1. Exhaust Catalog Voltage Filter Exhaust Connection Airflow cfm 3970000 115 Volt 60 Hz HEPA None 280 3970001 115 Volt 60 Hz Carbon None 280 3970002 115 Volt 60 Hz HEPA Thimble to outside 280 3970003 115 Volt 60 Hz HEPA Carbon None 220 3970004 115 Volt 60 Hz Carbon Carbon None 220 3970020 230 Volt 50 Hz HEPA None 280 3970021 230 Volt 50 Hz Carbon None 280 3970022 230 Volt 50 Hz HEPA Thimble to outside 280 3970023 230 Volt 50 Hz HEPA Carbon None 220 3970024 230 Volt 50 Hz Carbon Carbon None 220 HEPA Filter Part 3707900 is 99 99 efficient on particles 0 3 micron HEPA Filter Bag In Bag Out Bag Part 3776002 helps contain hazardous particulate matter during filter changing operations Carbon Filter Provides granular activated carbon or impregnated carbon Filter Classification Part x Pounds Organic 3923400 12 0 Ibs activated carbon Formaldehyde 3923401 14 0 Ibs impregnated carbon Ammonia 3923402 16 0 lbs impregnated carbon 8 Storage Cabinets SOLVENT ACID 48 9902000 9901000 36 9902100 9901100 30 9902200 9901200 24 9902300 9902400 9901300 9901500 18 9901400 9901600 12 48 w Self Closing Doors 9903000 36 w Self Closing Doors 9903100 30 w Self Closing Doors 9903200 24 w Self Closing Doors 9903300 9903400 24 ADA 9906000 9906100 9
2. 4 Zoned Rear Perforated Baffle directs horizontal laminar air streams to the three zoned sections of the perforated baffle in a single pass The three zoned sections have increasingly more airflow at the bottom that helps form the laminar airflow This smooth horizontal flow minimizes the potential for air to roll forward preventing contamination from moving toward the sash opening The concentrations of materials are largely removed on the first pass through the contaminated chamber The zoned rear perforated baffle may be removed or tilted down for cleaning See Figure 4 5 Product Service 1 800 522 7658 21 Chapter 4 Performance Features and Safety Precautions 22 Figure 4 5 Side Entry Air Foils allow turbulence free air to enter the enclosure from the sides and allow clean air to sweep the interior walls of the enclosure Ergonomic Slope provides maximum visibility and comfort reduces glare thereby minimizing operator fatigue Internal Depth of 23 provides necessary depth to support modern balances and other auxiliary equipment without extending outside the enclosure or resting on the lower air foil Two Enclosure Heights and Depths available as standard enclosures in 22 75 or 32 Taller height XVS enclosures are typically used for pipet operations titration or taller auxiliary equipment Unit depth of 29 is available on 2 through 8 enclosures and 35 5 depth is available on 5 6 and 8 enclosu
3. 2 The attraction and ultimate accumulation of powders hazardous or nuisance are minimized on the inside of the enclosure To correct or ensure against electrostatic issues the following additional measures may be prescribed to improve weighing operations e Maintain a humidity level between 45 and 60 The ability to sustain this humidity range can be challenging depending upon the regional climate and HVAC system e onizers in various forms guns bars and blowers are effective ways to flood an area with ions and essentially neutralize electrostatic electricity 4 Product Service 1 800 522 7658 Chapter 2 Prerequisites Background on Electrostatics or Static Electricity Electrostatic charges on a surface such as the wall of a balance enclosure are not created by moving air Gases do not cause the charge Impurities within the air impinging upon surfaces dictate the polarity and magnitude of the charge The process triboelectrification occurs when the friction of the dust particles contact the surface and electrons move across the interface The ability of a material to become polarized is a property known as permittivity On highly insulative materials like acrylic ions or charged molecules are strongly bound to the surface by polarization forces The higher the force the higher is the permittivity value of the material It is suggested that the use of high permittivity materials such as plastic be avoided S
4. Chapter 7 Accessorizing your XPert or XVS Enclosure Optional Equipment for the Enclosure Figure 7 1 42 Product Service 1 800 522 7658 Chapter 7 Accessorizing your XPert or XVS Enclosure 1 Work Surfaces An optional work surface is available to attach to the enclosure Dished work surfaces are contoured to fit the dimensions of the XPert Balance Enclosures and Protector XVS Ventilation Stations to contain spills Catalog Description Dimensions W x D x H 3908400 Black 2 foot wide Epoxy 24 x 26 66 x 1 3908401 Black 3 foot wide Epoxy 36 x 26 66 x 1 3908402 Black 4 foot wide Epoxy 48 x 26 66 x 1 3908403 Gray 2 foot wide Epoxy 24 x 26 66 x 1 3908404 Gray 3 foot wide Epoxy 36 x 26 66 x 1 3908405 Gray 4 foot wide Epoxy 48 x 26 66 x 1 3908406 Black 5 x 26 7 Epoxy 60 x 26 7 x 1 3908407 Black 6 x 26 7 Epoxy 72 x 26 7 x 1 3908408 Black 8 x 26 7 Epoxy 96 x 26 7 x 1 3908409 Black 5 x 35 5 Epoxy 60 x 35 5 x 1 3908410 Black 6 x 35 5 Epoxy 72 x 35 5 x 1 3908411 Black 8 x 35 5 Epoxy 96 x 35 5 x I 3975500 5 Stainless w Trough 60 x 26 7 x 0 8 3975501 5 Deep Stainless w Trough 60 x 35 5 x 0 8 3975600 6 Stainless w Trough 72 x 26 7 x 0 8 3975601 6 Deep Stainless w Trough 72 x 35 5 x 0 8 3975800 8 Stainless w Trough 96 x 26 7 x 0 8 3975801 8 Deep Stainless w Trough 96 x 35 5 x 0 8 2 Guardian Digital 1000
5. To complete the CAL CONFIGURATION be sure to enter DONE To start the calibration mode use the Labconco 1000 Manual and enter CALIBRATION mode on the display from the SETUP menu Follow the instructions on the display and alter the low exhaust volume with the speed control on the FilterMate or exhaust damper Measure the average face velocity and enter the low value on the display Be careful not to block the opening The low exhaust volume calibration will take about 5 seconds Now alter the high exhaust volume with the speed control on the FilterMate or exhaust damper Measure the average face velocity and enter the high value on the display The high value must be at least 20 fpm greater than the low value The high exhaust volume calibration will take about 5 seconds Product Service 1 800 522 7658 37 Chapter 6 Maintaining your XPert or XVS Enclosure 9 Besure to enter DONE after successfully completing the low and high calibration set points 10 Once calibration is completed go to RUN and hit ENTER The value should read close to the high calibration set point 11 To lower the face velocity to the operating point simply alter the exhaust volume with the speed control on the FilterMate or exhaust damper Then recheck the face velocity with an anemometer to confirm the display on the digital airflow monitor Digital 1000 Alternate Calibration Procedure Constant Volume Conditions 1 To successfu
6. per Chapter 6 Figure 4 1 Product Service 1 800 522 7658 19 Chapter 4 Performance Features and Safety Precautions l 2 20 Aerodynamic Clean Sweep Air Foil has a unique low profile shape that allows air to sweep the work surface for maximum containment The Clean Sweep openings create a constant protective barrier from contaminants In addition should the operator inadvertently block the airflow entering the airfoil air continues to pass under the air foil and through the Clean Sweep openings See Figure 4 2 Figure 4 2 Containment Enhancing Upper Sash Foil includes a perforated air passage directly atop the sash foil to bleed air into the hood chamber and direct chemical and powder concentrations away from the sash opening The radiused sash foil sweeps airflow into the hood with minimal turbulence See Figure 4 3 Figure 4 3 Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions 3 Upper Dilution Air Supply provides bypass air from above the work area This feature constantly bathes the sash interior with clean air and reduces powders and chemical fumes along the sash plane near the critical breathing zone Five to seven percent of the required air volume is introduced through the upper dilution air supply to ensure maximum containment Additionally the upper dilution air supply reduces stagnant pockets of air in the upper interior See Figure 4 4 Figure 4 4
7. 0 A 240 240 96 0 96 0 96 0 120 120 480 480 48 0 C 216 21 6 93 6 93 6 93 6 D 286 378 28 6 37 8 37 8 E 228 320 228 320 32 0 F 223 31 5 22 3 31 3 31 3 G 190 27 7 19 0 277 277 H 290 290 29 0 290 355 I 234 23 4 234 234 30 0 J K 5 and 6 enclosures include five covers two Accessory Exhaust Transition A on top and three on the Adapter 3912401 Centerline s bottom 8 endosures include four covers one on top and three on the bottom TOP 2 3 4 MODELS Accessory Exhaust Tronsition Adapter 3912400 4 88 OD Hose or 3912401 6 037 ID Duct 1 5 dio Monitor Waste Chute on some om 355 Depth Accessory Exhaust Transition Adapter FRONT 5 6 8 MODELS AIRFLOW MONITOR AND EXHAUST TRANSITION ADAPTERS NOT SHOWN 3912403 6 03 ID Duct FRONT 2 3 4 MODELS 2 3 for 3912401 SIDE ALL MODELS 52 Product Service 1 800 522 7658 Appendix C Specifications Environmental Conditions e Indoor use only e Maximum altitude 6562 feet 2000 meters e Ambient temperature range 41 to 104 F 5 to 40 C e Maximum relative humidity 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C e Main supply voltage fluctuations not to exceed 10 of the nominal voltage e Transient over voltages according to Installation Categories II Over voltage
