Contents
1. 5 kV amp 10 kV Insulation Resistance Testers USER MANUAL A SAFETY WARNINGS Safety warnings must be observed during use The circuit under test must be switched off de energised isolated and checked to be safe before insulation test connections are made Make sure the circuit is not reenergised whilst the instrument is connected Only Megger supplied test leads must be used Circuit terminals must not be touched during an insulation test or before suitable grounding of the unit under test is in place in line with safe working practices After completing a test capacitive circuits must be completely discharged before disconnecting the test leads Capacitive charges can be lethal Tested items must be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby raising the voltage to potentially dangerous levels The voltage indicator and automatic discharge features must be regarded as additional safety features and not a substitute for normal safe working practice It is rare but in certain circumstances breakdown of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner possibly causing a loss of display while the circuit remains energised In this event the unit must be turned off and the circuit discharged manually Test leads including crocodile clips mu
2. You will be asked to provide the following information to enable the Megger Service Department to prepare in advance for receipt of your instrument and to provide the best possible service to you e Model e g MIT1025 e Serial number to be found on the underside of the case or on the calibration certificate e Reason for return e g calibration required or repair e Details of the fault if the instrument is to be repaired 2 Make a note of the RA number A returns label can be emailed or faxed to you if you wish 3 Pack the instrument in the original packing box to prevent damage in transit 4 Ensure the returns label is attached or that the RA number is clearly marked on the outside of the package and on any correspondence before sending the instrument freight paid to Megger Copies of the original purchase invoice and packing note should be sent simultaneously by airmail to expedite clearance through customs In the case of instruments requiring repair outside the warranty period an immediate quotation can be provided when obtaining the RA number 5 You may track the progress of your return on line at www megger com Approved Service Centres A list of Approved Service Centres may be obtained from the UK address above or from Megger s website at www megger com Accessories Equipment and Spares Accessories all models Part Included Accessories all models Part Spare Li ion Battery 1002 552 Number User g
3. General operation The MIT515 MIT525 MIT1025 Insulation resistance testers IRTs are primarily controlled by two rotary switches and a TEST button used to start and stop a test see section entitled Instrument Control and Indicators The central rotary switch includes an OFF position the instrument switches on by rotating the switch either clockwise or anticlockwise from this position A range of test voltages for insulation resistance tests up to 5 kV for MIT515 MIT525 and 10 kV for MIT1025 are available including a user selectable voltage range which can be set between 100 V and 5000 V or 10000 V depending on model The lockable test voltage range can be adjusted in the settings function The settings function is indicated by a spanner symbol and facilitates lock voltage low resistance alarm temperature time date adjustment A light blue coloured section of the rotary switch denotes memory functions open records download via USB and delete records A dedicated save button is provided on the MIT525 and MIT1025 models and all models have a backlight button A second rotary switch controls the insulation test mode providing for the following tests e All models have basic insulation resistance IR timed insulation resistance IR t Dielectric Absorption Ratio DAR and Polarisation Index PI tests e MIT525 MIT1025 have additional tests Dielectric Discharge DD Step Voltage SV and ramp test A cluster of d
4. Maintenance Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA Care should be taken near the terminals IEC power and USB sockets Care of the instrument The instrument should always be handled with care and not dropped Always ensure that the instrument is secured when being transported to prevent mechanical shock Leads Leads are silicone insulated and perform well in all weather conditions Always keep the leads in the clip on lead pouch supplied with the instrument Regular inspection of leads is recommended to ensure they are not damaged in any way Damaged leads could affect insulation resistance readings and are a safety hazard Battery Care The battery should be charged on a routine basis at an absolute minimum of once a year However more frequent charging i e once per quarter is preferable Never attempt to charge the battery below 0 C or above 40 C The battery is charged by connecting line power at the instrument IEC power socket Store the instrument in a cool dry location to improve battery life Storage temperatures below freezing should be avoided Replacing the battery Read and fully understand the warnings on the Li ion battery in the Safety Warnings section of this document 21 The battery pack contains Lithium ion cells and should be replaced when it no longer holds a charge A new battery is available as a spar
5. Oberursel Andheri w Mumbai 400053 Germany Maharashtra T 06171 92987 0 India F 06171 92987 19 T 91 22 26740468 F 91 22 26740465 Megger products are distributed in 146 countries worldwide This instrument is manufactured in the United Kingdom Megger Z A Du Buisson de la Couldre 23 rue Eug ne Henaff 78190 TRAPPES France T 33 0 1 30 16 08 90 F 33 0 1 34 61 23 77 E infos megger com Megger AB Eldarvagen 4 Box 2970 SE 187 29 TABY Sweden T 46 8 510 195 00 F 46 8 510 195 95 The company reserves the right to change the specification or design without prior notice Megger is a registered trademark Part No MIT515_525_ 1025 UG en_V01 0312 Www megger com 2 Megger Pty Limited Unit 1 11 21 Underwood Road Homebush NSW 2140 Australia T 61 0 2 9397 5900 F 61 0 2 9397 5911 E ausales megger com Megger AG Ob Haselweg 630 5727 Oberkulm Aargau Switzerland T 41 62 768 20 30 F 41 62 768 20 33 Megger Megger Limited Unit 106 550 Alden Road Markham ON L3R 6A8 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com
