1. Note 1 Pass level for CLASS 1 is 5 mA Pass level for CLASS 2 is 1 mA Note 2 Leakage test may be performed if for any reason a standard CLASS 1 or CLASS 2 is not possible RCD Test gt Trip Current 2 to 500mA in 1mA steps User selectable gt Trip Time 0 to 3 000ms at 001sec resolution gt Current Ramp Trip Test 0 to 500mA in 1mA increments Power Measurement Power test available in 1 phase only Leakage Current 0 to 30 0mA at 200 to 265V gt Load Current 0 0 to 20Amp 1 Phase Socket gt Voltage 200 to 265VAC gt Power Factor 0 00 to 1 00 gt Apparent Power 0 to 4800VA gt Power 0 to 4800W 33 Wavecom Instruments 3PLM User Manual WinPATS Professional Asset Management Software Full Version of WinPATS Software is included with models 3PLM WinPATS must be installed on a Laptop or PC from the Autorun menu on the CD provided The USB USB Driver supplied on the CD must also be installed Brief Installation steps Firstly make sure you have registered WinPATS and have the correct REGISTRATION KEY Install WinPATS Install USB USB Driver Locate Interface panel under Keyboard USB PC socket R H side TE Connect from USB Port on 3PLM to Laptop or PC USB Check Device Manual for Comms Port Setting See full WinPATS Manual for full details Go to file option in WinPATS set Comm Port to same as the Silabs Driver Comm Port setting for communication to PC Laptop You now have the ab
2. e Wall GPO a Main Switchboard with RCD 3PLM Tester To do this test select the second menu press and release the Return Enter button This menu will now display 3 functions Main Menu E Power Test To enter the RGD menu press and release the F3 button RCD TESTS Time Test Ramp Test Press F1 to select the time test Trip Time Testing This principal is designed to trip RCD devices at a fixed current and to determine the trip time of the RCD device This function is factory set to 30mA for fast testing the user can set the current to X0 5 X1 0 X 5 using the RCD Multiplier LE 30mA X 0 5 15mA 24 Wavecom Instruments 3PLM User Manual 30mA X 1 0 30mA this also is effective on any set test current of the RCD tester from 5mA to max 500mA output 30mA X 5 0 150mA These tests should result in no trip trip amp fast trip times respectively F1 0 deg This is the positive half of the mains supply cycle 50HZ Aust NZ Press F1 The preset mA test current will then be used in the following test and begin the trip test from the positive half of the sine wave F2 180 deg This is the negative half of the mains cycle 50HZ Aus NZ Press F2 The preset mA test will then be used in the following test and begin the trip test in the negative half of the sine wave Performing a Time Test Timed Test 3S mi O phase 160 phase settings The displayed trip time is
3. First Test VISUAL INSPECTION IS TO OCCUR BEFORE ANY OTHER TEST IS CARRIED OUT BEFORE USING ANY OF THE RANGE OF 3PLM APPLIANCE TESTERS e There is no damage or component defects to the accessories plugs outlet sockets or connectors physical e There are no cracks amp or abrasions e There are no exposed inner cores or conductors flexible and the supply cords are not twisted or distorted e Any Fuse Over load protection components if fitted are checked e All labels markings and warning indicators of the maximum load to be connected to the device are legible and intact e The insulation is not damaged in any way i e melted cuts or abrasions There are no iron filings in the insulation There is no insulation tape on the lead Any flexible cords and or leads are effectively anchored glands and grommets intact All covers or guards are in place and secure as intended by the supplier manufacturer All safety devices and systems are in good working order i e overload latches amp buttons No dust amp or dirt obstructs any exhausts or ventilation outlets All controls are working properly and are secure and aligned Important If result is a FAIL If any Equipment FAILS ANY of the above it should be deemed to have FAILED the Visual Test and therefore no other tests need be performed lf this is the case the Equipment should be tagged with a DANGER TAG and removed from service It is recommended by the manufactu
4. Earth Bond Current Selection In all 3PLM models the tester has the ability to adjust the earth bond current amps This can be selected between 0 2 1 2 5 10 15 20 25 Amps e Press key to decrease the current e Press the key to increase the current e Press the to select the displayed current Label Options Only on Selected Memory Models TnT Options Label Options Changing the label options changes the way the printer behaves or how the tag is printed When selecting this option you will be prompted with 3 different options Enable Disable Choosing a label set and modify the due date Enable Disable SET LABEL PRINT FOR TESTS DISABLE EHLELE selecting this option will enable or disable the tag information being sent to the wavecom printer The printer can also be disabled by switching the printer off from the power switch located on the side of the printer Once the printer is switched back on it will not print out the records that were created while the printer was switched Off e Press the e Press the Goren key to enable key to disable a f w 30 Wavecom Instruments 3PLM User Manual Choose Label Set LABELO If you have more than one logo in the printers memory multiple company logos this function will allow you to select the various logos If you have one logo stored in the printer then you only need to choose Label 1 If you choose anything else a blank tag will
5. If the kKeypad is present then the new barcode screen will be loaded Powering on the Tester When the tester is powered on the scanner will beep a few times indicating the scanner is also powered Initially the printer s status light will be orange but after about 5 10 seconds it will change to green The tester will not print unless the printer light is solid green A flashing red status light can indicate a media or general printer fault Note 1 If the 3PLM stays locked on the logo screen please check the connectivity of keyboard amp scanner cables The 3PLM unit will not work with scanner only connected The 3PLM system requires either Keyboard only or both Keyboard amp Scanner to be connected Note 2 If no devices are connected to the interface the unit will function as a 3PLM basic tester only Adding New Equipment The following must be done before testing e Create a site or use and existing site e Create a user name to under the users list e lf auto generate number is used the initial number must be set correctly e Ifyou are adding equipment to an existing site please make sure that the correct site is displayed If the wrong site is displayed the equipment will be assigned to the wrong site When the appliance testers memory has been erased it will look like the following screen lt F1 SELECT SITE New Barcode EPS When the initial screen in memory mode operation is displayed A site and user oooooda sooo
6. an RCD in the main switchboard upstream it s most likely to trip this upstream RCD When performing RCD trip time or Ramp current tests on any portable RCD devices the RCD in the switchboard may trip faster This is due to increased upstream levels of leakage current from the additional circuits and devices connected to it The fixed RCD s can also have better connectivity sensitivity and mechanical mechanisms To avoid tripping large areas in the work place monitored by the switchboard RCD it is suggested that an RCD Isolation Transformer be used These are designed specifically for the purposes of field RCD tripping DO NOT use these Transformers for any other purpose Ratings 240VAC in 240VAC Out 30VA Fuse protected Primary winding 500mA To perform RCD tests the 3PLM supply lead needs to be plugged in to the RCD device to be trip tested For tripping switchboard mounted Fixed RCDs Plug 3PLM into GPO marked RCD protected or if known to be protected circuit Diagram 2 Isolation Transformer Required If testing portable RCD devices on power boards or extension leads plug 3PLM into power board or lead Isolation Transformer Required Diagram 1 me f Diagram 1 For Testing Portable RCD Wavecom Isolation Transformer Wall GPO Maybe RCD Protected Portable RCD Example May Vary Wavecom Instruments 3PLM User Manual Diagram 2 For Testing Fixed RCD at Switchboard 3PLM Power Lead Y