8. 46 47 47 48 49 52 53 54 55 Chapter 1 Introduction Congratulations on your purchase of a Labconco XPert Balance Enclosure or Protector amp XVS Ventilation Station Your high performance enclosure is designed to protect you by providing superior containment while conserving energy at OSHA approved low flow velocities as low as 60 feet per minute It is the result of Labconco s more than 50 years experience in manufacturing fume hoods The XPert Balance Enclosure or Protector XVS Ventilation Station has been engineered to provide maximum containment It will effectively contain toxic noxious or other harmful materials when properly installed The XPert or XVS offers many unique features to enhance safety performance and energy savings To take full advantage of them please acquaint yourself with this manual and keep it handy for future reference If you are unfamiliar with how high performance vented enclosures operate please review Chapter 4 High Performance Features and Safety Precautions before you begin working in the enclosure Even if you are an experienced user please review Chapter 5 Using Your Vented Enclosure which describes the XPert or XVS features so that you can use the vented enclosure efficiently 3 XPert Filtered Balance Enclosure 3930302 is shown with Black Solid Epoxy Work Surface 3908401 and Protector Standard Storage Cabinet 9900100 Work Surface base cabinet blower and ductwo
9. Airflow Monitor or Guardian 500 Airflow Monitor The Guardian Digital 1000 Airflow Monitor or Guardian 500 Airflow Monitor allows you to continuously monitor airflow through the enclosure The rear exhaust plenum is factory prepared to mount either monitor Catalog Description 3944700 Guardian 500 Airflow Monitor 115V 60 Hz 3944701 Guardian 500 Airflow Monitor 230V 50 Hz 3908800 Guardian Digital 1000 Airflow Monitor 115V 60 Hz 3908801 Guardian Digital 1000 Airflow Monitor 230V 50 Hz 3 Fluorescent Light Kit A fluorescent light which rests on the top of the enclosure is available Catalog Description 3909201 3909200 18 25 W x 10 12 D x 3 00 H light kit 115V 60 Hz 18 25 W x 10 12 D x 3 00 H light kit 230V 50 Hz Product Service 1 800 522 7658 43 Chapter 7 Accessorizing your XPert or XVS Enclosure 4 Exhaust Transition Adapters and Dual Exhaust Connectors A 2 x 10 adapter connects to the enclosure from either the top or the bottom of the rear plenum so the duct can be routed either up or down respectively The transition is available for either 5 00 I D hose or 6 OD duct The 5 00 I D hose upper connection is standard and included with the FilterMate Multiple Exhaust Connectors are used for multiple FilterMates or connections for 8 exhaust outlets Catalog Description Material 3912400 Upper connection 5 Hose Coated
10. Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed e Used in an environment of Pollution degrees 2 1 where normally only non conductive atmospheres are present Occasionally however a temporary conductivity caused by condensation must be expected in accordance with IEC 664 Product Service 1 800 522 7658 53 Appendix D Balance Stability Chemir Analytical Services conducted balance stability tests on the XPert Filtered Balance System mounted on a Solid Epoxy Work Surface No marble slab was used The enclosure was tested while operating at three face velocities 0 78 fpm factory setting and 100 fpm Samples of 5 mg 50 mg and 100 mg were weighed on an analytical balance located inside the enclosure Test results were excellent and confirm that the XPert Filtered Balance System including its built in blower and the airflow through the enclosure caused no detectable instability to the balance The weight of each sample was stable to five decimal places Listed next are some hints on balance stability 1 Balance is not calibrated or at least not recently All balances should be calibrated at least annually or after they have been shipped or moved around frequently If the balance has just been moved around the lab it probably will not require recalibration Most new balances have an internal mechanism that compensa
11. No audible alarm Wrong alarm set point Constant audible alarm Continuous alarm Monitor alarms air way to airflow monitor sensor is blocked by insects dust or debris Audible disable will not stay ON Precautions Chapter 4 sections in the manual Ref Appendix D Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed See Location Requirements Chapter 2 Certifying the Enclosure Chapter 3 and Safety Precautions Chapter 4 sections of this manual Ref Appendix D Have enclosure certified and check exhaust system Check FilterMate filters for loading Adjust FilterMate speed control Enclosure should have average face velocity of 60 100 fpm Power supply is not plugged into proper voltage plug in power supply Verify that airflow monitor interface cables are connected Check fuses on FilterMate or accessory light Alarm has been temporarily silenced using test reset or enter buttons Airflow monitor was not properly adjusted Repeat calibration steps outlined in the airflow monitor manual Check airflow and calibration of airflow monitor Check the face velocity of the enclosure as the airflow of the system may have changed If face velocity is correct calibrate the airflow monitor outlined in the Users Manual Lightly clean the airway with clean air Be careful not to touch sensitive electrical components An alarm condition must be continuously present before
12. Only validation through your Operational Qualification protocol can define what is acceptable To minimize the potential for temperature variations Product Service 1 800 522 7658 3 Chapter 2 Prerequisites e Never install balances near heating sources such as radiators and hot plates e Avoid incandescent lighting of the enclosure where radiant heat produced will raise temperatures See Chapter 7 for adding an accessory fluorescent light kit e Do not place the balance and enclosure on a bench that would receive direct sunlight Humidity and Static Electricity Requirements XPert Electrostatics can be troublesome in a balance enclosure It is important to understand and to the extent possible control static charges An electrostatic charged vessel sample or enclosure can apply forces and lead to errors in weighing The repulsion or attraction can be detected with micro semi micro and analytical balances Static charges can also lead to particulates being attracted to surfaces within the balance enclosure Containment of harmful powders prevention of cross contamination and clean up is all enhanced when static attraction of powders is minimized The construction of the XPert Balance Enclosure avoids the use of plastics which are highly insulative The advantages to the glass and epoxy coated metal construction are twofold 1 The enclosure does not contribute high electrostatic forces affecting the precision of the balance
13. SETTINGS DISPLAY UNITS FPM 48 FPM 60 fpm Operatng inflow LOW AIR ALARM RED LED 64 FPM 80 fpm Operatng inflow 80 FPM 100 fpm Operating inflow LOW AIR CUTOFF OFF 5 FPM 60 fpm Operating inflow WARNING AIR ALARM YELLOW LED 67 FPM 80 fpm Operatng inflow 83 FPM 100 fpm Operating inflow WARNING AIR RESET 3 FPM HIGH AIR ALARM OFF LOWER AIR SAMPLE FLUCTUATIONS OFF HIGHER AIR FLUCTUATIONS 396 LOWER HIGHER AIR SAMPLE DIFFERENCE 3 WARN TO ALARM AIR TIME 10 SECONDS ALARM TO WARN AIR TIME 3 SECONDS SHOW AIR FLOW ON SHOW TIME LINE OFF DISPLAYS BAR OFF GRAPH AUDIBLE ALARM ENABLED SENSOR DIFFERENCE 296 SENSITIVITY 80 Suggested Air Alarm Settings Refer to the chart on the previous page for a range of air velocity settings that may be used 5 To complete the CAL CONFIG be sure to enter DONE You are returned to the Main Menu 40 Product Service 1 800 522 7658 Chapter 7 Accessorizing your XPert Balance Enclosure or Protector XVS Ventilation Station There are several ways to accessorize the enclosure for your individual requirements These include the addition of accessory work surfaces airflow monitors fluorescent light kits exhaust transition adapters remote blowers exhaust dampers FilterMate Portable Exhausters filters storage cabinets and utility shelf kit Product Service 1 800 522 7658 41
14. View and Side View of Concealed Bottom Connection to Outside Exhaust 14 Product Service 1 800 522 7658 Chapter 3 Getting Started Exhaust Transition Adapter Accessory Fluorescent Light Accessory Airflow Monitor Moved forward per Chapter 6 5 ID Hose included with FilterMate FilterMate Portable Exhauster Figure 3 4 Top Connection to FilterMate Portable Exhauster Exhausted to Room Product Service 1 800 522 7658 15 Chapter 3 Getting Started To Outside Exhaust Via Thimble Connection 6 nominal Figure 3 5 Concealed Bottom Connection to FilterMate With Outside Thimble Exhaust Connection 16 Product Service 1 800 522 7658 Chapter 3 Getting Started Connecting the Electrical Supply A standard duplex receptacle should be nearby for connecting the FilterMate Portable Exhauster airflow monitor and accessories Please review Chapter 2 for electrical prerequisites ALL WIRING FOR THE BUILDING DUPLEX OUTLETS SHOULD BE PERFORMED BY A LICENSED ELECTRICIAN AND CONFORM TO ALL LOCAL CODES Sealing the Vented Enclosure to the Work Surface When the enclosure has been set in place ducted and wired it may be sealed at the work surface to prevent spilled materials from collecting under the walls A bead of silicone sealant is recommended to seal the enclosure to the work surface Validating the Vented Enclosure The exhaust damper ductwork exhaust blower or FilterMate gives you the flexibility to change the a
15. can be adjusted with the speed control on the FilterMate or by using an adjustable damper on the exhaust blower Typical calibration conditions are set at face velocity air sample differences of at least 20 feet per minute The airflow monitor is factory set to be calibrated with a difference of at least 50 fpm and can be changed by changing the lower higher air sample difference The following suggested in flow face velocity speeds are recommended to successfully calibrate Typical low air alarms are set 10 20 fpm below operational speeds Follow Step 2 below and review the Labconco 1000 Alarm User s Manual that comes with the airflow monitor Low Air Enclosure Low High Alarm Operating In Calibration Calibration Set Point flow Speed Set Point Set Point 40 50 fpm 60 fpm 40 60 fpm 100 120 fpm 60 70 fpm 80 fpm 50 90 fpm 100 150 fpm 80 90 fpm 100 fpm 50 110 fpm 100 170 fpm Go to setup and then CAL CONFIG MENU and change the lower higher air sample difference to 20 fpm This will allow you to successfully calibrate with values of a minimum of 20 fpm difference While in CAL CONFIG MENU change the sensor difference from 1096 to 396 While in CAL CONFIG MENU adjust the red low air alarm to the desired setting such as 55 fpm Then adjust the yellow CAUTION or WARNING to 59 fpm Then adjust the CAUTION or WARNING air reset to 3 fpm This sets the alarm condition
