Turbi Quick - Frank`s Hospital Workshop

image

Contents

1. 6 2 CONTROL MODE Few seconds after the warming up process of the lamp the Turbi Quick will automatically enter in the Control Mode and on the display a Control sample will be required the instrument now is ready to analyze one or more Control PT Control n 1 insert cuvette and press button When this indication Stand by mode appears on the display the Operator has to press rapidly the button On the display will appear PT Control n 1 when T 0 add PT Rea T 5 A 5 seconds countdown will start when the T zero seconds on the display will appear the following message PT Control n 1 add PT Rea Now the operator has 3 seconds to inject the starter reagent During these three seconds on the display will appear PT Control n 1 mixing And immediately after PT Control n 1 Time XX Once the analysis is finished the clot is formed the timer stops and on the display appears Control n 1 PT 26 3 INR 2 43 PT 222 Press the button Pag 12 24 Vital Diagnostics TURBI QUICK USER S MANUAL If the a printer is connected the same information will be also printed out Once the analysis is finished the clot is formed the timer stops and on the display appears Control n 1 PT 26 3 INR 2 43 PT 222 Press the button If the a printer is connected the same information will be also printed out Now the operator can a Press the button rapidly less than 2 seconds
2. Turbi Quick USER S MANUAL Manual code MAN 037 Revision 5 Revision date 05 26 2008 TURBI QUICK USER S MANUAL Vital Diagnostics Pag 2 24 Vital Diagnostics TURBI QUICK USER S MANUAL INSTRUMENT NAME Turbi Quick INSTRUMENT MANUFACTURER Vital Diagnostics S r l Via Balzella 41 g 4 47100 Forl ITALY INSTRUMENT DISTRIBUTOR Please read this manual completely before using the instrument Pag 3 24 TURBI QUICK USER S MANUAL Vital Diagnostics 1 INTRODUCTION E NEE E E T E E a cubase tasse E sonia 5 Eli MAAND D US E mura es er ti tm reti tnt ich rode iu aim hr tute 5 1 2 IMPROPER US E iii EN Ia E E E E einen ten 5 2 INSTRUMENT DESCRIPTION cccsssssssscccccccscsssssccccccccccssssccecccccecsssscccecceecessscccceccceeecssscsscccceseesscees 5 2L SCUVECIESHOEDER SEA a M o ec dea e den aio a dee bn e oo i Miao pei 5 2 2 SMART CARD na ANNE A A A USNJEN VR EE 6 2 3 RESET BUTTON eee 7 24 FILTERSSEEECTOR 5 2 did 7 3 INE NS aka zaka AE eve okove N 8 Sed TURBEOUICK POSTITONINGE lalla AA aje A A teak Re Sees ates A ae 8 3 2 POWER O NEDENARNE ESS K SN NE UN NE eti 8 3 3 WARM UP AND CALIBRATION kk 9 3 4 LAMP WARMING UP PROCESS dinas enee AR Up ob ad 9 4 OPERATING RECOMMENDATION G cccccccssssssssscccccccssssssccccccccccssssscccccccecesssccccecceecccssscccccceeeeessscees 10 5 SAFE IA O 10 5 1 USER PRECAUTIONS ceri eee 10 5 2 ELECTRICAL
3. In this case the analyzer will enter in the stand by mode and is ready for a new control b Press the button more than 2 seconds If the operator wants to start to analyze patients sample he can press more than 2 second the button the instrument will PT enter to Sample mode and on the display will Control n 2 appear insert cuvette PT and press button Sample n 1 insert cuvette and press button Now follow instructions described at point 6 3 6 3 SAMPLE MODE PRESS THE BUTTON MORE THAN 2 SECONDS In case the operator does not want to perform any control he can start immediately to analyze patient s samples To pass directly to Sample mode press the button for more than 2 seconds then on the display will appear PT Sample n 1 insert cuvette and press button When this indication Stand by mode appears on the display the Operator must insert the cuvette into the reading channel and then press rapidly less than 2 seconds the button In this way a 5 seconds countdown will start when the T zero seconds on the display will appear the following message PT Sample n 1 add PT Rea Now the operator has 3 seconds to inject the starter reagent During these three seconds on the display will appear PT Sample n 1 mixing And immediately after PT Sample n 1 Time XX Pag 13 24 TURBI QUICK USER S MANUAL Vital Diagnostics Once the analysis is finished the clot is formed the tim
