Home
FURNACE DOCTOR BASIC USER MANUAL
Contents
1. 10 0 It is recommended that you select span gas with values approximately 20 higher than typical observed process value Span gas values may be higher or lower but should not exceed the ranges of the measuring cells of 30 1 0 10 30 depending on the option purchase Be sure to keep the second stage regulator pressure around 1 2 PSI Your Furnace Doctor is equipped with good internal pressure regulation but too much pressure may override the regulator s ability to control the pressure in the cells It is important to understand the I R cells are very pressure sensitive by nature Realize that for a given species of gas there are more molecules of that gas present in a cell with a pressure of 1 5 PSI than in a cell with a cell pressure of 0 5 PSI will lead to measurement error if the unit is calibrated at a high pressure and then applied to low pressure measurements You may observe these phenomena by closing of the exhaust of the instrument while measuring a furnace Note the quick rise in the CO value Calibration Frequency Furnace Control Corp recommends the following procedure e Connect the span bottle to the instrument and measure it on a weekly basis If the instrument measures the span bottle within 2 of the analyzed values on the bottle it is not necessary to calibrate the instrument e f calibration is required do it in the environment that the instrument will be used in If y
2. or nitrogen methanol system Be sure that the probe control system is working properly Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 13 of 28 OPERATIONS e Connect one end of the sample tube to the inlet Connect the other end to the furnace sample port Do not use oxygen probe burn off ports or connect to a port with a lambs wool or angel hair filter already installed Sample inlet hookup point Any discoloration in the white sample element requires the element to be replaced P N T FIL 01 The 3 micron green filter should be changed annually P n T FIL 02 Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 14 of 28 FURNACE CONTROL Afurmace toctor Heat Treat Atmosph gea Power switch Power the unit on by depressing the power switch Depress the Pump button to start sampling e The LCD display will illuminate and the internal pump will begin after pump switch is depressed drawing a sample Allow adequate time for the sample line to purge Moisture could be retained in the sample line At least two minutes for each 10 feet of sample line is recommended e Ensure that sample flow is adequate If you are sampling a generator with
3. reserved FURNACE DR BASIC Version 005 Page 16 of 28 e The instrument displays the carbon dewpoint for the measured CO COs CH gasses and entered probe temperature Additionally it displays calculated probe millivolts e instrument will take at least 60 seconds from cold start to display any meaningful information Ensure that sample flow is adequate e Look at the flow meter on the front of the instrument and make sure that it indicates at least 6 CFH e Meaningful process values are displayed on the LCD after initialization and the unit has obtained a good sample from the furnace and process temperature is entered A note about sample ports In a large percentage heat treats we have found that the furnace sample ports are either plugged or leaking Even if you can look into the port and it appears unobstructed the port may leak somewhere behind the hot face of the refractory This may cause the sample to be drawn from between the brick and the furnace shell or even outside of the furnace If you are getting numbers do not make sense I E Extremely low CO values or high try taking a sample from the oxygen probe burn off port not the reference airport Only use this method of no other working port is available This method will affect the probe reading f the port extends for a large distance outside of the furnace wall moisture can accumulate in the bottom of the pipe leading to in correct readings Y
4. whether you are producing gas by means of an endothermic generator or using prepared gasses such as nitrogen and methanol Ratios are slightly different if propane is used rather than methane for the production of carrier gas Obviously any increase in volume in one of the gasses means a decrease in one or more of the other gasses We often get asked the question Why is my generator producing 20 CO but am only measuring 17 at my furnace It is even lower when add ammonia One part of the answer to this question is simple volumetric displacement There is only room in the furnace for 100 of the atmosphere Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 7 of 28 How can the Furnace Doctor help me with my generator The basic chemical reaction taking place in an endothermic generator is as follows 2 4 gt 2 4 Ignoring the nitrogen Remember you are mixing gas and air together and heating it up to produce the above described gas composition The reaction takes place in two stages First some of the methane burns with air and makes heat The by products of this combustion are water vapor and In the second stage the remaining methane reacts with the and H20 It becomes obvious that we want this reaction to be as complete as possible This is the reason for t
