Home
User`s Manual(Alternative)-EN
Contents
1. The Rightest Meter The Rightest Meter Up Button A Down Button 9 Press to review stored test value in memory and average Select the function mode Display Window Show your test result messages and relative information 10 Smart Code Key For calibrating your meter New one comes with every box of test strips Refer to page 17 Contain the test strip port for inserting strip Assemble Smart Code Key into code key base Test Strip Port Insert test strip here until you hear a click Main Button Press to turn on off meter Switch different function screen Press and hold down until enter to the setting mode There will be a power off period during this process 11 The Rightest Meter The Rightest Meter Please follow the step and step Smart Code Key Base Assemble Smart Code Key here 12 BIONING Rightest cmoo IVD A Ml Bionime Corporation cE 0537 MI T 1140052 Smart Code Key Release Button Press down to release and switch the Smart Code Key Battery Cover Slide cover to change batteries Model Number Meter Serial Number The Rightest Test Strip The Rightest Test Strip Noble Metal Electrodes
2. Recalling Your Test Result Recalling Your Test Result The Rightest meter is able to store 300 test results with time and date automatically If your test results are up to 300 sets which is the max memory of the meter your newest test results will replace the old ones starting from the oldest one To recall your test memory start with the meter without test strip inserted 1 Press the main button to switch the screen to memory function it will show the latest testing result with time and date The latest result is with sequence number 1 2 Under the memory screen use either the A or 9 button on the side to review all previous results with date and time You will see results from the most recent Sequence no 1 to the oldest Sequence no 300 at the down right corner of the screen 44 3 If a test result is not yours or you think it is not suitable to be calculated by average you can mark it as non average data by pressing and firmly holding the A or Q button for over 2 seconds until the AVG and NO appears on the screen then press the button to confirm it The quality control solution test result can be recalled from the memorized data When you see data with the CTL it is testing data with control solution The result will not be used for average calculation If the ambient temperature is out of 10 C 40 C 60 F 104 F the test result will show w
3. Consult with your healthcare professionals whenever you want to conduct a sampler from palm or forearm As the blood flow taken from forearm is slower than fingertip or palm we recommend using a special lancing device with Clear Cap for testing sites other than finger If you use normal lancing device for palm or forearm the blood sample might be not enough for meter 3 You can start measurement by the blood sample 4 The minimum blood sample size of Rightest Glucose Monitoring System is 1 4L 1 0ul 1 4ul 2 0ul 3 0ul 4 0ul e e e J We suggest you to take 1 4 2 5ul to do the test on glucose monitoring system Blood sample size above 4 0ul is too much which will contaminate the Smart Code Key PRECAUTION Don t take too much blood sample such as 4 0uL please see how much it is in the example picture or more to do the test Too much blood might contaminate the Smart Code Key Please clean it up right away if your Smart Code Key gets blood stain or moisture Refer to page 56 for cleaning Smart Code Key Sample Size Example 37 Performing a Test in Order Performing a Test in Order Applying blood sample to test strip View Window Appearance y JTY the whole area of the View Window to get an accurate test result v v Insufficient blood sample will cause O O O O inaccurate test result Repeat the Make sure your blood sample covers a i test
4. Preface Preface Thank you for selecting the Rightest Blood Glucose Monitoring System Please read this manual thoroughly before you start testing It provides all the information you need to use the product You will get a correct test result by the following instruction of this manual It is quite important for you to monitor blood glucose regularly A derived complication can thus be effectively reduced Through the assistance of Rightest Blood Glucose Monitoring system which provides considerate design and accurate results your diabetes management can become more reliable and ease to use The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use For self testing amp professional use only outside the body The testing result is calibrated to be plasma equivalent for test with fresh capillary whole blood samples by finger stick laboratory instrument You may consult your healthcare professional for instructions on how to use the system correctly Our Customer Support staff is available to assist you as well The Rightest Blood Glucose Monitoring System was manufactured and supported by Bionime Corporation and its authorized representative If you have any questions or concerns please contact your Bionime authorized representative or call us by phone number 886 4 24951268 GMT 08 00 08 30 17 30 toll free at 1 888 4952568 Monday through Friday 08 00 16 30 EST in USA
