokkert av l seflensen m pumpen aldri startes opp Lukkekontakten J1 m v re tilkoplet aktuell brygge dersom det ikke skjer en annen ekstern tilkopling Forvakuumpumpen og TwisTorr 304 FS pumpen m fungere sammen 116 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instruksjon Manual 9 Indikatorer og Kontakter p Styreenheten Nedenunder er frontpanelet og bakpanelet p styreenheten illustrert For ytterligere informasjon vises det til avsnittet Accessories and Spare Parts i vedlegget Technical Information Beskrivelse av frontpanelet Figur 2 Frontpan
186. r will contain a string of 5 characters with the following parameters Tab 11 BYTES MEANING 1 Current 0 255 scaled 2 Voltage 0 255 scaled 3 Rotational speed KRPM 4 Pump temperature C 0 254 temperature reading 255 fail 5 CRC When the Reguest is K Counters readings the Answer will contain a string of 11 characters with the following parameters Tab 12 BYTES MEANING 1 4 Cycle time 5 8 Pump life 9 10 Cycle number 11 CRC On request a sample program in QBasic language is available by Agilent TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 237 301 17 Technical Information Window Protocol Description Communication Format 8 data bit a no parity 1 stop bit a baud rate 600 1200 2400 4800 9600 programmable Communication Protocol The communication protocol is a MASTER SLAVE type where Host MASTER a Controller SLAVE The communication is performed in the following way 1 the host MASTER send a MESSAGE CRC to the controller SLAVE 2 the controller answer with an ANSWER CRC to the host The MESSAGE is a string with the following format lt STX gt lt ADDR gt lt WIN gt lt COM gt lt DATA gt lt ETX gt lt CRC gt Where When a data is indicated between two quotes it means that the indicated data is the corresponding ASCII character 238 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01
187. ressing the MEASURES key three times with the GAUGE connected and functioning the following screen page is displayed 7 4 E 05m BJA tr X X X X P 5 0 W 1 0 1 0 H z O O VACUUM TECHNOLOGIES TwisTorr 304 FS AG O Figure 35 Where mBar shows the pressure value measured XXXX shows the kind of gauge connected 260 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 The following messages may be displayed instead of the pressure GAUGE ERROR indicates that the GAUGE is not identified PRESSURE UNDER indicates that a pressure below the minimum threshold set has been measured PRESSURE OVER indicates that a pressure exceeding the maximum threshold set has been measured Pressing the COUNTERS key the following is obtained CIY CIL JE X X X X T I MIE XIX X X X m VACUUMTECHNOLOGIES MEASURES next OD TwisTorr 304 FS AG Figure 36 Where CYCLE XXXX Number of start stop cycles TIME XXXXXm Time of last cycle minutes Pressing again Where PUMP LIFE XXXXXX total operating time hours of the pump Twis
188. rge Valve Command 145 0 Close 1 Open Valid only if Par Buff Offset 20 Bo 1 B2 Not used Bo1 Operating Mode 008 These bits define from which source the controller has to get commands readings are always available In order to operate the controller by Profibus these bits must be set to 3 0 Remote 1 0 1 Serial 2 Front panel 3 Profibus 1 2 Hz 1100 High Speed Value 120 1350 Set the high rotational speed value 3 2 Hz 1100 Low Speed Value 117 1350 Set the low rotational speed value TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 285 301 17 Technical Information Input Buffer Tab 20 Input buffer OFFSET SIZE WIN BYTE BYTE UNIT RES RANGE DESCRIPTION SERIAL 0 1 V 0 100 Voltage 201 Voltage provided to the motor 1 1 C 0 100 Pump Bearing Temperature 204 Read the pump upper bearing temperature If gt 60 C the controller goes into fail status and the pump is stopped 2 1 C 0 100 Controller Power Supply Temperature 216 Read the controller power section temperature if gt 75 C the controller goes in fail status controller over temperature 3 1 C 0 100 Controller CPU Section Temperature 211 Read the controller CPU section temperature if gt 80 C the controller goes in fail controller overtemperature 4 1 0 6 Bo 3 Status 205 0 Stop 1 Waiting for interlock interlock connections on remote 1 0 connector missing 2 Ramp th
