Contents

1. 60V DC 24V AC 2 EX830 EU EN V5 0 3 09 CAUTIONS e Improper use of this meter can cause damage shock injury or death Read and understand this user manual before operating the meter e Always remove the test leads before replacing the battery or fuses e Inspect the condition of the test leads and the meter itself for any damage before operating the meter Repair or replace any damage before use e Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC These voltages are considered a shock hazard e Always discharge capacitors and remove power from the device under test before performing Diode Resistance or Continuity tests e Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts Other means should be used to ensure that the terminals are not live e If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired e This device is not a toy and must not reach children s hands It contains hazardous objects as well as small parts that the children could swallow In case a child swallows any of them please contact a physician immediately e Do not leave batteries and packing material lying around unattended they can be dangerous for children if they use them as toys e Incase the device is going to be unused for an extended period
2. and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is not to be used for 60 days or more remove the battery and store it separately Battery Replacement 1 Remove the Phillips head screw that secures the rear battery door 2 Open the battery compartment 3 Replace the 9V battery 4 Secure the battery compartment You as the end user are legally bound Battery ordinance to return all used batteries and accumulators disposal in the household garbage is prohibited You can hand over your used batteries accumulators at collection points in your community or wherever batteries accumulators are sold Disposal Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle 10 EX830 EU EN V5 0 3 09 Specifications Function Range amp Resolution Accuracy of reading digits AC Current 400 0 AAC 2 5 8d 50 60 Hz 1000 AAC 2 8 5d 400 0 ADC 2 5 5d DC Current 1000 ADC 2 8 5d 400 0 mVAC 1 0 10d AC Voltage _ 4 000 VAC 50 60Hz 40 00 VAC 1 5 5d 400 0 VAC 600 VAC 2 0 5d 400 0 mVDC 0 8 2d 4 000 VDC DC Voltage 40 00 VDC 1 5 2d 400 0 VDC 600 VDC 2 0 2d 400 00 1 0 4d 4 000kQ 40 000kQ 1 5 2d Resistance 400 0kQ 4 000MQ 2 5 3d 40 00MQ 3 5 5d 4 000nF 5 0 30d 40 00nF 5 0 20d 400 0
3. 41 F to 104 F 5 C to 40 C 4 F to 140 F 20 C to 60 C Max 80 up to 87 F 31 C decreasing linearly to 50 at 104 F 40 C lt 80 7000ft 2O0Ometers maximum One 1 9V Battery NEDA 1604 After approx 25 minutes 10 6x4 3x2 270x110x50mm 13 6 oz 886g For indoor use and in accordance with the requirements for double insulation to IEC1010 1 2001 EN61010 1 2001 Overvoltage Category III 600V and Category II 1000V Pollution Degree 2 U S Patent 7163336 Copyright 2009 Extech Instruments Corporation All rights reserved including the right of reproduction in whole or in part in any form 12 EX830 EU EN V5 0 3 09
4. User s Guide EXT INSTRUMENTS A FLIR COMPANY Extech EX830 True RMS 1000 Amp Clamp Meter with IR Thermometer COM YenicaP 20 eao CE Patented Introduction Congratulations on your purchase of the Extech EX830 True RMS 1000A Clamp Meter This meter measures AC DC Voltage AC DC Current Resistance Capacitance Frequency Duty Cycle Diode Test Continuity Type k thermocouple thermometer plus Non Contact IR Temperature Proper use and care of this meter will provide many years of reliable service Safety International Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information A This symbol adjacent to a terminal indicates that under normal use hazardous voltages may be present I Double insulation SAFETY NOTES e Do not exceed the maximum allowable input range of any function e Do not apply voltage to meter when resistance function is selected e Set the function switch OFF when the meter is not in use e Remove the battery if meter is to be stored for longer than 60 days WARNINGS e Set function switch to the appropriate position before measuring e When measuring volts do not switch to current resistance modes e Do not measure current on a circuit whose voltage exceeds 600V e When changing ranges always disconnect the test leads from the circuit under test Function Maximum Input Type K Temperature
5. ac or Voc position Use the MODE bution to select AC or DC Voltage EX830 only Connect the test leads in parallel to the circuit under test a oe w oy Read the voltage measurement on the LCD display 5 EX830 EU EN V5 0 3 09 Resistance Measurements Note Remove power before making resistance measurements 1 Insert the black test lead into the negative COM terminal and the red test lead into the Q positive terminal 2 Set the function switch to the Q position 3 Touch the test probe tips across the circuit or component under test 4 Read the resistance on the LCD display Capacitance Measurements WARNING To avoid electric shock discharge the capacitor under test before measuring If dlISC appears in the display remove and discharge the capacitor 1 Set the function switch to the 1 capacitance position 2 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the CAP positive jack Press MODE to zero any stray capacitance 3 4 Touch the test probe tips across the part under test 5 Read the capacitance value in the display 6 The display will indicate the proper decimal point and value Note For very large values of capacitance measurement time can be several minutes before the final reading stabilizes Frequency Measurements 1 Set the function switch to the V Hz Positio