5. Customer Service Parts of Critical Component We sincerely like to provide complete considerate services to our customers Please review all the instructions to make sure you are performing the steps correctly You are always welcome to contact us by call 886 4 24951268 GMT 08 00 08 30 17 30 toll free at 1 888 4952568 Monday through Friday 08 00 16 30 EST in USA amp Canada only or by e mail at info bionime com If you have any question or advice 64 Blood Glucose Meter Test Strip and Control Solution Manufacturer Bionime Corp Product complied with In Vitro Diagnostic Medical Device Directive 98 79 EC CE0537 EU Rep BIONIME GmbH Heinrich Wild Strasse 202 CH 9435 Heerbrugg Switzerland Lancing Device Manufacturer Suzhou SPR Medical Devices Development Co Ltd Product complied with Medical Device Directive 93 42 EEC EU Rep EMDAR BY lJsselburcht 3 Postbus 5486 6802 El Arnhem Netherlands Disposable Sterile Lancets Manufacturer Suzhou SPR Medical Devices Development Co Ltd Product complied with Medical Device Directive 93 42 EEC CE0197 EU Rep EMDAR BY lJsselburcht 3 Postbus 5486 6802 El Arnhem Netherlands 65
6. 2 With the Meter off follow and direction to put the new Smart Code Key into the track on code key base 3 Push down the Smart Code Key until it snaps into the Smart Code Key base A PRECAUTION Assemble the Smart Code Key into your meter is necessary before starting your test Be sure to change the Smart Code Key with each new box of Rightest Test Strips Use only the Smart Code Key that is packaged with the box of strips you are using Please don t remove this Smart Code Key before you run out of these new strips Batteries Installation Batteries Installation Your meter kit comes with two AAA 1 5volt batteries uninstalled New set of batteries will provide power to perform about 1000 tests under normal use The low battery symbol 83 will keep appearing on your display if your batteries run low Please keep spares on hand and replace the batteries soon when a battery symbol appears The meter will automatically turn off after 3 minutes of nonuse to save battery power Your test memories are still saved 1 Turn the meter over 2 Install both batteries at once 3 Slide the battery cover back Press and push battery Be sure to put battery in until it snaps into place cover to open correct direction 20 4 Meter will do electronic runing test and the whole screen will mCode Error keep blinking as soon as the batteries are installed H af a J ny 5 Press any but
7. amp Canada only we will make every effort to provide assistance and solutions for you Please forward your warranty card to Bionime authorized representative to activate your warranty coverage Precaution Precaution Before using the Rightest System to test your blood glucose please read all the instructions and practice the tests including quality control test Refer to page 46 Please do the quality control test regularly to make sure the test results are accurate The Rightest Blood Glucose Meter can only be used with the Rightest Blood Glucose Test Strips The use of any other brand Strips should not be used under any circumstances The use of other brands of strips may give inaccurate results The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use only The testing result is calibrated to be plasma equivalent for test with fresh capillary whole blood samples by finger stick The Rightest Blood Glucose Monitoring System is intended for self testing It should not be used to diagnose diabetes mellitus Rightest Blood Glucose Monitoring System has not been validated for use on neonates Therefore it s not intended for use of neonates 4 Do the test at least 30 minutes after while entering another site with different ambient temperature Pay attention to the environment protection when disposal of batteries Using appliance in tropical and
8. whichever comes first Example i Exp Date 2005 06 Replace the vial cap of control solution and close tightly right away after using of control solution sample Check the expiration date before you use the control solution Refer to Control Solution Package Insert 49 Performing a Quality Control Test Performing a Quality Control Test Starting with Check Key 1 Insert the Rightest Check Key with Check Key word facing up into test strip port The meter quickly goes through a series of internal checks 2 The check key allows you to make sure your meter is functioning well You can see zac symbol and 0K symbol blinking on the screen The check key also allows your meter proceeding quality control test with control solution NOTE The check key is only for functional testing of meter If you want to make sure strip and meter work properly please use control solution for the test see page 50 for detail 3 If both Ozman symbol and Error symbol blinking appear remove check key and insert it again If error screen appears again your meter is not working correctly Please contact Bionime authorized distributor or Customer Service Center 50 OK CHECK KEY o 7 Error RA CHECK KEY Mono Performing test with control solution CTL 1 Ple
