Endoscopes Inspection Report – Sale: 1521 Date: 13 / 10 / 2015
Contents
1. 113 PENTAX EC 3840JK Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition 1600x13 6mm Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Good physical condition 114 OLYMPUS CF 240L Videoscope 1600x12 8mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image quality Working Channel s Clear Leak Test Pass Overall Condition Very good overall condition 115 OLYMPUS CYF 240 Videoscope 350x5 8mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image
2. Functional Distal End Good condition Image Quality Good image Working Channel s Clear iinet Forceps Elevation Mechanism N A Leak Test Pass ep ae Overall Condition Good overall condition 127 OLYMPUS Insertion Tube Good condition CF 240AL Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Very good light output Videoscope Objective Lens and Focusing Poor optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Poor image quality dark film on image 1600x12 7mm Working Channel s Clear Leak Test Pass No carry case Overall Condition Poor image quality 128 PENTAX Insertion Tube Fair delaminated ED 3440T Bending mechanism amp Angulation Control Fair movement Video Light Guide Cable amp Light Transmission Good light output Duodenoscope Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Cap not included Image Quality Unable to check image quality no processor Working Channel s N A 1200x13 0mm Leak Test Fail No carey cane Oueralt Pondan Image quality not checked average physical condition with leak 129 PENTAX Insertion Tube Fair condition ED 3440 Bending mechanism amp Angulation Control Fair movement Video Light Guide Cable amp Light Transmission
3. Good light output Duodenoscope Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Cap not included Image Quality Unable to check image quality no processor Working Channel s N A 1200x13 0mm Leak Test Fail Nevcarty ease eyed Conan Image quality not checked average physical condition with leak 130 PENTAX Insertion Tube Completely delaminated ES 3840 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good light output delaminated fibre guide cable 900x13 7mm No carry case Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Ok Image Quality Unable to check image quality no processor Working Channel s N A Leak Test Pass Overall Condition Image quality not checked poor physical condition 131 PENTAX Insertion Tube Fair condition ED 3440 Bending mechanism amp Angulation Control Fair movement Video Light Guide Cable amp Light Transmission Good light output Duodenoscope Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Ok Image Quality Unable to check image quality no video processor Working Channel s N A 1200x13 0mm Leak Test Pass Notarioak Overall Condition Image quality not checked good physical condition 13
4. Fail Overall Condition Good image but leaking 141 OLYMPUS Insertion Tube Very good condition CF 240AL Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Very good light output Videoscope Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image quality eam Working Channel s Clear Leak Test Fail Overall Condition Good image quality but leaking 142 OLYMPUS Insertion Tube Good condition but ripped sleeve at the bending section BF 240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing and optics Remote Control Switches Fully functional 500x6 2mm Distal End Good Image Quality Good image Working Channel s Clear Leak Test Fail Overall Condition Good image but leaking 143 OLYMPUS Insertion Tube Good condition but ripped sleeve at the bending section BF 1T240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Poor light transmission Objective Lens and Focusing Fair focusing and optics Remote Control Switches Fully functional Suze dni Distal End Good Image Quality Blurry image
5. 4mm Arthroscope The rigid scope is in good physical condition and produces a clear image 208 Storz 70 2 7mm The rigid scope is in good physical condition and produces a Arthroscope clear image 209 Storz 0 4mm Autoclavable The rigid scope is in good physical condition and produces a Wide Angle Arthroscope clear image 210 Storz 0 4 x 175mm The rigid scope is in good physical condition and produces a Telescope clear image 211 Storz 0 3 5mm Wide Angle The rigid scope is in good physical condition and produces a Telescope clear image 212 Storz 30 2 7mm Telescope The rigid scope is in good physical condition and produces a clear image 213 Storz 0 10 5mm The rigid scope is in good physical condition and produces a Laparoscope clear image 214 Storz 70 4mm Arthroscope The rigid scope is in good physical condition and produces a clear image 224 Philips Agilent M3000A M3000A DE73107967 Refurbishment 05 2002 Module x6 M3015A Module M3015A DE02001141 07 2001 x5 and M3012A Module x1 M3000A DE94519509 12 2000 M3015A DE02000874 03 2001 M3000A DE94523646 04 2001 M3015A DE43526715 Refurbishment 10 2008 M3000A DE94519542 12 2000 M3012A DE34704340 04 2005 M3000A DE945 19506 03 2004 M3015A DE02000471 Refurbishment 08 2002 M3000A DE94521237 Refurbishment 09 2001 M3015A DE0200965 04 2001 Fair physical condition not tested 236 Karl Storz
6. Image Quality One with good image and second scope with average image Working Channel s Both clear Leak Test Not tested Overall Condition One of the scopes is in better overall condition with good image The second scope the bending section is completely flattened 152 OLYMPUS BF 240 Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing and optics Remote Control Switches Fully functional 500x6 2mm Distal End Good Image Quality Very good image Working Channel s Clear Leak Test Pass Overall Condition Very good overall condition 153 OLYMPUS Insertion Tube Good condition JF 240 Bending mechanism amp Angulation Control Fair movement and locks Duodeno Light Guide Cable amp Light Transmission Good light output Videoscope Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Good cap not included Image Quality Good image quality Working Channel s Clear 1100x12 0mm Leak Test Pass Overall Condition Good overall condition 154 OLYMPUS Insertion Tube Good condition JF 230 Bending mechanism amp Angulation Control Fair movement and locks Duodeno Light Guide Cable amp Light Transmission G
7. Oxylog 2000 metallic part of the unit with rust and the font panel is cosmetically poor in appearance O2 hose included unit not tested with medical gas Functions not fully tested 859 Dinamap XL Vital Signs Only one of the Pro 300 is working Dinamap XL amp 8100 all Monitor x 2 Dinamap Pro power up ok numerical displays and function keys are working 300 Monitor x 2 and Box of NIBP read was obtained when tested Spare parts are mostly Spare Parts consists of NIBP pumps solenoid valves printer assy and tubings 893 Mac 5500 ECG Recorder on The unit powers up ok system self test passed successfully Trolley with Leads The LCD screen keyboard and function soft keys are working fine The unit function menus are accessible The unit was tested with ECG simulator and good ECG traces was displayed and printout The patient leads and acquisition module are in good condition The system is in good physical condition 912 Oxford Sonicaid Foetal The CTG power up ok the LCD and function keys are working Monitors fine The US transducer is in good working order and gives a good graphic trace on display and on the printout as well The Toco transducer is responding fine when pressure is applied System is in fair physical condition US x2 and Toco x1 transducers included with spare printer paper 924 Hill Rom Isolette C2000 The incubator power up test ok and the temperature selector Incubator keys and LCD displays are working fine The air temperatur
8. Very good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Very good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Good image quality Working Channel s Clear Leak Test Pass Overall Condition Very good overall condition 122 OLYMPUS CF Q260DL Videoscope 1600x13 4mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Very good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image quality Working Channel s Clear Leak Test Fail Overall Condition Very good image quality but with leak 123 OLYMPUS CYF 240 Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement amp locks Videoscope Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Very good focusing amp optics Remote Control Switches Not working Distal End Good condition Image Quality Very good image Working Channel s Clear 350x5 7mm Forceps Elevation Mechanism N A Leak Test Fai
9. Working Channel s Clear Leak Test Fail Overall Condition Blur image and leaking 144 OLYMPUS Insertion Tube Good condition CYF 4 Bending mechanism amp Angulation Control Good movement Direct view Light Guide Cable amp Light Transmission Good light output Fibrescope Objective Lens and Focusing Good focusing with 19 fibres out Remote Control Switches N A Distal End Good 350x5 7mm Image Quality Good image Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 145 KARL STORZ Insertion Tube Fair with rip sleeve at the distal end 11101 RP1 Bending mechanism amp Angulation Control Good movement Direct View Scope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focus with 7 fibres out Remote Control Switches N A Distal End Good condition 250x3 8mm Image Quality Good image Working Channel s N A Forceps Elevation Mechanism N A Leak Test Not tested Overall Condition Good image and fair physical condition 146 KARL STORZ Insertion Tube Fair condition 11101 RP1 Bending mechanism amp Angulation Control Good movement Direct View Light Guide Cable amp Light Transmission Good light output Scope Objective Lens and Focusing Fair focus with over 30 fibres damaged Remote Control Switches N A Distal End Good conditi