189. roller will self diagnose the error and the following messages will be displayed If the pump is not connected the display will be as shown in the following figure CHECK 0 N TJO U M Check connection between controller and pump then press STOP push button twice to start the pump If the P1 input connector is not in position with the link or the external interlock connections are open when the START pushbutton is pressed the display will be as shown in the following figure W A T N G N T E L 0 C K Disconnect connector Pl and check the link or the external interlock then install the connector to start the pump If within each step of the soft start mode the rotational speed of the pump do not reach the planned value within 15 minutes the display will be as shown in the following figure 292 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 If in normal operation after the starting phase the current drawn by the pump is higher than programmed 1 8 A the pump and the inter connected devices are switched off and the display will be as shown in the following figure Verify that systems has no leaks than press STOP RESET pushbutton twice to start the pu
190. ronmental conditions must be respected temperature from 5 C to 45 C a relative humidity O 95 without condensation See the appendix Technical Information for detailed Information about the above mentioned and the other connections and about the options installation TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 199 301 16 Instructions for Use Use This paragraph describes the fundamental operating procedures Detailed information and operating procedures that envolve optional connection or option are supplied in the paragraph USE of the appendix Technical Information Make all vacuum manifold and electrical connections and refer to TwisTorr 304 FS pump instruction manual before operating the TwisTorr 304 FS AG Rack controller WARNING To avoid injury to personnel and damage to the eguipment if the pump is laying on a table make sure it is steady Never operate the TwisTorr 304 FS pump if the pump inlet is not connected to the system or blanked off The input signal J1 connector should be left in position including the shipping links if no external connections are made The forepump and TwisTorr 304 FS pump can be switched on at the same time 200 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instructions for Use 16 Controls Indicators and Connectors The following paragraph illustrates the Controller control panel and interconnection p
191. s not driven even if the threshold value has been reached This parameter is not available if SETPOINT TYPE is set to NORMAL R1 R2 OUTPUT LOGIC Selects the control logic of the SETPOINT output positive logic the relay is energized when the measurement exceeds the reference negative logic the relay is energized when the measurement is below the reference This parameter is not available if SETPOINT TYPE is set to TIME or NORMAL TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Gauge Menu This menu contains the parameters relating to functioning of the active gauge PRESSURE UNIT i I GAUGE gt I I I GAS CORRECTION Flow 5 6 Figure 42 PRESSURE UNIT selects the unit of measurement of the pressure used to show this on the display for display on the serial line and for setting of the setpoint and autopressure thresholds Possible options mbar a Pascal Torr GAUGE MODEL ID it shows the gauge model identified by the controller If there are more model with the same identification the user can select the right gauge connected TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 273 301 17 Technical Information GAS TYPE sets the gas type not configured Nitrogen Argon Idrogen Other bp O DN O CORRECTION if it is selecte
192. s in the input data are not fully coherent yet So if the first byte is equal to 0 the interface is working properly otherwise there is a problem Each nibble in the bytes following the interface status byte represents the status of the last attempt to write a parameter with following coding diag status byte 0 1 H o A N Write success Controller response was NACK Controller response was Unknown window Controller response was Bad Data Type Controller response was Over range Controller response was Bad operation Controller response was an unknown response Parameter not supported by the controller The ProfiBus external diagnosis services are used to report following unexpected situations Controller fail if the controller goes in fail mode an external diagnostic service is reguired and one byte in the diagnostic buffer signals the type of failure User parameter mismatch if the user sets one or more parameters out of their allowed range a ProfiBus external diagnostic service is reguired and some bits in the diagnosis frame signals which parameter is wrong i e over range or under range TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 289 301 17 Technical Information 290 301 Diagnostic Buffer Tab 21 Diagnostic buffer OFF SET MEANING 0 Bo station non exist set by master B1 station not ready slave not ready for data exchange B2 cfg fau
