Leica Angle Two™ User Manual

image

Contents

1. Leica Biosystems St Louis LLC 2009 41 Using the Instrument This chapter describes the whole process of using the Angle Two system to perform surgery from starting the software to successfully placing the probe This chapter has the following sections Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 Preparation on page 43 Selecting the Target Site on page 43 Positioning the Animal on page 43 Locating Bregma and Lambda on page 44 Locating the Target Site on page 45 Probe Placement on page 46 42 Using the Instrument Preparation 1 Ensure that the instrument is clean see Cleaning and Maintenance on page 47 2 Turn the computer on and start the Angle Two software see Figure 3 3 Align the instrument using the alignment kit see Alignment on page 33 Selecting the Target Site The software opens in Set Target mode 4 Enter the target site into the Target Red boxes beneath using one of the following options see Set Target on page 26 i Select the site on the atlas ii Open saved coordinates iii Manually type in the coordinates 5 If using offsets see Initial Settings on page 29 click Setup and check the Use Zero Offsets box see Figure 45 and enter the values 6 Unlock the rotation and tilt clamps and move the manipulator to the angle you want to use to approach the target site Relock the clamps when
2. Software Overview Atlas The Atlas mode shows the exact position of the probe including the rotation tilt and the deviation of the animal s skull from 0 skull flat The Skull value is only determined once the bregma and lambda points have been located To save the atlas coordinates click Export Position see Exporting on page 32 Vertical Vertical mode is available for users wanting to measure the vertical position from the top of the brain rather than the top of the skull In this mode the tilt and rotation must be zero and the virtual skull flat feature is unavailable To locate the vertical zero point a user locates bregma and from here moves the probe directly over the target site which must have already been determined in pilot studies The target site is marked on the skull and a hole drilled The probe tip is carefully lowered through the hole towards the brain until the blood fluid visible in the hole dips downwards This indicates that the probe is putting pressure on the dura membrane This is the vertical zero point To mark the location click Z to zero the instrument s coordinates Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 28 Software Overview 4 3 Initial Settings Before locating the target site you must do the following on the Angle Two software e Select the tilt direction e Zero the tilt and rotation angles e Enter any probe offsets e Enter the l
3. Leica Angle Two Computer Assisted Stereotaxic System User Manual Living up to Life Leica MICROSYSTEMS Intended Use Statement The Angle Two system is designed for stereotaxic surgery on small animals only and is not intended for use on humans Trademarks Leica Leica Microsystems and logo are trademarks of Leica Microsystems GmbH HRB 5187 and used under license Angle Two is a trademark of Leica Biosystems St Louis LLC 61 1561098 Other trademarks are the property of their owners Copyright Leica Biosystems St Louis LLC owns the copyright on this document and any associated software Leica Biosystems St Louis is part of the Leica Microsystems group of companies Under law our written permission is required before either the documentation or the software is copied reproduced translated or converted to electronic or other machine readable form in whole or in part Doc 95 8808 Rev A01 Leica Biosystems St Louis LLC 2009 Important Information for All Users The term Leica Microsystems when used in text in this document refers to Leica Biosystems St Louis LLC Due to a policy of continuous improvement Leica Microsystems reserves the right to change specifications without notice Warranty claims can be made only if the system has been used for the specified application and operated according to the instructions in this document Damage resulting from inappropriate handling and or misuse of the product wil
4. User Manual Rev A01 Leica Biosystems St Louis LLC 2009 34 Alignment Attach the clamp adaptor to the dial indicator by sliding the raised area of the dial indicator into the groove of the adaptor ij _ HE Clamp adaptor Dial indicator G Figure 36 The raised area of the dial indicator slides into the groove of the clamp adaptor Slide the clamp covering towards the middle of the clamp to reveal the clamp hole Push the top of the clamp adaptor into the clamp hole and release the covering to lock the adaptor in position Clamp screw p Clamp covering Figure 37 Slide the clamp covering to the middle to insert the clamp adaptor inside Attach the clamp to the bottom of the alignment shaft and tighten the clamp screw so that it cannot move Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 35 Alignment 8 Attach the other end of the alignment shaft into the manipulator and tighten the thumbscrew Figure 38 The dial indicator attached to the Angle Two 9 Remove the ear bars and install the T bar Lock it into position Figure 39 The T bar locked into position Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 36 Alignment Horizontal Alignment The ML horizontal path of the probe is controlled by the rotation of the instrument This part of the alignment procedure checks the ML horizontal path is perpendicular To test this the dial indicator
5. A single blue circle appears when the probe is 0 05 mm from the target A second blue circle appears when the probe is at the target site 34 Perform the operation and when complete withdraw the probe using the fine drive until the probe is clear of the brain If you are performing bilateral surgery i Select Set Target mode and click see Figure 25 to change the sign of the ML coordinate ii Move the probe to the target site and repeat steps 22 to 32 35 Remove the animal from the instrument and close up the incision Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 46 Cleaning and Maintenance Caution Do not autoclave The bushings and probe holder block can be damaged Caution A Do not lubricate the axis rod slides or threaded drive The plastic bushings can absorb oil and cause the instrument to seize Caution A Do not use alcohol to clean the surface of the probe holder block as it can cause cracking Wash the instrument with a cloth or paper towel dampened with water or dishwasher detergent If sterilization is needed use Cidex or a similar glutaraldehyde based sterilizer Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 47 Specifications Instrument Dimensions and Weight Base width Base length Base height U frame width U frame length U frame height Base and manipulator height Base weight no manipulator Resolution Linear Sc