9. Electrochemical sensor Indication Symbol Sample Entry Apply a drop of blood or control solution here The test request is only 1 4uL of blood l O 8 lt SS GO p Insert strip with indication symbol up and toward meter View Window This window is originally yellow before applying blood sample It will gradually become red after filled with blood drop 14 Hand Bar Where you can easily hand the strip to do the test Electrode Contacts Sensing signal output terminals The blood sample will be drawn into sample entry by capillarity automatically PRECAUTION Re cap the vial cap immediately and close tightly after taking test strip out from the vial Do not reuse test strips Do not use expired test strips Check the discard date you wrote on the vial label every time you use the strip Store the test strip in a cool dry place and avoid the exposure of direct sunlight and heat For detailed information please refer to the Rightest Test Strip Package Insert Dummy Code Key The Rightest Smart Code Key For package without test strip not for package with test strip A Dummy Code Key will be installed on the new Rightest meter in the package without test strip This Dummy Code Key is not intended to be used for testing When you open a new box of test strips you have to replace the Dummy Code Key by new S
10. of control solution which alteration or degeneration might happen If you approach the control solution vial to the sample entry on strip for 3 5 times you will find that the measurements will lower down for 10 20 Keep your Smart Code Key dry and clean Please clean it up right away if your Smart Code Key gets stain or moisture Refer to page 56 for cleaning Your control solution test results should fall within the control solution range That means your Rightest System is working correctly Control Soultion Range 90 130 mg dL 5 0 7 2 mmol L f control solution test results are working properly Repeat the qual Example of control solution range printed on your test strip vial label out of control solution range your Rightest System may not be ity control test If your control solution results outside the range still exist do not use the Rightest System to test your blood glucose And contact Bionime authorized distributor or Bionime Customer Service The reason your control solution results are out of the range Your control solution has expired or after 3 months since opened Your test strip has expired You leave the cap off the vial of test strips or control solution for a long time You didn t perform the test procedure correctly Malfunction of the meter 55 Caring for Your Meter Error Messages and Trouble Shooting Maintenance About Temperature Error Kee
11. on the top of the cap _ Error 52 8 When you hear a sound of beep leave the meter on the table while waiting the test result The screen will show the countdown mode 9 Tightly recap the cap of control solution 10 The control result appears after the measurement is completed Compare your control test result to the range printed on strip vial label Example 8 Your result should fall within the control solution range printed on the label of strip vial If result is not within the control range please refer to Understanding Control Test Results on page 55 Please follow the local regulation and discard the used strip properly Control Solution Range 90 130 mg dL 5 0 7 2 mmol L 53 Performing a Quality Control Test Understanding Control Test Results A PRECAUTION Your quality control test result by control solution test will not be calculated for average reading but still can be recalled The control solution test result will be shown with CTL 54 symbol on the screen You have to do check key test before contro solution test If you only use the control solution to do the quality control test your test result will be memorized and calculated for average same as real blood test Please don t touch the control solution by approaching the vial to the sample entry on strip see drawing The reagent might be sucked into the vial
12. with a new test strip Insufficient blood sample Enough blood sample 1 Touch the blood drop 2 Hold your finger on the sample entry until you hear a beep and to the sample entry of the view window is totally filled with blood If the view window is not A PRECAUTION the strip The blood totally filled with blood and or the test does not start discard the Check the expiration date printed on the package every time you use the strip Do not use will be drawn into the test strip and repeat the test with a new test strip When you repeat expired test strips strip automatically the test you may want to obtain the blood sample from a different aes siio imnmecleichy eiis Ang i Out iren ne viell Do not reuse test strips MOST AN SE AEREO Or PAGUES MOUGN DIGA SAMIS Assemble the Smart Code Key into your meter is necessary before starting your test PRECAUTION Be sure to change the Smart Code Key with each new box of Rightest Test Strips When you do a test please pay attention not to squeeze too much blood to pollute Smart Do the test at least 30 minutes after while entering another site with different ambient Code Key temperature 38 39 Performing a Test in Order Performing a Test in Order To ensure accurate test results make sure the code number on your meter matches the code number highlighted on the vial of test strips you are using Only place the blood drop to the sample entry of the strip Keep yo