4. 6 Inject the Starter reagent straight on the light bundle Do not insert the whole pipette tip into the cuvette but just the end of the tip 7 When insert the cuvette into the Reading channel push it down strongly until the cuvette reaches the bottom of the channel This operation is very important because an incorrect cuvette positioning will cause an erroneous sample s reading 8 Each sample and or control must be prepared following the instruction sheet inside each Turbi Kit Once the sample is ready transfer the cuvette into channels 1 4 for the pre incubation Maintain the cuvette into the pre incubation channel for the time shown on the instruction sheet Once the pre incubation time is finished transfer the cuvette into the reading channel and follow all the indication previously described to perform the test 5 SAFETY MEASURES 5 1 USER PRECAUTIONS Before using the analyzer the operator must know the rules for handling potentially infectious materials and electro mechanical systems 5 2 ELECTRICAL EQUIPMENT As with all electrical equipment the power supply is a potential source of danger To prevent the risk of electrical shock to the user and or damage the instrument the operator should not open any part electrical of the instrument Only authorized personnel such as service technicians may open the instrument to perform maintenance or repair 5 3 SAMPLES All biological fluids must be considered by the oper
5. Diagnostics S r l and supplied together with the instrument 1 2 IMPROPER USE Following uses are considered improper 1 The use of the device by not qualified personnel The use of cuvettes different from those specified on the manual and on the inserts The cuvette re use The use of reagents different from those specified on the the manual and on the inserts Every attempt to analyze samples different from human serum or plasma Every attempt to obtain different parameters form those specified on the inserts The exposure of the device and of the optical part to liquids which can soak into the reading system 8 The insertion of devices different form cuvettes into the reading channel 9 Disregard instructions given on this manual The above mentioned uses and every attempt to use the Turbi Quick with a purpose different from the intended use must be considered improper tO Or A 2 INSTRUMENT DESCRIPTION 2 1 CUVETTES HOLDER In the upper part of the Turbi Quick reader there is the cuvettes holder that is automatically thermostatized at 37 C It has with 5 channels Channel from 1 to 4 are pre incubation channels and can be used to reach the sample s temperature of 37 C The channel marked with V is instead the reading channel where cuvettes must be transferred after the pre incubation phase The reading channel is a photometer with two different filters A and B available Pag 5 24 TURBI QUICK USER S MANUAL Vital Di
6. In case the injection time has been expired and the reagent has not been injected on the display will appear PT Control n 1 PT Rea not detected repeat sample In this case press the button rapidly less than 2 sec on the display will appear PT Control n 1 insert cuvette and press button Now the instrument is ready again to perform the test press again the button and insert the starter reagent when the display requires the injection In case the reagent has been injected due to the small reagent volume and or the small reagent absorbance the analyzer can t recognize the injection the operator has to press the button within 2 seconds and on the display will appear consequently CRP Latex Control n 1 mixing CRP Latex Control n 1 reading CRP Latex Control n 1 Time 119 Once the reading time has expired on the display will appear CRP Latex Control n 1 conc 12 3 mg L press the button If the a printer is connected the same information will be also printed out Now the operator can run another sample as previously described Pag 17 24 TURBI QUICK USER S MANUAL Vital Diagnostics 8 2 SMART CARD CHANGE It is possible to change the smart card also without switching off the instrument This special feature is available only during the analyzer stand by mode The stand by mode can be displayed both during samples analyses and during controls analyses A during
7. control mode The stand by mode is displayed as follows PT Control n 1 insert cuvette and press button Under this condition to change the smart card it is necessary firstly to pass from control to sample mode pressing the button for more than 2 seconds the analyzer will enter in the sample mode PT Sample n 1 insert cuvette and press button Now it is possible to remove the smart card and insert a new one and pass from e g PT analysis to APTT analysis To do that follow the instructions described under point B Pag 18 24 B during sample mode The stand by mode is displayed as follows PT Sample n 1 insert cuvette and press button Under this condition it is possible to remove the smart card and insert a new one and pass from e g PT analysis to APTT analysis To do that remove the old smart card and insert the new one Then press more than 2 seconds the button so the instrument will start again with the calibration procedure Turbi Quick Reader V X X APTT Filter A Lot 9999 16 05 05 test counter 9999 Warmup to 37 0 C Temperature 34 6 Filter calibration please wait APTT Control n 1 insert cuvette and press button Now the instrument is in Control Mode ready to analyze the first APTT control Vital Diagnostics TURBI QUICK USER S MANUAL 8 3 HIGH LOW SAMPLE S CONCENTRATION For those analytes expressed in conc
