Home
AssayMaxTM Human ApoC
Contents
1. May be stored for up to 30 daysina vacuum desiccator e Diluent 1x may be stored for up to 30 days at 2 8 C e Store Standard at 2 8 C before reconstituting with Diluent and at 20 C after reconstituting with Diluent Other Supplies Required e Microplate reader capable of measuring absorbance at 450 nm Pipettes 1 20 ul 20 200 ul 200 1000 ul and multiple channel e Deionized or distilled reagent grade water Sample Collection Preparation and Storage e Plasma Collect plasma using one tenth volume of 0 1 M sodium citrate as an anticoagulant Centrifuge samples at 3000 x g for 10 minutes Dilute samples 1 4000 into EIA Diluent and assay The undiluted samples can be stored at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles EDTA can also be used as an anticoagulant e Serum Samples should be collected into a serum separator tube After clot formation centrifuge samples at 3000 x g for 10 minutes and remove serum Dilute samples 1 4000 into EIA Diluent and assay The undiluted samples can be stored at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles e Cell Culture Media Centrifuge cell culture media at 3000 x g for 10 minutes to remove debris Collect supernatants and assay Store samples at 20 C or below Avoid repeated freeze thaw cycles e Cell Lysate Rinse cell with cold PBS and then scrape the cell into a tube with 5 ml cold PBS with 0 5 M EDTA Centrifuge suspension at 1
2. equal to 400 ul 1 1000 1 100000 4 ul sample 396 ul buffer 100x 24 ul of A 216 ul buffer 10x 1000 fold dilution Assuming the needed volume is less than or equal to 240 ul 4 ul sample 396 ul buffer 100x 4 ul of A 396 ul buffer 100x 24 ul of B 216 ul buffer 10x 100000 fold dilution Assuming the needed volume is less than or equal to 240 ul Reagent Preparation e Freshly dilute all reagents and bring all reagents to room temperature before use e EIA Diluent Concentrate 10x If crystals have formed in the concentrate mix gently until the crystals have completely dissolved Dilute the EIA Diluent Concentrate 1 10 with reagent grade water Store for up to 30 days at 2 8 C e Standard Curve Reconstitute the 1 75 ug of Human ApoC lll Standard with 3 5 ml of EIA Diluent to generate a 0 5 ug ml standard stock solution Allow the standard to sit for 10 minutes with gentle agitation prior to making dilutions Prepare duplicate or triplicate standard points by serially diluting the standard stock solution 0 5 ug ml 1 4 with EIA Diluent to produce 0 125 0 0313 0 0078 0 0019 and 0 0005 ug ml solutions EIA Diluent serves as the zero standard 0 ug ml Any remaining solution should be frozen at 20 C and used within 30 days Standard pes ApoC lll Point Dilution ug ml 1 part Standard 0 5 ug ml 0 5000 ESI 1 part P1 3 parts EIA Diluent 0 1250 P3 1 part P2 3 parts EIA Diluen
3. times with 300 ul of Wash Buffer and then invert the plate decanting the contents hit 4 5 times on absorbent material to completely remove the liquid e Add 50 ul of Biotinylated Human ApoC Ill Antibody to each well and incubate for 1 hour e Wash the microplate as described above e Add 50 ul of Streptavidin Peroxidase Conjugate to each well and incubate for 30 minutes Turn on the microplate reader and set up the program in advance e Wash the microplate as described above e Add 50 ul of Chromogen Substrate per well and incubate for 12 minutes or till the optimal blue color density develops Gently tap plate to ensure thorough mixing and break the bubbles in the well with pipette tip e Add 50 ul of Stop Solution to each well The color will change from blue to yellow e Read the absorbance on a microplate reader at a wavelength of 450 nm immediately If wavelength correction is available subtract readings at 570 nm from those at 450 nm to correct optical imperfections Otherwise read the plate at 450 nm only Please note that some unstable black particles may be generated at high concentration points after stopping the reaction for about 10 minutes which will reduce the readings Data Analysis e Calculate the mean value of the duplicate or triplicate readings for each standard and sample e To generate a standard curve plot the graph using the standard concentrations on the x axis and the corresponding mean 450 nm absorban