10. camera head are in good working order and the system produces clear images The light source and the monitor are CLV S30 Light Source Olympus OTV S6 Camera Processor with Olympus Camera Head and Sony Colour Video Printer also functional The printer powers up and appear to be working Overall the units are in reasonable physical condition The strain relief of the camera head cable has been damaged 9 Haag Streit Slit Lamp Powers up ok and the lamp and brightness control are in BM900 on Table Goldman working order The optical system is clear and produces Tonometer Fixed satisfactory images Chin rest and table height adjustment are operational Overall the unit is in reasonable physical condition with signs of wear and tear The fixation light is not working 12 Drager Portable Ventilator The Oxylog 3000 power up ok the system self test completed Oxylog 3000 successfully The LCD screen function keys rotary selector and setting knobs are working fine Available ventilator modes are IPPV SIMV CPAP amp BIPAP Oxygen hose is included but unit was not tested with medical gas Therefore functions were not fully tested The lot is in good physical condition Serial No SRUL 0031 Year 2004 13 Stryker L9000 LED Light The unit powers up but the touch screen LCD does not respond Source and therefore the lamp cannot be switched on not working The unit is in good physical condition 16 Tomey UD 6000 Ultr
11. gas hoses Air O2 amp N20 vacuum and AGSS hose are in place SN AMRJ01845 Version 4 5 Ventilation modes Pressure and Volume Drager Isolette C2000 Infant Incubator Canopy Damaged The unit does not power up Motorised height adjustment is functional Part of the canopy is broken at the rear of the unit and the two hand port seals are missing from both end Otherwise the unit is in satisfactory physical condition Drager Evita 2 Dura Ventilator The unit powers up and completes the system self test successfully Display panel and controls are in good working order but the unit has not been fully tested Overall the unit is in good physical condition Air and O2 medical gas hoses in place but the expiratory block and flow sensor are not included SN ARRF 0450 Version 4 24 Working hours 47907 Ventilation modes IPPV SIMV BIPAP CPAP ASB MMV and BIPAP Assist Hitachi EUB 525 Ultrasound System The ultrasound powers up ok and the display function buttons and trackball are all in good working order All modes including B M Colour and PW are fully functional The probe is exhibiting satisfactory image quality The printer is functional Overall the ultrasound is in reasonable physical condition Attachment Ultrasound Probe convex 3 5MHz EUP C314G Olympus TC CI Clinical Trolley Including Sony Trinitron Monitor Olympus All units in the stack system power up ok Video processor and
12. good BP reading was 506DX Spo2 Monitor on obtained Similarly the 5 06DXN is working fine SpO2 is Stand with Finger Clip working on both units Both units are in fair physical condition 553 Valleylab Ligasure The unit powers up and completes the system self test Diathermy on Trolley with successfully Display panel and controls are fully functional Footswitch and the unit is in reasonable physical condition No further checks carried out Includes one footswitch YOM 2004 555 Seca CT61 ECG Machine The unit powers up and completes the system self test with Leads successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested because there were no papers Tested with patient simulator Overall the unit is in satisfactory physical condition 556 Plinth Company Hydraulic Couch s hydraulic lift back rest adjustment brake and castors Patient Couch x 2 are working fine Upholster are in good condition Lot is in good overall physical condition 561 Verathon Bladderscan on The bladder scanner and the probe are in good physical and Roll Stand cosmetic condition The battery is flat and the charger is not included therefore the unit could not be tested Model BVI 9400 563 Haag Streit 900 Slit Lamp The power supply for the slit lamp is coming on but the lamp on Table 90050955 and the lamp asse
13. panel and controls are in working order Both manual and automatic ECG recording modes are functional and the printer is working fine Tested with patient simulator Overall the unit is in satisfactory physical condition 720 Philips Pagewriter Trim III ECG Recorder with Leads The unit powers up and completes the system self test successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional and the printer is working fine Tested with patient simulator Overall the unit is in satisfactory physical condition 738 Zoll M Series Defib with Battery and Accessories The Biphasic defibrillator 200 Joules Max power up ok The LCD and the function soft keys are functional Both monitor and defib was tested with defib analyser and all simulated wave forms were displayed accurately The max energy discharge was tested and delivered accordingly for both manual and AED The NIBP was also tested ok Voice prompt and printer are working fine The unit comes with 10 wires ECG leads NIBP hose and cuff and two battery packs which needs charging S N T03I50352 DOM September 2003 There are no charger or external paddles with the lot 739 Zoll M Series Defib with Battery and Accessories The Biphasic defibrillator 200 Joules Max power up ok The LCD and the function soft keys are functional Both monitor and defib was tested with defib analyser and all simulated wave
14. quality Working Channel s Clear Leak Test Pass Overall Condition Very good overall condition 116 PENTAX Cystoscope E01150 400x5 8mm Insertion Tube Fair condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Fair light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Fair condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Fail Overall Condition Average physical condition and leaking 117 PENTAX ECY 1570 400x5 7mm Insertion Tube Fair condition with two dents on tube and bending section Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Fair light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Fair condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Good physical condition 118 OLYMPUS CF Q260DL Videoscope Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Good focusing amp optic system Remote Control Switc
15. s hydraulic lift back rest adjustment brake and castors are working fine Upholster are in good condition Lot is in good overall physical condition
16. system self test 8500 ECG on Stand with successfully Display panel and controls are in good working 259 ECG Leads order both manual and automatic ECG recording modes are 260 functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains 261 adaptor charger and ECG lead set are included Y O M 2012 Lot 259 has a small crack on the outer casing near the printer door 262 Drager Airshields The unit powers up and completes the system self test Resuscitaire with Fisher amp successfully Display panel and controls are in good working Paykel Humidifier order The heater and examination light are also functional The baby skin mode is not tested temperature sensor not included No further checks carried out Overall the unit is in satisfactory condition Air and O2 hoses are included The lot also includes a Fisher ad Paykel MR700 humidifier which also powers up ad appear to be functional 278 Mortara Instrument ELI The unit powers up and completes the system self test 250 ECG Machine with Leads successfully Display panel and controls are in working order some of the buttons are slow to respond Both manual and automatic ECG recording modes appear to be functional ECG signals can be viewed on the LCD screen but the printer is not fully working Tested with patient simulator Overall the unit is in satisfactory physical condition with minor damage to t
17. the system self test completed ok The display and functions keys are working Batteries are working fine Units were not tested for flow accuracy amp occlusion pressure Lot is in good physical condition 497 Bair Hugger Patient Warmer x 5 36642 28911 28905 28897 All five warming units power up and the temperature selection controls are in good working order Overall the units are functional and in satisfactory physical condition some the front panel buttons have been damaged Hoses are included with four units 503 Datex Ohmeda Anaesthesia Monitor Cardiocap 5 The unit power up ok system self test completed LCD and function keys are working fine The NIBP ECG and SpO2 parameters were tested and all values were within set limits The gas monitor was tested for CO2 detection only and was working fine The backup battery needs charging otherwise unit is in good overall condition No accessories included with lot S N FBWG03413 527 Urodyn 1000 Urine Analyser Powers up and appears to be working unable to carry out any further checks out of paper Transducer funnel and stand are included The unit is in reasonable physical condition The transducer cable has been damaged slightly 529 Seca CT 30001 ECG Only the ECG machine from this lot was tested The unit Recorder with Leads and powers up and completes the system self test successfully Sonicaid MD200 Ther
18. 11 Olympus CV 200 Endoscope All units power up ok Two of the unit s front panel is not Processors x 3 For working Third unit s all function buttons are responding Spares Repairs Video output not tested All are in reasonable physical condition 318 Respironics BiPap Focus Both units power up ok the LCD and function keys are Units x 2 working fine The turbines are working and the parameter and alarm setting are displays The batteries are charging when the system is connected to mains supply Both units are in good physical condition 322 Trumpf TITAN Operating The table has no remote control and it not powering up Unable Table to fully test all the functions The slit leg support is included and the couchions are in fair condition 323 Huntleigh Baby Dopplex Powers up ok Display panel and function keys are in good Foetal Monitor with Toco working order Both US and TOCCO channels are functional and US Leads and the printer is also working Includes US and TOCCO transducers and Event marker which are also in good working order Overall the unit is in good physical condition User manual is also included 325 EMS Lithoclast with The unit power up ok footswitch can be heard activating Unit footswitch has not been fully tested with medical air Unit is in good physical condition with no probes No indication of manufactured date 328 Bardscan Bladder Scanner Model BVI 2500 and PA 00145 and Bard Bladderscan Both bladder scanners