8. etici 23 13 DISPOSAL AND RECYCLING iesssdssccssecessessscccectdecsseessssecosssedascedsncecesssocssocsestesesssecssncteotedeessondcustestesesseses 24 Pag 4 24 Vital Diagnostics TURBI QUICK USER S MANUAL da INTRODUCTION Dear User we hope you will be satisfied with your Turbi Quick analyzer The Turbi Quick analyzer is a reader especially developed for coagulation and immunochemistry analyses The coagulation tests are performed using the optical detection of the clot once the clot is formed there is an absorbance shift which allows the detection of the end of the reaction For what concerns instead Immunochemistry a complete panel of quantitative immunoturbidimetric analyses is available for the Turbi Quick reader all the tests have been studied by Vital Diagnostics and optimized for the Turbi Quick reader All the tests are based on the Turbidimetry principle the attenuation of the intensity of the light through a solution containing Antigen Antibody aggregates is proportional to the Antigen concentration This manual describes the Turbi Quick way of working Please read it completely to understand the capabilities of your instrument 1 1 INTENDED USE The Turbi Quick analyzer is a semi automated photometer developed for the analysis of coagulation and immunochemistry parameters performed on human serum or plasma The complete list of tests available on the instrument is supplied and updated only by Vital
9. 20 24 Vital Diagnostics TURBI QUICK USER S MANUAL Picture 4 Picture 5 Unscrew the screw blocks the lamp as shown in picture 5 and then remove the lamp Insert a new lamp block it whit the screw connect the cable and mount the back panel of the photometer 10 PERFORMANCE CRITERIA AND LIMITATIONS 10 1 PERFORMANCE CRITERIA A Optical detection precision 0 001 O D Photometric Measurement B Analysis Reproducibility C V lt 6 test depending C Cuvette Holder block thermostatized 37 C Automatic room temperature between 15 32 C D Photometric zeroing Automatic E measuring points 2 F Measuring range 0 1 50 D 10 2 LIMITATIONS A Strongly lipemic or haemolytic samples may alter instrument s reading capability B Samples with a concentration gt or lt than linear range Pag 21 24 TURBI QUICK USER S MANUAL Vital Diagnostics 11 TECHNICAL SPECIFICATIONS Area of application Instrument size Weight Voltage Reading Channel Pre incubation Channels Thermostatization Analysis time Loading pattern Results Measuring Principle Filters Reading resolution Display interface Interface Software CE Marked Pag 22 24 Immunochemistry and coagulation analyses Height 145 mm Width 210 mm Depth 120 mm 1100 gr external power supply 240 V 110 V A C 50 60 Hz output 12 V dc 3 5 A 1 4 37 C kit s dependent One test a time Shown on the d
10. 24
11. EOUIPMENT e iii 10 SN A NA 10 5 4 NOTES ON SAFETY MEASURES ii I e rea 11 5 5 RESIDUALRISKS eee 11 6 COAGULATION OPERATING PROCEDURE ccccccccssssssssscccccccccsssssccccccccecssssssceccceeecssccccccceseeees 12 Gal FOREWORDS tt eet I ce ieee wean Tdi 12 6 2 CONTROL MODE rire 12 6 3 SAMPLE MODE PRESS THE BUTTON MORE THAN 2 SECONDS 13 6 4 READING ERROR MESSAGE kakec 14 7 IMMUNOTURBIDIMETRIC OPERATING PROCEDURE sccssscccccccssssssssscccccceccsssscccccceeeecssscees 14 TEL KONRAD da 14 7 2 SAMPLE MODE PRESS THE BUTTON MORE THAN 2 SECONDS aaa 15 8 TROUBLESHOOTING oreesa ieat ROTAIA RATA TIRA ici 17 8 1 THE INSTRUMENT DOESN T RECOGNIZE THE STARTER INJECTION cscccccceseccccessceccueccecuueceseueecsseueceeeaners 17 622 SMART CARD CHANGE Sci 18 8 3 HIGH LOW SAMPLE S CONCENTRATION erre 19 0 47 READING ERRORS tit is ti ld a E eta ba Baia eS 19 9 MAINTENANCE AA O 20 9 1 GLEANING INSTRUGTIONS sni ne nee aa tai 20 9 2 LAMPREPLACEMENT erre rie eee 20 10 PERFORMANCE CRITERIA AND LIMITATIONS cccccsssssssssccccccccscsssscsccccccecssssccscccceeeccssscsceeees 21 10 1 PERFORMANCE CRITERIA 2 50 en a ente 21 10 2 MEAN NS a dit i de di o 21 11 TECHNICAL SPECIFICATIONS iisscccscccssssssescuscsccsscessssccosssscsscossetesessscsssonscesssssescsssacsseesveeesoaccosseseesessenes 22 12 EGCDEGEARATION cnc RR