4. 500 rpm for 10 minutes at 4 C and aspirate supernatant Re suspend pellet in ice cold Lysis Buffer 10 mM Tris pH8 0 130 mM NaCl 1 Triton X 100 protease inhibitor cocktail For every 1 x 10 cells add approximately 100 uL of ice cold Lysis Buffer Incubate on ice for 60 minutes centrifuge at 13000 rpm for 30 minutes at 4 C and collect supernatant for assay e Urine Collect urine using sample pot Centrifuge samples at 800 x g for 10 minutes and assay Store samples at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles e Milk Collect milk using sample tube Centrifuge samples at 800 x g for 10 minutes and assay Store samples at 20 C or below for up to 3 months Avoid repeated freeze thaw cycles e CSF Collect cerebrospinal fluid CSF using sample pot Centrifuge samples at 3000 x g for 10 minutes Dilute samples 1 2 into EIA Diluent and assay The undiluted samples can be stored at 80 C for up to 3 months Avoid repeated freeze thaw cycles Refer to Sample Dilution Guidelines below for further instruction Guidelines for Dilutions of 1 100 or Greater for reference only please follow the insert for specific dilution suggested 1 100 1 10000 A 4ulsample 396 ul buffer 100x 100 fold dilution Assuming the needed volume is less than or equal to 400 ul 4 ul sample 396 ul buffer 100x 4 ul of A 396 ul buffer 100x 10000 fold dilution Assuming the needed volume is less than or
5. A assarono AssayMax Human ApoC Ill ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St Charles MO 63301 T 636 447 9175 F 636 395 7419 WWW assaypro com For any questions regarding troubleshooting or performing the assay please contact our support team at support assaypro com Thank you for choosing Assaypro Assay Summary Step 1 Add 50 ul of Standard or Sample per well Incubate 2 hours Step 2 Wash then add 50 ul of Biotinylated Antibody per well Incubate 1 hour Step 3 Wash then add 50 ul of SP Conjugate per well Incubate 30 minutes Step 4 Wash then add 50 ul of Chromogen Substrate per well Incubate 12 minutes Step 5 Add 50 ul of Stop Solution per well Read at 450 nm immediately Symbol Key cli Consult instructions for use Assay Template 12 11 10 Human Apolipoprotein C III ELISA Kit Catalog No EA8133 1 Sample insert for reference use only Introduction Apolipoprotein C III ApoC Ill is a surface component of chylomicrons very low density lipoproteins and high density lipoproteins It consists of 79 amino acids with a molecular mass of 8 8 kDa 1 ApoC lll is synthesized mainly in the liver and to a lesser degree in the intestine It plays a key role in triglyceride rich lipoprotein metabolism It is an inhibitor of lipoprotein lipase and hepatic lipase and interferes with binding of lipopr
6. ce on the y axis The best fit line can be determined by regression analysis using log log or four parameter logistic curve fit e Determine the unknown sample concentration from the Standard Curve and multiply the value by the dilution factor Typical Data e The typical data is provided for reference only Individual laboratory means may vary from the values listed Variations between laboratories may be caused by technique differences Standard Point Average OD P1 0 5000 P2 0 1250 P3 0 0313 P4 0 0078 P5 0 0019 P6 0 0005 P7 0 0000 Sample Pool Normal Sodium Citrate Plasma 4000x Standard Curve e The curve is provided for illustration only A standard curve should be generated each time the assay is performed Human ApoC4ll Standard Curve E E o D N Q o 0 010 ApoC Ii ug ml Reference Value e Normal human ApoC Ill plasma levels range from 60 to 160 ug ml e Human Plasma and Serum samples from healthy adults were tested n 40 On average ApoC lll level was 97 ug ml Sample Average Value ug ml Human Pool Normal Plasma 97 Human Normal Plasma 86 Human Pool Normal Serum Performance Characteristics e The minimum detectable dose of ApoC Ill as calculated by 25D from the mean of a zero standard was established to be 0 0004 ug ml e Intra assay precision was determined by testing replicates of three plasma samples in one assay
