Fluke 2700G Series Reference Pressure Gauge

image

Contents

1. m 2700G Series Users Manual Set Sample Rate NISUS LEES PUNCHON LOCK co stikki kann a kisa 9 SUPETVISOTY M de e RR Available Pressure Ranges essei aka annan kaan esent nante nuts NR Hrd How to Set a Custom Engineering Unit or Scale Battery LIfe rfe m Maintenance muss aon M How to Clean the Product esse sere eee eee How to Change the Batteries ACCOSSOMES maiss kaa asta asi ui LE exer ag a a a s Aa lata AR A r SRA ANA ga USB Interface Cable Power Module en annan a aa a aa n aan aa aa nannaa naan annan Serial Interface Instructions Initiating Communication List of Commands Parameter Units jode RD Units Conversion Specifications peel teas Media Compatibility ENVIRONMENT L i nir ina M Mechanical Gpechhications nennen Available Pressure Ranges List of Tables Title Page 23111 870 E em E M 4 WIEN EE ee LEE iii 2700G Series Users Manual List of Figures Figure Title Page 1 niz ess io S 2 Change the ST E 3 USB Power Module and Universal Connectors neee 2700G Series Users Manual vi Introduction The 2700G Series Reference Pressure Gauges the Product are high accuracy digital pres
2. Calibration 2 00G Series Reference Pressure Gauge Users Manual September 2012 2012 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service The warranty period is one year and begins on the date of shipment Parts product repairs and services are warranted for 90 days This warranty extends only to the original buyer or end user customer of a Fluke authorized reseller and does not apply to fuses disposable batteries or to any product which in Fluke s opinion has been misused altered neglected contaminated or damaged by accident or abnormal conditions of operation or handling Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non defective media Fluke does not warrant that software will be error free or operate without interruption Fluke authorized resellers shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of Fluke Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international pr
3. The subsequent list of units is used with the Product psi bar mBar kg cm2 inH204C inH2020C inH2O60F mH204C mH2020C cmH204C cmH200C ftH204C ftH2020C Pressure in pounds per square inch Pressure in bars Pressure in millibars Pressure in kilograms per centimeter squared Pressure in inches of water at 4 C Pressure in inches of water at 20 C Pressure in inches of water at 60 F Pressure in meters of water at 4 C Pressure in meters of water at 20 C Pressure in centimeters of water at 4 C Pressure in centimeters of water at 20 C Pressure in feet of water at 4 C Pressure in feet of water at 20 C Reference Pressure Gauge Error Codes ftH2O60F iNHgOC mmHg0C kpal Far Cel MSW ftw MPA torr Pressure in feet of water at 60 F Pressure in inches of mercury at 0 C Pressure in millimeters of mercury at0 C Pressure in kilopascals Temperature in Fahrenheit Temperature in Celsius Pressure in meters of SeaWater Pressure in feet of SeaWater Pressure in MegaPascal Pressure in Torr mmHG0C Error Codes The gauge uses the subseguent error codes 101 A non numeric entry was received where it should be a numeric entry 102 Too many significant digits entered 103 Invalid units or parameter value received 105 Entry is above the upper limit of the allowable range 106 Entry is below the lower limit of the allowable range 108 A reguired command par
4. EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE OR THEORY Since some countries or states do not allow limitation of the term of an implied warranty or exclusion or limitation of incidental or consequential damages the limitations and exclusions of this warranty may not apply to every buyer If any provision of this Warranty is held invalid or unenforceable by a court or other decision maker of competent jurisdiction such holding will not affect the validity or enforceability of any other provision Fluke Corporation Fluke Europe B V P O Box 9090 P O Box 1186 Everett WA 98206 9090 5602 BD Eindhoven U S A The Netherlands 11 99 Table of Contents Title Page niger How to Contact Fluke Cabratton mee Sinite M Safety Information ax a Special Conditions for Safe Use urra jer tee Display and le WI e How to Setup the Product ene eee Engineering Units ite eee PITO A etie eb GE ipee reden EN SeLAULO ic T nr mate i Show Battery Voltage ia ina sir inkai eeercl Display Actual Temperature enm nennen nnnnl SOE DAM DUNG sez
