1. Solution 1 Place the finger correctly inside and try again 2 Let the patient keep calm The device can not turn on 1 The batteries are drained or almost drained 2 The batteries are not inserted properly 3 The device is malfunction 1 Change batteries 2 Reinstall batteries 3 Please contact the local service center The indicator suddenly 1 The device will power off automatically when it gets no signal for 8 seconds 2 The batteries are almost drained 1 Normal 2 Change batteries User Manual for Fingertip Oximeter Appendix Key of Symbols Symbol Description With Type BF applied part A Warning See User Manual Spo The pulse oxygen saturation vhpm Pulse rate beats per minute Low battery voltage SN Serial number 7 Disposal of this device 5 according to WEEE regulations 14 User Manual for Fingertip Oximeter Common Knowledge 1 Meaning of SpO SpO is the saturation percentage of oxygen in the blood so called O2 concentration in the blood it is defined by the percentage of oxyhemoglobin HbO in the total hemoglobin of the arterial blood SpO is an important physiological parameter to reflect the respiration function it is calculated by the following method SpO HbO HbO2 100 HbO gt are the oxyhemoglobins oxygenized hemoglobin Hb are those hemoglobins which
2. Please take out the batteries if the oximeter will not be used for a long time D The recommended storage environment of the device ambient temperature 20 C 60 C relative humidity 10 95 atmospheric pressure 50kPa 107 4kPa E The oximeter is calibrated in the factory before sale there is no need to calibrate it during its life cycle However if it is necessary to verify its accuracy routinely the user can do the verification by means of SpO simulator or it can 11 User Manual for Fingertip Oximeter be done by the local third party test house High pressure sterilization cannot be used on the device Do not immerse the device in liquid A It is recommended that the device should be kept in a dry environment Humidity may reduce the life of the device or even damage it 7 2 Cleaning and Disinfecting Instruction A Surface clean sensor with a soft gauze by saturating with a solution such as 75 isopropyl alcohol if low level disinfection is required use a 1 10 bleach solution B Then surface clean with clean water and dry with a piece of cloth Caution Do not sterilize by irradiation steam or ethylene oxide Do not use the sensor if it is damaged 12 User Manual for Fingertip Oximeter 8 Troubleshooting Trouble The SpO2 and Pulse Rate display instable Possible Reason 1 The finger is not placed far enough inside 2 The finger is shaking or the patient is moving
3. to maintain it by themselves The oximeter cannot be used together User Manual for Fingertip Oximeter with devices not specified in User Manual Cautions e e Explosive hazard DO NOT use the oximeter in environment with inflammable gas such as some ignitable anesthetic agents DO NOT use the oximeter while the testee is under MRI or CT scanning Warnings e An uncomfortable or painful feeling may appear if using the oximeter continuously on the same place for a long time especially for poor microcirculation patients It is recommended that the oximeter should not be applied to the same location for longer than 2 hours If any abnormal condition is found please change the position of oximeter DO NOT clip this device on edema or tender tissue The light the infrared light is invisible emitted from the device is harmful to the e User Manual for Fingertip Oximeter eyes so service technician or testee should not stare at the light The local law must be followed when disposing of the device Attentions Keep the oximeter away from dust vibration corrosive substances explosive materials high temperature and moisture The device should be kept out of the reach of children If the oximeter gets wet please stop using it and do not resume operation until it is dry When it is carried from a cold environment to a warm and humid environment please do not use it immediately DO NOT o
4. Hanging Hole Figure 2 Rear View 1 2 User Manual for Fingertip Oximeter Features The Fingertip Oximeter can be used to measure pulse oxygen saturation SpOz value and pulse rate value PR Simply put your finger into sensor and press the start button SpO value and pulse rate value will be displayed on screen 1 3 1 4 It is lightweight small in size and easy to carry Large bright LCD Accurately measure SpO value and pulse rate value Automatically start measuring after putting finger into rubber cushions Power off automatically without signal for less than 5 seconds Visible alarm function Low voltage indication Name Name Fingertip Oximeter Applications and Scope Fingertip Oximeter is intended for measuring the 2 User Manual for Fingertip Oximeter pulse rate and functional oxygen saturation SpO gt through patients finger It is applicable for spot checking SpO and pulse rates of adult and pediatric patients in home and clinics 2 Battery Installation Figure 3 Battery Installation 1 Refer to Figure 3 insert two AAA size batteries into the battery compartment properly in the right direction 2 Replace the cover A Please take care when you insert the batteries or the improper insertion may make the device not work User Manual for Fingertip Oximeter 3 Operation 1 Open the clip as shown in Figure 4 c Figure 4 Put Finger into the Sensor 2