16. connections with either the top exhaust open and the bottom exhaust closed or vise versa The bottom exhaust connection is primarily used in installations where the hose or ductwork needs to be concealed Labconco offers accessory transition adapters for the top or bottom exhaust connection to either a 5 hose or 6 duct See Chapter 7 for ordering any of these accessories For house exhaust only one exhaust connection is required for the 2 through 6 enclosures and two exhaust connections are required on the 8 enclosure Data for the exhaust volume and enclosure static pressure loss are listed for each enclosure model at face velocities of 60 80 and 100 fpm When using the accessory FilterMate Portable Exhausters it requires the 5 Transition Adapter 3912400 which is sold with the FilterMate Multiple FilterMate Portable Exhausters require the airflow to be properly balanced The 5 6 and 8 enclosures require two or three FilterMate Portable Exhausters dependent on face velocity and size Face Exhaust Static Pressure Number of Enclosure Enclosure Velocity Volume Loss in w g Required Width Height Model Description fpm CFM FilterMates 60 70 02 1 2 e 2 XPert 2 XVS 80 95 03 1 100 120 05 1 F 60 110 04 1 3 SE 3 XPert 3 XVS 80 145 06 1 100 185 10 1 d 60 150 06 1 4 n 4 XPert 4 XVS 80 200 10 1 100 250 16 1 32 60 90 03 1 2 10 Sash Tall 2 XPert 2 XVS 80 120 0
17. surface XPert and XVS e When setting up a chemical station not used for weighing powders a mobile stand or cart is allowable The ability for analytical balances to accommodate vibration varies with type and brand More advanced balances have improved reliability however in the preparation of a balance enclosure site please consider the following e Avoid tubular stands or carts that have the potential of moving when touched e Work surfaces should be of a thick rigid material that remains stable when buttons are pressed An epoxy benchtop or accessory work surface is a minimum requirement e bench that is rigidly mounted to the floor or fixed to the wall but not both may be appropriate e corners of a building typically have less vibration than the center e bench with the balance enclosure should not contain any vibration producing equipment such as shakers or pumps e Marble granite or epoxy balance tables are generally recommended by the manufacturers of analytical and microbalances See the installation instructions in Chapter 3 e Marble slabs within the balance enclosure with dampening pads are also an effective low cost means of controlling vibration Temperature Variation Requirements XPert The extent the balance readings are influenced by temperature variations is a function of the balance design Most manufacturers would suggest that a temperature drift of 1 2 C is generally tolerable
18. with audible alarm for airflow conditions The audible alarm can be permanently muted if desired The Guardian Digital 1000 Airflow Monitor displays a face velocity value provides an RS232 communications port to a PC or building computer system can be configured for external input connections such as night setback or external alarm and provides up to three output relays that can be configured For complete detailed information please refer to the separate Labconco 1000 Alarm User s Manual provided with the enclosure Guardian Digital 1000 Installation Procedure 1 The enclosure comes prepared to accept the Guardian Digital airflow monitor system 2 First remove the large 1 19 dia gray hole plug See Figure 6 5 See Figure 6 6 only to reference internal assembly of the airflow monitor Locate the elbow locknut and washer and install it in the 1 19 dia hole per Figure 6 5 and Figure 6 6 The enclosure baffle pivots down to install the elbow washer and locknut 3 Cut the 1 hose supplied with the kit to 10 5 approximate length and install it between the airflow sensor and the elbow 4 Secure the Guardian Digital alarm to the enclosure with double stick tape as shown in Figure 6 7 The airway passage between the alarm module and the enclosure is now complete 5 Locate the metal hose cover and install with double stick tape per Figure 6 6 6 Locate the power supply transformer One end should already be connected to the two pin conn
19. 0 50 fpm 8 Note If the low airflow volume or sash opening cannot be adjusted then a 1 4 to 1 3 of a turn counterclockwise can be adjusted to set the airflow alarm condition at 10 25 below normal operating levels Guardian 500 Alarm Activation The audio and visual alarm will activate approximately 10 30 seconds after an alarm condition is detected To temporarily mute the audible alarm press and release the test reset button NOTE After an alarm condition has been detected the red light will stay on The audible alarm will remain muted until airflow returns to normal levels 32 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure Accessory Light Enclosure Hose Cover Monitor Figure 6 4 Guardian 500 Installation Product Service 1 800 522 7658 33 Chapter 6 Maintaining your XPert or XVS Enclosure Guardian Digital Kit No 3908800 or 3908801 Guardian Digital 1000 Operation The Guardian Digital Airflow Monitor consists of the airflow sensor the Alarm Unit and the 15 VDC power supply For 115V operation the alarm unit is powered by plugging the power supply into the factory prepared digital airflow monitor socket For 230V operation the Alarm Unit is powered by plugging the power supply into a building outlet The alarm has Enter and buttons to program the monitor There is also a green LED SAFE yellow LED CAUTION and red LED LOW
20. 3905600 Air Foil 2 10B 1 3905601 Air Foil 3 10C 1 3905602 Air Foil 4 10D 1 3969200 Air Foil 5 10E 1 3969201 Air Foil 6 10F 1 3969202 Air Foil 8 11 1 3905300 Header 2 11 1 3905301 Header 3 11C 1 3905302 Header 4 11D 1 3905303 Header 5 11 1 3905304 Header 6 1 3905305 Header 8 12 1 3913100 Cover Plate Exhaust Top or Bottom 13A 1 3912400 Upper Exhaust Transition 5 Hose 13B 1 3912401 Upper Exhaust Transition 6 Duct 13C 1 3912402 Lower Exhaust Transition 5 Hose 13D 1 3912403 Lower Exhaust Transition 6 Duct 14A 1 1934601 Bushing Heyco with Flex Shutter 1 50 dia 14B 1 1936800 Bushing Heyco Closed 1 50 dia 15A 2 7868402 Spacer Bushing 15B 2 1912108 Washer 194 ID x 38 OD x 03 T Plastic 15C 2 1889316 Screw 10 24 x 1 00 PH SS 16A 2 3915400 Side Air Foil Short 16B 2 3915401 Side Air Foil Tall 16C 4 1889912 Screw 6 32 x 75 PH Type F Oval Head 17 1 1595619 Hole Plug 500 dia Gray 18 1 1595621 Hole Plug 1 187 dia Gray 19A 1 3901100 Label Ventilation Station 19B 1 3901200 Label Balance Enclosure 50 Product Service 1 800 522 7658 Appendix A Replacement Parts 13 Not Shown 12 Not Shown 19 Not Shown 7 Not Shown Product Service 1 800 522 7658 51 Appendix B Dimensions 2 Foot 2 Foot 3 Foot 3 Foot 4 Foot 4 Foot Stoot 5 Foot 5 Foot 6 Foot 6 Foot 6 Foot Std Ht Std Ht Tall Tall Deep 8 Foot 8 Foot 8 Foot 8 0 100 12 0 24 0 240 24
21. 5 1 100 150 08 1 32 60 140 06 1 3 10 Sash Tall 3 XPert 3 XVS 80 185 10 1 100 230 15 1 32 60 185 09 1 4 10 Sash Tall A XPert 4 XVS 80 250 16 1 100 310 24 N A 60 190 08 1 5 Se 5 XPert 80 250 15 1 100 315 KEN 2 Se 60 230 12 1 6 EUM 6 XPert 80 305 21 2 100 380 32 2 60 310 07 2 8 m 8 XPert 80 410 12 2 100 510 18 3 32 60 290 20 2 5 12 Sash Tall 5 XPert 80 385 1897 2 100 480 54 3 32 60 350 2 2 6 12 Sash Tall 6 XPert 80 465 48 3 100 580 15 3 32 60 465 al 3 8 12 Sash Tall 8 XPert 80 620 27 3 100 775 42 3 Product Service 1 800 522 7658 Chapter 2 Prerequisites Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss The enclosure must be connected to a dedicated blower a house exhaust system or a dedicated filtered exhauster such as a Labconco FilterMate When using multiple FilterMates the airflow must be balanced Labconco offers accessory remote blowers listed next and in Chapter 7 Remote Blowers Roof Mounted A 1 4 hp direct drive Corrosion resistant epoxy coated steel housing and wheel Blower inlet is 6 ID Outlet dimensions are a rectangular 4 25 x 7 38 OD See dimensional data in Chapter 7 CFM Static Pressure Inches of H O S P 0 0 0 125 0 25 0 50 0 75 0 87 CFM 595 560 515 420 300 167 Shipping Catalog Description Wt Ibs k
22. 50 62 75 88 1 00 7068200 7068300 1 4 jmi 4 Remote Blower 115V 60 Hz 4 4 A Explosion Proof Blower 115V 60 Hz 45A 540 800 720 950 410 800 760 1050 350 870 390 970 430 1060 710 1060 620 1060 7068400 1 3 Remote Blower 115V 60 Hz 6 1 A 7068500 3 Explosion Proof Blower 115V 60 Hz 64A 76021050 850321130 710 1060 620 21060 430 1060 460 1150 825 1160 790 1190 770 1220 740 1260 500 1220 640 1260 TOP 6 Exhaust Dampers 375 DIA MTG HOLE 4 13 76 _ 1 50 16 75 SIDE Exhaust dampers allow an adjustment required to maintain proper airflow for roof mounted blowers or house exhaust systems Catalog Description 3924000 6 Epoxy Coated Steel In Line adjustable damper 4724200 6 PVC In line adjustable damper 5983400 10 PVC In line adjustable damper FilterMate Portable Exhausters and Filters For filtered exhaust Labconco offers FilterMate Portable Exhausters capable of exhausting up to 280 cfm of HEPA filtered air or up to 220 cfm of combination HEPA Carbon filtered air when connected to the enclosure For additional exhaust volume multiple FilterMates with balanced airflow can be used Product Service 1 800 522 7658 45 Chapter 7 Accessorizing your XPert or XVS Enclosure