6. and LED do not flash turn off the instrument before disconnection Instrument Control Initial setup It is important to setup the Real Time Clock RTC on models MIT525 and MIT1025 to ensure that records saved in the instrument are time date stamped correctly The MIT515 does not require time date setting The RTC has a separate battery to maintain settings even when the primary battery is removed To set the clock and date select the settings function on the central Rit s rotary switch and turn the mode rotary switch to IR Navigate using the left right arrows to where the time and date is displayed PI DAR DD V Set the time using the up and down arrows Change the hours and minutes then press OK to save 24 80 g0 go0 Select the day month format required i e d m for day month or m d for month day and press the right arrow button then set the date and press OK to save A tick on the left of the display indicates that a setting is saved a cross is displayed during adjustment indicates that it is not set Exit settings by changing the central rotary switch to a different position mod Jia o0 jie Se Lock Voltage The user selectable lock voltage range is set by adjusting the displayed voltage using the up and down arrow buttons When the desired voltage is displayed it is saved by pressing the 10 OK button The setting does not change even if the instrument is switched off Alarm setting A
7. breakdown amp Burn Left and right arrow buttons toggle between burn and breakdown mode when a voltage range is selected Press and hold left arrow burn to activate burn mode In breakdown mode the test will automatically terminate on detection of a breakdown to prevent damage to the insulation Burn mode disables the normal breakdown detection and test voltage continues after breakdown of the insulation This enables the location of the failure to be seen and detected acoustically but it is a destructive test Due to the potential damage that could occur the unit produces two long beeps when starting a test with burn mode activated Running an insulation test Before testing any reactive load the insulation must be fully discharged The required resistance test is selected on the mode rotary switch at the bottom left of the instrument keypad The test voltage is selected using the preconfigured voltage ranges on the central rotary switch or the V amp user settable lockable voltage range It is possible to adjust the test voltage using the up and down arrow buttons either before or during a test Once a test has begun it is advisable to only adjust the voltage in the first 10s of the test to prevent interference with the capacitive and absorptive currents in the insulator A test is started by pressing and holding down the TEST button for approximately 3 seconds A timer will be displayed to indicate elapsed
8. low resistance alarm sounds when the resistance level of an insulator reaches the alarm assuming alarm has been activated The default alarm setting is 500 kQ and inactive X is displayed on the right of the display Set central and mode rotary switches to the settings and IR positions respectively Press the right arrow button once The low resistance alarm can be set at the default value by simply pressing the OK button or changed to a different alarm resistance level using the up down arrow buttons and save it by pressing OK COOK S e Recording temperature The MIT525 and MIT1025 are able to record insulation temperature measured by an independent thermometer If you do not wish to record temperature do not change the default setting or reset it if it was previously set Move the central rotary switch to point to settings and press the right left arrow buttons until t is displayed The default setting is no temperature record This can be changed by pressing up or down arrows to select either F or C temperature entry Pressing OK will confirm the setting and result in a prompt for temperature to be entered whenever the save button is pressed after completing any test Up and down arrows facilitate temperature entry in 1 C increments decrements O o 0 Breakdown mode burn mode The insulation resistance IR test operates in either Breakdown or Burn mode 4 elegncevun Default mode is
9. time during the test The test is stopped by pressing the TEST button As soon as the test is stopped a discharge of the insulator is automatically initiated An StP indication informs the user that the test is terminating and after a few seconds the voltage on the terminals will be displayed Left and right arrows can be used to scroll between terminal voltage last test voltage and the set range voltage In the event of a terminal voltage of 250 V a voltage and warning will be displayed Do not disconnect instrument leads or clamps until the LED and display warnings are switched off indicating that the unit under test is discharged Significant current can be stored in reactive loads which act as capacitors or inductors which can be lethal The display shows the final resistance result capacitance test current and Time Constant TC in addition to test duration On MIT525 and MIT1025 models the result can be saved by pressing the dedicated save Fi button after a resistance or voltage test is complete The save button will appear momentarily to confirm the data is saved If a full test curve is required the user must select logging by pressing the save button before starting the test Data will be logged every 5 seconds for the duration of a resistance test It is not possible to log voltages in voltmeter mode If temperature entry has been activated a prompt will appear for the user to enter a temperature reading after
10. 0 V steps 1 kV to 5 kV in 25 V steps MIT1025 5 kV to 10 kV in 25 V steps Accuracy 23 C MIT515 MIT525 MIT1025 5 to 1 TO 20 to 10 TO 5 to 2 TO 20 to 20 TO Guard 2 error guarding 500 kQ leakage with 100 MQ load Display range analogue 100 kQ to 10 TQ Display range digital MIT515 MIT525 10 KO to 10 TO MIT1025 10 KO to 20 TO 23 Short circuit current 3 mA nominal Maximum power regulation technology ensure maximum power transfer throughout all loads not just at short circuit outperforming many 5 mA testers Insulation alarm 100 KO to 1 GO Capacitor charge MIT515 MIT525 MIT1025 lt 3 s UF at 3 mA to 5 kV lt 5 s uF at 3 mA to 10 kV MIT1025 Capacitor discharge MIT515 MIT525 MIT1025 lt 250 ms uF discharge from 5000 V lt 500 ms uF to discharge from 10000 V to 50 V Capacitance range above 500 V 10 nF to 25 uF dependant on measurement voltage Capacitance accuracy 23 C 10 5 nF Voltage output accuracy 0 C to 30 C 4 0 10 V nominal test voltage at 1 GO Current measurement range 0 01 nA to 6 mA Current measurement accuracy 23 C 2 0 5 nA at all voltages Interference noise rejection MIT515 MIT525 1 mA per 250 V up to a maximum of 3 mA MIT1025 1 mA per 600 V up a maximum of 3 mA Voltmeter range 30 V to 660 V ac or dc 50 60 Hz Voltmeter accuracy 3 3 V Timer range Up to 99 minutes 15 second minimum setting Me