7. be created If you have multiple logos then the next one to choose would be label1 and so on If you are unsure which one to choose select the Label 0 e Read the onscreen message and then press the key e Press the key to increase the label number e Press the key to decrease the labeknumber Press the owen key to select the displayed label number CEE Customize Next Test Due Dates USE CUSTOM TEST DUE DATES WHEN TESTING ENABLE DISABLE Enabling this feature will allow customize next test due dates when testing equipment When this feature is disabled default setting the next test due date is determined by the test date plus the frequency of test When enabled a custom date can be set after visual test in 3PLM models e Press the GC key to enable Setup printer Darkness Please leave this at default setting if unsure this only needs be changed if the print quality has degraded due to printer head wear or if using different label media eg UV rated tags that need higher head temp rating SET PRINTER DARENESS O DECREASE INCREASE Press the key to disable 31 Wavecom Instruments 3PLM User Manual Stop Labels Spooling On Power On STOF PRINTER SPOOLING ON a LAaRTOP TES CANCEL lf the printer is spooling tags on power on then it has reverted back to the factory default settings To stop this choose the F3 key when prompted The printer must be switched on and have a solid green light If not t
8. be no numbers entered in to switch between the two modes Wew Barcode OOO00000 SOOO Loc Desc Make Loc Desc Make OFFICE LEAD HPM A selection of details can be entered for an item Details include location description make model serial number asset number frequency of test and notes Each heading relates to an F e g Loc F1 Desc F2 and Make F3 Only 3 items are displayed at one time To scroll through the various details press enter key By 0 Tests continually pressing enter you can loop back to the first 3 items Please make sure location description are filled out and check to see that the frequency is correct Once the details are entered hold the enter key for 2 seconds to move on to the visual check screen Model SN isset Freg Motes Tests 1 Years 11 Wavecom Instruments 3PLM User Manual ALPHA MODE Entering an items details is the same process as entering sites and users please refer to the Adding a Site User steps above Ensure that test frequency for the item is set in accordance with the AS NZ 3760 Standard After the item has been visually inspected there is 1 of 3 options F2 is VISUAL CHECK Other Status F3 is Fail Enter is Pass if F2 is selected then under a Other Status You have the following choices F2 New to Service and 30 03 2015 17 46 l ag F3 Out of Service If Out of Service or Fail is selected the result will be OTHER STATUS FAIL PASS saved and the 3PL
9. cool This requires the 3PLM device to be best left unplugged for several minutes allowing unit to cool 3PLM will display results for 5 seconds after mains supply is tripped If the power is not reset by the time the unit loses power then the result will be displayed on power on A Caution Continuous ramp current testing will cause unit to over heat 27 Wavecom Instruments 3PLM User Manual 3PLM Options Menu Introduction Here you can adjust the way the 3PLM behaves under certain tests and other functions Some 3PLM options may not be available to some 3PLM versions To access the 3PLM options menu TuT Options From either the Enter barcode screen memory unit or the main menu A screen hold down the enter button and then press the F2 key You should now be in the 3PLM options screen The screen displays 1 option at a time e You can select this option by pressing the enter button e To select the next item press the F2 button e Pressing the F3 button escapes out of the 3PLM options menu The 3PLM Memory model options list Change Insulation Voltage Leakage Test Time Audio Option 3PLM Earth Bond Current Selection Label Options Barcode Options Change Insulation Voltage All 3PLM models Some DUT contain MOV s metal oxide varistor or commonly knowneas spike protection devices These limit the mains voltage to about 280VAC If you attempt to perform a 500VDC insulation test on one of these
10. devices you will get a failed result To fix this problem select the 250V option INSULATION TEST 250 DC 500 VDC 1000 VDC e Press the Gs key to select the 250V option e Press the key to select 500V option You can also set the unit back to 500V by switching the mains power off and on again The default insulation voltage is 500V 28 Wavecom Instruments 3PLM User Manual Leakage Test Time TaT Optians LELELGE DURATION Leak Test Time Up 205 Mext Down Esc ESC The leakage test time is USER adjustable You may need to adjust timing in this feature if the DUT has a long power up time or requires special power on procedure Default time for Leakage Test is 20 seconds From this menu you can change the test time from 5 seconds to 28800 seconds Adjusting to any less than 5 seconds will put the test time to infinite e Press the key to increase the test time e Press the key to decrease the time e Press the key to set the new leakage power test time e Press the Goren key to escape out of the menu No changes saved CEE Audio Option All 3PLM models SET AUDIO BUZZER FOR TESTS OR TD ISAALE ENABLE Enabling the audio option will make the 3PLM use a sound to indicate a pass or a fail at the end of a test If the 3PLM unit has never been altered then the audio beep is disabled e Press the GC key to disable e Press the key to enable 29 Wavecom Instruments 3PLM User Manual 3PLM
11. in milliseconds This is the time taken for the RCD device to trip once the injected fault current has been applied The 3PLM injects a true fault current value using a real time compensation calculation of the actual voltage at the time of test hence delivering a true and accurate trip current RCD Test Options Options Set Current met RCD type F3 Change to select tested This allows the user to set the trip current level 5mA to 500mA The RCD type can also be select here depending whether the unit is a type or type Il See the next section for explanation From the options menu press F2 to change the current level and F3 to change the RCD type 25 Wavecom Instruments 3PLM User Manual Adjusting the current level RCD TEST CURRENT Up 3 Orn Down SET ESC The 3PLM displays and maintains the last set trip current value lf the user wishes to change the value of the trip current the following steps enable the changes Press and release F2 from the options section to display test current Up This button raises the trip current in 1mA increments to 500mA Hold the button and the value will scroll faster the longer depressed Once 500mA limit is reached the value will then loop over and start again from OmA Down This button decreases the trip current in 5mA increments Hold the button and the value will scroll faster the longer depressed Once OmA limit is reached the value will then loop over and start agai
12. must be entered before testing can be done When using the 3PLM for the first time lt F1 Select Site will be flashing in the top region of the screen lt F1l SELECT SITE New Barcode OOO0000 S000 SITES 1 Edit Site 1 ALES oh WAVE COM VAVECOM New Barcode OOO0000 S000 VAVECOM New Barcode OOO00000 S000 USERS 1 3 Edit User 1 GEORGE USERS 1 3 GEORGE Wavecom Instruments 3PLM User Manual Adding a Site From the initial screen press F1 Don t hold the F key this will select a different function The 3PLM can hold up to 16 sites To scroll through the sites press enter 3 sites will be shown on each screen Each site corresponds to an F key for example Site 1 F1 To edit a site hold the corresponding F key for 2 seconds If the Site has a symbol next to the name then the site name cannot be altered This site was uploaded from WinPATS and needs to remain the same Using the keypad enter the name of the site The Keypad operates in 3 modes Alpha Numeric and Predictive When an editable screen is entered there will be an ALP NUM or PRE in the top right hand corner of the screen Leave the text on ALP mode Once the details have been typed in press enter to return Note Any reference in this instruction manual to keypad should be replaced by qwerty keyboard Once a site has been entered the F buttons are used to select the site For example to select site 1 F1 would be pres