23. 53 51 40 50 48 27 23 60 fpm 63 73 67 60 70 64 37 30 80 fpm 83 93 83 80 90 80 47 37 100 fpm Because of airflow fluctuations in a typical laboratory environment Labconco suggests setting the RED LED low alarm set point to 2096 below the enclosure s operating speed 38 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure 5 To complete the CAL CONFIGURATION be sure to enter DONE If needed refer to the Configuration procedure provided on the following page for additional details 6 To start the calibration mode enter CALIBRATION mode on the display from the SETUP menu 7 Follow the instructions on the display and simulate the low exhaust volume by fully opening the sash You may measure the average face velocity for the low calibration set point or utilize the calculated value provided in the chart above The average face velocity for the low set point is accurately measured by dividing the opening of the enclosure into equal area grids consisting of at least 9 data collection points and measuring the velocity at the center of each grid with a calibrated thermo anemometer Enter the low value on the display Be careful not to block the opening The low exhaust volume calibration will take about 5 seconds 8 Now simulate the high exhaust volume by fully closing the sash to its normal operating position Measure the average face velocity for the high ca
24. 658 25 Chapter 5 Using your XPert or XVS Enclosure Work Techniques e Keep all materials inside the lower air foil and perform all contaminated operations as far to the rear of the work area as possible Segregate all clean and contaminated materials in the work area e Avoid using techniques or procedures that disrupt the airflow patterns of the enclosure Final Purging e Upon completion of work the enclosure should be allowed to operate for two to three minutes undisturbed to purge airborne contaminants from the work area before shutting down the blower Unloading Materials and Equipment e Objects in contact with contaminated material should be surface decontaminated before removal from the enclosure e All open trays weigh vessels or containers should be covered before being removed from the enclosure Shutdown e Turn off the exhaust system and accessory light 1f provided 26 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert Balance Enclosure or Protector XVS Ventilation Station Now that you have an understanding of how to work in the enclosure we will review the suggested maintenance schedule and the common service operations necessary to maintain your enclosure for peak performance Only trained and experienced certification technicians should perform some of the service operations after the enclosure has been properly decontaminated DO NOT attempt to perform these operations if you are not proper
25. 68021 4868022 8 4868030 4868031 4868032 4868041 4868042 4868060 4868061 4868062 4868071 4868072 Protecting your y AN laboratory environment 7 LABCONCO Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 2696 800 821 5525 816 333 8811 Fax 816 363 0130 E Mail Labconco labconco com Home Page www labconco com To receive important product updates complete your product registration card online at register labconco com Please read the User s Manual before operating the equipment Copyright 2004 2007 2009 2010 Labconco Corporation All rights reserved The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change Warranty Labconco provides a warranty on all parts and factory workmanship The warranty includes areas of defective material and workmanship provided such defect results from normal and proper use of the equipment The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco whichever is sooner except the following e Purifier amp Logic Biological Safety Cabinets and PuriCare Lab Animal Research Stations carry a three year warranty from date of installation or four years from date of shi
26. 905000 9905200 24 ADA w Self Closing 9906200 9906300 Doors 18 ADA 9905100 9905300 12 ADA 46 Product Service 1 800 522 7658 Chapter 7 Accessorizing your XPert or XVS Enclosure STANDARD BASE VACUUM PUMP Size Description Dual Doors Right Hinge Left Hinge Dual Doors Right Hinge Left Hinge 48 9900000 36 9900100 30 9900200 24 9900300 9900600 18 9900400 9900700 9907000 9907100 12 9900500 9900800 48 w Self Closing Doors 36 w Self Closing Doors 30 w Self Closing Doors 24 w Self Closing Doors 24 ADA 9904000 9904300 24 ADA w Self Closing Doors 18 ADA 9904100 9904400 12 ADA 9904200 9904500 9 Utility Shelf Kit and Holders Labconco offers a utility shelf kit part number 3925000 consisting of three shelves to hold items inside the enclosure One shelf may be used for the printer one shelf for spatulas and weigh brushes and one miscellaneous shelf The shelves hang from slots in the rear baffle The printer shelf can be inverted and rest on top of the enclosure as an alternate location The 3927700 Bottle Holder holds a squirt bottle and has a 3 diameter opening The 3927800 Tissue Holder permits tissues to be dispensed from the bottom 10 Hoses Hose C
27. 967101 Foil Sash 6 4F 1 3967102 Foil Sash 8 5 2 3 Ft 6913700 Wiper Sash 6A 1 3907500 Stop Right Sash 6B 1 3907501 Stop Left Sash 1 1927403 Spring Compression Sash Latch HS 1 3906700 Latch Bracket 7C 1 6916500 Latch Sash 7D 2 1893206 Screw 8 32 x 38 Phil SS Type F Product Service 1 800 522 7658 49 Appendix A Replacement Parts Item Qty Part Number Description 7E I 3925300 Tall Latch Bracket 7F I 6916501 Tall Latch Sash 7G 1 3974300 Gas Spring 5 20 7H 1 3974301 Gas Spring 6 30 1 3974302 Gas Spring 8 402 7 1 3973200 Gas Spring 5 Tall 302 7K I 3973201 Gas Spring 6 Tall 402 7L 1 3973202 Gas Spring 8 Tall 50 8A 1 3986500 Sash Assembly 2 8B 1 3986501 Sash Assembly 3 8C 1 3986502 Sash Assembly 4 8D 1 3987100 Sash Assembly Tall 2 8E 1 3987104 Sash Assembly Tall 3 8F 1 3987102 Sash Assembly Tall 4 8G 1 3972300 Sash Assembly 5 8H 1 3972301 Sash Assembly 6 SI 1 3973202 Sash Assembly 8 8J 1 3977200 Sash Assembly 5 Tall 8K 1 3977201 Sash Assembly 6 Tall 8L 1 3977202 Sash Assembly 8 Tall 9A 1 3907400 Bracket Air Foil Right 9B 1 3907401 Bracket Air Foil Left 9C 2 1932401 Washer Shoulder Plastic 9D 2 1912108 Washer 194 ID x 38 OD x 03 thick plastic 9E 2 1909217 Cap Nut 10 24 SS 10A 1
28. Steel Epoxy Coated Included with FilterMate 3912401 Upper connection 6 Duct Coated Steel Epoxy Coated 3912402 Lower connection 5 Hose Coated Steel Epoxy Coated 3912403 Lower connection 6 Duct Coated Steel Epoxy Coated 3981600 Dual Exhaust 8 x 6 x 6 Coated Steel Epoxy Coated Use with 8 unit only to house exhaust 5 Remote Blowers Has a 1 4 hp direct drive motor and corrosion resistant epoxy coated steel housing and wheel with blower inlet of 6 00 ID Outlet dimensions are 4 25 x 7 38 OD CFM Static Pressure Inches of HO S P 0 0 0 125 0 25 0 50 0 75 0 87 CFM 595 560 515 420 300 167 id Shipping Wt Catalog Description Ibs kg 4863500 Remote Blower 115 V 60 Hz 4 4 amps 35 16 4863501 Remote Blower 115 230 V 50 Hz 5 6 2 8 amps 35 16 7053501 Explosion Proof Remote Blower 115 V 60 Hz 4 4 amps 40 18 7 313 10 13 4 HOLES bk 127 4 425 9 38 1 4 hp or 1 3 hp belt drive with adjustable sheave to change RPM Corrosion resistant epoxy coated steel housing and wheel Blower inlet is 10 87 ID Outlet dimensions are a rectangular 5 5 x 10 OD See dimensional data in Chapter 7 44 Product Service 1 800 522 7658 Chapter 7 Accessorizing your XPert or XVS Enclosure Catalog HP Description CFM RPM Ranges at Static Pressure in H20 Std EP 25 38
29. User s Manual XPert Filtered Balance Enclosures amp Protector XVS Ventilation Stations XPert Balance Enclosures Models 3930200 3930201 3930202 3930221 3930222 3930230 3930231 3930232 3930241 3930242 N 3930300 3930301 3930302 3930321 3930322 3930330 3930331 3930332 3930341 3930342 3930400 3930401 3930402 3930421 3930422 3930430 3930431 3930432 3930441 3930442 3971500 3971501 3971502 3971521 3971522 5 3971530 3971531 3971532 3971541 3971542 3971560 3971561 3971562 3971571 3971572 3971600 3971601 3971602 3971621 3971622 6 3971630 3971631 3971632 3971641 3971642 3971660 3971661 3971662 3971671 3971672 3971800 3971801 3971802 3971821 3971822 8 3971830 3971831 3971832 3971841 3971842 3971860 3971861 3971862 3971871 3971872 XVS Ventilation Stations oe 4862010 4862011 4862012 4862021 4862022 4862030 4862031 4862032 4862041 4862042 4863010 4863011 4863012 4863021 4863022 4863030 4863031 4863032 4863041 4863042 4864010 4864011 4864012 4864021 4864022 4864030 4864031 4864032 4864041 4864042 4865010 4865011 4865012 4865021 4865022 5 4865030 4865031 4865032 4865041 4865042 4865060 4865061 4865062 4865071 4865072 6 4866010 4866011 4866012 4866021 4866022 4866030 4866031 4866032 4866041 4866042 4866060 4866061 4866062 4866071 4866072 4868010 4868011 4868012 48
30. apter 3 Getting Started Now that the site for your XPert or XVS vented enclosure is properly prepared you are ready to unpack inspect install and validate your system Read this chapter to learn how to e Unpack and move the enclosure e Setup the enclosure with the proper supporting structure and work surface e Connect to an exhaust system Connect the electrical supply e Seal the enclosure to the work surface e Arrange validation for the enclosure Depending upon which model you are installing you may need common plumbing and electrical installation tools in addition to 5 16 3 8 7 16 and 1 2 wrenches ratchets sockets a nut driver set a flat blade screwdriver a Phillips screwdriver and a carpenter level to complete the instructions in the chapter Note The enclosure models weigh between 50 to 200 Ibs each 22 to 90 kg The shipping container allows for lifting with a mechanical lift truck or floor jack if you must lift the enclosure manually follow safe lifting guidelines Do not lift by the front Air Foil 10 Product Service 1 800 522 7658 Chapter 3 Getting Started Unpacking the Enclosure Carefully remove the shrink wrap or carton on the enclosure and inspect it for damage that may have occurred in transit If damaged notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier e Do not return goods without the prior authorization of Labconc
