Home
4000 CMS Intermittent Therapy
Contents
1. PUMP ALARMS amp RESOLUTIONS LCD DISPLAY INFUSION COMPLETE ALARM AIR IN LINE ALARM DOWN OCCLUSION ALARM UP OCCLUSION ALARM HIGH UP PRESSURE ALARM SET NOT INSTALLED ALARM UNATTENDED PUMP ALARM REPLACE SET 3 ALARM DOOR OPEN ALARM EMPTY BATTERY RESOLUTION Press PAUSE add another IV bag select REPEAT Rx and confirm fields or turn the pump off Remove air from administration set and IV bag Disconnect set from patient s access site prior to using prime function Check administration set from the pump to patient s access site for cause of occlusion Correct obstruc tion per protocol Consider changing the pressure setting from Low to High When occlusion is resolved the alarm will stop and pumping will resume Check administration set from IV bag to pump for cause of occlusion When ready to begin infusion press PAUSE select RESUME and press RUN Check for excessive pressure on IV bag When ready to begin infusion press PAUSE select RESUME and press RUN Install appropriate Curlin Medical administration set Use ONLY Curlin Medical administration sets See User s Manual or IV set directions for proper installation instructions To resolve press RUN and resume the therapy or continue with operating procedure Replace set prime set if necessary select RESUME and press RUN when ready to begin Check placement of administration set and close pump door pro
2. UME to continue therapy without interruption Highlight and press YES to REPEAT Rx when hanging a new bag Note Hang a new bag prior to the administration of a dose Interrupting a dose when hanging a new bag will lead to the dose dropping when REPEAT Rx is selected When the infusion is complete the following screen displays sg TOT VOL 2000 8 mil KVORATE malhe INFUSION COMPLETE H 9 To stop a therapy press PAUSE then turn pump off
3. e prescription menu Meas Vol ERE ral BAG ol GE mol BAG ol EAE rol AMT DOSE LO ral AMTIDOSE D 0 m49 AMTIDOSE 0 BQDOSERATE omi hr DOSERATE mathr I0SERATE DOSE Time FOG DOSE Tine OBA HH DOSE Time O HAMM DOSE Fregi O DD HH I MEMORATE L ral Ae I DOSEIEAG REGQVo mi a TOTTime GQ HH DONE YES 8 Make selection from the Run Options menu Press NO to review prescription OPTIONS to view options menu PRIME to prime set or press RUN key to start infusion PRESS to START to FEVIEM ae PLE PRIMING THE SET CAUTION Disconnect set from patient s access site before priming Sets with in line ASV must be primed on the pump By Gravity 1 Twist and remove breakaway tab from Flow Stop 2 Squeeze Flow Stop between thumb and forefinger to allow fluid to flow through tubing By Pump Note Set must be installed into the pump 1 Release any clamps on the administration set 2 Press and release PRIME key from Run Options menu 3 Press and hold the PRIME key until set is free of air release key to stop each prime cycle allows up to 6 mL of fluid 4 When prime is complete press YES ENTER to exit the prime screen Once pump is running While INTERMITTENT infusion is running the display will alternate between various information screens To interrupt or pause an infusion press RUN PAUSE key Select from the pause menu Press YES to RES
4. les at the top of the battery compartment Slide the battery compartment door back into place USING THE HELP FUNCTIONS When programming press the HELP key While running press PAUSE then press HELP key When a HELP screen is displayed and there is more text to read that is not displayed on the screen or icons will appear scroll up and down the text using the ARROW keys When finished press the HELP key again to exit the HELP screen MOO C Moog Medical Devices Group tf 800 970 2337 4314 Zevex Park Lane ph 801 264 1001 Salt Lake City Utah 84123 USA fx 801 264 1051 www moog com medical Moog is a registered trademark of Moog Inc and its subsidiaries All trademarks indicated herein are the property of Moog Inc and its subsidiaries 2011 Moog Inc All rights reserved 46254 Rev A MOO C MOOG MEDICAL DEVICES GROUP Curlin 4000 Series Ambulatory Infusion Pumps Clinician s Guide INTERMITTENT THERAPY REFER TO USER S MANUAL FOR DETAILED INSTRUCTIONS Pump Door Pump Door Instruction Label Display Door Latch f Screen Run Pause On Off Key Key Help Option Data Interface Key Bolus Cord Down Arrow LED Status Indicator Yes Enter Key Up Arrow Prime Bolus No Change Key Key Decimal Point Numeric Keys i Silence Key KEYPAD FUNCTION ON OFF Turns pump on and off When the pump is running the PAUSE key must be used to stop infusion first Then press the OFF key to turn pu