19. 2 OLYMPUS Insertion Tube Very good condition GIF XQ240 Bending mechanism amp Angulation Control Fair movement and locks Videoscope Light Guide Cable amp Light Transmission Good light output 1000x9 7mm Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Good image quality Working Channel s Clear No carry case Leak Test Pass Overall Condition Good overall condition 133 OLYMPUS Insertion Tube Good condition PCF 240S Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good Remote Control Switches Fully functional Distal End Good condition Image Quality Good image quality Working Channel s Clear eee Leak Test Pass Overall Condition Good overall condition 134 OLYMPUS Insertion Tube Good condition ENF P4 Bending mechanism amp Angulation Control Good movements Direct view Light Guide Cable amp Light Transmission Good light output fibrescope Objective Lens and Focusing Fair with more than 30 fibres out and sign of moisture on the lens Remote Control Switches N A Distal End Good condition Image Quality Fair image quality Working Channel s N A PS0 Omm Leak Test Pass Overall Condition Avera
20. 8500 ECG on Stand with successfully Display panel and controls are in good working 603 ECG Leads order both manual and automatic ECG recording modes are 604 functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains 605 adaptor charger and ECG lead set are included Y O M 2012 Lot 603 and 604 have minor scratch marks on the LCD protectors 613 Storz 27330520 Uromat and Both the Uromay amp Calcuson power up ok indicators and soft Storz 27610020 Calcuson key are working No probes with the units unable to fully with Leads and check functions The footswitch jars x3 and various tubings are Consumables available Both units are in good physical condition No date of manufacture 615 Draeger Scio Four Gas The lot includes 3x Drager Scio Four Oxi gas modules and a Agent Monitors x 3 and Drager Infinity dock station The mains adaptor is not included Draeger Docking Station therefore no further checks could be carried out The unit is in good physical condition 636 Patient Monitors All the monitors in the lot power up OK The modules TRAM RAC 4A X7 TRAM 451 SL Module X 5 provided with this lot are capable of monitoring ECG Temp CO 3 X IBP NIBP and SpO2 Following parameters were tested using patient simulator ECG amp Temp Other parameters including NIBP amp SpO2 were tested but not TRAM 451 N Module X 6 CDA15R Monitor X 8 Solar 8000
21. 932 Laserscope KTP 532 Laser with Footswitch No Key Unable to test this item and it has no ON OFF key or not accessories Item details Surgical Laser Laserscope KTP 532 Type No 0010 2774 Serial No 22063292 Date of Manufacture June 1994 Voltage 240V Single phase 30Amps 941 Seward Medical Light Source Keymed Light Source and Ellman Surgitron IEC Diathermy Both of the Keymed and Seward medical light source are working ok the bulbs and brightness controls are working fine The Electrosurgical unit power up ok the numerical displays and function keys for mode and RF output selector are in good working order The patient plate monitoring circuit and footswitch are working fine The RF output indicator and buzzer is active when the footswitch is depressed for both monopolar and bipolar modes However the RF output was not verified with ESU analyser Unit is in fair physical condition 969 Huntleigh Akron Electric Patient Couch The motorised height adjustment is working fine Backrest leg support adjustment are function and upholster is in good condition Castors are ok but rubber legs for the brakes with 2 missing 978 Marquette 2000 Treadmill with Marquette Hellige Cardio Sys Workstation The hard drive for the workstation has been removed and hence it does not power up No further tests were carried out The unit is in reasonable physical condition 979 Oxford So
22. 94534986 08 2002 M3000A DE94525924 08 2001 Fair physical condition not tested and one set faulty marked on it 792 GE Vivid Five Ultrasound Unit power up but does not completed the system initialisation with 2 5MHz Probe and and the CRT monitor goes in sleep mode since there are no Sony VCR Hard drive video signals Unit coned with ECG lead and one cardiac probe Removed Doesn t Power FPA 2 5MHz Cl System is in fair physical condition Date of Up manufacture 2001 800 Brandon 5 Spot Operating The Brandon light is ceiling mounted however the flange and Light mounting plate is short and it appear to have been modified to fit a low ceiling The unit is not complete the handles and transformer are not included The lamp assembly is in good physical condition but not tested 801 2000 KK4 Fluid Warmer Unit power up ok LCD and menu access keys are working The programme keys are not accessible the unit need password which is not available The unit is in good physical condition but not fully tested Cycle counter 00694 PAC version 2 05 Year 09 06 2003 804 Dinamap Pro Care 300 BP Only one of the units is powering up but NIBP is leaking Monitor on Stand with Physical condition fair but poor working order Leads x 6 805 Hill Rom Isolette C2000 The incubator power up test ok and the temperature selector Incubator keys and LCD displays are working fine The air temperature mode wa
23. Endoscopes Inspection Report Sale 1521 Date 13 10 2015 Condition Reports for other items at the end of the Scopes Report These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Condition 100 OLYMPUS e Insertion Tube Very good condition GIF Q240X e Bending mechanism amp Angulation Control Very good movement amp Videoscope on e Light Guide Cable amp Light Transmission Very good e Objective Lens and Focusing Very good focusing amp optics e Remote Control Switches Functional e Distal End Very good condition e Image Quality Very good image 1000x10 6mm e Working Channel s Clear e Forceps Elevation Mechanism N A e Leak Test Pass e Overall Condition Very good overall condition 101 OLYMPUS e Insertion Tube Very good condition CF Q240AL e Bending mechanism amp Angulation Control Very good movement amp locks Videoscope e Light Guide Cable amp Light Transmission Very good e Objective Lens and Focusing Very good focusing amp optics e Remote Control Switches Functional e Distal End Very good condition e Image Quality Very good image 1600x13 7mm Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass
24. G on Stand with ECG Leads The unit powers up and completes the system self test successfully Display panel and controls are in good working order both manual and automatic ECG recording modes are functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains adaptor charger and ECG lead set are included Y O M 2012 Lot 287 has minor scratch marks on the LCD protector 306 307 308 309 Cardiac Science Burdick 8500 ECG on Stand with ECG Leads The unit powers up and completes the system self test successfully Display panel and controls are in good working order both manual and automatic ECG recording modes are functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains adaptor charger and ECG lead set are included Y O M 2012 Lot 307 has scratch marks on the LCD protector Lot 308 has scratch marks and minor signs of wear and tear on the LCD protector 310 Olympus CV 200 Endoscope All three units were able to power up ok One of the unit s Processors x 3 For buttons are fully responding Second unit s some of front panel Spares Repairs buttons are not working and the membrane is peeled off Third unit s front panel is not functional Video output for the three units were not tested All are in reasonable physical condition 3
25. M Processors X 8 Printer PRN 50 M X 1 verified Monitors also have a slot for IBP but were not tested There are SpO2 leads finger probes and a few blood pressure cuffs with the units but were not tested No other accessories provided with the unit 3 X module racks from the lot are slightly damaged but are in good working order 3 X Monitors ON OFF switch is damaged but they are all in good working order The printer has not been tested Overall the units are in fair physical and cosmetic condition 649 Agilent Series 50 Fetal Powers up and completes the system self test successfully Monitor on Trolley Display panel and function keys are in good working order The US channels is functional and the printer is also working Includes US transducers and a scalp ECG cable Overall the unit is in reasonable physical condition 656 Cardiac Science Burdick The unit powers up and completes the system self test 8500 ECG on Stand with successfully Display panel and controls are in good working 657 ECG Leads order both manual and automatic ECG recording modes are functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains adaptor charger and ECG lead set are included Y O M 2012 Lot 657 has minor scratch marks on the LCD protector 658 Marquette Series 2000 The treadmill is in reasonable physical condition but the power Treadmill cable is not included t
