Lymph 12 Pro User Manual
Contents
1. Place fit the garment s onto the limb s requiring treatment Turn the unit on A visual mains indicator will illuminate 3 4 5 Select the required pressure using the pressure control 6 Select the desired treatment protocol by the mode selector 7 Note the time the treatment starts so that the unit can be switched off when the desired time has elapsed 8 DO NOT unzip the garment when the unit is switched on This may cause damage to the garment Always turn off the power before unzipping the garment 9 Extra comfort can be obtained by first placing a Pulse Press Garment Liner on the limb under the pressure garment TECHNICAL SPECS Compressor 20 Litres per Minutes Pressure 20 100 mmHg Power 220 240v AC 50Hz Dimensions 200mm x 160mm x 160mm Weight 3 1Kg Treatments Peristattic Sequential Pre drainage Remote 30min 60min Continuous Start Stop Barton Business Park 44 0 1582 344 970 Barton le Clay 44 0 1582 882 176 Bedfordshire sales mjsgrp com MK45 4RP England MJS Healthcare Ltd Reg No 2217463 Vat No 491191245 www mjsgrp com
2. PUESE PHESS DTIVAMIC COMPRESSION THERAPY COTU TENU TS ENGLISH SAFETY INSTRUCTIONS ELECTROMAGNETIC INTERFERENCE DESCRIPTION OF CONTROLS CARE AND MAINTENANCE INDICATIONS AND CONTRA INDICATIONS TECHNICAL SPECIFICATIONS FRANCAIS INSTRUCTIONS DE SECURITE INTERFERENCE ELECTROMAGNETIQUE DESCRIPTION DES COMMANDES ENTRETIEN ET MAINTENANCE INDICATIONS ET CONTRE INDICATIONS SPECIFICATIONS TECHNIQUES DEUTSCH SICHERHEITSHINWEISE ELEKTROMAGNETISCHE STORUNG BESCHREIBUNG DER STEUERELEMENTE PFLEGE UND WARTUNG ANZEIGEN UND GEGENANZEIGEN TECHNISCHE DATEN ITALIANO ISTRUZIONI DI SICUREZZA INTERFERENZA ELETTROMAGNETICA DESCRIZIONE DEI COMANDI CURA E MANUTENZIONE INDICAZIONI E CONTROINDICAZIONI SPECIFICHE TECNICHE PORTUGUESE INSTRUCOES DE SEGURANCA INTERFERENCIA ELECTROMAGNETICA DESCRICAO DOS CONTROLOS CUIDADOS E MANUTENC O INDICAC ES E CONTRA INDICAC ES ESPECIFICACOES T CNICAS ESPANOL INSTRUCCIONES DE SEGURIDAD INTERFERENCIA ELECTROMAGNETICA DESCRIPCION DE LOS MANDOS CUIDADO Y MANTENIMIENTO INDICACIONES Y CONTRAINDICACIONES ESPECIFICACIONES TECNICAS PULSEPRESS DYVAMIC COMPRESSION THERAPY SAFETY IUSTRUC TIOMS Read all of these instructions and save them for later reference Follow all warnings and instructions marked on the devices These units must only be operated by qualified medical staff Operation of these units by any other person is entirely the responsibility of that person This
3. ent suitable for use 0 50 humidity Range 30 60 100mmHg 5mmHg LCD Display Range 0 150mmHg 5mmHg 25W Class Il Double Insulated Device OPE ITIUG IUSTI UC TIOIUS WARNING These units must only be operated by qualified medical staff Operation of these units by any other person is entirely the responsibility of that person WARNING This manual must be read and fully understood before these units are applied to patients WARNING Refer to section on indications and contra indications prior to the unit being applied to patients Application to Patient These units are designed to provide compression to the Pulse Press range of Solo and Multi pressure garments Each unit can be connected to one or two garments simultaneously If only one garment is being used then the other outlet is sealed For the Multi pumps the blanking plug must be removed to use a second garment The hoses connection is unable to be the wrong way round The other end of the air feed hose simply pushes onto the garment For the Solo pumps the hose latches into the outlet with a push and twist action The other end of the air feed hose pushes into the garment inlet elbow Operational Use 1 Connect the mains lead to the IEC power inlet socket Insert the mains plug of the lead into a standard 220 240 V mains socket 2 Fitthe air feed hose s to the air outlets as described above Ensure that all connections to the pump and garment make a good seal
4. ers greatest level of limb drainage Peristaltic Mode This Mode offers a gentle wave of inflation working in a traditional Distal to Proximal manner This method in lymph drainage is preferred by some patients as it offers greater comfort than sequential inflation lt should be noted that the overall effectiveness of reducing the limb size Is lower Pre Drainage Mode This mode has been designed to copy the methods used Manual Lymph Drainage MLD to open the lymph nodes Once the nodes have been opened it allows easier drainage of excessive lymph Combination Therapy The lymph 12 Pro has two modes which automatically combine the Pre Drainage mode and the Peristaltic or Sequential The benefit is that dur ing a single treatment the limb is initially prepared by opening the lymph nodes and then automatically switches into the chosen lymph drainage mode PUMP Before cleaning the pump unit always ensure it has been disconnected from the mains power supply Clean the unit with a light detergent ensuring that only a damp and not wet cloth is used If in doubt as to the action of the detergent try cleaning the underside of the unit first where damage will be unseen There are no service items which need replacing or cleaning out on a regular basis All technical information can be found at www mjsgrp com GARMENT The garments can be sponged with a detergent Allow to dry fully before use DO NOT FORCE DRY In the e
