SPX MGD-2002 Multigas Leak Locator User Manual
Contents
1. gt lt INSTRUCTION MANUAL Model MGD 2002 Multi Gas Leak Detector MODEL MGD 2002 LEAK LOCATOR IB 417 REV G PART NO 84671 Ye gt s ielectric User Safety Warnings IMPORTANT Before operating the device for the first time read all of the following safety guidelines To prevent personal injury property damage or damage to the detector operate it only in accordance with these safety guidelines e Do not operate the device if it has been damaged Contact SPX Radiodetection for advice e Do not allow any part of this device to come in contact with an ener gized high voltage source Injury or death may result e Do not disassemble the device as it will void your warranty Repairs should only be performed by qualified factory service personnel e Do not allow the Detector to draw up water Water will adversely affect the device s operation and will cause internal damage Water found inside the unit will void your warranty e Take sensible precautions to prevent the ingress of moisture when using the detector in rain snow or other adverse conditions Although the detector is somewhat water resistant it is not waterproof Do not allow water to accumulate on the detector faceplate as it could mi grate into the electronics e Do not use the probe tip to dig in the soil or use it in any manner other than as directed in this manual e Although this device will operate with a low battery voltage do n
2. e Frozen ground may cause the tracer gas to migrate a consider able distance from the actual leak e It may be necessary to make holes at curbings or road expan sion joints to help the tracer gas to rise to the surface e When tracer gas is detected continue to sample the adjacent area until you determine the highest concentration of the tracer gas Boring holes with a t bar drill may be helpful at this point The location of highest tracer gas concentration will be directly over the system fault e Being able to reasonably repeat a low reading 100 to 1000ppm is significant Readings above this are generally significant and easily repeatable e Spot reading below 100ppm if not repeatable may be due to equipment sensitivity to certain movements or other factors 38 13 0 Troubleshooting The pump seems to be straining and no air is exiting from the out let port brass on the right side of the control unit e Check that the handle section and the ground probe or needle probe is properly installed e Check to see if the humidity filter is clogged e Check to see if the filter at the boot end of the ground probe is clogged e Contact customer service This is an indication of a clogged internal filter The detector continuously counts up without helium hydrogen present e Check to see that the handle section and the ground probe or needle probe is properly installed and not clogged e Ifthe unit was recently
3. 4 0 Tracer Gas 4 1 Helium vs Hydrogen as a tracer gas The MGD 2002 is a Helium Hydrogen detector calibrated to detect Helium However it will indeed give readings if hydrogen is present As this detector is based on the thermal conductivity of gasses and with Helium and Hydrogen being similar in this regard the detector will not be able to tell the difference Helium is the preferred tracer gas Helium is a noble gas and does not react or combine with any other gas or substance It is safe to use in all types of leak locating at any concentration up to 100 Hydrogen is not generally present in higher quantities in uncontrolled situations as it is very reactive and if in quantities greater then 4 it is very flammable and even explosive To use it as a tracer gas you need to purchase it mixed to 4 or less with the remainder being nitrogen Thus starting with an already diluted tracer gas making it that much harder to find as it dilutes even further when you add it to the system being tested not a problem when working with 100 helium We certainly DO NOT recommend Hydrogen be used as a tracer gas for these reasons Note It would take more then 20 tanks of 4 hydrogen mix to equal the amount of tracable gas in one 100 helium tank Mixing the helium down to 4 is certainly an option and can be done by adding compressed air to the system as it is being filled 4 2 Other Gases the MGD 2002 will react too When considering all ot
4. continually used in high humidity conditions or if water is present To maintain optimum filter performance we recommend that you change it or clear it of moisture when the color changes to light pink The useful life of the filter depends upon the relative humidity RH of the working environment For example on an average dry day the RH is roughly 20 and in this case the filter should last between 4 6 hours If the relative humidity increases to 50 RH the single cartridge might last only 2 3 hours If the humidity increases to between 50 90 RH the filter may only last one hour With the MGD 2002 field replaceable system filters can be changed within seconds or cleared and reused within minutes Refer to page 34 for instructions on how to change the filter 6 we Spent Cartridge Dry Cartridge The internal filtration material has a limited useful life and SPX Radiodetection strongly recommends that the MGD is returned for calibration once a year During factory recalibration the moisture cartridges are cleared or replaced and other filters have their filtra tion media changed Refer to page 35 for a list of all the value added benefits that are included in a factory recalibration 6 4 Power supplies The MGD 2002 is powered by a replaceable rechargeable nickel metal hydride NiMH battery and comes supplied with a universal AC DC adapter AC charging lead and 12 volt DC charging lead for use with a vehicle cigaret
5. 33 7cm x 12 3cm x 8 3cm Weight MGD 2002 only 3 1 Ibs 1400 grams Shipping 15 0 Ibs 6800 grams Temperature Range 40 F to 80 F 4 4 C to 26 6 C Humidity Range 20 RH to 50 RH Sensitivity Min 25 PPM Max 100 1 000 000 PPM Increment 25 PPM 0 to 20 000 PPM 1 2 to 100 Tracer gasses Helium He Hydrogen H Resolution Low range 25 PPM High Range 0 2 Response Time 15 Seconds approx Audio Out Pulse Width Modulation LCD 128 Bit X 64 Bit Dot Matrix With long life backlight Battery Voltage 7 2 VDC Type Nickel Metal Hydride Weight 383 0 grams Run Time 6 8 Hours Continuous Charging Cycles 300 500 Charging Volts 9 5 14 VDC 2A Power Supply AC Input 110 240V 1 6A 60 50Hz Charger DC Output 12V 4 0A Pump Volume 1 05 Liters Min approx 0 85 to 1 00 L min 75 to 85 L min 60 to 70 L min Compliance CE Description 2 0 List of kit Contents MGD 2002 gas detector 12VDC Auto Adaptor charger Ground probe with Collection cup Spare moisture filter cartridge Sampling handle Needle probe Battery NiMH Sholder strap OMONOAaABRWND gt Backflush connector 10 110 240 VAC universal AC charger 11 Carry case Optional equipment not included Headphones European power cord Serial port Part No 79766 163 0005 01 90664 882 0007 241 0003 00 requires factory installation 3 0 Unpacking your New