31. arm It is adjusted with a small screwdriver GUARDIAN AIRFLOW MONITOR sre SAFE LOW x SASH LJ SETBACK HIGH Ls Figure 6 1 Component Identification Product Service 1 800 522 7658 29 Chapter 6 Maintaining your XPert or XVS Enclosure Guardian 500 Installation Procedure 30 The enclosure comes prepared to except the Guardian 500 Airflow Monitor First remove the large 1 19 dia gray hole plug See Figure 2 See Figure 3 only to reference internal assembly of the airflow monitor Locate the elbow locknut and washer and install it in the 1 19 dia hole per Figure 6 2 and Figure 6 3 The enclosure baffle pivots down to install the elbow washer and locknut Cut the 1 hose supplied with the kit to 15 5 approximate length and install it between the airflow sensor and the elbow Secure the Guardian 500 Airflow Monitor to the enclosure with double stick tape as shown in Figure 6 4 The airway passage between the alarm module and the enclosure is now complete Locate the metal hose cover and install with double stick tape per Figure 6 4 Locate the power supply transformer One end should already be connected to the two pin connector labeled 15 VDC on the back of the alarm module and through the strain relief bushing If disconnected then reconnect to power the airflow monitor Plug the 115V power supply into a standard 115V duplex receptacle the back of the accessory FilterMate portable exhauster or t
32. closure interior Do not overload the work surface with apparatus or work material The safe operation of the enclosure is based upon having proper airflow through the structure Do not place large bulky objects such as block heaters directly on the work surface Instead elevate the object 3 4 on blocks to allow a flow of air under the object and into the rear baffle exhaust slots Ensure blocks are level and secured in place Blocking large portions of the rear baffle will change the airflow pattern in the enclosure causing turbulence Do not store containers or supplies against the rear baffle as this will affect airflow Always work with your hands as far back in the enclosure as possible It is best to keep all powders chemicals and apparatus inside the lower air foil of the enclosure Do not work with chemicals in this enclosure without the exhaust system running Perchloric acid use in this enclosure is prohibited High level radioisotope materials are prohibited in this enclosure Avoid cross drafts and limit traffic in front of the enclosure Air disturbances created may draw contaminants out of the enclosure A qualified certification technician should test the XPert Balance Enclosure or Protector XVS Ventilation Station before it is initially used The enclosure should be validated whenever it is relocated serviced or at least annually thereafter The use of safety goggles protective clothing gloves and any other pe
33. e speed control on the FilterMate or by using an adjustable damper on the exhaust blower Typical alarm conditions are set at face velocities of 10 to 20 feet per minute below the normal operating conditions due to supply air and exhaust air fluctuations as well as room air cross drafts See note 8 if the low airflow volume or sash opening cannot be adjusted 5 Using a properly calibrated thermoanemometer determine the velocity through the face of the enclosure by taking a detailed velocity traverse Divide the face area into equal increments One reading per square foot of face area is normally recommended for an accurate traverse Compute the average velocity for this area 6 Ifthe red light alarm is on slowly turn the adjustment screw counterclockwise until the green light is activated If the green light is on slowly turn the adjustment screw clockwise until the red light comes on Slowly turn the adjustment screw back until the red light is activated It is important that these adjustments be done in small increments at intervals about 30 seconds apart to allow for delayed reaction of the alarm itself The alarm low flow set point should now be set and the red light activated 7 Readjust the enclosure airflow to its normal operating levels The green light should now be activated Calibration is now complete Enclosure Operating In Flow Speed Alarm Condition Set Point Speed 100 fpm 80 90 fpm 80 fpm 60 70 fpm 60 fpm 4
34. ector labeled 15 VDC on the back of the alarm module and through the strain relief bushing If disconnected then reconnect to power the airflow monitor Plug the 115V power supply into a standard 115V duplex receptacle the back of the accessory FilterMate portable exhauster or the back of the accessory light For 230V plug into a standard receptacle with your specific outlet plug It is recommended that the airflow monitor be connected directly to the FilterMate switched auxiliary outlet so the airflow monitor is powered at the same time 7 Installation is now complete 34 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure Remove Large 1 19 Dia Hole Plug Install Elbow Locknut and Washer Figure 6 5 Hole Plug and Mounting Screws Location Airflow Sensor Display Digital 1000 Monitor Figure 6 6 Digital 1000 Airflow Monitor Details Product Service 1 800 522 7658 35 Chapter 6 Maintaining your XPert or XVS Enclosure Accessory Light Enclosure Hose Cover Monitor Figure 6 7 Digital 1000 Airflow Monitor Installation 36 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure Digital 1000 Calibration l Calibrate the airflow monitor according to the instruction manual that comes with the kit To successfully calibrate it will be necessary to change the face velocity by adjusting the airflow exhaust volume The exhaust volume
35. es Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen 15 days of delivery Limitation of Liability The disposal and or emission of substances used in connection with this equipment may be governed by various federal state or local regulations All users of this equipment are required to become familiar with any regulations that apply in the user s area concerning the dumping of waste materials in or upon water land or air and to comply with such regulations Labconco Corporation is held harmless with respect to user s compliance with such regulations Contacting Labconco Corporation If you have questions that are not addressed in this manual or if you need technical assistance contact Labconco s Customer Service Department or Labconco s Product Service Department at 1 800 821 5525 or 1 816 333 8811 between the hours of 7 00 a m and 6 00 p m Central Standard Time Part 3901000 Rev F ECO F773 TABLE Or CONTENTS CHAPTER 1 INTRODUCTION CHAPTER 2 PREREQUISITES Support Vibration and Movement Requirements Temperature Variation Requirements Humidity and Static Electricity Requirements Location and Air Current Requirements Exhaust and Blower Requirements Remote Blowers Roof Mounted FilterMate Portable Exhausters Electrical Requirements Space Requirements CHAPTER 3 GETTING STARTED Unpacking the Enclosure Install the Enclosure on a Supporting Structure amp Work Su
36. g 4863500 Remote Blower 115 volts 60 Hz 4 4 amps 35 16 4863501 Remote Blower 115 230 volts 50 Hz 5 6 2 8 amps 35 16 77053501 Explosion Proof Remote Blower 115 volts 60 Hz 40 18 4 4 amps B 1 4 hp or 1 3 hp belt drive with adjustable sheave to change RPM Corrosion resistant epoxy coated steel housing and wheel Blower inlet is 10 87 ID Outlet dimensions are a rectangular 5 5 x 10 OD See dimensional data in Chapter 7 Catalog HP Description CFM RPM Ranges at Static Pressure in H20 Std EP 38 50 62 75 88 1 00 Remote Blower 7068200 1 4 115V 60 Hz 54000800 410 800 3502870 3902970 430 1060 44A Explosion Proof 7068300 1 4 Blower 115V 720 950 760 1050 710 1060 620 1060 60 Hz 4 5 Remote Blower 7068400 1 3 115V 60 Hz 760004050 710 1060 620 1060 430 1060 460 1150 500 1220 6 1A Explosion Proof 7068500 1 3 Blower 115V 850 1130 825 1160 790 1190 770 1220 740 1260 640 1260 60 Hz 64A If the enclosure is connected directly to a house exhaust system an adjustable damper or valve must be installed to control the airflow properly This is equally important when a house exhaust system is controlling multiple balance enclosures or ventilation stations See Chapter 7 for ordering an accessory adjustable damper Product Service 1 800 522 7658 Chapter 2 Prerequisites FilterMate Portable Exhausters Filtered Room Exhaust or O
37. g Started Connecting to the Vented Enclosure Exhaust System WARNING The weight of the exhaust ductwork system must be supported independently of the enclosure superstructure or damage may occur The exhaust system should be installed by a qualified HVAC contractor The exhaust connection on the 2 through 6 enclosures has been designed to accept a 2 x 10 5 1cm x 25 4cm nominal transition adapter The 8 enclosure has two 2 x 10 connections See Chapter 7 for ordering accessory transition adapters Labconco manufactures transition adapters for either top or bottom exhaust and for both 5 dia hose or 6 dia duct Review Chapter 2 for remote blower and FilterMate Portable Exhauster exhaust prerequisites and review Chapter 7 for ordering blower exhaust equipment For your convenience several exhaust options are shown in Figures 3 2 3 3 3 4 and 3 5 Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and system static pressure loss To ensure compatibility the selected exhaust duct material should match the enclosures procedures and chemical applications Figure 3 2 Top Connection to Outside Exhaust 6 03 ID Exhaust Transition Adapter Airflow monitor moved forward per Chapter 6 Product Service 1 800 522 7658 13 Chapter 3 Getting Started 6 03 ID EXHAUST To Outside Exhaust Airflow monitor moved forward per Chapter 6 Figure 3 3 Front