193. see orderable parts table CLOSED f i OPEN gt lt gt lt gt lt gt lt gt i i i i STOP RAMP i NORMAL Vent Valve Relay Vent Valve Open i Win 126 Time Win 147 i i e o START COMMAND STOP COMMAND Figure 8 Vent Valve diagram in Auto mode 214 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 P1 Input Technical Information 17 P1 IN MALECONNECTOR O N N W o Aalo u SOFT START COM SYSTEM OVERRIDE COM INTERLO CK COM LOW SPEED COM START Figure 9 P1 input connector All the logic input to the controller must be connected at J1 mating connector With the provided Pl mating connector shipped with pin 3 and pin 8 shorted make the connections with AWG 24 0 24 mm2 or smaller wire to the pins indicated in the figure to obtain the desired capability The following table describes the signals available on the connector Tab 3 Pins Name Description Electrical Note 1 com Start Stop Contact closed lt Start Optically isolated Active only in Contact open lt Stop Vlow max lt 10v Remote Mode Imax 5mA operation Imin on 2mA 2 com Low Speed Contact closed Low Speed Optically isolated Active only in Contact open High Speed Vlow max 10v Remote Mode Imax 5mA operation 3 com Interlock Contact closed Active Optically isolated Active in all Contact open Inac
194. selected as pressure the gauge is always on 3 gt gauge normally off It turns on when the set point 1 is activated If the set point 2 type win 171 is selected as pressure the gauge is always on 4 gt gauge normally off It turns on when the set point 1 is activated If the set point 3 type win 101 is selected as pressure the gauge is always on The windows from 136 to 140 manage an additional set point set point 3 dedicated to turn ON or OFF the gauge win 267 4 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 225 301 17 Technical Information Pump Driving Typical pump driving chart start and normal operation 1200 160 1000 800 Power W Frequency Hz o S S EO o K OEM Re Es Freq Hz Power W 50 100 150 200 250 300 350 400 Time s Figure 24 Diagram 4 Stop Speed Reading and Active Stop This functions can be activated by Front Panel or serial line Win 107 167 Active Stop brakes the pump using the motor The Stop Speed reading reads the pump speed after a stop command Figure 25 Diagram 5 226 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 J2 Serial Connector Technical Information 17 This connector provides the connection for RS 232 and RS 485 serial line J2 Pin out 5 160 PA 000 ooo O
195. skontakt for logiske signaler tilkoplings kontakten leveres med spesiell lukkebrygge 2 Port for logiske signaler ut og monitoring av pumpefrekvens og analogisk utgang programmerbar 3 Port for mating ut 24 Vdc p kjolevifte p pumpen vent valve og kaliber 4 Inngangsmodul til mating av styreenheten som best r av sikringer matekontakt og ENC filter 5 Dreiebryter for innstilling av Profibusadresse 6 Profibus valgfritt 7 Serialport for fjernkontroll 8 Pumpeport 118 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instruksjon Manual 9 Instruksjoner for bruk Starte styreenheten Styreenheten startes ved sette stromkabelen i veggkontakten Starte pumpen For starte pumpen m man preparere interlockinngangen og trykke p knappen START p frontpanelet Stoppe pumpen Pumpen stoppes ved trykke p knappen STOPP p frontpanelet Vedlikehold TwisTorr 304 FS AG Rack seriens styreenheter er vedlikeholdsfrie Alt arbeid p styreenheten m kun utf res av autorisert personell Dersom styreenheten stanser m du ta kontakt med Agilents reparasjonservice eller med Agilents avanserte bytteservice som kan tilby overhalte styreenheter til erstatning for den delagte styreenheten WONINGEN F r noe arbeid utf res p styreenheten m den frakoples str mnettet Dersom en styreenhet skal kasseres m dette skje i henhold til nasjonale bestemmelser TwisTorr 304 FS AG Rack
196. specific national standards The TwisTorr 304 FS AG Rack controllers are micro processor controlled solid state freguency converter with self diagnostic and self protection features Controller features Front Remote Serial Operation 24Vdc pump fan cooling drive Vent valve drive Pump speed reading after stop command stop speed reading a Pressure reading Input voltage auto setting Profibus interface optional The following paragraphs contain all the information necessary to guarantee the safety of the operator when using the eguipment Detailed information is supplied in the appendix Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instructions for Use 16 This manual uses the following standard protocol WARNING The warning messages are for attracting the attention of the operator to a particular procedure or practice which if not followed correctly could lead to serious injury CAUTION The caution messages are displayed before procedures which if not followed could cause damage to the eguipment The notes contain important information taken from the text Storage When transporting and storing the controllers the following environmental reguirements should be satisfied temperature from 20 C to 70 C a relative humidity O 95 without condensation TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 197 30