6. e prefixes nano milli micro mega and kilo Continuity test Diode test Dee Laser pointer 4 EX830 EU EN V5 0 3 09 Operation NOTES Read and understand all Warning and Caution statements in this operation manual prior to using this meter Set the function select switch to the OFF position when the meter is not in use AC Current Measurements WARNING Ensure that the test leads are disconnected from the meter before making current clamp measurements 1 Set the Function switch to the Aac or ADc range 2 Press the trigger to open the jaw Fully enclose only one conductor For optimum results center the conductor in the jaw 3 The clamp meter LCD will display the reading DCA ZERO The DC Zero feature removes offset values and improves accuracy for DC current measurements To perform a zero select ADC and with no conductor in the jaw 1 Press the DC ZERO button to zero the display ZERO will appear in the display The offset value is now stored and removed from all measurements 2 To view the stored value press the DC ZERO button ZERO will flash and the stored value will be displayed Correct Incorrect 3 To exit this mode press and Hold the ZERO button until ZERO is no longer in the display AC DC Voltage Measurements 1 Insert the black test lead into the negative COM terminal and the red test lead into the positive V terminal Set the function switch to the V
7. etc can obscure measurements 6 To find a hot spot aim the meter outside the area of interest then scan across in an up and down motion until the hot spot is located 8 EX830 EU EN V5 0 3 09 Data Hold To freeze the LCD reading press the HOLD button While data hold is active the HOLD icon appears on the LCD Press the HOLD button again to return to normal operation Peak Hold The Peak Hold function captures the peak AC or DC voltage or current The meter can capture negative or positive peaks as fast as 1 millisecond in duration Turn the function switch to the A or V position Use the MODE button to select AC or DC Allow time for the display to stabilize Press and Hold the PEAK button until CAL appears in the display This procedure will zero the range selected Press the PEAK button Pmax will display ee The display will update each time a higher positive peak occurs Press the PEAK button again Pmin will display The display will now update and indicate the lowest negative peak 8 To return to normal operation press and hold the PEAK button until the Pmin or Pmax indicator switches off Note If the Function switch position is changed after a calibration the Peak Hold calibration must be repeated for the new function selected MAX MIN 1 Press the MAX MIN key to activate the MAX MIN recording mode The display icon MAX will appear The meter will display and hold the maximum reading and will u
8. lack test lead into the negative COM terminal and the red test lead into the Q positive terminal 2 Set the function switch to the position 3 Use the MODE button to select continuity The display icons will change when the MODE button is pressed 4 Touch the test probe tips across the circuit or component under test 5 If the resistance is lt 40Q a tone will sound Diode Test 1 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the gt positive jack 2 Turn the function switch to gt F position Use the MODE button to select the diode function if necessary diode symbol will appear on the LCD when in Diode test mode 3 Touch the test probe tips to the diode or semiconductor junction under test Note the meter reading 4 Reverse the test lead polarity by reversing the red and black leads Note this reading 5 The diode or junction can be evaluated as follows If one reading displays a value typically 0 400V to 0 900V and the other reading displays OL the diode is good If both readings display OL the device is open f both readings are very small or 0 the device is shorted 7 EX830 EU EN V5 0 3 09 Non Contact InfraRed Temperature Measurements 1 Set the function switch to the IR Temp position 2 Aim the infrared sensor rear of the meter at the surface to be measured 3 Press the button in the center of the rotary function
9. n 2 Press and hold the MODE button to select the Frequency Hz function k Hz will appear in the display 3 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the Hz positive jack erie Touch the test probe tips across the part under test Read the Frequency value on the display The display will indicate the proper decimal point and value S Oo mM eR Press and hold the MODE button again to return to the voltage mode 6 EX830 EU EN V5 0 3 09 Type K Temperature Measurements WARNING To avoid electric shock be sure the thermocouple probe has been removed before changing to another measurement function Note An open input or a temperature overrange the meter will display Set the function switch to the K Temp position Insert the Temperature Probe into the negative COM and the positive TEMP jacks observing polarity Touch the Temperature Probe tip to the device under test Continue to touch the part under test with the probe until the reading stabilizes Read the temperature on the display The digital reading will indicate the proper decimal point and value OL and beep Note See the Temperature Units paragraph to select F or C Note The temperature range of the supplied thermocouple probe is 20 to 250 C 4 to 482 F Continuity Measurements 1 Insert the b