5. Put finger into the rubber cushions of the clip make sure the finger is in the right position and then clip the finger 3 The device will power on automatically in 2 seconds and start to measure 4 Measurement result will be displayed on the screen as shown in Figure 5 User can read SpO and HR values from display screen SpO2 SpO symbol Wy Pulse beat icon BPM the unit of pulse rate beats per minute n eet Bargraph for pulsatile indication User Manual for Fingertip Oximeter 99 85 56 98 Figure 5 Figure 6 5 Change display direction Two directions display alternately Press Display Key once at a time display screen will be flipped 180 each time in a cyclic manner See Figure 6 6 Alarm Indicator When measuring if SpO value or pulse rate value exceeds the preset alarm limit the value which exceeds limit on the screen will flash refer to Technical Specifications for alarm limits User Manual for Fingertip Oximeter 4 Instructions for Operation The finger should be put in properly and correctly Do not shake the finger and keep at ease during using Do not put wet finger directly into sensor Avoid placing the device on the same limb which is wrapped with a cuff for blood pressure measurement or during venous infusion Do not let anything block the emitting light from device Vigorous exercise and electrosurgical device interference may affect the measuring accur
6. User Manual for Fingertip Oximeter Instructions to User Dear Users Thank you very much for purchasing our product Please read the manual very carefully before using this device Failure to follow these instructions may cause measuring abnormality or equipment damage The Manual is published in English and we have the ultimate right to explain the Manual No part of this manual may be photocopied reproduced or translated into another language without the prior written consent We reserve the right to improve and amend it at any time without prior notice Version of This Manual Ver 1 1 Revised date Feb 17 2012 All rights reserved 3502 2290046 User Manual for Fingertip Oximeter Notes The contents contained in this manual are subject to change without notice Information furnished by our company is believed to be accurate and reliable However no responsibility is assumed by us for its use or any infringements of patents or other rights of third parties that may result from its use Instructions for Safe Operations Check the device to make sure that there is no visible damage that may affect users safety measurement performance with regard to sensors and Clips It is recommended that the device should be inspected minimally once a week When there is obvious damage stop using the device Necessary maintenance must be performed only by qualified service technicians Users are not permitted
7. acy Using enamel or other makeup on the nail may affect the accuracy of measurement If the first reading is unlikely true the more stable value is expected by waiting for a while or a restart is needed when necessary User Manual for Fingertip Oximeter 5 Technical Specifications A Display mode LCD Display B Power supply requirement 2 x LRO3 AAA alkaline batteries Supply voltage 3 0VDC Operating current lt 40mA C SpO Measurement Transducer dual wavelength LED Measurement wavelength Red light 663 nm Infrared light 890 nm Maximal optical output power less than 1 5mW maximum average Measuring range 35 99 Measuring accuracy Not greater than 3 for SpO range from 70 to 10096 NOTE Accuracy defined as root mean square value of deviation according to ISO 9919 D Pulse Rate Measurement Measuring range 30bpm 240bpm User Manual for Fingertip Oximeter Accuracy 2bpm or 2 whichever is greater E Preset alarm limits SpOz alarm Lower limit 90 Pulse Rate alarm Upper limit 120bpm Lower limit 50bpm F Update rate 8 beats moving average for Pulse Rate and SpOz readings G Environment requirement Operating Temperature 5 40 C Operating Humidity 30 80 Atmospheric pressure 70kPa 106kPa H The performance under low perfusion condition The accuracy of SpO2 and PR measurement still meet the precision described above when the modulation amplitude is as low as 0 6 I Resistance to i