6. Mode button controls what is being displayed on the LCD By default the unit is set to automatic mode A and will constantly show what the sensing elements have detected in both concentration as well as time If the unit is placed in manual M mode the device will continue to draw in samples as normal However when the sample button is pressed the LCD no longer shows time but will only show what the sensors are detecting at that particular moment Only available in manual mode When the sample button is pressed one sample read ing is taken The device calculates the time it should take for the sample to reach the detec tor takes the sample and then displays the result The sample button must be pressed for every sample needed 11 6 10 Liquid Crystal Display LCD The LCD is a near real time display of the tracer gas concentration the unit has detected The refresh rate is slightly greater than ten times every second After the unit is switched on This screen will remain displayed while the processor conducts a power on self test which lasts for approximately 60 seconds The LCD displays the aproach ing zero symbol upon successful completion of the power on self test or when the unit has just been reset This display is show ing that the unit is setting a zero point of reference The display upon successful completion of the initial zeroing routine This screen will be displayed until the sensor elements
7. and the amount of time the equipment is used throughout the year If you frequently use this device or extremely high accuracy is required the unit should be calibrated a minimum of once per year If however the device is used infrequently or the accuracy is not as critical factory recalibration can be delayed to as much as every two years Sending the unit back for calibration is not limited to the annual or biannual calibration cycle chosen For example the user may suspect inaccurate operation that may be the result of the filters or plumbing being blocked with oil residue oil vapors water dirt petroleum vapors grass and insects or some other foreign material During the factory re calibration the following value added steps and procedures will be accomplished 1 Inspection 2 Light cleaning as required 3 Moisture cartridge filters replaced 2 4 All internal filter media replaced 5 All internal tubing is replaced 6 Air volume checked 7 Software updates installed as available 8 Calibration procedure completed 35 12 0 Frequently asked questions FAQ s Q Why do sudden movements cause the detector to react A Sudden movements such as rapidly swing the unit from a hori zontal position to a vertical position can make the detector tem porally display between 25 amp 50ppm This is caused by air move ment disruptions inside the control unit and is normal Q Directly following charging the detec
8. charged allow it approx 15 min of cool off time e Allow the unit to run for approx 10 min and then press Reset to zero the detector e Contact customer service This is an indication of a clogged internal filter or an internal leak Erratic readings when first turned on no tracer gas is present The readings just bounce all over the place from 0 to several hundred ppm and back in rapid sucession e Contact customer service the sensor is fouled with condensation or is damaged Requires factory service to correct Working on a busy street with heavy traffic the detector is giving false reading e Place the boot end of the ground probe up in the air as far away from Helium Hydrogen Auto emissions and press the re set button This may be an indication that the catalyst filter built into the base of the handle is no longer functioning properly Contact customer service 39 14 0 Service Information Should you need to contact us please call our Customer Service Department on 207 655 8525 or toll free at 877 247 3797 When returning a unit for factory service or calibration call the customer service department for a service return authorization number SRA The device should be boxed securely and contain contact information contact telephone number billing information and return shipping information If the device is being sent to the factory for service a written statement of the problem or symptoms should be inclu
9. gas Helium or Hydrogen mix The gas will escape from the system at the leak and be easily detectable at the surface of the ground 8 3 Locate the system Locate and mark the cable path in the projected search area not ing any recent construction or ground shifting in the area that may have caused damage to the cable system There are a large va riety of pipe cable locators available from Radiodetection Please contact Radiodetection sales or customer service to purchase a locator for your application A Caution Never pressurize any system above its working pressure 8 4 Charging the cable with tracer gas Buried cables are charged with helium hydrogen at the valve points closest to the leak on the air source side of the leak 21 Charging cables We suggest that the cables be charged from the central office unless a section leak has been identified by graphing the leak Attach a bottle to the manifold or the pipe distribution panel and meter the gas into the cable as directed for a minimum of several hours or overnight when possible This will allow the gas to completely saturate the cables and rise to the surface where it can be readily detected The tracer gas can then be detected with the MGD 2002 without the need to bore test holes Note Tracer gas will diffuse to the surface through loose soil faster than through compacted or frozen soil The highest concentration of helium hydrogen will be along the cable path wi
10. of the cable section from the tracer gas feed point to the bleed valve If analysis cannot be accomplished easily due to the lack of recent cable pressure monitoring data consider DL and TL to be the same length usually the end of the cable run Asecond method for determining cable filling times uses the chart page 32 This method does not account for the gauge and resistance a Determine the amount of flow in the cables you wish to charge by using one of the following methods i Observe the panel mounted flow rater in the central of fice ii Manually read the flow rate into a cable with a portable flow rater at the central office distribution panel b Determine the outside diameter of the cable from the cable charts or by manually measuring the cable diameter c Find the cable diameter on the left hand column of the table If the exact diameter is not shown use the next larger one d Find the flow time per thousand feet of cable by looking down the Flow rate column and across the cable diameter row If the flow rate into the cable is greater than 10 standard cubic feet 24 per hour SCFH divide the number found in the column for one SCFH by the flow rate The result is the number of hours per thousand feet for that flow rate and cable diameter Multiply the time found by the length of cable you want to charge This is done in number of thousands of feet The result is the total amount of time required for the hel