197. ssaggi di errore Istruzioni per uso In alcuni casi di guasto la circuiteria di autodiagnosi del controller presenta alcuni messaggi di errore elencati nella tabella seguente Tab 1 MESSAGGI0 DESCRIZIONE AZIONE CORRETTIVA 1 CHECK CONNECTION TO PUMP Malfunzionamento nel collegamento tra pompa e controller Oppure La pompa ha una temperatura inferiore a 0 C Verificare che il cavo di collegamento tra pompa e controller sia ben fissato da entrambe le estremit e non sia interrotto Premere due volte il pulsante START per riavviare la pompa WAITING INTERLOCK E attivo il segnale di interlock presente sul connettore P1 a causa dell interruzione del corto circuito tra il pin 3 ed il pin 8 del connettore J1 o a causa dell apertura del segnale di interlock esterno Ripristinare il corto circuito tra il pin 3 ed il pin8 del connettore P1 o chiudere il segnale di interlock esterno PUMP OVERTEMP La temperatura del cuscinetto superiore o della pompa ha superato i 60 C Attendere che la temperatura ritorni al di sotto della soglia Premere due volte il pulsante START per riavviare la pompa CONTROLLER OVERTEMP La temperatura dell ambiente del controller ha superato i 70 C Oppure La temperatura del radiatore del controller superiore a 60 C Attendere che la temperatura ritorni al di sotto della soglia Premere due volte il pulsante START per riavviare la pompa
198. t valve a Lectura velocidad bomba despu s de activaci n mando de parada lectura velocidad de parada Lectura de la presi n a Interfaz Profibus opcional Configuraci n autom tica tensi n de entrada En los apartados siguientes se facilita toda la informaci n necesaria para garantizar la seguridad del operador durante el uso del equipo Una informaci n m s detallada se facilita en el Suplemento Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Manual de istrucciones 4 Este manual utiliza los s mbolos convencionales siguientes LVI IHI Los mensajes de advertencia atraen la atenci n del operador sobre un procedimiento o una operaci n espec fica que al no realizarse correctamente podr a provocar graves lesiones personales ATENCI N Los mensajes de atenci n se visualizan antes de procedimientos que al no respetarse podr an provocar da os al equipo Las notas contienen informaci n importante extra da del texto Almacenamiento Durante el transporte y el almacenamiento de los controlers se deber cumplir con las condiciones ambientales siguientes temperatura de 20 C a 70 C humedad relativa 0 95 no condensadora TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 53 301 4 Manual de istrucciones Preparaci n para la instalaci n El controler se suministra en un embalaje de protecci n especial si se observan se
199. the entire slave status For this purpose one Input from slave to master and one Output from master to slave buffers are provided All the needed parameters are mapped inside each buffer The following tables summarise the ProfiBus parameter access related to the corresponding windows serial protocol for this protocol please refers to the serial interface paragraph The interface doesn t any scaling operation on the read written parameters It acts only as a gateway between the ProfiBus network and the pump s controller serial interface As a conseguence of this the scale and the meaning of some parameters depends directly to the specific model of Turbo Pump controller connected to the interface TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 281 301 17 Technical Information Parameter Buffer Tab 18 Parameter buffer OFFSET BYTE SIZE BYTE UNIT RES RANGE DESCRIPTION WIN SERIAL 0 1 0 99 Gauge gas load type The value is divided by 10 entering 12 the controller will apply 1 2 Enter here the correction factor for gas necessary only if gauge is connected 161 2 mA Hz sec 0 10000 Setpoint 1 value Enter here the threshold value for the R1 output valid only if Bo B2 of Parameter Buffer Offset 21 is egual to 0 1 2 or 3 102 0 30000 Setpoint 1 Mask Time Set here the time measured from the start pump event while the R1 output is masked not activated 10
