1. solvents such as acetone chloroform or methylene chloride Assistance After troubleshooting all components of your measurement system contact Technical Support Within the United States call 1 800 225 1480 and outside the United States call 978 232 6000 or fax 978 232 6031 In Europe the Middle East and Africa contact your local authorized dealer For the most current contact information visit www thermo com contactwater For the latest application and technical resources for Thermo Scientific Orion products visit www thermo com waterapps Warranty For the most current warranty information visit www thermo com water 12 RDO Optical Dissolved Oxygen Sensor User Guide WEEE Compliance This product is required to comply with the European Union s Waste Electrical amp Electronic Equipment WEEE Directive 2002 96 EC It is marked with the am Symbol on the left We have contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further information on compliance with these directives the recyclers in your country and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS directive are available at www thermo com WEEEROoHS Ordering Information Cat No Description 087010MD RDO optical dissolved oxygen sensor with 3 meter cable RDO cap calibration sleeve and st
2. the temperature sensor is covered by at least a inch of the solution If any salt or mineral deposits are observed on the RDO sensor immediately rinse them off with distilled or deionized water 1 Rinse the RDO sensor with distilled water blot it dry with a lint free tissue and insert it into the sample 2 Follow the instructions in the meter user guide for performing a measurement 3 Remove the RDO sensor from the sample rinse it with distilled water blot it dry with a lint free tissue insert the sensor into the next sample and repeat step 2 or refer to the RDO Sensor Storage section for the recommended storage procedures RDO Sensor Storage Short Term Storage Between measurements and overnight leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve making sure to wet the sponge with distilled water or in a beaker with distilled water Keep the RDO sensor away from direct sunlight during storage Long Term Storage Leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve making sure to wet the sponge with distilled water or in a beaker with distilled water Keep the RDO sensor away from direct sunlight during storage RDO Optical Dissolved Oxygen Sensor User Guide 7 RDO Sensor Maintenance Figure 7 RDO Sensor and RDO Cap Temperature Sensor O rings Spring Contacts Optical Window Cleaning the RDO Cap 1 Make sure that the R
3. 1 Remove the cap from the calibration sleeve and remove the sponge from the cap 2 Saturate the sponge with distilled water and squeeze the excess water out of the sponge 3 Reassemble the calibration sleeve 4 Unscrew and remove the thread ring from the RDO sensor See Figure 5 Slide the calibration sleeve over the front of the RDO sensor and screw the sleeve onto the sensor 5 Wait at least five minutes for the RDO sensor to equilibrate 6 Follow the instructions in the meter user guide for performing a water saturated air calibration 4 RDO Optical Dissolved Oxygen Sensor User Guide Zero Oxygen Solution for Zero Point Calibration 1 Perform a water saturated air calibration 2 Prepare a sodium sulfite solution by dissolving about 15 0 g of Na SO in about 250 mL of distilled water Transfer the solution to a BOD bottle or flask Note A small amount of cobalt salt can be added to the sodium sulfite solution The cobalt salt will act as an indicator and change color when the sodium sulfite solution no longer has a zero oxygen content 3 Adda stir bar to the bottle or flask Insert the RDO sensor into the bottle or flask and use parafilm to seal the open area between the bottle or flask and the sensor 4 Place the bottle or flask on a magnetic stir plate and gently stir the solution 5 Wait at least five minutes for the RDO sensor to equilibrate 6 Follow the instructions in the meter user guide for perfo