5. FURNACE WWW GROUP UPC COM CONTROL Nh FURNACE DOCTOR BASIC USER MANUAL Version Number 005 Afurnace toctor Heat Treat Atmosphere Analyzer PUMP Pnited AFFILIATED MEMBERS PROCESS CONTROLS Furnace Control Corp Marathon Monitors Inc Process Electronic Waukee Engineering Co FURNACE DR BASIC Version 005 Page 2 of 28 COPYRIGHT No part of this publication may be reproduced transmitted transcribed stored a retrieval system or translated into any language or computer language in any form or by any means electronic mechanical magnetic optical chemical manual or otherwise without prior written permission of FURNACE CONTROL CORP DISCLAIMER The FD BASIC is to be used by the industrial operator under his her direction Furnace Control Corp is not responsible or liable for any product process damage or injury incurred while using the FD BASIC FURNACE CONTROL CORP makes no representations or warranties with respect to the contents hereof and specifically disclaims any implied warranties or merchantability or fitness for any particular purpose For assistance please contact Furnace Control Corp TEL 1 513 772 1000 FAX 1 513 326 7090 Toll Free North America 1 800 547 1055 erika leeds group upc com www group upc com Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Pa
6. and the following screen will appear Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 22 of 28 Enter the correct gas values as displayed on you Span Gas Bottle Use the Left or Right arrows to move the cursor under the desired number you wish to change Use the or buttons to change the number that the cursor is under until the desired value is displayed Move the cursor to the next number and change the value in the same manner When done push the enter button you will return to the main menu Display Gas Values will be highlighted Scroll down until CH4 Span Gas Values is highlighted Press the Enter Key and the following screen will appear Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 23 of 28 Enter the correct CH gas values as displayed on you Span Gas Bottle Use the Left or Right arrows to move the cursor under the desired number you wish to change Use the or buttons to change the number that the cursor is under until the desired value is displayed Move the cursor to the next number and change the value in the same manner When done push the enter button you will return to the main menu Display Gas Values will be highlighted Once the correct CO an
7. ation Hardening Annealing Normalizing Stress Relieving call FCC if your process is not listed NH3 ammonia is strictly prohibited in the sample Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 6 of 28 THEORY Why Infrared By purchasing your new Furnace Doctor you have obviously made the decision that your current process measurement methods require some enhancement This may be due to customer requirements for back up process verification to your existing oxygen probe based system or simply a matter of answering the question is that probe really working You might be looking for more complete information of what is happening with respect to your furnace atmosphere Whatever your reasons the use of I R analysis and the Furnace Doctor will satisfy these requirements Since most Furnace Doctor owners are either performing gas carburizing or neutral hardening in an endothermic atmosphere the following discussion will help to understand the meaning of the measured values Endothermic atmospheres in use for these applications have this basic composition CO 20 carbon monoxide 40 Hydrogen carbon dioxide CH Trace methane usually from natural gas Trace oxygen H20 Trace water vapor Balance nitrogen These ratios are typical
8. charger to the charging port on the side e Red Status Indicator Light battery is charging e Green Status Indicator Light battery is charged Note Unit is designed to operate with a charged battery Using the charger as an AC adapter will not allow the battery to charge properly Eventually the battery will not be able to sustain operation of the analyzer Please connect the charger with the unit s power off and charge only when batt low light turns on Charging Port Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 28 of 28 RESTORE FACTORY CALIBRATION e You Furnace Doctor has the ability to restore to the last factory calibration if an error occurs during a user calibration e From the main menu scroll down until Restore Factory Calibration is highlighted e Press the enter button the following screen will appear e This will return calibration values to the last certified factory calibration Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved
9. d CH calibration gas values have to be entered from the main menu use the down arrow until Calibrate Analyzer is highlighted Press the enter button and the following screen will be displayed Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 24 of 28 Select 1 to start calibration 2 to exit and cancel the calibration by using the or buttons Once 1 has been entered the calibration will start The following screen will be displayed Connect and turn on the nitrogen gas be sure that you have adequate flow and press any key The unit will then start to count up to 60 When this is reached the following screen will be displayed Connect and turn on the span gas be sure that you have adequate flow and press any key Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 25 of 28 Once the unit has counted up to 60 the main menu will be displayed Display Gas Values will be highlighted Press the enter key to return to the main gas display screen The instrument should read the values in the bottle within 2 of the cell range i e 30 CO should be 6 CO Correction Factor This allows you to adjust the computed carbon to match shim stock analysis or actual carbon resul