200. tik af internationalt godkendt type For at undg risiko for elektrisk st d og opfylde CE kravene skal denne denne ledning altid bruges og s ttes ind i en stikkontakt med en passende jordforbindelse Sp nding frembragt i styreenheden kan n h je v rdier og for rsage stor skade og d dsfald Frakobl altid str mkablet inden der udf res installations eller vedligeholdelsesarbejde p styreenheden Styreenheden kan installeres p et bord eller et velegnet stativ I begge tilf lde skal der v re plads nok til at luft kan cirkulerer frit omkring apparatet Install r og anvend ikke styreenheden i milj er der uds ttes for p virkninger fra atmosf ren regn sne is damp aggressive gasser og ligeledes ikke i eksplosivt eller brandfarligt milj F lgende krav til omgivelsesforholdene g lder ved drift temperatur fra 5 C til 45 C relativ luftfugtighed 0 95 ikke kondenserende For vrige tilslutninger og installation af tilbeh r henvises til afsnittet Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 91 301 7 Istruktionsbog Anvendelse Dette afsnit beskriver de vigtigste driftsprocedurer For en detaljeret beskrivelse samt procedurer der involverer tilslutninger eller tilbeh r henvises til afsnittet Use i bilag Technical Information Inden styreenheden anvendes b r samtlige elektriske og pneumatiske tilslutninger udf res L s brugsanvisningen f
201. tive Vlow max 10v modes If active allows the pump running if Imax 5mA inactive stops the pump Imin on 2mA 4 com System Contact closed Active Optically isolated Active in all Override Contact open Inactive Vlow max 10v modes If active allows the pump running if Imax bmA active stops the pump and all the Imin on 2mA others connected devices Fan Vent TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 215 301 17 Technical Information Pins Name Description Electrical Note Valve b com Soft Start Contact closed Active Optically isolated Active only in Contact open Inactive Vlow max 10v Remote Mode If active before the start command Imax bmA operation and activates the spft start procedure Imin on 2mA in STOP To be used only at first pump run status up The following figure shows a typical contact logic input connection and the related simplified circuit of the controller I CONTROLLER Figure 10 Typical logic input connection J1 Output J1 OUT FEMALE CONNECTOR FAULT FAULT 24VDC R2 R2 24VDC START START 24VDC LOW SPEED LOW SPEED 24VDC R1 R1 24VDC PUMP CURRENT MONITOR 2V 1A PUMP FREOUENCY MONITOR SEE NOTE MONITOR COM PUMP FREQUENCY MONITOR 10V 56KRPM Figure 11 Logic output connector 216 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Control
202. tness for a particular purpose Agilent shall not be liable for errors or for incidental or conseguential damages in connection with the furnishing use or performance of this document or of any information contained herein Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted computer software as defined in FAR 52 227 19 June 1987 or any equivalent agency regulation or contract clause Use duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms and non DOD Departments and Agencies of the U S Government will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Government users will receive no greater than Limited Rights as defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015
203. trarna pump current pump temperature pump power och rotational speed p displayen Tryckknappen fungerar alltid oavsett det valda styrs ttet N r tryckknapparna 3 och 1 h lls samtidigt intryckta i minst 2 sekunder startas ett program som till ter att programmera vissa driftparametrar 4 Tryckknapp f r verf ring av START STOP RESET Tryckknappen fungerar endast n r styrs tt fr n frontpanelen har valts Tryck p knappen en g ng f r att starta pumpen tryck p knappen igen f r att stanna pumpen Om pumpen stannat automatiskt i samband med ett fel tryck p denna knapp en g ng f r att terst lla styrenheten och ytterligare en g ng f r att starta om pumpen 5 LCD display med siffror och bokst ver punktmatris 2 rader med 16 tecken TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 105 301 8 Bruksanvisning El OA LJ D EI Figur 3 Bakre panel p Styrenheten TwisTorr 304 FS AG Rack 1 Ing ngskontakt f r logiska signaler anslutnings kontakten levereras med speciell st ngningsbrygga 2 Kontakt f r utg ende logiksignaler och kontroll av pumpfrekvens plus programmerbar analog utg ng Uttag 24 Vdc f r kylfl kt vent valve och manometer Inmodul f r styrenhet omfattande s kringar str muttag och ENC filter Vred f r inst llning av Profilbus adressen Profibus
