User manual CuroCell® 4 AD
Contents
1. Care of sweden Supporting Life User manual 2015 06 05 CuroCell 4 AD Alternating pressure system Item number user manual 95 001061 WUN CE 7 Sal 345 0 Contents Important safety precautions Introduction Product description Assembly Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information oOo won A WW 11 12 13 14 15 17 18 19 19 IMPORTANT SAFETY PRECAUTIONS Read all instructions prior to use Warning To minimize the risk of fire electric shocks or bodily injury The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts e g pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has undergone fire tes
2. in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or your local dealer if you have questions regarding warranty and service 12 3 Life expectance This product s expected life is estimated to be 5 years 13 Other information 13 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information please contact Care of Sweden or your local dealer 13 2 Returns and complaints Complaints are to be reported to the supplier Return of a product sha
3. 0 1 4 Air filter 2 Control panel 2 Power cord connection 5 Bed hooks 3 Fuses 2 pcs 2 3 Pump functions control panel 2 Xo Gypole Tims 340 j Camie 6 1 COMFORT SETTING Suitable air pressure is set with the comfort dial 1 8 Guidelines based on user weight is given in the table on the pump NOTE In all uses where the head of the bed is raised the pressure must be increased at least one 1 step 2 CYCLE TIME There are four selectable cycle times Press the button to select cycle time 3 PROGRAM THERAPY Change the mode by pressing the button the chosen mode is indicated by the green light The modes order is Autofirm Alternate Static Alternate seat inflation Static seat inflation A Autofirm Autofirm is used to fill the mattress completely with air for firm support The pump always starts in this mode When the correct pressure is reached it changes automatically to the previous setting or to the presetting outlined below Presetting features for Autofirm process 1 Program Therapy Select the required program 2 Comfort setting Set the required air pressure 1 10 3 Cycle time Select from 10 15 20 or 25 minutes Note The preset Cycle Time Comfort Level and Therapy mode will be illuminated to indicate the settings during the Autofirm process Autofirm LED is also illuminated After mattress is inflated Autofirm LED will be turned off automatically B Alternate In this mod
4. local dealer for spare parts 7 2 Changing fuses 3 G Dy 7 3 Changing the air filter a s G Pull the plug out of the electrical socket before replacing fuses Remove the cover from the fuses with a screwdriver Pull out the fuse holder Fit anew fuse of the same type T1AL 250 V and screw the protective cover back in place There are two fuses Unscrew the round safety plate on the back of the pump Lean the pump backwards in order for the filter to fall out from the holder Place the new filter in the filter holder and replace the safety plate If the pump is used in a dirty environment the filter should be checked regularly 13 8 Troubleshooting Problem Solution Pump does not start Check that the power cord is connected to the power socket Check the fuses Alarm is on Check that the quick connectors to the air tubes mattress is deflating on the side of the pump are properly connected Check that the CPR valve is closed Open the mattress and check for leaks from air cells or tubes Check that all tube connectors are properly fitted and connected No air in head cells When the mattress is inflated the 4 static head cells are filled last Patient is bottoming The comfort setting may be set incorrectly Increase the falls through pressure and wait a few minutes Do a Hand Check Check the mattress as described in the section Alarm is on above Re
5. 20 240 V 4 Operation Note Always read the user manual prior to use 4 1 Start up 1 Set the power switch on the side of the pump to 1 On and start the pump by pressing the power button on the front panel 2 Air starts to inflate the mattress It takes at least 20 minutes until the mattress is ready for use Note The pump always starts in the Autofirm mode 3 The light for Low pressure is lit until the mattress has reached the set pressure When the light goes out the system changes mode to previous settings and the mattress are ready to be used 4 2 Program selection There are four different programs Please refer to section 2 3 for more details As a standard setting we recommend an alternating mode and a 10 minute cycle time 4 3 Comfort setting kg il kg 0 30 1 100 110 6 The air pressure in the mattress can be 40 5 120 130 7 adjusted according to the patient s desire for zo 43 140 150 8 comfort Always consult the prescriber orother C6 4 160 170 9 authorised personnel for appropriate setting 80 90 5 180 200 10 Note The pump always starts in the Autofirm mode When the mattress has reached the required pressure the pump is set automatically to the previously used setting or to the settings made in the Autofirm program Please refer to section 2 3 for more details The mattress is now ready for use NOTE e Werecommend the Seat inflation mode for all situations where the head of t