5. lue arrow positioning the tubing in front of the door hinge Center the tubing in the middle of the pumping fingers covering the yellow dot Holding the yellow Integral Flow Stop by the handle insert it on an angle into the receptacle on the left side of the pump in the direction of the yellow arrow CAUTION Do not press down on the top of the Integral Flow Stop when inserting it into the receptacle This action could inadvertently open the Integral Flow Stop and allow fluid to flow through the tubing Open the slide clamp on the administration set Push the tubing slightly into the black Air Detector slot Fully close the door of the pump and latch it securely Air Detector Important Notification A lack of compliance with proper positioning of the tubing could result in infusion inaccuracies CAUTION Please center the tubing in the pumping area when installing the administration set PROGRAMMING FOR INTERMITTENT THERAPY Turn the pump on Observe the self test the LED lights and the power source screen Select PROGRAM from the first selection menu Select NEW PROGRAM to program a new therapy manually If multiple therapies are available select INTERMITTENT from the therapy selection menu Complete the pre prescription menu I UNITS ie UNITS ma m DELAY OFF CONCEN D A mami W NExT YES DELAY OFF NEXT YES UNITS MCS CONCEN F DELAY OF NEXT YES Then complete th
6. mp off RUN PAUSE Starts or pauses the pump When the pump is paused pressing this key will resume pumping UP ARROW DOWN ARROW Used to scroll through the display screens NO CHANGE Rejects highlighted data YES ENTER Accepts highlighted data HELP OPTIONS Used to request and exit help screen when pump is paused or go to options screen when pump is running JSILENCE Used to silence alarm for 1 minute intervals and as a decimal point PRIME BOLUS Used to prime fluid through the tubing to clear it of air LED INDICATORS Green pump is running Yellow standby Red pump in alarm status 4 5 SETTING UP THE PUMP Install 2 new C size alkaline batteries Duracell ProCell or connect Battery Pack or AC adapter to port at bottom of pump Prepare IV bag and attach a new administration set Eliminate all air from the bag and the administration set see PRIMING THE SET Install set and close pump door Attach set to venous access device using agency protocol Note f a Check Valve Adapter is provided with your admin istration set remove it from the package and connect it to the distal end of the set 00 INSTALLING THE ADMINISTRATION SET Fully open the door of the pump Close the slide clamp on the administration set Activate the Integral Flow Stop by twisting and removing the breakaway tab Insert the blue tubing guide into the receptacle on right side of pump in the direction of the b
7. perly See User s Manual for instructions Press the pause key and turn pump off Install 2 new C size alkaline batteries If an error code occurs turn pump off and consult provider or manufacturer NOOR WD ND OPTIONS LOCK Press NO to change Lock Setting may need access code Press YES to accept new setting HOURLY TOTAL Press NO then YES to view Scroll to review each hour s totals Press NO to exit CLEAR HOURLY Press NO then YES to view Press YES to clear NO to retain and exit DOWN OCCLUSION Press NO to change YES to accept AIR SENSITIVITY Press NO to change YES to accept NEXT DOSE HH MM until next dose When correct press YES to accept AUDIO Press 1 9 to hear level press YES to accept POWER CHECK Press NO then YES to check level ACCEPT OPT Press YES to accept any changes made HOW TO CHANGE THE LOCK SETTING Lock may be initially selected in BIOMED SETUP When infusing press OPTIONS key Scroll to LOCK and press NO to change Enter access code at prompt Select NO until the desired lock setting is obtained Press YES ENTER to select Exit Options menu by pressing YES at ACCEPT OPT field INSTALLING BATTERIES Turn pump OFF or connect to AC adapter Open the battery door on the back of pump by depressing the release pin with a small coin and sliding itin a downward motion Insert 2 new C size alkaline Duracell ProCell batteries bottom side first with both positive po
Download Pdf Manuals
Related Search
Related Contents
TAS Steering Gear Service Manual Rendimiento mejorado con UASP Manual de instruções Muvit MUEAF0146 mobile phone case Abacus Installation and User Manual Cómo cargar papel - Downloaded from ManualsPrinter.com Manuals TeleBUS Protocols - Control Microsystems AVerKey iMicro User Manual Homecheck Professional Flood Report User Guide Copyright © All rights reserved.
Failed to retrieve file