1. X Cite Fluorescence Illumination In Control X Cite exacte Quick Start Guide SIDE 1 Start Up Install the lamp module see side 2 Install the light guide see below Plug the unit into a properly grounded AC outlet Turn on the Power switch at the front of the unit The lamp will automatically strike when power is turned on to the unit The LAMP indicator will be illuminated Allow the lamp to warm up approximately 4 minutes During the warm up period the LED display and indicators will flash RERE m Installing the Light Guide 1 Remove the protective end caps from the light guide 2 Insert the light guide into the light guide port on the front panel until it seats with a positive click 3 Insert the light emitting end of the light guide into the collimating adapter after it has been installed on the microscope Front Panel Controls 1 SHUTTER button will open close the shutter OPEN LED will indicate when shutter is open 2 MODE button is used to cycle between Coarse Fine and Hours mode when pressed and then released within 1 second The LED s to the right of the display indicate the current mode of operation e Coarse Mode display shows intensity in increase decrease in 10 steps using UP DOWN buttons e Fine Mode display shows intensity in increase decrease in 1 steps using UP DOWN buttons e Hours Mode display shows how many hours the lamp has been in operation 3 If the unit is cali
2. brated MODE button is used to toggle between Relative XXX and Absolute XX XX Watts power display modes when pressed and then released after more than 1 second The display number formatting will indicate the current power display mode Using the Footpedal Remote Device 1 Connect Footpedal remote device to 3 5mm jack at rear of unit 2 Depressing the footpedal or providing a momentary contact closure provides the same function as pressing the SHUTTER button gt 4 WARNING Never look into the light emitting end of the light guide The light could severely damage the cornea and retina of the eye oO Zed Setting the Intensity Level 1 2 4 Set your X Cite exacte to Coarse or Fine Intensity Mode by pressing the MODE button Press the UP or DOWN button to increase or decrease the light output Note No light will be emitted if 0 is selected Press the SHUTTER button to open the shutter The OPEN indicator will illuminate Warning UV light and high levels of visible light will be emitted from the end of the light guide Press the SHUTTER button to close the shutter The OPEN indicator will turn off Closed Loop Feedback CLF Ls To enable CLF Set the intensity to the desired level Press and hold the MODE button then press and release the UP button The CLF light will turn on Note UP must be released before MODE treat MODE like a shift key To disable CLF Press and hold the MODE bu
3. e AC power cord is disconnected from the unit and connect it to its mate located on the top of the lamp housing wall 2 Remove the screw from the lamp access side panel using the 3mm hex Tip the Intelli Lamp connector will only attach in the correct orientation If tool located on the bottom of the unit and remove the panel from the you are having difficulty attaching the connector try rotating it by 180 unit cover 3 Carefully remove the lamp module from its container holding only the ceramic component and glass rim 4 As illustrated below position the lamp facing towards the front of the unit with the power leads facing towards you Caution Do not touch the bulb s glass envelope or the inside surface of the reflector Skin oils can cause the lamp module to fail prematurely clus 7 Locate the Power connector with two leads and connect it to its mate located on the bottom of the lamp housing wall Tip the Lamp Power connector will only attach in the correct orientation If you are having difficulty inserting the connector try rotating it by 180 s a Center 5 Make sure the middle of the lamp is in position to fit into the spring lamp clamp Slide the lamp until it snaps into the spring clamp The leading power edge of the reflector should fit snugly into the lamp holder recess lead as shown 8 Itis very important to ensure that the lamp anode cooling fin bar and lamp power lead at the front of the lamp ar
4. e centered within the lamp holder cutout Rotate the lamp as required 9 Replace the lamp housing side panel and tighten the fastening screw 10 Replace the hex tool in the clips underneath the system for future use Note if the lamp housing panel is not secured completely the lamp will not strike and the bulb message will display when power is turned on to the unit Lumen Dynamics Group Inc Tel 1 905 821 2600 2260 Argentia Rd Toll Free 1 800 668 8752 Mississauga Ont Fax 1 905 821 2055 Canada LSN 6H7 www ldgi xcite com
5. tton then press and release the UP button The CLF light will turn off Calibration Radiometry optional To Measure Output Power 1 2 3 4 Set the X Cite to the desired intensity Close X Cite shutter using SHUTTER button Remove light guide from microscope adapter Snap the 3mm red light guide adapter into the top of radiometer if not already in place Loosen thumb screw on red light guide adapter fully insert and fasten light guide into holder Turn radiometer on using the ON button Open X Cite shutter the power reading will show on the display To Calibrate the X Cite exacte 1 Set up the radiometer to obtain a power reading as described above Connect radiometer cable to port on the right side of the radiometer under flap in rubber boot and the port on the right side of the X Cite exacte Press and release the X CITE CAL button The X Cite will display CAL and the radiometer will display a moving dash while the calibration sequence is running Printed in Canada if the light is observed directly Eye shielding must be used at all times as well as clothing to protect exposed skin Refer to X Cite exacte User s Guide Section 2 035 00284 Rev 1 X Cite exacte Quick Start Guide SIDE 2 A Warning Refer to User Manual for all safety precautions prior to use Installing the Lamp Module 6 Locate the 4 pin Intelli Lamp sensor connector at the rear of the lamp module 1 Be sure th