733HC-E Steam Sterilizer - Getinge Infection Control
Contents
1. Exp Temp 00 03 13 DAILY COUNTER 7 Parameter Check 73 5 CYCLE COMPLETE AIR IN The door unseals and remains closed ready to open INCOMPLETE AND UNSUCCESSFUL The Silence Alarm button VAN displays Tap it to silence the pulsing alarm The Process Failure status indicator displays and a red diagnostic message displays the reason for the failure When the cycle ends STANDBY displays in the cycle information area The RED Process Incomplete status indicator and the More screen display CYCLE ABORTED prints O 00 03 00 The doors remain sealed Correct the cause of the failure Tap the CLEAR button Then unseal the door 14 24 89 1 Before contacting Getinge USA Service have qualified maintenance personnel check the e steam supply Is pressure adequate strainer clogged e door gasket Is it worn Replace as necessary e chamber Is drain screen clogged screen missing e water supply Is pressure adequate strainer clogged e air filter Is it dirty clogged Replace as necessary e electrical supply Is circuit breaker lit If the problem persists contact your Getinge USA Service representative by e Telephone 800 950 9912 e Facsimile 800 950 2570 e E mail service getingeusa com Getinge USA Inc 1777 East Henrietta Road Rochester New York USA 14623 3133 2012 2013 Copyright Getinge USA Inc
2. GETINGE GETINGE GROUP 733HC E Steam Sterilizer QUICK REFERENCE POSTER Jacket Chamber Cycle Information Status Indicators Pressure Pressure a aT 00 00 vac iagnostic messages Gauge Gauge 00 24 15 01 STANDBY display in blue or red N GETINGE GETINGE GROUP 00 00 Time y Remaining in Cycle Touch Screen u oao e Command Buttons that Operate the Sterilizer CONTROLS Printer OFF ON NE 00 04 00 08 00 12 USB Port CONTROLS i behind panel DISABLE ENABLE Spn Start Cycle Cycle L Login Menu More Open Close on right side with key TURNING THE STERILIZER ON OFF Turn the key RIGHT to ENABLE the controls EFFECTIVENESS OF THE CYCLE Before you unload the sterilizer determine the effectiveness of the cycle The 733HC E Steam Sterilizer is designed to process hospital supplies For detailed operation information read your User Manual Use this poster for quick reference A CONTROLS DISABLE ENABLE O O Press CONTROLS OFF ON to turn the sterilizer ON illuminated or OFF not illuminated CONTROLS OFF ON O O If the button is ILLUMINATED but the display is OFF touch the screen to turn it ON Turn the key LEFT to DISABLE the controls Remove the key to prevent unauthorized operation OPERATING THE DOOR WARNING BURN HAZARD Steam released from the sterilizer chamber can cause serious burns Stand away while the door is moving 1 gt Op
3. en 1 5 Close Tap to open a closed door Tap to unseal a sealed door Tap to close an open door Tap to seal a closed door This status indicator displays when all doors are closed but at least one is not sealed ja This status indicator displays when all doors are sealed SELECTING A PROCESS SCREEN 1 To change the process screen tap the Menu button Screen WARNING RISK OF BURNS INJURY OR NONSTERILE ITEMS This poster is a supplement to the User Manual It does not list detailed instructions or the hazards associated with operating this equip ment For more instructions and warnings see the User Manual PN 61301606344 61301606345 REV B US 2 In the main menu tap the Screen button RISK OF NONSTERILE DEVICE You are responsible for any cycles with settings other than factory set parameters Use biological indicators BI to verify cycle performance RISK OF NONSTERILE DEVICE If the cycle is canceled it must be considered incom plete and the load must be reprocessed unwrapped goods or liquids or repacked and reprocessed wrapped goods only STERILIZING THE LOAD P1 1 Tap the Cycles button to P2 display the list of available Cycles cycles 2 Tap the button labeled with the correct cycle for the load If necessary edit the cycle s parameters password required Check that the Start button is enabled green If not correct any failure condi
4. tions and close all doors 5 Tap the Start button The In Process indicator 4 D Ye displays in the status area while the cycle runs If necessary you can cancel the cycle a Tap the Stop button b In the More screen tap the Abort button Circle Graph Bar Graph Plot Graph Details Graph HOT SURFACES The chamber door and accessories may be hot enough to cause burns Wear personal protective equipment PPE when loading and unloading the chamber POSSIBILITY OF INJURY Persons entering the cham ber could become trapped Before entering the cham ber turn the CONTROLS DISABLE ENABLE key to DISABLE remove it and keep it with you 3 In the Screen menu tap the button for the desired process screen POSSIBILITY OF INJURY If a WATER IN DRAIN message displays DO NOT operate the chamber door Hot water in the chamber may spill if the door is opened Wait until the message clears before attempting to open the door RISK OF INFECTION The Liquids Cycle is not intended for the sterilization of liquids intended for direct patient contact COMPLETE AND SUCCESSFUL The Cycle Complete button displays Tap it to silence the alert sound The GREEN Process Complete status indicator displays CYCLE COMPLETE displays in the cycle information area CYCLE COMPLETE prints below the CYCLE RESULTS PROCESS END 11 15 06 CYCLE RESULTS Exp Temp Max 135 5C Exp Temp Min 135 3C Time at
Download Pdf Manuals
Related Search
Related Contents
SolaWash 37 LED User Manual MANUAL SOFTWARE ZL100FP Manual de Usuario SubDrive con Conectividad WiFi 2 Versión 4.7 - Elipse Software Copyright © All rights reserved.
Failed to retrieve file