1. normal to feel less current after a few minutes of stimulation with specific parameters In such a case feel free to increase the intensity in order to get the best results e Frequency and pulse width Frequencies and pulse widths differ depending on the programs Therefore the intensity cannot be set at the same lever for all the programs even if the electrodes are placed in the same position The higher the frequency Is set the lower you will be able to set the intensity The higher the pulse width is set the lower you will be able to set the Intensity 13 schwa medico SPORECUP XTR2 General information Adjustments modifications and repairs The manufacturer is only responsible for the safety and performance of the XTR2 when readjustments alterations and repairs are carried out by authorised persons and when the XTR2 Is used in accordance with the operating instructions Warranty Legal right of warranty is applied according to German Civil Code Guarantee The manufacturer issues a guarantee of 12 months from the date of purchase The guarantee does not apply in the following cases e damage due to improper handling e defects the customer is aware of on the date of purchase e damage caused by the customer e for wearing parts and consumable supplies like for instance battery cables and electrodes Maintenance and cleaning No special cleaning or care agents are required for the XTR2 Clean the stimulator with a so
2. 0 60 40 resistance Q 500 1000 1500 2000 2500 schwa medico SPORECUP XTR2 Operating the SPORECUP XTR2 Connecting cables and electrodes e Connect the electrodes with the cable s always 2 electrodes per cable e Connect the cable s to the XTR2 no 6 e Place the electrodes on the desired area see examples on chapter Electrode Placement Switching ON the XTR2 e Switch the commutator no 8 to the ON position e Switch on the stimulator by pressing the key no 5 The program shown on the display is the last which was used during the previous session If the display shows a key symbol on the upper left side the XTR2 has previously locked a specific program see chapter locking the key pad Selecting a program Select a program by pressing the key To scroll through the programs press the key as often as necessary until you have selected the program of your cholce lo return to the previous program press the left keys Selecting a program is only possible if the keypad has not been locked on one specific program see chapter locking the key pad P D SET TIMER Hip Sos SVT n AUT UU Selecting an area of stimulation Depending on the selected program you can choose between upper and lower limbs Press the E key to change the pulse width of the selected programs The pulse width is increased for lower limbs Programs with a possibility of upper and lower limbs stimulation P1 P5 P6 P7 P9
3. P10 P15 P29 Recommendation Higher pulse width for lower limb stimulation Lower pulse width for upper limb stimulation schwa medico SPORECUP XTR2 Setting the intensity Start the stimulation by increasing the intensity with the key of the channel s connected to the electrodes key to increase the intensity key to decrease the Intensity The intensity of each channel is displayed on the screen The intensity can be set between O and 100 mA By keeping the intensity keys pressed the intensity will increase decrease more rapidly P 0 ser rIMER Di Sus ILI E Terminating the stimulation You can terminate the stimulation at any time by pressing the key When the therapy time Is completed the stimulation terminates automatically If you press the key the stimulation stops and the XTR2 switches off Locking the key pad Choose the desired program by pressing the key see chapter selecting a program Press and hold for 3 seconds the key and the right key to lock the stimulator on the chosen program The XTR2 can now only be used with the chosen program A key symbol appears on the display adjacent to the program number All keys except the intensity and the ON OFF keys are locked into this mode Use the same key combination to unlock the stimulator Switching OFF the timer You can use the XTR2 with or without timer All programs run for 30 minutes Press and hold for 3 seconds the key and the left key The total the
4. SPORECUP XTR2 Muscle stimulator for sports and fitness Item no 104064 Instruction Manual Item no 101478 schwa medico MEDICAL EQUIPMENT 3 schwa medico Contents Intended use ccccssescssscccsesesssseersesseees Safety instructions Important safety Contra indication Side effects Operating elements Description of the symbols Technical specification Operating the SPORECUP XTR2 Connecting cables and electrodes Switching ON the XTRZ selecting a program selecting an area of Setting the intensity stimulation Terminating the stimulation Locking the key pad Switching OFF the timer Swltching OFF the stimulator Charging the battery SPORECUP XTR2 Programs ouverte eessen 10 HERE GE 10 Setting of intensity Jovels 12 General information 13 Adjustments modifications and repairs 13 EE 13 Guarantee 13 Maintenance and deanmg 13 Classification o n 13 Technical check 13 Delivery content 14 ACCESSOFIES 1essseneenennennnnanneto 15 Electrode placement 17 Electrode placement for pain treatMent 17 Electrode pla