11. same way as for direct buried cables Note To determine the number of hours of tracer gas available from a tank divide the number of standard cubic feet SCF remaining in the tank by the flow rate in SCFH For example If the tank regulator shows 150 SCF and you are using a flow rate of 5SCFH the number of hours of flow would be 150 50 30 hours Tracer gas will travel to the end of a normal pipe run approximately 30 000 feet in five or six hours Additional time should be allowed for the helium hydrogen air mixture to permeate the pipe system beyond the manifold points We recommend that the charging is started at least 18 hours before samples are taken 27 If you want to survey a complete system charge the pipe from the central office manifold This allows the tracer gas to completely permeate the cables between the central office and the first mani fold location 8 10 Underground system If you are surveying a complete pipe system begin your sampling at the central office vault It is advisable to check the central office vault plugs and other plant before proceeding with the outside survey At the manholes set up work area protection according to local practice Without removing the manhole cover do the following a Switch on the MGD 2002 and following the startup cycle insert the extension tip of the probe through one of the holes in the manhole cover b Take a sample and record the result c Take a
12. sense the presence of a tracer gas The real time graph is updated once per second and shows 60 seconds of history 12 6 11 Display status icons There are five small icons shown at the bottom of the LCD screen The icon as well as its function is displayed below The icons are listed in the order in which they are displayed on the LCD from left to right NOTE A function that has been turned off will have an X through its icon ICON qd aK IS one BATTERY STATUS This icon gives a visual indicator of the relative life span remaining on the battery When the battery level drops below 1 4 a warning beep will sound The unit will automatically shut off when the empty battery icon is shown RECHARGE STATUS This icon is either displayed or crossed out If the unit is recharging the icon will change to a lightning bolt and back to a plug when charging is complete MODE Shows whether automatic A or manual M mode is selected The default is automatic SOUND STATUS By default the speaker is disabled to conserve battery power and the speaker symbol will initially have an X through it Pressing the speaker button once removes the X and enables the audio out circuitry PUMP SPEED Pump speed can be adjusted for use in certain sampling situation Pressing repeatedly will slow the pump until it returns to normal speed 13 7 0 Tracer gas leak locating for piping systems 7 1 Assumptions e You ha
13. MGD 2002 3 1 Upon delivery 1 Open the shipping package and inspect the MGD 2002 for any physical damage that may have been caused during shipping Keep the packing material in case you ever need to return the device to SPX Radiodetection NOTE The Manufacturer s warranty does not cover damage caused in transit You must notify your carrier immediately for any damage claims 2 Position the device face down to open the battery compart ment door Remove the battery to expose the device informa tion placard The serial number is located under the bar code 3 Connect the battery plug to the main logic board Close the battery compartment door and secure with the quarter turn fastener NOTE Connecting the battery incorrectly even for a second will destroy the circuit board 4 First plug the universal power supply into a convenient AC wall outlet then plug the other end into the device NOTE To ensure optimum battery working life charge battery for 2 hours before operating the MGD 2002 detector for the first time See page 7 for battery charging instructions 5 Check the moisture cartridge filter and ensure it is blue in color Refer to page 34 for filter changing instructions 6 Before switching the device on connect either the needle probe or ground probe to the handle assembly and then con nect the handle assembly to the detector using the quick con nect fitting on the end of the black tubing Refer to page 5
14. OL 1 0 ON OFF BACKLIGHT SOUND BUTTON DESCRIPTION The ON OFF button supplies power to the device Once switched on the device will initialize then conduct a 60 second power on self test POST The light button is used to switch the LCD backlight on and off By default the LCD backlight is turned off to conserve battery life When the button is pressed the back light will illuminate and will remain on until the button is pressed again or the unit is turned off The sound button is used to switch the audio output of the internal speaker on and off allowing the operator to listen for the detection of the tracer gas and not have to watch the display By default the audio output is turned off 10 Each button has a raised PUMP SPEED RESET A M A M SAMPLE Allows manual selection of various pump speeds to increase accuracy of the sample reading in situations with very small leaks At slower speeds the unit draws in less ambient air with the sample The Reset button allows the unit to remain powered up while clearing all the internal logic circuitry When the button is pressed the device goes through its zeroing routine to determine a new baseline for future samples Using this button to reset the device is preferred to turning the device off and back on as it maintains power to the sensing elements thus maintaining sensor stability Only press this button in a zero tracer gas environment The
15. apid movement of the control unit may produce false reading hold the control unit basically horizontal e The ground probe has an additional filter inside the yellow sec tion protecting the unit the needle probe does not e Continuous counting up of the detector is an indication of a clogged internal filter and will require factory replacement of inter nal filtering e If a constant reading of 25 50 is displayed with no helium present it may be necessary to press the Zero button to re zero the detector e The humidity moisture filter cartridges can be dried out and re used Placing them in an oven at 200 F for about an hour will turn the crystals blue again Or passing dry Nitrogen through them for about 10 min using the handle needle probe and moisture clear ing adapter 20 8 0 Tracer gas leak locating for pressurised ca bles 8 1 Assumptions e You have read and understand all safety information ex pressed or implied in this MGD 2002 Instruction Manual as well as your local safety codes and requirements e You have pinpointed the location of the system under ground e You are able to inject helium at the central office or valve points closest to the leak e You can drill holes in the concrete pavement every 3 feet 1me ter or so e You are using 100 Helium or 4 hydrogen with 95 nitrogen mix as a tracer gas 8 2 Applicability This method can be used on any pressurized cable system that can be fed with tracer