38. he back of the accessory light For 230V plug into a standard receptacle with your specific outlet plug It is recommended that the airflow monitor be connected directly to the FilterMate switched auxiliary outlet so the airflow monitor is powered at the same time Installation is now complete Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure Remove Large 1 19 Dia Hole Plug Install Elbow Locknut and Washer Figure 6 2 Hole Plug and Mounting Screws Location Display Guardian 500 Monitor Figure 6 3 Guardian 500 Details Product Service 1 800 522 7658 31 Chapter 6 Maintaining your XPert or XVS Enclosure Guardian 500 Calibration Procedure Each alarm module and enclosure fume hood is unique and needs to be individually calibrated in the field The procedure for the adjustment is as follows 1 Double check the installation to make sure that monitor and power supply are properly installed 2 Allow 10 minutes for the monitor to warm up once power has been connected 3 Determine the low flow set point for your monitor This is the value where the monitor will first indicate a low flow condition The red light will be on for this value Refer to your industrial hygiene officer for the proper low flow set point or consult the table below 4 Adjust your enclosure fume hood airflow to the low flow set point as previously determined The exhaust flow can be lowered by adjusting th
39. iculate matter during filter changing operations Carbon Filters Provides granular activated carbon or treated carbon Filter Classification Part Carbon lbs Organic Vapors 3923400 12 activated Formaldehyde 3923401 14 treated Ammonia 3923402 16 treated 8 Product Service 1 800 522 7658 Chapter 2 Prerequisites Electrical Requirements Standard duplex electrical receptacles should be nearby for connecting the FilterMate Portable Exhauster an airflow monitor accessory fluorescent light or other equipment such as a balance for weighing operations For your convenience both the accessory FilterMate and accessory fluorescent light kit have auxiliary outlet receptacles It is recommended that the airflow monitor be connected directly to the auxiliary switched outlet on the rear of the FilterMate so the airflow monitor is ON when the blower is ON The FilterMate auxiliary switched outlet is rated for 8 amps For other blower exhaust systems it is recommended that the airflow monitor be switched by the same circuit as the blower exhaust If this is not possible then the airflow monitor may be connected to the switched auxiliary outlet on the accessory fluorescent light kit The fluorescent light kit auxiliary outlet can be used for any accessory under 8 amps Space Requirements The dimensions for the different models are shown in Appendix B Enclosure Dimensions Product Service 1 800 522 7658 9 Ch
40. ides an RS232 output a night setback mode and several auxiliary relay ports See Chapter 7 for ordering information See Chapter 6 for detailed information installation and calibration 14 Accessory Fluorescent Light Kits provide excellent illumination with an auxiliary outlet plug for supplying power to auxiliary equipment such as a balance or printer See Chapter 7 for ordering information 15 Accessory Roof Mounted Blower sized to provide adequate airflow to XPert XVS enclosures when used with an exhaust damper See Chapters 2 3 and 7 16 Accessory FilterMate Portable Exhauster provides up to 280 cfm for HEPA filtration or up to 220 cfm for combination HEPA Carbon filtration when connected to XPert XVS enclosures Multiple FilterMates must be balanced and can be used on 5 6 and 8 units with multiple exhaust connections See Chapters 2 3 and 7 Product Service 1 800 522 7658 23 Chapter 4 Performance Features and Safety Precautions Safety Precautions Although the enclosure has been engineered to maintain optimum operator safety caution should always be used while working Prior to using the enclosure check to make sure that the exhaust blower is operating and that air is entering the enclosure at its specified face velocity The use of an airflow monitor is recommended to alert the user if there is a problem with airflow Use good housekeeping in the hood at all times Clean up spills immediately Periodically clean en
41. ince static electricity is a surface phenomenon materials can also be classified by their surface resistivity measured in ohms per square The table below lists the surface resistivity of various classes of material Surface Resistivity Table Material Surface Resistivity Example Conductive 0 10 per square Skin Metals Static dissipative 10 10 per square Glass Antistatic 10 gt 10 Q per square Polyethylene bag Insulative 10 per square Acrylic Packing foam Styrofoam Location and Air Current Requirements The XPert Balance Enclosure and XVS Ventilation Station both have been designed to contain hazards by negating typical cross drafts and movements within the opening Air movement does not affect most modern balances with draft shields However as a precautionary measure of safety and a higher level of quality management it is recommended that the enclosure be placed in such an area to avoid e High traffic areas where walking might cause an air disturbance or be a nuisance to balance readings e Overhead or wall HVAC diffusers fans radiators or other lab equipment producing air currents e Next to doorways or windows that may be opened Product Service 1 800 522 7658 5 Chapter 2 Prerequisites Exhaust and Blower Requirements The exhaust connection has been designed to accept a 2 x 10 5 1 x 25 4cm nominal exhaust collar The enclosure has two possible exhaust
42. irflow at the sash opening of your enclosure To determine the actual face velocity at the sash opening airflow velocity readings will need to be taken This should be done across the sash opening of the enclosure in accordance with the Industrial Ventilation Manual See Appendix D The average face velocity is achieved by taking readings in two rows across the enclosure with the readings 6 from the ends and evenly spaced every 12 the first row is 3 down from the upper sash foil and the second row is 3 up from the work surface Labconco recommends an average face velocity at the sash opening of 60 to 100 feet per minute for high performance XPert and Protector XVS enclosures Refer to Chapter 2 for proper airflow volumes for your particular model Your XPert or XVS enclosure has been tested at the factory per ASHRAE 110 1995 All enclosures achieve an as manufactured rating of less than 0 05 part per million ppm at 4 liters per minute Ipm AM 0 05 Consult Labconco for individual ratings For field use ASHRAE testing contact Labconco for a certified on site contractor For additional validation Labconco had independent testing performed to validate the enclosures for sodium naproxen powder containment Labconco also had independent ASHRAE testing performed to validate the enclosures for gas containment For copies of these independent validation reports contact Labconco Customer Service NOTE Face velocity profiles and smo
43. is permanently installed device provides both visual and audible alarms to alert the user of abnormal airflow conditions A green light on the front of the monitor indicates normal flow conditions When flow conditions lower than the set point are encountered a flashing red light is activated along with an audible alarm To temporarily mute the audible alarm press and release the test reset button The Guardian 500 Airflow Monitor has a built in sensor 3 relay inputs and 1 relay output The relay inputs can be configured for night setback external alarm and sash high alarm The night setback features disables the alarm The relay output is configured for Low Air Alarm 28 Product Service 1 800 522 7658 Chapter 6 Maintaining your XPert or XVS Enclosure Guardian 500 Component Identification 1 Air Inlet A portion of the air coming into the enclosure passes through the air inlet and across the flow sensors Normal Flow Indicator This green light indicates normal flow conditions Alarm Indicator Mute Button This red light is activated approximately 10 30 seconds after the low flow set point is reached Low flow set points are 10 20 fpm below normal flow If no alarm is present this button will cause the red lamp to light and the audible alarm to sound If an alarm is present the button will silence the audible alarm Adjustment for Alarm This potentiometer is used to set the low flow Set Point indicators for the al
44. ke testing should be performed frequently per your organizations quality system to ensure safe performance Product Service 1 800 522 7658 17 Chapter 4 High Performance Features and Safety Precautions High Performance Features The patented U S Patent No 6 461 233 XPert Balance Enclosure and Protector XVS Ventilation Station are designed to meet the needs of the laboratory scientist and provide superior containment while conserving energy at OSHA approved low flow velocities as low as 60 feet per minute The XPert XVS Enclosure has been designed to effectively contain toxic noxious or other harmful materials when properly installed and operated What makes the XPert XVS Enclosure so unique is the revolutionary way it directs air into and through the contaminated air chamber Labconco engineered the XPert XVS to minimize the effects of turbulence The containment enhancing and aerodynamic designs of the upper sash foil side air foils lower air foil upper dilution air supply and rear perforated baffle all work in concert to produce horizontal airflow patterns that significantly reduce both powder and chemical concentrations through the work area These concentrations of materials are predominantly removed on the first pass of airflow through the chamber resulting in high performance containment 18 Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions Airflow monitor moved forward