6. Leica Biosystems St Louis LLC 2009 20 Components Probe clamp screw No Figure 20 The probe shaft screw Attaching a Probe At the bottom of the standard probe shaft is a plastic block with three grooved faces and one flat face Insert the probe into one of the grooves grooves are 1 mm apart and tighten the metal clamp to secure Probe shaft Grooves on the plastic block Clamp screw Metal clamp Figure 21 The probe holder at the bottom of the probe shaft Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 21 Components 3 5 Vernier Scales All Leica stereotaxic instruments have vernier scales for manual measurements While the instrument software makes all calculations for you some manual measurements might be needed for example to record the exact position of the ear bars for repeat studies To use the scales follow the example below The main scale graduated in millimeters is on the ear bar and the vernier is on the U frame To measure the position of the ear bar 1 Locate the zero on the vernier scale and read below on the ear bar scale to get the first measurement in the example it measures 1 2 cm 2 Use the vernier scale to measure the second decimal place To do this locate the line of the vernier scale that directly aligns with a millimeter line on the main scale below only one can align the rest are slightly off The third line of the vernier scale al
7. deep Using the Fine Drive 1 Push the fine drive towards the cog to engage it Witt Figure 10 Engaging the fine drive Figure 11 Disengaging the fine drive 2 Turn the fine drive to move in the DV direction 3 When finished pull the drive away from the cog to disengage Note When engaging the fine drive if no click is heard slowly turn the coarse DV drive until you hear it click into place Immediately stop moving the coarse DV drive Caution AN Do not use the coarse DV drive when the fine drive is engaged as it will damage the instrument Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 15 Components 3 3 Manipulator Tilt Rotation and Swing Out Tilt rotation and swing out allow surgery to be performed at any angle To change the tilt or rotation or use the swing out unlock the clamps indicated Swing Out Tilt Rotation gt 7 Figure 12 The swing out rotation and tilt clamps Tilt changes the manipulator angle and in turn the probe s angle of approach The primary tilt can be in either the AP or ML planes Rotation rotates the plane of tilt between AP and ML Figure 14 Plane of the tilt with rotations from the basic AP and ML directions Swing out is used to swivel the manipulator 90 to allow access to the animal Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 16 Components Tilt The primary tilt plane c
8. is moved horizontally along the length of the T bar 10 Use the AP drive to position the dial indicator spring against the T bar 3 lt q AP drive Dial indicator spring T bar A Figure 40 The dial indicator spring against the T bar 11 Continue to move the AP drive so that the dial indicator arm completes one full rotation Stop moving the drive when the dial reads zero Figure 41 Dial indicator at zero Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 37 Alignment 12 Use the ML drive to move the spring horizontally along the T bar Move the indicator to one side and then across the total length of the bar to the other side Figure 42 Move the dial across the T bar to check the horizontal alignment The instrument is horizontally aligned if the dial arm only moves several divisions back and forth but stays roughly in the same position The instrument is not horizontally aligned if the dial arm continuously moves in one direction To select a new horizontal position i Loosen the rotation clamp see Figure 35 and carefully adjust the manipulator into a new position ii Retighten the clamp and repeat steps 1 3 NOTE if you cannot horizontally align the instrument contact Leica Microsystems Vertical Alignment The vertical path of the probe is controlled by the tilt of the instrument This part of the alignment procedure is to confirm that the vertical pa
9. its exact position later so you do not have to relocate the target site 27 Set up the drill and select the drill bit We recommend using a manual drill with settable depth stops This reduces the risk of drilling too far and into the brain tissue below Select the smallest possible drill bit 28 Drill into the skull at the marked spot Gently rotate the drill taking care not to drill too far Caution AN Drill bits have a spherical head so the center of the drilled area is deeper than the edges Be careful that the center of the drill does not go too far and strike a blood vessel beneath the bone Caution A After drilling the hole bone might remain hidden around the outer edges of the drill site Be careful when inserting the glass probe as the bone can easily break the probe tip Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 45 Using the Instrument Probe Placement 29 30 31 32 33 Unlock the swing out clamp and reposition the manipulator and probe over the animal Relock the swing out clamp Use the coarse DV drive to approach the target site Engage the fine drive when about to enter the brain A single blue circle appears around the blue crosshairs when the probe is 0 05 mm from the target site see Figure 50 Stop moving the probe once a second blue circle appears around the blue crosshairs This indicates that the target site has been reached Figure 50
10. loading installing executing or displaying the Software You may not modify the Software or disable any licensing or control features of the Software Ownership The Software is owned and copyrighted by Leica Microsystems Your license confers no title to or ownership in the Software and is not a sale of any rights in the Software Copies and Adaptations You may only make copies or adaptations of the Software for archival purposes or when copying or adaptation is an essential step in the authorized Use of the Software You must reproduce all copyright notices in the original Software on all copies or adaptations You may not copy the Software onto any public network No Disassembly or Decryption You may not disassemble or decompile the Software unless Leica Microsystem s prior written consent is obtained In some jurisdictions Leica Microsystem s consent may not be required for limited disassembly or decompilation Upon request you will provide Leica Microsystems with reasonably detailed information regarding any disassembly or decompilation Transfer Your license will automatically terminate upon any transfer of the Software Upon transfer you must deliver the Software including any copies and related documentation to the transferee The transferee must accept these License Terms as a condition to the transfer Termination Leica Microsystems may terminate your license upon notice for failure to comply with any of these License Te