13. About Meter Error Er2 Meter has malfunctioned Do the quality control test or reinstall the batteries to see if the meter works properly If error screen still appears contact Bionime Customer Service About Signal Error Er3 When you do the test the meter find the signal is unreasonable Please do a new test again _ Error Er Ere frj 59 Limitations Specification Store the strips in the original capped vial at temperatures between 4 C to 30 C 39 F to 86 F and relative humidity below 90 Do not freeze Rightest Blood Glucose Test strips are designed for using with capillary whole blood and venous whole blood samples Do not use serum or plasma samples Inaccurate test results may be obtained at high altitude more than about 10000 feet 3048 meters above sea level Hematocrit below 30 may cause higher results and hematocrit above 55 may cause lower results Severe dehydration and excessive water loss may cause inaccurately low results Rightest Blood Glucose Monitoring System has not been validated for use on neonates The glucose test may be interfered under abnormal concentration of Uric acid gt 0 54 mmol L gt 9 0 mg dL L Dopa gt 0 076 mmol L gt 1 5 mg dL Methyldopa gt 0 071 mmol L gt 1 5 mg dL Cholesterol gt 6 5 mmol L gt 250 mg dL 60 Measurement Technology Oxidase Electrochemical Sensor Sample Capi
14. Order Performing a Test in Order 5 While the symbol flashing you are ready to apply the blood sample The Obtaining blood sample Alternative site testing code number will appear on the right down corner of the screen f a on The Rightest blood glucose monitoring system gives you an opportunity to obtain blood samples 7 4 from fingertip palm or forearm A PRECAUTION Before appearing please don t touch your blood drop to the sample Fingertip Blood Sampling entry on strip because meter is doing a internal test If you do so the 1 Hold the lancing device to the soft side of your fingertip and press the meter will show and Error and you will waste a strip release button to get a tiny blood sample A PRECAUTION A PRECAUTION To ensure accurate test results make sure the code number on your meter matches the Before you take the blood sample please make sure your fingertip for code highlighted on the vial of test strips you are using sampling is dry thoroughly Water or alcohol might affect the accuracy If the code number does not match change the Smart Code Key to the right one and redo a of testing test If the meter turns off before you apply the blood sample take out the unused test strip and 2 Obtain a small blood drop on your finger reinsert it You have three minutes to apply the blood sample before the meter turns off In order to get a more accurate measurement suggest to discar
15. ase run check key test first After see I symbol and OK appear on the screen Remove check key from meter The screen shows the o 3 symbol blinking and ett appears Then you need to put a new strip into the meter 2 Remove test strip from vial and Re cap the vial cap immediately 3 Insert the test strip with view window facing up into test strip port The code number will appear 4 The test strip symbol stops blinking You will see blinking symbol and cTL symbol on the screen Then you need to approach your control solution to 4 en the meter i aa D83 51 Performing a Quality Control Test Performing a Quality Control Test PRECAUTION Before you do control solution test you have to do check key test first Before appears please don t touch the control solution to the sample entry on strip because meter is stilling in a internal test If you do so the meter will show Error and 5 Shake the bottle of control solution well before opening the cap Then open the cap and put it on the table 6 Drip a drop of control solution on the top of the cap NOTE J Don t touch the control solution by approaching the vial to the sample entry on strip Don t let control solution touch anywhere of your body In case please use water to clean down 7 Gently touch sample entry of the strip with the control solution