5. Pressure Ranges Model 2030 2030 2030 2030 2030 2030 2030 2030 2030 Number BG100K BG200K BG700K BG2M BG3 5M BG7M G20M G35M G70M Pressure deg 30 100 300 500 1000 3000 5000 10000 Range psi Pressure Range 0 1 0 2 0 7 2 3 5 7 20 35 70 MPa Vacuum Range psi 15 15 12 12 12 12 0 0 0 Vacuum Range kPa 100 100 80 80 80 80 0 0 0 Burst Pressure 45 90 1000 2000 2000 10000 10000 10000 15000 psi Burst Pressure 0 3 0 6 7 14 14 70 70 70 100 MPa Proof Pressure 30 60 200 600 1000 2000 6000 8000 13000 psi Proof Pressure 0 2 0 4 1 4 4 7 14 40 55 90 MPa 20
6. NPT Connector 2 Pressure Display 3 Engineering Units 4 Bargraph Operation The subsequent sections tell you how to operate the Product Push 9 to turn on the Product The analog bar graph at the bottom of the display shows the applied pressure level relative to the full range of the gauge Note If you record a Tare value the pressure shown is not the actual pressure applied How to Setup the Product Before you use the product it is necessary to configure it for your application Push to go to the Setup menu Each time is pushed the display goes to the subsequent function Push 4 or W to change the parameter value When a parameter is set push to exit the configuration menu or to move to the next parameter Engineering Units The Product s default engineering unit shows psi To change this push 4 and W to move through the standard engineering units plus one custom unit scale When the necessary unit shows push or ES Pressure now shows in the chosen engineering units See the Specifications section for a list of available Reference Pressure Gauge Operation engineering units See the Supervisory Mode section for instructions to set up custom units Set Auto Off Auto Off can be set in 1 minute increments from 1 to 30 minutes or you can turn off the function for continuous Product operation The Product is configured for 30 minutes Push A and W to set the necessary interva
7. ameter was missing 109 An invalid pressure unit was received 117 An unknown command was received 120 The serial input buffer overflowed 121 Too many entries in the command line 122 Pressure module not connected 15 2700G Series Users Manual Units Conversion See Table 3 for units and their conversion factors Table 3 Units Conversion Unit Name Conversion Factor from kPa Conversion Factor to kPa psi 0 14503773773 6 894759 bar 0 01 100 MPa 0 001 1000 kgf cm2 0 010197162130 98 06652 inH2O 4 C 4 014742 0 249082 inH2O 20 C 4 021845 0 248642 inH2O 60 F 4 0185886 0 248844 ftH20 4 C 0 33456183 2 988984 ftH2O 20 C 0 33515375 2 983705 ftH2O 60 F 0 33488238 2 986123 16 Reference Pressure Gauge Units Conversion Table 3 Units Conversion cont Unit Name Conversion Factor from kPa Conversion Factor to kPa cmH20 4 C 10 19744 0 09806383 cmH20 20 C 10 21549 0 09789056 mH20 4 C 0 1019744 9 806383 mH20 20 C 0 1021549 9 789056 kPa 1 1 mbar 10 0 1 inHg 0 C 0 2952998 3 386387 mmHg 0 C 7 500618 0 133322 Torr 7 500618 0 133322 ftSW 0 C 0 325408 3 073062 mSW 0 09918444 10 08222 17 2700G Series Users Manual Specifications Accuracy Positive PressUre ii ni eant ictu e doe 0 02 FS VACUUM mE 0 05 FS Tempera
8. arning identifies conditions and procedures that are dangerous to the user A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test AAWarning To prevent injury only assemble and operate high pressure systems if you know the correct safety procedures High pressure liquids and gases are hazardous and the energy from them can be released without warning To prevent possible electrical shock fire or personal injury Read all safety Information before you use the Product Use the Product only as specified or the protection supplied by the Product can be compromised Do not use the Product around explosive gas vapor or in damp or wet environments Do not use and disable the Product if it is damaged Remove the batteries if the Product is not used for an extended period of time or if stored in temperatures above 50 C If the batteries are not removed battery leakage can damage the Product Reference Pressure Gauge Safety Information Replace the batteries when the low battery indicator shows to prevent incorrect measurements The battery door must be closed and locked before you operate the Product ACaution To prevent possible damage to Product or to equipment under test The display reads OL when the pressure source is above the Product range limit The pressure source must immediately be removed Do not apply more than the maximum to