26. Overall Condition Very good overall condition 102 OLYMPUS Insertion Tube Very good condition CYF 260A Bending mechanism amp Angulation Control Very good movement amp Videoscope PE Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Good focusing amp optics system Remote Control Switches Functional Distal End Very good condition Image Quality Very good image 350x5 7mm Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Very good overall condition 103 PENTAX Insertion Tube Poor with several large dents FCY 15P2 Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good light output Fibrescope Objective Lens and Focusing Fair focusing with several fibres out on 400x5 8mm one side of the lens Remote Control Switches N A Distal End Fair condition Image Quality Fair image quality Working Channel s Clear Leak Test Pass Overall Condition Fair overall condition with dents 104 PENTAX EC 3830LK 1600x13 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition Im
27. Ureteroscope The rigid scope and the fibre optic cable are in reasonable with Working Channel and physical condition but the image is not very clear and the Fibre Optic Cable insertion tube is also slightly bent The focusing control is working 237 Karl Storz Ureteroscope The rigid scope and the fibre optic cable are in good physical with Working Channel and condition and produces a clear image The focusing control is Fibre Optic Cable also working 252 Sonosite iLook 25 Portable The ultrasound powers up and the display panel controls are Ultrasound 35409 functional Transducer is included and the system produces satisfactory images Overall the unit is in reasonable physical condition with minor damage to the transducer housing Mains adaptor charger is included 255 Stack Trolley Including All units on the stack system power up ok The Storz video Sony Trinitron Monitor processor and camera head are in good working order and Storz Halogen 250 Twin produces clear images and the light source is also functional Light Source Storz Telecam The Litechnica and Leica camera controllers do not include SL PAL Camera Processor camera heads Overall the units are in reasonable physical with Storz Telecam condition 20212030 Camera Head Litechnica System 4000 Camera Processor Leica 2 Camera Controller and Sony Video Cassette Recorder 258 Cardiac Science Burdick The unit powers up and completes the
28. age Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Good physical condition 105 PENTAX EC 3830LK 1600x13 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Good physical condition 106 OLYMPUS GIF XQ230 Videoscope Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement amp locks Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Very good focusing amp optics 1000x10 4mm Remote Control Switches Functional Distal End Very good condition Image Quality Very good image Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Very good overall condition 107 OLYMPUS CYF 240A Videoscope 350x5 7mm Insertion Tube Very good condition Bending mechanism amp Angulation Control Very good movement amp locks Light Guide Cable amp Light Transmission Very good Objective Le
29. al condition but no tests have been carried out 1109 KMS Martin Marlux H6 The operating light is a ceiling mounted and it is one system Double Op Light and KMS with the main lamp H8 and two satellite lamps H6 The lamp Martin Marlux H8 Single has individual integrated control panel one the support arms Op Light one with membrane damaged The units come with flange tube and mount plate at one end The mounting plate cover and transformer power supply are not included System in fair physical condition 1169 Sidhill x 1 and Akron Couch s hydraulic lift back rest adjustment brake and castors Hydraulic Patient Couch x 1 are working fine Upholster are in good condition Lot is in good overall physical condition 1172 Fresenius Medical Care The unit powers up ok system self test passed successfully 4008S Dialysis Machine The LCD screen LED bar graph and function soft keys are working fine The unit function menus are accessible The full function of the system was not tested The following modules are attached BVM Bibag Air bubbles Detector Blood Pump and Online Plus system Unit is in fair physical condition with water and drain hose attached Operation counter 3772 Lhrs Software version 4 3 1175 Hydraulic Patient Couch x 2 Couch s hydraulic lift back rest adjustment brake and castors are working fine Upholster are in good condition Lot is in good overall physical condition 1176 Hydraulic Patient Couch x 2 Couch
30. apy Display panel and controls are in working order Both manual Unit and automatic ECG recording modes are functional and the printer is working fine Tested with patient simulator Overall the unit is in satisfactory physical condition 537 Cardiac Science Burdick The unit powers up and completes the system self test 8500 ECG on Stand with successfully Display panel and controls are in good working 220 ECG Leads order both manual and automatic ECG recording modes are 539 functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains 540 adaptor charger and ECG lead set are included Y O M 2012 2013 Lot 537 the printer roller gears have been damaged slightly but the printer is still functional Lot 539 has minor scratch marks on the LCD protector 541 Medela Dominant 50 The unit is in good working order with good flow of suction Suction Unit on Roll Stand The pressure gauge and suction regulator are also operational The original suction jar is not present but the unit includes another suction container 542 Trumpf TITAN Operating Theoperating table with remote control The table failed to Table power unable to test functions Some of the coushions pad in place but the leg suppots are missing Average physical condition 543 Dinamap Pro 400 BP The Pro 400 power up and the displays and function keys are Monitor on Stand Criticare working fine NIBP was tested and
31. are in satisfactory physical condition but the mains adaptors chargers are not included to test the units Only one of the units PA 00145 includes a transducer 329 Drager Oxylog 2000 The unit powers up and completes the system self test Portable Ventilator Drager Resuscitaire Heater Element successfully Display panel and alarm system are in good working order but no further checks has been carried out Charger and O2 hose are included Overall the unit is in good physical condition Software version 3 10 331 GE Dash 3000 Patient Monitor The patient monitor powers up successfully Display screen function keys and rotary control knob are in good working order All parameters appear to be available but have not been tested The unit does not include any accessories but overall it is in good physical condition 357 Cardiocap II and GE Dash 3000 Patient Monitors The Cardiocap II monitor powers up and appears to be functional not fully tested but the Dash 3000 monitor does not power up The rotary control knob and the LED indicator with the cover are missing from the handle of the Dash 3000 The units are in satisfactory physical condition the lot includes ECG lead and SpO2 probe for the Cardiocap II monitor only 358 Kamplex AD27 Audiometer with Leads Powers up ok Display panel and function keys are in good working order and the unit is in reasonable physical condition Includes a few accessor
32. asonic The unit power up ok display LCD touch screen and functions A B Scanner and Bio Meter key rotary knobs are working Biometer and A B scanner probe 203546 with Footswitch included The A B scan probe is oscillating and an air scan image is displayed However there are cracks on the transducer s head and oil is leaking The biometer function screen comes when selected but not fully tested The system is in fair overall condition Footswitch included Software version M20 G 0b year of manufacture 2005 23 Bair Hugger Patient All three warming units power up and the temperature Warmers x 3 selection controls are in good working order Overall the units are functional and in satisfactory physical condition some the front panel buttons have been damaged Hoses are included 25 Sovereign Phaco Machine Unit power up and system initialization completed ok The with Hand Pieces x10 and Footswitch x 1 LCD soft keys and controller are in good working order The pole height adjustment is working Mode available Phaco I A VIT amp Diathermy The system self test was carried out and all tests were fine The footswitch test passed System is in good physical condition and comes with the following hand pieces Allergan Phaco hand piece x10 30 Topcon CL 100 The unit powers up but displays initial error message and Computerised Lensmeter does not proceed any further LCD screen and functi
33. condition Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good light transmission Objective Lens and Focusing Good Remote Control Switches Functional Distal End Good condition Image Quality Good quality images Working Channel s Clear Leak Test Pass Overall Condition Good overall condition 150 OLYMPUS Insertion Tube Discoloured and delaminated BF 1T30 Direct View Scope 500x6 2mm Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Fair with 3 fibres out Remote Control Switches N A Distal End Fair condition Image Quality Average image quality Working Channel s Clear Leak Test Pass Overall Condition Fair overall condition 151 Richard Wolf 7305 006 Direct View Flexible Scope x2 350x5 3mm Insertion Tube Dents on both scopes at the bending section One with delaminated insertion tube Bending mechanism amp Angulation Control One working ok the other with loose lever Light Guide Cable amp Light Transmission Good light output fibre light guide cable not included Objective Lens and Focusing One with good optic and the second scope with several fibres damaged Remote Control Switches N A Distal End Fair condition