4. 14 RDO Optical Dissolved Oxygen Sensor User Guide Thermo Fisher Scientific Environmental Instruments Water Analysis Instruments North America 166 Cummings Center Beverly MA 01915 USA Toll Free 1 800 225 1480 Tel 1 978 232 6000 Dom Fax 1 978 232 6015 Int l Fax 978 232 6031 Europe P O Box 254 3860 AG Nijkerk Wallerstraat 125K 3862 BN Nijkerk The Netherlands Tel 31 033 2463887 Fax 31 033 2460832 Asia Pacific Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 257262 001 Rev A 09 08 www thermo com water Thermo SCIENTIFIC
5. 2 Figure 2 4 Slide and press the RDO cap onto the RDO sensor until the cap is flush with the sensor body See Figure 3 Note Do not twist the RDO cap Once the cap is installed the cap should not be removed from the RDO sensor until a new cap needs to be installed or troubleshooting needs to be performed Ji 5 Connect the RDO sensor to the 9 pin miniDIN input on the meter See Figure 4 Figure 3 Figure 4 Note An Err 880 message on the meter indicates that the RDO sensor is not fully connected to the meter If this error appears reconnect the sensor by firmly pushing the connector into the 9 pin miniDIN meter input Note An Err 882 message on the meter indicates that the RDO cap is missing or not properly installed on the sensor RDO Optical Dissolved Oxygen Sensor User Guide 3 RDO Sensor Calibration Calibration of an RDO optical dissolved oxygen sensor can be done quickly and conveniently by performing a water saturated air calibration Under equilibrium conditions the partial pressure of oxygen in airsaturated water is equal to the partial pressure of oxygen in watersaturated air This means that a sensor calibrated in watersaturated air will correctly read the partial pressure of oxygen in water samples When measuring low concentration samples less than 1 ppm a second calibration with a zero oxygen standard is recommended Calibration Sleeve Preparation for Water Saturated Air Calibration
6. DO cap is properly installed on the RDO sensor and rinse the cap and temperature sections of the sensor with distilled water from a squirt bottle or spray bottle and blot dry with a lint free soft cloth 2 Inspect the RDO cap for scratches or discoloration and replace the cap if any flaws are observed If algal growth is present on the RDO cap gently clean it with a soft brush Note Do not use solvents or soaps to clean the RDO cap and do not rub the cap with abrasive material Use of these materials will damage the RDO cap and void the warranty Do not let salt or mineral deposits form on the RDO cap To remove salt or mineral deposits from the cap refer to step 2 in the Cleaning the RDO Sensor Body section Cleaning the RDO Sensor Body 1 Make sure that the RDO cap is properly installed on the sensor and gently scrub the sensor body with a soft brush 2 Calcareous fouling can normally be dissolved with household vinegar Disconnect the RDO sensor from the meter and soak the assembled sensor with the RDO cap installed on the sensor in vinegar overnight If the marine growth remains use clean cotton swabs to gently wipe the growth off after it has been softened by soaking in vinegar 3 After cleaning the RDO sensor thoroughly rinse the sensor in distilled water and blot it dry with a lint free tissue before using or storing it After cleaning the RDO sensor a water saturated air calibration should be performed 8 RDO Optic
7. User Guide RDO Optical Dissolved Oxygen Sensor ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc AQUAfast Cahn ionplus KNIpHE No Cal ORION perpHect PerpHecT PerpHecTion pHISA pHuture Pure Water Sage Sensing the Future SensorLink ROSS ROSS Ultra Sure Flow Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher 1 888 pHAX ION A All in One Aplus AQUAsnap AssuredAccuracy AUTO BAR AUTO CAL AUTO DISPENSER Auto ID AUTO LOG AUTO READ AUTO STIR Auto Test BOD AutoEZ Cable Free CERTI CAL CISA DataCOLLECT DataPLUS digital LogR DirectCal DuraProbe Environmental Product Authority Extra Easy Extra Value FAST QC GAP GLPcal GLPcheck GLPdoc ISEasy KAP LabConnect LogR Low Maintenance Triode Minimum Stir Requirement MSR NISS One Touch One Touch Calibration One Touch Measurement Optimum Results Orion Star Pentrode pHuture MMS pHuture Pentrode pHuture Quatrode pHuture Triode Quatrode QuikchekK rf link ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Star Navigator 21 Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are trademarks of Thermo Fisher Guaranteed Success and The Technical Edge are service marks of Thermo Fisher PerpHecT meters are protected by U S patent 6 168 707 PerpHecT ROSS are protected by U S patent 6 168 707 ORION Series A meters and 900A printer ar