13. or incorrect The 3PLM will determine the power source by the plug attached to the mains power supply in 3PLM front diagram Wavecom Instruments 3PLM User Manual Menu System Main Menu SPH Earth Test Main Menu A Main Menu E Earthed Test Power Test DY Insulated Test Leakage Test Ext Lead Test RCD Test 3PH Ex Lead Test 3PH Leakage Test The menu system is divided into 3 menus Menu 1 amp 2 are related to single phase testing Menu 3 is related to 3 phase testing When performing 1 phase test only the 1 phase DUT socket will be used In menu 3 only the 3 phase DUT socket will be used The 3PLM has an internal contactor that switches between the single and 3 phase outlet You must select the correct type of test to select the correct outlet For example if a DUT is plugged into the single phase outlet and a 3 phase test is chosen then a NCNT message will appear This is because the 3PLM has selected the 3 phase outlet and is not detecting any appliance Certain test will be available depending on what power source the 3PLM is plugged into The 3 different test modes are 1 Phase 3 Wire All tests except 3 Phase leakage test 3 Phase 4 Wire All tests except single phase power test 3 Phase 5 Wire All tests are available These modes are displayed when the 3PLM is first powered on If a test is not available based on the type of power input then a message will be displayed on the screen Other 3PLM Features e
14. M will revert back to the barcode entry screen If you have a wavecom printer a tag will be printed If the test is a visual check only then press enter hold See below for details Main Menu Earthed Test If the item passes a visual inspection and pass is selected testing can begin Select the correct test based on the appliance type D Insulated Test Ext Lead Test EARTH BOND TEST Pass lt 1 0 Ohms 0 0 Ohms Pass 15 Once test has been completed the results are displayed INSULATION TEST sOOVDC 1 Homes 38 To test the item multiple times press F3 This will take you back to WHAT NEXT the main menu A screen From here you can choose a test class You would use this for example if testing power boards To complete the test press enter This will take you back to the new barcode screen lf you have a wavecom printer a barcode will be printed once enter is pressed Visual Check Only Option VISUAL CHECK 30 03 2015 17 46 Items can be passed with a visual check only no testing required OTHER STATUS You must be atthe visual check screen shown on the left To do a visual check only press and hold enter for 2 seconds FAIL Pasa VISUAL CHECK ONLY SAWE AND EXIT YES HO Pressing enter will take you back to the visual check screen Pressing F3 will complete the test Print out a tag if you have a wavecom printer Then go back to the new barcode screen 12 Wavecom Instruments 3PLM
15. The 3PLM contains two fuse holders show in side view diagram These are used to protect the RCD circuitry of the 3PLM Only replace these fuses with M205 1A ceramic slow blow fuses Under normal condition these fuse should never have to be replaced e An RS 232 port is located in the under the keyboard tray This is for connecting to the Wavecom printer VIA a serial cable e USB ancserial socket is provided to connect a scanner and keyboard e A USB port that connects to the PC for uploading downloading records and general testing e 3PLM allows up to 1000VDC insulation tests The default is always 500VDC If the voltage needs to be changed then please read the change in the 3PLM options section Wavecom Instruments 3PLM User Manual 3PLM Test n Print Unit 3PLM 20A amp 3PLM 32A Introduction to Memory functions The 3PLM units have Memory There is inbuilt flash memory and a real time clock This gives the ability to store up to 5 000 records which can include items from 16 different sites and 6 different users This will help to increase the efficiency and pace of testing The 3PLM units with memory have two operating modes e 3PLM tester mode only no devices attached e Test and Tag mode all devices connected amp on allowing testing scanning retrieval amp printing The 3PLM mode is determined if a keyboard is connected If there is no keypad detected then the 3PLM will load the main menu screen and function as a standard unit
16. User Manual Retesting Equipment The following must be done before retesting e A site must exist in the tester with records A site is usually uploaded from WinPATS e The correct site must be displayed on the first line of the new search barcode screen If the incorrect site is displayed no equipment will be reported when scanned To change see the add sites section The correct user must be selected from the users list A user is important because it give a reference to the test results in WinPATS It is also the user that is displayed on the tag If there is no users create a user as described in the add equipment section or upload the users list from WinPATS WAVECOIM Search Barcode OOO0001 ggg Loc Desc Make OFFICE LEAD HPM VISUAL CHECK 30 03 2015 17 49 OTHER STATUS FAIL Main Menu A Earthed Test D Insulated Test Ext Lead Test EARTH BOND TEST Pass lt 1 0 Ohms 0 0 Ohms Pass 13 INSULATION TEST SOOO 1 Homes 38 WHAT MEXT Pasa Type in or scan the barcode number to be searched and press enter If the screen displays new barcode press enter to switch over The screen can only be switched if there is no entry The existing record details will be shown Any changes made here will be updated once the site is downloaded in WinPATS If nothing needs changing hold the enter key to proceed to the next screen After all item details have been entered hold the enter key for 2 seconds to proceed to t
17. V in this these situations can cause the surge protection devices to trip therefore conducting the applied voltage to earth thus showing a failure of insulation In these instances the test voltage should be changed to 250V then retest If DUT still fails check with the DUT Operators Manual or an electrician for details see Double Insulation Test 250 500VDC to change test voltage Under these circumstances it would be difficult for any damage to occur to either the surge protection device or the DUT as there is insufficient current generated by the 3PLM test unit Leakage Test 15 Wavecom Instruments 3PLM User Manual lf there are any doubts with insulation testing of the equipment the standard AS NZ3760 since 2001 allows for an alternative test method A Leakage Test can be performed instead The 3PLM Series are designed to perform these tests NOTE 20Amps MAXIMUM Resistive Load only Standard 3PLM 20A units 32Amps MAXIMUM Resistive Load only Standard 3PLM 32A units A Leakage Test applies power to the Device Under Test DUT and measures the imbalance of leakage current from the DUT between the active and neutral conductors The leakage is tested to the limits specified in the standard and a Pass Fail result as well as a digital reading is provided to ensure that the user gains as much information as necessary Earth Continuity Test commonly called EARTH BOND TEST Note Part of normal Class 1 test proc
18. WAVECOM INSTRUMENTS ov PTY LTD 3PLM User Manual 3PLM 204mp Model 3P LM 32Amp Model Australian Designed amp Manufactured to comply with testing requirements of AS NZS 3760 2010 Standards Note if required 3PLM Firmware can be updated by Wavecom only to comply with potential Changes in Testing Standards Head Office Wavecom Instruments Pty Ltd Western Australia Melbourne Office 257 Grange Road Findon SA 5023 Unit 2 17 Casino Street Welshpool WA 6106 772A Station Street Box Hill Victoria 3128 Phone 61 08 8243 3500 Phone 61 08 9353 1943 Phone 1300 793 301 Fax 08 8243 3501 Fax 61 08 9353 4319 Fax 61 03 9897 4766 Email sales wavecom com au Email saleswa wavecom com au Email salesvic wavecom com au Web www wavecom com au Web www wavecom com au Web www wavecomrentals com au Wavecom Instruments 3PLM User Manual coM ae O gt INSTRUMENTS Wavecom Instruments 3PLM User Manual Table of Contents Important Information Contents 3PLM Diagram Plug Adaptor and Cord Information Manufacturer Recommendations Safety Warning Competent Person Disclaimer Limited warranty Fully Integrated 3PLM Unit Powering on the Tester Menu System Adding new Site Auto generate barcode options Entering items Retesting Equipment Deleting Equipment Technical Information Class 1 Earthed Appliance Construction Class 2 Double Insulated Construction Testing of Elec
