Home
LCR-915/916 Series User Manual
Contents
1. Press the D Q ESR O button to select the measurement function 10 GW INSTEK LCR 915 916 Series User Manual Selecting the Test Frequency Ne 1kHz 40kHz ees nnm 100kHz SORTING ENTER Press 09889 wa 8 SERPAL 20000000 ZERO Hor ber Gub HCUR LCUR UD Press the FREQ button to select the test frequency The 100kHz test frequency is only for the LCR 916 11 GW INSTEK LCR 915 916 Series User Manual Selecting the Series Parallel Measurement Function b Auto Nord A i FERRERI ER Series SLL Parallel Press HcUR ZA DISCHARGE CAPACITOR BEFORE TESTING At the L C R measuring function it defaults to Auto Series Parallel measuring function Press the SER PAL button to select the measuring function 12 GW INSTEK LCR 915 916 Series User Manual Selecting the Display Count Press IL 106 R oe R UMANE 2000 20000 Press the 2000 20000 button to select the display count Zero ZERO Press ji m us ANN OMM MAII nz rn ZERO The Zero mode records the current input value as a reference and displays the reference
2. LCR 915 916 Series User Manual Meter Description Front Panel Illustration 1 LCD display 20000 2000 counts 2 Function buttons 3 5 Wire input terminal for SMD test probe or DIP part 4 2 Wire input terminal for Alligator Clips HOLD Auto LCR Range 1 lut y Sonat Ho Display ml ZEE 2 Buttons SEP 9 5 Wire Terminal 2 Wire Terminal Z N DISCHARGE CAPACITOR BEFORE TESTING v N GW INSTEK LCR 915 916 Series User Manual Package Contents 1 5V AC Adapter only LCR 916 2 USB Cable only LCR 916 3 Shorting Cube 4 SMD Test Probe only LCR 916 5 Alligator Clip Set 6 Hanging Kit GW INSTEK LCR 915 916 Series User Manual Measuring Principles noc L z E C ie E R j X X 2 si X X JR X amp tan m Z X 2afL oL X 2mfC aC X X oin z R 1 tan Q D an Series Measurement Rs jXs Z R 4JjX Parallel Measurement Rp GW INSTEK LCR 915 916 Series User Manual Phase Drawing XL XL Y XL XC gt 0 7 XL XC 0 XL XC 0 GW INSTEK LCR 915 916 Series User Manual Making 5 Wire Measurements with the SMD Test Probe VOLDA IDA Rana A gl a Hi ja zs M a eos pmr UCIRI
3. 0 1 0 2d 0 2 2d 2 0 2d 0 5 20d 5 0 2 2d 1 The 100kHz test frequency is only for the LCR 916 2 The measuring time is 2 seconds Input Protection 30 30V rms Minimum Resolution 0 001pF in the 20 000pF range Measuring Time 800ms Note If D gt 0 1 the accuracy should be multiplied by 414 D 24 GWINSTEK 5 916 Series User Manual 6 Resistance Frequency 200 000 2 5d i 0 0 20 000 0 596 2d 200 00 1 0 2d 20 0000 0 5 5d 200 000 2 0000kQ 20 000kQ 200 00kQ 0 2 2d 2 0000MQ 20 000MQ 2 0 2d 5 0 2 2 0000MQ 2 0 2d 20 000 5 0 2d 20 0000 0 5 5d 200 000 1 2 0000kQ 2d ura 20 000kO 0 2 200 00kQ 2 0 2d 2 0000MO 5 096 2d 1 The 100kHz test frequency is only for the LCR 916 2 The measuring time is 2 seconds 200 00MQ 5 0 2d 20 0000 0 5 5d 200 000 2 0000kQ 20 000kQ i 200 00kQ hz Input Protection 30V or rms Minimum Resolution 0 0010 in the 20 0000 range Measuring Time 800ms Note If Q gt 0 1 the accuracy should be multiplied by 41 4 Q 25 GWINSTEK 5 916 Series User Manual 7 DCR Range Resolution 00 000 0 2 5d 0 0000kQ 0 000kQ 00 00kQ 0 2 2d 0000MQ 0 0 000 0 596 2d 00 00MQ 1 0 2d 1