16. control solution to check if the strip and meter work together properly A PRECAUTION The complete quality check should include check key test and control solution test Before you do control solution test you have to do check key test first 47 About Quality Control Test About Quality Control Test When should perform a quality control test Before doing a blood glucose test with your meter for the first time When you open and start using a new vial of test strips When you replace a new Smart Code Key please check your meter by CHECK KEY When your meter is dropped or splashed with liquids Whenever you think your test result does not consistent with the way you feel Whenever you want to check if your system is working properly or not Whenever you want to practice testing and check correct procedure Required items for quality control tests To perform a quality control test you have to prepare the items below Rightest Meter with Smart Code Key installed Rightest Blood Glucose Test Strips Rightest Control Solution Rightest Check Key S Ed g 48 A PRECAUTION When you open a new bottle of Rightest Control Solution please write the discard date on the label Rightest Control Solution is good for 3 months after opening the bottle or until the expiration date printed on the label
17. d the first automatically drop of blood The first drop of blood from your fingertip might include Please follow the local regulation and discard the used strips properly some body liquid which might affect the testing accuracy 34 Performing a Test in Order Performing a Test in Order Alternative site testing palm or forearm blood sampling To do the alternative site testing please install the clear adjustable tip for your lancing device For detailed information of installing clear adjustable tip of lancing device please check the instruction manual for the lancing device To increase the blood flow please massage the puncture area of palm or forearm for a few seconds Immediately after massaging the puncture area press and hold the lancing device with the clear adjustable tip against palm or forearm Then press the release button Continue holding the lancing device against palm or forearm and gradually increase pressure for a few seconds until the blood sample size is sufficient A PRECAUTION The glucose test results of blood samples taken from different site might be varied in certain conditions such as your glucose is changing rapidly following a drink a meal insulin dose or exercise DO NOT test on the palm or forearm if you are testing for hypoglycemia Low blood glucose Fingertip samples can show the rapid change of glucose faster than palm or forearm samples 36
18. er 25 Turning On Off the Meter Function Screens and Button Operations 1 Auto Power Off The Meter will power off automatically if you don t operate it for over 3 minutes 2 Manual Power Off If you want to turn off the Glucose Meter please keep pressing the button for 2 seconds 3 How to turn on the Power 1 Press the button 2 Insert one piece of strip 26 Switching the Function Screens 1 Press the button to switch function screens including Memory Average and Time in sequence Memory Average 2 Under Memory function press the A or button to search your memorized testing data 3 Under Average function press the A or 9 button to search your average data of 1 day 7 days 14 days or 30 days in sequence l m AVG j m AVG l m AVG l m AVG IU IL IUI ILI e g w g ne g6 iw ie 4 When the display on meter screen shows time memory or average keep pressing button for 5 seconds to enter setting mode 27 Easily Handling the Rightest Test Strip Easily Handling the Rightest Test Strip How to easily handle the Rightest test strip Inserting the test strip 1 Hold the test strip between thumb and middle finger with indication symbol W facing up 2 Put forefinger on the side of strip as shown NOTE This pose is the most force saving way for inserting the test strip 3 Insert the test strip into test strip po
19. ith gt symbol blinking The result will not be used for average calculation K AVG Pala nus 1 Recalling Your Average Test Result About Quality Control Test The Rightest meter provides you several average test results You can view the 1 day 7 day 14 day and 30 day average of test results for better blood monitoring of your blood glucose SA 185 185 185 re T g en 26 30 ie 1 Press the main button to switch the screen to average function 2 Under the average of screen use either the A or Q for the option of 1 day 7 day 14 day or 30 day average of test results 3 The number showed on the down right corner means how many test results are calculated 4 The to Hi results the control solution results the non average test results and the test result made out of normal temperature range lt 10 C gt 40 C are not calculated in the average A PRECAUTION You have to set the time and date to activate the average function 46 What is quality control test To make sure the monitoring system will work properly It is necessary to do two kinds of quality controls 1 Check Key Test Please check your meter function by check key the detail is shown in Starting with Check Key section on page 50 2 Control Solution Test If you finish the check key test without error that only makes sure the meter is working properly You still have to use