34. condition Version 4 5 PSV Pro Vent Mode Volume amp Pressure control SIMV amp PSV Pro mode 430 Draeger Evita 4 Ventilator Vent power up and system self test completed successfully The LCD and function keys are working fine System was not tested with medical gas thus not fully checked Rotary selector knob missing Expiratory valve and flow sensor included Unit is in fair physical condition with gas hose attached Operation hour 109222hrs software version 04 10 432 433 Cardiac Science Burdick 8500 ECG on Stand with ECG Leads The unit powers up and completes the system self test successfully Display panel and controls are in good working order both manual and automatic ECG recording modes are functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains adaptor charger and ECG lead set are included Y O M 2012 2013 Lot 433 the printer roller gears have been damaged slightly but the printer is still functional 434 Philips Pagewriter Trim II ECG Recorder with Leads The unit powers up and completes the system self test successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 438 Datascope Spe
35. ctrum Patient Monitors x 3 Only two of the Spectrum monitors are working properly Unit power up ok and self test completed The LCD and function keys are working fine The parameters available are ECG NIBP SpO2 Temp CO2 and IBP The system was not fully tested with patient leads and accessories since they are not available with lot The third unit has its power indication on but the LCD screen remained blank Monitors are in fair physical condition Backup battery packs are not working needs charging 439 Olympus Light Source Unit powers up ok halogen lamp and brightness control are CLK 4 working fine The air pump is working as well Unit is in good condition except for one rubber leg missing 439 Draeger Evita 4 Ventilator Vent power up and system self test failed to complete The unit is going on and off half way in the self test phase Otherwise in fair physical condition 440 Aesculap Elan E Drill Unit The unit power up ok the drill and footswitch are working with Foot Switch The rotation direction is functional Irrigation pump is not rotating when activated Fair physical condition 442 Datascope Passport 2 All four Passport 2 monitors are powering up ok and self test Patient Monitor x 4 completed The LCD and function keys are working fine Two of the monitors have its soft key membrane worn out but functional The parameters available are ECG NIBP SpO2 Temp and IBP Printers are work
36. d physical condition and produces a Telescope clear image 196 Storz Hopkins II 30 4mm The rigid scope is in good physical condition and produces a Telescope clear image 197 The rigid scope is in good physical condition and produces a clear image 198 Storz 30 5mm Autoclavable The rigid scope is in good physical condition and produces a Laparoscope clear image 199 Storz 0 4mm Hopkins II The rigid scope is in good physical condition and produces a Telescope clear image 200 Storz 30 6 5mm The rigid scope is in good physical condition and produces a Autoclavable Laparoscope clear image 201 Storz 30 2 9mm Telescope The rigid scope is in good physical condition and produces a clear image 202 Storz Hopkins II 12 4mm The rigid scope is in good physical condition and produces a Autoclavable Telescope clear image 203 Storz 70 2 7mm Telescope The rigid scope is in good physical condition and produces a clear image 204 Storz 70 4mm Autoclavable The rigid scope is in good physical condition but the lens at the Arthroscope end of the insertion tube appears to have small crack and therefore affects the image slightly 205 Storz 30 4mm Autoclavable The rigid scope is in good physical condition and produces a Wide Angled Arthroscope clear image 206 Storz 0 4mm Tele Otoscope The rigid scope is in good physical condition and produces a clear image 207 Storz 70
37. e mode was tested and the set value was not being maintained accurately High temperature alarm was coming on Humidifier function not installed and tray is in place The oxygen sensor not installed Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 925 Hill Rom Isolette C2000 Incubator The control was coming on intermittently but after a while it remained on the power up test ok and the temperature selector keys and LCD displays are working fine The air temperature mode was tested and the set value was maintained accordingly Humidifier function and tray is in place The oxygen sensor installed but failed to calibrate Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 926 Hill Rom Isolette C2000 Incubator The main control box has been disconnected and removed from the unit System not tested 930 Respironics Bipap System on Stand Powers up and completes the system self test successfully Display screen and function keys are in good working order Both S T and CPAP modes are selectable and functional Overall the unit is in good physical condition Mains adaptor charger is included
38. e was tested and the set value was not maintained accurately high temperature alarm comes on Humidifier function and tray is in place The oxygen sensor not installed Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 773 Hill Rom Air Shield Incubator C2000 The incubator power up ok and the temperature selector keys and LCD displays are working fine The air temperature mode was tested and the set value was not maintained accurately high temperature alarm comes on Humidifier function and tray is in place The oxygen sensor is installed but failed to calibrate Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 784 Philips Agilent M3000A Module x 6 M3001A Module x1 M3016A Module x1 and M3015A Module x 4 M3000A DE94542354 Refurbishment 12 2005 M3015A DE90500483 05 2000 M3001A DE22725847 02 2004 M3016A DE90601130 Refurbishment 04 2003 M3000A DE94519510 12 2000 M3015A DE02001308 11 2001 M3000A DE94519541 12 2000 faulty M3015A DE43535753 09 2008 faulty M3000A DE945 14520 03 2000 M3015A DE43503625 06 2005 M3000A DE
39. eeded to start the unit which is not available The equipment and the probe s are in very good overall condition but not fully tested Date of manufacture 2004 software and firmware version 1 02 Large LED Cluster probe 950 amp 640nm output 10x15 amp 9x19mW The small LED Cluster probe 905nm and 100mW Battery fitted and charging ok 1036 Philips Heartstart XL All five units power up and complete the system self test Defibrillator x 5 with ECG successfully Display screens function keys and rotary control and Patient Leads knobs are all in good working order All units include Manual and AED option sync and printer One unit has pacer as well ECG leads x4 and therapy cables x4 are included No further checks carried out The units are in reasonable physical condition Batteries and hard paddles are not included YOM 2003 2004 1037 Olympus CLK 4 Light Both light source powers up and the halogen lamp is working Source x2 The brightness controls and air pumps are fully functional Units are in very good overall condition 1041 GE Marquette Monitoring All of the Dash monitors power up successfully The LCD of Equipment to Include Dash one of the Dash 4000 monitor is very dark and the information 3000 x 3 Dash 4000 x 2 is not readable Display screens and controls of the other Flatscreen Monitor x 8 monitors are in good working order but no further tests have Solar 9500 8000m Base Unit been carried out The rotary control kn
40. ement The lifePak 1000 power up ok the voice prompt and LCD are working The unit was tested with defib analyser and good detection of VF and NSR wave form The delivered shock was within the set limit Unit is good overall condition with battery at 34 capacities 1018 Philips Gas Module M1026B The gas module power up ok but there is no monitor to fully test the system Water tap not included Year 12 2010 1031 Heraerus Centrifuge Labofuge 400e x2 Both units power up ok numerical display speed amp time selector lid lock and indicators are working fine Both centrifuges were test at maximum speed of 3000rpm and both were rotating and the time brake working fine The compartment is in fair condition rotors in place but no buckets or test tube holders The units are in good physical condition with one of the centrifuges with a cracked front plastic panel 1034 Seca CT61 ECG with Leads The unit powers up and completes the system self test successfully Display indicators and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery and no paper Tested with patient simulator Overall the unit is in satisfactory physical condition 1035 EMS Laser 755 with LED Unit power up ok Display and function keys are working Cluster Probes With the laser probe connected security code is n
41. er Light ECG Holters have ECG leads present with Digital Holters with Heads them There is also 2x Seer Light Connect and 2 x Seer Light Extend Controller There is no CD present All the units in the lot are in fair physical condition 416 Alpha Laser Electro Vollastar Alpha Laser Power up but with audible alarm when Therapy Unit Novafon output switch is activated output light not lighting up System Massager Biolas Low Level is in an average physical condition Laser Biolas No battery but in fair physical condition not tested Novafon SK Fair physical condition not fully tested 421 Heidelberg 10PAC Both devices are not powering up but good physical condition Advanced Portable Optical Scanner with Probes x 2 424 Datex Ohmeda Aestiva 5 Unit powers up ok The pressure gauge and flow meters are Anaesthetic Trolley with working fine Basal flow is functional Anti hypoxic feature Smart Vent 7900 and Oz flush are working The Oz sensor failed to calibrate The ventilator powers up and the initial self test completed successfully The display is functional however it has some pixels missing The re breathing circuit and absorber are all working Unit s minute volume was verified against the tidal volume and frequency BPM All the set parameters and alarm system is functional Back up battery needs charging up Gas hoses included for O2 Air and N20 AGSS receiver not included Overall the lot is in good physical