4. 20 to 60 C 0 to 80 batteries not fitted Temperature Coefficient 0 15 x Spec Accy lt 18 or 28 Measurement Samples 1 25 times per second normal Altitude 6561 7 ft 2000m Weight 6309 including battery Dimensions W x H x D 95mm x 207mm x 52mm with holster Accessories 6 SMD Test Probe Standard Optional 8 USB Cable Standard Optional 9 PC Software CD Standard Optional Power Requirements 1 5V x 4 IEC LR6 or AA size External Power Requirements DC 5V USB cable or AC adapter Pollution Degree 2 Safety Complies with EN 61010 1 IEC 61010 1 EMC EN 61326 1 Shock Vibration Sinusoidal vibration per MIL T 28800E 5 55 Hz 3g maximum Drop Protection 4 feet drop to hardwood on concrete floor Indoor Use 21 GW INSTEK LCR 915 916 Series User Manual Electrical Specifications 1 Test Frequency Range 100 00 Hz 0 01 Hz 0 01 120 00 Hz 0 01 Hz 0 01 10 000 kHz 0 01 100 00 kHz 0 01 1 The 100kHz test frequency is only for the LCR 916 2 Test Signal AC Signal Level 600mVrms AC Signal Accuracy 10 DC Bias Level 1V DC Bias Accuracy 10 1 0000 kHz 0 01 3 Test Cable Model Length Bandwidth SMD Test Probe 4 Wire Test Probe Alligator Clip Set Accuracy x of reading A Basic Accuracy as specified by B Test Cable Accuracy 1 LXF x T L m Cable length F MHz T
5. DCR AUTO TEST GWINSTEK 5 916 Series User Manual Making 4 Wire Measurements with the 5 Wire Terminal HPOT LROT GUARD HcUR LouR GUARD 1 DISCHARGE CAPACITOR BEFORE TESTING Making 2 Wire Measurements with the Alligator Clip Set GUARD LouR _ GUARD DUT GW INSTEK LCR 915 916 Series User Manual Measuring L C R DCR o 857 Auto LCR HOLD Auto LCR Range 11 N 11 cmm B Z DCR T pmF EMI somone Press T Press gt 2sec amp N ZA i SORTING FREQ mxm ES 2000 20000 ZERO HOT LPOT GUARD I GUARD Press the L C R DCR button to select the measurement function Press the L C R DCR button for 2 seconds to enter the Auto L C R function GW INSTEK LCR 915 916 Series User Manual Measuring D Q ESR 0 HOLD Auto LCR Range T Z DCR iorting 12057648096 SRM 5 em EPIRI SORTING JQJESR G FREQ MAXMIN SERAL 2000 2000 ZERO
6. Hand held LCR Meter LCR 915 916 Series USER MANUAL GW INSTEK PART NO 82CR 91600Mo1 Test Equipment 99 Washington Street Depot Melrose MA 02176 Phone 781 665 1400 1 800 517 8431 Toll Free 1 800 517 8431 EMM ETEK This manual contains proprietary information which is protected by copyright All rights are reserved No part of this manual may be photocopied reproduced or translated to another language without prior written consent of Good Will company The information in this manual was correct at the time of printing However Good Will continues to improve products and reserves the right to change specification equipment and maintenance procedures at any time without notice Good Will Instrument Co Ltd No 7 1 Jhongsing Rd Tucheng Dist New Taipei City 236 Taiwan GW INSTEK LCR 915 916 Series User Manual Read First Safety Information Understand and follow the operating instructions carefully Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired WARNING Identifies hazardous conditions and actions that could cause BODILY HARM or DEATH CAUTION Identifies conditions and actions that could DAMAGE the meter or equipment under test WARNING e When using test leads or probes keep your fingers behind the finger guards e Remove test leads from the Meter before opening the battery door or the Meter case Use t
7. est frequency T Cable type If the cable is 5 wire type then T is 40 else T is 4 000 When measuring by basic accuracy the following conditions must be met 1 Ambient temperature 23 5 C 80 RH 2 Test cable length 3 Open and short calibration has been performed 4 Ds lt 0 1 for C L measurements Q lt 0 1 for R measurements See the operation manual for additional details 22 GWINSTEK 5 916 Series User Manual 4 Inductance Frequency 20 000mH 0 5 5d 200 00mH 100Hz a0 oo 0 2 2d 120Hz 200 00H 2000 0H 0 5 2d 20 000 1 0 2d 2000 0uH 0 5 5d 20 000mH 200 00mH 1 2 2000 0mH 20 000H 200 00H 0 5 2d 2000 0H 2d 1 0 200 00uH 0 5 5d 2000 0uH 20 000mH 0 2 2d 200 00mH 2 0 2000 0mH 20 000H 5 0 2d 20 000uH 0 5 59 200 00uH 100kHz 2000 0uH 10 2 20 20 000mH 2 096 2d 200 00mH 5 0 2d 1 The 100kHz test frequency is only for the LCR 916 2 The measuring time is 2 seconds Input Protection 30 30V rms Minimum Resolution 0 001uH in the 20 000UH range Measuring Time 800ms Note If D gt 0 1 the accuracy should be multiplied by 1 D 23 GWINSTEK 5 916 Series User Manual 5 Capacitance Frequency 2000 0pF 0 5 5d 20 000nF 200 00nF Ji 5000 OnE 0 2 2d 20 0000 1 0 2d 0 2 2