5. cement for dynamic stimulation 19 Electrode placement for muscle stimulation 20 Electrode placement for aestethics programs 22 schwa medico SPORECUP XTR2 Intended use The SPORECUP XTR2 has been designed as a device for the generation of transcutaneous electrical stimulation of the nerves and muscles on humans and may not be used for any other purposes Please read the operating instructions carefully before using the SPORECUP XTR2 Safety instructions Important safety The XTR2 is only to be used for stimulation purposes on one patient at a time Only use the XTR2 with original accessories The size of the electrodes should to be not less than 2 cm The electrodes should be positioned in such ways that the current flow will not cross through the heart Keep the XTR2 away from water or other liquids Do not drop the XTR2 do not use It inappropriately or expose It to extreme temperatures or high levels of humidity not less than 10 C or more than 40 C or a relative humidity of more than 90 90 Do not use the XTRZ If it is not working properly or If it has been damaged in any way Be careful when using the XTR2 near or on children Keep the unit away from children Always store the XTR2 in its case to protect it from damage and dust To avoid reciprocal interference the XTRZ should not be operated in the vicinity of other electronic devices I
6. e code number of the conformation evaluation process is given in accordance with the CE labelling of the involved notified body Application part type BF This product provides protection against electric shock on an acceptable leakage current This symbol indicates the serial number of the device This symbol indicates the article of the device This equipment is marked with the recycling symbol It means that at the end of the working life of the equipment you must dispose of it separately at an appropriate collection point and not place it in the normal domestic unsorted waste disposal In the interests of environmental protection please observe those laws and regulations of your country which relate to accumulator disposal schwa medico SPORECUP XTR2 Technical specification Two channel nerve and muscle stimulator with electrically insulated channels constant current characteristic output short circuit control element AKS 30 integrated pre set programs Output current 100 mA with an actual resistor of 1kQ Frequency range 0 5 120 Hz Impulse width 50 400 us Nominal current 15 mA Power supply integrated rechargeable battery Ni mH 4 8 V Dimensions 11 4x 5 9 x 2 7 cm Weight approx 300 g Pulse forms with 1 kQ load real with ANSI AAMI standard load change in output currentdepending on load current voltage time voltage time current time current mA 8
7. e stimulation This treatment method also has an enhanced effect on lymphatic drainage stimulation The placement of electrodes for dynamic stimulation depends on the direction of stimulation rising falling stimulation or sideways stimulation Rising or falling stimulation Sideways stimulation for back or legs for gluteus or stomach muscle gt ae ar Channel 1 de o ee __ Channel 2 Y ebe r K Channel 1 Channel a 20 schwa medico SPORECUP XTR2 Rising wave Falling wave channel 2 channel 1 channel 1 channel 2 Electrode placement for muscle stimulation Hand flebo Hand extensor Biceps 21 xu schwa medico SPORECUP XTR2 Back Quadriceps Leg flebo hamstrings 22 w schwa medico SPOR CUP XTR2 Electrode placement for aestethics programs P27 P30 Thighs Buttocks Abs schwa medico RECOVERY P1 Recovery after endurance training P2 Recovery after competition P3 Recovery after force training P4 Recovery with dynamic stimulation P5 Cramp treatment P6 Next day regeneration P7 Venous return dynamic stimulation P8 Heavy legs P9 Treatment of hematoma RELAXATION P10 Slow dynamic massage PAIN TREATMENT P11 Cervical pain P12 Low back pain dynamic P13 Lumbosciatica dynamic P14 Tendinitis P15 Joint pain P16 Chronic pain endorphinic 23 SPORECUP XTR2 SPORT P17 Resistance P18 Endurance P19 Strength P20 Explosive strength P21 Hypertr
8. el and thanks to the vasodilatation effect on the capillary vessel This way the athlete avoids the next days heavy legs The active recovery with Spor cup XTR2 allows a faster and better recuperation of the muscles with no supplement charge for the body compared to traditional recovery sessions Use recovery programs within 3 hours after the efforts Increase blood flow Emphasize release endorphin Eliminate muscle contraction Accelerate relax muscles P1 Recovery after endurance training P2 Recovery after competition P3 Recovery after force training P4 Recovery with dynamic stimulation P5 Cramp treatment P6 Next day regeneration P7 Venous return dynamic stimulation P8 Heavy legs P9 Treatment of hematoma RELAXATION P10 Slow dynamic massage PAIN TREATMENT P11 Cervical pain P12 Low back pain 11 schwa medico SPORECUP XTR2 Pils P14 ule P16 Lumbosciatica Tendinitis Joint pain Chronic pain endorphinic SPORT P17 P18 P19 P20 P21 Resistance Endurance Strength Explosive strength Hypertrophy body building REHABILITATION P22 Amyotrophy prevention CAPILLARIZATION E P24 Capillarization 1st phase Capillarization 2nd phase FITNESS P25 P26 Muscle starter Toning muscle building AESTHETICS P27 P28 Pas P30 NO shaping Cellulolipolyse Vascularization adipose tissue Buttocks amp abs training 12 schwa medico SPORECUP XTR2 Setting of inten