42. est Fail Overall Condition Good image The ultrasound probe not tested 138 OLYMPUS Insertion Tube Good condition BF 240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing and optics Remote Control Switches Fully functional 500x6 2mm Distal End Good condition Image Quality Good image Working Channel s Clear Leak Test Fail Overall Condition Good 139 OLYMPUS Insertion Tube Good condition TJF 240 Bending mechanism amp Angulation Control Fair movement and locks Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Cap not included Image Quality Good image quality Working Channel s Clear soak cas Leak Test Fail Overall Condition Good image but with leak 140 OLYMPUS Insertion Tube Good condition but sleeve at the distal damaged BF 1T240 Bending mechanism amp Angulation Control Good movement Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing and optics Remote Control Switches Fully functional 500x6 6mm Distal End Good Image Quality Good image Working Channel s Clear Leak Test
43. forms were displayed accurately The max energy discharge was tested and delivered accordingly for both manual and AED The NIBP was also tested ok Voice prompt and printer are working fine The unit comes with 10 wires ECG leads NIBP hose and one battery pack which need charging S N T04D58003 DOM April 2004 There are no charger or external paddles with the lot 767 Nihon Kohden Neuropack M1 EMG Workstation No The system power up but failed to initialise since the hard drive has been removed No further test was conducted Consists of Hard Drive Footswitch Electric Simulator SOMATO 1 amp 2 Shuttle Voltage stabilizer LCD monitor Neuropack M1 Trolley keyboard mouse headphone and Cannon printer System in fair physical condition 771 Hill Rom Air Shield Incubator C2000 The incubator power up ok and the temperature selector keys and LCD displays are working fine The air temperature mode was tested and the set value was maintained accordingly Humidifier function and tray is in place The oxygen sensor not installed Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 772 Hill Rom Air Shield Incubator C2000 The incubator power up ok and the temperature selector keys and LCD displays are working fine The air temperature mod
44. g order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 383 Philips Pagewriter Trim II The unit powers up and completes the system self test ECG Recorder with Leads successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 399 Cardiac Science Burdick The unit powers up and completes the system self test 8500 ECG on Stand with successfully Display panel and controls are in good working a ECG Leads order both manual and automatic ECG recording modes are 401 functional and the printer is also working tested with patient simulator Overall the unit is in good physical condition Mains 402 adaptor charger and ECG lead set are included Y O M 2012 2013 Lot 399 the printer roller gears have been damaged and therefore does not feed the paper continuously 404 Trumpf TITAN Operating Operating table with remote control The table power up ok Table battery is ok and all the motorised movement and positioning are working Some of the coushions pad are missing Fair physical condition 409 Quantity of Seer Compact All the Se
45. ge overall condition 135 OLYMPUS Insertion Tube Good condition TJF 240 Bending mechanism amp Angulation Control Fair movement and locks Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Cap not included Image Quality Good image quality Working Channel s Clear 1200x13 2mm Leak Test Fail Overall Condition Good image but with leak 136 PENTAX Insertion Tube Fair condition FB 18X Bending mechanism amp Angulation Control Fair movement and locks Direct View Light Guide Cable amp Light Transmission Good light output Fiberscope Objective Lens and Focusing Good optics with 4 fibres out Remote Control Switches N A Distal End Good condition Image Quality Fair image quality Working Channel s Clear 600x6 4mm Leak Test Pass Overall Condition Good overall condition 137 PENTAX Insertion Tube Good condition amp probe membrane intact FG 38UX Bending mechanism amp Angulation Control Good movement Direct View Scope 1200x13 7m Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing with 5 fibres out Remote Control Switches N A Distal End Good condition Image Quality Good image Working Channel s Clear Leak T
46. he printer door ECG lead set is included but one of the leads has been damaged 279 Mortara Instrument ELI 250 ECG Machine with Leads The unit powers up and completes the system self test successfully Display panel and controls are in working order some of the buttons are slow to respond Both manual and automatic ECG recording modes are functional and the printer is also working Tested with patient simulator Overall the unit is in satisfactory physical condition with minor damage to the printer door ECG lead set is included 282 Kodak CR950 Direct View CR System plus Qty of Assorted Cassettes The UPS is not working therefore the CR system and the touch screen monitor could not be tested The computer unit does not power up and the hard drive has been removed No further checks could be carried out Overall the CR system is in reasonable physical and cosmetic condition with minor signs of wear and tear Barcode scanner is included CR Cassettes 35x43cm x6 24x30cm x4 and 18x24cm x2 Manufacture Date 2005 283 Philips BV26 Mobile X Ray with Intensifier with Key The unit is in satisfactory physical condition with damages to the cover of the diaphragm cover on the tube head and signs of wear and tear The interconnecting cable between the c arm and the workstation does not fit into the c arm connector and therefore the system cannot be tested YOM 1994 286 287 Cardiac Science Burdick 8500 EC
47. hes Fully functional Distal End Good condition Image Quality Very good image quality Working Channel s Clear Leak Test Slight leak observed 1600x13 4mm Overall Condition Very good overall condition 119 OLYMPUS Insertion Tube Very good condition CYF 240A Bending mechanism amp Angulation Control Very good movement amp Videoscope oS Light Guide Cable amp Light Transmission Very good Objective Lens and Focusing Very good focusing amp optics Remote Control Switches Functional Distal End Very good condition Image Quality Very good image 350x5 7mm Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Very good overall condition 120 OLYMPUS Insertion Tube Very good condition CF 230L Bending mechanism amp Angulation Control Very good movement and locks Videaseape Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Very good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image quality Working Channel s Clear 1600x13 4mm Leak Test Pass Overall Condition Very good overall condition 121 OLYMPUS PCF 240L Videoscope 1600x11 7mm Insertion Tube Very good condition Bending mechanism amp Angulation Control
48. ical condition 586 Philips Pagewriter Trim III The unit powers up and completes the system self test ECG Recorder with Leads successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 587 Philips Pagewriter Trim III The unit powers up and completes the system self test ECG Recorder with Leads successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 590 Maquet Electric Operating The leg section is missing but remain sections are in fair Table Incomplete with condition The hand held control s casing has been damaged Remote Control Missing but most of the function keys are working except for the End Section trendelenberg position when pressed the table go up The reverse trendelenberg position is working Brake and battery ok 594 Spacelabs Eclipse plus ECG The unit fails to power up No further tests were carried out Machine with Leads The unit is in fair physical condition 602 Cardiac Science Burdick The unit powers up and completes the system self test
49. ies such as patient connection module system loop test box handheld push button and headsets 366 Olympus Suction Unit KV 4 Both unit power up ok and vacuum pump is working fine The manometer and vacuum regulator is in good working order and good suction power measured Both in good physical condition without accessories 368 Illumina Miseq Personal Sequencing System M000689 Unit powers up ok Display touch screen are working No further tests carried out YOM 2012 380 Philips Pagewriter Trim II ECG Recorder with Leads The unit powers up and completes the system self test successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 381 Philips Pagewriter Trim II ECG Recorder with Leads The unit powers up and completes the system self test successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery Tested with patient simulator Overall the unit is in satisfactory physical condition 382 Philips Pagewriter Trim II ECG Recorder with Leads The unit powers up and completes the system self test successfully Display panel and controls are in workin
50. ing fine The system was not fully tested with patient leads and accessories since they are not available with lot Monitors are in fair physical condition Backup battery packs are not working needs charging 446 Interacoustics AC30 Clinical Both units power up ok Display panel and function keys are in Audiometer x 1 and good working order and the units are in good physical Kamplex AC33 Clinical condition No accessories included with the lot Model AC 30 Audiometer and AC 33 455 Alcon Surgical Accurus The Accurus power up and the screen is working but was 800CS Phaco Machine with unable to perform any test since the unit requires air supply Footswitch and Hand Pieces System is in fair physical condition and comes with hand pieces x4 remote controller and footswitch Hand pieces Ref 8065740242 x amp Ref Turbosonic 375 x2 Accurus model no 800CS S N 0201102701X Year 01 2002 456 Datex Ohmeda Aestiva 5 Unit powers up ok The pressure gauge and flow meters are Anaesthetic Trolley with Smart Vent 7900 and Datex Ohmeda Anaesthesia Monitor S 5 working fine Basal flow is functional Anti hypoxic feature and Oz flush are working The Oz sensor completed the 21 calibration The ventilator powers up and the initial self test completed successfully The display is functional however it has some pixels missing The re breathing circuit and absorber are all working but a leak was detected during test run Unit s minute vol