8. he ENTER button to save the setup value and enter the next step 3 Press A V 1 and gt button to setup the standard value Then press the ENTER button to save the setup value and enter the next step 4 Press lt and gt button to setup the tolerance value Then press the ENTER button to save the setup value and exit this mode 19 GW INSTEK LCR 915 916 Series User Manual Battery Replacement Refer to the following figure to replace the batteries o t wu S sane Caution Replace the batteries as soon as the low battery indicator appears to avoid false reading 1 5V x 4 alkaline batteries External Power Source Save battery power by using the external power source _ Jr 3 Caution Use the 5V AC adapter only as specified in this manual Do not use other power sources with the meter 20 GW INSTEK LCR 915 916 Series User Manual Specifications General Specifications Maximum voltage applied to any terminal 30V oc or 30V rms Display 2000 20000 counts Overrange Indication OL Battery Life 80 hours Low Battery Indicator is displayed when the battery voltage drops below the operating voltage Low battery voltage Approx 4 5V Auto Power Off 10 minutes Operating Temperature No condensation x 10 C 11 30 x 8096 RH 30 40 x 7596 RH 40 C 50 C x 4596RH Storage Temperature
9. he Meter only as specified in this manual or the protection by the Meter might be impaired e Always use the correct terminals switch positions and ranges for measurements Do not apply more than the rated voltage as marked Meter between terminals or between any terminal and earth ground Use caution with voltages above 30 Vac rms 42 Vac peak or 60 Vdc These voltages pose a shock hazard To avoid false readings that can lead to electric shock and injury replace the battery when the low battery indicator appears e Discharge all high voltage capacitors before testing Do not use the Meter around explosive gas or vapor reduce the risk of fire or electric shock do not expose this product to rain or moisture GW INSTEK LCR 915 916 Series User Manual CAUTION e Never connect a source of voltage that could result in damage to the meter and the equipment under test Do not expose the Meter to extremes in temperature or high humidity Symbols as Marked on the Meter and the Instruction Manual Ee C contos to EU Conforms to EU directives to EU directives Do not discard this product or throw away Maintenance Do not attempt to repair this Meter It contains no user serviceable parts Repair or servicing should only be performed by qualified personnel Cleaning Periodically wipe the case with a dry cloth and detergent Do not use abrasives or solvents GW INSTEK
10. lt 50991 rolling Input Protection 30 or rms Minimum Resolution 0 010 in the 200 000 range Measuring Time 2 seconds Range 8 D amp Q Definition Q 1 D Range 2 000 2000 Minimum Resolution 0 001 in the 2 000 range Accuracy Accuracy of Main Reading x 1 D Input Protection 30 or rms 9 ESR The specification for ESR is the same as Resistance 10 6 90 0 90 0 0 2 59 Input Protection 30 or 30V rms