16. ater is encountered check the color change indicator in the wand handle 33 10 2 Moisture Filter replacement In order to replace the Moisture silica gel filter firmly grip the filter handle with one hand and turn the knurled cap assembly counter clockwise until they separate Once the cap assembly is removed push the filter out by using the viewing holes in the handle To restore these cartridges to a dry deep blue useful condition Method 1 Backflush the hand held probe with compressed Dry Air or Dry Nitrogen Use the Needle Probe connected to the front end of the probe handle and the clearing adaptor assembly connected to the hose The needle probe must be installed in order to defeat the one way check valve and allow air to flow back through the filter This process will take approximately 10 minutes per moisture filter Cartridge to turn them back to a deep blue color Method 2 Place the moisture filter cartriges in an oven at a maximum temperature of 200 F for approximately 60 minutes This method shortens the useable life of the cartridge Note Overheating will deform the cartridge 34 11 0 Calibration The MGD 2002 is a precision measuring device that requires factory calibration in order to ensure continued and sustained optimum operating performance The amount of time between calibration cycles is dependant upon the environment in which the device is to be used the end user s criteria for the equipment
17. ay concrete pave ment vinyl sheeting affect the time for the tracer gas to reach the surface of the ground Unfortunately with all these variables it is difficult to determine how long it will take for the tracer gas to reach the surface of the ground 7 11 How long to wait The system is filled with tracer gas and we know that the leak is allowing tracer gas to rise to the surface How long this will take is dependent on several factors such as the depth of the piping system what material loose earth clay concrete pavement vinyl sheeting is used as backfill and in the way slowing the tracer gas as it rises to the surface of the ground The denser the mate rial the slower the progress of the tracer gas to the surface It is virtually impossible to give equations that will help you calculate the time needed as the conditions in the ground are seldom known well enough to make calculations useful Although there are some basic rules and the majority of cases the waiting time is between 20 min 2 3 feet deep beach sand and 2 hours gt 6 feet deep gravel with additional time added for concrete or other dense materials e Materials like vinyl sheeting are non porous and are very difficult to permeate and may require holes to allow a measurable amount of tracer gas to escape e Thick pavement or concrete may add more than one hour to waiting time This extra time can be avoided by drilling holes to al low the tracer gas to vent to th
18. ded The SRA number must be on the outside of the package or indicated on the shipping label NOTE Do not ship equipment contaminated with any type of hazardous harmful substance SPX Dielectric 28 Tower Road Raymond ME Phone 207 655 8525 Toll Free 877 247 3797 Fax 207 655 8535 Email rd sales us spx com 40 WARRANTY The Manufacturer warrants that all goods supplied hereunder whether or not of its own manufacture will be of the kind described herein or in any specification and drawing approved by the Manufacturer and free from defects in material or workmanship under normal use and prescribed maintenance for a period of one 1 year with the exception of air dryers utilizing water sealed compressors as well as the compressors themselves which shall be for two 2 years Neither this warranty nor any other expressed or implied shall apply to goods delivered hereunder which have been damaged or subjected to alteration or negligence after delivery The Manufacturer s only obligation for breach of this warranty shall be the repair without charge or the furnishing EX Works Raymond Maine of a similar part to replace any part which within one 1 year with the exception as noted above from date of shipment is proven to have been defective provided that i the Purchaser shall have notified the Manufacturer within ten 10 days of the discovery of such defect and not later than ten 10 days after the last day of this warran
19. e surface e The deeper the pipe is buried in the ground the longer it takes to come to the surface e Higher water content in the soil will slow down the gas 7 12 Leak locating basics The highest concentration of tracer gas will be along the system path with the highest reading directly over the leak However cracks in concrete or pavement and differences in material density 18 soil will allow the gas to escape more readily in those areas A concentration of gas found several inches away from the marked system is due to the gas taking the path of least resistance But is an indicator that you are getting close The tracer gas will rise from the leak in most soil materials in an inverted cone shape When it reaches a denser material such as concrete it will slow its up ward travel and spread plume thus too much Helium too fast can make pinpointing more difficult 7 13 Preparing the MGD 2002 detector Assemble the black tube of the foam covered handle to the con trol unit attach the ground probe yellow stick to the handle Turn the detector ON and allow it to complete its 60 sec self test The detector should now read 0 Using the detector at its default set tings is best for this type of leak locating Changing pump speed or using manual mode is not recommended in this situation 7 14 Looking for the leak Start at the tracer gas feed point and begin to survey the system by placing the rubber cup at the end of the gr