5. ing measures Re orientate or relocate the receiving antenna Increase the separation between the equipment and the receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the manufacturer of the device PULSEPRESS DYVAMIC COMPRESSION THERAPY PUMP CON TROLS 1 10 PUMP CONTROLS 1 External fuse holder Replace with 500mA Slo Blo Power Inlet Use with 220 240v AC 50 Hz Air outlet for use with single garment Air bung Remove for inflating two garments at a time or trousers Power Switch Pressure Gauge Shows treatment pressure setting Start button Push to start treatment Mode Selector Used to select different treatment protocols O o AD 9 Y 1 Inflation cycle indicator Shows which chambers are inflated 10 Pressure selection control 30 to 100 mmHg adjustment IMODE SELEC y m MODE SELECTION EXPLANATION Mode 1 Fast Sequential x1 Arm or x1 Legging Mode 2 Slow Sequential x2 arms x2 Leggings or x1 Lymph Legging Mode 3 Pre Drainage 15 mins then Slow Sequential Mode 4 Fast Peristaltic x1 Arm or x1 Legging Mode 5 Slow Peristaltic x2 arms x2 Leggings or x1 Lymph Legging Mode 6 Pre Drainage 15 mins then Slow Peristaltic Mode 7 Pre Drainage only Gradient Sequential Mode The garment is inflated fully in a distal to proximal manner Once fully inflated the entire garment is deflated before the cycle repeats This method off
6. manual must be read and fully understood before these units are applied to patients Equipment not suitable for use in the presence of Flammable Anesthic mixture with air or with oxygen or nitrous oxide Unplug the pumps before cleaning Only ever use a damp cloth Do not spill liquid onto the pump enclosure Always place pumps on a flat stable surface before operation The pumps will not operate properly if tilted or at an angle Do not cover the pumps whilst in operation Use only the type of power source indicated on the pump s rear label Do not use a damaged or frayed power cord Unplug the pumps and refer servicing to qualified personnel under the following conditions o Ifthe power cord or plug is damaged o Ifliquid has entered the pumps o Ifthe plastic outer enclosure is damaged o If the pump is not operating normally continuous running noisy running or low performance Always switch off the pump before unzipping the garments ELECTROMAGUETIC IU TERFEREUCE This equipment has been constructed and tested to comply with EN60601 1 2 EMC regulations for Medical devices This test was designed to provide reasonable protection against harmful interference This equipment may generate and radiate radio frequency energy and if not installed correctly may cause interference to radio communications If this equipment does cause interference then the user is encouraged to try to correct the interference by one or more of the follow
7. vent of heavy soiling the garments can be machine washed See wash label Always insert the air entry blanking plug to prevent water ingress and turn the garment inside out prior to machine washing NUDICA TIOfUS AMD COM TRA IUDICA TIOMS INDICATIONS VENOUS Post phlebetic syndrome varicose veins varicosities of pregnancy stasis dermatitis leg ulcers WOUND RECOVERY Venous ligation or stripping hand surgery total knee replacement D V T For prevention of deep vein thrombosis ARTERIAL Raynaud s disease arteriosclerosis acro cyanosis blue leg EDEMA Gravitational lymphatic post mastectomy inguinal lymphadenectomy after cast removal rheumatoid multiple sclerosis FLEXION Chronic injury burns and skin graft CONTRACTURES CONTRA INDICATIONS Acute thrombophlebitis acute pulmonary edema congestive heart failure known or suspected deep vein thrombosis PULSEPRESS DYUAMIC COMPRESSION THERAPY Power Source Frequency Internal Fuse Caution A Water Resistance Mode Storage Operation Pressure Regulation Pressure Meter Power Consumption im PULSEPRESS DYUAMIC COMPRESSION THERAPY 220 250V AC 50Hz T500mA Type BF Equipment not suitable for use in the presence of Flammable Anesthic mixture with air or oxygen or nitrous oxide Read instructions prior to using equipment Ordinary Equipment Intermittent Operation Equipment to be stored used between 15 and 30 degrees C Equipm