11. positioned Positioning the Animal If repeat surgery is planned the position of the animal must be easily reproducible Always use ear bars and an incisor bar suitable for the animal Installing the Ear Bars 7 Set the left ear bar tip 1 cm left of the center position and lock it in place 8 Remove the right ear bar and leave the lock open Locked Unlocked Figure 48 The left ear bar locked 1 cm from the center position and the right ear bar removed with the lock open 9 Unscrew the incisor bar s horizontal screw see Figure 5 and slide the incisor bar to the rear of the instrument 10 Hold the animal s body with one hand and the head with the other hand Move the animal so that the left ear bar is inside the ear canal Point the animal s nose slightly to the right to help with insertion Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 43 Using the Instrument 11 Support the head and insert the right ear bar into the right ear 12 Lower the right ear bar into the instrument and lock it in place 13 Check that both ear bar scales have the same readings If not loosen both ear bar locks and move the animal until both ear bar scales agree then relock the ear bars 14 Gently move the animal s nose from side to side and up and down to check the ear bars are correctly positioned If the nose moves reposition the ear bars further inside the ear canal Ens
12. pull or attempt to lift the instrument by any of its cables Caution The computer must be connected to an earthed mains power outlet socket Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 8 Components This chapter describes the parts of the Leica Angle Two system and their functions The components are divided into the following sections e U Frame on page 10 includes the ear bars nose clamp and incisor bar e Manipulator Linear Movements on page 13 the three coarse drives that move the manipulator in the AP ML and DV directions and the fine drive for small movements in the DV direction e Manipulator Tilt Rotation and Swing Out on page 16 how to tilt and rotate the probe and swing out for moving the probe away in order to get clear access to the animal e Probe Shaft on page 20 how to change the shaft s vertical and horizontal positions and attach a probe e Vernier Scales on page 22 how to use the vernier scales e Computer on page 23 lists the hardware and software provided e Ancillary Products on page 23 other products available from Leica Microsystems for stereotaxic surgery Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 9 Components 3 1 U Frame Figure 4 shows the Angle Two U frame with a rat adapter and 45 ear bars attached Some of the part names in the figures below link to sec
13. rar A AE A ARAS a A A ae 2 Contacting Leica Microsystems s se Lere rai 004se e e ee eee eee A aa 2 1 INTrodUucHO Nina ada a A L1 iBaSiC Steps OF USB cu a it do ra io ae 5 2 Instrument Installation is 503 aa a O Za ENCON E wists Se ae eee eee Pa gp aad ae Sk ee a Re a Re ce ae ee ee No Cos 6 22 CSUN DIV ices ater cae a ane a ak sy ais ee as aa ge a Cea ae 7 gt COMPONents 4 ica Ex iS Sl Ade MAG osc aes reste os aa rs sara te tes Si aa E 10 3 2 Manipulator Linear Movements 0 000 ee ee eee eee ees 13 3 3 Manipulator Tilt Rotation and Swing Out 2 saaa aasa eaaa es 16 23A IPODS Smalls saws snes A ERA SOS ee he eee eee Ee Ae ee ee eee ee 20 2 9 VerAlelSCaleS era o 4 tet bo ee AS Ga ROMS Oe whe a od Hake GE 22 9 0 COMPU sews vu ist aha oa dao Ba ee eee heehee 23 37 ANIV Products cc oa a DR Me Wee ee ee ee 23 4 Software OvervieW iia a ida 24 Mele AMAS a ear di a dr a Sd a 25 a2 A E ER eee 26 ta INtalSetnds diia dido dci 29 AE ICONS arr n rA es o rc a o o ee es 31 5 Alignment E E E E E E E E E E E E E E E E E E E E E _ E E E _ E E A E E A E E 33 5 1 CHECKISE 73 4 5 5 ac Ac tee aaaea a a he ae hae a ad ba 33 5 2 Procedu oe oh Se Ee rtd ie eee ee ero ee eae tte ee tee he erg of dl acl is Boned 34 6 Using the Instrument sicario da iaa AZ Ja 7 Cleaning and Maintenance o SpeciiCations lt s inieraiiiaaaraaai ds Software Licence TerMS 0c0cco
14. s angle of variance from flat 09 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 44 Using the Instrument Locating the Target Site 21 Click To Target see To Target on page 27 to view the distances to the target site 22 Use the ML and AP drives to position the probe directly above the target As the target site is approached the ML and AP distances count down 23 Continue to move the probe until the AP and ML axis displays read zero A line appears on screen showing the path of descent DV towards the target Left Mode Probe Set Target j a crosshair C Atlas 1 C Vertical Line of ML d AP t To Target descent an a ML 0 00 mm j NN ooa Target site DV 434 mm position p Setup i 0 2 i C IToML IToAP Select Atlas Rat Coronal Save Named Point Export Position Figure 49 When the ML and AP coordinates read zero the probe crosshair blue shows a line of descent to the target crosshair red 24 Stop moving the drives Note you can preview the probe s path through the brain to the target site by removing the probe and lowering the shaft using the DV drive Drilling 25 Lower the probe using the DV drive so that the probe is just above the skull and mark the spot on the skull 26 Unlock the swing out clamp see Swing Out on page 19 and move the manipulator out of the way Using the swing out ensures the probe is returned to
15. 23 The Angle Two screen with the left manipulator attached Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 24 Software Overview 4 1 Atlas The instrument comes with either one mouse or one rat brain atlas installed Contact Leica Microsystems to see what other animal atlases are available Click Select Atlas to choose an atlas The atlas consists of a series of coronal plates displayed in the center of the screen The position of the current plate relative to the rest of the brain is shown in the top left corner Click Rostral and Caudal to move through the coronal plates Figure 17 Plate position in the brain Coronal plate 6 Atlas selection Bregma 0 20 mm 7 6 5 4 2 1 0 1 2 3 4 5 6 7 Rostral button SP 20111 Select Atlas Rat Coronal v Save Named Point Export Position Cauda Caudal button Figure 24 The coronal plate is displayed with an insert in the top left corner showing its position in the brain Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 25 Software Overview 4 2 Modes Four modes are used during surgery Set Target To Target Atlas and Vertical Each mode is used for a specific function and has its own set of coordinate boxes that appear beneath the mode selection area Click the appropriate mode to activate it for use le Left Mode Set Target Modes E C To Target C Atlas C Vertical Target Red l ML 35m a Coordi