45. lamps and Hose Kits Provides Alternatives for ducting Catalog 4868600 1921000 3927500 3927600 1921500 Description 8 Feet of 5 ID gray flexible polypropylene hose included with FilterMate General purpose chemical resistant hose suitable in pharmaceutical applications 5 T Bolt Hose Clamp two included with FilterMate 8 Feet of 5 ID clear smooth bore static dissipation hose with two hose clamps Suitable in clean rooms or pharmaceutical applications 8 Feet of 6 ID black thermoplastic hose with two 6 09 ID cuffs Include two 6 hose clamps Suitable for connection to house exhaust in pharmaceutical applications 6 T Bolt Hose Clamp Product Service 1 800 522 7658 47 Cha pter 8 Troubleshooting Refer to the following table if your XPert Balance Enclosure System or Protector XVS Ventilation Station fails to operate properly If the suggested corrective actions do not solve your problem contact Labconco for additional assistance PROBLEM CAUSE CORRECTIVE ACTION Contaminants Improper user techniques See Certifying the Hood Chapter 3 and Safety outside of enclosure Airflow monitor malfunction for the enclosure Restriction of the baffle air slots or blockage of the exhaust outlet External factors are disrupting the enclosure airflow patterns or acting as a source of contamination Enclosure has improper face velocity No power No lights No display
46. libration set point to confirm that the source of constant air volume is providing the desired face velocity for the enclosure The average face velocity for the high set point is accurately measured by dividing the opening of the enclosure into equal area grids consisting of at least 3 data collection points and measuring the velocity at the center of each grid with a calibrated thermo anemometer Enter the high value set point on the display The high value must be at least 10 fpm greater than the low value The high exhaust volume calibration will take about 5 seconds 9 Be sure to enter DONE after successfully completing the low and high calibration set points 10 Once calibration is completed go to RUN and hit ENTER The value should read close to the high calibration set point 11 With the sash fully open the Digital 1000 monitor should go into RED LED low air alarm if successfully calibrated Product Service 1 800 522 7658 39 Chapter 6 Maintaining your XPert or XVS Enclosure Note Enter Button stores information and Buttons allow for scrolling 1 Push the ENTER Button on the face of the alarm until the SET UP Menu is displayed 2 Scroll to SET UP and hit ENTER 3 The PASSWORD MENU displays The Password is 0000 Press the ENTER button repeatedly until the CAL CONFIG MENU is displayed 4 In CAL CONFIG MENU set the following CALIBRATION CONFIGURATION MENU
47. lly calibrate it will be necessary to change the face velocity by opening and closing the enclosure s sash The airflow monitor is factory set to be calibrated with a difference of at least 50 fpm and can be changed by adjusting the lower higher air sample difference The in flow face velocity speeds provided in the chart below are suggested to successfully calibrate the Digital 1000 2 Before proceeding with calibration it will first be necessary to configure the airflow monitor Go to the setup and then CAL CONFIG MENU and adjust the lower higher air sample difference to 10 fpm This will allow you to successfully calibrate with minimum difference values of 10 fpm 3 While in CAL CONFIG MENU change the sensor difference from 10 to 3 4 While in CAL CONFIG MENU adjust the red low alarm to the desired setting See the chart below for range and suggested settings Then adjust the yellow CAUTION or WARNING to the desired setting See the chart below for range and suggested settings Then adjust the CAUTION or WARNING air reset to 3 fpm This sets the alarm condition CAUTION WARNING LOW ALARM Low Calibration High Calibration YELLOW LED Setting RED LED Set Point Set Point Setting Sash Open or Enclosure Range Suggested Range Suggested fpm Operating user defined Set Point user defined Set Point 8 Sash 10 Sash Inflow Speed fpm fpm fpm fpm Height Height Sash Closed 43
48. ly trained The wrench icon precedes the service operations that require qualified technicians Product Service 1 800 522 7658 27 Chapter 6 Maintaining your XPert or XVS Enclosure Routine Maintenance Schedule Weekly e Clean the enclosure interior appropriate for the application e Operate the exhaust system noting the airflow velocity through the enclosure using a source of visible smoke Airflow alarms are recommended for constant monitoring Monthly or more often as required e Determine the actual face velocity through the sash opening of the enclosure where the average reading should be at the specified velocity Use calibrated thermal anemometer or other approved apparatus Airflow alarms are recommended for constant monitoring e Using cloth and glass cleaner clean the exterior surfaces of the enclosure particularly the front of the enclosure to remove any accumulated dust e The enclosure rear baffle should be checked for any blockage to ensure that the enclosure is maintaining proper airflow e All weekly activities Annually e Replace the fluorescent lamps on accessory light kit e Have the enclosure validated by a qualified certification technician See Validating the Vented Enclosure in Chapter 3 e All monthly activities Guardian 500 Airflow Monitor Kit No 3944700 or 3944701 The Guardian 500 Airflow Monitor is designed to continuously monitor airflow through enclosures and fume hoods Th
49. o Unauthorized returns will not be accepted e fenclosure was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages e The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen 15 days of delivery e Do not discard the packing material until you have checked all of the components and tested the unit e Werecommend that you do not remove the enclosure from its shipping container until it is ready to be placed into its final location Move the unit by placing a flat low dolly under the shipping skid or by using a floor jack NOTE DO NOT MOVE THE ENCLOSURE BY TILTING IT ONTO A HAND TRUCK Install the Enclosure on a Supporting Structure and Work Surface Note Use caution when lifting or moving the enclosure When installing the enclosure onto a work surface or benchtop ensure that the structure can safely support the combined weight of the enclosure and any related equipment The work surface should be at least as wide as the enclosure to properly support it The front of the enclosure should be aligned within 1 8 of the front of the work surface Mounting holes are provided in the Labconco accessory work surfaces to secure the enclosure Product Service 1 800 522 7658 11 Chapter 3 Getting Started Work Surface Specifications The work surface should be sm
50. ooth rigid and durable such as a chemically resistant epoxy resin or stainless steel The surface should be non porous and resistant to the acids solvents and chemicals used in conjunction with the XPert or XVS enclosure The work surface should also contain a dished recessed area for containing primary spills Work Surface and Vented Enclosure Installation 1 Level the base cabinets and the work surface Work surface should be placed flush with the front of the base cabinet as shown in Figure 3 1 2 Position the work surface in its intended location and with the front of the work surface wide flange or spill trough towards you 3 Secure the work surface to the base cabinet with a structural adhesive or silicone sealant 4 Insert the supplied mounting screws in the four holes for epoxy work surfaces For stainless steel work surfaces use 10 24 x 3 8 hex machine screws Labconco P N 1885806 Allow a minimum of 1 8 clearance under the head of the screw for positioning the enclosure 5 Place the enclosure on the work surface and slide the rear flange and front air foil flanges under the mounting screw heads 6 Tighten the four screws to complete the installation Airflow monitor moved forward per Chapter 6 2878 ENCLOSURE DEPTH 2900 TOTAL DEPTH WITH WORK SURFACE Figure 3 1 e Side View of Typical Installation 12 Product Service 1 800 522 7658 Chapter 3 Gettin
51. pment from Labconco whichever is sooner SteamScrubber amp FlaskScrubber Glassware Washers carry a two year warranty from date of installation or three years from date of shipment from Labconco whichever is sooner Blood Drawing Chairs carry a ten year warranty Carts carry a lifetime warranty Glassware is not warranted from breakage when dropped or mishandled This limited warranty covers parts and labor but not transportation and insurance charges In the event of a warranty claim contact Labconco Corporation or the dealer who sold you the product If the cause is determined to be a manufacturing fault the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation Under no circumstances shall Labconco Corporation be liable for indirect consequential or special damages of any kind This statement may be altered by a specific published amendment No individual has authorization to alter the provisions of this warranty policy or its amendments Lamps and filters are not covered by this warranty Damage due to corrosion or accidental breakage is not covered Returned or Damaged Goods Do not return goods without the prior authorization from Labconco Unauthorized returns will not be accepted If your shipment was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages The United Stat
52. re 2005 and subsequent I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s See individual Declaration of Conformity which will be signed by the importer for your country Place Signature Date Full Name Position Labconco P N 3696044 Rev A Product Service 1 800 522 7658 55