11. 30 minute Acceptable fA period can be adjusted but care should be taken to ensure that God full absorption will take place in the insulation test period DD should be selected on the test mode rotary switch and settings PI results gt 1 5 are regarded as acceptable by IEC60085 on the central rotary switch Timer t1 is set using the up and 01 1984 for thermal class rating A and PI results gt 2 0 for down arrow buttons and OK confirms the setting thermal class ratings B F and H The DD test requires the instrument to measure the discharge current 1 minute after the removal of the test 15 voltage which is greater than the primary time constant of the testing If a PI test was performed to verify the winding s discharge On completion of the test the instrument uses this suitability for over voltage testing the winding must be measurement along with the test voltage and calculated completely discharged before the overvoltage test is capacitance to produce a figure of merit indicating the quality performed of the insulation The SV test is based on the principle that an ideal insulator will DD limin V x C produce identical readings at all voltages while an insulator which is being over stressed will show lower insulation values where hmin is the discharge current in mA one minute after at higher voltages removal of the test voltage V in Volts and C is the capacitance in Farads During the test the appl
12. C and DC voltage from 30 V to 660 V Frequency Hz is measured and displayed for AC voltages The voltmeter can be used to confirm that the insulation is discharged and for general purpose use Positive and negative terminals are used for the voltmeter function do not connect the GUARD terminal in voltmeter mode Before starting a test the voltmeter detects any voltage between the positive and negative terminals and will display the voltage If the voltage is above 50 V at the start of an insulation test a voltage is displayed on the analogue and digital readouts to alert the user The voltage displayed is developed by current flowing through the internal discharge resistance of the instrument If the source impedance is high the current will remain constant regardless of the value of the internal discharge resistance In high electrical noise environments a lockout activates when induced currents exceed the capability of the instrument to remain within specified accuracies The voltmeter reports the voltage across the internal discharge resistor Voltage during discharge of a reactive load will be displayed in addition to warning symbols Non reactive loads will discharge too rapidly to show the voltage drop In the unlikely event of a fault during a test or during the discharge process the user is advised to check for flashing high voltage warning symbols and the red LED on the instrument panel will flash If the warning symbols
13. EST to start the test DAR and PI Insulation Tests DAR and PI tests are measurements of resistance over time expressed as a ratio of resistance at time t2 divided by resistance at time t1 The assumption is that insulation temperature does not vary widely over the duration of the test so the resulting DAR and or PI value are temperature independent i Testing should be done at or below 40 C 104 F for this assumption to hold DAR and PI timers t1 and t2 are set when DAR or PI is selected on the test mode rotary switch with the central rotary switch in the settings position Timer t1 is set first followed by t2 Up and down arrow buttons are used to change the t1 and t2 default values and OK confirms each setting DAR and PI insulation test voltages are selected on the central Dielectric Discharge test rotary switch by simply aligning the switch opposite to the required insulation test voltage Press and hold TEST to start a DAR PI test The Dielectric Discharge DD or re absorption r current test operates g a during the discharge of the dielectric under test r Originally developed by EDF France s power utility company itis a DAR IRsos IRaos diagnostic insulation test that allows ageing deterioration and voids in the insulation to be assessed The result is dependent DAR result on the discharge characteristic so the internal condition of the insulation is tested largely independent of any surface conta
14. IR and IR t insulation tests DAR Pl SV ramp and DD tests will not prompt for temperature input Display backlight is activated by pressing the button The backlight button can be pressed a second time to deactivate the backlight Automatic deactivation will occur after a preset timeout period if not deactivated manually Memory functions and downloading Models MIT525 and MIT1025 have advanced storage recall and download functions to facilitate documentation of insulation tests Recall results Setting the central rotary switch to open folder position enables the user to recall saved results beginning with the most recent result Up and down arrow buttons enable the user to scroll through results based on a sequential four digit index Left and right arrow buttons scroll through a single result showing all saved test data including time date Where logging has been enabled only the final result is displayed on screen The full result can be viewed by downloading to PowerDB PowerDB Lite In saved results the test mode is identified by the icon or abbreviation of each test on the display In addition the open folder icon is displayed to indicate recall memory mode Download results PowerDB Pro Advanced and Lite are Megger s asset and data management software packages with integrated forms for MIT525 and MIT1025 instruments The default download on the instrument is a single test log or 12 summary
15. O C to 40 C or that the battery has failed On screen error reporting Should an error be detected during the operation of the MIT515 MIT525 or MIT1025 an error code is reported preceded by Err with the read handbook warning Error codes are given in the following table 4 5 HV board mismatch with control board setup 2 Battery low error Control board detected inter board communication failure Test button stuck 8 Measurement board i2c failed Measurement board detected inter board communication failure Isolation power supply cut out lf an error occurs do not attempt to repair the instrument Obtain a repair number from Megger Instruments Limited carefully pack in a suitable box and send the faulty instrument to the nearest Megger Approved Service Centre if possible noting the error that was reported Measurement Modes Spot IR test The spot insulation resistance test IR is selected on the test mode rotary switch Select the IR setting and then the required test voltage using the preconfigured voltage ranges on the central rotary switch or the Vf user settable lockable voltage range All preconfigured voltage ranges but not Vi are adjustable using up and down arrow buttons before and during the test but the latter should be limited to the first 10 seconds of IR and IR t tests Press and hold TEST to start the test To set the user defined lock voltage V amp turn the central ro