16. Select Atlas Rat Corona Saved Bons Bee coordinates Figure 26 Target coordinates can be entered manually into the coordinate boxes loaded from saved coordinates or by positioning the mouse on the screen and clicking at the position If performing bilateral surgery after the first target site has been located and probe placed click beside the ML coordinates box The target site is moved to the opposite side of the coronal plate and the To Target coordinates are automatically updated To Target The To Target mode shows the distance that the probe must move in the ML AP and DV axes to reach the target site The probe s current position is shown as a blue crosshair and the target site as a red crosshair The To Target coordinate values reduce as the manipulator moves closer to the target site until they read zero indicating the target site has been reached Figure 30 Probe a position C Set Target To Target C Atlas C Vertical Distance to o Targe ta rg et ML 0 00mm AP 1 99 mm Aga i DV 585mm 2 g l r m ms Target site rar tee at Bregma at Lambda a Interaural 6 44 mm Setup 3 2 1 0 1 2 3 a 5 3 7 7 6 5 4 C ITOML EToAP Select Atlas Rat Coronal gt Save Named Point Export Position Figure 27 The distance the probe needs to move to reach the target site is shown in To Target mode Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 27
17. ales Vernier Digital display Internal scale Drive Advancement per Revolution Linear axis drive screw pitch Fine drive Manipulator Movement AP ML DV Tilt Rotation Rotary encoder resolution 253 mm 10 0 355 mm 14 0 20 mm 0 8 165 mm 6 5 200 mm 8 0 55 mm 2 15 381 mm 155 6 1 kg 13 5 Ib 0 1 mm 0 01 mm 0 005 rev 5 mm rev 0 2 rev 0 25 mm rev 0 01 rev 80 mm 3 15 80 mm 3 15 62 5 mm 2 5 90 ML or AP axis or combined 360 1000 lines rev 0 36 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 48 Tooth Bar Movement Ranges DV AP Probe Holder Diameter Shaft length Probe holder block Standard clamp holds Side clamp holds C clamp holds Set screw on shaft holder Specifications 20 mm 0 8 44 5 mm 1 8 7 88 mm 0 31 200 mm 7 9 11 mm 0 4 square 10 grooves spaced at 1 mm on 3 sides 0 2 to 2 mm shaft 0 008 to 0 08 0 2 to 4 5 mm shaft 0 008 to 0 2 0 2 to 4 5 mm shaft 0 008 to 0 2 3 32 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 49 Software Licence Terms Software Licence Terms The following License Terms govern your use of the accompanying software unless you have a separate signed agreement with Leica Microsystems License Grant Leica Microsystems grants you a license to Use one copy of the Software Use means storing
18. an be in either the AP or ML directions see Figure 15 Always remember to update the software with the primary tilt direction Tilt Direction on page 29 To adjust the tilt loosen the tilt clamp and move the manipulator to the desired angle then retighten the clamp A scale below the clamp shows the amount of tilt Figure 15 Tilt movement in the AP left and ML right directions Changing From AP to ML Tilt 1 Unlock the swing out clamp and swivel the manipulator 90 so the probe faces the front of the instrument see Swing Out Swing out clamp Tilt encoder Rotation clamp Figure 16 The manipulator swung out to the front of the instrument Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 17 Components 2 Unlock the rotation clamp and rotate the base of the manipulator 90 in the opposite direction to the swing out in the preceding step so the tilt encoder is facing the front of the instrument and the probe is back over the U frame The tilt is now in the ML direction see Figure 15 3 Retighten the clamps Changing From ML to AP Tilt 1 Unlock the rotation clamp and rotate the base of the manipulator 90 so the tilt encoder is facing away from the U frame and the probe faces the front of the instrument 2 Unlock the swing out clamp and swivel the manipulator 90 in the opposite direction to the rotation so the probe is back over the U frame The tilt is now in
19. anipulator has three linear movements e AP Anterior Posterior moves horizontally along the length of the instrument e ML Medial Lateral moves horizontally across the width of the instrument e DV Dorsal Ventral moves vertically towards the animal AP ML DV Figure 8 The three linear movements of the manipulator AP ML and DV Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 13 Components Each linear movement is controlled by a coarse drive and in the DV direction by an additional fine drive When locating the target site first use the ML and AP coarse drives to move the probe into place above the target Then lower the probe until it reaches the brain using the coarse DV drive Finally engage the fine drive to continue DV movements to the target within the brain it DV drive ________y ys y Il a lE ML drive E Vm Fine drive AP drive Figure 9 The AP ML DV and fine drives move the probe in linear movements Caution A Do not continue to turn the drives at the end of their movement range as it will damage the nut below Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 14 Components Fine Drive The fine drive see Figure 9 provides greater control and smoother movement in the DV direction when placing or removing the probe from the brain The fine drive helps prevent inserting the probe too
20. ccoccco ee 50 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 3 Introduction The Leica Angle Two stereotaxic system is designed to perform surgery at any angle and position Reaching the target site is easier and more accurate than manual systems as the Angle Two is computer assisted all Calculations are done for you meaning no user errors and no pilot studies are needed Probe position is set with three linear movements in anterior posterior medial lateral and dorsal ventral axes and two rotary movements tilt and rotation Encoders on the manipulator measure these positions and relay them to the Angle Two software Using this information the software calculates the exact distance the manipulator must move to reach the target site at any tilt or rotation The Virtual Skull Flat feature in the software uses the bregma and lambda reference points to calculate skull tilt You don t need to move the animal to align the skull horizontally or perform difficult calculations to incorporate tilt into the target coordinates the software does it for you automatically setting appropriate distance to target figures To use the Angle Two system choose one of the fully integrated animal atlases available and then select the coronal plate that shows the target site Click on the screen to mark the site or alternatively use previously saved target coordinates Once the target site is entered move the pr