51. ith Leads Both the units power up and complete system self test successfully Display indicators and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery on one the CT61 the printer for CT 3000B is in good working order Tested with patient simulator Overall the unit is in satisfactory physical condition 850 Metron Infusion Pump Tester 08556 Bio Tek QUED 6 Defibrillator The Infusion Pump Tester power up ok display and function keys are working fine The units are selectable and can do flow measurement and occlusion test The unit is in fair physical Analyser EME006724 condition but has not been tested with infusion device Calibration was due 10 09 2015 The defibrillator analyser power up ok display and function keys are working fine The tester was tested with a defibrillator and the wave forms simulator and energy measurement were in good working order The battery compartment at the back is missing its cover otherwise in fair physical condition Calibration was due 21 08 2015 Software version 2 11 851 Olympus High Definition LCD power up ok and the menu and setting are working The LCD Monitor and Sony VHS recorder power up not fully tested LCD in good VHS Recorder condition 853 Draeger Portable Ventilator Unit power up on battery mains adapter not included The
52. l Overall Condition Very good image quality but leaking 124 OLYMPUS Insertion Tube Fair condition CHF P20 Bending mechanism amp Angulation Control Average movement amp locks Choledochoscope lover loose Light Guide Cable amp Light Transmission Good light output few dents on guide cable Objective Lens and Focusing Good focusing with 4 fibres out Remote Control Switches N A Distal End Good condition Image Quality Good image 350x5 4mm Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Fail Overall Condition Image ok but leaking at the bending section 125 MEDTRONIC Insertion Tube Fair with tear at bending section XOMED Bending mechanism amp Angulation Control Good movement 1880500 Direct View Scope Light Guide Cable amp Light Transmission Light guide cable not included Objective Lens and Focusing Focusing ok Remote Control Switches N A Distal End Fair condition Image Quality Average image quality Working Channel s N A Leak Test Not tested 300x4 0mm Overall Condition Damaged bending section sleeve 126 OLYMPUS Insertion Tube Good condition GIF 1T240 Bending mechanism amp Angulation Control Good movement amp locks Videoscope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focusing amp optics Remote Control Switches
53. mbly cover are missing The fixation light is loose and one of the roller covers from the base of the slit lamp is missing The eyepieces are included and the optics appear to be clear 564 Aloka SSD 1700 Ultrasound The US system power up and the system initialization with Aloka UST 5524 5 completed successfully The CRT monitor and the control Probe and Aloka UST 979 console functions knobs and keys are working fine The gain 3 5 Probe trackball and slider pot are functional The convex and linear array probes produces good air scan and the probe is intact and in fair condition Available modes are B M and Doppler B amp W printer power up but failed to print out The system is in average overall physical condition No endo rectal probe included 568 Haag Streit Octopus 301 Unit failed to power up but in good physical condition Field Analyser on Trolley 569 Draeger Babytherm 8000 The unit power up ok temperature controller is working with display and temperature selector keys in good working order The motorised height adjustment is functional Perspex cover and the unit itself is in good physical condition 576 Medi Plinth Electric Patient The motorised height adjustment is not working Backrest Couch upholster is in fair condition and castors brake are functional 584 HNE Patient Couch x 1 Couch s hydraulic lift back rest adjustment brake and castors are working fine Upholster are in good condition Lot is in good overall phys
54. nicaid Foetal Monitors The CTG power up ok the LCD and function keys are working fine The US transducer is in good working order and gives a good graphic trace on display and on the printout as well The Toco transducer is responding fine when pressure is applied System is in fair physical condition US x1 and Toco x1 transducers included with spare printer paper 980 Oxford Sonicaid Foetal Monitors The CTG power up ok the LCD and function keys are working fine The US transducer is in good working order and gives a good graphic trace on display and on the printout as well The Toco transducer is not included but one was used to test the system but not responding when plugged in System is in fair physical condition Transducer included US x1 with spare printer paper 1003 Iris Medical OcuLight SLx Ophthalmic Laser System The unit power up and the numerical display test and system self test completed ok However system with not start without a fibre being connected Unable to fully test the laser system Unit is in good physical condition and comes with footswitch and ON OFF key 1006 Medtronic CR Plus and Lifepak 1000 Defibrillators The LifePak CR Plus power up ok the voice prompt is working The unit was tested with defib analyser and good detection of VF and NSR wave form The delivered shock was within the set limit Unit is in good overall condition with the battery at 1 4 capacities will need replac
55. ns and Focusing Very good focusing amp optics Remote Control Switches Functional Distal End Very good condition Image Quality Very good image Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Very good overall condition 108 PENTAX ECY 1570 350x5 8mm Insertion Tube Fair condition with crush on bending section Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Poor light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Fair physical condition 109 PENTAX EC 3830LK 1600x13 7mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Fail Overall Condition Good physical condition with slight leak 110 OLYMPUS GIF XQ240 Videoscope 1000x10 2mm Insertion Tube good condition Bending mechanism amp Angulation C
56. o test the unit Workstation is also not included 680 Quantity of W amp H Bien Air Lot consists of NSK ER4 x8 and E4R x1 W7H WD073M x8 NSK Dental Hand pieces WP 64M x12 WI 75E KM x2 Synea WA 99A x1 and 9 others I could not identify all in good physical condition but not tested 683 Kavo Intraflex LUX 2 2307 Lot consists of Kavo handpieces Intraflex Lux 2313L x2 LN Dental Handpiece New Intraflex Lux 2 2307LN x1 Kavo intra prophylactic head with Assorted Dental Heads 31GR x3 IntraLux Head 31LR x2 amp Intro Endo Head 3GB x1 x 7 Kavo Intraflex As new LUX2313L Dental Handpiece x 2 New 709 Hydraulic Patient Couch Couch s hydraulic lift back rest adjustment brake and castors Black x 2 are working fine Upholster are in good condition Lot is in good overall physical condition 716 Philips Series 50XM Foetal Monitor with Leads 3650G19402 Powers up ok Display panel and function keys are in good working order All monitoring channels which include US TOCCO SpO2 and NIBP are functional and the printer is also working All accessories are included 4x US and 4x TOCCO transducers 2x SpO2 probes 2x Event markers and NIBP cuffs Overall the unit is in reasonable physical condition The corner of the printer door has been damaged slightly 719 Philips Pagewriter Trim HI ECG Recorder with Leads The unit powers up and completes the system self test successfully Display
57. ob is missing from one x 8 Tram Rac with 450SL of the monitors and the Dash 3000 monitors have small cracks Module x 7 and Qty of on the casing The other units within the lot have not been Assorted Leads and tested No accessories included with the monitors Mounting Brackets 1042 Philips Pagewriter Trim II The unit powers up and completes the system self test ECG Machine with Leads successfully Display panel and controls are in working order Both manual and automatic ECG recording modes are functional but the printer could not be tested due to a bad backup battery When tested with patient simulator the ECG was not recording any waveforms Overall the unit is in satisfactory physical condition 1043 Oxford Sonicaid Foetal The CTG power up ok the LCD and function keys are working Monitors fine The US transducer is in good working order and gives a good graphic trace on display and on the printout as well The Toco transducer is not responding when plugged in System is in fair physical condition Transducer included US x1 amp Toco x1 with spare printer paper 1047 Sidhill Hydraulic Patient Couch s hydraulic lift back rest adjustment brake and castors Couch x 1 are working fine Upholster are in good condition Lot is in good overall physical condition 1081 Viasys Micro Loop The spirometer has no CD with software or a cable to connect Spirometer to PC The unit is in good physic
58. on 250x3 8mm Image Quality Fair image Working Channel s N A Forceps Elevation Mechanism N A Leak Test Not tested Overall Condition Fair image and fair physical condition 147 KARL STORZ Insertion Tube Poor condition with crushed insertion tube amp sleeve damage at the bending section 11101 RP1 A Bending mechanism amp Angulation Control Good movement View Scope Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Poor focus with liquid damage to the lens 250x3 8mm Remote Control Switches N A Distal End Fair condition Image Quality Poor image Working Channel s N A Forceps Elevation Mechanism N A Leak Test Not tested Overall Condition Poor image and poor physical condition of the insertion tube 148 OLYMPUS GIF 1T240 Videoscope 1000x12 0mm Insertion Tube Fair condition with dent next to control box Bending mechanism amp Angulation Control Good movement and locks Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Good focus amp optics Remote Control Switches Fully functional Distal End Fair condition Image Quality Good image quality Working Channel s Clear Leak Test Fail Overall Condition Good image and optics but leaking 149 OLYMPUS PCF 240S Videoscope 900x12 8mm Insertion Tube Fair