20. llary whole blood Minimum Sample Volume 1 4 microliter Measuring Range 10 600 mg dL 0 6 33 3 mmol L Test Time 8 seconds Memory Capacity 300 blood glucose test results with date and time Power Saving Turn off automatically after 3 minutes no use Press the button for 2 seconds Operating Temperature 10 40 C 50 104 F Operating Relative Humidity lt 90 61 Specification Warranty Hematocrit 30 55 Power Supply Battery Life Two 1 5V AAA batteries About 1000 tests eter Dimension 85 0 mm x 58 0 mm x 22 0 mm eter Weight 85 0 g with batteries onitor LCD display Display Area 39 0 mm x 38 0 mm eter Storage Conditions 10 60 C 14 140 F Test Strip Storage Conditions 62 4 30 C 39 86 F lt 90 relative humidity Bionime Corporation warrants that your Rightest Meter will be free from defects in materials and workmanship for five years from the date of purchase This warranty dose not apply to the performance of a Rightest Meter that has been altered misused tampered with or abused in any way This warranty applies only to the original purchaser of the meter Please complete and return the enclosed warranty card to Bionime authorized representative Different models have different specifications Some of the models are not included with the warranty card 63
21. lts and Messages Blood glucose test results are shown on the meter as mmol L or mg dL depending on which unit of measurement you have chosen Consult your healthcare professional before making any changes to your diabetes medication program If your blood glucose result is unusually high or low or if you question your testing results repeat the test with a new test strip You can also run a QUALITY CONTROL TEST with your Rightest Check Key and Rightest Control Solutions to check your meter and strip Refer to page 47 If the test result still remains unusually high or low contact your healthcare professional immediately If you are experiencing symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions of this manual contact your healthcare professional immediately 42 The Rightest Meter displays results between 0 6 and 33 3 mmol L or 10 and 600 mg dL If your test result is below 0 6 mmol L 10 mg dL 9 will appear on the screen Please repeat your test again by a new strip If you still get L o result you should immediately contact your healthcare professional If your test result is above the high end of the system s detective range 33 3 mmol L or 600 mg dL H will appear on the screen Please repeat your test again by a new strip If you still get Hi result you should immediately contact your healthcare professional 43
22. mart Code Key in the box Do not throw away this Dummy Code Key It can be used to keep the Smart Code Key Base clean while the Smart Code Key is not installed 16 What is Smart Code Key Smart Code Key is designed for calibrating your meter to match the strips of different production lot since each lot of strips owns different parameter Each new box of strips contains its new Smart Code Key The Smart Code Key stores calibrated information that your meter needs to give you accurate blood glucose results You must change your Smart Code Key with each new box of Rightest Test Strips Smart Code Key also contains the test strip port Although you can still review the memory and other settings it s necessary to assemble the Smart Code Key into meter before actually starting your test This procedure makes sure you do calibrate the meter and assures you of getting accurate test results 17 Smart Code Key Installation Smart Code Key Installation How to switch Smart Code Key Every new box of test strips should come with a Smart Code Key inside Before you install the Smart Code Key please check that the code number on the strip vial and the code number on the bottom of Smart Code Key should be the same If you find they are not the same please return them to the store where you bought from 1 Turn the meter over Remove old Smart Code Key by pressing down Smart Code Key release button if one is installed
23. o the test and the Error symbol will blink in this condition Please move the meter to environment with temperature between 10 40 C 50 104 F and redo the test after 30 minutes 56 57 Error Messages and Trouble Shooting Error Messages and Trouble Shooting About Battery Error 1 The xa symbol is blinking when the battery power is low Please change batteries as soon as you can You can still do the test 2 The 4 and Error symbol are blinking when the battery is too low Meter can not do the strip test Please change the batteries immediately About Check Key Error When you see xiene and Error symbols showing on screen please do the check key test again If this error message appears again it would be Meter or Check Key defect Please return your Meter and Check Key to the store where you bought About Smart Code Key Error The Smart Code Key is faulty or not assembled well to the meter Reinsert the Smart Code Key If the error screen appears again contact Bionime Customer Service 58 Error ACEEA CodeError About early applying blood sample or control solution The blood sample or control solution are applied before the meter is ready Repeat the test with a new test strip Add sample after the symbol appears About Strip Error Er1 Test strip inserted has been used or damaged Please use a new test strip from vial