20. ely 8 hours 240 30 8hr Once the system has been filled and tracer gas is detected at the vent Close the vent and begin to build pressure Continue to add tracer gas to the system as it will keep the concentration elevated 7 9 Very large piping system 12 or larger or tanks There are additional challenges in larger piping systems and tanks Getting enough tracer gas to saturate the system can be a difficult one This can be easily accomplished by emptying a full tank of tracer gas into the system to get the concentration of gas up rapidly Then provide additional tracer gas along with com pressed air to build the desired pressure Remember it is not nec essary to achieve 100 concentration in the system That would be a waste of tracer gas 5 10 is certainly adequate for leak locating as the MGD 2002 is capable of reading Helium as low as 25ppm 1 10 000ppm Another challenge for very large diameter pipe gt 2ft and tanks is that it may be necessary to provide some form of air movement inside the piping system as the tracer gas will want to concentrate near the top of the pipe This can be accomplished using a fan inside the piping to provide mixing 17 7 10 Estimating charging times The charging times pressure used and the amount of tracer gas needed are dependent on the size of the pipe and the expanse of the piping system being tested Additionally the size of the leak and the material over the pipe loose earth cl
21. ete and pavement slowly and this can cause a plume to build up under the concrete and when it does come through it will be in a much larger area In cases where there is substantial dense material such as concrete or pavement over the piping system it may be necessary to drill holes every two to three feet over the pipe to improve the search results For longer runs of piping with larger leaks under concrete or pavement every 5 8 feet will narrow the search area These holes become vents and allow for better pinpointing Systems that are imbed ded in concrete will not require holes However it will require time depending on the depth of the concrete Keep in mind that the tracer gas is lighter than air and wants to rise Start in the lowest point of the system as possible to allow 15 it to go up hill Additionally you will need to vent the other end of the system until tracer gas can be detected at this vent This will ensure you that the tracer gas has reached the leak Once detected at the vent the vent can be closed and the system can be pressurized with tracer gas Approximately 3 5 PSI to start but can be increased safely up to 10 of its normal working pressure if the leak is not found The helium molecule is much smaller than a water molecule and will exit a leak faster and easier than water will The best approach is to start with lower pressures and increase slowly Remember too much helium to start can make a leak easy
22. f equipped A serial port can be added to the MGD 2002 Contact customer service for more information Data is sent at 57600 n 8 1 in ASCII format with a carriage return 0x0D and line feed Ox0A after each value so it reads out fine in Hyper Terminal The values are bounded by spaces so a file cap ture can be done and then imported into a spreadsheet with space delimiting Values are sent in Automatic mode every 100ms 10 0 Maintenance 10 1 Cleaning amp inspection Instrument Clean away any debris or moisture accumulation with a clean soft damp cloth being careful not to scratch the protective LCD lens cover Check the exhaust ports on the top right side of the unit small brass fitting for any obstructions Ground Probe Collection Boot Clean with compressed air or with soap and water to remove mud and debris Dry the collection boot thoroughly when done Clean the small filter in the collection boot with a small brush as neces sary Small Probe Tip Clean with compressed air or with soap and water to remove mud and debris Dry the tip thoroughly when done Moisture Filter cartridge Check the filters in the handle before every use If the cartridge is pink or whitish in color replace the cartridge following the instructions below The handle can be cleaned with soap and water as long as the silica filter is in place and the knurled cap section is tightened firmly Note Do not allow water to enter the wand if w
23. g system and give you an idea of how big the leak is Pressurized air systems may show a sig nificant pressure drop or difficulty maintaining desired pressure Other systems may require other methods of leak determination 14 7 5 Locate the system Locate and mark the piping system path in the projected search area noting any recent construction or ground shifting in the area that may have caused damage to the piping system There are a large variety of pipe cable locators available from SPX Radio detection Please contact SPX Radiodetection sales or customer service to purchase a locator for your application 7 6 Empty the system In the case of a liquid filled system Drain the liquid from the sys tem It is not necessary to have the system completely dry but if the leak is under liquid the tracer gas will need to push the liquid out prior to exiting the system Smaller piping such as heating sys tems approx 1 inch or smaller may require using the tracer gas or compressed air to push the liquid out A Caution Compressed Gas Contains Energy Be careful around pressurized systems loose parts may be ejected with high velocity if the pressurized system bursts 7 7 Filling the system with tracer gas Attach the tracer gas The key words from here on are Don t Rush Saturating the ground or the underside of a concrete pad with tracer gas will make it difficult to pinpoint the leak The tracer gas will filter through concr
24. her gasses there are some that would normally present false reading when in high concentrations such as CO CO2 and Ozone on a busy street with helium detectors using this above mentioned sensor type However we looked at this situation based on our experience with older models of our detectors and through the development of some patented filter media combinations and custom software we have eliminated or captured these gasses eliminating the majority of false readings Unfortunately methane even in very low concentrations can cause the MGD 2002 to react but this is generally only seen during certain environmental tests 5 0 Assembling the MGD 2002 for use When assembling the device refer to the List of Kit Contents on page 2 of this manual 1 Clip the black sampling tube quick release connector of the sampling handle 5 onto the connector on end of the MGD 2002 detector 1 2 Attach either the needle probe 6 or the Ground probe 3 To attach the needle probe snap it into the quick connect coupling of the sampling handle cap adaptor To attach the ground probe Unscrew the cap adaptor from the sampling handle and screw the sampling handle into the extendable ground probe 3 Do not overtighten Hand tight Note applying a small amount of silicone based lubricant to the seal on both ends of the moisture cartridge will improve its performance 3 To extend the ground probe press the lock button on the sample probe and wi