21. ch e A brain atlas in book form appropriate to the species you have selected e Angle Two Support DVD e Computer with Windows XP Angle Two software v 3 0 0 and a rat and or mouse brain atlas installed e Monitor e Keyboard e Mouse Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 6 Instrument Installation 2 2 Assembly When unpacked the Angle Two instrument has either one or two manipulators folded over the U frame 1 Unlock the tilt clamp and gently lift each manipulator upright into position The rotation encoder will point down and the tilt encoder to the left away from the U frame Figure 1 Unlock the tilt clamp highlighted and gently lift the manipulator upright into position Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 7 Instrument Installation 2 Connect the five instrument cables into the ports at the rear of the computer Each cable and its corresponding port has the same label If you have a second manipulator insert the cables into the second card provided S a 0 0 000 00 8 a lt o Figure 2 Angle Two cables and computer ports Connect the mouse and keyboard into the computer s USB ports 4 Connect the monitor and computer power cables into an outlet Turn on the computer The Angle Two icon is displayed on the desktop Te we Angle Two Figure 3 The Angle Two desktop shortcut Caution Do not
22. ck Setup and click Zero Angles to zero the Tilt and Rotation m Set Up Left Set Up Left M Use Zero Offsets ML spf 00 Dv TILT 0 00 ROTATION 0 00 Figure 45 The set up dialog and Zero Angles button NOTE if you cannot vertically align the instrument contact Leica Microsystems ML Tilt 13 If the tilt is in the ML direction position the dial indicator spring against the side face of the T bar see Figure 47 Figure 46 Tilt 1 is in the ML direction 2 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 40 Alignment 14 Move the spring to the bottom and then up using the DV drive S LEN t i L Ts Figure 47 Vertical alignment when the tilt is in the ML direction The instrument is aligned in the vertical axis if the dial arm only moves several divisions back and forth but stays roughly in the same position The instrument is not vertically aligned if the dial arm continuously moves in one direction To select a new vertical position i Loosen the tilt clamp see Figure 35 and slightly adjust the manipulator into a new position ii Retighten the clamp and repeat step 14 15 When the ML tilt is aligned select Il To ML on the computer screen see Figure 28 16 Click Setup and click Zero Angles to zero the Tilt and Rotation NOTE if you cannot vertically align the instrument contact Leica Microsystems Leica Angle Two User Manual Rev A01
23. e coordinates are saved in the Rat Coronal folder Figure 32 The Export dialog Saved Coordinates Saved coordinates are displayed across the bottom of the interface Double clicking saved coordinates loads them into the Set Target boxes Right click on saved coordinates to open a menu with the following options see Figure 33 e Load Target loads the coordinates into the Set Target boxes e Edit Target opens Notepad which displays the coordinates which you can edit e Export Target exports the coordinates to an external location on the computer When selected the Export dialog opens Enter in a description of the coordinates and click Save e Delete Target deletes the coordinates from the saved area Click Yes to confirm 210509 Edit Target n Export Target Delete Target Figure 33 Options for saved coordinates Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 32 Alignment Align the instrument whenever the software is started Alignment is necessary to check the horizontal ML and vertical DV planes are perpendicular To align the instrument the rotation and tilt must be set so the scales read zero The alignment kit is installed onto the instrument with the dial indicator positioned where the probe would normally be Then the dial indicator is moved across and up the T bar mounted in the ear bar slots If the dial indicator reading remains unchanged th
24. em e Nanoinjector motorized stepper for stereotaxic use e Impact One stereotaxic impactor for CCI e Spinal holders For details of these and a full line of gas anesthesia equipment probes and species adapters see www myneurolab com Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 23 Software Overview This chapter describes the Angle Two software and its functions The topics covered are e Atlas on page 25 how to select an atlas and view the coronal plates e Modes on page 26 explains what Set Target To Target Atlas and Vertical are used for e Initial Settings on page 29 describes how to set the tilt direction zero the rotation and tilt angles enter offsets and indicate bregma and lambda to the software e Coordinates on page 31 how to save and export coordinates and how to use saved coordinates Figure 23 shows the software interface with the atlas coronal plate in the center and the right and left manipulator controls on either side Clicking anywhere on the left or right side of the screen activates the respective manipulator s controls for use If you have only one manipulator then only one side of the screen will be active Angle Two Atlas Left M od es poa Target C To Target C Atlas Vertical Target Red at Bregma at Lambda Initial Settings C IToML IToAP Interaural 9 20 mm Coordinates Figure
25. en the instrument is aligned and ready for surgical use However if the dial indicator reading changes then the instrument is not aligned and the rotation or tilt must be adjusted If you are unable to align the instrument contact Leica Microsystems 5 1 Checklist Figure 34 shows the alignment kit contents check that you have all five items Alignment o anai Dial C Clamp gt age indicator T bar tte th 7 v Clamp adaptor Figure 34 The alignment kit Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 33 Alignment 5 2 Procedure The primary tilt of the instrument can be in either the ML or AP direction so there are two procedures for the vertical alignment Select the correct instruction set for your instrument set up Set the Rotation and Tilt The instrument s rotation and tilt must be moved so that the scales read zero 1 Unlock the tilt and rotation clamps and move the instrument so the scales read zero as shown in Figure 35 2 Relock the clamps when finished Tilt clamp ulil So Koy 78 a S i Zero tilt Zero rotation Rotation clamp ga e Figure 35 Both the tilt and rotation are at zero Installing the Alignment Kit The alignment kit needs some assembly before being installed 3 Double click the Angle Two icon on the desktop to start the program 4 Remove the ear bars and probe shaft from the instrument Leica Angle Two