10. nF Capacitance 4 000uF 3 0 5d 40 00uUF 400 0uF 4 0 10d 4 000mF 10 10d 4 000kHz 1 5 2d Sensitivity 100V lt 50Hz 50V 50 to 400Hz 5V 401Hz to 4000Hz Frequency 11 EX830 EU EN V5 0 3 09 Function Temperature type K Range and Resolution Accuracy of reading digits 4 to 1400 F 3 rdg 9 F 20 to 760 C 3 rdg 5 C 4 to 518 F F C 20 to 270 Clamp jaw opening Display Continuity check Diode test Low Battery indication Over range indication Measurement rate PEAK Thermocouple sensor IR Spectral response IR Emissivity IR distance ratio Input Impedance AC bandwidth AC response Crest Factor Operating Temperature Storage Temperature Operating Humidity Storage Humidity Operating Altitude Battery Auto power OFF Dimensions amp Weight Safety Patent notice 2 0 reading or 4 F whichever is gt 2 0 reading or 2 C whichever is gt 1 7 43mm approx 3 3 4 digits 4000 counts backlit LCD Threshold 409 Test current lt 0 5mA Test current of 0 3mA typical Open circuit voltage lt 3VDC typical Battery symbol is displayed OL display 2 readings per second nominal Captures peaks gt 1ms Type K thermocouple required 6 to 16um 0 95 fixed 8 1 10MQ VDC and VAC 50 to 400Hz AAC and VAC True rms AAC and VAC 3 0 in 40A and 400A ranges 1 4 in 1000A range 50 60Hz and 5 to 100 of range
11. of time remove the batteries to prevent them from training e Expired or damaged batteries can cause cauterization on contact with the skin Always therefore use suitable hand gloves in such cases e See that the batteries are not short circuited Do not throw batteries into the fire e Do not directly view or direct the laser pointer at an eye Low power visible lasers do not normally present a hazard but may present some potential for hazard if viewed directly for extended periods of time 3 EX830 EU EN V5 0 3 09 Description Meter Description Current clamp Clamp opening trigger 1 Data Hold Button Mode Peak Range DCA Zero EX830 only MIN MAX Backlit LCD Display Test lead input jacks IR thermometer and laser pointer rear Backlight Button Laser pointer button Function switch en oe eS ee ee eee m Oo eh fo oO OO Ww Display icons Description HOLD Data Hold Minus sign Negative reading display 0 to 3999 Measurement display digits ZERO Zero s Aaa IR TEMP MAX MIN P Pmn gt MAX PMIN AUT Auto R d o ia fir cat anaes AUTO HOLD ZERO numF Mk Q Hz Current So Gp P lt p C MAX Max reading E all E Ii j MIN Min reading an A Low battery mV or V Milli volts or Volts Voltage mV ee 4 Ga a Q Ohms Resistance A Amperes Current F Farad Capacitance Hz Hertz Frequency F and C Fahrenheit and Celsius units Temperature n m u M k Unit of measur
12. pdate only when a new max occurs 2 Press the MAX MIN key and MIN will appear The display icon MIN will appear The meter will display and hold the minimum reading and will update only when a new min occurs 3 Press the MAX MIN key and a blinking MAX MIN will appear The meter will display the present reading but will continue to update and store the max and min readings 4 To exit MAX MIN mode press and hold the MAX MIN key for 2 seconds Temperature Units F C Units Switch The temperature units selection switch is located in the battery compartment To change the units remove the battery door lift m out the battery and set the switch for the desired units LCD Backlight Button The LCD is equipped with backlighting for easier viewing especially in dimly lit areas Press the backlight button to turn the backlight on Press again to turn the backlight off Automatic Power OFF In order to conserve battery life the meter will automatically turn off after approximately 25 minutes To turn the meter on again turn the function switch to the OFF position and then to the desired function position 9 EX830 EU EN V5 0 3 09 Maintenance WARNING To avoid electrical shock disconnect the meter from any circuit remove the test leads from the input terminals and turn OFF the meter before opening the case Do not operate the meter with an open case Cleaning
13. switch to turn on the laser pointer and identify the surface spot to be measured 4 The area of the surface to be measured must be larger than the spot size as determined by the distance to spot size specification 5 Read the temperature in the display Note See the Temperature Units paragraph to select F or C WARNING Do not directly view or direct the laser pointer at an eye Low power visible lasers do not normally present a CAUTION LASER RADIATION GO NOT STARE INTO BEAM DIODE LASER lt 1mW Output at 630 to 670nm CLASS II LASER PRODUCT hazard but may present some potential for hazard if viewed directly for extended periods of time IR Spot to Distance Diagram The 8 1 spot to distance ratio determines the size of the measured surface area with respect to the distance the meter is held away from the surface Diameter of spot 7 5cm 1 25em ZSM ny S 60cm gt Distance to object IR Measurement Notes 1 The object under test should be larger than the spot target size calculated by the field of view diagram 2 Ifthe surface of the object under test is covered with frost oil grime etc clean before taking measurements 3 Ifan object s surface is highly reflective apply masking tape or flat black paint to the surface before measuring 4 The meter may not make accurate measurements through transparent surfaces such as glass 5 Steam dust smoke

Download Pdf Manuals

image

Related Search

Related Contents

Cables Direct B6LZ-603 networking cable  電子音目覚まし時計 取扱説明書 保 証 書  Equip Cat.5e U/UTP Patch Cable  Samsung 1000P Bruksanvisning  Cobalt Qube 3 Disaster Recovery  Gigabit Ethernet - KnowledgeBase Manager Pro  DeLOCK 95001  SOLDERING IRON & ACCESSORIES SET  Pro-Vent® Plus Luer avec aiguille Needle  PISO-P8R8U/P16R16U User`s Manual  