19. ame plate details of operating voltage operating current and power factor etc should the appliance exceed the said values on the name plate it could be deemed faulty and require service Please note Because the power test is not required in the electrical testing standards there is no pass fail value in the 3PLM It is up to the user to determine if the item is a pass or a fail based on the compliance name plate If the result was a pass Tag with the appropriate tag including next test due date and return to service If the result was a fail Tag with a DANGER TAG and remove the unit from service The Power Test allows the user to turn the appliance on and measure its performance as a digital wattmeter Displayed parameters are Volts AC Current Volt Amp Power Factor Watts A Caution Before operating ensure the equipment is firmly secured to eliminate the possibility of causing injury or damage This function turns powers the Unit on Please make sure the drills or any blades of any rotating devices have been removed prior to any testing and have clearance from body or object Appliances may jump off benches and may need to be secured by clamping them or they may cause serious damage if not secured properly 22 Wavecom Instruments 3PLM User Manual RCD Test Main menu B Main Menu E Power Test Leakage Test RCD Test Please Note Where ANY RCD testing is to be carried out any circuit that is protected by
20. ch all sites is enabled If a barcode is found in another site the site name will change when you return the new barcode screen SEARCH ALL SITES JJ ALLOW ADVANCED ON BARCODE SEARCH METHODS SEARCH ON RECORDS Enable Disable ENABLE DISABLE 32 Wavecom Instruments 3PLM User Manual Other Features Barcode Printing Function Only on Selected Memory Models This utility will print out barcode labels from what is typed on the screen The idea is to save time by storing these barcodes in a book or collection From here you can quickly scan a barcode without having to retype the text back into the 3PLM unit To access this feature e Hold the G key from the new barcode search barcode screen Type Barcode Type Barcode 1234567 Exit Print You will be required to select your printer model The Wavecom printer has a different print orientation to the Wavecom model Selecting the wrong printer model will make the tag print incorrectly e Press the key if you have a Wavecom or plus model printer e Press the key if you have a Wavecom model printer OO Type the text you want for your barcode If you go past the end of the screen the text will wrap A maximum of 32 characters can be entered e Press the key to print the barcode e Press the xren key to return back to the new barcode screen 0O 33 Wavecom Instruments 3PLM User Manual Meter Mode Independent of all other Tests Meter M
21. ed tested and tagged correctly regulations require that a competent person such as a Licensed Electrician be employed to perform the required tests Please refer to the above definition as described in the current AS NZ 3760 2010 Standard and in addition to any other local legislation or jurisdictions as may be relevant in your State EXAMPLE A person competent to undertake Inspection and Testing of electrical equipment must have Knowledge and practical experience of electricity and its hazards A clear understanding of precautions to avoid danger The ability to recognise at all times whether or not it is safe for work to continue The ability to carry out visual examinations of electrical equipment The ability to distinguish between electrical equipment that is double insulated and equipment that is earthed as well as being able to identify the appropriate test for each type e The competency to safely carry out the Earthing Continuity Insulation Resistance or Leakage Test and RCD tests on electrical equipment e The knowledge of how to use the relevant testing instruments interpret and record the results for compliance with the Standard Workplace requirements e The knowledge to be able to correctly recommend the frequency of testing required Due to the potential hazards of electrical testing due care must be taken at all times Wavecom Instruments 3PLM User Manual Disclaimer Limited Warranty The Manufactur
22. edure Use supplied IEC Croc clip lead or optional accessory WCM Probekit To Change Test Voltage To change the test voltages select the change ins from the options menu Please refer to the special functions section of this manual Please Note The 3PLM Series wilkautomatically default to 500VDC when restarted 16 Wavecom Instruments 3PLM User Manual Integrated Tests Supply Mains Test The Supply Mains Test checks the polarity and connectivity of the mains supply by LED s This test is also a part of all the testing functions of this unit If the N E red light is on and you need to conduct load leakage tests DO NOT CONTINUE If you are carrying out standard Insulation and Earth Bond tests it is generally safe to continue This light will glow if a voltage difference lies between the neutral and the earth or if no earth is connected to the 3PLM Series supply If working with a generator or inverter this is most likely to occur and you may need to consult an electrician before proceeding If both the A E amp A N green lights are on but not the N E red mains supply test is ok continue to test If both of the N E red amp A E green lights are on consult an Electrician as there is a fault with the Mains Supply NCNT Test No Connection No Test The 3PLM Series appliance testers ensure that the appliance is plugged in and switched on This test is also a part of all of the testing functio
23. ents 3PLM User Manual If the result was a pass Tag with PASS tag showing next test due date and return the device to service If the result was a fail Tag the DUT with a DANGER tag and remove the device from service Please Note Ensure that the device is isolated from any ground loop m Class 2 Test Double Insulation Test F2 Please Note A Visual Inspection test must be carried out before any others refer to First Test Section The Double Insulation Test completes the following test sequences as part of its procedure 1 Integrated Supply Mains Test Refer to Integrated Test section 2 Integrated NCNT Test Refer to Integrated Test section 3 Double Insulation Test 250V or 500V pass level greater than 1MQ Procedure p lt Complete a Visual Inspection Note Only proceed if test passed Plug the appliance into appliance test socket Connect earth clip to any exposed metal on the device if any or device can be wrapped in foil or use metal mesh braid Part No WCM ES 500 Press the F2 key and wait for results ensure the device is powered on Read and record results appropriately Unplug device Tester will return to Main menu A automatically aN dis If the result was a pass Tag with PASS tag showing next test due date and return the device to service If the result was a fail Tag with a DANGER tag and remove the device from service In some situations if the device is labelled with Surge P