12. agnostic use only A t i 4 ELECTROSTATIC DISCHARGE SENSITIVE DEVICE ESDS The device could be damaged by electrostatic potentials 5 5 RESIDUAL RISKS Despite of the measures taken in the designing of the machine to guarantee a safe use of it there might happen reasonably predictable occurrences whose risk was possible to reduce but not to eliminate completely RESIDUAL RISKS PROTECTION MEASURES Biological contamination The operator must wear always gloves and protection glasses as prescribed by laboratory regulations Do not ever open tubes Cuvettes breaking and or overthrow In case of cuvettes accidental broken and or overthrow clean the instrument with a sodium hypochlorite solution Cuvettes re use Cuvettes are disposable so the re use of cuvettes can false the analysis result Pag 11 24 TURBI QUICK USER S MANUAL Vital Diagnostics 6 COAGULATION OPERATING PROCEDURE When a Coagulation smart card is inserted in the Turbi Quick follow these instructions 6 1 FOREWORDS The cuvettes required to perform coagulation test have a volume of 1 5 mL and a particular shape studied to reduce the volume of reagent needed code ACC 16 047 Only this kind of cuvette is suitable to perform coagulation tests All the cuvettes different from ACC16 047 must be discarded Coagulation tests are very sensible to the temperature Strictly respect pre incubation time to obtain reproducible results
13. agnostics Pre incubation channels from 1 to 4 Reading Channel 2 2 SMART CARD The Turbi Quick reader works only in association with Turbi kits Turbi kits are the reagents needed to perform analyses on the instrument and inside each Turbi kit box there is a Smart Card The smart card must be inserted in the Smart Card Slot located in the rear panel of the instrument Without the smart card the instrument doesn t work m ss Smart Card In the Smart Card s chip in fact the following data are stored Calibration Curve in each card the calibration curve of the test is stored in this way the Operator doesn t need to perform the calibration curve but he can start immediately with patient s samples Test Settings reading time mixer behavior and kind of filter are already stored in the card in order to avoid incorrect readings due to wrong filter mixing or reading time setting errors Expiry Date the Expiry Date of the kit is stored on the card in order to avoid the use of expired kits Number of test the number of tests performable with the kit is stored in the card this counter will decrease each time the user performs one test Pag 6 24 Vital Diagnostics TURBI QUICK USER S MANUAL Smart Card Slot Power ON OFF Smart Card Slot and Power ON OFF switch 2 3 RESET BUTTON On the upper part of the cuvettes holder between channel 2 and 3 is located the Reset Button whic
14. ample T 5 Pag 14 24 Vital Diagnostics TURBI QUICK USER S MANUAL A 5 seconds countdown will start when the T zero seconds on the display will appear the following message CRP Latex Control n 1 add sample Now the operator has 3 seconds to inject the starter reagent Once the sample has been injected on the display will appear CRP Latex Control n 1 mixing And immediately after CRP Latex Control n 1 reading CRP Latex Control n 1 Time 119 Once the reading time has expired on the display will appear CRP Latex Control n 1 conc 12 3 mg L press the button If the a printer is connected the same information will be also printed out Now the operator can a press the button more than 2 b press the button rapidly less than 2 seconds seconds If the operator wants to start to analyze patient s In this case the analyzer will enter in the sample and not controls he can press more stand by mode and is ready for a new control than 2 second the button the instrument will enter to Sample mode and on the display will CRP Latex appear Control n 2 insert cuvette CRP Latex and press button Sample n 1 insert cuvette Now the operator can run another control as previously described Now follow instructions described at point 7 2 7 2 SAMPLE MODE PRESS THE BUTTON MORE THAN 2 SECONDS In case the operator does not want to perform any control he can start immediately to analyze patie
15. ator as potentially infectious For this reason the operator must adopt the national and international standards of precautions to avoid the biological danger Specimens patient samples and controls and liquid waste should be considered potentially infectious and capable of transmitting human immunodeficiency virus HIV hepatitis B virus HBV and other Pag 10 24 Vital Diagnostics TURBI QUICK USER S MANUAL blood borne pathogens The handling of these substances must be performed in accordance with established laboratory safety regulations in order to minimize risk for laboratory staff This includes wearing of gloves splash protection etc Contact with skin and mucous membranes must be avoided This also applies to all components of the instrument that are exposed to these substances If any specimen is spilled on the instrument wipe it up immediately and clean the contaminated surface with a disinfectant of 0 5 sodium hypochlorite solution Compliance with local regulations pertaining to the disposal of waste is the responsibility of the operator Refer to local sources for additional information on correct biohazardous waste disposal Qualified technical operators must apply the same warning procedures for instrument maintenance 5 4 NOTES ON SAFETY MEASURES On the instrument to assure a correct use of the instrument may be placed the following symbols Attention read use instruction A IVD For in vitro di