25. ium hydrogen to flow from the entry point to the desired end point Example 1 The cable flow rate at the bypass is 7 SCFH This will push the helium hydrogen 4200 feet from this reading point 2 Cable diameter is 2 35 inches 3 Go to the line for 2 4 inches in the left column of the table 4 Go across to the column titled 7 Your flow time per 1000 feet is 2 24 hours 5 Multiply 2 24 hours by 4 2 number of thousands of feet to find your total flow time of 9 41 hours This is the time it will take the helium hydrogen to flow from the reading point to a point 4200 feet distant in the cable 25 8 7 Direct Buried Cables a Introduce helium hydrogen at the valve point closest to the area of the suspected leak i Attach a model 530 helium flow controller to the helium hydrogen tank regulator see the 530 helium flow con troller operations manual ii Attach the inlet hose of a model 526 portable flow rater to one of the valves at the end of the model 530 iii Attach the outlet hose of a model 526 portable flow rater to the pressure valve of the system to be charged If the hose of the model 526 is not long enough to reach from the regulator to the valve on the cable add additional lengths of pressure hose iv Set the pressure gauge on the tank regulator to 50 PSI The resulting flow into the system will be approximately five SCFH of tracer gas This will register on the model 560 portable flow rater as app
26. minate it Sev eral inches of pavement will substantially slow the permeation of Helium Hydrogen to the surface which can cause pluming of gas under the pavement making it more difficult to pinpoint the leak Q The detector displays a negative zero shortly after taking a reading with high helium levels What does the mean A The sensor is recovering after being cooled by the high concen tration of Helium Hydrogen in a few moments it will again stabilize and automatically zero usually within one minute If it remains at a negative zero for more than three minutes the detector may need to be serviced Contact customer service 37 12 1 Hints e You can use a standard Helium balloon latex to practice leak locating The MGD 2002 will detect helium leaking through the surface of the balloon e A sample of 100 helium will not damage the sensor e Ensure the rubber cup is placed directly on the ground over the path of the system during sampling e Start with any suspected areas first Look for locations where obvious ground shifting of construction work has occurred For concrete embedded systems look for cracks over or near the system e Samples should be taken approximately every three feet over normal soil and less then three feet over compacted soil pave ment or concrete e Pipe that is close to the surface lt 1 ft may require samples to be taken at shorter intervals The gas has little time to spread in verted cone
27. ot operate the detector when the battery indicator displays 1 4 battery or less False readings may occur when operated under this condi tion Refer to pages 7 through 9 for instructions on how to recharge or replace the battery e When charging This unit should be in an indoor dry location e Do Not plug the battery in backwards Circuit board damage will result A CAUTION Refer to accompanying information Table of Contents Section Description Page User safety warnings i Table of contents ii 1 0 Specifications 1 2 0 List of kit contents 2 3 0 Unpacking your New MGD 2002 3 4 0 Tracer gas 4 4 1 Helium vs Hydrogen as a tracer gas 4 2 Other Gasses the MGD 2002 will react too 5 0 Assembling the MGD 2002 for use 5 6 0 Operation of the MGD 2002 6 6 1 Overview 6 2 Humidity vs Accuracy 6 3 Moisture filter 6 4 Power supplies 6 5 Charging the battery 6 6 Charging the battery using the universal power supply 6 7 Charging the battery from a 12VDC auto adaptor 6 8 Replacing the battery 6 9 Button function and operation 6 10 Liquid crystal display LCD 6 11 Display status icon 7 0 Tracer gas leak locating for piping systems 14 8 0 Leak locating for pressurized cables 21 9 0 Serial port if equiped 33 10 0 Maintenance 33 11 0 Calibration 35 12 0 Frequently asked questions FAQ 36 13 0 Troubleshooting 39 14 0 Service information 40 Warranty Back cover 1 0 Specifications Dimensions LxWxD 13 3in x 4 9in x 3 3in
28. ound probe directly on the ground over the system and allow a few seconds 15 20 for the sample to reach the internal sensor Then proceed along the system surveying at intervals of 3 feet or less If you have not found the leak s on the first pass of the system turn around and survey the system in the other direction 7 15 Repair the leak Once found take appropriate action to repair the leak and then test the system to see if there are additional leaks You should not as sume there is only one leak It will save a lot of time in the long run by taking the time to check 7 16 Empty the system of tracer gas Warning Relieve the pressure prior to disconnecting equipment Remove the gas injection equipment and open the valve used as the injection point Slowly open the isolation valves at the lower end of the system to let the system fill slowly Leave the injection point valve open until all the tracer gas has left the system Follow proper safety procedures for filling empty piping 19 7 17 Things to Note e Do not suck up water into the unit damage will result e Do not plug the battery in backwards damage will result e Allow the MGD 2002 to normalize to the temperature of the search area before turning it on Rapid changes in temperature can cause condensation to build up on the sensor shorting it out and requiring factory repair in most cases e Sweep or vacuum dirt dust away from drilled holes in pavement or concrete e R
29. r located on the right hand side of the MGD 2002 6 8 Replacing the battery A Caution Do not mutilate puncture or dispose of bat teries by placing them in a fire The batteries can burst or ex plode releasing hazardous chemicals Discard used batteries in accordance with the battery manufacturer s instructions and your local regulations 1 Lay the device face down on a flat surface and using a flat headed screwdriver turn the battery cover retaining screw 1 4 of a turn counterclockwise and remove the battery housing cover 2 Lift out the battery and gently pull up the battery connector lead 3 Before installing the new battery attach the three pin con nector to the circuit board ensuring it is connected in the correct orientation Note Connecting the battery incorrectly even for a second will destroy the circuit board 4 Ensure that the battery is fitted in the correct position and that it lies flat in the housing 5 Install the battery cover taking care to avoid pinching the wires Tighten the retaining screw 1 4 turn clockwise 6 9 Button functions and operation The MGD 2002 features membrane buttons and has a large multi function liquid crystal display LCD embossed edge that is designed to be felt by a user even when wearing protective gloves When they are pressed the buttons make a noticable click that is both felt and heard The following is a description of each key on the MGD CONTR