8. ainless steel protective guard 087020MD RDO optical dissolved oxygen sensor with 6 meter cable RDO cap calibration sleeve and stainless steel protective guard 087030MD RDO optical dissolved oxygen sensor with 10 meter cable RDO cap calibration sleeve and stainless steel protective guard 087050MD RDO optical dissolved oxygen sensor with 15 meter cable RDO cap calibration sleeve and stainless steel protective guard 087100MD RDO optical dissolved oxygen sensor with 30 meter cable RDO cap calibration sleeve and stainless steel protective guard 087001 Replacement RDO cap 087002 Stainless steel protective guard 087003 Calibration sleeve RDO Optical Dissolved Oxygen Sensor User Guide 13 Specifications Dissolved Oxygen Range Dissolved Oxygen Accuracy Dissolved Oxygen Resolution Dissolved Oxygen Response Time Temperature Range Temperature Accuracy Minimum Immersion Depth Sensor Length Tip Sensor Diameter Tip Total Sensor Length Total Sensor Diameter Cable Length RDO Cap Expected Lifespan 0 to 20 mg L 0 to 200 0 1 mg L up to 8 mg L 0 2 mg L from 8 to 20 mg L 0 01 mg L 90 of final value in 30 seconds 95 of final value in 90 seconds 5 to 45 C 0 3 C 40 mm 70 mm 20 mm 190 mm 35 mm 087010MD 3 meters 087020MD 6 meters 087030MD 10 meters 087050MD 15 meters 087100MD 30 meters 1 year Specifications are subject to change without notice
9. al Dissolved Oxygen Sensor User Guide Cleaning the Optical Window 1 Remove the RDO cap and gently wipe the optical window with a clean lens cloth Note Do not wet the optical window area with water or any other solution Yearly Replacement of the RDO Cap The RDO cap must be replaced every 365 days The meter will display an Err 881 message when the RDO cap needs to be replaced To view the days remaining before the RDO cap must be replaced refer to the meter user guide 1 Turn the power to the meter off 2 Remove the old RDO cap from the RDO sensor Pull the cap straight off of the sensor and do not twist the cap 3 Remove the existing two O rings from the RDO sensor 4 Remove any moisture from the RDO sensor body with a lint free tissue Do not touch the optical window or spring contacts with anything other than a lint free tissue Make sure that there is no moisture in the O ring grooves 5 Apply a thin film of silicone grease around the O ring grooves and install the two new O rings on the sensor Only use the silicone grease and O rings that are supplied with the RDO cap 6 Lubricate the two O rings on the RDO sensor with a thin film of silicone grease Do not allow any silicone grease to touch the optical window or spring contacts 7 Examine the optical window for scratches or dirt and clean it if necessary To clean the optical window gently wipe it with a clean lens cloth 8 Install a new RDO cap onto t
10. at the temperature sensor on the RDO sensor is immersed in the solution Contact Technical Support The RDO sensor must be in an oxygen free solution for at least five minutes Make sure that the open area between the sensor and bottle is covered with parafilm Soak the RDO sensor in distilled water for 30 minutes use a wash bottle to thoroughly rinse the sensor with distilled water and then soak the sensor for another 30 minutes in fresh distilled water The RDO sensor is not fully connected to the meter Reconnect the RDO sensor The RDO cap has expired Install a new RDO cap Verify that the RDO cap is fully seated on the sensor body Install a new RDO cap RDO Optical Dissolved Oxygen Sensor User Guide 11 Interfering Substances Alcohols greater than 5 hydrogen peroxide H O greater than 3 sodium hypochlorite NaCIO commercial bleach greater than 3 gaseous sulfur dioxide SO and gaseous chlorine Cl may interfere with the dissolved oxygen measurements If the measurements are unstable and other troubleshooting procedures do not correct the problem there may be interfering substances present in the solution Carbon dioxide CO ammonia NH3 pH any ionic species like sulfide S gt sulfate SO chloride Cl and hydrogen sulfide HS do not interfere with the dissolved oxygen measurements Substances That Damage the RDO Sensor Do not use the RDO sensor in solutions that contain organic