10. earth or it may be reacting directly with the surface of the material gt 2H2 How much of the above reaction is happening is dependent on how much methane is present in your furnace The reaction also happens faster at higher temperatures but is still slow compared to the equilibrium reactions For example at 1600 with an equilibrium carbon potential of 4 and methane content of 5 the theoretical carbon potential is actually 44 In reality the effect is usually about 6 10 of what the theoretical calculations deliver making the carbon potential closer to 42 By contrast if we were to up the methane content to 5 the actual carbon potential might be closer to 6 a significant difference Whether the work really ends up seeing this is dependent on a number of issues temperature governing the rate of reaction the surface area of the work and the level of saturation at the surface atmosphere circulation etc Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 10 of 28 Using the Furnace Doctor to analyze furnace atmosphere Probe condition Carbon is computed using actual gas values so it is more accurate than the assumptions made by the probe Compare the C calculated by the Furnace Doctor to C displayed on the Carbon Control instrument Your carbon controller is set up with some as
11. etor to obtain the desired dew point How will the Furnace Doctor help me with my Furnace atmosphere If you are using oxygen probes the Furnace Doctor can supply independent traceable verification of your probes accuracy It will also give you a much more complete picture of what is happening in your furnace atmosphere The primary reactions involved in carbon transfer are well understood as 2CO Cy 0 gt Cy Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 9 of 28 These reactions are assumed to be close to equilibrium The theory behind this is that a reaction sometimes called the water gas reaction is busy keeping the H2 and Oz in balance CO lt H2 The measurement of Carbon in your atmosphere with an oxygen probe is based on the assumption that the above is true CO gt Cy Note that it is assumed that the primary mechanism of carbon transfer is CO dependant What is really happening is that the ratio of CO to is much more accurate means of determining equilibrium carbon One problem is that we use CH enriching gas What happens to the CH once it enters the furnace It decomposes mostly near the hot catalytic surfaces in your furnace It may be replenishing depleted CO based on equilibrium rules it may be making soot crumbs on your h
12. ge 3 of 28 TABLE OF CONTENTS SAFETY 4 EQUIPMENT 5 5 TRIE OB seen ces 6 OPERATIONS 18 SYSTEM ate tees 17 CALIBRATION eae 19 BATTERY CHARGER eee 27 RESTORE FACTORY 28 Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 4 of 28 SAFETY Please read the instructions before operating the instrument This instrument complies with accepted industrial safety standards Do NOT operate this instrument with the internal top cover removed FURNACE CONTROL Afurnace toctor Heat Treat Atmosphere Analyzer POWER Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 5 of 28 EQUIPMENT RATINGS Supply Voltage 110 240 VAC Supply Frequency 50 60 Hz Power Consumption 3 2 1 8 Amps Range Resolution Accuracy CO 0 30 0 02 2 of range CO2 0 2 20 1 of range CH4 0 10 0 02 2 of range Operating Temperature 40 F 120 F Relative Humidity 5 90 non condensing Atmospheres Suitable for Carburizing Carbonitriding Carbon Correction Neutral Hardening Austempering Martempering Precipit
13. he nickel catalyst in the generator Assuming that the main air gas ratio is set somewhere between 2 7 and 2 85 parts air to 1 part gas there are two considerations 1 The catalyst must be clean and free from soot If soot is present the efficiency of the above reactions goes down 2 You must have adequate temperature to keep the rate of the reactions high enough So how can the Furnace Doctor help First by measuring the generator s methane content can determine whether or not the catalyst is operating efficiently 8 CH is about the upper limit Above this level it is possible that the catalyst is laden with soot or the catalyst is simply spent Another indication of this condition would be a high COs or dew point The Furnace Doctor provides a convenient way to obtain dew point CO and Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 8 of 28 Check to see that the basic gas composition is as expected CO should be around 20 Check the CH level and be sure that it is less than 8 observe the and the dew point readings Note the relationship between dew point and COs Carbon Dioxide VS Dewpoint 1 60 SEE i 1 20 1 00 0 80 0 60 0 40 0 20 0 00 Carbon Dioxide Dewpoint Make adjustments to either your oxygen probe control system or the generator carbur