30. roximately two SCFH of air or five SCFH of tracer gas These readings indicate that the charging process is correct b Refer to Appendix A or the cable filling times chart at Appendix B to determine the time needed for the helium hydrogen to flow to the point of the suspected leak Note When charging direct buried cables allow extra time for the tracer gas to escape from the cables and rise up through the ground Atypical tracer gas tank contains 240 standard cubic feet of helium or hydrogen mix when full At a constant flow rate of five SCFH a full tank will keep adding tracer gas to the system for approximately 48 hours 26 8 8 Underground cables a Introduce tracer gas at the meter panel in the central office that feeds the leaking cable b The tracer gas will flow into the underground cable from manhole locations or riser pole valves instead of the central office This may reduce the charging time c Introducing tracer gas directly into an underground cable from a central office is accomplished as outlined on page 22 8 9 Charging air pipe systems Introduce tracer gas into the pipe system at the pipe panel in the central office Find the outlet side of the pipe panel and flow he lium hydrogen into the pipe at five SCFH a Place the tracer gas tank close to the pipe panel to be charged Ensure the tank is secure b Use a model 530 helium flow controller and model 526 por table flow rater as described in the
31. second sample on the opposite side of the cover and record the result d Proceed to the next manhole and repeat the procedure Survey all the manholes in the projected search area and record the results Return to the manholes with the highest percentages of tracer gas readings When entering manholes always follow local safety practices and search for leaks using the SPX Radiodetection Model1805 Multisonic translator detector If no faults are found in the manhole check the ducts using the MGD 2002 Do this using the needle probe extension tip Insert the probe into the duct as far as you can taking care not to get moisture into the unit Take a sample and record the result Examine and clean the probe tip if necessary after each sample 28 Using duct probe with multisonic tip to pinpoint leaks in conduit If tracer gas is detected in a duct the SPX Radiodetection Model 1802 mini duct probe may be used to establish the exact location of leaks between manholes fig 9 If pressure leaks are not found in the manholes or adjacent ducts the pressure leaks may be in riser cables or riser cable plugs 29 8 11 Leak Locating tips e Locate and mark the cable path in the projected search area e Ensure the rubber collection cup is placed on the ground directly over the path of the cable during the sample and read cycles e Samples should be taken approximately every three feet over normal soil and less than three feet o
32. te lighter socket 6 5 Charging the battery Charging from a completely dead battery will take approx 2 hours and should last for approximately 6 8 hours of run time While charging the battery will heat up and become warm As the sensor is sensitive to temperature changes it will be necessary to wait approx 20 min for the battery temperature to return to normal before using the device Note It is not possible to use the MGD 2002 during battery charging as the detection capability is automatically dis abled when the battery charger is plugged in Note The battery is equipped W a temperature sensing element to ensure the battery will not overheat during the charging cycle If the ambient temperature is too high bat tery charging may shut off prematurely Ensure that the MGD 2002 is charged away from any heat sources 6 6 Charging the battery using the universal power supply 1 Plug the AC charging lead into the AC DC adapter and plug the other end into a standard power outlet 2 Plug the DC adapter lead into the charging connector located on the right hand side of the MGD 2002 Note When the charger is plugged into the MGD 2002 the LCD will illuminate and display the battery recharging symbol Once charging is complete the symbol will change back to a plug 6 7 Charging the battery from a 12VDC auto adapter 1 Plug the auto adapter into the vehicle cigarette light socket 2 Plug the output jack into the charging connecto
33. th it pressed gently pull on the sample collection cup When the sample probe is at the required length release the lock button ensuring it engages in a locking position 4 Turn the MGD 2002 on The MGD 2002 will perform a self check and zeroing routine 1 min approx You are now ready to look for Helium Hydrogen 6 0 Operation of the MGD 2002 6 1 Overview The MGD 2002 gas detector is a portable leak locating and pinpointing device that can be used in a multitude of pressurized systems to detect both hydrogen and helium when used as a tracer gas Because they are both lighter than air hydrogen and helium penetrate small leakage points rapidly The gasses permeate through the densest soils and pavements quickly enabling the leak to be pinpointed easily with the detector 6 2 Humidity vs Accuracy The MGD will function correctly at high relative humidity RH levels of 50 or greater but its accuracy and sensitivity may be reduced Lower humidity is prefered 6 3 Moisture filter The replaceable moisture cartridge filter removes much of the moisture from the sample and is designed to enable users to quickly asses the filter daily life span by visual inspection A new dry cartridge will be bright blue in color and is shown on page 7 The moisture cartridge will turn pink as it removes moisture from the air As it changes color the filter cartridge is still within its useful life but will degrade further to a clear white color if
34. th the highest reading directly over the cable leak Before using the MGD 2002 mark the path of the cable Precise cable location can be accomplished while the cable is being charged with tracer gas 8 5 Charging underground cable systems with tracer gas from the central office To find leaks in underground systems feed the tracer gas from the central office pipe or distribution panel 22 Charging a cable or air pipe system When applying the tracer gas to a distribution panel use the Model 526 amp 530 for panel applications If you do not use the model 526 amp 530 or a locally assembled equivalent the cables may not all be charged with the tracer gas Tracer gas will escape from faults in manhole or duct structures once the system is properly charged Tracer gas will rise to the top of the manhole and out through the lid where it can be detected Searching for tracer gas 23 8 6 Directly charging the cable When applying the tracer gas directly to a cable refer to the chart page 31 This chart shows the amount of time needed for the helium hydrogen to flow to the suspected leak location This chart should be used whenever tracer gas is introduced directly into the cable This applies to both direct buried and underground cable types You will need the following information before using the chart a Cable type and gauge b Estimated distance to the leak DTL established through analysis c Total length TL