26. igns with the main scale beneath therefore the measurement is 1 23 cm The third line of the vernier scale on the U frame aligns with the ear bar scale 10 5 First measurement on the ear bar scale is 1 2 cm Figure 22 Using the vernier scale Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 22 Components 3 6 Computer The computer consists of a desktop case monitor mouse and keyboard The computer comes with the Angle Two software that includes an integrated rat or mouse brain atlas The computer should only be used to operate the Angle Two instrument If your computer is to be networked we recommend installing antivirus or firewall software only If you require additional species atlases contact Leica Microsystems see Contacting Leica Microsystems on page 2 Note if you do not want to have a keyboard connected you can use the virtual keyboard Click start All Programs Accessories Accessibility On Screen Keyboard 3 7 Ancillary Products Leica Microsystems has a wide range of accessories available for use with the Angle Two instrument Some useful accessories include e Video microscope for Angle Two Product 39464678 e Rat Autoclip suture clips Product 39465004 e Manual drill with stop Product 39463018 e Large probe holder Product 39464350 Other instruments that can be used in conjunction with the Angle Two are e Perfusion Two automated pressure perfusion syst
27. in its new position 3 Retighten the swing out clamp Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 19 Components 3 4 Probe Shaft A clamp attaches the probe shaft to the manipulator The clamp has two holes for positioning the clamp onto the manipulator and you can rotate the clamp to the left or right of the manipulator arm to give a total of four possible probe positions in the AP direction Figure 19 Four positions for attaching the probe shaft to the manipulator The Angle Two comes with a standard probe holder Contact Leica Microsystems if you require different probe holders see Contacting Leica Microsystems on page 2 Changing the Horizontal Position of the Probe Shaft 1 Unlock the clamp thumbscrew and remove it from the clamp 2 Select the second hole to reattach the clamp screw OR Rotate the clamp 180 to select from two new positions If you rotate the clamp i Use a 3 32 Allen wrench to undo the probe shaft screw see Figure 20 ii Remove the probe shaft from the clamp iii Rotate the shaft 180 degrees and reposition it back through the clamp and tighten the Screw 3 Screw the clamp back onto the manipulator Changing the Vertical Position of the Probe Shaft 1 Use a 3 32 inch Allen wrench to undo the probe shaft screw see Figure 20 2 Move the probe to the new position and retighten the screw when finished Leica Angle Two User Manual Rev A01
28. l invalidate the warranty Leica Microsystems cannot assume liability for any such damage Persons operating the Angle Two must be adequately trained and warned of any potential hazards or hazardous procedures before operating the instrument Only trained staff are to remove any parts from the instrument and only if instructed within this manual Repairs must only be carried out by qualified service personnel authorized by Leica Microsystems Safety Safety notices in this manual are cautions Cautions are notifications of hazards that could lead to damage to the instrument or the animal undergoing surgery Cautions use symbols with a yellow border as illustrated below Revision Record Rev Issued Sections affected Detail A01 September 2009 First release Contacting Leica Microsystems For enquiries related to the Angle Two stereotaxic system 1 866 444 0944 enone 1 314 522 0300 Email myneurolab leica microsystems com Web site www myneurolab com For other Leica Microsystems products see the web site www leica microsystems com Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 2 Contents Leica Angle TWO sra AAA AER AA Ee eee 1 Intended Use Statement yere bart rra son A a A RA ee a A A ee 2 A e 0 21g ots lt a eee ee eee eae a a A eee ee ee ee 2 COPY E o pd A A a A AAA Ee ee 2 Important Information for All Users Wi E RA AAA AREA A A AAA 2 Sie carrasco enseres 2 REVISION RECO
29. nate sam 7 26 MM boxes for the selected mode at Bregma at Lambda Bregma 0 40 mm __ Setup LE E E Oi ie a TEN E E a C IToML EToAP Rostral Select Atlas Rat Coronal Save Named Point Export Position Caudal 200509 Moo uy 240509 Bev Mos Figure 25 The four modes and coordinate boxes Set Target Select Set Target mode to enter the position of the target site where the probe will go The target site appears as a red crosshair on the coronal plate and its ML AP and DV coordinates are displayed in the Target Red coordinate boxes see Figure 25 Target coordinates can be edited at any time and saved To save target coordinates click Save Named Point see Saving on page 31 There are three ways to enter the coordinates 1 Select the coronal plate showing the target site and click on a position in the map to set it as the target 2 Use saved coordinates double click on saved coordinates located across the bottom of the screen The coordinates are automatically entered and a red crosshair appears see Figure 26 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 26 Software Overview 3 Manually enter the coordinates into the Target Red boxes Left Mode Set Target C To Target Atlas C Vertical Target Red ML 135mm s Ta rg et E 049mm y coordinates EE Target position at Bregma at Lambda Setup IToML BToAP Rostral
30. nates P gt Ee Me Figure 30 Saving coordinates exporting coordinates and saved coordinates Saving The target site probe position or any other coordinates can be saved by clicking Save Named Point see Figure 30 In the Saved Name Point dialog Save Named Point Name Save Description Cancel Choose Source Point e Target Position 2 35 0 40 7 26 C Device Position 1 83 6 19 1 08 Cine fo hbo hbo Figure 31 The Saved Name Point dialog Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 31 Software Overview Enter a name and description 2 In Choose Source Point select which coordinates you want to save Target Position saves the current target site Device Position saves the current position of the probe Input saves a new target position which you must provide coordinates for 3 Click Save Once the coordinates are saved they will appear in the area beneath the atlas with the entered save name Exporting To save the probe s current position to an external location on the computer click Export Position In the Export dialog the coordinates are displayed including the instruments tilt rotation and the deviation of the animal s head from skull flat 0 Enter a description of the coordinates and click Save As By default the coordinates are saved to folders located at c Program Files AngleTwo Targetpoints If using the rat brain atlas th