6. Olivia Item no Size CC 906911 80x200 cm CC 906912 80x210 cm CC 906913 85x200 cm CC 906914 85x210 cm CC 906916 90x200 cm CC 906917 90x210 cm CC 906910 105x200 cm CC 906920 120x200 cm Item no Name Size Item no Name Size CC 901310 Cell PU 80x21 cm CC 901330 Heel cell PU 80x21 cm CC 901311 Cell PU 85x21 cm CC 901331 Heel cell PU 85x21 cm CC 901312 Cell PU 90x21 cm CC 901332 Heel cell PU 90x21 cm CC 901313 Cell PU 105x21 cm CC 901333 Heelcell PU 105x21 cm CC 901314 Cell PU 120x21 cm CC 901334 Heelcell PU 120x21 cm Please make a note on your order when ordering ventilating cells CC 821 Fuse set of 2 CC 8202 Filter disposable If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance 17 11 Options and accessories 11 1 Optionals Comfort fabric For patients who are sensitive for the fluctuations in the mattress a padded inner cover can be placed over the air cells for increased comfort Heel section There are two options available to relieve sensitive heels Emptying the air cells with release valves or lower air cells NOTE These options are only available with new orders Cable holder A pocket along one long side of the mattress that holds the power cord in place to minimise the risk of tripping over the cord or the cord becoming trapped under the wheels of the bed NOTE This option is only available with new orders Sheet holder Specialised clips to w
7. The manufacturer retains the right to modify the product specification at any time WARNING Follow operating and maintenance instructions carefully to prevent injury as well as damage to the product or other equipment 9 2 Labelling The mattresses are labelled with product name production date article number size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot 15 9 3 Symbol Definitions Article No Counteract shear force Machine wash at 95 C m Drip dry Class Il equipment Manufacturer GROY LINNA x OLN 7 S Yuna Recommended user weight i Heel function gt lt 1 Chlorine X Do not iron j 3 gt gt gJ nm XXXX XX XX Year Month Day CE Patient information category TX Do not wash Tumble dry Do not dry clean IP__ IP class Vv Foot placement ye Y a Read the user manual Group of function CE marked acc to Recycling MDD 93 42 EEC X XXX Category i Not in domestic refuse follow recycling instructions Directly on bed base Machine wash at 70 C Si Do not tumble dry Wipe off Temperature range Double Insulated Ingress Protection Rating lam The mattress should only be used lying in the lengthwise direction on the mattress 16 10 Spare parts Top part
8. ad more in sections 4 2 and 4 3 Mattress moves around Check that the air mattress is fastened to the bed frame with the straps underneath two at the head end and two at each side of the mattress Some cells have less air This is normal as air flow alternates between the cells for a set cycle period depending on the designated cycle period one cycle varies from 10 to 25 minutes depending on the setting used Pump is noisy Check how the pump is hung on the bed Resonance can occur vibrations are felt in parts of the bed Remove the pump and listen whether this makes a difference Can be resolved by putting the pump ona flat steady surface or by placing a towel between the pump and bed If your questions are not answered in the above information please contact Care of Sweden or your local dealer 14 9 Technical description 9 1 Technical specification PUMP SPECIFICATION Model Pf Croce Vari S Voltage 220 240 50 6067 S Power consumption OFA Typeoffuse Taa Cycleperiods Variable 10 15 20or 25min Dimension LxWxH 29 1 K20x1FeM o Weit o ee po Storage strc oo Shipping C 70 C Po Humidity O OS Operation 10 90 non condensing ee eee Storage 10 90 non condensing Atmospheric pressure 700hPa 1060hPa _ O Classification J Class i Type BF PAD MATTRESS SPECIFICATION Model Croce AAD Dimension WxtxH 80 85 90 105 120x200 210x21cem Weight Tf 1O5kg 85x200x21em i O NOTE
9. agnostic errors 1 National Pressure Ulcer Advisory Panel European Pressure Ulcer Advisory Panel and Pan Pacific Pressure Injury Alliance Prevention and Treatment of Pressure Ulcers Quick Reference Guide Emily Haesler Ed Cambridge Media Osborne Park Western Australia 2014 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Sweden s range CuroCell 4 AD is a mattress in function group C This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se 2 Product description 2 1 Pump and mattress 1 CPR rapid deflation 2 Pump 3 Mattress 4 Pump connector 4 static head cells If the product is equipped with ventilated cells they are located here 9 cells Annannanaananananan W Y ANEN NEN NEN NEN NEN NEN NEN N e LX Y n E With the optional heel section is the four cells in the foot end a bit lower to in order to relief the heels gt 2 2 Pump Front Back 1 Quick connector unit air 1 Off On
10. e the pressure alternates between the cells and there are four different cycle times to choose from C Static When Static is chosen the alternating cycle is discontinued and the pressure is equalized over the cells This function makes all patient movements as well as nursing and care of the patient easier The function can be used in association with patient care repositioning transferring in and out of bed Press button to turn off Otherwise this is automatically done after about 20 minutes D Seat inflation This mode gives extra support when sitting up and can be set for both alternating and static mode 4 PANEL LOCK Press down the Panel lock button for about 2 seconds to lock or unlock the panel 5 ALARM MUTE Press the ALARM MUTE button to temporary turn off the alarm Should the situation not be resolved within 5 minutes the alarm will be activated again Please refer to section 8 to troubleshoot any problems with the alarm A Power failure alarm During a power failure both the sound and light alarm will go on Turn it off by pressing the ALARM MUTE button B Low pressure alarm When the low pressure light is on the pressure in the mattress is below normal Both the light and sound alarm will activate The alarm will sound until the ALARM MUTE button is pressed If the problem remains see chapter 8 for trouble shooting 6 SERVICE This light indicates a technical failure in the pump First check that the air filt