24. er warrants its products against defects in materials and workmanship for a period of 12 months from the date of purchase During the warranty period the manufacturer will repair or at its option replace at no charge the product that proves to be defective This warranty does not apply if the product has been damaged by accident abuse misuse or mis application or as a result of service or modification by anyone other than manufacturer of the 3PLM appliance tester The 3PLM product range of devices or its manufacturer IS NOT RESPONSIBLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM THE BREACH OF ANY EXPRESS OR IMPLIED WARRANTY INCLUDING DAMAGE TO PROPERTY AND TO THE EXTENT PERMITTED BY LAW and DAMAGES FOR PERSONAL INJURY The Distributors of this product cannot assume liability or responsibility for any loss or damage resulting from the use of this device The 3PLM manufacturer reserves the right to discontinue models change specification price or design at any time without notice or obligation Power Up Checking When the 3PLM is turned on it will determine the type power source This is either 1 Phase Detected 3 Phase Detected 3 Phase Detected 3 Wire Detected 4 Wire Detected 5 Wire Detected 1 Phase 3 Wire 3 Phase 4 Wire No Neutral wire 3 Phase 5 Wire With Neutral wire If the power source is 3 phase the tester will do an additional phase rotation test The test will indicate that the phase rotation is correct
25. f readings on a 12 monthly basis or as prescribed under any additional local Regulatory requirements Note Only Wavecom Instruments or its Authorised Service Agents are permitted to repair and Calibrate this Instrument Failure to perform unauthorised service or repair may void all warranties and Calibration Status Safety Warning AN The 3PLM series of products have been designed to meet stringent safety requirements however no device can completely protect persons from the consequences of incorrect use The testing of Electrical appliances requires that extra care and caution is taken at all times to ensure personal safety The Manufacturer also advises that appliance testing should be conducted by a Competent and suitably trained person as referred to under the current Standard AS3760 2010 as well as any additional legislation or rulings in different states If in doubt the manufacturer suggests the user contact their responsible Authority For maximum safety always ensure thatthe following advice is followed e The equipment being tested is in good condition visual check All user instructions are followed Double check power supply connections note LED status Always use specified fuses and protection devices Do not use leads that require repair or are damaged If you are unsure call a licensed Engineer Electrician See P 6 AS NZ 3760 2010 def Competent Person To ensure that all electrical equipment or devices are inspect
26. g and remove the device from service Please Note e Extension leads should always be uncoiled before using or testing e Please ensure that the IEC Adaptor amp the IEC socket are inserted firmly or it may result ina continuity polarity fail Leakage Run Test Main Menu LEAEAGE TEST EARTH LEAKAGE SPH Earth Test Earthed Devices 3 PHASE TEST SPH Ex Lead Testil Insulated Neutral 5 Pin 3PH Leakage Test IECL Protected No NWeutral 4 Pin The Leakage Test is an alternate method to perform insulation resistance tests Leakage testing is a major function of the 3PLM Series Please Note A Visual Inspection test must be carried out before any others refer to First Test Section This test determines errors of leakage not otherwise detected in a normal insulation test If there are any doubts with insulation testing of the equipment the Standard AS NZS 3760 since 2001 allows for a leakage test to be carried out instead The 3PLM Series appliance testers have been designed to perform these tests The Leakage Test applies power to the Device Under Test and measures any imbalance or leakage current The leakage is tested to the limits of the class types specified in the Standard AS NZS 3760 i e Class 1 gt 5mA as Fail The Limit of imbalance measured on the 3PLM appliance tester will read well in excess of the limits set in mA However should the supply circuit be protected by an RCD this device will trip anywhere between 10 to 30
27. he printer won t receive the command e Press the GC key to enable e Press the key to disable Barcode Options Only on Selected Models BARCODE LENGTH SET BARCODE OOOOL USE LEADING WAVE COM ZEROS New Barcode JET Enable Disable QO00001 49596 Barcode options refers to modifying the auto generate barcode function The barcode options can be selected in two ways Either holding F3 down when on the new barcode screen or by the 3PLM options menu Both have the same menu options expect when entering barcode options VIA 3PLM options these menu options are Set the number of digit for barcodes Set the auto generate barcode number Use leading zeros for the auto generate number Search all sites on barcode search only available from 3PLM options menu Search All Sites Allows the 3PLM to search all sites listed in the site names section for the given barcode If this option has not been altered then by default search all site is disabled When disabled the 3PLM can only search for barcodes that are in the site selected Indicated by the site name in new barcode search barcode screen Search all sites is useful disabled and enabled For example you may have many items with the same barcode number in different sites Disabling this feature allow you isolate a particular barcode with ease If you have a barcode but you are unsure which site it is in then you should enable search all sites Be careful when sear
28. he Visual Check screen After the item has been visually inspected there is 1 of 3 options that can be selected Out of Service F2 Fail F3 and Pass enter If Out of Services or Fail is selected the result will be saved and the 3PLM will revert back to the barcode entry screen If you have a wavecom printer a tag will be printed If the test is a visual check only then press enter hold See add equipment for details If the item passes a visual inspection and pass is selected testing can begin Once test has completed the results are displayed To test the item multiple times press F3 this will take you back to the main menu A screen From here you can choose a test class You would use this for example if testing power boards To complete the test press enter This will take you back to the search barcode screen If you have a wavecom printer a barcode will be printed once enter is pressed To download records using WinPATS software please refer to the Downloading Records section of the WinPATS manual 13 Wavecom Instruments 3PLM User Manual Deleting Equipment If for some reason a record needs to be deleted it can be removed from the search barcode screen This feature removes the whole record and not just the test results If a record is deleted and replaced it might create a new record when downloaded in WinPATS This is because the record details didn t match the original ones Multiple items cannot be deleted at t
29. he ratings and compare them to that on the compliance plate etc Quick test of Power outlet Voltages in rooms or locations split from multiple circuits 34 Wavecom Instruments 3PLM User Manual SPECIFICATIONS 3PLM Series Mains Supply Test Checks Polarity and continuity of mains supply by LED indicators A N A E and N E Flashing red LED indicates test fail CLASS 1 Earthed Appliance Appliance Check Ensures appliance is plugged in and turned on gt Earth Bond Test 0 2 1 2 5 10 15 20 25A Pass level Less than 1 0 QO Measurement 0 10 to 10 Q gt Insulation Test 1000VDC 500VDC 250VDC Pass level Greater than 1M Q Measurement 0 10 to 10MQ See also Leakage Test CLASS 2 Double Insulated Appliance gt Appliance Check Ensures appliance is plugged in and turned on Insulation Test 500VDC 250VDC Pass level greater than 1MQ Measurement 0 10 to 10MQ See also Leakage Test Extension Leads gt Earth Bond Test 0 2 1 2 5 10 15 20 25A Pass level Less than 10 Measurement 0 10 to 10 00s gt Insulation Test 500VDC 250VDC Pass level Greater than 1MQ Measurement 0 10 to 10 0MQ Continuity Polarity Test 250VAC check continuity and polarity of leads Displays Pass Fail Leakage Test Leakage Current 0 to 30 0mA at 200 to 265V gt Earth Leakage Test 240VAC Mains Pass level 1 2 5 5 0 mA leakage test levels with up to 20 32 Amp load operation Depending on 20A or 32A model