59. on keys are in working order The unit is in good physical condition No attachments accessories included 38 Seca CT3000I ECG Both the units power up and complete system self test Recorder with Leads Seca successfully Display panel and controls are in working order CT8000i ECG Recorder No Both manual and automatic ECG recording modes are Leads functional and the printer is also working Tested with patient simulator Overall the unit is in satisfactory physical condition ECG lead set is included 55 Storz 20131520 Xenon Nova The light source is in reasonable physical condition and Light Source functional Lamp and brightness control in good working order The lamp indicator is red 63 X Ray Cassettes x 20 Fuji Cassette sizes 18x24cm and 24x30cm 3 boxes of x ray films CR 3 Box Tray Film 100 per box sizes 35x35cm and 30x40cm Overall the cassettes are in reasonable physical condition 67 Olympus WM SC All units power up ok The LCD monitor and the light source Workstation Stack System are in good working order The video processor appears to be with Visera OTVS7 functional but a camera head is not included to test the system Processor CLV S 40 Light Overall the units are in reasonable physical condition Source Olympus Flat Screen Monitor and Panasonic DVD Player 69 Datex Patient Monitor AS 3 Both monitors power up ok and the LCDs function keys and rotary selectors are working fine One of the LCD has some pixel is
60. ontrol Good movement amp locks Light Guide Cable amp Light Transmission Good with dent on the fibre guide cable Objective Lens and Focusing Very good focusing amp optics Remote Control Switches Functional Distal End Good condition Image Quality Good image Working Channel s Clear Forceps Elevation Mechanism N A Leak Test Pass Overall Condition Good overall condition 111 OLYMPUS Insertion Tube Very good condition CF 240L Videoscope 1600x12 7mm Bending mechanism amp Angulation Control Very good movement and locks Light Guide Cable amp Light Transmission Very good light output Objective Lens and Focusing Good focusing amp optic system Remote Control Switches Fully functional Distal End Good condition Image Quality Very good image quality Working Channel s Clear Leak Test Pass Overall Condition Very good overall condition 112 PENTAX ECY 1570 350x5 8mm Insertion Tube Good condition Bending mechanism amp Angulation Control Good movement Light Guide Cable amp Light Transmission Good light output Objective Lens and Focusing Not tested Remote Control Switches Not tested Distal End Good condition Image Quality Image quality not checked no processor Working Channel s Clear Leak Test Pass Overall Condition Good physical condition
61. ood light output Videoscope Objective Lens and Focusing Good optic system Remote Control Switches Fully functional Distal End Good Image Quality Good image quality Working Channel s Clear 1100x12 0mm Leak Test Pass Overall Condition Good overall condition 174 OLYMPUS Insertion Tube Good condition PCF 240S Bending mechanism amp Angulation Control Good movement and locks Videoscope Light Guide Cable amp Light Transmission Good light transmission Objective Lens and Focusing Good focus and optics Remote Control Switches Functional Distal End Good condition 900x12 2mm Image Quality Good quality images Working Channel s Clear Leak Test Pass Overall Condition Good overall condition Inspection Report Sale 1521 Date 13 10 2015 These reports are compiled by our Medical Engineering Team amp are for guidance only all lots are sold as seen as per our Conditions of Sale We recommend that wherever possible you inspect lots prior to purchase Lot Description Condition Datex Ohmeda Aestiva 5 Anaesthetic Machine The unit powers up and completes the system self test successfully Ventilator display panel and controls are in good working order Breathing system absorber and bellows are all in place Overall the unit is in reasonable physical condition The unit has not been fully tested All medical
62. ress and canopy in good condition 809 Draeger Evita 4 Ventilator Front Screen Missing Unit in fair physical condition with display module missing Come with gas hoses Year 2005 813 Fresenius Medical Care 4008S Dialysis Machine The unit powers up ok system self test passed successfully The LCD screen LED bar graph and function soft keys are working fine The unit function menus are accessible The full function of the system was not tested The following modules are attached BVM Bibag Air bubbles Detector Blood Pump and Online Plus system Unit is in fair physical condition with water and drain hose attached Operation counter 30320hrs Software version 4 3 User manual included 814 Fresenius Medical Care 4008S Dialysis Machine The unit powers up ok system self test passed successfully The LCD screen LED bar graph and function soft keys are working fine The unit function menus are accessible The full function of the system was not tested The following modules are attached BVM Bibag Air bubbles Detector Blood Pump and Online Plus system Unit is in fair physical condition with water and drain hose attached Operation counter 27772hrs software version 4 3 820 Draeger Oxylog Portable Ventilator Unit is in good physical condition with O2 hose attached the selector knobs are intact Not tested with medical gas 830 Seca CT61 ECG Machine with Leads and Seca CT 3000B ECG Machine w
63. s tested and the set value was maintained accordingly Humidifier function and tray is in place The oxygen sensor installed but failed to calibrate Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 806 Hill Rom Isolette C2000 The incubator power up but LCD failed to come on Incubator 807 Hill Rom Isolette C2000 The incubator power up test ok and the temperature selector Incubator keys and LCD displays are working fine The air temperature mode was tested and the set value was maintained accordingly Humidifier function and tray is in place The oxygen sensor installed but failed to calibrate Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with mattress and canopy in good condition 808 Hill Rom Isolette C2000 Incubator The incubator power up test ok and the temperature selector keys and LCD displays are working fine The air temperature mode was tested and the set value was maintained accordingly Humidifier function not installed and tray is in place The oxygen sensor not installed Skin temperature probe not included Skin temperature mode not tested The motorised height adjustment is working Incubator is in fair overall physical condition with matt
64. sues but the displayed informations are all readable Battery packs are included ok will need charging ECG SpO2 and NIBP are working fine Also included SpO2 finger probe x1 and ECG trunk cable x1 and NIBP hose x3 and mains adapter x2 The monitors are in fair physical condition 179 Linvatec C3282 Camera Video camera heads are in good physical condition but unable Head x 3 to test for image quality no compatible video processor 181 Toshiba Linear Array Model PLT 704SBT S N 99B13X3312 The connector Transducer 11L4 cable probe head and membrane intact Probe is in good physical condition but not tested for its image quality 181 Toshiba Linear Array Model PLT 704SBT S N 99B13X3312 The connector Transducer 11L4 cable probe head and membrane intact Probe is in good physical condition but not tested for its image quality 183 Richards 30 Rigid Scope The telescope is in good physical condition but with poor optics The lens is blurry Dismeter 2 6mm x length 140mm 187 Philips L17 5 Ultrasound The transducer is in good physical condition but has not been Transducer tested 193 Storz 30 6 5mm The rigid scope is in good physical condition and produces a Autoclavable Laparoscope clear image 194 Storz 30 1 9mm The rigid scope is in good physical condition and produces a Arthroscope clear image 195 Storz Hopkins II 0 4mm The rigid scope is in goo
65. ume was verified against the tidal volume and frequency BPM All the set parameters and alarm system is functional Back up battery needs charging up Gas hoses included for O2 Air and N20 AGSS receiver not included Overall the lot is in good physical condition Software version 3 5 Vent Mode Volume amp Pressure control The patient monitor S 5 is working fine SpO2 and gas monitor tested ok Other parameters were not fully checked System is in fair physical condition 476 Ohmeda Excel 210 SE Anaesthetic Trolley Unit s pneumatic system startup ok The pressure gauge and flow meters are working fine Basal flow Anti hypoxic safety feature and O flush are working There is a leak at the ON OFF switch No ventilator attached Gas hoses included for O2 Air and N20 AGSS receiver included System is in average physical condition 483 GE Dash 3000 x 1 and Dash 4000 Patient Monitors x 1 Both monitors power up successfully Display screens and control panels are functional but no further tests have been carried out Available parameters include ECG SpO2 NIBP and Temp Dash 4000 includes IBP as well Some of the front panel controls of the Dash 3000 monitor are slow to respond and has to be pressed hard to activate them Otherwise the monitors are in good physical condition No accessories included with the lot 484 Braun Perfusor Compact 3 Syringe Pumps x 8 All 8 syringe pumps powers up ok and