31. obe to known reference points typically bregma and lambda which the software uses to align the atlas plate with the animal s brain Then simply move the manipulator drives and watch the AP ML and DV displays count down to zero indicating you have reached the target site 1 1 Basic Steps of Use After unpacking and installing the instrument Align the instrument see Alignment on page 33 Select the target site see Selecting the Target Site on page 43 Position the animal see Positioning the Animal on page 43 Locate the zero reference points see Locating Bregma and Lambda on page 44 Move the manipulator over the target site see Locating the Target Site on page 45 and Oe OP ce ee a Drill and insert the probe see Probe Placement on page 46 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 5 Instrument Installation The Leica Angle Two system needs to be assembled before use Use the information below to check you have received all parts and to assemble the system 2 1 Checklist The list below has the standard items supplied in the Angle Two system Additional or alternative items may be included depending on the configuration ordered e Angle Two instrument U frame with 45 and 18 ear bars and species adaptor One or two manipulators with one standard probe holder each e Stereotaxic Alignment Kit dial indicator T bar clamp and Allen wren
32. ocations of the reference points bregma and lambda Zero tilt and at Bregma Bregma button rotation and at Lambda Lambda button set offsets setup Tilt direction 3 gt IToML amp IToAP Figure 28 The initial settings buttons Tilt Direction After aligning the instrument click II To ML or Il To AP to indicate the primary plane of the tilt as either in the ML or AP direction see Tilt on page 17 Zero Rotation and Tilt After aligning the instrument and entering the tilt direction the software s tilt and rotation values must be zeroed Click Setup and Zero Angles in the Set Up dialog check both angles read 0 00 Set Up Left Set Up Left M Use Zero Offsets ML Ap 00 DV TILT 0 00 ROTATION 0 00 Figure 29 The set up dialog and Zero Angles button Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 29 Software Overview Enter Offsets Offsets are used when performing surgery on animals that do not fall within the weight range of the atlas being used Offsets are a distance in millimeters that is added or subtracted from the probe s path to the target site The offsets are determined in pre experiments by locating the target site using the atlas coordinates The amount that the atlas coordinates miss the actual target site is determined and used as the offset values Once the offsets are entered and the bregma and lambda points have been located the offset
33. of the form of action whether in contract tort including negligence strict product liability or any other legal or equitable theory even if such party has been advised of the possibility of such damages In no event will Leica Microsystems aggregate cumulative liability for all claims out of or related to this agreement and software exceed the amounts you paid for software Some jurisdictions do not allow the exclusion or limitation of incidental or consequential damages so this limitation and exclusion may not apply to you Warranty disclaimer and limited liability are fundamental elements of the basis of the bargain between Leica Microsystems and you Leica Microsystems would not be able to provide the software without such limitations Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 50
34. ove the nose clamp by undoing the thumbscrew and insert the nose cone Ear Bars The ear bars are tapered at one end to fit deep inside the ear canal The instrument comes with two rat ear bars 18 and 45 Other animal ear bars and adaptors can be purchased from Leica Microsystems After the bars are positioned in the animal s ears the animal is centered in the middle of the U frame using the scales on each bar Use this measurement to reproduce the exact position of the animal for repeat studies Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 11 Components The ear bars are locked into position using the locking levers The right lever has a right hand thread and the left lever a left hand thread The levers only need a quarter turn to lock the ear bars Figure 7 Turn the lever to lock the ear bar into place Caution AN Use 45 ear bars if post surgical testing is planned Do not use 18 ear bars as they can easily break the tympanic membrane of an animal Caution A Do not use the 45 or 18 ear bars on a mouse The mouse s breathing passages run between the ears and as the skull is very thin the mouse can easily suffocate if too much pressure is applied by the ear bars Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 12 Components 3 2 Manipulator Linear Movements The manipulator is attached to the U frame and is used to move the probe The m
35. rms Upon termination you must immediately destroy the Software together with all copies adaptations and merged portions in any form Export Requirements You may not export or re export the Software or any copy or adaptation in violation of any applicable laws or regulations Any person to whom this software is transferred should be aware that use of the software is subject to any Leica Microsystems software license terms distributed with the software Use of the software indicates acceptance of these terms If the user does not accept these terms the software should be returned Disclaimer of Warranty Leica Microsystems and its licensors disclaim all warranties express implied and statutory including without limitation the implied warranties of merchantability fitness for a particular purpose title and non infringement of third party rights No oral or written information or advice given by Leica Microsystems its dealers distributors agents or employees shall in any way increase the scope of this warranty Some jurisdictions do not allow the limitation or exclusion of implied warranties or how long an implied warranty may last so the above limitations may not apply to you Limitation of Warranty Leica Microsystems shall not be liable for any incidental special consequential or indirect damages of any kind including damages for interruption of business procurement of substituted goods loss of profits or the like regardless