7. e Inter assay precision was determined by testing three plasma samples in twenty assays Intra Assay Precision Inter Assay Precision Sample 1 2 3 1 2 3 n 20 20 20 20 20 20 CV 96 Average CV 96 Recovery Standard Added Value Recovery Average Recovery 0 002 0 125 pg ml 82 113 103 Linearity e Plasma and serum samples were serially diluted to test for linearity Average Percentage of Expected Value Sample Dilution Plasma Serum 1 2000 106 94 1 4000 99 97 1 8000 101 105 Cross Reactivity Species Cross Reactivity Canine None Bovine None Monkey lt 5 Mouse None Rat None None Cross Reactivity lt 20 Swine Protein ApoC ll e _ No significant cross reactivity observed with ApoA l ApoA Il ApoB ApoC l ApoE ApoH and ApoM Troubleshooting Issue Causes Course of Action Use of expired e Check the expiration date listed before use components e Do not interchange components from different lots e Check that the correct wash buffer is being used S e Check that all wells are dry after aspiration 7 Improper wash step e Check that the microplate washer is dispensing properly 9 e If washing by pipette check for proper pipetting a technique 3 Splashing of reagents e Pipette properly in a controlled and careful manner 2 while loading wells e Pipette properly in a cont
8. epeat the assay Non optimal sample e Competitive ELISA If samples generate OD values lower dilution than the highest standard point P1 dilute samples further and repeat the assay e User should determine the optimal dilution factor for samples Contamination of e A new tip must be used for each addition of different reagents samples or reagents during the assay procedure Contents of wells e Verify that the sealing film is firmly in place before placing evaporate the assay in the incubator or at room temperature e Pipette properly in a controlled and careful manner Improper pipetting e Check pipette calibration Deficient Standard Curve Fit e Check pipette for proper performance e Thoroughly agitate lyophilized components after reconstitution e Thoroughly mix dilutions Insufficient mixing of reagent dilutions References 1 Jong MC et al 1999 Arterioscler Thromb Vasc Biol 19 472 484 2 Wang CS etal 1985 J Clin Invest 75 384 390 3 Dammerman M et al 1993 Proc Natl Acad Sci U S A 90 4562 4566 Version 4 5R 10 Related Products e ERA9133 1 AssayMax Rat Mouse ApoC Ill ELISA Kit Plasma Serum and Cell Culture samples www assaypro com e e mail Support assaypro com
9. idic solution e Thekit should not be used beyond the expiration date Reagents e Human ApoC lll Microplate A 96 well polystyrene microplate 12 strips of 8 wells coated with a polyclonal antibody against human ApoC lll e Sealing Tapes Each kit contains 3 precut pressure sensitive sealing tapes that can be cut to fit the format of the individual assay e Human ApoC lll Standard Human ApoC lll in a buffered protein base 1 75 ug lyophilized e Biotinylated Human ApoC Ill Antibody 80x A 80 fold concentrated biotinylated polyclonal antibody against ApoC III 100 pl e EIA Diluent Concentrate 10x A 10 fold concentrated buffered protein base 30 ml e Wash Buffer Concentrate 20x A 20 fold concentrated buffered surfactant 30 ml 2 bottles e Streptavidin Peroxidase Conjugate SP Conjugate A 100 fold concentrate 80 ul e Chromogen Substrate A ready to use stabilized peroxidase chromogen substrate tetramethylbenzidine 8 ml e Stop Solution A 0 5 N hydrochloric acid to stop the chromogen substrate reaction 12 ml Storage Condition e Upon arrival immediately store components of the kit at recommended temperatures up to the expiration date e Store SP Conjugate and Biotinylated Antibody at 20 C e Store Microplate Diluent Concentrate 10x Wash Buffer Stop Solution and Chromogen Substrate at 2 8 C e Unused microplate wells may be returned to the foil pouch with the desiccant packs and resealed