35. to find large area but hard to pinpoint Also note that helium will travel in the path of least resistance It will find the easiest path it can and that path may not be where you would expect A Caution Never pressurize any system above its working pressure 7 8 How much tracer gas Calculate the volume Area Tt R 2 note Radius R in ft Volume of system cu ft Vsys Area sqft X Length ft Example 500 of 2 dia pipe 3 142 0 084 42 0 022 sq ft 022 500 11 cu ft In this example 11cu ft is the total volume of air inside the pipe Note again it is not necessary to get the pipe to 100 tracer gas concentration In most cases 5 10 is adequate for leak locating Now lets calculate the volume of air inside the pipe when we increase the pressure to 5 PSIG Building upon the previous ex ample Total cubic feet required at 5 PSIG V 5PSIG Vsys 14 7 5 14 7 16 V 5PSIG 11 19 7 14 7 V 5PSIG 14 74 cuft with no leaks At 5 PSIG you would need only 1 47 Cu ft to reach 10 helium in 500ft of 2 dia pipe with no leaks In a small system such as this building the pressure using the tracer gas is perhaps the easi est way to increase and maintain the pressure while keeping the tracer gas levels elevated A typical tracer gas tank contains 240 standard cubic feet of tracer gas when full At a constant flow rate of 30 SCFH a full tank will keep adding tracer gas to the system for approximat
36. tor gives false readings where there is no Helium Hydrogen A The battery heats up as it charges raising the temperature of the air inside the detector above room air causing false readings Allowing the unit to cool for approx 15 20 min after charging will correct this problem Opening the battery compartment during charging can reduce this cooling time Q How do dry the Humidity cartridge without a source of dry air A A tank of Dry Nitrogen can be used in place of Dry air As another alternative the humidity cartridge can be baked in an oven at no more than 200 F for approx 30 60 min We do recom mend in application where dry air is not available that you pur chase extra humidity cartridges Q Will gasses other than Helium Hydrogen effect my readings A Yes other gasses can affect the readings but most are ab sorbed within the internal filters or have a lower thermal conduc tivity as compared to air which will have little or no effect on the readings See section 4 2 on page 4 Q Will 100 helium Hydrogen burn out the sensor A No This sensor is not damaged by high concentrations of he lium hydrogen Q Can my MGD be used to locate leaks in natural gas lines A Yes Contact our customer eaog department for information on proper usage in these applications Q Will this detector completely eliminate the need to drill holes in pavement or concrete A No although it can reduce this need it will not eli
37. ty and ii the Manufacturer shall have the option of requiring the return of the defective material transportation prepaid to establish the claim The Manufacturer shall not in any event be liable for the Purchaser s manufacturing costs loss of profits good will or any other special consequential incidental or other damages resulting from such defects THERE ARE NO OTHER WARRANTIES EXPRESSED OR IMPLIED WHICH EXTEND BEYOND THE WARRANTY SET FORTH HEREIN SP 28 Tower Road Raymond Maine 04071 USA Tel 207 655 8525 Toll Free 1 877 247 3797 Fax 207 655 8535 Email rd sales us spx com www radiodetection com 2015 Radiodetection Ltd All rights reserved Radiodetection is a subsidiary of SPX Corporation SPX the green gt and X are trademarks of SPX Corporation Inc Radiodetection is a trademark of Radiodetection in the United States and or other countries Due to a policy of continued development we reserve the right to alter or amend any published specification without notice This document may not be copied reproduced transmitted modified or used in whole or in part without the prior written consent of Radiodetection Ltd
38. ve read and understand all safety information ex pressed or implied in this MGD 2002 Instruction Manual as well as your local safety codes and requirements e You have pinpointed the location of the system under ground e You have emptied the system of liquid e You are able to inject helium at the low end and provide a temporary vent at the other end to allow rapid filling e You can drill holes in the concrete every 3 feet 1meter or so e You are using 100 Helium or 4 Hydrogen as a tracer gas 7 2 Applicability This method can be used on any piping system large or small that can be emptied and filled with tracer gas The gas will escape from the system at the leak and be easily detectable at the surface of the ground 7 3 Effects on subscribers With this method of leak locating the effects on the subscribers is dependant on the type of system being checked In the case of a municipal water or gas system subscribers should be notified according to normal procedures Additionally on water systems subscribers along this system should be instructed not to open a tap during the leak detection process There is no danger from the gas but large amounts of tracer gas would be lost Other system types may not require any notification 7 4 Preparing to leak locate Leak test Prior to leak locating you must actually determine you have a leak Performing a hydrostatic leak test on water systems will confirm that you have a leak in the pipin
39. ver compacted soils e Frozen ground may cause tracer gas to migrate a consid erable distance from the actual leak e It may be necessary to make holes at curbings or road expansion joints to help the tracer gas to rise to the surface e When tracer gas is detected continue to sample the adja cent area until you determine the highest concentration of the tracer gas Boring holes with a t bar drill may be help ful at this point This location will be directly over the cable fault 30 isos gt gt GAUGE SSS APPENDIX A 2000 3 TL FEET INSTRUCTIONS 1 Locate cable type and gauge on upper horizontal axis 2 Locate DL on lower horizontal axis 3 Draw line connecting points and o 4 Locate TL on left vertical axis 5 Draw a line from point through the intersection of the line connect ing and and the fixed diagonal line to intersect the right vertical axis 6 Read waiting time from the right vertical axis 4000 3000 31 2000 1000 DL FEET N a WAITING TIME HOURS 20 15 10 APPENDIX B Approximate Cable fill times Approximate cable fill times based upon head end flow rates stated in hours per 1000 feet Cable diameter in inches aetna ea oe 35 fea s aa fos fo fefe 1 8 8 85 4 4 2 95 1 8 1 5 1 25 1 1 1 0 P22 ra 66 44 93 205 22 105 ses 15 126 3 0 24 55 12 3 8 2 6 15 4 9 4 1 3 5 3 1 2 75 2 45 32 9 0 Serial port i