16. anced memory storage includes time date stamping of results logging of data and recall of results to screen A fully isolated USB device interface is used for safe transfer of test records to all versions of Megger s data management software PowerDB or to popular third party spreadsheet packages The MIT515 MIT525 and MIT1025 can be powered from the mains supply or by it s rechargeable battery which provides for 4 2 to 6 hours of continuous testing with a 100 MQ load depending on model A battery level indicator on the LCD display indicates battery capacity Connecting power to line mains supply will automatically charge the battery whether the instrument is switched on or off except during testing Features e 10 TQ max resistance 20 TQ for MIT1025 e MIT515 with IR IR t Pl and DAR e MIT525 MIT1025 diagnostic and over voltage tests PI DAR DD SV and ramp test e Operate and charge on line power mains except during actual test e Light weight Li ion battery e CATIV 600 V safety rating e Advanced memory with time date stamp e DC and AC voltmeter 30 V to 660 V e Large LCD display with backlight e Download of saved results and logs via 10 kV isolated USB cable MIT525 MIT1025 only e PowerDB Lite software included with MIT525 and MIT1025 Instrument Controls and Indicators one O25 Positive terminal GUARD terminal Negative terminal USB devi
17. cation ccc cece eceeeeceeeeeeseeeeeeeeeseeeeeeeeenees 7 CS Al Pell eanacane ate odenas each tats eam eucking neniencedcasa ea osniaeimaaeeatatiabas 7 NOV AG S see a E a rennecatis acc useacsadaraass 7 Operating INStrUCTIONS ccccccececeeeceeeceeeceeceecaeeseeeeeeeseeeas 8 General OPEatlon erinan a a A 8 Breakdown vs burn MOE ccccceceeceeceeceeseeaecaeseeseeeeees 9 VOLMER 9 INNES eUD corian e a ER 10 LOCK VONAGO rrna 10 PAAR IMS OUI an acta tncseetecie teeta tites 10 Recording temperature cccccccsecceeceeeeeceeeeeseeseeees 10 Breakdown mode burn MOE ccceceeeeeeeeeeeeeeeneees 11 Running an insulation test cccccsecseeseeseeseeseeeeeseeees 11 Memory functions and downloading c scccseeeseeees 12 Real time output during insulation tests 12 Battery INGICAUOR enian eE E mens 13 On screen error reporting ccccceceseeceeecceeceeeeeeeseeeeees 13 Measurement Modes cccceccececeeceeceeceeceeeeseeseeaeeaes 14 SPOF IRES en nee een Ee ee 14 AMIN SGN ES ei E E 14 DAR and PI Insulation Tests ccceccecceeeeceeeeeeeeeeees 14 Dielectric Discharge test ccccsecsecceeceeseeeeeseesenees 15 Step Voltage TeSt cccccccsecccececeeeceeecseeseeeeeeeeeeeneeeses 16 Ramp voltage TEST sip o2 sete a Se lel oediot cee eee eens 17 Measurement TEChKniqQues cccccceceececceeeeeeeeecee
18. ce interface Four arrow buttons and OK button TEST button with associated HV warning lamp Backlight button Central rotary switch Save button on MIT525 and MIT1025 10 Test mode rotary switch 11 LED indicating line power mains 12 Display 13 Power socket Coto T S gt PIRDDARSV amp a OHEYT Sm mmi 4906 310G 416 5100M Grom 1 Ea 2 xioov o SO TITT rook ALLN TH ons gt l a m a MOkOQ VY 1 0 0 0 GOTO x h S RE e e m User lock 1 Delete records VA voltage Timer V Download via ee i USB USB LJ Save Filter in Open records Alarm ANNI Battery Breakdown el mode Burn mode d Ramp test hp Danger HV e Refer to manual Noise detected epek SP Preparations for Use e The battery must be charged between 0 C and 40 C ambient temperature If the battery detects a eno temperature outside this range the battery symbol will Initial instructions eee j Ai Remove instrument power lead and pouch from the packing box Power lead connection table UK International Earth Ground Yellow Green Clip the test lead pouch to the lid Open the lid and familiarise yourself with the layout and position of the IEC 60320 power inlet on the left side of the panel An isolated USB socket is found on the right side of the instrument Unpack leads and pack them into the lead pouch Read the product manual especially the warnings A quick reference is provided in the instrume
19. e the user should ground the Guard lead where possible otherwise parallel leakage paths may occur Alternatively screened leads are available as an optional accessory from Megger When using a screened lead the screen is plugged into the Guard terminal diverting any leakage currents This considerably improves measurements made with a floating output where the leads might touch each other or another object other than the test piece Terminals There are three test terminals marked and GUARD These terminals are designed to accept only genuine Megger test leads Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock with the shutters and are released by rotating the test lead plug by a quarter turn The GUARD terminal as explained below is only used in cases where surface leakage currents need to be eliminated Most measurements use just the and terminals The instrument s internal voltage generator drives the terminal with respect to the terminal current being measured in the terminal GUARD terminal screened leads For basic insulation tests and where there is ittle possibility of surface leakage affecting the measurement it is unnecessary to use the guard terminal i e if the insulator is clean and there are unlikely to be any adverse current paths However in cable testing for example there may be surface leakage paths across the insulation betwee
20. e part from Megger Genuine Megger battery packs must be used Failure to use genuine parts may affect product safety performance and will invalidate your warranty Danger Electric Shock Hazard Removing the lower case to change the battery exposes the AC supply wiring which will be at a hazardous voltage if the equipment is connected to the AC supply Replacement involves removal of four screws from the bottom of the instrument after which the base can be lifted away from the front panel and internal moulded assembly Care should be taken to keep the front panel and moulding assembly together The battery pack is housed within a grey moulded cover secured by four screws On no account must the transparent inner case be opened Hazardous voltages up to 10 kV will be exposed No user serviceable parts are inside Refer servicing to qualified service personnel If there is any damage to the inner transparent casing or its lid then confirm that the Central Rotary Switch is in the OFF position disconnect the old battery and DO NOT connect the new one Refit the bottom cover and contact Megger Instruments for service Instructions 1 Remove the lid switch the central rotary switch to OFF 2 Disconnect the IEC AC power lead and all test leads before inverting the lower case resting the front panel on a soft surface so as not to damage the keypad 3 Remove the four case fixing screws and lift off case bottom 4 Carefully unclip the ba