9. e accessed or changed from its current condition until you go to Supervisory Mode and unlock it Available Pressure Ranges Available pressure ranges are shown in the Specifications section Reference Pressure Gauge Battery Life How to Set a Custom Engineering Unit or Scale The last menu selection in Supervisory Mode is SET FACTR You can set a multiplier factor from 0 001 to 100 to make a custom scale The set factor is multiplied by the psi measured and the result is shown Example 40 psi is the equivalent of 1000 Ibs of product in a tank It is necessary to show the product weight with a 100 psi gauge If you set a factor of 25 40 psi pressure shows as 1000 40 x 25 The engineering unit shown is Cust custom Battery Life Battery life is approximately 75 hours of operation with the backlight off When the battery voltage is low the low battery icon B shows on the top left of the display To replace the batteries see the How to Change the Batteries section Maintenance How to Clean the Product Clean the Product with a soft cloth dampened with water or water and weak soap A Caution To prevent possible damage to the Product do not use solvents or abrasive cleansers A Caution For safe operation and maintenance of the product e Repair the Product before use if the battery leaks e Remove batteries to prevent battery leakage and damage to the Productif it is not used for an extended period e Be sur
10. e parameters above can be turned off to prevent unauthorized configuration changes This is done with password protection in Supervisory Mode Push to access Supervisory Mode or to go back to normal operation Supervisory Mode If necessary each user configurable parameter can be edited when you receive the Product Some parameters are locked and must be unlocked to configure them Use Supervisory mode to do this When you are in the Configure menu and FUnC LOCK is shown it means that there are locked parameters To disable function lock 1 Push W QB 0 PWRD is then shown 2 The password 101 is required to unlock Supervisory mode Push 4 to put in the password entry Hold 4 or W down to move faster through the selections by a factor of 10 When you stop the counter push 4 and W again to move forward or backward by a factor of 1 The password is factory set and cannot be changed 3 Push E From this point each parameter can be locked or unlocked Push 4 and W to select UnLOC or LOC for 8 each parameter To move to the subsequent parameter push ED You can access lock or unlock these functions e Zero function enable disable e Set pressure units enable disable e Auto shutdown adjustment enable disable e Damping settings enable disable e Sample rate setting enable disable e Tare setting enable disable e Custom engineering units set scale factor When a function is locked it cannot b
11. e that the battery polarity is correct to prevent battery leakage e Have an approved technician repair the Product 2700G Series Users Manual How to Change the Batteries AAWarning To prevent possible electrical shock fire or personal injury have an approved technician repair the Product To change the batteries see Figure 2 1 N 29 a s o 10 Pull off the Product cover Use a Phillips screwdriver to loosen the captive screw on the battery door Remove the battery door Replace the three AA batteries Install the battery door again Tighten the captive screw on the battery door Return the Product to its Holster Figure 2 Change the Batteries gsn002 eps Reference Pressure Gauge Accessories Accessories USB Interface Cable The Product includes an USB interface cable The input jack is on the back of the Product You can use serial communication to configure and calibrate the Product and move measurement data from the Product to a PC For specifications on the interface see the Specifications section Power Module The Product includes a USB power module See Figure 3 11 2700G Series Users Manual gsn003 eps Figure 3 USB Power Module and Universal Connectors 12 Reference Pressure Gauge Serial Interface Instructions Serial Interface Instructions The subsequent sections tell you how to set up the Product for serial communication Initiating Com