30. he same time WAVECOM Search Barcode From the search barcode screen press and hold the enter key Then press the F1 key The feature cannot be selected when new barcode is displayed OOO0001 4999 WAvVECOM Records Del The search barcode will change to S Record Del Scan or type the barcode that needs to be deleted If the record does not exist or is in the wrong site a message aradan re will be displayed saying nothing was found Delete Record O000000 When a match is found a last warning message will be displayed To delete the Hold F3 to Del record hold the F3 key for 2 seconds To cancel press enter DELETE QUIT Technical Information Class 1 Earthed Appliance Construction Single basic insulated and protectively earth equipment This type of product design provides two safety barriers between all live conductors at dangerous voltages and the equipment user The provision of basic insulation between exposed metal parts and live parts is the first barrier to provide basic protection against electric shock The second safety barrier is by the connection of exposed accessible conductive metal parts to the protective earthing conductor earth wire in the fixed wiring of the device Installation The protective earthing terminal of the equipment must be marked with the word earth or the symbol E or the symbol for Earth Terminal or Protective To perform this test a continuous earth loop must be made between the expo
31. ility to upload download data to amp from the 3PLM Unit NOTE Please familiarise yourself with and refer to the WinPATS manual for full details 36 Wavecom Instruments 3PLM User Manual Disclaimer E amp OE All specifications may be subject to be change by Wavecom Pty Ltd without prior notice Updated Specifications amp Model changes may be found on the Wavecom website www wavecom com au At the time of developing this manual all care and consideration for accuracy has been implemented Wavecom accepts no responsibility for any errors or omissions in this document This is partly based on the fact that Electronics amp Electrical testing and specifications worldwide are constantly changing and that Local State and National Regulatory Authorities may also have differing or additional requirements It is strongly recommended the Purchaser check Local Regulatory Standards that may be applicable in your region No part of this document may be Reproduced Copied Modified or Utilised in any way or form without the permission of Wavecom Instruments in writing The 3PLM Series is Designed and Manufactured by Wavecom Instruments Pty Ltd Australia Head Office 257 Grange Road Findon Adelaide South Australia 5023 www wavecom com au Tel 61 0 8 8243 3500 Fax 61 0 8 8243 3501 Email Enquiries office wavecom com au Product sales wavecom com au Technical support wavecom com au 37 Wavecom Inst
32. insulation or reinforced insulation provided there being no reliance on precautions in the fixed wiring of installation Class Il equipment is marked with the words DOUBLE INSULATION or the symbol o Note 1 Double Insulation is insulation comprising both basic and supplementary insulation For double insulated under Safety Symbols Note 2 Reinforced Insulation is a single insulation system with a degree of protection against electric shock which is equivalent to double insulation Testing of Electrical Equipment Many testing personnel have some reservations in testing sensitive electronic equipment using a 500V DC insulation test There is a perceived fear of causing internal damage from over voltage With the introduction on the 3PLM Range of appliance testers these concerns are alleviated The 8PLM Range of electrical portable appliance testers is safe to test electronic equipment as the tests are carried from Active Neutral shorted by a relay inside the tester to Earth No dangerous voltages pass through in this mode to the internal components of the DUT Device Under Test If these tests are done using an Insulation Tester only and the user tests Active to Neutral this would be a cause of potential damage this is why the 3PLM product range is far safer to use Some changes may be required in certain configurations where fitted surge protection devices MOV s in the DUT may cause a failed test result Applying 500
33. lay breaker Enter Test Function F1 Test Function F2 Test Function F3 Power indicato Supply Indicator IEC Socket 1 Phase DUT Socket Wireless Scanner Keyboard 3PL Mains Supply 3 Phase DUT Socket RCD fuse holders Power socket Wavecom Instruments 3PLM User Manual JP LM Tes t n P r in t Uni t 20A ver shown Plug Apaptor and Cord Information Single Phase Power Supply Cord 3 Phase Power Supply Cord 4 and 5 pin types 3 Phase Extension Lead Testing Cord Single Phase Extension Lead amp Alligator Clip Warning Do not connect the 3 phase extension cable to a 3 phase wall outlet Damage may result This cable should only be used for testing extension leads Similar testing procedures are done with this cable as described in Ext Lead Test Serial Male Scanner Connector USB data port Suits Wavecom Scanner Only USB interface for data transfer to laptop or PC 3PLM Interface Panel Located under Keyboard Keyboard Socket For Wavecom Keyboard Onl A Printer Port Optional printer port if fitted serial lead required Warning Do not plug any other leads or devices into the L H USB or R232 sockets These are proprietary connectors specifically wired for 3PLM models only Wavecom Instruments 3PLM User Manual Manufacturer Recommendations Calibration The AS NZS3760 2010 Standard recommends a routine calibration verification of this unit to ensure the accuracy o
34. ly altered in firmware A Pass Fail will also be displayed at the end of the test The run time period can be adjusted by 5sec increments The value can be changed by selecting the leakage test time in the options menu See the special functions section of the manual for more details The factory default setting is 20sec The value for the leakage runtime is also used for the power test plus models only Caution Before operating ensure the equipment is firmly secured to eliminate the possibility of causing injury or damage This function will power the Unit on Please make sure the drills or any blades of any rotating devices have been removed prior to any testing and have clearance from body or object Appliances may jump off benches and may need to be secured by clamping them or they may cause serious damage if not secured properly KILL SWITCH Goren To stop operation press the return enter key at any time The unit will power down after 20 seconds as factory default A Caution This test turns the Equipment ON The leakage test can handle up to 20 32 Amp Resistive 20A Models amp 32A Models Load MAX Exceeding this value will either trip the front thermal fuse oraninternal auto reset fuse Care should be taken when testing inductive loads i e AC motors with large start up currents Caution DO NOT OPEN 3PLM Series case as there are no serviceable parts inside and you can void your Warranty 21 Wavecom Instrume
35. mA and trip the mains supply switch OFF Procedure 1 Complete a Visual Inspection For details refer to the First Test section If the device passed the Visual Inspection continue with the following instructions if not refer to the First Test section 2 Press the return key to view Main menu B for single phase leakage test or Menu C for 3 phase leakage test 3 Press a function key as follows to start the test s Press F1 F2 or F3 depending on the type of device you are testing F1 Earth Leakage limit set to MAX then 5mA fail F2 D insulated leakage and ext leads limit set to MAX then 5mA fail F3 RCD leakage CONFIRM IF SAFE TO CONTINUE OK QUIT IF OK DEVICE WILL POWER SWITCH ON ENSURE ITEM AND ENVIRONMENT FOR SAFE OPERATION 4 Results are displayed read and record appropriately 5 Unplug DUT 6 Press Enter until you return to the main menu If the result was a pass Tag with the appropriate tag including next test due date and return to service If the result was a fail Tag with a DANGER TAG and remove from service 20 Wavecom Instruments 3PLM User Manual The Leakage Test allows the user to operate the appliance in normal 240v operation conditions and measure its Operating Leakage current The displayed parameter is mA A predefined value for individual class types is programmed into your 3PLM These limits are set according to the AS NZ3760 Should these values change in future it can be simp