204. trukcji INFORMACJA OPIS AKCJA KORYGUJACA TOO HIGH Absorpcja pradu przez pompe jest Sprawdzic czy wirnik nie ma LOAD wyzsza od zaprogramowanej przeszkod w swobodnym obrocie Nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy SHORT Podczas normalnej pracy po fazie Sprawdzic polaczenie pompy z CIRCUIT uruchomienia na wyjsciu sa kontrolerem zwarcia Nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy SYSTEM Pompa sie zatrzymala na skutek Odlaczyc kabel zasilania OVERRIDE sygnalu zagrozenia pochodzacego kontrolera i sprawdzic od pierwotnego kontaktu przyczyne powstania sygnalu zagrozenia Podlaczyc ponownie kabel zasilajacy i nacisnac dwa razy przycisk START dla ponownego uruchomienia pompy OVERVOLTAGE Nastapilo uszkodzenie w sekcji Nacisnac dwa razy przycisk zasilania kontrolera lub kontroler START dla ponownego otrzymal sygnal nieprawidlowy uruchomienia pompy Jezeli sygnal sie powtorzy zwrocic sie do Agilent o przeprowadzenie konserwacji POWER FAIL Uszkodzenie w sekcji zasilania Zwrocic sie do Agilent w celu pompy kontrolera naprawy TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 13 N vod k Pou it V eobecn informace 160 Uskladn n 161 P prava k instalaci 162 Instalace 163 Pou iti 164 Ovladani indikatory a konektory 165 Popis elniho panelu 165 Pou van procedury 167 Zapalov n K
205. turbo controller internal link The interface provides input data measures and output data commands to the controller and in addiction provides standard and user related diagnostics in order to allow the user to manage the exceptions Installation This option is factory installed The ProfiBus connector pin out is reported in the following figure Please in order to obtain best performance use ProfiBus certified cables SIGNAL GROUND RxD TxD P CNTR P SIGNAL GROUND 5V 24V RxD TxD N O 00 IS JE fo w IND FEMALE CONNECTOR Figure 45 ProfiBus connector 278 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Technical Information 17 Configuration Before power on the controller set the ProfiBus address by the 2 rotary switches Up to 126 different addresses from 0 to 125 can be selected The address value sets by the 2 switches is expressed in hexadecimal notation so value from 00 to 7D are permitted this setting is read by the interface during the power on phase so any change of the switch position after the power on is ignored Software Operation and State Machine From the software point of view the interface can be seen as two devices one for ProfiBus and the other for the internal link linked together Each device has its own state machine to manage the device operation Profibus link state ma
206. tyrenheten som endast avsetts f r inomhusbruk f rs rjas med en elkabel med 3 ledare se tabellen ver reservdelar som kan best llas med en kontakt av internationell godk nd typ F r att undvika risk f r elst tar och uppfylla kraven CE anv nd alltid denna elkabel och s tt i kontakten i ett jordat eluttag Sp nningen inuti styrenheten kan n h ga v rden och f rorsaka allvarliga skador och d dsfall Fr nkoppla alltid stt mkabeln innan n got installations eller underh llsmoment utf rs p styrenheten Styrenheten kan installeras p ett bord eller inuti ett d rtill avsett rack samtliga fall m ste dock kylluften kunna cirkulera fritt kring apparaten Installera och anv nd inte styrenheten i milj er som uts tts f r p verkan fr n atmosf ren regn sn is damm aggressiva gaser och inte heller i explosiv eller brandfarlig milj F ljande krav p omgivningsf rh llanden g ller vid drift temperatur fran 5 C till 45 C relativ luftfuktighet 0 95 utan kondens Betr ffande vriga anslutningar och installation av tillbeh r h nvisas till avsnittet Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 103 301 8 Bruksanvisning Anv ndning Detta avsnitt beskriver de viktigaste driftmomenten F r en detaljerad beskrivning samt betr ffande moment som involverar anslutningar eller tillbeh r h nvisas till avsnittet Use i bilaga Technical In
207. ustriavfall men att d remot ett differentierat uppsamlingssystem m ste uppr ttas Vi rekommenderar d rf r att slutanv ndaren tar kontakt med leverant ren av anordningen oberoende om det handlar om moderf retaget eller terf rs ljaren f r att kunna starta uppsamlings och bortskaffningsprocessen detta efter l mplig kontroll av kontraktsenliga tidsgr nser och f rs ljningsvillkor TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Felmeddelanden Bruksanvisning 8 N r vissa fel uppst r visar styrenhetens sj lvdiagnoskrets aktuellt felmeddelande p displayen De m jliga meddelandena listas i f ljande tabell Tab 1 MEDDELANDE BESKRIVNING TG RD CHECK Anslutningen mellan pump och Kontrollera att CONNECTION styrenhet r defekt anslutningskabeln mellan TO PUMP eller ocks pump och styrenhet r Pumpens temperatur ligger under 0 ordentligt monterad samt att oc inget avbrott f rekommer Tryck tv g nger p knappen START f r att starta om pumpen WAITING Interlock signalen finns p kontakt tg rda kortslutningen mellan INTERLOCK P1 p grund av kortslutning mellan stift 3 och stift 8 p kontakt P1 stift 3 och stift 8 i kontakten J1 eller st ng den externa eller p grund av att den externa interlock signalen interlock signalen r ppen PUMP Temperaturen p det vre lagret V nta tills temperaturen OVERTEMP eller pumpen verskrider 60 C sjunker under tr skelv rdet