12. ice Fluke reserves the right to invoice Buyer for importation costs of repair replacement parts when product purchased in one country is submitted for repair in another country Fluke s warranty obligation is limited at Fluke s option to refund of the purchase price free of charge repair or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period To obtain warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assumes no risk for damage in transit Following warranty repair the product will be returned to Buyer transportation prepaid FOB Destination If Fluke determines that failure was caused by neglect misuse contamination alteration accident or abnormal condition of operation or handling including overvoltage failures caused by use outside the product s specified rating or normal wear and tear of mechanical components Fluke will provide an estimate of repair costs and obtain authorization before commencing the work Following repair the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges FOB Shipping Point THIS WARRANTY IS BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES
13. l The off position is at the low end of the selections less than 1 minute Show Battery Voltage Actual battery voltage and a percent of life bargraph show the battery charge No adjustments are made in this parameter Display Actual Temperature The Product is temperature compensated This parameter shows the temperature measured by the internal sensor Push A or W to show degrees F or C Set Damping Selections are on A and off W Damping smooths readings from pulsating pressure sources Set Sample Rate This function finds how many times pressure is sampled and the display is updated Selections are 0 5 1 3 and 10 samples second Note that 10 second gives the fastest response time Set Tare Use this function to set a constant offset value that is then subtracted from the measured pressure For example if a tare is set at 30 psi and the measured pressure is 37 psi 7 psi is shown A pressure of 27 psi is shown as 3 psi Push 4 and W to set the tare value The value is in relation to the engineering units and resolution selected for display Tare value can be set to the maximum range of the gauge For safety the bar graph always shows the actual pressure in relation to the full range of the gauge regardless of the tare position This is done to make sure that with a 0 reading pressure is being applied to the gauge 2700G Series Users Manual Function Lock When set access to each of the settabl
14. munication Terminal communication can be setup with terminal communication Software on a PC The terminal must be set as follows e Bits per second 9600 e Data bits 8 e Parity None e Stop bits 1 e Flow control None e Local echo on List of Commands Use the subsequent list of commands to communicate with the Gauge CLS FAULT IDN TARE TARE PRES UNIT PRES UNIT ZERO MEAS ZERO MEAS MINMAX RST Clears the error gueue Returns an error code from the error gueue Identification guery Returns the manufacturer model number and firmware revision level of the Calibrator Tares the offset pressure of the reading on the calibrator Returns the current tare value Returns the pressure unit for the upper display Sets the pressure unit for the display Zeros pressure of the calibrator Returns the current zero offset value Resets the minimum and maximum recorded values 13 2700G Series Users Manual MIN MAX TEMP CAL_STORE CUST_MULT STREAM_OFF STREAM_ON VAL TEMP_UNIT TEMP_UNIT 14 Returns the minimum recorded value Returns the maximum recorded value Returns temperature in the chosen units Stores calibration data Sets the multiplier for the custom unit type Turns off streaming data Turns on streaming data Returns the measured pressure value in selected units Used to set temperature unit Returns temperature unit Parameter Units