9. es the electrode gel or the electric current impulses The physician should be consulted in cases of reddening burning itching or blistering under the electrodes or in the vicinity of the electrodes Slight skin reddening of short duration in the area of the electrodes following stimulation Is quite normal because blood circulation has been improved by the effects of the stimulation e Muscular pain If the stimulation has been too intense or too long muscular aches in terms of sore muscle can occur Shorten the treatment time and the intensity of stimulation at the beginning of muscle stimulation to avoid muscle ache Operating elements The XTR2 has been designed for the generation of nerve and muscle stimulation on human beings All parameters can be adjusted via the press keys and are Indicated on the large display Display Menu key Program choice Key to set different functions Modification keys to adjust parameters and to modify the intensity ON OFF key of el GE Output sockets for the cables 7 Qutput socket for the battery charger 8 ON OFF commutator schwa medico SPORECUP XTR2 SPORECUP Description of the symbols Attention Read accompanying documents especially user manual C 0197 By labelling with CE certificate the manufacturer states that the product meets e Ae P all requirements of the respective EU Directive A conformation evaluation process has been successfully completed Th
10. f this is not possible the various unit functions should be closely monitored during operation so that proper use in accordance with the regulations is ensured The simultaneous connection to the patient of a high frequency surgical unit can lead to burning under the electrodes or around the probe Operating close for example 1m to short wave and microwave units or mobile telephone systems can cause fluctuations in the output values of the electric current stimulation unit Do not use the machine during the simultaneous operation of other machines when driving or when asleep Contra indication Please check with your physician before using the XTR2 if one of these applies to you e Patients with electronic implants such as heart pacemakers defibrillators or pumps e Patients with cardiac rhythm disorders e During pregnancy schwa medico SPORECUP XTR2 e Patients subject to seizure disorder epilepsy e Patients with skin disorders in the vicinity of the electrodes e Patients with malignant diseases in the region of application Side effects e Pain caused by the stimulation The stimulation can be perceived as uncomfortable painful if the electric current is too intensive or if the electrodes have been positioned unfavorably Unpleasant side effects can be avoided by adjusting the intensity using another program with other parameters and the possible re positioning of the electrodes e Skin intolerance which can result from the electrod
11. ft fluff free cloth Please ensure that no moisture permeates the stimulator If moisture does permeate the stimulator a technical check must be carried out before re use Classification In accordance with the Law on Medical Devices the XTR2 is classified as being a Class lla medical device Technical check We recommend a technical check on the XTR2 every 24 months This includes 1 Confirming that user instructions have been included in the accompanying documentation 2 Checking the equipment for completeness 3 Visual check for mechanical damage for damage to all cables and sockets schwa medico SPORECUP XTR2 4 Functional safety Checking the output signals with a load resistance of 1 KQ real current and voltage Checking the frequency Checking the pulse width These technical checks may only be performed by suitably qualified persons The results must be documented with the date and name of the person carrying out the check Delivery content Transportation bag 101478 Instruction manual Item no Article Quantity 104064 SPORECUP XTR2 1 106351 Cable Type 5 15 2 283400 Self adhesive electrodes 50x50 mm 4 pieces 1 101062 Charger WK 112 for XTR2 1 1 1 schwa medico Accessories SPORECUP XTR2 The SPORECUP XTR2 may only be used together with all accessories referred below The manufacturer cannot guarantee the security of the stimulator if it is not used with original accessor
12. grated electrodes The pelotte with its massage nubs intensifies the pain relieving effect MedicoBack P type may be used for light or acute pain as well as for chronic pain in the lower back area Item no Article Size Length 107034 MedicoBack P type S 80 90 cm 107036 MedicoBack P type M 90 100 cm 107037 MedicoBack P type L 100 110 cm 107038 MedicoBack P type XL 110 120 cm eN 107039 MedicoBack P type XXL 120 130 cm MedicoBack P Type is including AA 1 low back support bandage e 1 pelotte 1 cover for the pelotte with integrated electrodes 1 cover pelotte without integrated electrodes 8 gel pads schwa medico Electrode placement Electrode placement for pain treatment P11 P16 Back pain Epicondylitis ulnaris SPORECUP XTR2 17 18 schwa medico SPORECUP XTR2 Arthrosis hip sciatic pain Gonarthrosis knee pain Ankle joint pain Polyarthritis 19 schwa medico SPORECUP XTR2 Carpal tunnel syndrome Pain of achilles tendon heel pain Electrode placement for dynamic stimulation P4 P7 P8 P10 P12 P13 The stimulation is not administered simultaneously on both channels The current flows from one electrode to the other just like a wave undulating on the body This kind of stimulation is far more comfortable than traditional neuro muscular stimulation Dynamic stimulation can be used with high frequency parameters for an improved pain control or with low frequency parameters for optimized muscl