11. e protected by U S patents 5 198 093 D334 208 and D346 753 ionplus electrodes and Optimum Results solutions are protected by US Patent 5 830 338 ROSS Ultra electrodes are protected by US patents 6 793 787 Orion ORP Standard is protected by US Patent 6 350 367 Orion NoCal electrodes are protected by US Patent 7 276 142 2008 Thermo Fisher Scientific Inc All rights reserved RDO is a registered trademark of In Situ Inc Fort Collins CO USA All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries The specifications descriptions drawings ordering information and part numbers within this document are subject to change without notice This publication supersedes all previous publications on this subject Table of Contents Introduction 0 ask ees Required Equipment j j j j Important Notes Read Before Assembling or Operating the Equipment RDO Sensor Preparation j j RDO Sensor Calibration Calibration Sleeve Preparation for WaterSaturated Air Calibration Zero Oxygen Solution for Zero Point Calibration Stainless Steel Protective Guard RDO Sensor Measurements RDO Sensor Storage cece kk ees RDO Sensor Maintenance Cleaning the RDO Gap v 23 pros ee Ceca eee es Cleaning
12. he RDO sensor by aligning the arrow on the cap with the arrow on the sensor and pressing the cap onto the sensor until the cap is flush with the sensor body Do not twist the RDO cap Make sure that the O rings are not pinched or rolled between the cap and the sensor 9 Turn the power to the meter on 10 After replacing the RDO cap a watersaturated air calibration should be performed RDO Optical Dissolved Oxygen Sensor User Guide 9 Troubleshooting The most important principle in troubleshooting is to isolate the components of the system and check each in turn The components of the system include the meter RDO sensor sample and technique Meter The meter is the easiest component to eliminate as a possible cause of error Thermo Scientific Orion meters include an instrument checkout procedure in the meter user guide Consult the meter user guide for complete instructions and verify that the meter operates correctly and is stable in all steps RDO Sensor 1 Rinse the RDO sensor thoroughly with distilled water blot it dry with a lint free tissue and inspect the RDO cap for scratches or discoloration 2 Remove the RDO cap from the RDO sensor and make sure that there is no water inside the cap the optical window is clean and clear the O rings are intact and have a thin coating of silicone grease and the spring contacts are clean and undamaged 3 If readings continue to be erratic and unstable the RDO cap may need
13. rming a zero point calibration 7 Thoroughly rinse the RDO sensor under running water and blot it dry with a lint free tissue Note If the RDO sensor is sluggish or inaccurate after a zero point calibration not all of the sodium sulfite was removed from the sensor A very thorough soaking and rinsing of the RDO sensor in distilled water is required to remove all of the sodium sulfite solution and restore the sensor performance Soak the RDO sensor in distilled water for 30 minutes blot it dry with a lint free tissue use a wash bottle to rinse the sensor with distilled water soak the sensor for another 30 minutes in fresh distilled water and blot it dry with a lint free tissue RDO Optical Dissolved Oxygen Sensor User Guide 5 Stainless Steel Protective Guard The RDO optical dissolved oxygen sensor can be used with the stainless steel protective guard Cat No 087002 The guard protects the RDO sensor from damage and adds weight to the sensor for low depth measuring applications Installation of the Protective Guard 1 Unscrew and remove the thread ring from the RDO sensor See Figure 5 Figure 5 Thread Ring 2 Slide the protective guard over the front of the RDO sensor and screw the guard onto the sensor See Figure 6 Figure 6 RDO Sensor with Protective Guard 6 RDO Optical Dissolved Oxygen Sensor User Guide RDO Sensor Measurements For accurate measurements the RDO sensor must be immersed in solution until