208. utput with relay R1 and R2 if it is connected a gauge it will present also the settings for the set point 3 I SET POINT R1 R2 SETTING pe gt R1 R2 TYPE amp gt FREQUENCY El pa A 4 z la K ca TIME NORMAL PRESSURE v R1 R2 SET VALUE more PT 1 1 A 9999 Y R1 R2 HYSTERESYSI 00 lt gt fa aa en 99 v R1 R2 MASK TIME lt gt 00 00 00 00 SEC Ki 00 SEE set i 1 zas x fsi 99 99 99 v R1 R2 OUTPUT gt l R LOGIC _ 7 gt Flow 4 6 Figure 41 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 271 301 17 272 301 Technical Information R1 R2 TYPE selects the type of reference measurement to drive the relay Possible types Frequency driving frequency of the pump Power power supplied to the pump Time time from start Normal achievement of normal condition Pressure pressure read by the gauge if present R1 R2 SET VALUE Sets the reference value at which the relay Rl R2 is activated This parameter is not available if TIME or NORMAL previous parameter is selected as SETPOINT TYPE R1 R2 HISTERESYS Sets in percentage the hysteresis of the relay R1 R2 MASK TIME sets the masked time in hh mm ss measured from start within which the SETPOINT R1 R2 i
209. v znovu obr tte sa na firmu Agilent pre dr bu POWER FAIL Vada v nap jacom syst me Pre dr bu sa obr te na erpadla controllera Agilent TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr 304 FS AG Rack Controller User Manual 15 Priro nik za Navodila Splo ne informacije 184 Skladi enje 185 Priprava za monta o 186 Monta a 187 Uporaba 188 Komande indikatorji in konektorji189 Opis prednje plo e 189 Postopki uporabe 191 Vklopitev Controllera 191 Zagon rpalke 191 Zaustavitev rpalke 191 Vzdr evanje 191 Odlaganje opadkov 192 Obvestilo o napaki 193 Prevod navodil v izvirniku tle Agilent Technologies 183 30 15 Priro amp nik za Navodila Splo ne informacije 184 301 Naprava je namenjena samo za strokovno rabo Uporabnik mora temeljito prebrati priro nik in pri tem upo tevati vsako dodatno informacijo ki jo nudi Agilent Agilent ne odgovarja v slu aju celotnega ali delnega nespo tovanja navodil pri nepravilni uporabi nestrokovnega osebja v slu aju nedovoljenih posegov ali pri zanemarjenju specifi nih nacionalnih meril Controllerji serije TwisTorr 304 FS AG Rack so frekven ni pretvorniki na katere nadzor ima mikrokrmilnik Sestavljajo jih deli v solidnem stanju ki imajo zna ilnosti avtodiagnoze in avtoza ite Zna ilnosti controllerja Osprednja oddaljena serijska operativnost Upravljanje na 24 Vdc ventila ohlajevalne rpalke Uprav
210. vlesing av pumpehastighet etter stoppekommando avlesing av stoppehastighet a Avlesing av trykk Grensesnitt Profibus valgfritt a Automatisk innstilling av inngangsspenning De f lgende avsnittene inneholder all informasjon som er n dvendig for sikre brukeren n r utstyret er i bruk For mer detaljert bruk vises det til tillegget Technical Information TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Instruksjon Manual Denne veiledningen bruker folgende standard protokoll DONA GARA Disse meldingene skal tiltrekke seg brukerens oppmerksomhet til en spesiell fremgangsm te eller praksis som hvis den ikke f lges kan medf re alvorlige skader NS ete Denne advarselen vises foran fremgangsm ter som dersom de ikke f lges kan fore til at utstyret skades Merknadene inneholder viktig informasjon som er hentet fra teksten Lagring N r styreenhetene transporteres eller lagres m f lgende forhold v re oppfylt temperatur fra 20 C til 70 C relativ fuktighet 0 95 uten kondens TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 113 301 9 Instruksjon Manual Forberede installasjonen Styreenheten leveres i en spesiell beskyttelsesemballasje Viser denne tegn p skader som kan ha oppst tt under transporten m du ta kontakt med det lokale salgskontoret N r styreenheten pakkes ut m du passe p at den ikke slippes ned eller utsettes for noen form for st
211. weede maal om de pomp weer op te starten 5 Alfanumeriek display met vloeibare kristallen LCD puntjespatroon 2 lijnen x 16 karakters TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 81 301 6 Bedrijfshandleiding 8 E 6 O Bo i 2 i m E EI Figuur 3 Achterpaneel van de controllers TwisTorr 304 FS AG Rack 1 Connector voor ingang logische signalen de te koppelen connector wordt met de speciale sluitklem geleverd 2 Connector voor uitgang logische signalen en freguentiecontrole pomp plus programmeerbare analogische uitgang 3 Uitgangsconnector voeding 24 Vdc voor de koelventilator van de pomp vent valve en kaliber 4 Ingangsmodule voeding controller die de zekeringen het voedingscontact en het ENC filter bevat 5 Draaischakelaar om het adres van de Profibus in te stellen 6 Profibus optie 7 Seri le connector voor afstandsbesturing 8 Connector pomp 82 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Bedrijfshandleiding 6 Gebruiksprocedures Inschakelen van de controller Om de controller in te schakelen de voedingskabel in de netcontactdoos inbrengen Starten van de pomp Om de pomp te starten de interlock ingang inschakelen en op de START knop op het frontpaneel drukken Stoppen van de pomp Voor het stoppen van de pomp de STOP knop op het frontpa