36. n from 500mA Set This button sets the selected current for the next trip time test The 3PLM will then return to the current trip time test screen Changing the RCD type set RCD Type 40m 10m lt 3 00e lt 3 Ori Depending on the RCD the RCD type needs to be selected from the options menu These options change the pass fail values when performing RCD tests Please make sure that you have the correct RCD type selected The RCD types are Type I Has a trip time of lt 40mS and a trip current of lt 10mA These types of RCD s are mainly used on sites containing medical equipment These types of RCD s must be compliant with AS3551 Type ll Has a trip time of lt 300mS and a trip current of lt 30mA Unless specified on the RCD device nearly all RCD will be this type This is the default setting on new units Selects type II and saves to memory Selects type I and saves to memory 26 Wavecom Instruments 3PLM User Manual RCD Timed Test continued Press the F1 key to select the 0 deg test and the F2 key to select the 180 deg test Use the F1 F2 keys to scroll through the multipliers X0 5 X1 0 X5 0 of the set current Maximum output current 500mA I E if set test current were 100mA then 100 X 5 0 500mA If set test current 200mA then maximum output 5 x 200mA 1A is out of range Unit will not deliver this output current and display on Screen OUT of RANGE Press F3 to start test 3PLM
37. n specify how many digits they want in the barcode The auto generate function allows a minimum of 2 and a maximum of 7 digits in the barcode to be generated To increase the number of digits in the barcode press F1 To decrease the number of digits in the barcode press F2 To proceed press enter or to return to the Barcode Entry screen press F3 VWAVECOM Hew Barcode OOOO000 500d BARCODE LENGTH 7 DIGITS lf enter was pressed the user can now set the barcode that will begin the sequence Key ina barcode and press enter to continue or F3 to return the Barcode Entry screen lf enter was pressed the user can specify whether or not to use leading zeros in USE LEADING the auto generated barcode For example If the user selected 7 digits in the ZEROS barcode and entered 1000 as the barcode the barcode will be 0001000 if leading zeros are enabled To enable leading zeros press F3 To disable press Enable Disable enter The current barcode that will be generated is displayed in the bottom left hand corner of the Barcode Entry screen WAvVECOM New Barcode The current barcode that will be generated is displayed in the bottom left hand corner of the Barcode Entry screen 0001000 spog Entering Items WAVECOM Type in the new barcode or press F3 quickly to choose the auto generate number Once a barcode has been specified items can be entered If the screen displays search barcode press enter to switch to new barcode here must
38. ns of this unit This test function ensures that the appliance is plugged into the 3PLM appliance tester and that it is switched on If the device is not plugged in and the 3PLM appliance tester detects that no device is present plug in to continue the test or confirm QUIT to return to the main menu If for some reason the NCNT circuit does not detect the device but it is actually plugged in and turned on the operator will need to override the NCNT function To do this Over Ride Press F3 With an emphasis in the Standard AS NZ3760 2010 for carrying out the live testing the 3PLM Series appliance testers will indicate for you to check if the device is plugged in and switched on If the device is not plugged in and or recognised it may require a live test therefore making it necessary for the operator to carry out a full functional Leakage Test Available on all models of 3PLM Series This function is to ensure that correct testing procedures are carried out in accordance with the Standard AS NZ3760 2010 Optional Note When using 3 Phase adaptors the NCNT function will need to be over ridden by pressing the OK key prior to the 3PLM appliance tester performing the assigned test Some single Phase appliances controlled by contactors will also require manual over ride In some instances holding the ON button will enable the NCNT function to work normally 17 Wavecom Instruments 3PLM User Manual Visual Inspection
39. nts 3PLM User Manual Power Test Main menu B Main Menu B WARN ING POWER RESULTS Power Test DEVICE POWERED ada OW Continue If SAFE 0 0974 aa i QE QUIT O 0O35PF The Power Test is for the purposes of monitoring and performance of equipment The power test feature is for single phase appliances only Please Note A Visual Inspection test must be carried out before any others refer to First Test Section The Power TEST completes the following sequence as part of its comprehensive testing procedure 1 Integrated Supply Mains Test Refer to Integrated Test section Power Test Procedure nd Complete a Visual Inspection For details refer to the First Test section If the unit passed the Visual Inspection continue with the following instructions If not refer to the First Test section Plug male end of appliance into 3PLM Single Phase Series Appliance socket Press F1 to start the power test If the appliance is safe to test press F3 on safety warning message The unit will power on Read and record results appropriately Once the appliance has powered down and the test is complete Unplug the unit Press Enter to return to the main menu oe eae ee This test is ideal for service agents and electricians The user can plugin the appliance and turn it on with real time measurements displayed on the display This is useful when testing an appliance with a compliance name plate on it The operator can compare the n
40. ode Screen METER MODE 242V Run Exit How to Activate Meter Mode e Hold F3 from main screen display for 2 seconds Main Menu A Volts Screen Main Menu METER MODE it Class 1 Test fe Class 2 Test 241 Y D Lead Test Run Exit Screen will change to show volts by default when socket is not powered Plug appliance into testers test socket Press F3 to go into power mode and will show Warning device will be powered Press F3 again to continue and power the device If you continue the A N LED on the tester will flash green O and device will power up Otherwise press enter to exit Warning Screen Amps Screen Watts Screen Volts Amps Screen METER MODE METER MODE METER MODE WARN ING DEVICE 366VA otop Hest otop WILL BE POWERED 4 0 9 9 6 OW OK QUIT Wext stop Wext Press the enter key to stop the power mode and return to volts Pressing F3 key will start power mode again TnT stays in Meter Mode _ regardless of power on off until you exit Pressing Return key will exit meter mode and return the main menu lf the TnT is reset in meter mode it will return to meter mode when powered up again Meter Mode Uses Meter Mode is a new and versatile function of the new Gen 4 tester that will allow you to display Electrical Parameters from a Power outlet source or from the appliance under test Typical Applications Example Measuring the Power of a bar heater or electric power drill check t
41. rer and distributor of this product that it SHOULD NOT BE RETURNED TO SERVICE To do so would be considered unsafe Class 1 Earthed Appliance Test F1 Please Note A Visual Inspection test must be carried out before any others refer to First Test Section The Class 1 Earth Appliance Test completes the following sequences as part of its procedure 1 Integrated Supply Mains Test Refer to ntegrated Test section 2 Integrated NCNT Test Refer to Integrated Test section 3 Earth Bond Test User Selectable current level pass level less than 1 4 Insulation Test 250V or 500V pass level greater than 1MQ Please Note 250V insulation testing applies to Class 1 appliances if selected refer to Technical Information for details to change test voltage In some situations if the DUT is labelled with Surge Protection Fitted or if it contains MOV s Metal Oxide Varistors conduct a 250V insulation test If unsure refer to the AS3760 Standard Should it still fail remove it from service Refer to the Technical Information section to change test voltage Procedure Complete a Visual Inspection Only proceed if test passed Plug device into appliance socket Connect earth clip to any exposed metal on the device Press the F1 key and wait for results ensure that device is powered on Read and record results appropriately Unplug device Tester will return to Main menu A automatically i Re 18 Wavecom Instrum