8. nterference of surrounding light The difference between the value measured in the condition of indoor natural light and that of darkroom is less than 1 8 User Manual for Fingertip Oximeter J Resistance to 50Hz 60Hz interference SpOz and PR are precise which have been tested by SpO simulator from Fluke Biomedical Inc K Dimensions 60mm L 33mm W 30mm H Gross Weight 60g including batteries L Classification The type of protection against electric shock Internally powered equipment The degree of protection against electric shock Type BF applied parts The degree of protection against harmful ingress of liquids Ordinary equipment without protection against ingress of water Electro Magnetic Compatibility Group I Class B User Manual for Fingertip Oximeter 6 Accessories oom E A lanyard Two batteries A pouch A User Manual Quality Certificate Note The accessories are subject to change Detailed items and quantity see the Packing List 10 User Manual for Fingertip Oximeter 7 Repair and Maintenance 7 1 Maintenance The service life not a warranty of this device is 5 years In order to ensure its long service life please pay attention to the use of maintenance A Please change the batteries when the low voltage indicator lightens B Please clean the surface of the device before using Wipe the device with alcohol first and then let it dry in air or wipe it dry C
9. perate the button on the front panel with sharp materials DO NOT use high temperature or high pressure steam disinfection on the oximeter Refer to Chapter 7 for instructions of cleaning and disinfection IV User Manual for Fingertip Oximeter Declaration of Conformity The manufacturer hereby declares that this device complies with the following standards IEG 60601 1 ISO 9919 and follows the provisions of the council directive MDD93 42 EEC User Manual for Fingertip Oximeter Table of Contents TOVERVIEW PR 1 1 1 ASSER 1 1 2 Features vu uides retet eie 2 1 3 Namo sessess it 2 1 4 Applications and 2 2 BATTERY INSTALLATION X 3 OPERATION 4 4 INSTRUCTIONS FOR OPERATION 6 5 TECHNICAL SPECIFICATIONS 7 6 ACCESSORIES eerte 10 7 REPAIR AND MAINTENANCE 11 7 1 11 7 2 Cleaning and Disinfecting Instruction 12 8 TROUBLESHOOTING ee 13 APPENDIX rrt ttt iter 14 Key of Symbols 14 Common 15 User Manual for Fingertip Oximeter 1 Overview 1 1 Appearance Rubber Cushions Display Key T E Display Screen ES Figure 1 Front View Battery Cover Nameplate
10. r Fingertip Oximeter will allow finding the hypoxemia patient earlier thereby preventing or reducing accidental death caused by hypoxia effectively 3 Factors affecting SpO2 measuring accuracy interference reason Intravascular dyes such as indocyanine green or methylene blue lt gt Exposure to excessive illumination such as surgical lamps bilirubin lamps fluorescent lights infrared heating lamps or direct sunlight lt Vascular dyes or external used color up product such as nail enamel or color skin care Excessive patient movement lt Placement of a sensor on an extremity with a blood pressure cuff arterial catheter or intravascular line lt Exposure to the chamber with High pressure oxygen lt There is an arterial occlusion proximal to the 17 User Manual for Fingertip Oximeter sensor lt Blood vessel contraction caused by peripheral vessel hyperkinesias or body temperature decreasing 4 Factors causing low SpO Measuring value pathology reason Hypoxemia disease functional lack of HbO Pigmentation or abnormal oxyhemoglobin level Abnormal oxyhemoglobin variation lt Methemoglobin disease Sulfhemoglobinemia or arterial occlusion exists near sensor lt Obvious venous pulsations lt Peripheral arterial pulsation becomes weak lt Peripheral blood supply is not enough User Manual for Fingertip Oximeter 1141 Budapest Fogarasi t 77 1095 Budapest Mester
11. release oxygen 2 Principle of Measurement Based on Lamber Beer law the light absorbance of a given substance is directly proportional with its density or concentration When the light with certain wavelength emits on human tissue the measured intensity of light after absorption reflecting and attenuation in tissue can reflect the structure 15 User Manual for Fingertip Oximeter character of the tissue by which the light passes Due to that oxygenated hemoglobin 2 and deoxygenated hemoglobin Hb have different absorption character in the spectrum range from red to infrared light 600nm 1000nm wavelength by using these characteristics SpO2 be determined SpO measured by this oximeter is the functional oxygen saturation a percentage of the hemoglobin that can transport oxygen In contrast hemoximeters report fractional oxygen saturation a percentage of all measured hemoglobin including dysfunctional hemoglobin such as carboxyhemoglobin or metahemoglobin Clinical application of pulse oximeters SpO is an important physiological parameter to reflect the respiration and ventilation function so SpO monitoring used in treatment has become more popular For example such as monitoring patients with serious respiratory disease patients under anesthesia during operation and premature and neonatal infants The status of SpO can be determined in timely manner by measurement and 16 User Manual fo
12. u 34 Tel 220 7940 220 8881 220 7959 Tel 218 5542 215 9771 215 7550 220 7814 364 3428 Fax 220 7940 216 7017 216 7018 Fax 218 5542 Mobil 30 531 5454 30 939 9989 Mobil 30 940 1970 20 949 2688 E mail delton delton hu Web www delton hu www medipek hu