212. y 169 14 N vod na Obsluhu 171 V eobecn inform cie 172 Usklad ovanie 173 Priprava pre in tal ciu 174 In tal cia 175 Pou itie 176 Postup pri pou it 179 dr ba 179 Likvid cia 180 Oznamy v d 181 15 Priro nik za Navodila 183 Splo ne informacije 184 Skladi enje185 Priprava za monta o 186 Monta a 187 Uporaba 188 Postopki uporabe 191 Vzdr evanje 191 Odlaganje opadkov 192 Obvestilo o napaki 193 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 11 301 Contents 12 301 16 Instructions for Use 195 General Information 196 Storage 197 Preparation for Installation 198 Installation 199 Use 200 Use Procedure 203 Maintenance 203 Disposal 204 Error Messages 205 17 Technical Information 207 TwisTorr 304 FS AG Rack Controller Description 209 Controller Specifications 211 Controller Outline 212 Controller Connection 213 R1 R2 Output Diagrams 220 Programmable Analog Output Diagrams 222 J2 Serial Connector 227 Letter Protocol Description 231 Examples 241 How to Use by Front Panel 249 Use 255 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Contents Programming 262 How to Use by Remote 1 0 276 How to Use in Serial Mode 276 Profibus Option 278 Error Messages 292 Orderable Parts 295 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 13 301 Contents 14 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 TwisTorr
213. ycisk STOP znajdujacy sie na pulpicie frontalnym Konserwacja Kontroler z serii TwisTorr 304 FS AG Rack nie wymaga zadnej konserwacji Jakakolwiek interwencja moze byc dokonywana wylacznie przez osoby autoryzowane W przypadku uszkodzenia mozliwe jest korzystanie z serwisow naprawczych Agilent lub Agilent advance exchange service w ktorym mozna otrzymac kontroler zregenerowany zastepujac go tym uszkodzonym Przed wykonaniem jakiejkolwiek interwencji na kontrolerze nalezy odlaczyc kabel zasilajacy W przypadku decyzji zlomowania nalezy przeprowadzic operacje respektujac krajowe specyficzne normy TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 155 301 12 Podrecznik Instrukcji Przetworstwo odpadow Objasnienie znajdujacego sie na etykiecie znaczenia logo WEEE Uzyty ponizej symbol jest zgodny z wymogiem zarzadzenia WEEE Unii Europejskiej Symbol ten prawomocny tylko w krajach Unii Europejskiej oznacza ze wyrob ktory nim zostal oznaczony NIE moze byc przetworzony jako odpad razem z innymi domowymi lub przemyslowymi natomiast musi byc skladowany w miejscu przeznaczonym dla odpadow do przerobki zroznicowanej Dlatego tez poleca sie uzytkownikowi po uprzedniej weryfikacji terminu i warunkow zawartych w kontrakcie sprzedazy nawiazac kontakt z dostawca lub sprzedawca urzadzenia w celu uruchomienia procesu zbiorki i przerobu 156 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029
214. yez sre a szem lyekben s a berendez sben val k rok elker l se rdek ben biztos tani kell az asztal stabilit s t Soha nem szabad a szivatty t a bementi peremnek a rendszerhez val csatlakoztat sa vagy a z r peremmel val lez r sa n lk l zemeltetni A J1 lez r csatlakoz ssze kell maradjon k tve a megfelel tk t ssel jumper ha nincs semmif le k ls sszek ttet s Az el v kuum s a TwisTorr 304 FS szivatty egyidej leg is be lehetnek kapcsolva 140 301 TwisTorr 304 FS AG Rack Controller User Manual 87 901 029 01 Felhaszn l i K zik nyv 11 A vez rl parancsai kijelz i s csatlakozoi Az al bbiakban bemutat sra ker l a vez rl el s h tlapja Tov bbi r szletek tal lhat k a Technical Information c m szakaszban Az el lap leir sa 5 LS R1 AG R2 1 VACUUMTECHNOLOGIES 3 COUNTERS MEASURES increase ol next r 1 menu 2 decrease Nf previous 4 LOW SPEED START STOP TwisTorr 304 FS AG Abra 2 A TwisTorr 304 FS AG Rack sz m vez rl el lapja 1 Nyom gomb a param tereknek ciklusszam ciklusid szivatty lettartama a kijelz re val lehivasara 2 A LOW SPEED kis sebess g zemm d kiv laszt s ra szolg l nyom gomb Csak akkor aktiv amikor az el lapr l a parancsm d

Download Pdf Manuals

image

Related Search

Related Contents

On reconnaît ceux dont le cœur bat depuis toujours  Samsung SEW-DX105 User Manual  Perlick DP32S Beverage Dispenser User Manual  Innovator Web verstehen und installieren  User Manual: swinglet CAM  IPU 40318 IRI 4015 Data Sheet - Termo  FICHA DE DATOS DE SEGURIDAD  fully autonomous pipeline cleaning robot safuan naim bin mohamad  トルネードポテト取扱説明書並びに組み立て説明書 《 組立方法 》  Triarch 32325 User's Manual  