42. rotection Fitted or if it contains MOV s Metal Oxide Varistors conduct a 250V insulation test Hint always read the compliance plates during Visual If unsure refer to the AS3760 Standard Should it still fail remove it from service Refer to the Technical Information section to change test voltage Lead Test Ext Lead Test F3 Please Note A Visual Inspection test must be carried out before any others refer to First Test Section The Lead TEST completes the following sequence as part of its comprehensive testing procedure 1 Integrated Supply Mains Test Refer to Integrated Test section 2 Earth Bond Test User Selectable current level pass level less than 10 3 Insulation Test 250V or 500V pass level greater than 1MQ 4 Continuity and Polarity Test 240VAC 2mA Checks continuity amp polarity of leads Procedure 1 Complete a Visual Inspection Only proceed if test passed 2 Plug in supplied orange IEC adaptor lead supplied into front IEC socket 3 Plug male end of extension lead or power board in to 3PLM Series appliance socket 4 Press the F3 key and wait for results 5 Read and record results appropriately 6 Unplug extension lead power board 7 Tester will return to Main menu A automatically 19 Wavecom Instruments 3PLM User Manual If the result was a pass Tag with PASS tag showing next test due date and return the device to service If the result was a fail Tag with a DANGER ta
43. ruments 3PLM User Manual WAVECOM TT040 50 Thermal Transfer Tag Printer Please Note there is a master POWER SWITCH at the rear r h side of the printer This should be left ON under all normal operation conditions Full instructions on the operation of the Tag Printer can be found in the Owner s Manual located on the Main Software and Drivers CD supplied with the 3PLM series unit Insert the disc into any PC then from the Auto Run Menu select the tab for the Printer Manual On this disc is information and operating instructions for the Tag Printer including steps with images on how to change the printer Media Ribbon and general maintenance tips The user may either read thismanual directly from the CD or print a hard copy for reference Wavecom Instruments 3PLM User Manual Fully Integrated 3PLM UNIT 3PLMe USER NOTES WWW Wavecom com au V100913 Note if required 3PLM Firmware can be updated by Wavecom only to comply with potential Changes in Testing Standards 39
44. sed Once a site has been selected the 3PLM will return to the initial screen and the site that was selected will be shown in the top left hand corner of the screen Adding a User From the initial screen press F2 Don t hold the F key this will select a different function The 3PLM can hold up to 6 users To scroll through the users press enter 3 sites will be shown on each screen Each site corresponds to an F key for example User 1 F1 To edit a user hold the corresponding F key for 2 seconds Entering a user is the same process as entering a site please refer to the Adding a Site steps above Note The user name will not appear on the initial screen once selected Once a user has been entered the F buttons are used to select the user For example to select user 1 F1 would be pressed 10 Wavecom Instruments 3PLM User Manual Auto Generate Barcode Function If required the auto generate barcode number can be set before testing The auto generate is always displayed in the bottom left hand corner of the new barcode screen The 3PLM incorporates an auto generate barcode function which allows barcodes to be generated sequentially after a barcode has been specified for example the specified barcode is 1000 so the next barcode to be generated will be 1001 1002 1003 etc This can be accessed by holding F3 for 2 seconds on the Barcode Entry screen Once the auto generate menu has been accessed the user ca
45. sed conductive material metal and the 3PLM appliance tester This is done by means of connecting the earth lead with the crocodile clip probe attached to a GOOD earth point paint amp coatings will not provide effective connections and the DUT Device Under Test plugged into the 3PLM appliance testers DUT socket The Maximum allowable limit is less than1 0 ohm The 3PLM has a user selectable current range for earth bond test It is up to the user to no exceed the current rating of the device under test For single phase appliances and extension cords 10A should be exceeded For 3 the correct current level should be selected based on the device under test To Change the Earth Current Level 14 Wavecom Instruments 3PLM User Manual To change the earth current select the E Bond Amps Set from the options menu Please refer to the special functions section of this manual Class 2 Double Insulated Construction Double insulated equipment This method of construction employs two safety barriers comprising two layers of insulation between dangerous voltages and the user of the equipment The first layer of insulation is formed around the live conductor and is termed the Functional Insulation The second layer of insulation is termed the Supplementary Insulation In Class II equipment protection against electric shock does not rely on basic insulation only but has additional supplementary insulation such as double
46. trical Equipment To Change Test Voltage Integrated Tests Supply mains Test NCNT Test First Test Visual Inspection test Class 1 Earth Appliance Test Class 2 Double Insulation Test Lead Test Leakage Run Test Kill Switch Power Test Main Menu B Diagram 1 For Testing Portable RCD Diagram 2 For Testing Fixed RCD at Switchboard Time Test Changing RCD Type RCD Time Test 3IPLIM Options Menu Introduction Leakage Test Time Earth Bond Current Selection Choose Labels Stop Labels Spooling Other Features Meter Mode Specifications WinPATS Professional Asset Management Software Short Guide Disclaimer WAVECOM TTO40 50 Thermal Transfer Tag Printer NN DOOD AAA 36 37 38 Wavecom Instruments 3PLM User Manual Important Information Contents 3PLM Appliance Tester IEC Test Lead 500mm Orange User Manual USB PC Cable Wavecom Keyboard Standard Earth Alligator Lead 1800mm Black Wavecom Scanner Standard 1 Phase Power Supply Cable Wavecom Tag Printer Optional 3 Phase Power Supply Cable Serial Printer Cable Optional 3 Phase Extension Test Lead WinPATS Software Standard Manufacturer s 1 Calibration Certificate Valid for 2 Rolls 500 Labels 2 x Resin Ribbons Optional 12 Months after Purchase Please Register your WinPATS Software and 3PLM Unit with the provided documentation or online via our website www wavecom com au 3PLM Diagram _ ES eee 15 amps resist circuit Disp
47. will display results for 5 seconds after mains supply is tripped If the power is not reset by the time the unit loses power then the result will be displayed on power on A Caution Pressing F3 at this point will cause tripping if RCD fitted to circuit RAMP CURRENT TEST This testing principal is designed to trip RCD devices using a ramping up current value to determine the trip current of the RCD device This useful test allows the user to determine circuit leakage load pre loading of RCD circuit This can assist in determining nuisance tripping issues or determining RCD performance if suspected faulty or inconsistent in performance The 3PLM has a nominal leakage current of 2mA which should be added to the result of test E G if RCD tripped at 22mA 2mA 3PLM 24mA trip current RCD Trip Current Press F2 to show the Trip current screen RCD RAMP TEST set RCD type Start Exit A Caution Pressing F3 at the RCD ramp test screen will cause tripping if RCD fitted to circuit F2 can be pressed to change the RCD type at this point See Changing the RCD type for details The Trip Current Test will ramp the mA current up until the RCD breaker trips Current range 1 500mA This test can go for up to 10 sec to scroll through full range if RCD faulty or not fitted Repeated testing in this mode will cause heating of 3PLM Should over heating occur the internal temperature sensor will cause display to indicate over temp allow to