11. er is clean and that the pump is not covered by anything such as a pillow or a blanket If the problem remains please contact the manufacturer or your local dealer for further information 7 POWER Power on off 3 Assembly Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact the manufacturer or your local dealer before using the product 1 J 5 i gaa ae e EY TEZ i 4 Place the mattress on the bed frame Secure the mattress to the bed with straps on the mattress bottom Note e Make the bed with sheets for added comfort e Check that the CPR valve is closed e The mattress must be attached to the bed s moving parts if it is height adjustable e Check the cells and snaps to make sure that they have been attached correctly Hang the pump at the foot end of the bed or place it on a flat steady surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels or getting crushed when the bed is raised or lowered If a cable holder is used optional open the snaps position the power cord in the pocket and close the snaps Connect the air tubes from the mattress to the side of the pump There is an audible click sound when the connectors are properly attached Check that the power switch at the side of the pump is set to O off Connect the plug to an approved electrical socket 2
12. g to the alternating pressure principle a technique that is well proven and been used for a long time Alternating pressure is when the pressure regularly alternates between the mattress cells to replicate body movement Variable pressure cycles are set up where there is little or no pressure This is optimal for skin improvement or to prevent skin damage Always read the user manual prior to use 1 1 General information This CuroCell mattress is a medical device in accordance to MDD 93 42 EEC According to statutory regulations made by the authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell 4 AD is an alternating pressure mattress that is placed directly on the bed frame The mattress can be used in all kinds of caring environments to prevent pressure ulcers as well as treatment up to Category 4 pressure ulcers The mattress is meant for use for persons of a recommended minimum length of 120 cm The recommended user weight is up to 200 kg CuroCell 4 AD is a replacement mattress and s
13. hall be placed directly on the bed frame We recommend that position changes are made regularly The time interval must be evaluated by the responsible personnel depending on the patient s status diagnosis and general condition cil The mattress should only be used lying in the lengthwise direction on the mattress with the feet at the end marked with the feet symbol The mattress must always be cleaned between patients _e eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee ee eee eee NOTE e For certain patients such as amputees are the recommended minimum weight not valid These users may however have other requirements as the entire mattress is not being loaded Please refer to sections 4 2 and 4 3 for operating control e When used together with positioning pillows read the user s manual carefully to ensure correct positioning of the patient e If you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become trapped or damaged e The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for use during x ray examinations because of the risk of blurred images or artefacts that may lead to di
14. he bed is raised e When only parts of the mattress are under load e g if the patient is an amputee the pressure should be increased 1 or 2 levels e Always check the set pressure with a hand check see 4 4 for more information e Minimum maximum user weight up to 200 kg 4 4 Hand check Hand check is used to ensure that the mattress is properly inflated This should be done regularly once per shift and after changing the position in the bed or changing the comfort settings Insert your hand between the air cells directly underneath the user s pelvic region use a vertical hand There should be a clearance of 2 3 fingers between mattress base and the patient If the distance is not sufficient increase the comfort setting to the next highest step Wait 10 minutes and perform a new hand check 4 5 CPR In case of an emergency where CPR cardiopulmonary resuscitation is necessary Rapidly pull the CPR band on the tubes between the pump and the mattress Remove the tube connections on the side of the pump for even faster deflation NOTE Make sure that the CPR connector is properly connected after use 4 6 Alarms When the pressure drops to an abnormally low level the Low Pressure diode lights up together with an audio alarm Please refer to section 8 for troubleshooting The alarm is deactivated by pressing the ALARM MUTE button on the pump s control panel If the problem is not resolved within 5 minutes