36. th is straight To test this the dial indicator is moved vertically up and down the T bar As the primary plane of the tilt can be either in the AP or ML direction select the appropriate instruction set below For AP tilt tilt is along the length of the instrument see AP Tilt on page 39 For ML tilt tilt is along the width of the instrument see ML Tilt on page 40 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 38 Alignment AP Tilt 13 If the tilt is in the AP direction use the ML drive to position the dial indicator spring in the middle of the T bar Do not move the AP drive Figure 43 Tilt 1 is in the AP direction 2 14 Move the spring to the bottom of the bar and then all the way up using the DV drive Figure 44 Move the dial up the T bar to check the vertical alignment The instrument is vertically aligned if the dial arm only moves several divisions back and forth but stays roughly in the same position The instrument is not vertically aligned if the dial arm continuously moves in one direction To select a new vertical position i Loosen the tilt clamp see Figure 35 and carefully move the manipulator into a new position ii Retighten the clamp and repeat step 14 Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 39 Alignment 15 When the AP tilt is aligned select II To AP on the computer screen see Figure 28 16 Cli
37. the AP direction see Figure 15 3 Retighten the clamps Caution A Do not position the tilt encoder towards the instrument U frame or the closed end of the U as this reverses the sign of the tilt angle Rotation Rotation moves the tilt plane as shown in Figure 14 Unlock the rotation clamp to horizontally rotate the manipulator Scales on the base of the manipulator above the rotation clamp indicate the amount of rotation The rotation encoder updates the software with the instrument s position Rotation clamp Rotation encoder Caution AN Always rezero the instrument if the rotation or tilt are changed by returning the probe to bregma and clicking at Bregma Small changes to the rotation and tilt cause large changes to the positioning of the probe Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 18 Components Swing Out Use swing out to change the primary tilt plane by 90 between the ML and AP directions see Figure 13 or to remove the probe from the area within the U frame and return it later to the same position To change the primary tilt plane see Tilt on page 17 To swing out the manipulator 1 Unlock the swing out clamp and push the manipulator up with your thumb just enough to clear the raised edge on the base of the manipulator Swing out clamp Figure 18 Raising the manipulator using swing out 2 Swivel the manipulator around 90 then lower it back down
38. tions of the manual where the part is discussed in more detail Nose Clamp Incisor Bar Ear Bars U Frame Figure 4 The Angle Two U frame The animal is held in the instrument U frame with the head at the closed end of the U and the tail at the open end The ear bars nose clamp and incisor bar attached to the U frame secure the animal during surgery Adapters for a range of species are available see www myneurolab com Incisor Bar The animal s incisor teeth are placed over the incisor bar The bar can be adjusted horizontally and vertically using the thumbscrews indicated below F SS Vertical control J N ya y y Incisor Bar e Horizontal control Figure 5 The incisor bar can be moved vertically and horizontally Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 10 Components Caution Do not use the rat incisor bar on a mouse A mouse adaptor is available from Leica Microsystems Note Hamsters and gerbils can use the rat ear bars and incisor bar Nose Clamp The nose clamp is placed over the animal s nose and gently lowered using the thumbscrew indicated The nose clamp helps to keep the animal in position when injectable anesthesia is used if ly Vertical control SN iy a Nose clamp gt Removes the nose clamp Wero REA E E dr ES Figure 6 The nose clamp If using gas anesthesia with a nose cone rem
39. ure the ear bars have the same reading Installing the Incisor Bar and Nose Clamp 15 Slide the incisor bar into the animal s mouth and hook the incisor teeth over the bar 16 Push the animal s nose down and slide the bar away from the animal so that the bar is directly behind the incisor teeth 17 Tighten the horizontal thumbscrew see Figure 5 so the bar cannot move 18 If using an injectable anesthetic lower the nose clamp so it is secure against the animal s nose If using gas anesthetic remove the nose clamp and install the nose cone Locating Bregma and Lambda After the skull is exposed locate the bregma and lambda points and mark them We recommend using the video microscope or a stereomicroscope to accurately determine each point and reduce error The bregma and lambda points are used by the software to calculate the distance the manipulator must move along each axis to reach the target site As well the bregma and lambda points provide the variance angle of the animal s skull from flat 0 which is corrected for by the software so no manual adjustment of the instrument is needed NOTE Always position the probe at the bregma point first To identify the bregma and lambda points to the software 19 Place the probe on the marked bregma point and click at Bregma The Atlas mode will open showing the probe s coordinates 20 Reposition the probe tip at lambda and click at Lambda The Skull box displays the skull
40. values are automatically factored into the To Target coordinates and the display is updated to show the probe approaching the correct target site To enter offsets click Use Zero Offsets in the Set Up dialog and enter the values Click Close when finished Bregma and Lambda Typically the bregma and lambda points of the skull are used as zero reference points to determine the location of the target site Once entered into the software they are used in the skull flat feature which automatically incorporates any tilt in the animal s skull into the calculation of the to target values To set the bregma point carefully place the probe tip on the bregma point of the animal s skull and click at Bregma To set the lambda point gently lower the probe tip on the lambda point of the animal s skull and click at Lambda Always locate the bregma point first Leica Angle Two User Manual Rev A01 Leica Biosystems St Louis LLC 2009 30 Software Overview 4 4 Coordinates Coordinates can be saved or exported at any time while the instrument is being used Double click on saved coordinates to load the target coordinates from a previous study rather than relocating the target position Figure 17 4 r A 4 Saving j coordinates 2 8 Exporting coordinates Interaural 9 20 mm Bregma 0 20 mm 7 6 5 4 3 2 1 0 1 2 3 4 6 7 Saved Rostral Select Atlas Rat Coronal Save Named Point Export Position Caudal Coordi