14. the RDO Sensor Body Cleaning the Optical Window Yearly Replacement of the RDO Cap Troubleshooting Troubleshooting Chart WR KK Interfering Substances Substances That Damage the RDO Sensor Assistante orne kw EW xwaka ener Warranty e 3 xy n Ku n a Ea Ke PA wk ele ees WEEE Compliange sk kk kk kk Ordering Information Specifications 0 0 e cece eens RDO Optical Dissolved Oxygen Sensor User Guide RDO Optical Dissolved Oxygen Sensor User Guide Introduction This user guide contains information on the preparation operation and maintenance of the Thermo Scientific Orion RDO optical dissolved oxygen sensors These dissolved oxygen sensors are compatible with a wide range of applications including rugged field applications water and wastewater analysis beer and wine testing and BOD biochemical oxygen demand analysis All of the RDO sensors feature automatic temperature compensation and convenient press on RDO caps Orion Star RDO Optical Dissolved Oxygen Sensors The Orion Star RDO optical dissolved oxygen sensors have a 9 pin miniDIN connector that ensures a waterproof and secure connection when used with the Orion Star Plus RDO optical dissolved oxygen meters Figure 1 RDO Optical Dissolved Oxygen Sensor Thread Ring Tempera
15. to be replaced Refer to the RDO Sensor Maintenance section Sample If the RDO sensor works properly in standards but not in the sample look for possible interferences or substances in the sample that could alter the sensor response or physically damage the sensor If possible determine the composition of the sample and check for issues Technique Check if the method of analysis is compatible with your sample 10 RDO Optical Dissolved Oxygen Sensor User Guide Troubleshooting Chart Issue Unable to calibrate the RDO sensor Dissolved oxygen measurements are unstable Dissolved oxygen measurement is too low Wrong temperature displayed Bad zero point calibration Bad reading after zero point calibration Err 880 error message Err 881 error message Err 882 error message Recommended Action Verify the calibration setup and procedure Make sure that no water droplets are on the surface of the RDO cap Perform RDO sensor maintenance Refer to the Troubleshooting and RDO Sensor Maintenance sections Measurements may take longer if the solution temperature is unstable Perform RDO sensor maintenance Refer to the Troubleshooting and RDO Sensor Maintenance sections Interfering substance may be present Salt may be present in the sample Set the salinity factor in the meter Perform RDO sensor maintenance Refer to the Troubleshooting and RDO Sensor Maintenance sections Verify th
16. ture Sensor O rings Spring Contacts Optical Window RDO Optical Dissolved Oxygen Sensor User Guide 1 Required Equipment 1 Thermo Scientific Orion Star Plus RDO optical dissolved oxygen meter 2 Thermo Scientific Orion RDO optical dissolved oxygen sensor 3 RDO cap Cat No 087001 4 Calibration sleeve Cat No 087003 5 Stainless steel protective sensor guard optional Cat No 087002 6 Magnetic stirrer or stirrer probe Cat No 096019 The stirrer probe can be used with 3 Star 4 Star and 5 Star Plus benchtop RDO meters 7 Distilled or deionized water Important Notes Read Before Assembling or Operating the Equipment Note Do not touch the bottom of the RDO cap optical window or spring contacts on the RDO sensor Note The RDO sensor has an internal clock that counts down the 365 day lifespan of anew RDO cap The countdown begins when the RDO cap is installed on the RDO sensor the sensor is connected to the meter and the first measurement is taken The 365 day countdown cannot be reset or changed once the first measurement is taken 2 RDO Optical Dissolved Oxygen Sensor User Guide RDO Sensor Preparation 1 Remove the protective shipping cap from the end of the RDO sensor 2 Lubricate the two O rings on the RDO sensor with a thin film of the silicone grease that is supplied with the sensor 3 Align the arrow on the RDO cap with the arrow on the RDO sensor See Figure