10. oteins to cell surface heparan sulfate proteoglycans and receptors 2 3 Principle of the Assay The AssayMax Human Apolipoprotein C III ELISA Enzyme Linked Immunosorbent Assay kit is designed for detection of human ApoC lll in plasma serum urine milk CSF and cell culture samples This assay employs a quantitative sandwich enzyme immunoassay technique that measures human ApoC lll in less than 4 hours A polyclonal antibody specific for human ApoC Ill has been pre coated onto a 96 well microplate with removable strips ApoC lll in standards and samples is sandwiched by the immobilized antibody and biotinylated polyclonal antibody specific for ApoC lll which is recognized by a streptavidin peroxidase conjugate All unbound material is then washed away and a peroxidase enzyme substrate is added The color development is stopped and the intensity of the color is measured Caution and Warning e This product is for Research Use Only and is Not For Use In Diagnostic Procedures Prepare all reagents working diluent buffer wash buffer standard biotinylated antibody and SP conjugate as instructed prior to running the assay e Prepare all samples prior to running the assay The dilution factors for the samples are suggested in this insert However the user should determine the optimal dilution factor e Spin down the SP conjugate vial and the biotinylated antibody vial before opening and using contents e The Stop Solution is an ac
11. rolled and careful manner Inconsistent volumes 5 NS Check pipette calibration loaded into wells n e Check pipette for proper performance e Thoroughly agitate lyophilized components after reconstitution e Thoroughly mix dilutions e Check the microplate pouch for proper sealing Insufficient mixing of reagent dilutions Improperly sealed e Check that the microplate pouch has no punctures microplate e Check that three desiccants are inside the microplate pouch prior to sealing Microplate was left e Each step of the procedure should be performed T unattended between uninterrupted 5 steps n Omission of step e Consult the provided procedure for complete list of steps 5 Steps performed in e Consult the provided procedure for the correct order incorrect order 5 z Insufficient amount of e Check pipette calibration 3 Pa reagents added to Check pipette for proper performance 2 9 wells gt Wash step was skipped e Consult the provided procedure for all wash steps E Improper wash buffer e Check that the correct wash buffer is being used 9 Improper reagent e Consult reagent preparation section for the correct E preparation dilutions of all reagents u Insufficient or e Consult the provided procedure for correct incubation gt prolonged incubation time periods e Sandwich ELISA If samples generate OD values higher than the highest standard point P1 dilute samples further and r
12. t 0 0313 Pa Apart P3 3partsEIADiluent_ 00078 Ps 1partP5 3 parts EIA Diluent 0 0005 e Biotinylated Human ApoC Ill Antibody 80x Spin down the antibody briefly and dilute the desired amount of the antibody 1 80 with EIA Diluent Any remaining solution should be frozen at 20 C e Wash Buffer Concentrate 20x If crystals have formed in the concentrate mix gently until the crystals have completely dissolved Dilute the Wash Buffer Concentrate 1 20 with reagent grade water e SP Conjugate 100x Spin down the SP Conjugate briefly and dilute the desired amount of the conjugate 1 100 with EIA Diluent Any remaining solution should be frozen at 20 C Assay Procedure e Prepare all reagents standard solutions and samples as instructed Bring all reagents to room temperature before use The assay is performed at room temperature 20 25 C e Remove excess microplate strips from the plate frame and return them immediately to the foil pouch with desiccants inside Reseal the pouch securely to minimize exposure to water vapor and store in a vacuum desiccator e Add 50 ul of Human ApoC IlI Standard or sample per well Cover wells with a sealing tape and incubate for 2 hours Start the timer after the last addition e Wash five times with 200 ul of Wash Buffer manually Invert the plate each time and decant the contents hit 4 5 times on absorbent material to completely remove the liquid If using a machine wash six
Download Pdf Manuals
Related Search
Related Contents
TP UP-PN Portable Dot Matrix Printer User's Manual Mikrowelle Bedienungsanleitung TMW 175 S Page11 - 岡山大学 工学部 1503-IN050 - Rockwell Automation Keralastic Mapei Bedienungsanleitung - Zusätzliche Funktionen RVBR4 v2 Operating Instructions P7831 Bird Repeller Congratulations on your Projet « Les jeunes et leur alimentation » Copyright © All rights reserved.