15. hich the sheet can be attached Six required per mattress Sturdy transport bag Transport and storage bag 11 2 Accessories Top part CuroCell LF low friction CuroCell Comfort fabric Item no Size Item no Size CC 916210 80x200 cm CC 907410 80x200 cm CC 916211 80x210 cm CC 907411 80x210 cm CC 916212 85x200 cm CC 907412 85x200 cm CC 916213 85x210 cm CC 907413 85x210 cm CC 916214 90x200 cm CC 907414 90x200 cm CC 916215 90x210 cm CC 907415 90x210 cm CC 916216 105x200 cm CC 907416 105x200 cm CC 916218 120x200 cm CC 907418 120x200 cm Top part CuroCell LF low friction Top part with increased sliding function can be used for heavy immobile patients Art No Description and size CC 900220 Heel function with valve 4 cells CC 900222 Heel function with valve 3 cells CC 9094 Transport bag 80 90 cm CC 9075 Sheet holder CuroCell one of 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 18 12 Warranty 12 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 12 2 Service and maintenance This CuroCell mattress is a medical device
16. ll follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 19 n Manufactured by Care of sweden Supporting Life Contact Phone 46 771106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns vag 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care of Sweden AB 2015 20
17. merge it in any liquid LT eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee eee Pee eee eee eee eee eee eee eee eee eee Wipe off the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the pump housing MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner 357 Q ll Bet RY Multi layer covers should be divided before washing Wash with like colours Chlorine and phenol based cleaners could have a negative impact on the PU surface zy and should be avoided If chlorine is used we recommend a mixture of max 1 11 INNER COVER AND COMFORT FABRIC accessories Tes Q ll Bx Bi Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Ilsopropanol 70 oxidating solutions or chlorine max 1 5 2 Reconditioning PUMP See section 5 1 for instructions MATTRESS 1 2 Disconnect the tubes and let the air out also open the CPR valve Remove the cover See instructions in chapter 5 1 for cleani
18. ng covers If the product is equipped with a foam core this should be removed and cleaned separately Move the mattress to a clean area which is suitable for cleaning Wipe off and disinfect the cells all tubes and the CPR valve with an alcohol based disinfectant cleaner intended for this purpose Perform Virkon or Desisoft can also be used Let it dry Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides lsopropanol 70 oxidating solutions or chlorine max 1 Put the mattress together If cells for any reason have loosened from the tubes these must be replaced according to drawing in chapter 2 1 NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 6 Storage The mattress and pump are best stored in the product bag optional extra Handle the packed product with care Do not put anything heavy on top otf it 12 7 Maintenance 7 1 General 1 Check the power cord and plug for damage and scratches 2 Check that the mattress cover is intact and that the cover and air cells are correctly assembled 3 Check that the air flow alternates between each outlet on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your
19. the alarm will start again The diode remains on until the pressure level is remedied and reset The pump is also fitted with a Power Failure alarm An alarm sounds and flashes when the pump is disconnected from the electric power Turn off the alarm by pressing the ALARM MUTE button WARNING If the air pressure remains low check the mattress and tubes for damage Repair if necessary or contact the manufacturer or your local dealer for help 4 7 Transport function If the patient must be moved in bed the tubes can be connected together detached from the pump and the mattress put in transport mode We recommend using this mode only for short periods of time 10 4 8 Heel function with check valves option This feature allows pressure relief of the heels Disconnect the air cell which is directly under the heel This air cell will then be emptied for air Press the bottom 1 and disconnect the plug Let the cell remain in its place in the mattress The air cell is activated by assembling the connection again 5 Cleaning and reconditioning It is important to follow the instructions below before using the product for another user Consult the hygiene responsible or the supplier for help and instructions when unsure 5 1 Cleaning PUMP be VARNING Do not open the pump housing risk of electric shock Service and maintenance must be performed by authorised technicians Do not get the pump wet or sub
20. ting it could become damaged if it comes into contact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 4 4 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover Please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk vapour permeable can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects 11 If the cover are supplied with handles on the long sides theses handles are only to be used for lifting and moving the mattress itself All other use will be at your own responsibility and is not covered by the product warranty COP eee E E ETF Ui 1 Introduction CuroCell air mattresses operate accordin