53. res Unique Flush Sash has a wiping seal to contain contaminants and features a spring loaded latch or gas spring for loading auxiliary equipment The sash pivots down for normal operation Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions 10 Electrical Pass through Iris allows electrical cords and data cords to pass through the back of enclosure without leaving a large hole for contaminants to escape The unit ships with solid plugs and the iris plugs are included with the instruction manual for your convenience 11 Shipped fully assembled to eliminate the need for costly onsite assembly Accessories such as the exhaust transition connections work surfaces airflow monitors and fluorescent light kits are easily installed 12 Accessory Exhaust Connections The XPert XVS enclosures feature two exhaust locations One is located on the top of the rear plenum and the other is located on the bottom of the rear plenum Exhaust transition connectors are available for either a 6 OD outside exhaust or a 5 ID hose 5 6 and 8 enclosures have multiple exhaust connections All exhaust transition connectors are reviewed in Chapters 2 and 3 and accessories displayed for order in Chapter 7 13 Accessory Guardian Airflow Monitors continuously monitor airflow An audio visual alarm alerts the user to low airflow conditions The Guardian Digital Airflow Monitor also displays a face velocity value prov
54. rface Connecting to the Vented Enclosure Exhaust System Connecting the Electrical Supply Sealing the Vented Enclosure to the Work Surface Validating the Vented Enclosure CHAPTER 4 HIGH PERFORMANCE FEATURES AND SAFETY PRECAUTIONS High Performance Features Safety Precautions CHAPTER 5 USING YOUR ENCLOSURE OO CL gt 18 18 24 25 CHAPTER 6 MAINTAINING YOUR ENCLOSURE Routine Maintenance Schedule Guardian 500 Airflow Monitor Kit No 3944700 or 3944701 Guardian 500 Component Identification Guardian 500 Installation Procedure Guardian 500 Calibration Procedure Guardian 500 Alarm Activation Guardian Digital 1000 Kit No 3908800 or 3908801 Guardian Digital 1000 Operation Guardian Digital 1000 Installation Procedure Guardian Digital 1000 Calibration Digital 1000 Alternate Calibration Procedure Constant Volume Conditions CHAPTER 7 ACCESSORIZING YOUR ENCLOSURE Work Surfaces Guardian Digital Airflow Monitor or Guardian 500 Fluorescent Light Kit Exhaust Transition Adapters amp Dual Exhaust Connectors Remote Blowers Exhaust Dampers FilterMate Portable Exhausters and Filters Storage Cabinets Utility Shelf Kit Hose Hose Clamps and Hose Kits CHAPTER 8 TROUBLESHOOTING APPENDIX A REPLACEMENT PARTS APPENDIX B DIMENSIONS APPENDIX C SPECIFICATIONS APPENDIX D BALANCE STABILITY DECLARATION OF CONFORMITY 27 28 28 29 30 32 32 34 34 34 37 38 4l 43 43 43 44 44 45 45
55. rk or FilterMate Portable Exhauster must be ordered separately Product Service 1 800 522 7658 1 Chapter 2 Prerequisites Before you install the vented enclosure you need to prepare your site for installation Carefully examine the location where you intend to install the vented enclosure You must be certain that the area is level and of solid construction In addition a dedicated source of electrical power should be located near the installation site to power the accessory FilterMate Portable Exhauster balance or other apparatus Additionally a balance should be strategically placed in the lab to provide efficient workflow Carefully read this chapter to learn the requirements for your installation site e The support vibration and movement requirements e The temperature variation requirements e The humidity and static electricity requirements e The location and air current requirements e The exhaust and blower requirements e The electrical power requirements e The space requirements Refer to Appendix B XPert XVS Dimensions for complete enclosure dimensions Refer to Appendix C XPert XVS Specifications for complete vented enclosure electrical and environmental conditions specifications and requirements 2 Product Service 1 800 522 7658 Chapter 2 Prerequisites Support Vibration and Movement Requirements At a minimum the supporting structure usually consists of a base cabinet and chemically resistant work
56. rsonal protective equipment recommended by your safety officer should be used The sash should remain in the down position while using the enclosure Proper operation of the enclosure depends largely upon its location and the operator s work habits Consult the references in Appendix D 24 Product Service 1 800 522 7658 Chapter 5 Using Your Xpert Balance Enclosure or Protector XVS Ventilation Station Planning e Thoroughly understand procedures and equipment required before beginning work e Arrange for minimal disruptions such as room traffic or entry into the room while the enclosure is in use Start up e Turn on exhaust system and accessory light if so equipped Only raise the sash for loading Check the baffle air slots for obstructions Allow the enclosure to operate unobstructed for 1 minute Wear a long sleeved lab coat and rubber gloves Use protective eyewear Wear a protective mask if appropriate Consult your Safety Officer for additional personal protective equipment recommendations Loading Materials and Equipment e Only load the materials required for the procedure Do not overload the enclosure e Do not obstruct the air foil or rear baffle slots e Large objects should not be placed close together and should be elevated above the work surface to permit airflow to sweep under the equipment e After loading wait one minute to purge airborne contaminants from the work area Product Service 1 800 522 7
57. tes for drift and vibration If in doubt schedule recalibration 2 Shield is not tight on the balance If the balance shield is loose air can get into the weighing chamber and cause fluctuation in sensitive weigh applications In other words air can get into the balance and cause the weigh plate to fluctuate The best way to address the problem is to have the balance servicer fix the balance shield 3 There is no shield on the balance Sometimes open balances without shields fluctuate Try running the ventilated balance enclosure at slower speeds such as 60 70 feet per minute getting a shield placed around the balance or using a balance with a tight shield 54 Product Service 1 800 522 7658 DECLARATION OF CONFORMITY Application Council Directive s 73 23 89 336 EEC 2002 95 EC ROHS 2002 96 EC WEEE Standard s to which conformity is declared EN61010 1 EN61326 1 Manufacturer s Name Labconco Corporation Manufacturer s Address 8811 Prospect Avenue Kansas City MO 64132 USA Importer s Name See Shipping Customs Documents Importer s Address See Shipping Customs Documents for your equipment Type of Equipment Laboratory Equipment XPert Bulk Powder Enclosures 3963500 3963600 3963800 Series XPert Bulk Powder Enclosures 3961500 3961600 Series XPert Bulk Powder Filtered Systems 3962500 3962600 Series XPert Bulk Powder Filtered Stations Serial No Various See Individual Declaration Year of Manufactu
58. the audible alarm can be silenced If flow conditions fluctuate near the alarm set point the airflow monitor will automatically reset itself Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower 48 Product Service 1 800 522 7658 Appendix A Replacement Parts The following illustrations indicate the replacement parts Item Qty Part Number Description 1A 1 3906500 Glass Side Short 1B 1 3926500 Glass Side Tall 1 1 3976400 Glass Side Deep 1D 1 3976401 Glass Side Deep w hole 2A 1 3906800 Glass Top 2 2B 1 3906801 Glass Top 3 2C 1 3906802 Glass Top 4 2D 1 3906803 Glass Top 5 2E 1 3906804 Glass Top 6 2F 1 3906805 Glass Top 8 2G 1 3906806 Glass Top 5 Deep 2H 1 3906807 Glass Top 6 Deep 21 1 3906808 Glass Top 8 Deep 3A 1 3906100 Glass Sash 2 Short 3B 1 3906101 Glass Sash 3 Short 3C 1 3906102 Glass Sash 4 Short 3D 1 3926100 Glass Sash 2 Tall 3E 1 3926101 Glass Sash 3 Tall 3F 1 3926102 Glass Sash 4 Tall 3G 1 3967203 Glass Sash 5 3H 1 3967204 Glass Sash 6 31 1 3967205 Glass Sash 8 3J 1 3967206 Glass Sash 5 Tall 3K 1 3967207 Glass Sash 6 Tall 3L 1 3967208 Glass Sash 8 Tall 4A 1 3907600 Foil Sash 2 4B 1 3907601 Foil Sash 3 4C 1 3907602 Foil Sash 4 4D 1 3967100 Foil Sash 5 4E 1 3
59. utside For filtered exhauster selection Labconco offers FilterMate Portable Exhausters capable of exhausting up to 280 cfm of HEPA filtered air or up to 220 cfm of combination HEPA Carbon filtered air These airflows are possible when connected to the enclosure with 8 of 5 diameter hose HEPA and Carbon filters are available and detailed below as well as in Chapter 7 For airflows greater than these listed use multiple FilterMates Labconco offers FilterMate Portable Exhausters listed below and in Chapter 7 FilterMate Portable Exhausters Number Voltage Filter Airflow cfm Connection 3970001 None Canopy to 3970002 115 Volt 60 Hz HEPA 280 outside 3970003 115 Volt 60 Hz HEPA Carbon None 3970004 115 Volt 60 Hz Carbon Carbon None 3970020 230 Volt 50 Hz HEPA 280 3970021 230 Volt 50Hz 280 Noe None None asovowsone am za _ 3970022 230 Volt 50 Hz HEPA 280 outside None None 3970023 230 Volt 50 Hz HEPA Carbon 220 Nom 3970024 230 Volt 50 Hz Carbon Carbon 220 Nom If the outlet of the FilterMate Exhauster is connected to the house exhaust system it is recommended that 3970002 115 Volt or 3970022 230 Volt be ordered as a thimble connection is built into the exhaust with a 6 nominal exhaust HEPA Filter Part 3707900 is 99 99 efficient for particles 0 3 micron HEPA Filter with Bag In Bag Out Bag Part 23779002 Helps contain part
Download Pdf Manuals
Related Search
Related Contents
MANUALE Delta Electronics Programmable Logic Controller DVP-PLC User's Manual Steren 253-200BK Installation Manual 2 and 4 Flavor FCB Pinnacle NEC P42XP10-BK User's Manual Philips PET703 User's Manual Philips Candle Bent tip energy saving bulb 872790092664400 Manuel du propriétaire en ice maker fr machine à glaçons es productor de hielo Copyright © All rights reserved.