15. on product Do not dispose of this product as unsorted municipal waste Go to Fluke s website for recycling information Reference Pressure Gauge Display and Buttons Display and Buttons The Display and Buttons are shown in Figure 1 The Buttons are in Table 2 2700G REFERENCE PRESSURE GAUGE O Calibra tion E CAL UNITS SET RATE AUTO OFF MIN MAX 189080 Oo s m mm ms mm ms mmes a v EJ EI EB E INTER 300 PSI ON gt gsn001 eps Figure 1 The Product Table 2 Display and Buttons Item Function Push to turn on the Product Push again to turn off the Product ZERO Zeros the display In Configure Mode push the button to move forward through the menus MIN MAX MIN MAX records minimum and maximum pressure values and saves them to memory Push to show maximum MAX indication Push again to show minimum MIN indication After 2 seconds the gauge goes back to live operation To clear the MIN MAX memory values push and hold for 2 seconds until CLr is shown In Configure Mode push Eg V to move backward through the menus 2700G Series Users Manual Table 2 Display and Buttons cont Item Function em Push to go to setup and configuration CONFIG menus Push to make a selection When the Product is not in Configuration mode push to turn on the backlight Push again to turn off the backlight n
16. rque specified Maximum torque specified is 20 Nm 15 ft lb Special Conditions for Safe Use Misuse If the Product is exposed to overpressure or sudden physical shock such as being dropped examine it for damage that can cause a safety issue If necessary send the Product for examination to Fluke Refer to the How to Contact Fluke Calibration section A Warning To prevent possible fire or personal injury e Donotuse the Product with flammable substances e The Product is intended for installation only in locations with protection against the entry of solid unwanted objects or water that can compromise safety 2700G Series Users Manual Symbols Symbols used on the Product and in this manual are in Table 1 Table 1 Symbols Symbol A Meaning Risk of danger Important information See manual Symbol CE Meaning Conforms to European Union directives A Hazardous voltage Risk of electrical shock G Conforms to relevant North American Safety Standards N10140 Conforms to relevant Australian standards lx This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE Directive Annex this product is classed as category 9 Monitoring and Control Instrumentati
17. sure test gauges Accurate to 0 02 FS the Product can be used as a calibration reference or in applications where high accuracy pressure measurement is required The Product features user configurable functions that include e Sampling rate e Tare e Damping e Auto off e Min Max When the Product is configured you can lock its settings and use password protection to prevent configuration changes See the Supervisory Mode section How to Contact Fluke Calibration To contact Fluke Calibration call one of the following telephone numbers e Technical Support USA 1 877 355 3225 e Calibration Repair USA 1 877 355 3225 e Canada 1 800 36 FLUKE 1 800 363 5853 e Europe 31 40 2675 200 e Japan 81 3 6714 3114 e Singapore 65 6799 5566 e China 86 400 810 3435 e Brazil 55 11 3759 7600 e Anywhere in the world 1 425 446 6110 To see product information and download the latest manual supplements visit Fluke Calibration s website at www flukecal com To register your product visit http flukecal com register product 2700G Series Users Manual Standard Equipment The Product ships with e Protective Cover installed e Three AA Alkaline Batteries installed e Safety Information printed e Report of Calibration e Manuals CD ROM with translated Users Manuals e USB Cable e USB Power Adapter e NPT to 74 BSP Male Adapter e NPT to M20 x 1 5 Male Adapter Safety Information A W
18. ture Compensation sse 18 C to 28 C 65 F to 82 F to rated accuracy Note For temperatures from 0 C to 18 C and 28 C to 50 C add 003 FS C Media Compatibility 15 30 e aceeseraeten eae any clean dry non corrosive gas 100 300 500 1000 pS 1 cise ed eli ele eee any liquids or gases compatible with 316 stainless steel Above 1000 EE any non flammable non toxic non explosive non oxidizing liquid or gas compatible with 316 stainless steel Environmental Operating Temperature em 0 C to 50 C 32 F to 122 F Slorage pe eiie e hinta apa cere st 20 C to 70 C 4 F to 158 F Humidity TED 10 96 to 90 RH Non condensing AU 2000 meters 6561 68 ft Pollution Degree 2 2 D e diced sn hang ERR e ne Eed 2 Agency APprovalS kanada akan cc 2 8 18 Reference Pressure Gauge Specifications Mechanical Specifications DIMENSIONS k e 11 4 x 12 7 cm depth 3 7 cm 4 5 x 5 in depth 1 5 in Without Protective Cover Pressure Soles mtem in NPT Male M SINGI ec P Cast ZNAL DISPLAY siss va 5 1 2 Digits 16 53 mm 0 65 in high 20 Segment bar graph 0 to 100 Power Battery Me T three size AA alkaline batteries cruce 75 hours typical without backlight 19 2700G Series Users Manual Available