16. ee the Operating Procedures of this manual Pag 9 24 TURBI QUICK USER S MANUAL Vital Diagnostics 4 OPERATING RECOMMENDATIONS These recommendations are essential to work correctly with Turbi Quick reader and with Turbi Kits Please before performing analyses read carefully these recommendations 1 Once the cuvette has been inserted into the Reading Channel don t touch the cuvette until the reading time is finished the cuvette s movement could cause an erroneous sample s reading 2 Once the cuvette has been inserted into the Reading Channel don t remove the cuvette until the reading time is finished the cuvette s removing causes an erroneous reading process 3 To obtain reproducible results with Turbi Quick reader it is necessary to use calibrated pipettes an erroneous volume of reagents will cause an incorrect ratio between reagents and consequently results won t be reproducible Vital Diagnostics recommends to calibrate pipettes at least every six months 4 The Smart Card can be removed from the Turbi Quick only when the reader is switched off or is in stand by mode Removing the Smart Card while the instrument is working can damage Turbi Quick and cause an erroneous sample s reading 5 During the injection of the Starter reagent avoid bubble formation Bubbles can create interference with the light source and cause an erroneous reading In case of bubbles we suggest to prepare a new sample
17. entration e g Fibrinogen if a sample has a concentration lower than the minimum value of the linear range on the display will appear Fibrinogen Sample n 1 conc lt 10 0 mg L press the button While if a sample has concentration higher than the maximum value of the linear range on the display will appear Fibrinogen Sample n 1 conc gt 300 mg L press the button In this case the operator can repeat the test after sample s dilution follow insert instructions 8 4 READING ERRORS In case of reading errors the following messages can appear Reading error Check r channel Check filter pos land press button In this case the operator can a check if the filter selected is the right one if not move the filter lever on the right position and press the button The instrument will check the reading system again b check if the reading channel is empty In case the reading channel is not empty remove the cuvette and press the button If these 2 operations won t solve the problem call the technical assistance White error t Check r channel land press button In this case the operator can a check if the filter selected is the right one if not move the filter lever on the right position and press the button The instrument will check the reading system again b change the lamp as described on this manual If these 2 operations won t solve the prob
18. er stops and on the display appears Sample n 1 PT 26 3 INR 2 43 PT 222 Press the button If the a printer is connected the same information will be also printed out Now the operator should press the button rapidly less than 2 seconds in this way the analyzer will enter in the stand by mode and is ready for a new sample PT Sample n 2 insert cuvette and press button 6 4 READING ERROR MESSAGE In some particular tests like PT or APTT can happen that the clot doesn t form so the instrument can t fix the end of the reaction When this happens the following message will appear on the display PT Sample n 1 Reading error Press the button In this case remove the cuvette from the reading channel and press the button rapidly less than 2 sec in this way the instrument will return in the stand by mode PT Sample n 2 insert cuvette and press button 7 IMMUNOTURBIDIMETRIC OPERATING PROCEDURE When an Immunoturbidimetric smart card is inserted in the Turbi Quick follow these instructions 7 1 CONTROL MODE Few seconds after the warming up process of the lamp the Turbi Quick6 will automatically enter in the Control Mode and on the display a Control sample will be required the instrument now is ready to analyze one or more Control CRP Latex Control n 1 insert cuvette and press button The operator now has to press rapidly the button on the display will appear CRP Latex Control n 1 when T 0 add s