13. ies Stimex self adhesive electrodes Item no 281000 282000 283400 281032 283600 283000 283100 281027 Description Quantity Stimex 32 mm round 4 stimex 50 mm round 4 stimex 50x50 mm 4 stimex 50x50 mm 20 stimex 50x90 mm 2 Stimex 50x130 mm 2 Stimex 80x130 mm 2 Stimex sensitive 50x50 mm 4 Silicone electrodes to be used with contact gel Item no 107090 107075 107011 107010 107050 107070 108000 SIZE Quantity 20 mm round 2 40x28 mm 2 38x45 mm 2 48x48 mm 2 70x65 mm 2 70x140 mm 2 contact gel tube 60 g 1 Garment electrodes gloves and socks Stimex electrode gloves and socks used in combination with a schwa medico electrotherapy unit are ideal for home treatment They provide stimulation of the entire hand and or foot and ankle and avoid the sometimes tiresome attachment of self adhesive electrodes to the hand or the foot A 15 16 schwa medico SPORECUP XTR2 K Gloves A 7 Item no Article Size Quantity N gt UU 107014 Stimulation gloves S 1 pair V x D 107021 Stimulation gloves M 1 pair EN J 107022 Stimulation gloves L 1 pair G Socks y Item no Article Size Quantity N 107023 Stimulation socks M 1 pair A L 107024 Stimulation socks L 1 pair S he MedicoBack P Type Low back support bandage with pelotte and integrated electrodes to be used with the XTR2 MedicoBack P type offers patients lumbar vertebrae stabilisation and support as well as easy nerve stimulation thanks to the inte
14. ophy body building REHABILITATION P22 Amyotrophy prevention CAPILLARIZATION P23 Capillarization 1st phase P24 Capillarization 2nd phase FITNESS P25 Muscle starter P26 Toning muscle building AESTHETICS P27 Shaping P28 Cellulolipolyse P29 Vascularization adipose tissue P30 Buttocks amp abs training schwa medico MEDICAL EQUIPMENT schwa medico Export Department Wetzlarer Strasse 41 43 35630 Ehringshausen Germany Tel 49 6443 8333 113 Fax 49 6443 8333 119 export schwa medico de www schwa medico com Manufacturer Pierenkemper GmbH Hoernsheimer Eck 19 35578 Wetzlar Germany REF 451600 0345 C 0197 Stand 2011 03 16
15. rapy time appears on the display Press the key to switch off the timer The symbol appears on the display Press the O key to save the changes Switching OFF the stimulator Press the O key to turn off the XTR2 If the battery level is too low or if the stimulator has not been used for the previous 2 minutes it turns off automatically schwa medico SPORECUP XTR2 Charging the battery The battery level is shown on the display with a battery symbol with 4 lines If the battery level is too low the stimulator switches off automatically and cannot be turned on again The battery needs to be re charged e Switch the commutator to the OFF position no 8 e Connect the charger with the XTR2 socket no 7 e Connect the charger to the mains circuit The diode on the charger illuminates in red e Charge the stimulator until the diode on the charger illuminates in green e The battery is fully charged if the light on the charger appears in green Important Do not forget to switch the commutator ON again after charging a g Commutator Charger 10 schwa medico SPORECUP XTR2 Programs overview RECOVERY sessions of recovery stimulation are highly recommended after strong muscular efforts Using Spor cup XTR2 and its dynamic stimulation after a competition or an intensive training session offers a better recovery of your muscles thanks to the faster reduction of the lactid acid blood lev
16. sity levels Do not try to achieve a higher and higher level of intensity Set the intensity in such a way that you get a comfortable feeling during the stimulation Increase the intensity carefully to a maximum tolerable and then reduce It slightly step by step to a comfortable level Note that the intensity settings change depending on the stimulation area and the time This is normal and can be explained as follows e Resistance of the skin Dry skin has a lower conductivity level than damp skin due to sweating The skin does not show the same resistance to linear current on all parts of the body For example the resistance level of horny skin can be twice that of the skin in the hollow of the knee The innervation differs greatly according to the area of the body This results ina difference of sensitivity e Volume of the muscle mass If you stimulate muscles with a high volume you will automatically need to set a higher intensity than for the stimulation of small muscles e Muscle fatigue A tired muscle will only tolerate a lower intensity e Age of electrodes self adhesive electrodes do not last forever You can use them up for up to 60 stimulation sessions This depends on the quality of the electrodes but also on the skin of the user Old electrodes have a higher resistance and their conductivity levels decline It is Important to replace electrodes regularly e Adaption to the current The nerves adapt rapidly to the current It is