11. on the sub display Any inputs after this will be subtracted from the reference value and displayed on the main display To use the Zero mode follow the steps below 1 Press the Zero button to enter Zero mode The A symbol appears on the display 2 Press the Zero button again to record a new input value as a reference 3 Hold the Zero button for 2 seconds to exit this mode 13 GW INSTEK LCR 915 916 Series User Manual Display Hold HOLD Press mn HOLD AU gt irre 400 m gung ad ME MM abe HOLD RON Press the HOLD button to hold the reading on the meter press the button again to return Display MAX MIN MAX MIN Press ae Dee m ana 40 m png M Press LE E It abel gt 2Sec ES abel Current MAXIMIN Reading The MAX MIN mode records the maximum and the minimum input values When the inputs go below the recorded minimum value or above the recorded maximum value the meter beeps and records the new value To use the MAX MIN mode follow the steps below 1 Press the MAX MIN button to enter the MAX MIN mode The MAX icon appears on the display The maximum value is shown on the main display and the current value is shown on the sub display 2 Press the MAX MIN button to select the MAX or MIN mode 3 Press the MAX MIN button for 2 seconds to exit this mode Note This function is only for LCR 916 14 GW INSTEK LCR 915 916 Series U
12. ser Manual Calibratation 9 HOLD Auto LOR Range I CAL S amp P or 4 e DUI Mr Bess E 48 Q 2 ai AUTO TEST Press s wm 0 SERIPAL 200020000 ZERO HPOT GUARD HCUR LCUR 2 2 In order to achieve the best measurement results calibration is essential To calibrate the meter press the CAL button 15 GW INSTEK LCR 915 916 Series User Manual Press gt Cal Open LPOT GUARD LcuR _ GUARD j ZINDISCHARGE CAPACITOR BEFORE TESTING fi Open When OPEn appears on the sub display make the terminal or the SMD test probe open and press the CAL button to start the open calibration About 30 seconds later the result of the open calibration appears on the main display If the result is pass press the CAL button to proceed to the next step If the result is fail press the CAL button to exit the function 16 GW INSTEK LCR 915 916 Series User Manual Press Cal e Shorting Cube EN ___1 When Srt appears on the sub display short the terminals or the SMD test probe and then press
13. the CAL button to start the short calibration About 30 seconds later the result of the short calibration appears on the main display If the result is pass press the CAL button to complete the calibration If the result is fail press the CAL button to exit the function 17 GW INSTEK LCR 915 916 Series User Manual Sorting Sorting Tolerance 40 1 10 2 0 25 40 5 51 0 12 096 55 0 10 0 20 0 80 20 Press HCUR GUARD j To check the accuracy of a part press the SORTING button to enter the sorting mode The sorting result appears on the main display and the current value appears on the sub display 18 GW INSTEK LCR 915 916 Series User Manual SQ Pes o gu SORTING The default sorting standard value is the current value and the default tolerance is 1 0 Setup Sorting Standard v Range P Range eu uc m MINI cant Te wann Sorting Tolt 1 uF SETUP Press ENTER hue Sorting Press Sorin Tolt 1 Tolerance Position To setup the sorting standard value follow the steps below 1 Press the SETUP button to enter the setup mode 2 Press and gt button to setup the range of the standard value Then press t
Download Pdf Manuals
Related Search
Related Contents
Washing Machine - Trail Appliances Nodor IPS 37 1 Índice - Lidl Service Website Samsung S1050 Brugervejledning bachdd-1 14,4v bc bachdd-1 18v bc bachdd-1 18v Kenwood DPC-41 Portable CD Player User Manual Word Pro - MTLCNVITA1.lwp Sommaire Massive Floor lamp 42028/17/10 User- and installation manual Cupido 50, Cupido 70, Optica, Bioptica Copyright © All rights reserved.