19. h can be used to reset the instrument The complete functioning of this button is explained in this manual Please read it carefully before use the Turbi Quick Reset Button 2 4 FILTERS SELECTOR On the front side of the cuvettes holder just in front of the reading channel is located the Filters Selector This selector allows the user selecting between 2 different filters filter A and filter B Each test requires a specific filter the filter needed appears on the Turbi Quick s display as soon as the instrument with a Smart Card inserted into the slot is switched on To change the filter move the lever to the left for filter A or to the right for filter B The filter should be selected before switching on the instrument The filter required is also written in the internal insert of each kit Pag 7 24 TURBI QUICK USER S MANUAL Vital Diagnostics 3 INSTALLATION 3 1 TURBI QUICK POSITIONING The Turbi Quick shouldn be placed near centrifuges oscillating agitators or other vibrating instruments that might cause movement of the bench It should be placed far from windows or direct light sources to avoid Turbi Quick overheating 3 2 POWER ON Connect power supply outlet to the instrument Insert the power supply plug in a socket with an earth connection If the instrument has an optional printer it should be connected to the Turbi Quick with the appropriate cable and plugged in Insert the Smart Card in the
20. he new Smart Card Now you can switch on again the Turbi Quick 3 3 WARM UP AND CALIBRATION Once the instrument is switched on with a correct Smart Card it will start the warm up procedure during this time the Turbi Quick will heat the cuvette holder until the temperature of 37 C is reached On the display will appear Warmup to 37 0 C Temperature 34 6 Once 37 C are reached the Turbi Quick will perform the auto calibration and the filter checking On the display will appear Filter calibration please wait If some calibration problems occur on the display will appear Reading error Check r channel Check filter pos land press button This message can be displayed when The filter selected is wrong in this case move the filter s selector to the other filter and press the Reset Button The Turbi Quick will repeat again the calibration with the right filter The reading channel is not empty in this case remove the cuvette from the reading channel and then press the Reset Button The Turbi Quick6 will repeat the calibration 3 4 LAMP WARMING UP PROCESS Once the calibration is finished the Turbi Quick reader starts the Lamp warming up process and on the display will appear Lamp warmup please wait During this phase the lamp reaches the correct operating temperature Once this process has been completed the instrument is ready to work To know how to proceed s
21. isplay and printed out if a printer is connected Expressed in concentration Photometric Filter A near UV Filter B UV 0 001 O D LCD 20 x 4 with back light RS 232 for printer Integrated yes Vital Diagnostics TURBI QUICK USER S MANUAL 12 EC DECLARATION DICHIARAZIONE DI CONFORMIT CE EC DECLARATION OF CONFORMITY conforme all Allegato III della Direttiva 98 79 CE Dispositivi Medico Diagnostici In Vitro conforme all Allegato Il della Direttiva 2006 42 CE Direttiva Macchine according to Annex III of the Directive 98 79 CE In Vitro Diagnostic Medical Devices according to Annex II of the Directive 2006 42 CE labbricante Vital Diagnostics S r l manufacturer Via Balzella 41 G 4 indirizzo address 47100 FORLI ITALIA telefono fax phone 0039 0543 721220 fax 0039 0543 796001 Lettore per analisi immunochimiche effettuate con metodo Identificazione dei prodotti turbidimetrico Product identification Reader for immunochemistry analyses performed by turbidimetric method Nome commerciale TURBI QUICK Brand name Numero i di catalogo PRD IMP XXX Part number s classificazione dei prodotti dispositivi diversi da quelli elencati nell Allegato Il della Direttiva 98 79 CE product identification devices other then those mentioned in Annex Il of the Directive 98 79 EC Si dichiara sotto la propria responsabilit che i dispositivi sopraelencati rispettano le disposizioni applicabili delle seguenti dirett
22. ive Hereby we declare under our sole responsibility that the above mentioned devices meet the applicable provisions of the following Directives Direttiva 98 79 CE Direttiva 2006 42 CE Direttiva 2004 108 CE Compatibilit Elettromagnetica Direttiva 2006 95 CE Bassa Tensione Direttiva 2002 96 CE e 2003 108 CE RAEE Direttiva 2002 95 CE RoHs Tutta la documentazione tecnica comprovante il rispetto dei requisiti applicabili delle Direttive elencate conservata a cura del Fabbricante All the technical documents required to demonstrate the conformity to the listed Directives are kept by the Manufacturer anno di immissione in commercio year of introduction on the market luogo e data place and date Forl 10 03 2008 2005 firma Mar yA yan signature timbro della Societa Pag 23 24 TURBI QUICK USER S MANUAL Vital Diagnostics 13 DISPOSAL AND RECYCLING Herewith we declare that this instrument is subject to the European Directive 2002 96 EC RAEE Directive Therefore the instrument must be disposed separately not as urban waste and delivered to the specific collection center in according to the Directive 2002 96 EC The user can ask to the dealer the collection of the instrument if a new instrument is ordered to replace the old one On the instrument there is a label with the symbol shown in this page The symbol means that the instrument cannot be disposed as urban waste Pag 24