24. or temperate climate Not to be exposed dripping or splashing The minimum blood sample size of Rightest Glucose Monitoring System is 1 4uL 1 0ul 1 4ul 2 0ul 3 0ul 4 0ul Sample Size Example e J We suggest you to take 1 4 2 5ul to do the test on glucose monitoring system Blood sample size above 4 Qul is too much which will contaminate the Smart Code Key PRECAUTION The unit of measure must be fixed to mg dl When your display shows mmol L during setting or test please contact customer service Use of the wrong unit of measure may cause you to incorrect treatment Table of Contents Table of Contents About Rightest System The Rightest Blood Glucose Monitoring System The Rightest Meter The Rightest Test Strip Dummy Code Key The Rightest Smart Code Key Being Ready for Test Smart Code Key Installation Batteries Installation Setting Up Your Meter Turning On Off the Meter Function Screens and Button Operations Easily Handling the Rightest Test Strip Testing Procedure Getting Ready for Testing Performing a Test in Order 6 30 31 Comprehending Test Results and Messages Review Your Test Memory Recalling Your Test Result Recalling Your Average Test Result Quality Control About Quality Control Test Performing a Quality Control Test Understanding Control Test Results Caring for Your Meter Error Messages and Trouble Shooting Limi
25. p your meter and test strip free of dust water or any other liquid Store the meter in the carrying case when not in use If meter is dropped or damaged perform a quality control test with the check 1 In order to get accurate testing please do testing between 10 40 C 50 104 F key and Control SolUYon DEINE donga Plead glucose lash 2 When the ambient temperature is 0 9 C 32 48 F or 41 50 C 106 122 F the e warning symbol will blinking you still can do the test but the result obtained is only for reference because the test result under these range of temperature might not be correct Repeat the test at an area with temperature between operating range 10 40 C or 50 104 F Cleaning Meter Clean the outside of the meter with a damp cloth and mild soap detergent Prevent the test strip port and Smart Code Key base from moisture Cleaning Smart Code Key 3 When you move from the area with temperature outside the operating range of If your Smart Code Key is stained with blood control solution or any liquid please use a dry tissue or test strip to another area with temperature under operating range of test strip alcohol swab to clean it up immediately Do not use any thing wet to clean it Perform a check key please wait for 30 minutes before you do the test and quality control test to ensure the system is working properly 4 When the temperature is below 0 C 32 F or over 50 C 122 F Meter can not d
26. rt until it snaps and firmly stops 28 Removing the test strip 1 Hold the test strip same as insert it 2 Rotate the test strip counterclockwise and pull up at the same time NOTE Rotating motion will be easier to move out the strip 3 Take the test strip out of test strip port totally Please follow the local regulation and discard used strip properly Getting Ready for Testing Performing a Test in Order Before performing a blood glucose test you have to prepare the items below Rightest Meter with Rightest Smart Code Key installed Rightest Blood Glucose Test Strips Please check the expired date on the vial Don t use the expired test strip Lancing device Sterile lancet Alcohol swab optional A PRECAUTION Release button Test Strips Discard 3 months afr frst opening Plunger ca N k Comfort dial depth selection J Assembling the Smart Code Key into your meter is necessary before starting your test 30 Each time of using a new box of test strip you have to assemble the Smart Code Key from the new strip package Cleaning your hands Before starting your test use warm soapy water to wash your hands and dry thoroughly Warm water helps to increase the blood flow to your fingertips You can also u
27. se alcohol swab to clean your finger to get accurate measurement 1 Unscrew the lancing device 2 Load a lancet into the lancing device Performing a Test in Order Performing a Test in Order 3 Twist off the protective cover of lancet and keep it you will need it after testing 4 Screw back the head of lancing device 5 Twist the comfort dial depth selection of lancing device to select the lancet depth Start at 2 or 3 For tougher skin dial to a higher number 6 Pull the plunger of lancing device until the end and release your hand Your Lancing device is ready Then set it aside you will need it in the next step PRECAUTION Load the lancing device Refer to Instruction manual for the lancing device Be careful to load the lancing device to avoid being pierced by the needle 32 Preparing your meter and strip 1 Remove new test strip from 2 Insert the test strip with 3 Push the test strip in until it vial and re cap the vial cap indication symbol facing up immediately into test strip port 4 Check that the code number blinking on your meter matches the code highlighted on the vial of test strips currently in use snaps and stops The meter turns on automatically When you put the strip into the strip port you will hear a sound of beep ir a 053 a 33 Performing a Test in