21. erence standard for step voltage testing is IEEE 95 Step Voltage Test 2002 The SV test is a controlled overvoltage test that can be applied to stator and rotor windings on synchronous and asynchronous AC motors and the armature and field windings on DC motors It is advisable to perform a PI test before an SV test to determine if the insulation is suitable for overvoltage e r o 16 Ramp voltage test The ramp voltage test is an overvoltage test similar to the SV test but with improved control and warning of potential insulation failure The slow continuous voltage ramp is less likely to result in unpredictable damage to the insulation than the rapid step increases employed in SV test If a Pl test was performed to verify the winding s suitability for over voltage testing the winding must be completely discharged before the over voltage test is performed The typical voltage ramp o dV dt is 1 kV min ai which is the default for IR MIT525 MIT1025 This value is user adjustable from the settings function with the mode rotary switch set to ramp Up and down buttons are used to adjust dV dt to the required rate and OK confirms the setting Press and hold TEST to start IR The test will ramp the voltage until it reaches the selected test voltage unless a breakdown or sudden rise in current is detected The result displayed after the test is the final insulation resistance voltage and current If the result is
22. he user the second year being conditional on registration of the product on www megger com Any unauthorised prior repair or adjustment will automatically invalidate the warranty These products contain no repairable parts with the exception of the user replaceable battery and if defective should be returned to your supplier in original packaging or packed so that it is protected from damage during transit Damage in transit is not covered by this warranty and replacement repair is chargeable Calibration Service and Spare Parts For service requirements for Megger Instruments contact Megger Instruments Limited Archcliffe Road Dover Kent CT17 9EN Megger Valley Forge Corporate Centre 2621 Van Buren Avenue Norristown PA 19403 U S A Tel 1 610 676 8579 Fax 1 610 676 8625 England Tel 44 0 1304 502 243 Fax 44 0 1304 207 342 Megger operate fully traceable calibration and repair facilities ensuring your instrument continues to provide the high standard of performance and workmanship you expect These facilities are complemented by a worldwide network of 25 approved repair and calibration companies to offer excellent in service care for your Megger products Returning product to Megger UK amp USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA number must first be obtained from one of the addresses shown above
23. ied voltage steps incrementally by one fifth of the final test voltage each minute for 5 minutes taking DD results can identify excess discharge currents that arise successive measurements when a layer of multi layer insulation is damaged or contaminated a condition that will be missed by both the IR Resistance readings for the first four steps are displayed and PI tests Discharge current will be higher for a given value under consecutive time designators 1m to 4m The 9 minute of voltage and capacitance if an internal layer is damaged reading is displayed by the main display If the default 5 _ The time constant of this individual layer will mismatch the minute test duration is changed by the user the four readings other layers giving rise to a higher value of current than for will not show the respective 1m to 4m indicators insulation that is good in this respect The SV test duration PI Homogenous insulation will have a DD value of 0 while good can be adjusted if ag multi layer insulation will have a value up to 2 The following desired from the 5 i table is a guide to DD test results minute default value R using the up down arrows and OK to save the setting Bad Insulation Condition The step timer will always be set to total test time divided by five Too short a step time may result in incorrect readings and too long a step time may over stress a motor Questionable The ref
24. irectional buttons and an OK button are used in settings and memory functions The up down arrows also enable the test voltage to be adjusted during a test Prior to the start of an IR or IR t test holding down the left arrow button with a voltage level selected on the central rotary switch will activate burn mode Burn mode is deactivated if the voltage range or mode is changed or by pressing the right arrow breakdown button PI DAR DD Instrument controls are simple to operate The central rotary switch incorporates the OFF position The left hand rotary switch selects insulation test type test mode switch The TEST button starts and stops a test Four arrow buttons and OK facilitate adjustment and selection Backlight Save of settings voltages and modes Breakdown burn modes are set using the left and right arrow buttons Backlight and save functions are dedicated buttons All models have backlight and the MIT525 and MIT1025 have the Save button Breakdown vs burn mode In breakdown mode insulation tests are automatically stopped when a fault causes the applied voltage to drop rapidly Burn mode IR tests ignore breakdown and continue to test the insulation and are therefore destructive tests Burn mode is used to purposely create a carbon track in insulation to facilitate fault location Burn mode only operates at test voltages of 500 V and above Voltmeter A voltmeter is incorporated in the instrument and measures A
25. literature for future reference NOTE THE INSTRUMENT MUST ONLY BE OPERATED BY SUITABLY TRAINED AND COMPETENT PERSONS Users of this equipment and or their employers are reminded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits Symbols used on the instrument A Caution refer to user manual AN Caution risk of electric shock fl Equipment protected throughout by Double Insulation e Line Power mains C Equipment complies with current EU directives N13117 Equipment complies with current C tick requirements m Do not dispose of in the normal waste stream L Earth o Universal Serial Bus USB CAT IV 600 V Overvoltage category IV equipment installed at or near the origin of the electrical supply to a building 600 V refers to the rms phase to earth voltage that this instrument can withstand to the overvoltage category IV rating Table of contents and index Generald scripio Nnna oe n si tevbncekannaubessdiaussenmes 5 FEAOG anii E tected 5 Instrument Controls and Indicators cceccecceeee seca eeeeeeeeeees 6 Preparations TOMI SGre ban tarcetucaasiicasceeconas T MIHarNSTUCIONS cae T Power lead and battery charging scceeeseeeeseeeeeeeeeeees T Functional verifi