23. lem call the technical assistance Black error t Check r channel land press button In this case the operator can a check if there is a direct source of light on the instrument Remove the direct source of light from the instrument If this operation won t solve the problem call the technical assistance Pag 19 24 TURBI QUICK USER S MANUAL Vital Diagnostics 9 MAINTENANCE Its simplicity and component parts mean that the Turbi Quick does not require special maintenance The most sensitive part is the lamp which can be easily replaced in case of damage see further Care and attention are required above all regarding the cleanliness of the upper part reading part which must be kept dry and clean The soaking of the upper part can damage the reading system Do not clean the upper plate with liquids or damp cloths 9 1 CLEANING INSTRUCTIONS Dust can be removed using an ordinary vacuum cleaner In case of cuvettes accidental broken and or overthrow clean the instrument with a sodium hypochlorite solution 9 2 LAMP REPLACEMENT In each Turbi Quick package is inserted a lamp as spare part In case of blown lamp follow these instructions to replace the lamp Picture 1 Picture 2 Unscrew the 3 screw as shown in pictures 1 and 2 then remove the back panel of the photometer as shown in picture 3 Now disconnect the cable of the lamp white cable as shown in picture 4 Picture 3 Pag
24. nt s samples To pass directly to Sample mode press the button for more than 2 seconds while the instrument is in stand by mode then on the display will appear CRP Latex Sample n 1 insert cuvette and press button Pag 15 24 TURBI QUICK USER S MANUAL Vital Diagnostics When this indication Stand by mode appears on the display the Operator must insert the cuvette into the reading channel and then press rapidly less than 2 seconds the button In this way a 5 seconds countdown will start when the T zero seconds on the display will appear the following message CRP Latex Sample n 1 add sample Now the operator has 3 seconds to inject the starter reagent During these three seconds on the display will appear CRP Latex Sample n 1 mixing And immediately after CRP Latex Sample n 1 reading CRP Latex Sample n 1 Time 119 Once the reading time has expired on the display will appear CRP Latex Sample n 1 conc 12 3 mg L press the button If the a printer is connected the same information will be also printed out Now the operator should press the button rapidly less than 2 seconds in this way the analyzer will enter in the stand by mode and is ready for a new sample CRP Latex Sample n 2 insert cuvette and press button Pag 16 24 Vital Diagnostics TURBI QUICK USER S MANUAL 8 TROUBLESHOOTING 8 1 THE INSTRUMENT DOESN T RECOGNIZE THE STARTER INJECTION
25. slot Select the Filter Switch on the instrument Once the Turbi Quick is switched on on the display you can see Turbi Quick Reader V X X If the printer is connected to the instrument the same information will be also printed out If the Smart Card is not recognized by the instrument or is not inserted in the slot on the display will appear Invalid card turn off and replace the card In this case the instrument must be switched off the Smart Card changed or inserted into the slot and then the instrument can be switched on again Now on the display will appear this message XX Filter A Lot 9999 16 05 05 test counter 9999 and if the printer is connected the following information will be also printed out Analysis XX Filter A Lot nr 9999 Exp date 16 05 05 T counter 9999 Where First Line Name of the test Second Line Filter reguired Third Line Kit s lot number Fourth Line Kit s expiry date Fifth Line Test counter This counter will decrease each time the user performs one test Pag 8 24 Vital Diagnostics TURBI QUICK USER S MANUAL If the test counter is zero on the display will appear test counter 0 turn off land replace card This message will appear when the maximum number of test per kit have been achieved and kits reagents are finished Once will appear this message switch off the instrument open a new kit and insert into the slot t