28. tations Specification Warranty Customer Service Parts of Critical Component 42 44 46 47 50 55 56 57 60 61 63 64 65 The Rightest Blood Glucose Monitoring System The Rightest Blood Glucose Monitoring System Package of Meter Kit Your Rightest Blood Glucose Monitoring System consists of several items Please identify each item of your system and learn what they re called and how they re used These items are included in your Rightest Blood Glucose Monitoring System Getting Started Guide User s Manual Rightest Blood Glucose Test Strip Package Insert Control Solution Package Insert Rightest Meter Rightest Dummy Code Key for the package without strip or Smart Code Key pre installed for the package with strip 7 Rightest Test strips 0 10 25 pcs 8 Rightest Control Solution 9 Rightest Check Key 10 Lancing Device 11 Clear Adjustable Tip 12 Disposable Sterile Lancets 10 pcs 13 Carrying Case not shown k aonrk WwW Nh 14 Log Book not shown 15 Warranty Card not shown 16 Two AAA Batteries not shown 17 Emergency Card not shown 18 Instruction for the lancing device not shown x Different packages have different bundled items Some of packages might not include items Rightest GETTING STARTED GUIDE BIONING
29. ton to close the runing test of the meter and ava enter setting mode CTL CI LL ay 6 Please set time and date when the batteries are replaced x CHECK KEY FM re D PAY gap TELE ee See Setting the Time Date and Unit on page 23 A o Om E a r Paa m I Dm j pat pme Dax AN PRECAUTION Danger of explosion might happen if battery is incorrectly replaced replace only with the same or equivalent type Please follow the local regulation and discard used battery properly 21 Setting Up Your Meter Setting Up Your Meter Entering Setting Mode You can enter Setting Mode by below two ways ils 2 22 Reload batteries After removing the battery please press button for several times until no signal on screen then follow the steps of page 20 21 to load batteries With batteries Press button for 5 seconds till hearing a sound of beep The display screen will show setting data PRECAUTION When you keep pressing button for 2 seconds the display on screen will turn off Please don t care it Keep pressing the button till entering Setting Mode Setting the Time Date and Unit a Year setting With the year format blinking press the A or button on the side to adjust it Then press the button to confirm the year adjustment and move to month setting Month setting With the month blinking press the A or Y button
30. until the current month appears Then press the button to confirm it and move to day setting Day setting With the day blinking press the A or Y button until the current day appears Then press the button to confirm it and move to time format setting Setting Up Your Meter Setting Up Your Meter 4 Time Format 12H 24H Selection fa With the time format blinking press the A or Y icH button on the side to adjust it Then press the button 1 to confirm it and move to hour setting at 26 BoC 1 ou 76 Bog 5 Hour Setting With the hour blinking press the A or 9 button until the current hour appears Then press the button to confirm it and move to minute setting 6 Minute setting With the minute blinking press the A or amp button until the current minute appears Then press the button to confirm it and move to the setting mode of measurement unit 24 ICH s AM 9 6 BoD 2H 6 s AM rr KEIR Hig 7 Ending Setting Then press the button to complete the setting and go back to time screen The settings you have entered are saved 8 When you do not do any settings of meter for over 3 minutes the meter will leave setting mode and power down automatically NOTE Fast Settings When you do the settings you can press the or amp buttons and hold them for two seconds to select or adjust fast
31. ur Smart Code Key dry and clean Please clean it up right away if your Smart Code Key gets blood stain or moisture Refer to page 56 for cleaning Pose the finger and the strip to the way you feel most comfortable Please follow the local regulation and discard the used strip and lancet properly Appearance of result 1 When blood is applied to the strip you see the countdown mode appearing on A your screen This means the meter is measuring your glucose value D63 2 Your blood glucose result appears after the measurement is completed The nG result is displayed in mmol L or mg dL At the same time the result will be E memorized with time and date es h oo 40 The testing result with time and date is automatically stored in your meter s memory You may also record the glucose value in your Log Book Rotate to remove the test strip Please follow the local regulation and discard the used strip properly To remove the lancet unscrew the head of lancing device and put the protective cap of lancet on the table Stick the lancet into the protective cap by holding the lancing device as a pen Then unload the lancet and discard it into a puncture proof container PRECAUTION The meter will power off automatically after 3 minutes no use You can also turn off the meter by pressing the main button for 2 seconds 41 Comprehending Test Results and Messages Comprehending Test Resu