26. mination DAR is defined as the ratio of insulation resistance at 1 minute divided by insulation resistance at 30 seconds although a 1 minute 15 second DAR is also popular Insulation Condition Acceptable The insulator must first be charged for a sufficient time to be stable i e charging and polarization are complete and the only remaining component of current is leakage current due to the insulation On discharge the capacitive component of the discharge current decays from a high value with a relatively short time constant of a few seconds The released absorption PI IR TIR current decays from a lower value with a relatively long time aaa constant of up to several minutes IEEE standard 43 2000 Recommended Practice for Testing Insulation Resistance for Rotating Machines defines PI as the ratio of insulation resistance at 10 minutes divided by insulation resistance at 1 minute If IRamin gt 5000 MQ the PI may or may not be an indication of The DD timer defaults to 30 minutes of charging which is insulation condition and is therefore not recommended by generally sufficient time for full absorption to take place in an IEEE std 43 insulation material The default test voltage is set to 500 V so Insulation Condition Poor the primary rotary switch must be set at or above 500 V The default DD test duration t1 is 30 minutes insulation test Questionable Oo 1 2 followed by a fixed 1 minute discharge The initial
27. mory capacity 5 2 hours continuous logging every 5 s or 33 logged PI tests or 350 logged IR tests Test regimes MIT515 MIT525 MIT1025 IR IR t DAR PI IR IR t DAR Pl SV DD ramp test Interface USB type B device Real time output USB 1 reading second resistance current and voltage Environmental Conditions Altitude 3000 m test lead CAT rating valid to 2000 m safe working practices must be applied and clips must not be handled until discharge is complete Operating temperature 20 C to 50 C Storage temperature 25 C to 65 C Humidity 90 RH non condensing at 40 C Ingress protection IP65 lid closed IP40 lid open 24 GENERAL SPECIFICATION Safety Meets the requirements of IEC 61010 1 CATIV 600 V to 3000 m Instrument must be operated with all test leads connected above 2000 m EMC Meets the requirements of IEC61326 1 Dimensions L 315 mm x W 285 mm x H 181 mm Weight 4 5 kg Repair and Warranty If the protection of an instrument has been impaired it should not be used but sent for repair by suitably trained and qualified personnel The protection is likely to be impaired if for example the instrument shows visible damage fails to perform the intended measurements has been subjected to prolonged storage under unfavourable conditions or has been exposed to severe transport stresses New instruments are covered by a two year warranty from the date of purchase by t
28. n the bare cable and the external sheathing due to the presence of moisture or dirt Where it is required to remove the effect of this leakage particularly at high testing voltages a bare wire may be bound tightly around the insulation and connected via the third test lead to the guard terminal G The guard terminal is at the same potential as the negative terminal Since the leakage resistance is effectively in parallel with the resistance to be measured the use of the guard causes the current flowing through surface leakage to be diverted from the measuring circuit The instrument therefore reads the leakage of the insulator ignoring leakage across its surface The display will show a warning and fuse terminal symbol if the internal guard terminal fuse has blown The instrument must be switched off to clear the message before further testing is permitted The fuse should be replaced by an authorised repairer The instrument may be used in the mean time if the guard terminal not used Refer to notes regarding measurements above 100 GQ above Leakage current This diagram illustrates GUARD terminal used to prevent surface leakage on cable insulation from affecting a high resistance measurement Screened leads are available for the complete range of insulation testers They are useful in HV switchyards where induced currents are an issue The screen connects to GUARD and prevents induced currents in the lead 20 Preventive
29. nt lid Keep the original packaging for re use Power lead and battery charging If the power lead supplied is not suitable for your line mains connection do not use an adaptor Always use a power lead fitted with the correct plug The ground connection is not used as the instrument is double insulated If using a fused plug ensure that it is fitted witha 3A fuse Supply voltage 90 to 265 V rms ac at 50 60 Hz A green LED illuminates when line power mains is present The battery will charge as long as the mains supply is connected except when a test is in progress For optimum battery life charge the battery after each use Full charge duration is up to 2 hours but a first charge time of 3 hours is advised Functional verification Simply turning on the instrument will initiate a start up process and the display will respond If an error is detected Err will be displayed with an associated error number Calibration The MIT515 MIT525 and MIT1025 are supplied with a calibration certificate which is automatically generated as part of Megger s final test procedure UKAS accredited calibration certificates are available from Megger but this service is chargeable Storage Instruments should be stored in storerooms which meet the storage temperature and humidity specifications listed in this document If charging is incorporated in the storeroom the room must be well ventilated Operating Instructions
30. paths Sharp points at the test lead connections should also be avoided since this will encourage corona discharge The output is isolated and so will float relative to ground such that the positive terminal is at plus half of the test voltage and the negative terminal is at minus half of the test voltage with respect to ground Leakages therefore occur between the positive terminal and ground between the negative terminal and ground and directly between the positive and negative terminals These leakages have a significant effect and can occur through air IO Test V 7 77nn os A Ground If the guard lead is grounded and since the negative terminal is at the same voltage as the guard terminal the leakage into the negative terminal will be considerably reduced This will improve accuracy because the current flowing into the negative terminal is measured by the instrument and used to calculate resistance This technique is only permissible if the item under test is isolated from ground In this context isolated means insulated by a resistance of at least 5 MQ for the positive terminal or at least 10 kQ for the negative terminal Ground Conversely if the positive terminal is grounded then the negative terminal will be at a voltage equal to the test voltage relative to ground which will result in an increase in leakage current and worsening of measurement accuracy 19 When making measurements above 100 GQ therefor