14. ou calibrate the instrument in an office that is 70 degrees and then go use the instrument in a 110 degree heat treat you will probably experience some error in measurement Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 20 of 28 To begin calibration e The procedure for Span gas values only needs to be performed when a new span bottle is to be used e From the default gas value display hold down the Enter button until the main menu is displayed Scroll down the menu by using the Down Arrow key until the CO Span Gas Value is highlighted e Press the Enter Key and the following screen will appear Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 21 of 28 e Enter the correct CO gas values as displayed on you Span Gas Bottle Use the Left or Right arrows to move the cursor under the desired number you wish to change Use the or buttons to change the number that the cursor is under until the desired value is displayed Move the cursor to the next number and change the value in the same manner When done push the enter button you will return to the main menu Display Gas Values will be highlighted Scroll down until CO2 Span Gas Values is highlighted e Press the Enter Key
15. ou may also purchase a non metallic sample probe that will penetrate the hot zone adequately Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 17 of 28 MENU SYSTEM Main Menu Screen Pressing and holding down the enter button from the default gas value display will cause the main menu to be displayed From this menu you may enter process temperatures for the computation of carbon and dew point calibrate the instrument choose between the endo generator calculation of dew point and the furnace temperature dependant calculation Entering a process temperature e Hold down the Enter button until the main menu is displayed e Use the down arrow until Enter Process Temp is highlighted Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 18 of 28 Press the enter button the following screen will be displayed Use the Left or Right arrows to move the cursor under the desired number you wish to change Use the or buttons to change the number that the cursor is under until the desired value is displayed Move the cursor to the next number and change the value in the same manner When done push the enter button you will return to the main menu Display Gas Values will be highlighted Pres
16. re and gas composition It tracks inversely to carbon potential At 1600F 4 C will be about 0 62 assuming 20 achieve 4 carbon at 1750 will be around 25 If the carbon potential is lower than expected the should be higher than expected Check for air leaks Check for burner tube leaks Check for furnace fan water jacket leaks Check generator or nitrogen methanol system If the probe system is in agreement with the Furnace Doctor be sure there is adequate carrier gas flow and adequate enriching gas flow Is carbon controller turning on the natural gas when it should Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 12 of 28 CH Most batch furnaces will operate with a CH content ranging from 1 to 4 5 At the start of the cycle this number is higher than when the atmosphere and the work approach equilibrium with each other Continuous furnaces performing carburizing will usually operate at higher methane levels In fact in some belt furnace applications this is necessary for the process to work High free methane content is usually a product of the control system calling for too much natural gas due to an air leak unusually high surface are in the load Check for air leaks Check for burner tube leaks Check for furnace fan water jacket leaks Check generator
17. s Enter to return to the gas values screen Selecting the Dew point Calculation Mode The instrument will compute dew point using either a temperature based calculation used when measuring furnace atmospheres where you can identify the process temperature or a simplified table look up method based to be used when measuring the dew point of endothermic generators If you use the furnace calculation when measuring your endothermic generator and enter a process temperature of 1900F or so you will get meaningless values for dew point If you have the generator calculation selected it is not necessary to enter a process temperature From the main menu using the Down Arrow key scroll until Switch Dewpoint Mode is highlighted Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 19 of 28 CALIBRATION You may calibrate your unit with certified grade span gas and nitrogen If you have plant nitrogen available make sure to regulate approximately 1 PSI Span bottle gas must have a bottle top regulator that will be capable or regulation at 1 PSI Be sure that your gas supplier provides you with a certified analysis of the gas in the bottle You may purchase span gas bottles from Furnace Control Corp Please contact your local Furnace Control Corp representative Values for span gas should be 25 0 CO 1 0