31. result To download all results press an arrow button Detailed instructions on how to interface with PowerDB are available on the product CD on a document entitled Interfacing MIT525 MIT1025 to PowerDB PowerDB offers instructions specific to MIT525 and MIT1025 regarding the download procedure When results are downloaded the IRT can be disconnected from the PC after the application releases the port Deleting results There are two delete functions delete a single result and delete all results Select the bin icon on the central rotary switch The first record indicated contains the result of the last test performed Up down arrows navigate through records and the OK button is used to select delete where the X changes to a tick and the on screen bin icon flashes A subsequent press of the OK button activates the deletion The default delete is a single test result press the right arrow button to select delete all test results from memory Real time output during insulation tests PowerDB or PowerDB Lite can be used to record real time data output from the MIT525 and MIT1025 models Voltage current and resistance data is sent at a rate 1 Hz from the IRT and displayed in real time on a graph e g a plot of current uA versus voltage kV for the ramp test Before running a test where a real time output is required attach a PC running PowerDB Pro PowerDB Advanced or PowerDB Lite via a USB cable Check the produc
32. saved a complete curve of current uA and voltage kV is recorded and can be read into PowerDB PowerDB Lite or converted to a spreadsheet so that the current vs voltage curves can be compared to published curves in IEEE 95 2002 17 Measurement Techniques Understanding Measurement Currents Insulation resistance is defined as the dc test voltage divided by the total current flowing in an insulator The total current has four components capacitive current absorption current conductance current and leakage current 100 I total Relative Current I leakage I capacitive I absorption l l l l l I l l l l l l l l l l l l I l l I l l I conductance l bw 1 5 10 Test Duration min In the case of dry insulation conductance current may be negligible and the leakage current may be low in which case the absorption current will dominate the total current measured 18 Insulation measurements above 100 GQ 100 Relative Current D I total I capacitive I absorption I leakage 5 Test Duration min Measurements up to 100 GQ can be made without any special precautions assuming that the test leads are reasonably clean and dry The guard lead can be used to remove the effects of surface leakage if necessary When measuring resistances above 100 GQ the test leads should not be allowed to touch each other or any other object since this will introduce leakage
33. seeseeees 18 Understanding Measurement Currents 0c0000 18 Insulation measurements above 100 GOQ e 18 ASR AMIINGNS ect aachie se nse cio enlace nto cnnacteuauadtneteccdse abso eedviine 19 GUARD terminal screened leads cccceeeeeeeeeeees 19 Preventative MEASUFEMENL cccccecceceeceeceeseeaeeeeceeeeeaeeaes 21 Cleaning sec then nce hisses once cone hag tinw detinrecnesunladulaiardane eecoines 21 WS ccs scrip rec sre a ce acer A oe oma AERAN E 21 Battery Care ial cid ed cos vatiea a 21 Replacing the battery ccccccccecceeceeeneeeeeneeeeeneeeeeeees 21 Technical Specification siaii naa 23 Electrical SD CIFICATION enra a 23 Environmental Conditions ccccecceceeseeeeeseeeeeeeeseenes 24 GENERAL SPECIFICATION recien 24 Repar and Warnanlys reenienoeie a Ea 24 Calibration Service and Spare Parts ccceceeeeeeeeeees 25 Returning product to Megger UK amp USA service centres 25 Approved Service Centres ccccccecceeeceeeeeeeeeteeeteeeaeees 26 Accessories Equipment and Spares cccccceeeeeeeeeeeees 28 General description The new range of Insulation Resistance Testers IRT are smaller and lighter than previous models yet offer advanced features improved battery life and rapid charge capability The range consists of three models an entry level 5 kV and two fully featured units one 5 kV the other 10 kV Resistance meas
34. st be in good order clean and with no broken or cracked insulation The instrument must not be used if any part of it is damaged Insulation testing in wet weather conditions might be hazardous It is recommended that this instrument is not used in these circumstances If this is unavoidable the user must take all necessary precautions This instrument is not intrinsically safe and must not be used in hazardous atmospheres If this equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired BATTERY WARNINGS Do not disassemble or modify the battery The battery contains safety and protection devices which if damaged may cause the battery to generate heat rupture or ignite Never heat the battery in a fire or otherwise Do not pierce or damage the battery in any way Do not subject the battery to strong impacts shocks Do not expose the battery to water salt water or allow the battery to get wet Never short circuit reverse polarity or disassemble the battery pack In the event of a battery cell leaking do not allow the liquid to come into contact with the skin or eyes If contact has been made wash the affected area with copious amounts of water and seek medical advice Keep cells and batteries out of reach of children Seek medical advice if a cell or battery has been swallowed Do not leave a battery on prolonged charge when not in use e Retain the original product