18. significant pressure at the sample port it is recommended that you partially close the port valve so that the internal pump is not damaged The pump is rated at 4 LPM 1 PSI Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 15 of 28 e The following display is shown on the LCD when power is applied to the unit e Measurement approximately 1 min after start up and is continuous after the process temperature is entered The unit will not correctly display O2mv Carbon or Dewpoint until temperature is entered The display includes actual CO as readings calculated theoretical carbon and dewpoint and miilli volts for the computed Carbon and manually entered temperature e f you desire to compute the correct C associated with the CO CH4 measured gasses press and hold the enter button The main menu will be displayed e Press the enter button to access the process temperature screen The arrow will point under the digit in the temperature number Change the indicated digit by using the keys until the desired digit appears Use the right or left pointing arrow keys to move from digit to digit until the desired temperature is displayed Press the enter key to store Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are
19. sumptions about the content of the furnace atmosphere Most controllers assume a CO to be 20 out of the box you may use the Furnace Doctor to correct this situation by adjusting this factor to the correct CO Observe the probe millivolts as computed by the Furnace Doctor and displayed by the carbon controller These should be within 1 2 of each other If not and you have entered the process temperature into the furnace doctor correctly you probably have a problem with the probe or the controller First check the millivolt reading at the back of the probe with a meter to eliminate any possible instrument problems Next be sure that the probe is free of soot by burning the probe off If the numbers still do not agree perform an impedance test on the probe to help determine the electrode condition You will also want to check to see how fast the probe recovers after this test Recovery time should be in seconds not minutes Finally disconnect the probes reference air tube and see if the millivolts reading on the controller changes if it does you have a leak in the probe substrate If you change the probe or if you are sure that the probe is ok check to be sure you are obtaining a good sample at the Furnace Doctor Many times sample ports are the cause of the problem A good non metallic sample probe helps solve this One trick is to use the probe burnoff port to obtain a sample This has a few problems associated with it in f
20. ts obtained from analysis of the work The number has a range of 0 255 An entry of 0 disables the calculation The calculated carbon is affected as follows C Corr fact 100 x Computed C For example if the computed carbon from the gas analysis and temperature is 1 00 C and the Corr Fact 90 then the displayed C 90 Degrees F C The unit has the capability of displaying the process temperature in degrees F or C From the main menu scroll down until Degrees F C is highlighted Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 26 of 28 Press the enter button the following screen will be displayed Choose 1 for degrees F and 2 for degrees C Push the enter button to accept the change The unit will then be in the main menu Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 27 of 28 BATTERY CHARGER Your Furnace Doctor includes a battery charger Charging e Unit will operate continuously for approximately 8 hours without recharging A full charge will take approximately 8 hours to complete It is not necessary to completely discharge the battery before charging e To charge the unit turn the Furnace Doctor Power switch to the off position e Connect the supplied battery
21. urnaces running high methane contents The methane will crack near the alloy surface of the probe creating a local reaction not representative of the overall furnace atmosphere Copyright 2010 Furnace Control Corp a member of United Process Controls All rights to copy reproduce and transmit are reserved FURNACE DR BASIC Version 005 Page 11 of 28 Interpreting the gas value readings CO For most atmospheres in use this number should be around 18 20 If it is lower than 18 then observe the following Check generator for proper operation If you are using nitrogen methanol check the ratio of nitrogen to methanol Also be sure that the atmosphere sparger is in good condition If you are using nitrogen to push the methanol into the furnace be sure that the methanol is not getting nitrogen bubbles in it If you are using ammonia the ammonia additions will dilute the CO in the furnace Remember there is only room for 100 of the gas in the furnace Check for leaks This can be confirmed by a high If you are asking for a high carbon potential high methane content may also dilute atmosphere mix Check for leaks Air leaks typically cause the control system to add more natural gas to the furnace High surface area loads will be able to draw more carbon out of the atmosphere Typically the methane content will go up because of this This number can be anywhere between 05 to 1 and is dependent on temperatu
Download Pdf Manuals
Related Search
Related Contents
Buddy Phone® - Ocean Technology Systems ST7570 S-FSK power line networking System-on-Chip Notice d`utilisation お買い上げありかとうございます。 この 「取扱説明書」 を HART Transmitter Calibration AD-EN-M-0002V1 Service Manual for distributor EN version Copyright © All rights reserved.
Failed to retrieve file