35. t CD provided with the instrument for a folder named Megger USB If this folder exists use it when starting PowerDB for the first time to find the driver if not allow the operating system to search the internet for the driver Check the serial port allocation on Device Manager and enter the serial port number allocated when starting PowerDB PowerDB offers instructions specific to MIT525 and MIT1025 regarding the real time capture procedure Start the application and activate real time data capture in the form of choice As soon as the test is started real time data output will begin When the test is complete ensure that the form is saved in PowerDB Pro Advanced Lite Battery indicator The battery symbol on the LCD display contains four pairs of segments The battery is monitored continuously when the instrument is turned on The charge remaining in the battery is indicated by segment pairs as follows Fully charged battery THT Ae 50 charged battery MN Tests cannot be started and the battery HHI may fail at any time Symbol flashes when there is not enough charge for a test and the instrument will HIE turn itself off S 7 When mains power is present the indicator shows the battery is being charged by animating the segments of the bar graph A blinking full battery icon indicates that the battery is prevented from charging due to the temperature being out of 13 the allowable charge temperature range
36. tary switch to settings and the mode switch to IR The preset voltage 5000 V will flash and can be changed using the up down buttons When the required maximum voltage is displayed press the OK button to save the setting This setting will remain until it is reset PI DAR DD IR t Whenever V amp is selected the set voltage is shown on the display The voltage lock is useful when for example testing insulation of XLPE cables that should not be tested above 5000 V The lock function will ensure it does not exceed the Vil voltage within the stated output voltage accuracy On test completion insulation capacitance C and the Time Constant TC associated with it is calculated and displayed Time Constant TC Rinsutation X Cinsutation 14 Timed IR test A timed test IR t will automatically terminate ie M an insulation test after a preset time Default timer is set to 1 minute and is IR adjustable within the settings function This is a useful feature which saves the user watching the display for the full duration of the test and the possibility of missing the 1 minute reading IR t Turn the central rotary switch to the settings position Select IR t on the test mode rotary switch The default time of 1 00 minute will flash prompting the user to select a new time using the up down arrow buttons Press OK to set test duration and turn central rotary switch to desired test voltage Press and hold T
37. ttery cable connector leading from the main printed circuit board to the battery and remove the cables from recesses designed to hold them in place Remove the four screws and lift off the battery cover Remove the used battery and replace with a genuine spare battery ordered from Megger ensuring correct orientation of cable exit Route the battery cables via the recesses and clip the battery connector to the printed circuit board battery receptacle ensuring correct orientation Replace the battery cover and secure with the four screws Ensure the alignment of the instrument panel and high voltage moulding then replace the lower case and secure with the retaining screws Take especial care not to leave foreign bodies inside the casing 10 Check and verify instrument operation 22 Technical Specification Electrical specification Voltage input range 85 265 V rms 50 60 Hz 100 VA Battery 11 1 V 5 2 A hour safety rated to IEC 62133 2003 Battery life MIT515 MIT525 Typical capacity is 6 hours continuous 5 kV with a 100 MQ load Typical capacity is 4 5 hours continuous 10 kV with a 100 MQ load MIT1025 Battery charge time 2 5 hours from deep discharge 2 hours normal discharge 30 min charge 1 hour continuous test at 100 MQ 5 kV Test voltages MIT515 MIT525 MIT1025 250 V 500 V 1000 V 2500 V 5000 V 900 V 1000 V 2500 V 5000 V 10000 V User defined test voltage MIT515 MIT525 100 V to 1 kV in 1
38. uide CD ROM Power lead 3 m leadset x 3 medium insulated clips 1002 531 Included Accessories MIT525 MIT1025 USB cable 25970 041 PowerDB Lite software 3 m leadset x 3 large insulated clips MIT1025 only 1002 534 Optional Accessories HV test lead sets 3 m leadset x 3 medium insulated clips 1002 531 10 m leadset x 3 medium insulated clips 1002 532 15 m leadset x 3 medium insulated clips 1002 533 3 m leadset x 3 large insulated clips 1002 534 10 m leadset x 3 large insulated clips 1002 535 15 m leadset x 3 large insulated clips 1002 536 3 m leadset x 3 bare clips 8101 181 8 m leadset x 3 bare clips 8101 182 15 m leadset x 3 bare clips 8101 183 Screened HV test lead sets 3 m 5 kV screened un insulated small clips 6220 835 15 m 5 kV screened un insulated small clips 6311 080 3 m 10 kV screened un insulated small clips 6220 834 10 m 10 kV screened un insulated small clips 6220 861 15 m 10 kV screened un insulated small clips 6220 833 Other CB101 5 kV Calibration Box 6311 077 Calibration Certificate CB101 1000 113 UKAS calibration Certificate CB101 1000 047 26 Megger Limited Archcliffe Road Dover Kent CT17 9EN England T 44 0 1 304 502101 F 44 0 1 304 207342 E uksales megger com Megger 4271 Bronze Way Dallas Texas 75237 1019 USA T 1 800 723 2861 USA ONLY T 1 214 333 3201 F 1 214 331 7399 E ussales megger com Megger Megger GmbH 501 Crystal Paradise Mall Obere Zeil 2 Off Veera Desai Road 61440
39. urement capability is up to 10 TQ for the 5 kV models and 20 TQ for the 10 kV model A key productivity feature is the ability to operate with a discharged battery when line mains power is connected An intelligent battery charger ensures the optimum charge rate as a function of battery level resulting in extended battery life and minimum charge times The tough case provides ultimate protection for a portable instrument and a clip on lead pouch ensures that supplied leads remain with the instrument at all times The case lid is removable for improved lead access IP rating is IP65 with case closed ensuring no moisture ingress High reliability is built in and all models are safety rated to CATIV 600 V Five preset voltage ranges are provided in insulation test mode plus a user settable lock voltage range which is useful for repetitive measurements at specific test voltages All test voltages can be adjusted before and during the test using up down buttons Preconfigured diagnostic tests like Polarisation Index PI Dielectric Absorption Ratio DAR Stepped Voltage SV and a ramp test are available along with basic and timed insulation resistance tests Simplicity of operation is achieved with two rotary switches whereby test voltage range and test mode are self evident The large backlight display enables multiple results to be displayed for example the step voltage test shows five step voltage resistance results simultaneously Adv
Download Pdf Manuals
Related Search
Related Contents
JIRA User Manual - ACS - Advanced Computer Systems LG-A130 Manual do Usuário BT Community Web Kit User Guide Tiptel 142 User Manual POLYMETRON 9126 ILX34-AENWG ProSoft Wireless Designer User Manual Enterprise+ - Manual de Instalação e Operação Samsung ES95 Vartotojo vadovas Copyright © All rights reserved.